Fig. 11:Interconnection with IC system without suppression...................................20
Fig. 12:Interconnection with IC system with suppression with 766 as "Master"......... 20
Fig. 13:Interconnection with IC system with suppression with PC as "Master".......... 21
Fig. 14:Interconnection with IC system for anion/cation system .............................. 21
Fig. 15:Interconnection with IC system with dialysis............................................... 22
Fig. 16:Connection possibilities for the RS232 interface ........................................ 23
Fig. 17:Schematic representation of the instrument dialog .....................................47
Fig. 18:Installation of the preconcentration column ............................................... 86
766 IC Sample Processor
1 Introduction
1.1 Instrument description
The 766IC Sample Processor can be used for automating ion chromatographic determinations, especially in combination with the
Metrohm IC system instruments. The 127 sample tubes with a volume
of up to 11 mL are arranged on the sample rack in three rows, which
guarantees easy access and unrestricted programming. Two additional
rinsing positions allow sample introduction free from crosscontamination even with widely varying sample matrices. If the 766 IC
Sample Processor is used together with the 732 IC Detector and the
733 IC Separation Center, various injections are possible from a single
sample tube, depending on the sample size. Sample tubes made of
polypropylene are standard. To protect the samples from external contamination, the tubes can be hermetically sealed.
1.1 Instrument description
Sample introduction from the 766 IC Sample Processor is achieved by
means of its integrated peristaltic pump. The sample is conveyed by
the pump through the capillary into the sample loop of the injector located within the 733 IC Separation Center. A steel needle for sealed or
a PEEK needle for open sample tubes can be used alternatively. Owing
to the relatively large volume of the sample tubes the 766 IC Sample
Processor can also be used for applications with enrichment columns,
dialysis (with the 754 Dialysis Unit) or simultaneous anion/cation determinations.
The sequences for the processing of each sample are freely definable
within broad boundaries. The same is true for the start and final sequences that are executed once either before or at the end of a sample
series. With the help of the LEARN mode, which is provided for creating
process methods, methods can be created easily and stored as user
methods.
Preprogrammed standard methods for the most common modes of
operation allow to use the 766 IC Sample Processor directly, with only
little programming effort. The time frame can be given by the IC Sample
Processor; alternatively, the whole sequence can be managed by the
«IC Metrodata» remote control software via the «Queue» (sample table). The RS232 interface built in as standard allows remote control of
the IC Sample Processor from a PC.
766 IC Sample Processor
1
1 Introduction
1.2 Parts and controls
5566778899
44
33
22
11
Always install splash
protection!
1010
1111
1212
1313
1414
1515
1616
1717
1111
2828
2727
2626
2525
2424
MetrohmMetrohm
DES FUSIBLES DE 250V
USE ONLY WITH 250V
100-120V
UNIQUEMENT AVEC
FUSES/EMPLOYER
220-240V
Always install plug cover!
Fig. 1: Side view of the 766 IC Sample Processor
1818
1919
2020
2121
2222
2323
2
766 IC Sample Processor
1.2 Parts and controls
11Splash protection
(6.2751.040)
Must be installed always in
operation!
22Needle
Steel needle (6.2624.000) or
PEEK needle (6.1835.000)
33Lift
With swing head attached
44Needle holder
(attached)
55PEEK compression fitting
(4.766.4070)
For peek capillary 88
66PEEK compression fitting
(6.2744.010)
1515Contact pressure lever
For adjusting the contact pressure
1616Holding clamp
For locking the tubing cartridge into
place
1717Snap-action lever
For releasing the tubing cartridge
1818PEEK capillary (6.1831.060; 1 m)
For conveying the sample to the
injection valve of the 733 IC Separation Center
1919Tower
2020PE bottle (6.1608.080; 300 mL)
77Swing head
(attached)
88PEEK capillary (6.1831.050;
40 cm)
Connection needle – pump
tubing
99Guide chain
For fixing tubings and cables
1010Pump tubing (6.1826.040)
For conveying the sample
1111Screw thread for splash
protection
1212Mounting pin
For attaching the tubing
cartridge
2121Connection cable for swing head
Attached, incl. branch plug for remote
connection
2222Remote connection
2323Plug cover (6.2752.010)
2424Fuse holder
Changing the fuses, see section 2.2.2
2525Mains connection plug
Mains connection, see section 2.2.3
2626Mains switch
For switching the instrument on/off:
I = ON 0 = OFF
Please read through these Instructions for Use carefully before operating the 766 IC Sample Processor. The Instructions for Use contain
information and warnings to which the user must pay attention in order
to assure safe operation of the instrument.
