Getinge servoair_operman_6882040_r03_v2dot1_170223-en-non_us n Ventilator Systems Cleaning and Maintenance User's Manual

x
User's Manual
SERVO-air Ventilator System
Cleaning and Maintenance

Table of contents

| Table of contents |
1 2 3 4 5 6 7 8
4|Introduction
7|Patient unit and user interface 11|Expiratory cassette 15|O2 cell (maintenance) 17|Inspiratory channel (maintenance) 20|Aerogen Pro nebulizer 23|Accessories 24|Recommended intervals and disinfectants
SERVO-airVentilator System, User's Manual Infologic 1.18
3
| Introduction |
| 1

1 Introduction Table of contents

5|Introduction1.1
5|General information1.2
5|Safety guidelines1.3
6|Symbols1.4
4
SERVO-airVentilator System, User's Manual
| Introduction |
1 |

1.1 Introduction

Unless otherwise stated, the information in this User's Manual is valid for the SERVO-air ventilator system.
Here you will find the information needed to safely clean and maintain the ventilator system.

1.2 General information

As cleaning practices vary widely among health care institutions, it is not possible for Maquet to specify particular practices that will meet all needs, or to be responsible for the effectiveness of cleaning procedures carried out in the patient care setting.
Maquet recommends methods that have been validated using the specified equipment and procedures outlined in this manual. Other methods may work but are not covered by the warranty unless Maquet has given written permission.

1.3 Safety guidelines

WARNING! Indicates critical information about a potential serious outcome to the patient or the user.
CAUTION: Indicates instructions that must be followed in order to ensure the proper operation of the equipment.
Important: Indicates information intended to help you operate the equipment or its connected devices easily and conveniently.
Note: Indicates information requiring special attention.
CAUTIONS:
All personnel should be aware of the risk
of parts being infected when disassembling and cleaning the ventilator system.
The expiratory channel and expired gas
from the exhaust port may be contaminated.
All disposable parts must be discarded
according to hospital routines and in an environmentally safe way.
SERVO-airVentilator System, User's Manual
5
| Introduction |
| 1
Important:
Follow your hospital's routines for
handling infectious material when handling any part of the ventilator system.
If possible, cleaning should be performed
immediately after use and always before steam autoclaving/disinfection. Blood or other residues should not be allowed to dry onto the devices.
Water quality affects
cleaning/disinfection. Maquet recommends drinking water quality as the minimum quality level.
Maquet recommends the use of Servo
Duo Guard bacterial filter or equivalent to reduce the transmission of bacteria from the patient via the expiratory cassette to the ambient air. This reduces the risk of infection spread and prolongs the life of the expiratory cassette due to reduced cleaning requirements.
The use of a washer-disinfector for
cleaning/disinfection is recommended if bacterial filters are not used.
Steam autoclaving is normally not
necessary for the expiratory cassette, as it is not an invasive instrument, but when applied, use processes recommended by Maquet.
Notes:
Single use (i.e. disposable) items shall
not be reused. Cleaning of single use items can damage the item and compromise its intended use. Reuse of single use items increases the risk of spreading infections and compromises the durability of the item.
Maquet recommends that the equipment
is inspected regularly.
A preventive maintenance must be
performed by authorized personnel at least once every year as long as the unit is not used more than normal. Normal operation during one year is estimated to correspond to approximately 5000 hours of operation. The current operating time and time to next preventive maintenance is presented in SERVICE &
SETTINGS/BIOMED/STATUS/General
on the user interface.

1.4 Symbols

Hazardous wasteThe device
contains parts which must not be disposed of with ordinary waste.
Special waste
This product contains electronic and electrical components. Discard disposable, replaced and left-over parts in accordance with appropriate industrial and environmental standards.
6
SERVO-airVentilator System, User's Manual

2 Patient unit and user interface

| Patient unit and user interface |
2 |

2.1 Preparations

Note: Cleaning should be done after each patient or according to hospital routines.
Switch off the ventilator system using the
on/off switch on the patient unit.
Disconnect the ventilator system from the
mains power and gas supply.
Disconnect any optional equipment from
the mains power and from the ventilator system.
Discard disposable items according to
hospital routines and environmental regulations.
- Servo Duo Guard viral/bacterial filter
- SERVO humidifier/HME
- disposable patient tubing
- disposable airway adapter
- Aerogen Solo nebulizer
Remove the expiratory cassette.

