Datex-Ohmeda S-5 Service Manual Datex S5

Page 1
S/5
S/5 S/5
Datex-Ohmeda
Hemodynamic Modules
TM
PRESTN Module, M-PRESTN (Rev. 01)
TM
RESTN Module, M-RESTN (Rev. 01)
TM
PRETN Module, M-PRETN (Rev. 01)
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licenced practitioner.
Datex-Ohmeda, Inc.
P.O. Box 7550, Madison WI 53707-7550, USA Tel: 1-608-221-1551 Fax: 1-608-222-9147
www.us.datex-ohmeda.com
mailto:product.support.ussub@us.datex-ohmeda.com

All specifications are subject to change without notice.

Outside the USA, check local laws for any restriction that may apply.

Document No. 8005571-1

Datex-Ohmeda Division, Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel: +358 10 394 11 Fax: +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.

April 2004

Page 2
Page 3

Table of contents

TABLE OF CONTENTS
S/5 Hemodynamic Modules
TABLE OF CONTENTS i
TABLE OF FIGURES iii
Introduction 1
1 Specifications 3
1.1 General specifications.......................................................................................................................... 3
1.2 Typical performance ............................................................................................................................. 3
1.2.1 NIBP ........................................................................................................................................... 3
1.2.2 ECG.............................................................................................................................................3
1.2.3 Pulse oximetry .............................................................................................................................4
1.2.4 Temperature................................................................................................................................4
1.2.5 Invasive blood pressure................................................................................................................5
1.2.6 Respiration.................................................................................................................................. 5
1.3 Technical specifications........................................................................................................................ 6
1.3.1 NIBP ........................................................................................................................................... 6
1.3.2 ECG.............................................................................................................................................6
1.3.3 Pulse oximetry .............................................................................................................................6
1.3.4 Temperature................................................................................................................................7
1.3.5 Invasive blood pressure................................................................................................................7
1.3.6 Respiration.................................................................................................................................. 7
2 Functional Description 8
2.1 Measurement principle......................................................................................................................... 8
2.1.1 NIBP ........................................................................................................................................... 8
2.1.2 ECG.............................................................................................................................................8
2.1.3 Pulse oximetry .............................................................................................................................8
2.1.4 Temperature..............................................................................................................................10
2.1.5 Invasive blood pressure..............................................................................................................10
2.1.6 Respiration................................................................................................................................ 10
2.2 Main components ..............................................................................................................................11
2.2.1 M-PRESTN/-RESTN/-PRETN modules .........................................................................................11
2.2.2 NIBP board................................................................................................................................ 12
2.2.3 ECG board in 12-lead measurement...........................................................................................15
2.2.4 ECG filtering...............................................................................................................................17
2.2.5 STP board..................................................................................................................................18
2.3 Connectors and signals ......................................................................................................................23
2.3.1 Module bus connector (on the NIBP board) ................................................................................. 23
2.3.2 Front panel connectors...............................................................................................................25
2.3.3 Test points on boards.................................................................................................................26
3 Service Procedures 29
3.1 General service information ................................................................................................................29
3.2 Service check..................................................................................................................................... 29
3.2.1 Recommended tools.................................................................................................................. 29
3.2.2 Recommended parts..................................................................................................................29
3.3 Disassembly and reassembly..............................................................................................................39
3.3.1 M-PRESTN/-RESTN/-PRETN modules .........................................................................................39
3.4 Adjustments and calibrations ............................................................................................................. 40
Document No. 8005571-1
i
Page 4
Datex-Ohmeda S/5 monitors
3.4.1 NIBP calibrations....................................................................................................................... 40
3.4.2 Temperature calibration............................................................................................................. 42
3.4.3 Invasive pressure calibration...................................................................................................... 42
4 Troubleshooting 43
4.1 Troubleshooting charts ....................................................................................................................... 43
4.1.1 NIBP ......................................................................................................................................... 43
4.1.2 NIBP error code explanation....................................................................................................... 46
4.1.3 ECG .......................................................................................................................................... 47
4.1.4 Pulse oximetry (SpO2) ................................................................................................................ 47
4.1.5 Temperature.............................................................................................................................. 48
4.1.6 Invasive blood pressure ............................................................................................................. 49
4.1.7 Impedance respiration............................................................................................................... 50
4.2 Troubleshooting flowcharts................................................................................................................. 51
4.2.1 M-PRESTN module troubleshooting for NIBP parameter............................................................... 51
4.2.2 M-PRESTN module troubleshooting for parameters ESTPR........................................................... 52
5 Service Menu 53
5.1 NIBP service menu ............................................................................................................................. 54
5.1.2 NIBP demo menu....................................................................................................................... 55
5.1.3 NIBP calibration menu ............................................................................................................... 56
5.1.4 NIBP safety valve menu.............................................................................................................. 57
5.1.5 NIBP pulse valve menu .............................................................................................................. 58
5.1.6 NIBP buttons/leds menu............................................................................................................ 59
5.1.7 NIBP pneumatics menu ............................................................................................................. 60
5.1.8 NIBP watchdog menu................................................................................................................. 61
5.1.9 ECG service menu...................................................................................................................... 62
5.1.10 ECG setup menu.................................................................................................................... 64
5.2 STP service menu............................................................................................................................... 65
5.2.2 STP calibration menu................................................................................................................. 67
6 Spare Parts 68
6.1 Spare parts list................................................................................................................................... 68
6.1.1 M-PRESTN rev. 01, M-RESTN rev. 01, M-PRETN rev. 01 ............................................................... 68
7 Earlier Revisions 71
APPENDIX A 73
Service Check Form 1
ii
Document No. 8005571-1
Page 5
Table of contents

TABLE OF FIGURES

Figure 1 S/5 PRESTN Module, M-PRESTN........................................................................................................... 1
Figure 2 Absorption of infrared light in the finger probe parts layout and schematic diagram ................................. 9
Figure 3 Front panel of M-PRESTN....................................................................................................................11
Figure 4 NIBP board functional block diagram .................................................................................................. 12
Figure 5 NIBP pneumatics diagram ..................................................................................................................14
Figure 6 12-lead ECG measurement block diagram .......................................................................................... 15
Figure 7 STP board block diagram ....................................................................................................................18
Figure 8 Temperature measurement principle ...................................................................................................19
Figure 9 Pressure measurement principle.........................................................................................................20
Figure 10 Pulse oximetry measurement block diagram....................................................................................21
Figure 11 Serial communication of NIBP board............................................................................................... 22
Figure 12 Serial Communication and Isolation of STP board............................................................................22
Figure 13 Serial Communication and Isolation of ECG board ...........................................................................22
Figure 14 Module bus connector (X1) pin layout ............................................................................................. 23
Figure 15 ECG board connectors and test points............................................................................................. 26
Figure 16 NIBP board connectors and test points............................................................................................ 27
Figure 17 STP board connectors and test points ............................................................................................. 28
Figure 18 M-PRESTN module troubleshooting flowchart for NIBP Parameter..................................................... 51
Figure 19 M-PRESTN Module Troubleshooting Flowchart for Parameters ESTPR................................................52
Figure 20 Exploded view of M-PRESTN Module ...............................................................................................68
iii
Document No. 8005571-1
Page 6
Datex-Ohmeda S/5 monitors
iv
Document No. 8005571-1
Page 7

