Datex-Ohmeda S-5 Service Manual Datex S5

S/5
S/5 S/5
Datex-Ohmeda
Hemodynamic Modules
TM
PRESTN Module, M-PRESTN (Rev. 01)
TM
RESTN Module, M-RESTN (Rev. 01)
TM
PRETN Module, M-PRETN (Rev. 01)
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licenced practitioner.
Datex-Ohmeda, Inc.
P.O. Box 7550, Madison WI 53707-7550, USA Tel: 1-608-221-1551 Fax: 1-608-222-9147
www.us.datex-ohmeda.com
mailto:product.support.ussub@us.datex-ohmeda.com

All specifications are subject to change without notice.

Outside the USA, check local laws for any restriction that may apply.

Document No. 8005571-1

Datex-Ohmeda Division, Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel: +358 10 394 11 Fax: +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.

April 2004

Table of contents

TABLE OF CONTENTS
S/5 Hemodynamic Modules
TABLE OF CONTENTS i
TABLE OF FIGURES iii
Introduction 1
1 Specifications 3
1.1 General specifications.......................................................................................................................... 3
1.2 Typical performance ............................................................................................................................. 3
1.2.1 NIBP ........................................................................................................................................... 3
1.2.2 ECG.............................................................................................................................................3
1.2.3 Pulse oximetry .............................................................................................................................4
1.2.4 Temperature................................................................................................................................4
1.2.5 Invasive blood pressure................................................................................................................5
1.2.6 Respiration.................................................................................................................................. 5
1.3 Technical specifications........................................................................................................................ 6
1.3.1 NIBP ........................................................................................................................................... 6
1.3.2 ECG.............................................................................................................................................6
1.3.3 Pulse oximetry .............................................................................................................................6
1.3.4 Temperature................................................................................................................................7
1.3.5 Invasive blood pressure................................................................................................................7
1.3.6 Respiration.................................................................................................................................. 7
2 Functional Description 8
2.1 Measurement principle......................................................................................................................... 8
2.1.1 NIBP ........................................................................................................................................... 8
2.1.2 ECG.............................................................................................................................................8
2.1.3 Pulse oximetry .............................................................................................................................8
2.1.4 Temperature..............................................................................................................................10
2.1.5 Invasive blood pressure..............................................................................................................10
2.1.6 Respiration................................................................................................................................ 10
2.2 Main components ..............................................................................................................................11
2.2.1 M-PRESTN/-RESTN/-PRETN modules .........................................................................................11
2.2.2 NIBP board................................................................................................................................ 12
2.2.3 ECG board in 12-lead measurement...........................................................................................15
2.2.4 ECG filtering...............................................................................................................................17
2.2.5 STP board..................................................................................................................................18
2.3 Connectors and signals ......................................................................................................................23
2.3.1 Module bus connector (on the NIBP board) ................................................................................. 23
2.3.2 Front panel connectors...............................................................................................................25
2.3.3 Test points on boards.................................................................................................................26
3 Service Procedures 29
3.1 General service information ................................................................................................................29
3.2 Service check..................................................................................................................................... 29
3.2.1 Recommended tools.................................................................................................................. 29
3.2.2 Recommended parts..................................................................................................................29
3.3 Disassembly and reassembly..............................................................................................................39
3.3.1 M-PRESTN/-RESTN/-PRETN modules .........................................................................................39
3.4 Adjustments and calibrations ............................................................................................................. 40
Document No. 8005571-1
i
Datex-Ohmeda S/5 monitors
3.4.1 NIBP calibrations....................................................................................................................... 40
3.4.2 Temperature calibration............................................................................................................. 42
3.4.3 Invasive pressure calibration...................................................................................................... 42
4 Troubleshooting 43
4.1 Troubleshooting charts ....................................................................................................................... 43
4.1.1 NIBP ......................................................................................................................................... 43
4.1.2 NIBP error code explanation....................................................................................................... 46