2.1.1 Remove the expiratory cassette

Important: The expiratory cassette is a
precision instrument and must be handled carefully.
Tilt the user interface backwards.
Press the button on the expiratory cassette.
Tilt the cassette upwards and remove.
Important: After removing the expiratory cassette, do not pour any fluid into the expiratory cassette compartment. Avoid contact with electrical connectors.
SERVO-airVentilator System, User's Manual
7
| Patient unit and user interface |
| 2

2.2 Wiping and discarding

Wipe all parts with cleaning wipes or a soft lint-free cloth moistened in soap & water or detergent-based disinfectant.
In case of more contaminated surfaces, use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical connectors.
Important: Do not use abrasive or sharp tools to clean the user interface.

2.3 Assembling

2.3.1 Insert the expiratory cassette

Hinge a clean expiratory cassette and press
it firmly down into lock position.
Important: Make sure that the cassette clicks into position. Check that it cannot be moved upwards and that the button on top of the cassette is completely ejected.
Note on a log sheet that a routine cleaning
has been performed. Refer to hospital routines.
8
SERVO-airVentilator System, User's Manual
| Patient unit and user interface |
2 |

2.3.2 Pre-use check

After cleaning, always perform a pre-use check. Refer to the ventilator system User's Manual.
If the pre-use check fails:
Carefully shake/tilt the cassette (5-7 times).
Run the cassette in a ventilator system with
a test lung for 10 minutes.
Re-do the pre-use check.

2.4 Preventive maintenance

2.4.1 Check fan filters

Regularly check that the fan filter on both sides looks clean (i.e. black). If dusty, remove (snap off/snap on) and rinse in water. Shake out, ensuring that the filter is free from excess water.

2.4.3 Replace the dust filter

Check the dust filter and replace it once a month or more often when necessary.
Note: The dust filter must be replaced. If it is washed it will lose its characteristics.
Open the air inlet filter lid by pressing the
two latches towards each other.

2.4.2 Check air inlet filter

The air inlet filter must be exchanged once a year during preventive maintenance or more often when necessary.
SERVO-airVentilator System, User's Manual
Replace dust filter.
9
| Patient unit and user interface |
| 2
Close the lid.
10
SERVO-airVentilator System, User's Manual

3 Expiratory cassette

3.1 Preparations

Remove the expiratory cassette
Refer to section Remove the expiratory cassette on page 7.

3.2 Disinfection

| Expiratory cassette |
3 |
There are two disinfection procedures for the expiratory cassette and associated parts when no bacterial filter is used:
Washer-Disinfector (ISO 15883-1)
(recommended)
Disinfectant

3.2.1 Washer-disinfector

If rinsing before disinfection is not included in the washer disinfector program:
Rinse the parts thoroughly in water
(<35°C/95°F).
Let the water flow through the parts.
Important: Immediate rinsing in water (<35°C/95°F) can remove particles and reduces the risk of cross-contamination between patients.
Wash the parts with water only in a
washer-disinfector at a temperature of 85-95°C (185-203°F).
Refer to section Recommended intervals and disinfectants on page 24.
Place the expiratory cassette on its side
with the electrical connector uppermost.
Notes:
The water pressure in the
washer-disinfector should not exceed
1.5 bar.
The maximum water flow should not
exceed 10 l/min.
SERVO-airVentilator System, User's Manual
The water must have free passage
through the cassette.
11
| Expiratory cassette |
| 3

