Xylem Lab 850 User Manual

Lab 850
Operating manual
6
2
9
4
8
9
H
2
°
C
TP
Lab 850
Laboratory pH meter
ba75553e03 01/2006
DRAFT:February 6, 2006
Lab 850
Accuracy when
going to press
Warranty We guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our instruments are the result of continuous development. This may result in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions.
purchase. The instrument warranty covers manufacturing faults that are discovered within the warranty period. The warranty does not cover components that are replaced during maintenance work, e. g. batteries.
The warranty claim extends to restoring the instrument to readiness for use but not, however, to any further claim for damages. Improper handling or unauthorized opening of the instrument invalidates any warranty claim.
To ascertain the warranty liability, return the instrument and proof of purchase together with the date of purchase freight paid or prepaid.
Copyright
© 2006, SCHOTT Instruments
Reprinting - even in the form of excerpts - is only allowed with the explicit written authorization of SCHOTT Instruments. Printed in Germany.
ba75553e03 01/2006
Lab 850
KONFORMITÄTSERKLÄRUNG DECLARATION OF CONFORMITY DÉCLARATION DE CONFORMITÉ
Wir erklären in alleiniger
Verantwortung, dass das
Produkt
pH-Meter
Lab 850
auf das sich diese Erklärung
bezieht, übereinstimmt mit
den Angaben im Kapitel
We declare under our sole
responsibility that the
product
pH meter
Lab 850
to which this declaration
relates is in conformity with
the specifications in the
chapter
Technische Daten
pH-Meter
30. Oktober 2005
Lab 850
Nous déclarons sous notre seule responsabilité que le
produit
pH-mètre
Lab 850
auquel se réfère cette
déclaration est conforme aux
indications du chapitre
30. Oktober, October 30, 30 octobre 2005 AGQSF 0000-A100-00/051030
ba75553e03 01/2006
SCHOTT Instruments GmbH
Hattenbergstr. 10
D-55122 Mainz
Deutschland, Germany, Allemagne
3
Lab 850
4
ba75553e03 01/2006
Lab 850 Contents
Lab 850 - Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.3 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 12
3 Commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.1 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2.1 Z865 stand (S4D). . . . . . . . . . . . . . . . . . . . . . . . 16
4 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1 Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2 General operating principles . . . . . . . . . . . . . . . . . . . . . . 18
4.2.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . 18
4.2.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.3 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.3.1 Measuring the pH value . . . . . . . . . . . . . . . . . . . 20
4.3.2 Measuring the ORP . . . . . . . . . . . . . . . . . . . . . . 21
4.4 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.4.1 Calibration interval (Int.C) . . . . . . . . . . . . . . . . . 26
4.4.2 Automatic calibration (AutoCal) . . . . . . . . . . . . . 27
4.4.3 Downloading calibration data . . . . . . . . . . . . . . . 30
4.5 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4.5.1 System settings . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.5.2 Measurement settings . . . . . . . . . . . . . . . . . . . . 34
4.6 Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.6.1 Resetting calibration values . . . . . . . . . . . . . . . . 38
4.6.2 Resetting all meter settings . . . . . . . . . . . . . . . . 39
5 Maintenance, cleaning, disposal. . . . . . . . . . . . . . . . . 41
5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.1.1 Replacing the batteries . . . . . . . . . . . . . . . . . . . 41
5.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.3 Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
ba75553e03 01/2006
5
Contents Lab 850
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.2 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 49
8 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6
ba75553e03 01/2006
Lab 850 Overview

1Overview

The compact Lab 850 precision pH meter enables you to perform pH measurements rapidly and reliably. The Lab 850 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications. The proven calibration procedures and special stability control function (SC) support your work with the pH meter.
pH
6
9
9
2
2
4
L
a
8
b
8
°
C
5
TP
0
3
2
1
1 Keypad
2 Display
3 Socket field
Note
The meter is also available as part of individual Sets of equipment. You will find information on this and other accessories in the SCHOTT Instruments laboratory catalog or via the Internet.
Note
If you need further information or application notes, you can obtain the following material from SCHOTT Instruments:
! Application reports
! pH primers
! Safety datasheets.
ba75553e03 01/2006
7
Overview Lab 850

1.1 Keypad

In this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <OK>) indicates a keystroke in this operating manual.
MODE
CAL
SC
<On/Off>: <On/Off
<MODE>: <MODE
<CAL>: <CAL
<SC>: Activate / deactivate stability control
<>: Increment values, scroll
<>: Decrement values, scroll
_>:
_>:
_>:
Switch meter on / off Reset calibration values
Select measured parameter Open setting menu for calibration and measurement
Call up calibration procedure Display calibration data
<OK>:
_>:
<OK
8
Confirm entries Open setting menu for system settings
ba75553e03 01/2006
Lab 850 Overview

1.2 Display

Status display indicator
CalClock (calibration interval and sensor evaluation)
Measured value display
Function and temperature display indicator
1
Time
Day.Month
Year
LoBat
SET
8
AutoCal
U pH
8
Baud
No.
Ident
88
Auto
BUFFER
8
8
Store
8
8
CalError
SC
% mV/pH
°F
°
C
RCL
TP
Function display
indicators
SET During calibration: number of the buffer set
BUFFER During calibration: number of the buffer
AutoCal Calibration with automatic buffer recognition
CalError An error occurred during calibration
LoBat With battery operation: batteries almost empty
SC Stability control is active
TP Temperature measurement active
ba75553e03 01/2006
9
Overview Lab 850