INTRODUCTION

This Technical Reference Manual Slot provides information for the maintenance and service of the hemodynamic modules S/5 M-PRESTN/-RESTN/-PRETN. The modules are double width modules designed for use with S/5 monitors. Later in this manual modules may be referred to w/o the system name S/5 for simplicity.
Please also refer to Technical Reference Manual of the S/5 monitor for information regarding system specific information e.g. related documentation, conventions used, symbols on equipment, safety precautions, system description, system installation, interfacing, functional check and planned maintenance.
The M-PRESTN/-RESTN/-PRETN modules provide general hemodynamic parameters.
Auto
On/Off
NIBP
ECG
Start
Cancel
T1
SpO
P1
Zero
P1
S/5 M-PRESTN Modules
NOTE: Do not use identical modules in the same
Zero
P2
2
monitor simultaneously. The following modules are considered identical: M-ESTP/-EST/-ETP, M-ESTPR/-ESTR/-ETPR, M-NESTPR/-NESTR/-NETPR, M-NE12STPR/-NE12STR/-NE12TPR M-PRESTN/M-RESTN/M-PRETN
T2
P2
Figure 1 S/5 PRESTN Module, M-PRESTN
Table 1 Options of S/5 hemodynamic modules
Parameter PRESTN RESTN PRETN
P
R
E
S
T
N
Two invasive blood pressures
Impedance respiration
ECG
Pulse oximetry
Two temperatures
NIBP
NOTE: 12-lead ECG measurement requires Display Controller, B-DISP or B-DISPX.
Intended purpose (Indications for use)
The Datex-Ohmeda PRESTN module (model family M-PRESTN) and accessories are indicated for monitoring of hemodynamic parameters of all hospital patients. The hemodynamic parameters of the module comprise ECG including ST-segment and arrhythmia, Impedance respiration, NIBP, Temperature,SpO2 (including monitoring during conditions of clinical patient motion),and invasive blood pressure.
1
Document No. 8005571-1
Page 8
Datex-Ohmeda S/5 monitors
Impedance respiration measurement is indicated for patients aged 3 and up. The NIBP measurement is indicated for patients who weigh 5kg (11 lb.) and up. This device is indicated for use by qualified medical personnel only.
Monitor software compatibility
Datex-Ohmeda PRESTN rev. 01 module is designed for use with Datex-Ohmeda monitors using software as follows: AM: L-ANE01(A) or later versions; CCM: S-00C01 rev. 10.5, S-00C02 rev. 10.5 or newer versions; CAM: S-00A05 rev. 10.9, S-00A06 rev. 10.9, L-00A07 rev. 10.9, L-00A08 rev. 10.9 or newer versions and CCCM: S-00C03 rev. 10.9, S-00C04 rev. 10.9 or newer versions.
2
Document No. 8005571-1
Page 9
1 SPECIFICATIONS
1.1 General specifications
Module size 75 × 180 × 112 mm W × D × H 3.0 × 7.1 × 4.4 in Module weight 0.7 kg / 1.5 lbs
Power consumption about 6 W Operation temperature 10 to 40 °C / 50 to 104 °F
1.2 Typical performance
1.2.1 NIBP
NOTE: Non-invasive blood pressure measurement is intended for patients weighing over 5 kg (11 lb.)
S/5 M-PRESTN Modules
1.2.2 ECG
Oscillometric measurement principle.
Measurement range adult 25 to 260 mmHg child 25 to 195 mmHg infant 15 to 145 mmHg
Pulse rate range accepted 30 to 250 bpm
Measurement interval STAT (continuous 5 min), 1, 2.5, 3, 5, 10, 15, 30 and 60 min (1 h),
2 and 4 h
Typical measuring time adult 23 s infant 20 s
Initial inflation pressure adult 170 ±10 mmHg child 150 ±10 mmHg infant 120 ±10 mmHg
Venous stasis adult 40 ±5 mmHg / 2 min child 40 ±5 mmHg / 2 min infant 30 ±5 mmHg / 1 min
Cuff widths please see User’s Guide
Lead selection, 12-lead ECG I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 Lead selection, other modules I, II, III, aVR, aVL, aVF, V Sweep speeds 12.5, 25, 50 mm/sec
DISPLAY FILTER
Diagnostic, 12-lead ECG 0.05 to 150 Hz Diagnostic, other modules 0.05 to 100 Hz
Monitoring 0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
3
Document No. 8005571-1
Page 10
Datex-Ohmeda S/5 monitors
ST filter 0.05 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.05 to 40 Hz (-3 dB, with 60 Hz reject filter)
HEART RATE FROM ECG
Range 30 to 250 bpm Accuracy ±5 bpm or ±5 %, whichever is greater Resolution 1 bpm Update interval 5 s Averaging time 10 s
ST LEVELS (in main software)
ST level range -9 to +9 mm (-0.9 to +0.9 mV) Resolution 0.1 mm (0.01 mV) Averaging calculated from 8 QRS
SYNCHRONIZATION
Direct ECG analog output of ECG, 1 V/1 mV Pacer 5 V and 0.5 to 2.5 ms pulse, < 30 ms after pacer peak Defibrillator 5 V and 10 ms pulse, < 35 ms after R-point synchronization
1.2.3 Pulse oximetry
Measurement range 0 to 100 % Calibration range 70 to 100 %
Accuracy
1 100 to 70 %, ±2 digits
±3 digits during clinical patient motion 69 to 0 %, unspecified
Display resolution 1 digit = 1 % of SpO
2
Display averaging time 20, 10 sec, beat-to-beat Pulse beep pitch varies with SpO
The monitor is calibrated against functional oxygen saturation SpO
2 level
func.
2
PULSE RATE FROM PLETH
Measurement range 30 to 250 bpm Accuracy 30 to 100, ±5 bpm, 100 to 250, ±5 % Resolution 1 bpm Display averaging 10 s
Adjustable pulse beep volume.
PLETH WAVEFORM
Scales 2, 5, 10, 20, 50 mod%, Auto
Start up scale is 20 mod% if AUTO is not selected to be the default setting.
1.2.4 Temperature
Measurement range 10 to 45 °C (50 to 113 °F) Measurement accuracy ±0.1 °C (25 to 45.0 °C)
1 Accuracy is based on deep hypoxia studies with volunteered subjects during motion and non-motion conditions over a wide range of arterial
blood oxygen saturations as compared to arterial blood CO-Oximetry.
4
Document No. 8005571-1
Page 11
±0.2 °C (10 to 24.9 °C) Display resolution 0.1 °C (0.1 °F) Temperature test automatic (every 10 min) Probe type compatible with YSI 400 series
Single use sensors ±0.2 °C (25 to 45.0 °C) ±0.3 °C (10 to 24.9 °C)
1.2.5 Invasive blood pressure
Measurement range -40 to 320 mmHg Measurement accuracy ±2 mmHg or ±5 % Zero adjustment range ±150 mmHg Calibration range ±20 % Scales upper limit is adjustable between 10 and 300 mmHg in steps of
10. Lower limit is 10 % of selected upper limit below zero.
Sweep speed 12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range -40 to 320 mmHg Resolution ±1 mmHg
S/5 M-PRESTN Modules
WAVEFORM DISPLAY
Range -30 to 300 mmHg
PULSE RATE FROM ARTERIAL PRESSURE
Measurement range 30 to 250 bpm Resolution 1 bpm Accuracy ±5 bpm or ±5 % whichever is greater
1.2.6 Respiration
NOTE: The respiration measurement is intended for patients over three years old
Measurement range 4 to 120 bpm Accuracy ±5 bpm or ±5 % Resolution 1 bpm Averaging time 30 s Update interval 10 s
RESPIRATION WAVEFORM
Sweep Speeds 6.25 mm/s and 0.625 mm/s
Document No. 8005571-1
5
Page 12
Datex-Ohmeda S/5 monitors
1.3 Technical specifications
1.3.1 NIBP
Deflation rate, PR dep. 3 to 8 mmHg/s Inflation time 20 to 185 mmHg, 1 to 5 s
Automatic software control, max. inflation pressure adult 280 ±10 mmHg child 200 ±10 mmHg infant 145 ±5 mmHg
Over pressure limit, stops measurement after 2 seconds adult 320 mmHg child 220 mmHg infant 160 mmHg
The safety circuit limits the maximum cuff pressure to 320 mmHg in adult/child mode or 160 mmHg in infant mode. Independent timing circuit limits pressurizing (>15 mmHg) time to 3 minutes maximum in adult/child mode, and 90 seconds at (>5mmHg ) in infant mode.
Zeroing to ambient pressure is done automatically.
1.3.2 ECG
Inflation pressure is adjusted according to the previous systolic pressure, typically 40 mmHg above. If the systolic pressure is not found, inflation pressure is increased typically 50 mmHg.
Max. measurement time adult 120 s child 120 s infant 75 s
Pressure transducer accuracy is better than ±3 mmHg or ±2 % whichever is greater. Max. error ±4 mmHg.
Protection against electrical shock Type BF defibrillation proof
Defibrillation protection 5000 V, 360 J Recovery time 5 s Input impedance >2.5 M (10 Hz) CMRR 95 dB (ST) System noise <30 µV (p-p, RTI)
Allowable offset ±800 mVDC Gain range 0.2 to 5.0 cm/mV Pacemaker pulse detection 2 to 700 mV, 0.5 to 2 ms pulses
Protection against electrical shock Type CF defibrillator proof
1.3.3 Pulse oximetry
Protection against electrical shock Type BF defibrillation proof
6
Document No. 8005571-1
Page 13
1.3.4 Temperature
Measurement accuracy ±0.1 °C (25.0 to 45.0 °C) ±0.2 °C (10.0 to 24.9 °C) Protection against electrical
shock Type CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on transducer/probe used. Please refer to the transducer/probe specification.
1.3.5 Invasive blood pressure
DIGITAL DISPLAY AVERAGING
Digital displays Art and P1 are averaged over 5 seconds and updated at 5 seconds intervals. All other pressures have respiration artifact rejection.
Accuracy ±5 % or ±2 mmHg, whichever is greater Transducer and input sensitivity 5 µV/V/mmHg, 5 VDC, 20 mA max current Filter 0 to 4 - 22 Hz adjustable Zero set accuracy ±1 mmHg Calibration resolution ±1 mmHg Zero time less than 15 s Protection against electrical shock Type CF defibrillation proof
S/5 M-PRESTN Modules
NOTE: The accuracy of the measurement may be different from the specified, depending on transducer/probe used. Please refer to the transducer/probe specification.
1.3.6 Respiration
Excitation frequency, 12-lead ECG 31.25 kHz
Breath detection automatic, range 0.3 to 6 manually adjustable minimum detection: 0.2, 0.4, 0.6, 0.8, 1.0
Input dynamic range 0.2 to 32 Input impedance range 100 to 5000 Respiration Rate min. 4 bpm
max. 120 bpm Lead off detection >3 M
Document No. 8005571-1
7
Page 14
Datex-Ohmeda S/5 monitors
2 FUNCTIONAL DESCRIPTION
2.1 Measurement principle
2.1.1 NIBP
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a speed based on the patient’s pulse, collecting data from the oscillations caused by the pulsating artery. Based on these oscillations, values for systolic, mean, and diastolic pressures are calculated.
The following parts are necessary for the NIBP measurement:
M-PRESTN/-RESTN/-PRETN module
twin hose (adult or infant model)
blood pressure cuffs (various sizes)
2.1.2 ECG
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical potential produced with electrodes placed on the surface of the body.
ECG reflects:
electrical activity of the heart
normal/abnormal function of the heart
effects of anesthesia on heart function
effects of surgery on heart function
See the User's Guide or the User’s Reference Manual for electrodes positions and other information.
2.1.3 Pulse oximetry
A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near infrared (about 900 nm) and the other in the red region (about 660 nm) of the light spectrum. These wavelengths are emitted by LEDs in the SpO tissue and is finally detected by a PIN-diode opposite the LEDs in the probe. The pulse oximeter derives the oxygen saturation (SpO relative absorption at the two wavelengths and the arterial oxygen saturation SaO
In order to measure the arterial saturation accurately, pulse oximeters use the component of light absorption giving variations synchronous with heart beat as primary information on the arterial saturation.
probe, the light is transmitted through peripheral
2
) using an empirically determined relationship between the
2
.
2
8
Document No. 8005571-1
A general limitation of pulse oximetry is that due to the use of only two wavelengths only two hemoglobin species can be discriminated by the measurement.
The modern pulse oximeters are empirically calibrated either against fractional saturation SaO
frac;
2
Page 15
2
2
fracSaO
= Formula 1
HbO
2
++
binDyshemogloHbHbO
S/5 M-PRESTN Modules
or against functional saturation SaO
2
funcSaO
HbO
= Formula 2
Functional saturation is more insensitive to changes of carboxyhemoglobin and methemoglobin concentrations in blood.
The oxygen saturation percentage SpO against functional saturation SaO measurement relative to SaO carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeters are not able to correctly measure oxygen content of the arterial blood at elevated carboxyhemoglobin or methemoglobin levels.
Plethysmographic pulse wave
The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation at the measuring site. Thus the amplitude of the waveform represents the perfusion.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The signals are filtered to reduce noise and checked to separate artifacts.
func;
2
2
HbHbO
2
+
measured by the Datex-Ohmeda module is calibrated
2
func. The advantage of this method is that the accuracy of SpO2
2
func can be maintained even at rather high concentrations of
2
Probe
The standard probe is a finger clamp probe which contains the light source LEDs in one half and the photodiode detector in the other half. Different kinds of probes are available from Datex-Ohmeda.
SpO2 sensor connector
6
GND
7
ILED
4
GND
5
DEF_A
8
IS
9
DET_C
PRSTN_absorption_of_infrared.vsd
Emitter
Detector
SpO2 sensor cable
IRED
RED
Figure 2 Absorption of infrared light in the finger probe, parts layout and schematic
diagram
Document No. 8005571-1
9
Page 16
Datex-Ohmeda S/5 monitors
2.1.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature changes, called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
Applying a constant voltage across the resistor and measuring the current that flows through it
Applying a constant current through the resistor and measuring the voltage that is generated across it.
In Datex-Ohmeda modules the two methods are combined in the form of a voltage divider. The NTC­resistor is connected in series with a normal resistor and a constant voltage is applied across them. The temperature dependent voltage can be detected at the junction of the resistors, thus producing the temperature signal from the patient. The signal is amplified by analog amplifiers and further processed by digital electronics.
2.1.5 Invasive blood pressure
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal saline all connected together by stopcocks, is attached to the catheter. The transducer is placed at the same level with the heart, and is electrically zeroed.
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
2.1.6 Respiration
Impedance respiration is measured across the thorax between ECG electrodes. The respiration signal is made by supplying current between the electrodes and by measuring the differential current from the electrodes. The signal measured is the impedance change caused by breathing. From these impedance changes, respiration rate is calculated, and the respiration waveform is displayed on the screen.
10
Document No. 8005571-1
Page 17
2.2 Main components
2.2.1 M-PRESTN/-RESTN/-PRETN modules
Auto
Start
Zero
On/Off
NIBP
ECG
Cancel
T1
T2
SpO
P1
P2
Zero
P1
P2
2
S/5 M-PRESTN Modules
Figure 3 Front panel of M-PRESTN
The M-PRESTN, M-RESTN, and M-PRETN modules contain three main PC boards, the STP board, the ECG board, and the NIBP board. They work independently. Each of these has their own processor and software in the processor flash memory.
There are two small boards, the SP input and the ECG input board attached to the front panel of the module. The front panel has seven connectors and four keys. The connectors are two for temperature measurement, two for invasive blood pressure measurement, one for ECG, one for NIBP, and one for SpO
measurement. The keys are for NIBP Auto On/Off, NIBP Start/Cancel, P1
2
zero, and P2 zero.
11
Document No. 8005571-1
Page 18
Datex-Ohmeda S/5 monitors
2.2.2 NIBP board
PATIENT AND NIBP CUFF
NIBP TUBING
NIBP CONNECTOR
NIBP PNEUMATICS
NIBP CONTROL KEYS
NV MEMORY
