SERVO-U/n Ventilator Systems
Cleaning and Maintenance

TABLE OF CONTENTS

| TABLE OF CONTENTS |
1 2 3 4 5 6 7 8 9 10 11
4|Introduction
7|Patient unit and user interface 10|Expiratory cassette 14|O2 cell (maintenance) 17|Inspiratory channel (maintenance) 22|CO2 analyzer 24|Aerogen Pro nebulizer 27|Y sensor 29|NAVA system 30|Hardware accessories 31|Recommended intervals and disinfectants
SERVO-U/n Ventilator Systems, User's Manual Infologic 1.27
3
| Introduction |
| 1

1 Introduction Table of contents

5|Introduction1.1
5|Maintenance1.2
5|Safety guidelines1.3
6|Symbols1.4
4
SERVO-U/n Ventilator Systems, User's Manual
| Introduction |
1 |

1.1 Introduction

As cleaning practices vary widely among health care institutions, it is not possible for Maquet to specify particular practices that will meet all needs, or to be responsible for the effectiveness of cleaning procedures carried out in the patient care setting.
Maquet recommends methods that have been validated using the specified equipment and procedures outlined in this manual. Other methods may work but are not covered by the warranty unless Maquet has given written permission.
Unless otherwise stated, the information in this User´s Manual is valid for all SERVO-U and SERVO-n ventilator systems.

1.2 Maintenance

A preventive maintenance must be performed by authorized personnel at least once every year as long as the unit is not used more than normal. Normal operation during one year is estimated to correspond to approximately 5000 hours of operation.

1.3 Safety guidelines

WARNING! Indicates critical information about a potential serious outcome to the patient or the user.
CAUTION: Indicates instructions that must be followed in order to ensure the proper operation of the equipment.
Important: Indicates information intended to help you operate the equipment or its connected devices easily and conveniently.
Note: Indicates information requiring special attention.
CAUTIONS:
All personnel should be aware of the risk
of parts being infected when disassembling and cleaning the ventilator system.
The expiratory channel and expired gas
from the exhaust port may be contaminated.
All disposable parts must be discarded
according to hospital routines and in an environmentally safe way.
SERVO-U/n Ventilator Systems, User's Manual
5
| Introduction |
| 1
Important:
Follow your hospital‘s routines for
handling infectious material when handling any part of the ventilator system.
If possible, cleaning should be performed
immediately after use and always before steam autoclaving/disinfection. Blood or other residues should not be allowed to dry onto the devices.
Water quality affects
cleaning/disinfection. Maquet recommends drinking water quality as the minimum quality level.
Maquet recommends the use of Servo
Duo Guard bacterial filter or equivalent to reduce the transmission of bacteria from the patient via the expiratory cassette to the ambient air. This reduces the risk of infection spread and prolongs the life of the expiratory cassette due to reduced cleaning requirements.
If a Servo Duo Guard bacterial filter (or equivalent) has been used only wiping off the ventilator and discarding filter is necessary.
The use of a washer-disinfector for
cleaning/disinfection is recommended if bacterial filters are not used.
Steam autoclaving is normally not
necessary for the expiratory cassette, as it is not an invasive instrument, but when applied, use validated processes only.
Note: Single use (i.e. disposable) items shall not be reused. Cleaning of single use items can damage the item and compromise its intended use. Reuse of single use items increases the risk of spreading infections and compromises the durability of the item.

1.4 Symbols

Hazardous wasteThe device
contains parts which must not be disposed of with ordinary waste.
Special waste
This product contains electronic and electrical components. Discard disposable, replaced and left-over parts in accordance with appropriate industrial and environmental standards.
6
SERVO-U/n Ventilator Systems, User's Manual

2 Patient unit and user interface

2.1 Preparations

Note: Cleaning should be done after each patient or according to hospital routines.
Switch off the ventilator system using the
on/off switch on the patient unit.
Disconnect the ventilator system from the
mains power and gas supply.
Disconnect any optional equipment from
the mains power and from the ventilator system.
Discard disposable items according to
hospital routines and environmental regulations.
- Servo Duo Guard viral/bacterial filter
- SERVO humidifier/HME
- disposable Y sensor
- disposable patient tubing
- disposable airway adapter
- Edi catheters
- Aerogen Solo nebulizer
Remove the expiratory cassette.

