Datex-Ohmeda Inc.
3030 Ohmeda Drive
53707-7550 MADISON, WIS
USA
Tel. +1-608-221 1551,Fax +1-608-222 9147
www.us.datex-ohmeda.com
All specifications are subject to change without notice.
Document No. 800 1008-2
June 2001
Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.
Table of contents
TABLE OF CONTENTS
HEMODYNAMIC MODULES
TABLE OF CONTENTSi
Table of figuresiii
Introduction1
1Specifications2
1.1 General specifications ..............................................................................................................................2
2.2 Main components...................................................................................................................................10
2.3 Connectors and signals...........................................................................................................................24
2.3.1 Module bus connector....................................................................................................................24
2.3.2 Front panel connectors...................................................................................................................25
2.3.3 Test points on boards .....................................................................................................................26
3Service Procedures28
3.1 General service information.....................................................................................................................28
3.2 Service check .........................................................................................................................................28
5.1 NIBP service menu .................................................................................................................................54
5.2 ECG service menu ..................................................................................................................................62
5.2.1 ECG setup menu ............................................................................................................................64
5.3 STP service menu ...................................................................................................................................65
5.3.1 STP calibration menu .....................................................................................................................67
6Spare Parts68
6.1 Spare parts list.......................................................................................................................................68
This Technical Reference Manual Slot provides information for the maintenance and service of the
hemodynamic modules. Please see also related Technical Reference Manual for information
related to system e.g. related documentation, conventions used, symbols on equipment, safety
precautions, system description, system installation, interfacing, functional check and planned
maintenance.
The S/5 M-ESTPR/-ESTR/-ETPR and S/5 M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/NESTR/-NETPR are double width modules designed for use with S/5 monitors. The modules
provide general hemodynamic parameters. Later in this manual modules can be called w/o
system name S/5.
S/5 Hemodynamic modules
NOTE: Do not use identical modules in
the same monitor simultaneously.
The following modules are considered
identical:
M-ESTP/-EST/-ETP
M-ESTPR/-ESTR/-ETPR
M-NESTPR/-NESTR/-NETPR
M-NE12STPR/-NE12STR/-NE12TPR
M-MRI/-MRIP
child60 ±10 mmHg / 2 min.
infant40 ±10 mmHg / 1 min.
2
Document No. 800 1008-2
1.2.2 ECG
S/5 Hemodynamic modules
Lead selection @ 12-lead ECGI, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Lead selection @ other modulesI, II, III, aVR, aVL, aVF, V
Sweep speeds 12.5, 25, 50 mm/sec
DISPLAY FILTER
Diagnostic @ 12-lead ECG0.05...150 Hz
Diagnostic @ other modules0.05...100 Hz
Monitoring0.5...30 Hz (-3 dB, with 50 Hz reject filter)
0.5...40 Hz (-3 dB, with 60 Hz reject filter)
ST filter0.05...30 Hz (-3 dB, with 50 Hz reject filter)
0.05...40 Hz (-3 dB, with 60 Hz reject filter)
HEART RATE FROM ECG
Range30...250 bpm
Accuracy±5 bpm or ±5 %, whichever is greater
Resolution1 bpm
Update interval5 s
Averaging time10 s
ST LEVELS (in main software)
ST level range-9...+9 mm (-0.9...+0.9 mV)
Resolution0.1 mm (0.01 mV)
Averagingcalculated from 8 QRS
SYNCHRONIZATION
Direct ECGanalog output of ECG, 1 V/1 mV
Pacer5 V and 0.5...2.5 ms pulse, < 30 ms after pacer peak
Defibrillator5 V and 10 ms pulse, < 35 ms after R-point synchronization
Display resolution1 digit = 1 % of SpO
Display averaging time20, 10 sec, beat-to-beat
Pulse beep pitchvaries with SpO
The monitor is calibrated over the measurement range against functional saturation SpO
HEART RATE FROM PLETH
Measurement range 30...250 bpm
Accuracy30...100, ±5 bpm,
Resolution1 bpm
Display averaging10 s
2±1 SD)
1
80...50 %, ±3 digits
50...40 %, unspecified
2
2 level
func.
