INFICON PSG55x User Manual

Pirani Standard Gauge
PSG550 PSG552 PSG554
tina60e1-a (2011-04) 1
Operating Manual
Incl. EC Declaration of Conformity
Product Identification
In all communications with INFICON, please specify the infor­mation given on the product nameplate. For convenient refer­ence copy that information into the space provided below.
INFICON AG, LI-9496 Balzers
Model: PN: SN: V W ; LPS
2
tina60e1-a (2011-04)
Validity
This document applies to products of the PSG55x series.
Part numbers of standard products are indicated below. OEM products have other part numbers and different parameter settings (e.g. factory setting of setpoint) as defined in the corresponding ordering information.
3PIx-0xx-xxxx
Flange
1 ⇒ DN 16 ISO-KF
2 ⇒ DN 16 ISO-KF long tube
4 ⇒ DN 16 CF-F
5 ⇒ DN 16 CF-R long tube 6 ⇒ DN 25 ISO-KF D ⇒ 4 VCR female E 8 VCR female
F 1/8" NPT
Flange 29×29 mm
K
M ⇒ 4 VCR 90° female N 7/16-20 UNF male
Unit
0 mbar 1 Torr 2 Pa
3 ⇒ micron
Filament
Tungsten
1 2 ⇒ Nickel
coated
Al
3
2O3
6
7 ⇒ Nickel, galvanically isolated 8 ⇒ Al
galvanically isolated
Tungsten,
coated, galvanically isolated
2O3
tina60e1-a (2011-04) 3
3PIx-0xx-xxxx
Measuring
signal
Interface
Receptacle
Display
Switching function
*) Solid state relays **) Mechanical relays
The part number (PN) can be taken from the product nameplate. If not indicated otherwise in the legends, the illustrations in this
document correspond to gauges with the DN 16 ISO-KF vacuum connection and display. They apply to gauges with other vacuum connections by analogy.
We reserve the right to make technical changes without prior notice.
0.61 ... 10 V
0 ⇒ 1 ⇒ 1.2 ... 8.5 V
2 ⇒ 0.375 ... 5.529 V 3 ⇒ 1.57 ... 8.875 V
0 ⇒ None (analoge) 1 ⇒ DeviceNet 2 Profibus
0 FCC 1 ⇒ D-Sub, 9-pin 2 ⇒ D-Sub, 15-pin HD
D-Sub, 15-pin HD, RS485 INF
4
0 ⇒ None 1 ⇒ Display 2 ⇒ 2 switching functions
4 Display & 2 switching functions *) 6 ⇒ 2 switching functions **)
*)
4
tina60e1-a (2011-04)
Intended Use
The Pirani Standard Gauge PSG55x has been designed for vacuum measurement of gases in the pressure range of
-5
5×10
… 1000 mbar.
They must not be used for measuring flammable or combustible gases in mixtures containing oxidants (e.g. atmospheric oxygen) within the explosion range.
The gauge is intended for operation in connection with an INFICON Vacuum Gauge Controller VCG401, 402, 403 or with another suitable controller.
Trademark
VCR® Swagelok Marketing Co.
Patents
EP 0689669 B1, 0689670 B1 US Patent 5608168
Scope of Delivery
1× gauge 1× pin for adjusting settings via buttons 1× Operating Manual English 1× Operating Manual German
tina60e1-a (2011-04) 5
Contents
Product Identification 2 Validity 3 Intended Use 5 Trademark 5 Patents 5 Scope of Delivery 5
1
Safety 8
1.1 Symbols Used 8
1.2 Personnel Qualifications 8
1.3 General Safety Instructions 9
1.4 Liability and Warranty 9
2 Technical Data 10
2.1 Output Signal vs. Pressure 18
2.2 Gas Type Dependence 23
3 Installation 24
3.1 Vacuum Connection 24
3.2 Power Connection 27
3.2.1 FCC 68, 8-pin Connector 28
3.2.2 D-Sub, 9-pin Connector 29
3.2.3 D-Sub, 15-pin HD Connector 30
3.2.4 D-Sub, 15-pin HD, RS485 INF Connector 31
3.2.5 DeviceNet Connector 32
3.2.6 Profibus Connector 33
4 Operation 34
4.1 Status Indication and Displays 34
4.2 Gas Type Dependence 38
4.3 Switching Functions SP1, SP2 38
4.4 Diagnostic Port (RS232C Interface) 44
4.5 DeviceNet Operation 45
4.6 Profibus Operation 48
5 Deinstallation 49
6
tina60e1-a (2011-04)
6 Maintenance, Repair 51
6.1 Adjusting the Gauge 51
6.2 Replacing the Sensor 53
6.3 Troubleshooting 54
7 Returning the Product 55 8 Disposal 56 9 Accessories 57 10 Spare Parts 58
Further Information 61
ETL Certification 62 EC Declaration of Conformity 63
For cross-references within this document, the symbol ( XY) is used, for cross-references to further documents, listed under "Further Information", the symbol ( [Z]).
tina60e1-a (2011-04) 7
1 Safety
1.1 Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and environ­mental damage.