1.3.1 Organization
These 8.766.1003 Instructions for Use for the 766 IC Sample Processor provide a comprehensive overview of the installation, startup procedure, operation, fault rectification and technical specifications of this
instrument. The Instructions for Use are organized as follows:
Section 1 Introduction
Description of the instrument, parts and controls,
safety notes
Section 2 Installation
Mains connection, attachment of accessories,
connection of external devices
Section 3 Operating tutorial
Introduction to the operation
Section 4 Operation
Detailed description of display, keyboard, methods,
manual operation, sample racks, standard methods
Section 5 Interfaces
Remote interface, RS232 interface and
remote control language
Section 6 Appendix
Error messages, technical data, maintenance
and servicing, diagnosis, warranty, declaration
of conformity, standard equipment, optional
accessories, index
To find the information you require about the instruments please use
either the Table of contents or the Index at the back. The 8.766.1013Quick Reference Guide is suitable for use as a reference work for
daily use as it explains the most important parameters and key functions.
766 IC Sample Processor
5
1 Introduction
1.3.2 Notation and pictograms
The following notations and pictograms (symbols) are used in these Instructions for Use:
<PUMP>Switch or key
1515Part or control of 766
8989Part or control of 732/733
2626Part or control of 754
'rack number'Parameter or entry value
at 766 IC Sample Processor
******** counter 0/127
PUMP- ready
Display
Text in keyboard display of 766
IC Sample Processor
Hazard
This symbol draws attention to
a possible danger to life or
injury if the associated
directions are not followed
correctly.
Warning
This symbol draws attention to
possible damage to instruments
or instrument parts if the associated directions are not followed correctly.
Caution
This symbol marks important
information. Read these directions before continuing.
Comment
This symbol marks additional
information and tips.
6
766 IC Sample Processor
1.4 Support documentation
1.4.1 Application Bulletins
The «Application Bulletin» is a collection of analytical methods, application examples and literature references. Of Metrohm's approximately
200 Application Bulletins, 34 currently refer to ion chromatography. All
these Application Bulletins are available on request free of charge from
your Metrohm supplier.
You will find an updated list of the Application Bulletins at any time under «http://www.metrohm.ch».
1.4.2 Application Notes
The «Application Notes» present application information in concentrated form, i.e. on maximum 2 pages. There are currently 88 Application Notes (in English) in the field of ion chromatography. You can order
these free of charge from your Metrohm supplier or view them in the
Internet under «http://www.metrohm.ch» and copy them from there.
1.4 Support documentation
1.5 Safety notes
1.5.1 Electrical safety
While electrical safety in the handling of the 766 IC Sample Processor is
assured in the context of the specifications IEC 1010-1 (protection class
1, degree of protection IP40), the following points should be noted:
• Mains connection
Set the mains voltage and check the mains fuse and mains
connection in accordance with the instructions in section 2.2.
• Opening the 766 IC Sample Processor
To avoid all danger of coming into contact with live components do
not open the instrument or remove any parts when the 766 IC Sample
Processor is connected to the power supply. Always disconnect the
instrument from all voltage sources before you open it and ensure that
the mains cable is disconnected from mains connection 2525 !
766 IC Sample Processor
7
1 Introduction
• Protection against static charges
Electronic components are sensitive to static charging and can be
destroyed by discharges. Before you touch any of the components
inside the 766 IC Sample Processor, you should earth yourself and
any tools you are using by touching an earthed object (e.g. housing of
the instrument or a radiator) to eliminate any static charges which
exist.