2
100...250, ±5 %
1
1 SD (standard deviation) = 68 % of all readings in the specified range in stable conditions.
3
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
Adjustable pulse beep volume.
PLETH WAVEFORM
Scales2, 5, 10, 20, 50 mod%, Auto
Start up scale is 20 mod% if AUTO is not selected to be the default setting.
1.2.4 Temperature
Measurement range10...45 °C (50...113 °F)
(In rev. ESTP 03/ EST 02/ETP 02 or earlier: 15...45 °C (59...113 °F))
Measurement accuracy±0.1 °C (25...45.0 °C)
Display resolution0.1 °C (0.1 °F)
Temperature testautomatic (every 10 min)
Probe typecompatible with YSI 400 series
1.2.5 Invasive blood pressure
Measurement range-40...320 mmHg
Measurement accuracy±2 mmHg or ±5 %
Zero adjustment range±150 mmHg
Calibration range±20 %
Scalesupper limit is adjustable between 10 and 300 mmHg in steps of
±0.2 °C (10...24.9 °C)
10. Lower limit is 10 % of selected upper limit below zero.
Sweep speed12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range-40...320 mmHg
Resolution±1 mmHg
WAVEFORM DISPLAY
Range-30...300 mmHg
HEART RATE FROM ARTERIAL PRESSURE
Measurement range30...250 bpm
Resolution1 bpm
Accuracy±5 bpm or ±5 % whichever is greater
1.2.6 Respiration
Measurement range4...120 bpm
Accuracy±5 bpm or ±5 %
Resolution1 bpm
Averaging time30 s
Update interval10 s
RESPIRATION WAVEFORM
Sweep Speeds6.25 mm/s and 0.625 mm/s
4
Document No. 800 1008-2
1.3 Technical specifications
1.3.1 NIBP
Deflation rate, PR dep.5...13 mmHg/s
Inflation time20...185 mmHg, 1...5 s
Over pressure limit, stops measurement after 2 seconds
Safety valve limits the maximum cuff pressure to 320 mmHg in adult/child mode or 165 mmHg in
infant mode. Independent timing circuit limits pressurizing (>5 mmHg) time to 2 minutes 10
seconds maximum in adult/child mode, and 1 minute 5 seconds in infant mode.
Zeroing to ambient pressure is done automatically.
Inflation pressure is adjusted according to the previous systolic pressure, typically 40 mmHg
above. If the systolic pressure is not found, inflation pressure is increased typically 50 mmHg.
Max. measurement timeadult2 min
child2 min
infant1 min
Pressure transducer accuracy is better than ±3 mmHg or ±2 % whichever is greater.
Max. error ±4 mmHg.
Protection against electrical
shockType BF defibrillation proof
Defibrillation protection5000 V, 360 J
Recovery time2 s
Input impedance>2.5 MΩ (10 Hz)
CMRR>100 dB (ST)
System noise<40 µV (p-p, RTI)
Allowable offset±300 mVDC
Gain range0.2...5.0 cm/mV
Pacemaker pulse detection2...500 mV, 0.5...2 ms pulses
Protection against electrical
shockType CF defibrillator proof
5
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
1.3.3 Pulse oximetry
Protection against electrical
shockType BF defibrillation proof
1.3.4 Temperature
Measurement accuracy±0.1 °C (25.0...45.0 °C)
Protection against electrical
shockType CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on
transducer/probe used. Please check the transducer/probe specification.
1.3.5 Invasive blood pressure
DIGITAL DISPLAY AVERAGING
Digital displays Art and P1 are averaged over 5 seconds and updated at 5 seconds intervals. All
other pressures have respiration artifact rejection.
