The issue date for the M Series Service Manual (REF 9650-0450-01 Rev. R) is March, 2013.
If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to determine if additional product information updates
are available.
Masimo is a trademark or registered trademark of Masimo Corporation in the United States
and/or other countries.
9650-0450-01 Rev R
M Series Service Manual
TABLEOF CONTENTS
PREFACE ....................................................................................................................................................................................... V
Additional Reference Material ........................................................................................................................................................... vi
Service Policy Warranty .................................................................................................................................................................... vii
Technical Service ..............................................................................................................................................................................vii
Technical Service for International Customers ................................................................................................................................viii
Before You Begin the Maintenance Tests ............................................................................................................................................2
Equipment that You Need to Perform the M Series Maintenance Tests ..............................................................................................3
Equipment You Need for the M Series Options Maintenance Tests ...................................................................................................4
1.0 Physical Inspection of the Unit...................................................................................................................................................... 5
2.0 Front Panel Button Test................................................................................................................................................................. 7
3.0 3, 5, and 12 Leads Test................................................................................................................................................................ 10
4.0 Power Supply Test (Optional)..................................................................................................................................................... 11
5.0 Leakage Current Test................................................................................................................................................................... 15
6.0 Paddles Test (If applicable)......................................................................................................................................................... 16
8.0 Calibrating Pulses on Strip Chart Test ........................................................................................................................................ 19
13.0 Shock Test ................................................................................................................................................................................. 26
15.0 Advisory Message Test ............................................................................................................................................................. 30
16.0 Pacer Test .................................................................................................................................................................................. 31
i
M Series Service Manual
17.0 SpO2 Monitor Test for SpO2 Option........................................................................................................................................ 33
18.0 EtCO2 Monitor Test (for EtCO2 Option).................................................................................................................................. 36
19.0 Temperature Test........................................................................................................................................................................ 38
23.0 NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer............................................................................................. 49
24.0 Bluetooth Test............................................................................................................................................................................ 51
Zoll M Series Error Messages ............................................................................................................................................................58
Parts That May Need Replacing After Disassembly ..........................................................................................................................76
Overview of Modules .........................................................................................................................................................................78
1. Removing the ZIF Keeper ............................................................................................................................................81
2. Removing the Front Panel ............................................................................................................................................82
2A.Removing the Display .................................................................................................................................................83
2B.Removing the Control Board ......................................................................................................................................84
3. Removing the Upper Housing Assembly .....................................................................................................................86
4. Removing the System Board Assembly .......................................................................................................................87
5. Removing the Battery Interconnect Board Assembly ..................................................................................................90
6. Removing the High Voltage/Charger Assembly .........................................................................................................91
7. Removing the High Voltage Module Assembly ..........................................................................................................94
8. Removing the High Voltage Capacitor Assembly .......................................................................................................96
9. Removing the System Interconnect Board ...................................................................................................................98
10.Removing the Printer/Recorder Motor ......................................................................................................................100
11.Removing the Lower Housing Assembly ..................................................................................................................102
12.Removing the Print Head Assembly ..........................................................................................................................103
13.Removing the PCMCIA Card Slot Assembly ...........................................................................................................104
ii
M Series Service Manual
14.Removing the Paddle Release Latch .........................................................................................................................105
CHAPTER 4 REPLACEMENT PARTS .........................................................................................................................................107
Replacement Parts ............................................................................................................................................................................108
Field Replacement Parts ...................................................................................................................................................................112
Main System Board ..........................................................................................................................................................................116
Main System Board Functions .........................................................................................................................................................118
Power Supply ................................................................................... .................................................................................................120
ECG Front End .................................................................................................................................................................................121
Multifunction Electrode (MFE)/PADS (System Board and High Voltage Module) .......................................................................121
CPU and EPU ...................................................................................................................................................................................121
High Voltage Module .......................................................................................................................................................................122
Defibrillator Charging and Discharging ...........................................................................................................................................123
High Voltage Capacitor Monitor ......................................................................................................................................................124
Pacer/Defibrillator Control Signals ..................................................................................................................................................125
System Interconnect Module ............................................................................................................................................................127
Front Panel and Controls PWBA ......................................................................................................................................................128
M Series Options ..............................................................................................................................................................................128
Isolated Power Supply Module .........................................................................................................................................................129
12 Lead Option .................................................................................................................................................................................129
Interconnect Diagram for the M Series Monophasic Unit ...............................................................................................................136
Interconnect Diagram for the M Series Biphasic Unit .....................................................................................................................137
Interconnect Diagram for the M Series CCT Biphasic Unit ............................................................................................................138
iii
M Series Service Manual
INDEX ..........................................................................................................................................................................................139
iv
M Series Service Manual
Overview
ZOLL Medical Corporation’s M Series Service Manual is intended for the service technician whose responsibility
is to identify malfunctions and/or make repairs at the subassembly level. The Zoll M Series Service Manual has five
main sections and one appendix.
Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to review this section
thoroughly before attempting to use or service the M Series unit.
Chapter 1—Maintenance T ests explains how to check the defibrillator’s performance using a series of recommended
checkout procedures to be conducted every six months.
Chapter 2—Troubleshooting provides a listing of the procedures and error messages to help the service technician
detect faults and repair them.
Chapter 3—Disassembly Procedures describes step-by-step procedures for removing subassemblies from the
M Series unit.
Chapter 4—Replacement Parts List displays a complete list of ZOLL part numbers for field replaceable parts
available for the M Series unit, allowing the service person to identify and order replacement parts from ZOLL.
Chapter 5—Functional Description provides technical descriptions for the M Series major subassembly modules.
Appendix A—M Series Operator’s Manual.
Preface
Safety Considerations
The following section describes general warnings and safety considerations for operators and patients. Service
technicians should review the safety considerations prior to servicing any equipment and read the manual carefully
before attempting to disassemble the unit. Only qualified personnel should service the M Series unit.
Federal (U.S.A.) law restricts this unit for use by or on the order of a physician.
Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug Administration (FDA)
under section 510(K) of the Medical Device Act to obtain approval to market is based upon the use of ZOLL
accessories such as disposable electrodes, patient cables and batteries. The use of external pacing/defibrillation
v
M Series Service Manual
electrodes and adapter units from sources other than ZOLL is not recommended. ZOLL makes no representations or
warranties regarding the performance or effectiveness of its products when used in conjunction with pacing/
defibrillation electrodes and adapter units from other sources. If unit failure is attributable to pacing/defibrillation
electrodes or adapter units not manufactured by ZOLL, this may void ZOLL's warranty.
Only qualified personnel should disassemble the M Series unit.
WARNING! This unit can generate up to 4500 volts with sufficient current to cause lethal shocks.
All persons near the equipment must be warned to “STAND CLEAR” prior to discharging the defibrillator.
Do not discharge the unit’s internal energy more than three times in one minute or damage to the unit may result.
Do not discharge a battery pack except in a Base PowerCharger
Do not use the M Series in the presence of flammable agents (such as gasoline), oxygen-rich atmospheres, or
flammable anesthetics. Using the unit near the site of a gasoline spill may cause an explosion.
Do not use the unit near or within puddles of water.
NOTEThe M Series is protected against interference from radio frequency emissions typical of two-way radios and cellular
phones (digital and analog) used in emergency service/public safety activities. Users of the M Series should assess the
unit’s performance in their typical environment of use for the possibility of radio frequency interference from
high-power sources. Radio Frequency Interference (RFI) may be observed as shifts in monitor baseline, trace
compression, or transient spikes on the display.
4x4
or compatible ZOLL Battery Charging/Testing unit.
Additional Reference Material
In addition to this guide, there are several other components to the Zoll M Series documentation. They include:
• Operator’ s Guide - A comprehensive reference work that describes all the user tasks needed to operate the M Series.
• Configuration Guide - Describes the M Series features and functions whose operation can be customized by
authorized users.
vi
M Series Service Manual
Conventions
WARNING!Warning statements describe conditions or actions that can result in personal injury or death.
CAUTIONCaution statements describe conditions or actions that can result in damage to the unit.
NOTENotes contain additional informationon using the defibrillator.
Service Policy Warranty
In North America: Consult your purchasing agreement for terms and conditions associated with your warranty. Outside
of North America, consult ZOLL authorized representative.
In order to maintain this warranty , the instructions and procedures contained in this manual must be strictly followed.
For additional information, please call the ZOLL Technical Service Department 1-800-348-9011 in North America.
Technical Service
If the ZOLL M Series unit requires service, contact the ZOLL Technical Service Department:
Telephone: 1-978-421-9655; 1-800-348-9011
Fax 1-978-421-0010
Have the following information available for the Technical Service representative:
• Unit serial number.
• Description of the problem.
• Department where equipment is used.
• Sample chart recorder strips documenting the problem, if applicable.
• Purchase Order to allow tracking of loan equipment.
• Purchase Order for a unit with an expired warranty.
vii
M Series Service Manual
If the unit needs to be sent to ZOLL Medical Corporation, obtain a service order request number from the Technical
Service representative. Return the unit in its original container to:
Technical Service for International Customers
International customers should return the unit in its original container to the nearest authorized ZOLL Medical
Corporation Service Center. To locate an authorized service center, cont ac t the International Sales Department at
ZOLL Medical at the above address.
Units are available on loan while your unit is being repaired.
ZOLL Medical Corporation
269 Mill Road
Chelmsford, Massachusetts 01824-4105
Attn: Technical Service Department
Telephone: 1-800-348-9011; 1-978-421-9655 FAX: 978-421-0010
viii
M Series Service Manual
Overview
The M Series has two checkout procedures: the operator’s shift checklist and the extensive six-month maintenance
tests checkout procedures.
Because the M Series units must be maintained ready for immediate use, it is important for users to conduct the
Operator’s Shift Checklist procedure at the beginning of every shift. This procedure can be completed in a few minutes
and requires no additional test equipment. (See the ZOLL M Series Operator’s Guide for the Operator’s Shift
Checklist.)
A qualified biomedical technician must perform a more thorough maintenance test checkout every six months to
ensure that the functions of the M Series unit work properly. This chapter describes the step by step procedures for
performing the six month maintenance test checkout. Use the checklist at the back of this document (ZOLL M Series
Maintenance Tests Checklist) to record your results of the maintenance tests.
Chapter 1
Maintenance Tests
This chapter describes the following M Series maintenance tests:
1. Physical Inspection of the Unit
2. Front Panel Button
3. 3, 5, and 12 Leads
4. Power Supply
5. Leakage Current
6. Paddles
7. Heart Rate Display
8. Calibrating Pulses on Strip Chart
9. Notch Filter
10. Heart Rate Alarm
1
M Series Service Manual
11. Defibrillator Self Test
12. Synchronized Cardioversion
13. Shock
14. Summary Report
15. Advisory Message
16. Pacer
17. SpO
18. EtCO
19. Temperature Test
20. IBP Monitoring Test
21. NIBP Transducer Calibration Test
22. NIBP Monitor Test
23. NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer
24. Bluetooth Test
Monitor
2
Monitor
2
Before You Begin the Maintenance Tests
• Assemble the tools or specialized testing equipment listed in the “Equipment You Need to Perform the Maintenance
Tests” section shown below.
• Keep an extra fully charged ZOLL M Series compatible battery available.
• Schedule an hour to conduct the entire maintenance test.
• Photocopy the checklist at the back of this document and use the copy to record your results. As you con duct each
step of a procedure, mark the
Pass/Fail/NA check boxes on your checklist and then save it for your maintenance file.
• Perform the tests in the order presented.
• Perform all the steps of each test procedure.
• Complete all the steps of the procedure before evaluating the test results.
2
M Series Service Manual
Equipment that You Need to Perform the M Series Maintenance
This section lists equipment that we use to perform the maintenance tests that we describe in this chapter. You can
substitute an equivalent device for a listed device; however, not all simulators and analyzers will produce the same results. Be sure to follow the manufacturer’s recommendations for conducting the maintenance tests.
We recommend the use of the following equipment when performing M Series Maintenance Tests:
• ZOLL Medical Electrode Adapter from Fluke Biomedical (ZOLL Part Number 3010-0378).
• QED 6 Defibrillator Analyzer
• MFC Test Port Connector, 1004-0053-99
• Fluke Biomedical
• ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201).
• Stop watch.
• Standard series II PC flash memory cards.
• 1 red miniature alligator to miniature alligator test lead.
• 1 black miniature alligator to miniature alligator test lead.
• DC power supply (15 amp minimum).
Tests
601 Pro Series International Safety Analyzer.
• 0.1 ohm 1% resistor (1/4 watt or greater).
• 1000 ohm 1% resistor (1/4 watt or greater).