4.1.3 ECG .......................................................................................................................................... 47
4.1.4 Pulse oximetry (SpO2) ................................................................................................................ 47
4.1.5 Temperature.............................................................................................................................. 48
4.1.6 Invasive blood pressure ............................................................................................................. 49
4.1.7 Impedance respiration............................................................................................................... 50
4.2 Troubleshooting flowcharts................................................................................................................. 51
4.2.1 M-PRESTN module troubleshooting for NIBP parameter............................................................... 51
4.2.2 M-PRESTN module troubleshooting for parameters ESTPR........................................................... 52
5 Service Menu 53
5.1 NIBP service menu ............................................................................................................................. 54
5.1.2 NIBP demo menu....................................................................................................................... 55
5.1.3 NIBP calibration menu ............................................................................................................... 56
5.1.4 NIBP safety valve menu.............................................................................................................. 57
5.1.5 NIBP pulse valve menu .............................................................................................................. 58
5.1.6 NIBP buttons/leds menu............................................................................................................ 59
5.1.7 NIBP pneumatics menu ............................................................................................................. 60
5.1.8 NIBP watchdog menu................................................................................................................. 61
5.1.9 ECG service menu...................................................................................................................... 62
5.1.10 ECG setup menu.................................................................................................................... 64
5.2 STP service menu............................................................................................................................... 65
5.2.2 STP calibration menu................................................................................................................. 67
6 Spare Parts 68
6.1 Spare parts list................................................................................................................................... 68
6.1.1 M-PRESTN rev. 01, M-RESTN rev. 01, M-PRETN rev. 01 ............................................................... 68
7 Earlier Revisions 71
APPENDIX A 73
Service Check Form 1
ii
Document No. 8005571-1
Table of contents

TABLE OF FIGURES

Figure 1 S/5 PRESTN Module, M-PRESTN........................................................................................................... 1
Figure 2 Absorption of infrared light in the finger probe parts layout and schematic diagram ................................. 9
Figure 3 Front panel of M-PRESTN....................................................................................................................11
Figure 4 NIBP board functional block diagram .................................................................................................. 12
Figure 5 NIBP pneumatics diagram ..................................................................................................................14
Figure 6 12-lead ECG measurement block diagram .......................................................................................... 15
Figure 7 STP board block diagram ....................................................................................................................18
Figure 8 Temperature measurement principle ...................................................................................................19
Figure 9 Pressure measurement principle.........................................................................................................20
Figure 10 Pulse oximetry measurement block diagram....................................................................................21
Figure 11 Serial communication of NIBP board............................................................................................... 22
Figure 12 Serial Communication and Isolation of STP board............................................................................22
Figure 13 Serial Communication and Isolation of ECG board ...........................................................................22
Figure 14 Module bus connector (X1) pin layout ............................................................................................. 23
Figure 15 ECG board connectors and test points............................................................................................. 26
Figure 16 NIBP board connectors and test points............................................................................................ 27
Figure 17 STP board connectors and test points ............................................................................................. 28
Figure 18 M-PRESTN module troubleshooting flowchart for NIBP Parameter..................................................... 51
Figure 19 M-PRESTN Module Troubleshooting Flowchart for Parameters ESTPR................................................52
Figure 20 Exploded view of M-PRESTN Module ...............................................................................................68
iii
Document No. 8005571-1
Datex-Ohmeda S/5 monitors
iv
Document No. 8005571-1

INTRODUCTION

This Technical Reference Manual Slot provides information for the maintenance and service of the hemodynamic modules S/5 M-PRESTN/-RESTN/-PRETN. The modules are double width modules designed for use with S/5 monitors. Later in this manual modules may be referred to w/o the system name S/5 for simplicity.
Please also refer to Technical Reference Manual of the S/5 monitor for information regarding system specific information e.g. related documentation, conventions used, symbols on equipment, safety precautions, system description, system installation, interfacing, functional check and planned maintenance.
The M-PRESTN/-RESTN/-PRETN modules provide general hemodynamic parameters.
Auto
On/Off
NIBP
ECG
Start
Cancel
T1
SpO
P1
Zero
P1
S/5 M-PRESTN Modules
NOTE: Do not use identical modules in the same
Zero
P2
2
monitor simultaneously. The following modules are considered identical: M-ESTP/-EST/-ETP, M-ESTPR/-ESTR/-ETPR, M-NESTPR/-NESTR/-NETPR, M-NE12STPR/-NE12STR/-NE12TPR M-PRESTN/M-RESTN/M-PRETN
T2
P2
Figure 1 S/5 PRESTN Module, M-PRESTN
Table 1 Options of S/5 hemodynamic modules
Parameter PRESTN RESTN PRETN
P
R
E
S
T
N
Two invasive blood pressures
Impedance respiration
ECG
Pulse oximetry
Two temperatures
NIBP
NOTE: 12-lead ECG measurement requires Display Controller, B-DISP or B-DISPX.