1.3 Socket field

Connectors:
1 pH electrode
2 Reference electrode
3 Temperature sensor
4 Power pack
5 Service interface
Caution
Only connect sensors to the meter that cannot return any voltages or currents that are not allowed (> SELV and > current circuit with current limiting). Almost all sensors - in particular SCHOTT Instruments sensors ­fulfill these conditions.
10
ba75553e03 01/2006
Lab 850 Safety

2 Safety

This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the meter. Consequently, all responsible personnel must read this operating manual before working with the meter. The operating manual must always be available within the vicinity of the instrument.
Target group The meter was developed for work in the laboratory.
Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to take when handling chemicals.
Safety instructions Safety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g. "Caution") indicates the level of danger:
Further notes
Warning
indicates instructions that must be followed precisely in order to avoid possibly great dangers to personnel.
Caution
indicates instructions that must be followed precisely in order to avoid the possibility of slight injuries or damage to the instrument or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
ba75553e03 01/2006
11
Safety Lab 850

2.1 Authorized use

This meter is authorized exclusively for pH and ORP measurements in the laboratory. The technical specifications as given in chapter 7 T
ECHNICAL DATA
(page 47) must be observed. Only the operation and running of the meter according to the instructions given in this operating manual is authorized. Any other use is considered unauthorized.

2.2 General safety instructions

This instrument is constructed and tested in compliance with the IEC 1010 safety regulations for electronic measuring instruments. It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the meter can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the meter can only be guaranteed under the environmental conditions that are specified in chapter 7 T
ECHNICAL DATA (page 47).
If the instrument was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the instrument reaches room temperature before putting the instrument back into operation.
Caution
The meter is only allowed to be opened by personnel authorized by SCHOTT Instruments.
12
ba75553e03 01/2006
Lab 850 Safety
Safe operation If safe operation is no longer possible, the instrument must be taken out
of service and secured against inadvertent operation! Safe operation is no longer possible if the meter:
! has been damaged in transport
! has been stored under adverse conditions for a lengthy period of
time
! is visibly damaged
! no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and guidelines are observed when using dangerous substances:
! EEC directives for protective labor legislation
! National protective labor legislation
! Safety regulations
! Safety datasheets of the chemical manufacturers.
ba75553e03 01/2006
13
Safety Lab 850
14
ba75553e03 01/2006
Lab 850 Commissioning

3 Commissioning

3.1 Scope of delivery

! Lab 850 laboratory meter
! Power pack
! 4 batteries 1.5 V Mignon type AA
! Transparent cover
! Stand with stand base (S4D)
! Operating manual
! CD-ROM

3.2 Initial commissioning

Switching on the meter
Setting the date and
time
Connecting the power
pack
Perform the following activities:
! Switch on the meter
! Set the date and time
! Connect the power pack (for line power operation only).
1 Switch on the meter with <On/Off>.
A display test is briefly displayed.
2 See page 32.
Caution
The line voltage at the operating site must lie within the input voltage range of the original power pack (see page 47).
ba75553e03 01/2006
Caution
Use original power packs only (see page 47).
3 Insert the plug into the socket of the pH meter.
4 Connect the original power pack to an easily accessible power
outlet.
15
Commissioning Lab 850
Note
You can carry out measurements without the power pack.

3.2.1 Z865 stand (S4D)

The meter can be inserted in the stand base on the right or left side of the electrode holder.
1 Insert the stand rod in the stand base.
2 Insert the meter in the stand base.
An arrangement of meter with stand looks, for example, as follows:
16
ba75553e03 01/2006
Lab 850 Operation

4 Operation

4.1 Switching on the meter

1 Place the meter on a flat surface and protect it from intense
light and heat.
2 Press the <On/Off>
A display test is briefly displayed. Subsequently, the meter switches to the measuring mode (measured value display).
Note
The meter has an energy saving feature to avoid unnecessary battery depletion during battery operation. The energy saving feature switches off the meter if no key was pressed during the specified interval (setting the switch-off interval, see page 32).
key.
ba75553e03 01/2006
17
Operation Lab 850

4.2 General operating principles

This section contains basic information on the operation of the Lab 850.

4.2.1 Operating modes

The instrument has the following operating modes:
! Measurement
The display indicates the measurement data in the measured value display
! Calibration
The display guides you thru a calibration procedure with calibration information
! Configuration
The system menu or a sensor menu with submenus, settings and functions is displayed

4.2.2 Operation

Keys The meter is operated via keys. The keys can have different functions
with long or short keystrokes.
Functions Generally, with a short keystroke a function is carried out. A long
keystroke opens a setting menu.
In a setting menu, settings are selected with the <▲><▼> keys. A setting is confirmed with <OK>. With confirming, the setting is finished and the next setting is displayed.
Representation In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) generally indicates a short keystroke (under 2 sec) in this operating manual. A long keystroke (approx. 2 sec) is indicated by the underscore behind the key symbol (e.g.
<OK
_>).
18
ba75553e03 01/2006
Lab 850 Operation