• Fluke 75 Multimeter or equivalent.
•Printer Paper.
• Battery.
• AC line cord.
• 3 lead, 5 lead and 12 lead ECG cables. (12 lead cable needed if 12 lead option is installed.
3
M Series Service Manual
Equipment You Need for the M Series Options Maintenance
• Fluke Biomedical Index 2 SpO2 Simulator or equivalent. (For SpO2 units only.)
•SpO
•EtCO
• Novametrix Capnostat Simulator Yb 1265/7100
• Paddles (if used).
•)
• Fluke Biomedical BP Pump 2 NIBP Monitor Analyzer (For NIBP units only) with NIBP cable and cuff (if NIBP
• Fluke Biomedical Cufflink Analyzer (if NIBP option is installed)
• IBP Temperature Simulator 9100-0402-TF
• DNI Nevada 214B Patient SImulator, with M Series interface cable
Tests
cable and sensor (if option is installed).
2
cable, and CAPNOST AT 3 Mainstream cable with airway adapter, or CAPNOSTAT 3 Sidestream cable with
2
cannula (if option is installed).
option is installed), or
NOTE The Fluke Biomedical BP Pump NIBP Monitor Analyzer and the Fluke Biomedical BP Pump NIBP Analyzer
use different technologies for testing NIBP monitors and therefore, the manual provides two different procedures for
performing the NIBP Volume Leak test with each of these types of test equipment.
4
M Series Service Manual
1.0 Physical Inspection of the Unit
Tools Needed Battery.
Test Setup None.
Observe this...Pass/Fail/NA
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
1.10
1.11
Housing
Is the unit clean and undamaged?
Does the unit show signs of excessive wear?
Does the handle work properly?
Does the recorder door open and close properly?
Are input connectors clean and undamaged?
Are there any cracks in the housing?
Do the front panel or selector switches have any damage or cracks?
Are there any loose housing parts?
Cables
Are all cables free of cracks, cuts, exposed or broken wires?
Are all cable bend/strain reliefs undamaged and free of excessive cable
wear?
Paddles
Do the adult and pedi plates have major scratches or show signs of damage?
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo o
1.12
1.13
Do the adult shoes slide on and off easily to expose the covered pedi plates?
Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
5
oo o
oo o
M Series Service Manual
Observe this...Pass/Fail/NA
1.14
1.15
Record your results on the Maintenance Test Checklist.
Battery
Place battery in battery well.
Is the battery seated in the battery well correctly?
oo
oo
6
M Series Service Manual
2.0 Front Panel Button Test
Tools Needed QED 6 Defibrillator Analyzer
Test Setup• Install strip chart paper into the recorder tray.
• Install the battery in the unit or connect the A/C power cord to the unit and then plug the cord into an electri cal outlet.
• Connect the universal cable and ECG cable (3 lead, 5 lead, or 12 lead) to the QED 6 Analyzer (or equivalent).
2.1
2.2
2.3
2.4
2.5
Do this...Verify that...Pass/Fail/
Turn the selector switch to
MONITOR.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Set the QED 6 analyzer to NSR
of 120 BPM. T o check the size of
the ECG waveform, press the
SIZE button.
Press the ALARM SUSPEND
button.
Press the RECORDER button.The strip chart paper moves out of the unit from the
Open the paper compartment
door, then press the
RECORDER button.
The unit sounds 4 beep tones. PADS and MONITOR
display on the monitor.
Note: PADS is a factory default setting.
As you press the SIZE button five times (0.5, 1.0, 1.5,
2.0, 3.0), the size of the ECG waveform appropriately
changes on the display.
Alarm symbol changes from disabled to enabled. If the
alarm sounds, press the ALARM SUSPEND button to
turn it off. The alarm will only be suspended for 90
seconds at this point. Press and hold the ALARM SUSPEND button for 3 seconds to disable alarms.
paper tray. Check that the correct time, date, ECG
lead annotation and waveform are recorded on the
paper. (Set Time and Date, if necessary.)
The CHECK RECORDER message appears on the
monitor.
NA
oo
oo
ooo
ooo
ooo
2.6
Close the paper compartment
door, then press the
RECORDER button.
The strip chart paper flows out of paper tray and the
monitor no longer displays the CHECK RECORDER
message.
7
ooo
M Series Service Manual
2.7
2.8
2.9
2.10
2.11
2.12
Do this...Verify that...Pass/Fail/
Press the RECORDER button.The strip chart paper stops flowing out of the paper
tray.
Press the VOLUME softkey.The volume bar graph displays.
NA
ooo
oo
Note: The QRS tone is on or off. There is no gradual
change in volume. If equipped, voice prompts
are gradual. Note: The voice volume has 5
settings. Setting 3 is in the mid-range.
Press the INC softkey.The bar graph appears on the display, indicating an
increase in volume.
oo
Note: This action does not increase the volume of the
unit’s audio prompts.
Press the DEC softkey.The bar graph appears on the display, indicating a
decrease in volume.
Verify that pressing the DEC softkey until only the last
bar appears silences the unit’s audio prompts,
otherwise the volume of the audio prompts does not
decrease.
Press the CONTRAST button. The monitor screen displays the Contrast Menu.
Press the INC softkey. The screen contrast and bar graph increases on the
display.
oo
oo
ooo
2.13
2.14
2.15
2.16
Press the DEC softkey.The screen contrast and bar graph decreases on the
display.
Press the SUMMARY button.The monitor screen displays the Summary menu,
showing the summary report options.
Press the CODEMARKER
button.
Connect to AC or DC current
and install the battery. Turn the
unit off.
The monitor screen displays the Code marker menu.
The CHARGER ON indicator lights (either the Green
or Amber indicator lights).
8
oo
ooo
ooo
ooo
M Series Service Manual
2.17
2.18
2.19
2.20
2.21
2.22
2.23
Do this...Verify that...Pass/Fail/
Remove the battery.The charger light alternately flashes green and amber.
Replace the battery and the turn
unit on.
Press the ANALYZE button.The SELECT DEFIB MODE message appears on the
Connect the universal cable to
the QED 6. Move the selector
switch to DEFIB. Select 2J.
Press the CHARGE button.
Press and hold the ENERGY SELECT down arrow.
Press and release the ENERGY SELECT up arrow 19 times.
Press the CHARGE button.The display shows the unit charged up to 200J (360J
The amber charger light illuminates.
monitor (for manual devices).
The display shows that the unit is charging. The
SHOCK button lights when the unit is charged, and
the Ready tone for DEFIB sounds.
The unit discharges internally and selected energy
decrements to 1J.
The following selections appear: Verify the following
energy amounts display incrementally:
Record your results on the Maintenance Test Checklist.
Press the SHOCK button.The unit discharges and the SHOCK button is no
longer illuminated.
A 15 second strip chart automatically prints, displaying
the number of joules delivered (if equipped with
recorder and configured to print event).
9
ooo
M Series Service Manual
3.0 3, 5, and 12 Leads Test
Tools Needed: 3 lead, 5 lead, and 12 lead cables. Test each cable separately. QED 6 Defibrillator Analyzer.
Test Setup: The M Series unit must be configured to display ECG LEAD OFF message.
Connect the lead wires appropriate to each test to the QED 6 Defibrillator Analyzer.
Do this...Observe this...Pass/Fail/
3.1
3.2
3.3
3.4
Turn the selector switch to
MONITOR. Select leads.
Disconnect one lead from the
simulator.
Reconnect the lead. Repeat
step 3.2 with the remaining
leads.
Repeat 3.2 and 3.3 for 5 lead
and 12 lead cables.
The monitor displays the NO ECG LEADS OFF
message.
The monitor displays the ECG LEAD OFF message
within 3 seconds (if configured).
The ECG LEAD OFF message appears when the
lead is disconnected and clears the lead is
reconnected.
Note: If heart rate alarm sounds, press and hold the
ALARM SUSPEND button for 4 seconds to
disable the alarms.
Note: When testing the 12 lead cable, the ECG
LEAD OFF message displays when you pull
off a limb lead. When you pull off a V lead, the
ECG VX LEAD OFF message displays, where
“X” is the number between 1 and 6.
Record your results on the Maintenance Tests Checklist.
NA
ooo
ooo
ooo
ooo
10
M Series Service Manual
4.0 Power Supply Test (Optional)
Tools Needed:• 1 red miniature alligator to miniature alligator test lead.
• 1 black miniature alligator to miniature alligator test lead.
• DC power supp ly (15 Amp min imu m).
• 0.1 ohm
• 1000 ohm 1% resistor (
• Fluke 75 multimeter or equivalent.
Test Setup:Make sure the unit and power supply are turned off.
Connect one end of the black lead to the “-” terminal in the battery well.
Connect the other end of the black lead to the “-” terminal of the power supply.
Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it.
Connect the other end of the red lead to the “+” terminal of the power supply.
Set the power supply voltage to 7V.
CAUTIONBe sure to connect the power supply properly to the M Series battery well terminals or damage to the unit may result. Do NOT raise
the power supply voltage above 12V.
resistor (¼W or greater).
¼W or greater).
11
M Series Service Manual
?
Battery Well
15 Amp
Supply
?
Red
Black
Do this...Verify that...Pass/
Turn the selector switch to
4.1
MONITOR.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Turn the selector switch to the
4.2
OFF position.
Adjust the power supply voltage to
4.3
10.3V and turn the selector switch
to MONITOR (for AED units, turn
the selector switch to the ON
position).
Low Battery Test
4.4
Set voltage to 9.8V.
Set voltage to 9.4V .The unit displays the LOW BATTER Y message within 30
4.5
Shut Down Voltage Test
4.6
Set voltage to 8.5V.
Record your results on the Maintenance Tests Checklist.
The unit does not turn on.
The unit turns on.
The unit does not display the LOW BATTERY message.
seconds.
The unit shuts off within 30 seconds.
Fall
oo
oo
oo
oo
oo
12
M Series Service Manual
Test Setup
Remove red lead from power supply and connect to 0.1ohm resistor.
Connect other end of resistor to “+” terminal of power supply using a second red label.
Connect multimeter across the resistor.
Set voltage scale (if DVM is not autoranging) to 220mV.
Red
Black
+
_
Battery Well
_
+
+
_
DMM
15 Amp
Supply
4.7
Do this....Observe this...Pass/
Fail/
System Current Test
Set power supply to 10.3V.
13
M Series Service Manual
Do this....Observe this...Pass/
4.8
4.9
Record your results on the Maintenance Test Checklist
Turn the selector switch to
MONITOR. (For AED units, turn
the selector switch to ON and
select Manual mode.
Turn unit off.
Voltage across resistor should be 80 mV or less (<800
mA of ON current). Note: Without optional parameters.
a) With green screen or LCD and no options <80mV.
b) With yellow screen and no options <81mV.
c) With yellow screen and SpO2 <104mV.
d) With yellow screen and voice recording <91mV.
e) With yellow screen, voice recording, and SpO2
<114mV.
f) All devices with EtCO2 <121mV.
Off Current Test
Fail/
oo
oo
Test SetupRemove 0.1 ohm resistor and replace with 1K
Connect DMM across resistor.
Set voltage scale to DCV.
Measure voltage across resistor.
Measure across resistor with unit
4.10
turned off.
Record your results on the Maintenance Test Checklist
Voltage should be less than 450mV (<450uA of current).
14
oo
M Series Service Manual
5.0 Leakage Current Test
Tools Needed: See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Test Setup: See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with
accessories: MFC, external paddles, and anterior/posterior paddles.
Table 1:
Maximum Leakage Acceptance Limits
Normal ConditionSingle Fault Condition*
ECG1050
Universal100100
Earth5001000
*Single fault considered AC mains on applied part.
15
M Series Service Manual
6.0 Paddles Test (If applicable)
Tools Needed: QED 6 Defibrillator Analyzer .
Test Setup: If applicable, connect the universal cable to the paddles and place the paddles in paddle wells.
6.1
6.2
6.3
6.4
6.5
6.6
Do this...Verify that...Pass/Fail/
Turn the selector switch to
DEFIB (For AED units,turn the
selector switch to ON and
select Manual mode.) Press
and hold the ENERGY DOWN
button on the sternum paddle.
Press and release the ENERGY UP button on the sternum paddle
for each setting.
Press and release the
RECORDER button on the
sternum paddle.
Press and release the
RECORDER button again.
Select 30J using the paddle
ENERGY button. Press the
CHARGE button on the Apex
paddle.
Press and release the APEX SHOCK button.
The energy selection decreases to 1J.
The joules selection increases as follows: Biphasic 110, 15, 20, 30, 50, 75, 100, 120, 150, 200J.
