Zoll AED+ Service manual

Service Manual
9650-0308-01 Rev. A
March 2005
An issue date and revision level for this manual appear on the front cover.
If more than three years have elapsed since this date, contact ZOLL Medical Corporation to determine if additional product information updates are available.
ZOLL and stat•padz are registered trademarks, and AED Plus, AED Pro, CPR-D•padz, pedi•padz, pro•padz, PowerCharger, Preconnect, RescueNet, Smart Alarms, and v•pak are trademarks of ZOLL Medical Corporation. All other trademarks and registered trademarks are property of their respective owners.
Copyright © 2005 ZOLL Medical Corporation. All rights reserved.

Table of Contents

Preface
Related Documentation .........................................................................................................iv
Conventions...........................................................................................................................iv
Symbols ..................................................................................................................................v
Safety Considerations............................................................................................................vi
Warranty ................................................................................................................................vi
Technical Service.................................................................................................................. vii
Returning a Unit for Service ..........................................................................................vii
Chapter 1 Maintenance Tests
Preparations for Testing...................................................................................................... 1-2
Test Equipment ................................................................................................................... 1-2
1.0 Physical Inspection.................................................................................................. 1-3
2.0 Preventive Maintenance Test .................................................................................. 1-4
3.0 Manual Self-Test ..................................................................................................... 1-9
Chapter 2 Troubleshooting
Troubleshooting the AED Plus Unit .................................................................................... 2-2
AED Plus Error Log Messages ........................................................................................... 2-4
Appendix A Maintenance Tests Checklist
Instructions .........................................................................................................................A-2
ZOLL AED Plus Maintenance Tests Checklist....................................................................A-3
Index
9650-0308-01 Rev. A ZOLL AED Plus Service Manual i
TABLE OF CONTENTS
ii www.zoll.com 9650-0308-01 Rev. A

Preface

ZOLL Medical Corporation provides this manual for technicians who service the AED Plus™ device or diagnose malfunctions. This manual contains the following sections:
Preface — Provides safety warnings and an overview of the manual’s contents. Review this section thoroughly before servicing an AED Plus unit.
Chapter 1, "Maintenance Tests" — Provides the recommended procedures for evaluating the condition and performance of the AED Plus unit.
Chapter 2, "Troubleshooting" — Describes possible problems and solutions and lists the error log messages that indicate the unit requires technical service.
Appendix A, "Maintenance Tests Checklist" — Provides a checklist for recording the results of maintenance tests.
Note: The AED Plus contains no user-replaceable parts.
9650-0308-01 Rev. A ZOLL AED Plus Service Manual iii
PREFACE

Related Documentation

In addition to this service manual, the following manuals are available for the AED Plus unit:
AED Plus Operator’s Guide (9650-0300-01) — Describes what the rescuer sees, hears, and
does while operating the unit.
AED Plus Administrator’s Guide (9650-0301-01) — Describes the features and functions of
the unit and the procedures for configuring and maintaining it.
AED Plus Simulator Operator’s Guide (9651-0800-01) — Describes the AED Plus
simulator, which you can use to test the AED Plus unit’s response to various simulated cardiac rhythms.

Conventions

This manual uses uppercase italics for audible prompts and for text messages displayed on the screen (for example DON’T TOUCH PATIENT, ANALYZING).
!
Within test or troubleshooting procedures, this symbol indicates an audible prompt or message; for example:
! UNIT OK
WARNING! Warning statements alert you to conditions or actions that can result in personal injury
or death.
Caution Caution statements alert you to conditions or actions that can result in damage to the unit.
iv www.zoll.com 9650-0308-01 Rev. A

Symbols

Symbols
Symbols used in this manual or on the equipment include the following:
Class II equipment
Defibrillation-protected Type BF patient connection
ATTE NTION: Refer to the operator’s guide for more information
WARNING: Dangerous voltage
New battery cells
Not new battery cells
VARTA
Push button
Do not push button
Recommended manufacturer
Do not use this manufacturer
Unit equipped to treat adult and pediatric patients
Conformité Européenne Complies with medical device directive 93/42/EEC.
9650-0308-01 Rev. A ZOLL AED Plus Service Manual v
PREFACE

Safety Considerations

Only qualified personnel should service or disassemble an AED Plus unit. Before using this device, read the AED Plus Administrator’s Guide. Before servicing or disassembling any equipment, review these safety considerations and read this manual carefully.
WARNING! The AED Plus unit can generate as much as 2250 volts with sufficient current to cause
lethal shocks.
Before discharging the defibrillator, warn everyone near the equipment to “STAND CLEAR.”
Use only Type 123A lithium manganese dioxide batteries from manufacturers recommended by ZOLL. Discard used batteries properly.
Do not recharge, disassemble, or dispose of batteries in fire. Batteries might explode if mistreated.
This device is protected against interference from radio frequency emissions typical of the two-way radios and cellular telephones used in emergency service and public safety activities. You should assess the device’s performance in your typical operating environment to determine the likelihood of radio frequency interference (RFI) from high-power sources.