±0.2 °C (10.0...24.9 °C)
Accuracy±5 % or ±2 mmHg, whichever is greater
Transducer and input sensitivity
Filter0...4 - 22 Hz adjustable
Zero set accuracy±1 mmHg
Calibration resolution±1 mmHg
Zero timeless than 15 s
Protection against electrical
shockType CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on
transducer/probe used. Please check the transducer/probe specification.
1.3.6 Respiration
Excitation frequency @ 12-lead ECG62.5 kHz
Excitation frequency @ other modules31.25 kHz
Breath detectionautomatic, range 0.3...6 Ω manually adjustable minimum
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is
inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a
speed based on the patient’s pulse, collecting data from the oscillations caused by the pulsating
artery. Based on these oscillations, values for systolic, mean, and diastolic pressures are
calculated.
The following parts are necessary for the NIBP measurement:
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical
potential produced with electrodes placed on the surface of the body.
ECG reflects:
• electrical activity of the heart
• normal/abnormal function of the heart
• effects of anesthesia on heart function
• effects of surgery on heart function
See the User's Reference Manual for electrodes positions and other information.
2.1.3 Pulse oximetry
A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near
infrared (about 900 nm) and the other in the red region (about 660 nm) of light spectrum. These
wavelengths are emitted by LEDs in the SpO
tissue and is finally detected by a PIN-diode opposite to LEDs in the probe. Pulse oximeter derives
the oxygen saturation (SpO
absorption at the two wavelengths and the arterial oxygen saturation SaO
probe, the light is transmitted through peripheral
2
) using empirically determined relationship between the relative
2
.
2
In order to measure the arterial saturation accurately, pulse oximeters use the component of light
absorption giving variations synchronous with heart beat as primary information on the arterial
saturation.
Document No. 800 1008-2
7
Datex-Ohmeda S/5 monitors
A general limitation of the above pulse oximetry principle is that due to only two wavelengths used
only two hemoglobin species can be discriminated by the measurement.
The modern pulse oximeters are empirically calibrated either against fractional saturation
frac;
SaO
2
2
fracSaO
2
=
HbO
2
++
binDyshemogloHbHbO
Formula 1
or against functional saturation SaO
HbO
=
2
funcSaO
which is more insensitive to changes of carboxyhemoglobin and methemoglobin concentrations in
blood.
The oxygen saturation percentage SpO
the functional saturation SaO
measurement relative to SaO2func can be maintained even at rather high concentrations of
carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeter is not able to
correctly measure oxygen content of the arterial blood at elevated carboxyhemoglobin or
methemoglobin levels.
Plethysmographic pulse wave
The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation at
the measuring site. Thus the amplitude of the waveform represents the perfusion.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The
signals are filtered to reduce noise and checked to separate artifacts.
func;
2
2
HbHbO
2
+
measured by Datex-Ohmeda module is calibrated against
2
func. The advantage of this method is that the accuracy of SpO
2
Formula 2
2
Probe
8
Document No. 800 1008-2
The standard probe is a finger clamp probe which contains the light source LEDs in one half and
the photodiode detector in the other half. Different kinds of probes are available from DatexOhmeda.
Intensity of
transmitted
light
I
max (DC-component)
S/5 Hemodynamic modules
I
max
AC-component
I
min
Transmitted
light
Incident light
Emitter
RED
Detector
No pulsation
IRED
Pulsatile blood
SpO sensor cable
2
Variable absorption
due to pulse added
volume of arterial
blood
Arterial blood
Venous blood
Tissue
Time
SpO sensor connector
2
6
GND
7
I
LED
4
5
GND
8
V
B
R
C
1
I
S
9
GND
Figure 2Absorption of infrared light in the finger probe parts layout and schematic
2.1.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature changes,
called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
• Applying a constant voltage across the resistor and measuring the current that flows through
diagram
it
9
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
• Applying a constant current through the resistor and measuring the voltage that is generated
across it.
In Datex-Ohmeda modules the two methods are combined in a form of a voltage divider. The NTCresistor is connected in series with a normal resistor and a constant voltage is applied across them.
