Parr Instrument 6200 User Manual

5 (1)
Operating Instruction Manual
9
No. 442M
6200
Oxygen Bomb Calorimete
p
6200Calorimeter O
erating Instruction Manual

TABLE OF CONTENTS

Preface I
Scope............................................................................................................I
Explanation of Symbols.................................................................................II
Safety Information.........................................................................................II
Intended Usage.............................................................................................III
Cleaning & Maintenance...............................................................................III
General Specifications..................................................................................III
Environmental Conditions.............................................................................III
Provisions for Lifting and Carrying................................................................III
Getting Started..............................................................................................IV
Concept of Operation 1-1
Overview..................................................................................................... 1-1
Dynamic Operation ..................................................................................... 1-2
Full Microprocessor Based Process Control............................................... 1-2
Full Microprocessor Based Data Acquisition and Handling......................... 1-2
Flexible Programming ................................................................................. 1-2
Installation 2-1
Filling the Jacket Reservoir......................................................................... 2-1
Power Connection....................................................................................... 2-1
Jacket Cooling Water Connection............................................................... 2-1
Oxygen Filling Connection.......................................................................... 2-2
Printer and Balance Connections................................................................ 2-2
Swagelock Tube Fittings............................................................................. 2-3
Retightening Swagelock Tube Fittings........................................................ 2-3
6200 External Plumbing.............................................................................. 2-5
Standardizing the Calorimeter..................................................................... 2-7
Instrument Description 3-1
Types of Controls........................................................................................ 3-1
Menu Keys.................................................................................................. 3-1
Control Keys................................................................................................ 3-1
Program Installation & Control 4-1
Software Installation.................................................................................... 4-1
Default Settings........................................................................................... 4-1
Revising Default Settings............................................................................ 4-2
Factory Default Settings.............................................................................. 4-3
3
p
TABLE OF CONTENTS
Operating Instructions 5-1
Operating the 1108 Oxygen Bomb.............................................................. 5-1
Operating the Filling Connection................................................................. 5-1
Operating the Calorimeter........................................................................... 5-2
Samples and Sample Holders..................................................................... 5-5
Combustion Aids......................................................................................... 5-5
Oxygen Charging Pressure......................................................................... 5-5
Combustion Capsules................................................................................. 5-6
Foodstuffs and Cellulosic Materials ............................................................ 5-6
Coarse Samples.......................................................................................... 5-6
Corrosive Samples...................................................................................... 5-7
Explosives and High Energy Fuels ............................................................. 5-7
Volatile Sample Holders.............................................................................. 5-7
Poor Combustion ........................................................................................ 5-8
Corrections & Final Reports 6-1
Entering Corrections and Obtaining the Final Report ................................. 6-1
Manual Entry............................................................................................... 6-1
Fixed Corrections........................................................................................ 6-1
Reporting Instructions 7-1
Report Option Selection.............................................................................. 7-1
Report Generation....................................................................................... 7-1
File Management 8-1
Clearing Memory......................................................................................... 8-1
Removable Compact Flash Memory........................................................... 8-1
Maintenance & Troubleshooting 9-1
Oxygen Bomb ............................................................................................. 9-1
Jacket Temperature Troubleshooting ......................................................... 9-1
Fuses .......................................................................................................... 9-1
6200 Calorimeter Error List......................................................................... 9-2
Appendix A - Menu Operating Instructions A-1
Calorimeter Operation Menu....................................................................... A-1
Temperature vs. Time Plot Screen ............................................................. A-2
Temperature Plot Setup Menu.................................................................... A-2
Operating Controls Menu............................................................................ A-4
Program Information and Control Menu...................................................... A-5
Calibration and Data Controls Menu........................................................... A-6
Thermochemical Calculations Menu........................................................... A-8
Data Entry Controls Menu........................................................................... A-10
Reporting Controls Menu............................................................................ A-11
Communication Controls Menu.................................................................. A-12
File Management Menu ............................................................................. A-14
Diagnostics Menu....................................................................................... A-15
6200Calorimeter O
erating Instruction Manual
4
p
6200Calorimeter O
erating Instruction Manual
TABLE OF CONTENTS
Appendix B – Calculations B-1
Calculating the Heat of Combustion............................................................ B-1
General Calculations................................................................................... B-1
Appendix C- Standardization C-1
Appendix D - Communication Interfaces D-1
Appendix E - Technical Service E-1
Appendix F – 6200 Calorimeter Parts & Diagrams F-1
Thermochemical Corrections...................................................................... B-2
ASTM and ISO Methods Differ.................................................................... B-3
Acid and Sulfur Corrections ........................................................................ B-4
ASTM Treatment for Acid and Sulfur .......................................................... B-6
ISO Calculations ......................................................................................... B-6