1.5.2 General safety rules
• Install splash protection
To avoid any danger of injury by the needle, the splash protection
must always be installed when operating the 766 IC Sample Processor!
• Install plug cover
To prevent any contamination of the mains and remote connection by
spilled solvents or chemicals, the plug cover must always be installed
when operating the 766 IC Sample Processor !
• Do not use caps together with the PEEK tube
If you are using the 6.1835.000 PEEK needle, the sample tubes may
not be sealed with caps because they cannot be pierced by the PEEK
needle and the needle may be damaged thereby.
• Handling of solvents
Check the pump tubing and all input and output leads periodically for
possible leaks. Follow the relevant instructions regarding the handling
of flammable and/or toxic solvents and their disposal.
• Regular exchange of pump tubings
Pump tubings constitute consumable material and must be replaced
from time to time (see section 6.3.3). Suitable measures must be
taken so that any leak which might occur in the pump tubing or
connections during unattended operation will cause no damage
(collection device for any liquid which may leak out).
8
766 IC Sample Processor
2 Installation
2.1 Setting up the instrument
2.1.1 Packaging
The 766 IC Sample Processor is supplied together with the separately
packed accessories in special packagings containing shock-absorbing
foam linings designed to provide excellent protection. The actual instrument is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as only they
can assure damage-free transport of the instrument.
2.1 Setting up the instrument
2.1.2 Check
After receipt, immediately check whether the shipment is complete and
undamaged (compare with delivery note and list of accessories in sec-
tion 6.8). In the case of transport damage, see instructions in section
6.7.1 "Warranty".
2.1.3 Location
Position the instrument in the laboratory at a location convenient for operation, free from vibrations and protected against a corrosive atmosphere and contamination by chemicals.
Do not operate the 766 IC Sample Processor without splash protection 11 and plug cover 2323 being installed!
2.1.4 Arrangement of the instruments
To make the connection between needle 22 and the injection valve at
the 733 IC Separation Center with the peek capillary 1818 (length 1 m)
supplied, position the 766 IC Sample Processor immediately on the left
side of the IC system.
766 IC Sample Processor
Take precautions to ensure that any leaks from pump tubings or
connections cannot cause more damage.
9
2 Installation
2.2 Mains connection
Follow the instructions below for connecting to the power supply. If
the instrument is operated with the mains voltage set wrongly and/or
wrong mains fuse there is a danger of fire!
2.2.1 Setting the mains voltage
Before switching on the 766 IC Sample Processor for the first time,
check that the mains voltage set on the instrument (see Fig. 2) matches
the local mains voltage. If not, reset the mains voltage on the instrument as follows:
Disconnect mains cable
1
Disconnect mains cable from mains connection plug 2525 of the
766 IC Sample Processor.
Remove fuse holder
2
Using a screwdriver, loosen fuse holder 2424 beside the mains
connection 2525 and take out completely.
Check fuse
3
Carefully take the fuse installed for the desired mains voltage out
of fuse holder 2424 and check its specifications (the position of the
fuse in the fuse holder is marked by the white arrow imprinted
next to the mains voltage range):
100……120 V 0.5 A (slow-blow)Metrohm-Nr. U.600.0013
220……240 V 0.25 A (slow-blow) Metrohm-Nr. U.600.0010
Insert fuse
4
Change fuse if necessary and reinsert in fuse holder 2424.
Install fuse holder
5
Depending on the desired mains voltage, insert fuse holder 2424 in
the 766 IC Sample Processor so that the corresponding mains
voltage range can be read normally and the adjacent white
arrow points to the white bar imprinted on the right side of the
fuse holder (see Fig. 3).
