4.3Control on the dentist element ........................................................................................................................ 13
4.4Configure for highspeed handpiece ................................................................................................................ 14
4.5Configure for motor ......................................................................................................................................... 15
4.6Control panel on the dentist element - MOTOR CLOCKWISE/COUNTERCLOCKWISE ROTATION ............ 16
4.7Control panel on the dentist element - keys without function .......................................................................... 16
5Putting the system into operation ..................................................................................... 17
5.1Starting an instrument ..................................................................................................................................... 17
5.2Adjusting the driving air pressure .................................................................................................................... 18
6.3Electric motor .................................................................................................................................................. 22
6.4SIROSONIC L scaler ..................................................... ... .. ........................................ .. ................................... 23
6.5SIROSONIC L endodontic scaler .................................................................................................................... 24
7Care and cleaning ............................................................................................................ 25
Sirona Dental Systems GmbH1 Explanation of the symbols used
273,2 °F
134 °C
93 °C
199,4 °F
+
1
40
6
0
8
0
E
N
Operating Instructions Simulation Unit C8+ Cabinet S
1 Explanation of the symbols used
SymbolExplanationSymbolExplanation
ONCan be sterilized in the autoclave
at 134°C/273.2°F
OFFCan be thermodisinfected in the
thermodisinfector at 93°C/
199.4°F
CAUTION3-way syringe – water
Backup3-way syringe – air
Clean with water3-way syringe – spray
ЗЙмнлЕЬ
Caution!
General warning
Can be set only by the service
engineer
Read the documentation
Wrongor
RightCW rotation of electric micromo-
tor
Maximum loadCCW rotation of electric micro-
motor
WrongPress button / switch
Factory settingsProtective earth conductor
Speed increase
64 72 695 D3625
D3625.201.01.03.02
Speed decrease
5
2 Warning and safety information Sirona Dental Systems GmbH
Operating Instructions Simulation Unit C8+ Cabinet S
2 Warning and safety information
Intended useThis Simulation Unit C8+ Cabinet S is used
for training purposes. It is a laboratory
workstation that may not be used to treat
living persons!
This unit is not intended for operation in
areas subject to explosion hazards.
General safety informationThe “on-site installation” must have been
performed according to our requirements.
As manufacturers of simulation units and in
the interest of the operational safety of
your system, we stress the importance of
having maintenance and repair of your
simulation unit performed only by ourselves or by agencies expressly authorized by us. Furthermore, safety-critical
system components must always be
replaced with original spare parts upon failure.
We suggest that you request a certificate
showing the nature and extent of the work
performed from those who carry out such
work; it must contain any changes in rated
parameters or working ranges (if applicable), as well as the date, the name of the
company and a signature.
This type of appliance complies with the
requirements of EN 61010/1.
Modifications to this system which might
affect the safety of the system operator or
parties are prohibited by law!
For reasons of product safety, this product
may be operated only with original Sirona
accessories or third-party accessories
expressly approved by Sirona. The user is
responsible for any damage resulting from
the use of non-approved accessories.
If any devices not approved by Sirona are
connected, they must comply with the
applicable standards: EN 61010/1.
64 72 695 D3625
6D3625.201.01.03.02
Sirona Dental Systems GmbH2 Warning and safety information Warning and safety information
i
Operating Instructions Simulation Unit C8+ Cabinet S
Warning and safety information
Water quality complianceA safety unit as specified by EN1717/DIN
1988 must be installed on site.
Please ask your specialist dealer about
the national requirements and relevant
measures to be taken.
Maintenance of the simulation unitDespite the outstanding quality of your
simulation workstation and regular care
by the practice team, it is required in the
interest of operational safety that preventive maintenance be performed at scheduled intervals.
In order to ensure the operational
safety and reliability of your simulation
unit and to avoid damage due to natural wear, as the system owner you
must have your system checked regularly by an authorized service engineer. Furthermore, safety checks must be
performed.
ЗЙмнлЕЬ
Highlighting of warning and safety informationTo avoid personal injury and material dam-
age please also observe the warning and
safety information provided in the present
operating instructions. All such information
is highlighted by the signal words
NOTE, CAUTION or
WARNING.
