The issue date for the X Series Service Manual (REF 9650-001356-01 Rev. D) is November, 2013.
0123
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA USA
01824-4105
ZOLL International Holding B.V.
Newtonweg 18
6662 PV ELST
The Netherlands
If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to det ermine if
additional product information updates are available.
9650-001356-01 Rev. DZOLL X Series Service Manualiii
TABLEOF CONTENTS
iv9650-001356-01 Rev. D
Overview
Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to
review this section thoroughly before attempting to use or service the X Series unit.
Chapter 1—Maintenance Tests describes step-by-step procedures for various maintenance
tests.
Chapter 2—Troubleshooting provides a listing of error messages to help the service
technician detect faults and repair them.
Chapter 3—Disassembly Procedures describes step-by-step procedures for disassembling
modules in the X Series unit.
Chapter 4—Replacement Parts List displays a complete list of ZOLL part numbers for field
replaceable parts available for the X Series unit, allowing the service person to identify and
order replacement parts from ZOLL.
Preface
The ZOLL® X Series™ Service Manual is intended for the trained biomedical technician
whose responsibility is to identify malfunction s and/or make repairs at the subassembly level.
The X Series Service Manual has five main sections and two appendixes.
Chapter 5—Functional Description provides technical descriptions for the X Series major
subassembly modules.
Appendix A—X Series Electrical Hardware Interconnect Diagram and X Series Signal and
Power Interconnect Schematic.
Appendix B—Contains simulators and settings that may be used to assess the performance of
the NIBP module.
9650-001356-01 Rev. DZOLL X Series Service Manual1
PREFACE
Safety Considerations
The following section describes general warnings and safety considerations for operators and
patients. Service technicians should review the safety considerations prior to servicing any
equipment and read the manual carefully before attempting to disassemble the unit. Only
qualified personnel should service the X Series unit.
Federal (U.S.A.) law restricts this unit for use by or on the order of a physician.
Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug
Administration (FDA) under section 510(K) of the Medical Device Act to obtain approval to
market is based upon the use of ZOLL accessories such as disposable electrodes, patient cables
and batteries. The use of external pacing/defibrillation electrodes and adapter units from
sources other than ZOLL is not recommended. ZOLL makes no representations or warranties
regarding the performance or effectiveness of its products when used in conjunction with
pacing/defibrillation electrodes and adapter units from other sources. If unit failure is
attributable to pacing/defibrill ation e lectrodes or adapter units not manufactured by ZOLL, this
may void ZOLL's warranty.
Only qualified personnel should disassemble the X Series uni t.
WARNING! This unit can generate up to 2775 volts with sufficient current to cause lethal shocks.
All persons near the equipment must be warned to “STAND CLEAR” prior to dischargin g the
defibrillator.
Do not discharge the unit’s internal defibrillator energy more than three times in one minute or
damage to the unit may result.
Do not discharge a battery pack except in a ZOLL SurePower
Do not use the X Series in the presence of flammable agents (such as gasoline), oxygen-rich
atmospheres, or flammable anesthetics. Using the unit near the site of a gasoline spill may
cause an explosion.
Do not use the unit near or within puddles of water.
Additional Reference Material
In addition to this guide, there is a X Series Operator’s Guide which is a comprehensive
reference work that describes all the user tasks needed to operate the X Series.
TM
Charging Station.
29650-001356-01 Rev. D
Conventions
Conventions
WARNING! Warning statements describe conditions or actions that can result in personal injury or death.
CautionCaution statements describe conditions or actions that can result in damage to the unit.
Note:Notes contain additional information on using the defibrillator.
Service Policy Warranty
In North America: Consult your purchasing agreement for terms and conditions associated with
your warranty. Outside of North America, consult a ZOLL authorized representative.
In order to maintain this warranty, the instructions and procedures contained in this manual
must be strictly followed. For additional information, please call the ZOLL Technical Service
Department 1-800-348-9011 in North America.
Technical Service
If the ZOLL X Series unit requires service, contact the ZOLL Technical Service Department:
Telephone: 1-978-421-9655; 1- 800-348-9011
Fax: 1-978-421-0010
Email: techsupport@zoll.com
Have the following information available for the Te chnical Service representative:
• Unit serial number.
• Description of the problem.
• Department where equipment is used.
• Sample chart recorder strips or electronic log files documenting the problem, if applicable.
• Purchase Order to allow tracking of loan equipment.
• Purchase Order for a unit with an expired warranty.
If the unit needs to be sent to ZOLL Medical Corporation, obtain a service request number
(SR#) from the T e chn ic al Service representative. Return the unit in its original container to:
ZOLL Medical Corporation
269 Mill Road
Chelmsford, Massachusetts 01824-4105
Attn: Technical Service Department (SR#)
Telephone: 1-800-348-9011; 1-978-421-9655 FAX: 978-421-0010
9650-001356-01 Rev. DZOLL X Series Service Manual3
PREFACE
Technical Service for International Customers
International customers should return the unit in its origin al co nt ain er to the neare s t aut hor ize d
ZOLL Medical Corporation Service Center. To locate an authorized service center, contact the
International Sales Department at ZOLL Medical at the above address.
49650-001356-01 Rev. D
X Series Overview
This chapter includes step-by-step procedures for various maintenance tests. At a minimum, a
trained biomedical technician must perform a CO
once per year. A CO
calibration. It is necessary to perform maintenanc e tests after a module is repaired to ensure that
the functions of the X Series unit work properly . Use the checklist at t he back of this document
(ZOLL X Series Maintenance Tests Checklist) to record your results of the maintenance tests.
Because the X Series units must be maintaine d ready for immediate use, it is important for
users to conduct the Operator’s Shift Checklist procedure at the beginning of every shift. This
procedure can be completed in a few minutes and requires no additional test equipment. (See
the ZOLL X Series Operator’s Guide for the Operator’s Shift Checklist.)
Chapter 1
Maintenance Test s
calibration and an NIBP calibration check
calibration check may be performed at any time to verify the CO2
2
2
9650-001356-01 Rev. DZOLL X Series Service Manual5
CHAPTER 1MAINTENANCE TESTS
This chapter describes the following maintenance tests:
• 1.0 Physical Inspection of the Unit
• 2.0 Operational TestsOperational Tests
• 3.0 Leads T e st
• 4.0 Heart Rate Display Test
• 5.0 Heart Rate Alarm Test
• 6.0 Power Supply Test
• 7.0 Leakage Current Test
• 8.0 Shock Test
• 9.0 Advisory Test
• 10.0 Synchronized Cardioversion Test
• 11.0 Pacer Test
• 12.0 Printer Test
• 13.0 Treatment Report Test
• 14.0 SpO2 Test (for SpO
• 15.0 SpO2, SpCO, and SpMet Test (for SpO
• 16.0 EtCO2 Calibration (required for annual calibration)
• 17.0 EtCO2 Calibration Check
• 18.0 EtCO2 Test
• 19.0 IBP Test
• 20.0 NIBP Calibration Check
• 21.0 NIBP Leak Test
• 22.0 Deflation Test
• 23.0 Inflation Test
• 24.0 NIBP Test
• 25.0 Temperature Test
• 26.0 Wi-Fi Test (Optional)
• 27.0 Audio Recording Test (Optional)
feature without SpCO and SpMet)
2
feature with SpCO, or SpCO and SpMet)
2
Before You Begin the Maintenance Tests
• Assemble the tools listed in following section, “Equipment You Need to Perform the
Maintenance Tests.”
• Keep an extra fully charged X Series battery available.
• Schedule an hour to conduct the entire maintenance test.
• Perform the tests in the order presented.
• Perform all the steps of each test procedure.
• Complete all the steps of the procedure before evaluating the test results.
69650-001356-01 Rev. D
Equipment You Need to Perform the Maintenance Tests
The equipment listed below is used in the maintenance procedures in this chapter. You can
substitute an equivalent device; however, not all simulators and analyzers will produce the
same results. Be sure to follow the manufacturer’s recommendations for conducting the
maintenance tests.
BP Pump/Dynatech Nevada Inc. part number 3010-0378)
• ECG simulator
• Fluke Index 2 SpO
• IBP/Temp simulator
• NIBP simulator
• 5% CO
• CO
• Stop watch
• Battery Current Test Fixture 9100-3055-TF
• Agilent E3633A Power Supply
• Agilent 34401 Digital Multi-meter
calibration gas cylinder with CO2 sample line (kit available from Scott Medical
2
Products, PN 4653ORF-BD)
sampling line airway adapter
2
simulator
2
X Series Overview
Accessories You Need for the X Series Options Maintenance Tests
• 3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is
installed.)
• Dual lumen NIBP hose
• IBP cable
• SpO
• EtCO
• Paddles
• Printer paper
• Battery
• Auxiliary power source
• USB thumb drive
• PC w/ Windows media player and speakers
cable and sensor
2
cable and sensor
2
9650-001356-01 Rev. DZOLL X Series Service Manual7
CHAPTER 1MAINTENANCE TESTS
Maintenance Tests
1.0 Physical Inspection of the Unit
Observe this...Pass/Fail
Housing
1.1
Is the unit clean and undamaged?
oo
1.2 Does the unit show signs of excessive wear?
1.3 Does the handle work properly?
1.4 Does the recorder drawer open and close properly?
1.5 Are input connectors clean and undamaged?
1.6 Are there any cracks in the housing?
1.7 Does the front panel have any damage or cracks?
1.8 Are there any loose housing parts?
1.9 Paddles (if applicable)
Do the adult and pedi plates have major scratches or show signs of damage?
