Datex-Ohmeda S-5 Service Manual M-COP

S/5
Datex-Ohmeda Cardiac Output Modules
TM
Cardiac Output and SvO2 Module, M-COPSv (Rev. 01)
TM
S/5
Cardiac Output Module, M-COP (Rev. 03)

Technical Reference Manual Slot

Datex-Ohmeda Inc. 3030 Ohmeda Drive 53707-7550 MADISON, WIS USA Tel. +1-608 -221 1551, Fax +1-608-222 9147 www.us.datex-ohmeda.com

Document No. 800 1012-1

June 2001

Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.

Table of contents

TABLE OF CONTENTS
Cardiac Output Modules, M-COP and M-COPSv
TABLE OF CONTENTS i
TABLE OF FIGURES ii
Introduction 1
1 Specifications 2
1.1 General specifications ..............................................................................................................................2
1.2 Typical performance .................................................................................................................................2
1.2.1 C.O..................................................................................................................................................2
1.2.2 SvO
1.2.3 InvBP...............................................................................................................................................2
1.3 Technical specifications............................................................................................................................3
2 Functional Description 4
2.1 Measurement principle .............................................................................................................................4
2.1.1 Cardiac output and REF....................................................................................................................4
2.1.2 SvO
2.1.3 Invasive blood pressure measurement ..............................................................................................6
2.2 Main components.....................................................................................................................................6
2.2.1 COP board........................................................................................................................................7
2.2.2 COPSv board..................................................................................................................................10
2.3 Connectors and signals...........................................................................................................................11
2.3.1 Module bus connector....................................................................................................................11
2.3.2 Front panel connectors...................................................................................................................12
................................................................................................................................................2
2
measurement ..........................................................................................................................5
2
3 Service Procedures 14
3.1 General service information.....................................................................................................................14
3.2 Service check .........................................................................................................................................15
3.2.1 Recommended tools ......................................................................................................................15
3.3 Disassembly and reassembly..................................................................................................................20
3.4 Adjustments and calibrations..................................................................................................................20
3.4.1 Cardiac output calibration...............................................................................................................20
3.4.2 Invasive pressure calibration...........................................................................................................20
4 Troubleshooting 22
4.1 Troubleshooting charts............................................................................................................................22
4.1.1 Cardiac Output...............................................................................................................................22
4.1.2 SvO
4.1.3 InvBP.............................................................................................................................................23
4.2 Troubleshooting flowchart .......................................................................................................................24
..............................................................................................................................................22
2
5Service Menu 25
5.1 COP Menu ..............................................................................................................................................26
5.1.1 COP calibration menu.....................................................................................................................28
6 Spare Parts 29
6.1 Spare parts list .......................................................................................................................................29
Document No. 800 1012-1
i
Datex-Ohmeda S/5 monitors
6.1.1 Cardiac Output Module, M-COP Rev. 00.......................................................................................... 29
6.1.2 Cardiac Output Module, M-COP Rev. 01.......................................................................................... 29
6.1.3 Cardiac Output Module, M-COP Rev. 02.......................................................................................... 30
6.1.4 Cardiac Output Module, M-COP Rev. 03.......................................................................................... 30
6.1.5 Cardiac Output and SvO
6.1.6 Cardiac Output and SvO
Module, M-COPSv Rev. 00 .......................................................................30
2
Module, M-COPSv Rev. 01 .......................................................................31
2
6.1.7 Front panel stickers for AS/3 modules (square buttons)...................................................................31
6.1.8 Front panel stickers for S/5 modules (round buttons).......................................................................31
7 Earlier Revisions 32
APPENDIX A 33
Service check form A-1

