ZOLL AutoPulse User Manual

Resuscitation System Model 100

AutoPulse® Power System User Guide

P/N. 12457-001 Rev. 2
Power System User Guide

About this Guide

The information in this User Guide applies to the ZOLL AutoPulse® Power System designed for the AutoPulse Resuscitation System Model 100. The AutoPulse Power System consists of the AutoPulse Multi-Chemistry Battery Charger and the AutoPulse Li-Ion Battery and/or the AutoPulse NiMH Battery .
ZOLL shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Copyright

© Copyright ZOLL 2012. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form
or by any means, mechanical, electronic, photocopying, recording, or otherwise, without prior written permission of ZOLL.
AutoPulse and LifeBand® are trademarks of ZOLL. All other trademarks mentioned herein belong to their respective owners.
0344
USA ZOLL Circulation 650 Almanor Avenue Sunnyvale, CA 94085 USA
t: +1.408.541.2140 f: +1.408.541.1030
EU Authorized Representative ZOLL International Holding B.V. Newtonweg 18 6662 PV ELST The Netherlands
t: +31 481 366 410
Page ii P/N. 12457-001 Rev. 2
Power System User Guide

Table of Contents 0

Figures ......................................................................................................................v
Tables .......................................................................................................................vi
Preface ....................................................................................................................vii
Who Should Read this Guide ................................................................................vii
General Warnings and Precautions ......................................................................vii
Symbols ...................................................................................................................ix
1 Introduction of the AutoPulse Power System ................................................1-1
1.1 Recommended Battery, Charger and AutoPulse Ratios ....................................................1-2
2 AutoPulse Li-Ion Battery ..................................................................................2-1
2.1 Expected AutoPulse Li-Ion Battery Life .............................................................................2-2
2.2 Handling New AutoPulse Li-Ion Batteries .........................................................................2-2
3 AutoPulse NiMH Battery ...................................................................................3-1
3.1 Expected AutoPulse NiMH Battery Life .............................................................................3-2
3.2 Handling New AutoPulse NiMH Batteries .........................................................................3-2
4 The AutoPulse Multi-Chemistry Battery Charger ..........................................4-1
4.1 Setting Up the AutoPulse Multi-Chemistry Battery Charger ..............................................4-1
4.2 Operating the AutoPulse Multi-Chemistry Battery Charger ..............................................4-2
4.2.1 AutoPulse Multi-Chemistry Battery Charger Test-Cycle mode ...................................4-5
5 Using the AutoPulse Battery ............................................................................5-1
5.1 AutoPulse Li-Ion Battery Status Check .............................................................................5-1
5.2 AutoPulse Li-Ion Battery Installation .................................................................................5-2
5.3 AutoPulse NiMH Battery Status Check .............................................................................5-3
5.4 AutoPulse NiMH Battery Installation ................................................................................5-4
5.5 Battery Rotation .............................................................................................................. 5-6
6 Managing the AutoPulse Power System .........................................................6-1
6.1 AutoPulse Battery Management ......................................................................................6-1
6.2 AutoPulse Battery Maintenance ....................................................................................... 6-1
6.2.1 Cleaning the AutoPulse Battery ................................................................................6-1
6.2.2 AutoPulse Battery Inspection ....................................................................................6-2
6.2.3 Storing AutoPulse Batteries ......................................................................................6-2
6.2.4 Reaching the End of AutoPulse Li-Ion Battery Service Life .........................................6-3
6.2.5 Reaching the End of AutoPulse NiMH Battery Service Life .........................................6-3
6.2.6 Disposing of AutoPulse Batteries ..............................................................................6-3
7 AutoPulse Multi-Chemistry Battery Charger Maintenance ...........................7-1
7.1 Cleaning the AutoPulse Multi-Chemistry Battery Charger ................................................7-1
P/N. 12457-001 Rev. 2 Page iii
Power System User Guide
7.2 Replacing a AutoPulse Multi-Chemistry Battery Charger Fuse ...........................................7-1
Appendix A AutoPulse Daily Checklist ..............................................................A-1
Appendix B Troubleshooting .............................................................................. B-1
Appendix C Technical Specifications ................................................................. C-1
C.1 Li-Ion Battery Physical and Environmental ....................................................................... C-1
C.2 NiMH Battery Physical and Environmental ....................................................................... C-2
C.3 Multi-Chemistry Battery Charger Physical And Environmental ......................................... C-3
C.4 FCC Statement ............................................................................................................... C-4
C.5 Limited Warranty for AutoPulse Resuscitation System ..................................................... C-5
Index ....................................................................................................................... I-1
Page iv P/N. 12457-001 Rev. 2
Power System User Guide

Figures 0

Figure 1-1 AutoPulse System .................................................................................................1-1
Figure 1-2 Recommended component ratios .........................................................................1-2
Figure 2-1 The AutoPulse Li-Ion Battery .................................................................................2-2
Figure 3-1 AutoPulse System NiMH Battery ............................................................................3-2
Figure 4-1 AutoPulse Multi-Chemistry Battery Charger ..........................................................4-1
Figure 4-2 Sliding the Battery into a Multi-Chemistry Battery Charger Charging Bay ..............4-3
Figure 4-3 Multi-Chemistry Charger Control Panel .................................................................4-3
Figure 5-1 AutoPulse Li-Ion Battery Status Check Button and Status LEDs ..............................5-1
Figure 5-2 AutoPulse Li-Ion Battery Installation and Removal .................................................5-2
Figure 5-3 Battery Status Check Button and Status LEDs ........................................................5-3
Figure 5-4 AutoPulse NiMH Battery Installation and Removal .................................................5-4
Figure 5-5 Four Battery rotation .............................................................................................5-6
Figure 5-6 Three Battery rotation ...........................................................................................5-7
Figure 6-1 Examples of a damaged Battery connector and housing ........................................6-2
Figure 7-1 AutoPulse Multi-Chemistry Battery Charger Fuse Location ....................................7-1
P/N. 12457-001 Rev. 2 Page v
Power System User Guide

Tab le s 0

Table 4-1 AutoPulse Multi-Chemistry Battery Charger Status LEDs ....................................... 4-4
Table 5-1 AutoPulse Li-Ion Battery Status LEDs ...................................................................... 5-1
Table 5-2 Battery Status LEDs ................................................................................................ 5-3
Table B-1 Battery Troubleshooting Procedures ....................................................................... B-1
Table B-2 Multi-Chemistry Battery Charger Troubleshooting Procedures ................................ B-2
Table C-1 Li-Ion Battery Specifications ...................................................................................C-1
Table C-2 NiMH Battery Specifications ...................................................................................C-2
Table C-3 Multi-Chemistry Battery Charger Specifications ......................................................C-3
P/N. 12457-001 Rev. 2 Page vi
Power System User Guide

Preface 0

This document describes the operating steps and maintenance requirements for the AutoPulse Power System for use as part of the AutoPulse Resuscitation System Model 100. The AutoPulse Power System consists of the AutoPulse Multi-Chemistry Battery Charger and the AutoPulse Li-Ion Battery and/or the AutoPulse NiMH Battery.
Proper use of the AutoPulse Power System requires a thorough understanding of the Power System, and appropriate training and practice using the Power System.
Please read the entire AutoPulse Power System User Guide and AutoPulse Us er Guide before using the AutoPulse Power System.

Who Should Read this Guide 0

This document should be read by personnel who are tasked with the care and maintenance of the Power System used to operate the AutoPulse System.

