«ICNet 2.0» is the name of the data acquisition and control software for PC-controlled ion chromatographic systems consisting
of Metrohm and Bischoff instruments. The current version is a
true 32-bit application. It was designed specially to operate under Windows 95, Windows 98 and Windows NT and uses all its
benefits.
The «ICNet 2.0» program can be used to create systems for recording and evaluating chromatograms. Time programs can also
be created in which a large number of instrument functions can
be triggered for each program step. It is also possible to use
programmable signals to control external instruments.
1.1 Purpose of program
The operating software meets all the requirements you could
place today on a modern integration software: single or multipoint calibration, internal or external standard, selectable algorithms for non-linear calibration, various integration modes with
integration parameters and integration events, different methods
for peak recognition, peak editor, free scaling, superimposing
several chromatograms, use of sample tables and batch reprocessing; a powerful and GLP-conform report generator with output
interfaces for monitor, printer and external databases.
1.2 Information about the software manual
This 8.110.8193Software Manual provides a comprehensive
overview of the operation of the «IC Net» program. The manual is
organized as follows:
Section 1Introduction
Installation
Section 2Fundamentals of the operation
Program elements and features
IC Net 2.0
Section 3Software settings
Fonts, security system, global settings
Section 4Systems
Creation of new systems, interfaces and devices,
determinations
1
1 Introduction
Section 5Interfaces
Installation, control and parameter settings of interfaces
Section 6Devices
Installation, control and parameter settings of devices
To find the required information, you will find it an advantage to
use either the Table of contents or the Index at the back.
This Software Manual describes only the installation and operation of Metrohm instruments. For details concerning the
Bischoff instruments, please refer to the on-line help in the program and to the Bischoff instruction manuals.
Additional publications about ion chromatographic analyses are
available on request free of charge from your Metrohm agency.
The 8.732.2003 Metrohm Monograph «Ion chromatography»
provides an introduction to the theoretical fundamentals and
general information on separating columns and sample pretreatment.
The 8.732.2013 IC Applications Collection contains all the
Application Notes on the subject of ion chromatography and
can be updated at any time by downloading the latest applications from the Internet under «www.metrohm.ch». Last, but not
least, you will find detailed information on the separating columns
available from Metrohm and on special IC applications in the
relevant "Application Bulletins".
1.3 Notation and pictograms
The following notations are used in this software manual:
Range
SYSTEM STATE
2
Menu item, parameter or entry value
Program window
IC Net 2.0
1.4 Installation
<OK>
Button
[ Ctrl ]Key
Hazard
This symbol draws attention to a possible
danger to life or of injury if the associated
directions are not followed correctly.
Warning
This symbol draws attention to possible damage to instruments or instrument parts if the
associated directions are not followed correctly.
Caution
This symbol marks important information.
First read the associated directions before
you continue.
Comment
This symbol marks additional information and
tips.
1.4 Installation
1.4.1 Hardware requirements for the PC
ComputerPentium II with 200 MHz or higher
Operating systemWindows® 95, Windows® 98, or Win-
Free space on hard disk12 MB for program files
Working memory RAM32 MB for Windows® 95 and Win-
Graphics resolution800×600, better 1024×768 or more
PrinterAny printer supported by the operat-
1.4.2 Software Installation
Insert installation CD into CD drive. Select <Start> and Run. Find
the file setup.exe on the CD and click on <OK>. Click on IC Net in
the selection menu and follow the setup program instructions.
dows® NT
50 MB recommended for data files
dows® 98, 64 MB for Windows® NT
ing system
IC Net 2.0
3
1 Introduction
The two software packages «IC Net 2.0» and «Autodatabase
1.0» will be installed in the desired directory. Icons are created in
the program folder, in the startup folder, and on the desktop. In
addition to the program files of the «IC Net» program, the follow-
ing folders are installed:
DataFolder for storage of chromatogram files (*.chw)
and batch reprocessing files (*.bar) with several
examples.
DevicesFolder for storage of device files.(*.dev).
LogFolder for storage of exception files.(*.exc),
history files (*.hst), and log files (*.log).
MethodsFolder for storage of data acquisition method
files (*.mtw) and sample queue files (*.que) with
several examples.
ReportsFolder for storage of report files (*) and graphic
files (*.wmf).
SystemsFolder for storage of system files (*.smt).
1.4.3 Software deinstallation
Select <Start> / Settings / Control panel. Double-click the Software
icon. Select Metrohm IC Net 2.0 in the list and click on
<Add/remove>. All program files and icons should be removed.
1.4.4 Demo version
If the «IC Net Software 2.0» is installed on a PC without installa-
tion of IC system components, this software can be used as a
demo version which is restricted to the display and recalculation of already recorded chromatograms.
1.4.5 Registration
Please send us your 8.110.8207 Registration card as soon as
possible. Only registered users will get updated program versions at a special price.
4
IC Net 2.0
2.1 Starting/closing the program
2 Fundamentals of the operation
2.1 Starting/closing the program
2.1.1 Start the «IC Net» program
Start the program
Double-click this icon or the ICNet.exe file to start the «IC Net 2.0»
program. The IC Net login window appears:
Enter your password and click on <Log In>.
After software installation, the program can be started without
entering a Password. For the definition of users, see section 3.2.
2.1.2 Close the «IC Net» program
IC NET / File / Exit (Alt+F4)
Exit the «IC Net 2.0» program.
The program is also quit by clicking on in the upper right part
of the main window.
IC Net 2.0
5
2 Fundamentals of the operation
2.2 Glossary
Batch reprocessing
Batch reprocessing is understood to be the subsequent reprocessing of a series of chromatograms which have been loaded in
a batch reprocessing queue. During reprocessing with a selected
method the settings for calibration, integration, passport, appearance and report can be altered at will (for details, see section
10).
Calibration
Calibration is used to describe the method of determining the relationship between the peak height or peak area found for one
component and its concentration in the sample. The result of the
calibration is a calibration function (calibration curve), which
shows the relationship between the amount of sample and the
evaluated quantity.
The determination of the calibration function with reference solutions can be carried out as a one-point or as a multiple-pointcalibration. The calibration method which is mainly used in ion
chromatography is the external standard calibration (absolute
calibration); calibration with an internal standard (relative calibration) or tabulated calibration are also possible (for details,
see section 4).
Chromatogram
A chromatogram is a graphic plot of the elution curve (signal vs.
time) recorded following a chromatographic separation on a
separating column.
Chromatograms are stored as chromatogram files (*.chw) in the
Data directory. As well as the measuring data the chromatogram
files also contain the method parameters and system settings
which have been used for data recording, data processing and
remote control (for details, see section 8).
Determination
In order to carry out a determination a suitable system must be
selected for the separating problem. The result of the determination is a chromatogram, in which the measuring data and results of the determination are stored (for details, see section 8).
Device
The «IC Net» program supports remote control for Metrohm and
Bischoff instruments. Each instrument within a system is a device. Devices are stored as device files (*.dev) in the Devices directory. All Metrohm devices start with Me*.dev and all Bischoff
devices start with Bi*.dev (for details, see section 6).
Integration
Integration is to be understood as being the method for determining the peak area and peak height with the aid of approximate
6
IC Net 2.0
Interface
Method
2.2 Glossary
baselines. The integration algorithm included in the program is
influenced by the integration parameters and the optionally
programmable integration events which are defined in the
method. In addition, the integration can be manually corrected
later with the aid of the peak editor (for details, see section 7).
The «IC Net» program supports remote control for interfaces
available from Metrohm and Bischoff. Interfaces generally convert
analog signals to digital form, which can be handled by the computer. Interfaces are stored as device files (*.dev) in the Devices
directory. All Metrohm interfaces and Metrohm devices start
with Me*.dev; all Bischoffinterfaces and Bischoff devices start
with Bi*.dev (for details, see section 5).
A method contains all information necessary for dataacquisi-tion, integration, peakevaluation and quantification. It can
be considered as the chromatogram template, i.e. chromatogram
without raw data. Methods are stored as method files (*.mtw) in
the Methods directory.
Each system is linked to a method. This method is called proc-essing method and is opened automatically at the start of a new
determination (for details, see section 7).
Sample queue
A sample queue is used for the automated processing of series
of samples, particularly in combination with a sample changer
(for details, see section 9).
System
The combination of Metrohm and Bischoff devices connected to
an interface is called a system. The system includes all the settings of the devices, their time program, the data acquisition parameters and the processing method which have been optimized
for the specific separating column and the determination to be
carried out with it. A system is used to start single determinations
or determinations with the help of a sample queue.
Systems are stored as system files (*.smt) in the Systems directory (for details, see section 4).
Workplace
The workplace contains all interfaces and devices connected
directly to the PC COM ports and all systems connected to the
interfaces. The icons of all these active interfaces, devices and
systems are shown on the toolbar. They can also be displayed in
the Workplace window opened by selecting the Options / Devices
setup menu item (for details, see section 3.3).
IC Net 2.0
7
2 Fundamentals of the operation
2.3 Overview of program windows
«IC Net 2.0» consists of different windows whose functionality is
linked together. The different windows are:
IC NETMain program window with menus for file
administration, printing, method modification, options, login and user rights, window
handling.
CHROMATOGRAMChromatogram window for graphic plot of
running or recorded chromatograms.
SYSTEMSystem window for control of interfaces and
devices.
SYSTEM STATESystem state window for status messages.
WATCH WINDOWWatch window for live display of instrument
values.
QUEUE EDITORQueue editor window for edition of sample
2.4 Main window elements
The elements of the IC NET main window are the menu bar, the
toolbar and the status bar, indicating prompts and logged-in
user.
Menu bar
Toolbar
queue tables and batch reprocessing tables.
Status bar
Logged-in userPrompts, information
8
IC Net 2.0
2.5 Icons of the main window
The following icons are displayed in the IC NET main window:
Open chromatogram
Save chromatogram
Open last batch reprocessing file
Make report
Print preview
Send chromatograms to Autodatabase
Print chromatogram
Passport
2.5 Icons of the main window
Method setup
Integration
Components
Cascade all opened chromatogram windows
Vertical tiling of open chromatogram windows
Horizontal tiling of open chromatogram windows
Appearance
Enable/disable peak editor mode. If the Peak editor
mode is enabled, the peak editor icons appear.
View all
Help
Lock system
IC Net 2.0
762 IC Interface
771 Compact Interface
761 Compact IC
DAD 157
Connected system
Global timer
9
2 Fundamentals of the operation
2.6 Overview of file types
The following file types are produced by the «IC Net» software:
*.adbAutodatabase file
Contains chromatogram and method data of all determinations.
*.barBatch reprocessing file (binary file)
Contains batch reprocessing data. The *.bar file is
stored automatically in the Data folder.
*.calCalibration file (binary file)
Contains calibration data, which can be exported with
IC Net / Method / Calibration / Export calibration.
The *.cal file is stored automatically in the Methods
folder.
*.chwChromatogram file (binary file)
Contains chromatogram, system and method data of a
determination.
The *.chw file is stored automatically in the Data folder.
*.excException file (ASCII file)
Contains exceptions from normal running and error
messages.
The *.exc file is stored automatically in one of the day
subfolders of the Log folder.
*.hstHistory file (ASCII file)
Contains history of commands and program actions.
The *.hst file is stored automatically in one of the day
subfolders of the Log folder.
*.logLog file (ASCII file)
Contains log file of data communication between PC
and instruments.
The *.log file is stored automatically in one of the day
subfolders of the Log folder.
*.mtwMethod file (binary file)
Contains the data acquisition method, which can be
linked to a system.
The *.mtw file is stored automatically in the Methods
folder.
10
*.queSample queue file (binary file)
Contains a sample data table.
The *.que file is stored automatically in the Systems
folder.
IC Net 2.0
*.rttReport file (ASCII file)
Contains a report template.
The *.rtt file is stored in the program folder.
*.smtSystem file (ASCII file)
Contains the system settings.
The *.smt file is stored automatically in the Systems
folder.
*.devDevice file (ASCII file)
Contains drivers for devices.
The *.dev file is stored in the Devices folder.
2.7 Context-sensitive menus
Some of the menu functions of the program windows are also
accessible by clicking on the desired window or item and pressing the right mouse button. The pop up windows have different
contents and functions depending on the selected active window
or item type.
