WarrantyWe guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result in
differences between this operating manual and your instrument. Also, we
cannot guarantee that there are absolutely no errors in this manual.
Therefore, we are sure you will understand that we cannot accept any legal
claims resulting from the data, figures or descriptions.
purchase.
The instrument warranty covers manufacturing faults that are discovered
within the warranty period.
The warranty does not cover components that are replaced during
maintenance work, e. g. batteries.
The warranty claim extends to restoring the instrument to readiness for use
but not, however, to any further claim for damages. Improper handling or
unauthorized opening of the instrument invalidates any warranty claim.
To ascertain the warranty liability, return the instrument and proof of
purchase together with the date of purchase freight paid or prepaid.
The compact Lab 850 precision pH meter enables you to perform pH
measurements rapidly and reliably. The Lab 850 provides the
maximum degree of operating comfort, reliability and measuring
certainty for all applications.
The proven calibration procedures and special stability control function
(SC) support your work with the pH meter.
pH
6
9
9
2
2
4
L
a
8
b
8
°
C
5
TP
0
3
2
1
1Keypad
2Display
3Socket field
Note
The meter is also available as part of individual Sets of equipment.
You will find information on this and other accessories in the SCHOTT
Instruments laboratory catalog or via the Internet.
Note
If you need further information or application notes, you can obtain the
following material from SCHOTT Instruments:
! Application reports
! pH primers
! Safety datasheets.
ba75553e0301/2006
7
OverviewLab 850
1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) indicates a keystroke in this operating
manual.
MODE
CAL
SC
<On/Off>:
<On/Off
<MODE>:
<MODE
<CAL>:
<CAL
<SC>:Activate / deactivate stability control
<▲>: Increment values, scroll
<▼>: Decrement values, scroll
_>:
_>:
_>:
Switch meter on / off
Reset calibration values
Select measured parameter
Open setting menu for calibration and
measurement
Call up calibration procedure
Display calibration data
<OK>:
_>:
<OK
8
Confirm entries
Open setting menu for system settings
ba75553e0301/2006
Lab 850Overview
1.2Display
Status display
indicator
CalClock (calibration
interval and sensor
evaluation)
Measured value
display
Function and
temperature display
indicator
1
Time
Day.Month
Year
LoBat
SET
8
AutoCal
U pH
8
Baud
No.
Ident
88
Auto
BUFFER
8
8
Store
8
8
CalError
SC
%
mV/pH
°F
°
C
RCL
TP
Function display
indicators
SETDuring calibration: number of the buffer set
BUFFERDuring calibration: number of the buffer
AutoCalCalibration with automatic buffer recognition
CalErrorAn error occurred during calibration
LoBatWith battery operation: batteries almost empty
SCStability control is active
TPTemperature measurement active
ba75553e0301/2006
9
OverviewLab 850
1.3Socket field
Connectors:
1pH electrode
2Reference electrode
3Temperature sensor
4Power pack
5Service interface
Caution
Only connect sensors to the meter that cannot return any voltages
or currents that are not allowed (> SELV and > current circuit with
current limiting).
Almost all sensors - in particular SCHOTT Instruments sensors fulfill these conditions.
10
ba75553e0301/2006
Lab 850Safety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating
manual before working with the meter.
The operating manual must always be available within the vicinity of the
instrument.
Target groupThe meter was developed for work in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.
Safety instructionsSafety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the level of danger:
Further notes
Warning
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.
Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the instrument
or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.
ba75553e0301/2006
11
SafetyLab 850
2.1Authorized use
This meter is authorized exclusively for pH and ORP measurements in
the laboratory.
The technical specifications as given in chapter 7 T
ECHNICALDATA
(page 47) must be observed. Only the operation and running of the
meter according to the instructions given in this operating manual is
authorized.
Any other use is considered unauthorized.
2.2General safety instructions
This instrument is constructed and tested in compliance with the IEC
1010 safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during
operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICALDATA (page 47).
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty
functioning of the instrument. In this event, wait until the temperature of
the instrument reaches room temperature before putting the instrument
back into operation.
Caution
The meter is only allowed to be opened by personnel authorized
by SCHOTT Instruments.