General Warnings and Precautions 0

Warning:
Always charge a new Battery. Failure to charge a Battery may cause reduced Battery
performance.
Always charge a stored Battery before placing the Battery in active operation. Battery may
self-discharge when not in use. Failure to charge a Battery before use may cause device power failure. In no case should any Battery be used if it has not been charged within 2 days.
No modification of the AutoPulse Multi-Chemistry Battery Charger, the AutoPulse Li-Ion
Battery, or the AutoPulse NiMH Battery is allowed.
Caution: United States federal law restricts this device to sale by or on the order of a licensed physician. Caution: The AutoPulse System is designed to be used only with ZOLL-approved accessories. The
AutoPulse System will perform improperly if non-approved accessories are used.
Caution: Do not short the Battery leads. Electrical connection (short) between the Battery power leads
on the connector permanently damages the Battery and renders the Battery inoperable.
Caution: Always charge AutoPulse Batteries at temperatures between 41ºF (5ºC) and 95ºF (35ºC).
Charging Batteries at temperatures below 41ºF (5ºC) or above 95ºF (35ºC) will prevent the Battery from reaching its full capacity (operational time) and may lead to irreve rsible Battery damage.
Caution: Do not block the Multi-Chemistry Battery Charger’s ventilation slots.
P/N. 12457-001 Rev. 2 Page vii
Power System User Guide
Caution: Do not operate the Multi-Chemistry Battery Charger in a confined space. Caution: Do not position the Multi-Chemistry Battery Charger so that it is difficult to unplug the power
cord.
Caution: Keep the Multi-Chemistry Battery Charger away from moisture. Caution: Do not remove the Multi-Chemistry Battery Charger cover. The Multi-Chemistry Battery
Charger has no internal user-serviceable parts.
Caution: Use the Multi-Chemistry Battery Charger only with ZOLL AutoPulse Multi-Chemistry
Battery Charger Power Cord, as supplied.
Caution: The OPERA TOR shall not touch simultaneously any Battery Charger conductive parts and the
PATIENT
Caution: The AutoPulse System is designed to be used only with ZOLL approved Batteries. The
AutoPulse System will perform improperly if non-approved batteries are used. The use of other batteries may cause permanent damage to the AutoPulse and will void the warranty.
Caution: ZOLL Batteries are to be used only with the AutoPulse Platform or with ZOLL Chargers. Use
of a Battery in other applications may damage the Battery and will void warranty.
Caution: It is strongly recommended that a Battery not be stored in AutoPulse when AutoPulse is not
in active service (shift deployment) or is in extended storage. Storage in AutoPulse longer than a week may result in irreversible damage to a Battery.
Caution: Always inspect a Battery for damage prior to insertion into either the AutoPulse or the Multi-
Chemistry Charger. Never place a damaged Battery into the AutoPulse Platform or Charger. If damage to a Battery is found, contact ZOLL Technical Service.
Caution: ZOLL AutoPulse Batteries are mechanically keyed to the AutoPulse Platform and Battery
Charger to facilitate correct installation. Insert a Battery, connector first, into the AutoPulse Battery Bay or Multi-Chemistry Charger until it properly latches into position. Do not force a connection if you cannot easily connect Battery to either the Charger or the AutoPulse. Doing so may result in damage to the Battery, Charger, and/or AutoPulse.
Caution: Remove the protective plastic cap from a Battery before attempting to charge the Battery. Caution: The Battery is intended to be used by trained professionals. Keep out of the reach of children. Caution: Do not attempt to swallow the Battery in whole or in part. Caution: Do not use a Battery that has cracks in the Battery case exposing internal components. Do not
strike or throw a Battery. Do not use a Battery to strike another object. Mishandling of a Battery may lead to physical damage and present a fire or shock hazard.
Caution: Do not immerse any portion of a Battery in water or other fluids. Do not allow fluids to enter
a Battery or a Battery Connector. Fluid immersion or spillage may permanently damage the Battery or present a fire or shock hazard.
Page viii P/N. 12457-001 Rev. 2
Power System User Guide
Caution: If Battery pack leaks, do not allow the liquid to come into contact with skin or eyes. If contact
has been made, do not rub. Rinse skin or eyes with clean running water and immediately seek medical attention.
Caution: Do not heat, burn, or incinerate a Battery . Never expose a Battery to an open flame. Exposure
to heat above 158ºF (70ºC) may irreversibly damage the Battery . Incinerating the Battery may result in flames or explosion.
Caution: If the Battery gives off an odor, generates heat, becomes discolored or deformed, or in any
way appears abnormal during use, recharging, or storage, immediately remove it from the AutoPulse or Charger and stop using it. Otherwise, the problematic Battery may develop acid and/or electrolyte leakage, overheating, smoke emission, bursting and/or ignition.
Caution: Do not throw your Batteries away or send them to municipal dumps. Call you local waste
management officials for proper disposal instructions.
Caution: Do not transport or store Battery pack together with metal objects such as necklaces, keys,
zippers, etc. Contact with these and other similar metal objects may cause the Battery to short and generate high heat and burns.
Caution: Do not attempt to open a Battery. AutoPulse Batteries have no user-serviceable parts. Caution: Clean the Battery Connector and contacts only with a clean dry cloth and/or a non-conductive
brush.
Caution: Do not autoclave the AutoPulse Battery or the Multi-Chemistry Battery Charger. Caution: Retain the original product literature for future reference.

Symbols 0

The symbols below may be found in this User Guide, on the AutoPulse Multi-Chemistry Battery Charger, or Li-Ion or NiMH Battery.
Follow instructions for use
Indoor Use Only
Class II Equipment
P/N. 12457-001 Rev. 2 Page ix
Date of Manufacture
-
F
(-20ºC)
+
113
ºF
(
+
45ºC)
Manufacturer
EU Authorized Representative
Power System User Guide
SN
Serial Number
Recycle
Temperature Limitations
Do not discard
Rechargeable Battery
Do Not Incinerate
Fuse
DC Voltage
AC Voltage
Charging
Ready
Fail
Test Cycle
Power
Page x P/N. 12457-001 Rev. 2
Power System User Guide
LifeBand
AutoPulse Platform
User Control Panel
Li-Ion Batteries
(Primary and Spare)
NiMH Batteries
(Primary and Spare)
or
Multi-Chemistry Battery Charger

1 Introduction of the AutoPulse Power System

The AutoPulse Power System consists of a battery (either the AutoPulse Li-Ion Battery or AutoPulse NIMH Battery) and battery charger (the AutoPulse Multi-Chemistry Battery Charger).
Battery: Both the AutoPulse Li-Ion Battery and the AutoPulse NIMH Battery are proprietary, rechargeable, removable batteries that are specifically designed to supply power for AutoPulse operation.
Battery Charger: The AutoPulse Multi-Chemistry Battery Charger is a stand alone unit designed to charge and automatically maintain both the AutoPulse Li-Ion Battery and the AutoPulse NIMH Battery.
The AutoPulse Platform is intended to be deployed with other emergency equipment and should always be kept in a state of high-readiness. Customers should integrate an AutoPulse and Battery check into their daily equipment check procedures. Like other life supporting equipment, good battery management practices are essential to provide proper operation, and to avoid problems during use.
P/N. 12457-001 Rev. 2 Page 1-1

Figure 1-1 AutoPulse System

Power System User Guide

1.1 Recommended Battery, Charger and AutoPulse Ratios

An AutoPulse unit in general should be equipped with three to four Batteries to allow for two with the device (one for operation and one as a spare) and one to two being charged to support the next shift change, exchange or replacement after a patient use (see Figure 1-2 below).
AutoPulse Platform
LifeBand
Multi-Chemistry Battery Charger
Four Batteries

Figure 1-2 Recommended component ratios

A one-to-one AutoPulse to Charger ratio is recommended. It is important that there be sufficient capability to both charge and Test-Cycle Batteries while supplying the recommended minimum of two Batteries for use.
Page 1-2 P/N. 12457-001 Rev. 2
Power System User Guide

