Datex-Ohmeda Inc.
3030 Ohmeda Drive
53707-7550 MADISON, WIS
USA
Tel. +1-608-221 1551, Fax +1-608-222 9147
www.us.datex-ohmeda.com
All specifications are subject to change without notice.
Document No. 800 1005-1
June 2001
Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.
Table of contents
TABLE OF CONTENTS
S/5 Airway modules and S/5 Gas Interface Board, B-GAS
TABLE OF CONTENTSi
TABLE OF FIGURESiii
INTRODUCTION1
1Specifications2
1.1 General specifications..............................................................................................................................2
2.2 Main components.....................................................................................................................................9
2.2.1 Gas sampling system......................................................................................................................10
2.2.8 Gas mother boar d...........................................................................................................................24
2.2.9 Gas interface board........................................................................................................................26
2.3 Connectors and signals...........................................................................................................................26
2.3.1 Module bus connector....................................................................................................................26
, N2O and Agent measurement....................................................................................................5
2.3.2 Gas mother board connectors.........................................................................................................27
3Service Procedures31
3.1 General service information....................................................................................................................31
3.2 Service check.........................................................................................................................................32
5.1 Gas mother board...................................................................................................................................56
5.2 ACX service menu...................................................................................................................................59
5.3 PVX service menu...................................................................................................................................61
5.3.2 Temp & Hum service menu.............................................................................................................63
5.4 ASX service menu...................................................................................................................................64
6Spare parts65
6.1 Spare parts list.......................................................................................................................................65
The S/5 Airway Modules, G-AO, G-AiO, G-AOV and G-AiOV are designed for use with the S/5
Anesthesia Monitor and provide airway and respiratory parameters. Later in this manual modules
can be called w/o system name S/5.
This Technical Reference Manual Slot provides information for the maintenance and service of the
airway modules. Please see also related
system e.g. related documentation, conventions used, symbols on equipment, safety precautions,
system description, system installation, interfacing, functional check and planned maintenance.
Letters in the name stand for:
G = Side mountable gas module
O = CO
V = Patient Spirometry
A = Anesthetic agents
i = Agent identification
, Patient O2, and N2O
2
Technical Reference Manual
S/5 Airway modules
for information related to
Table 1Options of Parameter Modules
G-AO
G-AiO
G-AOV
G-AiOV
NOTE: The Airway Modules and Compact Airway Modules cannot be used simultaneously in the
same monitor.
Gas Interface Board
Gas Interface Board, B-GAS is used for connecting the airway module to the central unit. The
connection can also be made through the Interface Board, B-INT.
CO
N2OPatient O
2
••••
•••••
•••••
••••••
AgentsAgent idSpirometry
2
Document No. 800 1005-1
1
Datex-Ohmeda S/5 Anesthesia Monitor
1 SPECIFICATIONS
1.1 General specifications
Module size, W × D × H135 × 410 × 135 mm/5.3 × 15.0 × 5.3 in
Module weight6 kg/13 lbs.
Automatic compensation for pressure, CO
Warm-up time 3 min for operation, 30 min for full specifications.
Auto-zeroing is performed at start-up, after 5 min + 5 min + 5 min + 15 min + 15 min + 15 min,
and after that every 60 min at regular intervals.
1.2.1 CO
2
Measurement range0 to 10 %, (0 to 10 kPa), (0 to 76 mmHg)
Extended range10 to 15 %, (10 to 15 kPa), (76 to 114 mmHg) (unspecified)
concentration is below 0.1 %, 0.0 % is displayed.
If CO
2
1.2.2 Respiration rate
Breath detection1 % change in CO2 level
Measurement range4 to 60 breaths/min
1.2.3 O
2
Measurement range0 to 100 % O
1.2.4 N2O
O, and CO2-O2 collision broadening effect.
2-N2
2
Measurement range0 to 100 % N2O
1.2.5 Hal, Iso, Enf
Measurement range0 to 5 %
Extended range5 to 15 % (unspecified)
1.2.6 Sev
Measurement range0 to 8 %
Extended range8 to 15 % (unspecified)
2
Document No. 800 1005-1
1.2.7 Des
Measurement range0 to 18 %
Extended range18 to 30 % (unspecified)
Resolutiontwo decimals when the AA concentration below 1.0 %
If AA concentration is below 0.10 %, 0.00 % is displayed.