2.7 Context-sensitive menus
2.8 Keyboard and mouse functions
The mouse can be used to carry out the normal program operating functions such as the selection of menu items and fields. It
can additionally be used for magnifying a section of a chromatogram (zooming). To zoom a portion of the plot it is necessary
to place the mouse cursor to the upper left corner of the square
to zoom, press the left mouse button and drag the cursor to the
lower right corner of the rectangle. After releasing of the left
mouse button the selected region will be zoomed full-screen. If
the cursor is active in the peak editor mode then it can be moved
by pressing down the right-hand mouse key.
The keyboard can also be used to scale a chromatogram in the
window, as described below.
Keyboard quick reference
Cursor is inactive:
[ up ] Increases sensitivity on the Y axis.
[ down ] Reduces sensitivity on the Y axis.
[ right ] Expands a chromatogram on the X axis.
IC Net 2.0
[ left ] Shrinks a chromatogram on the X axis.
[ Ctrl ] + [ Home ] Autoscale procedure on the X axis (shows
all on X).
[ Ctrl ] + [ End ] Autoscale procedure on the Y axis (shows
all on Y).
[ PageUp ]Shifts a chromatogram on 1/10 part of a
screen upwards.
11
2 Fundamentals of the operation
[ PageDown ]Shifts a chromatogram on 1/10 part of a
[ Shift ] + [ up ]Increases a distance between channels of
[ Shift ] + [ down ]Reduces a distance between channels of a
[ 0 (Zero) ] Adjusts a zero on the last point of a chro-
Only part of the chromatogram is on screen:
[ Ctrl ] + [ right ] Moves one window right (without change of
[ Ctrl ] + [ left ] Moves one window left (without change of
[ Home ] Shows the beginning of a chromatogram
[ End ] Shows the end of a chromatogram (without
screen downwards.
a chromatogram.
chromatogram.
matogram (running chromatogram) or its
lowest level (finished run).
scale on X and Y axes).
scale on X and Y axes).
(without change of scale on X and Y).
change of scale on X and Y).
2.9 Help
[ 0 (Zero) ] Adjusts a zero on the lowest level in the
window.
Cursor is active:
[ 0 (Zero) ] Adjust a zero in site of the cursor.
[ right ] Moves cursor left to right.
[ Shift ] + [ right ] Quickly moves cursor left to right.
[ left ] Moves cursor to the left.
[ Shift ] + [ left ] Quickly moves cursor to the left.
[ Home ] Moves cursor to beginning of a window.
[ End ] Moves the cursor to end of a window.
[ Shift ] + [ End ] Sets the beginning of a window in site of
the cursor.
[ Shift ] + [ Home ] Sets the end of a window in site of the cur-
sor.
By clicking on , by clicking on , by selecting the
Help / Contents menu item, or by pressing the [ F1 ] key you can
get on-line help on the current topic anywhere in the program.
12
Green textscan be clicked to jump to a different Help
topic.
Violet textsidentify the dialog item, parameter or button
in the corresponding window.
Blue textsidentify important information.
IC Net 2.0
3 Software settings
3.1 Fonts
IC NET / Options / Fonts
This option allows the selection of fonts used by the system.
Font for dialogs...Selection of font used for dialog boxes.
Default setting: MS Sans Serif / Standard / 8 pt.
Font for reports...Selection of font used for report output to the
screen or printer.
Default setting: Courier New / Standard / 10 pt.
3.1 Fonts
Font for tables...Selection of font used for data presentation in
Font for plots...Selection of font used for labels on chroma-
Save fonts configuration
3.2 Security system
The «IC Net» program has a security system based on the list of
users. Every user has his unique password and one of the following access levels:
tables on the screen. It is not used for presentation of tables in the report.
Default setting: MS Sans Serif / Bold / 8 pt.
togram plots and calibration curves.
Default setting: Times New Roman / Bold / 10 pt.
Save chosen font configuration.
If you want to reset the modified font settings to the
initial default settings, delete the fonts.cfg file in the
program path.
IC Net 2.0
NoviceRestricted access to program functions. Al-
lows only start and stop of determinations using existing system and method files and
manual control of the devices. Modifications
of system, method and data files are not allowed.
MasterAccess to all program functions with few ex-
ceptions: the user cannot set Global prefer-
13
3 Software settings
ences, open the Workplace window, change
hardware settings of interfaces and devices,
and change Security options.
AdministratorAccess to all program functions. This level
should be switched on only while installing the
system or in the case of configuration change.
The administrator user is authorized to change
access levels, names and passwords of other
users
It is recommended to make user lists and enter passwords as a
first action after software installation. So, select the menu item IC
NET / Options / Security and click on <Log In> in the IC Net login
window without entering a password. The Security options window
appears:
Enter user name, password and access level for all users. Don't
forget to make one of the user an Administrator, otherwise this
window will not open again. At the end, click on <OK>.
After configuration of the security system the program prompts
for the password every time the system starts. This user name
stamps all methods, chromatograms and reports. It is possible at
any time to change the user with the menu item IC NET / Options /
Lock system.
3.3 Global settings
IC NET / Options / Global preferences
This window is used for global program settings. It can be
opened only by the user with Administrator access level.
14
IC Net 2.0
3.3 Global settings
Data file overwriteGlobal settings for overwriting of chroma-
togram files:
NeverChromatogram files cannot be overwritten.
A modified chromatogram is saved as new
file with the file name number raised by 1.
AskThe user is asked if the chromatogram
should be overwritten.
AlwaysChromatogram files are always overwritten
without confirmation.
If method changedGlobal settings for saving of method files:
Don't ask to saveThe method file is not saved automatically.
It can be saved only with File / Save /
Method.
Ask to saveThe user is asked if the method should be
saved.
Overwrite without asking
Method files are overwritten without confirmation.
If disk method is newer
Global settings for overwriting of method
files:
IC Net 2.0
Ask to overwriteThe user is asked if the method should be
overwritten.
Overwrite without asking
Method files are overwritten without confirmation.
Opening chromatograms
Global settings for opening chromatograms:
15
3 Software settings
Ignore last data directory
If this option is enabled, the directory is
opened where the current selected chromatogram is saved.
If this option is disabled, the directory is
opened where the last opened chromatogram is saved.
GLP OnSwitch on/off global GLP settings. The fol-
lowing parameter settings are automatically
set:
Data file overwrite= Never
If method changed = Don't ask to save
If disk method is newer= Ask to
overwrite
System face
Stacked recorder icons
If this option is enabled, only one recorder
icon with different tabs appears in the system window if several data recorders are
installed.
OK, Apply buttons mean
Save object to fileIf this option is enabled, the <Apply> button
in system settings windows is replaced by
the <Save> button. The system settings are
saved if the <Save> or <OK> button is
clicked.
with file name confirmation
If this option is enabled, the <Apply> button
in system settings windows is replaced by
the <Save as> button. The system settings
can be saved in a new file if the <Save as>
or <OK> button is clicked.
Chromatogram unitsUnits for chromatograms:
FlowUnit for flow rate: µµL/min, mL/min
PressureUnit for pressure: MPa, psi, bar, atm
Print via print spooler Switch on/off printing via print spooler.
Switch off this option if problems with
printing occur.
16
Measuring prioritySetting the priority of program execution.
realtime means that the «IC Net» program
has the highest priority, normal means that
all active programs have the same priority.
<Default colors>Default colors for chromatographic win-
dows (details see section 8.3.1).
IC Net 2.0
4 Systems
The combination of Metrohm and Bischoff devices connected to
an interface is called a system. The system includes all the settings of the devices, their time program, the data acquisition parameters and the processing method which have been optimized
for the specific separating column and the determination to be
carried out with it. A system is used to start single determinations
or determinations with the help of a sample queue. Systems are
stored as system files (*.smt) in the Systems directory.
4.1 System creation
4.1 System creation
4.1.1 System wizard
IC NET / File / New / System
The New system wizard window and a new SYSTEM window are
opened. The New system wizard guides you trough the installation
routine step-by-step to build a new systemsystem. The next step of the
system installation is reached with <Next>, the previous step with
<Back> and the installation is completed with <Finish>.
The first step of the system installation is to enter a name for a
new system folder within the «IC Net» program path
...\ICNet\Systems. For every instrument combination, a new folder
should be created.
IC Net 2.0
After clicking on <Next>, the window for addition of interfaces to
the toolbar appears. This step can be skipped if the desired interface has been already added to the toolbar (see section 5.1.2).
If this is not the case, the interface has to be selected from the
Metrohm Interfaces or Bischoff Interfaces groups. After selecting the
serial port where the interface is connected to and clicking on
17
4 Systems
<Add to workplace>, the interface icon will appear on the toolbar
(PC Boards will not appear on toolbar).
After clicking on <Next>, the window for addition of devices (see
6.1.1) to the system window appears.
For every device to be controlled with «IC Net», open the group
which contains this device, select the device and the Serial port of
the interface where the device is connected to. After clicking on
<Add to system>, the device is added to the SYSTEM window. If
desired, an additional data recorder and a timer of the More mod-
ules group can be added to the system.
18
After clicking on <Next>, the window for selection of processing
method linked to the system and data source appears (details
see section 6.12.2).
IC Net 2.0
4.1 System creation
Watch window
Data recorder
Devices
After clicking on <Finish>, the user is asked to save the new system as a system filesystem file (*.smt) in the new system folder. The new
system is now ready to start a determination (see section 4.3.3).
4.1.2 System window
A system window is opened with IC NET / File / Open / System and
the selection of the desired system file. It contains icons for Datarecorder, Watch window (screen) and all the devices installed
with the system wizard.
4.1.3 Add devices to an existing window
SYSTEM / Setup / New devices/ Install new device
Add new devices to the open SYSTEM window. For every device
to be added to the system, open the group which contains this
device and select the device and the serial port of the interface
where the device is connected to. After clicking on <Add to sys-
tem>, the device is added to the SYSTEM window. If desired, an
additional data recorder and a timer of the More modules group
can be added to the system.
IC Net 2.0
19
4 Systems
4.2 System file handling
The following menu items are used for opening, changing, saving
and closing of systems:
IC NET / File / Open / System
Load an existing system file (*.smt) from the Systems directory
and open the corresponding SYSTEM window. The name of the
folder and of the system file are displayed in the title bar of the
SYSTEM window. A star ( * ) at the end of the name indicates that
the system settings have been changed since the last saving.
SYSTEM / System / Open other
Load an existing system file (*.smt) from the desired system folder
in the Systems directory and open a new SYSTEM window for parameter settings. The old system remains connected.
SYSTEM / System / Change
Open the selected new system in the desired system folder. The
current system is disconnected and closed, and the new system
is connected.
SYSTEM / System / Save
Save the current system file (*.smt) in the desired folder of the
Systems directory.
SYSTEM / System / Close
Close the selected system.
Delete a system
To delete an existing system, delete the *.smt file in the Systems
folder.
4.3 System functions
4.3.1 Connect and disconnect system
To make possible manual control of the system devices and start
of determinations, the selected system must be connected to the
workplace. Systems are connected and disconnected as follows:
20
SYSTEM / Control / Connect to workplace
Connect selected system to the workplace. Only one system from
the same system folder can be connected at the same time.
The icon appears on the toolbar. If this icon is clicked, the
SYSTEM window is always displayed in front of all other windows.
IC Net 2.0
SYSTEM / Control / Disconnect from workplace
Disconnect selected system from the workplace. If a system is
disconnected, manual control is not available for this system, but
all other system settings can be modified and saved. Start deter-
mination and Startup hardware is only possible if no other system
from the same system folder is connected. In this case, the system is automatically connected.
4.3.2 Start/stop hardware and record baseline
SYSTEM / Control / Startup hardware (Measure Baseline)
With this menu item the following actions are executed automatically:
• Send System startup values to the devices.
• Start Metrohm solvent delivery systems and Bischoff solvent
delivery systems.
4.3 System functions
• Start Metrohm additional modules and Bischoff additional
modules.
• Start Bischoff detectors.
• Switch Metrohm Autosamplers and Bischoff Autosamplers to
remote control.
• Start recording of the measurement signal using the method of
the connected system. Independently of the set chromatogram
Duration, the measurement signal is recorded until the data
acquisition is stopped if a new determination is started with
Start determination or by clicking the icon of the chroma-
togram window. In this case the user is asked if the recorded
baseline should be saved or not.
SYSTEM / Control / Shutdown hardware
With this menu item the devices are stopped as follows:
• Stop Metrohm solvent delivery systems and Bischoff solvent
delivery systems.