2.1.1 Remove the expiratory cassette

Important: The expiratory cassette is a
precision instrument and must be handled carefully.
| Patient unit and user interface |
1
2
1. Press the button on the expiratory cassette.
2. Tilt the cassette upwards and remove.
Important: After removing the expiratory cassette, do not pour any fluid into the expiratory cassette compartment. Avoid contact with electrical connectors.

2.2 Wiping and discarding

Important:
Do not use abrasive or sharp tools to
clean the user interface.
If a Servo Duo Guard bacterial filter (or
equivalent) has been used only wiping off the ventilator and discarding filter is necessary.
2 |
SERVO-U/n Ventilator Systems, User's Manual
7
| Patient unit and user interface |
| 2
Cleaning should be done after each patient or according to hospital routines.
Wipe all parts with cleaning wipes or a soft lint-free cloth moistened in soap & water or detergent-based disinfectant.
Note: In case of more contaminated surfaces, use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical connectors.
Refer to section Recommended intervals and disinfectants on page 31.

2.3 Assembling

2.3.1 Insert the expiratory cassette

1
2
Hinge the expiratory cassette and press it
firmly down into lock position.
Important: Make sure that the cassette clicks into position. Check that it cannot be moved upwards and that the button on top of the cassette is completely ejected.
Note on a log sheet that a routine cleaning
has been performed. Refer to hospital routines.
8
SERVO-U/n Ventilator Systems, User's Manual
| Patient unit and user interface |
2 |

2.3.2 Pre-use check

After cleaning, always perform a pre-use check.
Refer to the ventilator system User's Manual.
If the pre-use check fails:
Carefully shake/tilt the cassette (5-7 times).
Run the cassette in a ventilator system with
a test lung for 10 minutes.
Re-do the pre-use check.

2.4 Preventive maintenance

2.4.1 Check fan filters

Regularly check that the fan filter looks clean (i.e. black). If dusty, remove (snap off/snap on) and rinse in water. Shake out, ensuring that the filter is free from excess water.
Regularly check that the emergency air intake is not occluded.
SERVO-U/n Ventilator Systems, User's Manual
9
| Expiratory cassette |
| 3

3 Expiratory cassette

3.1 Preparations

Remove the expiratory cassette
Refer to section Remove the expiratory cassette on page 7.

3.2 Disinfection

Pre-disinfect according to hospital routines.
Refer to section Recommended intervals and disinfectants on page 31.
There are two disinfection procedures for the expiratory cassette and associated parts when no bacterial filter is used:
Washer-disinfector (ISO 15883-1)
(recommended)
Disinfectant

3.2.1 Washer-disinfector

If rinsing before disinfection is not included in the washer disinfector program:
Rinse the parts thoroughly in water
(<35°C/95°F).
Let the water flow through the parts.
Important: Immediate rinsing in water (<35°C/95°F) can remove particles and reduces the risk of cross-contamination between patients.
Wash the parts with water only in a
washer-disinfector at a temperature of 85-95°C (185-203°F).
Place the expiratory cassette on its side
with the electrical connector uppermost.
Notes:
The water pressure in the
washer-disinfector should not exceed
1.5 bar.
The maximum water flow should not
exceed 10 l/min.
The water must have free passage
through the cassette.
10
SERVO-U/n Ventilator Systems, User's Manual
| Expiratory cassette |
3 |