Spiking Samples ......................................................................................... B-7
Conversion to Other Moisture Bases .......................................................... B-7
Conversion to Net Heat of Combustion....................................................... B-7
Standardizing the Calorimeter..................................................................... C-1
Standard Materials...................................................................................... C-1
Automatic Statistical Calculations............................................................... C-2
Printer Port.................................................................................................. D-1
Balance Port................................................................................................ D-1
Mettler 011/012 Balance Interface.............................................................. D-1
Sartorius Balance Interface......................................................................... D-2
Generic Interface......................................................................................... D-3
Run Date Template..................................................................................... D-4
Ethernet Interface........................................................................................ D-5
Bar Code Port ............................................................................................. D-10
Network Data Devices................................................................................. D-10
Return for Repair......................................................................................... E-1
Principal Assemblies in Calorimeter............................................................ F-1
Parts List for A1279DD Controller Assembly .............................................. F-1
Parts List for Temperature Control Assembly ............................................. F-2
Parts List for A1284DD2 Stirrer Hub Assembly .......................................... F-2
Spare and Installation Parts List ................................................................. F-2
Parts List for Water Tank Assembly............................................................ F-3
Parts List for Cooling Water Supply............................................................ F-3
Parts List for Oxygen Filling System........................................................... F-3
Parts List for 6200 Stirrer Motor and Drive.................................................. F-3
F-1 6200 Oxygen Bomb Calorimeter Cutaway Front ............................. F-4
F-2 6200 Oxygen Bomb Calorimeter Cutaway Rear.............................. F-5
F-3 A1279DD Control Schematic........................................................... F-6
F-4 Oxygen Solenoid Assembly............................................................. F-7
F-5 Water Tank Assembly...................................................................... F-8
F-6 A1311DD Circulating Pump Assembly ............................................ F-9
F-7 Temperature Control Assembly.......................................................F-10
F-8 A1284DD2 Stirrer Hub Assembly ....................................................F-11
F-9 Stirrer Motor Assembly ....................................................................F-12
5
p
TABLE OF CONTENTS
Figures
2-1 6200 Calorimeter Back Panel ........................................................... 2-4
2-2 Closed Loop Configuration with 6510............................................... 2-5
2-3 Closed Loop Configuration with 1563............................................... 2-5
2-4 Open Loop Configuration with Tap Water......................................... 2-6
2-5 Open Loop Configuration with 1552 ................................................. 2-6
4-1 6200 Factory Default Settings........................................................... 4-3
5-7 Combustion Capsule with Adhesive Tape Seal............................... 5-7
F-1 6200 Oxygen Bomb Calorimeter Cutaway Front ............................. F-4
F-2 6200 Oxygen Bomb Calorimeter Cutaway Rear.............................. F-5
F-3 A1279DD Control Schematic........................................................... F-6
F-4 Oxygen Solenoid Assembly............................................................. F-7
F-5 Water Tank Assembly...................................................................... F-8
F-6 A1311DD Circulating Pump Assembly ............................................ F-9
F-7 Temperature Control Assembly.......................................................F-10
F-8 A1284DD2 Stirrer Hub Assembly ....................................................F-11
F-9 Stirrer Motor Assembly ....................................................................F-12
Tables
B-1 Settings for ISO & BSI Methods......................................................... B-5
C-1 Calorimeter Control Limits.................................................................. C-2
D-1 6200 Calorimeter Serial Ports Pin-Out............................................... D-3
D-2 Calorimeter Run Data Template......................................................... D-4
6200Calorimeter O
erating Instruction Manual
6
p
6200Calorimeter O
erating Instruction Manual

PREFACE

This manual contains instructions for installing and operating the Parr 6200 Calorimeter. For ease of use, the manual is divided into nine chapters.
Concept of Operation Installation Instrument Description Program Installation & Control Operating Instructions Corrections & Final Reports Reporting Instructions File Management Maintenance & Troubleshooting
Subsections of these chapters are
identified in the Table of Contents. To assure successful installation and
operation, the user must study all instructions carefully before starting to use the calorimeter to obtain an understanding of the capabilities of the equipment and the safety precautions to be observed in the operation.
Customer Service:
Questions concerning the installation or operation of this instrument can be answered by the Parr Customer Service Department: 309-762-7716 800-872-7720 Fax: 309-762-9453 www.parrinst.com parr@parrinst.com
Additional instructions concerning the
installation and operation of various
component parts and peripheral items used with the 6200 Calorimeter have been included and made a part of these instructions.
No. Description 201M Limited Warranty 483M Introduction to Bomb Calorimetry 205M 1108 Oxygen Combustion Bomb 207M Analytical Methods for Oxygen Bombs 230M Safety Precautions
Additional instructions for the printer, cooler, and water handling systems are found in the respective package and should be made a part of this book.
Note: The unit of heat used in this manual is the international calorie, which is equal to 4.1868 absolute joules
.
Scope
I
7
p
PREFACE
Explanation of Symbols
6200Calorimeter O
erating Instruction Manual
I
O
~
On position
Off position
Alternating Current (AC)
This CAUTION symbol may be present on the Product Instrumentation and literature. If present on the product, the user must consult the appropriate part of the accompanying product literature for more information.
ATTENTION, Electrostatic Discharge (ESD) hazards. Observe precautions for handling electrostatic sensitive devices.
Protective Earth (PE) terminal. Provided for connection of the protective earth (green or green/yellow) supply system conductor.
Chassis Ground. Identifies a connection to the chassis or frame of the equipment shall be bonded to Protective Earth at the source of supply in accordance with national and local electrical code requirements.
Earth Ground. Functional earth connection. NOTE: This connection shall be bonded to Protective earth at the source of supply in accordance with national and local electrical code requirements.
II
8
p
6200Calorimeter O
erating Instruction Manual
PREFACE
Safety Information
To avoid electrical shock, always:
1. Use a properly grounded electrical outlet of correct voltage and current handling capability.
2. Ensure that the equipment is connected to electrical service according to local national electrical codes. Failure to properly connect may create a fire or shock hazard.