10
766 IC Sample Processor
2.2 Mains connection
100 – 120 V
2626
2525
2424
2.2.2 Fuses
220 – 240 V
220
-
- 240 V
-
- 120 V
100
100
-
- 120 V
-
- 240 V
220
2424 Fuse holder
2525 Mains connection
plug
2626 Mains switch
Fig. 3: Setting the mains voltage
One of the two fuses 0.5 A/slow-blow for 100…120 V or 0.25 A/slowblow for 220…240 V is installed in fuse holder 2424 of the 766 IC Sample
Processor as standard.
Ensure that the instrument is never put into operation with fuses of
another type, otherwise there is danger of fire!
For checking or changing fuses, proceed as described in section 2.2.1.
2.2.3 Mains cable and mains connection
Mains cable
The instrument is supplied with one of three mains cables
• 6.2122.020 with plug SEV 12 (Switzerland, …)
• 6.2122.040 with plug CEE(7), VII (Germany, …)
• 6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin. If a
different plug has to be fitted, the yellow/green lead (IEC standard)
must be connected to protective earth (protection class 1).
Any break in the earthing inside or outside the instrument can make it
a hazard!
Mains connection
Plug the mains cable into mains connection plug 2525 of the 766 IC
Sample Processor (see Fig. 3).
2.2.4 Switching the instrument on/off
The 766 IC Sample Processor is switched on and off using mains
switch 2626.
766 IC Sample Processor
11
2 Installation
2.3 Attaching the accessories
For attaching the accessories at the 766 IC Sample Processor, proceed in the order described below.
2.3.1 Connecting the swing head
Plug in the branch plug of the connection cable 2121 permanently at-
tached to the swing head into the remote connection socket at the 766
IC Sample Processor and screw it onto this connection using a screwdriver (see Fig. 1).
2.3.2 Connecting the keyboard
Connecting the keyboard
1
Connect the 6.2142.010 keyboard
to the keyboard connection 3131
"Keyboard". For disconnection,
press the plug together slightly on
both sides.
Switch on instrument
2
Switch on the 766 IC Sample
Processor with mains switch 2626.
The keyboard display lights up.
The instrument is initialized and
the lift is raised completely.
2.3.3 Installing the plug cover
Fig. 4: Keyboard connection
12
To prevent any contamination of the mains and remote connection by
spilled solvents or chemicals, the 6.2752.010 plug cover must always
be installed when operating the 766 IC Sample Processor !
Install the plug cover 2323 in the corresponding guide groove above
mains connection plug 2525 and remote connection 2222 (see Fig. 1 andFig. 2).
766 IC Sample Processor
2.3.4 Installing the splash protection
To avoid any danger of injury by the needle, the 6.2751.040 splash
protection must always be installed when operating the 766 IC Sample Processor!
Remove holding screws
1
Remove the holding screws 2929 and the washer mounted on the
screw threads 1111 at tower 1919 using the 6.2621.100 Allen key.
Remove protective film from splash protection
2
Remove the plastic film glued on both sides of the splash protection 11.
Install splash protection
3
Attach splash protection 11 onto the screw threads 1111 at the
tower 1919 and fix it with the holding screws 2929 and the washer
using the 6.2621.100 Allen key.
Insert needle 22 (6.2624.000 steel needle or 6.1835.000 PEEK
needle) completely into the opening of the needle holder 44.
2929
11
2929
Fig. 5: Installing the splash protection
766 IC Sample Processor
Fix needle
3
Tighten compression fitting 55 in needle holder 44 by hand (never
use tools!).
13
2 Installation
22 Steel needle (6.2624.000) or
PEEK needle (6.1835.000)
88
77
44 Needle holder
66
55 PEEK compression fitting
(4.766.4030)
66 PEEK compression fitting
(6.2744.010)
77 Swing head
88 PEEK capillary (6.1831.050)
If you are using the 6.1835.000 PEEK needle, the sample tubes may
not be sealed with caps because they cannot be pierced by the PEEK
needle and the needle may be damaged thereby.
55
44
22
Fig. 6: Needle installation
2.3.6 Placing the sample rack
Place sample rack
1
Place sample rack 2727 on the
turntable of the 766 IC Sample
Processor acc. to Fig. 7.