Intended purpose (Indications for use)
The Datex-Ohmeda PRESTN module (model family M-PRESTN) and accessories are indicated for monitoring of hemodynamic parameters of all hospital patients. The hemodynamic parameters of the module comprise ECG including ST-segment and arrhythmia, Impedance respiration, NIBP, Temperature,SpO2 (including monitoring during conditions of clinical patient motion),and invasive blood pressure.
1
Document No. 8005571-1
Datex-Ohmeda S/5 monitors
Impedance respiration measurement is indicated for patients aged 3 and up. The NIBP measurement is indicated for patients who weigh 5kg (11 lb.) and up. This device is indicated for use by qualified medical personnel only.
Monitor software compatibility
Datex-Ohmeda PRESTN rev. 01 module is designed for use with Datex-Ohmeda monitors using software as follows: AM: L-ANE01(A) or later versions; CCM: S-00C01 rev. 10.5, S-00C02 rev. 10.5 or newer versions; CAM: S-00A05 rev. 10.9, S-00A06 rev. 10.9, L-00A07 rev. 10.9, L-00A08 rev. 10.9 or newer versions and CCCM: S-00C03 rev. 10.9, S-00C04 rev. 10.9 or newer versions.
2
Document No. 8005571-1
1 SPECIFICATIONS
1.1 General specifications
Module size 75 × 180 × 112 mm W × D × H 3.0 × 7.1 × 4.4 in Module weight 0.7 kg / 1.5 lbs
Power consumption about 6 W Operation temperature 10 to 40 °C / 50 to 104 °F
1.2 Typical performance
1.2.1 NIBP
NOTE: Non-invasive blood pressure measurement is intended for patients weighing over 5 kg (11 lb.)
S/5 M-PRESTN Modules
1.2.2 ECG
Oscillometric measurement principle.
Measurement range adult 25 to 260 mmHg child 25 to 195 mmHg infant 15 to 145 mmHg
Pulse rate range accepted 30 to 250 bpm
Measurement interval STAT (continuous 5 min), 1, 2.5, 3, 5, 10, 15, 30 and 60 min (1 h),
2 and 4 h
Typical measuring time adult 23 s infant 20 s
Initial inflation pressure adult 170 ±10 mmHg child 150 ±10 mmHg infant 120 ±10 mmHg
Venous stasis adult 40 ±5 mmHg / 2 min child 40 ±5 mmHg / 2 min infant 30 ±5 mmHg / 1 min
Cuff widths please see User’s Guide
Lead selection, 12-lead ECG I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 Lead selection, other modules I, II, III, aVR, aVL, aVF, V Sweep speeds 12.5, 25, 50 mm/sec
DISPLAY FILTER
Diagnostic, 12-lead ECG 0.05 to 150 Hz Diagnostic, other modules 0.05 to 100 Hz
Monitoring 0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
3
Document No. 8005571-1
Datex-Ohmeda S/5 monitors
ST filter 0.05 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.05 to 40 Hz (-3 dB, with 60 Hz reject filter)
HEART RATE FROM ECG
Range 30 to 250 bpm Accuracy ±5 bpm or ±5 %, whichever is greater Resolution 1 bpm Update interval 5 s Averaging time 10 s
ST LEVELS (in main software)
ST level range -9 to +9 mm (-0.9 to +0.9 mV) Resolution 0.1 mm (0.01 mV) Averaging calculated from 8 QRS
SYNCHRONIZATION
Direct ECG analog output of ECG, 1 V/1 mV Pacer 5 V and 0.5 to 2.5 ms pulse, < 30 ms after pacer peak Defibrillator 5 V and 10 ms pulse, < 35 ms after R-point synchronization
1.2.3 Pulse oximetry
Measurement range 0 to 100 % Calibration range 70 to 100 %
Accuracy
1 100 to 70 %, ±2 digits
±3 digits during clinical patient motion 69 to 0 %, unspecified
Display resolution 1 digit = 1 % of SpO
2
Display averaging time 20, 10 sec, beat-to-beat Pulse beep pitch varies with SpO
The monitor is calibrated against functional oxygen saturation SpO
2 level
func.
2
PULSE RATE FROM PLETH
Measurement range 30 to 250 bpm Accuracy 30 to 100, ±5 bpm, 100 to 250, ±5 % Resolution 1 bpm Display averaging 10 s
Adjustable pulse beep volume.
PLETH WAVEFORM
Scales 2, 5, 10, 20, 50 mod%, Auto
Start up scale is 20 mod% if AUTO is not selected to be the default setting.
