These instructions include procedures for a planned maintenance (PM) for Datex-Ohmeda S/5
Anesthesia Monitor and S/5 Critical Care Monitor. The Planned maintenance should be performed
once a year.
Planned maintenance instructions
These instructions include a Check Form
corresponding procedures.
The symbol
check form.
The procedures should be performed in ascending order, bypassing those that are not applicable
for a particular monitor.
If you need further information how to perform a certain PM procedure, please refer to the
corresponding slot in the Product Service Guide in the Technical Reference Manual.
All menu selections related to
e.g.
As you enter the service menus, you need following passwords:
Monitor Setup
In case you evaluate the measurement accuracy with a patient simulator, add simulator’s accuracy
specification to the one of the monitor.
The procedures are designed for monitors with S/5 monitor software L-ANE01(A) or L-ICU01(A).
However, most of the procedures also apply to monitors with older monitor software.
? in the instructions means that the performed procedure should be signed in the
Parameters - More... - TONO
- Install/Service
(APPENDIX A)
Datex-Ohmeda products are written in following typeface:
(password 16-4-34) -
to be filled in when performing th e
Service
(password 26-23-8)
Document No. 8002949
1
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
NOTE: Corresponding PM kit is available with Order No. 8001761.
Planned maintenance instructions
5
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
2 PLANNED MAINTENANCE CHECKLIST
2.1 Visual inspection/preparation
2.1.1 General
NOTE: Wear a static control wrist strap when handling PC boards. Electrostatic discharge may
damage components on the board.
• Make sure that the monitor is switched to standby.
Disconnect the mains power cord from the 8-Module Frame, F-CU8. If the monitor is
connected to the Datex-Ohmeda network, disconnect the Mon-Net cable from the Network
Board, B-NET or UPI4NET board, B-UPI4NET. If the Memory Module, M-MEM is connected,
remove any memory cards.
1. Check all units visually. Check e.g. that all parts are intact, the cables and screws are
connected and tightened properly.
Especially check the following parts:
Video displays : the display power cord is locked to the display.
F-CU8: the equipotential tap and all the screws are tightened properly.
Airway Module, G-XXXXX: the equipotential tap is tightened properly.
Check that modules go in smoothly and lock up properly in all module slots.
2. The lead-acid battery in the power supply unit is recommend to be replaced after every 4
years. Replace the battery, if necessary.
The SRAM/Timekeeper battery on the CPU board is recommend to be replaced after every 8
years. Replace the battery, if necessary.
NOTE: The Factory Reset should be performed if the SRAM/Timekeeper battery, or the
SRAM/Timekeeper chip is replaced.
3. Clean or replace the F-CU8 fan filter.
6
Document No. 8002949
4. Check that the fuses are of the correct rating.
?
2.1.2 Extension Frame, F-EXT4
1. If the F-EXT4 contains a fan, clean or replace the fan filter.
?
2.1.3 Airway Module, G-XXXXX
1. Remove the Airway Module case, the top protection cover and the bronze plate from the
side of the O2 sensor, if installed.
Detach the PVX board with the support plate, if installed, and leave the board detached
until further notice.
Check that all cables and tubes are connected properly and there are no loose objects
inside the module.
NOTE: Make sure that none of the tubes are in contact with the sampling pump or with the
O2 sensor.
Planned maintenance instructions
2. Install the PM Kit
Replace the special tubes (Nafion™).
Replace or install the OM ref. filter, order code 86901.
Check the D-fend O-rings and replace them, if necessary
Replace the D-fend and the sampling lin e.
Clean or replace the fan filter.
NOTE: Use only Datex-Ohmeda sampling lines to ensure proper functioning.
3. The CO
necessary.
absorber is recommend to be replaced after every 4 years. Replace the absorber, if
2
?
Connect the Airway Module to the F-CU8 with a long gas interface cable.
7
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
2.1.4 Compact Airway Module, M-CXXXXX
1. Detach the module box.
Check that all cables and tubes are connected properly and there are no loose objects
inside the module.
NOTE: The tubes connected to the Oxygen board pressure transducers should not be
pressed too deep.
NOTE: Make sure that tubes are not in contact with the sampling pump or the O
its springs.
2. Install the PM Kit:
Replace the special tubes (Nafion™).
Replace the Ref. filter assembly.
Replace the filters in the pneum atic unit (2 pcs).
Check the D-fend O-rings and replace them, if necessary.
