In all communications with INFICON, please specify the information given on the product nameplate. For convenient reference copy that information into the space provided below.
INFICON AG, LI-9496 Balzers
Model:
PN:
SN:
V W ; LPS
2
tina60e1-a (2011-04)
Validity
This document applies to products of the PSG55x series.
Part numbers of standard products are indicated below. OEM
products have other part numbers and different parameter
settings (e.g. factory setting of setpoint) as defined in the
corresponding ordering information.
3PIx-0xx-xxxx
Flange
1 ⇒ DN 16 ISO-KF
2 ⇒ DN 16 ISO-KF long tube
4 ⇒ DN 16 CF-F
5 ⇒ DN 16 CF-R long tube
6 ⇒ DN 25 ISO-KF
D ⇒ 4 VCR female
E ⇒ 8 VCR female
F ⇒ 1/8" NPT
⇒ Flange 29×29 mm
K
M ⇒ 4 VCR 90° female
N ⇒ 7/16-20 UNF male
Unit
0 ⇒ mbar
1 ⇒ Torr
2 ⇒ Pa
3 ⇒ micron
Filament
Tungsten
⇒
1
2 ⇒ Nickel
⇒
coated
Al
3
2O3
⇒
6
7 ⇒ Nickel, galvanically isolated
8 ⇒ Al
galvanically isolated
Tungsten,
coated, galvanically isolated
2O3
tina60e1-a (2011-04)3
3PIx-0xx-xxxx
Measuring
signal
Interface
Receptacle
Display
Switching function
*) Solid state relays
**) Mechanical relays
The part number (PN) can be taken from the product nameplate.
If not indicated otherwise in the legends, the illustrations in this
document correspond to gauges with the DN 16 ISO-KF vacuum
connection and display. They apply to gauges with other vacuum
connections by analogy.
We reserve the right to make technical changes without prior
notice.
The Pirani Standard Gauge PSG55x has been designed for
vacuum measurement of gases in the pressure range of
-5
5×10
… 1000 mbar.
They must not be used for measuring flammable or combustible
gases in mixtures containing oxidants (e.g. atmospheric oxygen)
within the explosion range.
The gauge is intended for operation in connection with an
INFICON Vacuum Gauge Controller VCG401, 402, 403 or with
another suitable controller.
Trademark
VCR® Swagelok Marketing Co.
Patents
EP 0689669 B1, 0689670 B1
US Patent 5608168
Scope of Delivery
1× gauge
1× pin for adjusting settings via buttons
1× Operating Manual English
1× Operating Manual German
tina60e1-a (2011-04)5
Contents
Product Identification 2
Validity 3
Intended Use 5
Trademark 5
Patents 5
Scope of Delivery 5
1
Safety 8
1.1 Symbols Used 8
1.2 Personnel Qualifications 8
1.3 General Safety Instructions 9
1.4 Liability and Warranty 9
2 Technical Data 10
2.1 Output Signal vs. Pressure 18
2.2 Gas Type Dependence 23
3 Installation 24
3.1 Vacuum Connection 24
3.2 Power Connection 27
3.2.1 FCC 68, 8-pin Connector 28
3.2.2 D-Sub, 9-pin Connector 29
3.2.3 D-Sub, 15-pin HD Connector 30
3.2.4 D-Sub, 15-pin HD, RS485 INF Connector 31
3.2.5 DeviceNet Connector 32
3.2.6 Profibus Connector 33
4 Operation 34
4.1 Status Indication and Displays 34
4.2 Gas Type Dependence 38
4.3 Switching Functions SP1, SP2 38
4.4 Diagnostic Port (RS232C Interface) 44
4.5 DeviceNet Operation 45
4.6 Profibus Operation 48
5 Deinstallation 49
6
tina60e1-a (2011-04)
6 Maintenance, Repair 51
6.1 Adjusting the Gauge 51
6.2 Replacing the Sensor 53
6.3 Troubleshooting 54
7 Returning the Product 55
8 Disposal 56
9 Accessories 57
10 Spare Parts 58
Further Information 61
ETL Certification 62
EC Declaration of Conformity 63
For cross-references within this document, the symbol (→ XY)
is used, for cross-references to further documents, listed under
"Further Information", the symbol (→ [Z]).
tina60e1-a (2011-04)7
1 Safety
1.1 Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and environmental damage.
Caution
Information on correct handling or use. Disregard can lead to
malfunctions or minor equipment damage.
Notice
<…> Labeling
1.2 Personnel Qualifications
Skilled personnel
All work described in this document may only be carried out by
persons who have suitable technical training and the necessary experience or who have been instructed by the end-user
of the product.
8
tina60e1-a (2011-04)
1.3 General Safety Instructions
• Adhere to the applicable regulations and take the necessary
precautions for the process media used.
Consider possible reactions with the product materials.
Consider possible reactions (e.g. explosion) of the process
media due to the heat generated by the product.
• Adhere to the applicable regulations and take the necessary
precautions for all work you are going to do and consider the
safety instructions in this document.
• Before beginning to work, find out whether any vacuum components are contaminated. Adhere to the relevant regulations
and take the necessary precautions when handling contaminated parts.
Communicate the safety instructions to all other users.
