Datex-Ohmeda Tec7 Service manual

4.8 (4)
Tec 7 Vaporizer
User’s Reference Manual

User Responsibility

w
CAUTION
This Product will perform in conformity with the description thereof contained in this User’s Reference manual and accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. Should repair or replacement become necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be made to the nearest Datex-Ohmeda Customer Service Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex-Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered without the prior written approval of Datex-Ohmeda. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Datex-Ohmeda.
U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the U.S.A, check local laws for any restriction that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code. The year of manufacture and a sequential unit number for identification.
AAA A 12345
T
his alpha character indicates the year of product manufacture and when the serial number was assigned; “C” = 1999, “D” = 2000, “E” = 2001, etc. “I” and “O” are not used.

1 Introduction

2 Description

3 Setup and Mounting Procedure

Table of Contents

Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
What is a Tec 7 Vaporizer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Control dial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Safety interlocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Vaporizer identification label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Vaporizer mounting procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Mounting the vaporizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

4 Operating Instructions

Checking the vaporizer for correct mounting . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Removing the vaporizer from a manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Setting the dial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Filling and draining the vaporizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Filling procedure with funnel filler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Draining procedure with funnel filler . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Filling procedure with Easy-Fil
Draining procedure with Easy-Fil . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Filling procedure with Quik-Fil™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Draining procedure with Quik-Fil . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
1175-0013-000
i
Tec 7 Vaporizer

5 Maintenance

User maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
External cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Internal contamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Output concentration check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Checking the calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Analytical techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5

6 Principle of Operation

7 Specifications

Interlock mechanism . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Delivery of gas/agent vapor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Bypass circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Vaporizing chamber circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Weight and dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Flow characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Effects of variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Anesthetic agent consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Barometric pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Back pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Carrier gas composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-10

Warranty

ii
Time out of service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-10
Effects of variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
1175-0013-000

1 Introduction

Precautions

ww
ww
WARNING Do not fill the vaporizer with any agent other than the agent specified on
the front label. The vaporizer is designed for that agent only. If any substance other than that specified is used, patient injury could occur.
United States (U.S.) Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the U.S., check local laws for any restrictions that may apply.
Do not attempt to use a vaporizer that has been dropped. A dropped vaporizer MUST be sent to the nearest Datex-Ohmeda Field Operations Unit for servicing.
Do not use malfunctioning equipment. Make all necessary repairs or have the equipment serviced by an authorized Datex-Ohmeda service center. After repair, test the equipment to ensure that it is functioning properly in accordance with the manufacturer’s published specifications.
1175-0013-000
1-1
y
Tec 7 Vaporizer

Symbols

Important
ww
ww
WARNING Warnings tell about a condition that can cause injury to the operator or
European Standard EN 740 ­that an appropriate gas monitor is used to monitor the concentration of anesthetic agent vapor in the inspiratory gas when the vaporizer is in operation in order to provide protection against hazardous output in the event of a device malfunction.
Datex-Ohmeda strongly recommends the use of anesthesia gas monitoring with this equipment. Refer to local standards for mandatory monitoring.
Requests for servicing facilities, advice or assistance must be addressed to a local Datex-Ohmeda office.
Additional copies of this manual, can be requested from a local Datex-Ohmeda Field Operations Unit or a Datex-Ohmeda Authorized Distributor.
Datex-Ohmeda strongly recommends that you keep all relevant documentation, including this manual and accompanying labels, immediately available to all users.
Warnings and Cautions tell you about conditions that can occur if you do not follow all instructions in this manual.
the patient.
Anesthetic Workstations and Their Modules
requires
onl
SEV
ISO
ENF
HAL
ww
ww
CAUTION Cautions tell about a condition that can cause damage to the
equipment. Read and follow all warnings and cautions.
Caution: federal law prohibits dispensing without prescription.
Sevoflurane
Isoflurane
% v/v
OFF symbol/OFF setting
Percentage of anesthetic vapor per total volume.
Lock
z
Enflurane
Unlock
Z
Halothane
Maximum Agent Level
Minimum Agent Level Stock number
Systems with this mark agree with the European Council Directive (93/42/EEC) for Medical Devices when they are used as specified in their User’s Reference manuals. The xxxx is the certification number of the Notified Body used by Datex-Ohmeda's Quality Systems.
w
NN
NN
Caution or Warning
Direction of flow
Serial number
1-2
1175-0013-000

2 Description

What is a Tec 7 Vaporizer?