4.3 Measuring

Preparatory activities Perform the following preparatory activities when you want to measure:
1 Connect an electrode to the meter.
2 Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a temperature sensor.
3 Calibrate or check the meter with the electrode.
4 Select the measured parameter with <MODE>.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values. Calibrate regularly before measuring.
Caution
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.
Temperature sensor You can measure with or without a temperature sensor. If a
temperature sensor is connected, it is indicated on the display by TP.
Note
The pH meter automatically recognizes the type of the temperature sensor used. Therefore, you can connect electrodes with an NTC30 or Pt1000.
The temperature measurement is absolutely essential for a reproducible pH measurement. If the measurement is made without a temperature sensor, proceed as follows:
1 Measure the current temperature using a thermometer.
2 Set the temperature value with <▲><▼>.
Note
When calibrating without temperature sensor, also set the current temperature of each buffer solution manually (see <▲><▼>).
ba75553e03 01/2006
19
Operation Lab 850

4.3.1 Measuring the pH value

1 Perform the preparatory activities according to page 19.
2 Immerse the pH electrode in the test sample.
3 Using <MODE>, scroll as necessary until the measured
parameter pH is displayed.
pH
Stability control SC
(drift control)
9
4
2
8
TP
°
C
9
6
2
The stability control function (drift control) checks the stability of the measurement signal. The stability has a considerable effect on the reproducibility of the measured value.
For identical measurement conditions, the following criteria apply:
Drift within 15 sec < 0.02 pH units.
1 If necessary, call up the measured variable pH with <MODE>.
2With <SC>, activate the stability control function.
The SC function display indicator appears. The current measured value is frozen (hold function).
3 Start measurement with stability control with <OK>.
SC flashes until a stable measured value is reached. This measured value is downloaded to the interface.
20
4 If necessary, start the next measurement with stability control
with <OK>.
5 To terminate the stability control function: Press the <MODE>
or <SC>
Note
The current measurement with stability control can be terminated at any time (accepting the current value) by pressing <OK>.
key.
ba75553e03 01/2006
Lab 850 Operation

4.3.2 Measuring the ORP

The meter can, in conjunction with an ORP electrode, measure the ORP (mV) of a solution.
1 Perform the preparatory activities according to page 19.
2 Immerse the ORP electrode in the test sample.
3 If necessary, call up the measured parameter U with <MODE>.
4 Wait for a stable measured value.
U
4
8
2
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution.
6
1
3
mV
TP
°
C
ba75553e03 01/2006
21
Operation Lab 850

4.4 Calibration

Why calibrate? pH electrodes age. This changes the zero point (asymmetry) and slope
of the pH electrode. As a result, an inexact measured value is displayed. Calibration determines the current values of the zero point and slope of the electrode and stores them. Thus, you should calibrate at regular intervals. The calibration data is stored in the meter.
When do you have to
calibrate?
Buffer sets for
calibration
! After connecting another electrode
! When the CalClock on the display flashes:
after the calibration interval has expired
after a voltage interruption (empty batteries)
You can use the buffer sets quoted in the table for an automatic calibration. The pH values are valid for the specified temperature values. The temperature dependence of the pH values is taken into account during calibration.
No. Buffer set* pH values at
1 SCHOTT Instruments DIN buffers
according to DIN 19266/NBS
1.679
4.006
25 °C
6.865
9.180
12.454
2 SCHOTT Instruments Technical
buffers according to DIN 19267
2.000
4.010
25 °C
7.000
10.011
22
3 Merck1* 4.000
7.000
9.000
4 Merck2 * 1.000
6.000
8.000
13.000
5 Merck3 * 4.660
6.880
9.220
6 DIN 19267 * 1.090
4.650
6.790
9.230
ba75553e03 01/2006
20°C
20°C
20°C
25 °C
Lab 850 Operation
No. Buffer set* pH values at
7 Mettler Toledo USA * 1.679
4.003
7.002
10.013
8 Mettler Toledo TEC * 1.995
4.005
7.002
9.208
9 Fisher * 2.007
4.002
7.004
10.002
10 Fluka BS * 4.006
6.984
8.957
11 Radiometer * 1.678
4.005
7.000
9.180
12 Baker * 4.006
6.991
10.008
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
13 Metrohm * 3.996
25 °C
7.003
8.999
14 Beckman * 4.005
25 °C
7.005
10.013
15 Hamilton Duracal * 4.005
25 °C
7.002
10.013
16 Precisa * 3.996
25 °C
7.003
8.999
* Brand names or trade names are trademarks of their respective
owners protected by law (see page 55).
ba75553e03 01/2006
23
Operation Lab 850
Note
The buffer set is selected in the menu for measurement settings (see page 34).
Calibration points Calibration can be performed using one, two or three buffer solutions in
any order (single-point, two-point or three-point calibration). The meter determines the following values and calculates the calibration line as follows:
Determined
Displayed calibration data
values
1-point ASY ! Zero point = ASY
! Slope = Nernst slope
(-59.16 mV/pH at 25 °C)
2-point ASY
SLO
3-point ASY
SLO
! Zero point = ASY
! Slope = SLO
! Zero point = ASY
! Slope = SLO
The calibration line is calculated by linear regression.
Note
You can display the slope in the units, mV/pH or %. You can display the zero point in the units, mV or pH.
AutoCal is adapted to the permanently programmed buffer solutions as a fully
automatic two-point or three-point calibration. The buffer solutions are automatically recognized by the meter.
24
Stability control The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be terminated at any time (accepting the current value) by pressing <OK>.
Calibration record When finishing a calibration, the new calibration values are first
displayed as an informative message and stored. )
ba75553e03 01/2006
Lab 850 Operation
Displaying the
You can display the data of the last calibration (see page 30).
calibration data
Calibration evaluation After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of both is taken into account.The evaluation appears on the display as the CalClock and in the calibration record.
CalClock Zero point
[mV]
Slope [mV/pH]
-15 ... + 15 -60.5 ... -58
-20 ... + 20 -58 ... -57
-25 ... +25 -61 ... -60.5 or
-57 ... -56
-30 ... +30 -62 ... -61 or
-56 ... -50
Preparatory activities
Clean the electrode according to the sensor operating manual
CalError
< -30 or > 30
... -62 or ... -50
Eliminate the error according to chapter 6 W
HAT TO DO IF...
(page 43)
1 Switch on the meter with <On/Off>.
2 Connect a pH electrode to the meter.
3 Keep the buffer solutions ready.
4 Adjust the temperature of the solutions and measure the
current temperature if the measurement is made without temperature sensor.
ba75553e03 01/2006
5 Set the buffer set to be used for calibration as necessary.
25
Operation Lab 850