The temperature dependent voltage can be detected at the junction of the resistors, thus producing
the temperature signal from the patient. The signal is amplified by analog amplifiers and further
processed by digital electronics.
2.1.5 Invasive blood pressure
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive
pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal
saline all connected together by stopcocks, is attached to the catheter. The transducer is placed at
the same level with the heart, and is electrically zeroed.
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor
interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
2.1.6 Respiration
Impedance respiration is measured across the thorax between ECG electrodes. The signal of the
respiration is made by supplying current between electrodes and by measuring the differential
current from the electrodes. The signal measured is the impedance change caused by breathing.
From these impedance changes, respiration rate is calculated, and respiration waveform is
displayed on the screen.
2.2 Main components
2.2.1 M-ESTPR/-ETPR/-ESTR modules
T1
T2
ECG+
Resp
SpO
P1
P2
2
ECG
Lead
Start
Wedge
Zero
P1
Zero
P2
10
Document No. 800 1008-2
Figure 3Front panel of M-ESTPR
The M-ESTPR, M-ETPR, and M-ESTRmodules contain two main PC boards, the STP board and the
ECG board. They work independently. Both of them have their own processor and software EPROM.
Some components on the boards are not used in ETPR and ESTR modules.
In M-ESTPR module, additionally, there are two small boards, the SP input and the ECG input
S/5 Hemodynamic modules
boards, attached to the front panel of the module. The front panel has six connectors and four keys.
The connectors are two for temperature measurement, two for invasive blood pressure
measurement, one for ECG, and one for SpO
measurement. The keys are for ECG lead, Start
2
Wedge, P1 zero, and P2 zero.
In M-ETPR module, there are two small boards, the ECG input board and the 2P input board
attached to the front panel of the module. The front panel has five connectors and four keys. The
connectors are two for temperature measurement, two for invasive blood pressure measurement,
and one for ECG
measurement. The keys are for ECG lead, Start Wedge, P1 zero, and P2 zero.
In M-ESTR module, there are two small boards: the S input board and the ECG input board,
attached to the front panel of the module. The front panel has four connectors and one key. The
connectors are two for temperature measurement, one for ECG, and one for SpO
The key is for ECG lead select.
NOTE: M-ESTP rev. 03, M-ETP rev. 02 and M-EST rev. 02 and all earlier revisions have separate T
and SP input boards.
The M-NESTPR, M-NETPR, and M-NESTRmodules contain three main PC boards, the STP board,
the ECG board, and the NIBP board. They work independently. Each of these has their own
processor and software EPROM.
The M-NE12STPR, M-NE12TPR, and M-NE12STR contain three main PC boards, The STP board,
the ECG board and the NIBP board. They work independently. Each of them has their own
processor. The STP board and NIBP board have software EPROM. In the ECG board the software is
in flash memory. The STP and NIBP boards are the same as in M-NESTPR module but the ECG
board and ECG input board are different.
In M-NESTPR module, there are two small boards, the SP input and the ECG input board attached
to the front panel of the module. The front panel has seven connectors and four keys. The
connectors are two for temperature measurement, two for invasive blood pressure measurement,
one for ECG, one for NIBP, and one for SpO
measurement. The keys are for NIBP Auto On/Off,
2
NIBP Start/Cancel, P1 zero, and P2 zero. The structure of M-NE12STPR is similar except the ECG
board and ECG input board are different.
Document No. 800 1008-2
11
Datex-Ohmeda S/5 monitors
In M-NETPR module, there are two small boards, the 2P input board and the ECG input board,
attached to the front panel of the module. The front panel has six connectors and four keys. The
connectors are two for temperature measurement, two for invasive blood pressure measurement,
one for ECG, and one for NIBP. The keys are for Auto On/Off, Start/Cancel, P1 zero, and P2 zero.
The structure of M-NE12TPR is similar except the ECG board and ECG input board are different.