3.2.2 Disinfectant

Let the parts soak in disinfectant.
Refer to section Recommended intervals and disinfectants on page 24.
Important: Follow the disinfectant manufacturer’s recommendations and instructions, otherwise the cassette may be damaged or not disinfected properly.
Rinse after disinfection
Important:
Follow the disinfectant manufacturer's recommendation and instructions regarding rinsing.
Mineral deposits from disinfectant on the expiratory cassette affect the function. It is important to rinse the expiratory cassette thoroughly. Residues from chemicals can affect the patient, cause leakage and extra stress on the material.
12
Rinse the parts thoroughly in water to
remove all traces of disinfectant. Let the water flow through the parts.
Carefully shake and tilt the cassette, turn
upside down and repeat at least 5-7 times.
SERVO-airVentilator System, User's Manual
| Expiratory cassette |
3 |

3.3 Drying alternatives

Important: The expiratory cassette must be dried before use (if not dry, the expiratory cassette may not pass the pre-use check).
Always shake/tilt the cassette carefully (5-7 times).
There are several drying alternatives for the expiratory cassette:
Drying cabinet 1 hour in maximum 70°C
(158°F); if available, connect a 22 mm air hose to the expiratory cassette for extra drying effect, or
Drying the cassette in room air 12-24 hours,
depending on surrounding conditions.

3.3.1 Recommended position in a drying cabinet

3.3.2 Recommended position in room air

If the pre-use check is not passed after drying then:
Run the cassette in a ventilator system with
a test lung for 10 minutes, or
Dry the cassette in room air (if no drying
cabinet is available) 12-24 hours, depending on surrounding conditions.
Notes:
Never dry the cassette by applying
high-pressure air as the internal tubing may be damaged.
Drying may not be necessary if the
washer-disinfector has a drying phase.
SERVO-airVentilator System, User's Manual
13
| Expiratory cassette |
| 3

3.4 Steam autoclaving

Note: Steam autoclaving procedures are not recommended.
Steam autoclaving will reduce the lifetime of the expiratory cassette and is not recommend as a cleaning method.
Steam autoclaving is normally not necessary for the expiratory cassette, as it is not an invasive instrument, but when applied, use validated processes only.
Refer to section Recommended intervals and disinfectants on page 24.

3.4.1 Cleaning

Clean before steam autoclaving.
Refer to section Washer-disinfector on page 11.

3.4.4 Steam autoclave

Refer to section Recommended intervals and disinfectants on page 24.

3.4.5 Drying after steam autoclaving

Refer to section Drying alternatives on page 13.

3.5 Assembling

Refer to section Assembling on page 8.

3.4.2 Rinse

Rinse the parts thoroughly in water
(<35°C/95°F).
Let the water flow through the parts.

3.4.3 Drying before steam autoclaving

Before placing the expiratory cassette in a steam autoclave, make sure that no water remains inside the cassette. Refer to section Drying alternatives on page 13.
14
SERVO-airVentilator System, User's Manual

4O2 cell (maintenance)

4.1 Preventive maintenance

Regularly check the cell status in quick menu SYSTEM STATUS/O2 cell. The O2 cell should be exchanged if <10% estimated remaining capacity is indicated.
Refer to the ventilator system User's manuals for O2 cell adjustment.
cell (maintenance) |
| O
2
4 |

4.2 Preparations

WARNING! The sealed unit of the O2 cell contains a caustic liquid which may cause severe burns to the skin and eyes. In case of contact, immediately wash the affected area continuously with water for at least 15 minutes and seek medical attention, especially if the eyes are affected.
Important:
Make sure the O2 cell is for the intended
ventilator model.
Replacement of the O2 cell and filter and
extended cleaning must only be performed by trained personnel.
Unpack the O2 cell at least 15 minutes
before replacement. Turn the ventilator system off using the on/off switch on the patient unit.
Disconnect the ventilator system from the
power and gas supply.
Remove the expiratory cassette.
Remove the front cover.
Remove the O2 cell.
WARNING! The inspiratory channel cover shall be kept in place once maintenance has been performed. Otherwise the gas modules can be affected and deliver the wrong flow.