12
ba75553e0301/2006
Lab 850Safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
of service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
! has been damaged in transport
! has been stored under adverse conditions for a lengthy period of
time
! is visibly damaged
! no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
! EEC directives for protective labor legislation
! National protective labor legislation
! Safety regulations
! Safety datasheets of the chemical manufacturers.
ba75553e0301/2006
13
SafetyLab 850
14
ba75553e0301/2006
Lab 850Commissioning
3Commissioning
3.1Scope of delivery
! Lab 850 laboratory meter
! Power pack
! 4 batteries 1.5 V Mignon type AA
! Transparent cover
! Stand with stand base (S4D)
! Operating manual
! CD-ROM
3.2Initial commissioning
Switching on the meter
Setting the date and
time
Connecting the power
pack
Perform the following activities:
! Switch on the meter
! Set the date and time
! Connect the power pack (for line power operation only).
1Switch on the meter with <On/Off>.
A display test is briefly displayed.
2See page 32.
Caution
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see page 47).
ba75553e0301/2006
Caution
Use original power packs only (see page 47).
3Insert the plug into the socket of the pH meter.
4Connect the original power pack to an easily accessible power
outlet.
15
CommissioningLab 850
Note
You can carry out measurements without the power pack.
3.2.1Z865 stand (S4D)
The meter can be inserted in the stand base on the right or left side of
the electrode holder.
1Insert the stand rod in the stand base.
2Insert the meter in the stand base.
An arrangement of meter with stand looks, for example, as follows:
16
ba75553e0301/2006
Lab 850Operation
4Operation
4.1Switching on the meter
1Place the meter on a flat surface and protect it from intense
light and heat.
2Press the <On/Off>
A display test is briefly displayed.
Subsequently, the meter switches to the measuring mode
(measured value display).
Note
The meter has an energy saving feature to avoid unnecessary battery
depletion during battery operation.
The energy saving feature switches off the meter if no key was pressed
during the specified interval (setting the switch-off interval, see
page 32).
key.
ba75553e0301/2006
17
OperationLab 850
4.2General operating principles
This section contains basic information on the operation of the Lab 850.
4.2.1Operating modes
The instrument has the following operating modes:
! Measurement
The display indicates the measurement data in the measured value
display
! Calibration
The display guides you thru a calibration procedure with calibration
information
! Configuration
The system menu or a sensor menu with submenus, settings and
functions is displayed
4.2.2Operation
KeysThe meter is operated via keys. The keys can have different functions
with long or short keystrokes.
FunctionsGenerally, with a short keystroke a function is carried out. A long
keystroke opens a setting menu.
In a setting menu, settings are selected with the <▲><▼> keys.
A setting is confirmed with <OK>. With confirming, the setting is
finished and the next setting is displayed.
RepresentationIn this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<OK
_>).
18
ba75553e0301/2006
Lab 850Operation
4.3Measuring
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect an electrode to the meter.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3Calibrate or check the meter with the electrode.
4Select the measured parameter with <MODE>.
Note
Incorrect calibration of pH electrodes leads to incorrect measured
values. Calibrate regularly before measuring.
Caution
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.
Temperature sensorYou can measure with or without a temperature sensor. If a
temperature sensor is connected, it is indicated on the display by TP.
Note
The pH meter automatically recognizes the type of the temperature
sensor used. Therefore, you can connect electrodes with an NTC30 or
Pt1000.
The temperature measurement is absolutely essential for a
reproducible pH measurement. If the measurement is made without a
temperature sensor, proceed as follows:
1Measure the current temperature using a thermometer.
2Set the temperature value with <▲><▼>.
Note
When calibrating without temperature sensor, also set the current
temperature of each buffer solution manually (see <▲><▼>).
ba75553e0301/2006
19
OperationLab 850
4.3.1Measuring the pH value
1Perform the preparatory activities according to page 19.
2Immerse the pH electrode in the test sample.
3Using <MODE>, scroll as necessary until the measured
parameter pH is displayed.
pH
Stability control SC
(drift control)
9
4
2
8
TP
°
C
9
6
2
The stability control function (drift control) checks the stability of the
measurement signal. The stability has a considerable effect on the
reproducibility of the measured value.
For identical measurement conditions, the following criteria apply:
Drift within 15 sec < 0.02 pH units.
1If necessary, call up the measured variable pH with <MODE>.
2With <SC>, activate the stability control function.
The SC function display indicator appears.
The current measured value is frozen (hold function).
3Start measurement with stability control with <OK>.
SC flashes until a stable measured value is reached.
This measured value is downloaded to the interface.