In M-NESTR module, there are two small boards, the ECG input board and the S input board,
attached to the front panel of the module. The front panel has five connectors and two keys. The
connectors are two for temperature measurement, and one for SpO
and one for NIBP. The keys are for Auto On/Off, Start/Cancel. The structure of M-NE12STR is
similar except the ECG board and ECG input board are different.
2.2.3 NIBP board
CUFF
ZERO
VALVE
EXHAUST
VALVE 2
JOIN ING
CHAMBER
EXHAUST VALVE 1
BLEED VALVE
CHECK VALVE
measurement, one for ECG,
2
PUMP and
PUMP
M
VALVE
DRIVER
B1B2
t
o
/
f
r
o
m
m
o
d
u
l
e
b
u
s
X1
PRESSURE
S
TRAN
-
D
UCERS
AD-CONVERTER
POWER-UP
RESET
RS485
INTER F.
EEPRO
(C alibration
Data)
Figure 5NIBP board functional block diagram
Pressure transducers
OVERPRESSURE CONTROL
M
Write protect switch
CPU
80C51FA
NTERNAL
I
WATCHDOG
ADDRESS BUS
ADDRESS
DECODER
ADDRESS
LATCH
WATCHDO G
TIMER
RAM
DATABUS
+
1
5
EPROM
D
V
FRONT
PANEL
KEYS
SOFT WAR E
CON TRO L
12
Document No. 800 1008-2
The NIBP board contains two pressure transducers. They are of piezoresistive type. One is used for
measuring the pressure of the blood pressure cuff and the pressure fluctuations caused by arterial
wall movement (B1). The other is used for detection of cuff hose type, cuff loose and cuff occlusion
situations etc. (B2). The transducers are internally temperature compensated. They are supplied by
a constant voltage and their output voltage changes up to 40 mV max. (50 kPa, 375 mmHg).
Signal processing
Two signals from the pressure transducers are amplified and sent to A/D converter. After the
converter, digitized signals are sent to microprocessor for data processing. Before the converter,
one of the signals is used to adjust the offset to the pressure safety level.
The NIBP board is controlled with 80C51FA microprocessor at 16 MHz oscillator frequency.
Memory
NIBP program memory (EPROM) size is 128k × 8. RAM size is 32k × 8 bit and it stores variable
values in NIBP measurement. EEPROM is size 64 × 16 bit and is used to store the calibration
values for the pressure transducers, the pulse valve constants gained during measurements, the
PC board identification, and module serial number.
Software control
Software controls valves and pump. In addition to the individual on/off signals for each component
there is a common power switch for the valves and the pump that can be used at pump/valve
failures.
S/5 Hemodynamic modules
In addition to external RS485 reset line the microprocessor system is equipped with its own powerup reset. See the section in ECG board’s description: “RS485 communication”
Watchdog timer
The NIBP board is equipped with software independent safety circuit to disconnect supply voltages
from the pump and the valves if the cuff has been pressurized longer than preset time. As soon as
the cuff pressure rises over a specifiedpressure limit, timer starts counting. The timer is adjusted to
stop the pump and open the valves in 2 minutes 10 seconds in adult/child mode and in 1 minute
5 seconds in infant mode.
Valves
Exhaust valves are used for emptying the cuff and the joining chamber after the measurement.
Exhaust valve 1 is also used as safety valve in infant mode. Valve opens at 165 mmHg. Exhaust
valve 2 is also used as safety valve in adult mode and opens at 320 mmHg.
Bleed valve is used for emptying the cuff during measurement. Zero valve is used for connecting
the pressure transducer B1 to open air.
Power supply section
All connections are established via 25-pin connector (D-type, female). The module needs +5 V,
±15 V, and +15 VD (dirty) power supply to operate. The pump and the valves use separate +15 VD
power line. The supply voltages are generated in the power supply section of the S/5 monitor. The
reference voltages ±5 V
and +10 V
ref
are generated on the NIBP board.
ref
13
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
2.2.4 ECG board in 3-and 5-lead measurement
Patient signals are connected to overload protection circuits (resistors and gas-filled surge
arresters) and analog switches to instrumentation amplifiers. Then the signals are amplified by 480
and limited by slew rate. Then they are A/D-converted, analyzed and transferred to module bus in
digital form.