3. For continued protection against possible hazard, replace fuses with same type and rating of fuse.
4. Disconnect from the power supply before maintenance or servicing.
To avoid personal injury:
1. Do not use in the presence of flammable or combustible materials; fire or explosion may result. This device contains components which may ignite such material.
2. Refer servicing to qualified personnel.
Intended Usage
If the instrument is used in a manner not specified by Parr Instrument Company, the protection provided by the equipment may be impaired.
Cleaning & Maintenance
Periodic cleaning may be performed on the exterior surfaces of the instrument with a lightly dampened cloth containing mild soap solution. All power should be disconnected when cleaning the instrument.
There are no user serviceable parts inside the product other than what is specifically called out and discussed in this manual. Advanced troubleshooting instructions beyond the scope of this manual can be obtained by calling Parr Instrument Company in order to determine which part(s) may be replaced or serviced.
General Specifications
Electrical Ratings
120VAC, 6.0 Amps. 50/60 Hz 240VAC, 3.0 Amps, 50/60 Hz
Before connecting the calorimeter to an electrical outlet, the user must be certain that the electrical outlet has an earth ground connection and that the line, load and other characteristics of the installation do not exceed the following limits:
Voltage: Fluctuations in the line voltage should not exceed 10% of the rated nominal voltage shown on the data plate.
Frequency: Calorimeters can be operated from either a 50 or 60 Hertz power supply without affecting their operation or calibration.
Current: The total current drawn should not exceed the rating shown on the data plate on the calorimeter by more than 10 percent.
Environmental Conditions
Operating: 15ºC to 30ºC; maximum relative humidity of 80% non-condensing. Installation Category II (overvoltage) in accordance with IEC 664. Pollution degree 2 in accordance with IEC 664. Altitude Limit: 2,000 meters. Storage: -25ºC and 65ºC; 10% to 85% relative humidity.
Provisions for Lifting and Carrying
Before moving the instrument, disconnect all connections from the rear of the apparatus. Lift the instrument by grabbing underneath each corner.
III
9
p
PREFACE
6200Calorimeter O
erating Instruction Manual
Getting Started
These steps are offered to help the user
become familiar with, install, operate and develop the full capabilities of the Parr 6200 Calorimeter.
1. Review the Concept of Operations,
Chapter 1, to get an understanding of the overall capabilities of the calorimeter and microprocessor control.
2. Unpack and install the calorimeter in
accordance with the Installation Instructions, Chapter 2. This simple, step-wise procedure will acquaint the user with the various parts of the calorimeter and make it easier to understand the operating instructions which follow.
3. Turn the power switch ON (located on
the back). Turn to the Instrument Description, Chapter 3, to review the touch screen controls.
4. Review the Program Installation and
Control, Chapter 4, to match the factory settings to the intended mode of operation. Any required changes can be made to the program parameters located in the Main Menu.
5. Review the Reporting Instructions,
Chapter 7, to become familiar with the manner in which calorimetry corrections are entered. Also discussed are generating final reports, editing and clearing memory.
6. Turn to the Menu Operating Instructions,
Appendix A, to review the menu
functions used to modify the program contained in the 6200 Calorimeter. A review of the menus will provide a good idea of the capabilities and flexibility designed into this instrument.
7. Review the Calculations, Appendix B. This provides information about calculations performed by the 6200 Calorimeter.
8. Review Standardization, Appendix C. This will serve two important functions. First, it provides instructions on generating the energy equivalent factor required to calculate the heat of combustion of unknown samples. Secondly, it will give the user the opportunity to run tests on a material with a known heat of combustion to become familiar with the instrument and confirm that the instrument and operating procedures are producing results with acceptable precision. Most 6200 Calorimeters will have an energy equivalent of approximately 2400 calories per ºC. The runs for standardization and determinations are identical, except for the setting of the instrument to the standardization or determination mode.
9. Review the Communication Interfacing, Appendix D, for the correct installation of any peripherals connected to the 6200 Calorimeter.
10. After successful standardization, the 6200 Calorimeter should be ready for testing samples.
IV
10
p
6200Calorimeter O
erating Instruction Manual

1 CONCEPT OF OPERATION

The 6200 Calorimeter has been designed to provide the user with:
A traditional design calorimeter with
removable oxygen bomb and bucket.
A moderately priced calorimeter which uses real time temperature measurements to determine heat leaks using a controlled calorimeter jacket.
A high precision calorimeter capable of exceeding the repeatability and reproducibility requirements of all international standard test methods.
A compact calorimeter requiring minimum laboratory bench space.
A modern intuitive graphical user interface for ease of operation and
training.
A calorimeter with up to date digital hardware, software and communications capabilities.
A calorimeter that is cost effective and which can incorporate a user’s current bombs, buckets, and accessories.
Removable Bomb
The Model 6200 calorimeter utilizes the Parr 1108 oxygen bomb. More than 20,000 of these reliable oxygen combustion bombs have been placed in service on a world wide basis. This bomb features an automatic
inlet check valve and an adjustable needle valve for controlled release of residual gasses following combustion. They are intended for samples ranging from 0.6 to 1.2 grams with a maximum energy release of 8000 calories per charge.
The 1108 oxygen bomb is made of high­strength, high nickel stainless steel designed to resist the corrosive acids produced in routine fuel testing. An alternative 1108CL bomb is available, constructed of an alloy containing additional cobalt and molybdenum to resist the corrosive conditions produced when burning samples containing chlorinated compounds.