4.4.1 Calibration interval (Int.C)

The calibration interval and calibration evaluation are indicated on the display as the CalClock.
CalClock
The remaining time of the calibration interval is indicated by the segmented ring around the calibration evaluation. This segmented ring reminds you to calibrate regularly.
After the specified calibration interval (Int.C) has expired, the outer ring of the CalClock flashes. It is still possible to measure.
Note
To ensure the high measuring accuracy of the meter, calibrate after the calibration interval has expired.
Setting the calibration
interval
The calibration interval (Int.C) is set to 7 days (d7) in the factory. You can change the interval (1 ... 999 days):
1 Open the menu for measurement settings with <MODE_>.
2 Confirm all settings with <OK> until Int.C is displayed.
pH
N
1
Time
T
D
3 Set the calibration interval with <▲><▼>.
4 Confirm the setting with <OK>.
C
8
26
ba75553e03 01/2006
Lab 850 Operation

4.4.2 Automatic calibration (AutoCal)

For this procedure, use two or three buffer solutions of the selected buffer set in any order.
Note
The steps 2, 6 and 10 are not necessary if you use a temperature sensor.
1 Start the calibration with <CAL>.
The number of the specified buffer set is displayed (SET 1 ...
16). The first buffer (BUFFER 1) of the buffer set (SET_x) is requested.
2
BUFFER
1
8
5
TP
°
C
SET
pH
1
AutoCal
2 If necessary, set the temperature of the buffer solution with
<▲><>.
3 Immerse the pH electrode in the first buffer solution.
4 Start the measurement with <OK>.
The SC display indicator flashes. Depending on the setting, the display shows the nominal pH value for the recognized buffer or the electrode voltage (mV). As soon as a stable value is recognized, the next buffer (BUFFER 2) is requested.
ba75553e03 01/2006
27
Operation Lab 850
Continue with two-point
calibration
2
BUFFER
2
8
5
TP
°
C
SET
pH
1
AutoCal
Note
Here you can cancel the calibration procedure with <MODE>. This corresponds to a single-point calibration. The value of the zero point (Asy) is displayed for 10 seconds, then the value of the slope (Slo) is displayed for 10 seconds. After this the meter switches to the measuring mode.
5 Thoroughly rinse the electrode with distilled water.
6 If necessary, set the temperature of the second buffer solution
with <▲><▼>.
7 Immerse the pH electrode in the second buffer solution.
8 Start the measurement with <OK>.
The SC display indicator flashes. Depending on the setting, the display shows the nominal pH value for the recognized buffer or the electrode voltage (mV). As soon as a stable value is recognized, the next buffer (BUFFER 3) is requested.
2
Buffer
5
3
8
TP
°
C
SET
1
AutoCal
DIN
pH
28
ba75553e03 01/2006
Lab 850 Operation
Note
Here you can cancel the calibration procedure with <MODE>. This corresponds to a two-point calibration. The value of the zero point (Asy) is displayed for 10 seconds, then the value of the slope (Slo) is displayed for 10 seconds. After this the meter switches to the measuring mode.
Continue with three-
point calibration
9 Thoroughly rinse the electrode with distilled water.
10 If necessary, set the temperature of the third buffer solution
with <><>.
11 Immerse the pH electrode in the third buffer solution.
12 Press the <OK>
The SC display indicator flashes. Depending on the setting, the display shows the nominal pH value for the recognized buffer or the electrode voltage (mV). As soon as a stable value is recognized, the value of the zero point (ASY) is displayed for 10 seconds, then the value of the slope (SLO) is displayed for 10 seconds. After this the meter switches to the measuring mode.
Note
While the zero point (ASY) is being displayed, you can change the unit of the zero point with
While the slope (SLO) is being displayed, you can change the unit of the slope with The % display refers to the Nernst slope of 59.2 mV/pH at 25° C (100 x determined slope/Nernst slope).
<▲><▼> .
key.
<▲><▼> .
ba75553e03 01/2006
The unit of zero point and slope can be permanently changed in the measurement settings (see page 34).
29
Operation Lab 850

