Datex-Ohmeda S-5 Service Manual M-NE12STPR

S/5
S/5 S/5
Datex-Ohmeda Hemodynamic modules
TM
NE12STPR Module, M-NE12STPR (rev. 01)
TM
NE12STR Module, M-NE12STR (rev. 01)
TM
NE12TPR Module, M-NE12TPR (rev. 01)
TM
S/5
NESTPR Module, M-NESTPR (rev. 01)
TM
S/5 S/5
S/5
NESTR Module, M-NESTR (rev. 01)
TM
NETPR Module, M-NETPR (rev. 01)
TM
TM
S/5 S/5
ESTR Module, M-ESTR (rev. 04)
TM
ETPR Module, M-ETPR (rev. 04)

Technical Reference Manual Slot

Datex-Ohmeda Inc. 3030 Ohmeda Drive 53707-7550 MADISON, WIS USA Tel. +1-608-221 1551,Fax +1-608-222 9147
www.us.datex-ohmeda.com
All specifications are subject to change without notice.

Document No. 800 1008-2

June 2001

Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.

Table of contents

TABLE OF CONTENTS
HEMODYNAMIC MODULES
TABLE OF CONTENTS i
Table of figures iii
Introduction 1
1 Specifications 2
1.1 General specifications ..............................................................................................................................2
1.2 Typical performance .................................................................................................................................2
1.2.1 NIBP................................................................................................................................................2
1.2.2 ECG.................................................................................................................................................3
1.2.3 Pulse oximetry..................................................................................................................................3
1.2.4 Temperature.....................................................................................................................................4
1.2.5 Invasive blood pressure ....................................................................................................................4
1.2.6 Respiration ......................................................................................................................................4
1.3 Technical specifications............................................................................................................................5
1.3.1 NIBP................................................................................................................................................5
1.3.2 ECG.................................................................................................................................................5
1.3.3 Pulse oximetry..................................................................................................................................6
1.3.4 Temperature.....................................................................................................................................6
1.3.5 Invasive blood pressure ....................................................................................................................6
1.3.6 Respiration ......................................................................................................................................6
2 Functional Description 7
2.1 Measurement principle .............................................................................................................................7
2.1.1 NIBP................................................................................................................................................7
2.1.2 ECG.................................................................................................................................................7
2.1.3 Pulse oximetry..................................................................................................................................7
2.1.4 Temperature.....................................................................................................................................9
2.1.5 Invasive blood pressure ..................................................................................................................10
2.1.6 Respiration ....................................................................................................................................10
2.2 Main components...................................................................................................................................10
M-ESTPR/-ETPR/-ESTR modules...............................................................................................................10
M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules .................................................11
2.2.3 NIBP board ....................................................................................................................................12
2.2.4 ECG board in 3-and 5-lead measurement........................................................................................14
2.2.5 ECG board in 12-lead measurement ...............................................................................................16
2.2.6 ECG filtering...................................................................................................................................18
2.2.7 STP board ......................................................................................................................................19
2.3 Connectors and signals...........................................................................................................................24
2.3.1 Module bus connector....................................................................................................................24
2.3.2 Front panel connectors...................................................................................................................25
2.3.3 Test points on boards .....................................................................................................................26
3 Service Procedures 28
3.1 General service information.....................................................................................................................28
3.2 Service check .........................................................................................................................................28
Document No. 800 1008-2
i
Datex-Ohmeda S/5 monitors
3.2.1 Recommended tools......................................................................................................................28
3.2.2 Recommended parts......................................................................................................................29
3.3 Disassembly and reassembly..................................................................................................................39
3.3.1 M-ESTPR, M-ESTR, and M-ETPR modules........................................................................................39
3.3.2 M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules.......................................39
3.4 Adjustments and calibrations..................................................................................................................40
3.4.1 Pressure safety level detection “OFFSET”.........................................................................................40
3.4.2 NIBP calibrations ...........................................................................................................................40
3.4.3 Temperature calibration .................................................................................................................42
3.4.4 Invasive pressure calibration ..........................................................................................................42
4 Troubleshooting 43
4.1 Troubleshooting charts ...........................................................................................................................43
4.1.1 NIBP..............................................................................................................................................43
4.1.2 NIBP error code explanation ...........................................................................................................46
4.1.3 ECG...............................................................................................................................................47
4.1.4 Pulse oximetry (SpO
4.1.5 Temperature ..................................................................................................................................48
4.1.6 Invasive blood pressure..................................................................................................................49
4.1.7 Impedance respiration ...................................................................................................................50
4.2 Troubleshooting flowcharts .....................................................................................................................51
4.2.1 M-NE12STPR and M-NESTPR module troubleshooting.....................................................................51
4.2.2 M-ESTPR, M-ESTR, and M-ETPR module troubleshooting .................................................................52
).....................................................................................................................47
2
5Service Menu 53
5.1 NIBP service menu .................................................................................................................................54
5.1.1 NIBP demo menu...........................................................................................................................55
5.1.2 NIBP calibration menu....................................................................................................................56
5.1.3 NIBP safety valve menu..................................................................................................................57
5.1.4 NIBP pulse valve menu...................................................................................................................58
5.1.5 NIBP buttons/leds menu................................................................................................................59
5.1.6 NIBP pneumatics menu..................................................................................................................60
5.1.7 NIBP watchdog menu.....................................................................................................................61
5.2 ECG service menu ..................................................................................................................................62
5.2.1 ECG setup menu ............................................................................................................................64
5.3 STP service menu ...................................................................................................................................65
5.3.1 STP calibration menu .....................................................................................................................67
6 Spare Parts 68
6.1 Spare parts list.......................................................................................................................................68
6.1.1 M-ESTP rev. 01, M-ETP rev. 00, M-EST rev. 00 .................................................................................68
6.1.2 M-ESTP rev. 02, M-ETP rev. 01, M-EST rev. 01 .................................................................................69
6.1.3 M-ESTP rev. 03, M-ETP rev. 02, M-EST rev. 02 .................................................................................69
6.1.4 M-ESTP rev. 04, M-ETP rev. 03, M-EST rev. 03 .................................................................................69
6.1.5 M-ESTP rev. 05, M-ETP rev. 04, M-EST rev. 04 .................................................................................69
6.1.6 M-ESTPR rev. 01, M-ETPR rev. 01, M-ESTR rev. 01...........................................................................70
6.1.7 M-ESTPR rev. 02, M-ETPR rev. 02, M-ESTR rev. 02...........................................................................70
6.1.8 M-ESTPR rev. 03, M-ETPR rev. 03, M-ESTR rev. 03...........................................................................70
6.1.9 M-ESTPR rev. 04, M-ETPR rev. 04, M-ESTR rev. 04...........................................................................70
6.1.10 M-NESTPR rev. 00, M-NETPR rev. 00, M-NESTR rev. 00 ...............................................................71
6.1.11 M-NESTPR rev. 01, M-NETPR rev. 01, M-NESTR rev. 01 ...............................................................72
ii
Document No. 800 1008-2
Table of contents
6.1.12 M-NE12STPR rev. 00, M-NE12STR rev. 00, M-NE12TPR rev. 00...................................................72
6.1.13 M-NE12STPR rev. 01..................................................................................................................72
6.1.14 Front panel stickers....................................................................................................................73
6.1.15 Front panel stickers for S/5 modules...........................................................................................74
7 Earlier Revisions 76
APPENDIX A 77
Service Check Form A-1