Read magnetic code
2
Press <RESET> to move the
rack to the home position, in
which the magnetic code can
be read (details see section
4.5).
2.3.7 Adjusting the sample rack
If a new sample rack is placed on the 766 IC Sample Processor for the
first time, it must be adjusted on a rack position in the middle row
(example: position 37) as follows:
Fig. 7: Sample rack placing
14
Place sample rack
1
Place empty sample rack 2727 on the 766 IC Sample Processor
and press <RESET> (see section 2.3.6).
766 IC Sample Processor
2.3 Attaching the accessories
22 Steel neddle
(6.2624.000) or PEEK
needle (6.1835.000)
2727 Sample rack
(6.2041.430)
3232 Adjusting position 37
3333 Adjusting screw
Move to adjusting position
2
Press <MOVE>, enter number '37' and confirm with
<ENTER>. Sample rack and swing head are turned until needle
22 is above the adjusting position 3232 (position 37: first opening of
the middle row).
Check needle position
3
• Press <ê> until the needle 22 is ca. 1 cm above the upper
level of the sample rack 2727.
• Check needle position: If the needle 2 cannot be lowered
unhindered through the upper hole of the adjusting position
32, continue directly with point 4.
• Continue lowering the needle by pressing <ê> until the
needle 22 is ca. 1 cm above the lower level of the sample rack
2727.
• Check needle position: If the needle 2 cannot be lowered
unhindered through the lower hole of the adjusting position
32, continue directly with point 4.
• Lower needle completely by pressing <ê>.
• Check needle position: If the needle 22 is in the middle of the
lower hole, the sample rack must not be adjusted (continue in
this case with point 5).
22
3232
3333
2727
Fig. 8: Adjusting the sample rack
766 IC Sample Processor
Adjust sample rack
4
• Loosen the three adjusting screws 3333 on the lower level of the
sample rack using the 6.2621.100 Allen key
• Carefully turn the two upper levels of the sample rack 2727 by
hand until the lowered needle 22 is exactly in the middle of the
lower hole at the adjusting position 3232.
• Tighten the adjusting screws 3333.
Move to rest position
5
Press <RESET> to move the sample rack to the rest position.
15
2 Installation
2.3.8 Tubing connections 766 – 733
For transferring the sample from the 766 IC Sample Processor to the
injection valve of the 733 IC Separation Center the following tubing
connections must be made:
Pump tubings are consumable material with a lifetime which depends
on the contact pressure (see section 6.3.3). This is why the tubing
cartridges should be raised completely by loosening snap-action lever
1717 on the right-hand side if the pump is to remain switched off for a
considerable length of time (the set contact pressure remains unchanged).
The 6.1826.0X0 pump tubing is made of PVC and must not be used
for rinsing with solutions which contain acetone. In such cases, rinse
with different pump tubing or a different pump.
Install pump tubing
1
• Release tubing cartridge 1313 from holding clamp 1616 by
pressing down snap-action lever 1717 and remove from
mounting pin 1212 on the 766 IC Sample Processor (see Fig.
1).
• Press contact pressure lever 1515 on the tubing cartridge down
as far as it will go.
• Insert pump tubing 1010 in the tubing cartridge 1313 as shown inFig. 9. The black-black stopper 3636 must click into the corre-
sponding holder on the left-hand side of the tubing cartridge.
• Place the tubing cartridge 1313 on mounting pin 1212 and press
down on the right-hand side until snap-action lever 1717 clicks
into position on holding clamp 1616. Take care that no kinks are
formed in the pump tubing.
Connection needle – pump tubing
2
• Mount the PEEK compression fitting 66 on one end of the
PEEK capillary 88 and the PEEK compression fitting 3434 on the
other end.
• Screw the PEEK capillary 88 with the PEEK compression fitting
66 on to the PEEK compression fitting 55 already mounted onneedle holder 44 (see Fig. 6).