1.10 Do the adult shoes slide on and off easily to expose the covered pedi plates?
1.11 Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
1.12 Cables
Are all cables free of cracks, cuts, exposed or broken wires?
1.13 Are all bend/strain reliefs undamaged and free of excessive cable wear?
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo
1.14 Battery
Is the X Series battery fully charged?
1.15 Is the battery seated in the battery well correctly?
1.16 Is battery housing free of cracks and damage?
1.17 Are connectors free of damage and excessive wear?
89650-001356-01 Rev. D
oo
oo
oo
oo
2.0 Operational Tests
The X Series unit comes equipped with routine operational tests that can be performed
periodically. With the exception of the Physical Inspection of the Unit test, these tests contain
instructions within them that provide guidance during the tests.
The operational tests on the unit can be accessed by doing the following:
1. Turn on the X Series unit.
Note:If the unit is in AED mode, press the Manual quick access key to go into manual
mode. Use the navigation keys to sel ect th e fou r di gits in the Manual Mode pass code.
Press SAVE when you are finished. If the unit has not been config ured to enter a pass
code, the message Exit to Manual Mode is displayed. Use the na vigation ke ys to select Yes.
2. Press the More () then the Setup () quick access keys.
3. Use the navigation keys to select Supervisor>Service>Device Tests. The operational tests
are displayed. Use the navigation keys to select one of the tests.
Note:You need a supervisor passcode to enter the Supervisor menu .
Maintenance Tests
9650-001356-01 Rev. DZOLL X Series Service Manual9
CHAPTER 1MAINTENANCE TESTS
2.1 Keypad Test
This test checks all the front panel buttons on the unit to make sure they are working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
Note:To exit screen after test is complete, press the “Enter” button 3 times to display arrow.
2.2 LED Test
This test checks all the Light Emitting Diodes in the unit to make sure they are working
correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
Note:The speed of the flash varies by color.
109650-001356-01 Rev. D
Maintenance Tests
2.3 LCD Test
This test checks the colors of the Liquid Crystal Display on the unit.
Follow the instructions and use the navigation keys to move around the display and make
selections.
2.4 RFU Indicator Test
This test checks the Ready For Use Indicator on the unit to make sure it is working correctly.
Follow the instructions shown on the display.
9650-001356-01 Rev. DZOLL X Series Service Manual11
CHAPTER 1MAINTENANCE TESTS
2.5 Audio Test
This test checks the audio tones on the unit to make sure they are working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
2.6 Printer Test
This test checks the printer to make sure it is working correctly.
Follow the instructions shown on the display.
129650-001356-01 Rev. D
2.7 Battery Test
This test checks the battery to make sure it is working correctly.
Follow the instructions on the display.
Maintenance Tests
2.8 Fan Test
This test checks the fan in the unit to make sure it is working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
9650-001356-01 Rev. DZOLL X Series Service Manual13
CHAPTER 1MAINTENANCE TESTS
2.9 USB Test (Optional)
Note:Before starting this test, contact the ZOLL Technical Service Department to obtain a
test file, or create a file (see instructions below). Once you have obtained a copy, save
it on a USB thumb drive.
This test allows you to perform the USB test to make sure it is working correctly.
Follow the instructions and use the navigation keys to move around the display and make
selections.
When these tests are completed, use the navigation keys to exit out of the Service menu.
Creating a Test File
1. Create a Text file using Microsoft Notepad.
2. Open new file in Notebook.
3. Click in the first line of the file and type “Thunderbolt Testfile".
4. Save file as Tbolt_usb.txt.
149650-001356-01 Rev. D
Maintenance Tests
3.0 Leads Test
Equipment3-lead, 5-lead, and 12-lead cables, Fluke Impulse 7000DP Analyzer
Note:Test each cable separately.
Test SetupConnect the lead wires appropriate for each test to the Fluke Impulse 7000DP Analyzer or
equivalent, then connect the ECG cable to the X Series unit.
Do this...Observe this...Pass/Fail
3.1 Turn on the X Series unit.
Note: Make sure the lead is not
displayed.
3.2 Disconnect one lead from the
simulator.
3.3 Reconnect the lead. Repeat step 2
with the remaining leads.
3 .4 Repeat steps 2 and 3 for 5 lead and 12
lead cables.
No ECG LEAD FAULT message
displayed.
The ECG LEAD F AULT message displays
within 3 seconds (if configured).
Wa it for ECG LEAD FAULT message to
clear from the display (if configured).
Note: If heart rate alarm sounds, on the
X Series unit:
Press the More quick access key () key
to access the second set of quick access
keys.
Press the Alarm quick access key ().
Press the Alarm Cancel quick access key
().
oo
oo
oo
oo
9650-001356-01 Rev. DZOLL X Series Service Manual15
CHAPTER 1MAINTENANCE TESTS
4.0 Heart Rate Display Test
EquipmentFluke Impulse 7000DP Analyzer
ECG Cable (3-lead, 5-lead or 12-lead)
Test SetupPress Lead quick access key until II displays.
Connect the ECG leads to the ECG studs on Analyzer.
Do this...Observe this...Pass/Fail
4.1 Set the ECG Simulator to 120BPM.The Heart Rate displays as 120 +/- 4 bpm.
oo
169650-001356-01 Rev. D
5.0 Heart Rate Alarm Test
I, II,
III...
EquipmentFluke Impulse 7000DP Analyzer
Do this...Observe this...Pass/Fail
5.1
Press the Leads quick access key
() on the X Series unit and
select lead II.
5.2 On the X Series unit:
• Press the More quick
access key () key to
access the second set of
quick access keys.
• Press the Alarm quick
access key ().
• Press the Limits softkey.
5.3 On the X Series unit, select HR/PR
Alarm in the Alarm Settings menu
and use the navigation keys to do
the following:
• Set the Upper Limit Enable
to ON.
• Set the Lower Limit
Enable to ON.
• Set the Lower limit to 35.
• Set the Upper limit to 150.
Use the navigation keys to exit the
Alarm Settings menu.
5.4 On the X Series unit, press the
Alarm Cancel quick access key
().
5.5
Remove a lead wire from the
simulator
.
ECG
Maintenance Tests
II displays next to the Leads quick
access key.
The Alarm Settings menu displays.
The red alarm suspension timer
displays at the top of the screen next
to the message area.
T
he ECG LEAD FAULT message
displays.
oo
oo
oo
5.6
On the X Series unit, press .
5.7
Reattach ECG Lead wire to
ECG
simulator.
5.8 Set simulator to 30 BPM or lower.
5.9
On the X Series unit, press .
The Alarm Suspend key flashes
The ECG LEAD FAULT message no
longer displays.
Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR LOW ALARM message displays.
The Alarm Suspend key flashes.
.
The
oo
oo
red alarm suspension timer displays
at the top of the screen next to the
message area.
5.10 Set the simulator to 120 BPM.
5.11
On the X Series unit, press .
9650-001356-01 Rev. DZOLL X Series Service Manual17
Alarm is enabled. The alarm suspend
key flashes. The alarm suspension
timer disappears.
oo
CHAPTER 1MAINTENANCE TESTS
Do this...Observe this...Pass/Fail
5.12 Set simulator to 160 BPM or higher.
5.13
On the X Series unit, press
the
Alarm Silence/Reset button
()
.
Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR HIGH ALARM message displays.
Alarm is suspended for 90 seconds.
The alarm suspension timer appears
in black.
oo
oo
189650-001356-01 Rev. D
Maintenance Tests
6.0 Power Supply Test
EquipmentBattery Current Test Fixture (P/N 9100-3055-TF), Power Supply, Digital Multi-meter.
Test Setup
1. Set the Power Supply to 12V VDC 10 A.
2. Turn off the Power Supply output.
3. Connect the positive (PS+) post of the Power Supply to t he p ositiv e (PS+) post (PS Red) of
the Battery Current Test Fixture.
4. Connect the negative (PS-) post of the Power Supply to the negative (PS-) post (PS Black) of
the Battery Current Test Fixture.
5. With the Battery Current Test Fixture out of the battery well, turn on the Power Supply
output.
6. Set the Digital Multi-meter to measure DC I DC Amps.
7. Set the rear/front switch on the Digital Multi-meter to rear.
8. Connect the Digital Multi-meter (I+) terminal of the Battery Current Test Fixture to the red
terminal (I) on the back of the Digital Multi-meter.
9. Connect Digital Multi-meter (I-) terminal of the Battery Current Test Fixture to the black
terminal (LO) on the back of the Digital Multi-meter.
Do this...Observe this...Pass/Fail
6.1 Install the Battery Current Test
Fixture into battery well and wait
about five seconds.
6.2 Make sure the X Series unit is
turned off.
6.3 Press the I Measure switch on the
Battery Current Test Fixture.
6.4 Turn on the X Series unit, and wait
about ten seconds until the power
has stabilized.
Verify that the reading on the digital
multi-meter is between
0.583mA-0.770mA.
oo
9650-001356-01 Rev. DZOLL X Series Service Manual19
CHAPTER 1MAINTENANCE TESTS
Do this...Observe this...Pass/Fail
6.5 On the X Series unit, do the
following:
• Press the More () then the
Setup () quick access keys.
• Use the navigation keys to
select Supervisor>Display/
Configuration>Device Info
and select Restore All Factory
Defaults.