PATIENT
OVERLOAD PROTECTION
DEFIBRILLATION
DETECTOR
uP
SLEWRATE
DETECTOR
HP
Filter
uP
ANTIALIAS
RL
ANALOG SWITCHES
SLEWRATE
DETECTOR
uP
ANTIALIAS
SUPPLY
VOLTAGES
FCN
SLEWRATE
DETECTOR
uPuPuP
uP
HP
Filter
ANTIALIAS
A/D
MICROPROCESSOR
(uP)
HP
Filter
SLEWRATE
DETECTOR
uP
uP
HP
Filter
ANTIALIAS
LEAD SELECTION LOGIC
Controlled by ECG lead
selection signals from uP
Resp
amplifier
31 kHz
from uP
Sync. Rectifier
Amplifier
Figure 63- and 5- lead ECG board block diagram
Analog ECG section
ECG cable is connected to connector pins E1 to E6 on the input board which contains an overload
protection circuit. Leads are connected to amplifiers via analog switches. States of the switches
depend on the cable type. Lead-off, noise and pacemaker are detected by a slew rate detector.
14
Document No. 800 1008-2
ISOLATION
POWER
SOURCE
POWER
SOURCE
OPTOCOUPLER
OPTOCOUPLER
to STP Board
DIRECT
ECG
DIRECT
ECG
S/5 Hemodynamic modules
Lower frequency is determined by high pass (HP) filter 0.5 Hz (monitor bandwidth) or 0.05 Hz
(diagnostic or ST- bandwidth).
Respiration section
3-lead cableThe analog switches control the current supply source of the impedance respiration measurement,
and the lead selection for the 3-lead cable can be seen from the following table:
Table 2Lead selection and coding for the 3-lead cable
Selected leadCurrent source betweenSignal measured from
IR - LN
IIR - NL
IIIL - NR
Position on
IEC standard codingAAMI standard coding
body surface
right armR = redRA = white
left armL = yellowLA = black
left legF = greenLL = red
5-lead cableWhen the 5-lead cable is used, the current source is between L-F and the signal is measured from
the N, independently on the lead selection.
The respiration amplifier consist of the operational amplifiers, and the components around them.
There is an analog switch for controlling the gain of the first stage of the preamplifier. Synchronous
rectifier consists of the analog switches, which are used for detecting the respiration signal from 31
kHz amplitude modulated raw signal. The amplifier stage consists of the differential amplifier and
the last amplifier. The differential amplifier consists of the operational amplifiers and the
components around them. This stage is AC-coupled on both sides for minimising the offset
voltages. The last amplifier is used for amplifying the signal derived from differential amplifier
stage. The respiration signal is zeroed at the beginning of the measurement. Zeroing is also used
for fast recovering the measurement after the motion artefact. This is done in amplifier section.
NOTE: The respiration measurement is switched OFF for 20 seconds when defibrillation is detected
at the defibrillation detector.
Microprocessor section
Microprocessor contains RAM and EPROM memories. The processor uses external EEPROM
memory. The microprocessor’s internal 8-channel A/D-converter converts the ECG-signals to
digital form. See the section in ECG board’s description: “RS485 communication
Serial communication
Communication with the module bus is made through RXD and TXD pins. See the section in STP
board’s description: “Serial communication”.
15
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
Isolated section
The patient isolation of ECG is 5 kV.
NOTE: The isolation has been changed from the earlier revisions.
WARNING Do not touch battery operated monitor during defibrillation procedure.
See the “Isolated section” in STP board description.
Power supply section
See the “Power supply section” in STP board description.
There is a test connector (X20) on the board for voltages +5 VREF, +5 V, +12 V, GND and -12 V.
2.2.5 ECG board in 12-lead measurement
The 12-lead ECG measurement consists of the functions, which are shown in the figure 7. All
functions are located in the ECG board except the front panel connector and the ECG input board.