TABLE OF FIGURES

Figure 1 Cardiac Output Module, M-COP, and Cardiac Output and SvO2 Module, M-COPSv ..................................1
Figure 2 Cardiac output measurement curve.......................................................................................................5
Figure 3 Front panels of Cardiac Output Modules, M-COP and M-COPSv.............................................................. 6
Figure 4 COPSv board block diagram, on COP board the SvO
section is excluded................................................ 8
2
Figure 5 Pressure transducer principle of operation.............................................................................................8
Figure 6 Serial communication and opto isolation............................................................................................... 9
Figure 7 Cardiac Output Module troubleshooting flowchart................................................................................24
Figure 8 Module box and Cardiac Output Module, M-COP .................................................................................29
Figure 9 Cardiac Output and SvO
Module, M-COPSv........................................................................................ 30
2
ii
Document No. 800 1012-1

INTRODUCTION

This section provides information for the maintenance and service of Cardiac Output Modules, M­COP and M-COPSv. Cardiac Output Modules, M-COP and M-COPSv are single width plug-in modules designed for use with the S/5 monitors. Later in this manual modules can be called w/o system name S/5.
Both modules provide
Cardiac output (C.O.)
Right ventricular ejection fraction (REF)
Invasive blood pressure (InvBP) measurement
Cardiac Output Modules, M-COP and M-COPSv
Additionally, the COPSv module provides venous oxygen saturation (SvO
) measurement.
2
NOTE: Do not use identical modules in the same monitor simultaneously.These modules are considered as identical and would cause communication errors if used in the same system.
Figure 1 Cardiac Output Module, M-COP, and Cardiac Output and SvO
COPSv
Document No. 8001012-1
Module, M-
2
1
Datex-Ohmeda S/5 monitors

1 SPECIFICATIONS

1.1 General specifications
Module size (W × D × H) 37 × 180 × 112 mm / 1.5 × 7.1 × 4.4 in Module weight 0.35 kg / 0.8 lbs
Power consumption, M-COP Approximately 3.5 W Power consumption, M-COPSv Approximately 5 W
1.2 Typical performance
1.2.1 C.O.
Measurement range 0.1...20 l/min Display resolution 0.01 l/min Repeatability ±2 % or ±0.02 l/min whichever is greater
(measured from electrically generated flow curves)
1.2.2 SvO
2
1.2.3 InvBP
Max. change in blood temp 2.99 °C
Injectate temp range (with Edward's compatible probes)
0...25.5 °C ±0.3 °C
25.5...27.0 °C ±0.5 °C
Blood temp range (with Edward's compatible catheters)
17.5...30.9 °C ±0.5°C
31.0...43.0 °C ±0.3°C
Protection against electric shock type CF defibrillation proof
REF
Repeatability ±2 % (Measuring range 10-60 %)
Accuracy ±2 % (Measuring range 30-95 %) Equal to standard deviation when using in-vivo calibration.
Measurement range -40...+320 mmHg Zero adjustment range ±150 mmHg Calibration range ±20 %
2
Document No. 800 1012-1
Scales Upper limit is adjustable between 10 and 300
Sweep speed 12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range -40...+320 mmHg Resolution ±1 mmHg
WAVEFORM DISPLAY
Range -30...+300 mmHg
PULSE RATE
Measurement range 30...250 bpm Resolution 1 bpm Accuracy ±5 % or ±5 bpm whichever is greater Respiration artifact rejection
1.3 Technical specifications
Cardiac Output Modules, M-COP and M-COPSv
mmHg in steps of 10. Lower limit is 10 % of selected upper limit below zero.
The digital display is averaged over 5 seconds and updated at 5 second intervals.
Accuracy ±5 % or ±2 mmHg, whichever is greater
Transducer and 5 µV/V/mmHg, 5 VDC input sensitivity 20 mA max current
Nonlinearity <1 %, 0 to 200 mmHg
<2 %, -40 to 0 and 200 to 320 mmHg
Filter 0...22 Hz (-3 dB), adjustable upper limit 4...22 Hz
Zero set accuracy ±1 mmHg Calibration resolution ±1 mmHg Zero time < 15 sec
Protection against electric shock type CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified accuracy, depending on the transducer/probe used. Please check the transducer/probe specification.
Document No. 8001012-1
3
Datex-Ohmeda S/5 monitors