1.2.4 Temperature
Measurement range 10 to 45 °C (50 to 113 °F) Measurement accuracy ±0.1 °C (25 to 45.0 °C)
1 Accuracy is based on deep hypoxia studies with volunteered subjects during motion and non-motion conditions over a wide range of arterial
blood oxygen saturations as compared to arterial blood CO-Oximetry.
4
Document No. 8005571-1
±0.2 °C (10 to 24.9 °C) Display resolution 0.1 °C (0.1 °F) Temperature test automatic (every 10 min) Probe type compatible with YSI 400 series
Single use sensors ±0.2 °C (25 to 45.0 °C) ±0.3 °C (10 to 24.9 °C)
1.2.5 Invasive blood pressure
Measurement range -40 to 320 mmHg Measurement accuracy ±2 mmHg or ±5 % Zero adjustment range ±150 mmHg Calibration range ±20 % Scales upper limit is adjustable between 10 and 300 mmHg in steps of
10. Lower limit is 10 % of selected upper limit below zero.
Sweep speed 12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range -40 to 320 mmHg Resolution ±1 mmHg
S/5 M-PRESTN Modules
WAVEFORM DISPLAY
Range -30 to 300 mmHg
PULSE RATE FROM ARTERIAL PRESSURE
Measurement range 30 to 250 bpm Resolution 1 bpm Accuracy ±5 bpm or ±5 % whichever is greater
1.2.6 Respiration
NOTE: The respiration measurement is intended for patients over three years old
Measurement range 4 to 120 bpm Accuracy ±5 bpm or ±5 % Resolution 1 bpm Averaging time 30 s Update interval 10 s
RESPIRATION WAVEFORM
Sweep Speeds 6.25 mm/s and 0.625 mm/s
Document No. 8005571-1
5
Datex-Ohmeda S/5 monitors
1.3 Technical specifications
1.3.1 NIBP
Deflation rate, PR dep. 3 to 8 mmHg/s Inflation time 20 to 185 mmHg, 1 to 5 s
Automatic software control, max. inflation pressure adult 280 ±10 mmHg child 200 ±10 mmHg infant 145 ±5 mmHg
Over pressure limit, stops measurement after 2 seconds adult 320 mmHg child 220 mmHg infant 160 mmHg
The safety circuit limits the maximum cuff pressure to 320 mmHg in adult/child mode or 160 mmHg in infant mode. Independent timing circuit limits pressurizing (>15 mmHg) time to 3 minutes maximum in adult/child mode, and 90 seconds at (>5mmHg ) in infant mode.
Zeroing to ambient pressure is done automatically.
1.3.2 ECG
Inflation pressure is adjusted according to the previous systolic pressure, typically 40 mmHg above. If the systolic pressure is not found, inflation pressure is increased typically 50 mmHg.
Max. measurement time adult 120 s child 120 s infant 75 s
Pressure transducer accuracy is better than ±3 mmHg or ±2 % whichever is greater. Max. error ±4 mmHg.
Protection against electrical shock Type BF defibrillation proof
Defibrillation protection 5000 V, 360 J Recovery time 5 s Input impedance >2.5 M (10 Hz) CMRR 95 dB (ST) System noise <30 µV (p-p, RTI)
Allowable offset ±800 mVDC Gain range 0.2 to 5.0 cm/mV Pacemaker pulse detection 2 to 700 mV, 0.5 to 2 ms pulses
Protection against electrical shock Type CF defibrillator proof
1.3.3 Pulse oximetry
Protection against electrical shock Type BF defibrillation proof
6
Document No. 8005571-1
1.3.4 Temperature
Measurement accuracy ±0.1 °C (25.0 to 45.0 °C) ±0.2 °C (10.0 to 24.9 °C) Protection against electrical
shock Type CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on transducer/probe used. Please refer to the transducer/probe specification.
1.3.5 Invasive blood pressure
DIGITAL DISPLAY AVERAGING
Digital displays Art and P1 are averaged over 5 seconds and updated at 5 seconds intervals. All other pressures have respiration artifact rejection.
Accuracy ±5 % or ±2 mmHg, whichever is greater Transducer and input sensitivity 5 µV/V/mmHg, 5 VDC, 20 mA max current Filter 0 to 4 - 22 Hz adjustable Zero set accuracy ±1 mmHg Calibration resolution ±1 mmHg Zero time less than 15 s Protection against electrical shock Type CF defibrillation proof
S/5 M-PRESTN Modules
NOTE: The accuracy of the measurement may be different from the specified, depending on transducer/probe used. Please refer to the transducer/probe specification.