The Model 6200 can also be equipped with a variety of special purpose oxygen bombs for unusual samples and/or applications. The 1104 high strength oxygen bomb is designed for testing explosives and other potentially hazardous materials. The 1109/1109A semimicro oxygen bombs can be fitted along with its unique bucket to test samples ranging from 25 to 200 mg.
Removable Bucket
The A391DD removable bucket has been designed to hold the bomb, stirrer and thermistor with a minimum volume of water and to provide an effective circulating system which will bring the calorimeter to rapid thermal equilibrium both before and after firing.
Overview
11
1-1
p
g
6200Calorimeter O
erating Instruction Manual
1
Dynamic Operation
curve matching technique to compare the temperature rise with a known thermal curve to extrapolate the final temperature rise without actually waiting for it to develop.
Full Micro­processor Based Process Control
A. Generate all temperature readings in B. Monitor jacket as well as bucket C. Confirm equilibrium conditions.
D. Fire the bomb. E. Confirm that ignition has occurred.
Full Micro­processor Based Data Acquisition and Handling
Flexible
ramming
Pro
user to customize the operation of the calorimeter to meet a wide variety of operating conditions including:
A. A large selection of printing options. B. Choice of accessories and peripheral
C. Multiple options in regard to handling D. Choice of ASTM or ISO correction E. A variety of memory management and F. Complete freedom for
In its Dynamic Operating Mode, the calorimeter uses a sophisticated
The microprocessor controller in this calorimeter has been preprogrammed to automatically prompt the user for all required data and control input and to:
the calorimeter. temperature.
In addition to its process control functions, the microprocessor in the calorimeter has been preprogrammed to:
The fifth generation software
built into this calorimeter and accessed through the screen menus permit the
equipment. thermochemical corrections. procedures. reporting procedures. reagent concentrations and
calculations.
Repeated testing, and over 20 years of routine use in fuel laboratories, has demonstrated that this technique can cut the time required for a test by one-half without significantly affecting the precision of the calorimeter.
F. Determine and apply all necessary
heat leak corrections.
G. Perform all curve matching and
extrapolations required for dynamic operation.
H. Terminate the test when it is
complete.
I. Monitor the conditions within the
calorimeter and report to the user whenever a sensor or operating condition is out of normal ranges.
A. Collect and store all required
test data.
B. Apply all required corrections for
combustion characteristics.
C. Compute and report the heat of
combustion for the sample.
G. Unlimited choice of reporting
units.
H. Automatic bomb usage
monitoring and reporting.
I. A choice of Equilibrium or
Dynamic test methods.
J. Automatic statistical treatment of
calibration runs.
K. Enhanced testing and trouble
shooting procedure.
The 6200 Calorimeter is equipped with two RS232C connections for direct communication with its printer and an attached balance. It is also equipped with an Ethernet network connection for connections to laboratory computers.
1-2
12
p
r
r
r
6200Calorimeter O
erating Instruction Manual
2 INSTALLATION
The 6200 Calorimeter is completely assembled and given a thorough test before it is shipped from the factory. If the user follows these instructions, installation of the calorimeter should be completed with little or no difficulty. If the factory settings are not disturbed, only minor adjustments will be needed to adapt the calorimeter to operating conditions in the user’s laboratory.
The water reservoir of the calorimeter must be filled with approximately 1.4 liters of water (distilled or de-ionized preferred). This must be done prior to turning on the heater and the pump. The reservoir is filled through
Plug the power line into any grounded outlet providing proper voltage that matches the specification on the nameplate of the calorimeter. The calorimeter will draw approximately 300 watts of power. Grounding is very important not only as a safety measure, but also to ensure satisfactory controller performance. If there is any question about the reliability of the ground connection through the power cord, run a separate earth ground wire to the controller chassis.
It becomes necessary to use the jacket cooling water connection only if the calorimeter operating room temperature exceeds 24 °C (75°F).
When required, an external water source is used to cool the jacket of the 6200 Calorimeter. This is done in either of the following ways:
1. Tap water is used for cooling and then run to a drain.
2. Cooling water is re-circulated to the calorimeter from a Parr water handling system.
The water that provides the cooling goes through a heat exchanger and does not mix with the water in the jacket and its reservoir. There is a very low cooling load and tap water up to a temperature of 27°C should be adequate.
This apparatus is to be used indoors. It requires at least 4 square feet of workspace on a sturdy bench or table in a well-ventilated area with convenient access to an electric outlet, running water and a drain. The supply voltage must be within ± 10% of marked nominal voltage on the apparatus. The supply voltage receptacle must have an earth ground connection.
the tank fill elbow on the back of the calorimeter. The tank is full once water stands in the horizontal run of the filling elbow.
Turn the power switch to the on position. After a short time, the Parr logo will appear on the LCD display followed by a running description of the instrument boot sequence. When the boot sequence is complete, the calorimeter Main Menu is displayed. Go to the Calorimeter Operation page and turn the heater and pump on. This begins circulating and heating the calorimeter jacket water. Add water to the filling elbow at the rear of the instrument as required in order to keep it full.