The unit charges to 30J, then the red LED charge
indicator illuminates and the charge tone sounds.
Note: The front panel shock button does not
illuminate.
The unit does not discharge.
NA
oo
oo
ooo
oo
oo
oo
6.7
6.8
Press and release the
STERNUMSHOCK button.
Press and hold both paddles
SHOCK buttons.
The unit does not discharge.
The unit discharges. The TEST OK message displays
and the red LED turns off. If configured, the recorder
prints a strip chart.
16
oo
oo
M Series Service Manual
Do this...Verify that...Pass/Fail/
Record your results on the Maintenance Tests Checklist.
NA
17
M Series Service Manual
7.0 Heart Rate Display Test
Tools Needed:• ECG Cable (3, 5 or 12 leads).
• QED 6 Defibrillator Analyzer.
Test Setup:Turn the selector switch to MONITOR (For AED units, turn the selector switch to ON and select Manual override.)
Press LEAD button until “I” displays.
Connect the ECG leads to the QED 6 Defibrillator Analyzer.
Connect the ECG cable to the unit.
Do this...Verify that...Pass/Fail/
7.1
Record your results on the Maintenance Tests Checklist.
Set the QED 6 Defibrillator
Analyzer to 120BPM.
The Heart Rate displays as 120 +/- 2 bpm.
NA
ooo
18
M Series Service Manual
8.0 Calibrating Pulses on Strip Chart Test
Tools Needed: None
Test Setup: None.
Do this...Verify that...Pass/Fail/
8.1
8.2
Record your results on the Maintenance Tests Checklist.
Press the RECORDER button.
Press and hold SIZE button to
activate the calibration signal.
The strip chart displays a signal of 300 ppm with an
amplitude of 10 mm +/- 1 mm. The signal also appears
on the video display.
NA
ooo
19
M Series Service Manual
9.0 Notch Filter Test
Tools Needed: QED 6 Defibrillator Analyzer.
Test Setup:Connect the ECG cable to the QED 6 Defibrillator Analyzer.
Connect the ECG cable to the unit
Do this...Verify that...Pass/Fail/
9.1
9.2
9.3
9.4
Record your results on the Maintenance Tests Checklist.
Turn the selector switch to MONITOR mode.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Select lead I, size 3x.
Select 60Hz (or 50 Hz for a 50Hz
unit) on
the QED 6 Defibrillator Analyzer.
Press RECORDER button.Verify that the waveform amplitude on the strip chart is
Press RECORDER button to
stop printing.
.
less than 1.5 mm.
NA
ooo
20
M Series Service Manual
10.0 Heart Rate Alarm Test
Tools Needed: QED 6 Defibrillator Analyzer
10.1
10.2
10.3
10.4
10.5
10.6
10.7
Do this...Verify that...Pass Fail/
Turn the selector switchto
MONITOR mode.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Connect the ECG leads to the
QED 6 Defibrillator Analyzer
the QED 6 Analyzer to 120 BPM
and the defibrillator to lead II.
Press ALARMS button.The monitor displays the alarm menu.
Press SELECT PARAM softkey
until ECG HR displays.
Press INC> for state.The cursor scrolls through ENABLE, AUTO and
Press DEC>for state.The cursor scrolls through ENABLE, DISABLE, AND
Press INC> until ENABLE
displays.
Press ENTER softkey, then press
the NEXT FIELD softkey to
select the heart rate limit.
. Set
The monitor screen displays the Lead II message and
the waveform for NSR ECG at 120 BPM +/- 2
.
The cursor scrolls through parameters.
DISABLE.
AUTO.
ENABLE appears.
The cursor scrolls to Low field. Press RETURN softkey
to exit out of Alarm Menu.
NA
ooo
ooo
ooo
ooo
ooo
ooo
ooo
10.8
Press ALARM SUSPEND
button.
No alarm sounds.
21
ooo
M Series Service Manual
10.9
10.10
10.11
10.12
10.13
10.14
Do this...Verify that...Pass Fail/
Remove an EEG lead wire from
the QED 6 analyzer.
Reattach ECG Lead wire to the
QED 6 Analyzer and hold the
ALARM SUSPEND button on
unit for 4 seconds.
Press the ALARM SUSPEND
button to enable the alarms.
Set QED 6 Analyzer to 160 BPM
or higher.
Press the ALARM SUSPEND
button on the M Series unit.
Press and hold ALARM SUSPEND for 4 seconds to
disable alarms.
The alarm symbol flashes and the heart symbol stops
flashing. The ECG LEAD OFF alarm tone sounds.
Recorder prints a stripchart showing a low heart rate, if
enabled.
The unit displays the alarm symbol with an “X” through
it.
The heart symbol flashes with each QRS wave.
The alarm symbol does not have an “X” through it.
The Heart Rate Value is highlighted, the alarm tone
sounds, and the alarm and the heart symbol both flash.
The alarm is suspended for 90 seconds. The unit
displays the alarm symbol with an “X” through it, and
the heart symbol flashes with each QRS wave.
Note: Software version 35.00-36.XX, the alarm is
suspended until the alarm button is press again, or the
fault is cleared.
NA
ooo
ooo
ooo
ooo
ooo
ooo
Record your results on the Maintenance Tests Checklist.
22
M Series Service Manual
11.0 Defibrillator Self Test
WARNING!
SHOCK HAZARD!
TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY
BEFORE YOU START CONDUCTING THE DEFIBRILLATOR TESTS.
Keep hands and all other objects clear of the multi-function cable connections and defibrillator
analyzer when discharging the defibrillator.
Before you discharge the defibrillator, warn everyone near the equipment to STAND CLEAR.
CAUTIONDo NOT internally discharge the unit more than 3 times in 1 minute. Note that the rapid repetition of internal discharges at
more than 30 Joules may damage the unit.
Tools Needed:• MFC Test Port Connector 1004-0053-99 with universal cable.
• MFC Test Adaptor Connector (Fluke Biomedical Part Number 3010-0378 or equivalent).
• QED 6 Defibrillator Analyzer.
• ECG Cable.
• Stop watch.
Test Setup: Ensure the unit is turned off
Ensure the ECG cable is connected to the unit and analyzer.
23
M Series Service Manual
NOTE: The universal cable should not be connected to any equipment at the beginning of this test.
11.1
11.2
11.3
11.4
11.5
11.6
11.7
Do this...Verify that...Pass/
Turn the selector switch to DEFIB
mode.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Set leads to PADS.
Connect the universal cable to the
MFC test port.
Select energy level of 100J and
press the CHARGE button.
Press the SHOCK button.The unit does not discharge. The unit displays the
Set energy level to 30J.The unit discharges internally.
Press the CHARGE button.The unit charges to 30J and displays message, DEFIB
Press and hold SHOCK button. The unit discharges and displays the number of joules
The unit displays the CHECK PADS/POOR PAD
CONTACT message.
DEFIB PAD SHORT message displays.
The charge time is greater than 2 seconds and less than
10 seconds and the unit displays the message, SELECT 30J FOR TEST.
message, DEFIB PAD SHORT.
30J READY. The charge ready tone sounds.
delivered and the TEST OK message. For example, the
message at the top of the printed strip chart would read
as follows: 30 JOULES TEST OK. TEST_CUR=10-14A
DEFIB_IMPED=0.
Fail
oo
oo
oo
oo
oo
oo
oo
The impedance value may range from 0 to 5 ohms.
If configured, the recorder prints a strip chart.
Record your results on the Maintenance Tests Checklist.
24
M Series Service Manual
12.0 Synchronized Cardioversion Test
Tools Needed: QED 6 Defibrillator Analyzer .
Test Setup:Connect the universal cable via the adapter (D.N.I #3010-0378) to the defibrillator analyzer.
Select Sync on analyzer and set ECG on analyzer to 60 -120 BPM.
Do this...Verify that...Pass/
12.1
12.2
12.3
Record your results on the Maintenance Tests Checklist.
Press LEAD button to select
PADS and Size X1.
Press the SYNC softkey on the
defibrillator. Enter synchronized
cardioversion timing test mode on
the defibrillator analyzer.
Select 200J (360J for Monophasic
units), and then press the
CHARGE button. When the
SHOCK button lights, press and
hold the SHOCK button.
Sync markers appear on display.
Note: Sync markers display on the monitor. The sync
marker appears as a down arrow over the ECG
R-wave peaks on strip chart and display.
On the analyzer display, the R-wave to shock delay is
less than60 milliseconds.
Fail
oo
oo
25
M Series Service Manual
13.0 Shock Test
Tools Needed: QED 6 Defibrillator Analyzer, Stop Watch.
Test Setup:Connect the universal cable via the adapter (D.N.I #3010-0378) to the defibrillator analyzer.
Ensure that a fully charged battery is installed in the unit.
NOTE If your M Series AED does not have manual override capability, do not perform this test.
13.1
13.2
13.3
13.4
13.5
13.6
Do this...Verify that...Pass/Fail/
Turn the selector switch to
DEFIB mode.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Press the ENERGY SELECT
down arrow until 5J displays.
Press the CHARGE button. Wait
for the SHOCK button to
illuminate.
Press the SHOCK button.The unit discharges 3J-5J into the simulator.
The unit displays DEFIB 5J SEL.
The unit displays DEFIB 5J RDY.
NA
ooo
ooo
ooo
Note: The displayed rhythm may change shape for 30
seconds before it returns to an original rhythm.
This is caused by the operation of the adaptive
bandwidth defibrillator recovery circuit.)
Press the ENERGY SELECT up
arrow until the unit displays 50J.
Press the CHARGE button. Wait
for the SHOCK button to
illuminate.
The unit displays the message, DEFIB 50J SEL.
The unit displays the message, DEFIB 50J RDY.
ooo
ooo
13.7
Press the SHOCK button.The unit discharges 41J-55J (monophasic 43J-57J)
into the simulator.
26
ooo
M Series Service Manual
13.8
13.9
13.10
13.11
13.12
13.13
13.14
13.15
Do this...Verify that...Pass/Fail/
Press the ENERGY SELECT up
arrow until the unit displays 100J.
Press the CHARGE button. Wait
for the SHOCK button to
illuminate.
Press the SHOCK button.The unit discharges 83J-122J (monophasic 85J-115J)
Press the ENERGY SELECT up
arrow until the unit displays 200J.
Press the CHARGE button. Wait
for the SHOCK button to
illuminate.
Press the SHOCK button.The unit discharges 182J-247J (monophasic 170J-
(Monophasic Units) Press the
ENERGY SELECT up arrow unit
the unit displays 300J.
Press the CHARGE button. Wait
for the SHOCK button to
illuminate.
The unit displays the message, DEFIB 100J SEL.
The unit displays the message, DEFIB 100J RDY.
into the simulator.
The unit displays the message, DEFIB 200J SEL.
The unit displays the message, DEFIB 200J RDY.
230J) into the simulator.
The unit displays the message, DEFIB 300J SEL.
The unit displays the message, DEFIB 300J RDY.
NA
ooo
ooo
ooo
ooo
ooo
ooo
ooo
ooo
13.16
13.17
13.18
13.19
Press the SHOCK button.The unit discharges 255J-345J into the simulator.
(Monophasic Units) Press the
ENERGY SELECT up arrow unit
the unit displays 360J.
Press the CHARGE button. Wait
for the SHOCK button to
illuminate.
Press the SHOCK button.The unit discharges 306J-414J into the simulator.
The unit displays the message, DEFIB 360J SEL.
The unit displays the message, DEFIB 360J RDY.
27
ooo
ooo
ooo
ooo
M Series Service Manual
13.20
13.21
13.22
13.23
13.24
Do this...Verify that...Pass/Fail/
(Biphasic Units) Press the
CHARGE button and start timing
with a stopwatch. Stop timing
when the SHOCK button lights.
Press the SHOCK button.On the strip chart, the Patient Current is between 20-24
(Monophasic Units) Press the
CHARGE button and start timing
with a stopwatch. Stop timing
when the SHOCK button
illuminates.
Press the SHOCK button.On the strip chart, the Patient Current is between 20-24
(AED units only.) Disconnect the
universal cable from the
analyzer.
The charge time is between 3-6 seconds.
Amps, and the Defib Impedance is between 46-54
Ohms.
Note: This will occur if configured to Auto-Generate
Strip s on event.
The charge time is between 4-8 seconds.
Amps, and the Defib Impedance is between 46-54
Ohms.
Note: This will occur if configured to Auto-Generate
Strip s on event.
The unit will display CHECK PADS, and give anaudio
prompt.if configured for audio.
NA
ooo
ooo
ooo
ooo
ooo
Record your results on the Maintenance Tests Checklist.