• Stop Metrohm additional modules and Bischoff additional
modulesVariotherm.
• Stop Bischoff detectors.
IC Net 2.0
• Switch Metrohm Autosamplers and Bischoff Autosamplers to
local control.
• Stop running determination.
• Stop active sample queue.
SYSTEM / Control / Stop data acquisition
Stop recording of the baseline.
21
4 Systems
4.3.3 Start/stop determinations
SYSTEM / Control / Start determination
Start determination using the settings of the selected system.
At this start command, the System startup values are set at the devices. The time programs of the different devices and the data
recording are started either immediately (Start with determination)
or after switching the injection valve to the "Inject" position (Start
with inject) as set in the Start mode window.
SYSTEM / Control / Stop determination
Stop running determination. Data acquisition and time pro-
grams are terminated immediately. The recorded chromatogram
is saved automatically if the Save chromatogram after the run option
on the Passport / Processing tab is enabled.
Alternatively the determination can be stopped by clicking the
icon of the chromatogram window. In this case the user is asked
always if the determination should be saved or not.
If the Auto restart option is enabled or if a sample queue
is active, a new determination is started immediately
after the running determination is stopped. If you want
to avoid this, disable the Auto restart option or Pause the
sample queue.
SYSTEM / Control / Stop data acquisition
Stop data acquisition of the running determination immediately
and save the recorded chromatogram automatically if the Save
chromatogram after the run option on the Passport / Processing tab
is enabled. The time programs of the running determination are
continued normally.
4.3.4 Options for determinations
SYSTEM / Control / Auto restart
If this option is enabled, a new determination is started automatically using the current system after the preceding determination
has been finished normally or stopped manually.
22
The Auto restart option is disabled if determinations are made
with an active sample queue. This option allows to make an infinite batch cycle with the current system
SYSTEM / Control / Verify sample
If this option is enabled, the Edit sample description window is
opened automatically at the start of each determination for entry
of the following sample information:
IC Net 2.0
4.3 System functions
IdentUser defined identifier (title) for the chro-
matogram to be displayed in the title bar of
the chromatogram window and in the
Info 1 / Info 2Sample description.
VolumeInjected volume in µL.
DilutionDilution of the sample.
Vial numberAutosampler vial position to take sample
from.
AmountSample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:
Cs = Cc •Amounts / Amount
Internal standard amount
c
Concentration of the internal standard
component for relative concentration calculations.
Date/time when...Date and time of sample collection (the
default values are equal to the date and
time when the chromatogram starts).
IC Net 2.0
The Verify sample option is disabled if determinations are made
with an active sample queue.
23
4 Systems
4.4 System settings
4.4.1 Modify system window
SYSTEM / Setup / Drag icons
If this option is enabled, the device icons can be resized and
moved in the SYSTEM window and the SYSTEM window itself can
be resized.
To resize an icon or the window, move the cursor to the desired
object until appears. Press the left mouse button and resize
the object to the desired size.
To move an icon, move the cursor to the desired object until
appears. Press the left mouse button and move the object to the
desired place.
4.4.2 Watch window display
SYSTEM / Setup / Watch window
If this option is enabled, the WATCH WINDOW is automatically
opened if a determination or measure baseline is started (see
section 6.2).
4.4.3 Set start mode
SYSTEM / Setup / Start mode
This menu item opens the Start mode window for definition of start
mode for time programs and data acquisition.
This window contains the following two fields:
24
Start with determination
The time program of the loaded object or
the data acquisition (RECORDER) is started
at the moment the determination isstarted.
IC Net 2.0
Start with injectThe time program of the loaded object or
The objects can be moved from one field to the other using the
or buttons.
4.4.4 Install new devices
SYSTEM / Setup / New devices/ Install new device
Add new devices to the open SYSTEM window. For every device
to be added to the system, open the group which contains this
device and select the device and the serial port of the interface
where the device is connected to. After clicking on <Add to sys-
tem>, the device is added to the SYSTEM window. If desired, an
additional data recorder and a timer of the More modules group
can be added to the system.
4.4 System settings
the data acquisition (RECORDER) is started
at the moment the sample is injected.
4.4.5 Install existing devices
SYSTEM / Setup / New devices/ Link to existing device
4.4.6 Print system parameters
SYSTEM / Setup / Parameters / Print
A report of the system parameters including the time program is
created and opened using the «Wordpad» program. The *.txt file
opened can be printed, saved and exported into other programs.
The system report includes the following elements:
STARTUP HARDWARE
Add devices to the open SYSTEM window which
are already present on the toolbar. This is mostly
used to add the 761 Compact IC or 762 IC Interface
events. For every device to be added to the sys-
tem, select the device and click on <OK>.
Name of the method linked to the system
Main measurement channel
Configuration settings for the devices
System startup values
IC Net 2.0
START WITH DETERMINATION
Objects defined in the Start mode window.
The time programs are printed if the programs are enabled.
START WITH INJECTObjects defined in the Start mode window.
The time programs are printed if the programs are enabled.
25
4 Systems
4.4.7 Show system parameters
SYSTEM / Setup / Parameters / Startup hardware
Display of the name of the method linked to the system, the main
measurement channel, the configuration settings for the devices,
and the system startup values.
SYSTEM / Setup / Parameters / Start with determination
Display of the objects defined in the Start mode window. The time
programs are displayed if the programs are enabled.
SYSTEM / Setup / Parameters / Start with inject
Display of the objects defined in the Start mode window. The time
programs are displayed if the programs are enabled.
4.4.8 Links
SYSTEM / Setup / Links
Display of all devices of the system (including watch window and
data recorder).
4.5 System state window
The SYSTEM STATE window is automatically opened if a system is
connected. It shows status and error messages for this system.
Messages concerning a device are followed by ["device name"],
messages concerning the loaded system are followed by ["folder
name"] (name of the folder who contains the system file). The fol-
lowing messages can appear:
4.5.1 Status messages
Checking on-line
On-line
UploadStartupValues
Checking connection between PC and device.
Connection between PC and device ok.
Hardware or system startup values have
been loaded to the device.
26
Initialisation
Ready
Starting
Hardware or system initialization.
Device is ready.
Starting program or chromatogram data
acquisition.
IC Net 2.0
4.5 System state window
Running
Running program
(xxx min left)
Waiting for INJECT
INJECT done
Stopping
Finished
SHUTDOWN
4.5.2 General error messages
Running program or chromatogram data
acquisition.
Running time program with time display.
Waiting for "INJECT" to start program
and/or chromatogram data acquisition as
defined in the Start mode.
Injection valve has been switched to the
"INJECT" position.
Determination has been stopped.
Program or chromatogram data acquisition
has been finished.
System is shutdown.
Detection of hardware failed
Bad connection between PC and device or
device switched off (check connecting cable or switch on instrument).
4.5.3 Error messages for 761 Compact IC
LEAK DETECTED
The leak detector has detected a leak
(check IC system and connections).
Storage of maintenance information failed
(repeat last action; if error reappears, call
Metrohm service).
28
IC Net 2.0
5 Interfaces
The «IC Net» program supports remote control for interfaces
available from Metrohm and Bischoff. Interfaces generally convert
analog signals to digital form, which can be handled by the computer. Interfaces are stored as device files (*.dev) in the Devices
directory. All Metrohm interfaces and Metrohm devices start
with Me*.dev; all Bischoffinterfaces and Bischoff devices start
with Bi*.dev.
5.1 Interface installation
5.1 Interface installation
5.1.1 Add interfaces with system wizard
IC NET / File / New / System
The New system wizard window and a new SYSTEM window are
opened. The New system wizard guides you trough the installation
routine step-by-step to build a new systemsystem. The next step of the
system installation is reached with <Next>, the previous step with
<Back> and the installation is completed with <Finish>.
The first step of the system installation is to enter a name for a
new system folder within the «IC Net» program path
...\ICNet\Systems. For every instrument combination, a new folder
should be created.
IC Net 2.0
After clicking on <Next>, the window for addition of interfaces to
the toolbar appears. The interface has to be selected from the
Metrohm Interfaces or Bischoff Interfaces groups. After selecting the
serial port where the interface is connected to and clicking on
<Add to workplace>, the interface icon will appear on the toolbar
(PC Boards will not appear on toolbar).
29
5 Interfaces
Details to the other steps of the system wizard procedure (installation of devices, processing method, and data source) see
section 4.1.1.
5.1.2 Add interfaces in workplace window
IC NET / Options / Devices setup
Open the WORKPLACE window which shows all interfaces and
devices installed. This window can be opened only by the user
with Administrator access level.
Click on <Install device> to open the Adding interfaces to your work-
place window and proceed as described in section 5.1.1.
5.1.3 Delete interfaces
IC NET / Options / Devices setup
30
Open the WORKPLACE window which shows all interfaces and
devices installed. This window can be opened only by the user
with Administrator access level. Select the desired interface and
click on <Delete device> to delete the interface from the workplace.
IC Net 2.0
5.2 762 IC Interface
5.2.1 762 IC Interface features
The Metrohm 762 IC Interface is available as a one system version (2.762.0010) or a two system version (2.762.0020). Each
system can control a complete ion chromatography or HPLC
equipment and contains the following parts:
• Two 24-bit Analog-to-Digital Converters (often abbreviated
as ADC) to acquire two analog signals from two different detectors simultaneously (for example conductivity detection and
UV detection).
• • One external start input.
• One start button at the interface front panel.
• RS232 communication ports to control up to eight devices
of a complete ion chromatography equipment.
• Seven Event output lines.
The 762 IC Interface is normally connected to a PC COM port
using the 6.2134.100 connecting cable.
5.2 762 IC Interface
For detailed information about the 762 IC Interface see 762 Instructions for Use.
5.2.2 762 icon on the toolbar
The 762 icon is available on the toolbar if a 762 IC Interface has
been installed with the new system wizard or by using the <Install
device> option of the WORKPLACE window (see section 5.1.2). By
clicking this icon the 762 IC Interface window for parameter settings opens.
5.2.3 762 icons in the system window
The 762 System 1and762 System 2 iconsare available in the SYS-
TEM window if the 762 System [1] or 762 System [2] have been in-
stalled with the Install existing devices option (see section 6.1.3).
IC Net 2.0
By clicking this icon using the right mouse button, the following
menu items can be selected:
OpenOpen the 762 System [#] window for programming the
event output lines (this window can also be opened by
double-clicking the icon).
UnlinkDelete the 762 icon from the system.
31
5 Interfaces
5.2.4 762 IC Interface window
762 icon / Open
The 762 IC Interface window for parameter settings is opened by
selecting this menu option with the right mouse button or by
clicking the 762 icon on the toolbar. It consists of three tabs Init,
RS settings, and Links.
Init
The Init tab of the 762 IC Interface window contains data acquisition parameters for the 762 IC Interface.
LabelOptional label to name the interface with
maximum 8 characters.
ROM versionInstrument program version number.
Measuring per second Number of data points measured per s.
Second press on RUN/STOP means 'stop' when measuring
If this option is enabled, the running data
acquisition will be stopped if the
RUN/STOP button on the 762 IC Interface
front panel is pressed. This is equal to Stop
determination.
If this option is disabled, the running data
acquisition will not be stopped if the
RUN/STOP button on the 762 IC Interface
front panel is pressed.
32
IC Net 2.0
<ASK>Read current parameters from 762 IC In-
<SET>Send current parameters to 762 IC Inter-
RS settings
The RS settings tab of the 762 IC Interface window contains RS232
parameters for the 762 IC Interface.
5.2 762 IC Interface
terface.
face.
Baud RateBaud rate of the device.
Selection: 1200...115200
Default value: 115200
ParityParity check.
Selection: None, Even, Odd
Default value: None
HandshakeEnable/disable software handshake mode.
Selection: Swchar, none
Default value: SWchar
FIFO sizeIntermediate memory in byte.
Selection: 1...16
Default value: 16
<ASK>Read current parameters from 762 IC In-
terface.
<SET>Send current parameters to 762 IC Inter-
face.
IC Net 2.0
33
5 Interfaces
Links
The Links tab of the 762 IC Interface is used for COM port selection
and settings.
AliasName of the interface or device.
COM #If this entry is clicked with the right mouse
button, the following menu appears:
Put on desktopPossibility for setting COM port parameters
(details see on-line help).
ChangePossibility for changing the COM port. The
following window is opened, where the
COM port can be changed by clicking on
the desired entry.