Caution
Information on correct handling or use. Disregard can lead to malfunctions or minor equipment damage.
Notice
<…> Labeling
1.2 Personnel Qualifications
Skilled personnel
All work described in this document may only be carried out by persons who have suitable technical training and the neces­sary experience or who have been instructed by the end-user of the product.
8
tina60e1-a (2011-04)
1.3 General Safety Instructions
Adhere to the applicable regulations and take the necessary precautions for the process media used.
Consider possible reactions with the product materials. Consider possible reactions (e.g. explosion) of the process
media due to the heat generated by the product.
Adhere to the applicable regulations and take the necessary precautions for all work you are going to do and consider the safety instructions in this document.
Before beginning to work, find out whether any vacuum com­ponents are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contamin­ated parts.
Communicate the safety instructions to all other users.
1.4 Liability and Warranty
INFICON assumes no liability and the warranty becomes null and void if the end-user or third parties
disregard the information in this document
use the product in a non-conforming manner
make any kind of interventions (modifications, alterations etc.)
on the product
use the product with accessories not listed in the product documentation.
The end-user assumes the responsibility in conjunction with the process media used.
Gauge failures due to contamination or wear and tear, as well as expendable parts (e.g. filament) are not covered by the warranty.
tina60e1-a (2011-04) 9
2 Technical Data
For further technical data for gauges with serial inter-
faces [3], [4], [5].
Measurement principle thermal conductance acc. to
Measurement range 5×10 Accuracy (N
5×10
1×10
)
2
-4
… 1×10-3 mbar
-3
… 100 mbar
100 … 1000 mbar
Resolution (at 1000 mbar)
Repeatability (N
1×10
)
2
-3
… 100 mbar
Output signal (measurement signal)
Voltage range 3PIx-0xx-xxx0
3PIx-0xx-xxx1 0 … +8.5 V 3PIx-0xx-xxx2 0 … +5.529 V 3PIx-0xx-xxx3 0 … +8.875 V
Measurement range 3PIx-0xx-xxx0
3PIx-0xx-xxx1 +1.2 … +8.5 V 3PIx-0xx-xxx2 +0.375 … +5.529 V 3PIx-0xx-xxx3 +1.57 … +8.875 V
Error signal 0 … 0.05 V
Voltage vs. pressure 3PIx-0xx-xxx0
3PIx-0xx-xxx1 1 V/decade, logarithmic 3PIx-0xx-xxx3 1 V/decade, logarithmic 3PIx-0xx-xxx2
Pirani
-5
… 1000 mbar
±50% of reading
±15% of reading
±50% of reading
±0.15% of reading
±2% of reading
0 … +10 V
+0.61 … +10 V
1.286 V/decade, logarithmic
20
10
tina60e1-a (2011-04)
Output impedance
Load impedance
Response time
Gauge identification
FCC 68 (+0.61 … +10 V)
HV adjustment at <10-5 mbar
2 × 4.7 Ω, short circuit-proof
10 kΩ <10 ms
27 kΩ
Solid state relays switching functions SP1,
SP2, ATM Setting range (N Hysteresis Switching characteristics
) 5.0×10-5 … 1000 mbar
2
1)
10% of threshold
1)
Low Trip Point
Contact rating <30 VAC/DC, 0.3 A
resistive
closed LED lit solid open LED off
Switching time <30 ms
Mechanical relays switching functions SP1,
SP2, ATM Setting range (N Hysteresis Switching characteristics
) 5.0×10-5 … 1000 mbar
2
1)
10% of threshold
1)
Low Trip Point
Type 1 floating contact (n.o.) per
switching function Contact rating <30 VAC/DC, 1 A
resistive
closed LED lit solid open LED off
Switching time <30 ms
Diagnostic port Jack connector 2.5 mm,
3-pin
1
)
The hysteresis and the switching characteristics can be
programmed via the serial interface or the diagnostic port.
tina60e1-a (2011-04) 11
Supply
DANGER
The gauge may only be connected to power sup­plies, instruments, or control devices that conform to the requirements of a grounded protective extra­low voltage (SELV) and limited power source (LPS), Class 2. The connection to the gauge has to be fused.