4.2.1 Remove the O2 cell

Unfasten and turn out the locking latch.
Disconnect the O2 cell connector.
SERVO-airVentilator System, User's Manual
15
| O2 cell (maintenance) |
| 4
Remove the O2 cell.

4.3 Assembling

4.3.1 Insert the O2 cell

Connect the O2 cell connector.
Fasten the locking latch.
Put the lower part of the front cover in
position.
Press the upper part of the front cover into
position.
Perform a pre-use check.
Put the O2 cell in position.
16
SERVO-airVentilator System, User's Manual

5 Inspiratory channel (maintenance)

Press the latch and gently pull out the

5.1 Preventive maintenance

Important: It is recommended that a regular cleaning and an extended cleaning of the inspiratory channel should be performed before carrying out preventive maintenance.

5.2 Preparations

inspiratory channel.
Disconnect the tube and remove the
bacterial filter. Discard the filter to hazardous waste.
| Inspiratory channel (maintenance) |
5 |
Remove the expiratory cassette.
Remove the front cover.
Remove the O2 cell.
Refer to section Remove the O2 cell on page 15.

5.2.1 Remove inspiratory channel and tube

5.3 Disinfection

5.3.1 Rinse before disinfection

If rinsing before disinfection is not included in the washer disinfector program:
Rinse the parts thoroughly in water
(<35°C/95°F).
Let the water flow through the parts.
Important: Immediate rinsing in water (<35°C/95°F) can remove particles and reduces the risk of cross-contamination between patients.
SERVO-airVentilator System, User's Manual
17
| Inspiratory channel (maintenance) |
| 5

5.3.2 Washer-disinfector

Place the inspiratory channel so that the
water flows through it.
Clean the inspiratory channel and tube
the same way as the expiratory cassette.
Refer to section Disinfection on page 11.

5.4 Assembling

5.4.1 Insert inspiratory channel and tube

Put a new bacterial filter in position and
connect the filter to the tube.
18
SERVO-airVentilator System, User's Manual
Push the inspiratory channel until it clicks
into locked position.
Re-insert the O
Refer to section Insert the O
cell.
2
cell on
2
page 16.
Mount the front cover.
Refer to section Assembling on page 16.
Note on a log sheet that an extended
cleaning of the inspiratory channel has been performed.
Insert the expiratory cassette.
Perform a pre-use check.
Refer to the ventilator system User's Manual.
| Inspiratory channel (maintenance) |
5 |
SERVO-airVentilator System, User's Manual
19
| Aerogen Pro nebulizer |
| 6

6 Aerogen Pro nebulizer

6.1 Preparation

Important: Do not use abrasive or sharp tools to clean the nebulizer unit.
Note: The Aerogen Solo nebulizer unit is intended for single patient use only, do not re-use, clean or sterilize.
Cleaning should be done after each patient or according to hospital routines.
Disconnect the nebulizer unit from the control cable, and remove the nebulizer unit, T piece and adapters from the patient circuit or mask.

6.2 Wiping and discarding

6.3 Disinfection

6.3.1 Cleaning

Ensure there is no medication remaining in
the nebulizer.
Disassemble the nebulizer unit and adapters
into individual components.
Remove the filler cap from nebulizer unit.
Clean all parts with warm water and mild
liquid detergent.
Rinse all parts in sterile water.
Shake excess water from parts and allow
parts to air dry.
Assemble the nebulizer.
Refer to the ventilator system User's Manual.
Wipe the control cable with a damp cloth.
Do not autoclave.
Check for exposed wiring, damaged
connectors, or other defects and replace if any are visible.
Note: In case of more contaminated surfaces, use ethyl alcohol (70%).
20