5.2.5 762 Event output lines
762 System [#] icon / Open
The 762 System [#] window with programs and settings for the
seven event output lines is opened by selecting this menu option
with the right mouse button or by double-clicking the 762 icon in
the SYSTEM window. It consists of the tabs Program and Events
setup.
34
IC Net 2.0
Time program
On the Program tab of the 762 System [#] window a user-defined
time program for the event output lines of the 762 IC Interface
can be entered. This program is started automatically as defined
in the Start mode window (see section 4.4.3) either at the moment
the determination is started (Start with determination) or at the
moment the sample is injected (Start with inject).
5.2 762 IC Interface
Program
The Program tab contains the two following subpages:
ProgramMain time program with all program steps.
Events configurationPossibility for creation of user-defined re-
mote commands.
On the Program subpage, program steps including time, program
instruction and parameter can be entered.
First columnTime at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruction is applied together with the last instruction with time entry.
Second columnProgram instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the Events configura-
tion tab.
IC Net 2.0
Third columnParameter for program instruction (see
below).
ENABLEDEnable program start (a disabled program
is not started).
35
5 Interfaces
<Add>Add new program instruction.
<Delete>Delete selected program instruction.
<Verify>Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time program on the Program subpage:
InstructionParameter entryMeaning
Events0, 1, p, *Set event output lines 1...7 to the
desired values. For entry of the first
value, enter 1, 0, p or *. For entry of
the other values, move the cursor in
front of the value to be changed
and enter 1, 0, p or *.
ProgramEND, RESETThe END flag can be used to end a
Events configuration
On the Events configuration subtab user-defined event commands
can be defined, which can be inserted into a time program after
being activated with <Activate>.
program, especially if the program
time should be longer than the
chromatogram duration. Additional
steps after this flag are not allowed.
The RESET flag is used to reset the
parameters to the system startup
values.
36
Name (1
st
column)User-definable name of the event com-
mand (e.g. Start_766).
IC Net 2.0
5.2 762 IC Interface
Event command (2nd column)
Setting the event output lines 1…7.
Selection:
0(line off, inactive, open)
1(line on, active, 0 V)
p(pulse, pulse length 150 ms)
*(leave line in current status)
For entry of the first value, enter 1, 0, p or *.
For entry of the other values, move the cursor in front of the value to be changed and
enter 1, 0, p or *.
<Add>Add new remote command.
<Delete>Delete selected remote command.
<Activate>Activate the defined remote commands for
insertion into the time program.
Events setup
The Events setup tab of the 762 System [#] window contains the
startup values for the seven event output lines of system 1 or
system 2 of the 762 IC Interface. These startup values are sent
each time the system is connected or a determination is started.
Each event output line can be set to the following values:
0Line off, inactive (open)
IC Net 2.0
1Line on, active (0 V)
For technical information see 762 Instructions for Use.
37
5 Interfaces
5.3 771 Compact Interface
5.3.1 771 Compact Interface features
The Metrohm 771 Compact Interface is a two channel Analogto-Digital Converter to acquire two analog signals from two different detectors simultaneously or time independent. It contains the
following parts:
• Two 24-bit Analog-to-Digital Converters (often abbreviated
as ADC) to acquire two analog signals from two different detectors simultaneously (for example conductivity detection and
UV detection).
• • Two external start inputs.
The 771 Compact Interface is normally connected to a PC COM
port using the 6.2134.100 connecting cable. The 6.2128.170 cable is used to acquire the analog signal from the 732 IC Detector
(or another detector) and to start the data acquisition with contact
closure from the 733 IC Separation Center.
For detailed information about the 771 Compact Interface see
771 Instructions for Use.
5.3.2 771 icon on the toolbar
The 771 icon is available on the toolbar if a 771 Compact Interface has been installed with the New system wizard or by using the
<Install device> option of the WORKPLACE window (see section
5.1.2). By clicking this icon the 771 Compact Interface window for
parameter settings opens.
5.3.3 771 Compact Interface window
771 icon / Open
The 771 Compact Interface window for parameter settings is
opened by selecting this menu option with the right mouse button
or by clicking the 771 icon on the toolbar. It consists of three tabs
Init, RS settings, and Links.
38
Init
The Init tab of the 771 Compact Interface window contains data acquisition parameters for the 771 Compact Interface.
LabelOptional label to name the interface with
maximum 8 characters.
ROM versionInstrument program version number.
Measuring per second Number of data points measured per s.
The RS settings tab of the 771 Compact Interface window contains
RS232 parameters for the 771 Compact Interface.
pact Interface.
Interface.
IC Net 2.0
BaudrateBaud rate of the device.
Selection: 1200...115200
Default value: 57600
ParityParity check.
Selection: None, Even, Odd
Default value: None
39
5 Interfaces
HandshakeEnable/disable software handshake mode.
Selection: Swchar, none
Default value: SWchar
<ASK>Read current parameters from 771 Com-
pact Interface.
<SET>Send current parameters to 771 Compact
Interface.
Links
The Links tab of the 771 IC Interface is used for COM port selection
and settings (details see section 5.2.4).
5.4 761 Compact IC
5.4.1 761 Compact IC features
The Metrohm 761 Compact IC is a PC-controlled system for ion
chromatographic analyses. The two following versions are available:
• 2.761.0010 Compact IC without suppressor module
• • 2.761.0020 Compact IC with suppressor module
All the parts of the 761 Compact IC like injection valve, highpressure pump, conductivity detector, suppressor module and
peristaltic pump, are normally fully controlled by the «761 Com-pact IC» program. However, the 761 Compact IC can also be
operated using the «IC Net» program. This program also enables
the operation of more than one 761 Compact IC with one PC.
The 761 Compact IC is normally connected to a PC COM port
using the 6.2134.100 connecting cable.
For detailed information about the 761 Compact IC see 761 In-structions for Use.
5.4.2 761 icon on the toolbar
The 761 icon is available on the toolbar if a 761 Compact IC has
been installed with the new system wizard or by using the <Install
device> option of the WORKPLACE window (see section 5.1.2). By
clicking this icon the 761 Compact IC window for parameter settings opens.
40
IC Net 2.0
5.4.3 761 icon in the system window
System disconnectedSystem connectedSystem connected
The 761 icon is available in the system window if a 761 Compact IC
has been installed with the Install existing devices option (see sec-tion 6.1.3). If the system is connected, the 761 icon contains two
buttons for manual control of the injection valve at the 761
Compact IC:
5.4 761 Compact IC
Injection valve inInjection valve in
"INJECT" position"FILL" position
<Inject>Switch the injection valve to the "Inject" position.
<Fill>Switch the injection valve to the "Fill" position.
If the system is connected and the 761 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 761 Compact IC window for parameter
HardwareOpen the hardware settings window.
DiagnosticsOpen the diagnostics window (for service per-
UnlinkDelete the 761 icon from the system.
5.4.4 761 Compact IC window
761 icon / Open
The 761 Compact IC window for parameter settings is opened by
selecting this menu option with the right mouse button or by
clicking the 761 icon on the toolbar. It consists of the five tabs
Control, Initial, Program, Configuration, and Links.
settings (this window can also be opened by
double-clicking the icon).
sonal only).
IC Net 2.0
Control
The Control tab of the 761 Compact IC window is only available for
a connected system. It allows manual control of the 761
Compact IC functions and setting of startup values to be sent to
the instrument.
41
5 Interfaces
Conductivity, µµS/cmLive display of measured conductivity.
Pressure, MPaLive display of measured pressure.
The color settings for this two fields can be
changed by clicking the fields with the right
mouse button (see section 6.2.2).
Actual valuesDisplay of actual values.
Flow, mL/minDisplay of current flow rate of the high-
pressure pump.
Pressure max, MPa Display of current maximum pressure limit
for high-pressure pump.
Pressure min, MPa Display of current minimum pressure limit
for high-pressure pump.
Full scale, µµS/cmDisplay of current full scale range.
Remote linesDisplay of current remote line settings.
ValveInjection valve.
<Inject>Switch injection valve to "INJECT" position.
<Fill>Switch injection valve to "FILL" position
42
SuppressorSuppressor module.
<Step>Switch the suppressor module to the next
position. The time since the last switching
of the suppressor module is displayed in
the field beside the <Step> button.
System startup values The system startup values are sent and
applied to the 761 Compact IC each time
the system is connected, a determination is
started, or the values are sent manually
with <Send to unit>.
Flow, mL/minStartup value for flow rate of the high-
pressure pump.
Entry range: 0.20 ... 2.50 mL/min
Pressure max, MPa Startup value for maximum pressure limit
for high-pressure pump. This limit is controlled even without connection to the PC.
Entry range: 0.0 ... 25.0 MPa
Initial
Pressure min, MPa Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range: 0.0 ... 25.0MPa
Full scale, µµS/cmStartup value for full scale range.
Selection: 50, 250, 1000 µµS/cm
Remote linesStartup value for remote line settings 1…8.
Selection: 0, 1
The Initial tab of the 761 Compact IC window for settings of system
startup values is only available for a disconnected system.
IC Net 2.0
Flow, mL/minStartup value for flow rate of the high-
pressure pump.
Entry range: 0.20 ... 2.50 mL/min
43
5 Interfaces
Pressure max, MPaStartup value for maximum pressure limit
for high-pressure pump.
Entry range: 0.0 ... 25.0 MPa
Pressure min, MPaStartup value for minimum pressure limit for
high-pressure pump.
Entry range: 0.0 ... 25.0MPa
Full scale, µµS/cmStartup value for full scale range.
Selection: 50, 250, 1000 µµS/cm
Remote linesStartup value for remote line settings 1…8.
Selection: 0, 1
Time program
On the Program tab of the 761 Compact IC window a user-defined
time program for the remote output lines of the 761 Compact IC
can be entered. This program is started automatically as defined
in the Start mode window (see section 4.4.3) either at the moment
the determination is started (Start with determination) or at the
moment the sample is injected (Start with inject).
44
The Program tab contains the two following subpages:
ProgramMain time program with all program
steps.
Remote configurationPossibility for creation of user-defined
remote commands.
Program
On the Program subpage, program steps including time, program
instruction and parameter can be entered.
First columnTime at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruction is applied together with the last instruction with time entry.
IC Net 2.0
5.4 761 Compact IC
Second columnProgram instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the Remote configura-
tion tab.
Third columnParameter for program instruction (see
below).
ENABLEDEnable program start (a disabled program
is not started).
<Add>Add new program instruction.
<Delete>Delete selected program instruction.
<Verify>Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time program on the Program subpage:
InstructionParameter entryMeaning
ValveInject, FillSwitch injection valve to "INJECT"
FullScale50, 250, 1000
µµS/cm
ICPumpon, offSwitch on or off the high-pressure
Flow0.2 ... 2.5 mL/minSet flow rate of the high-pressure
Pmax0.0 ... 25.0 MPaSet maximum pressure limit for
Pmin0.0 ... 25.0 MPaSet minimum pressure limit for
or "FILL" position.
Set full scale range to the se-
lected value.
pump.
pump to the desired value.
the high-pressure pump to the desired value.
the high-pressure pump to the desired value.
IC Net 2.0
Remote0, 1, *, pSet remote output lines 1...8 to
the desired values. For entry of the
first value, enter 1, 0, p or *. For entry of the other values, move the
cursor in front of the value to be
changed and enter 1, 0, p or *.
45
5 Interfaces
ProgramEND, RESETThe END flag can be used to end a
program, especially if the program
time should be longer than the
chromatogram duration. Additional
steps after this flag are not allowed.
The RESET flag is used to reset the
parameters to the system startup
values.
SuppressorSwitch suppressor module to the
next position.
Peristalticon, offSwitch on or off the peristaltic
pump.
Remote configuration
On the Remote configuration subtab user-defined remote commands can be defined, which can be inserted into a time program after being activated with <Activate>.
46
Name (1
st
column)User-definable name of the remote com-
mand (e.g. Start_766).
Remote command (2nd column)
Setting the remote output lines 1…8.
0(line off, inactive, open)
1(line on, active, 0 V)
p(pulse)
*(leave line in current status)
For entry of the first value, enter 1, 0, p or *.
For entry of the other values, move the cursor in front of the value to be changed and
enter 1, 0, p or *.
<Add>Add new remote command.
<Delete>Delete selected remote command.