1.2.8 Agent identification
Identified agentsHAL, ENF, ISO, SEV, DES
Identification time30 seconds (typical value with pure agents)
Identification threshold0.15 vol% (typical)
Mixture warning when minor component concentration > 0.3 vol% and > 15 %of total agent
concentration
1.2.9 Patient Spirometry
Values are valid when:
Respiratory rate adult 4...30pedi 4...50 breaths/min
I:E ratio1:3 - 1:0.5
Inner diameter of ET tube is ≥ 5.5 mm (adult) or 3 to 6 mm (pediatric).
S/5 Airway modules
1.2.10 Airway Pressure (Paw)
Accuracy±1.5 cmH2O
Resolution1 cmH
Measuring range-20 to +80 cmH
1.2.11 Tidal Volume (TV)
Accuracy±6 % or 30 ml (adult); ±6 % or 4 ml (ped)
Resolution1 ml
Measurement range150 to 2000 ml (adult)
1.2.12 Minute Volume (MV)
Resolution0.1 l/min
Measurement range2 to 15 l/min (adult)
1.2.13 Airway flow
Measurement range1.5 to 100 l/min for both directions (adult)
O
2
O
2
15 to 300 ml (ped)
0.5 to 5 l/min (ped)
0.25 to 25 l/min for both directions (ped)
Document No. 800 1005-1
3
Datex-Ohmeda S/5 Anesthesia Monitor
1.3 Technical specification
1.3.1 CO
1.3.2 O
2
1.3.3 N2O
2
Measurement rise time<360 ms (from 10 to 90 %)
Gain stability≤0.2 %CO
≤0.4 %CO
Gain temperature drift≤0.2 %CO
≤0.4 %CO
Nonlinearity error≤0.2 %CO
≤0.4 %CO
/24 h (0 to 8 %)
2
/24 h (8 to 10 %)
2
/10 °C (0 to 8 %)
2
/10 °C (8 to 10 %)
2
(0 to 8 %)
2
(8 to 10 %)
2
Measurement rise time<480 ms (from 10 to 90 %)
Gain drift≤2 % O
Gain temperature drift≤3 % O
Nonlinearity error≤2 % O
/24 h
2
/10 °C
2
2
Measurement rise time<360 ms (from 10 to 90 %)
Gain drift≤2 % N
Gain temperature drift≤3 % N
Nonlinearity error≤2 % N
O/24 h
2
O/10 °C
2
O
2
1.3.4 AA
Measurement rise time<520 ms (from 10 to 90 %)
Gain drift≤0.4 % AA/24 h
Gain temperature drift≤0.4 % AA/10 °C
Nonlinearity error≤0.2 % AA
Protection against electrical shock
Type BF
4
Document No. 800 1005-1
2 FUNCTIONAL DESCRIPTION
2.1 Measurement principle
2.1.1 CO2, N2O and Agent measurement
The CO2, N2O, and anesthetic agent gas measurements are based on absorption of infrared light as
it passes through the gas sample in measuring chamber in the photometer. The light absorption is
measured at three wavelengths using an infrared detector. One of the wavelengths is that of the
absorption peak at 4.3 micrometers, the second is that of the N2O absorption peak at 3.9
CO
2
micrometers, and the third is that of the anesthetic agent absorption peak at 3.3 micrometers. The
signal processing electronics receive the signals from the IR detector and demodulate it to get DC
components out of these signals which correspond to the content of each gas in the sample.
S/5 Airway modules
Figure 1CO2/N2O/AA gas absorption spectra
5
Document No. 800 1005-1
Datex-Ohmeda S/5 Anesthesia Monitor
2.1.2 O2 measurement
The differential oxygen measuring unit uses the paramagnetic principle in a pneumatic bridge
configuration. The signal picked up with a differential pressure transducer is generated in a
measuring cell with a strong magnetic field that is switched on and off at a frequency of 110 Hz.
The output signal is a DC voltage proportional to the O
gases to be measured.
Microphone
2
Electromagnet
concentration difference between the two
Mixture
out
Switched
magnetic
field
Sample in
Reference in
Figure 2O2 measurement principle
6
Document No. 800 1005-1
2.1.3 Agent identification
The anesthetic agent identification bench identifies Halothane, Enflurane, Isoflurane, Desflurane
and Sevoflurane.