Note: You need a supervisor
passcode to enter the Supervisor
menu. Also the unit will reboot
during the of restoring factory
defaults.
6.6 Press the I Measure switch on the
Battery Current Test Fixture.
Observe the current measurement on
the DMM, and verify that the reading
is between 0.47A-1.25A.
oo
209650-001356-01 Rev. D
Maintenance Tests
7.0 Leakage Current Test
EquipmentSee the manufacturer’s instructions or supplied specifications for the leakage tester you use.
SetupSee the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/
posterior paddles. Perform these tests at the line-power voltage and frequency used in your
installation.
Maximum Leakage Acceptance Limits
Patient LeakageMains on Applied Part
NormalSingle FaultSingle Fault
ECG10
MFC10
Earth Leakage2500
50 100
50 100
NormalSingle Fault
5000
9650-001356-01 Rev. DZOLL X Series Service Manual21
CHAPTER 1MAINTENANCE TESTS
I, II,
III...
8.0 Shock Test
WARNING!Take the necessary precautions to guard against shock or injury before you start
conducting the defibrillator tests.
Keep hands and all other objects clear of the multi-function cable connections and
defibrillator analyzer when discharging the defibrillator.
Before you discharge the defibrillator, warn everyone near the equipment to STAND
CLEAR.
8.4 Connect the X Series MFC connector to the
pads connection.
8.5 On the X Series unit, apply power to the unit
using a battery or an AC adapter.
8.6 Connect the MFC cable to the X Series unit.
8.7 Turn on the X Series unit, and press the
Leads quick access key
PADS.
8.8 On the X Series unit, set energy level to 5J
by pressing the Energy Select DOWN arrow .
8.9 On the X Series unit, press the CHARGE
button.
8.10 On the X Series unit, press the Shock button.
8.11 On the Analyzer, observe and record the
energy measurement.
() to select
The unit charges to 5J, the red
Shock button illuminates and a
ready tone sounds.
Energy is delivered at 5J. Verify
that the delivered energy is within
15% based on the values in the
table “Delivered Energy at 200J
Defibrillator Setting into a Range
of Loads” on page 24.
oo
oo
8.12 On the X Series unit, set the energy level to
50J by pressing the Energy Select UP arrow.
8.13 On the X Series unit, press the CHARGE
button.
229650-001356-01 Rev. D
The unit charges to 50J, the red
Shock button illuminates and a
ready tone sounds.
oo
Maintenance Tests
Do This...Observe This...Pass/Fail
8.14 On the X Series unit, press the Shock button.
8.15 On the Analyzer, observe and record the
energy measurement.
8.16 On the X Series unit, set energy level to 100J
by pressing the Energy Select UP arrow.
8.17 On the X Series unit, press the CHARGE
button.
8.18 On the X Series unit, press the Shock button.
8.19 On the Analyzer, observe and record the
energy measurement.
8.20 On the X Series unit, set energy level to 200J
by pressing the Energy Select UP arrow.
8.21 On the X Series unit, press the CHARGE
button.
Energy is delivered at 50J. Verify
that the delivered energy is within
15% based on the values in the
table “Delivered Energy at 200J
Defibrillator Setting into a Range
of Loads” on page 24.
The unit charges to 100J, the
Shock button illuminates and a
ready tone sounds.
Energy is delivered at 100J.
Verify that the delivered energy is
within 15% based on the values in
the table “Delivered Energy at
200J Defibrillator Setting into a
Range of Loads” on page 24.
The unit charges to 200J, the red
Shock button illuminates and a
ready tone sounds.
oo
oo
oo
oo
8.22 On the X Series unit, press the Shock button.
8.23 On the Analyzer, observe and record the
energy measurement.
8.24 On the X Series unit, press the CHARGE
button.
8.25 When the X Series unit is charged, start the
stop watch.
8.26 Turn off the X Series unit.
Energy is delivered at 200J.
Verify that the delivered energy is
within 15% based on the values in
the table “Delivered Energy at
200J Defibrillator Setting into a
Range of Loads” on page 24.
The unit charges to 200J, the red
Shock button illuminates, and a
ready tone sounds.
Verify that after 50 seconds the
tone changes and within 10
seconds the unit does an internal
discharge and displays
DISARMED TIMEOUT.
oo
oo
oo
9650-001356-01 Rev. DZOLL X Series Service Manual23
CHAPTER 1MAINTENANCE TESTS
Delivered Energy at 200J Defibrillator Setting into a Range of Loads
sensor to the X Series unit. Verify that SpO2 numeric
2
displays SEARCHING.
SpO
SpO
sensor to the simulator.
2
SpO
and 60 PPM.
2
and 60 PPM.
2
Verify that after a brief period of
initialization, the
displays an
67% and 73%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
Verify that after a brief period of
initialization, the
displays an
77% and 83%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
SpO
SpO
SpO
numeric
2
value between
2
SpO
numeric
2
value between
2
oo
oo
oo
oo
14.5
Set the simulator to 90%
14.6
Disconnect the
309650-001356-01 Rev. D
SpO
SpO
simulator.
2
and 60 PPM.
2
Verify that after a brief period of
initialization, the
displays an
87% and 93%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
SpO
SpO
value between
2
numeric
2
oo
oo
Maintenance Tests
15.0 S pO2, SpCO, and SpMet Test
(for Sp O
feature with SpCO, or SpCO and SpMet)
2
Note:Use this test if your unit has SpO2 and SpCO, or SpO2, SpCO and SpMet. If you do n ot
have the SpMet feature on the unit, disregard the SpMet values in the test below.
EquipmentFluke Index 2 SpO
Simulator (or equivalent), Masimo Rainbow 3 ft BioTek Test Sensor
2
(Masimo P/N 3379) or Masimo Rainbow 3 ft Direct Connect Optical Test Sensor
Test SetupDisconnect any ECG cable or IBP probes from the X Series unit.
Do This...Observe This...Pass/Fail/N/A
15.1 Power on the X Series unit without the
SpO
sensor attached.
2
15.2
Connect the
Series unit.
15.3
Set the simulator to 75%
BPM. Attach the
simulator.
SpO
sensor to the X
2
SpO
SpO
and 50
2
sensor to the
2
Verify that the
displays INITIALIZING, then
goes blank.
Verify that SpO2 numeric
displays SEARCHING.
Verify that after a brief period of
initialization,
The
SpO
SpO
value between 72.1% and
2
78%.
The
SpCO numeric displays an
SpO
numeric displays an
2
SpCO value between 13.5 and
21.5.
SpMet numeric displays an
The
SpMet value between 5.5 and
9.5.
Verify that the displayed pulse
rate reads between 44.5 and 55.5
BPM.
numeric
2
oo
oo
ooo
oo
15.4
Set the simulator to 75%
BPM.
SpO
and 230
2
Verify that after a brief period of
initialization,
SpO
The
SpO
78%.
The
numeric displays an
2
value between 72.1% and
2
SpCO numeric displays an
oo
SpCO value between 12.8 and
20.8.
The
SpMet numeric displays an
oo
SpMet value between 5.2 and
9.2.
Verify that the displayed pulse
rate reads between 222.7 and 236
BPM.
9650-001356-01 Rev. DZOLL X Series Service Manual31
ooo
oo
CHAPTER 1MAINTENANCE TESTS
Do This...Observe This...Pass/Fail/N/A
15.5
Set the simulator to 95%
BPM.
15.6
Set the simulator to 95%
BPM.
SpO
SpO
and 230
2
and 50
2
Verify that after a brief period of
initialization,
The
SpO
numeric displays an
2
SpO
value between 91.3% and
2
98.8%.
SpCO numeric displays an
The
SpCO value between -.5 and 7.5.
The
SpMet numeric displays an
SpMet value between -.5 and 3.5.
Verify that the displayed pulse
rate reads between 222.7 and 236
BPM.
Verify that after a brief period of
initialization,
The
SpO
numeric displays an
2
SpO
value between 91.3% and
2
98.8%.
The
SpCO numeric displays an
SpCO value between -.5 and 7.5.
The
SpMet numeric displays an
SpMet value between -.5 and 3.5.
Verify that the displayed pulse
rate reads between 44.5 and 55.5
BPM.
oo
oo
ooo
oo
oo
oo
ooo
oo
15.7
Disconnect the
SpO
simulator.
2
329650-001356-01 Rev. D
16.0 EtCO2 Calibration (required for annual calibration)
Maintenance Tests
Equipment5% CO
Products, PN 4653ORF-BD)
16.1
16.2
16.3
16.4
16.5
1 6.6 Continue to apply the calibration gas until the
calibration gas cylinder with CO2 sample line (kit available from Scott Medical
2
Do This...Observe This...Pass/Fail
Turn on the X Series unit.
On the X Series unit, select
Supervisor>Service>Diagnostic>CO
Calibration to enter the CO
Calibration
2
>
2
control panel.
Connect the X Series unit to the 5% CO2
calibration gas cylinder using the sample
line included with the calibration kit.
From the CO2 Calibration Menu, select
Calibration.
Activate the calibration gas flow by
pressing the top valve on the calibration gas
canister.
oo
status in the CO
displays: CALIBRATION GAS CAN BE
REMOVED.
Calibration control panel
2
• When the calibration is complete, the
status in the CO
displays: CALIBRATION DONE-OK
Calibration control panel
2
The calibration date will be updated upon
exiting the CO
Te st control panel.