MAIN CPU
PRESSURE SENSORS
RS 485 COMMUNICATION POWER SUPPLY
DRIVERS FOR PUMP &
VALVES
SAFETY CPU
NIBP BLOCK DIAGRAM
Figure 4 NIBP board functional block diagram
Signal processing
Two signals from the pressure transducers are amplified and sent to the A/D converter. After the converter, digitized signals are sent to the microprocessor for data processing.
The NIBP board is controlled with a H8/3052 microprocessor at 16 MHz oscillator frequency.
12
Document No. 8005571-1
MODULE BUS CONNECTOR
Page 19
Memory
NIBP program memory (processor flash memory) size is 512k × 8. The processor has 4 kBytes RAM and there is also an external RAM memory the size of which is 128k x 8. Variable values of the NIBP measurement are stored into the external RAM. The EEPROM size is 512 x 8 and it is used to store the calibration values for the pressure transducers, the pulse valve constants gained during measurements, the PC board identification, and module serial number.
Software control
Software controls valves and pump. In addition to the individual on/off signals for each component there is a common power switch for the valves and the pump that can be used at pump/valve failures.
In addition to external RS485 reset line the microprocessor system is equipped with its own power­up reset. See the section in the ECG board’s description:
Safety circuit
The NIBP board is equipped with an independent safety circuit to disconnect supply voltages from the pump and the valves if the cuff has been pressurized longer than the preset maximum measurement time, or if the pressure of the cuff is inflated over the specified pressure limit. The maximum measurement time values and pressure limits for different measurement modes have been specified in the technical specification section of this manual.
S/5 M-PRESTN Modules
“RS485 communication”
Pneumatics
Pneumatics of PRESTN module has the following parts:
Intake air filter; for preventing dust and other parts to enter the air pump and the
valves.
Air pump: for pumping the measuring pressure of the cuff.
(
Pulse) Valve; for producing a linear pressure fall (bleeding) in order to measure the
blood pressure of the patient. Note that there has been used also two other names designate pulse valve in service menu.
Safety valve; The safety valve has been intended to be used for deflating the cuff in
single fault case, i.e. to prevent too long measurement time or too high inflation pressure of the cuff. Note that there has been used also service menu.
Main pressure sensor; for measuring the pressure of the blood pressure cuff and
the pressure fluctuations caused by arterial wall movement.
Safety pressure sensor for detection of cuff hose type, cuff loose, cuff occlusion
situations etc.and recognising the pressure sensor fault.
Exh2 valve to designate the Safety valve in
Valve and Set valve to
Cuff connector; for connecting the cuff.
13
Document No. 8005571-1
Page 20
Datex-Ohmeda S/5 monitors
Cuff connector
Safety pressure sensor
Proportional valve
Air pump
Main pressure sensor
Figure 5 NIBP pneumatics diagram
Power supply section of the NIBP board
All connections are established via 25-pin connector (D-type, female). The module needs +15 V (dirty) power supply to operate. The supply voltage (+15V) is generated in the power supply section of the S/5 monitor. The other voltages needed for the operation of the NIBP measurement are made on the NIBP board.
Dump valve
Intake air filter
NIBP_pneum_diagr.vsd
14
Document No. 8005571-1
Page 21
2.2.3 ECG board in 12-lead measurement
The 12-lead ECG measurement consists of the elements shown in Figure 6. All functions are located in the ECG board except the ECG input unit .
PATIENT AND ECG ELECTRODES
ECG CABLE
- ECG LEAD SET
- ECG TRUNK CABLE
ECG INPUT UNIT
- ECG CONNECTOR
- INPUT PROTECTION RESISTORS
INPUT PROTECTION DIODES FOR ECG & RESPIRATION MEASUREMENT
INPUT FILTERING FOR ECG & RESPIRATION MEASUREMENTS
S/5 M-PRESTN Modules
RESPIRATION
MEASUREMENT
AMPLIFIERS
RS 485 COMMUNICATION POWER SUPPLY
ISOLATION ISOLATION
MODULE BUS CONNECTOR
RESPIRATION
MEASUREMENT CURRENT
SUPPLY
ECG CPU
ECG PREAMPLIFIERS &
RLD CIRCUIT
BASELINE
RESTORATION
NV MEMORY
ECG BLOCK DIAGRAM
LEADS OFF & PACER &
DEFIBRILLATION
DETECTION
Figure 6 12-lead ECG measurement block diagram
15
Document No. 8005571-1
Page 22
Datex-Ohmeda S/5 monitors
ECG input unit
The ECG input unit consists of the front panel connector and the ECG input connector board with the high voltage protection resistors. The connector for the 12-lead ECG cable is a green 11-pin rectangle shaped connector.
Input protection and filtering
The input protection is implemented with high voltage protection resistors in the ECG input unit and with protection diodes in the ECG board. The input filtering for ECG measurement is done with passive RC filtering.
ECG preamplifiers
The buffer amplifiers are used for each lead. The “Leads off” detection is implemented by measuring the output level of the input buffer amplifiers with A/D converter of CPU. The ECG signals are measured using differential amplifiers.
ECG amplifiers and baseline restoration
The function of the ECG amplifiers and baseline restoration is to amplify the signal and to restore the baseline of the signal in the middle of the display after the change of the signal level e.g. after the change of the DC offset voltage.
Pacer detection
Pacer detection has been made by using three slew rate detector circuits. The pacer detection amplifiers have been realized at the front of the slew rate detectors independently from the ECG measuring channels.
Respiration impedance supply
The 31.25 kHz sine wave generator is used as the respiration measurement signal supply. Analog switches are used for connecting the sine wave to the ECG leads to be measured.
Respiration impedance amplifiers
Buffer amplifiers are used in respiration measurement. Analog switches are used for selecting the measurement leads. There are also additional amplifiers for increasing the respiration signal gain. When ECG measurement is 5/12-lead, the respiration measurement is always done between R and F, independently on the ECG lead selection. When ECG measurement is 3-lead, then the respiration measurement is done at the same lead as the ECG measurement (I, II or III).
ECG CPU
16
Document No. 8005571-1
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory and 4 kbytes of RAM. The clock frequency is 16 MHz.
Page 23
RS485 communication
The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver circuits are optically isolated from the processor of the module. PWM signal is used for direct ECG signal. Direct ECG signal is available from the X2 connector of the UPI board or from the PT module.
Power supply
The ECG board has a driver controlled half bridge switching power supply with 5 kV isolation. The supply voltages have been regulated with linear regulators.
2.2.4 ECG filtering
The S/5 monitors have three ECG filtering modes:
MONITORING 0.5 to 30 Hz (with 50 Hz reject filter)
0.5 to 40 Hz (with 60 Hz reject filter)
DIAGNOSTIC 12-lead ECG 0.05 to 150 Hz
ST FILTER 0.05 to 30 Hz (with 50 Hz reject filter)
0.05 to 40 Hz (with 60 Hz reject filter)
S/5 M-PRESTN Modules
The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory) movement artifacts.
Monitor filter is used in normal monitoring. Diagnostic filter is used if more accurate diagnostic information is needed. ST filter gives more accurate information of ST segment, but reduces high frequency noise.
The high-pass filters 0.5 Hz and 0.05 Hz are done with software. The monitor sends a command to the hemodynamic module determining which of the corner frequencies 0.5 Hz or 0.05 Hz is to be used.
The 50 Hz and 60 Hz reject filters are both low-pass filters with zero at 50 Hz or 60 Hz correspondingly. They are software based filters used for the mains supply filtering. With these filters the 3 dB value for low-pass filter is 30 Hz or 40 Hz.
In diagnostic mode the upper frequency is 150 Hz and it is limited by software.
17
Document No. 8005571-1
Page 24
Datex-Ohmeda S/5 monitors
2.2.5 STP board
PATIENT AND SpO2 PROBE
SpO2 TRUNK CABLE
SpO2 CONNECTOR
SpO2 PROBE RECOGNITION & LED
DRIVE SELECTION MATRIX
SpO2 LED DRIVE
SpO2 AMPLIFIER
PATIENT AND TEMPERATURE
SENSOR
TEMPERATURE CONNECTOR
TEMPERATURE
MEASUREMENT
AMPLIFIER
A/D CONVERSION
SENSOR
SIGNAL
CURRENT
SOURCE
PATIENT AND INVASIVE CANNULA
PRESSURE SENSOR WITH DOME
INV: PRESSURE
MEASURE MENT
OR CATHETER
FLUSHING KIT & INVASIVE
INV.PRESSURE CONNECTORINPUT PROTECTION CIRCUITRY
SENSOR SIGNAL
AMPLIFIER
VOLTAGE
SOURCE
STP CPU
NV MEMORY
RS 485 COMMUNICATION POWER SUPPLY
ISOLATION ISOLATION
MODULE BUS CONNECTOR
STP BLOCK DIAGRAM
Figure 7 STP board block diagram
18
Document No. 8005571-1
Page 25
Microprocessor unit
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory and 4 kbytes of RAM. The clock frequency is 16 MHz.
High speed I/O is used to obtain pulse control sequence necessary for pulse oximetry measurement. Timing for the clock is from the oscillator.
Temperature measurement unit
The NTC-resistor value in the probe depends on the patient’s temperature. It is measured with the following principle described below.
The constant current source is supplied about 7µA current through the temperature sensor (YSI 400-series NTC resistor). The constant current is caused a voltage over the temperature sensor (NTC resistor). The voltage over the temperature sensor is amplified in a differential amplifier stage. The amplified voltage is transferred to a controller of the STP board through an A/D converter.
Defibrillation/ESD protection
resistors and diodes
S/5 M-PRESTN Modules
constant
current source
Temperature
sensor
Figure 8 Temperature measurement principle
100k
10k
0C:7k36 15C:3k54 25C:2k53 38C:1k30 45C: 984
10k
100k
Differential
amplifier
d/dt
0
0
0
0
To A/D
converter
temp_meas_principle.vsd
19
Document No. 8005571-1
Page 26
Datex-Ohmeda S/5 monitors
Invasive blood pressure measurement unit
An isolated +5 V voltage is supplied to the pressure transducer. The differential voltage, which depends on the pressure and the supplied voltage, is calculated from the bridge connection (see the formula below).
U
= Uin × pressure × 5 V, where Uin is 5 V
out
U
= 25 V × pressure [mmHg]
out
Pressure amplification is realized in the instrumentation amplifier. The gain of the amplifier is set to keep the level of the signal transferred to A/D converter within the measurement range even when there are circumstantial offsets or offsets caused by the transducer. There is a filter before the amplifier to attenuate high frequency disturbances.
Vin
Pressure transducer
Vout
Figure 9 Pressure measurement principle
Pulse oximetry measurement section
LED control signals
The D/A converters of the microcontroller on STP board set the LED intensity adjustment values for the infrared- and red LEDs of SpO2 probe. The microcontroller on the STP board switches ON ( to the adjusted intensity ) and OFF the SpO2 probe LEDs according to the predetermined sequence.
Input Filter
Instrumentation amplifier
G
to AD converter
pressure_meas_principle.vsd
LED driving circuit
Differential amplifiers measure the LED currents (LED current indication) of SpO2 probe over the shunt resistors placed in the LED current paths. The LED driving voltages (LED voltage indication) are measured from the driver circuitry. The LED driving circuits also have MOSFET transistor matrix to enable the use of different probe configurations.
Measured signal preamplification
The preamplifier is bipolar/single-ended current-to-voltage converter with adjustable gain. A higher gain is used for measuring thin tissue. The preamplification stage has also ambient light reduction and second amplifier stage.
20
Document No. 8005571-1
Page 27
S/5 M-PRESTN Modules
SpO2 Probe
Ambient reduction
Preamplifier: Current-to-voltage type Bipolar/single-ended modes Adjustable gain
Amplifier: Gain = 2
LED Driving circuit 1
LED Driving circuit 2
DE-MUX
LED Intensity adjustment 1
LED ON/OFF control 1
LED voltage indication 1
LED current indication 1
LED Intensity adjustment 2
LED ON/OFF control 2
LED voltage indication 2
LED current indication 2
LP
LP
LP
LP
Gain=7.5
suppression
Gain=7.5
suppression
Gain=7.5
suppression
Amplifiers
Oximeter channel 1
DC-
Oximeter channel 2
DC-
Oximeter channel 3
DC-
Oximeter channel 4
Figure 10 Pulse oximetry measurement block diagram
Red and infrared channel separation
The detector signal is possible to multiplex to four different channels depending on the content of the signal. The detector signal must at least multiplex into infrared- and Red signals. Other channels are for e.g. diagnostic purposes.
Analog Digital
Spo2_measurement_blck_diagr.vsd
21
Document No. 8005571-1
Page 28
Datex-Ohmeda S/5 monitors
Serial communication
An RS485 type bus driver makes the serial communication between the module and the frame. Data transmission rate is 500kbps.
S R E
L L
O R
T
RECEIVE DATA
N O C
SEND DATA
D R A O
SEND/RECEIVE
B P
B
I N
NIBP BOARD
RS 485
DRIVER
DATA
NDATA
Figure 11 Serial communication of NIBP board
R E
L L
O R
T
RECEIVE DATA
N O C
SEND DATA
D R A O
SEND/RECEIVE
B P
T S
STP BOARD
RS 485
DRIVER
PATIENT ISOLATION
D R A O B
G C E
/ P T S
O T
R O
T C E N N O C
DATA
NDATA
MODULE BUS CONNECTOR
R O
T C E N N O C
D R A O B
P T S
NIBP_communication.vsd
STP_communication.vsd
22
Document No. 8005571-1
Figure 12 Serial Communication and Isolation of STP board
R E
L
DIRECT ECG (PWM)
L O R
T
RECEIVE DATA
N O C
D
SEND DATA
R A
O B
SEND/RECEIVE
G C E
ECG BOARD
PATIENT ISOLATION
DIRECT ECG (PWM)
DATA
RS 485
DRIVER
NDATA
Figure 13 Serial Communication and Isolation of ECG board
R O
T C E N N O C
D R A O B
G C E
ECG_communication.vsd
Page 29
Signals and isolation barrier
The communication signals transfer over the isolation barrier by using high isolation voltage (6kV) opto isolators.
Power supply section
The power for the electronics on the floating part of the STP and the ECG boards is made on each board with the switching power supplies connected to a high voltage isolated transformer. The switching power supplies on the STP and ECG boards are synchronized to the frequency, about 340kHz of the switching power supply on the NIBP board. The PRESTN module uses only +15VD voltage of the frame. The other voltages of the measuring boards are made by the switching power supplies and regulators or the linear regulators. Each measuring board is protected against overloading with PTC type automatic fuses.
2.3 Connectors and signals
2.3.1 Module bus connector (on the NIBP board)
S/5 M-PRESTN Modules
13
25
1
14
Figure 14 Module bus connector (X1) pin layout
23
Document No. 8005571-1
Page 30
Datex-Ohmeda S/5 monitors
Table 2 Module bus connector description
Only the shaded signals of the table below are valid for the PRESTN module
Pin No I/O Signal Note
1 I RESET_RS485
2 I -15 VDC
3 I +15 VDIRTY
4 I +15 VDC
5 I/O NDATA_RS485
6 I/O DATA_RS485
7 Ground & Shield
8 I -RESET_RS485
9 I CTSB
10 O RTSB
11 I RXDB
12 O TXDB
13 Ground & Shield
14 I +32 VDIRTY
15 I GroundDIRTY
16 I CTSC
17 O RTSC
18 I RXDC
19 O TXDC
20 ON/STANDBY
21 O PWM_ECG
22 RXDD_RS232
23 TXDD_RS232
24 I +5 VDC
25 I +5 VDC
24
Document No. 8005571-1
Page 31
2.3.2 Front panel connectors