1.3.6 Respiration
Excitation frequency, 12-lead ECG 31.25 kHz
Breath detection automatic, range 0.3 to 6 manually adjustable minimum detection: 0.2, 0.4, 0.6, 0.8, 1.0
Input dynamic range 0.2 to 32 Input impedance range 100 to 5000 Respiration Rate min. 4 bpm
max. 120 bpm Lead off detection >3 M
Document No. 8005571-1
7
Datex-Ohmeda S/5 monitors
2 FUNCTIONAL DESCRIPTION
2.1 Measurement principle
2.1.1 NIBP
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a speed based on the patient’s pulse, collecting data from the oscillations caused by the pulsating artery. Based on these oscillations, values for systolic, mean, and diastolic pressures are calculated.
The following parts are necessary for the NIBP measurement:
M-PRESTN/-RESTN/-PRETN module
twin hose (adult or infant model)
blood pressure cuffs (various sizes)
2.1.2 ECG
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical potential produced with electrodes placed on the surface of the body.
ECG reflects:
electrical activity of the heart
normal/abnormal function of the heart
effects of anesthesia on heart function
effects of surgery on heart function
See the User's Guide or the User’s Reference Manual for electrodes positions and other information.
2.1.3 Pulse oximetry
A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near infrared (about 900 nm) and the other in the red region (about 660 nm) of the light spectrum. These wavelengths are emitted by LEDs in the SpO tissue and is finally detected by a PIN-diode opposite the LEDs in the probe. The pulse oximeter derives the oxygen saturation (SpO relative absorption at the two wavelengths and the arterial oxygen saturation SaO
In order to measure the arterial saturation accurately, pulse oximeters use the component of light absorption giving variations synchronous with heart beat as primary information on the arterial saturation.
probe, the light is transmitted through peripheral
2
) using an empirically determined relationship between the
2
.
2
8
Document No. 8005571-1
A general limitation of pulse oximetry is that due to the use of only two wavelengths only two hemoglobin species can be discriminated by the measurement.
The modern pulse oximeters are empirically calibrated either against fractional saturation SaO
frac;
2
2
2
fracSaO
= Formula 1
HbO
2
++
binDyshemogloHbHbO
S/5 M-PRESTN Modules
or against functional saturation SaO
2
funcSaO
HbO
= Formula 2
Functional saturation is more insensitive to changes of carboxyhemoglobin and methemoglobin concentrations in blood.
The oxygen saturation percentage SpO against functional saturation SaO measurement relative to SaO carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeters are not able to correctly measure oxygen content of the arterial blood at elevated carboxyhemoglobin or methemoglobin levels.
Plethysmographic pulse wave
The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation at the measuring site. Thus the amplitude of the waveform represents the perfusion.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The signals are filtered to reduce noise and checked to separate artifacts.
func;
2
2
HbHbO
2
+
measured by the Datex-Ohmeda module is calibrated
2
func. The advantage of this method is that the accuracy of SpO2
2
func can be maintained even at rather high concentrations of
2
Probe
The standard probe is a finger clamp probe which contains the light source LEDs in one half and the photodiode detector in the other half. Different kinds of probes are available from Datex-Ohmeda.
SpO2 sensor connector
6
GND
7
ILED
4
GND
5
DEF_A
8
IS
9
DET_C
PRSTN_absorption_of_infrared.vsd
Emitter
Detector
SpO2 sensor cable
IRED
RED
Figure 2 Absorption of infrared light in the finger probe, parts layout and schematic
diagram
Document No. 8005571-1
9
Datex-Ohmeda S/5 monitors
2.1.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature changes, called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
Applying a constant voltage across the resistor and measuring the current that flows through it
Applying a constant current through the resistor and measuring the voltage that is generated across it.
In Datex-Ohmeda modules the two methods are combined in the form of a voltage divider. The NTC­resistor is connected in series with a normal resistor and a constant voltage is applied across them. The temperature dependent voltage can be detected at the junction of the resistors, thus producing the temperature signal from the patient. The signal is amplified by analog amplifiers and further processed by digital electronics.
2.1.5 Invasive blood pressure
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal saline all connected together by stopcocks, is attached to the catheter. The transducer is placed at the same level with the heart, and is electrically zeroed.