2
oo
9650-001356-01 Rev. DZOLL X Series Service Manual33
CHAPTER 1MAINTENANCE TESTS
17.0 EtCO2 Calibration Check
Equipment5% CO
Products, PN 4653ORF-BD)
Note:If you have previously run the CO2 Calibration procedure, start at step 3.
17.1
17.2
17.3
17.4
17.5
calibration gas cylinder with CO2 sample line (kit available from Scott Medical
2
Do This...Observe This...Pass/Fail
On the X Series unit, select
Supervisor>Service>Diagnostic>CO
Calibration to enter the CO
Calibration
2
>
2
control panel.
Connect the X Series unit to the 5% CO2
calibration gas cylinder using the sample
line included with the calibration kit.
From the CO2 Calibration control panel,
select Verify Calibration.
Activate the calibration gas flow by
pressing the top valve on the calibration gas
canister.
Continue to apply the calibration gas until
the status in the CO
Calibration control
2
oo
panel displays: CALIBRATION GAS CAN BE REMOVED.
17.6
Verify that the unit displays:
VERIFICATION DONE - OK.
17.7
Verify that the measured gas reading is
between 4.8 and 5.2 (Vol%).
17.8
Exit out of calibration by pressing the Back
arrow key () three times or pressing the
Display/Home button ().
oo
oo
349650-001356-01 Rev. D
18.0 EtCO2 Test
Maintenance Tests
EquipmentCO
18.1 On the X Series unit, install the airway
18.2
18.3 On X Series unit, gently blow into the
sampling line airway adapter
2
Do this...Observe this...Pass/Fail
adapter set and sampling line in the
connector at the side of the unit.
CO
CO
2
Press the
CO
2
activate the CO
quick access key () to
trace.
2
2
airway adapter set while observing the
The CO
The CO2 waveform changes on the
display.
trace is on the display.
2
oo
oo
display.
9650-001356-01 Rev. DZOLL X Series Service Manual35
CHAPTER 1MAINTENANCE TESTS
19.0 IBP Test
Tools NeededIBP Simulator (with accuracy better than +/- 1 mmHg), or equivalent
SetupConnect the test cables to IBP patient simulator.
Do This...Observe This...Pass/Fail
1 9 . 1 Connect the IBP patient simulator to the X
Series P1 channel.
1 9.2 Set the IBP patient simulator to 0 mmHg.
19.3 On the X Series unit, press > IBP.
19.4 On the X Series unit, Press > P1->0<-.Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).
The waveform for channel 1 is
present.
19.5 Exit th e menu to see the waveform.
19.6 Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).
1 9.7 Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.
19.8 Disconnect IBP cable.
Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).
oo
oo
oo
369650-001356-01 Rev. D
Maintenance Tests
20.0 NIBP Calibration Check
EquipmentNIBP Simulator, dual lumen hose w/ NIBP test fixture cuff (9100-3072-TF) or equivalent
Note:If ECG simulator is connected to the device, either disconnect ECG simulator or
disable the Smart Cuff feature in the Supervisor Menu.
SetupAttached the hose connected to the T-fitting of the NIBP test fixture cuff to t he pressure port o f
the NIBP simulator. Set simulator to Static Pressure.
Do This...Observe This...Pass/Fail
20.1 Turn on the X Series uni t.
Note: If the unit is in AED mode, press
the Manual quick access key to go into
manual mode.
20.2 Connect a dual lumen hose from the NIBP
simulator to the NIBP port of the X
Series.
20.3 On the X Series unit, do the following to
enter the NIBP Test control panel:
• Press the More () then the Setup
() quick access keys.
• Use the navigation keys to select
Supervisor>Service>Device
Tests>NIBP Test.
Note: You need a supervisor passcode to
enter the Supervisor menu.
20.4 From the X Series NIBP Test control
panel, use the navigation keys to select:
• Close Valve
• Turn Pump On
20.5 Monitor the pressure reported by the
external pressure standard.
20.6 Once the pressure reaches the target
pressure of 50 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.
20.7 Allow the system to stabilize for at least
10 seconds.
20.8 Verify that the primary transducer agrees
with the reading on the NIBP simulator.
Reading should be within
3mmHg.
oo
20.9 Verify that the safety transducer agrees
with the reading on the NIBP simulator.
9650-001356-01 Rev. DZOLL X Series Service Manual37
Reading should be within
4mmHg.
oo
CHAPTER 1MAINTENANCE TESTS
Do This...Observe This...Pass/Fail
20.10 From the NIBP Test control panel, select
Open Valve to release the pressure.
20.11 From the X Series NIBP Test control
panel, use the navigation keys to select:
• Close Valve
• Turn Pump On
20.12 Monitor the pressure reported by the
external pressure standard.
20.13 Once the pressure reaches the target
pressure of 150 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.
20.14 Allow the system to stabilize for at least
10 seconds.
20.15 Verify that the primary transducer agrees
with the reading on the NIBP simulator.
20.16 Verify that the safety transducer agrees
with the reading on the NIBP simulator.
20.17 From the NIBP Test control panel, select
Open Valve to release the pressure.
20.18 From the X Series NIBP T est control
panel, use the navigation keys to select:
• Close Valve
• Turn Pump On
20.19 Monitor the pressure reported by the
external pressure standard.
20.20 Once the pressure reaches the target
pressure of 250 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.
20.21 Allow the system to stabilize for at least
10 seconds.
20.22 Verify that the primary transducer agrees
with the reading on the NIBP simulator.
Reading should be within
3mmHg.
Reading should be within
4mmHg.
Reading should be within
3mmHg.
oo
oo
oo
20.23 Verify that the safety transducer agrees
with the reading on the NIBP simulator.
20.24 From the NIBP Test control panel, select
Open Valve to release the pressure.
389650-001356-01 Rev. D
Reading should be within
4mmHg.
oo
Maintenance Tests
Do This...Observe This...Pass/Fail
2 0. 25 Once complete, use the navigation keys to
select Update NIBP Cal Check Date to
save the date of the most recent
calibration check.
Note: This will reset the date. Nothing
displays on the screen.
20.26
Press twice and select
Diagnostic>NIBP>Calibration Info.
Verify that the last NIBP Cal date is
today’s date.
20.27 Exit out of the Supervisor menu.
9650-001356-01 Rev. DZOLL X Series Service Manual39
NIBP Test control panel or the external
pressure standard, and select Turn Pump Off as soon as the pressure reaches the
target pressure of 265 +/- 5 mmHg.
22.3 Allow the system to stabilize for 5
seconds.
22.4 Use the navig ation keys to select Open Valve and immediately start the timer.
Maintenance Tests
22.5 Stop the timer as soon as pressure drops
below 15 mmHg.
The pressure should drop within 10
seconds.
oo
9650-001356-01 Rev. DZOLL X Series Service Manual41
23.1 Verify that the pressure reported by the
NIBP Test control panel or the external
pressure standard system is less than 5
mmHg.
23.2 From the NIBP Test panel, use the
navigation keys to select Close Valve.
23.3 Select Turn Pump On and immediately
start the timer.
2 3.4 Stop the timer as soon as pressure exceeds
210 mmHg.
23.5 Select Turn Pump Off, then Open Valve.
23.6 Exit th e Service Menu and turn off the
unit.
Verify the time is less than 7
seconds.
oo
429650-001356-01 Rev. D
24.0 NIBP Test
The primary purpose of an NIBP simulator is to reproduce a pressure profile similar to a live
patient to be used for testing repeatability and functionality of the system. There are many
different NIBP simulators, and manufacturers use different methods to develop their algorithm,
which can cause varying readings from different simulators. In order to test for repeatability,
you should first establish the offset of your simulator. The offset is a value that is produced on
the NIBP monitor that is shifted from the simulator’s setting. This value must be established
based on the a statistical sample of monitors and readings.
Note:Refer to Appendix B for information on simulators and settings that you may use to
assess the performance of the NIBP module.
EquipmentDual lumen NIBP hose, NIBP simulator
Do This...Observe This...Pass/Fail
24.1 On the X Series unit, ensure the patient
mode is set to Adult.
24.2 Set the simulator to the 80/50 simulator
preset.
Maintenance Tests
24.3 Take a manual reading.Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
24.4 S et the sim ulator to the 120/80 simulator
preset.
24.5 Take a manual reading.Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
2 4. 6 Set the simulator to the 200/150 simulator
preset.
24.7 Take a manual reading.Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
24.8 Disconnect hose.
oo
oo
oo
9650-001356-01 Rev. DZOLL X Series Service Manual43
CHAPTER 1MAINTENANCE TESTS
25.0 Temperature Test
EquipmentIBP Temperature Simulator (or equivalent)
Do this...Observe This...Pass/Fail
25.1 Connect the temperature simulator to the
X Series T1 channel.
25.2 On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.
25 . 3 Repeat steps 1 and 2 of this procedure for the
T2 channel.
25.4 Disconnect cable.
The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.
oo
449650-001356-01 Rev. D
26.0 Wi-Fi Test (Optional)
EquipmentDual-Band Router (2.4GHz and 5.0GHz)
SetupSet the router network name for each of the frequency bands.
• SSID= (your network name, i.e. WATS24, WATS50)
• Authentication= (your authentication mode, i.e. WPA-PSK, WPA2-PSK, Enterprise)
• Security Key= (enter your security key, i.e. 123456789)
Do This...Observe This...Pass/Fail
Maintenance Tests
26. 1 Turn on the X Series unit.Verify that the amber wireless icon
appears at the top of the display
screen within one minute.
26.2 O n the X Series unit , do the following to
test at 2.4GHz:
• Press the More () then the Setup
() quick access keys.