3.2.2 Disinfectant

Let the parts soak in disinfectant.
Refer to section Recommended intervals and disinfectants on page 31.
Important: Follow the disinfectant manufacturer’s recommendations and instructions, otherwise the cassette may be damaged or not disinfected properly.
Rinse after disinfection
Important:
Follow the disinfectant manufacturer's
recommendation and instructions regarding rinsing.
Mineral deposits from disinfectant on the
expiratory cassette affect the function. It is important to rinse the expiratory cassette thoroughly. Residues from chemicals can affect the patient, cause leakage and extra stress on the material.
SERVO-U/n Ventilator Systems, User's Manual
Rinse the parts thoroughly in water to
remove all traces of disinfectant. Let the water flow through the parts.
Carefully shake and tilt the cassette, turn
upside down and repeat at least 5-7 times.
11
| Expiratory cassette |
| 3

3.3 Drying alternatives

Important: The expiratory cassette must be dried before use (if not dry, the expiratory cassette may not pass the pre-use check).
Always shake/tilt the cassette carefully (5-7 times).
There are several drying alternatives for the expiratory cassette:
Drying cabinet 1 hour in maximum 70°C
(158°F); if available, connect a 22 mm air hose to the expiratory cassette for extra drying effect, or
Drying the cassette in room air 12-24 hours,
depending on surrounding conditions.

3.3.1 Recommended position in a drying cabinet

3.3.2 Recommended position in room air

If the pre-use check is not passed after drying then:
Run the cassette in a ventilator system with
a test lung for 10 minutes, or
Dry the cassette in room air (if no drying
cabinet is available) 12-24 hours, depending on surrounding conditions.
Notes:
Never dry the cassette by applying
high-pressure air as the internal tubing may be damaged.
Drying may not be necessary if the
washer-disinfector has a drying phase.
12
SERVO-U/n Ventilator Systems, User's Manual
| Expiratory cassette |
3 |

3.4 Steam autoclaving

Note: Steam autoclaving procedures are not recommended.
Steam autoclaving will reduce the lifetime of the expiratory cassette and is not recommend as a cleaning method.
Steam autoclaving is normally not necessary for the expiratory cassette, as it is not an invasive instrument, but when applied, use validated processes only.
Refer to section Recommended intervals and disinfectants on page 31.

3.4.1 Cleaning

Clean before steam autoclaving.
Refer to section Washer-disinfector on page 10.

3.4.4 Drying after steam autoclaving

Refer to section Drying alternatives on page 12.

3.5 Assembling

Refer to section Assembling on page 8.

3.4.2 Rinse

Rinse the parts thoroughly in water
(<35°C/95°F).
Let the water flow through the parts.

3.4.3 Drying before steam autoclaving

Before placing the expiratory cassette in a steam autoclave, make sure that no water remains inside the cassette. Refer to section Drying alternatives on page 12.
SERVO-U/n Ventilator Systems, User's Manual
13
| O2 cell (maintenance) |
| 4

4O2 cell (maintenance)

4.1 Preventive maintenance

Regularly check the cell status in quick menu SYSTEM STATUS/O2 cell. The O2 cell should be exchanged if <10% estimated remaining capacity is indicated.
Refer to the ventilator system User's manuals for O2 cell adjustment.

4.2 Preparations

WARNING! The sealed unit of the O2 cell contains a caustic liquid which may cause severe burns to the skin and eyes. In case of contact, immediately wash the affected area continuously with water for at least 15 minutes and seek medical attention, especially if the eyes are affected.
Important:
Make sure the O2 cell is for the intended
ventilator model.
Replacement of the O2 cell and filter and
extended cleaning must only be performed by trained personnel.
Unpack the O2 cell at least 15 minutes
before replacement. Turn the ventilator system off using the on/off switch on the patient unit.
Disconnect the ventilator system from the
power and gas supply.
Turn the inspiratory channel lock one
quarter of a turn to release.
Lift off the inspiratory channel cover.
WARNING! The inspiratory channel cover shall be kept in place once maintenance has been performed. Otherwise the gas modules can be affected and deliver the wrong flow.
14
SERVO-U/n Ventilator Systems, User's Manual