The Tec 7 Vaporizer is designed for use in continuous flow techniques of inhalation anesthesia. Each vaporizer is agent specific and is clearly labeled with the anesthetic agent that it is designed for.
The vaporizer is temperature, flow and pressure compensated so that its output remains relatively constant despite cooling due to evaporation, variations in inlet flow and fluctuating pressures as described in Section 7, Effects of Variables.
The vaporizer is designed to be used on Selectatec vaporizer can be installed on other Selectatec Manifolds but the interlock system is designed to function on Selectatec Series Mounted Manifolds only. Mounting a Tec 7 Vaporizer on a Selectatec 7 Compatibility Block is not recommended.
®
Series Mounted Manifolds. The
1175-0013-000
2-1
Tec 7 Vaporizer
AB80008
Figure 2-1 • Tec 7 Vaporizer
w WARNING Improper use may result in patient injury.
This manual and its associated documentation must be studied before any attempt is made to install, operate or clean any part of the Tec 7 Vaporizer.
The performance of the anesthesia machine and vaporizer can be degraded if the machine and vaporizer are mis-matched.
Only operate the vaporizer with dry medical gases.
If a vaporizer containing agent in the sump has been inverted, connect it to a gas scavenging system, set the dial to 5% and purge the vaporizer with the carrier gas at 5 liters/minute for 5 minutes.
ww
ww
CAUTION The vaporizer is intended to be operated in its upright position.
Turn the vaporizer to when it is not in use.
2-2
1175-0013-000

Components

2 Description

Control dial

Safety interlocks

w WARNING Earlier versions of the Selectatec Series Mounted Manifold that provide
A single control dial with a concentration scale calibrated in percentage of anesthetic agent vapor per total volume (% v/v) sets the desired concentration of the anesthetic agent.
A dial release in the dial assembly helps prevent accidental displacement of the control dial from the position. To select an ON setting, squeeze the dial release and simultaneously rotate the dial counter-clockwise.
The dial and dial release are designed to enable an ON setting to be selected using only one hand.
The vaporizer incorporates an interlock mechanism. This mechanism also interfaces with the Selectatec
The vaporizer must be locked onto the manifold before it can be turned ON.
Only one vaporizer at a time can be turned ON when two or more vaporizers are fitted on a Selectatec
The gas flow enters only the vaporizer that is turned ON.
Any unwanted anesthetic trace vapor is minimized after a vaporizer is turned to
.
®
Series Mounted Manifold to help satisfy the following criteria:
®
Series Mounted Manifold.
mounting positions for three vaporizers require that if only two vaporizers are fitted, then the center position must be occupied. If the center position is not occupied, the interlock that helps ensure that only one vaporizer at a time can be turned ON is ineffective.
Vaporizer identification
label
1175-0013-000
Later versions of the Selectatec Series Mounted Manifold that provide mounting positions for three vaporizers incorporate an additional interlock that helps ensure that only one vaporizer at a time can be turned ON even if the center position is not occupied.
A vaporizer identification label is affixed to the back panel of the vaporizer as illustrated on Fig. 2-2.
An anesthesia system fitted with a vaporizer identification unit uses this label to identify the vaporizer type.
2-3
Tec 7 Vaporizer
w WARNING Do not affix any additional labels or markings to the back panel. They
may adversely affect the operation of the vaporizer identification unit.
1
H
A
L
O
T
H
A
N
E
1. Vaporizer identification label
Figure 2-2 • Vaporizer identification label
AB80011
2-4
1175-0013-000

