Datex-Ohmeda S-5 Technical reference manual (2009)

GE Healthcare

Datex-Ohmeda S/5
Anesthesia Monitor
TM
TM
Datex-Ohmeda S/5
Critical Care Monitor

Technical Reference Manual

Conformity according to the Council Directive 93/42/EEC concerning Medical Devices
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the USA, check local laws for any restriction that may apply.
All specifications subject to change without notice.
Order code M1162897
st
edition
1
June 25, 2009
GE Healthcare Finland Oy Helsinki, Finland P.O. Box 900 FI-00031 GE, FINLAND Tel: +358 10 39411 Fax: +358 9 1463310
www.gehealthcare.com
Copyright © 2009 General Electric Company. All rights reserved.
Madison, WI 53707-7550, USA
Datex-Ohmeda Inc.
Tel: +1 608 221 1551
Fax: +1 608 222 9147
P.O. Box 7550
Intended purpose (Indications for use)
The Datex-Ohmeda S/5 Anesthesia Monitor with L-ANE07 or L-ANE07A software is intended for multiparameter patient monitoring with optional patient care documentation.
The S/5 Anesthesia Monitor with L-ANE07 and L-ANE07A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Entropy (State Entropy and Response Entropy) and neurophysiological status of all hospital patients.
When the BIS module is used with the S/5 Anesthesia Monitor with L-ANE07 and L-ANE07A, it is intended for use by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The Bispectral index (BIS), a processed EEG variable, and one component of the BIS module, may be used in adults as an aid in monitoring the effects of certain anesthetic agents. The Bispectral index is a complex technology, intended for use only as an adjunct to clinical judgement and training. In addition, the clinical utility, risk/benefit, and application of BIS have not undergone full evaluation in the pediatric population.
The S/5 Anesthesia Monitor with L-ANE07 and L-ANE07A software is also indicated for documenting patient care related information.
The S/5 Anesthesia Monitor with L-ANE07 and L-ANE07A software is indicated for use by qualified medical personnel only.
The Datex-Ohmeda S/5 Critical Care Monitor with L-ICU07 or L-ICU07A software is intended for multiparameter patient monitoring.
The S/5 Critical Care Monitor with L-ICU07 and L-ICU07A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion and neurophysiological status of all hospital patients.
When the BIS module is used with the S/5 Critical Care Monitor with L-ICU07 and L-ICU07A, it is intended for use by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The Bispectral index (BIS), a processed EEG variable, and one component of the BIS module, may be used in adults as an aid in monitoring the effects of certain anesthetic agents. The Bispectral index is a complex technology, intended for use only as an adjunct to clinical judgement and training. In addition, the clinical utility, risk/benefit, and application of BIS have not undergone full evaluation in the pediatric population.
The S/5 Critical Care Monitor with L-ICU07 and L-ICU07A software is indicated for use by qualified medical personnel only.
Classifications
In accordance with IEC 60601-1
Class I and internally powered equipment – the type of protection against electric shock. Type BF or CF equipment. The degree of protection against electric shock is indicated by a symbol on
each parameter module. Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen
or nitrous oxide. Continuous operation according to the mode of operation.
In accordance with IEC 60529
With F-CU8 Central Unit: IPX0 - the degree of protection against harmful ingress of water. With F-CU5(P) Central Unit: IPX1 - the degree of protection against harmful ingress of water.
In accordance with EU Medical Device Directive
The Datex-Ohmeda S/5 Anesthesia Monitor is classified as IIb. The Datex-Ohmeda S/5 Critical Care Monitor is classified as IIb. In accordance with CISPR 11: Group 1, Class A
• Group 1 contains all ISM (industrial, scientific and medical) equipment in which there is intentionally generated and/or used conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself.
• Class A equipment is suitable for use in all establishments other than domestic and those directly connected the public low voltage power supply network which supplies buildings used for domestic purposes.
Responsibility of the manufacturer
GE Healthcare Finland Oy (GE) is responsible for the effects on safety, reliability and performance of the equipment only if:
• assembly, extensions, readjustments, modifications, servicing and repairs are carried out by personnel authorized by GE.
• the electrical installation of the monitor room complies with appropriate requirements.
• the equipment is used in accordance with the "User's Guide" and serviced and maintained in accordance with the “Technical Reference Manual”.
Trademarks
Datex, Ohmeda, S/5, D-lite, D-lite+, Pedi-lite, Pedi-lite+, D-fend, D-fend+, Mini D-fend, OxyTip+, MemCard, ComWheel, ComBar, EarSat, FingerSat, FlexSat, PatientO
are trademarks of GE Healthcare Finland Oy. All other product and company names are property of their respective owners.
A portion of the Entropy software is derived from the RSA Data Security, Inc. MD5 Message-Digest Algorithm.
, Patient Spirometry, Entropy and Tonometrics
2
Masimo SET
Masimo SET is a licensed trademark of Masimo Corporation.
Product availability
Some of the products mentioned in this manual may not be available in all countries. Please, consult your local representative for the availability.
Datex-Ohmeda S/5TM Anesthesia and Critical Care Monitors
Technical Reference Manual, Order code: M1162897
Document No. Updated Description
1st edition
Part I, General Service Guide
M1137263-004 Introduction, System description, Installation,
Interfacing, Functional check, General
troubleshooting
M1125633-006 Planned Maintenance Instructions 2
Part II, Product Service Guide
Document No. Updated Description
M1137266 - 002 Service Menu, L-ANE07(A), L-ICU07(A) 1
M1125635 - 005 8-Module Frame, F-CU8 2
M1125636 - 005 5-Module Frame, F-CU5(P) 3
M1125637 - 004 CPU Board, B-CPU6 4
M1137269 - 002 UPINET Board, B-UPI4NET 5
M1137272 - 002 Displays and Display Controller Boards 6
M1125639 - 003 Command Bars 7
M1125641 - 003 Extension Frame, F-EXT, Extension Module, E-EXT 8
M1137274 - 001 AM, CCM Spare Parts 9
1
1
Document no. M1137263-04
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
2
Document no. M1137263-04