Warranty

For warranty information, refer to the AED Plus Administrator’s Guide.
To maintain the warranty, strictly follow the instructions and procedures in this service manual. Unit failure that is attributable to the use of accessories that are not manufactured by ZOLL might void the ZOLL warranty.
vi www.zoll.com 9650-0308-01 Rev. A

Technical Service

If the AED Plus unit requires service, contact the ZOLL Technical Service Department or your local distributor.
Telephone: 1-800-348-9011 (U.S.A. customers only)
1-978-421-9655
Fax: 1-978-421-0010
When requesting service, please provide the following information to the service representative:
Unit serial number
Description of the problem
Department where the equipment is used and the name of the person to contact
Purchase order to allow tracking of loan equipment
Purchase order for a unit with an expired warranty

Returning a Unit for Service

Tec hn ic al Se rv ic e
Before sending a unit to the ZOLL Technical Service Department for repair, obtain a service request (SR) number from the service representative.
Remove all batteries from the unit. Pack the unit in its original container or equivalent packaging. Be sure the assigned service request number appears on the package.
For customers Return the unit to
In the U.S.A. ZOLL Medical Corporation
269 Mill Road Chelmsford, MA 01824-4105
Attention: Technical Service Dept. [ SR number ]
In Canada ZOLL Medical Canada
Unit #15 5266 General Road Mississauga, Ontario L4W 1Z7
Attention: Technical Service Dept. [ SR number ]
Telephone: 1-866-442-1011
In other locations The nearest authorized ZOLL Medical Corporation representative.
To locate an authorized service center, contact the International Sales Department at
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105
Telephone: 1-978-421-9655
9650-0308-01 Rev. A ZOLL AED Plus Service Manual vii
PREFACE
viii www.zoll.com 9650-0308-01 Rev. A
Chapter 1
Maintenance Tests
This chapter describes the procedures for routine testing of the AED Plus unit.
A test typically consists of multiple steps, each of which requires a pass/fail rating. For each step, perform the specified action and then verify that the expected results occur. If all expected results occur, assign a “Pass” rating to the step; otherwise, assign it a “Fail” rating.
This chapter includes the following inspection and test procedures:
“1.0 Physical Inspection” on page 1-3
“2.0 Preventive Maintenance Test” on page 1-4
“3.0 Manual Self-Test” on page 1-9
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 1–1
CHAPTER 1MAINTENANCE TESTS

Preparations for Testing

Gather the required equipment listed in the next section.
Perform all the steps within each test procedure in the order specified.
If testing results indicate that the batteries must be replaced, you will need a set of 10 fully charged Type 123A batteries.

Test Equipment

For some tests, you need one or more of these items:
ZOLL universal adapter cable
Fluke
The following items are optional:
ZOLL AED Plus Simulator —
ZOLL Administration Software (ZAS) —
For ZOLL part numbers, refer to the accessories list in the AED Plus Administrator’s Guide.
®
Impulse 4000 Defibrillator/Pacer Performance Analyzer (Fluke Biomedical/
DNI Nevada) with 1.06 software or higher
When connected to an AED Plus unit, the simulator can send it various simulated cardiac rhythms (ventricular fibrillation, normal sinus rhythm, or asystole) and optional simulated chest compression readings (rate and depth), so that you can test the unit’s response.
When installed on a personal computer or personal digital assistant that has an IrDA® (wireless infrared) interface, the ZAS application can communicate with an AED Plus unit in nonrescue mode to retrieve its usage history, hardware and software configuration, and error log. You can also use ZAS to view or modify the unit’s configurable settings and to retrieve event data.
1–2 www.zoll.com 9650-0308-01 Rev. A