2 AutoPulse Li-Ion Battery

The AutoPulse Li-Ion Battery (see Figure 2-1) is a proprietary, rechargeable, removable Lithium-Ion battery that is a power source for the AutoPulse Platform.
The AutoPulse Li-Ion Battery is mechanically keyed to the AutoPulse Platform and Multi-Chemistry Battery Charger to facilitate correct installation. One end of the AutoPulse Li-Ion Battery contains connections for power and communications. A Battery Status Check button illuminates the AutoPulse Li-Ion Battery’s status light-emitting diodes (LEDs).
ZOLL recommends that users change AutoPulse Li-Ion Batteries daily or after each use. Charged AutoPulse Li-Ion Batteries left for an extended period in the AutoPulse or as a spare may not have sufficient capacity to operate effectively.
Warning: Always charge a stored Battery before placing the Battery in active operation. Battery may
self-discharge when not in use. Failure to charge a Battery before use may cause device power failure. In no case should any Battery be used if it has not been charged within 2 days.
Caution: Remove the protective plastic cap from the Battery before attempting to charge the Battery. Caution: Only use ZOLL Batteries specifically designed for use with the AutoPulse System. The use
of other batteries may cause permanent damage to the AutoPulse Platform and will void the warranty.
P/N. 12457-001 Rev. 2 Page 2-1
Power and Communications
Power System User Guide
Remove Protective Plastic Cap before use
Status Check Button
Figure 2-1 The AutoPulse Li-Ion Battery
2.1 Expected AutoPulse Li-Ion Battery Life
The expected service life of a properly maintained AutoPulse Li-Ion Battery is three years from its date of manufacture. ZOLL recommends that customers plan to purchase batteries on intervals that reduce the likelihood that they will have all batteries due for replacement simultaneously. A specific replacement interval may be difficult to establish until use and charging patterns are well established. Some customers may want to replace batteries on a preventative basis regardless of the Battery’s capacity or remaining life.
Note: The Battery will not operate after five years from its date of manufacture.
2.2 Handling New AutoPulse Li-Ion Batteries
New AutoPulse Li-Ion Batteries should be unpacked immediately and not stored for any extended period. Before putting new AutoPulse Li-Ion Batteries into service, they must be placed into the AutoPulse Multi-Chemistry Battery Charger. The Multi-Chemistry Charger will charge and test the Battery, and may automatically initiate a Test-Cycle. If a test cycle is initiated, do not remove the AutoPulse Li-Ion Battery from the Charger until the Test-Cycle is complete. This process could take up to 12 hours.
Page 2-2 P/N. 12457-001 Rev. 2
Power System User Guide

3 AutoPulse NiMH Battery

The AutoPulse NiMH Battery (see Figure 3-1) is a proprietary, rechargeable, removable Nickel-Metal Hydride battery that is a power source for the AutoPulse Platform.
The AutoPulse NiMH Battery is mechanically keyed to the AutoPulse Platform and the AutoPulse Multi-Chemistry Battery Charger to facilitate correct installation. One end of the AutoPulse NiMH Battery contains connections for power and communications. A Battery Status Check button illuminates the AutoPulse NiMH Battery’s status light-emitting diodes (LEDs).
Batteries experience a loss of capacity when left standing. ZOLL recommends AutoPulse NiMH Batteries be changed daily or after each use. Charged AutoPulse NiMH Batteries left for any extended period in the AutoPulse or as a spare may not have sufficient capacity to operate effectively.
Warning: Always charge a stored Battery before placing the Battery in active operation. Battery may
self-discharge when not in use. Failure to charge a Battery before use may cause device power failure. In no case should any Battery be used if it has not been charged within 2 days.
Caution: Remove the protective plastic cap from the Battery before attempting to charge the Battery. Caution: Only use ZOLL Batteries specifically designed for use with the AutoPulse System. The use
of other batteries may cause permanent damage to the AutoPulse Platform and will void the warranty.
P/N. 12457-001 Rev. 2 Page 3-1
Power and Communications
Green LED
Amber LED
Red LED
Power System User Guide
Remove Protective Plastic Cap before use
Status Check Button
Figure 3-1 AutoPulse System NiMH Battery
3.1 Expected AutoPulse NiMH Battery Life
The expected service life of the AutoPulse NiMH Battery is 100 full charge/discharge cycles or 2 to 4 years depending on battery maintenance and usage patterns. ZOLL recommends that customers plan to purchase batteries on intervals that reduce the likelihood that they will have all batteries due for replacement simultaneously. A specific replacement interval may be difficult to establish until use and charging patterns are well established. Some some customers may want to replace batteries on a preventative basis regardless of the Battery’s capacity or remaining life.
Note: The NiMH Battery will not operate after 100 full charge/discharge cycles.
3.2 Handling New AutoPulse NiMH Batteries
New AutoPulse NiMH Batteries should be unpacked immediately and not stored for any extended period. Before putting new AutoPulse NiMH Batteries into service they must be placed into the AutoPulse Multi-Chemistry Battery Charger , and run through a Test-Cycle. A Test-Cycle is initiated by the Charger automatically, and no user action is required. Once a test cycle is initiated, do not remove the AutoPulse NiMH Battery from the Charger until the Test-Cycle is complete. This process could take up to 12 hours.
Page 3-2 P/N. 12457-001 Rev. 2
Power System User Guide
User Control Panel
Battery Charging Bays
AC Power Cord, as supplied