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
2.1.6 Respiration
Impedance respiration is measured across the thorax between ECG electrodes. The respiration signal is made by supplying current between the electrodes and by measuring the differential current from the electrodes. The signal measured is the impedance change caused by breathing. From these impedance changes, respiration rate is calculated, and the respiration waveform is displayed on the screen.
10
Document No. 8005571-1
2.2 Main components
2.2.1 M-PRESTN/-RESTN/-PRETN modules
Auto
Start
Zero
On/Off
NIBP
ECG
Cancel
T1
T2
SpO
P1
P2
Zero
P1
P2
2
S/5 M-PRESTN Modules
Figure 3 Front panel of M-PRESTN
The M-PRESTN, M-RESTN, and M-PRETN modules contain three main PC boards, the STP board, the ECG board, and the NIBP board. They work independently. Each of these has their own processor and software in the processor flash memory.
There are two small boards, the SP input and the ECG input board attached to the front panel of the module. The front panel has seven connectors and four keys. The connectors are two for temperature measurement, two for invasive blood pressure measurement, one for ECG, one for NIBP, and one for SpO
measurement. The keys are for NIBP Auto On/Off, NIBP Start/Cancel, P1
2
zero, and P2 zero.
11
Document No. 8005571-1
Datex-Ohmeda S/5 monitors
2.2.2 NIBP board
PATIENT AND NIBP CUFF
NIBP TUBING
NIBP CONNECTOR
NIBP PNEUMATICS
NIBP CONTROL KEYS
NV MEMORY
MAIN CPU
PRESSURE SENSORS
RS 485 COMMUNICATION POWER SUPPLY
DRIVERS FOR PUMP &
VALVES
SAFETY CPU
NIBP BLOCK DIAGRAM
Figure 4 NIBP board functional block diagram
Signal processing
Two signals from the pressure transducers are amplified and sent to the A/D converter. After the converter, digitized signals are sent to the microprocessor for data processing.
The NIBP board is controlled with a H8/3052 microprocessor at 16 MHz oscillator frequency.
12
Document No. 8005571-1
MODULE BUS CONNECTOR
Memory
NIBP program memory (processor flash memory) size is 512k × 8. The processor has 4 kBytes RAM and there is also an external RAM memory the size of which is 128k x 8. Variable values of the NIBP measurement are stored into the external RAM. The EEPROM size is 512 x 8 and it is used to store the calibration values for the pressure transducers, the pulse valve constants gained during measurements, the PC board identification, and module serial number.
Software control
Software controls valves and pump. In addition to the individual on/off signals for each component there is a common power switch for the valves and the pump that can be used at pump/valve failures.
In addition to external RS485 reset line the microprocessor system is equipped with its own power­up reset. See the section in the ECG board’s description:
Safety circuit
The NIBP board is equipped with an independent safety circuit to disconnect supply voltages from the pump and the valves if the cuff has been pressurized longer than the preset maximum measurement time, or if the pressure of the cuff is inflated over the specified pressure limit. The maximum measurement time values and pressure limits for different measurement modes have been specified in the technical specification section of this manual.
S/5 M-PRESTN Modules
“RS485 communication”
Pneumatics
Pneumatics of PRESTN module has the following parts:
Intake air filter; for preventing dust and other parts to enter the air pump and the
valves.
Air pump: for pumping the measuring pressure of the cuff.
(
Pulse) Valve; for producing a linear pressure fall (bleeding) in order to measure the
blood pressure of the patient. Note that there has been used also two other names designate pulse valve in service menu.
Safety valve; The safety valve has been intended to be used for deflating the cuff in
single fault case, i.e. to prevent too long measurement time or too high inflation pressure of the cuff. Note that there has been used also service menu.
Main pressure sensor; for measuring the pressure of the blood pressure cuff and
the pressure fluctuations caused by arterial wall movement.
Safety pressure sensor for detection of cuff hose type, cuff loose, cuff occlusion
situations etc.and recognising the pressure sensor fault.
Exh2 valve to designate the Safety valve in
Valve and Set valve to
Cuff connector; for connecting the cuff.
13
Document No. 8005571-1
Datex-Ohmeda S/5 monitors
Cuff connector
Safety pressure sensor
Proportional valve
Air pump
Main pressure sensor
Figure 5 NIBP pneumatics diagram
Power supply section of the NIBP board
All connections are established via 25-pin connector (D-type, female). The module needs +15 V (dirty) power supply to operate. The supply voltage (+15V) is generated in the power supply section of the S/5 monitor. The other voltages needed for the operation of the NIBP measurement are made on the NIBP board.
