WarrantyWe guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result
in differences between this operating manual and your instrument.
Also, we cannot guarantee that there are absolutely no errors in this
manual. Therefore, we are sure you will understand that we cannot
accept any legal claims resulting from the data, figures or descriptions.
purchase.
The instrument warranty covers manufacturing faults that are
discovered within the warranty period.
The warranty does not cover components that are replaced during
maintenance work, e. g. batteries.
The warranty claim extends to restoring the instrument to readiness for
use but not, however, to any further claim for damages. Improper
handling or unauthorized opening of the instrument invalidates any
warranty claim.
To ascertain the warranty liability, return the instrument and proof of
purchase together with the date of purchase freight paid or prepaid.
The Lab 960 compact precision conductivity meter enables you to
perform conductivity measurements quickly and reliably. The Lab 960
provides the maximum degree of operating comfort, reliability and
measuring certainty for all applications.
The proven procedures for determining or adjusting the cell constant
support your work with the conductivity meter.
1Keypad
2Display
3Socket field
Note
The meter is also available as part of an individual Set of equipment.
You will find information on this and other accessories in the SI
Analytics GmbH laboratory catalog or via the Internet.
ba75557e0511/2009
7
OverviewLab 960
STO
RCL
MODESC
CAL
MODE
CAL
SC
STOSTO
RCL
1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<OK
_>).
<On/Off>:
<On/Off
<MODE>:
<MODE
<CAL>:
<CAL
<SC>:
<SC
<STO>:
<STO
_>:
_>:
_>:
_>:
_>:
Switch the meter on/off
Resetting the determined cell constant
Select measured parameter
Open setting menu for calibration and
measurement
Call up calibration procedure
Call up calibration data
Activate / deactivate stability control
Set the interval for data transmission
Store measured value
Set the interval for automatic storing
processes
8
<RCL>:
<RCL
_>:
<▲>: Increment values, scroll
Display / transmit measured values
and calibration records
Erase stored measured values
ba75557e0511/2009
Lab 960Overview
Measured variable
Measured value
display
Temperature
display, function
display indicators,
status display
indicators
<▼>: Decrement values, scroll
<OK>:
<OK
1.2Display
_>:
Confirm entries
Open setting menu for system settings
Temperature display,
function display
indicators, status
display indicators
ba75557e0511/2009
[AR]Stability control is active
[ARng]Automatic range switch-over: meter measures with
highest possible resolution
[Auto][Store]Automatic storing is switched on
[Cal]Calibration
[Lin]Linear temperature compensation
[LoBat]With battery operation: Batteries almost empty
[nLF]Nonlinear temperature compensation
[RCL]Memory display / memory download
[TP]Temperature measurement active
[Tref20]Reference temperature 20 °C
[Tref25]Reference temperature of 25 °C
9
OverviewLab 960
1.3Socket field
Connectors:
1Conductivity measuring cell
2Temperature sensor
3
4USB interface
5Power pack
6RS 232 interface
CAUTION
Only connect sensors to the meter that cannot return any voltages
or currents that are not allowed (> SELV and > current circuit with
current limiting).
Almost all measuring cells - in particular SI Analytics GmbH
measuring cells - fulfill these conditions.
10
ba75557e0511/2009
Lab 960Safety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating
manual before working with the meter.
The operating manual must always be available within the vicinity of the
instrument.
Target groupThe meter was developed for work in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.
Safety instructionsSafety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the level of danger:
Further notes
WARNING
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.
CAUTION
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the instrument
or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.
ba75557e0511/2009
11
SafetyLab 960
2.1Authorized use
Authorized use of the meter consists exclusively of the measurement
of conductivity, temperature, salinity and TDS (total dissolved solids) in
a laboratory.
The technical specifications as given in chapter 7 T
ECHNICALDATA
(page 63) must be observed. Only the operation and running of the
meter according to the instructions given in this operating manual is
authorized.
Any other use is considered unauthorized.
2.2General safety instructions
This instrument is constructed and tested in compliance with the IEC
1010 safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during
operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICALDATA (page 63).
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty
functioning of the instrument. In this event, wait until the temperature of
the instrument reaches room temperature before putting the instrument
back into operation.
CAUTION
The meter is only allowed to be opened by personnel authorized
by SI Analytics GmbH.
12
ba75557e0511/2009
Lab 960Safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
of service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the meter.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
ba75557e0511/2009
13
SafetyLab 960
14
ba75557e0511/2009
Lab 960Commissioning
3Commissioning
3.1Scope of delivery
z Lab 960 laboratory meter
z Power pack
z 4 batteries 1.5 V Mignon type AA
z Z875 USB cable with A plug on B plug
z Transparent cover
z Operating manual
z CD-ROM with USB driver
3.2Initial commissioning
Insert batteries
Perform the following activities:
z Insert batteries
z Switch on the meter
z Set the date and time
z Connect the power pack (for line power operation only).
1Open the battery compartment (1) on the underside of the
meter.
2Place four batteries (type Mignon AA) in the battery
compartment.
3Close the battery compartment (1).
The date (day) flashes in the display.
4Set the date and time according to page 43.
ba75557e0511/2009
15
CommissioningLab 960
CAUTION
Make sure that the poles of the batteries are the right way round.