<Activate>Activate the defined remote commands for
insertion into the time program.
IC Net 2.0
Configuration
The Configuration tab in the system settings window contains
configuration settings for the 761 Compact IC.
5.4 761 Compact IC
761 unit versionSelection of instrument version:
761.0010 = 761 Compact IC without sup-
pressor
761.0020 = 761 Compact IC with suppres-
sor
SuppressorSuppressor module:
Autostep withAutomatic switching to the next position
each time the injection valve is switched
either to the Fill or to the Inject position.
Use peristaltic pumpIf this option is disabled, the peristaltic
pump is not switched on automatically at
the start of a determination or with Startup
Measure channelDisplay of the data source selected in the
Data source window (see section 6.12.2).
IC Net 2.0
47
5 Interfaces
5.4.5 Hardware settings
Selection of the Hardware menu item of the 761 icon using the right
mouse button opens the hardware settings window consisting of
the three tabs Hardware, Outputs and Alarm stops.
Hardware
The Hardware tab of the hardware settings window defines general parameters which are set automatically at power on of the
761 Compact IC.
Remote lines after power on
The remote output lines 1…8 are set to this
values after power on or a manual stop with
Shutdown hardware.
Selection: 0, 1
ThermostatOperating temperature of the conductivity
cell.
Selection: 25, 30, 35, 40, 45 °C, off
Thermostatting functions only if the ambient temperature is at least 5 °C lower than the operating temperature. It normally takes 30…60 min after power on
until a temperature stability of ± 0.01 °C is attained.
Flow correctionFactor for correction of the difference be-
tween displayed and actual flow rate of the
high-pressure pump.
Range: 0.9 ... 1.09
The flow correction is determined by measurement of the actual flow rate with the aid
of a measuring cylinder as follows:
48
Flow correction =
Displayed flow rate
Measured flow rate
IC Net 2.0
5.4 761 Compact IC
Cell constantCell constant of the conductivity cell for
correct display of the absolute conductivity.
Enter the cell constant printed on the detector block into this field.
Range: 0.1 ... 1000 /cm
For a precise determination of the cell constant, pump a calibration solution of known
conductivity through the IC system, observe the displayed conductivity and
change the cell constant until the correct
conductivity value is displayed.
<Break times>Possibility for changing the break times for
injection valve Valve and suppressor module Suppressor.
These values should only be changed after consulting the Metrohm Service.
Outputs
The Outputs tab of the hardware settings window defines remote
output signals to be set automatically if specific events occur.
IC Net 2.0
EventEvents for automatic remote signal output:
InjectSwitching of the injection valve to the
"INJECT" position.
FillSwitching of the injection valve to the "FILL"
position.
Suppressor stepSwitching of the suppressor module to the
next position.
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
49
5 Interfaces
Hardware errorHardware error detected at the 761 Com-
pact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
Alarm leak detector Leak detector has detected solvent in the
instruments interior.
Remote lineSet remote output lines 1…8.
Selection:0 (line off, inactive, open)
1 (line on, active, 0 V)
p (pulse output)
* (leave line in current status)
PowerOn valuesDisplay of the power on startup values for
remote output lines set on the Hardware tab.
Pulse lengthLength of a pulse in ms.
Alarm stops
The Alarm stops tab of the hardware settings window defines the
events for which the 761 Compact IC is stopped immediately. At
an alarm stop, high-pressure pump and peristaltic pump are
stopped immediately, the running determination and the active
sample queue are also stopped.
50
EventEvents for alarm stop:
Pump stopPump stopped because pressure limits are
exceeded.
Hardware errorHardware error detected at the 761 Com-
pact IC (high-pressure pump, injection
valve, or suppressor not working correctly).
LeakLeak detector has detected solvent in the
instruments interior. This information is also
stored in the instrument itself, so that it is
stopped automatically even without connection to the PC.
Power onTemporary power failure at the 761 Com-
pact IC hardware.
IC Net 2.0
5.5 Metrohm PC Board
The 2.714.0310 Metrohm PC Board Interface is a two channel
Analog-to-Digital Converter (often abbreviated as ADC) to ac-
quire two analog signals from two different detectors simultaneously or time independent (e.g. conductivity detection and UV
detection).
One or two PC Boards can be installed in the same PC: PCBoard 1 is the first interface, PC Board 2 is the second interface
installed in the PC. They differ in different jumper settings on the
board. The «IC Net» program is the successor for the former «IC
Metrodata for Windows» program used to work with the PC
Boards.
The 6.2128.140 cable is used to connect the analog output of the
732 IC Detector (or another detector) and the contact closure
output of the 733 IC Separation Center to the Metrohm PC Board.
The Metrohm PC Board must be installed with the new system
wizard or by using the <Install device> option of the Workplace
window (see section 5.1.2). However, there is no PC Board icon
on the toolbar after installation. The PC Board can only be seen in
the Workplace window opened with Options / Devices setup. By
double-clicking the MeBoard1 or MeBoard2 item, the MeBoard#
window opens with the following tabs:
5.5 Metrohm PC Board
Device settingsManual control and startup values for PC
LinksWindow for COM port selection and set-
For detailed information about the Metrohm PC Board see on-line
help and Instructions for Use for «IC Metrodata for Windows».
5.6 Global timer
5.6.1 Timer icon
The global timer is used for programming system tasks which are
started automatically daily or once at the desired time. The Global
timer icon is available on the toolbar if the Timer has been installed
with the new system wizard or by using the <Install device> option
of the Workplace window (see section 5.1.2). By clicking this icon
the Timer window for parameter settings opens.
Board (it is recommended not to change
any of these parameters!).
tings (it is recommended not to change any
of these parameters!).
IC Net 2.0
51
5 Interfaces
5.6.2 Timer program
The Timer window contains the user-defined program for system tasks. These tasks are started automatically at the day and
time defined in the program.
The title line of the Timer window shows the current time left in
days:hours:minutes:seconds until the next task is automatically
started.
<New task>Add a new task to the timer program.
<Daily>Selecting this option opens a subwindow for
selection of the days at which the task should
be started.
<Once>Selecting this option opens a subwindow for
selection of time and day at which the task
should be started. After the selection of the
desired program instruction the system file for
which this instruction should be applied must
be selected in the Systems folder.
<Add subtask>Add a new subtask for the selected daily task.
The time and program instruction for this
subtask has to be entered in the opened subwindow and the system file to which this instruction should be applied must be selected
in the Systems folder.
<Del>Delete the selected task from the timer pro-
gram.
52
The following program instructions can be used for userdefined event program tasks in the Timer program window:
OpenOpen the selected system.
CloseClose the selected system and disconnect it
from the workplace.
StartupStartup hardware of the selected system.
StartStart determination using the selected system.
StopStop determination using the selected system.
ShutdownShutdown hardware of the selected system.
IC Net 2.0
6 Devices
The «IC Net» program supports remote control for Metrohm and
Bischoff instruments. Each instrument within a system is a de-vice. Devices are stored as device files (*.dev) in the Devices directory. All Metrohm devices start with Me*.dev and all Bischoffdevices start with Bi*.dev.
6.1 Device installation
6.1 Device installation
6.1.1 Install devices at system creation
IC NET / File / New / System
The New system wizard window and a new SYSTEM window are
opened. The New system wizard guides you trough the installation
routine step-by-step to build a new system. The next step of the
system installation is reached with <Next>, the previous step with
<Back> and the installation is completed with <Finish>.
The first step of the system installation is to enter a name for a
new system folder within the «IC Net» program path
...\ICNet\Systems. For every instrument combination, a new folder
should be created.
IC Net 2.0
After clicking on <Next>, the window for addition of interfaces to
the toolbar appears (details see section 4.1.1).
After clicking on <Next>, the window for addition of devices to the
system window appears.
53
6 Devices
For every device to be controlled with «IC Net», open the group
which contains this device, select the device and the Serial port of
the interface where the device is connected to. After clicking on
<Add to system>, the device is added to the SYSTEM window. If
desired, an additional data recorder and a timer of the More mod-
ules group can be added to the system.
Details to the other steps of the system wizard procedure
(processing method, data source) see section 4.1.1.
6.1.2 Install new devices in system window
SYSTEM / Setup / New devices/ Install new device
Add new devices to the open SYSTEM window using the New
system wizard window (see above). For every device to be added
to the system, open the group which contains this device and
select the device and the serial port of the interface where the
device is connected to. After clicking on <Add to system>, the device is added to the SYSTEM window. If desired, an additional
data recorder and a timer of the More modules group can be
added to the system.
6.1.3 Install existing devices in system window
SYSTEM / Setup / New devices/ Link to existing device
Add devices to the open SYSTEM window which
are already present on the toolbar. This is mostly
used to add the 761 Compact IC or 762 IC Interface
events. For every device to be added to the sys-
tem, select the device and click on <OK>.
6.1.4 Delete devices
54
Icon / Unlink
Devices in a SYSTEM window can be deleted by clicking on the
device icon using the right mouse button an select the Unlink
menu item.
IC Net 2.0
6.2 Watch window
6.2.1 Watch window icon
The watch window icon is one of the components always present in a new SYSTEM window. If the system is connected and this
icon is clicked with the right mouse button, the following menu
item appears:
OpenOpen the WATCH WINDOW for live display of
UnlinkDelete the WATCH WINDOW from the SYS-
6.2 Watch window
instrument values (this window can also be
opened by double-clicking the watch window icon).
TEM window (not recommended).
6.2.2 Watch window settings
Watch window icon / Open
The WATCH WINDOW displays all live parameters of the interfaces
and devices included in the connected system. It can be set to
be opened automatically if a determination or measure baseline
is started with SYSTEM / Setup / Watch window (see section 4.4.2).
The WATCH WINDOW can be configured manually by clicking the
frame titles inside the window with the right mouse button. This
opens the following menu:
Manual arrangePossibility to select the parameters which
should be shown or to rearrange the sequence of the displayed parameters. This
option opens the Data source window for
selection of the parameters to be displayed
in the watch window. The Connected field
shows the parameters to be displayed, the
Available field shows the parameters not to
be displayed. The parameters can be
moved from one field to the other using the
IC Net 2.0
or button.
Auto arrangeAutomatic arrangement of all important
parameters.
55
6 Devices
The colors of the watch window fields and of the SYSTEM STATE
window can be changed by clicking the fields with the right
mouse button. This offers the following possibilities:
Choose color / Foreground / Out of range
Set display color for field numbers if the
measurement value is out of range
(overflow or out of limits).
Choose color / Foreground / In range
Set display color for field numbers if the
measurement value is inside the normal
range.
Choose color / Foreground / Active
Set display color for field characters in the
SYSTEM STATE window for running tasks.
Choose color / Foreground / Passive
Set display color for field characters in the
SYSTEM STATE window for finished tasks.
Choose color / Background / Out of range
Set display color for field background if the
measurement value is out of range
(overflow or out of limits).
Choose color / Background / In range
Set display color for field background if the
measurement value is inside the normal
range.
Choose color / Background / Active
Set display color for field background in the
SYSTEM STATE window for running tasks.
Choose color / Background / Passive
Set display color for field background in the
SYSTEM STATE window for finished tasks.
The colors set for this fields are also applied to the measurement
display fields in the control tabs of some interfaces and devices.
56
IC Net 2.0
6.3 732 IC Detector
6.3.1 732 IC Detector features
The Metrohm 732 IC Detector is a conductivity detector for ion
chromatography. The associated thermostatable detector block
can be used for recording of chromatograms with and without
chemical suppression and is normally installed in the 733 IC
Separation Center. The two following versions are available:
• 2.732.0010 IC Detectorwith standard detector block
• 2.732.0110 IC Detectorwith metal-free detector block
The 732 IC Detector, which is normally operated using the keypad of the instrument, can be fully operated using the «IC Net»
program. The 732 IC Detector is normally connected to the 762
IC Interface using the 6.2134.080 or 6.2134.090 cable. Examples
for interconnections with the 732 IC Detector can be found in the
762 Instructions for Use.
6.3 732 IC Detector
In order to operate the 732 IC Detector with «IC Net», the following RS settings have to be made using the [ CONFIG ] key on
the 732 IC Detector panel:
>CONFIG/RS settings/handshake: Swchar
>CONFIG/RS settings/RS control: on
For detailed information about the 732 IC Detector see 732/733
Instructions for Use.
