Datex-Ohmeda S-5 Service Manual M-PP

S/5

Datex-Ohmeda

TM
Dual Pressure Module, M-PP (Rev. 01)

Technical Reference Manual

www.us.datex-ohmeda.com
All specifications are subject to change without notice.

Document No. 8001014-1

June 2001

Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.

Table of contents

TABLE OF CONTENTS
Dual Pressure Module, M-PP
TABLE OF CONTENTS i
TABLE OF FIGURES ii
Introduction 1
1 Specifications 2
1.1 General specifications ..............................................................................................................................2
1.2 Typical performance .................................................................................................................................2
1.3 Technical specifications............................................................................................................................2
2 Functional Description 3
2.1 Measurement principle .............................................................................................................................3
2.2 Main components.....................................................................................................................................3
2.2.1 PP board..........................................................................................................................................4
2.3 Connectors and signals.............................................................................................................................7
2.3.1 Module bus connector......................................................................................................................7
2.3.2 Front panel connectors.....................................................................................................................8
2.3.3 Other connectors..............................................................................................................................8
3 Service Procedures 9
3.1 General service information.......................................................................................................................9
3.2 Service check .........................................................................................................................................10
3.2.1 Recommended tools ......................................................................................................................10
3.3 Disassembly and reassembly..................................................................................................................13
3.4 Adjustments and calibrations..................................................................................................................14
3.4.1 Invasive pressure calibration...........................................................................................................14
4 Troubleshooting 15
4.1 Troubleshooting chart .............................................................................................................................15
4.2 Troubleshooting flowchart .......................................................................................................................16
5Service Menu 17
5.1 PP menu................................................................................................................................................18
5.2 PP calibrations menu..............................................................................................................................19
6 Spare Parts 20
6.1 Spare parts list .......................................................................................................................................20
6.1.1 Dual Pressure Module, M-PP, Rev. 00 .............................................................................................20
6.1.2 Dual Pressure Module, M-PP, Rev. 01 .............................................................................................21
6.1.3 Front panel stickers for AS/3 modules (square buttons) ...................................................................21
6.1.4 Front panel stickers for S/5 modules (round buttons).......................................................................21
7 Earlier Revisions 23
APPENDIX A 25
Service check form A-1
Document No. 8001014-1
i
Datex-Ohmeda S/5 monitors

TABLE OF FIGURES

Figure 1 Dual Pressure Module, M-PP.................................................................................................................1
Figure 2 Dual Pressure Module, M-PP, front panel...............................................................................................3
Figure 3 PP board block diagram........................................................................................................................4
Figure 4 Pressure transducer principle of operation.............................................................................................5
Figure 5 Serial communication and opto isolation sections .................................................................................6
Figure 6 Module bus connector (X1) ...................................................................................................................7
Figure 7 M-PP troubleshooting flowchart ..........................................................................................................16
ii
Document No 8001014-1

INTRODUCTION

This section provides information for the maintenance and service of the Dual Pressure Module, M­PP. The M-PP module is a single width plug-in module designed for use with the S/5 monitors. The Dual Pressure module provides invasive blood pressure (InvBP) measurement.
Dual Pressure Module, M-PP
Figure 1 Dual Pressure Module, M-PP
1
Document No 8001014-1
Datex-Ohmeda S/5 monitors

1 SPECIFICATIONS

1.1 General specifications
Module size, W × D × H 37 × 180 × 112 mm / 1.5 × 7.1 × 4.4 in Module weight 0.35 kg / 0.8 lbs Power consumption approximately 3.5 W
1.2 Typical performance
Measurement range -40...+320 mmHg Accuracy ±5 % or ±2 mmHg, whichever is greater Zero adjustment range ±150 mmHg Calibration range ±20 % Scales Upper limit is adjustable between 10 and 300 mmHg in steps of
Sweep speed 12.5, 25, 50 mm/s
10. Lower limit is 10 % of selected upper limit below zero.
DIGITAL DISPLAY Range -40...+320 mmHg Resolution ±1 mmHg
WAVEFORM DISPLAY Range -30...+300 mmHg
HEART RATE FROM ARTERIAL PRESSURE Measurement range 30...250 bpm Resolution 1 bpm Accuracy ±5 bpm or ±5 %, whichever is greater
1.3 Technical specifications
Accuracy ±5 % or ±2 mmHg, whichever is greater Transducer and input sensitivity 5 µV/V/mmHg, 5 VDC, 20 mA max current Nonlinearity < 1 %, 0...200 mmHg
Filter 0...4 - 22 Hz adjustable Zero set accuracy ±1 mmHg Calibration resolution ±1 mmHg Zero time < 15 sec
Protection against electrical shock Type CF defibrillation proof
< 2 %, -40...0 and 200...320 mmHg
2
Document No 8001014-1
DIGITAL DISPLAY AVERAGING Art and P1 digital displays are averaged over 5 seconds and updated at 5 second intervals. All other pressures have respiration artifact rejection.
NOTE: The accuracy of the measurement may be different from that specified, depending on the transducer/probe being used. Please check the transducer/probe specification.