4.2.1 Remove O2 cell and/or bacterial filter

Lower the locking catch.
cell (maintenance) |
| O
2
Pb
Disconnect the O2 cell connector.
Lift and discard the O2 cell with the rubber
seal to special waste and/or remove and discard the bacterial filter to hazardous waste.
4 |
SERVO-U/n Ventilator Systems, User's Manual
15
| O2 cell (maintenance) |
| 4

4.3 Assembling

4.3.1 Insert the O2 cell

Ensure that the new rubber seal is intact,
then firmly put in a new bacterial filter.
Connect the O2 cell connector.
Put the O2 cell in position.
Close the locking catch.
Put the inspiratory channel cover in
position.
Tighten the inspiratory channel lock.
Perform a pre-use check.
Refer to the ventilator system User's Manual.
16
SERVO-U/n Ventilator Systems, User's Manual

5 Inspiratory channel (maintenance)

If an O2 sensor is installed:

5.1 Preventive maintenance

Important: It is recommended that a regular cleaning and an extended cleaning of the inspiratory channel should be performed before carrying out preventive maintenance.

5.2 Preparations

| Inspiratory channel (maintenance) |
5 |
Remove the inspiratory channel cover.
Refer to chapter O2 cell (maintenance), section Preparations on page 14.
Remove O2 cell/sensor.
If an O2 cell is installed:
Lower the locking catch.
Disconnect the O2 sensor and carefully
unlock the latches.
Lift the O2 sensor out of position.
Disconnect the O2 cell connector and lift
out the O2 cell.
SERVO-U/n Ventilator Systems, User's Manual
17
| Inspiratory channel (maintenance) |
| 5

5.2.1 Remove inspiratory channel and tube

Remove the inspiratory channel.
Press the latches and lift the inspiratory
channel upwards. Disconnect the connector muffs.
Disconnect the tube and remove the
bacterial filter. Discard the filter to hazardous waste.

5.3 Disinfection

Important: It is recommended that cleaning of the inspiratory channel should be performed during preventive maintenance. The cleaning must be done by trained personnel only.
Pre-disinfect according to hospital routines.
Refer to section Recommended intervals and disinfectants on page 31.

5.3.1 Rinse before disinfection

If rinsing before disinfection is not included in the washer disinfector program:
Rinse the parts thoroughly in water
(<35°C/95°F).
Let the water flow through the parts.
Important: Immediate rinsing in water (<35°C/95°F) can remove particles and reduces the risk of cross-contamination between patients.
18
SERVO-U/n Ventilator Systems, User's Manual
| Inspiratory channel (maintenance) |
5 |

5.3.2 Washer-disinfector

Place the inspiratory channel so that the
water flows through it.
Clean the inspiratory channel and tube the
same way as the expiratory cassette. Refer to section Disinfection on page 10.

5.4 Assembling

5.4.1 Insert inspiratory channel and tube

Put a new bacterial filter in position and
connect the filter to the tube.
SERVO-U/n Ventilator Systems, User's Manual
Put the connector muffs in position.
Press the latches and insert the inspiratory
channel.
19
| Inspiratory channel (maintenance) |
| 5
Important:
There should always be clearance
between the connector muffs and gas modules.
Make sure the latches are locked in
position.