Replace the D-fend and sampling line.
Clean or replace the fan filter.
NOTE: Use only Datex-Ohmeda sampling lines to ensure proper functioning.
Use a 2 m/7 ft sampling line with Compact Airway Modules M-COVX and M-CAiOVX.
3. The CO
necessary.
Note: All Compact Airway Modules do not contain a CO
absorber is recommend to be replaced after every 4 years. Replace the absorber if
2
absorber.
2
?
sensor, or
2
8
Document No. 8002949
Connect the Compact Airway Module to one of the F-CU8 module slots with a long gas
interface cable (the grounding plates of the cable should be removed).
2.1.5 Tonometry Module
1. Detach the module box.
Check that all cables and tubes are connected properly and there are no loose objects
inside the module.
2. Replace the special tube (Nafion™)
Reattach the module box and check that the latch moves properly. Plug the module back
into the F-CU8.
?
2.1.6 Recorder module
1. Clean the Recorder Module, M-REC.
Open the paper compartment hatch and remove the paper roll, if installed.
Remove any paper chaff from the paper compartment.
Planned maintenance instructions
Clean the thermal printhead and the small glass window in front of the static brush with a
cotton swab dipped in isopropyl alcohol, if necessary.
NOTE: Avoid contact with the rubber paper roller. Be careful to limit the application of
alcohol to the thermal printhead and the window.
Reinstall the paper roll.
?
2.1.7 Modules with NIBP measurement
1. Check the NIBP pump filter. Replace the filter, if necessary.
Note: M-NIBP Rev. 00-03 do not contain a NIBP pump filter.
Plug the module back into the frame.
?
9
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
2.2 Functional Inspection
2.2.1 General
1. Connect the mains power cord to the F-CU8.
Check that the stand-by LED is lit.
2. Switch the monitor on.
Check that the monitor starts up properly, i.e. the alarm LEDs turn on and off, normal start-
up sound is heard from the loudspeaker and monitoring screen appears.
No error messages should appear onto the screen.
3. Check that all of the connected modules are recognized, i.e. the required parameter
information is shown on the screen.
If the Recorder Module, M-REC, is connected, the module should record two lines of startup information.
Preset measurement settings for those parameters that are connected are, for example:
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
2.2.2 Disp lay(s)
1. Check that the pictur e on the screen is adjusted correct ly. Readjust the picture with the
adjustment knobs or with the display menu keys, if necessary.
?
2.2.3 Keyboard(s)
1. Tests with the Command Bar / Command Board:
Press the
menu moves correspondingly. Highlight
that the menu disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
Tests with the ARK Keyboard:
Enter the
Check functioning of the ComWheel.
Press all keys, except
loudspeaker, or the ‘Message count’ value in the service menu increases.
Press the Modify and Print keys and check that the corresponding menus appear on the
monitor screen.
Tests with the Remote Controller:
Enter the
Check functioning of the ComWheel.
Press all keys. Check that each key produces a sound from the loudspeaker, or the
‘Message count’ value in the service menu increases.
Help key. Turn the ComWheel in both directions and check that the cursor in the
Keyboard
Keyboard
Normal Screen
service menu
Modify and Print. Check that each key produces a sound from the
service menu
, press the ComWheel and check
?
2.2.4 8-Module Frame, F-CU8
1. Check that the fan is running.
2. Check that the clock on the screen shows correct time.
Readjust the time and date, if necessary.
3. Enter the
Check the content of the Service Log for possible problems.
Service Log
?
12
Document No. 8002949
service menu.
2.2.5 Extension Frame, F-EXT4
1. If the F-EXT4 contains a fan, check that the fan is running.
2. Check that the modules in the F-EXT4 are recogn ized, i.e. the required parameter
information is shown on the monitor screen.
Disconnect the Extension Module, M-EXT shortly, then reconnect the module back into the
F-CU8.
Check that the modules in t he F-EXT4 are still recognized.
?
2.2.6 Airway Module, G-XXXX
1. Check that the Airway Module fan is running.
2. Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter the
ACX
service menu.
Planned maintenance instructions
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ -values are not increasing
faster than by 50 per second.
3. If the module contains a membrane key on the front panel, press the key for at least one
second and check that it is identified.
4. Check that the ‘Calib zero’ value for N2O is less than 61000.
5. Check that the ‘Ambient’ value corresponds with the current ambient pressure
(±20 mmHg).