Filling the
Reservoi
Connection
Tap Water Cooling
Connect the tap water supply to the cold water inlet on the back of the calorimeter using either ¼” copper or nylon tubing (HJ0025TB035). A 196VB metering valve is provided with the calorimeter. This valve should be installed in this inlet line near the calorimeter. This valve is used to adjust the flow of water to the heat exchanger to compensate for differences in tap water temperatures and water line pressures. Once the calorimeter is operating at equilibrium, check the jacket temperature that is displayed on the operating page. If this temperature is cycling significantly, close down on the metering valve to reduce the flow of cooling water. If the jacket rises above its 30 °C set point, open this valve to increase the cooling. A flow rate of 100 ml / minute is generally all that is required.
Connection
13
Jacket
Powe
Jacket
Cooling
Wate
2-1
p
2
6200Calorimeter O
erating Instruction Manual
Jacket Cooling Water Connection
Continued
calorimeter will not be used for an extended period.
Oxygen Filling Connection
consists of an oxygen pressure regulator with a relief valve that mounts on an oxygen tank and a controlled solenoid inside the calorimeter. To install the regulator on the oxygen supply tank, unscrew the protecting cap from the oxygen tank and inspect the threads on the tank outlet to be sure they are clean and in good condition. Place the ball end of the regulator in the outlet and draw up the union nut tightly, keeping the gages tilted slightly back from an upright position. Connect the regulator to the oxygen inlet fitting on the back of the calorimeter case. This hose should be routed so that it will not kink or come in contact with any hot surface. Connect the high-pressure nylon hose with the push on connector to the back of the calorimeter.
Printer and Balance Connections
Connect the cooling water outlet on the back of the calorimeter to a drain using either nylon (HJ0025TB035) or copper ¼” tubing. A shut off valve in tap water supply line is also a good idea if the
The 6200 Calorimeter is equipped with an automatic bomb oxygen filling system. This system
Connect the printer to the calorimeter at this time. The Parr 1757 Printer is configured and furnished with a cord to connect
Cooling with the Water Handling System
If the calorimeter is to be operated with a Parr Water Handling System, connect the pump output to the cooling water inlet and connect the cooling water outlet to the return connection on the water handling system. With this installation it is neither necessary nor desirable to install the 196VB metering valve in the inlet line. It is a good idea to keep all water line runs as short as practical to avoid unwanted temperature changes in the water between the source and the calorimeter.
All connections should be checked for leaks. Any leaks detected must be corrected before proceeding. Instructions for operating the filling connection are in the Operating Instructions chapter.
Adjust the pressure regulator to deliver 450 psi of O without a charge and attach the filling hose to the bomb inlet valve. Press the O on the Calorimeter Operation page and observe the delivery pressure on the 0 – 600 psi gage while the oxygen is flowing into the bomb. Adjust the regulator, if needed, to bring the pressure to 450 psi. If there is any doubt about the setting, release the gas from the bomb and run a second check.
directly to the RS232C printer port on the back of the calorimeter.
The balance port connection, if used, should be made at this.
. Assemble the oxygen bomb
2
fill key
2
2-2
14
p
2
When Swagelok Tube Fittings are used, the instructions for installation are:
1. Simply insert the tubing into the Swagelok Tube Fitting. Make sure that the tubing rests firmly on the shoulder of the fitting and that the nut is finger-tight.
2. Before tightening the Swagelok nut, scribe the nut at the 6 o’clock position.
3. While holding the fitting body steady with a back-up wrench, tighten the nut 1-1/4 turns. Watch the scribe mark, make one complete revolution and continue to the 9 o’clock position.
4. For 3/16" and 4mm or smaller tube fittings, tighten the Swagelok nut 3/4 turns from finger-tight.
Swagelok tubing connections can be disconnected and retightened many times. The same reliable leak-proof seal can be obtained every time the connection is remade using the simple two-step procedure.
1. Insert the tubing with pre-swaged ferrules into the fitting body until the front ferrule seats.
2. Tighten the nut by hand. Rotate the nut to the original position with a wrench. An increase in resistance will be encountered at the original position. Then tighten slightly with a wrench. Smaller tube sizes (up to 3/16” or 4mm) take less tightening to reach the original position than larger tube sizes. The type of tubing and the wall
6200Calorimeter O
thickness also has an effect on the amount of tightening required. Plastic tubing requires a minimal amount of additional tightening while heavy wall metal tubing may require somewhat more tightening. In general, the nut only needs to be tightened about 1/8 turn beyond finger tight where the ferrule seats in order to obtain a tight seal.
Over tightening the nut should be avoided. Over tightening the nut causes distortion
(flaring) of the lip of the tube fitting where the ferrule seats. This in turn causes the threaded portion of the body to deform. It becomes difficult to tighten the nut by hand during a subsequent re-tightening when the fitting body becomes distorted in this manner.