20
4If necessary, start the next measurement with stability control
with <OK>.
5To terminate the stability control function: Press the <MODE>
or <SC>
Note
The current measurement with stability control can be terminated at
any time (accepting the current value) by pressing <OK>.
key.
ba75553e0301/2006
Lab 850Operation
4.3.2Measuring the ORP
The meter can, in conjunction with an ORP electrode, measure the
ORP (mV) of a solution.
1Perform the preparatory activities according to page 19.
2Immerse the ORP electrode in the test sample.
3If necessary, call up the measured parameter U with <MODE>.
4Wait for a stable measured value.
U
4
8
2
Note
ORP electrodes are not calibrated. However, you can check ORP
electrodes using a test solution.
6
1
3
mV
TP
°
C
ba75553e0301/2006
21
OperationLab 850
4.4Calibration
Why calibrate?pH electrodes age. This changes the zero point (asymmetry) and slope
of the pH electrode. As a result, an inexact measured value is
displayed. Calibration determines the current values of the zero point
and slope of the electrode and stores them. Thus, you should calibrate
at regular intervals.
The calibration data is stored in the meter.
When do you have to
calibrate?
Buffer sets for
calibration
! After connecting another electrode
! When the CalClock on the display flashes:
– after the calibration interval has expired
– after a voltage interruption (empty batteries)
You can use the buffer sets quoted in the table for an automatic
calibration. The pH values are valid for the specified temperature
values. The temperature dependence of the pH values is taken into
account during calibration.
No.Buffer set*pH valuesat
1SCHOTT Instruments DIN buffers
according to DIN 19266/NBS
1.679
4.006
25 °C
6.865
9.180
12.454
2SCHOTT Instruments Technical
buffers according to DIN 19267
2.000
4.010
25 °C
7.000
10.011
22
3Merck1*4.000
7.000
9.000
4Merck2 *1.000
6.000
8.000
13.000
5Merck3 *4.660
6.880
9.220
6DIN 19267 *1.090
4.650
6.790
9.230
ba75553e0301/2006
20°C
20°C
20°C
25 °C
Lab 850Operation
No.Buffer set*pH valuesat
7Mettler Toledo USA *1.679
4.003
7.002
10.013
8Mettler Toledo TEC *1.995
4.005
7.002
9.208
9Fisher *2.007
4.002
7.004
10.002
10Fluka BS *4.006
6.984
8.957
11Radiometer *1.678
4.005
7.000
9.180
12Baker *4.006
6.991
10.008
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
13Metrohm *3.996
25 °C
7.003
8.999
14Beckman *4.005
25 °C
7.005
10.013
15Hamilton Duracal *4.005
25 °C
7.002
10.013
16Precisa *3.996
25 °C
7.003
8.999
*Brand names or trade names are trademarks of their respective
owners protected by law (see page 55).
ba75553e0301/2006
23
OperationLab 850
Note
The buffer set is selected in the menu for measurement settings (see
page 34).
Calibration pointsCalibration can be performed using one, two or three buffer solutions in
any order (single-point, two-point or three-point calibration). The meter
determines the following values and calculates the calibration line as
follows:
Determined
Displayed calibration data
values
1-pointASY! Zero point = ASY
! Slope = Nernst slope
(-59.16 mV/pH at 25 °C)
2-pointASY
SLO
3-pointASY
SLO
! Zero point = ASY
! Slope = SLO
! Zero point = ASY
! Slope = SLO
The calibration line is calculated
by linear regression.
Note
You can display the slope in the units, mV/pH or %.
You can display the zero point in the units, mV or pH.
AutoCalis adapted to the permanently programmed buffer solutions as a fully
automatic two-point or three-point calibration. The buffer solutions are
automatically recognized by the meter.
24
Stability controlThe calibration procedure automatically activates the stability control
function.
The current measurement with stability control can be terminated at
any time (accepting the current value) by pressing <OK>.
Calibration recordWhen finishing a calibration, the new calibration values are first
displayed as an informative message and stored. )
ba75553e0301/2006
Lab 850Operation
Displaying the
You can display the data of the last calibration (see page 30).
calibration data
Calibration evaluationAfter calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account.The evaluation appears on the display as the
CalClock and in the calibration record.