2 FUNCTIONAL DESCRIPTION

2.1 Measurement principle
2.1.1 Cardiac output and REF
Cardiac output measurement is made using the principle of thermodilution. During measurement the catheter lies in the heart, with an injection port in the right atrium (RA) and a thermistor, which is to monitor blood temperature, in the pulmonary artery (PA). A small, known amount of thermal indicator is injected into the RA and is mixed with the blood on its way to the PA. The catheter thermistor measures the decrease in blood temperature as the blood flows past the thermistor in the PA.
The information is stored in the module and the cardiac output is calculated from the area beneath the time-temperature Cardiac Output Measurement Curve, as shown in figure 2.
The area under the time-temperature curve is inversely proportional to the flow rate which corresponds to cardiac output.
The cardiac output is calculated from the equation:
C.O.= (1.08 C
60 Vi(TB-Ti))/(TBdt + C)
T
where: C.O. = cardiac output in liters/minute
1.08 = factor comparing the density and specific heat of 5% dextrose solution in water to those of blood.
= correction factor for the injectate temperature rise as it passes through the
C
T
catheter and its dead space
60 = seconds/minute
= injectate volume in liters
V
i
= baseline blood temperature (°C)
T
B
Ti = injectate temperature
dt = area under time-temperature curve between time o and x, where x is the
T
B
time when the curve has dropped to 30% of its peak value.
C = area beneath time-temperature curve between x and the end of the
curve.
4
Document No. 800 1012-1
Cardiac Output Modules, M-COP and M-COPSv
%
-T area A
100%
80%
start of meas.
end of meas.
area C
30%
0%
Temperature change in blood [
m.s.
80% 30%
Time
Figure 2 Cardiac output measurement curve
A = area derived by integration of the time-temperature curve
C = area beneath the time-temperature curve between t based on an exponential fit to the curve between t
% and end of the curve. Computation
30
% of the peak and t30%.
80
REF (right ventricular ejection fraction) measurement is a part of the time-temperature (thermodilution) cardiac output measurement. Ejection fraction is determined using an exponential technique by synchronizing sensed R-waves with points of temperature changes on the time­temperature curve. Once ejection fraction, cardiac output, and heart rate are known, right ventricular volumes may be calculated. The measurement requires a Baxter-Edwards fast response thermistor catheter and an ECG module to synchronize R-wave detection to the time-temperature curves.
t
2.1.2 SvO2 measurement
The COPSv module measures SvO2 when coupled with a Baxter-Edwards OM-2E optical module and a Swan-Ganz oximetry catheter. To measure SvO technique involving the use of light emitting diodes (LEDs) that produce red (660 nm) and infrared (810 nm) light. The light is transmitted to the blood through a single plastic optical fiber in the oximetry catheter and reflected back through a separate optical fiber to a photodetector in the optical module. The light is electrically transmitted to the COPSv module and analyzed to determine SvO
The oximetry portion of the system measures SvO changes in the red blood cells. When pulses of red and infrared light are transmitted through the oximetry catheter, the light is reflected from the red blood cells and transmitted back through the catheter to the optical module. The amount of light reflected at each wavelength depends primarily on the color of the blood and the number of red blood cells. Since the number of red blood cells in the blood affects the amount of reflected light, the differences are compensated for when the patient’s total hemoglobin value is entered. The optical module stores and transfers SvO calibration data. SvO2 values can be affected by the presence of methemeglobin or carboxyhemoglobin which imitate the absorption characteristics of HbO methemeglobin or carboxyhemoglobin could then cause a falsely elevated SvO dysfunctional hemoglobins are suspected, SvO
, the system utilizes a spectrophotometric
2
.
2
in the pulmonary artery by detecting color
2
2
. Large concentrations of
2
. In cases where
2