erating Instruction Manual
Retightening
Tube Fittings
Swagelok
Tube
Fittings
Swagelok
2-3
15
p
6200Calorimeter O
erating Instruction Manual
2
6200 Calorimeter Back Panel
Figure 2-1
2-4
16
p
2
r
l
g
6200Calorimeter O
erating Instruction Manual
Line 1 & 2 – Maximum length of 10 feet
-¼” OD, Polyurethane (Part Number HJ0025TB035)
Line 3 - Maximum length of 25 feet
1
-
/8” OD, Nylon (Part Number HX0012TB024)
Closed Loop Configurations
Figure 2-2
6200
Calorimete
Externa
Plumbin
Figure 2-3
2-5
17
p
6200Calorimeter O
erating Instruction Manual
Open Loop Configurations
Figure 2-4
2
Figure 2-5
2-6
18
p
r
2
6200Calorimeter O
erating Instruction Manual
The calorimeter must be accurately standardized prior to actually performing calorimetric tests on sample materials. Review Appendix C - Standardization, in order to become familiar with the general procedure and calculations. The user should configure the calorimeter at this time to accommodate the desired sample weight entry mode. The calorimeter can be placed into standardization mode on the Calorimeter Operation Page, with the operating mode key. If two bombs and buckets are being used with the calorimeter to maximize sample throughput, the calorimeter can be configured to prompt for a Bomb ID at the start of each test. The Bomb ID can also be selected on the
Calorimeter Operations Page, using the Bomb key. All bomb and bucket combinations will need to be standardized separately. The end result of a standardization test is an energy equivalent value, or the amount of energy required to raise the temperature of the calorimeter one degree. Repeated standardization with any given bomb and bucket combination should yield an energy equivalent value with a range of 14 calories per degree, centered around the mean value for all tests using that bomb and bucket combination. The calorimeter is ready for testing samples after an energy equivalent value has been obtained.
Standardizing
the Calorimete
2-7
19
p
This page intentionally left blank.
6200Calorimeter O
erating Instruction Manual
2
2-8
20
p
6200Calorimeter O
erating Instruction Manual
3 INSTRUMENT DESCRIPTION
All calorimeter configurations and operations are handled by a menu-driven system operated from the bright touch screen display. The settings and controls are organized into nine main sections or pages which comprise the MAIN MENU.
Note:
Keys with a “double box” in the upper left hand corner lead to sub-menus.
The controls that change the data field information in the menus will be one of the following:
1. Toggles. These data fields contain
ON/OFF or YES/NO choices. Simply touching the key on the screen toggles the choice to the other option. The current setting is displayed in the lower right corner of the key.
2. Option Selection. These data fields
contain a list of options. Touching the key on the screen steps the user through the available choices. The current setting is displayed in the lower right corner of the key.
3. Value Entry Fields. These data fields
are used to enter data into the calorimeter. Touching the key on the screen brings up a sub menu with a key pad or similar screen for entering the required value.
There are five control keys which always appear in the right column of the primary displays. These keys are unavailable when they are gray instead of white.
1. Escape. This key is used to go up one
level in the menu structure.
2. Main Menu. This key is used to return
to the main menu touch screen from anywhere in the menu structure.
Some keys lead to multiple choices. Always clear the current value before entering a new value. Once entered the screen will revert to the previous menu and the new value will be displayed in the lower right corner of the key.
4. Data Displays. Most of these keys display values that have been calculated by the calorimeter and are informational only. Certain ones can be overridden by the user entering a desired value through a sub-menu. The value is displayed in the lower right corner of the key.
Note:
Some keys will respond with an opportunity for the user to confirm the specified action to minimize accidental disruptions to the program and/or stored data.
3. Start. This key is used to start a calorimeter test.
4. Report. This key is used to access the test results stored in the calorimeter, to enter thermochemical corrections and to initiate report on the display, printer or attached computer
5. Help. This key is used to access help screens related to the menu currently displayed on the touch screen.
Menu Keys
Control Keys
Types of Controls
3-1
21
p
This page intentionally left blank.
6200Calorimeter O
erating Instruction Manual
3
3-2
22
p
6200Calorimeter O
erating Instruction Manual
4 PROGRAM INSTALLATION & CONTROL
Software Installation
Default Settings
The program in the 6200 Calorimeter can be extensively modified to tailor the unit to a wide variety of operating conditions, reporting units, laboratory techniques, available accessories and communication modes.
In addition, the calculations, thermochemical corrections and reporting modes can be modified to conform to a number of standard test methods and procedures.
Units are preprogrammed with default settings. See pages 4-3 and 4-4 for a listing of the factory default settings.
These default settings remain in effect until changed by the user. Should the user ever wish to return to the factory default settings, go to the Program Information and Control Menu, then to User/Factory Settings, and then touch Reload Factory Default Settings and YES.
Numerous provisions are included to permit the use of other reagent concentrations, techniques, combustion aids and short cuts appropriate for the user’s work.
Note:
Changes to the program are made by use of the menu structure described in Appendix A of this manual. Any of these items can be individually entered at any time to revise the operating program.
Non-volatile memory is provided to retain any and all operator initiated program changes; even if power is interrupted or the unit is turned off. If the unit experiences an intentional or unintentional “Cold Restart”, the controller will return to its default settings.
4-1
23
p
6200Calorimeter O
erating Instruction Manual
4
The default parameters of the 6200 Calorimeter can be changed to guarantee that the 6200 Calorimeter, when cold restarted, will always be in the desired configuration before beginning a series of tests.
Users who wish to permanently revise their default settings may do so using the following procedure:
Establish the operating parameters to be stored as the user default settings.
Go to the Program Info and Control Menu, User/ Factory Settings, User Setup ID, and enter the desired User Setup ID.
Select Save User Default Settings
To re-load the user default settings, go to the Program Info and Control Page, User/Factory Settings, Re-load User Default Settings, and YES.