4 The AutoPulse Multi-Chemistry Battery Charger

The AutoPulse Multi-Chemistry Battery Charger is a stand alone unit intended to charge and automatically maintain both the AutoPulse Li-Ion Battery and the AutoPulse NIMH Battery. The AutoPulse Multi-Chemistry Battery Charger has two charging bays, each with its own indicators. The AutoPulse Multi-Chemistry Battery Charger is designed to charge, maintain, and test batteries for optimum performance in the AutoPulse.
Batteries should always be properly maintained and be fully charged so that they are ready for use before deploying the AutoPulse.
Figure 4-1 AutoPulse Multi-Chemistry Battery Charger
4.1 Setting Up the AutoPulse Multi-Chemistry Battery
The AutoPulse Multi-Chemistry Battery Charger must be installed near the wall outlet to which it is connected and that outlet must be easily accessible at all times. Unplug the power cord from the wall outlet to remove power from the AutoPulse Multi-Chemistry Battery Charger.
Caution: Do not block the AutoPulse Multi-Chemistry Battery Charger’s ventilation slots. Caution: Do not operate the AutoPulse Multi-Chemistry Battery Charger in a confined space. Caution: Do not position the Multi-Chemistry Battery Charger so that it is difficult to unplug the power
P/N. 12457-001 Rev. 2 Page 4-1
Charger
cord.
Power System User Guide
Caution: Keep the AutoPulse Multi-Chemistry Battery Charger away from moisture. Caution: Use the AutoPulse Multi-Chemistry Battery Charger only with a ZOLL AutoPulse Multi-
Chemistry Battery Charger Power Cord.
To prepare the AutoPulse Multi-Chemistry Battery Charger for use:
1. Plug the alternating current (AC) power cord into the power receptacle on the back of the AutoPulse Multi-Chemistry Battery Charger.
2. Plug the AC power cord into an appropriate wall outlet receptacle.
3. When the AutoPulse Multi-Chemistry Battery Charger is powered on, all the LED’s illuminate briefly as the Charger performs a self test.
Note: While the self test is being performed, if the indicator lights remain illuminated or if one of the
indicator lights fails to illuminate, contact ZOLL.
4. When the power indicator (green LED) on the AutoPulse Multi-Chemistry Battery Charger’s Control Panel illuminates, the AutoPulse Multi-Chemistry Battery Charger is ready for use. (If the green Power light on the Control Panel does not illuminate, read Section 7.2, “Replacing a AutoPulse Multi-Chemistry Battery Charger Fuse”)
Note: USB port is for technician use only. Note: The AutoPulse Multi-Chemistry Battery Charger is a Class II type equipment where no
protective earthing/grounding is provided.
4.2 Operating the AutoPulse Multi-Chemistry Battery
Charger
The AutoPulse Multi-Chemistry Charger charges and maintains both the AutoPulse Li-Ion Battery and the AutoPulse NiMH Battery.
To charge a Battery, follow these steps:
1. Slide the Battery into an available charging bay (see Figure 4-2). Make sure that the Battery locks into place (locking bar engaged).
Caution: Remove the protective plastic cap from the Battery Connector before attempting to charge the
Battery.
Note: For optimal charging, make sure that the Battery is at room temperature before insertion into
the AutoPulse Multi-Chemistry Battery Charger.
Note: If a Li-Ion Battery's internal temperature is below a nominal 41ºF (5ºC), the Battery will fail to
charge in the Multi-Chemistry Battery Charger. If a Battery is retrieved from cold storage or extensive exposure to cold weather, allow the Battery to warm to room temperature (may take up to three hours) before insertion into the Charger.
Page 4-2 P/N. 12457-001 Rev. 2
Power System User Guide
Note: Do not slam a Battery into a Charger because doing so may cause damage to the Battery’s
Connector.
Note: Both Batteries are mechanically keyed so that they can only be inserted in one orientation. Do
not force a Battery into a charging bay. If resistance is met, check for appropriate orientation, and check to ensure that there are no obstructions to battery insertion.
Figure 4-2 Sliding the Battery into a Multi-Chemistry Battery Charger Charging Bay
2. The Multi-Chemistry Battery Charger automatically detects the presence of a Battery within 5 seconds.
3. The Multi-Chemistry Battery Charger’s status will be indicated on the control panel (see Figure 4­3 and Table 4-1).
Figure 4-3 Multi-Chemistry Charger Control Panel
P/N. 12457-001 Rev. 2 Page 4-3
Power System User Guide
4. AutoPulse Li-Ion batteries placed in a charging bay are automatically charged and tested for minimal performance (Charging LED illuminated) in less than 4
¼ hours.
5. AutoPulse NiMH batteries placed in a charging bay are automatically charged and tested for minimal performance (Charging LED illuminated) in less than 6
¼ hours.
Note: Do not remove a Battery from the Multi-Chemistry Battery Charger until its charging
completes, or the Battery’s run time will be reduced.
6. The AutoPulse Multi-Chemistry Battery Charger automatically conducts a Performance Test on both the AutoPulse Li-Ion Battery and the AutoPulse NIMH Battery to ensure each Battery meets performance standards. The Performance Test occurs every time a Battery is placed into the AutoPulse Multi-Chemistry Battery Charger.
7. Once the green LED on the Charger illuminates the Battery is fully charged and has successfully passed
the Performance Test. The Battery is ready to use.
Table 4-1 AutoPulse Multi-Chemistry Battery Charger Status LEDs (Page 1 of 2)
Multi­Chemistry Battery Charger Mode
Charge Charging (yellow LED) The Battery is charging. Leave the Battery in the Charger until
Test-Cycle Charging (yellow LED)
Ready Ready (green LED) The Battery is fully charged and
Multi-Chemistry Battery Charger Status LEDs Used
Test (amber LED)
Definition Action
the Ready (green) LED illuminates.
T ypical Test-Cycles last up to 12 hours.
Leave the Battery in the Charger until the Test-Cycle completes and the ready (green) LED illuminates.
Leave the Battery in the Charger
has successfully passed the Performance Test. The Battery is ready to use.
to ensure that the Battery is fully charged when needed or
Install in the AutoPulse Platform
or
Store in a cool location.
Fail Fail (red LED)
The Charger was unable to
charge the Battery or
the Battery has failed the
Performance Test or
Remove and re-insert the Battery into the Charger. If the Fail LED remains illuminated, contact ZOLL.
the Battery has failed a test
cycle or
the Battery has reached its
end of life.
Page 4-4 P/N. 12457-001 Rev. 2
Power System User Guide
Table 4-1 AutoPulse Multi-Chemistry Battery Charger Status LEDs (Page 2 of 2)
Multi­Chemistry Battery Charger Mode
Idle The Battery Charger is unable to
Multi-Chemistry Battery Charger Status LEDs Used
Definition Action
Remove and re-insert the Battery. If
recognize the Battery.
the status is still Idle, refer to Appendix B, “Troubleshooting” for more information.
Note: Newly-charged Batteries can be warm to the touch. This is a consequence of normal operation.
4.2.1 AutoPulse Multi-Chemistry Battery Charger Test-Cycle mode
In addition to the Performance Test, the AutoPulse Multi-Chemistry Battery Charger periodically executes a more extensive test on the Battery. The “Test-Cycle” measures a Battery’s charge holding capability by cycling the Battery through a charge-discharge-recharge sequence. At the end of the sequence, the Battery is tested to ensure performance standards are met.
The AutoPulse Multi-Chemistry Battery Charger will automatically perform a Test-Cycle every 10th charge/discharge cycle or at a minimum of every 30 days. When a Battery is placed in the Multi­Chemistry Battery Charger under those conditions, the Test-Cycle yellow LED will illuminate and the Charger will automatically begin the Test-Cycle.
Note: Do not remove a Battery from the AutoPulse Multi-Chemistry Battery Charger during a Test-
Cycle, or the Battery’s run time may be reduced. If a Battery is removed during a Test-Cycle, the Charger will automatically restart the T est-Cycle the next time the Battery is inserted into it.
AutoPulse Batteries that pass the Test-Cycle remain ready for use. Once the Test-Cycle has been completed, the Battery is either restored to ready (green LED on the Control Panel) or has failed and should be replaced (red LED on the Control Panel). The normal Test-Cycle requires up to 12 hours.
If a Battery has failed the Test-Cycle (Fail LED), it should be considered out of service. In the U.S., contact ZOLL Technical Service at 1-800-348-9011. Outside the U.S., contact your local ZOLL representative.
P/N. 12457-001 Rev. 2 Page 4-5
Power System User Guide
[This page left intentionally blank.]
Page 4-6 P/N. 12457-001 Rev. 2
Power System User Guide