• Insert PEEK capillary 88 into the guide chain 99 (see Fig. 1 and
section 2.3.8).
• Screw a coupling 3535 on to the PEEK compression fitting 3434 at
the other end of PEEK capillary 88.
• Push coupling 3535 on to the inlet end of the pump tubing 1010
(see Fig. 6).
• At the 733 IC Separation Center, loosen the rotary nipple
screwed onto the interior side of connection 2222 or 2828.
• Take PTFE suction tubing 8484 (see Fig. 14 and Fig. 15 of the
732/733 Instructions for Use) completely out of connection 2222
or 2828 and unscrew from connection "1" of injection valve 6666.
• Pull the PEEK capillary 1818 through the opening 2222 or 2828 of
the 733 IC Separation Center and screw onto connection "1"
of injection valve 6666 using a 6.2744.010 PEEK compression
fitting.
• Retighten rotary nipple on the interior side of connection 2222
or 2828 to fix the capillary 1818.
766 IC Sample Processor
Tubing connection injection valve – waste
4
• Insert 6.2744.020 coupling (from 733 accessories) into con-
nection 2121 or 2727 of the 733 IC Separation Center.
• Screw PTFE suction tubing 8484 onto the 6.2744.020 coupling
attached to connection 2121 or 2727 and lead it into the waste
container.
In the case of the 733.0020 IC Separation Center with two injection
valves, it is possible to fill both sample loops from the same 766 IC
Sample Processor. For this, connection "1" of valve A (outlet of the
sample loop) must be connected to connection "2" of valve B (inlet of
the sample loop) using a 6.1803.040 PEEK capillary (15 cm).
17
2 Installation
2.3.9 Tubing connections 766 – 754
If the 766 IC Sample Processor is used for an IC system with sample
dialysis (see section 2.4.6), the peristaltic pump at the 754 Dialysis Unit
is used for sample conveying instead of the pump at the 766 IC Sample
Processor. The tubing connections between the 754 Dialysis Unit 754
and the 733 IC Separation Center 733 have to be made as shown in
Fig. 8 of the 754 Instruction for Use. The only change concerns the inlet
of the sample, which is installed as follows:
Install a PEEK compression fitting 6.2744.010 on the inlet end of the
PTFE tubing 2626 (6.1803.030, see Fig. 8 of the 754 Instruction for Use)
connected to the 754 Dialysis Unit and screw it on the PEEK compression fitting 55 (see Fig. 6).
2.3.10 Fixing tubing and cables
In order to fix tubing or cables in
the guide chain 99 any chain link
may be opened with a screw
driver or another appropriate tool.
Fig. 10: Opening chain links
18
766 IC Sample Processor
2.4 Connection of devices to the remote interface
732/1 Remote
2.4 Connection of devices to the
remote interface
2.4.1 General information on remote interface
The branch plug of cable 2121 leading from the swing head (see Fig. 1) is
plugged into the 25 pin remote interface (see section 2.3.1). Any external devices can be connected to remote connection 2222 of this branch
plug. The 766 IC Sample Processor can be remote controlled via the 8
input lines, the 14 output lines can be used to control external devices.
Before an external device is connected to remote connection 2222, the
766 IC Sample Processor must always be switched off using mains
switch 2626!
The pin assignment of the remote interface, its functions, the electrical
requirements and conditions are described in section 5.1.
2.4.2 Connection cables
Connecting peripheral instruments to the 766 IC Sample Processor requires Metrohm cables. Otherwise a safe data transmission may not be
guaranteed.
Metrohm cables are labeled with the type of the instrument which they
may be connected with and optionally with the particular socket. Mind
the cable ends.
Example:
766 IC Sample Processor
19
2 Installation
2.4.3 IC system without suppression
The 766 IC Sample Processor is connected to an IC system without
suppression consisting of 732 IC Detector, 733 IC Separation Center
and 709 IC Pump as shown in Fig. 11 using the 6.2141.110 cable. With
this interconnection the standard methods 'PC', 'PCSeg', 'SP' and
'SP Seg' can be used (see section 4.6).