CalClockZero point
[mV]
Slope
[mV/pH]
-15 ... + 15-60.5 ... -58
-20 ... + 20-58 ... -57
-25 ... +25-61 ... -60.5
or
-57 ... -56
-30 ... +30-62 ... -61
or
-56 ... -50
Preparatory activities
Clean the electrode
according to the
sensor operating
manual
CalError
< -30 or
> 30
... -62 or
... -50
Eliminate the error
according to chapter
6 W
HATTODOIF...
(page 43)
1Switch on the meter with <On/Off>.
2Connect a pH electrode to the meter.
3Keep the buffer solutions ready.
4Adjust the temperature of the solutions and measure the
current temperature if the measurement is made without
temperature sensor.
ba75553e0301/2006
5Set the buffer set to be used for calibration as necessary.
25
OperationLab 850
4.4.1Calibration interval (Int.C)
The calibration interval and calibration evaluation are indicated on the
display as the CalClock.
CalClock
The remaining time of the calibration interval is indicated by the
segmented ring around the calibration evaluation. This segmented ring
reminds you to calibrate regularly.
After the specified calibration interval (Int.C) has expired, the outer ring
of the CalClock flashes. It is still possible to measure.
Note
To ensure the high measuring accuracy of the meter, calibrate after the
calibration interval has expired.
Setting the calibration
interval
The calibration interval (Int.C) is set to 7 days (d7) in the factory.
You can change the interval (1 ... 999 days):
1Open the menu for measurement settings with <MODE_>.
2Confirm all settings with <OK> until Int.C is displayed.
pH
N
1
Time
T
D
3Set the calibration interval with <▲><▼>.
4Confirm the setting with <OK>.
C
8
26
ba75553e0301/2006
Lab 850Operation
4.4.2Automatic calibration (AutoCal)
For this procedure, use two or three buffer solutions of the selected
buffer set in any order.
Note
The steps 2, 6 and 10 are not necessary if you use a temperature
sensor.
1Start the calibration with <CAL>.
The number of the specified buffer set is displayed (SET 1 ...
16).
The first buffer (BUFFER 1) of the buffer set (SET_x) is
requested.
2
BUFFER
1
8
5
TP
°
C
SET
pH
1
AutoCal
2If necessary, set the temperature of the buffer solution with
<▲><▼>.
3Immerse the pH electrode in the first buffer solution.
4Start the measurement with <OK>.
The SC display indicator flashes.
Depending on the setting, the display shows the nominal pH
value for the recognized buffer or the electrode voltage (mV).
As soon as a stable value is recognized, the next buffer
(BUFFER 2) is requested.
ba75553e0301/2006
27
OperationLab 850
Continue with two-point
calibration
2
BUFFER
2
8
5
TP
°
C
SET
pH
1
AutoCal
Note
Here you can cancel the calibration procedure with <MODE>. This
corresponds to a single-point calibration. The value of the zero point
(Asy) is displayed for 10 seconds, then the value of the slope (Slo) is
displayed for 10 seconds. After this the meter switches to the
measuring mode.
5Thoroughly rinse the electrode with distilled water.
6If necessary, set the temperature of the second buffer solution
with <▲><▼>.
7Immerse the pH electrode in the second buffer solution.
8Start the measurement with <OK>.
The SC display indicator flashes.
Depending on the setting, the display shows the nominal pH
value for the recognized buffer or the electrode voltage (mV).
As soon as a stable value is recognized, the next buffer
(BUFFER 3) is requested.
2
Buffer
5
3
8
TP
°
C
SET
1
AutoCal
DIN
pH
28
ba75553e0301/2006
Lab 850Operation
Note
Here you can cancel the calibration procedure with <MODE>. This
corresponds to a two-point calibration. The value of the zero point
(Asy) is displayed for 10 seconds, then the value of the slope (Slo) is
displayed for 10 seconds. After this the meter switches to the
measuring mode.
Continue with three-
point calibration
9Thoroughly rinse the electrode with distilled water.
10If necessary, set the temperature of the third buffer solution
with <▲><▼>.
11Immerse the pH electrode in the third buffer solution.
12Press the <OK>
The SC display indicator flashes.
Depending on the setting, the display shows the nominal pH
value for the recognized buffer or the electrode voltage (mV).
As soon as a stable value is recognized, the value of the zero
point (ASY) is displayed for 10 seconds, then the value of the
slope (SLO) is displayed for 10 seconds.
After this the meter switches to the measuring mode.