5 Using the AutoPulse Battery

5.1 AutoPulse Li-Ion Battery Status Check
To determine if an AutoPulse Li-Ion Battery needs to be charged, press the Status Check button on the AutoPulse Li-Ion Battery (see Figure 5-1).
Status Check Button
Figure 5-1 AutoPulse Li-Ion Battery Status Check Button and Status LEDs
The AutoPulse Li-Ion Battery status light-emitting diodes (LEDs) will illuminate (refer to Table 5-1).
Table 5-1 AutoPulse Li-Ion Battery Status LEDs (Page 1 of 2)
Status LEDs Definition Action
Green The AutoPulse Li-Ion Battery is fully
charged.
Y ellow The AutoPulse Li-Ion Battery is not fully
charged.
Green flashing The AutoPulse Li-Ion Battery has
exceeded its expected service life of three years from its date of manufacture; nonetheless, it is fully charged.
Yellow flashing The AutoPulse Li-Ion Battery has
exceeded its expected service life, and is not fully charged.
The AutoPulse Li-Ion Battery is ready for use in the AutoPulse.
Charge the AutoPulse Li-Ion Battery. Refer to Chapter 4, “Operating the AutoPulse Multi-Chemistry Battery Charger” for more information.
ZOLL recommends replacement of Batteries that have exceeded their expected service life; however, this AutoPulse Li-Ion Battery is fully charged, and can be used in the AutoPulse.
ZOLL recommends replacement of Batteries that have exceeded their expected life. Charge the AutoPulse Li-Ion Battery to determine if it remains functional. Refer to Chapter 4, “Operating the AutoPulse Multi-Chemistry Battery Charger” for more information.
Red-flashing The AutoPulse Li-Ion Battery has
failed the Performance Test
failed a Test-Cycle
exceeded five years from its date of
manufacture
P/N. 12457-001 Rev. 2 Page 5-1
The AutoPulse Li-Ion Battery has failed and should not be used. Refer to Section 6.2.6, “Disposing of AutoPulse Batteries” for more information.
Power System User Guide
Table 5-1 AutoPulse Li-Ion Battery Status LEDs (Page 2 of 2)
Status LEDs Definition Action
None The AutoPulse Li-Ion Battery voltage is
too low to illuminate the LEDs.
Charge the AutoPulse Li-Ion Battery by placing it into the AutoPulse Multi-Chemistry Battery Charger.
5.2 AutoPulse Li-Ion Battery Installation
The AutoPulse Li-Ion Battery is mechanically keyed so that it can only be inserted in one orientation. If resistance is met, check for appropriate orientation, and check to ensure there are no obstructions to Battery insertion.
Finger Latch
Locking Bar
Battery Front
Battery Latch
Battery Latch Closed
Battery Latch Opened
Figure 5-2 AutoPulse Li-Ion Battery Installation and Removal
T o install the AutoPulse Li-Ion Battery, first make sure the Battery Latch is rotated out of the way so that the AutoPulse Li-Ion Battery can slide into the Battery compartment (see Figure 5-2). Then slide the AutoPulse Li-Ion Battery into the Battery compartment in the AutoPulse Platform. The AutoPulse Li­Ion Battery should snap into place and mount flush with the AutoPulse Platform. The Finger Latch for the AutoPulse Li-Ion Battery should also be flush with the AutoPulse Platform such that the red strip on the inside of the Finger Latch is not visible. Then rotate the Battery Latch into position.
Page 5-2 P/N. 12457-001 Rev. 2
Power System User Guide
T o remove the Battery, first rotate the Battery Latch into the open position as shown in Figure 5-2. Then hold the AutoPulse Platform firmly and grip the Battery while pulling the finger latch outwards to disengage the locking bar (see Figure 5-2), then pull the Battery straight out until it fully clears the battery compartment.
Caution: Only use ZOLL Batteries specifically designed for use with the AutoPulse. The use of other
batteries may cause permanent damage to the AutoPulse and will void the warranty.
5.3 AutoPulse NiMH Battery Status Check
T o determine if an Auto Pulse Batte ry needs to be char ged, press the S tatus Check button on the Battery (see Figure 8-1).
Status Check Button
Green LED
Amber LED
Red LED
Figure 5-3 Battery Status Check Button and Status LEDs
One of the following Battery status light-emitting diodes (LEDs) will illuminate (refer to Table 5-2).
Table 5-2 Battery Status LEDs
Status LEDs Definition Action
Green The Battery is fully charged. The Battery
is ready for use.
Amber The Battery is not fully charged. Charge the Battery in the Multi-Chemistry Battery
Red-flashing The Battery has
Failed the Performance Test or
No action is necessary.
Charger. The Battery has failed and should not be used.
Dispose of Battery properly.
Failed a Test-Cycle or
Has exceeded 100 charge/discharge
cycles
None The Battery voltage is too low to
illuminate the LEDs.
Place the Battery into the Multi-Chemistry Battery Charger.
P/N. 12457-001 Rev. 2 Page 5-3
Power System User Guide
5.4 AutoPulse NiMH Battery Installation
T o install the AutoPulse NiMH Battery , first make sure the Battery Latch is rotated out of the way so that the AutoPulse NiMH Battery can slide into the Battery compartment
Then slide the AutoPulse NiMH Battery into the Battery compartment in the AutoPulse Platform. The Battery should snap into place and mount flush with the AutoPulse Platform. Then rotate the Battery Latch into position.
Finger Latch
Locking Bar
Battery Latch
Battery Latch Opened
Figure 5-4 AutoPulse NiMH Battery Installation and Removal
Battery Front
Battery Latch Closed
The AutoPulse NiMH Battery is mechanically keyed so that it can only be inserted in one orientation. If resistance is met, check for appropriate orientation, and check to ensure there are no obstructions to Battery insertion and that the Battery Latch is open. Ensure that the AutoPulse NiMH Battery is securely latched (snaps into place) before moving the AutoPulse Platform or initiating chest compression. Power on the AutoPulse every time the AutoPulse NiMH Battery is installed to ensure it is properly seated and providing power to the AutoPulse Platform.
To remove the Battery, first rotate the Battery Latch into the open position. Then hold the AutoPulse Platform firmly and grip the Battery while pressing the finger latch upwards to disengage the locking bar (see Figure 5-4), then pull the Battery straight out until it fully clears the Battery compartment.
Page 5-4 P/N. 12457-001 Rev. 2
Power System User Guide
Caution: Only use ZOLL Batteries specifically designed for use with the AutoPulse. The use of other
batteries may cause permanent damage to the AutoPulse and will void the warranty.
P/N. 12457-001 Rev. 2 Page 5-5
Power System User Guide
5.5 Battery Rotation
After every use, at the beginning of a shift, or at least once every 24 hours, the Battery in the AutoPulse should be replaced with a fully charged Battery.
Recommended daily 4 Battery maintenance routine
1
Replace the spare Battery
with a freshly-charged Battery from the Charger.
Check the Battery
LED
s before
status placing a Battery in service.
2
Replace the Platform
Battery with a freshly charged Battery from the Charger.
3
Charge the Platform Battery
and the Spare Battery.
Figure 5-5 Four Battery rotation
Four Battery Rotation—to be done after every use and/or once per shift (see Figure 5-5 above).
Remove the Battery from the AutoPulse, and place it into the Charger.
Remove the spare Battery, and place it into the Charger.
T ake two fully charged Ba tteries from the Charger , check for green LEDs on each Battery, and then
place one into the AutoPulse, and use the second as a spare.
Power on the AutoPulse and ensure no faults are displayed.
Page 5-6 P/N. 12457-001 Rev. 2
Power System User Guide
Recommended daily 3 Battery maintenance routine
3 Place the Battery that
was in the AutoPulse into the Charger.
Check the Battery status
LEDs before
placing a Battery in service.
1 Remove fully charged
Battery from the Charger and store it with the AutoPulse as the spare.
2 Install the previous
spare Battery into the AutoPulse.
Figure 5-6 Three Battery rotation
Three Battery Rotation—to be done after every use and/or once per shift (see Figure 5-6 above).
Remove the Battery from the AutoPulse and place into the Charger.
Check the spare Battery for green LED’s , and then place it in the AutoPulse. (If the spare Battery is
not fully charged, place it in the Charger and use one that is fully charged.)
Remove a fully charged Battery from the Charger and use it as the spare Battery.
Power on AutoPulse and ensure no faults are displayed.
P/N. 12457-001 Rev. 2 Page 5-7
Power System User Guide
[This page left intentionally blank.]
Page 5-8 P/N. 12457-001 Rev. 2
Power System User Guide