6. Check the CO
Keep the tip of the sampling line away from you and let the monitor draw in room air. Check
that the ‘Insp CO
7. Perform a sampling system leak test.
8. Block the tip of the sampling line with your finger until the ‘Amb-Work’ value becomes
stable. Check that the value reaches 110.
absorber.
2
’ value on the
2
ACX
service menu is higher than 3.
9. Check the flow rates and adjust them, if necessary.
NOTE: If any of the constriction cassettes is replaced, the leak test should be repeated.
13
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
10. Perform a gas calibration:
NOTE: For maximum accuracy, a warm-up time of 30 minutes is recommended.
NOTE: If the module contains the agent identification unit, ASX-100 or ASX-200, keep
feeding gas at least 15 seconds after the message ‘Adjust’ appears on the menu. This gives
the agent identification unit enough time for calibration.
11. Switch the monitor to standby.
Install the PVX board, if inst alled originally, and reassemble the module.
Connect the Airway Module to the F-CU8 with the original gas interface cable.
Switch the monitor back on and wait until the message ‘Calibrating gas sensor’ disappears
from the screen.
?
For Airway Modules containing the Agent identification option
12. Enter the
Feed calibration gas. When a proper absorption spectrum is shown in the menu check that
the module identifies the gas and the ‘Peak normal’ value is close to 10.50 ( ±0.20). Check
also that the difference between the displayed ‘Peak normal’ and ‘Peak mirror’ values is not
greater than 0.30.
NOTE: The ASX-100 is not capable of identifying all calibration gases. The
menu values are not updated with the ASX-100.
ASX
service menu:
?
For Airway Modules containing the Patient Spirometry option
13. Connect a clean spirometry tube and D-lite to the module.
Perform the spirometry tubing leak test.
14. Connect the sampling line to the D-lite. Breathe through the wider side of the D-lite. Check
that the flow waveform moves downwards when you breathe in, and upwards when you
breathe out.
ASX
service
14
Document No. 8002949
?
For all Airway Modules, G-XXXX
15. Block the tip of the sampling line with your finger and check that the message ‘Sample line
blocked’ appears onto the monitor screen within 30 seconds.
16. Remove the D-fend and check that the message ‘Check D-fend’ appears on the screen
within 30 seconds.
?
2.2.7 Compact Airway Module
1. Check that the fan is running.
2. If the module contains membrane keys on the front panel, press each of the keys for at least
one second and check that their are identified.
NOTE: The Compact Airway Module membrane keys require S/5 main software.
Planned maintenance instructions
3. Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter the
Compact Airway Module
Check that the module configuration displayed corresponds with the Compact Airway
Module type being used.
4. Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second.
5. Enter the
the current ambient pressure (±20 mmHg).
6. Check that the flow measurement offset, i.e. the sample ‘Zero’ value displayed is within ±10
ml/min.
7. Perform a sampling system leak test.
8. Check the flow rates. Adjust the sampling flow, if necessary.
9. Perform a gas calibration:
Gases
General
service menu: Check that t he ‘Ambient’ value displayed corresponds with
service menu.
NOTE: For maximum accuracy, a warm-up time of 30 minutes is recommended.
NOTE: For correct measurement values, modules need different amounts of oxygen in the
calibration. If you do not use the recommended calibration gases, the calibration does not
succeed.
?
15
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
Compact Airway Modules w/ the Anesthesia Agent option
10. Enter the
Feed calibration gas (order code 755583) continuously for at least for 30 seconds and
check that the ‘ID’ in the service menu shows ‘DES’ and that the value for ‘ID unrel.’ is lower
than 50.
Gases
service menu.
?
Compact Airway Modules w/ the Patient Spirometry option
11. Enter the
Connect a clean spirometry tube and D-lite to the module.
Perform spirometry leak test.
12. Connect the sampling line. Breath through the wider side of the D-lite. Check that the flow
waveform moves downwards when you breathe in, and upwards when you breathe out.
Spirometry
service menu.
?
For all Compact Airway Modules
• Switch the monitor to standby and reassemble the module.
NOTE: Attach the plastic cover to the CPU Board before fitting the module box. Make sure
that the grounding claws in the front panel frame make good contact with the module box.
NOTE: When reassembling the module make sure that the tubes are not pinched between
the module box and internal parts.
Install the Compact Airway Module into the F-CU8. Switch the monitor back on and wait
until the message ‘Calibrating gas sensor’ disappears from the screen.