6.3.2 Washer-disinfector

Two different washing cycles are validated for the Aerogen Pro nebulizer:
Liquid alkaline cleaner (diluted as per
manufacturer instruction) and mains water.
Load the components in the
-
washer-disinfector.
- Pre-rinse the components for 3 minutes.
- Clean the components with liquid alkaline
cleaner at 55 °C (131 °F) for 10 minutes.
- Rinse for 1 minute.
- Rinse using thermal disinfection cycle at
93 °C (199.4 °F) for 10 minutes.
Mains water and without the use of a
detergent.
Load the components in the
-
washer-disinfector.
SERVO-airVentilator System, User's Manual
| Aerogen Pro nebulizer |
6 |
- Wash components for 10 minutes at 91 °C (195.8 °F).
- Drain the machine for 40 seconds.
- Rinse at 90 °C (194 °F) for 1 minute.
- Drain the machine for 40 seconds.
- Rinse at 90 °C (194 °F) for 1 minute.
- Drain the machine for 40 seconds.
- Dry at 90 °C (194 °F) for 15 minutes.

6.3.3 Disinfectant

Immerse the parts in appropriate
disinfecting agent.
Refer to section Recommended intervals and disinfectants on page 24.
Important: Follow the disinfectant manufacturer’s recommendations and instructions.

6.4 Steam autoclaving

Disassemble the nebulizer unit and adapters
into individual components.
Remove the filler cap from the nebulizer
unit.
Clean all parts with warm water and mild
liquid detergent. Rinse thoroughly and air dry.
Check for cracks or damage and replace if
any defects are visible.
Place the disassembled components into
appropriate sterilization wrapping.
Note: Do not reassemble parts prior to steam autoclaving
Only the Aerogen Pro nebulizer unit, T piece and neonate adapters can be autoclaved.
Refer to section Recommended intervals and disinfectants on page 24.
SERVO-airVentilator System, User's Manual
21
| Aerogen Pro nebulizer |
| 6
Note: Sterilization using the long autoclave cycle may cause some areas of the nebulizer to become discolored. This does not affect the performance of the nebulizer unit.
To sterilize with hydrogen peroxide gas plasma, place wrapped parts in a STERRAD® System, and use the long cycle.
Note: Users should refer to the product labelling for the STERRAD® 100S Sterilization System for specific instructions regarding its proper operation.

6.5 Assembling

Assemble the nebulizer. Perform a function test of the Aerogen Pro nebulizer after the cleaning.
Refer to the ventilator system User's Manual.
22
SERVO-airVentilator System, User's Manual

7 Accessories

7.1 Hardware accessories

7.2 Patient circuits

| Accessories |
7 |
Hardware accessories are:
Mobile cart
Humidifier holder
Support arm 179
Water bag/IV pole
Drawer
Cabinet
Gas cylinder restrainer kit
Shelf base
Hook and Y piece holder
Cable holder
Refer to section Recommended intervals and disinfectants on page 24.
Note: Discard the disposable patient circuits. The disposable patient circuits are for single patient use only.
Refer to the cleaning routines in the operating manual for the patient circuit.

7.3 Humidifier

Fisher & Paykel Humidifier MR850/MR810
Refer to the cleaning routines in the operating manual for the Fisher & Paykel Humidifier.
SERVO-airVentilator System, User's Manual
23
| Recommended intervals and disinfectants |
| 8