6 Managing the AutoPulse Power System

6.1 AutoPulse Battery Management
The AutoPulse System is intended to be deployed in a state of high-readiness. Therefore, daily AutoPulse System checks should be integrated into Emergency Medical Service (EMS) rig-check or hospital procedures. AutoPulse Batteries that are not fully charged (Battery Status LED is yellow/amber or there are less than four bars on the AutoPulse User Control Panel), will result in shorter AutoPulse run times. AutoPulse Batteries that are not fully charged should be replaced with ones that are fully charged (green Battery Status LED or has four bars on the AutoPulse User Control Panel). A recommended AutoPulse Daily Checklist is located in Appendix A.
The following essential elements of AutoPulse Battery management should be incorporated into a regular routine:
Leave a fully-charged AutoPulse Battery installed in the AutoPulse Platform at all times.
Have a fully-charged spare AutoPulse Battery available for use in the AutoPulse System.
Maintain one or two fully-charged AutoPulse Batteries in the AutoPulse Multi-Chemistry Battery
Charger.
Caution: Do not use Batteries that have cracks in the Battery case exposing internal components.
Mishandling of the Battery may lead to physical damage and present a fire or shock hazard.
Caution: Do not immerse any portion of the AutoPulse Battery in water or other fluids. Do not allow
fluids to enter the Battery or the Battery Connector. Fluid immersion or spillage may permanently damage the Battery or present a fire or shock hazard.
6.2 AutoPulse Battery Maintenance
6.2.1 Cleaning the AutoPulse Battery
Wipe all the surfaces of the Battery free of foreign matter and spills with a clean dry cloth or a disinfectant wipe such as Super Sani-Cloth (or equivalent).
Caution: Clean the Battery Connector and contacts only with a clean dry cloth and/or a nonconductive
brush.
Caution: Do not autoclave the AutoPulse Battery or the Multi-Chemistry Battery Charger. Clean the surfaces of the Battery with one of the following approved cleaning products:
70% Isopropyl alcohol
Chlorine bleach solution (3% bleach, 97% tap water)
Super Sani-Cloth
P/N. 12457-001 Rev. 2 Page 6-1
Power System User Guide
Wipe the Battery completely dry with a clean dry cloth. Ensure the Battery is completely dry prior to placement in the AutoPulse Platform or the Multi-Chemistry Battery Charger.
Inspect the Battery per Section 6.2.2.
6.2.2 AutoPulse Battery Inspection
The Battery should be physically and visually inspected on a regular basis to ensure it is in a state of operational readiness.
Caution: Do not use a Battery that has cracks in the Battery case exposing internal components. Do not
strike or throw a Battery. Do not use a Battery to strike another object. Mishandling of a Battery may lead to physical damage and present a fire or shock hazard.
Housing Damage
NiMH Battery
Connector Damage
Li-Ion Battery
Figure 6-1 Examples of a damaged Battery connector and housing
Note: If the Battery is damaged, do not attempt to place the Battery into the AutoPulse- this can cause
damage to the internal connector of the AutoPulse.
If damaged, do not use. In the U.S., contact ZOLL Technical Service at 1-800-348-9011. Outside the U.S., contact your local ZOLL representative.
6.2.3 Storing AutoPulse Batteries
You should always have a fully charged AutoPulse Battery installed in the AutoPulse Platform ready for use. Leave any additional AutoPulse Batteries in the AutoPulse Multi-Chemistry Battery Charger. This will ensure that they are fully charged when needed.
Caution: Always charge AutoPulse Batteries at temperatures between 41ºF (5ºC) and 95ºF (35ºC).
Charging AutoPulse Batteries at temperatures below 41ºF (5ºC) or above 95ºF (35ºC) will prevent the AutoPulse Battery from reaching its full capacity (operational time) and may lead to irreversible Battery damage.
If you cannot leave your charged AutoPulse Batteries in the AutoPulse Multi-Chemistry Battery Charger , store them in a cool dry place. Place stored batteries into the char ger prior to use to ensure they are fully charged and ready for use.
Page 6-2 P/N. 12457-001 Rev. 2
Power System User Guide
AutoPulse Batteries stored outside the Multi-Chemistry Charger for longer than 4 weeks may be subject to irreparable damage.
Warning: Always charge a stored Battery before placing the Battery in active operation. Battery may
self-discharge when not in use. Failure to charge a Battery before use may cause device power failure. In no case should any Battery be used if it has not been charged within 2 days.
6.2.4 Reaching the End of AutoPulse Li-Ion Battery Service Life
The expected service life of AutoPulse Li-Ion batteries is three years from its date of manufacture. Refer to Section 5.1, “AutoPulse Li-Ion Battery Status LEDs” for more information.
Note: The AutoPulse Li-Ion Battery will not operate after five years from its date of manufacture.
Once an AutoPulse Li-Ion Battery has reached the end of its service life, you should discontinue use of the AutoPulse Li-Ion Battery . Dispose of it properly. Refer to Section 6.2.6, “Disposing of AutoPulse Batteries” for more information.
Caution: Do not attempt to open the AutoPulse Li-Ion Battery. The AutoPulse Li-Ion Battery has no
serviceable parts.
6.2.5 Reaching the End of AutoPulse NiMH Battery Service Life
The expected service life of AutoPulse NiMH batteries is two to four years. Note: The AutoPulse NiMH Battery will not operate after 100 full charge/discharge cycles.Once a
AutoPulse NiMH Battery has reached the end of its service life, you should discontinue use of the AutoPulse NiMH Battery. Dispose of it properly. Refer to Section 6.2.6, “Disposing of AutoPulse Batteries” for more information.
Caution: Do not attempt to open the AutoPulse NiMH Battery. The AutoPulse NiMH Battery has no
serviceable parts.
6.2.6 Disposing of AutoPulse Batteries
Do not throw your batteries away or send them to municipal dumps. Call your local waste management officials for proper disposal instructions.
Caution: Do not heat, burn, or incinerate a AutoPulse Battery. Exposure to heat above 158ºF (70ºC)
may irreversibly damage the AutoPulse Battery.
P/N. 12457-001 Rev. 2 Page 6-3
Power System User Guide
[This page left intentionally blank.]
Page 6-4 P/N. 12457-001 Rev. 2
Power System User Guide
7 AutoPulse Multi-Chemistry Battery Charger
Maintenance
Warning: No modification of the AutoPulse Multi-Chemistry Battery Charger, the AutoPulse Li-Ion
Battery, or the AutoPulse NiMH Battery is allowed.
Caution: Do not remove the AutoPulse Multi-Chemistry Battery Char ger cover. The AutoPulse Multi-
Chemistry Battery Charger has no internal user serviceable parts.
7.1 Cleaning the AutoPulse Multi-Chemistry Battery Charger
Clean the external surfaces of the AutoPulse Multi-Chemistry Battery Charger at least once a month only with a lint-free cloth that is either dry or slightly damp with water.
7.2 Replacing a AutoPulse Multi-Chemistry Battery Charger Fuse
The only user-serviceable parts on the AutoPulse Multi-Chemistry Battery Charger are the alternating current (AC) power fuses. To check if the fuse has burned out, follow these steps:
1. Unplug the Power Cord from the wall outlet and from the receptacle on the back of the AutoPulse
Multi-Chemistry Battery Charger. Wait one minute before going to step 2.
2. The fuse holder is located directly under the power receptacle on the back of the AutoPulse Multi-
Chemistry Battery Charger (see Figure 7-1). Open the fuse holder by pressing down on the locking tab and pulling the fuse holder straight out.
Battery Charger Fuse Holder
Figure 7-1 AutoPulse Multi-Chemistry Battery Charger Fuse Location
P/N. 12457-001 Rev. 2 Page 7-1
Power System User Guide
3. Check both of the fuses. If a fuse must be replaced, follow these steps: a) Replace both of the original fuses with T2.5A 250V AC fuses. b) Push in the fuse holder until the locking tab clicks into place. c) Re-attach the Power Cord. d) When the AutoPulse Multi-Chemistry Battery Charger is powered on, all the LED’ s illuminate
briefly as the Charger performs a self test.
Note: While the self test is being performed, if the indicator lights remain illuminated or if one
of the indicator lights fails to illuminate, contact ZOLL.
Page 7-2 P/N. 12457-001 Rev. 2
Power System User Guide
Appendix A AutoPulse Daily Checklist
Date
Battery Serial Numbers
AutoPulse Spare
AutoPulse
Self Test
Performed
Perform AutoPulse Self Test - Install a LifeBand and power on the AutoPulse. On Power up, all LEDs on the display will momentarily light up, then only the Green POWER LED will remain lit. The Battery Charge Status Icon on the AutoPulse User Control Panel should also be visible with 4 bars, if not replace Battery with a fully charged Battery from the Multi-Chemistry Battery Charger. If the AutoPulse Platform red ALERT LED remains on, refer to the Chapter 5, “Troubleshooting Procedures” in the AutoPulse User Guide. If the RED LED can not be resolved, contact ZOLL.