The bench measures the spectrum of the gas between 3.24 µm and 3.39 µm. Because the
spectrum of each of the anaesthetic agents is different it is possible to identify them.
The bench consists of an infrared source, a measuring chamber, a rotating filter and a detector. The
peak wavelength of the narrow bandpass filter changes when the angle between the light path and
the filter is changed. When the filter rotates the required spectrum is scanned through. The agent or
a mixture of agents is identified by comparing the measured spectrum with stored reference
spectra.
In anesthesia, CMV (Controlled Mechanical Ventilation) is the mostly used ventilation mode. In this
mode, mechanical breaths are delivered to the patient by a ventilator with a proper tidal volume
(TV), respiration rate (RR), and inspiration/expiration ratio in time (I:E) determined by the settings of
the ventilator.
Delivery of life support gases is based on pressure. However, without knowing volume measured of
exhalation, one cannot be sure that a breath occurred. The ultimate goal of ventilation is to use the
least amount of pressure to generate the most appropriate volume for each breath.
The Patient Spirometry monitors ventilation in anesthesia. Both patient breathing circuit and the
function of the ventilator are monitored. The following parameters are displayed:
Expiratory and inspiratory tidal volume (TV) in ml.
Expiratory and inspiratory minute volume (MV) in l/min.
Expiratory volume in first second (V1.0) in per cent for adults and in 0.5 seconds for children.
Inspiration/expiration ratio in time (I:E)
Airway pressures: Peak pressure (P
pressure (PEEP), Real time airway pressure waveform (P
Flow: Real time flow waveform (V')
Compliance (C)
Pressure volume loop
Flow volume loop
), End inspiratory pressure (P
peak
)
aw
), Positive end expiratory
plat
Airway pressure
Airway flow
PEEP, P
peak
, and P
are measured by pressure transducer on the PVX board. Atmospheric pressure
plat
is used as a reference in measurement. The pressure measurement is made from the airway part
that is closest to the patient between patient circuit and intubation tube.
The measurement is based on measuring the kinetic gas pressure and is performed using Pitot
effect. Pressure transducer is used to measure the Pitot pressure. The obtained pressure signal is
linearized and corrected according to the density of the gas. Speed of the flow is calculated from
these pressure values and TV value is then integrated. MV value is further calculated a nd averaged
using TV and RR (respiratory rate) values.
8
Document No. 800 1005-1
Patient Spirometry sensor, D-lite
Patient Spirometry is measured with a specific sensor, D-lite or Pedi-lite.
D-lite and Pedi-lite sensors are designed to measure kinetic pressure by t wo-sided Pitot tube. The
pressure reduction caused by measuring cross is taken into account, too, especially in small flows.
Velocity is calculated from pressure difference according to Bernoulli's equation. Flow is then
Finally the volume information is obtained by integrating the flow signal.
2.2 Main components
The airway modules consist of ACX-200 and OM-101 gas measuring units, ASX-200 agent
identification unit (G-AiO/AiOV), PVX board (G-OV/AiOV/AOV), gas sampling system, ACX
measuring board and gas mother board.
Oxygen
Measuing
Unit
Serial I/O
X4
Supply
voltages
+5 V, GND
X1
Serial I/O
2
)
3
)
O2
signal
ACX
Board
2
Signal out
Supply
voltages,
Control
ACX
Supply
serial
voltages
I/O
Gas Module CPU
O)
ACX
Bench
X1
X2
Valve
controls
ACX, ASX,
PVX serial
I/O
+5 Vref
X8
+5 VL
GND
GNDD
Lamp
Supply
Voltage
Lamp
Driver
Lamp
control
Pump control
Mod key in
X9
Lamp
control
ASX
Unit
Supply
voltages
ASX
serial
i/O
Valve controls
+15 VD, GNDD
Pump control
+15 VD, GNDD
PVX
Board
X5
Supply
voltages
PVX
serial
i/O
Valve
Driver
Pump
Driver
X1
X3
Gas pressure
and flow in
X10
+24/+32 VD
X12
+15 VD
X7
+15 VD
Fan
Valves
Pump
+5 V
+5 VL
+15 VD
+/-15 V
+24 /+32 VD
GND
GNDD
+15 V
Regulator
Figure 4Airway module block diagram
Gas Mother Board
GND
Mod key in
X11
Module
Key
Document No. 800 1005-1
9
Datex-Ohmeda S/5 Anesthesia Monitor
2.2.1 Gas sampling system
The gas sampling system samples the measured air to the module, and removes water and
impurities from it. A sampling line is connected to the water trap on the front panel. The pump
draws gas through the sampling line to gas measuring units. After the measurements, the gas is
exhausted from sample gas out connector on the rear panel of the module.