• Use the navigation keys to select
Supervisor>Communications.
• Use the navigation keys to enable
Wi-Fi, then select Configure.
• Highlight Add, then in the Add AP
Profile window, select DHCP setting,
and enter the desired settings for
Profile Name (i.e. Test24), SSID (i.e.
WATS24), Authentication, and
Security Key .
• Press the Display/Home () key to
exit the menu.
26.3 On the main display screen, do the
following:
oo
oo
• Use the Display/Home () key and
then the navigation keys to highlight
the wireless icon.
• Use the navigation keys to highlight
Wi-Fi Access Point and select the
profile name you created in step 2.
26.4
Press to exit the menu.
9650-001356-01 Rev. DZOLL X Series Service Manual45
Verify that the Access Point name is
in the Selected Access Point field
Verify that the wireless icon turns
green.
oo
oo
CHAPTER 1MAINTENANCE TESTS
Do This...Observe This...Pass/Fail
26.5 O n the X Series unit , do the following to
test at 5.0GHz:
• Press the More () then the Setup
() quick access keys.
• Use the navigation keys to select
Supervisor>Communications.
• Use the navigation keys to enable
Wi-Fi, then select Configure.
• Highlight Add, then in the Add AP
Profile window, select DHCP setting,
and enter the desired settings for
Profile Name (i.e. Test50), SSID (i.e.
WATS50), Authentication, and
Security Key .
• Press the Display/Home () key to
exit the menu.
26.6 On the main display screen, do the
following:
• Use the Display/Home () key and
then the navigation keys to highlight
the wireless icon.
• Use the navigation keys to highlight
Wi-Fi Access Point and select the
profile name you created in step 2.
oo
Verify that the Access Point name is
in the Selected Access Point field
oo
26.7
Press to exit the menu.
26.8 On the main display screen, do the
following:
• Press the More () then the Setup
() quick access keys.
• Use the navigation keys to select
Supervisor>Communications.
• Use the navigation keys to enable
Wi-Fi, then select Configure.
• Highlight the Access Point Names
from step 2 and step 5, then select
Delete.
26.9 Turn off the unit.
Verify that the wireless icon turns
green.
Verify that the wireless icon changes
to amber.
oo
oo
469650-001356-01 Rev. D
27.0 Audio Recording Test (Optional)
EquipmentUSB2.0 Thumb Driive (1-4GB)
PC w/ Windows media player and speakers
Do This...Observe This...Pass/Fail
27.1 Power unit on , press the LOG softkey,
then press the TRASH icon softkey.
Highlight YES, and press enter to clear
the log.
Maintenance Tests
27.2 Turn the unit Off, and then back On.V erify “Self-T est Passed” message
is displayed.
27.3 Wait 15 seconds after the “Self-Test
Passed” message disappears, and speak
loud and clear into the unit’s microphone:
“Test, one two three”
27.4 Press the LOG softkey to access the log,
and then press the USB icon softkey to
access USB
27.5 Insert USB thumb drive onto USB port .Verify the USB symbol to display
at the top of a screen.
27.6 Highlight TRANSFER ALL CASES and
press Enter.
27.7 Remove USB thumb drive from the unit,
then highlight OK and press Enter, then
press HOME softkey to exit the Log
Transfer Menu.
27.8 Insert USB thumb drive onto PC, and
then navigate to USB thumb drive via
Windows Explorer.
Wa it for fi le transfer to USB
thumb drive to complete.
oo
oo
oo
27.9 Right click on the recently .OGG file and
select Play (Windows media player will
play this audio file).
27.10 Close the Windows media player.
27. 11 Using Windows Explorer, highlight all
files at USB thumb drive. Right click on
the highlight and select Delete
Click YES to confirm all files on USB
thumb drive to be deleted.
27.12 Remove USB thumb drive.
9650-001356-01 Rev. DZOLL X Series Service Manual47
Verify if the words that you spoke
(Tes t, One Two Three), are clearly
heard in the playback.
Verify file has been removed.
oo
oo
(This page intentionally left blank.)
1–48www.zoll.com9650-001356-01 Rev. D
Overview
This chapter contains a list of error messages that users may see if the unit is not operating
properly.
If the errors you encounter are not listed below, call ZOLL Medical Corporation’s Technical
Service Department for further assistance. (See “Technical Service” on page 3 for contact
information.)
X Series Error Messages
The following is a list of the X Series error messages that may appear on your display. The
“Technical Action” column describes what you as a technician can do to correct the situation.
Chapter 2
T roubleshooting
9650-001356-01 Rev. DX Series Service Manual49
CHAPTER 2TROUBLESHOOTING
Error MessageExplanationTechnical Action
ALARM SILENCE SHORTAlarm Silence/Reset key is
stuck.
BLUETOOTH PHONE
UNAVAILABLE
The X Series unit was
unable to establish a Wide
Area Network connection.
None of the Bluetooth
phones paired with the
device were detected.
CELLULAR CONNECT
FAILURE
The X Series unit was
unable to establish a Wide
Area Network connection. A
cellular modem or Bluetooth
cellular phone connection
was attempted, but failed.
Replace the front panel.
To correct the problem,
verify the following:
• The Bluetooth device is
paired with the X Series
unit.
• The phone is on and
near enough to the
device to permit
communication.
• The phone is configured
to allow data connection
requests (if the phone
has such a configuration
setting).
To correct the problem,
verify the following:
• The cellular provider
information, particularly
the call number, is
correctly configured.
• The cellular device has
an active plan.
• There is adequate
cellular network signal
coverage in the area.
• (For phones) The phone
is configured to allow
data connection
requests (if the phone
has such a configuration
setting).
• (For modems) The
cellular modem is a
supported device and is
securely connected to
the X Series unit.
If all the configurations are
correct, retry the operation.
509650-001356-01 Rev. D
X Series Error Messages
Error MessageExplanationTechnical Action
CHECK COMM
CONFIGURATION
CO2 BAROMETRIC OUT
OF RANGE
A configuration error
prevented successful
communication.
Altitude is too high for
module.
To correct the problem,
verify the following:
• The reports
configuration contains
valid values for the
Server, Port, Customer
ID, and Password fields.
• (Cellular modem) You
are using a valid cell
provider configuration
(including call number)
and the provider name is
specified as the USB
cell modem provider.
• (Bluetooth DUN phone)
A valid cell provider
configuration (including
call number) is present
and the provider name is
associated with the DUN
phone pairing.
Bring device into specified
altitude range. If problem
persists, replace the CO
2
module.
CO2 CAL FAILURECalibration failed.Retry calibration or
continue with old
calibration. Replace the
CO
module.
2
CO2 CAL REMINDERCalibration reminder.Calibrate the CO
module
2
soon.
CO2 CAL REQUIREDCalibration due.Calibrate the CO
module.
2
CO2 CHECK FLOWExhaust port blocked.Check for exhaust port
occlusion. If problem
persists, replace the CO
module.
CO2 COM ERROR CODECommunication failure.Check communication
cables. If problem persists,
replace the CO
CO2 DEVICE FAILUREMalfunction or device failure.Replace the CO
module.
2
module.
2
2
9650-001356-01 Rev. DX Series Service Manual51
CHAPTER 2TROUBLESHOOTING
Error MessageExplanationTechnical Action
CO2 GAS INPUT
OCCLUSION
CO2 inlet blocked.
Replace FilterLine
problem persists, check for
CO
inlet occlusion. If
2
®
. If
problem still persists,
module.
2
module soon.
2
module.
2
CO2 MAINTENANCE
REMINDER
CO2 MAINTENANCE
replace the CO
Periodic maintenance
Replace CO
reminder.
Periodic Maintenance due.Replace CO
REQUIRED
CO2 NO FILTERLINENo filterline is detected.If FilterLine is connected
and problem persists,
replace FilterLine. If
problem still persists,
DEFIB CAP LEAKAGECapacitor bank cannot hold
a charge.
replace the CO
Replace the capacitor bank
module.
module.
2
DEFIB CHARGE ERROROver voltage was detected.If problem persists, replace
the defibrillator board
module.
DEFIB CHARGE TIMEOUT
FAIL
The battery did not produce
enough current to charge
capacitor.
DEFIB DEVICE FAILUREOne of the following issues
occurred:
• Device could not charge
up the self-test energy.
• Discharge circuit failed to
discharge voltage within
specified time duration.
• Device failed to deliver
correct self-test energy.
DEFIB DISCHARGE TEMPDevice is in a hot
environment or the
discharge circuit is always
on.
DEFIB FATAL CHARGE
Charging error.Replace the defibrillator
ERROR
Charge the battery. If
voltages are okay (from
battery and aux), replace
the defibrillator board
module.
Check communication
cables. If problem persists,
replace defibrillator board
module.
Move device to cooler area.
If problem persists, replace
the defib board module.
board module.
529650-001356-01 Rev. D
X Series Error Messages
Error MessageExplanationTechnical Action
DEFIB FIRE FAILUREHigh impedance or low
impedance (short).
DEFIB LOW BATTERYBattery is too low to charge
defibrillator.
DEFIB NO ENERGY
DELIVERED
DEFIB PAC ER DEVICE
FAILURE
High impedance. No defib
energy was delivered.
One of the following issues
occurred:
• Error was detected on the
defib/pacer board.
• Interface self-test
communication failed.
• Pacer interface self-test
failed.
• Communication with
pacer hardware failed
after startup.