5.4.2 Insert O2 cell/O2 sensor

If an O2 cell is installed:
If an O
sensor is installed:
2
Put the O2 sensor in position.
2x
Connect the O2 cell connector and put the
O2 cell in position.
Close the locking catch.
Connect the O2 sensor (two "clicks" are
heard).
Connect the O2 sensor connector.
20
SERVO-U/n Ventilator Systems, User's Manual
Re-assemble the inspiratory channel cover
according to instructions given in section Assembling on page 16.
Note on a log sheet that an extended
cleaning of the inspiratory channel has been performed.
Perform a pre-use check.
Refer to the ventilator system User's Manual.
| Inspiratory channel (maintenance) |
5 |
SERVO-U/n Ventilator Systems, User's Manual
21
| CO2 analyzer |
| 6

6CO2 analyzer

6.1 Preparation

Cleaning should be done after each patient or according to hospital routines.
Important: Do not immerse the CO2 module or the CO2 sensor in fluid.
Disconnect the CO2 sensor and airway adapter from the patient circuit and the ventilator system.

6.2 Wipe and discard

6.2.1 CO2 sensor

Wipe the CO2 sensor with a soft cloth
moistened in disinfectant (Cidex OPA or isopropyl alcohol 70%).

6.2.2 Airway adapter

Note: Discard the disposable airway
adapter. The disposable airway adapter is for single patient use only.
Pre-disinfect according to hospital routines.
Refer to section Recommended intervals and disinfectants on page 31.
The reusable airway adapter can be disinfected or steam autoclaved.
Rinse the adapter thoroughly in water
(<35°C/95°F).
Let the adapter soak in Cidex OPA solution.
Rinse the adapter thoroughly in distilled
water.
Before reusing the adapter, the windows
must be dry and wiped off.
After cleaning, wipe the CO2 sensor with a
water-dampened clean cloth.
The CO2 sensor windows must be dried
after cleaning.
22
Important: Follow the disinfectant manufacturer's recommendations and instructions.
SERVO-U/n Ventilator Systems, User's Manual

6.3 Steam autoclaving

6.3.1 Airway adapter

The adult adapter can be autoclaved in a
validated process.
The adapters can be sterilized using EtO
(ethylene oxide) gas method.
After sterilization procedure the adapters
must be dried before use.
Refer to section Recommended intervals and disinfectants on page 31.
| CO
analyzer |
2
6 |
SERVO-U/n Ventilator Systems, User's Manual
23
| Aerogen Pro nebulizer |
| 7

7 Aerogen Pro nebulizer

7.1 Preparation

Important: Do not use abrasive or sharp tools to clean the nebulizer unit.
Note: The Aerogen Solo nebulizer unit is intended for single patient use only, do not re-use, clean or sterilize.
Cleaning should be done after each patient or according to hospital routines.
Disconnect the nebulizer unit from the control cable, and remove the nebulizer unit, T piece and adapters from the patient circuit or mask.

7.2 Wiping and discarding

7.3 Disinfection

7.3.1 Cleaning

Ensure there is no medication remaining in
the nebulizer.
Disassemble the nebulizer unit and adapters
into individual components.
Remove the filler cap from nebulizer unit.
Clean all parts with warm water and mild
liquid detergent.
Rinse all parts in sterile water.
Shake excess water from parts and allow
parts to air dry.
Assemble the nebulizer.
Refer to the ventilator system User's Manual.
Wipe the control cable with a damp cloth.
Do not autoclave.
Check for exposed wiring, damaged
connectors, or other defects and replace if any are visible.
Note: In case of more contaminated surfaces, use ethyl alcohol (70%).
24

7.3.2 Washer-disinfector

Two different washing cycles are validated for the Aerogen Pro nebulizer:
Liquid alkaline cleaner (diluted as per
manufacturer instruction) and mains water.
Load the components in the
-
washer-disinfector.
- Pre-rinse the components for 3 minutes.
- Clean the components with liquid alkaline
cleaner at 55 °C (131 °F) for 10 minutes.
- Rinse for 1 minute.
- Rinse using thermal disinfection cycle at
93 °C (199.4 °F) for 10 minutes.
Mains water and without the use of a
detergent.
Load the components in the
-
washer-disinfector.
SERVO-U/n Ventilator Systems, User's Manual
| Aerogen Pro nebulizer |
7 |
- Wash components for 10 minutes at 91 °C (195.8 °F).
- Drain the machine for 40 seconds.
- Rinse at 90 °C (194 °F) for 1 minute.
- Drain the machine for 40 seconds.
- Rinse at 90 °C (194 °F) for 1 minute.
- Drain the machine for 40 seconds.
- Dry at 90 °C (194 °F) for 15 minutes.