Note
While the zero point (ASY) is being displayed, you can change the unit
of the zero point with
While the slope (SLO) is being displayed, you can change the unit of
the slope with
The % display refers to the Nernst slope of 59.2 mV/pH at 25° C (100
x determined slope/Nernst slope).
<▲><▼> .
key.
<▲><▼> .
ba75553e0301/2006
The unit of zero point and slope can be permanently changed in the
measurement settings (see page 34).
29
OperationLab 850
4.4.3Downloading calibration data
1Press <CAL_> to display the calibration data.
The following data is displayed consecutively for 10 seconds
each:
date, zero point, slope.
While the calibration data is being displayed, you can:
!press <OK>to display further calibration data
(date, zero point, slope)
!press
<▲><▼>
!press
<▲><▼>
to switch over the unit of the zero point (Asy)
(while the zero point is being displayed)
to switch over the unit of the slope (Slo)
(while the slope is being displayed)
30
ba75553e0301/2006
Lab 850Operation
4.5Settings
You can adapt the meter to your individual requirements. The settings
are done in the following menus:
! System settings (<OK
– Switch-off interval (t.Off)
– Date (Day.Month)
– Date (Year)
– Time (Time)
! Settings for calibration and measurement (<MODE
– Number of the buffer set for pH calibration (Set 1 ... 16)
– Display of the buffer during calibration
(pH nominal value or measured voltage value in mV)
– Unit of the value for the slope (mV/pH or %)
– Unit of the value for the zero point (mV, pH)
– Resolution (Hi, Lo)
– Temperature unit (°C / °F)
– Calibration interval (Int.C [0 ... 999])
Note
You can exit the setting menu at any time by pressing <MODE>.
Settings already modified and confirmed with <OK> are stored.
Day.Month, the setting of the date is displayed.
The day display flashes.
Day.Month
0
30
32
ba75553e0301/2006
Lab 850Operation
4Set the date of the current day with <▲><▼> .
5Confirm with <OK>.
The month display flashes.
6Set the current month with <▲><▼> .
7Confirm with <OK>.
Year, the setting of the year is displayed.
8Set the year with <▲><▼> .
9Confirm with <OK>.
The setting of the time is displayed.
The hour display flashes.
10Set the current hour with <▲><▼> .
11Confirm with <OK>.
The minute display flashes.
12Set the current minute with <▲><▼> .
13Confirm with <OK>.
The system settings are completed.
The meter switches to the measuring mode.
ba75553e0301/2006
33
OperationLab 850
4.5.2Measurement settings
These settings apply to calibration and measurement (the default
setting is printed in bold).
Number of the buffer set for pH
calibration (SET)
Display during calibration (BUFFER)pH (buffer nominal value),
1 ... 16
U (electrode voltage)
Buffer set for pH
calibration
(SET)
Unit of the value for the zero point
(ASY)
Unit of the value for the slope (SLO)%, mV/pH
Temperature unit (UnIt)°C, °F
Calibration interval (Int.C) 0 ... 7 ... 999 d
1Open the menu for measurement settings with <MODE_>.
Set 1 ... 16, the specified buffer set is displayed.
SET
pH, mV
1
34
AutoCal
2Using <▲><▼>, select a buffer set Set 1 ... 16.
3Confirm with <OK>.
bUFF, the setting of the display during calibration is displayed.
ba75553e0301/2006
Lab 850Operation
Display during
calibration
(bUFF)
pH
Unit of the value for the
slope (SLO)
F
H
L
F
O
mV/pH
U
b
p
AutoCal
4Using <▲><▼>, select the display during calibration pH or U.
5Confirm with <OK>.
SLO, the unit of the value for the slope (mV/pH or %) is
displayed.
pH
S
ba75553e0301/2006
6Using <▲><▼>, select the unit for the slope.
7Confirm with <OK>.
ASY, the unit of the value for the zero point (mV or pH) is
displayed.
35
OperationLab 850
Unit of the value for the
zero point (ASY)
8Using <▲><▼>, select the unit for the zero point.
9Confirm with <OK>.
rES, the setting of the resolution is displayed.
pH
A
S
Y
mV
Resolution
(rES)
pH
e
H
s
I
r
10Using <▲><▼>, toggle between Hi and Lo.
11Confirm with <OK>.
Unit, the setting of the unit of the temperature value is
displayed.