Table of contents

Table of contents
Table of contents i
Table of figures iv
List of tables v
About this manual 1
1Introduction 3
1.1 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.1 Symbols on transport packaging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.2 Symbols on equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.3 Equipment safety symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1.4 Other symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.2.1 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.2.2 ESD precautionary procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
1.2.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2 System description 17
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.2 Bus structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.3 Distributed processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4 Module communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5 Software loading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.6 Parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3 System installation 21
3.1 Unpacking instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2 Choosing location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3 Central Unit; S/5 8-Module Frame, F-CU8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3.1 Connecting to mains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3.2 Connecting to the Datex-Ohmeda Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3.3 Inserting the parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3.4 E-PSM(P) Mounting Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.3.5 Positioning PC boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.6 Replacing PC boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.3.7 Performing Factory Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.4 Central Unit; S/5 5-Module Frame, F-CU5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.4.1 Mounting the Frame. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.4.2 Connecting to mains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.4.3 Connecting to the Datex-Ohmeda Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.4.4 Inserting the parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.4.5 Positioning PC boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.4.6 Replacing PC boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Document no. M1137263-04
i
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
3.4.7 Performing Factory Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.5 Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.5.1 Main displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.5.2 Secondary displays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.5.3 3rd display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.5.4 Display installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.5.5 15” Flat Panel Display, D-FPD15 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.5.6 12” LCD Display, D-LCC12A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.5.7 19” medical grade display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.6 Display controller boards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3.6.1 Jumper settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3.6.2 Resolution selection for optional B-DISPX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.6.3 Resolution selection for primary display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.7 S/5 Remote Controller, K-REMCO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.8 S/5 Airway Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3.8.1 S/5 Compact Airway Modules, E-xxxx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3.8.2 Sample gas exhaust. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3.9 Record Keeping Keyboard for Anesthesia, K-ARKB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
3.9.1 Connection to Central Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
3.9.2 Connection to LCD Display, D-LCC12A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
3.10 ARK Barcode Reader, N-SCAN (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3.10.1 Connection to Central Unit/LCD Display or D-LCC12A . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3.11 S/5 Extension Frame, F-EXT4, with F-CU8 only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.11.1 Mounting of Extension Frame, F-EXT4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.11.2 Connection to Central Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.11.3 Inserting the parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.11.4 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4Interfacing 47
4.1 Interfacing external monitors via Interface Module, E-INT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.1.1 Connection to external Datex-Ohmeda monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.1.2 Connection to Critikon Dinamap 1846SX, Abbott Oximetrix 3 and
Baxter Explorer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.1.3 Connection to Baxter Vigilance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.1.4 Connection to Nellcor N-100 and N-1000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.1.5 Connection to Nellcor N-200. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.2 Interfacing external bedside devices via Device Interfacing Solutions, N-DISxxx. . . . . . . . 51
4.2.1 Device Interfacing Solution components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.2.2 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.2.3 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.2.4 Selecting the external device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.2.5 Functional check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.2.6 Selecting the parameter data source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.3 Interfacing Datex-Ohmeda Anesthesia Delivery Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4.3.1 Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4.3.2 Setting interfacing parameters on the S/5 Anesthesia Delivery Unit . . . . . . . . . . . . . . 56
4.3.3 Setting interfacing parameters on the S/5 Anesthesia Monitor. . . . . . . . . . . . . . . . . . . 56
4.4 Interfacing Dräger Cicero, Cato, Julian and Narkomed 2C (by NAD) . . . . . . . . . . . . . . . . . . . . 57
4.4.1 Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
ii
Document no. M1137263-04
Table of contents
4.4.2 Setting communication parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
4.4.3 Setting interfacing parameters on the S/5 Anesthesia Monitor or
S/5 Critical Care Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
4.5 Interfacing printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.5.1 Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.5.2 Setting interfacing parameters on the printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.5.3 Setting interfacing parameters on the S/5 Anesthesia Monitor or
S/5 Critical Care Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.6 Interfacing computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.7 Output signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4.7.1 UPI4NET Board output signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4.7.2 Digital outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4.7.3 Analog outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
4.7.4 S/5 Pressure Temp Module, E-PT, output signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5 Functional check 63
5.1 Recommended tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.1.1 Hemodynamic patient simulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
5.2 Visual inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
5.3 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
5.3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
5.3.2 Display(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
5.3.3 Keyboard(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
5.3.4 5-Module Central Unit, F-CU5/ 8-Module Central Unit, F-CU8 . . . . . . . . . . . . . . . . . . . . 68
5.3.5 Extension Frame, F-EXT4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
5.3.6 Compact Airway Module, E-CXXXXX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
5.3.7 Single width Airway Module, E-miniC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
5.3.8 Multiparameter Hemodynamic Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
5.3.9 Pressure/Pressure Temp Modules, E-P, E-PT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
5.3.10 Dual pressure Module, E-PP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
5.3.11 Cardiac Output Modules, E-COP, E-COPSv. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
5.3.12 Masimo Compatible Saturation module, E-MASIMO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
5.3.13 Nellcor Compatible Saturation module, E-NSATX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
5.3.14 BIS Module, E-BIS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
5.3.15 Entropy Module, E-ENTROPY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
5.3.16 Memory Module, E-MEM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
5.3.17 Recorder module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
5.3.18 Network connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
5.3.19 Interface Module, E-INT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
5.3.20 Interface module for PSM, E-INTPSM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
5.3.21 Device Interfacing Solution, N-DISxxx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
5.3.22 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
6 General troubleshooting 75
6.1 Software troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Appendix A: Functional check form, Datex-Ohmeda S/5 AM, CCM A-1
Appendix B: ElectroMagnetic Compatibility B-1
Document no. M1137263-04
iii
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