732
733
6.2125.060 cable
709
766
6.2141.110 cable
Fig. 11: Interconnection with IC system without suppression
2.4.4 IC system with suppression with 766 as "Master"
The 766 IC Sample Processor is connected to an IC system with suppression consisting of 732 IC Detector, 733 IC Separation Center, 709
IC Pump and either 752 Pump Unit or 753 Suppressor Module as
shown in Fig. 12 using the 6.2125.120 adaptor. With this interconnection, in which the 766 IC Sample Processor is the "Master", the standard
methods 'SP' and 'SPSeg' can be used (see section 4.6).
6.2125.090 cable
20
732
733
709
766
6.2125.120 adaptor
6.2125.090 cable
6.2143.210 cable
6.2125.090 cable
6.2141.110 cable
6.2125.060 cable
752
753
6.2143.200 cable
Fig. 12:Interconnection with IC system with suppression with
766 as "Master"
766 IC Sample Processor
2.4 Connection of devices to the remote interface
2.4.5 IC system with suppression with PC as "Master"
The 766 IC Sample Processor is connected to an IC system with suppression consisting of 732 IC Detector, 733 IC Separation Center, 709
IC Pump and either 752 Pump Unit or 753 Suppressor Module as
shown in Fig. 13 using the 6.2125.120 adaptor. With this interconnection, in which the PC is the "Master", the standard methods 'PC',
'PC Seg' and 'Preconc' can be used (see section 4.6).
732
733
709
766
6.2141.110 cable
6.2125.120 adaptor
Fig. 13:Interconnection with IC system with suppression with
PC as "Master"
2.4.6 IC system for simultaneous anion/cation
determination
The 766 IC Sample Processor is connected to an IC system for simultaneous determination of anions and cations consisting of two 732 IC
Detectors, the 733 IC Separation Center, two 709 IC Pumps and (if
suppression is used) the 753 Suppressor Module as shown in Fig. 14
using the 6.2125.120 adaptor. With this interconnection the standard
methods 'AnCat' and 'AnCatSeg' can be used (see section 4.6).
6.2125.090 cable
6.2125.090 cable
6.2125.060 cable
752
753
6.2143.210 cable
6.2143.200 cable
6.2141.110 cable
766
6.2125.090 cable
766 IC Sample Processor
732/1
733
709/1
6.2125.120 adaptor
6.2143.210 cable
732/2
6.2125.090 cable
709/2
6.2125.060 cable
753
Fig. 14: Interconnection for anion/cation system
6.2125.090 cable
6.2125.060 cable
6.2141.120 cable
21
2 Installation
2.4.7 IC system with sample dialysis
The 766 IC Sample Processor is connected to an IC system with sample dialysis consisting of 732 IC Detector, 733 IC Separation Center,
709 IC Pump, 754 Dialysis Unit and (if suppression is used) the 753
Suppressor Module as shown in Fig. 15 using the 6.2125.120 adaptor.
With this interconnection the standard method 'Dialysis' can be used
(see section 4.6). If no suppression is used, the 754 Dialysis Unit can
be connected to the 6.2125.120 adaptor using the 6.2143.200 cable.
732
733
709
766
6.2125.120 adaptor
6.2125.090 cable
6.2125.090 cable
6.2125.060 cable
754
753
6.2143.220 cable6.2141.110 cable
Fig. 15: Interconnection with IC system with dialysis
22
766 IC Sample Processor
2.5 Connection of devices to the RS232 interface
6.2125.060 cable
Printer cables, see section 2.5.2
2.5 Connection of devices to the RS232
interface
2.5.1 General information on RS232 interface
Many different instruments may be connected via the serial RS232 interface 3232. In addition to all Metrohm instruments that support the
Metrohm remote control language (see section 5.2) any printer with serial interface (or parallel interface and parallel/serial converter) or a personal computer (PC) may be connected. Any other measuring instrument may be controlled via RS232 interface, as long as it supports
serial data transmission.
