Metrohm 767 User Manual

CH-9101 Herisau/Switzerland E-Mail info@metrohm.com Internet www.metrohm.com
767 Calibrated Reference
for mV, pH, S, :S, °C
8.767.1023 08.2005 / ars
Teachware Metrohm AG Oberdorfstrasse 68 CH-9101 Herisau teachware@metrohm.com
These instructions are protected by copyright. All rights reserved.
Although all the information given in these instructions has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please inform the author at the address given above.
Contents

Table of contents

1 Overview ............................................................... 1
1.1 Introduction ....................................................................................................1
1.2 Functional description ...................................................................................2
2 General instrument handling............................... 4
2.1 Storage ...........................................................................................................4
2.2 Maintenance ...................................................................................................4
2.3 Calibration ......................................................................................................4
2.4 ‘High-Impedance’, an important basic term’ ................................................4
2.5 Measurement of the insulation resistance ...................................................5
3 Procedure for checking instruments .................. 6
3.1 Basics .............................................................................................................6
3.2 pH Meters and Titrators.................................................................................6
3.2.1 U/mV, pH ..................................................................................................7
3.2.2 Polarization current and voltage source ..................................................8
3.2.3 Temperature (Pt 100/Pt 1000)..................................................................8
3.2.4 Tolerances ................................................................................................9
3.3 Conductivity meters .......................................................................................9
3.3.1 Conductance............................................................................................9
3.3.2 Temperature ...........................................................................................10
3.3.3 Tolerances ..............................................................................................10
3.4 Rancimat 617 and 679 ................................................................................ 10
4 Checking by means of the diagnosis
instructions ........................................................ 12
5 Appendix ............................................................. 13
5.1 Technical specifications ............................................................................. 13
5.1.1 Measuring source...................................................................................13
5.1.2 Temperature coefficient .........................................................................13
5.1.3
5.1.4 Ambient temperature..............................................................................14
5.1.5 Safety specifications ..............................................................................14
5.1.6 Electricity supply.....................................................................................14
5.1.7 Dimensions.............................................................................................14
5.2 Cables for connecting 767 – instrument X ................................................ 14
5.3 Standard equipment ................................................................................... 15
5.4 Warranty and conformity ............................................................................ 16
5.4.1 Warranty..................................................................................................16
5.4.2 Declaration of Conformity ......................................................................17
5.4.3 Quality Management Principles .............................................................18
Longterm stability (2 years)....................................................................13
6 Index ................................................................... 19
767 Calibrated Reference Instructions for Use I
Contents

List of illustrations

Fig. 1: 767 Calibrated Reference...............................................................................................1
Fig. 2 Functional scheme: Position "Voltage source ON" .........................................................1
Fig. 3: Label on cover ................................................................................................................1
Fig. 4: Unloaded potential source .............................................................................................5
Fig. 5: On-load potential source................................................................................................5
Fig. 6: Measurement of the insulation resistance versus earthing ...........................................5
Fig. 7: Cable arrangement for 1st G value (15.3 kS approx. 66 :S) ..................................11
Fig. 8: Cable arrangement for 2nd G value (14.3 kS approx. 69 :S) ................................11
II 767 Calibrated Reference Instructions for Use
1 Overview

1 Overview

1.1 Introduction

The measuring source 767.0010 Calibrated Reference for mV, pH, , µS, °C is a calibrated instrument for the quantities mentioned above.
It is connected instead of the electrodes and can be used for rapidly and easily checking of the functioning and the basic accuracy of most Metrohm instruments.
In addition the input resistance of high-impedance measuring amplifiers (pH Meters, Titrators) and, for separate amplifiers, the insulation of the reference point from the earth can be checked.
Fig. 1: 767 Calibrated Reference
U R G T U R G T
+U/1G
+U/direct
-U+/direct
(1)
(2)
(3)
(4)
(5)
(6)
solar cell
S1a
S1b
100 Ohm
1000 Ohm
1G
Pt 100
Pt 1000
Fig. 2 Functional scheme: Position "Voltage
source ON"
767 Calibrated Reference Instructions for Use 1
mV pH S :S °C mV pH S :S °C
100 10'000 0 100 10'000 0
1000 1000 0 1000 1000 0
(1200) 0 7 1G
1200 0 7 14'300 70
-341 12.7 0 7 460'000 2
Fig. 3: Label on cover
(exact values are given on the cover)
cover open ¦ cover closed
1 Overview