Dump valve
Intake air filter
NIBP_pneum_diagr.vsd
14
Document No. 8005571-1
2.2.3 ECG board in 12-lead measurement
The 12-lead ECG measurement consists of the elements shown in Figure 6. All functions are located in the ECG board except the ECG input unit .
PATIENT AND ECG ELECTRODES
ECG CABLE
- ECG LEAD SET
- ECG TRUNK CABLE
ECG INPUT UNIT
- ECG CONNECTOR
- INPUT PROTECTION RESISTORS
INPUT PROTECTION DIODES FOR ECG & RESPIRATION MEASUREMENT
INPUT FILTERING FOR ECG & RESPIRATION MEASUREMENTS
S/5 M-PRESTN Modules
RESPIRATION
MEASUREMENT
AMPLIFIERS
RS 485 COMMUNICATION POWER SUPPLY
ISOLATION ISOLATION
MODULE BUS CONNECTOR
RESPIRATION
MEASUREMENT CURRENT
SUPPLY
ECG CPU
ECG PREAMPLIFIERS &
RLD CIRCUIT
BASELINE
RESTORATION
NV MEMORY
ECG BLOCK DIAGRAM
LEADS OFF & PACER &
DEFIBRILLATION
DETECTION
Figure 6 12-lead ECG measurement block diagram
15
Document No. 8005571-1
Datex-Ohmeda S/5 monitors
ECG input unit
The ECG input unit consists of the front panel connector and the ECG input connector board with the high voltage protection resistors. The connector for the 12-lead ECG cable is a green 11-pin rectangle shaped connector.
Input protection and filtering
The input protection is implemented with high voltage protection resistors in the ECG input unit and with protection diodes in the ECG board. The input filtering for ECG measurement is done with passive RC filtering.
ECG preamplifiers
The buffer amplifiers are used for each lead. The “Leads off” detection is implemented by measuring the output level of the input buffer amplifiers with A/D converter of CPU. The ECG signals are measured using differential amplifiers.
ECG amplifiers and baseline restoration
The function of the ECG amplifiers and baseline restoration is to amplify the signal and to restore the baseline of the signal in the middle of the display after the change of the signal level e.g. after the change of the DC offset voltage.
Pacer detection
Pacer detection has been made by using three slew rate detector circuits. The pacer detection amplifiers have been realized at the front of the slew rate detectors independently from the ECG measuring channels.
Respiration impedance supply
The 31.25 kHz sine wave generator is used as the respiration measurement signal supply. Analog switches are used for connecting the sine wave to the ECG leads to be measured.
Respiration impedance amplifiers
Buffer amplifiers are used in respiration measurement. Analog switches are used for selecting the measurement leads. There are also additional amplifiers for increasing the respiration signal gain. When ECG measurement is 5/12-lead, the respiration measurement is always done between R and F, independently on the ECG lead selection. When ECG measurement is 3-lead, then the respiration measurement is done at the same lead as the ECG measurement (I, II or III).
ECG CPU
16
Document No. 8005571-1
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory and 4 kbytes of RAM. The clock frequency is 16 MHz.
RS485 communication
The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver circuits are optically isolated from the processor of the module. PWM signal is used for direct ECG signal. Direct ECG signal is available from the X2 connector of the UPI board or from the PT module.
Power supply
The ECG board has a driver controlled half bridge switching power supply with 5 kV isolation. The supply voltages have been regulated with linear regulators.
2.2.4 ECG filtering
The S/5 monitors have three ECG filtering modes:
MONITORING 0.5 to 30 Hz (with 50 Hz reject filter)
0.5 to 40 Hz (with 60 Hz reject filter)
DIAGNOSTIC 12-lead ECG 0.05 to 150 Hz
ST FILTER 0.05 to 30 Hz (with 50 Hz reject filter)
0.05 to 40 Hz (with 60 Hz reject filter)
S/5 M-PRESTN Modules
The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory) movement artifacts.
Monitor filter is used in normal monitoring. Diagnostic filter is used if more accurate diagnostic information is needed. ST filter gives more accurate information of ST segment, but reduces high frequency noise.
The high-pass filters 0.5 Hz and 0.05 Hz are done with software. The monitor sends a command to the hemodynamic module determining which of the corner frequencies 0.5 Hz or 0.05 Hz is to be used.