36
ba75553e0301/2006
Lab 850Operation
Temperature unit (Unit)
Calibration
interval
(Int.C)
i
o
t
C
C
TP
°
C
n
U
12Using <▲><▼>, toggle between °C and °F.
13Confirm with <OK>.
Int.C, the setting of the calibration interval is displayed.
N
I
Time
T
d
AutoCal
14Set the interval with <▲><▼>.
15Confirm with <OK>.
The measurement settings are completed.
The meter switches to the measuring mode.
1
4
ba75553e0301/2006
37
OperationLab 850
4.6Reset
4.6.1Resetting calibration values
This function resets the calibration values to the default condition. All
other meter settings are retained.
Calibration values in the
default condition
Resetting calibration
values
Zero pointpH 7.000 (0 mV)
Slope 100 % (-59.16 mV/pH)
Note
The measuring system is not calibrated after a reset. Before measuring
recalibrate the meter.
1Press <On/Off_> to open the menu for the reset of the
calibration data.
Ini.C is displayed.
pH
n
C
o
I
n
i
38
2Press <▲><▼> to display no or YES.
YES: Reset the calibration values.
no: Retain the calibration values.
3Confirm with <OK>.
The menu is finished.
The meter switches to the measuring mode.
ba75553e0301/2006
Lab 850Operation
4.6.2Resetting all meter settings
This function resets to the default condition all meter settings including
the calibration values. The corresponding values are given on the
following pages:
Calibration valuespage 38
System settingspage 32
Measurement settingspage 34
Resetting the meter
settings
1Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2During the display test, press <MODE> to open the menu for
the reset of the meter settings.
Init is displayed.
i
n
t
o
n
I
3Press <▲><▼> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.
ba75553e0301/2006
4Confirm with <OK>.
The menu is finished.
The meter switches to the measuring mode.
Note
The measuring system is not calibrated after a reset. Before measuring
recalibrate the meter.
39
OperationLab 850
40
ba75553e0301/2006
Lab 850Maintenance, cleaning, disposal
5Maintenance, cleaning, disposal
5.1Maintenance
The only maintenance activity required is replacing the batteries.
Note
For maintenance of the electrodes refer to the relevant operating
manuals.
5.1.1Replacing the batteries
1Open the battery compartment (1) on the underside of the
meter.
2Remove the four batteries from the battery compartment.
3Place four new batteries (type Mignon AA) in the battery
compartment.
4Close the battery compartment (1).
The date (day) flashes in the display.
5Set the date and time according to page 32.
1
ba75553e0301/2006
Caution
Make sure that the poles of the batteries are the right way round.
The
±signs on the batteries must correspond to the ± signs in the
battery compartment.
Only use leakproof alkaline manganese batteries.
41
Maintenance, cleaning, disposalLab 850
5.2Cleaning
Occasionally wipe the outside of the meter with a damp, lint-free cloth.
Disinfect the housing with isopropanol as required.
Caution
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain solvents.
Remove any splashes immediately.
5.3Packing
This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing
protects the instrument against damage during transport.
5.4Disposal
BatteriesThis note refers to the battery regulation that applies in the Federal
Republic of Germany. We would ask end-consumers in other countries
to follow their local statutory provisions.
Note
This instrument contains batteries. Batteries that have been removed
must only be disposed of at the recycling facility set up for this purpose
or via the retail outlet.
It is illegal to dispose of them in household refuse.
42
ba75553e0301/2006
Lab 850What to do if...
6What to do if...
Error message
Err1
Error message
Err2
CauseRemedy
pH electrode:
– Measured value outside the
measuring range
– Air bubble in front of the
junction
– Air in the junction– Extract air or moisten
– Cable broken– Replace the electrode
– Gel electrolyte dried out– Replace the electrode
CauseRemedy
– No electrode connected– Connect the electrode
– Setting time during calibration
too long
– Use a suitable electrode
– Remove air bubble
junction
– Adjust temperature if
necessary
– Recalibrate
Error message
Err4
Error message
CalError
CauseRemedy
– Temperature not stable during
calibration.
CauseRemedy
pH electrode:
– The values determined for zero
point and slope of the electrode
are outside the allowed limits.
– Junction contaminated– Clean junction
– Electrode broken– Replace the electrode
Buffer solutions:
– Adjust temperature if
necessary
– Recalibrate
– Recalibrate
ba75553e0301/2006
43
What to do if...Lab 850
CauseRemedy
– Incorrect buffer solutions– Change calibration
procedure
– Buffer solutions too old– Use only once.