Water trap, D-fend
The gas sample enters the monitor through the water trap, where it is divided into two flows, main
flow and side flow (see Gas sampling system block diagram). The main flow goes into the
measuring system through a hydrophobic filter.
The side flow creates a slight sub-atmospheric pressure within the water trap container. This
facilitates gathering th e fluid removed by the hydrophobic filter.
Sampling line
The sampling line is an integral part of the total sampling system. The resistance established by the
sampling line is significant when the software determines the occlusion a nd air-leak alarm limits
during the turn-on sequence.
The Nafion
Zero valve
The small inner diameter causes fluids such as blood or mucus not to propagate within the tube, so
that when the line is clogged, it is replaced.
TM
tube
1)
A nafion tube (tubes A or B, and C: see figure 5) is used to balance the sample gas humidity with
that of ambient air. The tube will prevent errors caused by the effect of water vapor on gas partial
pressure when humid gases are measured after calibration with dry gases. It is inserted between
the water trap and the zero valve (G-AiO/A iOV) or between the zero valve and ACX-200 measuring
unit (G-O/OV/AO/AOV). The tube is also inserted between the CO
absorber and the zero valve.
2
The main flow passes through a solenoid valve before proceeding to the ACX-200 measuring unit.
This valve is activated to establish the zero points for the A CX-200 and O
measuring units at start-
2
up, at 5 minutes, and after that at regular intervals. After 1-hour monitoring, the auto-zeroing is
performed once an hour. When the valve is activated, room air is drawn through the CO
into the internal system and the gas sensors.
absorber
2
1)
10
Document No. 800 1005-1
Nafion is a trademark of Du Pont
Gas measuring units, ACX-200 and O2 unit
After the zero valve, the gas passes through the ACX-200 and O2 measuring units. In the ACX-200
measuring unit, infrared light is passed through chambers containing the main flow gas
(measurement) and a chamber containing reference gas. The measurement is made by
determining the ratio between the two light intensities.
The oxygen sensor has two inputs. One input accepts the main flow and the other draws in room air
for reference. The sensor uses a differential pressure transducer to compare the pressure gradient
produced when both gases are exposed to an oscillating magnetic field. Both gas flows exit from a
single port.
In i model, the ASX agent identification unit is installed in parallel with the oxygen sensor . The task
of the ASX unit is to identify anesthesia agents by infrared light method used also in the ACX-200
unit.
Pressure valve
The pressure valve is used to meas ure the pressure gradient between the O2 measurement flow
and the O
filter.
reference flow. This pressure gradient reflects the condition of the D-fend water trap
2
S/5 Airway modules
Normally the pressure gradient between the O
approximately +8 mmHg. If the software detects the gradient to be between 0 and -5 mmHg, the
pressure valve will initiate pressure measurement of the reference flo w. If the gradient is greater
than -5 mmHg, the software triggers the message ’Replace Trap’.
Flow cassettes
The internal flow rates are set using flow cassettes. These cassettes are used to set the side flow
rate and the O
reference flow rate, the flow rates through the measuring units and the total flow
2
rate of the sampling system.
Sampling pump and damping chamber
The sampling pump is a vibrating membrane pump driven by a 50 Hz/12 V/0.4 A square wave
current.
The damping chamber is used to even out the pulsating flow and silence the exhaust flow.
measurement flow and the reference flow is
2
11
Document No. 800 1005-1
Datex-Ohmeda S/5 Anesthesia Monitor
Figure 5Gas sampling system block diagram
In G-AO, -AOV models, tube A is Teflon, B and C Nafion. In G-AiO, -AiOV models, tubes A and C are
Nafion, B is Teflon.