Table 3 Front panel connectors

ECG Connector Pin No Signal Name
S/5 M-PRESTN Modules
1 2 3 4 5 6 7 8 9 10 11
R/RA; Right arm electrode C2/V2; Chest electrode C3/V3; Chest electrode L/LA; Left arm electrode N/RL; Neutral/ Right Leg Drive electrode C1/V1; Chest electrode C4/V4; Chest electrode F/LL; Left Leg electrode C6/V6; Chest electrode C5/V5; Chest electrode Cable Shield
SpO2 connector (SpO2) Invasive blood pressure
connectors (P1, P2)
6
3
1
2
9
4
7
0
5
8
6
3
1
2
9
4
7
0
5
8
Temp connector
(T1, T2)
Pin No Signal Pin No Signal Pin No Signal
1
2
3
4
5
6
7
8
9
10
Feedback resistor
Ground
Not Connected
Cable shield + probe identification ground
Probe identification
LED drive ground
LED drive current
Input signal current
Ground
Ground
1
2
3
4
5
6
7
8
9
10
Pressure +
Pressure -
Polarisation - (ground)
Polarisation +
Not connected
Not connected
Not connected
Not connected
Ground
Cable detection
1
2
Temperature probe
Temperature probe
25
Document No. 8005571-1
Page 32
Datex-Ohmeda S/5 monitors
2.3.3 Test points on boards
12-lead ECG board
Test pins (strips) of X103:
Component side:
Solder side:
PROGR.
PIN1
6
X103
6: +5VD
1: GNDF
4
5
1
X102
TEST
Test pins (strips) of X102:
Component side:
Solder side:
1
2
3
2: 2.5V 4: 2.5Vref
1:+5VA 3: +2.5V_R 5: GNDF
ECG INPUT BOARD CONNECTOR
MODULE BUS CONNECTOR OF ECG BOARD
26
ISOLATIO N TRANSFORMER
PIN ORDER FOR MODULE BUS CONNECTOR OF ECG BOARD:
1: +15VD 2: GNDN 3: DATA 4: NDATA 5: GNDN
6: PS_OSC(340kHz) 7: NC 8: GNDN 9: PWM_ECG 10: +5VRS
Figure 15 ECG board connectors and test points
ECG_board_conn_testpoints_01.vsd
Document No. 8005571-1
Page 33
NIBP board
S/5 M-PRESTN Modules
CONNECTOR FOR NIBP FLEX
+5VGND
X103
PROGR
Pump
(Pulse)
MODULE BUS CONNECTOR
NIBP MANIFOLD
PIN1
PIN1
OF STP BOARD
X106
MOD.BUS CONNECTOR
Test connector X106 : 1: POS.PUMP VOLTAGE
2: (Pulse) VALVE VOLTAGE 4: NEG.PUMP VOLTAGE
6: Safety VALVE VOLTAGE 8,9,10: GND
Figure 16 NIBP board connectors and test points
valve
Safety
valve
Intake
NIBP_brd_conn_testpoints_01.vsd
Air Filter
27
Document No. 8005571-1
Page 34
Datex-Ohmeda S/5 monitors
STP board
ISOLATION TRANSFORMER
PIN1
MODULE BUS CONNECTOR
TEST CONNECTOR PIN ODER:
19: +2.3V(ref.voltage) 20: GNDF 21: +6V(non regulated) 22: GNDF 23: +4.5V(ADC ref. voltage) 24: GNDF
Figure 17 STP board connectors and test points
PROGR
X107
TEST CONNECT OR
PIN2 PIN34
PIN1
25: +5V 26: GNDF 27: +5V(analog) 28: GNDF 29: +5VLED 30: GNDF
STP INPUT CONNECTOR
STP_brd_conn_testpoints_01.vsd
28
Document No. 8005571-1
Page 35
3 SERVICE PROCEDURES
3.1 General service information
Field service of the hemodynamic modules is limited to replacing faulty printed circuit boards or mechanical parts. Faulty printed circuit boards should be returned to Datex-Ohmeda for repair.
Datex-Ohmeda is always available for service advice. Please provide the unit serial number, full type designation, and a detailed description of the fault.
Only trained personnel with appropriate equipment should perform the tests and repairs outlined in this section. Unauthorized service may void warranty of the unit.
3.2 Service check
These instructions include complete procedures for a service check. The service check should be performed after any service repair; however, the service check procedures can also be used for determining possible failures.
S/5 M-PRESTN Modules
The procedures should be performed in ascending order.
The instructions include a check form , which should be filled in when performing the procedures.
The mark
check form.
" in the instructions means that the performed procedure should be signed in the
3.2.1 Recommended tools
Tool Order No. Notes
Patient simulator -
Pressure manometer -
Temperature test set 884515
Multi-link 3-leadwire set
Multi-Link 5-leadwire set V2-V6
Multi-Link 5-leadwire set C2-C6
Multi-Link 12-lead ECG trunk cable
SpO2 finger probe OXY-F4-N or SAS-F4
InvBP transducer
Adult NIBP cuff & hose
Infant NIBP cuff & hose
Screwdriver
3.2.2 Recommended parts
Part Order No. Notes
NIBP pump filter 57142
29
Document No. 8005571-1
Page 36
Datex-Ohmeda S/5 monitors
All modules
Detach the module box by removing the two screws from the back of the module. Be careful with loose latch and spring pin for locking.
1. Check internal parts:
screws are tightened properly
cables are connected properly
all IC’s that are on sockets are attached properly
EMC covers are attached properly
there are no loose objects inside the module
"
2. Check external parts:
the front cover and the front panel sticker are intact
all connectors are intact and are attached properly
the module box, the latch and the spring pin are intact
"
3. Replace the NIBP pump filter in PRESTN/RESTN/PRETN modules, if necessary.
"
Reattach the module box and check that the latch is moving properly.
Switch the monitor on and wait until the monitoring screen appears. Configure the monitor
screen so that all the needed parameters are shown, for example as follows:
Monitor Setup - Waveform Fields - Field 1 - ECG1
Field 3 - P1 Field 4 - P2 Field 5 - Pleth Field 6 - Resp
Lower Field 3 - T1+T2
Field 2 - ECG2
Digit Fields - Lower Field 2 - NIBP
30
Document No. 8005571-1
Page 37
S/5 M-PRESTN Modules
4. Plug in the module. Check that it goes in smoothly and locks up properly
"
5. Check that the module is recognized, i.e. all needed parameter information, except invasive blood pressure, starts to show on the screen.
"
ECG measurement
6. Enter the service menu:
Preset ECG, Respiration, InvBP and SpO
ECG - ECG Setup - Hr Source - Auto
Pacemaker - Show
Others - Resp Setup - Size - 1.0
Resp Rate Source - Auto Measurement - On Detection Limit - Auto
Invasive Pressures - P1 ‘Art’ Setup - Label - Art
P2 ‘Cvp’ Setup - Label - Cvp
Pulse Oximetry - Pleth Scale - Auto
Monitor Setup - Install/Service (password 16-4-34) -
Service (password 26-23-8) - Parameters
Take down the information regarding module software by selecting Scroll Vers and turning the ComWheel.
measurement settings:
2
"
7. Enter the ESTP: ECG service menu:
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing faster than by 5 per second. Check also that the ECG/RESP board memories have passed the internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ state all OK.
"
8. Check that the power frequency value is set according to the current mains power
frequency. Change the setting by selecting
Power Freq, if necessary.
"
31
Document No. 8005571-1
Page 38
Datex-Ohmeda S/5 monitors
9. Connect a 12-lead ECG trunk cable without a lead set to the module. Check that the message "Leads off" is displayed on the screen.
"
10. Connect both 5-leadwire sets to the trunk cable. Connect limb lead electrodes and one electrode from the chest lead set to the same potential. Check that the ‘Cable type’ shows 10 lead.
"
11. Change the 3-leadwire set to the trunk cable. Check that all the electrodes show OFF in the service menu and the message ‘Leads Off’ is shown on the screen.
Connect all the leads together, for example to a suitable screwdriver. Check that all the electrodes show ON and the message ‘Asystole’ appears.
Disconnect one of the leads and check that the corresponding electrode in the service menu shows OFF within 10 seconds from the disconnection, and then reconnect the lead. Check the rest of the leads using the same method.
NOTE: When the ground lead (black) is disconnected all the electrodes should show OFF.
NOTE: The ‘Asystole’ and ‘Different leads off’ messages are shown using certain priority, so even when one of the leads is disconnected, the lead related ‘Leads off’ message may not appear onto the screen.
NOTE: When RA, LA or LL electrode is disconnected, all six V electrodes show OFF.
NOTE: With PRESTN/RESTN/PRETN modules and 5 lead cable the state of V2, V3, V4, V5 and V6 electrodes follow the state of the V electrode.
Connect the leads to a patient simulator. Perform the settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
32
Document No. 8005571-1
ECG - BASE - BPM - 160 PACE - WAVE - NSR
Check that normal ECG waveform is shown, the HR -value is 160 (±5) and the ‘Pacer count’ ­value is not increasing in the service menu.
ECG - PACE - WAVE - ASNC
Check that pacemaker spikes are shown on the ECG waveform, the ‘HR’ -value changes to 75 (±5) and the ‘Pacer count’ -value is increasing according to shown pacemaker spikes.
Set the pacemaker option off:
ECG - PACE - WAVE - NSR
"
Page 39
Respiration measurement
12. Check that the ‘Resp Available’ and ‘RESP Measurement’ show both ON in the ESTP: ECG service menu.
"
13. Check the respiration measurement with a patient simulator.
The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
BASELINE IMPEDANCE -switch LEAD SELECT-switch
RESP - WAVE - NORM RATE - 20 OHMS - 1.0 RATIO - 1/1 APNEA - OFF SHIFT - OFF
S/5 M-PRESTN Modules
- 500
- II/RL-LL
Check that the RESP waveform is shown and the ‘RR’ -value is 20 (±5). Change the position of the BASELINE IMPEDANCE -switch and check that appropriate RESP waveform and ‘RR’ ­value are shown again within 30 seconds.
RESP - APNEA - 32 S
Check that the monitor activates the APNEA -alarm.
NOTE: Make sure that only the ECG leads are connected to the simulator during the apnea ­test. If other cables are connected at the same time, the respiration signal from the simulator may be disturbed, and therefore, the APNEA -alarm may not be activated.
NOTE: When you have ECG service menu open, spikes will appear on the respiration waveform. These spikes represent the threshold level for detecting inspiration and expiration.
"
Temperature measurement
14. Enter the ESTP: STP service menu:
Parameters - ESTP : STP
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values do not increase faster than by 5 per second. Check also that the STP board memories have passed the internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ show all OK.
"
15. Check that the ‘Cable’ and ‘Probe’ show OFF for both channels, T1 and T2, when no probes
are connected.
Connect a temperature test plug into the connector T1. Check that the ‘Cable’ and ‘Probe’ for
Document No. 8005571-1
33
Page 40
Datex-Ohmeda S/5 monitors
T1 show ON and the corresponding temperature value appears onto the monitor screen. Perform the same check also for the channel T2.
"
16. Check the temperature calibrations using temperature test plugs. If the deviation on a temperature reading on the screen is more than 0.1 °C, calibrate the temperature channels according to the instructions in the chapter
calibration
"
3.4.2 Temperature
.
17. Activate the temperature test by selecting ComWheel twice. When the message ‘Performing temp test’ disappears from the digit field, check that no error messages appear and ‘Temp error’ shows OFF for both channels in the service menu.
"
18. Check that the module configuration has been set correctly. The configuration in use is shown beside the text ‘Configuration’ in the service menu and it can be either STP, ST or TP. Change the configuration in the
"
Invasive blood pressure measurement
19. Check the front panel membrane keys that are related to the InvBP measurement.
Press each of the keys at least for one second. Check that the pressed key is identified, i.e. one of the texts for ‘Buttons’ changes from OFF to ON in the service menu.
"
20. Check that the ‘Cable’ and ‘Probe’ for P1 show OFF. Plug a cable with an invasive blood pressure transducer into the front panel connector P1 and check that the ‘Cable’ and ‘Probe’ show ON and the corresponding pressure waveform appears onto the screen.
Perform the same check also for the InvBP channel P2.
Temp Test from the menu and pressing the
Calibrations menu, if necessary.
34
Document No. 8005571-1
"
21. Calibrate the InvBP channels P1 and P2 according to the instructions in the chapter
Invasive pressure calibration
"
22. Check the InvBP channels with a patient simulator.
The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
SENSITIVITY -switch - 5 µV/V/mmHg
ECG - BASE - BPM - 60 - BP - 1 - WAVE - ATM 2 - WAVE - ATM
.
Page 41
S/5 M-PRESTN Modules
SpO2 measurement
23. Check that the message ‘No probe’ is shown when no SpO2 sensor is connected to the
24. Connect the SpO
Restore the normal monitoring screen by pressing the key
Connect cables from the channels BP1 and BP2 to the module connectors P1 and P2. Zero the InvBP channels by pressing the keys ZERO P1 and ZERO P2 on the module front panel.
BP - 1 - WAVE - ART 2 - WAVE - CVP
Check that appropriate InvBP waveforms are shown and the InvBP values are approximately 120/80 (±3 mmHg) for the channel P1 and 15/10 (±2 mmHg) for the channel P2.
Check that HR- value is calculated from P1 when ECG is not measured (ECG cable disconnected).
Normal Screen.
"
module. Connect a SpO shown when the probe is not connected to a finger.
finger probe to the module. Check that the message ‘Probe off’ is
2
"
probe onto your finger. Check that the reading of
2
95-99 and SpO ECG and InvBP (P1) are not measured.
waveform appears. Check that HR- value is calculated from SpO2 when
2
"
Non Invasive Blood Pressure measurement
25. Enter the NIBP module service menu:
Parameters - NIBP
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing faster than by 5 per second. Check also that the NIBP board memories have passed the internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ show all OK.
"
26. Check the front panel membrane keys.
Select
Buttons/Leds.
Press each of the two NIBP related membrane keys at least for one second. Check that the pressed key is identified, i.e. the corresponding text changes from OFF to ON in the menu when the key is released back up again.
"
27. Check the pump and valves.
Highlight
Pneumatics from the NIBP menu. Connect a pressure manometer to the NIBP
35
Document No. 8005571-1
Page 42
Datex-Ohmeda S/5 monitors
module cuff connector.
Select pressure inside the tubing system starts to increase. Stop the pump by pressing the ComWheel again when the pressure reaches 280 mmHg.
Highlight system starts to drop then press the ComWheel again. If necessary, turn the pump on again for a moment to increase the pressure inside the tubing system.
Highlight number 150 by turning the ComWheel. Press the ComWheel again and check that the pressure inside the tubing system starts to drop. Finish the test by selecting
"
28. Check the NIBP tubing system for leakages.
Select
Connect the pressure manometer to the NIBP module cuff connector. Start the active leak test from the menu by pressing the ComWheel. The module pumps a pressure of about 265 mmHg and then the pump stops. Wait for 15 seconds for the pressure to stabilize then check that the pressure does not drop more than 6 mmHg per one minute. Release the pressure by pressing the ComWheel once more.
Start Pump and press the ComWheel. Check that the pump turns on and the
Open Exh2. Press the ComWheel and check that the pressure inside the tubing
Set Valve. Press the ComWheel and set the value under the text ‘Pulse Valve’ to
Previous Menu.
Calibrations from the NIBP service menu.
"
29. Calibration check.
Recalibrate the NIBP measurement. Remember to set the calibration protection back on after the calibration.
Disconnect the pressure manometer. Select