TABLE OF FIGURES

Figure 1 S/5 NE12STPR Module, M-NE12STPR ................................................................................1
Figure 2 Absorption of infrared light in the finger probe parts layout and schematic diagram................9
Figure 3 Front panel of M-ESTPR.....................................................................................................10
Figure 4 Front panel of M-NESTPR ..................................................................................................11
Figure 5 NIBP board functional block diagram.................................................................................12
Figure 6 3- and 5- lead ECG board block diagram ...........................................................................14
Figure 7 12-lead ECG measurement block diagram.........................................................................16
Figure 8 STP board block diagram...................................................................................................19
Figure 9 Temperature measurement principle .................................................................................20
Figure 10 Pressure measurement principle .......................................................................................20
Figure 11 Pulse oximetry measurement block diagram ......................................................................21
Figure 12 Serial communication and opto isolation of M-NESTPR/-NE12STPR ...................................22
Figure 13 Serial communication and opto isolation of M-ESTPR.........................................................23
Figure 14 Module bus connector (X1) pin layout ................................................................................24
Figure 15 M-NE12STPR and M-NESTPR module troubleshooting flowchart .........................................51
Figure 16 M-ESTPR Module Troubleshooting Flowchart ......................................................................52
Document No. 800 1008-2
iii
Datex-Ohmeda S/5 monitors
This page intentionally left blank.
iv
Document No. 800 1008-2

INTRODUCTION

This Technical Reference Manual Slot provides information for the maintenance and service of the hemodynamic modules. Please see also related Technical Reference Manual for information related to system e.g. related documentation, conventions used, symbols on equipment, safety precautions, system description, system installation, interfacing, functional check and planned maintenance.
The S/5 M-ESTPR/-ESTR/-ETPR and S/5 M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/­NESTR/-NETPR are double width modules designed for use with S/5 monitors. The modules provide general hemodynamic parameters. Later in this manual modules can be called w/o system name S/5.
S/5 Hemodynamic modules
NOTE: Do not use identical modules in the same monitor simultaneously. The following modules are considered identical: M-ESTP/-EST/-ETP M-ESTPR/-ESTR/-ETPR M-NESTPR/-NESTR/-NETPR M-NE12STPR/-NE12STR/-NE12TPR M-MRI/-MRIP
NOTE: 12-lead ECG measurement requires Display Controller, B-DISP.
Figure 1 S/5 NE12STPR Module, M-NE12STPR
Table 1 Options of S/5 hemodynamic modules
Parameter NE12STPR NESTPR NE(12)
STR
12
12-lead ECG (•)(•)
N
NIBP •••
E
ECG • •••••
S
Pulse oximetry •••
T
Two temperatures • •••••
P
Two invasive blood pressures ••
R
Impedance respiration • •••••
NE(12) TPR
ESTPR ESTR ETPR
Document No. 800 1008-2
1
Datex-Ohmeda S/5 monitors