Revising
Default
Settings
4-2
24
p
4
6200Calorimeter O
erating Instruction Manual
Factory Default Settings
Calorimeter Operation
Operating Mode Determination Bomb Installed/EE 1/2400.0 Heater and Pump OFF
Operating Controls
Method of Operation Dynamic Reporting Units Btu/lb Use Spiking Correction OFF “OTHER” Multiplier 4.1868 Calibrate Touchscreen LCD Backlight Timeout(s) 1200 S LCD Contrast 30% Print Error Messages ON Language English
Spike Controls
Use Spiking OFF Heat of Combustion of Spike 6318.4 Use Fixed Spike OFF Weight of Fixed Spike 0.0 Prompt for Spike before Weight OFF
Program Info and Control
Date XX/XX/XXXX Time XX:XX Software and Hardware Info Settings Protect OFF User/Factory Settings Feature Key Bomb Type Select User Function Setup Cold Restart
User / Factory Settings
User Setup ID 00000001 Reload Factory Default Settings Reload User Default Settings Save User Default Settings
Figure 4- 1
6200 Factory Default Settings
Calibration Data & Control
Calibration Run Limit 10 EE Max Std Deviation 0.0 Heat of Combustion of Standard 6318.4 Bomb Service Interval 500 Use Bomb 1
Bomb 1 Through 4
EE Value 2400.0 Protected EE Value OFF
Thermochemical Corrections Standardization
Fixed Fuse Correction ON 15.0 Fixed Acid Correction ON 10.0 Fixed Sulfur Correction ON 0.0
Determination
Fixed Fuse Correction ON 15.0 Fixed Acid Correction ON 10.0 Fixed Sulfur Correction OFF 0.0 Calculate Net Heat of Combustion OFF
Calculation Factors
Acid Value is Nitric Acid Only ON Acid Multiplier 0.0709 Sulfur Value is Percent ON Sulfur Multiplier 0.6238 Fuse Multiplier 1.0 Use Offset Correction (ISO) OFF Offset Value 0.0 Fixed Hydrogen OFF 0.0 Hydrogen Multiplier 50.68 Dry Calculation OFF
25
4-3
p
6200Calorimeter O
erating Instruction Manual
4
Data Entry Controls
Prompt for Bomb ID ON Weight Entry Mode Touch Screen Acid Entry Mode Touch Screen Hydrogen Entry Mode Touch Screen Auto Sample ID Controls ON Sample Weight Warning above 2.0 Spike Weight Entry Mode Touch Screen Sulfur Entry Mode Touch Screen Moisture Entry Mode Touch Screen Auto Preweigh Controls ON
Auto Sample ID Controls
Automatic Sample ID ON Automatic Sample ID Increment 1 Automatic Sample ID Number 1
Auto Preweigh Controls
Automatic Preweigh ID ON Automatic Preweigh ID Increment 1 Automatic Preweigh ID Number 1
Reporting Controls
Report Width 40 Automatic Reporting ON Auto Report Destination Printer Individual Printed Reports OFF Edit Final Reports OFF Recalculate Final Reports OFF Use New EE Values in Recalculation OFF
Communication Controls
Printer Port (RS232) Balance Port (RS232) Network Interface
Printer Destination Local (RS232) Bar Code Port (RS232) Network Data Devices
Printer Port Communications
Number of Data Bits 8 Parity None Number of Stop Bits 1 Handshaking Xon/Xoff Baud Rate 9600 Printer Type Parr 1757
Balance Port Communications
Balance Type Custom Customize Balance Settings
Balance Port Settings
Number of Data Bits 8 Parity None Number of Stop Bits 1 Handshaking None Baud Rate 9600 Data Characters from Balance 8 Data Precision 4 Transfer Timeout (seconds) 10 Balance Handler Strings
Data Logger
Data Logger OFF Data Log Interval 12s Data Log DestinationLog File and Printer Select Data Log Items Data Log Format Text Format
4-4
26
p
6200Calorimeter O
erating Instruction Manual
5 OPERATING INSTRUCTIONS
Operating the 1108 Oxygen Bomb
Operating the Filling Connection
calorimeter control panel. The calorimeter will then fill the bomb to the preset pressure and release the residual pressure in the connecting hose at the end of the filling cycle. It will take approximately 60 seconds to fill the bomb. During this time a countdown timer on the O2 fill button will display the remaining fill time. Pushing the O2 key a second time will stop the flow of oxygen at any time. Once the display returns to its normal reading, the user can disconnect the coupling and proceed with the combustion test.
If the charging cycle should be started inadvertently, it can be stopped immediately by pushing the O2 fill key a second time. During extended periods of inactivity,
Detailed Instructions for preparing the sample and charging the 1108 Oxygen Bomb are given in Operating Instructions No.
To fill the bomb, connect the hose to the bomb inlet valve and push the O2 button on the
205M. Follow these instructions carefully, giving particular attention to the precautions to be observed in charging and handling the bomb.
overnight or longer, close the tank valve to prevent leakage. When changing oxygen tanks, close the tank valve and push the O2 FILL key to exhaust the system. Do not use oil or combustible lubricants on this filling system or on any devices handling oxygen under pressure. Keep all threads, fittings, and gaskets clean and in good condition. Replace the two 394HCJE O-rings in the slip connector if the connector fails to maintain a tight seal on the bomb inlet valve.
The recommended filling pressure is 450 psig (3 MPa or 30 bar). This pressure is prescribed by most of the standard bomb calorimetric test methods. Higher or lower filling pressures can be used, but the bomb must never be filled to more than 600 psig (40 atm).