Replace pads. If problem
persists, check internal
connections. If problem still
persists, replace the
defibrillator board module.
Charge the battery. Check
battery contacts and clean
if necessary.
Replace pads. If problem
persists, check internal
connections. If problem still
persists, replace the
defibrillator board module.
Check communication
cables. If problem persists,
replace the defibrillator
board module.
DEFIB PHASE FAILUREPhase I or phase II
waveform dropout.
DEFIB USING DEFAULT
Calibration data lost.Replace the defibrillator
CALIBRATION
ECG DEVICE FAILUREECG software detected a
problem with the ECG data
that is most likely a hardware
issue.
FP DEFIB CHARGE
Keys are stuck.Replace the front panel.
SHORT
FP DEFIB ENERGY DOWN
SHORT
FP DEFIB ENERGY UP
SHORT
Defib energy Select Down
(Decrease) key is stuck.
Defib energy Select Up
(Increase) key is stuck.
FP DEFIB FIRE SHORTA key or a front panel button
is stuck.
Replace the defibrillator
board module.
board module.
Replace the main board.
Replace the front panel.
Replace the front panel.
Replace the front panel.
9650-001356-01 Rev. DX Series Service Manual53
CHAPTER 2TROUBLESHOOTING
Error MessageExplanationTechnical Action
HW SHUTDOWNVoltage threshold exceeded
the minimum working
threshold of the device.
There is no warning.
IBP DEVICE FAILUREIBP software detected a
problem with the IBP data
that is most likely a hardware
issue.
IBP PROBE1/PROBE2/
PROBE3 SHORTED
An A/D value was detected
that indicated the probe was
shorted.
IBP PROBE1/PROBE2/
PROBE3 CIRCUIT
FAILURE
An unsupported probe was
inserted or there is a circuit
problem.
NAMESERVICE FAILUREThe reports server hostname
could not be resolved via
DNS.
Charge the battery.
Replace the main board.
Replace the problematic
IBP probe. If the problem
persists, replace the right
side panel.
Unplug the problematic
probe, power cycle. If
problem persists, replace
main board.
To correct the problem,
verify the following:
• The server DNS name is
entered correctly and
completely in the reports
configuration.
• The DNS server(s) are
correctly configured (if
using a Wi-Fi profile
where static IP
information was
configured).
• The DHCP server on the
Wi-Fi network provides
valid and correct DNS
information in its
response (if using a
Wi-Fi profile with
DHCP).
If using any form of cellular
connectivity, retry the
operation.
NAV PAD DOWN SHORTDown/Counterclockwise
Replace the front panel.
Navigation key is stuck.
NAV PAD SELECT SHORTSelect key is stuck.Replace the front panel.
NAV PAD UP SHORTUp/Clockwise Navigation
Replace the front panel.
key is stuck.
549650-001356-01 Rev. D
X Series Error Messages
Error MessageExplanationTechnical Action
NIBP AIR LEAKMajor air leak or could not
reach target pressure.
Attempt to identify
problems with the hose
and/or cuff by doing the
following:
• Check the hose
connection to the
device. Make sure that
the hose is tightly
screwed onto the NIBP
connector.
• Ensure that the cuff is
appropriately wrapped
and is not too loose.
• Inspect the cuff and
hose for leaks. Inspect
the O-ring on the NIBP
hose connector.
Consider replacing the
cuff and/or hose.
If the problem persists,
perform the following
procedure: “NIBP Leak
Test” on page 40:
Inspect internal NIBP
tubing for leaks.
NIBP ARTIFA CTUnable to determine blood
pressure. Check
connections; restrict patient
movement.
NIBP COMMUNICATION
FAILURE
There is a communication
problem with the NIBP
module.
NIBP EQUIPMENT CHECKDevice is unable to obtain a
reading.
NIBP KINKED HOSEThe cuff inflation was too
rapid. There may be kinked
tubing or another air
obstruction.
Attempt to identify and
eliminate the source of
artifact. If problem persists,
replace the NIBP module.
Ensure that the J32
connection is secure. If
error persists, replace the
NIBP module.
Check the hose and cuff
connections. If problem
persists, perform one or
more of the following
procedures in Chapter 1:
• NIBP Leak Test
• NIBP Deflation Test
• NIBP Inflation Test
Check the external NIBP
hose for kinks. If problem
persists, check the routing
of the NIBP tubing inside
the device.
9650-001356-01 Rev. DX Series Service Manual55
CHAPTER 2TROUBLESHOOTING
Error MessageExplanationTechnical Action
NIBP LOW VOLTAGEInsufficient voltage is
supplied to the NIBP
module.
NIBP MAX PRESSURE
Pressure limit exceeded.If problem persists, do the
EXCEEDED
NIBP MONITORING
FAILURE
One of the following issues
occurred:
• A failure was reported by
the NIBP module.
• The two transducers are
not aligned. They are
either not calibrated,
there is a transducer
failure, or there is a
kinked tube.
• A calibration problem was
reported by the NIBP
module.
Check the module voltage
in the NIBP Test screen.
Inspect the unit or water
ingress. If evidence of
water ingress is found,
allow the water to
evaporate while periodically
checking the module
voltage. If problem persists,
replace the NIBP module.
following:
• Check the calibration via
the “NIBP Calibration
Check” on page 37.
• Check the routing of the
NIBP tubing inside the
device for kinks or
obstructions.
If the NIBP module is out of
calibration, replace the
NIBP module.
If problem persists, replace
the NIBP module.
NIBP NEONATAL
DETECTED
Device thinks a neonate cuff
is connected. If no neonate
cuff is connected, tubing
may be damaged or kinked.
Verify that neonate cuff is
not connected. If problem
persists, check the routing
of the NIBP tubing inside
the device.
NIBP TEMP OUT OF
RANGE
NIBP WEAK PULSEToo few pulses are detected
Module temperature is out of
its specified operating range.
during the reading attempt.
Move the device to a cooler
or warmer environment.
Tighten the cu ff and ensure
appropriate placement.
569650-001356-01 Rev. D
X Series Error Messages
Error MessageExplanationTechnical Action
PACER CURRENT FAULTLack of pacer current.Replace the defibrillator
board module.
PACER DEVICE FAILUREGeneration of pacer pulse
failed.
PRINTER OFFLINEPrinter module is off-line.If it remains off line, try
PRINTER OUT OF PAPERIf there is paper in the
device, probably a paper
sensor problem.
PRINTER PLATEN OPENIf printer door is closed,
probably a problem with the
micro switch.
PS DEFIB CHARGE
SHORT
PS DEFIB ENERGY DOWNPaddle set keys are stuck.Replace the paddle set or
PS DEFIB ENERGY DOWN
SHORT
One of the paddle set keys
are stuck.
Paddle set energy down
select button is stuck.
Replace the defibrillator
board module.
checking the connectors or
power to the printer. If
problem persists, replace
the printer module.
Replace the printer module.
Replace the printer module.
Replace the paddle set or
MFC.
MFC.
Replace the paddles.
PS DEFIB ENERGY UPPaddle set keys are stuck.Replace the paddle set or
MFC.
PS DEFIB FIRE SHORTPaddle set keys are stuck.Replace the paddle set or
MFC.
PS DEFIB RECORDER
SHORT
QUICK ACCESS SW 1(to
SW 7) SHORT
SCREEN CYCLE SHORTDisplay/home key is stuck.Replace the front panel.
One of the paddle set keys
are stuck.
One of the quick access
keys (key 1 to key 7) is
stuck.
Replace the paddle set or
MFC.
Replace the front panel.
9650-001356-01 Rev. DX Series Service Manual57
CHAPTER 2TROUBLESHOOTING
Error MessageExplanationTechnical Action
SERVER
AUTHENTICATION FAILED
The reports server rejected
the conncetion request from
the X Series device.
SERVER UNREACHABLEAlthough a WAN connection
was established, the reports
server connection could not
be established or the
connection was lost during a
transfer.
SERVICE ERROR
<CODE>
While communicating with
the remote server , there was
an error.
To correct the problem,
verify the following:
• The customer ID and
password are correctly
entered in the reports
configuration. (The
password is case
sensitive.)
• The reports
configuration has the
correct server hostname
and port.
• The communications
processor (CP) software
is up to date.
Retry the operation.
Retry the operation.
SNAPSHOT SHORTSnapshot key is stuck.Replace the front panel.
SPO2 COM ERROR CODECommunication error.Check communication
cables. If problem persists,
SPO2 DEVICE FAILURECommunication failure or
device malfunction.
replace the SpO
Check the SpO
communication cables. If
module.
2
2
problem persists, replace
SPO2 SENSOR FAILUREDefective sensor or sensor
the SpO
Replace the SpO
module.
2
sensor.
2
error.
SPO2 UNKNOWN
Recovery error .Replace the SpO
sensor.
2
RECOVERY NUM
SW SHUTDOWN
WARNING
Indication that the battery
has reached the software
Charge the battery.
shutdown capacity.
TEMP1 (or TEMP2)
CIRCUIT FAILURE
The software detected an
out-of-range A/D converter.
Remove probe then power
cycle. If problem persists,
replace the main board.
589650-001356-01 Rev. D
X Series Error Messages
Error MessageExplanationTechnical Action
TEMP DEVICE FAILURETemp software detected a
problem with the
temperature data that is
most likely a hardware issue.
TEMP PROBE1 (or
PROBE2) SHORTED
An A/D value was detected
that indicated the probe was
shorted.