1 SPECIFICATIONS

1.1 General specifications
Module size 75 × 180 × 112 mm W × D × H3.0 × 7.1 × 4.4 in Operation temperature 10...40 °C / 50...104 °F
@ M-ESTPR/-ETPR/-ESTR
Module weight 0.6 kg / 1.3 lbs Power consumption 6 W
@ M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR
Module weight 1 kg Power consumption about 9 W
1.2 Typical performance
1.2.1 NIBP
Oscillometric measurement principle.
Measurement range adult 25...260 mmHg
Pulse rate range accepted 30...250 bpm
Measurement interval from continuous to 1h, 2h, 4h
Typical measuring time adult 23 s
Initial inflation pressure adult 185 ±10 mmHg
Venous stasis adult 80 ±10 mmHg / 2 min.
Cuff widths please see User’s Guide
child 25...195 mmHg infant 15...145 mmHg
infant 20 s
child 150 ±10 mmHg infant 120 ±10 mmHg
child 60 ±10 mmHg / 2 min. infant 40 ±10 mmHg / 1 min.
2
Document No. 800 1008-2
1.2.2 ECG
S/5 Hemodynamic modules
Lead selection @ 12-lead ECG I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 Lead selection @ other modules I, II, III, aVR, aVL, aVF, V Sweep speeds 12.5, 25, 50 mm/sec
DISPLAY FILTER
Diagnostic @ 12-lead ECG 0.05...150 Hz Diagnostic @ other modules 0.05...100 Hz
Monitoring 0.5...30 Hz (-3 dB, with 50 Hz reject filter)
0.5...40 Hz (-3 dB, with 60 Hz reject filter)
ST filter 0.05...30 Hz (-3 dB, with 50 Hz reject filter)
0.05...40 Hz (-3 dB, with 60 Hz reject filter)
HEART RATE FROM ECG
Range 30...250 bpm Accuracy ±5 bpm or ±5 %, whichever is greater Resolution 1 bpm Update interval 5 s Averaging time 10 s
ST LEVELS (in main software)
ST level range -9...+9 mm (-0.9...+0.9 mV) Resolution 0.1 mm (0.01 mV) Averaging calculated from 8 QRS
SYNCHRONIZATION
Direct ECG analog output of ECG, 1 V/1 mV Pacer 5 V and 0.5...2.5 ms pulse, < 30 ms after pacer peak Defibrillator 5 V and 10 ms pulse, < 35 ms after R-point synchronization
1.2.3 Pulse oximetry
Measurement range 40...100 % Accuracy 100...80 %, ±2 digits
(% SpO
Display resolution 1 digit = 1 % of SpO Display averaging time 20, 10 sec, beat-to-beat Pulse beep pitch varies with SpO
The monitor is calibrated over the measurement range against functional saturation SpO
HEART RATE FROM PLETH
Measurement range 30...250 bpm Accuracy 30...100, ±5 bpm,
Resolution 1 bpm Display averaging 10 s
2 ±1 SD)
1
80...50 %, ±3 digits
50...40 %, unspecified
2
2 level
func.
2
100...250, ±5 %
1
1 SD (standard deviation) = 68 % of all readings in the specified range in stable conditions.
3
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
Adjustable pulse beep volume.
PLETH WAVEFORM
Scales 2, 5, 10, 20, 50 mod%, Auto
Start up scale is 20 mod% if AUTO is not selected to be the default setting.
1.2.4 Temperature
Measurement range 10...45 °C (50...113 °F) (In rev. ESTP 03/ EST 02/ETP 02 or earlier: 15...45 °C (59...113 °F))
Measurement accuracy ±0.1 °C (25...45.0 °C)
Display resolution 0.1 °C (0.1 °F) Temperature test automatic (every 10 min) Probe type compatible with YSI 400 series
1.2.5 Invasive blood pressure
Measurement range -40...320 mmHg Measurement accuracy ±2 mmHg or ±5 % Zero adjustment range ±150 mmHg Calibration range ±20 % Scales upper limit is adjustable between 10 and 300 mmHg in steps of
±0.2 °C (10...24.9 °C)
10. Lower limit is 10 % of selected upper limit below zero.
Sweep speed 12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range -40...320 mmHg Resolution ±1 mmHg
WAVEFORM DISPLAY
Range -30...300 mmHg
HEART RATE FROM ARTERIAL PRESSURE
Measurement range 30...250 bpm Resolution 1 bpm Accuracy ±5 bpm or ±5 % whichever is greater
1.2.6 Respiration
Measurement range 4...120 bpm Accuracy ±5 bpm or ±5 % Resolution 1 bpm Averaging time 30 s Update interval 10 s
RESPIRATION WAVEFORM
Sweep Speeds 6.25 mm/s and 0.625 mm/s
4
Document No. 800 1008-2
1.3 Technical specifications
1.3.1 NIBP
Deflation rate, PR dep. 5...13 mmHg/s Inflation time 20...185 mmHg, 1...5 s
Automatic software control, max. inflation pressure
Over pressure limit, stops measurement after 2 seconds
Safety valve limits the maximum cuff pressure to 320 mmHg in adult/child mode or 165 mmHg in infant mode. Independent timing circuit limits pressurizing (>5 mmHg) time to 2 minutes 10 seconds maximum in adult/child mode, and 1 minute 5 seconds in infant mode.
S/5 Hemodynamic modules
adult 280 ±10 mmHg child 200 ±10 mmHg infant 150 ±10 mmHg
adult 320 mmHg child 220 mmHg infant 165 mmHg
1.3.2 ECG
Zeroing to ambient pressure is done automatically.
Inflation pressure is adjusted according to the previous systolic pressure, typically 40 mmHg above. If the systolic pressure is not found, inflation pressure is increased typically 50 mmHg.
Max. measurement time adult 2 min
child 2 min infant 1 min
Pressure transducer accuracy is better than ±3 mmHg or ±2 % whichever is greater. Max. error ±4 mmHg.
Protection against electrical shock Type BF defibrillation proof
Defibrillation protection 5000 V, 360 J Recovery time 2 s Input impedance >2.5 M (10 Hz) CMRR >100 dB (ST) System noise <40 µV (p-p, RTI) Allowable offset ±300 mVDC Gain range 0.2...5.0 cm/mV Pacemaker pulse detection 2...500 mV, 0.5...2 ms pulses
Protection against electrical shock Type CF defibrillator proof
5
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
1.3.3 Pulse oximetry
Protection against electrical shock Type BF defibrillation proof
1.3.4 Temperature
Measurement accuracy ±0.1 °C (25.0...45.0 °C)
Protection against electrical shock Type CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on transducer/probe used. Please check the transducer/probe specification.
1.3.5 Invasive blood pressure
DIGITAL DISPLAY AVERAGING
Digital displays Art and P1 are averaged over 5 seconds and updated at 5 seconds intervals. All other pressures have respiration artifact rejection.
±0.2 °C (10.0...24.9 °C)
Accuracy ±5 % or ±2 mmHg, whichever is greater Transducer and input sensitivity
Filter 0...4 - 22 Hz adjustable Zero set accuracy ±1 mmHg Calibration resolution ±1 mmHg Zero time less than 15 s Protection against electrical shock Type CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on transducer/probe used. Please check the transducer/probe specification.
1.3.6 Respiration
Excitation frequency @ 12-lead ECG 62.5 kHz
Excitation frequency @ other modules 31.25 kHz
Breath detection automatic, range 0.3...6 manually adjustable minimum
Input dynamic range 0.2...6 Input impedance range 100...5000 Respiration Rate min. 4 bpm
Lead off detection >3 M
5 µV/V/mmHg, 5 VDC, 20 mA max current
detection: 0.2, 0.4, 0.6, 0.8, 1.0
max. 120 bpm
6
Document No. 800 1008-2