3 Setup and Mounting Procedure

Vaporizer mounting procedure

The vaporizer is designed to be used on Selectatec Series Mounted Manifolds. The vaporizer can be installed on other Selectatec Manifolds but the interlock system is designed to function on Selectatec Series Mounted Manifolds only.
Mounting a Tec 7 Vaporizer on a Selectatec 7 Compatibility Block is not recommended.
ww
ww
WARNING Do not lift or support the vaporizer by holding the control dial. Handle the
vaporizer with care at all times.
Before mounting a vaporizer onto the Selectatec Series manifold, ensure that each manifold port valve O-ring is intact and that there is no foreign matter around the mating surfaces. A damaged O-ring and/or foreign matter around the mating surfaces can cause leaks.
Earlier versions of the Selectatec Series Mounted Manifold that provide mounting positions for three vaporizers require that if only two vaporizers are fitted, then the center position must be occupied. If the center position is not occupied, the interlock that helps ensure that only one vaporizer at a time can be turned ON is ineffective.
Do not use a vaporizer if the liquid level decreases below the minimum level.
1175-0013-000
Before using a vaporizer allow it to attain the ambient temperature of the location in which it has to be used.
3-1
Tec 7 Vaporizer

Mounting the vaporizer

Step 1
Set the dial to .
Step 2
Unlock the locking lever.
Turn the lever counter-clockwise.
Make sure the lever releases.
Step 3
Prepare the manifold.
Remove any plugs fitted to the vaporizer interlock block ports.
Verify that each manifold port valve O-ring is intact. If necessary, remove the existing O-rings and fit one new O-ring to each port valve, as described in the relevant anesthesia system User’s Reference Manual. Replacement O-rings are supplied with each vaporizer.
n
e
b
Å
Figure 3-1 • Setting the concentration dial
Figure 3-2 • Unlocking the locking lever
AB80001
AB80002
1
2
AA13052
3-2
1. Vaporizer Interlock Block Port - ensure plugs removed
2. Replace Manifold Port Valve O-ring, if necessary
Figure 3-3 • Readying the manifold
1175-0013-000
Step 4
Install the vaporizer onto the manifold.
Hold the main body of the vaporizer in an upright position with both hands
Lower the vaporizer onto the manifold, ensuring that the vaporizer interlock block ports engage correctly with the manifold port valves.
.
ww
ww
CAUTION Push the locking lever all the way down before turning it. The mechanism
3 Setup and Mounting Procedure
AB80018
Figure 3-4 • Installing the vaporizer
can be damaged if an attempt is made to turn the lever before it is pushed all the way down.
Step 5
Lock the vaporizer onto the manifold.
Push the locking lever all the way down.
Turn it clockwise to the locked position to lock the vaporizer onto the manifold.
Step 6
Ensure that the vaporizer is correctly mounted (see instructions on the next page).
1
1
AB80004
1. Locking lever
Figure 3-5 • Locking the vaporizer onto a manifold
1175-0013-000
3-3
Tec 7 Vaporizer

Checking the vaporizer for correct mounting

ww
ww
WARNING To help ensure correct operation, do not use a vaporizer that is either
visibly out of line on the manifold or that can be lifted off the manifold when the locking lever is in the locked position.
If more than one vaporizer is fitted, visually check to make sure that the tops of the vaporizers are level. If the vaporizer is visibly out of line, perform steps 2 and 3 as described in Removing the vaporizer from a manifold and remount it correctly.
When the vaporizer appears to be level and the locking lever is in the locked position, attempt to lift the vaporizer straight up from the manifold. If the vaporizer can be lifted off the manifold, it is not correctly mounted. Remount the vaporizer (see Vaporizer
mounting procedure ).
Verify that the interlock rods are in alignment by making sure that only one vaporizer at a time can be turned ON.
Check the anesthesia system for leaks in accordance with the relevant User’s Reference Manual with the vaporizer dial turned to 0% and then repeat the check with the vaporizer dial turned to .

Removing the vaporizer from a manifold

Step 1
Set the dial to .
If the dial is not completely turned to
OFF
the position the vaporizer cannot be released from the manifold.
OFF
Figure 3-6 • Setting the dial
n
e
b
Å
AB80001
3-4
1175-0013-000
Step 2
Unlock the locking lever.
Turn the locking lever counter­clockwise.
Release the locking lever and check that the locking lever springs up to the unlocked position to release the vaporizer from the manifold.
3 Setup and Mounting Procedure
Locked
1
2
Unlocked
1. Locking lever
2. Dial
Figure 3-7 • Unlocking the locking lever
1
2
AB80003
Step 3
Carefully lift the vaporizer up from the manifold.
AB80019
Figure 3-8 • Lifting the vaporizer
1175-0013-000
3-5
Loading...
+ 33 hidden pages