Table of figures

Figure 1 Datex-Ohmeda S/5 Anesthesia Monitor system........................................................................................................... 3
Figure 2 General bus structure of S/5 system ................................................................................................................................17
Figure 3 Distributed processing in S/5 system ...............................................................................................................................18
Figure 4 Principle of UPI section operation ......................................................................................................................................19
Figure 5 Software loading ........................................................................................................................................................................19
Figure 6 General structure of parameter modules with patient isolation.........................................................................20
Figure 7 Central Unit: S/5 8-Module Frame, F-CU8 ......................................................................................................................22
Figure 8 Module insert................................................................................................................................................................................23
Figure 9 E-PSM(P) mounting accessories ..........................................................................................................................................24
Figure 10 Rear view and positioning, F-CU8 (-12 shown).............................................................................................................26
Figure 11 Service reset button..................................................................................................................................................................27
Figure 12 F-CU5 parts connected with cables..................................................................................................................................29
Figure 13 Two F-CU5s connected to one F-CPU ..............................................................................................................................29
Figure 14 PC boards.......................................................................................................................................................................................32
Figure 15 Service reset button..................................................................................................................................................................33
Figure 16 Display options............................................................................................................................................................................34
Figure 17 Address dip switch settings, B-DISPX ...............................................................................................................................38
Figure 18 Resolution dip switch settings, B-DISPX ..........................................................................................................................39
Figure 19 Compact Airway Module, E-XXXX .......................................................................................................................................40
Figure 20 Scavenging through ventilator reservoir ........................................................................................................................41
Figure 21 Connecting the gas module to the scavenging connector of S/5 Avance .....................................................41
Figure 22 Sample gas returned to patient circuit in ADU.............................................................................................................42
Figure 23 Barcode Reader connected to LCD Display...................................................................................................................43
Figure 24 N-SCAN Barcode Reader connection directly to the keyboard...........................................................................44
Figure 25 S/5 Extension Frame, F-EXT4................................................................................................................................................44
Figure 26 Connection cables and LED indicators ............................................................................................................................53
Figure 27 An example of interfacing external devices with Device Interfacing Solution..............................................54
Figure 28 S/5 AM, CCM general troubleshooting flowchart........................................................................................................75
iv
Document no. M1137263-04

List of tables

List of tables
Table 1 Transference of parameters, Datex-Ohmeda monitors ...................................................................... 48
Table 2 Transference of parameters, external monitors...................................................................................... 48
Table 3 DIS modules and interfaced devices............................................................................................................. 51
Table 4 Parameters transferred from S/5 Anesthesia Monitor to S/5 Anesthesia Delivery Unit....... 56
Table 5 Events transferred from S/5 Anesthesia Delivery Unit to the S/5 Anesthesia Monitor.......... 57
Table 6 Parameters transferred from Dräger Cicero monitor to S/5 Anesthesia Monitor or
S/5 Critical Care Monitor ..................................................................................................................................... 58
Table 7 Parameters transferred from Dräger Cato, Julian and Narkomed 2C (NAD) monitor to
the S/5Anesthesia Monitor or S/5Critical Care Monitor........................................................................ 59
Table 8 Coding element connector, X4......................................................................................................................... 61
Table 9 Defib & IABP sync connector, X7 (B-UPI4NET -02) and on the front of F-CPU ........................... 61
Table 10 Signal output connector pin assignments.................................................................................................. 62
Table 11 Patient simulators’ compatibility with each hemodynamic module.............................................. 65
Table 12 Adapter cables for hemodynamic patient simulators.......................................................................... 65
Table 13 Guidance and manufacturer’s declaration – electromagnetic emissions ................................ B-1
Table 14 Guidance and manufacturer’s declaration – electromagnetic immunity................................. B-2
Table 15 Guidance and manufacturer’s declaration – electromagnetic immunity................................. B-3
Table 16 Recommended separation distances between portable and
mobile RF communications equipment and the S/5™ AM or CCM.............................................. B-5
Document no. M1137263-04
v
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
vi
Document no. M1137263-04