1.0 Physical Inspection

Use this procedure to ensure that the unit shows no signs of damage or excessive wear.
Tools Needed
None.
Test Setup
None.
Test Procedure
Observe the following: Pass Fail
Physical Inspection
1.1 Is the unit clean?
1.2 Are all signs of wear reasonable and not excessive?
1.3 Are the top panel and housing undamaged and free of cracks?
1.4 Are input connectors clean and undamaged?
1.5 If electrodes are preconnected, is the cable free from damage, are the pads
sealed within the packaging, and are the electrodes within the expiration date shown on the packaging?
!!
!!
!!
!!
!!
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 1–3
CHAPTER 1MAINTENANCE TESTS

2.0 Preventive Maintenance Test

Use this test to verify that the AED Plus unit is working properly and ready for use.
For an alternative test using the AED Plus Simulator, refer to the AED Plus Administrator’s Guide.
Note: To verify that the delivered energy for the first, second, and third shocks corresponds
to the unit’s preconfigured adult energy levels, use the ZOLL Administration Software (ZAS) to obtain the current configuration of the unit.
Tools Needed
Fluke Impulse 4000 Defibrillator/Pacer Performance Analyzer or equivalent
Universal adapter cable
Test Setup
Turn off the AED Plus unit, remove the cover, and unplug the preconnected electrodes.
Keep the electrodes so they can be reconnected later.
Connect the universal adapter cable to the AED Plus unit and to the analyzer.
Turn on the analyzer and set its function to VF (ventricular fibrillation).
Variations in Messages
When verifying messages in the test results, please be aware of the following issues:
The wording of some messages is configurable. The following table shows equivalent
messages.
CHECK RESPONSIVENESS CHECK PATIENT
CHECK CIRCULATION CHECK PULSE
IF NO CIRCULATION START CPR IF NO PULSE START CPR
IF NO CIRCULATION CONTINUE CPR IF NO PULSE CONTINUE CPR
TREATMENT ADVISED SHOCK ADVISED
NO TREATMENT ADVISED NO SHOCK ADVISED
PRESS TREATMENT BUTTON PRESS SHOCK BUTTON
RELEASE TREATMENT BUTTON RELEASE SHOCK BUTTON
TREATMENT DELIVERED SHOCK DELIVERED
NO TREATMENT DELIVERED NO SHOCK DELIVERED
The message that reports the detected electrode type can be either ADULT PADS or
PEDIATRIC PADS.
Some messages are issued only if the unit is configured to do so. If an optional message is
not issued during a test, do not consider this a failure.
In some cases a text message on the display screen is abbreviated or otherwise differs
slightly from the associated voice message.
1–4 www.zoll.com 9650-0308-01 Rev. A
Preventive Maintenance Test
Test Procedure
Test Action Expected Result(s) Pass Fail
2.1 Turn on the AED Plus unit.
2.2 Press the Tre atment button
(first shock).
! UNIT OK
! ADULT PADS
! DON’T TOUCH PATIENT,
ANALYZING
! TREATMENT ADVISED
(or SHOCK ADVISED)
! DON’T TOUCH PATIENT.
! PRESS TREATMENT BUTTON
(or PRESS SHOCK BUTTON)
Analyzer indicates a defibrillator discharge at the configured first-shock energy level (120, 150 or 200 joules ±15%).
! TREATMENT DELIVERED
(or SHOCK DELIVERED)
An updated shock count appears on the display (SHOCKS: 1).
!!
!!
! DON’T TOUCH PATIENT,
ANALYZING
! TREATMENT ADVISED
(or SHOCK ADVISED)
! DON’T TOUCH PATIENT.
! PRESS TREATMENT BUTTON
(or PRESS SHOCK BUTTON)
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 1–5
CHAPTER 1MAINTENANCE TESTS
Test Action Expected Result(s) Pass Fail
2.3 Press the Tre atment button
(second shock).
2.4 Press the Tre atment button (third shock).
Analyzer indicates a defibrillator discharge at the configured second-shock energy level (120, 150 or 200 joules ±15%).
! TREATMENT DELIVERED
(or SHOCK DELIVERED)
An updated shock count appears on the display (SHOCKS: 2).
! DON’T TOUCH PATIENT,
ANALYZING
! TREATMENT ADVISED
(or SHOCK ADVISED)
! DON’T TOUCH PATIENT.
! PRESS TREATMENT BUTTON
(or PRESS SHOCK BUTTON)
Analyzer indicates a defibrillator discharge at the configured third-shock energy level (120, 150 or 200 joules ±15%).
!!
!!
! TREATMENT DELIVERED
(or SHOCK DELIVERED)
An updated shock count appears on the display (SHOCKS: 3).
These three prompts are optional:
OPEN AIRWAYCHECK BREATHINGGIVE TWO BREATHS
! CHECK CIRCULATION
(or CHECK PULSE)
! IF NO CIRCULATION START CPR
(or IF NO PULSE START CPR)
1–6 www.zoll.com 9650-0308-01 Rev. A
Preventive Maintenance Test
Test Action Expected Result(s) Pass Fail
Turn off the AED Plus unit.
Set the analyzer to NSR (normal sinus rhythm).
2.5 Turn on the AED Plus unit.
2.6 Lift the puck that is attached to the
universal adapter cable. Move the puck up and down very slightly (no more than 1 inch or 2 cm) and continue to do so.
! UNIT OK
! ADULT PADS
! DON’T TOUCH PATIENT,
ANALYZING.
! NO TREATMENT ADVISED
(or NO SHOCK ADVISED)
These three prompts are optional:
OPEN AIRWAYCHECK BREATHINGGIVE TWO BREATHS
! CHECK CIRCULATION
(or CHECK PULSE)
! IF NO CIRCULATION START CPR
(or IF NO PULSE START CPR)
Unit emits spaced tones as a guide for performing chest compressions at the proper rate.
! PUSH HARDER
!!
!!
If the unit has a display, prevent the bar on the chest compression gauge from extending into the area between the lower two lines.
2.7 Increase the vertical motion of the puck (about 2 to 3 inches or 5 to 8 cm).
If the unit has a display, make sure the bar on the chest compression gauge extends into the area between the lower two lines.
Unit emits spaced tones as a guide for performing chest compressions at the proper rate.
! GOOD COMPRESSIONS
The following prompt is optional:
IF NO CIRCULATION CONTINUE
CPR (or IF NO PULSE CONTINUE CPR)
!!
! STOP CPR
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 1–7
CHAPTER 1MAINTENANCE TESTS
Test Action Expected Result(s) Pass Fail
Put the puck down and turn off the AED Plus unit.
Disconnect the universal adapter cable from the AED Plus unit.
2.8 Wait 10 seconds and then turn on the AED Plus unit.
2.9 Check the expiration date on the previously connected defibrillation electrodes.
If the electrodes have not expired, reconnect them to the unit; otherwise, connect a fresh package of defibrillation electrodes.
! PLUG IN CABLE
This prompt is optional:
STAY CALM
! CHECK RESPONSIVENESS
(or CHECK PATIENT)
This prompt is optional:
CALL FOR HELP
These three prompts are optional:
OPEN AIRWAYCHECK BREATHINGGIVE TWO BREATHS
! CHECK CIRCULATION
(or CHECK PULSE)
! PLUG IN CABLE
! ADULT PADS
(or PEDIATRIC PADS)
!!
!!
Turn off the AED Plus unit.
2.10 After 90 seconds, check the Status indicator on the unit.
The Status indicator shows a green check.
!!
1–8 www.zoll.com 9650-0308-01 Rev. A