4.4.3 Downloading calibration data

1 Press <CAL_> to display the calibration data.
The following data is displayed consecutively for 10 seconds each: date, zero point, slope.
While the calibration data is being displayed, you can:
! press <OK> to display further calibration data
(date, zero point, slope)
! press
<▲><>
! press
<▲><>
to switch over the unit of the zero point (Asy) (while the zero point is being displayed)
to switch over the unit of the slope (Slo) (while the slope is being displayed)
30
ba75553e03 01/2006
Lab 850 Operation

4.5 Settings

You can adapt the meter to your individual requirements. The settings are done in the following menus:
! System settings (<OK
Switch-off interval (t.Off)
Date (Day.Month)
Date (Year)
Time (Time)
! Settings for calibration and measurement (<MODE
Number of the buffer set for pH calibration (Set 1 ... 16)
Display of the buffer during calibration
(pH nominal value or measured voltage value in mV)
Unit of the value for the slope (mV/pH or %)
– Unit of the value for the zero point (mV, pH)
– Resolution (Hi, Lo)
Temperature unit (°C / °F)
Calibration interval (Int.C [0 ... 999])
Note
You can exit the setting menu at any time by pressing <MODE>. Settings already modified and confirmed with <OK> are stored.
_>)
_>)
ba75553e03 01/2006
31
Operation Lab 850

4.5.1 System settings

The default setting is printed in bold.
Switch-off interval (t.Off) 10, 20, 30, 40, 50 min,
1, 2, 3, 4, 5, 10, 15, 20, 24 h
Date (Day.Month)Any
Date (Year) Any
Time (Time)Any
1 Open the menu for system settings with <OK_>.
The first system setting is displayed.
Switch-off interval
(t.OFF)
Date and time
f
f
0
3
O
t
Time
1
00
2 Set the switch-off interval with <▲><▼>.
3 Confirm with <OK>.
Day.Month, the setting of the date is displayed. The day display flashes.
Day.Month
0
30
32
ba75553e03 01/2006
Lab 850 Operation
4 Set the date of the current day with <▲><▼> .
5 Confirm with <OK>.
The month display flashes.
6 Set the current month with <▲><▼> .
7 Confirm with <OK>.
Year, the setting of the year is displayed.
8 Set the year with <><> .
9 Confirm with <OK>.
The setting of the time is displayed. The hour display flashes.
10 Set the current hour with <▲><▼> .
11 Confirm with <OK>.
The minute display flashes.
12 Set the current minute with <▲><▼> .
13 Confirm with <OK>.
The system settings are completed. The meter switches to the measuring mode.
ba75553e03 01/2006
33
Operation Lab 850

4.5.2 Measurement settings

These settings apply to calibration and measurement (the default setting is printed in bold).
Number of the buffer set for pH calibration (SET)
Display during calibration (BUFFER) pH (buffer nominal value),
1 ... 16
U (electrode voltage)
Buffer set for pH
calibration
(SET)
Unit of the value for the zero point (ASY)
Unit of the value for the slope (SLO) %, mV/pH
Temperature unit (UnIt) °C, °F
Calibration interval (Int.C) 0 ... 7 ... 999 d
1 Open the menu for measurement settings with <MODE_>.
Set 1 ... 16, the specified buffer set is displayed.
SET
pH, mV
1
34
AutoCal
2 Using <><>, select a buffer set Set 1 ... 16.
3 Confirm with <OK>.
bUFF, the setting of the display during calibration is displayed.
ba75553e03 01/2006
Lab 850 Operation
Display during
calibration
(bUFF)
pH
Unit of the value for the
slope (SLO)
F
H
L
F
O
mV/pH
U
b
p
AutoCal
4 Using <><>, select the display during calibration pH or U.
5 Confirm with <OK>.
SLO, the unit of the value for the slope (mV/pH or %) is displayed.
pH
S
ba75553e03 01/2006
6 Using <><>, select the unit for the slope.
7 Confirm with <OK>.
ASY, the unit of the value for the zero point (mV or pH) is displayed.
35
Operation Lab 850
Unit of the value for the
zero point (ASY)
8 Using <><>, select the unit for the zero point.
9 Confirm with <OK>.
rES, the setting of the resolution is displayed.
pH
A
S
Y
mV
Resolution
(rES)
pH
e
H
s
I
r
10 Using <><>, toggle between Hi and Lo.
11 Confirm with <OK>.
Unit, the setting of the unit of the temperature value is displayed.
36
ba75553e03 01/2006
Lab 850 Operation
Temperature unit (Unit)
Calibration
interval
(Int.C)
i
o
t
C
C
TP
°
C
n
U
12 Using <><>, toggle between °C and °F.
13 Confirm with <OK>.
Int.C, the setting of the calibration interval is displayed.
N
I
Time
T
d
AutoCal
14 Set the interval with <▲><▼>.
15 Confirm with <OK>.
The measurement settings are completed. The meter switches to the measuring mode.
1
4
ba75553e03 01/2006
37
Operation Lab 850