28
M Series Service Manual
14.0 Summary Report Test
Tools Needed: QED 6 Defibrillator Analyzer .
Test Setup: Connect the universal cable to the QED 6 analyzer.
If you are using paddles, place the paddles on the analyzer’s discharge plates.
14.1
14.2
14.3
14.4
Do this...Verify that...Pass/
Press and hold the SUMMARY
softkey for 4 to 8 seconds to
display erase options. Press the
ERASE SUMMARY softkey to
erase any previously stored data
erase any previously stored data.
Set selector switch to DEFIB.
Select 200J (360J for
Monophasic) using the ENERGY
SELECT button, and press the
CHARGE button. When charged,
press the SHOCK button to
discharge into the defibrillator
analyzer.
Wait 18 seconds, then press the
Code Marker softkey. Press the
CPR softkey.
Turn the unit off. Wait 10 seconds
and then turn the unit on. Press
the SUMMARY softkey, then
press the PRINT CHART softkey,
then press PRINT ALL.
The unit displays the message, ERASING REPORT.
The unit successfully discharges and prints a strip chart.
The Code Markers appear.
The summary report prints. The report displays the
correct date, time, 1 shock delivered and Code Marker
event.
Fail
oo
oo
oo
oo
Record your results on the Maintenance Tests Checklist.
29
M Series Service Manual
15.0 Advisory Message Test
Tools Needed:QED 6 Defibrillator Analyzer .
Test Setup: Connect the universal cable via the adapter (D.N.I #3010-0378).
Attach the M Series to the QED 6 Defibrillator Analyzer.
15.1
15.2
15.3
15.4
Do this...Verify that...Pass/
Turn the selector switch to DEFIB
mode.
(For AED units, turn the selector
switch to ON.)
Select VF (ventricular fibrillation)
on the QED 6, then press the
ANALYZE button.
Press the SHOCK button.The unit discharges.
Select the NSR (normal sinus
rhythm) on the simulator, then
press the ANALYZE button.
The unit displays the following messages:
ANALYZING ECG.
ST AND CLEAR .*
SHOCK ADVISED.*
PRESS SHOCK.*+
*AED’s audio prompts are standard. Advisory audio
prompts are user configurable.
+If configured for auto charge.
The unit displays the following messages:
ANALYZING ECG.
ST AND CLEAR .*
NO SHOCK ADVISED.*
*AED’s audio prompts are standard.
Fail
oo
oo
oo
Record your results on the Maintenance Tests Checklist.
30
M Series Service Manual
?
16.0 Pacer Test
Tools Needed:
QED 6 Defibrillator Analyzer .
NOTE The following tests are to be performed only on M Series units equipped with the optional pacing function.
The pacer output can be measured using an oscilloscope set to DC coupling connected across a load resistor. (See
diagram in column for universal cable connector polarity.) The load resistor is a 100 ohm, 5 watt or greater. The pacer
output is a positive going pulse, 40 +/- 2 ms duration with an amplitude of 0.1 volt per milliamp of selected output
(e.g., 40 milliamps of selected output has an amplitude of 4 +/- 0.5 volts the specified tolerance displayed on the
oscilloscope).
If an external non-invasive pacer analyzer is being used, then follow the manufacturer’s guidelines for measuring the
frequency (ppm), output (mA) and the pulse width measured in milliseconds. Note that the analyzer pace load resistor
must be less than 250 ohms.
Test Setup:Connect the universal cable from the M Series to the QED 6 Defibrillator Analyzer.
Do this...Verify that...Pass/
16.1
16.2
16.3
16.4
16.5
Turn the Main Selector knob of
the unit to Pacer Mode.Set the
PACEROUTPUT to 14 mA and
disconnect MFC connector from
the QED 6 analyzer.
Reconnect theuniversal cable to
QED 6 analyzer. Press Clear
the
Pace Alarm softkey.
Set rate to 180 ppm; output to
0mA.
Increase the output to 40mA.Output on the QED 6 analyzer is 40mA +/- 5mA.
Increase the output to 120mA.Output on the QED 6 analyzer is 120mA +/- 6mA.
The unit displays the CHECK PADS and POOR PAD CONTACT messages displays, and the pace alarm is
active.
CHECK PADS and POOR PAD CONTACT message
disappears. The pace alarm is cleared.
No output appears on the QED 6 analyzer.
31
Fail
oo
oo
oo
oo
oo
M Series Service Manual
16.6
16.7
16.8
16.9
16.10
16.11
16.12
16.13
Do this...Verify that...Pass/
Increase the output to 140mA.Output on the QED 6 analyzer is 140mA +/- 7 mA.
Check the pulse width.Pulse width is 40mS +/-2mS.
Decrease the output to 60mA.
Decrease the rate to 30ppm.
Increase the rate to 180ppm.Pacer rate on QED 6 an alyzer is 177-183 ppm.
Connect the ECG cable to the
M Series and QED 6 analyzer.
Select the ECG at 60 BPM on the
QED 6 analyzer.
Press the Async Pace softkey.The unit displays ECG at 60 BPM with stimulu s markers,
Turn off the QED 6 analyzer. Set
Pacer Rate to 100ppm. Press the
RECORDER ON button.
Press and hold 4:1 button.The unit displays pace stimulus markers every 60 mm+/-
Pacer rate on QED 6 analyzer is 29-31 ppm.
The unit displays ECG at 60 BPM with no stimulus
markers.
and displays the Async pace message.
The unit displays pace stimulus markers every 15mm +/-
1mm.
1.5 mm.
Fail
oo
oo
oo
oo
oo
oo
oo
oo
Record your results on the Maintenance Tests Checklist.
32
M Series Service Manual
17.0 SpO2 Monitor Test for SpO2 Option
Tools Needed:• Masimo
Reusable Sensor.
• Masimo Patient Cable
• Fluke Index Series 2 SpO
Test Setup:Connect the universal cable to the MFC test plug.
Simulator.
2
DO NOT connect the ECG cable to the simulator.
Connect the Masimo Patient Cable and attach the Masimo sensor to the patient cable.
Connect the Masimo sensor to the finger simulation post.
Place a fully charged battery into the battery well or connect to ac power (dc power, if equipped).
Ensure that the SpO
Do this...Verify that...Pass/
17.1
17.2
Turn the selector switch to
MONITOR.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Wait ten seconds.
Turn on the SpO2 simulator.
Press the SIM softkey on the
Index SpO
the MAN softkey.
Simulator is off.
2
Simulator, then press
2
The SpO2 saturation percentage appears as a dashed
line on the monitor.
The unit displays the SpO
PULSE SEARCH message.
2
Fail
oo
oo
17.3
Press the 02+ or 02- softkey of
the simulator until the SpO2
output is at 98%.
The SpO2 reading of 98 +/- 1% appears on the E Series
monitor.
Note: You may need to wait up to 2 minutes for the
information to appear on the ZOLL display.
33
oo
M Series Service Manual
17.4
17.5
17.6
17.7
Do this...Verify that...Pass/
Using the Index SpO2 Simulator,
press the BPM+ or BPM- softkey
until the heart rate is 230 BPM.
Using the Index SpO2 Simulator,
press the BPM- softkey until the
heart rate is 50 BPM.
Using the Index SpO2 Simulator,
press the 02+ softkey until the
output is at 72%.
SpO
2
Press Wave 2 softkey.
Select the SpO
For CCT units:
softkey, and then select SET
TRACE 2, and press Enter.
Select SPO2, press Enter.
waveform.
2
Press the Traces
The SpO2 simulator screen displays an SpO2 rate of 230
BPM.
Note that you may need to wait up to 2 minutes for the
information to appear on the ZOLL display.
The M Series monitor displays an SpO
100%.
The M Series monitor displays a heart
BPM.
The SpO2 simulator screen displays an SpO
of 96-100%.
The M Series monitor displays a heart rate of 46-54
BPM.
The SpO2 simulator screen displays an SpO
of 70-74%.
The M Series monitor displays a heart rate of 46-54
BPM.
Plethysmographic waveform appears on the ZOLL
display.
saturation of 96-
2
rate of 226-234
saturation
2
saturation
2
Fail
oo
oo
oo
oo
oo
oo
17.8
17.9
17.10
Press RECORDER. The plethysmographic waveform prints on the strip chart
paper.
Using the Index SpO2 Simulator,
press the BPM- softkey until the
heart rate is at 230 BPM.
Remove the Masimo patient
cable.
The SpO2 simulator screen displays an SpO
rate of 70-74%.
The M Series monitor displays a heart rate in the heart
position of
226-234 BPM.
saturation
2
34
oo
oo
M Series Service Manual
Do this...Verify that...Pass/
Record your results on the Maintenance Tests Checklist.
Connect the 2-channel Y adapter
to the temperature connector
(T1/T2) on the M Series CCT
Unit.
Connectt the Temperature Probe
Simulator cables to the 2-channel
Y adapter.
On the Temperature Simulator,
set the T1 temperature channel
to 59.2 degrees F(15 degrees C).
On the Temperature Simulator,
set the T1 temperature channel
to 112.8 degrees F (45 degrees
C).
On the Temperature Simulator,
set the T2 temperature channel
to 59.2 degrees F(15 degrees C).
Verify that the value displayed for the T1 temperature
channel betwwen 59.0 degrees to 59.4 degrees F (14.9
to 15.1 degrees C).
Verify that the value displayed for the T1 temperature
channel betwwen 112.6 degrees to 112.8 degrees F
(44.9 to 45.1 degrees C).
Verify that the value displayed for the T2 temperature
channel betwwen 59.0 degrees to 59.4 degrees F(14.9
to 15.1 degrees C).
1
)
NA
ooo
ooo
ooo
19.6
Record your results on the Maintenance Test Checklist.
On the Temperature Simulator,
set the T2 temperature channel
to 112.8 degrees F (45 degrees
C).
Verify that the value displayed for the T2 temperature
channel betwwen 112.6 degrees to 112.8 degrees F
(44.9 to 45.1 degrees C).
ooo
1. If you are using a T emperature Simulator that does have the same setting above, use the closest value available. Verify the
display temperature value is within 0.2 F or 0.1 C of the simulator setting
38
M Series Service Manual
20.0 IBP Monitoring Test
Tools Needed: DNI Nevada 214B Patient Simulator with M Series interface cables, FOGG Transducer Simulator
Test Setup: Set the BP output sensitivity scale to 5 uV/V/mmHg on the Patient Simulator
Connect the Patient Simulator to the IBP connector on the M Series CCT unit
20.1
20.2
20.3
20.4
20.5
20.6
20.7
20.8
Do this...Verify that...Pass/Fail/
Set the selector switch on the M
Series CCT unit to MONITOR and
press the PARAM softkey on the
front panel to display the
Parameters menu.
Select the IBP option and press the
Enter softkey.
Press the Label softkey to display
the IBP Label menu.
Select the Label P1 option and
press the Enter softkey.
Select the P1 option and press the
Enter softkey.
Press the Label softkey to display
the IBP Label menu.
Select the Label P2 option and
press the Enter softkey.
Select the ART option and press
the Enter sofkey.
Verify the parameter menu displayed.
Verify the unit displays the IBP menu.
Verify the unit displays the IBP label menu.
Verify the M Series CCT unit displays the Label Options
menu.
Verify the unit assigns the P1 label to the first pressure
channel (P1) and returns to the IBP menu.
Verify the unit displays the IBP Label menu.
Verify the M Series CCT unit display the Label Options
menu.
Verify the unit assigns the ART label to the second
pressure channel (P2) and return to the IBP menu.
Verify that the M Series CCT unit displays advisory
messages to zero the pressure channels.
NA
ooo
ooo
ooo
ooo
ooo
ooo
ooo
ooo
ooo
39
M Series Service Manual
20.9
20.10
Do this...Verify that...Pass/Fail/
On the Patient Simulator, set BP
channel 1 to an atmospheric
pressure of 0 mmHg (waveform
10) and press EXECUTE. Set BP
channel 2 to duplicate BP channel
1 (waveform 22) and press
EXECUTE.
On the unit go to the IBP menu
(press the Param softkey and
select the IBP option), then press
the Zero softkey to display the Zero
menu, select the Zero Both option
and press the Enter softkey.
Verify the unit zeroes both pressure channels and
displays the P1 ZEROED and P2 ZEROED messages
within 15 seconds.
Verify that the values displayed for the P1 and ART
pressure channels are accurate within ±3 mmHg of
atmospheric pressure (that is, 0 mmHg):
systolic pressure: -3 to 3 mmHg
diastolic pressure: -3 to 3 mmHg
mean pressure: -3 to 3 mmHg
Note: The unit displays pressure values for each pr essure
channel within 15 seconds.