3.0 Manual Self-Test

Use this test to initiate the unit’s self-diagnostic procedure, which verifies the following:
Battery usage since last reset.
Defibrillation electrodes are properly connected.
ECG signal acquisition and processing circuitry is functional.
Defibrillator charge and discharge circuitry (operational at 2 joules) is functional.
Microprocessor hardware and software are functional.
CPR monitoring circuitry and compression depth sensor are functional (when CPR-D•padz
are attached).
Audio output circuitry is functional.
Display screen (if present) is functional.
Tools Needed
None.
Test Setup
Turn off the AED Plus unit.
If they are not already in place, preconnect defibrillation electrodes to the unit.
Manual Self-Test
Test Procedure
Test Action Expected Result(s) Pass Fail
3.1 Press and hold the ON/OFF
button for at least 5 seconds.
Within 10 seconds after the unit is turned on, the Status indicator displays a green check.
On the top panel, the indicator lights for the rescue actions illuminate in sequence.
!!
! UNIT OK
If an IrDA connection is established, turn off the unit; otherwise, the unit shuts down automatically.
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 1–9
CHAPTER 1MAINTENANCE TESTS
1–10 www.zoll.com 9650-0308-01 Rev. A
Chapter 2
Troubleshooting
This chapter describes common technical issues encountered during routine maintenance or after a malfunction of the AED Plus unit. This chapter also lists critical AED Plus error log messages that indicate the unit requires technical service.
This chapter contains the following sections:
“Troubleshooting the AED Plus Unit” on page 2-2
“AED Plus Error Log Messages” on page 2-4
If this chapter does not help you resolve the problem, call the ZOLL Technical Service Department for assistance. For contact information, refer to page vii.
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 2–1
CHAPTER 2TROUBLESHOOTING