2 FUNCTIONAL DESCRIPTION

2.1 Measurement principle
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal saline all connected together by stopcocks, is attached to the catheter. The transducer is placed at the same level with the heart, and is electrically zeroed.
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
2.2 Main components
The main components of the Dual Pressure Module, M-PP, are a PP circuit board, a front panel and a box. The front panel includes two Nicolay-type connectors, P5 and P6, and two direct function keys, Zero P5 and Zero P6 for pressure zeroing.
Dual Pressure Module, M-PP
Communication between the Dual Pressure Module and the Central Unit is an RS485 serial interface.
The power supply voltages to the Dual Pressure Module are generated in the power supply section of the Central Unit. All electrical connections between the Dual Pressure Module and the Central Unit are via a 25-pin D-connector at the back of the module.
Zero
P5
P5
Zero
P6
P6
Figure 2 Dual Pressure Module, M-PP, front panel
Document No 8001014-1
3
Datex-Ohmeda S/5 monitors
2.2.1 PP board
RS485 Driver for data
Module bus connector
Module data
Module reset
RS485 Driver for module reset
Power for Communication
Power reset
Power for module
Power non­isolation section
Opto isolation
Non volatile memory
Reset
Module bus data
Patient isolation
µprocessor
RAM internal 2K
EPROM 48K
RS communication
AD-converter
-8 chn
-12bit
Press AD
INV PRESS measuring unit
external 16K
Isolation transformer
Power isolation section
Front panel keys
Figure 3 PP board block diagram
Microprocessor unit
The microprocessor uses the Intel 80C196KC-16 CPU which includes three A/D converters and a UART. The microprocessor uses external memories, an 8-bit data bus, a 16 MHz oscillator, an open collector reset, and a watchdog timer. The three A/D-converters within the CPU convert the analog input signals to digital. The internal UART communicates and transfers data between the module and the CPU board in the monitor.High speed I/O is used to obtain the pulse control sequence necessary for pulse oximetry measurement. It receives its timing clock signal from the 16 MHz oscillator.
4
Document No 8001014-1
Patient connectors
Invasive blood pressure measurement unit
An isolated +5 V supply is connected to the input of the pressure transducer bridge circuit. From the bridge circuit output a differential voltage, which depends on blood pressure and input supply voltage, is calculated using the following formula:
Uout = Uin × Pressure × 5 V
where Uin = 5 V
Uout = 25 V × Pressure [mmHg]
Pressure amplification is performed by the instrumentation amplifier. The gain of the amplifier is set so that the level of the signal transferred to the A/D converter stays within the measurement range even when there are circumstantial offsets or offsets caused by the pressure transducer. The input filter before the amplifier attenuates high frequency disturbances.
Dual Pressure Module, M-PP
Figure 4 Pressure transducer principle of operation
Serial communication
Serial communication between the Dual Pressure Module and the Central Unit is via an RS485 type bus. The communication bus drivers are powered from the Central Unit Module Bus. The module isolation section (+5 V) is powered from the isolated power supply.
The buffers of the serial communication drivers are controlled by a reset signal such that when the reset is active the drivers do not transfer data.
Vin
Pressure transducer
Vout
Input Filter
Instrumentation amplifier
G
to AD converter
In addition to the RS485 reset there is a logic power-up reset, which holds for approximately 500 ms regardless of the state of the RS485 reset. A time constant determines the power-up reset time. The power-up reset also prevents the module from sending data to the Module Bus. The data transmission rate is 500 kbps.
Document No 8001014-1
5
Datex-Ohmeda S/5 monitors
Isolated section
r o
Receive data
sess o
Send
r
send
p e
dat
/receive
Reset
a
Opto isolation
Receive data
Send data
Reset
RS485 Driver
send/r
RS485 Driver
Data
NDa
eceive
Reset in
NReset in
)
nit
s
t
a
u l
ra
bu e
nt ce
to
(modul
to modul
Patient
Isolation
Figure 5 Serial communication and opto isolation sections
There are two opto isolators, one for data and one for the reset signal. Signals are processed on logical high-low levels even though the output of the opto isolators in the isolation section are analog signals.
The reset line is an open collector type with a pull-up resistor so that the microprocessor is able to use its internal watch-dog function.
Power supply section
The module isolated supply voltage is developed from the +15 Vdirty (non-isolated) supply from the Central Unit power supply.
The isolated power supply is a switched-mode circuit, where an FET switch is controlled by an oscillator using a bipolar timer. The frequency of the oscillator is approximately 30 kHz with a pulse ratio 50 %; switching of the FET switch is slow to suppress spurious interference.
A special isolation pulse transformer is used in the circuit. The transformer secondary circuit uses normal linear regulators except for +5 V which uses a low drop type linear regulator.
6
Document No 8001014-1
Loading...
+ 22 hidden pages