NA
ooo
ooo
20.11
On the Patient Simulator, set to an
arterial pressure of 120/80 mmHg
(waveform 00) and press
EXECUTE.
Verify the values displayed for the P1 and ART pressure
channels:
Note: The unit displays pressure values for each pr essure
channel within 15 seconds.
40
ooo
M Series Service Manual
20.12
20.13
20.14
20.15
20.16
20.17
20.18
20.19
20.20
Do this...Verify that...Pass/Fail/
Zero the IBP channels (see step
20.9).
Press the Traces softkey.Verify the unit displays the Traces menu.
Select the Set Trace 2 option and
press the Enter softkey.
Verify the M Series CCT unit
displays the Trace Options menu.
Press the Traces softkey.Verify the unit displays the Traces menu.
Select the Set Trace 3 option and
press the Enter softkey.
Select the ART option and press
the Enter softkey.
Go to the IBP menu (press the
Param softkey and select the IBP
option).
Press the Range softkey.Verify the unit displays the IBP Range menu.
Verify the M Series CCT unit displays the Trace Options
menu.
Verify the unit assigns the P1 waveform to the second
trace display area and returns to the main menu.
Verify the unit displays the Trace Options menu.
Verify the unit assigns the ART waveform to the third trace
display area and returns to the main menu.
Verify the unit displays the IBP menu.
NA
ooo
ooo
ooo
ooo
ooo
ooo
ooo
ooo
20.21
20.22
20.23
20.24
20.25
Select the P1 Range option and
press the Enter softkey.
Select the 0-30 mmHg option and
press the Enter softkey.
Press the Range softkey.Verify the unit displays the IBP Range menu.
Select the ART option and press
the Enter softkey.
Select the 0-30 mmHg option and
press the Enter softkey.
Verify the unit displays the Range Options menu.
Verify the unit displays the P1 waveform on a 0-30 mmHg
scale and returns to the IBP menu.
Verify the unit displays the Range Options menu.
Verify the unit displays the ART waveform on a 0-30
mmHg scale and returns to the IBP menu.
41
ooo
ooo
ooo
ooo
ooo
M Series Service Manual
20.26
20.27
20.28
20.29
Do this...Verify that...Pass/Fail/
On the Patient Simulator, set BP
channel 1 to a static pressure of 20
mmHg (waveform 11) and press
EXECUTE.
Press the RECORDER button.Verify that there is no deviation from the baseline greater
than 1 mm for the waveforms printed on stripchart and the
unit displays the waveforms for 20mmHg on the monitor
screen, and prints a stripchart of the traces.
Set the input and output resistance
to 350 ohms on the Transducer
Simulator, and set the gain to x10.
Connect the Transducer Simulator
cable to the P1 connector on the
unit and on the Pressure
Transducer Simulator, set the
pressure polarity to positive, and
select a pressure of 200 mmHg.
Zero pressure channel 1.Verify the unit calibrates pressure channel 1 against a
positive pressure of 200 mmHg and displays the P1
ZEROED message.
NA
ooo
ooo
20.30
20.31
20.32
On the Pressure Transducer
Simulator, set the pressure polarity
to positive, and select a pressure
of 500 mmHg.
Connect the simulator cable to the
P2 connector on the unit. On the
Pressure Transducer Simulator, set
the pressure polarity to positive,
and select a pressure of 200
mmHg.
Zero ART channel.Verify the unit calibrates pressure channel 1 against a
Verify that the value displayed for the P1 pressure
channel is in the range of 291 - 300 mmHg. (Note: value
will be displayed within 15 seconds.)
positive pressure of 200 mmHg and displays the ART
ZEROED message.
42
ooo
ooo
M Series Service Manual
20.33
20.34
20.35
20.36
20.37
Do this...Verify that...Pass/Fail/
On the Pressure Transducer
Simulator, set the pressure polarity
to positive, and select a pressure
of 500 mmHg.
Connect the Pressure Transducer
Simulator to the first IBP channel
connector (P1). On the Pressure
Transducer Simulator, set the
pressure polarity to negative, and
select a pressure of -200 mmHg.
Zero the pressure channel.The unit calibrates the pressure channel against a
On the Pressure Transducer
Simulator, select a negative
pressure of -250 mmHg.
Connect the Pressure Transducer
Simulator to the second IBP
channel connector (P2). On the
Pressure Transducer Simulator, set
the pressure polarity to negative,
and select a pressure of -200
mmHg.
On the unit, verify that the value displayed for the ART P2
pressure channel is in the range of 291 - 300 mmHg.
(Note: value will be displayed within 15 seconds.)
negative pressure of -200 mmHg and displays the P1
ZEROED message.
On the unit, verify that the value displayed for the P1
pressure channel is in the range of -47 to -50 mmHg.
(Note: value will be displayed within 15 seconds.)
NA
ooo
ooo
ooo
20.38
20.39
Zero the pressure channel.The M Series CCT unit calibrates the pressure channel
against a negative pressure of -200 mmHg and displays
the P2 ZEROED message.
On the Pressure Transducer
Simulator, select a negative
pressure of -250 mmHg.
Verify that the value displayed for the P2 pressure
channel is in the range of -47 to -50 mmHg. Note: Value
will be displayed within 15 seconds.
43
ooo
ooo
M Series Service Manual
21.0NIBP Transducer Calibration Test
The NIBP module’s pressure transducers are factory-calibrated prior to shipment. However, you can perform a twopoint calibration procedure periodically to ensure accurate pressure measurements.
This procedure is optional at 6 months, but should be performed annually or every 10,000 readings, whichever comes
first.
Tools Needed: Fluke NIBP simulator (the values and procedure that we provide are specific to the BP Pump 2).
Test Setup:Connect the Analyzer’s simulator hose to the NIBP connector on the M Series unit.
Configure the NIBP Analyzer to simulate cuff pressure. For example, on the Fluke Biomedical BP Pump:
Press the MODE button three (3) times to go into Tests mode.
Press the SELECT button once to access the Pressure Simulator screen.
These instructions apply to the Fluke Biomedical BP Pump; for equivalent devices, follow the manufacturer’s
instructions.
Make sure the ECG cable is not connected to the M Series unit.
If the SpO2 option is installed, make sure that the Masimo cable is NOT connected to the M Series unit.
21.1
21.2
21.3
Do this...Verify that...Pass/
Turn the Selector Switch to OFF.
After 10 seconds, press and hold
the fourth softkey from the left and
turn the Selector Switch to
MONITOR (for AED units, turn the
Selector switch to ON).
.
Press the NIBP Calib softkey.The M Series unit displays the NIBP Transducer
On the NIBP simulator set the
pressure parameter to 0 mmHg.
The M Series unit powers on in the NIBP Service Mode.
Calibration Screen.
The NIBP Simulator displays a pressure reading of 0
mmHg.
44
Fail
oo
oo
oo
M Series Service Manual
21.4
21.5
21.6
21.7
21.8
Do this...Verify that...Pass/
On the M Series unit, press the
Set Low softkey to calibrate the
transducer to a 0 mmHg pressure
reading.
On the NIBP Simulator, set the
pressure parameter to
250 mmHg, and then press Start
on the NIBP simulator.
On the M Series unit, press the
Set High softkey to calibrate the
transducer to a 250 mmHg
pressure reading.
On the NIBP simulator, set the
pressure parameter to stimulate a
different cuff pressure (for
example, 205 mmHg).
On the M Series unit, press the
Read Cuff softkey when the
simulator’s Measured screen
reaches 205.
The NIBP pressure transducer registers its voltage
output at a known pressure of 0 mmHg. The field
adjacent to the 0 mmHg value changes to PASS.
Note: If the M Series unit displays a FAIL reading,
verify the NIBP simulator’s pressure setting and
connection to the M Series and repeat the step.
The NIBP Simulator displays a pressure reading of 250
mmHg.
The NIBP pressure transducer registers its voltage
output at a known pressure of 250 mmHg. The field
adjacent to the 250 mmHg value changes to PASS.
Note: If the M Series unit displays a FAIL reading,
verify the NIBP Analyzer’s pressure setting and
connection to the M Series and repeat the step.
The NIBP simulator displays the specified pressure
reading.
The value that the M Series unit displays is accurate
within +/- 3mmHg of the pressure parameter value set on
the NIBP simulator.
Fail
oo
oo
oo
oo
oo
21.9
Record your results on the Maintenance Tests Checklist.
On the M Series unit, press the
EXIT softkey twice.
The M Series returns to the main NIBP Service Mode
screen, then to normal Monitor mode operation.
45
oo
M Series Service Manual
Warning!NIBP transducer calibration can affect clinical readings of the NIBP parameter. Ensure that the NIBP T ransducer Calibration
procedure is performed correctly, followed by an NIBP Monitor Test for verify proper operation.
46
M Series Service Manual
22.0NIBP Monitor Test
The NIBP monitor test verifies the repeatability of the systolic, diastolic, and mean blood pressure measurements, as
well as the patient pulse rate calculation.
Tools Needed: Fluke NIBP simulator (the values and procedure that we provide are specific to the BP Pump 2).
Note: The primary propose of an NIBP simulator is to reproduce a pressure profile similar to a live patient to be
used for testing th e repeatability
the market, each manufacturer uses a different method to develop
different simulators may
vary .
offset value should then be used to determine the expected values. NIBP simulators cannot be used as a source
testing the accuracy of the non-invasive blood pressure measurements of devices such as the ZOLL
monitor/defibrillator.
Test Setup:Connect the Analyzer’s simulator hose to the NIBP connector on the M Series unit.
Set the following parameters on the NIBP Analyzer:
Table 1-1:
and
functionality of the system. There are many different NIBP simulators on
their
algorithm. Consequently, readings from
To test for repeatability, you should first establish
2
the
offset
1
of your simulator. The
for
M Series
ParameterValue
Systolic pressure120 mmHg
Diastolic pressure80 mmHg
Mean pressure
Heart pressure80 bpm
93 mmHg
3
Make sure the ECG cable is not connected to the M Series unit.
If the SpO2 option is installed, make sure that the Masimo cable is NOT connected to the M Series unit.
1
NIBP
Simulators may produce a reading on the
established based on
tance establishing the offset of the simulator and
2
If
you are using the
sure Curve Adjust
3
Not all simulators have a setting of 93mmHg. Check the simulators user's manual for
a
statistical sample of monitors and readings. Please contact ZOLL Technical Support
Fluke®
BiomedicalCuftlink,
Menu.
NIBP
monitor that is shifted from the simulator's setting. The offset value must be
if
you require assis-
test
set-up that you are
you must change the shift value of the Blood Pressure Envelope to +3 on the Pres-
utilizing.
recommendations.
47
M Series Service Manual
Do this...Verify that...Pass/
22.1
22.2
22.3
22.4
Record your results on the Maintenance Tests Checklist.
Turn the selector switch to MONITOR mode.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Ensure that the LEADS
parameters is set to PADS
(default).
If necessary, press the LEADS
button to cycle through the values
to select PADS.
Press the NIBP button on the
E Series front panel.
Press the SUMMARY softkey,
then press the TREND softkey,
followed by the NIBP Trend
softkey.
The M Series unit displays PADS in the Lead selection
field on the monitor.
The M Series displays the following measurements:
• Systolic pressure (115 - 125 mmHg)
• Diastolic pressure (75 - 85 mmHg)
• Mean pressure (88 - 98 mmHg).
The M Series unit displays a summary of the NIBP
measurements, including the pulse rate reading (in the
range of 77 - 83 bpm).
Fail
oo
oo
oo
4
These values only apply for test set-ups utilizing the BP Pump 2 Simulator. V ariations of the test set-up or different simulators may
produce readings
Please
contact ZOLL Technical Support i f you
outside
the provided values and will require end-user facility to establish the appropriate offset and tolerances.
require
assistance establishing the offset o f y o u r simulator and test set-up
.
48
M Series Service Manual
23.0NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer
The volume leak test verifies the integrity of the pneumatic system on the NIBP module. This test is optional at 6
months, but should be performed annually or every 10,000 readings , whichever comes first.
Tools Needed: Fluke NIBP simulator (the values and procedure that we provide are specific to the BP Pump 2).
Test Setup:Connect the Analyzer’s simulator hose to the NIBP connector on the M Series unit.
Configure the NIBP Analyzer for the volume leak test. For example, on the Fluke Biomedical BP Pump:
• Press the MODE button three times to go into Tests mode.
• Press the SELECT button twice to access the volume leak test.
Make sure the ECG cable is not connected to the M Series unit.
If the SpO2 option is installed, make sure that the Masimo cable is NOT connected to the M Series unit.