About this manual

Intended audience
This Technical reference manual is meant for service representatives and technical personnel who install, configure, maintain, administer, troubleshoot or repair Datex-Ohmeda S/5
Anesthesia and Critical Care Monitors.
Notes to the reader
As the monitor setup may vary, some functions described may not be available in the monitor you are using.
The order code for the printed manual is M1162897. The manual includes Technical
Reference Manual Slots and every slot has an individual document number. M1137263 is the document number of this first slot.
Part I gives the reader an overview of the S/5 Anesthesia Monitor and S/5 Critical Care
Monitor. It contains the information needed to install, interface and troubleshoot the monitors. Instructions for functional check and planned maintenance are also included. Read the manual through and make sure that you understand the procedures described before the installation of the monitor. To avoid risks concerning safety or health, strictly observe the warning indications. If you need any assistance concerning the installation, please do not hesitate to contact your authorized distributor.
Part II contains detailed descriptions of each component of the S/5 AM, CCM Monitor,
such as frame unit, parameter modules Remote Controller and Device Interfacing Solution. Service check for each product, service menus and all the spare parts information for the Monitor is included.
For information of parameter modules, Remote Controller and Device Interfacing Solution refer to the “S/5 E-Modules, Technical Reference Manual”. Service check for each of these products is included in these slots.
The manufacturer reserves the right to change product specifications without prior notice. Although the information in this manual is believed to be accurate and reliable, the manufacturer assumes no responsibility for its use.
Installation and service are allowed by authorized service personnel only. GE Healthcare Finland Oy (GE) assumes no responsibility for the use or reliability of its software
in equipment that is not furnished by GE.
Related documentation
S/5 Modules
S/5 E-Modules, Technical Reference Manual
S/5 Anesthesia Monitor
For instructions for daily use including cleaning and daily maintenance, clinical aspects and basic methods of measurement see: S/5 Anesthesia Monitor, User’s Guide S/5 Anesthesia Monitor, User’s Reference Manual
Document no. M1137263-04
1
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
S/5 Critical Care Monitor
For instructions for daily use including cleaning and daily maintenance, clinical aspects and basic methods of measurement: S/5 Critical Care Monitor, User’s Guide S/5 Critical Care Monitor, User’s Reference Manual
For more information about the iCentral, S/5 Arrhythmia Workstation and anesthesia record keeping solution, see the “Technical Reference Manuals” and “User’s Reference Manuals” for these products.”
Software options and default settings are described in the “Default Configuration Worksheet” delivered with each monitor.
Available accessories are described in the “Supplies and Accessories” catalog delivered with each monitor.
Conventions used
To help you find and interpret information easily, the manual uses consistent text formats:
Sign the check form after performing the procedure.
"
Hard Keys Names of the hard keys on the Remote Controller, Command Bar and modules are written in
the following way:
Menu Items Software terms that identify window parts or menu items are written in bold italic: ECG Setup.
Menu access is described from top to bottom. For example, the selection of the
Others.
Monitor
Setup hard key, the Screen 1 Setup menu item and the Waveform Fields menu item would be
shown as
‘Messages’ Messages (alarm messages, informative messages) displayed on the screen are written inside
single quotes: ‘Please wait’.
“Sections” When referring to different sections in this manual or to other manuals, manual names and
section names are enclosed in double quotes: See section "Cleaning and care."
Please refer to "User's Reference Manual: Alarms."
Hypertext links Hypertext links on PDF versions are written in blue color.
WARNING Warnings are written in the following way:
Monitor Setup - Screen 1 Setup - Waveform Fields.
WARNING This is a WARNING.
CAUTION Cautions are written in the following way:
CAUTION This is a CAUTION.
NOTE Notes are written in the following way:
NOTE: This is a NOTE.
In this manual, the word “select” means choosing and confirming.
Illustrations and names
All illustrations in this manual are only examples, and may not necessarily reflect your system
settings or data displayed in your system. If a particular selection is not available in your
system, the selection is shown grayed.
2
Document no. M1137263-04

1 Introduction

The Datex-Ohmeda S/5 Anesthesia Monitor is a modular multiparameter patient monitor
primarily used during anesthesia in operating rooms.
The Datex-Ohmeda S/5 Critical Care Monitor provides full patient profile throughout the care
period.
The modular design makes the system flexible and easy to upgrade. In addition to parameter
changes, the modularity includes an easy upgrade to anesthesia record keeping, monitor
networking and interfacing with other external devices.
NOTE: Your system may not include all these components. Consult your local representative for
the available components.
Introduction
Figure 1 Datex-Ohmeda S/5 Anesthesia Monitor system
(1) 12" LCD display, D-LCC12A
(2) 15” LCD display, D-FPD15
(3) 19" medical grade display
(4) F-CU5P, 5-module frame unit with E-PSM(P) support, and measurement modules
(5) Patient Side Module, E-PSM(P)
(6) F-CPU, central processor unit for F-CU5(P), and N-AC, power unit
(7) F-CU5, 5-module frame unit, and measurement modules
(8) Extension Frame, F-EXT4
(9) Extension Module, E-EXT
(10) F-CU8, 8-module Central Unit, and measurement modules
(11) Detachable Command Bar to be used with displays not having an integrated Command
Bar (12) Anesthesia record keeping keyboard for automated record keeping (13) Remote Controller, K-REMCO (14) Printer
3
Document no. M1137263-04
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