Supply voltage
at the gauge Ripple 1 V
Power consumption
without fieldbus DeviceNet Profibus
Fuse to be connected
Electrical connection
3PIx-0xx-x0xx
2)
Class 2 / LPS
+15 … +30 VDC
pp
2.5 W
3 W
2)
1 AT
3 W
FCC 68 3PIx-0xx-x1xx D-Sub 9-pin, male 3PIx-0xx-x2xx D-Sub 15-pin HD, male 3PIx-0xx-x4xx D-Sub 15-pin HD, RS485
INF, male
Sensor cable shielded
Cable length
RS232C operation
Grounding concept
Vacuum connection to signal common
2
0.14 mm
/conductor
100 m
30 m
"Power Connection"
connected via 10 kΩ, 10 nF
2)
INFICON controllers fulfill this requirement.
12
tina60e1-a (2011-04)
RS232C / RS485 interface
Transmission rate Data format
57600 baud (default)
binary
8 data bits
one stop bit
no parity bit
no handshake
"Power Connection"
For further information on the RS232C / RS485C interface [3].
DeviceNet interface
Specification, data format, communication protocol
[6]
Interface, physical CAN bus
Data rate (adjustable via <RATE> switch)
125 kBaud
250 kBaud
500 kBaud (default)
<P> (125 kBaud,
250 kBaud, 500 kBaud
programmable via
DeviceNet, [4])
Node address (MAC ID) (Adjustable via <ADDRESS>, <MSD>, <LSD> switches)
0 … 63dec (63dec default)
<P> (0 … 63 programmable
via DeviceNet, [4])
DeviceNet connector Micro-Style, 5-pin, male
Cable shielded, special DeviceNet
cable, 5 conductors
32, [7] Cable length, system wiring according to DeviceNet
specifications, [6], [7]
For further information on the DeviceNet interface [4]
tina60e1-a (2011-04) 13
Profibus interface
Specification, data format, communication protocol
[8] Interface, physical RS485
Data rate
Node address
Local (Adjustable via hexadecimal <ADDRESS>, <MSD>, <LSD> switches)
Default setting Via Profibus
(hexadecimal <ADDRESS> switches set to >7D
)
(>125
dec
hex
12 Mbaud ( [5])
00 … 7D
01C
00 … 7D
hex
(0 … 125
hex
(0 … 125
hex
Profibus connection D-Sub, 9-pin, female
Cable shielded, special Profibus
cable, 33, → [9]
Cable length, system wiring according to Profibus speci-
fications, [8], [9]
For further information on the Profibus interface [5]
Materials exposed to vacuum
Vacuum connection Filament 3PI1 / 6-0xx-xxxx 3PI2 / 7-0xx-xxxx 3PI3 / 8-0xx-xxxx Feedthrough
3)
Orifice Further materials
stainless steel 1.4435
W
Ni
ceramic coated
glass
stainless steel
Ni, NiFe, stainless steel
1.4301
dec)
dec
)
3)
Only versions DN 16 ISO-KF and DN 16 CF-F.
14
tina60e1-a (2011-04)
Internal volume
DN 16 ISO-KF DN 16 ISO-KF, long tube DN 16 CF-F DN 16 CF-R, long tube DN 25 ISO-KF
®
female
4 VCR 8 VCR® female 1/8" NPT Flange 29×29 mm
®
4 VCR
90°, female
7/16-20 UNF
4.7 cm
14.5 cm
3
8 cm
14 cm
5.5 cm
5.5 cm
3
7 cm
5.2 cm
4.9 cm
7.9 cm
5.2 cm3
3
3
3
3
3
3
3
3
Permissible pressure (absolute) 5 bar Bursting pressure (absolute) 10 bar
Permissible temperatures
Operation Vacuum connection long tube 4) Filament Storage
Relative humidity
Year's mean During 60 days
4)
+10 °C … +50 °C
80 °C
250 °C
<160 °C
–20 °C … +65 °C
65% (no condensation)
85% (no condensation)
Mounting orientation any Use indoors only, altitude up to
2000 m NN
Degree of protection IP 40
Weight
without fieldbus interface with fieldbus interface
115 g …130 g
230 g … 250 g
4)
For horizontal mounting orientation only. During bakeout,
measurement range, accuracy, and repeatability may deviate from specifications.