2 FUNCTIONAL DESCRIPTION

2.1 Measurement principle
2.1.1 NIBP
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a speed based on the patient’s pulse, collecting data from the oscillations caused by the pulsating artery. Based on these oscillations, values for systolic, mean, and diastolic pressures are calculated.
The following parts are necessary for the NIBP measurement:
M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR (or M-NIBP) module
S/5 Hemodynamic modules
twin hose (adult or infant model)
blood pressure cuffs (different sizes)
2.1.2 ECG
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical potential produced with electrodes placed on the surface of the body.
ECG reflects:
electrical activity of the heart
normal/abnormal function of the heart
effects of anesthesia on heart function
effects of surgery on heart function
See the User's Reference Manual for electrodes positions and other information.
2.1.3 Pulse oximetry
A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near infrared (about 900 nm) and the other in the red region (about 660 nm) of light spectrum. These wavelengths are emitted by LEDs in the SpO tissue and is finally detected by a PIN-diode opposite to LEDs in the probe. Pulse oximeter derives the oxygen saturation (SpO absorption at the two wavelengths and the arterial oxygen saturation SaO
probe, the light is transmitted through peripheral
2
) using empirically determined relationship between the relative
2
.
2
In order to measure the arterial saturation accurately, pulse oximeters use the component of light absorption giving variations synchronous with heart beat as primary information on the arterial saturation.
Document No. 800 1008-2
7
Datex-Ohmeda S/5 monitors
A general limitation of the above pulse oximetry principle is that due to only two wavelengths used only two hemoglobin species can be discriminated by the measurement.
The modern pulse oximeters are empirically calibrated either against fractional saturation
frac;
SaO
2
2
fracSaO
2
=
HbO
2
++
binDyshemogloHbHbO
Formula 1
or against functional saturation SaO
HbO
=
2
funcSaO
which is more insensitive to changes of carboxyhemoglobin and methemoglobin concentrations in blood.
The oxygen saturation percentage SpO the functional saturation SaO measurement relative to SaO2func can be maintained even at rather high concentrations of carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeter is not able to correctly measure oxygen content of the arterial blood at elevated carboxyhemoglobin or methemoglobin levels.
Plethysmographic pulse wave
The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation at the measuring site. Thus the amplitude of the waveform represents the perfusion.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The signals are filtered to reduce noise and checked to separate artifacts.
func;
2
2
HbHbO
2
+
measured by Datex-Ohmeda module is calibrated against
2
func. The advantage of this method is that the accuracy of SpO
2
Formula 2
2
Probe
8
Document No. 800 1008-2
The standard probe is a finger clamp probe which contains the light source LEDs in one half and the photodiode detector in the other half. Different kinds of probes are available from Datex­Ohmeda.
Intensity of transmitted light
I
max (DC-component)
S/5 Hemodynamic modules
I
max
AC-component
I
min
Transmitted light
Incident light
Emitter
RED
Detector
No pulsation
IRED
Pulsatile blood
SpO sensor cable
2
Variable absorption due to pulse added volume of arterial blood
Arterial blood
Venous blood
Tissue
Time
SpO sensor connector
2
6
GND
7
I
LED
4
5
GND
8
V
B
R
C
1
I
S
9
GND
Figure 2 Absorption of infrared light in the finger probe parts layout and schematic
2.1.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature changes, called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
Applying a constant voltage across the resistor and measuring the current that flows through
diagram
it
9
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
Applying a constant current through the resistor and measuring the voltage that is generated
across it.
In Datex-Ohmeda modules the two methods are combined in a form of a voltage divider. The NTC­resistor is connected in series with a normal resistor and a constant voltage is applied across them. The temperature dependent voltage can be detected at the junction of the resistors, thus producing the temperature signal from the patient. The signal is amplified by analog amplifiers and further processed by digital electronics.
2.1.5 Invasive blood pressure
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal saline all connected together by stopcocks, is attached to the catheter. The transducer is placed at the same level with the heart, and is electrically zeroed.
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
2.1.6 Respiration
Impedance respiration is measured across the thorax between ECG electrodes. The signal of the respiration is made by supplying current between electrodes and by measuring the differential current from the electrodes. The signal measured is the impedance change caused by breathing. From these impedance changes, respiration rate is calculated, and respiration waveform is displayed on the screen.