6.3.2 732 IC Detector icon
The 732 icon is available in the SYSTEM window if a 732 IC Detector
has been installed with the New system wizard or by using the
Setup/New devices/Install new device option of the SYSTEM window
(see section 6.1.2).
IC Net 2.0
If the system is connected and the 732 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 732 IC Detector window for parameter set-
tings (this window can also be opened by doubleclicking the icon).
UnlinkDelete the 732 icon from the system.
57
6 Devices
6.3.3 732 IC Detector window
732 icon / Open
The 732 IC Detector window for parameter settings is opened by
selecting this menu option with the right mouse button or by
double-clicking the 732 icon in the SYSTEM window. It consists of
the four tabs Display, Method parameter, Program, and Links.
Display
The Display tab of the 732 IC Detector window is only available for
a connected system. It displays the current measurement values and parameters settings of the 732 IC Detector.
58
ConductivityLive display of measured conductivity
(auto-zero value) in µS/cm. This field is
available for the WATCH WINDOW. The
watch window color is changed to the Out
of range color if the measured value is more
than 80% outside the Full Scale range.
abs. Cond.Live display of absolute conductivity value
in µS/cm. This field is available for the
WATCH WINDOW. The watch window color is
changed to the Out of range color if the
measured value is bigger or equal to 110%
of Range.
IonDisplay of ion settings:
Anion: polarity = +, temperature coeff. = 2.5
Cation: polarity = –, temperature coeff. = 1.5
IC Net 2.0
6.3 732 IC Detector
RangeDisplay of measuring range in µS/cm.
Full scaleDisplay of full scale range in µS/cm.
Cell const.Display of cell constant of the conductivity
cell in /cm.
ThermostatDisplay of temperature within in the detec-
tor block.
OffsetDisplay of offset of the zero point.
DampingDisplay of electronic damping of the analog
output.
Zero <ON>Triggers the auto-zero function.
Zero <OFF>Switch off the auto-zero function.
<Ask all>Update all display values.
<Import in method parameters>
Copy displayed parameters into the Method
parameter tab.
Method parameter
The Method parameter tab of the 732 IC Detector window contains
the method startup values for the instrument. These values are
sent and applied to the 732 IC Detector by Start determination and
Startup hardware, or if the values are sent manually with <SET>.
IC Net 2.0
59
6 Devices
RangeMeasuring range in µS/cm
Selection: 100, 200, 500 µµS/cm and
1, 2, 5, 10 mS/cm
Full scaleThe full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
in µS/cm depend on the preset measuring
Range.
IonCombination of polarity and temperature
coefficient:
Anionpolarity = +, temperature coeff. = 2.5
Cationpolarity = –, temperature coeff. = 1.5
Customindependent selection of polarity and tem-
perature coeff.
PolarityPolarity of the analog output signal:
+positive polarity (e.g. for anions)
–negative polarity (e.g. for cations)
Temperature coeff.Converts automatically the conductivity
from the operating temperature of the
measuring cell to the reference temperature
of 20°C.
Selection: 1.5, 2.5 %/°C
Cell constantCell constant of the conductivity cell in /cm.
ThermostatOperating temperature of the conductivity
OffsetOffset of the zero point of the conductivity.
DampingElectronic damping of the analog output
<SET>Send current parameters immediately to
For detailed information about the parameters of the 732 IC Detector see 732/733 Instructions for Use.
Time program
On the Program tab of the 732 IC Detector window a user-defined
time program for the remote output lines of the 732 IC Detector
can be entered. This program is started automatically as defined
in the Start mode window (see section 4.4.3) either at the moment
the determination is started (Start with determination) or at the
moment the sample is injected (Start with inject).
Entry range: 13 ... 21 /cm
cell within in the detector block.
Selection: 25, 30, 35, 40, 45 °C and off
Selection: 0, 10, 50 %fs
Selection: on, off
the 732 IC Detector. Parameters are not
stored in the system file (*.smt) as long as
the file is not saved.
60
IC Net 2.0
6.3 732 IC Detector
The Program tab contains the two following subpages:
ProgramMain time program with all program
steps.
Program
Remote configurationPossibility for creation of user-defined
remote commands.
On the Program subpage, program steps including time, program
instruction and parameter can be entered.
First columnTime at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruction is applied together with the last instruction with time entry.
Second columnProgram instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the Remote configura-
tion tab.
Third columnParameter for program instruction (see
below).
IC Net 2.0
ENABLEDEnable program start (a disabled program
is not started).
Will be executed bySelection if the time program will be exe-
cuted by the PC or the 732 unit.
<Add>Add new program instruction.
<Delete>Delete selected program instruction.
<Verify>Test the time program (error messages are
displayed if program is wrong).
61
6 Devices
List of program instructions
The following program instructions can be added to the time program on the Program subpage:
Instruction Parameter entryMeaning
Remote0, 1, *Set remote output lines 1...8 to
the desired values. For entry of
the first value, enter 1, 0 or *. For
entry of the other values, move
the cursor in front of the value to
be changed and enter 1, 0 or *.
Flagreset, endSet program flag. The end flag
can be used to end a program,
especially if the program time
should be longer than the chromatogram duration. Additional
steps after this flag are not allowed. The reset flag is used to
reset the parameters to the system startup values.
Range100, 200, 5000 µµS/cm
1, 2, 5, 10 mS/cm
FS0.05 µµS/cm ...
10 mS/cm
Set measuring range to the selected value.
Set full scale range to the selected value (depends on Range
setting).
Zeroon, offSwitch on or off the auto-zero
function.
Polarity+, –Reset the polarity of the output
signal.
MarkTrigger marking signal.
ValveAfill, injectSwitch injection valve A to "inject"
or "fill" position.
ValveBfill, injectSwitch injection valve B to "inject"
or "fill" position.
SuppressorSwitch suppressor module to the
next position.
62
PumpRSon, offSwitch on or off the high-pressure
pump connected to the 732 IC
Detector.
Flow0.01 ... 5 mL/minSet flow rate of the high-pressure
pump to the desired value.
IC Net 2.0
Pmax0.0 ... 25.0 MPaSet maximum pressure limit for
Pmin0.0 ... 25.0 MPaSet minimum pressure limit for the
Remote configuration
On the Remote configuration subtab user-defined remote commands can be defined, which can be inserted into a time program after being activated with <Activate>.
6.3 732 IC Detector
the high-pressure pump connected to the 732 IC Detector to
the desired value.
high-pressure pump connected to
the 732 IC Detector to the desired
value.
Links
Name (1
st
column)User-definable name of the remote com-
mand (e.g. Start_766).
Remote command (2nd column)
Setting the remote output lines 1…8.
0(line off, inactive, open)
1(line on, active, 0 V)
*(leave line in current status)
For entry of the first value, enter 1, 0 or *.
For entry of the other values, move the cursor in front of the value to be changed and
enter 1, 0 or *.
<Add>Add new remote command.
<Delete>Delete selected remote command.
<Activate>Activate the defined remote commands for
insertion into the time program.
The Links tab of the 732 IC Detector is used for COM port selection
and settings (details see section 5.2.4).
IC Net 2.0
63
6 Devices
6.4 733 IC Separation Center
6.4.1 733 IC Separation Center features
The Metrohm 733 IC Separation Center is a thermally and
electronically isolated wet part which accommodates injectors,
columns, detectors, suppressor module and pulsation dampener
and is controlled by and connected to the 732 IC Detector. The
following versions are available:
• 2.733.0010 IC Separation Center with 1 injector
• 2.733.0020 IC Separation Center with 2 injectors
• 2.733.0120 IC Separation Center with 2 injectors, metal-free
• 2.733.0030 IC Separation Center with 1 injector and 1 sup-
pressor module
• 2.733.0130 IC Separation Center with 1 injector and 1 suppressor module, metal-free
The 733 IC Separation Center, which is normally operated by the
732 IC Detector, can be fully operated using the «IC Net» program. The 733 IC Separation Center with one injection valve is
normally connected to the 732 IC Detector using the 6.2125.090
cable. For the operation of a 733 IC Separation Center with twoinjection valves, two individual 732 IC Detectors are needed.
Each of this detectors is connected to the 733 IC Separation
Center using a 6.2125.090 cable. For operation of a two-valve
system with «IC Net», two 732 icons and two 733 icons (one for
valve A and one for valve B) have to be installed in the SYSTEM
window.
For detailed information about the 733 IC Separation Center see
732/733 Instructions for Use. Examples for interconnections
with the 733 IC Separation Center, the 732 IC Detector and the
762 IC Interface can be found in the 762 Instructions for Use.
6.4.2 733 IC Separation Center icon
The 733 icon is available in the SYSTEM window if a 733 IC Separa-
tion Center has been installed with the New system wizard or by
using the Setup/New devices/Install new device option of the SYS-
TEM window (see section 6.1.2).
64
IC Net 2.0
If the system is connected and the 733 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 733 IC Separation Center window for parame-
ter settings (this window can also be opened by double-clicking the icon).
UnlinkDelete the 733 icon from the system.
6.4.3 733 IC Separation Center window
733 icon / Open
The 733 IC Separation Center window for parameter settings is
opened by selecting this menu option with the right mouse button
or by double-clicking the 733 icon in the SYSTEM window. It consists of the two tabs IC733 and Links.
IC733
The IC733 tab of the 733 IC Separation Center window is only available for a connected system. The parameters and buttons for
manual operation depend on the type of instrument connected.
6.4 733 IC Separation Center
IC Net 2.0
ConfigurationDisplay of current instrument configuration.
valve A733 IC Separation Center with 1 injector.
valve A + valve B733 IC Separation Center with 2 injectors.
valve A + suppressor
733 IC Separation Center with 1 injector
and 1 suppressor module.
Valve A
fill/injectDisplay of current valve position.
<FILL>Switch valve A to "fill" position.
<INJECT>Switch valve A to "inject" position.
Valve B
fill/injectDisplay of current valve position.
<FILL>Switch valve B to "fill" position.
<INJECT>Switch valve B to "inject" position.
65
6 Devices
Suppressor
Auto Step
Links
The Links tab of the 733 IC Separation Center is used for COM port
selection and settings (details see section 5.2.4).
6.5 709 IC Pump
in positionDisplay of current suppressor status.
<STEP>Switch suppressor module to next position.
fill, inject, offDisplay of current auto step status.
Entry fieldSet automatic switching of the suppressor
module after each "fill" or "inject" of valve A.
Selection: fill, inject, off
6.5.1 709 IC Pump features
The Metrohm 709 IC Pump is a serial dual piston high pressure
pump for ion chromatography. The two following versions are
available:
• 2.709.0010 IC Pump(standard version)
• 2.709.0110 IC Pump(metal-free version)
The 709 IC Pump, which is normally operated using the keypad
of the instrument or via 732 IC Detector, can be fully operated
using the «IC Net» program. It can be connected either to the 732
IC Detector using the 6.2125.060 cable, to the 762 IC Interface
using the 6.2134.080 or 6.2134.090 cable, or to a PC COM port
using the 6.2134.040 or 6.2134.100 cable. If it is connected to the
732 IC Detector, the Isocratic pump 709 for 732 device must be selected for the installation of the 709 IC Pump in a system. If it is
connected to the 762 IC Interface, the Isocratic pump 709 device
must be selected for the installation of the 709 IC Pump in a system.
Examples for interconnections with the 709 IC Pump can be
found in the 762 Instructions for Use.
66
The cable connection between 709 IC Pump and 732 IC Detector, 762 IC Interface or PC must be made before switching
on the instrument. In order to operate the 709 IC Pump with «IC
Net», the external control must be switched on using the [ Ext. ]
key.
For detailed information about the 709 IC Pump see 709 Instructions for Use.
IC Net 2.0
6.5.2 709 IC Pump icon
The 709 icon is available in the SYSTEM window if an Isocratic
pump 709 or an Isocratic pump 709 for 732 has been installed with
the New system wizard or by using the Setup/New devices/Install new
device option of the SYSTEM window (see section 6.1.2).
If the system is connected and the 709 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 709 IC Pump window for parameter settings
UnlinkDelete the 709 icon from the system.
6.5 709 IC Pump
(this window can also be opened by double-clicking
the icon).
6.5.3 709 IC Pump window
709 icon / Open
The 709 IC Pump window for parameter settings and manual control is opened by selecting this menu option with the right mouse
button or by double-clicking the 709 icon in the SYSTEM window. It
consists of the three tabs Settings, Links and Advanced.