tina60e1-a (2011-04) 15
Dimensions [mm]
45
16
21.531.5 DN 25 ISO-KF
29.2×29.2 mm
36.5 34
7/16-20 UNF
4-VCR 90°
24.4
36
14.5
28
1/8" NPT
DN 16 ISO-KF
4 VCR
30.9 42.7
130.8
DN 16 ISO-KF long
tina60e1-a (2011-04)
79.5 65
DN 16 CF-F
46.3
8 VCR
130.2
DN 16 CF-R long
DeviceNet
67
BUS RATE STATUS
ADDRESS
LSDMSD
I/O
DIA
RS485
67
ADDRESS
LSDMSD
SP1 SP2 ST
62
AE
DIA
BUS
Profibus
10.8
NET MOD
56
ST
SP1SP2
ADJ
RATE
62
AE62AE
56
ADJ
321ON
ADDRESS
26
08
EA
67
BUS
4
4
26 08 EA
C
C
LSDMSD
I/O
DIA
SP1SP2
56
ST
ADJ
tina60e1-a (2011-04) 17
2.1 Output Signal vs. Pressure
Measurement range 0.61 … 10 V
Pressure p
1E+06
1E+05
1E+04
1E+03
1E+02
1E+01
Error
1E+00
1E–01
1E–02
1E–03
1E–04
1E–05
0.0
0.61
p = 10
3.02.52.01.51.00.5 4.03.5 5.04.5 6.05.5 7.06.5 8.07.5 9.08.5 10.09.5 10.5
Output signal U [V]
0.778(U-c)
U = c +1.286log10 p
valid in the range 5×10-5 mbar <p< 1000 mbar
U p c U p c
[V] [mbar] 6.143 [V] [micron] 2.448
[V] [µbar] 2.287 [V] [Pa] 3.572
[V] [Torr] 6.304 [V] [kPa] 7.429
[V] [mTorr] 2.448
where p pressure
U output signal c constant (pressure unit dependent)
Pa
mbar
Torr
O
18
tina60e1-a (2011-04)
Measurement range 1.2 … 8.5 V
Pressure p
1E+05
1E+04
1E+03
1E+02
1E+01
1E+00
Error
1E–01
1E–02
1E–03
1E–04
1E–05
0.0 0.5 1.0 1.5 2.5 3.5 4.5 5.5 6.5 7.5 8.52.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0
1.2
p = 10
Output signal U [V]
(U-c)
U = c + log10 p
Pa
mbar
Torr
valid in the range 5×10-5 mbar <p< 1000 mbar
U p c U p c
[V] [mbar] 5.5 [V] [micron] 2.625
[V] [µbar] 2.5 [V] [Pa] 3.5
[V] [Torr] 5.625 [V] [kPa] 6.5
[V] [mTorr] 2.625
where p pressure
U output signal c constant (pressure unit dependent)
tina60e1-a (2011-04) 19
Measurement range 0.375 … 5.529 V
Signal U Pressure p
[V] [mbar] [Pa] [Torr]
0.375 <5×10-5 <6.65×10-3 <5×10-5
0.376 0.000133322 0.013332237 0.0001
0.377 0.000266645 0.026664474 0.0002
0.379 0.000666612 0.066661184 0.0005
0.384 0.001333224 0.133322368 0.0010
0.392 0.002666447 0.266644736 0.0020
0.417 0.006666118 0.66661184 0.0050
0.455 0.013332237 1.33322368 0.0100
0.523 0.026664474 2.66644736 0.0200
0.682 0.066661184 6.6661184 0.0500
0.876 0.133322368 13.3322368 0.1000
1.155 0.266644736 26.6644736 0.2000
1.683 0.66661184 66.661184 0.5000
2.217 1.33322368 133.322368 1.0000
2.842 2.66644736 266.644736 2.0000
3.675 6.6661184 666.61184 5.0000
4.206 13.3322368 1333.22368 10.0000
4.577 26.6644736 2666.44736 20.0000
4.846 66.661184 6666.1184 50.0000
4.945 133.322368 13332.2368 100.0000
5.019 266.644736 26664.4736 200.0000
5.111 399.967104 39996.7104 300.0000
5.224 533.289472 53328.9472 400.0000
5.329 666.61184 66661.184 500.0000
5.419 799.934208 79993.4208 600.0000
5.495 933.256576 93325.6576 700.0000
5.529 1000 100000 750.0637
20
tina60e1-a (2011-04)