766
732
Fig. 16: Connection possibilities for the RS232 interface
In order to guarantee safe data transmission, it is important to set the
same RS232 interface parameters correctly for both instruments connected (see section 2.5.2).
Control commands (examples)
CTL:RS &M;$Gstarts a Metrohm instrument
CTL:RS &M;$Sstops a Metrohm instrument
PRINT: configprints a configuration report to
a printer or PC
766 IC Sample Processor
Scanning input data (example):
SCN:RS : *R"waiting for readiness of a Metrohm
instrument
23
2 Installation
2.5.2 Connection of a printer
Printers with the following printer drivers may be connected:
IBMIBM Proprinter and printers with IBM emulation
EpsonEpson printers and printers with Epson emulation
SeikoSeiko printers DPU-411 and DPU-414
CitizenCitizen printer IDP562-RS
HPHP printers and compatibles with HP PCL3 emulation
If you connect other printers, ensure that these emulate a printer mode
supported by the 766 IC Sample Processor.
Most printers with a serial interface are connected using the 6.2125.050
Cable. Printers with a parallel interface need a serial/parallel converter
(e.g. 2.145.0300) and the 6.2125.020 Cable.
Before connecting a printer to the RS232 interface, switch off the 766
IC Sample Processor !
The interface parameters are set in the configuration submenu ">RS232
settings".
The following table provides information on a few selected printers.
PrinterCableRS232 settingsSettings on printer
IBM Proprinter6.2125.050
Seiko
DPU-411
Seiko
DPU-414
6.2125.020baud rate: 9600
6.2125.130baud rate: 9600
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: IBM
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: Seiko
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: Seiko
see printer manual
Settings of the DIP switches:
DIP01DIP02
on
off
1 2 3 4 5 6 7 81 2 3 4 5 6
The switchable 7-bit ASCII character set of
the printer will be automatically set by the
732 IC Detector to the national character
sets in accordance with the set dialog
language.
The switchable 7-bit ASCII character set of
the printer will be automatically set by the
732 IC Detector to the national character
sets in accordance with the set dialog
language.
24
766 IC Sample Processor
2.5 Connection of devices to the RS232 interface
PrinterCableRS232 settingsSettings on printer
Citizen
IDP562-RS
6.2125.050baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set:Citizen
Epson
with 6-pin round
connector
6.2125.040baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: Epson
Epson
with additional
serial interface
#8148
6.2125.050baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: Epson
Epson LX-3006.2125.050baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: Epson
Epson and
Canon printers
with parallel
interface
HP Deskjet
with serial
interface
HP Laserjet
with serial
interface
6.2125.020
+
2.145.0300
serial/
parallel
converter
6.2125.050
or adaptor
cable 25-pin
neg./9-pin
pos. (e.g. HP
C2933A)
Adaptor
cable 25-pin
neg./9-pin
pos. (e.g. HP
C2933A)
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: Epson
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: HP
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: HP
HP Deskjet/
Laserjet
with parallel
6.2125.020
+
2.145.0300
serial/
parallel
converter
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: Hws
character set: HP
Settings of the DIP switches:
ON
1 2 3 4 5 6 7 8 9
SSW1
The switchable 7-bit ASCII character set of
the printer can be changed to the national
character sets only by setting Dip switch 4
and 5:
4455Character setCharacter set
OFFOFFUSA
ONONGreat Britain
ONOFFFrance
OFFONGermany
Spanish does not have its own character set
(it is best to select French).
Settings of the DIP switches:
SW1SW2
on
off
1 2 3 4 5 6 7 81 2 3 4 5 6 7 8
Settings of the DIP switches on the interface:
SW1SW2
on
off
1 2 3 4 5 6 7 81 2 3 4 5 6
see printer manual
see printer manual
Settings of the DIP switches:
AB
on
off
1 2 3 4 5 6 7 81 2 3 4 5 6 7 8
see printer manual
see printer manual
10
766 IC Sample Processor
25
Loading...
+ 113 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.