P/N. 12457-001 Rev. 2 Page A-1
Power System User Guide
[This page left intentionally blank.]
Page A-2 P/N. 12457-001 Rev. 2
Power System User Guide
Appendix B Troubleshooting
This Appendix details symptoms, possible causes, and recommended actions for difficulties you might have with your AutoPulse Power System. T able B-1 provides troubleshooting procedures for the Battery . Table B-2 provides troubleshooting procedures for the Multi-Chemistry Battery Charger.
Table B-1 Battery Troubleshooting Procedures
Symptom Possible Cause Recommended Action
Battery Status Check LEDs do not illuminate.
Battery will not fully insert into the Multi-Chemistry Battery Charger.
The Battery’s status is unknown.
Protective plastic cap
was left on the Battery.
The Battery may be
damaged.
The Battery Charger’s
charging bay might be obstructed.
Place the Battery in one of the charging bays of the Multi-Chemistry Battery Charger.
1. If its Charging LED is illuminated, the Multi­Chemistry Battery Charger is attempting to restore the Battery. Refer to Chapter 4, “Operating the AutoPulse Multi-Chemistry Battery Charger” for more information.
2. If the Charger’s Fail LED is illuminated, the Battery has failed. Replace the Battery. Refer to Section 6.2.4, “Reaching the End of AutoPulse Li-Ion Battery Service Life” or Section 6.2.5, “Reaching the End of AutoPulse NiMH Battery Service Life” for more information.
Remove protective cap.
Inspect the guide rails around the Connector for
damage. If the guides are damaged, replace the Battery.
Inspect the Battery Connector for damage. If the
Connector is damaged, replace the Battery.
Unplug the alternating current (AC) from the
Multi-Chemistry Battery Charger. Check the charging bay to ensure that no debris has accumulated in the bay.
The Battery will not fully insert into the AutoPulse Platform.
P/N. 12457-001 Rev. 2 Page B-1
Protective plastic cap
was left on the Battery.
The Battery may be
damaged.
The AutoPulse
Platform’s Battery compartment might be obstructed.
Remove protective cap.
Inspect the guide rails around the Connector for
damage. If the guides are damaged, replace the Battery.
Inspect the Battery Connector for damage. If the
Connector is damaged, replace the Battery.
Check the AutoPulse Platform’s battery
compartment to ensure that no debris has accumulated in the bay.
Power System User Guide
Table B-2 Multi-Chemistry Battery Charger Troubleshooting Procedures (Page 1
of 2)
Symptom Possible Cause Recommended Action
Multi-Chemistry Battery Charger’s green Power LED is not illuminated.
Multi-Chemistry Battery Charger’s green Power LED is not illuminated.
Charging a Li-Ion Battery takes much longer than 4¼ hours, or charging an NiMH Battery takes much longer than 6¼ hours.
Multi-Chemistry Battery Charger’s red Fail LED is illuminated.
Multi-Chemistry Battery Charger’s alternating current (AC) power cord is not plugged in.
Blown fuse. Refer to Section 7.2, “Replacing a AutoPulse Multi-
The ambient temperature around the Multi-Chemistry Battery Charger is too warm.
Refer to Chapter 4, “The AutoPulse Multi-Chemistry Battery Charger” for more information.
Chemistry Battery Charger Fuse” for more information.
Make sure that the Charger is located in an
environment where temperatures do not reach above 95°F (35°C).
Make sure that the Charger’s vents are not
blocked.
Make sure that the Charger has adequate
ventilation.
The Battery has
Failed to charge or
Failed the
Performance Test or
Failed the Test-Cycle
or
Reached its end of life
Remove the Battery from the Charger. Perform a Battery status check:
1. If the Battery status LED is flashing red, the Battery has failed. Replace the Battery. Refer to Section 6.2.4, “Reaching the End of AutoPulse Li-Ion Battery Service Life” or Section 6.2.5, “Reaching the End of AutoPulse NiMH Battery Service Life” for more information.
2. If no status LEDs illuminate when you press the Battery’s Status Check button, the Battery has failed. Replace the Battery. Refer to Section
6.2.4, “Reaching the End of AutoPulse Li-Ion Battery Service Life” or Section 6.2.5, “Reaching the End of AutoPulse NiMH Battery Service Life” for more information.
3. If a Li-Ion Battery’s internal temperature is below a nominal 41°F (5°C), it will fail to charge. Remove from Charger, allow the Battery to warm to room temperature (may take up to 3 hours), and re-insert in Charger.
4. If the Li-Ion Battery status LED’s are green or yellow, or the NiMH Battery status LED’s are green or amber, remove and re-insert the Battery . If the Charger’s FAIL LED remains illuminated, contact ZOLL.
Page B-2 P/N. 12457-001 Rev. 2
Power System User Guide
Table B-2 Multi-Chemistry Battery Charger Troubleshooting Procedures (Page 2
of 2)
Symptom Possible Cause Recommended Action
One or both of the Battery Bay’s indicator lights are all illuminated.
The Multi-Chemistry Battery Charger has detected an internal error in one or both of the Battery Bays.
Remove the Battery from the Charger . Disconnect the power cord from the wall outlet and then plug the Charger back in. If the Indicator lights remain illuminated (the Charger has failed the self-test), contact ZOLL.
P/N. 12457-001 Rev. 2 Page B-3
Power System User Guide
[This page left intentionally blank.]
Page B-4 P/N. 12457-001 Rev. 2
Power System User Guide
Appendix C Technical Specifications
The specifications provided in this appendix apply to the AutoPulse Power System.
C.1 Li-Ion Battery Physical and Environmental
Table C-1 Li-Ion Battery Specifications (Page 1 of 2)
Category Specifications
Manufacturer ZOLL Circulation Model Number 8700-0752-01 Size (LWH) 11.5 in. by 3.2 in. by 2.2 in. (29.2 cm by 8.1 cm by 5.7 cm). Weight 3.0 lbs. (1.3 kg). Type Rechargeable Lithium-Ion (LiFePO Battery voltage (nominal) 36.3V Capacity 2300 mAh (typical) Current (maximum) 30 A continuous, 48 A pulse (96 ms max) Initial Battery run time
(nominal patient) Maximum Battery charge
time Battery Test-Cycle time Less than 12 hours per Test-Cycle session Recommended replacement
interval
Operating temperature +32º to +113ºF (0º to +45ºC) ambient installed in device Charge temperature +41º to +95ºF (5º to +35ºC) ambient (68º to 77ºF [20º to 25ºC] preferred) Storage temperature -4º to +113ºF (-20º to +45ºC) ambient for up to six months with charging every
Operating altitude 0 to 15,000 ft. (0 to 4,572 m) Enclosure protection Meets IP24 per IEC 60529
30 minutes (typical)
Less than 4¼ hours at 77ºF (25ºC)
3 years from date of manufacture Note: The Battery will not operate after 5 years from date of manufacture.
four weeks, starting with a fully charged Battery.
)
4
Shock Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)
Vibration Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
Free fall Meets IEC 60068-2-31 Basic Environmental Testing Procedures – Free Fall –
Electrostatic discharge Meets IEC 61000-4-2, Level 3
P/N. 12457-001 Rev. 2 Page C-1
2
10 m/s Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
Procedure 1.
)
Power System User Guide
Table C-1 Li-Ion Battery Specifications (Page 2 of 2)
Category Specifications
Radiated emissions Meets CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A
Radiated Immunity Meets IEC-61000-4-3, 80-2500 MHz, Level 3
Safety Meets IEC-60601-1 including UL310DV.1.1 for Lithium batteries
C.2 NiMH Battery Physical and Environmental
Table C-2 NiMH Battery Specifications (Page 1 of 2)
Category Specifications
Manufacturer Manufactured for ZOLL Circulation Model Number 8700-0702-01 Size (LWH) 11.5 in. by 3.2 in. by 2.2 in. (29.2 cm by 8.1 cm by 5.7 cm). Weight 5.1 lbs. (2.3 kg). Type Rechargeable Nickel-Metal Hydride (NiMH) Battery voltage (nominal) 32.4V Capacity 3200 mAh (typical) Initial Battery run time
(nominal patient) Maximum Battery charge
time Battery Test-Cycle time Less than 12 hours per Test-Cycle session Required replacement interval 100 full charge/discharge cycles.
Operating temperature +32º to +113ºF (0º to +45ºC) ambient installed in device Charge temperature +41º to +95ºF (5º to +35ºC) ambient (68º to 77ºF [20º to 25ºC] preferred) Storage temperature -4º to +95ºF (-20º to +35ºC) ambient for up to six months with charging every
Operating altitude 0 to 15,000 ft. (0 to 4,572 m) Enclosure protection Meets IP24 per IEC 60529
30 minutes (typical)
Less than 6¼ hours at 77ºF (25ºC)
Note: The Battery will not operate after 100 full charge/discharge cycles.
four weeks, starting with a fully charged Battery.
Shock Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
11ms pulse, half sine wave)