1.4 Liability and Warranty
INFICON assumes no liability and the warranty becomes null
and void if the end-user or third parties
• disregard the information in this document
• use the product in a non-conforming manner
• make any kind of interventions (modifications, alterations etc.)
on the product
• use the product with accessories not listed in the product
documentation.
The end-user assumes the responsibility in conjunction with the
process media used.
Gauge failures due to contamination or wear and tear, as well as
expendable parts (e.g. filament) are not covered by the warranty.
tina60e1-a (2011-04)9
2 Technical Data
For further technical data for gauges with serial inter-
faces → [3], [4], [5].
Measurement principle thermal conductance acc. to
Measurement range 5×10
Accuracy (N
5×10
1×10
)
2
-4
… 1×10-3 mbar
-3
… 100 mbar
100 … 1000 mbar
Resolution (at 1000 mbar)
Repeatability (N
1×10
)
2
-3
… 100 mbar
Output signal (measurement signal)
Voltage range
3PIx-0xx-xxx0
3PIx-0xx-xxx1 0 … +8.5 V
3PIx-0xx-xxx2 0 … +5.529 V
3PIx-0xx-xxx3 0 … +8.875 V
Measurement range
3PIx-0xx-xxx0
3PIx-0xx-xxx1 +1.2 … +8.5 V
3PIx-0xx-xxx2 +0.375 … +5.529 V
3PIx-0xx-xxx3 +1.57 … +8.875 V
SP2, ATM
Setting range (N
Hysteresis
Switching characteristics
) 5.0×10-5 … 1000 mbar
2
1)
10% of threshold
1)
Low Trip Point
Contact rating <30 VAC/DC, ≤0.3 A
resistive
closed LED lit solid
open LED off
Switching time <30 ms
Mechanical relays switching functions SP1,
SP2, ATM
Setting range (N
Hysteresis
Switching characteristics
) 5.0×10-5 … 1000 mbar
2
1)
10% of threshold
1)
Low Trip Point
Type 1 floating contact (n.o.) per
switching function
Contact rating <30 VAC/DC, ≤1 A
resistive
closed LED lit solid
open LED off
Switching time <30 ms
Diagnostic port Jack connector 2.5 mm,
3-pin
1
)
The hysteresis and the switching characteristics can be
programmed via the serial interface or the diagnostic port.
tina60e1-a (2011-04)11
Supply
DANGER
The gauge may only be connected to power supplies, instruments, or control devices that conform
to the requirements of a grounded protective extralow voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused.
Supply voltage
at the gauge
Ripple ≤1 V
Power consumption
without fieldbus
DeviceNet
Profibus
Fuse to be connected
Electrical connection
3PIx-0xx-x0xx
2)
Class 2 / LPS
+15 … +30 VDC
pp
≤2.5 W
≤3 W
2)
1 AT
≤3 W
FCC 68
3PIx-0xx-x1xx D-Sub 9-pin, male
3PIx-0xx-x2xx D-Sub 15-pin HD, male
3PIx-0xx-x4xx D-Sub 15-pin HD, RS485
INF, male
Sensor cable shielded
Cable length
RS232C operation
Grounding concept
Vacuum connection to
signal common
2
0.14 mm
/conductor
≤100 m
≤30 m
→ "Power Connection"
connected via 10 kΩ, 10 nF
2)
INFICON controllers fulfill this requirement.
12
tina60e1-a (2011-04)
RS232C / RS485 interface
Transmission rate
Data format
57600 baud (default)
binary
8 data bits
one stop bit
no parity bit
no handshake
→ "Power Connection"
For further information on the RS232C / RS485C interface
→ [3].
DeviceNet interface
Specification, data format,
communication protocol
→ [6]
Interface, physical CAN bus
Data rate (adjustable via <RATE>
switch)
125 kBaud
250 kBaud
500 kBaud (default)
<P> (125 kBaud,
250 kBaud, 500 kBaud
programmable via
DeviceNet, → [4])
Node address (MAC ID)
(Adjustable via <ADDRESS>,
<MSD>, <LSD> switches)
0 … 63dec (63dec default)
<P> (0 … 63 programmable
via DeviceNet, → [4])
DeviceNet connector Micro-Style, 5-pin, male
Cable shielded, special DeviceNet
cable, 5 conductors
→ 32, → [7]
Cable length, system wiring according to DeviceNet
specifications, → [6], [7]
For further information on the DeviceNet interface → [4]
tina60e1-a (2011-04)13
Profibus interface
Specification, data format,
communication protocol
→ [8]
Interface, physical RS485
Data rate
Node address
Local
(Adjustable via hexadecimal
<ADDRESS>, <MSD>,
<LSD> switches)
Default setting
Via Profibus
(hexadecimal <ADDRESS>
switches set to >7D
)
(>125
dec
hex
≤12 Mbaud (→ [5])
00 … 7D
01C
00 … 7D
hex
(0 … 125
hex
(0 … 125
hex
Profibus connection D-Sub, 9-pin, female
Cable shielded, special Profibus
cable, → 33, → [9]
Cable length, system wiring according to Profibus speci-
fications, → [8], [9]
For further information on the Profibus interface → [5]