13. Block the tip of the sampling line with your finger and check that the message ‘Sample line
blocked’ appears on the monitor screen within 30 seconds.
14. Detach the D-fend and check that the messages ‘Check D-fend’ appears on the monitor
screen within 30 seconds.
?
16
Document No. 8002949
2.2.8 Tonometry Module, M-TONO
Planned maintenance instructions
1. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second.
Check that the memories of the module have passed the internal memory test, i.e. ‘RAM’,
‘ROM’ and ‘EEPROM’ all state OK.
Check that the ‘general error status’, ‘module pneumatics error status’, ‘module hardware
error status’ and ‘testbit status’ are all zero.
2. Check functioning of the front panel
3. Perform the pressure sensor calibration.
4. Perform the system test.
5. Perform the gas calibration.
6. Connect the catheter to the module. Start measurement by pressing the
the module. Check that th e catheter fills up.
TONO
service menu:
Start-Stop and Lab Data membrane keys.
Start-Stop key on
Stop the measurement by pressing the
on the digit field.
?
2.2.9 Hemodynamic Modules
ECG and RESP measurements
1. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the ECG/RESP board memories have passed the
internal memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ a ll state OK.
2. Check the front panel membrane key ECG LEAD, if available.
3. Check that the ‘Power Freq’ value is set according to the mains power frequency. Correct the
setting, if necessary.
ESTP : ECG
Start-Stop key. Check that Meas. Off text appears
service menu.
4. Connect a 3-lead ECG trunk cable without a lead set to the module. Check that the
message "Leads off" is displayed on the screen.
Document No. 8002949
17
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
5. @ M-ESTPR, M-ETPR , M-ESTR, M-NESTPR, M-NETPR and M-NESTR modules: connect a 5lead ECG cable to the module. Connect the cable leads to a patient simulator.
Check that the ‘Cable type’ shows 5 lead.
@ M-NE12STPR, M-NE12TPR and M-NE12STR modules: connect a 10-leadwire ECG cable
to the module. Connect the cable leads to a patient simulator . Check that the ‘Cable type’
shows 10 lead.
6. Disconnect one of the leads and check that the corresponding electrode in the service
menu shows OFF within 10 seconds from the disconnection, then reconnect the lead.
Check the rest of the leads using the same method.
NOTE: When the ground lead (black) is disconnected all the electrodes should show OFF.
NOTE: The asystole and different leads off -messages are shown using certain priority, Even
though one of the leads is disconnected, the related leads off -message may not appear
onto the screen.
NOTE: When RA, LA or LL electrode is disconnected, all V electrodes show OFF.
NOTE: With NESTPR/ESTPR type modules and 5 lead cable the state of V2, V3, V4, V5 and
V6 electrodes follow the sta te of the V electrode.
7. Check that all ECG and impedance respiration information is shown on the monitor screen
as configured on the simulator.
Check that the pacer count value in the service menu is shown according the simulator
configuration.
Change baseline impedance on the simulator and check that appropriate RESP waveform
and RR -value are shown again within 30 seconds.
Turn the simulator off. Check that the ‘Asystole’ and ‘Apnea” messages are displayed.
?
Modules w/ temperature measurement
8. Enter the
ESTP : STP
service menu.
18
Document No. 8002949
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the STP board memories have passed the internal
memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all show OK.
9. Check the temperature measurement calibration using temperature test plugs.
NOTE: Make sure that the protection for temperature calibration is set on.
?
Modules w/ invasive blood pressure measurem ent
10. Check functioning of the front panel membrane keys.
11. Check the InvBP channels with a patient simulator.
Zero the InvBP channels, then check that the values and waveforms correspond to the
simulator settings.
?
Planned maintenance instructions
Modules w/ SpO2 measurement Check that the message ‘No probe’ is shown when no SpO
connected.
Connect SpO
when the probe is not connected to a finger.
13. Attach the SpO
appear onto the screen.
finger probe to the module. Check that the message ‘Probe off’ is shown
2
probe to your finger. Check that a reading of 95-99 and a pleth waveform
2
?
Modules w/ non invasive blood pressure measurem ent
14. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the NIBP board memories have passed the
internal memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all sh ow
15. Check functioning the front panel membrane keys.
16. Check the NIBP tubing system for leakages by performing
NIBP
module service menu.
OK
.