1.1 Symbols

1.1.1 Symbols on transport packaging
The contents of the transport package are fragile and must be handled with care.
Indicates the correct upright position of the transport package.
The transport package must be kept in a dry environment.
Indicates the temperature limitations within which the transport package should be stored.
This package can be recycled.
4
Document no. M1137263-04
1.1.2 Symbols on equipment
The battery contains lead acid, and in the event of disposal must be separated from other waste according to local regulations.
Pb
The separate collection symbol is affixed to a battery, or its packaging, to advise you that the battery must be recycled or disposed of in accordance with local or country laws. The letters below the separate collection symbol indicate whether certain elements (Pb=Lead, Cd=Cadmium, Hg=Mercury) are contained in the
Pb/Cd/Hg
battery. To minimize potential effects on the environment and human health, it is important that all marked batteries that you remove from the product are properly recycled or disposed. For information on how the battery may be safely removed from the device, please consult the technical or service manual, or equipment instructions. Information on the potential effects on the environment and human health of the substances used in batteries is available at this url: http://www.gehealthcare.com/euen/weee-recycling/index.html
Introduction
Pb
This battery contains lead and can be recycled.
Dangerous voltage.
When using the ARK Barcode Reader, N-SCAN, do not stare into beam. The N-SCAN Barcode Reader is a Class 2 laser product.
Document no. M1137263-04
5
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
1.1.3 Equipment safety symbols
- Attention, consult accompanying documents.
- When displayed next to the O is set below 21%.
- When displayed next to the HR value, indicates that the pacer is set on R.
- On the modules or frames indicates the following warning:
WARNING Do not use modules with identical measurements in
the same monitor. If such modules have been inserted, remove the module that has been most recently connected. You can also remove both modules and re-connect the new module after five seconds.
- On the 15” Flat Panel Display, D-FPD15-00, indicates the following warning:
WARNING The display must only be used together with the
original type of D-FPD15 power adapter. The display should be supplied from the mains outlet.
- On the 19” display, D-LCC19, indicates the following warning:
WARNING The display must only be supplied from the mains
outlet via an appropriate additional separating transformer and the original D-LCC19 power adapter, not from the Central Unit, F-CU8.
value, indicates that the FiO2 low alarm limit
2
- On the E-PRESTN, E-PRETN, E-RESTN, E-PSM, E-PSMP, E-P, E-PP, E-PT, E-COP and E-COPSv module indicates the following warning:
WARNING Protection against cardiac defibrillator discharge is
due in part to the accessories for pulse oximetry (SpO2), temperature (T) and invasive pressure (P) measurement.
- On the E-NMT module indicates the following warnings:
WARNING Do not place the NMT stimulating electrodes on the
patient’s chest.
WARNING Always stop the NMT measurement before handling
the stimulating electrodes.
WARNING Never subject a patient with an implanted electronic
device to electrical stimulation without consulting a medical specialist first.
- On the rear or bottom panel this symbol indicates the following warnings and caution:
WARNING Electric shock hazard. Do not open the cover or the
back. Refer servicing to qualified service personnel.
WARNING Disconnect from the power supply before servicing. WARNING Do not use the monitor without manufacturer
approved mounting attached.
CAUTION For continued protection against fire hazard, replace the
fuse only with one of the same type and rating.
6
Document no. M1137263-04
Introduction
- BIS: On the Aspect DSC indicates the following caution:
CAUTION The converter must not be opened for any reason or
autoclaved.
- On the Interface Module E-INT indicates that it is for connecting external
devices. Do not connect patient cables to the module.
Type BF (IEC 60601-1) protection against electric shock.
Type BF (IEC 60601-1) defibrillator-proof protection against electric shock.
Type CF (IEC 60601-1) protection against electric shock.
Type CF (IEC 60601-1) defibrillator-proof protection against electric shock.
1.1.4 Other symbols
When displayed in the upper left corner of the screen, indicates that the alarms are silenced. When displayed in the menu or digit fields, indicates that the alarm source has been turned off or alarm does not meet the alarm-specific activation criteria.
ESD warning symbol for electrostatic sensitive devices. Pins of connectors identified with the ESD warning symbol should not be touched. Connections should not be made to these connectors unless ESD precautionary procedures are used. For details, see section “1.2.2. ESD precautionary procedures”.
Symbol for non-ionizing electromagnetic radiation. Interference may occur in the vicinity of equipment marked with this symbol.
Equipotentiality. Monitor can be connected to potential equalization conductor.
Alternating current
Fuse. Replace the fuse only with one of the same type and rating.
SN, S/N
Serial Number
Connector for color display
7
Document no. M1137263-04
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
Signal/power output
Signal/power input
Signal/power input/output
Connector for defibrillator synchronization
Connector for the S/5 Device Interfacing Solution, DIS
Power input
Signal input
Power input
Submenu. Selecting an alternative marked with this symbol in a menu opens a new menu.
The monitor is connected to the Datex-Ohmeda Network (LAN).
Data Card (green) and/or Menu Card (white) is inserted.
Ethernet connector
A blinking heart next to the heart rate or pulse rate value indicates the beats detected.
8
Document no. M1137263-04
A lung next to the respiration rate value indicates that respiration rate is calculated from the impedance respiration measurement.
Gas inlet
Gas outlet
Do not reuse.
Use by. Indicates the last use day.
Date of manufacturer
Does not contain Latex.
Introduction
Do not immerse the sensor in liquids.
Document no. M1137263-04
9
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
IPX class:
IPX0 IPX1 IPX2 IPX3 IPX4 IPX7 IPX8
20
Degree of protection against harmful ingress of water as detailed in the IEC 60529:
- Ordinary equipment
- Protection against vertically falling water drops.
- Protection against vertically falling water drops when enclosure tilted up to 15 °.
- Protected against spraying water.
- Protected against splashing water.
- Protected against the effects of temporary immersion in water.
- Protected against the effects of continuous immersion in water.
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.
This symbol indicates the product contains hazardous materials in excess of the limits established by the Chinese standard SJ/T11364-2006 Requirements for Concentration Limits for Certain Hazardous Substances in Electronic Information Products. The number in the symbols is the Environment-friendly Use Period (EFUP), which indicates the period during which the toxic or hazardous substances or elements contained in electronic information products will not leak or mutate under normal conditions so that the use of such electronic information products will not result in any severe environmental pollution, any bodily injury or damage to any assets. The unit of the period is “Year”.
In order to maintain the declared EFUP, the product shall be operated normally according to the instructions and environmental conditions as defined in the product manual, and the periodic maintenance schedules specified in Product Maintenance Procedures shall be followed strictly.
Consumables or certain parts may have their own label with an EFUP value less than the product. Periodic replacement of those consumables or parts to maintain the declared EFUP shall be done in accordance with the Product Maintenance Procedures.
This product must not be disposed of as unsorted municipal waste, and must be collected separately and handled properly after decommissioning.
10
Document no. M1137263-04
Introduction