7.3.3 Disinfectant

Immerse the parts in appropriate
disinfecting agent.
Refer to section Recommended intervals and disinfectants on page 31.
Important: Follow the disinfectant manufacturer’s recommendations and instructions.

7.4 Steam autoclaving

Disassemble the nebulizer unit and adapters
into individual components.
Remove the filler cap from the nebulizer
unit.
Clean all parts with warm water and mild
liquid detergent. Rinse thoroughly and air dry.
Check for cracks or damage and replace if
any defects are visible.
Place the disassembled components into
appropriate sterilization wrapping.
Note: Do not reassemble parts prior to steam autoclaving
Only the Aerogen Pro nebulizer unit, T piece and neonate adapters can be autoclaved.
Refer to section Recommended intervals and disinfectants on page 31.
SERVO-U/n Ventilator Systems, User's Manual
25
| Aerogen Pro nebulizer |
| 7
Note: Sterilization using the long autoclave cycle may cause some areas of the nebulizer to become discolored. This does not affect the performance of the nebulizer unit.
To sterilize with hydrogen peroxide gas plasma, place wrapped parts in a STERRAD® System, and use the long cycle.
Note: Users should refer to the product labelling for the STERRAD® 100S Sterilization System for specific instructions regarding its proper operation.

7.5 Assembling

Assemble the nebulizer. Perform a function test of the Aerogen Pro nebulizer after the cleaning.
Refer to the ventilator system User's Manual.
26
SERVO-U/n Ventilator Systems, User's Manual

8 Y sensor

| Y sensor |
8 |

8.1 Preparation

Cleaning should be done after each patient or according to hospital routines.
Important: Do not immerse the Y sensor module in fluid.
Disconnect the Y sensor from the patient
circuit.

8.2 Wipe and discard

CAUTION: The disposable Y sensor is intended for single patient use only, do not re-use, clean or sterilize.
Discard the pressure line. The pressure line
is for single patient use only.
Wipe all parts with a soft lint-free cloth
moistened in soap and water or detergent-based disinfectant.

8.3 Disinfection

Pre-disinfect according to hospital routines.
Refer to section Recommended intervals and disinfectants on page 31.

8.3.1 Washer-Disinfector

Place the Y sensor in a washing basket.
Wash the parts with water only in a
washer-disinfector at a disinfection temperature of maximum 90°C (194°F).
Important: Do not clean the flow sensor with compressed air or a hard water jet, since the sensor wires can be destroyed.
SERVO-U/n Ventilator Systems, User's Manual
Refer to section Recommended intervals and disinfectants on page 31.
27
| Y sensor |
| 8

8.3.2 Cleaning

Clean the Y sensor with warm water and
mild liquid detergent.
Rinse the Y sensor in water.
Shake excess water from Y sensor and
allow to air dry.

8.3.3 Disinfectant

Immerse the parts in appropriate
disinfecting agent.
Refer to section Recommended intervals and disinfectants on page 31.
Important: Follow the disinfectant manufacturer’s recommendations and instructions.

8.4 Steam autoclaving

Clean the Y sensor with warm water and
mild liquid detergent. Rinse thoroughly and air dry.
Check for cracks or damage and replace if
any defects are visible.
Refer to section Recommended intervals and disinfectants on page 31.
28
SERVO-U/n Ventilator Systems, User's Manual

9 NAVA system

9.1 Preparation

Cleaning should be done after each patient or according to hospital routines.
Disconnect the Edi module from the ventilator system and remove Edi cable and Edi catheter.