Troubleshooting the AED Plus Unit

The following tables include common issues and their solutions. First, try the recommendations given under “Operator Action.” If these steps do not remedy the problem, follow the suggestions under “Technical Action.”
For information on contacting ZOLL Technical Service, refer to page vii.
Table 2-1. AED Plus Troubleshooting
Problem Operator Action Technical Action
Status indicator shows a red “X”.
! UNIT FAILED
Unit beeps while turned off.
! CHANGE BATTERIES
! PLUG IN CABLE
! ANALYSIS HALTED
KEEP PATIENT STILL
! NO SHOCK DELIVERED
(or NO TREATMENT DELIVERED)
— Press and hold the ON/OFF button for at least 5 seconds to perform a self-test.
— Verify that the electrode cable is properly plugged into the patient cable connector.
— Replace the electrodes.
—Turn the unit off and then press and hold the ON/OFF button for at least 5 seconds to perform a self-test.
— Verify that the electrode cable is properly plugged into the patient cable connector.
— Replace the electrodes.
— Verify that the electrode cable is properly plugged into the patient cable connector.
—Turn the unit off and then press and hold the ON/OFF button for at least 5 seconds to perform a self-test. Verify that a green check appears in the Status indicator.
— Replace all batteries with new ones. When prompted, press the battery reset button.
— Ensure that the electrode cable is properly plugged into the patient cable connector.
— Keep the patient still during ECG analysis. If transporting the patient by stretcher or vehicle, stop all patient movement during analysis.
The Treatment button was not pressed or an internal error occurred.
— When prompted to press the button, do so within 30 seconds.
If the unit continues to fail, remove it from service and contact Technical Service.
If the unit continues to fail, remove it from service and contact Technical Service.
If the unit continues to beep, remove it from service and contact Technical Service.
If the message persists, remove the unit from service and contact Technical Service.
If the message persists, remove the unit from service and contact Technical Service.
If the message persists, remove the from service and contact Technical Service.
If an internal error occurred (no shock was delivered when theTreatment button was pressed properly), remove the unit from service and contact Technical Service.
2–2 www.zoll.com 9650-0308-01 Rev. A
Troubleshooting the AED Plus Unit
Table 2-1. AED Plus Troubleshooting (continued)
Problem Operator Action Technical Action
! RELEASE SHOCK
BUTTON (or RELEASE TREATMENT BUTTON)
— Release the Treatment button. Do not press the button until the charge-ready tone sounds and the button begins flashing.
If the message persists, remove the unit from service and contact Technical Service.
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 2–3
CHAPTER 2TROUBLESHOOTING

AED Plus Error Log Messages

The AED Plus unit records internal errors in an error log, which you can retrieve by using a Windows-based personal computer with ZOLL Administration Software (ZAS).
To retrieve the error log, follow these steps:
Step Action
1 Turn off the AED Plus unit.
2 Start the ZOLL Administration Software (ZAS) on the computer.
3 Position the AED Plus unit so that its IrDA port has a clear line-of-sight with the computer’s
IrDA port and the ports are about 10 to 18 inches (25 to 45 cm) apart.
4 Press and hold the unit’s ON/OFF button for at least 5 seconds to start the unit in
nonrescue mode.
5 After the unit issues the message COMMUNICATIONS ESTABLISHED, refer to the ZAS
help for instructions on saving the unit’s history file and viewing the included error log.
Note: After saving the error log file on the computer, you can use ZAS to clear the error log
on the AED Plus unit.
Table 2-2 lists the critical error log messages.
Important: If any of these messages appears in the error log, contact ZOLL Technical Service
as described on page vii.
Table 2-2. AED Plus Error Log Messages
Error
No.
Error Message
1 The patient relay is not working properly
2 Attempting to charge the capacitor while the SAFELINE in the CPLD is set - should not
charge
3 The capacitor takes too long to charge
4 The H-bridge test failed while the bridge was disabled. There should have been 0 current
measured.
5 VCAP1 and VCAP2 are not within a defined range of each other
6 The H-bridge test failed while the bridge was enabled - an invalid current reading was
taken during the positive test on the H-bridge
7 The H-bridge test failed while the bridge was enabled - an invalid current reading was
taken during the negative test on the H-bridge
8 Defib did not charge during the power-on charge test
9 The Defib task is not receiving A2D vCap1 event samples - needed for charging
10 FnSafety has determined that the charge has dropped below a specified value during the
holding of charge.
11 The capacitor failed to charge
12 The capacitor over-charged
13 The capacitor contained an unsafe voltage in an idle state
2–4 www.zoll.com 9650-0308-01 Rev. A
AED Plus Error Log Messages
Table 2-2. AED Plus Error Log Messages (continued)
Error
No.
Note: Error numbers 256 and greater indicate noncritical conditions.
Error Message
14 Patient impedance experienced a short
15 The A2D is getting non-accurate readings during its test
17 The message queue overflowed
18 PowerOn task has timed out while waiting for a self test result to come back.
19 Calibration faults that occur during a clinical event.
20 An unexpected error has occurred when calling a Nucleus operation.
21 An unexpected error has occurred when calling a Rhapsody operation.
22 An unexpected coding error has occurred.
23 No Language installed - corrupt language.
24 The shock button is pressed/stuck.
25 The ECG Calibration data is invalid
26 The Patient Impedance Calibration data is invalid
27 A critical flash error has occurred
28 The WatchDog is not enabled during clinical event
29 The clock test failed
9650-0308-01 Rev. A ZOLL AED Plus Service Manual 2–5
CHAPTER 2TROUBLESHOOTING
2–6 www.zoll.com 9650-0308-01 Rev. A
Appendix A
Maintenance Tests Checklist
This appendix provides a sample maintenance checklist for the AED Plus unit. To ensure that the equipment is functioning properly and ready for emergency use, complete the checklist periodically according to local protocols.
Your organization can create custom checklists for your specific requirements.
9650-0308-01 Rev. A ZOLL AED Plus Service Manual A–1
APPENDIX AMAINTENANCE TESTS CHECKLIST