Symbols usedObserve accompanying docu-
ments (on connection cabinet rating plate)
Ventilation slotsUnder no circumstances may the ventila-
tion slots on the unit be covered, since
otherwise the air circulation will be
obstructed. Do not spray with disinfectants, etc. into ventilation slots.
Electric micromotors in dentist elementIntermittent operation:
1 min. ON – 7 min. OFF with a motor current of 1.5 A.
Wireless phone interference with equipmentTo ensure safe operation of equipment,
the use of mobile wireless phones (cell
phones) must be prohibited within the
vicinity of the system.
64 72 695 D3625
D3625.201.01.03.02
7
2 Warning and safety information Warning and safety informationSirona Dental Systems GmbH
Operating Instructions Simulation Unit C8+ Cabinet S
Warning and safety information
Disposal
On the basis of the Directive 2012/19/EU and the country-specific disposal
regulations on waste electrical and electronic equipment, we would like to
stress that such equipment must be disposed of accordingly within the
European Union (EU). These regulations demand environmentally friendly
recycling / disposal of waste electrical and electronic equipment. These parts
must not be disposed of as domestic refuse. This has been symbolized since
March 24, 2006 amongst other things by the symbol of the crossed out
garbage bin.
Disposal procedure
We feel responsible for our products from the initial idea to their disposal. That
is why we give you the option of taking back our waste electrical and electronic
equipment.
If disposal is required, please proceed as follows:
In Germany
In order to arrange return of the electrical equipment, please send a disposal
request to enretec GmbH. The following options are available for this purpose:
● On the homepage of enretec GmbH (www.enretec.de), click on the
"Return of electronic equipment" button under the "eom" menu item.
● Alternatively, you may also contact enretec GmbH directly.
enretec GmbH
Kanalstraße 17
16727 Velten
Tel.: +49 3304 3919-500
E-mail: eom@enretec.de
As manufacturers, we assume the costs for disposal of waste from electrical
and electronic equipment in accordance with the country-specific disposal
regulations (ElektroG). All disassembly, transport and packaging costs are to
be borne by the owner/operator.
Proper preparation (cleaning/disinfection/sterilization) of the equipment must
be carried out prior to disassembly/disposal.
Any nonpermanently installed equipment will be picked up at its installation
site in the practice. Permanently installed equipment will be picked up
curbside at your address by appointment.
Other countries
The dental dealers would be glad to provide you with country-specific
information.
SIROSONIC L
64 72 695 D3625
8D3625.201.01.03.02
Sirona Dental Systems GmbH3 Technical description Warning and safety information
Operating Instructions Simulation Unit C8+ Cabinet S
3 Technical description
Model designation
Power connection
Nominal current
Main unit fuse
Media
Connection values
Simulation Unit C8+ Cabinet S
230 V AC±10 % 50 / 60 Hz
1.5 A at 230 VAC
T 10 A H, 250 V AC, Order No. 10 77 460
Overvoltage categoryII
Protection classI
Water qualityDrinking water
Water hardness1.4 mmol/l (=8 °dH) - 2.1 mmol/l (=12 °dH)
Air qualityoil-free, dry, non-contaminated and free of
dirt
Power consumptionapprox. 350VA
On-site water pressure2.5 - 6 bar ( ≥ 2.5 l/min.)
On-site air pressure5.5 - 7.5 bar ( ≥ 50 l/min.)
On-site suction air vacuumP ≤ 0.18 bar ( ≥ 500 l/min.)
ЗЙмнлЕЬ
Operating conditions
Transport and storage conditions
Maximum water consumption per unit150 ml/min.
Maximum air consumption per unit70 l/min.
Maximum suction air per unit350 l/min
Intended useFor indoor use
Ambient temperature10°C - 40°C (50°F - 104°F)
Relative humidity30% - 75%
Air pressure500 hPa - 1060 hPa
Installation site≤ 3000 m above sea level.
This simulation unit is not suitable for operation in areas subject to explosion hazards.
Degree of contamination2 (acc. to IEC 60664-1)
Temperature-40°C - +70°C (-40°F - +158°F)
Relative humidity10% - 95%
Air pressure500 hPa - 1060 hPa
64 72 695 D3625
D3625.201.01.03.02
9