Active leak test.
sensor is
2
17. Check NIBP calibration by performing
18. Enter the
Check the NIBP watchdog timer activation pressure with a pressure manometer.
19. Check the NIBP watchdog timer by performing
20. Check the safety valve by performing
21. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check the
module identifies the cuff, i.e. the text ‘Adult’ appears in the NIBP digit field for a short time.
Pneumatics
service menu.
Calibration Check.
Watchdog –Test Adult
Safety Valve – Adult
and
Infant.
and
Test Infant
Document No. 8002949
.
19
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
Check that the module gives a reasonable measurement result.
?
2.2.10 EEG Module, M-EEG and EEG Headbox, N-EEG
1. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values in the module view
are not increasing faster than by 50 per second. Check that the memories of the module
have passed the internal memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all state OK.
Check that the ‘HB Mod Timeouts’, ‘HB Mod Bad Checksum’ and ‘Mod HB Bad Checksum’
values are not increasing faster than by 50 per second. Check that the memory of the
headbox has passed the internal memory test, i.e. ‘HB Rom Error’ in the headbox view
states 0.
2. Check the
headbox.
3. Connect the EEG simulator to the headbox . Select 10 kΩ as imped. pos. and imped. neg.
value on the simulator. Go to
headbox view check that the impedances in all four channels are 10 kΩ ±1 kΩ.
4. Select 2 kΩ as imped. pos. and imped. neg. value on the simulator. Select 10 Hz 200 µV
sinewave on the simulator and check that all the four waveforms have same form. Check
that the size of the waveforms is 200 µV
Check that the amp value is 71 µV ±3 µV.
EEG & EP
service menu:
EP Start/Stop and Imp. Check membrane keys both in the module and the
EEG & EP
service menu and select
±5 µV. Check that the MF value is 10 ±0.5 Hz.
pp
Check Electr
. From the
20
Document No. 8002949
5. Plug in the earphones to the headbox.
NOTE: Be careful with loud stimulation from the earphones when starting AEP stimulation.
Start AEP stimulation by pressing the EP Start/Stop button on the module. Check that the
clicking sound comes from the earphones in 1.1 Hz frequency. Stop the stimulation by
pressing again the
EP Start/Stop button on the module. Check that the clicking stopped.
?
2.2.11 Pressure/Pressure Temp Modules, M-P/-PT
Planned maintenance instructions
1. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the PT board memories have passed the internal
memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all show
P/P
module service menu:
?
Invasive blood pressure measurement
2. Check the functioning of the front panel Zero P3 membrane key.
3. Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
?
Modules w/ temperature measurement
OK
.
4. Check the temperature measurement calibration using temperature test plugs.
NOTE: Make sure that the protection for temperature calibration is set on
?
2.2.12 Dual pressure Module, M-PP
1. Enter the PP module service menu:
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the Dual Pressure module, M-PP, memories have
passed the internal memory test, i.e. ‘RAM’, ‘ROM’ a nd ‘EEPROM’ all show
2. Check the functioning of the front panel Zero P5 and Zero P6 membrane keys.
3. Check the InvBP channels with a patient simulator.
Zero the InvBP channels. Then check that the values and waveforms correspond to the
simulator settings.
OK
.
?
21
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
2.2.13 Cardiac Output Modules, M-COP/-COPSv
1. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the module’s memories have passed the internal
memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all show
COP
module service menu:
?
Invasive blood pressure measurement
2. Check the functioning of the front panel membrane Zero P4 key.
3. Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
?
SvO2 measurement
4. Check that the SvO2 values ‘Meas. state’, ‘OM fail’ and ‘OM temp.’ in the
service menu all show
NO OM
OK
.
COP
module
.
5. Checks with the SvO
Turn the SvO
pulse’. Connect the simulator to the module and check that the following messages appear
in the digit field for SvO
Initializing, please wait > Warming up > Not calibrated
Check that ‘Meas. state’ has changed to
NOTE: The ‘OM temp.’ may show
within a half a minute.
6. Perform an In Vitro calibration with the SvO
Check that the calibration date for In Vitro calibration was updated correctly and the SvO
reading on the screen is 81% (±2%).
simulator pulsation switch to ‘Medium’ and the range switch to ‘Normal
2
simulator:
2
:
2
NORMAL
UNSTABLE
2
and ‘OM fail’ and ‘OM temp.’ show OK.
at first, but the message should change to
simulator.
?