9.2 Wiping and discarding

Discard the Edi catheter. The Edi catheter is for single patient use only.
| NAVA system |
Wipe all parts with cleaning wipes or a soft lint-free cloth moistened in soap & water or detergent-based disinfectant.
Note: In case of more contaminated surfaces, use cleaning wipes, ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical connectors.
9 |
Note: The Edi catheter must not be
disposed of with ordinary waste.
Important: Do not immerse the Edi module or the cable in fluid.
SERVO-U/n Ventilator Systems, User's Manual
29
| Hardware accessories |
| 10

10 Hardware accessories

Hardware accessories are:
Mobile cart
Handle
Humidifier holder
Support arm 178
Water bag/IV pole
Drawer
Gas cylinder restrainer kit
Pendant/bed holder
Shelf base
Y piece holder
User interface holder
Cable holder for handle
Refer to section Recommended intervals and disinfectants on page 31.

10.1 Patient circuits

Note: Discard the disposable patient circuits. The disposable patient circuits are for single patient use only.
Refer to the cleaning routines in the operating manual for the patient circuit.

10.2 Humidifier

Fisher & Paykel Humidifier MR850/MR810
Refer to the cleaning routines in the operating manual for the Fisher & Paykel Humidifier.
30
SERVO-U/n Ventilator Systems, User's Manual
| Recommended intervals and disinfectants |