The
±signs on the batteries must correspond to the ± signs in the
battery compartment.
Only use leakproof alkaline manganese batteries.
Switching on the meter
Setting the date and
time
Connecting the power
pack
1Switch on the meter with <On/Off>.
A display test is briefly displayed.
2See page 49
You can either operate the measuring instrument with batteries or with
the plug-in power supply. The plug-in power supply supplies the measuring instrument with low voltage (12 VDC). This saves the batteries.
CAUTION
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see page 63).
CAUTION
Use original power packs only (see page 63).
3Insert the plug into the socket of the conductivity meter.
4Connect the original power pack to an easily accessible power
outlet.
16
Note
You can carry out measurements without the power pack.
ba75557e0511/2009
Lab 960Commissioning
ba75557e0511/2009
17
CommissioningLab 960
18
ba75557e0511/2009
Lab 960Operation
4Operation
4.1Switching on the meter
1Place the meter on a flat surface and protect it from intense
light and heat.
2Press the <On/Off>
A display test is briefly displayed.
Subsequently, the meter switches to the measuring mode
(measured value display).
Note
The meter has an energy saving feature to avoid unnecessary battery
depletion during battery operation.
The energy saving feature switches off the meter if no key was pressed
during the specified interval (setting the switch-off interval see
page 49).
The energy saving feature is not active:
z if the meter is supplied via the power pack or the USB interface
z if the automatic storing function is active (see page 40)
z if a PC is connected (with communication cable to RS232 interface)
z if the printer cable is connected (for external printers).
key.
ba75557e0511/2009
19
OperationLab 960
4.2General operating principles
This section contains basic information of the operation of the Lab 960.
4.2.1Operating modes
The instrument has the following operating modes:
z Measuring
The display indicates the measurement data in the measured value
display
z Calibration
The display guides you thru a calibration procedure with calibration
information
z Storage in memory
The meter stores measuring data automatically or manually
z Transmitting data
The meter transmits measuring data and calibration records to a
serial interface automatically or manually.
z Configuration
The system menu or a sensor menu with submenus, settings and
functions is displayed
4.2.2Operation
KeysThe meter is operated via keys. The keys can have different functions
with long or short keystrokes.
FunctionsGenerally, with a short keystroke a function is carried out. A long
keystroke opens a setting menu.
In a setting menu, settings are selected with the <▲><▼> keys.
A setting is confirmed with <OK>. With confirming, the setting is
finished and the next setting is displayed.
RepresentationIn this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<OK
_>).
20
ba75557e0511/2009
Lab 960Operation
4.3Measuring
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect a measuring cell to the meter.
2Adjust the temperature of the test solutions and measure the
current temperature if the measurement is made without a
temperature sensor.
3Calibrate or check the meter with the measuring cell.
4Select the measured parameter with <MODE>.
CAUTION
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result.
The RS232 and USB interfaces are not galvanically isolated.
Temperature sensorThe temperature measurement is absolutely essential for a
reproducible conductivity measurement. If a temperature sensor is
connected, it is indicated on the display by TP.
If you use a conductivity measuring cell without integrated temperature
sensor, we recommend to use an external temperature sensor.
Note
The conductivity meter automatically recognizes the type of the
temperature sensor used. Therefore, you can connect measuring cells
with an NTC30 or Pt1000.
ba75557e0511/2009
21
OperationLab 960
0
5
2
0
3
5
C
°
S/cm
µ
nLF
TP
Tref25
4.3.1Measuring the conductivity
1Perform the preparatory activities according to page 21.
2Immerse the conductivity measuring cell in the test sample.
3If necessary, scroll with <MODE> until the measured
parameter
4Wait for a stable measured value.
ӕ with the unit mS/cm or µS/cm is displayed.
5Measurement with stability control (see page 25).
4.3.2Measuring the resistivity
1Perform the preparatory activities according to page 21.
2Immerse the conductivity measuring cell in the test sample.
3If necessary, scroll with <MODE> until the measured
parameter
4Wait for a stable measured value.
ӕ with the unit MOhm is displayed.
22
ba75557e0511/2009
Lab 960Operation
2
1
2
3
2
3
C
°
Sal
nLF
TP
5Measurement with stability control (see page 25).
4.3.3Measuring the salinity
1Perform the preparatory activities according to page 21.
2Immerse the conductivity measuring cell in the test sample.
3Using <MODE>, scroll as necessary until the measured
parameter Sal is displayed.
4Wait for a stable measured value.
ba75557e0511/2009
5Measurement with stability control (see page 25).
23
OperationLab 960
0
0
1
3
5
3
TDS
mg/l
Tref25
nLF
7
8
0
3
1
3
TDS
mg/l
Tref25
nLF
4.3.4Measuring the total dissolved solids (TDS)
1Perform the preparatory activities according to page 21.
2Immerse the conductivity measuring cell in the test sample.
3Using <MODE>, scroll as necessary until the measured
parameter TDS is displayed.
4Using <▲><▼>, set the TDS factor (0.40 ... 1.00).
5Wait for a stable measured value.
6Measurement with stability control (see page 25).
24
ba75557e0511/2009
Loading...
+ 53 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.