. Press the ComWheel and take down the zero offset values for both pressure
Check
transducers, B1 and B2. The values should be within ±20 mmHg.
Connect the pressure manometer to the cuff connector and check the calibration with pressures 100 mmHg, 200 mmHg and 260 mmHg. The zero offset value must be added to the displayed pressure value in order to determine the real pressure.
Calibrations and then highlight Calibration
"
30. Check the safety valve.
Select
Safety Valve from the NIBP service menu.
Keep the pressure manometer connected to the cuff connector.
NOTE: Make sure your pressure manometer can be used to measure pressures over 300 mmHg. If such a pressure manometer is not available, perform the check with an adult cuff that is connected around some round object, for example a calibration gas bottle.
Highlight
Start Test. Start the adult safety valve test by pressing the ComWheel. Wait until
36
Document No. 8005571-1
Page 43
S/5 M-PRESTN Modules
the pump stops and the pressure is deflated. Check the pressure values ‘Max press’ and ‘2 s after stop’ for both transducers. All the values should be within 270 - 330 mmHg.
Highlight Select been updated. Check that the values ‘Max press’ and ‘2 s after stop’ are all now within 135 to 165 mmHg.
Return to the normal monitoring mode by pressing
ADULT. Press the ComWheel and check that the text changes now to INFANT.
Start Test and wait until the pump stops and the pressure values on the screen have
Normal Screen.
"
31. Connect an adult NIBP cuff to the cuff connector and disconnect one of its hoses. Start NIBP measurement by pressing the key the message ‘Cuff loose’ appears on the screen within 70 seconds.
Reconnect the hose and then bend it with your fingers. Restart the measurement and check that the message ‘Cuff occlusion’ appears on the screen within 70 seconds.
Start/Cancel on the module and check that
"
32. Check that automatic inflation limits are in use:
All modules
NIBP - NIBP Setup - Inflation Limits - Auto - Previous Menu
Connect the cuff onto your arm, highlight Start Ven.Stasis in the NIBP menu and press the ComWheel. Check the module identifies the cuff, i.e. the text ‘Adult’ appears into the NIBP digit field for a short moment.
Keep the pressure inside the cuff for about half a minute in order to find out that the cuff is not leaking, then press the ComWheel again. Select
Normal Screen.
"
33. Keep the cuff on your arm and perform one NIBP measurement. Check that the module gives
a reasonable measuring result.
"
34. Connect an infant cuff to cuff connector and wrap it around your fingers.
Start NIBP measurement and check that the module identifies the cuff, i.e. the text ‘Infant’ appears into the NIBP digit field. Cancel the measurement after the cuff identification.
"
35. Perform electrical safety check and leakage current test.
"
36. Check that the module functions normally after the performed electrical safety check.
"
Document No. 8005571-1
37
Page 44
Datex-Ohmeda S/5 monitors
37. Clean the module with suitable detergent.
"
Fill in all necessary documents.
38
Document No. 8005571-1
Page 45
S/5 M-PRESTN Modules
3.3 Disassembly and reassembly
CAUTION When reassembling the module, make sure that all cables are reconnected
properly.
3.3.1 M-PRESTN/-RESTN/-PRETN modules
Disassemble the M-PRESTN/-RESTN/-PRETN module in the following way. See the exploded view of the module in chapter
1. Remove the two screws from the back of the module.
2. Pull the module box slowly rearwards and detach it from the main body. Be careful with loose latch and spring pin for locking.
3. To detach the ECG board, detach four screws, disconnect ribbon cable from the STP board (supply voltage), and ribbon cable from the ECG input board.
4. When the ECG board is removed, the STP board can be detached by removing four screws, disconnecting the cable from the membrane keypad, the cable from the temperature connectors, and cables from the SP input board. Also disconnect the NIBP hoses and the ribbon cable from the NIBP board.
6.1.1.
5. When the ECG board and the STP board are removed, the NIBP board can be detached by removing four screws. Disconnect the hoses from the pressure transducers and the pump. If the filter for the air inlet of the pump is removed, it must be replaced.
39
Document No. 8005571-1
Page 46
Datex-Ohmeda S/5 monitors
3.4 Adjustments and calibrations
3.4.1 NIBP calibrations
The electronics of the NIBP pressure measurement is calibrated at the factory. The processor automatically maintains the zeroing pressure. If the zero point of the pressure transducer drifts more than specified, an error message is given and the NIBP board should be recalibrated or replaced.
Recalibrate the NIBP measurement once a year. The checking and recalibrating can be done in the NIBP service menu.
The calibration of the primary pressure channel can also be checked from the NIBP setup menu (
NIBP - NIBP Setup - Calibration Check). In this case the auto zeroing is performed at start -
remove hose before entering to ensure atmospheric pressure to the pressure transducers - the primary pressure is displayed. The zero-offset value should then be zero.
Check the intake air filter as part of the calibration check. Change the filter if it is visibly dirty.
Calibration check
1. Enter Calibration menu.
2. Select
3. Connect an external precision manometer to the module.
Calibration Check and press the ComWheel.
40
Document No. 8005571-1
Page 47
S/5 M-PRESTN Modules
4. Pump the following pressures to manometer and check the difference between the
manometer and monitor pressure display (The zeroing offset is automatically subtracted from the pressure readings).
Table 4 NIBP calibration check pressures
Pressure Max. error Example
Calibration
0 mmHg
100 mmHg
200 mmHg
±5 mmHg (=zero offset)
100 ±2 mmHg 100 ±2
200 ±3 mmHg 200 ±3
-1
If the error of pressure channel B1 is larger than specified above, the module should be recalibrated. The error of B2 is allowed to be even twice as large because it has no effect on blood pressure measurement accuracy. However, we recommend recalibrating the module when the error of B2 is larger than specified above to ensure best possible operation.
1. Enter Calibration menu.
2. Remove hoses from front panel connector to enable proper zeroing.
3. Select
Calibration. If it is not available, perform the steps A, B, and C.
NOTE: Do not pull out the NIBP module from the monitor frame. The module must be in the frame during the whole procedure.
A. Press NIBP module buttons
enable the calibration. This enables menu selection
Auto ON/OFF and Start Cancel simultaneously for 3 seconds to
Protection. The message ‘Calibration
switch ON!’ appears.
B. Select
C. Press the buttons again for 3 seconds. Menu selection
Protection OFF in the Calibration menu and press the ComWheel.
Calibration is now enabled, and
Protection is disabled. When the calibration is enabled, a message ‘Calibration not
protected’ appears.
Start Calibration by pressing the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in
the NIBP message field. After this a pressure bar and text ‘Calibrating’ will appear.
Connect an external mercury manometer with pump to module through the both tubes of the
hose - both transducers B1 and B2 must be calibrated simultaneously. Pump up to a pressure about 200 mmHg according to the manometer. Calibration is possible in the range 150 to 250 mmHg.
Verify that both pressure values in the prompt field match the manometer reading. If not,
adjust by turning the ComWheel. When the values of the pressure bar and the manometer are equal, press the ComWheel to confirm the calibration. The message ‘Calibrated’ will appear onto the NIBP digit field after a few seconds, which means that the calibration succeeded, and the new calibration data is saved in EEPROM.
NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the ComWheel, even in cases where the value would be correct. For example change the value one step higher and then back one step lower. “Calibrated” text should appear in the display. This ensures that the calibration procedure is correctly registered and stored by the module.
41
Document No. 8005571-1
Page 48
Datex-Ohmeda S/5 monitors
To set the protection on: Press NIBP module buttons
Protection ON and push the ComWheel. Then press the buttons again for three
Select seconds.
Remove the module from the frame and plug it back again. Then perform Calibration check
(see the preceding page) to verify the new calibration.
3.4.2 Temperature calibration
NOTE: For the temperature calibration, separate, accurate test plugs (25 °C and 45 °C) are needed. A test set of two plugs is available from Datex-Ohmeda, order code 884515.
Calibrate the temperature when measured test values differ for more than ±0.1 °C, and always after STP board replacement.
1. Enter ESTPR: STP service menu.
Auto ON/OFF and Start Cancel simultaneously for 3 seconds.
2. Enter
3. Choose
4. Select
5. Insert calibration plug (25 °C) into T1/T2 connector.
6. Press the ComWheel.
7. Insert calibration plug (45 °C) into T1/T2 connector.
8. Press the ComWheel.
9. Choose
Calibrations menu.
Protection OFF in protect mode.
Calibrate T1/Calibrate T2.
Protection ON in protect mode.
3.4.3 Invasive pressure calibration
Calibrate invasive pressure when the pressure transducer (probe) is replaced with a different type of transducer, and when STP board is replaced.
1. Enter ESTPR: the STP service menu.
(
Monitor Setup, Install/Service (password 16-4-34), Service (password 26-23-8),
Parameters).
2. Enter
3. Connect a pressure transducer with a pressure manometer to the P1/P2 connector. Choose
Calibrations menu.
Calibrate P1 or Calibrate P2. Leave the transducer to room air pressure.
42
Document No. 8005571-1
4. Press the ComWheel to start zeroing.
5. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
6. Set the pressure on the display to match the pressure reading on the manometer and press
the ComWheel. A tolerance of ±1 mmHg is allowed.
7. The message ‘Calibrated’ will appear on the display.
Page 49
S/5 M-PRESTN Modules
4 TROUBLESHOOTING
4.1 Troubleshooting charts
See also the User’s Reference Manual for more troubleshooting procedures.
4.1.1 NIBP
TROUBLE CAUSE TREATMENT
No NIBP value displayed NIBP not selected on screen. Check monitor setup.
NIBP menu fading No M-PRESTN module, module not
properly connected, or NIBP and PRESTN module connected at the same time.
Artifacts-message Unsuccessful measurement due to
patient movement, shivering, external artifact or weak signal.
Weak pulsation-message Weak or unstable oscillation pulses
due to:
artifacts
weak pulse pressure due to
arrhythmias
improper cuff position or attachment
too few pulses detected
weak or unusual blood
circulation
obese subject
Call service Error X-message
NIBP hardware error. X = error number.
Plug in the module.
Check patient condition and retry.
Check any leaks and retry.
Use proper size of cuff. Check attachment.
See the description of the error message code in solutions listed in the next chapter.
4.1.2, the causes and the
43
Document No. 8005571-1
Page 50
Datex-Ohmeda S/5 monitors
TROUBLE CAUSE TREATMENT
Cuff loose-message 1. Hose and/or cuff not connected. 1. Connect the hose and the cuff.
2. Hose and cuff connected. Reasons:
cuff loosely wrapped tighten the cuff
leakage in cuff or hose replace cuff/hose
leakage inside module check internal tubing and air
chamber, and fix if necessary
pump does not work check pump connector; if OK, replace
pump
Air leakage-message 1. Hose or cuff leaking. Reasons: 1. Replace cuff
cuff damaged replace cuff
cuff connector damaged replace cuff connector (if the fault is
in hose connector,)
O-ring damaged or missing replace O-ring
hose double connector
replace hose
damaged
2. Hose and cuff OK. Reasons: 2. Connect or replace tube
leakage inside the module replace the whole tubing
tube disconnected or damaged fix connections
tubes or valve(s) damaged replace tubes/valve(s)
Unable to measure Sys­message
Systolic blood pressure probably higher than the inflation pressure or
Automatic retrial with increased pressure.
artifacts.
44
Document No. 8005571-1
Page 51
S/5 M-PRESTN Modules
TROUBLE CAUSE TREATMENT
Cuff occlusion-message 1. Cuff and/or hose occluded.
Reason:
cuff tube kinked straighten tube
tube inside module kinked straighten tube
occlusion inside/outside
remove occlusion
module
2. Cuff, hose, and tubes OK. Reason:
fault in pressure transducer replace the NIBP board
fault in A/D converter replace the NIBP board
faulty calibration check calibration
Calibration switch on – message
EEPROM protection has been handled by pressing module buttons
Auto ON/OFF and Start
Enables setting the protection OFF in the Calibration menu. Press the buttons again if you are not going to calibrate.
Cancel simultaneously for 3
seconds.
Calibration not protected ­message.
Calibration protection is set to OFF.
Set the protection ON in the NIBP Calibration menu.
45
Document No. 8005571-1
Page 52
Datex-Ohmeda S/5 monitors
4.1.2 NIBP error code explanation
Code Explanation Treatment
0 RAM failure; memory failure Change the NIBP board.
1 ROM checksum error; memory failure Change the NIBP board.
2 Pump current failure Check short circuits. Change the NIBP
board.
3 Safety CPU internal test failure or pressure sensor
reference voltage failure
4 EEPROM protection error
5 Calibration not protected Protect calibration by selecting Protection
6 Pressure sensors give different readings Try to remeasure, if the problem persists
7 Calibration failure Reset module and recalibrate. If this does
8 Exhaust Valve occlusion Check tubing. If this does not help,
9 Measurement related SW error Automatic recovery.
10 EEPROM checksum error; memory failure Change the NIBP board.
11 Auto zero range exceeded Calibrate the NIBP.
12 Communication break; temporal break down of
communication from monitor detected
Change the NIBP board.
Press module buttons Start
Cancel simultaneously for 3
seconds.
ON in the NIBP calibration menu.
recalibrate. If the problem still persists change the NIPB board.
not help, change the NIBP board.
change the NIBP board.
Automatic recovery.
Auto ON/OFF and
13 Illegal neonate cuff with identifying magnet
14 -
15 Safety CPU pressure calibration error Recalibrate. If this does not help, change
16 Communication error between CPUs Change NIBP board.
17 Safety CPU has cut down power from pneumatics
46
Document No. 8005571-1
connected
due to repeating safety limit violations
Remove the cuff.
NIBP board.
Reset module. If problem persists change NIBP board.
Page 53
S/5 M-PRESTN Modules
4.1.3 ECG
TROUBLE CAUSE TREATMENT
HR numerical display shows ‘---’
If ECG waveform exists, check heart
Unacceptable ECG waveform Poor electrode or poor electrode
Poor electrode condition. Electrodes are dried out.
Improper site of electrodes. Check that electrodes are not placed
Improper skin preparation. Remove body hair. Clean attachment
Improper bandwidth filter. Check filter.
No ECG trace Waveform not selected on screen.
Module not plugged in correctly. Plug in.
No heart rate available. If no ECG waveform, check LEADS OFF
message and connect the leads.
rate source e.g. in the ECG Setup menu behind ECG key.