OK
2
22
Document No. 8002949
Cardiac Output measurement
Planned maintenance instructions
7. Enter the
Check the front panel
8. Enter the
Others
Connect a catheter connecting cable to module connector C.O.
If the module contains a C.O. Test connector (M-COP), attach the catheter connector of the
connecting cable to the C.O. Test connector. Check that the message ‘Cable OK’ appears in
the menu after the self-test.
No Catheter > Self Test in Progress > Cable OK
9. Check the C.O. measurement with a patient simulator.
?
COP
C.O. View
- C.O. View
2.2.14 NIBP module, M-NIBP
1. Enter the
NIBP
module service menu.
Start C.O. membrane key.
menu:
service menu:
Check that the Timeouts, Bad checksums and Bad c-s by mod values are not increasing
faster than by 50 per second. Check that the NIBP board memories have passed the
OK
internal memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all sh ow
2. Check the front panel LEDs and membrane keys using the
3. Check the NIBP tubing system for leakages by performing
4. Check NIBP calibration by performing
5. Enter the
Check the NIBP watchdog timer activation pressure with a pressure manometer.
6. Check the NIBP watchdog timer by performing
7. Check the safety valve by performing
8. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check the
module identifies the cuff, i.e. the text ‘Adult’ appears on the NIBP digit field for a short time.
Check that the module gives a reasonable measurement result.
Pneumatics
service menu.
Calibration Check
Watchdog –Test Adult
Safety Valve – Adult
Buttons/Leds
Active leak test.
.
and
.
Infant
service menu.
and
Test Infant
.
.
?
23
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
2.2.15 NeuroMuscular Module, M-NMT
1. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the module’s memories have passed the internal
memory test, i.e. ‘RAM’, ‘ROM’ and ‘EEPROM’ all state
2. Check the functioning of the front panel membrane keys Start-Up and Stop/Continue.
3. Check that the message ‘Cable off’ is shown in the digit field and that ‘Cable’ on the service
menu states
Plug the M-NMT Sensor Cable with the M-NMT ElectroSensor into the front panel connector
NMT. Check that the message in the digit field changes to ‘Measurement OFF’ and ‘Cable’ in
the service menu states
4. Place a 3 kΩ resistor between the ElectroSensor’s stimulus electrode leads (brown and
white). Perform
Check that the test was successful, i.e. the ‘Current test (mA):’ in the menu states
OK
and
NMT
module service menu:
OFF
.
Start Curr. Test.
70 OK
.
EMG
and
ELECTR. OFF
OK
.
.
30 OK, 50
5. Connect the M-NMT ElectroSensor leads to the NMT simulator. Set the switch on the
simulator to FADE OFF and turn the knob to MAX. Check ‘Cable’ in the service menu states
EMG
now only
Start the NMT measurement (TOF) by pressing the START-UP key on the module.
Check that the found supramaximal current is less than 70 mA, i.e. the ‘Current set’ value in
the service menu is less than 700.
6. Check that the digit field ‘TOF%’ value is within 95-105, ‘Count’ is 4 and ‘T1%’ is within 95-
105.
7. Check that the ‘Noise’ value in the service menu remains under 50.
1. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check also that the module’s ROM memory has passed the
internal memory test, i.e. the ‘ROM’ shows
2. Check that the message ‘No probe’ is displayed on the screen when no probe is connected.
M-SAT
service menu:
OK
2
.
3. Connect a suitable SpO2 finger probe to the module. Check that the message ‘Check probe’
is displayed on the screen within 30 seconds
4. Attach the SpO
waveform appear.
?
2.2.18 Memory Module, M-MEM
1. Enter the
Check that the module is recognized properly, i.e. ‘Present’ and ‘Active’ state
2. Check that the Memory board memories and the PCMCIA controller have passed the tests.
The status for each should be
3. Select
Check that ‘Interface status’ states
bottom part of the menu are stable.
MemCards
Communication
probe on your finger. Check that a reading of 95-100 and a proper SpO
2
service menu:
YES
.
OK
.
Active
continuously and the error counter values on the
2
Status
4. Select
Insert Menu card and Data card into their slots.
.
25
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
Wait until the information is fully updated in the service menu then check that the Card
types are correct and the ‘File system ’ states ATA
Check that the rest of the information is reliable and no errors have bee detected.
?
2.2.19 Recorder Module, M-REC
1. Open the paper compartment cover. Check that the message ‘Recorder: Cover open’
appears on the screen, then close the cover.