8 Recommended intervals and disinfectants

Patient unit and user interface
Surface cleaning
Expiratory cassette
Pre-disinfectant
Disinfectant
cycles Cleaning agent
washer-disinfector Steam autoclave
Number of cycles in steam autoclave
3M Respirator Cleaning Wipes
Clorox Bleach Germicidal Wipes (may cause discoloration)
Super Sani-Cloth Germicidal Wipes (may cause discoloration)
Virkon S
Accel TB (Virox 5)
In case of more contaminated surfaces, use ethyl alcohol (70%) or isopropyl
alcohol (70%) together with lint free cleaning wipes.
Aniosyme DD1
Gigazyme Plus
Salvanios pH7
Follow the disinfectant manufacturer's recommendation and instructions.
Alcohol (ethyl or isopropyl alcohol 70%)
Anioxyde 1000
Cidex OPA
Actanios HLD (prion decontamination agent), may cause discoloration
Follow the disinfectant manufacturer's recommendation and instructions. at least 100 cyclesNumber of immersion
neodisher® MediClean forte
Getinge MIS Detergent
at least 200 cycles.Number of cycles in
Instrument parts should be steam autoclaved in a validated process, at a
temperature of 134°C (273°F) for 3 minutes.
Rubber parts should be steam autoclaved in a validated process, at a
temperature of 121°C (250°F) for 15-20 minutes.
The expiratory cassette will last at least:
100 autoclaving cycles with 3 minutes sterilization time at 134ºC (273ºF).
30 autoclaving cycles with 18 minutes sterilization time at 134ºC (273ºF).
O2 cell
24
Replace the O2 cell when it is requested by the ventilator system.O2 cell
SERVO-airVentilator System, User's Manual
Inspiratory channel
Pre-disinfectant
Disinfectant
cycles Cleaning agent
washer-disinfector Number of cycles in
steam autoclave
Aerogen Pro nebulizer
Pre-disinfectant
Disinfectant
steam autoclave Steam autoclave
| Recommended intervals and disinfectants |
Aniosyme DD1
Gigazyme Plus
Salvanios pH7
Follow the disinfectant manufacturer's recommendation and instructions.
Alcohol (ethyl or isopropyl alcohol 70%)
Anioxyde 1000
Cidex OPA
Actanios HLD (prion decontamination agent), may cause discoloration
Follow the disinfectant manufacturer's recommendation and instructions. at least 10 cyclesNumber of immersion
neodisher® MediClean forte
Getinge MIS Detergent
at least 20 cycles.Number of cycles in
The inspiratory channel will last at least 10 autoclaving cycles with 3 minutes sterilization time at 134ºC (273ºF).
Hexanios G+R
Follow the disinfectant manufacturer's recommendation and instructions.
Isopropyl alcohol (70 %)
CIDEX
NU-CIDEX
Cidex OPA
at least 26 cyclesNumber of cycles in
Steam autoclaving can be performed using the following three methods:
Autoclave wrapped parts using steam autoclaving pre-vacuum cycle, a
minimum of 134°C (270°F - 275°F) for 3.5 minutes with drying cycle (134°C wrapped cycle).
Autoclave wrapped parts using steam autoclaving pre-vacuum cycle, a
minimum of 121°C (250°F) for 20 minutes with drying cycle (121°C wrapped cycle).
Autoclave wrapped parts using steam autoclaving pre-vacuum cycle, a
minimum of 134°C (270°F - 275°F) for 20 minutes with drying cycle (sometimes referred to as a “Prion cycle”).
Note: This cycle may cause some areas of the nebulizer to become discolored. This does not affect the performance of the nebulizer.
8 |
SERVO-airVentilator System, User's Manual
25
| Recommended intervals and disinfectants |
| 8
Accessories
Surface cleaning
Wipe the surface with a soft cloth moistened in soap and water or in disinfectant.
In case of more contaminated surfaces, use ethyl alcohol (70%) or isopropyl alcohol (70%) together with lint free cleaning wipes.
26
SERVO-airVentilator System, User's Manual
x
Manufacturer: Maquet Critical Care AB Röntgenvägen 2 171 54 Solna, Sweden Phone: +46 (0) 10 335 73 00 www.maquet.com
For local contact:
Please visit our website www.maquet.com
Getinge Group is a leading global provider of innovative solutions for operating rooms, intensive-care units, hospital wards, sterilization departments, elderly care and for life science companies and institutions. With a genuine passion for life we build quality and safety into every system. Our unique value proposition mirrors the continuum of care, enhancing efficiency throughout the clinical pathway. Based on our first-hand experience and close partnerships, we are able to exceed expectations from customers – improving the every-day life for people, today and tomorrow.
© Maquet Critical Care AB 2017. All rights reserved. • Maquet reserves the right to modify the design and specifications contained herein without prior notice.
• Order No. 68 82 040 • User's Manual • Printed in Sweden • 170223 • Rev: 03 English •
Loading...