Electrodes from different
skin contact.
manufacturers are used. /Too much/little gel is used.
over bones, active muscles, or layers of fat.
site carefully with alcohol.
Press the make adjustments.
Monitor Setup key and
Noise-message High frequency or 50/60 Hz noise. Isolate noise source.
4.1.4 Pulse oximetry (SpO2)
TROUBLE CAUSE TREATMENT
Message ‘NO PROBE’ No probe connected to the
monitor.
Probe faulty. Change the probe.
Message ‘PROBE OFF’ though probe properly attached to the patient
Probe faulty. Try another probe.
Probe connection cable not
Finger probe falls off 1. Probe is slippery. 1. Wipe with 70 % isopropyl alcohol
2. Finger is too thin or thick. 2. Try other fingers, or other probe
Unsuitable site. Try another site.
connected to probe.
Check probe connections.
Connect the cable to probe.
and allow drying.
types.
Weak signal artifacts Poor perfusion. Try another place.
47
Document No. 8005571-1
Page 54
Datex-Ohmeda S/5 monitors
TROUBLE CAUSE TREATMENT
Movement artifacts.
Shivering.
Message ‘NO PULSE’ Pulse search > 20 sec. and low
SpO
or low pulse rate.
2
Message ‘ARTIFACT’ Pulse modulation exceeds the
present scale.
Message ‘CHECK PROBE’ DC value not in balance. Try another probe.
Message ‘POOR SIGNAL’ Poor perfusion.
Modulation (Red or Ired) < 0.25 %
Message ‘FAULTY PROBE’ Probe is faulty. Change the probe.
No SpO2 No waveform selected on screen. Check selected SpO2 waveforms by
Wrong configuration setting. Check the configuration settings from
Try other fingers.
Try another place or another probe.
Check that the sensor is positioned correctly to the patient.
pressing selecting
the ESTPR:STP/Calibrations menu (
Monitor Setup - Install/Service -
Service - Parameters)
Monitor Setup key and
Modify waveforms.
4.1.5 Temperature
TROUBLE CAUSE TREATMENT
Message ‘TEMPERATURE ERROR’ Faulty calibration. Perform calibration. If it does not
help, check that front panel connector is properly connected to STP board.
No temperature displayed Wrong type of probe. Use correct probe.
Temperature out of measurable
range.
Temperature calibration not
protected.
The range is between 10 and 45 °C.
Set the protection ON in the Service Menu.
48
Document No. 8005571-1
Page 55
S/5 M-PRESTN Modules
4.1.6 Invasive blood pressure
TROUBLE CAUSE TREATMENT
Abnormally low pressure Transducer wrongly positioned. Check mid-heart level and reposition
transducer.
No pressure Defective transducer. Check transducer.
No pressure module plugged in. Check the module.
No waveform selected on screen. Check selected pressure waveforms
by pressing Monitor Setup key and selecting modify waveforms.
Check that pressure transducer is
open to patient.
Wrong configuration setting Check the configuration setting from
the ESTP:STP/Calibrations menu (
Monitor Setup - Install/Service -
Service - Parameters
)
Not zeroed -message Measurement on, channel not
zeroed.
Zeroing failed -message Unsuccessful zeroing of P1 /P2
(number field).
Offset is > 150 mmHg. Open the
Defective transducer. Replace it and
Calibration failed -message Unsuccessful calibrating of
P1/P2 (number field), possibly due to pulsating waveform
Out of range < 40 mmHg Measurement pressure is beyond
measurement range.
Out of range > 320 mmHg Measurement pressure is beyond
measurement range.
Zero the channel.
Possibly due to pulsating pressure waveform. Open the transducer to air and zero the channel.
transducer to air and zero the channel.
zero the channel.
Turn the transducer to sphygmomanometer and try again (zeroing takes place first).
Gain is beyond the limits (± 20 % of the default gain). Replace the transducer.
Check transducer level. Zero the channel.
Check transducer level. Zero the channel. The patient may also have high pressure.
Zero adj. > 100 mmHg Offset when zeroing is > 100
mmHg (but < 150 mmHg) from the absolute zero of the module (with default gain).
Check transducer. The waveform may hit the top and the numeric display not shown.
49
Document No. 8005571-1
Page 56
Datex-Ohmeda S/5 monitors
TROUBLE CAUSE TREATMENT
Out of range Measured pressure is beyond the
internal measurement range of the module.
The waveform hits the top and the numeric display not shown. Check transducer and its level. Zero the channel.
4.1.7 Impedance respiration
TROUBLE CAUSE TREATMENT
No resp trace Waveform not selected on the
screen
Module not plugged in correctly Re-plug the module.
Unacceptable resp waveform Poor electrode or poor electrode
skin contact
Poor electrode condition Electrodes are dried out.
Improper site of electrodes Check that electrodes are not placed
Improper skin preparation Remove body hair. Clean
Press the Monitor Setup key and make adjustments.
Electrodes from different manufacturers are used. Too much/little gel is used.
over bones, active muscles, or layers of fat.
attachment site carefully with alcohol.
Message: ‘SMALL RESP CURVE’ Respiration signal is very small With 3-lead cable in ESTPR/NESTPR
try another lead connection I, II, III or try 5-lead cable.
Message: ‘APNEA ALARM’, and respiration waveform normal
Respiration source is CO2 Check respiration source and
change it to correct one.
50
Document No. 8005571-1
Page 57
4.2 Troubleshooting flowcharts
4.2.1 M-PRESTN module troubleshooting for NIBP parameter
Possible faulty
START
module with NIBP
parameter.
S/5 M-PRESTN Modules
Insert NXXX module
and turn power on.
Does fault
still appear?
YES
Module ID
on screen?
YES
Error
message on
screen?
YES
See error code explanation in Service Manual and
fix it.
NO
NO
NO
Remove all
modules from
Monitor Frame.
Fault not in
NXXX module.
NO
Does another module work in same
slot?
Key check
in Service Menu
OK?
YES
YES
NO
Check and
replace
NIBP board.
Check keyboard
connector. If OK,
change front panel.
Replace
NXXX module.
NO
OK?
Check and
replace
NIBP board.
NO
OK?
Do
they work?
NO
Check tubes and
connectors. Find leak and
fix it.
Figure 18 M-PRESTN module troubleshooting flowchart for NIBP Parameter
Pump check
in Service Menu
OK?
Leak test
in Service Menu
OK?
NO
YES
NO
YES
Check pump
connector. If OK,
change pump.
NESTPR_module_troubles_flowch.vsd
51
Document No. 8005571-1
Page 58
Datex-Ohmeda S/5 monitors
4.2.2 M-PRESTN module troubleshooting for parameters ESTPR
Possibly fault in module with
ESTPR parameters.
Check module c onfiguration in Service menu (ESTPR/STP/Calibrations)
Is it
correct?
YES Remove all modules from
the frame
Insert ESTPR module and turn power on
Does fault
still appear?
YES
Enter the
Service menu
STP module ID on screen?
YES
NO
NO
NO
Select the correct configuration
Fault not in ESTPR module
NO
Does
another
modu le work
in same
slot?
YES
Remove
ECG board
Replace
STP board
NO
STP module ID on screen?
YES
ECG module ID on screen?
YES Check front panel key functions in Service menu
Do
they wo rk?
NO
Check front panel unit, ECG board, and STP board to find culprit. Replace if necessary.
Figure 19 M-PRESTN Module Troubleshooting Flowchart for Parameters ESTPR
52
Document No. 8005571-1
NO
YES
Has ECG
measurement
failed?
YES
Replace ECG board
NO
board failed?
Has STP
YES
Replace STP board
NO
Replace ECG board
Has inv BP
measurement
failed?
YES
Replace STP board
NO
Has Temp
measurement
failed?
YES
Repla ce STP board
NO
measurement
Has SpO
failed?
YES
Repla ce STP board
NESTPR_MESTPR_module_trou
bles_flowch.vsd
2
Page 59
5 SERVICE MENU
S/5 M-PRESTN Modules
NOTE: Parameter values in Service Data fields are for reference only on this chapter.
1. Press the
2. Select
3. Select
4. Select
Monitor Setup key.
Install/Service (password 16-4-34).
Service (password 26-23-8).
Parameters - NIBP.
53
Document No. 8005571-1
Page 60
Datex-Ohmeda S/5 monitors
5.1 NIBP service menu
Service Data
Pressure shows measured pressure multiplied by 10. This value is automatically zero-drift
compensated.
Zero shows difference between zeroing value in the permanent memory (stored when module is
calibrated) and the current automatic zero-drift compensation multiplied by 10. The value can change between +20 and -20 mmHg. If the zero drift exceeds ± 10 mmHg, the module should be recalibrated.
Protect handle indicates hardware protection for EEPROM memory. It should be ON all the time in
normal operation. If it is OFF data can not be read from or written to EEPROM, only the calibration protection can be set or reset by software. It can be turned to OFF by pressing NIBP module buttons
Auto ON/OFF and Start Cancel simultaneously for 3 seconds, which also enables Protection ON/OFF
Calibr. prot. shows software calibration protection and it should be OFF to enable calibration.
+15 V power refers to legacy NIBP modules. Not used in M-PRESTN/M-RESTN/M-PRETN.
AD0 to AD7 show the values of each eight channels of A/D converter. AD7 is not used in M-
PRESTN/M-RESTN/M-PRETN.
Timeouts
the monitor’s inquiry. communication from the module to monitor broke down.
Bad c-s by mod is a cumulative number that indicates how many communication errors the module
has detected. The monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) indicates either serial communication failure, or module not in place. Also other modules can cause communication errors that cause these numbers rise.
menu selection in the calibration menu.
is a cumulative number that indicates how many times the module has not responded to
Bad checksums is a cumulative number that indicates how many times
54
Document No. 8005571-1
RAM indicates the state of the RAM memory. ROM indicates whether the checksum in the EPROM is in accordance with the one the software has
Page 61
calculated.
EEPROM indicates if the values stored in the permanent memory are valid.
The state is either
5.1.2 NIBP demo menu
S/5 M-PRESTN Modules
OK, Fail or ? (module not in place or a communication error).
A service menu for demonstrating the oscillometric method of NIBP measurement. The menu shows the real-time pressure signals that are measured from the NIBP cuff. The measurement result is shown in the adjoining digit field.
Wave Recording Wave Recording is for selecting the recording option. If ON is
selected, the pressure signals are recorded in real-time onto the M-
Remove menu
REC paper.
Remove menu widens the displayed waveform area.
Previous Menu The menu can be closed by selecting the Previous Menu or just by
pressing the ComWheel if the
Remove menu was selected.
55
Document No. 8005571-1
Page 62
Datex-Ohmeda S/5 monitors
5.1.3 NIBP calibration menu
Active Leak Test Wrap an adult cuff around a pipe and connect the cuff to the module. Select the active leak test
(ON). The module automatically pumps a pressure of 260 mmHg into the cuff. Wait for several seconds until the pressure stabilizes. Then check that the pressure reading does not drop more than 6 mmHg per minute. If it does, leaking point(s) should be detected and fixed. Cancel the test by selecting Active leak test OFF.
Calibration Check
After the calibration check is selected (ON), the module zeroes the pressure transducers at the beginning of the calibration check. Do not pump pressure until the text ‘Calibrating’ appears to the NIBP digit field or the zeroing will fail. After zeroing is done, manually pump pressure into the module and make sure that the same pressure values are shown both on the display and on the manometer. Pressure of both pressure channels B1 and B2 are shown. Note difference to the legacy modules with NIBP. The pressure values are automatically zero-compensated, so the
Protection
readings of B1 and B2 should be the same as the manometer readings.
Software calibration protection (ON/OFF). Select OFF when calibrating. Protection selection becomes available in the menu after pressing NIBP module buttons
Cancel
simultaneously for 3 seconds.
Auto ON/OFF and Start
Calibration Calibration selection is available only when protection is OFF and module buttons Auto ON/OFF
and Start Cancel are pressed simultaneously for 3 seconds.
NIBP calibration can be performed in the NIBP Service menu as follows:
NOTE: Both channels B1 and B2 must be calibrated simultaneously.
1. If
Protection is ON change it to OFF by pressing NIBP module buttons Auto ON/OFF and
Start Cancel simultaneously for 3 seconds, which enables the Protection selection. Then
press the buttons again for 3 seconds to enable
Calibration.
56
Document No. 8005571-1
NOTE: The module must be in the frame during the whole procedure.
NOTE: When the buttons have been pressed, the NIBP field shows an error message ‘Calibration switch on!’.
NOTE: When calibration is enabled, a message ‘Calibration not protected’ appears.
Page 63
S/5 M-PRESTN Modules
2. For proper zeroing to take place, remove the hose from the front panel connector. Select
Calibration and push the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in the NIBP
message field. After this a pressure bar will appear beside the menu and the text ‘Calibrating’ will appear to NIBP digit field.
3. Connect an external mercury manometer with pump to module through the both tubes of the hose. Pump up to about 200 mmHg pressure (range of 150 to 300 mmHg allowed) according to the manometer. Verify that both pressure values in the prompt field match the manometer reading. If not, adjust by turning the ComWheel.
4. When the values are equal, push the ComWheel to confirm the calibration. After a few seconds ‘Calibrated’ message will appear to the NIBP digit field, which means that the calibration data has now been saved.
NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the ComWheel, even in cases where the value would be correct. For example change the value one step higher and then back one step lower. “Calibrated” text should appear in the display. This ensures that the calibration procedure is correctly registered and stored by the module.
5. Use the module buttons again to enable
Protection setting.
5.1.4 NIBP safety valve menu
Protection setting and set it ON, and finally disable
Start Test Start test is for starting and Stop test is for stopping the Safety Valve test.
Safety Valve Data
See NIBP Service menu in chapter 5.1 for information on general items Pressure, Zero, Protect
handle, Calibr. prot., +15 V power, AD0
Max. press and 2 s after stop show the measured values at Safety Valve test.
Safety Valve Test Adult/Infant
Wrap an adult cuff around a pipe and connect the cuff to the module. Highlight Start test and give the ComWheel a push. The test ends automatically or when
test
) is pushed.
to AD7 as well as Timeouts etc.
Stop test (appears in place of Start
57
Document No. 8005571-1
Page 64
Datex-Ohmeda S/5 monitors
Max. press indicates the pressure at which the safety valve opens and is normally 300 ±15 mmHg
for adult and 150 mmHg ±15 mmHg for infant. after the pump has stopped and is normally > 270 mmHg for adult and > 130 mmHg for infant. If the value is less, check leakage by the active leak test.
5.1.5 NIBP pulse valve menu
2 s after stop indicates the pressure at 2 seconds
Start Test Start test is for starting and Stop test is for stopping the test.
Set Valve Set Valve
lets you adjust the opening of the pulse valve.
Pulse Valve Data
See NIBP Service menu in chapter 5.1 for information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0
Pulse Valve Checking