2.2 Main components
2.2.1 M-ESTPR/-ETPR/-ESTR modules
T1
T2
ECG+ Resp
SpO
P1
P2
2
ECG
Lead
Start
Wedge
Zero
P1
Zero
P2
10
Document No. 800 1008-2
Figure 3 Front panel of M-ESTPR
The M-ESTPR, M-ETPR, and M-ESTR modules contain two main PC boards, the STP board and the ECG board. They work independently. Both of them have their own processor and software EPROM. Some components on the boards are not used in ETPR and ESTR modules.
In M-ESTPR module, additionally, there are two small boards, the SP input and the ECG input
S/5 Hemodynamic modules
boards, attached to the front panel of the module. The front panel has six connectors and four keys. The connectors are two for temperature measurement, two for invasive blood pressure measurement, one for ECG, and one for SpO
measurement. The keys are for ECG lead, Start
2
Wedge, P1 zero, and P2 zero.
In M-ETPR module, there are two small boards, the ECG input board and the 2P input board attached to the front panel of the module. The front panel has five connectors and four keys. The connectors are two for temperature measurement, two for invasive blood pressure measurement, and one for ECG
measurement. The keys are for ECG lead, Start Wedge, P1 zero, and P2 zero.
In M-ESTR module, there are two small boards: the S input board and the ECG input board, attached to the front panel of the module. The front panel has four connectors and one key. The connectors are two for temperature measurement, one for ECG, and one for SpO The key is for ECG lead select.
NOTE: M-ESTP rev. 03, M-ETP rev. 02 and M-EST rev. 02 and all earlier revisions have separate T and SP input boards.
2.2.2 M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules
measurement.
2
Auto
On/Off
NIBP
T1
ECG+ Resp
P1
Start
Cancel
P2
T2
Zero
SpO
2
P1
Zero
P2
Figure 4 Front panel of M-NESTPR
The M-NESTPR, M-NETPR, and M-NESTR modules contain three main PC boards, the STP board, the ECG board, and the NIBP board. They work independently. Each of these has their own processor and software EPROM.
The M-NE12STPR, M-NE12TPR, and M-NE12STR contain three main PC boards, The STP board, the ECG board and the NIBP board. They work independently. Each of them has their own processor. The STP board and NIBP board have software EPROM. In the ECG board the software is in flash memory. The STP and NIBP boards are the same as in M-NESTPR module but the ECG board and ECG input board are different.
In M-NESTPR module, there are two small boards, the SP input and the ECG input board attached to the front panel of the module. The front panel has seven connectors and four keys. The connectors are two for temperature measurement, two for invasive blood pressure measurement, one for ECG, one for NIBP, and one for SpO
measurement. The keys are for NIBP Auto On/Off,
2
NIBP Start/Cancel, P1 zero, and P2 zero. The structure of M-NE12STPR is similar except the ECG board and ECG input board are different.
Document No. 800 1008-2
11
Datex-Ohmeda S/5 monitors
In M-NETPR module, there are two small boards, the 2P input board and the ECG input board, attached to the front panel of the module. The front panel has six connectors and four keys. The connectors are two for temperature measurement, two for invasive blood pressure measurement, one for ECG, and one for NIBP. The keys are for Auto On/Off, Start/Cancel, P1 zero, and P2 zero. The structure of M-NE12TPR is similar except the ECG board and ECG input board are different.
In M-NESTR module, there are two small boards, the ECG input board and the S input board, attached to the front panel of the module. The front panel has five connectors and two keys. The connectors are two for temperature measurement, and one for SpO and one for NIBP. The keys are for Auto On/Off, Start/Cancel. The structure of M-NE12STR is similar except the ECG board and ECG input board are different.
2.2.3 NIBP board
CUFF
ZERO VALVE
EXHAUST VALVE 2
JOIN ING CHAMBER
EXHAUST VALVE 1
BLEED VALVE
CHECK VALVE
measurement, one for ECG,
2
PUMP and
PUMP
M
VALVE DRIVER
B1 B2
t o
/ f r o
m m
o d u l
e b
u s
X1
PRESSURE
S
TRAN
-
D
UCERS
AD-CON­VERTER
POWER-UP
RESET
RS485 INTER F.
EEPRO (C alib­ration Data)
Figure 5 NIBP board functional block diagram
Pressure transducers
OVERPRESSURE CONTROL
M
Write protect switch
CPU 80C51FA
NTERNAL
I WATCHDOG
ADDRESS BUS
ADDRESS DECODER
ADDRESS LATCH
WATCHDO G
TIMER
RAM
DATABUS
+
1
5
EPROM
D
V
FRONT PANEL
KEYS
SOFT ­WAR E
CON TRO L
12
Document No. 800 1008-2
The NIBP board contains two pressure transducers. They are of piezoresistive type. One is used for measuring the pressure of the blood pressure cuff and the pressure fluctuations caused by arterial wall movement (B1). The other is used for detection of cuff hose type, cuff loose and cuff occlusion situations etc. (B2). The transducers are internally temperature compensated. They are supplied by a constant voltage and their output voltage changes up to 40 mV max. (50 kPa, 375 mmHg).