2. Press the Record Wave key and check that the module starts recording the selected
waveforms. Press the
Press the Print Trends key and check that the module starts recording trends. Press the
Stop key on the module to stop recording.
Stop key on the module to stop recording.
3. Check that the quality of the recordings is acceptable.
?
2.2.20 Network Board, B-NET and UPINET board, B-UPI4NET
1. Check that the Mon-Net cable connector and the Identification plug are clean and intact,
then connect them to the Network/UPI4NET Board.
2. Check that the monitor connects to the S/5 Network, i.e. the network symbol appears under
the clock on the upper right-hand corner of the screen. Also a message regarding the
connected Central should appear in the message field on the screen.
NOTE: If necessary, the monitor's network communication should be reselected according
to the used network software in the
3. Enter the
Check that the counters for data errors (‘CRC’, ‘Frame’, ‘Transm.’) are stable.
Check that the counters for hardware errors (‘Intern.’, ‘Missed’, ‘FIFO’, ‘Overrun’) all show 0.
Network
service menu:
Network
service menu..
?
2.2.21 Interface Board/Module, B-INT/M-INT
1. Enter the
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check that the memories have passed the internal memory
26
Document No. 8002949
Interface
service menu:
test, i.e. ‘RAM’ and ‘ROM’ state
?
2.2.22 Device Interfacing Solution, N-DISxxx
OK
Planned maintenance instructions
.
2.2.23 General
1. Enter t he
Check that the DIS module ‘tout' and ‘cse’ values do not increase faster than by 50 per
second. Check also that the DIS module memories have passed the internal memory test, i.e.
‘Ram’, ‘Rom’ and 'EEPROM' state all OK.
Perform the same check for all connected DIS modules.
DIS Interfacing
service menu:
?
1. Check that the monitor is capable of storing the trend information and temporary settings in
a short (max. 15 minutes) standby situation with no power cord.
2. Check the Service Reset switch. Switch the monitor to standby and press the Service Reset
switch for at least five seconds. Switch the monitor back on and check that the monitor
performs a ‘Cold Start.’, i.e. all trend information is cleared.
3. Enter the
Check that the monitor restarts.
Set/Test
service menu and perform
WD by GSP
4. Enter the
Clear the content of the Service Log by selecting
Service Log
service menu:
Reset Log
from the menu.
?
5. Perform an Electrical safety check and a leakage current test.
Check that the monitor and all connected units function normally after the performed test
6. Save information about the performed planned maintenance into the
menu by performing
Plan. Maint - 1 year PM - Save
Maintenance
?
7. Switch the monitor to standby and perform final cleaning.
Fill in all necessary documents.
?
service
27
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
28
Document No. 8002949
APPENDIX A, Planned maintenance check form, S/5 AM, S/5 CCM
APPENDIX A
29
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
30
Document No. 8002949
APPENDIX A, Planned maintenance check form, S/5 AM, S/5 CCM
)
PLANNED MAINTENANCE CHECK FORM
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
Customer
Service
Service engineerDate
Monitor installation
F-CU8
S-L-M-MB-M-M-FB-M-M-NB-B-M-
OK = Test OKN.A. = Test not applicableFail = Test Failed
Visual inspection/preparation
2.1.1. General
2.1.2. Extension Frame, F-EXT4
2.1.3. Airway Module, G-XXXXX
2.1.4. Compact Airway Module, M-CXXXXX
D-K-G-
S/N
S/N
S/N
OKN.A.Fail
2.1.5. Tonometry Module
2.1.6. Recorder module
2.1.7. Modules with NIBP measurement
Notes
S/N
S/N
A-1(4
Document No. 8002949
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Critical Care Monitor
A
Functional Inspection
2.2.1. General
2.2.2. Display(s)
2.2.3. Keyboard(s)
OKN.A.Fail
2.2.4. 8-Module Frame, F-CU8
2.2.5. Extension Frame, F-EXT4
Notes
2.2.6. Airway Module, G-XXXX
Agent identification
Patient Spirometry
General
Notes
2.2.7. Compact Airway Module
Agent identification
S/N
Patient Spirometry
General
Notes
2.2.8. Tonometry Module, M-TONO
Notes
2.2.9. Hemodynamic Modules
ECG and RESP measurements
Temperature measurement
-2(4)
Document No. 8002949
S/N
APPENDIX A, Planned maintenance check form, S/5 AM, S/5 CCM