4.6 Reset

4.6.1 Resetting calibration values

This function resets the calibration values to the default condition. All other meter settings are retained.
Calibration values in the
default condition
Resetting calibration
values
Zero point pH 7.000 (0 mV)
Slope 100 % (-59.16 mV/pH)
Note
The measuring system is not calibrated after a reset. Before measuring recalibrate the meter.
1 Press <On/Off_> to open the menu for the reset of the
calibration data. Ini.C is displayed.
pH
n
C
o
I
n
i
38
2 Press <><> to display no or YES.
YES: Reset the calibration values. no: Retain the calibration values.
3 Confirm with <OK>.
The menu is finished. The meter switches to the measuring mode.
ba75553e03 01/2006
Lab 850 Operation

4.6.2 Resetting all meter settings

This function resets to the default condition all meter settings including the calibration values. The corresponding values are given on the following pages:
Calibration values page 38
System settings page 32
Measurement settings page 34
Resetting the meter
settings
1 Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2 During the display test, press <MODE> to open the menu for
the reset of the meter settings. Init is displayed.
i
n
t
o
n
I
3 Press <><> to display no or YES.
YES: Reset the meter settings. no: Retain the meter settings.
ba75553e03 01/2006
4 Confirm with <OK>.
The menu is finished. The meter switches to the measuring mode.
Note
The measuring system is not calibrated after a reset. Before measuring recalibrate the meter.
39
Operation Lab 850
40
ba75553e03 01/2006
Lab 850 Maintenance, cleaning, disposal

5 Maintenance, cleaning, disposal

5.1 Maintenance

The only maintenance activity required is replacing the batteries.
Note
For maintenance of the electrodes refer to the relevant operating manuals.

5.1.1 Replacing the batteries

1 Open the battery compartment (1) on the underside of the
meter.
2 Remove the four batteries from the battery compartment.
3 Place four new batteries (type Mignon AA) in the battery
compartment.
4 Close the battery compartment (1).
The date (day) flashes in the display.
5 Set the date and time according to page 32.
1
ba75553e03 01/2006
Caution
Make sure that the poles of the batteries are the right way round. The
± signs on the batteries must correspond to the ± signs in the
battery compartment. Only use leakproof alkaline manganese batteries.
41
Maintenance, cleaning, disposal Lab 850

5.2 Cleaning

Occasionally wipe the outside of the meter with a damp, lint-free cloth. Disinfect the housing with isopropanol as required.
Caution
The housing is made of synthetic material (ABS). Thus, avoid contact with acetone or similar detergents that contain solvents. Remove any splashes immediately.

5.3 Packing

This meter is sent out in a protective transport packing. We recommend: Keep the packing material. The original packing protects the instrument against damage during transport.

5.4 Disposal

Batteries This note refers to the battery regulation that applies in the Federal
Republic of Germany. We would ask end-consumers in other countries to follow their local statutory provisions.
Note
This instrument contains batteries. Batteries that have been removed must only be disposed of at the recycling facility set up for this purpose or via the retail outlet. It is illegal to dispose of them in household refuse.
42
ba75553e03 01/2006
Lab 850 What to do if...

6 What to do if...

Error message
Err1
Error message
Err2
Cause Remedy
pH electrode:
– Measured value outside the
measuring range
– Air bubble in front of the
junction
Air in the junction – Extract air or moisten
Cable broken Replace the electrode
Gel electrolyte dried out Replace the electrode
Cause Remedy
No electrode connected Connect the electrode
Setting time during calibration
too long
Use a suitable electrode
Remove air bubble
junction
– Adjust temperature if
necessary
– Recalibrate
Error message
Err4
Error message
CalError
Cause Remedy
– Temperature not stable during
calibration.
Cause Remedy
pH electrode:
– The values determined for zero
point and slope of the electrode are outside the allowed limits.
Junction contaminated Clean junction
Electrode broken Replace the electrode
Buffer solutions:
– Adjust temperature if
necessary
Recalibrate
Recalibrate
ba75553e03 01/2006
43
What to do if... Lab 850
Cause Remedy
– Incorrect buffer solutions – Change calibration
procedure
– Buffer solutions too old – Use only once.
Note the shelf life
– Buffer solutions depleted – Change solutions
No stable measured
Cause Remedy
value
pH electrode:
Junction contaminated Clean junction
Membrane contaminated Clean membrane
CalClock expired
Test sample:
– pH value not stable – Measure with air excluded if
necessary
– Temperature not stable – Adjust temperature if
necessary
Electrode + test sample:
Conductivity too low Use a suitable electrode
Temperature too high Use a suitable electrode
Organic liquids Use a suitable electrode
Cause Remedy
– Calibration interval expired – Recalibrate the measuring
system
44
Display,
LoBat
Cause Remedy
– Batteries almost empty – Replace the batteries (see
page 41)
ba75553e03 01/2006
Lab 850 What to do if...
Cause Remedy
Obviously incorrect
measured values
pH electrode:
– pH electrode unsuitable – Use a suitable electrode
Instrument does not
react to keystroke
You want to know which
software
version is in the
instrument
– Temperature difference
between buffer and test sample
– Adjust temperature of buffer
or sample solutions
too high
Measurement procedure not
Follow special procedure
suitable
Cause Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press and hold the <SC> key and switch the meter on
Cause Remedy
– E. g., a question by the service
department
– Switch on the meter.
During the display test, display the software version with <OK>.
ba75553e03 01/2006
45
What to do if... Lab 850
46
ba75553e03 01/2006
Lab 850 Technical data
Stand October 10, 2005Translation of the legally binding German version