Signal processing
Two signals from the pressure transducers are amplified and sent to A/D converter. After the converter, digitized signals are sent to microprocessor for data processing. Before the converter, one of the signals is used to adjust the offset to the pressure safety level.
The NIBP board is controlled with 80C51FA microprocessor at 16 MHz oscillator frequency.
Memory
NIBP program memory (EPROM) size is 128k × 8. RAM size is 32k × 8 bit and it stores variable values in NIBP measurement. EEPROM is size 64 × 16 bit and is used to store the calibration values for the pressure transducers, the pulse valve constants gained during measurements, the PC board identification, and module serial number.
Software control
Software controls valves and pump. In addition to the individual on/off signals for each component there is a common power switch for the valves and the pump that can be used at pump/valve failures.
S/5 Hemodynamic modules
In addition to external RS485 reset line the microprocessor system is equipped with its own power­up reset. See the section in ECG board’s description: “RS485 communication”
Watchdog timer
The NIBP board is equipped with software independent safety circuit to disconnect supply voltages from the pump and the valves if the cuff has been pressurized longer than preset time. As soon as the cuff pressure rises over a specifiedpressure limit, timer starts counting. The timer is adjusted to stop the pump and open the valves in 2 minutes 10 seconds in adult/child mode and in 1 minute 5 seconds in infant mode.
Valves
Exhaust valves are used for emptying the cuff and the joining chamber after the measurement. Exhaust valve 1 is also used as safety valve in infant mode. Valve opens at 165 mmHg. Exhaust valve 2 is also used as safety valve in adult mode and opens at 320 mmHg. Bleed valve is used for emptying the cuff during measurement. Zero valve is used for connecting the pressure transducer B1 to open air.
Power supply section
All connections are established via 25-pin connector (D-type, female). The module needs +5 V, ±15 V, and +15 VD (dirty) power supply to operate. The pump and the valves use separate +15 VD power line. The supply voltages are generated in the power supply section of the S/5 monitor. The reference voltages ±5 V
and +10 V
ref
are generated on the NIBP board.
ref
13
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
2.2.4 ECG board in 3-and 5-lead measurement
Patient signals are connected to overload protection circuits (resistors and gas-filled surge arresters) and analog switches to instrumentation amplifiers. Then the signals are amplified by 480 and limited by slew rate. Then they are A/D-converted, analyzed and transferred to module bus in digital form.
PATIENT
OVERLOAD PROTECTION
DEFIBRILLA­TION DETECTOR
uP
SLEW­RATE DETECTOR
HP Filter
uP
ANTI­ALIAS
RL
ANALOG SWITCHES
SLEW­RATE DETECTOR
uP
ANTI­ALIAS
SUPPLY VOLTAGES
FCN
SLEW­RATE DETECTOR
uPuPuP
uP
HP Filter
ANTI­ALIAS
A/D
MICRO­PROCESSOR (uP)
HP Filter
SLEW­RATE DETECTOR
uP
uP
HP Filter
ANTI­ALIAS
LEAD SELECTION LOGIC Controlled by ECG lead selection signals from uP
Resp amplifier
31 kHz from uP
Sync. Rectifier
Amplifier
Figure 6 3- and 5- lead ECG board block diagram
Analog ECG section
ECG cable is connected to connector pins E1 to E6 on the input board which contains an overload protection circuit. Leads are connected to amplifiers via analog switches. States of the switches depend on the cable type. Lead-off, noise and pacemaker are detected by a slew rate detector.
14
Document No. 800 1008-2
ISOLATION
POWER SOURCE
POWER SOURCE
OPTO­COUPLER
OPTO­COUPLER
to STP Board
DIRECT ECG
DIRECT ECG
S/5 Hemodynamic modules
Lower frequency is determined by high pass (HP) filter 0.5 Hz (monitor bandwidth) or 0.05 Hz (diagnostic or ST- bandwidth).
Respiration section
3-lead cable The analog switches control the current supply source of the impedance respiration measurement,
and the lead selection for the 3-lead cable can be seen from the following table:
Table 2 Lead selection and coding for the 3-lead cable
Selected lead Current source between Signal measured from
IR - L N
II R - N L
III L - N R
Position on
IEC standard coding AAMI standard coding
body surface
right arm R = red RA = white
left arm L = yellow LA = black
left leg F = green LL = red
5-lead cable When the 5-lead cable is used, the current source is between L-F and the signal is measured from
the N, independently on the lead selection.
The respiration amplifier consist of the operational amplifiers, and the components around them. There is an analog switch for controlling the gain of the first stage of the preamplifier. Synchronous rectifier consists of the analog switches, which are used for detecting the respiration signal from 31 kHz amplitude modulated raw signal. The amplifier stage consists of the differential amplifier and the last amplifier. The differential amplifier consists of the operational amplifiers and the components around them. This stage is AC-coupled on both sides for minimising the offset voltages. The last amplifier is used for amplifying the signal derived from differential amplifier stage. The respiration signal is zeroed at the beginning of the measurement. Zeroing is also used for fast recovering the measurement after the motion artefact. This is done in amplifier section.