TRANSMISSION FAILEDThe communication
operation failed. This
message displays until a
new operation is started.
Replace the main board.
Replace the temperature
probe or adapter cable. If
problem persists, replace
the right side panel.
To correct the problem,
verify the following:
• Wireless communication
is enabled on the X
Series unit.
• The Wi-Fi settings are
correct in the
Communications setup
menu.
• The ZOLL server is
configured correctly.
• Your cellular device is
configured correctly.
• The X Series unit is
within range of the
wireless server.
WAN CONNECTION
UNAVAILABLE
The X Series unit was
unable to establish a WAN
connection.
To correct the problem,
verify the following:
• Wireless communication
is enabled on the X
Series unit.
• The expected means of
coummunication is
enabled (for instance, if
using a cell modem,
make sure the USB cell
modem is set to
Enabled).
• (Bluetooth) It is paired
with the X Series unit.
• (Wi-Fi) The correct Wi-Fi
access point profile is
selected.
• (Cellular modem) It is
properly connected to
the X Series unit and
that the USB icon on the
display is illuminated.
9650-001356-01 Rev. DX Series Service Manual59
CHAPTER 2TROUBLESHOOTING
609650-001356-01 Rev. D
Overview
This chapter provides instructions on how to disassemble and reassemble the X Series unit, and
includes the following sections:
• Required Equipment
• Safety Precautions
• Modules
• Handle Disassembly
• Printer Disassembly
• Main Unit Disassembly
• Front Enclosure Disassembly
• Display Disassembly
• NIBP and EtCO
• Monitor Board and SpO
• CP Carrier Disassembly
• Defibrillator Board and Capacitor Bank Disassembly
• AUX Power Connector Disassembly
• Dock Connector Disassembly
Chapter 3
Disassembly Procedures
Disassembly
2
Disassembly
2
Required Equipment
• No. 1 Phillips screwdriver with torque capability
• No. 2 Phillips screwdriver with torque capability
• Wooden stick (Available from H.A. Stiles: 1-800-447-8537)
• Needle nose pliers
• 6 mm socket wrench
• 26 mm socket wrench
9650-001356-01 Rev. DX Series Service Manual61
CHAPTER 3DISASSEMBLY PROCEDURES
• Kapton tape
• 3M copper adhesive tape, or equivalent
• Large diagonal cutters
Safety Precautions
WARNING!SHOCK HAZARD!
CautionTAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY
BEFORE YOU CONDUCT DEFIBRILLATOR TESTS OR REPAIRS.
• Only properly trained technicians should service the unit.
• The unit can contain deadly voltages even if the unit is turned off.
• Make sure to discharge the unit before working with it.
• Make sure you take the necessary precautions when working with static sensitive units. For
example, you must wear a conductive wrist strap (which touches your skin) connected to a
grounding mat and to the earth ground. Y ou must remove the wrist strap when you discharge
high voltage or when you are working on energized equipment.
Modules
The X Series unit contains the following ten (10) modules or assemblies:
• Printer
• Front Enclosure
• Display
• NIBP
• EtCO
• Monitor Board
• Capacitor Bank
• Defibrillator Board
• SpO
• CP Carrier Module
(Oridion®)
2
2
These modules can be purchased as replacement parts from the ZOLL Technical Support
Department. See Chapter 4, "Replacement Parts" for information on ordering replacement
parts.
629650-001356-01 Rev. D
Removing the Handle
Tools Required
• #1 Phillips screwdriver with torque capability
• Wooden stick
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
To reinstall the handle assembly, reverse the steps.
Handle Disassembly
1Using a wooden stick, lift the
screw covers off the handle in
two locations.
Removing the Handle
2Remove four (4) screws (2 on
each handle) and remove the
handle.
Note: During reassembly,
torque screws to 8 in-lbs.
9650-001356-01 Rev. DX Series Service Manual63
CHAPTER 3DISASSEMBLY PROCEDURES
Removing the Printer Module
Tools Required
• #1 Phillips screwdriver with torque capability
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
To reinstall the printer module, reverse the steps.
Printer Disassembly
1Place the unit on the mat in a
horizontal position.
2Open the printer housing and
remove the printer paper.
3Place the unit screen down
and remove the three (3)
screws outside the printer
module in the illustrated
areas.
Note: During reassembly,
torque screws to 6 in-lbs.
649650-001356-01 Rev. D
4Slowly slide the printer from
the unit, being careful not to
strain the cables.
5Disconnect the two 2 cable
connectors from the printer
module, and pull cables out
of slot.
Removing the Printer Module
9650-001356-01 Rev. DX Series Service Manual65
CHAPTER 3DISASSEMBLY PROCEDURES
Disassembling the Main Unit
Tool Required
• # 1 Phillips screwdriver with torque capability
• Wooden stick
• Kapton tape
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
• Complete these procedures in the following order before performing this procedure:
1.“Removing the Handle” on page 63
2.“Removing the Printer Module” on page 64
To reassemble the main unit, reverse the steps.
Main Unit Disassembly
1Place tape on both sides of
the unit. (This will temporary
hold the sections of the unit
together).
2Gently place unit screen so
the display is facing down.
669650-001356-01 Rev. D
3Remove ten (10) screws: four
(4) on the top, four (4) on the
bottom, and two (2) in the
middle in the illustrated
areas. Take care to keep the
two parts of the unit together.
Note: During reassembly,
torque screws to 6 in-lbs.
4While holding the two parts of
the unit together, place the
unit vertically on the mat (feet
down).
Disassembling the Main Unit
5Separate the top part of the
unit by creating a small gap,
and remove any screws that
are in the way.
9650-001356-01 Rev. DX Series Service Manual67
CHAPTER 3DISASSEMBLY PROCEDURES
6Disconnect the battery
communication cable (yellow,
orange, brown, red)
connector from the monitor
board.
7Disconnect the monitor/
defibrillator flex cable and
MFC cable connectors from
the monitor board.
Note: Be careful to pull the
connector straight out.
Rocking the connector or
disconnecting it at an angle
may result in bent pins.
8Disconnect the three power
cable connectors (battery
(black), AUX (white), and
defibrillator (red)).
689650-001356-01 Rev. D
9Press the tabs on either side
of the two printer cables to
disconnect them.
10Use a wooden stick to
release the tabs on either
side of brown flex connector
between the dock connector
and the monitor board.
Disassembling the Main Unit
Note: Release one side at a
time.
11Separate the two parts of the
unit.
9650-001356-01 Rev. DX Series Service Manual69
CHAPTER 3DISASSEMBLY PROCEDURES
Removing the Front Enclosure Module
Tools Required
• #2 Phillips screwdriver with torque capability
• Wooden stick
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
• Complete these procedures in the following order before performing this procedure:
1.“Removing the Handle” on page 63
2.“Removing the Printer Module” on page 64
3.“Disassembling the Main Unit” on page 66
To reinstall the front enclosure module, reverse the steps.
Front Enclosure Disassembly
1Remove tape from both sides
of the unit.
2Separate front enclosure
assembly.
3Pivot top up with cables still
connected.
709650-001356-01 Rev. D
4Remove grounding screw as
shown in the illustration.
Note: During reassembly,
torque screw to 6 in-lbs.
5Disconnect the front panel
multi-wire power/speaker
cable in the area illustrated.
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
• Complete these procedures in the following order before performing this procedure:
1.“Removing the Handle” on page 63
2.“Removing the Printer Module” on page 64
3.“Disassembling the Main Unit” on page 66
4.“Removing the Front Enclosure Module” on page 70
To reinstall the display module, reverse the steps.
Display Disassembly
1Remove the grounding screw
and washer for EtCO
area illustrated.
in the
2
729650-001356-01 Rev. D
2Using the 6mm socket
wrench, remove the hex
standoff located underneath
the screw you just removed.
Note: During reassembly,
torque to
6 in-lbs.
3Place the back of the unit
down on the mat so that the
display screen is facing up.
4Remove the six (6) screws
from the front of the casting in
the areas illustrated.
Note: During reassembly,
torque screws to 6 in-lbs.
Removing the Display Module
9650-001356-01 Rev. DX Series Service Manual73
CHAPTER 3DISASSEMBLY PROCEDURES
5Lift the casting up while
pushing down on the rubber
boot that runs around the
edge of the LCD screen.
6Remove the copper tape from
the casting.
Note: During reassembly,
apply new copper tape for
grounding.
If casting is being replaced,
scuff the underside of the
casting where the tape is
being applied, with a
fiberglass brush.
Wipe area with lint free wipe
and alcohol.
749650-001356-01 Rev. D
7Disconnect the LCD display
cable connectors from the
back of the LCD screen to
release it.
Note: During reassembly,
make sure that when
taping the LED cable
assembly to the spider
frame, that the cable
is clear from the
display frame as
shown. (The use
mid-frame assembly
fixture 9100-3028-AF
is optional.)
Removing the Display Module
9650-001356-01 Rev. DX Series Service Manual75
CHAPTER 3DISASSEMBLY PROCEDURES
Removing the EtCO2 and NIBP Modules
Tool Required
• # 1 Phillips screwdriver with torque capability
• Wooden stick
• Large diagonal cutters
• Metal Pick
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
• Complete these procedures in the following order before performing this procedure:
1.“Removing the Handle” on page 63
2.“Removing the Printer Module” on page 64
3.“Disassembling the Main Unit” on page 66
4.“Removing the Front Enclosure Module” on page 70
To assemble the EtCO2 and/or NIBP modules, reverse the steps.