1.2 Safety

The following list contains general warnings and cautions you should know before installing, maintaining or servicing the system. Warnings and cautions specific to the use of the system can be found in the User’s Guide and User’s Reference Manual.
1.2.1 Safety precautions Warnings
WARNING A WARNING indicates a situation in which the user or the patient may be in
danger of injury or death.
The device is not able to withstand unpacked drops from a height of 1 m without
damaging the module latches. If the device is dropped, please service the device before taking it back into use.
Power connection
Always check that the power cord and plug are intact and undamaged.
Do not use the power cord delivered with this product for any other product or purpose.
Use only hospital-grade grounded power outlets and power cord. Do not remove the
grounding pin from the power plug.
Use only an intact power cord. Replace the power cord if it is cracked, frayed, broken or
otherwise damaged.
Do not apply tension to the power cord otherwise the cord may get damaged.
Do not use an additional multiple socket outlet, extension cord or adapters of any kind.
Before starting to use the system, ensure that the whole combination complies with the
international standard IEC 60601-1-1 and with the requirements of the local authorities. Do not connect any external devices to the system other than those specified.
When detaching Patient Side modules, be careful not to drop them. Always support with
one hand while pulling out with the other.
To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
Installation
Keep the monitor horizontal when the Compact Airway Module is used. T ilting the
monitor may cause erroneous results in the Compact Airway Module’s readings and damage the module.
The monitor or its components should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the monitor and its components should be observed to verify normal operation in the configuration in which it will be used.
Pins of connectors identified with the ESD warning symbol should not be touched.
Connections should not be made to these connectors unless ESD precautionary procedures are used. For details, see section “1.2.2. ESD precautionary procedures
After transferring or reinstalling the monitor, always check that it is properly connected and
all parts are securely attached. Pay special attention to this in case of stacked mounting.
Do not use the monitor in high electromagnetic fields (for example, during MRI.)
Never install the monitor or the displays so that they are above the patient.
Document no. M1137263-04
11
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
A secondary display and printer must always be supplied from an additional transformer
providing at least basic isolation (isolating or separating transformer.) Without an appropriate transformer the leakage current of the secondary display can be too high.
A printer or computer must be supplied from an additional transformer providing at least
basic isolation (isolating or separating transformer).
If you accidentally drop the monitor, modules or frames, have them checked by
authorized service personnel prior to clinical use.
To avoid explosion hazard, do not use the monitor in presence of flammable anesthetics.
The monitor measures only non-flammable anesthetics.
Do not touch the patient, table, instruments, modules or the monitor during defibrillation.
Laser radiation
When using the ARK Barcode Reader, N-SCAN, do not stare into the beam. The N-SCAN is
a Class 2 laser product.
External connection
Do not connect any external devices to the monitor other than those specified.
Explosion hazard
To avoid explosion hazard do not use the monitor in the presence of flammable
anesthetics.
Patient safety
Do not perform any testing or maintenance on the monitor while it is being used on a
patient.
PACEMAKER PATIENTS: The impedance respiration measurement may cause rate
changes in Minute Ventilation Rate Responsive Pacemakers. In this case, set the pacemaker rate responsive mode off or turn the monitor impedance respiration measurement off.
Never install the monitor or the displays so that they are above the patient.
The monitor must not be used without manufacturer approved mounting attached.
Operation of the monitor outside the specified values may cause inaccurate results.
Autoclaving and sterilizing
Do not autoclave any part of the system with steam or sterilize with ethylene oxide.
Cleaning and service
Only trained personnel with proper tools and test equipment should perform the tests
and repairs described in this manual. Unauthorized service may void the monitor warranty.
Always unplug the monitor before cleaning or service. After cleaning or service ensure
that every part of the monitor is dry before reconnecting it to the power supply.
Do not touch any exposed wire or conductive surface while any cover is removed and the
monitor is energized. The voltages present can cause injury or death.
Pins of connectors identified with the ESD warning symbol should not be touched.
Connections should not be made to these connectors unless ESD precautionary procedures are used. For details, see section “1.2.2. ESD precautionary procedures”.
12
Document no. M1137263-04
Introduction
Electrostatic discharge through the PC boards may damage the components. Before
handling PC boards, wear a static control wrist strap. Handle all PC boards by their non-conductive edges and use anti-static containers when transporting them.Do not break or bypass the patient isolation barrier when testing PC boards.
Always perform an electrical safety check and a leakage current test on the monitor after
service.
Handle the water trap and its contents as you would any body fluid. Infectous hazard
may be present.
Do not immerse any part of the device in any liquid, or allow liquid to enter the monitor or
modules.
If liquid has accidentally entered the system or its parts, disconnect the power cord from
the power supply and have the equipment serviced by authorized service personnel.