NOTE: The respiration measurement is switched OFF for 20 seconds when defibrillation is detected at the defibrillation detector.
Microprocessor section
Microprocessor contains RAM and EPROM memories. The processor uses external EEPROM memory. The microprocessor’s internal 8-channel A/D-converter converts the ECG-signals to digital form. See the section in ECG board’s description: “RS485 communication
Serial communication
Communication with the module bus is made through RXD and TXD pins. See the section in STP board’s description: “Serial communication”.
15
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
Isolated section
The patient isolation of ECG is 5 kV.
NOTE: The isolation has been changed from the earlier revisions.
WARNING Do not touch battery operated monitor during defibrillation procedure.
See the “Isolated section” in STP board description.
Power supply section
See the “Power supply section” in STP board description. There is a test connector (X20) on the board for voltages +5 VREF, +5 V, +12 V, GND and -12 V.
2.2.5 ECG board in 12-lead measurement
The 12-lead ECG measurement consists of the functions, which are shown in the figure 7. All functions are located in the ECG board except the front panel connector and the ECG input board.
Front panel connector and ECG input board
Input protection and filtering
Respiration impedance amplifiers
RS 485 communication
Respiration impedance supply
ECG CPU
Power supply
ECG preamplifiers
ECG amplifiers
and
Baseline
restoration
NV memory
Pacer
detection
16
Document No. 800 1008-2
Isolation
to STP board
Isolation
Figure 7 12-lead ECG measurement block diagram
Front panel connector and ECG input board
The connector for the 12-lead ECG cable is a green 12 pin Nicolay type connector. 3- or 5-lead cables with blue connector cannot be connected to this connector. The ECG input board contains high voltage resistors and a connector for ECG board.
Input protection and filtering
The input protection is implemented with protection diodes, which are connected to analog power supply voltage and ground. The input filtering for ECG measurement is done with discrete components. The measured signal is AC-coupled for respiration measurement. The signal from the respiration supply is AC coupled. There are also the overload protection diodes for respiration measurement supply.
ECG preamplifiers
The buffer amplifiers are used for each lead except N/RL. The leads off detection is implemented by measuring the level of the input buffer amplifiers with A/D converter of CPU. The ECG signals are measured using differential amplifiers.
S/5 Hemodynamic modules
ECG amplifiers and baseline restoration
The function of the ECG amplifiers and baseline restoration is to amplify the signal and to restore the baseline of the signal in the middle of the display after the change of the signal level e.g. after the change of the DC offset voltage.
Pacer detection
Pacer detection has been made by using two slew rate detector circuits. The pacer detection amplifiers have been realized at the front of the slew rate detectors independently from the ECG measuring channels.
Respiration impedance supply
The 62.5 kHz sine wave generator is used as the respiration measurement signal supply. Analog switches are used for connecting the sine wave to the ECG leads to be measured.
Respiration impedance amplifiers
Buffer amplifiers are used in respiration measurement. Analog switches are used for selecting the measurement leads. There are also additional amplifiers for increasing the respiration signal gain. Respiration is always measured between R and F, independently on the ECG lead selection.
ECG CPU
The CPU is a 16 bit H8/3048 single-chip microcomputer. It contains 128 kbytes of flash memory and 4 kbytes of RAM. The clock frequency is 16 MHz.
17
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
RS485 communication
The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver circuits are optically isolated from the processor of the module. PWM signal is used for direct ECG signal. Direct ECG signal is available from the X2 connector of the UPI board or from the PT module.
Power supply
The ECG board has a driver controlled half bridge switching power supply with 5 kV isolation. The supply voltages have been regulated with linear regulators.
2.2.6 ECG filtering
The S/5 monitors have three ECG filtering modes:
MONITORING 0.5...30 Hz (with 50 Hz reject filter)
DIAGNOSTIC @ 12-lead ECG 0.05...150 Hz DIAGNOSTIC @ other modules 0.05...100 Hz
0.5...40 Hz (with 60 Hz reject filter)
ST FILTER 0.05...30 Hz (with 50 Hz reject filter)
0.05...40 Hz (with 60 Hz reject filter)
The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory) movement artifacts.