Instructions

To maintain records for maintenance testing, photocopy the checklist shown on the following pages. The checklist includes an entry for each test described in this manual. Use these copies to record the results of your testing and then retain the copies for your records.
A–2 www.zoll.com 9650-0308-01 Rev. A
ZOLL AED Plus Maintenance Tests Checklist
Use this checklist to record the results of AED Plus maintenance tests and keep it for your records. For a description of each numbered test, refer to Chapter 1, "Maintenance Tests" in the AED Plus Service Manual.
Serial No.
Location
Date
Te st er
Signature
Result of Check:
" No action required " Minor problems
corrected
" Disposable supplies
replaced
" Major problems identified
(unit out of service)
Additional Remarks
1.0 Physical Inspection
PASS FAIL
1.1
""
1.2
""
1.3
""
1.4
""
1.5
""
2.0 Preventive Maintenance Test
PASS FAIL
2.1
""
2.2
""
2.3
""
2.4
""
2.5
""
2.6
""
2.7
""
2.8
""
2.9
""
2.10
""
3.0 Manual Self-Test
PASS FAIL
3.1
""
9650-0308-01 Rev. A ZOLL AED Plus Service Manual A–3
A–4 www.zoll.com 9650-0308-01 Rev. A

Index

A
Administration software, ZOLL 1-2, 2-4 Analyzer 1-2, 1-4
C
Checklists for maintenance tests A-1 Configurable prompts 1-4 Conventions used in manual iv
D
Defibrillator analyzer 1-2, 1-4 Documentation iv
E
Error log 2-4
H
History file, device 2-4
I
Impulse 4000 analyzer 1-2, 1-4 Inspection, physical 1-3 IrDA port 1-2
L
Log, error 2-4
M
Maintenance tests 1-1
checklists A-1 equipment needed 1-2 prerequisites 1-2
preventive 1-4 Manual self-test 1-9 Manuals, related iv Messages
error log
wording 1-4
P
2-4
Physical inspection 1-3 Port, IrDA 1-2 Preventive maintenance test 1-4 Prompts, configurable 1-4
9650-0308-01 Rev. A ZOLL AED Plus Service Manual Index–1
INDEX
R
Radio frequency interference vi Repairs vii Returns vii RFI vi
S
Safety considerations vi Self-test, manual 1-9 Service, technical vii Simulator iv, 1-2 Software, ZOLL Administration 1-2, 2-4
T
Technical Service vii
Tests, maintenance 1-1
checklist A-1
prerequisites 1-2 Troubleshooting 2-1
common problems 2-2
V
Voltage warning vi
W
Warranty vi Wording, configurable 1-4
Z
ZOLL Administration Software (ZAS) 1-2, 2-4
Index–2 www.zoll.com 9650-0308-01 Rev. A
Loading...