23.1
23.2
23.3
23.4
Do this...Verify that...Pass/
Turn the Selector Switch to OFF.
After 10 seconds, press and hold
the fourth softkey from the left and
turn the Selector Switch to
MONITOR (For AED units, turn
the selector switch to ON).
Press the Leak Test softkey.The M Series unit displays the NIBP Le ak Test Screen.
On the NIBP simulator, set the
pressure parameter to 200
mmHg.
On the M Series unit, press the
Close Valves softkey .
The M Series unit powers on in the NIBP Service Mode.
The NIBP simulator displays a pressure reading of 200
mmHg.
The Valves status changes from OPEN to CLOSED.
49
Fail
oo
oo
oo
oo
M Series Service Manual
Do this...Verify that...Pass/
23.5
Press the START TEST softkey
within 30 seconds after closing
the valve.
Note: This test takes
approximately 3 minutes.
23.6
23.7
Record your results on the Maintenance Tests Checklist.
On the NIBP Analyzer, press the
STOP TEST softkey.
On the M Series unit, press the
EXIT softkey twice.
After approximately 1 minutes, a number appears in the
upper middle area of the NIBP simulator display.
If the simulator:
• Displays a Volume Leak reading <5,
Series unit has passed the test.
• Displays a Volume Leak reading >5,
Series unit has failed the test.
• Displays no Volume Leak reading, but maintains a
stable pressure reading at or above 200 mmHg, then
the M Series unit has passed the test; there is no
volume leak.
In addition, the M Series unit displays the simulator’s
pressure reading in the “Cuff Pressure” field.
After approximately 3 minutes, the valves open on the M
Series unit.
The NIBP simulator terminates the Volume Leak Test.
The M Series unit returns to the main NIBP Service
Mode screen, then to normal Monitor mode operation.
Tools Needed: USB Bluetooth dongle with driver CD, Bluetooth application software BlueSoleil.
Test Setup: Verify that the M Series unit is configured as follows: Bluetooth INSTALLED, and Baud Rate set to 115200 or 38400.
24.1
24.2
24.3
24.4
24.5
Do this...Verify that...Pass/
Power up the M Series unit while
pressing in and holding the leftmost softkey.
Insert a PCMCIA card containing
data into the card reader.
Press the UPLOAD CARD
softkey, then press the softkey
UPLOAD CARD.
Press the SEND softkey.
Double click on the HyperT erminal
shortcut (BT115200) from the
desktop.
The M Series unit displays the System Utilities Menu.
Verify that the Bluetooth LED is on and displays a green
light.
Fail
oo
oo
Note: If setting up Hyperterminal
for the first time set COM
PORT programmed for
Bluetooth adapter,
1 15200bp s, 8, N, 1, N
24.6
24.7
Start Bluetooth Utility software
(Blue Soleil) on the computer.
In the IVT Corporation Blue Soleil
window under “View”, select
refresh devices.
51
M Series Service Manual
Do this...Verify that...Pass/
24.8
24.9
Record your results on the Maintenance Tests Checklist.
Find the unit under test Serial
Number and double click on it.
Click on the Serial Port Icon that is
highlighted at the bottom of the
window
Verify the following:
At the bottom of the window it states Connected,
The file transfer window pops up and the file upload
progress bar goes to 100%.
The unit displays the file upload message, Card
Uploaded.
Fail
oo
oo
oo
52
M Series Service Manual
•
Overview
This chapter describes the most common technical problems that biomedical technicians experience when checking
the M Series during routine maintenance or when there is a malfunction of the unit. It also contains a list of error
messages that users may see if the unit is not operating properly.
This chapter contains the following:
• Troubleshooting tables for ECG Leads Off Messages and Monitor Displays
• Zoll M Series Error Messages
Chapter 2
Troubleshooting
If the problems you encounter are not listed below, call ZOLL Medical Corporation’s Technical Service Department
for further assistance. (See page vii for contact information.)
53
M Series Service Manual
Troubleshooting
The following tables show the most common troubleshooting issues and their solutions.
First, attempt to solve the problem with “Recommended User Action.” If these steps do not solve the problem, follow
the steps listed in the “Recommended Technical Action” column.
Reported ProblemRecommended User ActionRecommended Technical Action
ECG LEAD OFF message
displays.
(3, 5, 12 lead cable)
• Check preparation of ECG electrode site by
cleaning the site, lightly abrading the patient’s skin
and/or clipping the patient’s hair at the electrode
site.
• If electrode gels are dry, replace electrodes with
new ones from a freshly opened package.
• Verify that all leads are attached.
• Set monitor to anotherlead.
• Verify that the electrodes have not exceeded their
expiration date.
• T ry to reproduce the problem using
a simulator.
• Inspect the ECG cables looking for
corrosion or broken connector pins.
• Check the cable for intermittent
connections by flexing the cable at
the yoke and snap connectors.
• Check the cable connection to the
defibrillator.
• Inspect the ECG input connector
and its pins. Replace it, if
necessary.
• Inspect the ECG cable connection
to the system board.
• Inspect the system board ECG
shielding.
• Remove and replace the system
board.
54
M Series Service Manual
Reported ProblemRecommended User ActionRecommended Technical Action
V LEADS OFF message
displays.
CHECK P ADS/POOR PAD
CONTACT message
displays.
• If the user is not using V leads, attach V lead
connector terminator plug to the cable’s V lead
connector.
• If a V1 lead wire metal snap comes in contact with
the patient’s skin, then the system will show all V
leads as OFF.
• Remove V1 leads and others away from the
patient. Turn off the unit and wait ten seconds
before turning it back on.
• Remove and reinsert PADS connector into the
universal cable.
• Check for damaged defibrillator pads, wires and or
connector.
• Check for dried out or expired defibrillator pads.
• Clip (not shave) the patient’s hair and wipe pad
contact area dry.
• Connec t the cable to the test plug. The DEFIB PAD
SHORT message displays to indicate that the cable
is functioning properly.
• If the DEFIB PAD SHOR T message displays, then
check the connections of the pads to the patient and
to the defibrillator cable.
• If the DEFIB PAD SHORT message does not
display, remove the defibrillator from service.
• Connect universal cable to the
shorting plug. The DEFIB PAD
SHORT message should display,
when you SELECT PADS. If the
message does not display , then:
• Try another universal cable.
• Chec k the cable from the universal
cable connector to the High
Voltage Module.
• Chec k the cable from the High
Voltage Module to the system
board.
• Remove and replace the High
Voltage Module.
• Remove and replace the system
board.
• Call ZOLL Technical Support for
assistance.
55
M Series Service Manual
Reported ProblemRecommended User ActionRecommended Technical Action
Flash or arcing under
defibrillator pad.
Displayed HR not accurate.
No artifact present.
• Avoiding using alcohol and betadine in and around
the treatment area because these skin preparations
may lead to increased conductivity and/or bonding
between the electrode’s adhesive and skin.
• Check for ge l droop. If the gel has leaked out of the
gel treatment area, replace the electrode.
• Ensure pads are coupling to the patient’s skin and
connected to the universal cable.
• Check for dried out gel on the defibrillator pad.
• Clip patient’s excessive hair. Do not shave hair.
• Check expiration date. Replace pad if date has
expired.
• Do not conduct chest compression through the
pads because the pads could be damaged leading to
the possibility of arcing and skin burns.
• Apply the back electrode first. If the front electrode
is already in place when the patient is being
maneuvered for placement on the back, the front
may become partially lifted, possibly causing
arching and skin burns.
Verify heart rate flashes with each QRS on display.
• Change lead selection.
• Change ECG size.
• Reposition ECG electrodes.
• Ensure that wet gel pads are stored
flat.
56
M Series Service Manual
Reported ProblemRecommended User ActionRecommended Technical Action
Displayed HR not accurate;
artifact present.
• Reduce or eliminate ECG artifact due to
electrode or patient cable movement. Route
cables so that they don’t pull on electrodes or
swing excessively .
• Ensure patient is motionless.
• Check for possible excessive radio frequency
interference.
• Verify a good connection of electrod es to the
patient.
• Prepare the patient’s skin prior to the electrode
attachment.
• Move patient cables away from other electrical
equipment, especially any RFI source.
• Ensure ECG cable fits snugly in unit.
• Change ECG cable.
• Replace/reposition ECG electrodes.
Wandering baseline.See “Displayed HR not accurate.” above. Note that
in 90% of electrode issues, size and lead changes
don’t help.
Electronic interference. Check for possible excessive radio frequency
interference.
Move patient cables away from other electrical
equipment.
• Chec k for contamination on snaps.
Ensure springs are intact.
• Check for intermittent ECG patient
cable or connector wiring.
• Replace ECG input connector.
• Replace ECG connector to the
system board cable.
• Replace system board.
Same as above example.
• Turn off sources of excessive RFI.
• Move M Series unit away from
RFI source.
57
M Series Service Manual
Zoll M Series Error Messages
The following is a list of Zoll M Series error messages that may appear on your display. The “User Advisory” column
informs you about an action in progress or provides feedback on a user correctable situation that typically does not
require further technical support. The “T echnical Action” column describes what you as a technician can do to correct
the situation. Note that these messages will sometimes overlap part of the waveform display.
First, attempt to clear the message by turning the Selector Switch to OFF for ten seconds, then back to the desired
operating mode. If the fault persists, call ZOLL Technical Service.
Error MessageExplanation
200J MAX BIPHASICUser attempted to set defibrillation energy >200J on
Biphasic Unit. No higher energy is available.
50J MAXEnergy < 50J for internal paddles. No higher energy
is available.
ADJUST ECGUnit is in sync mode and heart rate is < 20 BPM.
Or, QRS size set too small for proper
synchronization.
ANALYSIS HALTED• ECG analysis halted due to user interaction such
as:
• Lead/size change
• Analyze button was pressed again
• Impedance fault
• Charging error detected in auto defib mode
ANALYSIS RESTARTEDThis is a user prompt issued simultaneously with
ECG TOO LARGE or ECG TOO SMALL. Device
detected ECG signals out of range, automatically
adjusted ECG size and is now restarting its
shockable rhythm analysis sequence.
User
Advisory
Technical Action
58
M Series Service Manual
Error MessageExplanation
AUDIO FAULT 136Audio DSP hard ware error.Replace audio board.
AUDIO NOT
RECORDING
AUDIO QUEUE FULLIndicates that the audio out put queue is full.
BATT HIGH CURRENTBattery is charged and battery current is >.1 A or:
BATT HIGH VOLTAGEBattery voltage > 15.5 v.Replace battery and or
Audio is not recording.Install PCMCIA card.
Additional voice prompts can't be queued at this
time.
Battery is not charged and battery current is > 1.6 A.
User
Advisory
Technical Action
Replace system
board.Turn unit off and
back on again.
Replace system board.
None.
Unplug from A/C.
Remove the battery for
20 seconds. Reconnect
all above. If the
problem persists,
replace battery and or
charger.
charger
BATT LOW CURRENTBattery is not charged and battery current is <.35 A.Replace battery and or
charger.
BATT LOW VOLTAGEBattery voltage < 9.5 v.Replace battery and or
charger.
BATT OVERCHARGECharger on for > 4 hours.Replace battery and or
charger.
59
M Series Service Manual
Error MessageExplanation
BRIDGE SHORTCurrent higher than expected was detected during the
Biphasic bridge test or immediately following a
discharge.
BRIDGE TEST FAILEDBiphasic modul e not operating properly while
charging.
User
Advisory
Technical Action
Ensure pads/paddles
are used properly .
Attempt to clear the
message by turning the
Selector switch to off
then back to the desired
operating modes.
Replace bridge or high
voltage module.
Charge again.
Attempt to clear the
message by turning the
Selector switch to OFF ,
then back to the desired
operating mode.
Replace bridge or high
voltage module.
CABLE FAULT(Auto defib mode only.)
Incorrect A/D reading for paddle ID (similar to
PADDLE FAULT).
CANNOT CHARGECannot charge when charge button pressed.Replace high voltage
REPLACE CARDWrite errors during manual or semi-automated
modes.
60
Replace paddle set,
universal cable and/or
system board.
module or capacitor.
May have configuration
card installed or write
protection on.
M Series Service Manual
Error MessageExplanation
CARD FULLMemory Card Full.
CHECK CO2 SENSOREtCO2 Sensor is unplugged or defective.
CHECK CO
CHECK MEMORY CARDNo card detected during manual or semi-automated
CHECK PADSMessage displayed in conjunction with either POOR
ADAPTERAirway adapter is removed, occluded or adapter
2
zeroing needs to be performed or was performed
incorrectly.
modes.