The 50 Hz and 60 Hz reject filters are both low-pass filters with zero at 50 Hz or 60 Hz correspondingly. They are software based filters used for the mains supply filtering. With these filters the 3 dB value for low-pass filter is 30 Hz or 40 Hz.
In diagnostic mode the upper frequency is 150 Hz and it is limited by software.
17
Document No. 8005571-1
Datex-Ohmeda S/5 monitors
2.2.5 STP board
PATIENT AND SpO2 PROBE
SpO2 TRUNK CABLE
SpO2 CONNECTOR
SpO2 PROBE RECOGNITION & LED
DRIVE SELECTION MATRIX
SpO2 LED DRIVE
SpO2 AMPLIFIER
PATIENT AND TEMPERATURE
SENSOR
TEMPERATURE CONNECTOR
TEMPERATURE
MEASUREMENT
AMPLIFIER
A/D CONVERSION
SENSOR
SIGNAL
CURRENT
SOURCE
PATIENT AND INVASIVE CANNULA
PRESSURE SENSOR WITH DOME
INV: PRESSURE
MEASURE MENT
OR CATHETER
FLUSHING KIT & INVASIVE
INV.PRESSURE CONNECTORINPUT PROTECTION CIRCUITRY
SENSOR SIGNAL
AMPLIFIER
VOLTAGE
SOURCE
STP CPU
NV MEMORY
RS 485 COMMUNICATION POWER SUPPLY
ISOLATION ISOLATION
MODULE BUS CONNECTOR
STP BLOCK DIAGRAM
Figure 7 STP board block diagram
18
Document No. 8005571-1
Microprocessor unit
The CPU is a 16 bit H8/3052 single-chip microcomputer. It contains 128 kbytes of flash memory and 4 kbytes of RAM. The clock frequency is 16 MHz.
High speed I/O is used to obtain pulse control sequence necessary for pulse oximetry measurement. Timing for the clock is from the oscillator.
Temperature measurement unit
The NTC-resistor value in the probe depends on the patient’s temperature. It is measured with the following principle described below.
The constant current source is supplied about 7µA current through the temperature sensor (YSI 400-series NTC resistor). The constant current is caused a voltage over the temperature sensor (NTC resistor). The voltage over the temperature sensor is amplified in a differential amplifier stage. The amplified voltage is transferred to a controller of the STP board through an A/D converter.
Defibrillation/ESD protection
resistors and diodes
S/5 M-PRESTN Modules
constant
current source
Temperature
sensor
Figure 8 Temperature measurement principle
100k
10k
0C:7k36 15C:3k54 25C:2k53 38C:1k30 45C: 984
10k
100k
Differential
amplifier
d/dt
0
0
0
0
To A/D
converter
temp_meas_principle.vsd
19
Document No. 8005571-1
Datex-Ohmeda S/5 monitors
Invasive blood pressure measurement unit
An isolated +5 V voltage is supplied to the pressure transducer. The differential voltage, which depends on the pressure and the supplied voltage, is calculated from the bridge connection (see the formula below).
U
= Uin × pressure × 5 V, where Uin is 5 V
out
U
= 25 V × pressure [mmHg]
out
Pressure amplification is realized in the instrumentation amplifier. The gain of the amplifier is set to keep the level of the signal transferred to A/D converter within the measurement range even when there are circumstantial offsets or offsets caused by the transducer. There is a filter before the amplifier to attenuate high frequency disturbances.
Vin
Pressure transducer
Vout
Figure 9 Pressure measurement principle
Pulse oximetry measurement section
LED control signals
The D/A converters of the microcontroller on STP board set the LED intensity adjustment values for the infrared- and red LEDs of SpO2 probe. The microcontroller on the STP board switches ON ( to the adjusted intensity ) and OFF the SpO2 probe LEDs according to the predetermined sequence.
Input Filter
Instrumentation amplifier
G
to AD converter
pressure_meas_principle.vsd
LED driving circuit
Differential amplifiers measure the LED currents (LED current indication) of SpO2 probe over the shunt resistors placed in the LED current paths. The LED driving voltages (LED voltage indication) are measured from the driver circuitry. The LED driving circuits also have MOSFET transistor matrix to enable the use of different probe configurations.
Measured signal preamplification
The preamplifier is bipolar/single-ended current-to-voltage converter with adjustable gain. A higher gain is used for measuring thin tissue. The preamplification stage has also ambient light reduction and second amplifier stage.
20
Document No. 8005571-1
Loading...
+ 58 hidden pages