Since calibration gas contains anesthetic agents, always ensure sufficient ventilation of
the room during calibration.
Accessories
Use only accessories, including mounts and batteries, and defibrillator-proof cables and
invasive pressure transducers approved by GE Healthcare. For a list of approved supplies and accessories, see the “Supplies and Accessories” catalog delivered with the monitor. Other cables, batteries, transducers and accessories may cause a safety hazard, damage the equipment or the system, result in increased emissions or decreased immunity of the equipment or system or interfere with the measurement. Protection against cardiac defibrillator discharge is due in part to the accessories for pulse oximetry (SpO2), temperature (T) and invasive pressure (P) measurement.
Single use accessories are not designed to be reused. Reuse may cause a risk of
contamination and/or affect the measurement accuracy.
Special components
Special components are used in these monitors that are vital to assure reliability and safety. GE Healthcare assumes no responsibility for damage, if replacement components not approved by GE Healthcare are used.
Batteries
The battery packages in the central unit, F-CPU and in the power supply unit of F-CU8 contain lead acid (Pb) which is hazardous to the environment and therefore needs to be disposed of carefully according to local regulations.
Refresh the batteries completely every six months. To replace the batteries safely, please refer to the service instructions in this manual.
Do not short-circuit the battery terminals, this may produce a very high current, which will
damage the battery.
Do not dispose of the battery into open flame, nor put the battery near fire, as it may
explode.
Do not dismantle the battery. It contains electrolyte, which may damage clothing or
cause injury to skin or eyes. If exposed to electrolyte, wash the injured area with plenty of clean water and contact a doctor.
See also section “Symbols”.
13
Document no. M1137263-04
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
Cautions
CAUTION A CAUTION indicates a condition that may lead to equipment damage or
malfunction.
Installation
Leave space for air circulation to prevent the monitor from overheating.
Ensure that the module is properly orientated (i.e. module release latch facing downward)
before insertion.
Before connecting the power cord to the power supply, check that the local voltage and
frequency correspond with the rating stated on the device plate. See instructions for different displays in section “Displays”.
Turn off the power before making any rear panel connections.
Before use
Allow two minutes for warm-up and note any error messages or deviations from normal
operation.
Clean the rear panel fan dust filters once a month or whenever necessary.
Do not connect a sampling line to the female Patient Spirometry connector while the
other end of the sampling line is connected to the D-fend water trap. The pressure in the gas sampling system may cause damage to the PVX unit pressure transducers.
Fuse replacement
Replace a fuse only with one of the same type and rating.
Cleaning and service
Do not use hypochlorite-, acetone-, phenol- or ammonia -based cleaners, abrasive
material or harsh chemicals as they may damage the surfaces of the device.
Do not use abrasive cleaning compounds, instruments, brushes or rough-surface
materials.
Do not apply pressurized air to any outlet or tubing connected to the monitor. Pressure
may destroy sensitive elements.
Do not clean the spirometry tubes with high pressure air or O
spirometry tubes are connected to Patient Spirometry connector. High differential pressure may damage PVX unit pressure transducers.
flushing while the
2
Special components
A lithium battery on the CPU Board. Dispose of the faulty IC containing the battery
according to local regulations.
Storage and transport
Do not store or transport the monitor outside the specified temperature, pressure and humidity ranges:
Temperature -10...+50 °C/14...122 °F Atmospheric pressure 660...1060 hPa/500...800 mmHg/660...1060 mbar Relative humidity 10...90% noncondensing For display specific environmental requirements see specifications in the “Display” slot.
14
Document no. M1137263-04
1.2.2 ESD precautionary procedures
To avoid electrostatic charges building up, it is recommended to store, maintain and use
the equipment at a relative humidity of 30% or greater. Floors should be covered by ESD dissipative carpets or similar. Non-synthetic clothing should be used when working with the component.
To prevent applying a possible electrostatic discharge to the ESD sensitive parts of the
equipment, one should touch the metallic frame of the component or a large metal object located close to the equipment. When working with the equipment and specifically when the ESD sensitive parts of the equipment may be touched, a grounded wrist strap intended for use with ESD sensitive equipment should be worn. Refer to the documentation provided with the wrist straps for details of proper use.
ESD precautionary procedure training
It is recommended that all potential users receive an explanation of the ESD warning symbol and training in ESD precautionary procedures.
The minimum contents of an ESD precautionary procedure training should include an introduction to the physics of electrostatic charge, the voltage levels that can occur in normal practice and the damage that can be done to electronic components if they are touched by an operator who is electrostatically charged. Further, an explanation should be given of methods to prevent build-up of electrostatic charge and how and why to discharge one’s body to earth or to the frame of the equipment or bond oneself by means of a wrist strap to the equipment or the earth prior to making a connection.
Introduction
1.2.3 Disposal
Dispose of the whole device, parts of it and its packing material and manuals in accordance with local environmental and waste disposal regulations.
15
Document no. M1137263-04
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
16
Document no. M1137263-04