PAD CONTACT or DEFIB PAD SHORT.
User
Advisory
Technical Action
Check that sensor cable
is plugged in and seated
properly. Check that
sensor is not exposed to
excessive heat. If
problem persists,
replace the sensor.
Replace/Clean airway
adapter. Zeroing
performed
automatically.
Ensure pads are
coupled to patient.
Check /replace pads
and universal cable.
Replace system board.
CHECK PATIENTBackground ECG analysis detects shockable rhythm.
CHECK PULSEAlternate message for NO SHOCK ADVISED
message.Message also shown after delivering third
shock when auto analyze 3 times option is enabled.
CHECK RECORDERProduced when paper tray is empty, paper jams or
recorder door is opened.
61
Replace paper sensor
board, system
interconnect board,
and/or system board.
M Series Service Manual
Error MessageExplanation
CHECK SPO2 SITELow or no perfusion in monitored finger or toe.
CHECK SPO2 SENSORReposition SpO2 sensor on patient.
CLOCK FAULT 11Real time clock oscillator failure.Replace system board.
CLOCK FAULT 12Real time clock back-up power supply failure. Found
oscillator stopped at power-up, but oscillator now
running when the system is running. (Oscillator only
runs when main power is applied).
CLOCK FAULT 13One of the set time units (seconds, minutes, year,
etc.) is out of range.
COMM ERRORNo or invalid communication from the EtCO
CO
2
module.
SENSOR WARM UPEtCO2 Sensor warming up.
CO
2
CONFIRM MANUAL
MODE
Displayed when manual mode is entered. Alerts user
to confirm that manual mode is desired.
2
User
Advisory
Technical Action
Replace system board.
Replace system board.
Replace EtCO
and or system board.
W ait for sensor to warm
up. This process takes
up to approximately
one minute.
module
2
DEFIB DISABLEDUser prompt issued simultaneously with other faults
if defib is disabled.
DEFIB FAULT 71More than 50 internal dumps occurred in less than 20
minutes.
62
Possible configuration
problem. Replace high
voltage module. Call
ZOLL Technical
Support.
Turn the unit to OFF
and back on. If fault
persists, replace high
voltage module.
M Series Service Manual
Error MessageExplanation
DEFIB FAULT 72General defib error.Turn the unit to OFF
DEFIB FAULT 76Capacitor voltage too high for selected energy.Replace high voltage
DEFIB FAULT 77Capacitor voltage > than absolute rated max. Replace high voltage
DEFIB FAULT 78Unable to charge defib cap.Replace high voltage
DEFIB FAULT 79Defibrillator charging too slowly.Replace high voltage
DEFIB FAULT 804 defibrillator faults detected within 20 second
period.
DEFIB FAULT 81Discharge switch in undefined state.Replace high voltage
User
Advisory
Technical Action
and back on. If fault
persists, replace high
voltage module.
module or capacitor.
module or capacitor.
module or capacitor.
module or capacitor.
Replace high voltage
module or capacitor.
module or capacitor.
DEFIB FAULT 84“Upper” discharge transistor shorted (measured via
applicable A/D channel).
DEFIB FAULT 85“Lower” discharge transistor shorted (measured via
applicable A/D channel).
DEFIB FAULT 86One discharge switch closed during power up test.Replace paddles,
DEFIB FAULT 87Both discharge switches closed during power up test.Replace paddles,
63
Replace high voltage
module.
Replace high voltage
module.
control board or system
board.
control board or system
board.
M Series Service Manual
Error MessageExplanation
DEFIB FAULT 94Processor fault causing safety monitor port to be
non-functional.
DEFIB FAULT 95Safe or shutdown line is not functional.Replace high voltage
DEFIB FAULT 96XPATREL or XPAT_ENABLE is faulted or one of
the discharge transistors has shorted.
DEFIB FAULT 108VMON voltage is less than the target energy during
charging.
DEFIB FAULT 109Defib capacitor voltage is greater than selected
energy when defibrillator is charging or ready.
DEFIB FAULT 111Defib capacitor voltage has exceeded the absolute
maximum acceptable voltage.
DEFIB NOT CHARGEDDischarge button is pressed but the unit is not
charged.
User
Advisory
Technical Action
Replace system board,
high voltage module or
capacitor.
module.
Replace high voltage
module.
Replace high voltage
module or capacitor.
Replace high voltage
module or capacitor.
Replace high voltage
module, capacitor, and
or system board.
DEFIB PAD SHORTMeasured impedance between high voltage leads of
MFC.
DISABLE SYNCSync mode active when analyze pressed in defib.
DISCHARGE FAULTDefib capacitor voltage is not decreasing.Replace high voltage
64
Ensure pads are
coupled to patient.
Check /replace pads or
universal cable.
Replace system board.
module, capacitor, and/
or system board.
M Series Service Manual
Error MessageExplanation
ECG FAULT 4Communication fault between ECG processo r and
main processor.
ECG FAULT 5ECU RAM test failure, or ROM checksum test
failure.
ECG LEAD OFFOne or more ECG leads are not properly connected
when leads are selected as input.
ECG TOO LARGEECG signal too large for accurate shockable rhythm
analysis.
ECG TOO SMALLECG signal too small for accurate shockable rhythm
analysis.
ECG V1 LEAD OFFChest lead V1 is not properly att ach ed to pati ent.
User
Advisory
Technical Action
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
Check cable and patient
connection. Change
electrodes. Prepare
patient’s skin.
Reduce ECG size.
Increase ECG size.
Reattach V lead. Check
cable.
ECG V2 LEAD OFFChest lead V2 is not properly att ach ed to pati ent.
ECG V3 LEAD OFFChest lead V3 is not properly att ach ed to pati ent.
ECG V4 LEAD OFFChest lead V4 is not properly att ach ed to pati ent.
ECG V5 LEAD OFFChest lead V5 is not properly att ach ed to pati ent.
65
Reattach V lead.
Check cable.
Reattach V lead.
Check cable.
Reattach V lead.
Check cable.
Reattach V lead.
Check cable.
M Series Service Manual
Error MessageExplanation
ECG V6 LEAD OFFChest lead V6 is not properly att ach ed to pati ent.
ENTER ACCESS CODEManual mode access code needed.
ERASING REPORTSummary report being erased.
ECU CRC FAULTInvalid ECG samples detected over a one second
period.
EtCO2 COM ERRORNo or invalid communication from EtCO2 module. Return unit for service
FAX DIALINGPreparation for sending fax.
FAX DONETransmission complete.
FAX PREPARINGPreparing fax for transmission.
User
Advisory
Technical Action
Reattach V lead. Check
cable.
Enter access code to
enter manual mode
with AED.
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
to ZOLL Technical
Service Department.
FAX SENDINGTransmitting fax.
INSERT CARD
Check memory card
LOW BATTERYLow battery.
NO QRS DETECTUnit is in sync mode and heart rate is < 20 BPM or
No card installed in unit during manual or semiautomated modes.
QRS amplitude is too low for proper
synchronization.
66
Replace battery or plug
into AC power. Replace
charger.
Increase ECG size and/
or change lead.
M Series Service Manual
Error MessageExplanation
NO SHOCK ADVNo shock advised. Advisory message when analysis
finds non-shockable rhythm.
NOISY ECGNumber of noisy analysis intervals exceeds
threshold.
OPEN AIR DISCHARGECap voltage too high after discharge attempt, e.g.,
full energy discharge did not occur.
PACER DISABLEDUser prompt issued simultaneously with other pace
faults if pacing is disabled.
PACER FAULT 115Flyback pulse width cont rol circuit is not under
proper control of the processor and gate array.
PACER FAULT 116Failure to detect XPACE_ON.Replace high voltage
User
Advisory
Technical Action
Stop all patient
movement. Check
connections. Press
Analyze button again.
Replace paddles, and,
or high voltage module
and system board.
Replace high voltage
module or system
board.
Replace high voltage
module, capacitor, or
system board.
module, capacitor, and/
or system board.
PACER FAULT 117Pace relay is stuck closed.Replace high voltage
module, capacitor, and/
or system board.
PACER FAULT 121During pace, the pace pulse width <30ms or >50ms. Replace high voltage
module, or system
board.
PACER FAULT 122Pace current is more than 15mA above and below
selected value.
67
Replace high voltage
module, or system
board.
M Series Service Manual
Error MessageExplanation
PACER FAULT 123Measured pace rate is too fast compared to selected
rate.
PACER FAULT 126Issued in conjunction with message 122. Pace
current is more than 15mA and below selected value.
PADDLE FAULTCannot detect type of accessory attached to the
universal cable.
PERFORM CPRAdvisory message in AED auto defib mode.
PLACE ON ZERO CELLEtCO2 sensor cable plugged into unit for the first
time. Zeroing error or probe drift error detected.
POOR LEAD CONTACTOne or more ECG leads are poorly connected or not
connected to patient. (User configurable.)
User
Advisory
Technical Action
Replace high voltage
module or system
board.
Replace high voltage
module.
Replace paddles,
internal paddles, system
board, high voltage
module and/or
universal cable.
Zero sensor. Replace
sensor. Return to ZOLL
for service.
Check electrode
attachment to patient,
cable connector to
electrode, cable to unit
connector. Broken unit.
POOR PAD CONTACTElectrode impedance exceeds threshold. Ensure pads are
coupled to patient.
Check /replace pads or
universal cable.
Check impedance
circuit calibration.
Replace system board.
PRESS ANALYZEAlternate message for check patient prompt.
68
M Series Service Manual
Error MessageExplanation
PRESS CHARGEAdvisory message in conjunction with shock
advised.
PRESS SHOCKPrompt issued in AED auto defib mode when defib is
charged (ready).
RECORDER FAULT 142Strip chart system error.Check paper tray and
REPORT FULLSummary report memory full.
REPORT HALTEDSummary report stops printing unexpectedly.Turn unit off and then
RESERVED 1The watchdog timer is not functional in the unit. Turn off unit and then
RETRY ANALYSISAdvisory message in conjunction with noisy ECG.
SELECT 30J FOR TESTAttempt to run a self test at an energy other than 30J.
SELECT DEFIB MODEAnalyze button pressed in pace or monitor mode.
2
SENSOR WARM UP message displays for
EtC0
2
more than five minutes. Sensor defective.
Analysis halted.
User
Advisory
Technical Action
Replace sensor cable.
Erase summary report.
back on again. Print
Summary again. If fault
persists, replace system
board.
turn on to reset. If fault
persists, replace system
board.
SELECT LIMB LEADSPaddles or augmented ECG leads selected when
continuous analysis active or started.
SELECT PADSLead I, II, or III selected when analyze pressed.
SET CLOCKReal time clock failure: invalid date or time.Set date and time and/
SET PACE mAMultiple copy errors are the product of intended
software or memory errors. If error reoccurs other
than on entering pace the first time or after more than
10 minutes in other mode, the unit could be broken.
70
Select limb leads I, II,
III or MFE
or replace system
board.
Set pace current. If
broken, replace system
board.
M Series Service Manual
Error MessageExplanation
SET PACE RATEMultiple copy errors are the product of intended
software or memory errors. Multiple copies of pace
rate don't match. If error persists, unit could be
broken
SHOCK ADVISEDAdvisory message when analysis finds a shockable
rhythm.
SpO2 AMBIENT LIGHTAmbient light is too bright.Shield sensor from
SpO2 COMM ERRNo transmissions from SpO2unit received.
Communication error or no communication from
Sp0
module.
2
SpO2 PULSE SEARCHPulse search in progress.
SpO2 SENSOR F AULT 1Defective sensor.Replace Sp02 sensor.
STAND CLEAR(Auto defib mode only.) Single analysis mod e ju st
turned on and defib idle. Patient rhythm is being
analyzed.
SYSTEM FAULT 2MCU ROM checksum test failure or MCU RAM test
failure.
71
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
M Series Service Manual
Error MessageExplanation
SYSTEM FAULT 5ECU RAM test failure or ROM checksum test
failure.
SYSTEM FAULT 6No communications received from ECU for 4
seconds.
SYSTEM FAULT 7The A/D converter is not performing conversions in
a timely manner.
SYSTEM FAULT 36 P1MON is less than 412 counts or greater than 612
A/D counts. Pace/defib is disabled as long as
condition exits.
SYSTEM FAULT 37Disable pace/defib and MFE monitoring. Replace system board.
SYSTEM FAULT 38Failure to shutdown after “shutdown order” is
written to the RTC.
TEST FAILEDMCU performed ipeak test (defib peak current) and
unit failed during 30J self test.
User
Advisory
Technical Action
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
Replace system board.
Replace system board.
Replace system board.