Wrap an adult cuff around a pipe and connect the cuff to the module. Select the Start test and push the ComWheel. The pressure rises beyond 240 mmHg and stops. The pulse valve opens. The module counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg and displays it on the screen. The test can be manually stopped by selecting
The valve can be adjusted between 0 and 255 (0 for fully closed and 255 for fully open). First select Set Valve and push the ComWheel. See the pulse valve value and adjust it by turning the ComWheel. Then push the ComWheel to confirm the value.
The ‘ ‘150’ and less than 10 when fully opened (255). When fully closed (0), the system should be airtight and the pressure does not drop. Depending on an individual, the pulse valve may remain closed up to approx. value 100.
Interval 240 mmHg -> 50 mmHg’ time should be less than 60 seconds when the valve is
to AD7 as well as Timeouts etc.
Stop test.
58
Document No. 8005571-1
If the measured time deviates much from those above, then the pulse valve or its tubes are faulty.
Page 65
5.1.6 NIBP buttons/leds menu
S/5 M-PRESTN Modules
The selections effect on the module.
Buttons/Leds Data
See NIBP Service menu in chapter 5.1 for information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc.
Buttons Checking
The front panel keys function is confirmed by pressing and releasing the key and observing OFF turns to ON at Auto On/Off, and Start Cancel.
Auto ON/OFF, Manual ON/OFF, STAT ON/OFF, and Measur. ON/OFF have no
59
Document No. 8005571-1
Page 66
Datex-Ohmeda S/5 monitors
5.1.7 NIBP pneumatics menu
Start Pump/Stop Pump
A manual control for the pump. The selection changes to
Open Exh1/Close Exh1
Open Exh2/Close Exh2
Open Zero valve
Set Valve With Set Valve, the opening of the pulse valve is adjusted between 0 and 255 (0 for fully closed
Reset Clock Reset Clock will zero the time on the display.
No effect on the module.
A manual control for the exhaust valve 2. The selection changes to opened.
No effect on the module.
and 255 for fully open). First push the ComWheel, then turn it to adjust the value on screen and finally push to set the value.
Stop Pump when the pump turns on.
Close Exh2 when the valve is
Pneumatics Data field
See NIBP service menu in chapter 5.1 for information on general items Pressure, Zero, Protect
handle, Calibr. prot., +15 V power, AD0
Pump, Exh1 Valve, and Exh2 Valve show their states. Note that Exh1 Valve has no effect to the
module.
to AD7 as well as Timeouts etc.
60
Document No. 8005571-1
Pulse Valve shows how much the valve is opened (0 to 255) during Valve Setting.
Interval 20 mmHg -> 185 mmHg Checking
Select the Start pump at different combinations of the valves open/closed and push the ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to 185 mmHg and displays it. When all the valves are closed, the pump should be able to pump the pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does not stop without selecting the
Stop Pump by pushing the ComWheel.
Page 67
5.1.8 NIBP watchdog menu
S/5 M-PRESTN Modules
Test ADULT No effect on the module.
Test INFANT No effect on the module.
Stop Test No effect on the module.
Watchdog Data field
See NIBP Service menu in chapter 5.1 for information on general items Pressure, Zero, Protect
handle,
Watchdog Interval
Adult watchdog time testing
No effect on the module.
Infant watchdog time testing
No effect on the module.
Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc.
no effect on the module.
61
Document No. 8005571-1
Page 68
Datex-Ohmeda S/5 monitors
5.1.9 ECG service menu
Power freq Set power frequency; 50 Hz/60 Hz.
Filter low Set filter low frequency; 0.05 Hz/0.5 Hz.
Filter high Set filter high frequency; 30 Hz (40 Hz if power freq is 60 Hz) /100 Hz or 150 Hz @ NE12STPR.
Service Data field
Power freq, and Cable type show the values chosen or detected, Filter low and high defines the
selected filter (Monitor/Diagnostic/ST).
Quick zero
returned to optimal range using fast signal processing methods. All the the same time.
Cable shows ON when ECG cable is connected.
Electrode shows ON when each of these electrodes are connected. Pacer count is a running number for pacemaker users.
Button No effect on the module.
Resp Available indicates that ECG hardware is capable of measuring impedance respiration.
Measurement shows ON when the respiration measurement is on.
at PRESTN module is ON when the ECG signal is beyond scale, and therefore, is quickly
Quick zero bits are ON at
62
Document No. 8005571-1
Amp zero shows ON when zeroing of the respiration amplifier takes place.
Waveform
Timeouts
the monitor’s inquiry. communication from the module to monitor broke down.
Bad c-s by mod is a cumulative number that indicates how many communication errors the module
has detected.
VALUE will be updated in one second interval.
is a cumulative number that indicates how many times the module has not responded to
Bad checksums is a cumulative number that indicates how many times
Page 69
S/5 M-PRESTN Modules
The monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) indicates either serial communication failure, or module not in place. Also other modules can cause communication errors that cause these numbers rise.
RAM indicates the state of the RAM memory. ROM indicates whether the checksum at the EPROM is in accordance with the one the software has
calculated.
EEPROM indicates if the values stored in the permanent memory are valid.
The state is either
OK, Fail or ? (module not in place or a communication error).
63
Document No. 8005571-1
Page 70
Datex-Ohmeda S/5 monitors
5.1.10 ECG setup menu
Filter Filters the ECG signal high frequency noise and slow respiratory artefacts.
Monit (monitor) filter is used in routine monitoring. It effectively filters the artefacts caused by the
electrosurgery unit and respiration.
Diagn (diagnostic) filter is used if more accurate information of the waveform is needed (e.g., of P-
wave or AV block). The diagnostic filter is more susceptible to both high frequencies and baseline wander than monitor filter.
STfilt (ST filter) permits more accurate information of ST segment. It filters the high frequency
artefacts caused by electrosurgery unit but catches the slow changes in ST segment. The ST filter is more susceptible to baseline wander than the monitor filter.
Pacemaker Selects how to display the pacing pulse of cardiac pacemaker. The selections are Show, Hide, ON
R
and Sensit.
Hide, the pacing pulse is filtered away from ECG data.
Show, the pacer pulse is filtered away from ECG data but the pulse is displayed as a constant
height marker.
ON R, pacing pulses are not filtered away from ECG data. This improves ECG monitoring with A-V
pacemaker patients, as QRS complexes are counted even if the pacing pulse hits the QRS complex. However, during asystole the monitor may count pacing pulses as heart beats.
Sensit selection uses a more sensitive pacemaker detection. Pacemaker spike is displayed on
ECG.
64
Document No. 8005571-1
Page 71
5.2 STP service menu
S/5 M-PRESTN Modules
Record Data Record Data prints out the shown service data and board information (id, serial number and sw id)
onto the recorder module, M-REC.
Temp Test Temp Test activates the automatic temperature test for the temperature channels T1 and T2. The
result from the test is shown in the service data field.
NOTE: The Temp Test needs to be selected twice before the test starts.
Service Data field
Gain is a coefficient to compensate gain error. Usually the values for P1 and P2 are between
17000 and 25000 and for T1 and T2 between 13000 and 14300. compensation value of each parameter in A/D converter. Typically the values for P1 and P2 are within ±1000 and for T1 and T2 between -150 and +300. Calibrate if zero and/or gain value is outside the ranges.
shows ON when a corresponding cable is connected to the front panel and Probe shows ON
Cable
when a corresponding probe is connected to the cable.
Under
Value the measured numeric values are displayed simultaneously. Pressure values are real
time values and shown in mmHg. Temperature values are shown in degrees Celsius.
The front panel STP keys functions are confirmed by pressing each key and observing OFF turns to
Button.
ON at
SpO2 shows measured beat-to-beat SpO
ratio in the measured signal.
Hr is a pulse rate calculated from every beat.
value. Modpr is a modulation % that indicates AC/DC
2
Cable
and Probe can be either OFF or ON, and these indicate the state PROBE OFF.
Under them there is a message field for SpO
. It can be OK, PULSE SEARCH, NO PROBE, PROBE
2
OFF, NO PULSE, ARTEFACT, POOR SIGNAL, or CHECK PROBE.
Zero indicates offset
Balance between leds is adjusted by changing the intensity of red/infrared. Intensity of infrared
Ired int.) is in the range of 40 to 255 and red intensity (red int.)is in the range of 40 to 255.
(
DC gain shows the gain of DC signal adjusted by the module.
Document No. 8005571-1
65
Page 72
Datex-Ohmeda S/5 monitors
IDC is the value of infrared signal.
RDC is the dc value of red signal.
AC gain is the gain of infrared and red ac signals. AC gain values can be 1 or 0. Value 1 means high
ac gain and 0 means low gain.
Pre gain is a preamplifier gain for infrared and red signals. Pre gain values can be 1 or 0. Value 1
means normal operation. Value 0 means that signal levels are very low and extra gain is taken into use.
Temp error shows the status of the temperature test. No errors found show the status (OFF) and
errors found (ON).
Protect key
pressed.
Protect mode
calibration in Calibration Menu.
Configuration
Timeouts is a cumulative number that indicates how many times the module has not responded to
the monitor’s inquiry. communication from the module to monitor broke down.
Bad c-s by mod is a cumulative number that indicates how many communication errors the module
has detected. The monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) indicates either serial communication failure, or module not in place. Also other modules can cause communication errors that cause these numbers rise.
shows normally OFF but turns to ON when the button at the bottom of the module is
is normally ON. It turns to OFF when Protect is switched to OFF for the temperature
shows the chosen module configuration: TP, ST, or STP.
Bad checksums is a cumulative number that indicates how many times
RAM indicates the state of the RAM memory. ROM indicates whether the checksum at the EPROM is in accordance with the one the software has
calculated.
EEPROM indicates if the values stored in the permanent memory are valid.
The state is either
OK, Fail or ? (module not in place or a communication error).
66
Document No. 8005571-1
Page 73
5.2.2 STP calibration menu
S/5 M-PRESTN Modules
Protection Protection for the configuration and temperature calibrations can be set ON and OFF .
Set Config The module configuration should be set according to the module type. The setting is possible only
when the protection is set OFF. The available selections are TP, ST or STP. The configuration setting should be checked if the STP board is replaced.
Calibrate T1 / Calibrate T2
The functions are for calibrating the temperature channels T1 and T2. The calibrations are possible only when the protection is set OFF. The temperature calibration requires accurate test plugs of value 25 °C and 45 °C.
Calibration:
1. Select
Calibrate T1/Calibrate T2
2. Insert the test plug 25 °C into the T1/T2 connector
3. Press the ComWheel
4. Insert the test plug 45 °C into the T1/T2 connector
5. Press the ComWheel
Calibrate P1/Calibrate P2
The functions are for calibrating the invasive blood pressure channels P1 and P2. The calibrations require a pressure transducer (with an appropriate cable) and a pressure manometer.
1. Connect the pressure transducer with the pressure manometer to the P1/P2 connector.
Select
Calibrate P1/Calibrate P2. Leave the transducer to room air pressure.
2. Press the ComWheel to start zeroing.
3. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
4. Set the pressure on the display to match the pressure reading on the manometer and press
the ComWheel.
67
Document No. 8005571-1
Page 74
Datex-Ohmeda S/5 monitors
6 SPARE PARTS
6.1 Spare parts list
6.1.1 M-PRESTN rev. 01, M-RESTN rev. 01, M-PRETN rev. 01
Figure 20 Exploded view of M-PRESTN Module
Item Description Order No. Replaced by
1 Module box (wide) 886168
2 Spring pin 879182
3 Latch for module box 879181
4 ECG_120 board M1008909
5 ECG Input board holder M1021216
6 STP_120 board M1006153
7 NIBP unit M1012677
8 Air filter (NIBP) 57142
9 Silicon tube NIBP (2pcs, 100mm) 73375
10 Membrane keypad 8005215
11 Mask including NIBP flex board and ECG connector unit M1018978
12 NIBP Cuff connector 64654
68
Document No. 8005571-1
Page 75
S/5 M-PRESTN Modules
Item Description Order No. Replaced by
13 NIBP Mounting nut 8004036
14 TEMPERATURE input board 8005216
15 Screw; pan-head, TORX PT 3x10mm (7 pcs) 628728
16 PRESS / SpO2 input board, M-PRESTN 8005219
16 PRESS input board, M-PRETN 8005492
16 SpO2 input board, M-RESTN 8005490
17 Screw; Cross head recess PT 2.5x10 (6kpl) M1021234
18 Fitting plate (only M-RESTN, M-PRETN) 879510
19 STP cable (34 contacts, female) M1001763
20 Cable holder M1008749
21 Board holder M1021283
22 Module bus cable 8004009
23 Cable tie 64001
24 Screw; cross head recess M3x8 black (2 pcs) 616215
25 Screw; cross head cylinder M3x6 ( 6 pcs) 61721
26 Front panel, DA, M-PRESTN M1021293
26 Front panel, DE, M-PRESTN M1021294
26 Front panel, EN, M-PRESTN M1021440
26 Front panel, ES, M-PRESTN M1021295
26 Front panel, FI, M-PRESTN M1021296
26 Front panel, FR, M-PRESTN M1021297
26 Front panel, IT, M-PRESTN M1021298
26 Front panel, JA, M-PRESTN M1021299
26 Front panel, NL, M-PRESTN M1021300
26 Front panel, NO, M-PRESTN M1021301
26 Front panel, PL, M-PRESTN M1021302
26 Front panel, PT, M-PRESTN M1021303
26 Front panel, SV, M-PRESTN M1021305
26 Front panel, DA, M-PRETN M1021423
26 Front panel, DE, M-PRETN M1021424
26 Front panel, EN, M-PRETN M1021422
26 Front panel, ES, M-PRETN M1021425
26 Front panel, FI, M-PRETN M1021426
26 Front panel, FR, M-PRETN M1021427
26 Front panel, IT, M-PRETN M1021428
26 Front panel, JA, M-PRETN M1021429
26 Front panel, NL, M-PRETN M1021430
26 Front panel, NO, M-PRETN M1021431
26 Front panel, PL, M-PRETN M1021432
26 Front panel, PT, M-PRETN M1021433
26 Front panel, SV, M-PRETN M1021434
26 Front panel, DA, M-RESTN M1021441
26 Front panel, DE, M-RESTN M1021443
26 Front panel, EN, M-RESTN M1021440
26 Front panel, ES, M-RESTN M1021444
26 Front panel, FI, M-RESTN M1021445
26 Front panel, FR, M-RESTN M1021446
26 Front panel, IT, M-RESTN M1021447
26 Front panel, JA, M-RESTN M1021448
26 Front panel, NL, M-RESTN M1021449
26 Front panel, NO, M-RESTN M1041450
69
Document No. 8005571-1
Page 76
Datex-Ohmeda S/5 monitors
Item Description Order No. Replaced by
26 Front panel, PL, M-RESTN M1021451
26 Front panel, PT, M-RESTN M1021452
26 Front panel, SV, M-RESTN M1021453
70
Document No. 8005571-1
Page 77
7 EARLIER REVISIONS
No earlier revisions.
S/5 M-PRESTN Modules
71
Document No. 8005571-1
Page 78
Datex-Ohmeda S/5 monitors
For your notes:
72
Document No. 8005571-1
Page 79
S/5 M-PRESTN Modules