Vibration Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
Page C-2 P/N. 12457-001 Rev. 2
2
10 m/s Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
)
Power System User Guide
Table C-2 NiMH Battery Specifications (Page 2 of 2)
Category Specifications
Free fall Meets IEC 60068-2-31 Basic Environmental Testing Procedures – Free Fall –
Procedure 1. Electrostatic discharge Meets IEC 61000-4-2, Level 3 Radiated emissions Meets CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A
C.3 Multi-Chemistry Battery Charger Physical And
Environmental
Table C-3 Multi-Chemistry Battery Charger Specifications (Page 1 of 2)
Category Specifications
Manufacturer Manufactured for ZOLL Circulation Model Number 8700-0753-01 Size (LWH) 16.01 in. by 9.50 in. by 6.54 in. (40.6 cm by 24.1 cm by 16.6 cm). Weight 7.1 lbs. (3.23 kg) Operating input voltage 100 to 240V AC Operating input frequency 50/60 Hz Input current 2.0 Amps (maximum) Maximum Battery charge
time Fuses User-replaceable, T2.5A 250V AC (2 required) Operating temperature +41º to +95ºF (5º to +35ºC) (68º to 77ºF [20º to 25ºC] preferred) Storage temperature -40º to +158ºF (-40º to +70ºC) Relative humidity 5% to 95%, non-condensing. Operating altitude 0 to 10,000 ft. (0 to 3,048 m) Enclosure protection Meets IP22 per IEC 60529 Shock Meets IEC 60068-2-27 Basic Environmental Testing Procedures – Shock (50g,
Vibration Meets IEC 60068-2-6 Basic Environmental Testing Procedures (10 to 150 Hz,
Less than 6¼ hours (at 77ºF [25ºC])
11ms pulse, half sine wave)
2
10 m/s
Meets IEC 60068-2-64 Basic Environmental Testing Procedures – Random
Vibration Broad Band – General Requirements (f1:20, f2:2000, ASD 0.05)
)
Free fall Meets IEC 60068-2-31 Basic Environmental Testing Procedures – Free Fall –
Procedure 1. Electrostatic discharge Meets IEC 61000-4-2, Level 3
P/N. 12457-001 Rev. 2 Page C-3
Power System User Guide
Table C-3 Multi-Chemistry Battery Charger Specifications (Page 2 of 2)
Category Specifications
RF electromagnetic fields immunity
EFT/burst Meets IEC 61000-4-4, Level 2 Surge immunity Meets IEC 61000-4-5, Level 2 Conducted RF disturbances
immunity Dips, interruptions, and
variations Harmonics current emissions Meets IEC 61000-3-2, Class A Radiated emissions Meets CISPR 11/EN55011, Group 1, Class A
Safety Meets IEC/EN60601-1
Meets IEC 61000-4-3, Level 2
Meets IEC 61000-4-6, Class A
Meets IEC 61000-4-11
FCC part 15, Class A
Note: These requirements provide reasonable protection against harmful electromagnetic interference in a typical medical installation. However, high level of radio-frequency emissions from electrical devices, such as cellular phones, may disrupt the performance of this device. To mitigate disruptive electromagnetic interference, position this device away from radio frequency transmitters and other sources of electromagnetic energy.
C.4 FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Page C-4 P/N. 12457-001 Rev. 2
Power System User Guide
C.5 Limited Warranty for AutoPulse Resuscitation System
ZOLL Circulation warrants to the initial Purchaser only that the “Warranted Product” purchased hereunder will be free from defects in workmanship or materials, when given normal, proper, and intended usage, for a specified period (“Warranty Period”) from the date of its initial shipment to Purchaser. “Warranted Products” consist solely of those products whose description in this price list expressly states that the product includes a warranty for a specified time period (the W arranty Period for the product). Excluded from this warranty are expendable components and supply items such as the LifeBand® Load-distributing Band.
Warranty Period: The AutoPulse Resuscitation System Platform, the AutoPulse Li-Ion Battery, the AutoPulse NiMH Battery and the AutoPulse Multi-Chemistry Battery Charger (collectively and individually referred to as “Product”) are sold with a one year warranty period to the end-user. The warranty period begins at delivery.
ZOLL Circulation’s sole obligations under this warranty are to repair or replace, at its option, any Warranted Product (or part thereof) that ZOLL Circulation reasonably determines to be covered by this warranty and to be defective in workmanship or materials provided that the Purchaser has given notice of such warranty claim within the Warranty Period and the Purchaser has complied with ZOLL Circulation’s Return Material Authorization (“RMA”) procedures. Repair or replacement of Products under this warranty does not extend the Warranty Period.
T o request repair or replacement under this warranty , Purchaser should contact ZOLL Circulation at 650 Almanor Avenue, Sunnyvale, CA 94085, 1-800-321-4CPR or 1-408-541-2140. ZOLL Circulation will inform purchaser of its then-current RMA procedure. ZOLL Circulation shall determine whether to repair or replace Products and parts covered by this warranty and all Products or parts replaced shall become ZOLL Circulation’s property. In the course of warranty service, ZOLL Circulation may but shall not be required to make engineering improvements to the Warranted Product or part thereof.
Exclusions
This warranty does not extend to any Warranted Products or parts thereof that have (a) been subject to misuse, neglect or accident; (b) been damaged by causes external to the Warranted Product, including but not limited to failure of or faulty electrical power; (c) not been used in accordance with ZOLL Circulation’s instructions; (d) been affixed to any nonstandard accessory attachment; (e) had the serial number removed or made illegible; (f) been modified by anyone other than ZOLL Circulation; (g) been used with any software not provided by ZOLL Circulation; or (h) been disassembled, serviced, or reassembled by anyone other than ZOLL Circulation, unless authorized by ZOLL Circulation. ZOLL Circulation shall have no obligation to make repairs, replacements, or corrections which result, in whole or in part, from normal wear and tear.
ZOLL Circulation makes no warranty (a) with respect to any products that are not Warranted Products, (b) with respect to any products purchased from a person other than ZOLL Circulation or a ZOLL Circulation-authorized distributor or (c) with respect to any product sold under a brand name other than ZOLL Circulation.
P/N. 12457-001 Rev. 2 Page C-5
Power System User Guide
THIS WARRANTY IS THE SOLE AND EXCLUSIVE WARRANTY FOR ZOLL CIRCULATION’S PRODUCTS, EXTENDS ONLY TO THE PURCHASER AND IS EXPRESSLY IN LIEU OF ANY OTHER EXPRESS OR IMPLIED WARRANTIES INCLUDING WITHOUT LIMITATION ANY WARRANTY AS TO MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. ZOLL CIRCULATION’S MAXIMUM LIABILITY ARISING OUT OF THE SALE OF THE PRODUCTS OR THEIR USE, WHETHER BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE, SHALL NOT EXCEED THE ACTUAL PAYMENTS RECEIVED BY ZOLL CIRCULATION IN CONNECTION THEREWITH. ZOLL CIRCULATION SHALL NOT BE LIABLE FOR ANY INCIDENTAL, SPECIAL OR CONSEQUENTIAL LOSS, DAMAGE OR EXPENSE (INCLUDING WITHOUT LIMITATION LOST PROFITS) DIRECTLY OR INDIRECTLY ARISING FROM THE SALE, INABILITY TO SELL, USE OR LOSS OF USE OF ANY PRODUCT (HOWEVER CAUSED AND ON ANY THEOR Y OF LIABILITY), EVEN IF ZOLL CIRCULATION HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH LOSS. THE FOREGOING LIMITATIONS SHALL NOT APPLY TO ANY CLAIMS FOR BODILY INJURY OR DEATH TO THE EXTENT THAT LIMITATION OF DAMAGES FOR SUCH CLAIMS ARE UNENFORCEABLE OR AGAINST PUBLIC POLICY UNDER ANY APPLICABLE STATUTE OR RULE OF LAW
Page C-6 P/N. 12457-001 Rev. 2
Power System User Guide
Index X
A
AutoPulse Multi-Chemistry Battery Charger
cleaning 7-1 operating 4-2 replacing fuse 7-1 setting up 4-1 troubleshooting A-1, B-1
AutoPulse Power System
technical specifications C-1 troubleshooting B-1
AutoPulse System
warranty C-5
B
Battery
capacity C-1, C-2 charge time C-1, C-2 disposal 6-3 specifications C-1, C-2 status check 5-1, 5-3 status lights 5-1, 5-3 storage 6-2 test-cycle time C-1, C-2
Battery orientation
diagram 4-3
C
charging bay
status LEDs 4-4
F
FCC statement C-4 fuses, replacing 7-1
storing Batteries 6-2
O
operating the AutoPulse Multi-Chemistry Battery Charger 4-2
P
physical specifications C-3 Power System
setting up 4-1 technical specifications C-1 troubleshooting B-1
R
replacing a fuse 7-1
S
setting up the AutoPulse Multi-Chemistry Battery Charger 4-1 status check 5-1, 5-3 status LEDs
charging bay 4-4
status lights
Battery 5-1, 5-3
storing batteries 6-2
T
technical specifications C-1 test-cycle
time requirements C-1, C-2
troubleshooting A-1, B-1
W
warranty C-5
L
limited warranty C-5
M
maintenance
Battery disposal 6-3 Battery service life 6-3 cleaning AutoPulse Multi-Chemistry Battery
Charger 7-1
replacing fuse 7-1
P/N. 12457-001 Rev. 2 Page I-1
Power System User Guide
[This page left intentionally blank.]
Page I-2 P/N. 12457-001 Rev. 2
Loading...