Replace universal
cable, paddles or high
voltage module,
capacitor, or system
board.
TEST OKMCU performed ipeak test (defib peak current) and
unit passed 30J self test.
USE PADDLE DISCHGFront Panel discharge button is pressed when either
external paddles or internal spoons with discharge
buttons are connected.
72
M Series Service Manual
Error MessageExplanation
USE PADS(AUTO DEFIB MODE ONLY.)
Attempt to defib with paddles in auto defib (AED)
mode. Defib only allowed using PADS in AED
modes.
USE PADS TO PACEExternal paddles detected in pace mode.
USE ROOM AIR
ADAPTER
USER SETUP REQBoth copies of stored cal/config data are bad or have
VF ALARMS OFFVF alarms disabled in pace mode or when paddles
VX LEADS OFFV lead not properly attached to patient. “X” denotes
ZERO CO
ZERO CO
SENSORNew EtCO2 sensor needs to be zero calibrated.
2
ADAPTERNew EtCO2 airway adapter needs to be zero
2
Adapter zeroing started with EtCO2 in the adapter or
the adapter is on the REF or “0” cell.
never been programmed.
are selected as leads.
lead number.
calibrated.
User
Advisory
Technical Action
Place CO2 sensor on
adapter in room air.
Perform configuration
setup.
Reattach V lead.
Zero EtCO2 sensor.
Zero EtCO2 adapter.
73
M Series Service Manual
74
M Series Service Manual
Overview
This chapter provides instructions on how to disassemble and reassemble the M Series unit, and includes the following
sections:
• Required Equipment
• Parts That May Need Replacing After Disassembly
• Safety Precautions
• Overview of Modules
• 1. Removing the ZIF Keeper
Chapter 3
Disassembly Procedures
• 2. Removing the Front Panel
• 2A. Removing the Display
• 2B. Removing the Control Board
• 3. Removing the Upper Housing Assembly
• 4. Removing the System Board Assembly
• 5. Removing the Battery Interconnect Board Assembly
• 6. Removing the High Voltage/Charger Assembly
• 7. Removing the High Voltage Module Assembly
• 8. Removing the High Voltage Capacitor Assembly
• 9. Removing the System Interconnect Board
• 10. Removing the Printer/Recorder Motor
• 11. Removing the Lower Housing Assembly
• 12. Removing the Print Head Assembly
75
M Series Service Manual
• 13. Removing the PCMCIA Card Slot Assembly
• 14. Removing the Paddle Release Latch
Required Equipment
• No. 1 Phillips screwdriver.
• No. 2 Phillips screwdriver.
• Exacto-knife.
• Orange wooden stick. (Available from H.A. Stiles: 1-800-447-8537)
• 90° dental pick.
• Needle nose pliers.
• Kapton tape.
• 3M copper adhesive tape, or equivalent.
• 1/2” nut driver.
• Lar ge diagonal cutters.
• Strong glue, such as Loctite 420 or equivalent.
• Loctite needle tip dispenser.
Parts That May Need Replacing After Disassembly
If you are removing the Control Board from the Front Panel, you need to have :
• Main Selector knob replacement (ZOLL Part Number 9310-0521)
• Pacer/Output/Rate knob replacement (ZOLL Part Number 9310-0520)
If you are removing the Battery Interconnect Board, you may need to replace it with a new one, using ZOLL Part
Number 9301-0302, if connectors have been UV welded as in older M Series models.
76
M Series Service Manual
Safety Precautions
WARNING! SHOCK HAZARD!
CAUTIONTAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU CONDUCT
DEFIBRILLATOR TESTS OR REPAIRS.
• Only properly trained technicians should service the unit.
• The unit can contain deadly voltages even if the unit is turned off.
• Make sure to discharge the unit before working with it.
• Make sure you take the necessary precautions when working with static sensitive units. For example, you must wear
a conductive wrist strap (which touches your skin) connected to a grounding mat and to the earth ground. Y ou must
remove the wrist strap when you discharge high voltage or when you are working on energized equipment.
77
M Series Service Manual
Overview of Modules
The M Series unit contains 14 modules, as shown below.
Isolated Power Supply with EtCO
System Interconnect BoardBattery Interconnect Board Assembly with 3 Battery
2
SpO2 module with bracket for EtCO
Pin Gaskets
2
System Board Assembly
AC Charger Assembly
78
M Series Service Manual
Control Board (from Front Panel)High Voltage Module AssemblyHigh Voltage Capacitor Assembly
PCMCIA Card Slot Recorder MotorSpO
Isolated Power Supply for SpO
ModuleBiphasic Capacitor and Bridge Assembly
2
79
Module (without bracket)
2
M Series Service Manual
80
M Series Service Manual
Latch
Lift here
Latch
Keeper
1.Removing
the ZIF
Keeper
Tools Required
• Orange stick
Note: It is important to know this procedure
before you start disassembling the unit.
Removing the ZIF (Zero Insertion Force)
Keeper incorrectly can damage the unit’s
system board.
To reinstall the ZIF Keeper:
1. Place the ZIF Keeper over the laminate
cable and insert the flex cable into the
connector. Latch the connector.
Step 1: The connector must be facing you as
shown in the diagram.
Step 2: Angle and lift up the right end of the ZIF
Keeper from the connector and the board. Slide
the Keeper approximately 1 mm to the left, then
gently lift the left side to clear the connector.
Step 3: Gently pull the ZIF (Zero Insertion
Force) Keeper over the laminate cable and rotate
the latch upwards.
2. Lower the left end of the ZIF Keeper over
the connector end until it touches the
printed wire assembly (PWBA). The other
end of the connector should be angled.
3. Press the end of the ZIF Keeper down over
the end of the connector. Be careful that
the ZIF Keeper snaps over the end of the
connector.
81
M Series Service Manual
Multi-wire cable
2.Removing
the Front
Panel
Tools Required
• No. 2 Phillips screwdriver
• Orange stick
To reinstall the Front Panel:
1. Reinstall the laminate cable first with the
black band facing up and towards the system
circuit board.
2. Reconnect the multi-wire cable from the
display.
3. Reverse steps 1, 2, and 3 above to reinstall
the front panel.
Step 1: Remove the battery from the battery well
and place it in front of the unit.
Step 3: Place your thumbs in the main selector
switch cup and push up on the front panel to
release the panel from the unit. After the front
panel is removed, use the battery as a support for
the panel. Do not use the main selector switch as
point of leverage.
Step 2: Rotate unit on to its back side.Remove
the two Phillips head screws located on the left
and right sides on the bottom of the unit.
Step 4: Disconnect the multi-wire cable from the
system board by gently pulling the beige
connector by its sides towards the front of the
unit. Remove the ZIF keeper from the laminated
ribbon cable and then disconnect it. Lift the right
side first with the unit facing you.
82
M Series Service Manual
video display
assembly
2A.
Removing
the
Display
Tools Required
• No. 2 Phillips Screwdriver
• Exacto-Knife
• 3M Copper Adhesive Tape
To reinstall the Display
Reverse steps 1 through 3.
Step 1: Remove the grounding copper tape from
the outer display shield.
Step 3: Remove the Video Display Assembly.
Step 2: Remove the V ideo Display Assembly by
rotating the display upwards from the lower
portion of the display panel assembly.
83
M Series Service Manual
foam packaging
wooden stick
panel overlay circuit cable
2B.
Removing
the
Control
B o a r d
Tools Required
• No. 2 Phillips Screwdriver
• Orange (Wooden) Stick
• 1/2” Nut Driver
• Large Diagonal Cutters
• Strong Glue, such as Loctite 420 or
equivalent
Step 1: (Caution: The knob will be damaged
during this step.) Gently insert the cutters at the
edge of the main selector knob and pry outward
until the knob is removed. Then carefully remove
the 1/2” nuts without damaging the Selector
Switch.
Step 3: Rock the foam packaging back and forth
from the control panel board to remove the foam.
Step 2: Gently insert the cutters at the edge of the
pacer knobs and pry outward until the knobs are
removed. Then carefully remove the 1/2” nuts
without damaging the Pacer Switches and green
pace cups.
Step 4: Remove the panel overlay circuit cable
from the control board by lifting the side s of the
lock lever located under the front of the control
board. (See 1. Removing the ZIF Keeper.)
Note: If you are removing the Control Board
from the Front Panel you may need to replace
the following parts:
• Main Selector knob replacement (ZOLL Part
Number 9310-0521)
• Pacer Output/Rate knob replacement (ZOLL
Part Number 9310-0520).
84
M Series Service Manual
50 pin laminate cable
2B.
Removing
the
Control
Board
(Continued)
To reinstall the Control Board
1.Set the dip switches and attach the SHOCK
button LED actuator.
2.Place a small amount of strong glue on the
end of the mode selector switch. Slide the
replacement knob on. After the glue dries,
rotate the selector knob to ensure that it is
properly glued in place.
Step 5: Disconnect the speaker microphone cable.
Remove the Control Panel Board from the Front
Panel Assembly. Important: Note the position of
the dip switches because they must be in the same
position for reinstallation.
Step 6: Remove the 50 pin laminate cable from
the control board. See 1.0 Removing the ZIF
Keeper.
3.Replace the Pacer Output/Rate knobs, if
applicable.
85
M Series Service Manual
battery pin gasket
3.Removing
the Upper
Housing
Assembly
Tool Required
• No. 1 and 2 Phillips screwdriver
Set-up
Remove Front Panel (See Step 2.)
1. Remove the two screws securing the
universal cable.
2. Do not lose the O-ring when removing the
universal cable.
Step 1: Remove two screws from the back side of
the Upper Housing Assembly and three screws
from the front.
Step 3: Make sure that the rubber gaskets are
covering the battery contact pins. If the gaskets
are still seated in the housing, remove and place
them onto the contact pins. Before installing the
Upper Housing ensure that the battery pin gaskets
are set properly.
Step 2: Secure lower housing to the table by
pressing downward in the paddle well. Using the
carry handle, lift the Upper Housing upward.
3. Remove the three Phillips screws in the
front of the Upper Housing.
To reinstall the Upper Housing Assembly,
reverse the above steps.
Ensure the battery pin gaskets are properly set.
86
M Series Service Manual
4.Removing
the
System
Board
Assembly
Tools Required
• #1 and #2 Phillips Screwdriver
• Grounding Mat
• Grounding Wrist Strap
2. Remove the Front Panel Assembly.
3. Remove the Upper Housing Assembly.
To reinstall the System Board Assembly, reverse the steps.
• Needle Nose Pliers
• Orange Stick
WARNING! You can damage the hardware of
the unit. You must use ESD grounding before
you handle any printed circuit boards on the
unit.
Setup
Before you begin this procedure, make sure you
are grounded.
1. Know or review 1.0 Removing the ZIF Keeper procedure.
87
M Series Service Manual
battery interconnect laminate cable
Step 1: Using needle nose pliers, remove left and
right multi-wire cables on the back side of the
system board. To avoid damage to the cable, do
not pull the wires. Hold system board securely
with one hand. DO NOT let it fall forward to
prevent excess tension on the unit’s wires.
Step 3: Lower the system board with one hand
and remove the battery interconnect laminate
cable from the center of the system board. (See 1.
Removing the ZIF Keeper.)
Step 2: Remove the two wire speaker cable (if
applicable).
Step 4: Gently roll the system board forward and
rest it on the battery.
88
M Series Service Manual
Step 5: Remove the ZIF Keeper. (See 1.
Removing the ZIF Keeper.)
Step 7: For M Series units with the SpO
SpO
with EtCO2 only: Remove the 20 pin power
2
cable by lifting the slide locking tab upwards.
and/or
2
Step 6: For Biphasic M Series units only:
Remove the 20 pin power cable by lifting the slide
locking tab upwards.
89
M Series Service Manual
charger cable
10 pin laminate cable
5.Removing
the
Battery
Interconn
ect Board
Assembly
Tools Required
• Orange stick
Setup
Step 1: Identify the Battery Interconnect Board. Step 2: Remove the wide laminate cable from the
high voltage module connector by lifting the
cable vertically.
Step 3: Remove the push pin and small insert
collar.
Step 4: Rotate the unit around so that the rear of
the unit faces you. Rotate the Battery Interconnect
Board upwards toward the front of the device.
Hold the board vertically while removing the 10pin laminate cable from connectors by lifting the
connector lock.
1. Remove the Front Panel Assembly.
2. Remove the Upper Housing Assembly
3. Remove the System Board Battery Cable.
To reinstall the Battery Interconnect Board,
reverse the steps.
NOTEFor Step 4, remember to carefully remove
the cable from the Charger Assembly
when disassembling and reinstalling it.
90
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.