Monitor filter is used in normal monitoring. Diagnostic filter is used if more accurate diagnostic information is needed. ST filter gives more accurate information of ST segment, but reduces high frequency noise.
The high-pass filters 0.5 Hz and 0.05 Hz are done with hardware. The monitor sends a command to the hemodynamic module determining which of the corner frequencies 0.5 Hz or 0.05 Hz is to be used.
The 50 Hz and 60 Hz reject filters are both low-pass filters with zero at 50 Hz or 60 Hz correspondingly and they are done with software. They are for the mains supply filtering. When these filters are used, 3 dB value for low-pass filter is 30 Hz or 40 Hz.
In diagnostic mode the upper frequency is limited by hardware and the -3 dB frequency is 100 Hz for 3 or 5 lead ECG measurement. For 12 lead ECG the upper frequency is 150 Hz and it is limited by software.
18
Document No. 800 1008-2
2.2.7 STP board
S/5 Hemodynamic modules
Patient connectors
Front panel keys
Power isolation section
Isolation transformer
TEMP measuring unit
Temp AD
AD-converter
- 8 chn
- 12 bit
INV PRESS measuring unit
Press AD
Pox AD
Serial device communication
µprocessor unit
RAM internal 2K
external 16K
EPROM 48K
RS communication
Patient isolation
POX pre­amplifier
POX
gain
control
Re d
Module bus data
Reset
Opto isolation
ver
i dr
Ds LE
ox
g
rin
su
ed
measuring
ea
IR
m
Intensities
Non volatile memory
P
Figure 8 STP board block diagram
Microprocessor unit
As processor, Intel’s 80C196KC-16 is used. There are external memories, an 8-bit data bus, a 16 MHz oscillator, an open collector reset, and a watchdog timer. Three A/D-converters within the processor are used. The processor’s internal UART communicates with the CPU board.
High speed I/O is used to obtain pulse control sequence necessary for pulse oximetry measurement. It gets its timing clock from the oscillator.
Power non­isolation section
Power for module
Power reset
Power for Communication
Module bus connector or
connector to NIBP Board (NESTPR)
RS485 Driver for module reset
Module reset
RS485 Driver for data
Module data
19
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
Temperature measurement unit
Value of NTC-resistor in the probe depends on patient’s temperature. It is measured with the following principle.
The temperature signal(s) is produced by voltage dividers, part of which is the patient probe (YSI 400-series thermistor). The output is amplified by the calibrated amplifier(s) whose offset voltage makes its output spread on both sides of zero. Wider output range (measurement range) means better resolution.
0 °C => 7K357 15 °C => 3K541 25 °C => 2K253 38 °C => 1K301 45 °C => 984R1
+5Vreference
NTC
Figure 9 Temperature measurement principle
Invasive blood pressure measurement unit
Isolated +5 V voltage is supplied to the pressure transducer. From the bridge connection a differential voltage, which depends on pressure and supplied voltage, is calculated (see the formula below).
= Uin × pressure × 5 V, where Uin is 5 V
U
out
= 25 V × pressure [mmHg]
Þ U
out
Pressure amplification is realized in the instrumentation amplifier. Gain of the amplifier is set so that the level of the signal transferred to A/D converter stays within the measurement range even when there are circumstantial offsets or offsets caused by the transducer. There is a filter before the amplifier to attenuate high frequency disturbances.
A1
to AD converter
offset
20
Document No. 800 1008-2
Vin
Pressure transducer
Vout
Input Filter
Figure 10 Pressure measurement principle
Instrumentation amplifier
G
to AD converter
Pulse oximetry measurement section
S/5 Hemodynamic modules
I=5-350mA
Probe
Preamplifier Current - to - Voltage converter
LED driving circuit
Level of LED current measurement and feedback circuit
G =1/4096-1
Digitally controlled attenuator
Amplifier
IRed LED intensity adjustment
Red LED intensity adjustment
Level of LED current indication (to CPU)
IR DC level
G = 275
Two step AC amplifier
G =16 or 63
G =16 or 63
IRed AC signal
for NESTPR G =18 or 125
Red AC signal
Red DC level
Figure 11 Pulse oximetry measurement block diagram
LED control signals
The processor sends pulse width modulated signals, IRED intensity and RED intensity, which are converted to DC voltage and filtered. By switches either RED or IRED intensity is sent forward to amplifier in LED driving circuit.
LED driving circuit
Voltage difference which corresponds to LED current, is measured by the differential amplifier circuit and its output is sent back to the processor in 0 to 5 V level. There are feedback circuits from LED current measurement and LED intensity control.
Background light is measured by picking up a sample from the signal. The sample is modified to 0 to 5 V level and sent to the processor.
Measured signal preamplification
Preamplifier is current-to-voltage converter with gain selection. The higher gain is used for measuring of thin tissue.
Digitally controlled amplifier
D/A converter is a digitally controlled amplifier after which there is another constant amplifier.
Document No. 800 1008-2
21
Loading...
+ 61 hidden pages