Front panel connector and
ECG input board
Input protection and filtering
Respiration
impedance
amplifiers
RS 485
communication
Respiration
impedance
supply
ECG CPU
Power
supply
ECG preamplifiers
ECG amplifiers
and
Baseline
restoration
NV memory
Pacer
detection
16
Document No. 800 1008-2
Isolation
to STP board
Isolation
Figure 712-lead ECG measurement block diagram
Front panel connector and ECG input board
The connector for the 12-lead ECG cable is a green 12 pin Nicolay type connector. 3- or 5-lead
cables with blue connector cannot be connected to this connector. The ECG input board contains
high voltage resistors and a connector for ECG board.
Input protection and filtering
The input protection is implemented with protection diodes, which are connected to analog power
supply voltage and ground. The input filtering for ECG measurement is done with discrete
components. The measured signal is AC-coupled for respiration measurement. The signal from the
respiration supply is AC coupled. There are also the overload protection diodes for respiration
measurement supply.
ECG preamplifiers
The buffer amplifiers are used for each lead except N/RL. The leads off detection is implemented
by measuring the level of the input buffer amplifiers with A/D converter of CPU. The ECG signals are
measured using differential amplifiers.
S/5 Hemodynamic modules
ECG amplifiers and baseline restoration
The function of the ECG amplifiers and baseline restoration is to amplify the signal and to restore
the baseline of the signal in the middle of the display after the change of the signal level e.g. after
the change of the DC offset voltage.
Pacer detection
Pacer detection has been made by using two slew rate detector circuits. The pacer detection
amplifiers have been realized at the front of the slew rate detectors independently from the ECG
measuring channels.
Respiration impedance supply
The 62.5 kHz sine wave generator is used as the respiration measurement signal supply. Analog
switches are used for connecting the sine wave to the ECG leads to be measured.
Respiration impedance amplifiers
Buffer amplifiers are used in respiration measurement. Analog switches are used for selecting the
measurement leads. There are also additional amplifiers for increasing the respiration signal gain.
Respiration is always measured between R and F, independently on the ECG lead selection.
ECG CPU
The CPU is a 16 bit H8/3048 single-chip microcomputer. It contains 128 kbytes of flash memory
and 4 kbytes of RAM. The clock frequency is 16 MHz.
17
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
RS485 communication
The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver
circuits are optically isolated from the processor of the module. PWM signal is used for direct ECG
signal. Direct ECG signal is available from the X2 connector of the UPI board or from the PT module.
Power supply
The ECG board has a driver controlled half bridge switching power supply with 5 kV isolation. The
supply voltages have been regulated with linear regulators.
The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory)
movement artifacts.
Monitor filter is used in normal monitoring. Diagnostic filter is used if more accurate diagnostic
information is needed. ST filter gives more accurate information of ST segment, but reduces high
frequency noise.
The high-pass filters 0.5 Hz and 0.05 Hz are done with hardware. The monitor sends a command to
the hemodynamic module determining which of the corner frequencies 0.5 Hz or 0.05 Hz is to be
used.
The 50 Hz and 60 Hz reject filters are both low-pass filters with zero at 50 Hz or 60 Hz
correspondingly and they are done with software. They are for the mains supply filtering. When
these filters are used, 3 dB value for low-pass filter is 30 Hz or 40 Hz.
In diagnostic mode the upper frequency is limited by hardware and the -3 dB frequency is 100 Hz
for 3 or 5 lead ECG measurement. For 12 lead ECG the upper frequency is 150 Hz and it is limited
by software.