1.2 Functional description

As mentioned before the input resistance of high-impedance measuring amplifiers (pH Meters, Titrators) and, for separate amplifiers, the insula­tion of the reference point from the earth can be checked.
This is done using the potential of a reference diode (approx. 1200 mV) on the one hand at output socket (5) +U/direct and on the other hand a high-impedance resistor (1 G) at socket (4) +U/1 G. This potential is also switched to socket (6) U÷/direct by a divider. This means that a lower potential (approx. 341 mV) with inverted polarity is also avail­able; this can also be converted to a value within the pH scale (approx. pH 12.7).
The reference diode is fed by a solar cell. This means that neither a mains supply nor a battery is necessary and makes the instrument vir­tually maintenance-free. An internal potential monitor ensures that the output potential is switched off under inadequate lighting conditions be­fore the tolerance requirements are no longer fulfilled.
The solar cell can be covered and thus switched off. The potential monitor then switches a second electronic switch so that the internal resistance of the switched-off source is 14.3 k. This resistance can be used to check the current and voltage sources built into pH Meters and Titrators in a very simple manner. The voltage divider at socket (6)
U÷/direct gives a resistance of about 460 k, which can also be used for testing purposes.
For checking the temperature measuring amplifier the 0°C resistances of the temperature sensors Pt 100 and Pt 1000 are built-in, see sock­ets (1), (2), (3). These are separated from the other circuits in the test instrument. This means that no unwanted earthing loops can occur when they are used.
This means that 4 resistance values which can be used for checking conductivity meters.
For the 767 Calibrated Reference we have done without fine adjustment and instead have entered the resulting exact values in the table on the cover. In this way we have gained a considerable degree of accuracy and stability. In addition we have converted the resistance values into conductance (µS) and temperature (°C), and the potential into the exact pH value wherever this makes sense. This means that it is possible to compare the display of the instrument to be tested directly with the cor­responding value in this table. Two different tables are provided for the open and closed covers.
Practice has shown that problems are often caused by the electrode cable. They are subjected to mechanical stress (tension, pressure, tor­sion, etc.) and on the other hand they constantly and unavoidably come directly into contact with chemicals (spilt solutions, vapors, etc.). Never­theless their insulation value must always remain exactly as good as that at the measuring amplifier input. Such an exposed element must therefore be included in a test at all costs. This is why this measuring source is equipped with sockets which correspond to the plug head of
2 767 Calibrated Reference Instructions for Use
1 Overview
Metrohm electrodes and means that a test can be carried out very eas­ily:
Screw off cable at electrode plug-in head plug into Calibrated Reference measure
If this check produces a variation from the expected result then it is not immediately clear as to whether the error lies in the instrument to be checked or in the cable. Therefore specially labeled cables are avail­able in the accessories which are included with the instrument; these can be temporarily used instead of the original cable (see list of cables chap. 5.2). These accessory cables are also useful for sensors which do not have a plug-in head.
Note
It must be mentioned here that under no circumstances can or should the 767 Calibrated Reference replace the periodic maintenance of the instrument, but should only be used for determining whether an error is present or not if functional difficulties occur. In addition, the basic accuracy and high impedance of the instrument can be checked at regular intervals.
During maintenance the instrument is subjected to a far more strin­gent check (e.g. linearity of display and A/D converter, etc.). In addi­tion the switches, motors, mechanical components, etc. are also che­cked for corrosion and wear and tear.
767 Calibrated Reference Instructions for Use 3
Loading...
+ 16 hidden pages