2 System description

2.1 Introduction

Datex-Ohmeda monitors build up a freely configurable modular system. The architecture is designed to enable different module combinations so that the user is able to get the desirable parameter and feature set. This modular approach makes it possible to add new features when they are needed.

2.2 Bus structure

The operation of Datex-Ohmeda monitors is based on two communication channels, the CPU bus and module bus.
In the 5-Module Central Unit, PC boards receive power from the F-CPU power supply and the parameter modules receive power from a separate power supply in the 5-Module Frame unit. These power supplies are both fed by the N-AC Power Unit. In the 8-Module Central Unit, F-CU8, all PC boards connected to the CPU bus, as well as the parameter modules attached to the module bus receive power from the same power supply, which is an integral part of the Central Unit, F-CU8.
System description
UPI CPU
Parameter Module
CPU Bus
Module Bus
Parameter Module
Display Controller
Power Supply Unit
Parameter Module
Figure 2 General bus structure of S/5 system
The CPU bus is a communication channel used only for internal data transfer. It is based on the ISA bus used in IBM PC computers. Data is transferred on this 16 bit wide bus using the CPU clock frequency.
The module bus is used to connect the parameter modules to the Central Unit. The bus is based on the industry standard RS-485, which uses a differential serial method to transfer data. This type of bus is robust and it allows parameter modules to be inserted or removed while the power is on. The module bus uses a 500 kbps data transfer rate and can be used for longer distances than the CPU bus, e.g. for external frame connections.
The RS-485 type of serial communication supports so-called multidrop or party line connections. This means that all parameter modules connected to the module bus use exactly the same lines for communication. The advantage of this is that all bus connectors are identical and the modules can be connected in any order and position.
17
Document no. M1137263-04
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

2.3 Distributed processing

A system assembled from Datex-Ohmeda products is a multiprocessor system. All parameter modules have their own microprocessor, which performs functions such as module key control, waveform filtering, parameter related computing and pneumatic control, etc. At the same time the main CPU performs higher level tasks such as trending and alarm control. While the parameter modules and CPU are performing their tasks, the UPI (Universal Peripheral Interface) microprocessor handles all functions needed to transfer data between the parameter modules and the CPU. At the same time the Display controller microprocessor performs pixel calculations for graphics.
Module
RAM
Display
Command Board/Bar
UPI
Dual-port
RAM
Figure 3 Distributed processing in S/5 system
This kind of parallel processing gives one major advantage to centralized processing. When new parameter modules or PC boards are added to the system, the processing power is increased. As a result, the system does not slow down when new features are added.

2.4 Module communication

The communication master controlling data transfers between the CPU and parameter modules is called UPI processor. It sends data to each connected module 100 times a second. Modules respond to each data request immediately by sending a data package, whose length depends on the type of the module. This communication protocol ensures that each module receives and sends data every 10 ms. If a module does not respond to data requests, the UPI processor presumes that the module is disconnected.
Parameter modules may hold a static (fixed) or dynamic address, which the UPI processor uses when sending out data. Two parameter modules of the same type must not be fitted onto the same monitor since they might reply to a data request simultaneously, thus causing communication errors.
CPU
Display
Controller
System
memory
Display
memory
18
Document no. M1137263-04
Loading...
+ 423 hidden pages