18
Document No. 800 1008-2
2.2.7 STP board
S/5 Hemodynamic modules
Patient connectors
Front panel
keys
Power
isolation
section
Isolation
transformer
TEMP
measuring
unit
Temp AD
AD-converter
- 8 chn
- 12 bit
INV
PRESS
measuring
unit
Press AD
Pox AD
Serial device communication
µprocessor unit
RAM internal 2K
external 16K
EPROM 48K
RS communication
Patient isolation
POX preamplifier
POX
gain
control
Re d
Module
bus
data
Reset
Opto
isolation
ver
i
dr
Ds
LE
ox
g
rin
su
ed
measuring
ea
IR
m
Intensities
Non volatile
memory
P
Figure 8STP board block diagram
Microprocessor unit
As processor, Intel’s 80C196KC-16 is used. There are external memories, an 8-bit data bus, a 16
MHz oscillator, an open collector reset, and a watchdog timer. Three A/D-converters within the
processor are used. The processor’s internal UART communicates with the CPU board.
High speed I/O is used to obtain pulse control sequence necessary for pulse oximetry
measurement. It gets its timing clock from the oscillator.
Power
nonisolation
section
Power for
module
Power
reset
Power for
Communication
Module bus connector or
connector to NIBP Board (NESTPR)
RS485 Driver
for module
reset
Module reset
RS485 Driver
for data
Module data
19
Document No. 800 1008-2
Datex-Ohmeda S/5 monitors
Temperature measurement unit
Value of NTC-resistor in the probe depends on patient’s temperature. It is measured with the
following principle.
The temperature signal(s) is produced by voltage dividers, part of which is the patient probe (YSI
400-series thermistor). The output is amplified by the calibrated amplifier(s) whose offset voltage
makes its output spread on both sides of zero. Wider output range (measurement range) means
better resolution.
0 °C => 7K357
15 °C => 3K541
25 °C => 2K253
38 °C => 1K301
45 °C => 984R1
+5Vreference
NTC
Figure 9Temperature measurement principle
Invasive blood pressure measurement unit
Isolated +5 V voltage is supplied to the pressure transducer. From the bridge connection a
differential voltage, which depends on pressure and supplied voltage, is calculated (see the
formula below).
= Uin × pressure × 5 V, where Uin is 5 V
U
out
= 25 V × pressure [mmHg]
Þ U
out
Pressure amplification is realized in the instrumentation amplifier. Gain of the amplifier is set so
that the level of the signal transferred to A/D converter stays within the measurement range even
when there are circumstantial offsets or offsets caused by the transducer. There is a filter before the
amplifier to attenuate high frequency disturbances.
A1
to AD
converter
offset
20
Document No. 800 1008-2
Vin
Pressure
transducer
Vout
Input
Filter
Figure 10Pressure measurement principle
Instrumentation
amplifier
G
to AD converter
Pulse oximetry measurement section
S/5 Hemodynamic modules
I=5-350mA
Probe
Preamplifier
Current - to - Voltage
converter
LED driving circuit
Level of LED current
measurement and
feedback circuit
G =1/4096-1
Digitally controlled
attenuator
Amplifier
IRed LED intensity
adjustment
Red LED intensity
adjustment
Level of LED current indication (to CPU)
IR DC level
G = 275
Two step
AC amplifier
G =16
or 63
G =16
or 63
IRed AC signal
for NESTPR
G =18
or 125
Red AC signal
Red DC level
Figure 11Pulse oximetry measurement block diagram
LED control signals
The processor sends pulse width modulated signals, IRED intensity and RED intensity, which are
converted to DC voltage and filtered. By switches either RED or IRED intensity is sent forward to
amplifier in LED driving circuit.
LED driving circuit
Voltage difference which corresponds to LED current, is measured by the differential amplifier
circuit and its output is sent back to the processor in 0 to 5 V level. There are feedback circuits from
LED current measurement and LED intensity control.
Background light is measured by picking up a sample from the signal. The sample is modified to 0
to 5 V level and sent to the processor.
Measured signal preamplification
Preamplifier is current-to-voltage converter with gain selection. The higher gain is used for
measuring of thin tissue.
Digitally controlled amplifier
D/A converter is a digitally controlled amplifier after which there is another constant amplifier.
Document No. 800 1008-2
21
Loading...
+ 61 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.