Settings
The Settings tab of the 709 IC Pump window is only available for a
connected system.
IC Net 2.0
67
6 Devices
PressureLive display of current pressure in MPa.
This field is also available for the WATCH
WINDOW.
FlowLive display of current flow in mL/min. This
field is also available for the WATCH WIN-
DOW.
New flowA new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range: 0.01 ... 5.00 mL/min
Min/Max pressureThe minimum and maximum pressure limit
for the 709 IC Pump can be set in steps of
0.1 MPa.
Entry range: 0.0 ... 50.0 MPa.
The set maximum limit value should lie
between 5 to 10 MPa above the particular
operating pressure or the maximum admissible operating pressure of the column
used. If the pump exceeds the preset limit
value during operation, it is switched off
immediately. At the same time, the Pressure
field color in the 709 settings window and in
the WATCH WINDOW is changed to the value
set for Out of range.
The set minimum limit value should lie farenough below the particular operating
pressure. If the pump pressure falls below
this preset lower limit during operation and
this pressure drop persist for several revolutions of the pump cam due to leaks or
interrupted inflow of the eluent, the pump is
automatically switched off. At the same
time, the Pressure field color in the 709 settings window and in the WATCH WINDOW is
changed to the value set for Out of range.
<Set>Send current parameters immediately to
the 709 IC Pump. Parameters are not
stored in the system file (*.smt) as long as
the file is not saved.
<Start>Start 709 IC Pump drive.
<Stop>Stop 709 IC Pump drive.
68
Links
The Links tab of the 709 IC Pump is used for COM port selection
and settings (details see section 5.2.4).
IC Net 2.0
Advanced
6.6 Metrohm solvent delivery unit
The Advanced tab of the 709 IC Pump window is only available for a
connected system.
Correction factorFlow correction factor to adjust the dis-
played flow to the actual flow measured
with the aid of a measuring cylinder.
Entry range: 0.90 ... 1.10
<Ask>Read current Correction factor value.
<Set>Send current Correction factor value to the
709 IC Pump. This parameter is not stored
in the system file (*.smt) as long as the file
is not saved.
6.6 Metrohm solvent delivery unit
6.6.1 Metrohm SDU features
A Metrohm SDU (Solvent Delivery Unit) system can be used for
high-pressure gradient mixture of one to four eluents. For each
eluent, a 709 IC Pump is used. Each 709 IC Pump must be connected to the 762 IC Interface using a 6.2134.080 or 6.2134.090
cable. Examples for interconnections with the 709 IC Pump can
be found in the 762 Instructions for Use.
IC Net 2.0
The outlet capillaries of the 709 IC Pumps must be connected to
a mixing chamber from where one inlet capillary leads to the injection valve of the 733 IC Separation Center.
For detailed information about the 709 IC Pump see 709 Instruc-
tions for Use.
69
6 Devices
6.6.2 Metrohm SDU icon
The SDU icon is available in the SYSTEM window if an SDU
(Solvent Delivery Unit) has been installed with the New system wiz-
ard or by using the Setup/New devices/Install new device option of
the SYSTEM window (see section 6.1.2). The SDU icon consists of
one to four 709 IC Pump icons one above the other depending
on the SDU type installed.
If the system is connected and the SDU icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the SDU window for parameter settings (this
window can also be opened by double-clicking the
icon).
UnlinkDelete the SDU icon from the system.
6.6.3 Metrohm SDU window
SDU icon / Open
The Pump control system window for parameter settings and manual control is opened by selecting this menu option with the right
mouse button or by double-clicking the SDU icon in the SYSTEM
window. It consists of the four tabs Manual, Program, Interfaces,
and Links.
Manual
The Manual tab of the Metrohm SDU window is only available for a
connected system.
70
IC Net 2.0
6.6 Metrohm solvent delivery unit
PressureLive display of current pressure in MPa.
This field is also available for the WATCH
WINDOW.
FlowLive display of current flow in mL/min. This
field is also available for the WATCH WIN-
DOW.
New flowA new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range: 0.01 ... 20.00 mL/min (depending
on number of pumps)
Min/Max pressureThe minimum and maximum pressure limit
for the individual pump units can be set in
steps of 0.1 MPa.
Entry range: 0.0 ... 50.0 MPa.
%BIndividual flow rate for pump B in percent of
total flow.
Entry range: 0 ... 100 %
%CIndividual flow rate for pump C in percent of
total flow.
Entry range: 0 ... 100 %
%DIndividual flow rate for pump D in percent of
total flow.
Entry range: 0 ... 100 %
Stop afterStop time for all pump drives.
Entry range: 0 ... 99999 min
<Set>Send current parameters immediately to
the pump control system. Parameters are
not stored in the system file (*.smt) as long
as the file is not saved.
<Start>Start all pump drives.
<Stop>Stop all pump drives.
IC Net 2.0
71
6 Devices
Program
The Program tab of the SDU window contains a time program for
the pump control system to define eluent gradients.
FlowOverall flow of pump control system in
mL/min. This value is also available for the
WATCH WINDOW. This field is only available
if the Show flow column option is disabled.
Entry range: 0.01 ... 20.00 mL/min (depending
on number of pumps)
Stop pumps after the run
Stop pumps after the determination.
Program table
NumberRow number in the program table (read-
only).
TimeTime at which the flow and/or gradient is
changed.
Entry range: 0.0 ... 99999 min
FlowOverall flow of pump control system in
mL/min.
Entry range: 0.01 ... 20.00 mL/min (depending
on number of pumps)
%BIndividual flow rate for pump B in percent of
total flow.
Entry range: 0 ... 100 %
72
%CIndividual flow rate for pump C in percent of
total flow.
Entry range: 0 ... 100 %
IC Net 2.0
EnabledEnable time program start (a disabled pro-
Show flow columnShow Flow column in the program table.
<Add>Add a new line to the time program.
<Delete>Delete the selected line from the time pro-
<Start>Start all pump drives.
Interfaces
The Interfaces tab of the SDU window is only available for a connected system.
6.6 Metrohm solvent delivery unit
%DIndividual flow rate for pump D in percent of
total flow.
Entry range: 0 ... 100 %
gram is not started).
gram.
IC Net 2.0
Number of eluentsNumber of eluents used for gradient mix-
ing.
PumpsList of used pumps for gradient mixing.
Clicking on <A>, <B>, <C>, or <D> opens the
709 IC Pump window for parameter settings
and manual control of the selected pump.
Pressure checkingSelection of the pump whose pressure is
checked using the Min/Max pressure limits.
Links
The Links tab of the Metrohm SDU is used for COM port selection
and settings (details see section 5.2.4).
73
6 Devices
6.7 752 Pump Unit
6.7.1 752 Pump Unit features
The Metrohm 752 Pump Unit is a 2-channel peristaltic pump
which is normally used for operating the suppressor module in
the 733 IC Separation Center by supplying regeneration and
rinsing solutions.
The 752 Pump Unit, which is normally operated manually using
the keypad of the instrument or automatically via 732 IC Detector
via remote interface, can be fully operated using the «IC Net»
program. The 752 Pump Unit is normally connected to one of the
event line connections of the 762 IC Interface using the optional
6.2128.180 cable. Examples for interconnections with the 752
Pump Unit can be found in the 762 Instructions for Use.
In order to operate the 752 Pump Unit with «IC Net», the external
control must be switched on using the [ Remote ] key.
For detailed information about the 752 Pump Unit see 752 Instructions for Use.
6.7.2 752 Pump Unit icon
The 752 icon is available in the SYSTEM window if a 752 Pump Unit
has been installed with the New system wizard or by using the
Setup/New devices/Install new device option of the SYSTEM window
(see section 6.1.2).
If the system is connected and the 752 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 752 Pump Unit window for parameter settings
UnlinkDelete the 752 icon from the system.
(this window can also be opened by double-clicking
the icon).
6.7.3 752 Pump Unit window
752 icon / Open
The 752 Pump Unit window for parameter settings is opened by
selecting this menu option with the right mouse button or by
double-clicking the 752 icon in the SYSTEM window. It consists of
the three tabs Manual, Program, and Links.
74
IC Net 2.0
Manual
6.7 752 Pump Unit
The Manual tab of the 752 Pump Unit window for manual operation
is only available for a connected system.
Pump
<On>Start pump drive.
<Off>Stop pump drive.
Start pump with startup hardware
Automatic start of pump drive with Startup
hardware.
Time program
On the Program tab of the 752 Pump Unit window a user-defined
time program can be entered. This program is started automatically as defined in the Start mode window (see section 4.4.3) either at the moment the determination is started (Start with determi-
nation) or at the moment the sample is injected (Start with inject).
First columnTime at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
IC Net 2.0
If no time is entered, the program instruction is applied together with the last instruction with time entry.
75
6 Devices
Second columnProgram instruction (see below).
Third columnParameter for program instruction (see
below).
ENABLEDEnable program start (a disabled program
is not started).
<Add>Add new program instruction.
<Delete>Delete selected program instruction.
<Verify>Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time program on the Program page:
InstructionParameter entryMeaning
Links
ProgramRESET, ENDSet program flag. The END flag
can be used to end a program,
especially if the program time
should be longer than the chromatogram duration. Additional
steps after this flag are not allowed. The RESET flag is used to
reset the parameters to the system startup values.
Pumpon, offSwitch on or off the pump drive.
The Links tab of the 752 Pump Unit is used for COM port selection
and settings (details see section 5.2.4).
76
IC Net 2.0
6.8 753 Suppressor Module
6.8.1 753 Suppressor Module features
The Metrohm 753 Suppressor Module consists of a control unit
with a 2-channel peristaltic pump and a block containing a suppressor module.
The 753 Suppressor Module, which is normally operated manually using the keypad of the instrument or automatically via 732 IC
Detector via remote interface, can be fully operated using the «IC
Net» program. The 753 Suppressor Module is normally connected to two of the event line connections of the 762 IC Interface
using two optional 6.2128.180 cables. Examples for interconnections with the 753 Suppressor Module can be found in the 762Instructions for Use.
In order to operate the 753 Suppressor Module with «IC Net»,
the external control must be switched on using the [ Remote ]
key.
6.8 753 Suppressor Module
For detailed information about the 753 Suppressor Module see
753 Instructions for Use.
6.8.2 753 Suppressor Module icon
The 753 icon is available in the SYSTEM window if a 753 Suppressor
Module has been installed with the New system wizard or by using
the Setup/New devices/Install new device option of the SYSTEM window (see 6.1.2).
If the system is connected and the 753 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 753 Suppressor Module window for parameter
settings (this window can also be opened by doubleclicking the icon).
UnlinkDelete the 753 icon from the system.
6.8.3 753 Suppressor Module window
753 icon / Open
The 753 Suppressor Module window for parameter settings is
opened by selecting this menu option with the right mouse button
or by double-clicking the 753 icon in the SYSTEM window. It consists of the three tabs Manual, Program, and Links.
IC Net 2.0
77
6 Devices
Manual
The Manual tab of the 753 Suppressor Module window for manual
operation is only available for a connected system.
Pump
<On>Start pump drive.
<Off>Stop pump drive.
Suppressor
<Step>Rotate suppressor to the next position. The
Start pump with startup hardware
Time program
On the Program tab of the 753 Suppressor Module window a userdefined time program can be entered. This program is started
automatically as defined in the Start mode window (see 4.4.3) either at the moment the determination is started (Start with determi-
nation) or at the moment the sample is injected (Start with inject).
time since the last step action is displayed
in the field beside the <Step> button.
Automatic start of pump drive with Startup
hardware.
78
IC Net 2.0
6.8 753 Suppressor Module
First columnTime at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruction is applied together with the last instruction with time entry.
Second columnProgram instruction (see below).
Third columnParameter for program instruction (see
below).
ENABLEDEnable program start (a disabled program
is not started).
<Add>Add new program instruction.
<Delete>Delete selected program instruction.
<Verify>Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time program on the Program page:
InstructionParameter entryMeaning
ProgramRESET, ENDSet program flag. The END flag
Pumpon, offSwitch on or off the pump drive.
StepRotate suppressor to the next
can be used to end a program,
especially if the program time
should be longer than the chromatogram duration. Additional
steps after this flag are not allowed. The RESET flag is used to
reset the parameters to the system startup values.
position.
IC Net 2.0
Links
The Links tab of the 753 Suppressor Module is used for COM port
selection and settings (details see section 5.2.4).