7 Technical data

7.1 General data

Dimensions approx. 240 x 190 x 80 mm
Weight approx. 1.0 kg (without power pack, without stand)
Mechanical structure Type of protection IP 43
Electrical safety Protective class III Test certificates cETLus
Ambient
conditions
Power
supply
Service
interface
Storage - 25 °C ... + 65 °C
Operation 0 °C ... + 55 °C
Climatic class 2
Batteries 4 x 1.5 V alkali-manganese batteries, Type
AA
Operational life Approx. 1000 operating hours
Power pack (charging device)
This interface can be used for service purposes only.
Baud rate 4800 baud
Type RS232, bidirectional
Data bits 8
Stop bits 2
FRIWO FW7555M/09, 15.1432.500-00 Friwo Part. No. 1883259 Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA Output: 9 V = / 1.5 A Connection max. overvoltage category II Primary plugs contained in the scope of delivery: Euro, US, UK and Australian.
ba75553e03 01/2006
Parity None
Handshake RTS/CTS+Xon/Xoff
47
Technical data Lab 850
Stand October 10, 2005Translation of the legally binding German version
Guidelines
and norms used
EMC EC guideline 89/336/EEC
EN 61326 EN 61000-3-2 EN 61000-3-3 FCC Class A
Instrument safety EC guideline 73/23/EEC
EN 61010-1 ANSI/UL 61010-1 CAN/CSA-C22.2 No. 61010-1
Climatic class VDI/VDE 3540
IP protection EN 60529
48
ba75553e03 01/2006
Lab 850 Technical data
Stand October 10, 2005Translation of the legally binding German version

7.2 Measuring ranges, resolution, accuracy

Measuring ranges,
resolution
Variable Measuring range Resolution
pH - 2.000 ... + 19.999
- 2.00 ... + 19.99
U [mV] - 999.9 ... + 999.9
- 2000 ... + 2000
0.001
0.01
0.1 1
T [°C] - 5.0 ... + 120.0 0.1
T [°F] + 23.0 ... + 248.0 0.1
Manual
temperature input
Variable Range Increment
T
T
[°C] - 25 ... + 125 1
manual
[°F] - 13 ... + 257 1
manual
Accuracy (± 1 digit) Variable Accuracy Temperature of the test
sample
pH / range *
- 2.000 ... + 19.999 ± 0.005 + 15 °C ... + 35 °C
- 2.00 ... + 19.99 ± 0.01 + 15 °C ... + 35 °C
U [mV] / range
- 999.9 ... + 999.9 ± 0.3 + 15 °C ... + 35 °C
- 2000 ... + 2000 ± 1 + 15 °C ... + 35 °C
T [°C] / temperature sensor
NTC 30 ± 0.1
PT 1000 ± 0.3
* when measuring in a range of ± 2 pH around a calibration point
Note
The accuracy values specified here apply exclusively to the meter. The accuracy of the electrodes and buffer solutions has to be taken into account additionally.
ba75553e03 01/2006
49
Technical data Lab 850
Stand October 10, 2005Translation of the legally binding German version
50
ba75553e03 01/2006
Lab 850 Lists