APPENDIX A

73
Document No. 8005571-1
Page 80
Datex-Ohmeda S/5 monitors
74
Document No. 8005571-1
Page 81
APPENDIX A, Service check form, S/5 M-PRESTN Modules

SERVICE CHECK FORM

DATEX-OHMEDA M-PRESTN, M-RESTN, M-PRETN MODULES
Customer
Service
Service engineer
Module type S/N
Date
OK = Test OK N.A. = Test not applicable Fail = Test Failed
All modules OK N.A. Fail OK N.A. Fail
1. Internal parts
3. NIBP pump filter
5. Recognition
Notes
2. External parts
4. Installation
ECG measurement
S/N
6. Module software (serial numbers)
ECG/RESP
STP
NIBP
OK N.A. Fail OK N.A. Fail
7. Communication and memories
9. Cable recognition
11. Test with patient simulator
Notes
8. Power frequency
10. Lead detection
A- 1(4)
Document No. 8005571-1
Page 82
Datex-Ohmeda S/5 monitors
RESP measurement
S/N
OK N.A. Fail OK N.A. Fail
12. RESP measurement recognition
Notes
13. Test with patient simulator
TEMP measurement
S/N
OK N.A. Fail OK N.A. Fail
14. Communication and memories
16. Calibration check
18. Configuration STP/ST/TP
Notes
15. Temperature probe detection
17. Temp test -function
InvBP measurement
S/N
OK N.A. Fail OK N.A. Fail
19. Membrane keys
21. Calibration
Notes
20. Cable and transducer detection
22. Test with patient simulator
SpO
measurement
2
S/N
OK N.A. Fail OK N.A. Fail
23. SpO2 probe detection
Notes
24. Test measurement
A- 2(4)
Document No. 8005571-1
Page 83
APPENDIX A, Service check form, S/5 M-PRESTN Modules
NIBP measurement
S/N
OK N.A. Fail OK N.A. Fail
25. Communication and memories
27. Pump and valves
28. Leak test
26. Membrane keys
28. 6 mmHg/min
29. Calibration check Measured B1 Measured B2 Allowed range
0 mmHg
100 mmHg
200 mmHg
260 mmHg
± 9 mmHg
100 ± 2 mmHg
200 ± 3 mmHg
260 ± 4 mmHg
30. Safety valve functions
B1 B2 Allowed range
‘Max press’ ADULT 270 to 330 mmHg
‘2 s after stop’ ADULT 270 to 330 mmHg
‘Max press’ INFANT 130 to 165 mmHg
‘2 s after stop’ INFANT 130 to 165 mmHg
OK N.A. Fail OK N.A. Fail
31. Cuff related messages
33. Test measurement
32. Adult cuff detection
34. Infant cuff detection
Notes
A- 3(4)
Document No. 8005571-1
Page 84
Datex-Ohmeda S/5 monitors
All modules
OK N.A. Fail OK N.A. Fail
35. Electrical safety check
37. Final cleaning
Notes
Notes
Used Spare Parts
36. Functioning after electrical safety check
Signature
A- 4(4)
Document No. 8005571-1
Loading...