Note the shelf life
– Buffer solutions depleted– Change solutions
No stable measured
CauseRemedy
value
pH electrode:
– Junction contaminated– Clean junction
– Membrane contaminated– Clean membrane
CalClock expired
Test sample:
– pH value not stable– Measure with air excluded if
necessary
– Temperature not stable– Adjust temperature if
necessary
Electrode + test sample:
– Conductivity too low– Use a suitable electrode
– Temperature too high– Use a suitable electrode
– Organic liquids– Use a suitable electrode
CauseRemedy
– Calibration interval expired– Recalibrate the measuring
system
44
Display,
LoBat
CauseRemedy
– Batteries almost empty– Replace the batteries (see
page 41)
ba75553e0301/2006
Lab 850What to do if...
CauseRemedy
Obviously incorrect
measured values
pH electrode:
– pH electrode unsuitable– Use a suitable electrode
Instrument does not
react to keystroke
You want to know which
software
version is in the
instrument
– Temperature difference
between buffer and test sample
– Adjust temperature of buffer
or sample solutions
too high
– Measurement procedure not
– Follow special procedure
suitable
CauseRemedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press and hold the <SC>
key and switch the meter on
CauseRemedy
– E. g., a question by the service
department
– Switch on the meter.
During the display test,
display the software version
with <OK>.
ba75553e0301/2006
45
What to do if...Lab 850
46
ba75553e0301/2006
Lab 850Technical data
Stand October 10, 2005Translation of the legally binding German version
7Technical data
7.1General data
Dimensionsapprox. 240 x 190 x 80 mm
Weightapprox. 1.0 kg (without power pack, without stand)
Mechanical structureType of protectionIP 43
Electrical safetyProtective classIII
Test certificatescETLus
Ambient
conditions
Power
supply
Service
interface
Storage- 25 °C ... + 65 °C
Operation0 °C ... + 55 °C
Climatic class2
Batteries4 x 1.5 V alkali-manganese batteries, Type
AA
Operational lifeApprox. 1000 operating hours
Power pack
(charging device)
This interface can be used for service purposes only.
Baud rate4800 baud
TypeRS232, bidirectional
Data bits8
Stop bits2
FRIWO FW7555M/09, 15.1432.500-00
Friwo Part. No. 1883259
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1.5 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.
ba75553e0301/2006
ParityNone
HandshakeRTS/CTS+Xon/Xoff
47
Technical dataLab 850
Stand October 10, 2005Translation of the legally binding German version
Guidelines
and norms used
EMCEC guideline 89/336/EEC
EN 61326
EN 61000-3-2
EN 61000-3-3
FCC Class A
Instrument safetyEC guideline 73/23/EEC
EN 61010-1
ANSI/UL 61010-1
CAN/CSA-C22.2 No. 61010-1
Climatic classVDI/VDE 3540
IP protectionEN 60529
48
ba75553e0301/2006
Lab 850Technical data
Stand October 10, 2005Translation of the legally binding German version
7.2Measuring ranges, resolution, accuracy
Measuring ranges,
resolution
VariableMeasuring rangeResolution
pH- 2.000 ... + 19.999
- 2.00 ... + 19.99
U [mV]- 999.9 ... + 999.9
- 2000 ... + 2000
0.001
0.01
0.1
1
T [°C]- 5.0 ... + 120.00.1
T [°F]+ 23.0 ... + 248.00.1
Manual
temperature input
VariableRangeIncrement
T
T
[°C]- 25 ... + 1251
manual
[°F]- 13 ... + 2571
manual
Accuracy (± 1 digit)VariableAccuracyTemperature of the test
sample
pH / range *
- 2.000 ... + 19.999± 0.005+ 15 °C ... + 35 °C
- 2.00 ... + 19.99± 0.01+ 15 °C ... + 35 °C
U [mV] / range
- 999.9 ... + 999.9± 0.3+ 15 °C ... + 35 °C
- 2000 ... + 2000± 1+ 15 °C ... + 35 °C
T [°C] / temperature sensor
NTC 30± 0.1
PT 1000± 0.3
* when measuring in a range of ± 2 pH around a calibration point
Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the electrodes and buffer solutions has to be taken into
account additionally.
ba75553e0301/2006
49
Technical dataLab 850
Stand October 10, 2005Translation of the legally binding German version
50
ba75553e0301/2006
Lab 850Lists
8Lists
This chapter provides additional information and orientation aids.