8Lists

This chapter provides additional information and orientation aids.
Abbreviations The list of abbreviations explains abbreviations that appear on the
display or when dealing with the instrument.
Specialist terms The glossary briefly explains the meaning of the specialist terms.
However, terms that should already be familiar to the target group are not described here.
Trademarks used The list comprises the trademarks used in the present document and
their owners.
Index The index will help you to find the topics that you are looking for.
ba75553e03 01/2006
51
Lists Lab 850
Abbreviations
°C Temperature unit °Celsius
°F Temperature unit, °Fahrenheit
ASY Zero point (asymmetry)
AutoCal Automatic calibration using a selected buffer set
Cal Calibration
CalError Error message (see W
HAT TO DO IF...)
Err1 Overflow
Display range exceeded
Err2, Err4 Error message (see W
HAT TO DO IF...)
InI Initialization
Resets individual basic functions to the status they had on delivery
LoBat Low Battery (batteries almost empty)
mV Voltage unit
mV/pH Unit of the electrode slope
pH pH value
S Slope value
SC Stability control (drift control)
SELV Safety Extra Low Voltage
SLO Slope (slope on calibration)
52
TP Temperature Probe
Temperature measurement active
ba75553e03 01/2006
Lab 850 Lists
Glossary
Asymmetry see zero point
Resolution Smallest difference between two measured values that can be
displayed by a meter.
AutoRange Name of the automatic selection of the measuring range.
Junction The junction is a porous body in the housing wall of reference
electrodes or electrolyte bridges. It arranges the electrical contact between two solutions and makes the electrolyte exchange more difficult. The expression, junction, is also used for ground or junction­less transitions.
Adjusting To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or a value that is regarded as correct, or that the difference remains within the tolerance.
Calibration Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting).
Electromotive force of
an electrode
The electromotive force (voltage) U of the electrode is the measurable electromotive force of an electrode in a solution. It equals the sum of all the galvanic voltages of the electrode. Its dependency on the pH results in the electrode function which is characterized by the parameters, slope and zero point.
Measured variable The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or DO concentration.
Test sample Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample and original sample are identical if the test sample was not processed.
Measured value The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g
solvent.
Zero point The zero point of a pH electrode is the pH value at which the
pH value The pH is a measure of the acidic or basic effect of an aqueous
ba75553e03 01/2006
electromotive force of the pH electrode at a specified temperature is zero. Normally, this is at 25 °C.
solution. It corresponds to the negative decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality. The practical pH value is the value of a pH measurement.
53
Lists Lab 850
Potentiometry Name of a measuring technique. The signal (depending on the
measured parameter) of the electrode is the electrical potential. The electrical current remains constant.
ORP voltage The ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface (e. g. a gold or platinum surface).
Reset Restoring the original condition of all settings of a measuring system.
Stability control Function to control the measured value stability.
Standard solution The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Slope The slope of a linear calibration function.
54
ba75553e03 01/2006
Lab 850 Lists
Trademarks used
Trademark Owner of the trademark
Merck Merck KGaA
Mettler Toledo Mettler Toledo
Fisher Fisher Scientific Company
Fluka Fluka AG
Radiometer Radiometer
Baker Mallinckrodt Baker, Inc.
Metrohm Metrohm AG
Beckman Beckman Instruments, Inc.
Hamilton Hamilton Company Corporation
Precisa Precisa Instruments AG
Windows Microsoft Corporation
ba75553e03 01/2006
55
Lists Lab 850
56
ba75553e03 01/2006
Lab 850 Lists
Index
A
Authorized use . . . . . . . . . . . . . . . . . . . . . . 12
AutoCal . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
AutoRead . . . . . . . . . . . . . . . . . . . . . . . . . . 20
B
Battery compartment . . . . . . . . . . . . . . . . . 41
Baud rate setting . . . . . . . . . . . . . . . . . . . . 32
C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 22
Calibration evaluation pH . . . . . . . . . . . . . . 25
Calibration interval . . . . . . . . . . . . . . . . . . . 26
Calibration points pH . . . . . . . . . . . . . . . . . 24
Connect the plug-in power supply unit . . . . 15
D
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Drift control . . . . . . . . . . . . . . . . . . . . . . . . . 20
E
Energy saving feature . . . . . . . . . . . . . . . . 17
Error message . . . . . . . . . . . . . . . . . . . . . . 43
F
Firmware update . . . . . . . . . . . . . . . . . . . . 59
ORP electrode . . . . . . . . . . . . . . . . . . . . . . 21
ORP voltage . . . . . . . . . . . . . . . . . . . . . . . 21
P
pH buffer sets . . . . . . . . . . . . . . . . . . . . . . 22
Precautions . . . . . . . . . . . . . . . . . . . . . . . . 11
R
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
all meter settings . . . . . . . . . . . . . . . . . 39
Calibration values . . . . . . . . . . . . . . . . 38
Resolution setting . . . . . . . . . . . . . 34, 35, 36
S
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Scope of delivery . . . . . . . . . . . . . . . . . . . . 15
Setting the date . . . . . . . . . . . . . . . . . . 15, 32
Setting the time . . . . . . . . . . . . . . . . . . 15, 32
Single-point calibration . . . . . . . . . . . . . . . 28
Slope pH . . . . . . . . . . . . . . . . . . . . . . . 22, 29
Socket field . . . . . . . . . . . . . . . . . . . . . . . . 10
T
Temperature sensor . . . . . . . . . . . . . . . . . 19
Three-point calibration . . . . . . . . . . . . . . . . 24
Two-point calibration . . . . . . . . . . . . . . . . . 29
I
Initial commissioning . . . . . . . . . . . . . . . . . 15
Interval Calibration . . . . . . . . . . . . . . . . . . . 26
K
Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
L
LoBat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
M
Measurement accuracy . . . . . . . . . . . . . . . 26
O
Operation location . . . . . . . . . . . . . . . . . . . 17
Operational safety . . . . . . . . . . . . . . . . . . . 12
ba75553e03 01/2006
Z
Zero point of the pH electrode . . . . . . . 22, 29
57
Lists Lab 850
58
ba75553e03 01/2006
Lab 850

Appendix: Firmware update

General information With the Update_Labxxx_MxxxP program and a PC you can update
the firmware of the Lab 850 to the newest version.
For the update you need:
! a free RS232 interface on your PC
! the RS232 cable, Z390.
Program installation With the installation program,
"Install_Update_Labxxx_MxxxP_Vx_yy_English.exe", install the firmware update program on your PC.
Program start Start the "Update_Labxxx_MxxxP" program from the Windows start
menu.
You can change the language via the language menu.
Firmware update Proceed as follows:
1 Connect the Lab 850 to a serial interface (COM port) of the PC
with the aid of the interface cable Z390.
2 Make sure the Lab 850 is switched on.
3 To start the update process click the OK button.
The program automatically recognizes the used interface.
4 To go on, follow the instructions of the program.
During the programming process, a corresponding message and a progress bar (in %) appear. The programming process takes approx. two minutes. A terminatory message is displayed after a successful programming process. The firmware update is now completed.
5 Disconnect the meter from the PC.
The instrument is ready for operation.
After switching the meter off and on you can check whether the meter has taken over the new software version (see page 45).
ba75553e03 01/2006
59
Lab 850
60
ba75553e03 01/2006
SCHOTT Instruments GmbH Postfach 2443 D-55014 Mainz Hattenbergstr. 10 D-55122 Mainz
Telefon +49 (0) 61 31/66 5111 Telefax +49 (0) 61 31/66 5001 Internet: www.schottinstruments.com
Loading...