should be interpreted with caution.
2
5
Document No. 8001012-1
Datex-Ohmeda S/5 monitors
2.1.3 Invasive blood pressure measurement
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal saline all connected together by stopcocks, is attached to the catheter. The pressure transducer is placed at the same level with the heart, and electrically zeroed.
The pressure transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor interprets the voltage signal so that blood pressure data and blood pressure waveforms can be displayed.
2.2 Main components
The Cardiac Output Module, M-COP consist of a COP circuit board and two input boards - a CO input board and a P input board, attached to the front panel.
The Cardiac Output and SvO boards - a CO input board, a SvO
Module, M-COPSv, consist of a COPSv circuit board and three input
2
input board and a P input board, attached to the front panel.
2
The front panels are shown in figure 3.
M-COP
1
P
4
C.O.
2
3
4
5
M-COPSv
P4
SvO
2
C.O
.
6
Figure 3 Front panels of Cardiac Output Modules, M-COP and M-COPSv
6
Document No. 800 1012-1
(1) Key for pressure zeroing (Zero P4)
(2) Key for cardiac output measurement (Start C.O.)
(3) Connector for invasive blood pressure measurement
(4) Connector for C.O. self test
(5) Connector for C.O. measurement
(6) Connector for SvO
measurement
2
2.2.1 COP board
The COP board consists of the following functional sections.
Processor
Cardiac output measurement
Cardiac output self test
Invasive blood pressure measurement
Serial communication
Isolation
Power supply
Processor section
The microprocessor uses the Intel 80C196KC-16 CPU which includes three A/D converters and a UART. The microprocessor uses external memories, an 8-bit data bus, a 16 MHz oscillator, and a watchdog timer. The three A/D-converters within the CPU convert the analog input signals to digital. The internal UART communicates and transfers data between the module and the CPU board in the monitor.
Cardiac Output Modules, M-COP and M-COPSv
Cardiac output measurement section
The catheter and the probe contain an NTC resistor that reacts to temperature change.
The temperature dependent voltage across the NTC resistor is amplified and an offset value is added to it. The resultant signal is then regulated into a A/D converter.
Because the temperature measurements are calibrated digitally and non-linearity of catheter/probe is compensated by software, ambient temperature change after calibration is the only factor that may influence the measurement.
Cardiac output self test
The Cardiac Output Module, M-COP contains a C.O. Self Test connector.
When the cable is connected to the C.O. Self Test connector, the microprocessor starts the test program automatically. First, the microprocessor measures 30 °C, then it activates test circuits and measures 37 °C and 41 °C.
If the values are not correct ‘Cable fault’ is displayed and there is a fault is either in the module or in the catheter connecting cable.
±5 V range by voltage slicing and sent to an
Document No. 8001012-1
7
Datex-Ohmeda S/5 monitors
Modul e data
RS485 DRIVER FOR DATA
OPTO
ISOLATION
NON-
VOLAT ILE
MEMOR Y
SvO2 AD OM-2 control
To opti cal module
Module bus data
SvO2
MEAS URI NG
UNI T
Modul e reset
RS485 DRIV ER FOR MODU LE RESET
Res et
Serial device
communicati on
Power f or communication
POWER RESET
PATIENT ISOLATION
MIC ROPROC ESSOR RAM internal 256 external 16K EPROM 48K RS COMMUNICATI ON
AD­CONVERTER 8 chn 12 bit
C.O.
control
C.O.
AD
C.O.
MEASURI NG
UNIT
Press AD
P4 INV PRESSURE MEASURING UNIT
Power f or
module
POWER NON­ISOLATION SECTION
ISOLATION
TRANS-
FORMER
POWER
ISOLATION
SECTION
FRONT
PANEL
KEYS
Figure 4 COPSv board block diagram, on COP board the SvO2 section is excluded
Invasive blood pressure measurement section
Uin
Figure 5 Pressure transducer principle of operation
An isolated +5 V supply is connected to the input of the pressure transducer bridge circuit. From the bridge circuit output a differential voltage, which depends on blood pressure and input supply voltage, is calculated using the following formula:
Pressure transducer
Uout
Input Filter
Instrumentation amplifier
G
to AD converter
8
Document No. 800 1012-1
Loading...
+ 28 hidden pages