27
5-1
p
r
6200Calorimeter O
erating Instruction Manual
5
All operations required to standardize the 6200 Calorimeter, or test an unknown sample, should proceed step-wise in the following manner:
1. Turn on the calorimeter and activate the pump and heater using Calorimeter Operations. Allow at least 20 minutes for the calorimeter to warm up and the jacket temperature to stabilize. Once the jacket temperature comes within
0.5°C of 30°C and stays there for approximately 10 minutes, the calorimeter is ready to begin testing. The bomb parts should be wetted and then dried in the manner used at the conclusion of a test. This serves to wet all sealing parts as well as leaving the bomb with the same amount of residual water which will exist in all subsequent testing.
2. Prepare the sample weighing the material to 0.1 mg and charge the oxygen bomb as described in the section entitled, Operating the Filling Connection. Using an additional bomb and bucket can increase the throughput of the 6200 Calorimeter. With this arrangement, the calorimeter can operate almost continuously since the operator will be able to empty a bomb and recharge it while a run is in progress. A bomb and bucket for the next run will be ready to go into the calorimeter as soon as it is opened.
Each bomb and bucket combination will have to be standardized separately and the proper energy equivalent for each set must be used when calculating the heat of combustion.
3. Fill the calorimeter bucket by first taring the dry bucket on a solution or trip balance; then add 2000 (+/- 0.5) grams of water. Distilled water is preferred, but demineralized or tap water containing less than 250 ppm of dissolved solids is satisfactory. The bucket water temperature should be approximately 3 to 5 °C below the jacket temperature. It is not necessary to use exactly 2000 grams, but the amount selected must be duplicated within +/- 0.5 gram for each run. Instead of weighing the bucket, it can be filled from an automatic pipet, or from any other volumetric device if the repeatability of the filling system is within +/- 0.5 ml.
To speed and simplify the bucket filling process, and to conserve water and energy, Parr offers a closed circuit Water Handling System (No. 6510). This provides a water supply, cooled to the starting temperature and held in an automatic pipet ready for delivery in the exact amount needed to fill the bucket. Instructions for this automatic system are given in Operating Instruction No. 454M.
Operating the
Calorimete
5-2
28
p
5
6200Calorimeter O
erating Instruction Manual
Operating the Calorimeter
Continued
5. Close the calorimeter cover. This
6. Select determination or
7. The calorimeter will now take over
4. Set the bucket in the calorimeter. Attach the lifting handle (421A) to the two holes in the side of the
screw cap and partially lower the bomb in the water. Handle the bomb carefully during this operation so that the sample will not be disturbed. Push the two ignition lead wires into the terminal sockets on the bomb head. Orient the wires away from the stirrer shaft so they do not become tangled in the stirring mechanism. Lower the bomb completely into water with its feet spanning the circular boss in the bottom of the bucket. Remove the lifting handle and shake any drops of water back into the bucket and check for gas bubbles.
lowers the stirrer and thermistor probe into the bucket.
standardization as appropriate on the Calorimeter Operation menu by toggling the operating mode key. After pressing the start key, the calorimeter will now prompt the operator for Bomb ID number, sample ID number, sample weight and spike weight in accordance with the instructions set into the operating controls page.
and conduct the test. During the time it is establishing the initial equilibrium, it will display PREPERIOD on the status bar. Just before it fires the bomb, it will sound
a series of short beeps to warn the user to move away from the calorimeter. Once the bomb has been fired, the status bar will display POSTPERIOD. The calorimeter will check to make certain that a temperature rise occurs and will then look for the final equilibrium conditions to be met. If it fails to meet either the initial or final equilibrium conditions, or if it fails to detect a temperature rise within the allotted time, the calorimeter will terminate the test and advise the user of the error.
8. At the conclusion of the test, the calorimeter will signal the user.
9. Open the cover and remove the bomb and bucket. Remove the bomb from the bucket and open the knurled valve knob on the bomb head to release the residual gas pressure before attempting to remove the cap. This release should proceed slowly over a period of not less than one minute to avoid entrainment losses. After all pressure has been released, unscrew the cap; lift the head out of the cylinder and examine the interior of the bomb for soot or other evidence of incomplete combustion. If such evidence is found, the test will have to be discarded. Otherwise, wash all interior surfaces of the bomb, including the head, with a jet of distilled water and collect the washings in a beaker.
5-3
29
p
6200Calorimeter O
erating Instruction Manual
5
10. Remove all unburned pieces of fuse wire from the bomb electrodes; straighten them and measure their combined length in centimeters. Subtract this length from the initial length of 10 centimeters and multiply this burned length by 2.3 calories per cm (for Parr 45C10 Fuse Wire) to obtain the fuse correction. The scale on the fuse wire card can be used to obtain this value directly.
11. Titrate the bomb washings with a standard sodium carbonate solution using methyl orange, red or purple indicator. A 0.0709N sodium carbonate solution is recommended for this titration to simplify the calculation. This is prepared by
dissolving 3.76 grams of Na2CO3 in the water and diluting to one liter. NaOH or KOH solutions of the same normality may be used.
12. Analyze the bomb washings to determine the sulfur content of the sample if it exceeds 0.1%. for determining sulfur are discussed in Operating Instructions No. 207M.
13. At the end of the testing period, turn OFF the calorimeter at the power switch.
Methods
5-4
30
Loading...
+ 65 hidden pages