11 Recommended intervals and disinfectants

Patient unit and user interface
Surface cleaning
Expiratory cassette
Pre-disinfectant
Disinfectant
cycles Cleaning agent
washer-disinfector Steam autoclave
Number of cycles in steam autoclave
3M Respirator Cleaning Wipes
Clorox Bleach Germicidal Wipes (may cause discoloration)
Super Sani-Cloth Germicidal Wipes (may cause discoloration)
Virkon S
Accel TB (Virox 5)
In case of more contaminated surfaces, use ethyl alcohol (70%) or isopropyl
alcohol (70%) together with lint free cleaning wipes.
Aniosyme DD1
Gigazyme Plus
Salvanios pH7
Follow the disinfectant manufacturer's recommendation and instructions.
Alcohol (ethyl or isopropyl alcohol 70%)
Anioxyde 1000
Cidex OPA
Actanios HLD (prion decontamination agent), may cause discoloration
Follow the disinfectant manufacturer's recommendation and instructions. at least 100 cyclesNumber of immersion
neodisher® MediClean forte
Getinge MIS Detergent
at least 200 cycles.Number of cycles in
Instrument parts should be steam autoclaved in a validated process, at a
temperature of 134°C (273°F) for 3 minutes.
Rubber parts should be steam autoclaved in a validated process, at a
temperature of 121°C (250°F) for 15-20 minutes.
The expiratory cassette will last at least:
100 autoclaving cycles with 3 minutes sterilization time at 134ºC (273ºF).
30 autoclaving cycles with 18 minutes sterilization time at 134ºC (273ºF).
11 |
O2 cell
Replace the O2 cell when it is requested by the ventilator system.O2 cell
SERVO-U/n Ventilator Systems, User's Manual
31
| Recommended intervals and disinfectants |
| 11
Inspiratory channel
Pre-disinfectant
Aniosyme DD1
Gigazyme Plus
Salvanios pH7
Follow the disinfectant manufacturer's recommendation and instructions.
Disinfectant
Alcohol (ethyl or isopropyl alcohol 70%)
Anioxyde 1000
Cidex OPA
Actanios HLD (prion decontamination agent), may cause discoloration
Follow the disinfectant manufacturer's recommendation and instructions. at least 10 cyclesNumber of immersion
cycles Cleaning agent
neodisher® MediClean forte
Getinge MIS Detergent
at least 20 cycles.Number of cycles in
washer-disinfector Number of cycles in
steam autoclave
CO2 analyzer
Disinfectant
The inspiratory channel will last at least 10 autoclaving cycles with 3 minutes sterilization time at 134ºC (273ºF).
Cidex OPA
Isopropyl (70 %)
Follow the disinfectant manufacturer's recommendation and instructions. neodisher® MediClean forteCleaning agent
Steam autoclave (adapter)
Instrument parts should be steam autoclaved in a validated process, at a temperature of 134°C (273°F) for 3 minutes.
32
SERVO-U/n Ventilator Systems, User's Manual
Aerogen Pro nebulizer
Pre-disinfectant
Disinfectant
steam autoclave Steam autoclave
| Recommended intervals and disinfectants |
Hexanios G+R
Follow the disinfectant manufacturer's recommendation and instructions.
Isopropyl alcohol (70 %)
CIDEX
NU-CIDEX
Cidex OPA
at least 26 cyclesNumber of cycles in
Steam autoclaving can be performed using the following three methods:
Autoclave wrapped parts using steam autoclaving pre-vacuum cycle, a
minimum of 134°C (270°F - 275°F) for 3.5 minutes with drying cycle (134°C wrapped cycle).
Autoclave wrapped parts using steam autoclaving pre-vacuum cycle, a
minimum of 121°C (250°F) for 20 minutes with drying cycle (121°C wrapped cycle).
Autoclave wrapped parts using steam autoclaving pre-vacuum cycle, a
minimum of 134°C (270°F - 275°F) for 20 minutes with drying cycle (sometimes referred to as a “Prion cycle”).
Note: This cycle may cause some areas of the nebulizer to become discolored. This does not affect the performance of the nebulizer.
11 |
Y sensor
Surface cleaning
Pre-disinfectant
Disinfectant
washer-disinfector
Steam autoclave
Number of cycles in steam autoclave
3M Respirator Cleaning Wipes
Super Sani-Cloth Germicidal Wipes
Virkon S
Accel TB (Virox 5)
In case of more contaminated surfaces, use ethyl alcohol (70%) or isopropyl
alcohol (70%) together with lint free cleaning wipes.
Aniosyme DD1
Gigazyme Plus
Salvanios pH7
Follow the disinfectant manufacturer's recommendation and instructions.
Alcohol (ethyl or isopropyl alcohol 70%)
Cidex OPA
Follow the disinfectant manufacturer's recommendation and instructions. at least 20 cycles.Number of cycles in
neodisher® MediClean forteCleaning agent Instrument parts should be steam autoclaved in a validated process, at a
temperature of 134°C (273°F) for 3 minutes. The Y sensor will last at least 20 autoclaving cycles with 3 minutes sterilization
time at 134ºC (273ºF).
SERVO-U/n Ventilator Systems, User's Manual
33
| Recommended intervals and disinfectants |
| 11
NAVA system
Surface cleaning
Wipe the surface with a soft cloth moistened in soap and water or in disinfectant.
In case of more contaminated surfaces, use ethyl alcohol (70%) or isopropyl alcohol (70%) together with lint free cleaning wipes.
Accessories
Surface cleaning
Wipe the surface with a soft cloth moistened in soap and water or in disinfectant.
In case of more contaminated surfaces, use ethyl alcohol (70%) or isopropyl alcohol (70%) together with lint free cleaning wipes.
34
SERVO-U/n Ventilator Systems, User's Manual
x
Manufacturer: Maquet Critical Care AB Röntgenvägen 2 171 54 Solna, Sweden Phone: +46 (0) 10 335 73 00 www.maquet.com
For local contact:
Please visit our website www.maquet.com
Getinge Group is a leading global provider of innovative solutions for operating rooms, intensive-care units, hospital wards, sterilization departments, elderly care and for life science companies and institutions. With a genuine passion for life we build quality and safety into every system. Our unique value proposition mirrors the continuum of care, enhancing efficiency throughout the clinical pathway. Based on our first-hand experience and close partnerships, we are able to exceed expectations from customers – improving the every-day life for people, today and tomorrow.