See new tubing since autumn 1998 in figure 7.
12
Document No. 800 1005-1
S/5 Airway modules
A
Ref IN
Room AIR
Sample
Gas IN
Figure 6Gas sampling system layout
See new sampling system layout since autumn 1998 in figure 8.
Filter
CO
2
absorber
Pressure valve
Pressure
transducer
C
ASX-200
in G-AiO/AiOV
Sample pump
Zero valve
main flow
D-fend
B
ACX-200
O2 sensor
Damping
chamber
side flow
Constriction casse tte
Figure 7Gas sampling system block diagram
In G-AO, -AOV models, tube A is Teflon, B and C Nafion. In G-AiO, -AiOV models, tubes A and C are
Nafion, B is Teflon. Figure 7 is valid for modules manufactured since autumn 1998.
Sample
Gas OUT
13
Document No. 800 1005-1
Datex-Ohmeda S/5 Anesthesia Monitor
Figure 8Gas sampling system layout
Figure 8 is valid for modules manufactured since autumn 1998.
NOTE: The number on the cassette represents relative flow when a specific pressure is applied.
Therefore 50/26.0 presents the least resistance and 50/1.1 the most.
Document No. 800 1005-1
15
Datex-Ohmeda S/5 Anesthesia Monitor
2.2.2 ACX-200 measuring unit
The ACX photometer is of dual path type. The infrared light beam passes through a measuring
chamber containing the gas to be analyzed, and a reference chamber, which is free of CO
and AA. The measurement is made by determining the ratio between the two light intensities.
, N2O,
2
Figure 9ACX photometer (ACX-200 measuring unit)
A filter wheel is used to control the light from an incandescent lamp that passes through the
photometer. The filters are arranged so that the light is passed sequentially:
• first at the CO
• then through the measuring chamber
• finally it is blocked completely
The same sequence is repeated at the N
absorption wavelength through the reference chamber
2
O and anaesthetic agent gas absorption wavelengths.
2
After passing through the filters the light is reflected and focused by a mirror onto the infrared
detector. This detector measures the three light levels for each gas described above.
There is an optical sensor incorporated in the photometer which detects light from a reflective
surface on the filter wheel once every revolution. The pu lses from this sensor are us ed to
synchronize the electronics to the signal from the infrared detector. A stabilizing diode measures
the temperature, which is needed to compensate for thermal drifts. The infrar ed detector, the
optical sensor and the stabilizing diode are mounted on the preamplifier board.
16
Document No. 800 1005-1
S/5 Airway modules
Figure 10CO2/N2O/AA measurement block diagram
2.2.3 OM measuring unit
The oxygen measurement is based on the paramagnetic susceptibility, which is a unique property
of oxygen among all gases generally present in a breathing gas mixture. The gas to be measured
and the reference gas, which usually is room air, are conducted into a gap in an electromagnet with
a strong magnetic field switched on and off at a frequency of approximately 110 Hz.
An alternating differential pressure is generated between the sample and reference inputs due to
forces acting to the oxygen molecules in a magnetic field gradient.
The pressure is measured with a sensitive differential transducer, rectified with a synchronous
detector and amplified to produce a DC voltage proportional to the oxygen partial pressure
difference of the two gases.
17
Document No. 800 1005-1
Datex-Ohmeda S/5 Anesthesia Monitor
2.2.4 ACX measuring board
The measuring electronics can be divided into a few functional blocks, which are described below
(see the block diagram in figure 11).
The ACX Measuring board controls gas measurements. It converts the photometer signal into digital
data, calculates results and transmits it to Gas mother board. The board contains, in addition to
the 80C51FA processor, EPROM, RAM, and EEPROM, several analog and digital I/O functions.
Internal and external bus
The processor has access to the Measuring board peripherals (memory, A/D converter, D/A
converters, etc) via an internal bus. For communication between the Gas mother board and the
Measuring board, there is an external bus in connector X1.
Memory
Memory components include 64k × 8 bit EPROM program memory, 32k × 8 bit low current CMOS
RAM, and EEPROM for permanent calibration values and setup memory.
18
Document No. 800 1005-1
S/5 Airway modules
Figure 11ACX measuring board block diagram
19
Document No. 800 1005-1
Loading...
+ 58 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.