EtCO2 and/or NIBP Disassembly
1Unplug the power connector
by press on the tab and
puling straight out.
769650-001356-01 Rev. D
2Remove four (4) screws
around the EtCO2 unit.
Note: During reassembly,
torque screws to 6 in-lbs.
3Disconnect the EtCO2 sensor
from side panel by using a
metal pick and releasing the
clips on both sides.
Removing the EtCO2 and NIBP Modules
4Disconnect the exhaust tube
connector from the air vent at
area shown.
9650-001356-01 Rev. DX Series Service Manual77
CHAPTER 3DISASSEMBLY PROCEDURES
5Remove EtCO2 module by
lifting assembly out. If
necessary disconnect NIBP
hoses to make it easier to pull
connector through.
6Cut the tie wrap holding the
ferrite bead.
7Remove three (3) screws
from NIBP unit.
Note: During reassembly,
torque screws to 6 in-lbs.
789650-001356-01 Rev. D
8Disconnect the 5 NIBP tubes.
9Disconnect the 2 red
connectors, then remove the
4 screws (top 2 screws have
washers) securing the NIBP
MOD G board. Lift NIBP
MOD G board and pump out.
Note: During reassembly,
align connector on bottom of
NIBP board to monitor board.
Torque screws to 6 in-lbs.
Removing the EtCO2 and NIBP Modules
Note: During reassembly, put
the hose on left side of fan
through the middle slot, and
the hose on the right through
in the bottom slot.
9650-001356-01 Rev. DX Series Service Manual79
CHAPTER 3DISASSEMBLY PROCEDURES
Removing the Monitor Board and SpO2 Modules
Tool Required
• # 1 Phillips screwdriver with torque capability
• Wooden stick
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
• Complete these procedures in the following order before performing this procedure:
1.“Removing the Handle” on page 63
2.“Removing the Printer Module” on page 64
3.“Disassembling the Main Unit” on page 66
4.“Removing the Display Module” on page 72
5.“Removing the EtCO2 and NIBP Modules” on page 76
To reinstall the monitor board and/or SpO2 modules, reverse the steps.
Monitor Board and/or SpO2 Disassembly
1Remove 4 screws on the right
side of the unit in the
illustrated areas.
Note: During reassembly,
torque screws to 6 in-lbs.
809650-001356-01 Rev. D
2Disconnect the SpO2 ground
cable.
3Remove 4 screws on the left
side in the illustrated areas.
Note: During reassembly,
torque screws to 6 in-lbs.
Removing the Monitor Board and SpO2 Modules
4Turn the unit over.
5Remove 3 screws on the left
side of the unit in the
illustrated areas.
6Turn the unit over.
9650-001356-01 Rev. DX Series Service Manual81
CHAPTER 3DISASSEMBLY PROCEDURES
7Disconnect ECG flex cable
ZIF connector from the left
side.
8Disconnect the SpO2
connector.
9Remove the left side of the
unit.
829650-001356-01 Rev. D
10Disconnect the IBP
connector.
11Disconnect the USB ZIF
connector.
Removing the Monitor Board and SpO2 Modules
12Turn the unit over.
13Remove the 3 screws on the
right side of the unit in the
illustrated areas.
9650-001356-01 Rev. DX Series Service Manual83
CHAPTER 3DISASSEMBLY PROCEDURES
14Turn the unit over.
15Remove the right side panel
of the unit.
16Disconnect the LCD cable.
17Remove 7 screws securing
monitor board, then lift board
out.
849650-001356-01 Rev. D
Removing the CP Carrier Module
Tools Required
• #1 Phillips screwdriver with torque capability
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
• Complete these procedures in the following order before performing this procedure:
1.“Removing the Handle” on page 63
2.“Removing the Printer Module” on page 64
3.“Disassembling the Main Unit” on page 66
4.“Removing the Display Module” on page 72
5.“Removing the EtCO2 and NIBP Modules” on page 76
6.“Removing the Monitor Board and SpO2 Modules” on page 80
Removing the CP Carrier Module
CP Module Disassembly
1Remove the shield by pulling
it upwards over the CP carrier
board.
Note: During reassembly, be
sure to route antenna wires
through slot (red circle) in the
shield. Secure the shield to
the board by aligning with
slots on the board.
9650-001356-01 Rev. DX Series Service Manual85
CHAPTER 3DISASSEMBLY PROCEDURES
2Remove the 9-pin GPS and 2
antenna cables (gray cable
J7 and black cable J6) from
the CP carrier board.
3Turn board over, and remove
the 4 screws securing the CP
board to the monitor board.
Note: During reassembly,
torque screws to 6 in-lbs.
4Only lift the CP carrier board
(not the metal heat spreader)
upwards at location J60 on
the monitor board.
Note: During reassembly,
align connectors on board.
869650-001356-01 Rev. D
18Disconnect the top flex cable
connectors from SPO2
module, then remove cable
and ferrite.
19Remove the screw from the
flex cable, then disconnect
flex cable from the
connectors in two locations.
Note: During reassembly,
torque screws to 4 in-lbs.
Removing the CP Carrier Module
20Remove 4 screws from the
module and remove it.
SpO
2
Note: During reassembly,
torque screws to 4 in-lbs.
9650-001356-01 Rev. DX Series Service Manual87
CHAPTER 3DISASSEMBLY PROCEDURES
Removing the Defibrillator Boards and Capacitor Bank
Modules
Tools Required
• #1 Phillips screwdriver with torque capability
• #2 Phillips screwdriver with torque capability
• 26 mm open end wrench
• Large diagonal cutters
• 5k Ohm, 25 watt, resistor
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
• Complete these procedures in the following order before performing this procedure:
1.“Removing the Handle” on page 63
2.“Removing the Printer Module” on page 64
3.“Disassembling the Main Unit” on page 66
To reinstall the defibrillator boards and/or capacitor bank modules, reverse
the steps.
Defibrillator Boards and Capacitor Bank Disassembly
1Lay back end of separated
unit down with opened side
facing up.
2Pull the black snap rivet
straight out.
889650-001356-01 Rev. D
Removing the Defibrillator Boards and Capacitor Bank Modules
3Disconnect the defib-system
flex cable from the processor
board by pulling straight up
from board.
4Remove screw in the middle
of the plastic isolator (located
on top of the processor
board) in the area illustrated.
Note: During reassembly,
torque screw to 6 in-lbs.
5Remove the plastic isolator
shield.
9650-001356-01 Rev. DX Series Service Manual89
CHAPTER 3DISASSEMBLY PROCEDURES
WARNING! This unit may contain lethal voltages. You MUST completely discharge the high voltage capacitor
before removing from unit. DO NOT SHORT THE TRERMINAL ENDS OF THE CAPACITOR.
6Discharge the capacitor by
bleeding out the excess
voltage using a resistor with
values of approximately 5k
Ohms, (25 watts for 10-20
seconds). Counting from the
right connect, apply the
resistor across the second
and forth via as shown in
picture. Measure the voltage
of the capacitor on the same
two vias to verify that the
capacitor is discharged.
7Remove the four (4) screws
from the processor board in
the areas illustrated.
Note: During reassembly,
torque screws to 6 in-lbs.
909650-001356-01 Rev. D
Removing the Defibrillator Boards and Capacitor Bank Modules
8Lift up the processor board
and disconnect the brown flex
cable (there are two
connections).
9Lift out the processor board.
10Remove the black capacitor
cable by pressing the tab to
release.
9650-001356-01 Rev. DX Series Service Manual91
CHAPTER 3DISASSEMBLY PROCEDURES
11Cut the tie wrap securing the
Power Monitor to Pace cable
(black and red coupled wire)
and disconnect the cable (via
the red connector) in
illustrated area.
12Remove four (4) screws from
the frame support in the
areas illustrated.
Note: During reassembly,
torque screws to 6 in-lbs.
13Lift up on the ECG assembly
and remove red and white
MFE cable (via the white
connector) from the ECG
board. Remove ECG
assembly.
929650-001356-01 Rev. D
Removing the Defibrillator Boards and Capacitor Bank Modules
14Remove the 5 screws
securing the ECG board to
bracket, and then separate
the board from the bracket.
Note: During reassembly,
torque screws to 6 in-lbs.
15Remove the Bridge/Pace to
ECG cable from the ECG
board in the two areas
illustrated.
Note: During reassembly,
match the designator on the
flex cable and the ECG
board.
16Remove two (2) screws
securing the capacitor to
housing, then lift capacitor
out.
Note: During reassembly,
torque screws to 6 in-lbs.
9650-001356-01 Rev. DX Series Service Manual93
CHAPTER 3DISASSEMBLY PROCEDURES
Removing the AUX Power Connector
Tools Required
• Socket, 9100-3060-AF
• Large diagonal cutters
Setup
• This procedure should be conducted on a static mat.
• Wear a wrist strap to ground during the procedure.
• Complete these procedures in the following order before performing this procedure:
1.“Removing the Handle” on page 63
2.“Removing the Printer Module” on page 64
3.“Disassembling the Main Unit” on page 66
4.“Removing the Defibrillator Board s and Cap aci to r Ban k Mo du les” on page 88
AUX Power Connector
1Disconnect AUX Power
Connector cable from
harness.
2Using socket, 9100-3060-AF,
remove lock nut that is
securing the connector to the
housing.
Note: During reassembly,
apply a drop of loctite before
secure the lock nut.
949650-001356-01 Rev. D
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.