79
6 Devices
6.9 754 Dialysis Unit
6.9.1 754 Dialysis Unit features
The Metrohm 754 Dialysis Unit is a unit for on-line sample dialysis before sample injection. It consists of a 2-channel peristaltic pump and a dialysis cell.
The 754 Dialysis Unit, which is normally operated manually using
the keypad of the instrument or automatically via 732 IC Detector
via remote interface, can be fully operated using the «IC Net»
program. The 754 Dialysis Unit is normally connected to one of
the event line connections of the 762 IC Interface using the optional 6.2128.180 cable. Examples for interconnections with the
754 Dialysis Unit can be found in the 762 Instructions for Use.
In order to operate the 754 Dialysis Unit with «IC Net», the external control must be switched on using the [ Remote ] key.
For detailed information about the 754 Dialysis Unit see 754 Instructions for Use.
6.9.2 754 Dialysis Unit icon
The 754 icon is available in the SYSTEM window if a 754 Dialysis
Unit has been installed with the New system wizard or by using the
Setup/New devices/Install new device option of the SYSTEM window
(see section 6.1.2).
If the system is connected and the 754 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 754 Dialysis Unit window for parameter set-
tings (this window can also be opened by doubleclicking the icon).
UnlinkDelete the 754 icon from the system.
6.9.3 754 Dialysis Unit window
754 icon / Open
The 754 Dialysis Unit window for parameter settings is opened by
selecting this menu option with the right mouse button or by
double-clicking the 754 icon in SYSTEM window. It consists of the
three tabs Manual, Program, and Links.
80
IC Net 2.0
Manual
6.9 754 Dialysis Unit
The Manual tab of the 754 Dialysis Unit window for manual operation is only available for a connected system.
Pump
<On>Start pump drive.
<Off>Stop pump drive.
Start pump with startup hardware
Automatic start of pump drive with Startup
hardware.
Time program
On the Program tab of the 754 Dialysis Unit window a user-defined
time program can be entered. This program is started automatically as defined in the Start mode window (see section 4.4.3) either at the moment the determination is started (Start with determi-
nation) or at the moment the sample is injected (Start with inject).
First columnTime at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
IC Net 2.0
If no time is entered, the program instruction is applied together with the last instruction with time entry.
81
6 Devices
Second columnProgram instruction (see below).
Third columnParameter for program instruction (see
below).
ENABLEDEnable program start (a disabled program
is not started).
<Add>Add new program instruction.
<Delete>Delete selected program instruction.
<Verify>Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time program on the Program page:
Links
InstructionParameter entryMeaning
ProgramRESET, ENDSet program flag. The END flag
can be used to end a program,
especially if the program time
should be longer than the chromatogram duration. Additional
steps after this flag are not allowed. The RESET flag is used to
reset the parameters to the system startup values.
Pumpon, offSwitch on or off the pump drive.
The Links tab of the 754 Dialysis Unit is used for COM port selection and settings (details see section 5.2.4).
82
IC Net 2.0
6.10 750 Autosampler
6.10.1 750 Autosampler features
The Metrohm 750 Autosampler is a sampler for automatic sample injection for ion chromatography. It has a capacity of up to
128 samples each with an effective volume of ca. 700 µL.
The 750 Autosampler, which is normally operated using the keypad of the instrument, can be fully operated using the «IC Net»
program. The 750 Autosampler is normally connected to the 762
IC Interface using the 6.2134.000 cable and to the 733 IC Separation Center using the 6.2128.100 cable. Examples for interconnections with the 750 Autosampler can be found in the 762 In-structions for Use.
If the 750 Autosampler is connected to the 762 IC Interface, first
switch on the 762 IC Interface and then the 750 Autosampler. In
order to operate the 750 Autosampler with «IC Net», the remote
control via RS232 must be switched on using the [[ REM/REM/LOCLOC ] ]
key.
6.10 750 Autosampler
For detailed information about the 750 Autosampler see 750 Instructions for Use.
6.10.2 750 Autosampler icon
The 750 icon is available in the SYSTEM window if a 750 Autosam-
pler has been installed with the New system wizard or by using the
Setup/New devices/Install new device option of the SYSTEM window
(see section 6.1.2).
IC Net 2.0
If the system is connected and the 750 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 750 Autosampler window for parameter set-
tings (this window can also be opened by doubleclicking the icon).
UnlinkDelete the 750 icon from the system.
83
6 Devices
6.10.3 750 Autosampler window
750 icon / Open
The 750 Autosampler window for parameter settings is opened by
selecting this menu option with the right mouse button or by
double-clicking the 750 icon in the SYSTEM window. It consists of
the four tabs Manual, Injection, Diagnostics and Links.
Manual
The Manual tab of the 750 Autosampler window for manual operation is only available for a connected system.
84
Current data
Vial numberLive display of current vial number.
Needle positionLive display of current needle position.
Valve positionLive display of current valve position at the
733 IC Separation Center.
Valve position for rinse
Position of the injection valve at the 733 IC
Separation Center during rinsing.
Filter stopper for vials are used
If 6.2743.030 Filter stoppers are used, only
3 injections per sample vial are possible.
IC Net 2.0
6.10 750 Autosampler
Loop size isSize of the sample loop at the 733 IC
Separation Center. If the loop size is
> 20 µL, only 3 injections per sample vial
are possible.
<Move needle>Move transfer needle to the desired vial po-
sition (1...128 or Home).
<Lower needle>Lower transfer needle to the desired posi-
tion (1...4).
<Rinse needle>Start manual rinsing using the desired vial
(1...128).
<Change valve position>
Switch injection valve at the 733 IC Separation Center to the desired position (FILL or
INJECT).
<Disconnect>Enable manual operation of the 750 Auto-
sampler using the keypad.
For detailed information about the 750 Autosampler see 750 In-structions for Use.
Injection
The Injection tab of the 750 Autosampler window for performing
manual injections is only available for a connected system.
IC Net 2.0
85
6 Devices
Manual injection
Sample vial number
Number of sample vial to be used for manual injection.
<Select>Selection of the vial number (1...128).
Count of injections per vial
Number of injections per vial (1...4).
<Select>Selection of the injection number (1...4).
Needle lowering type
Injection numberPosition where the needle is moved to in
case of multiple injections (1...4).
Absolute positionPosition where the needle is moved to in
case of single injections (1...4).
Rinse type
Standard shiftUse previous or next vial for rinsing.
Nonstandard shiftFor every sample, skip this number of vials
and use this vial for rinsing.
No rinse between different sample vials
No rinsing.
Links
Data acquisition started by software
If this option is enabled, the data acquisition is started with "Inject". No cable is required between the injection valve and the
interface.
If this option is disabled, the 6.2115.070
cable is needed to start data acquisition.
<Inject>Start injection.
For detailed information about the 750 Autosampler see 750 Instructions for Use.
The Links tab of the 750 Autosampler is used for COM port selection and settings (details see section 5.2.4).
86
IC Net 2.0
6.11 766 IC Sample Processor
6.11.1 766 IC Sample Processor features
The Metrohm 766 IC Sample Processor is a sampler for automatic sample injection for ion chromatography. It has a capacity
of up to 127 samples each with an effective volume of up to
11 mL.
The 766 IC Sample Processor, which is normally operated using
the keypad of the instrument, can be fully operated using the «IC
Net» program. The 766 IC Sample Processor is normally connected to the 762 IC Interface using the 6.2134.080 or 6.2134.090
cable. If the injection valves at the 733 IC Separation Center are
to be controlled by the 766 IC Sample Processor, the 6.2141.110
cable is required for the connection of the 766 IC Sample Processor and the 732 IC Detector. Examples for interconnections
with the 766 IC Sample Processor can be found in the 762 In-structions for Use.
6.11 766 IC Sample Processor
In order to operate the 766 IC Sample Processor with «IC Net»,
the following RS settings have to be made using the [ CONFIG ]
key on the 766 keyboard:
>RS232 settings
handshake: SWChar
RS control: ON
For detailed information about the 766 IC Sample Processor see
766 Instructions for Use.
6.11.2 766 IC Sample Processor icon
IC Net 2.0
The 766 icon is available in the SYSTEM window if a 766 IC Sample
Processor has been installed with the New system wizard or by us-
ing the Setup/New devices/Install new device option of the SYSTEM
window (see section 6.1.2).
87
6 Devices
If the system is connected and the 766 icon is clicked with the
right mouse button, the following menu appears:
OpenOpen the 766 IC Sample Processor window for parame-
ter settings (this window can also be opened by double-clicking the icon).
UnlinkDelete the 766 icon from the system.
6.11.3 766 IC Sample Processor window
766 icon / Open
The 766 IC Sample Processor window for parameter settings and
manual control is opened by selecting this menu option with the
right mouse button or by double-clicking the 766 icon in the SYS-
TEM window. It consists of the four tabs Manual, Program, Configuration and Links.
Manual
The manual control of the 766 IC Sample Processor can be done
on the Manual subtab of the Manual tab which is only available for
a connected system.
88
Move
<Start>Start moving the rack to place the specified
vessel at the needle.
<Stop>Stop moving the rack to place the specified
vessel at the needle.
IC Net 2.0
6.11 766 IC Sample Processor
PositionSelection of the vessel to be placed at the
<Start>Switch on peristaltic pump at the tower for
the specified time.
<Stop>Switch off peristaltic pump at the tower.
TimeSelection of the pump time.
Selection: on (infinite), 1...999 s
Control
<Set>Send specified signal via remote control
lines of the 766 IC Sample Processor.
Control linesSelection of special remote line commands
or individual control line settings (details
see 766 Instructions for Use).
Program
<Start>Start program of the Program tab.
<Stop>Stop program of the Program tab.
Sample
<Set>Sample number for manual program start.
NumberSelection of the rack position.
Selection: 1...n
For detailed information about the 766 IC Sample Processor see
766 Instructions for Use.
IC Net 2.0
89
6 Devices
Status display
The current status of the 766 IC Sample Processor can be displayed on the Current state subtab of the Manual tab which is only
available for a connected system.
SampleDisplay of current sample position.
Beaker positionDisplay of current beaker position.
Lift positionDisplay of current lift position in mm.
Pump stateDisplay of current pump state (on, off).
Status of the 8 input lines
Status of the 14 output lines
<Ask>Read current values from 766 IC Sample
For detailed information about the 766 IC Sample Processor see
766 Instructions for Use.
Time program
On the Program tab of the 766 IC Sample Processor window a userdefined time program can be entered. This program is started
automatically as defined in the Start mode window (see section
4.4.3) either at the moment the determination is started (Start with
determination) or at the moment the sample is injected (Start with
inject).
Status of the remote input lines.
Status of the remote output lines.
Processor.
90
IC Net 2.0
6.11 766 IC Sample Processor
NumRow number (read-only).
TaskProgram instruction (see below).
ParameterParameter for program instruction (see
below).
EnabledEnable program start (a disabled program
is not started).
<Add>Add new program instruction.
<Delete>Delete selected program instruction.
List of program instructions
The following program instructions can be added to the time program on the Program tab:
Instruction Parameter entryMeaning
Movesample, sample+1,
spec.#, 1...n
Liftshift, work, rinse,
special,
0, 1...125 mm
Move rack to place the specified
vessel at the needle.
Raise or lower the lift to the specified
position.
IC Net 2.0
Pumpon, off, 1...999 sSwitch on/off peristaltic pump at
the tower.
Scan"scan command"Scan remote interface until the
specified signal is received. In addition to the normal scan commands (details see 766 Instruc-tion for Use) special commands
defined on the Configuration tab
can be used.
91
6 Devices
Ctrl"control
Wait1...9999 sWaiting time.
StartDAStart data acquisition.
Configuration
The Configuration tab of the 766 IC Sample Processor window contains parameters settings on the five subtabs Rack, Beakerposi-
tions, Needle positions, Control, and Scan.
Rack
command"
Set the specified remote output
signal. In addition to the normal
control commands (details see 766Instruction for Use) special
commands defined on the Configu-
ration tab can be used.
92
Rack definitions
CodeBinary ID code for rack identification (read-
only).
TypePredefined rack type (read-only).
Settings
Maximum lift wayDefinition of lowest lift position possible in
mm.
Entry range: 0...125 mm
<Apply>Send parameters to 766 IC Sample Proc-
essor.
<Ask>Read current parameters from 766 IC
Sample Processor.
IC Net 2.0
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.