AbbreviationsThe list of abbreviations explains abbreviations that appear on the
display or when dealing with the instrument.
Specialist termsThe glossary briefly explains the meaning of the specialist terms.
However, terms that should already be familiar to the target group are
not described here.
Trademarks usedThe list comprises the trademarks used in the present document and
their owners.
IndexThe index will help you to find the topics that you are looking for.
ba75553e0301/2006
51
ListsLab 850
Abbreviations
°CTemperature unit °Celsius
°FTemperature unit, °Fahrenheit
ASYZero point (asymmetry)
AutoCalAutomatic calibration using a selected buffer set
CalCalibration
CalErrorError message (see W
HATTODOIF...)
Err1Overflow
Display range exceeded
Err2, Err4Error message (see W
HATTODOIF...)
InIInitialization
Resets individual basic functions to the status they
had on delivery
LoBatLow Battery (batteries almost empty)
mVVoltage unit
mV/pHUnit of the electrode slope
pHpH value
SSlope value
SCStability control (drift control)
SELVSafety Extra Low Voltage
SLOSlope (slope on calibration)
52
TPTemperature Probe
Temperature measurement active
ba75553e0301/2006
Lab 850Lists
Glossary
Asymmetrysee zero point
ResolutionSmallest difference between two measured values that can be
displayed by a meter.
AutoRangeName of the automatic selection of the measuring range.
JunctionThe junction is a porous body in the housing wall of reference
electrodes or electrolyte bridges. It arranges the electrical contact
between two solutions and makes the electrolyte exchange more
difficult. The expression, junction, is also used for ground or junctionless transitions.
AdjustingTo manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains
within the tolerance.
CalibrationComparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct.
Often, this expression is also used when the measuring system is
adjusted at the same time (see adjusting).
Electromotive force of
an electrode
The electromotive force (voltage) U of the electrode is the measurable
electromotive force of an electrode in a solution. It equals the sum of
all the galvanic voltages of the electrode. Its dependency on the pH
results in the electrode function which is characterized by the
parameters, slope and zero point.
Measured variableThe measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or DO concentration.
Test sampleDesignation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
Measured valueThe measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
MolalityMolality is the quantity (in Mol) of a dissolved substance in 1000 g
solvent.
Zero pointThe zero point of a pH electrode is the pH value at which the
pH valueThe pH is a measure of the acidic or basic effect of an aqueous
ba75553e0301/2006
electromotive force of the pH electrode at a specified temperature is
zero. Normally, this is at 25 °C.
solution. It corresponds to the negative decadic logarithm of the molal
hydrogen ions activity divided by the unit of the molality. The practical
pH value is the value of a pH measurement.
53
ListsLab 850
PotentiometryName of a measuring technique. The signal (depending on the
measured parameter) of the electrode is the electrical potential. The
electrical current remains constant.
ORP voltageThe ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface
(e. g. a gold or platinum surface).
ResetRestoring the original condition of all settings of a measuring system.
Stability controlFunction to control the measured value stability.
Standard solutionThe standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Zero point of the pH electrode . . . . . . . 22, 29
57
ListsLab 850
58
ba75553e0301/2006
Lab 850
Appendix: Firmware update
General informationWith the Update_Labxxx_MxxxP program and a PC you can update
the firmware of the Lab 850 to the newest version.
For the update you need:
! a free RS232 interface on your PC
! the RS232 cable, Z390.
Program installationWith the installation program,
"Install_Update_Labxxx_MxxxP_Vx_yy_English.exe", install the
firmware update program on your PC.
Program startStart the "Update_Labxxx_MxxxP" program from the Windows start
menu.
You can change the language via the language menu.
Firmware updateProceed as follows:
1Connect the Lab 850 to a serial interface (COM port) of the PC
with the aid of the interface cable Z390.
2Make sure the Lab 850 is switched on.
3To start the update process click the OK button.
The program automatically recognizes the used interface.
4To go on, follow the instructions of the program.
During the programming process, a corresponding message
and a progress bar (in %) appear.
The programming process takes approx. two minutes.
A terminatory message is displayed after a successful
programming process. The firmware update is now completed.
5Disconnect the meter from the PC.
The instrument is ready for operation.
After switching the meter off and on you can check whether the meter
has taken over the new software version (see page 45).