Datex-Ohmeda S-5 Service Manual M-NMT

Page 1

Datex-Ohmeda

S/5TM NeuroMuscular Transmission Module, M-NMT (Rev. 02)

Technical Reference Manual Slot

USA Tel. +1-608-221 1551, Fax +1-608-222 9147
www.us.datex-ohmeda.com
All specifications are subject to change without notice.

Document No. 800 1018-1

June 2001

Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.
Page 2
Page 3

Table of contents

TABLE OF CONTENTS
NeuroMuscular Transmission Module, M-NMT
TABLE OF CONTENTS i
TABLE OF FIGURES ii
Introduction 1
1 Specifications 2
1.1 General specifications ..............................................................................................................................2
1.2 Technical specifications............................................................................................................................2
1.2.1 NMT.................................................................................................................................................2
1.2.2 Stimulator........................................................................................................................................2
1.2.3 Regional block mode........................................................................................................................2
2 Functional description 3
2.1 Measurement principle .............................................................................................................................3
2.1.1 Nerve stimulation.............................................................................................................................3
2.1.2 Response.........................................................................................................................................4
2.1.3 Regional block .................................................................................................................................4
2.2 Main components.....................................................................................................................................4
2.2.1 NMT board.......................................................................................................................................4
2.3 Connectors and signals.............................................................................................................................6
2.3.1 Module bus connector......................................................................................................................6
2.3.2 Front panel connector.......................................................................................................................7
3 Service procedures 8
3.1 General service information.......................................................................................................................8
3.2 Service check ...........................................................................................................................................9
3.2.1 Recommended tools ........................................................................................................................9
3.3 Disassembly and reassembly..................................................................................................................13
4 Troubleshooting 14
4.1 Troubleshooting chart .............................................................................................................................14
4.2 Troubleshooting flowchart .......................................................................................................................15
5Service Menu 16
5.1 NMT menu..............................................................................................................................................17
6 Spare parts 19
6.1 Spare part list.........................................................................................................................................19
6.1.1 NMT Module, M-NMT rev. 00...........................................................................................................19
6.1.2 NMT Module, M-NMT rev. 01...........................................................................................................19
6.1.3 NMT Module, M-NMT rev. 02...........................................................................................................19
6.1.4 Front panel stickers for AS/3 modules (square buttons) ...................................................................20
6.1.5 Front panel stickers for S/5 modules (round buttons).......................................................................20
7 Earlier revisions 21
APPENDIX A 23
Document No. 800 1018-1
i
Page 4
Datex-Ohmeda S/5 monitors
Service check form A-1

TABLE OF FIGURES

Figure 1 NeuroMuscular Transmission Module, M-NMT ....................................................................................... 1
Figure 2 Principle of response measurement ......................................................................................................4
Figure 3 Serial communication and opto isolation...............................................................................................5
Figure 4 Stimulator block diagram......................................................................................................................5
Figure 5 Module troubleshooting flowchart .......................................................................................................15
ii
Document No. 800 1018-1
Page 5

INTRODUCTION

This section provides information for the maintenance and service of the NeuroMuscular Transmission Module, M-NMT. The M-NMT module is a single width plug-in module designed for use with the S/5 Anesthesia and Compact Anesthesia Monitors. The module contains peripheral nerve stimulation and response measurement which supports electromyography EMG. The module can also be used as a nerve locator for regional nerve blocking with a regional block cable. However, in this case there is no response measurement.
NeuroMuscular Transmission Module, M-NMT
Figure 1 NeuroMuscular Transmission Module, M-NMT
1
Document No. 800 1018-1
Page 6
Datex-Ohmeda S/5 monitors

1 SPECIFICATIONS

1.1 General specifications
Module size, W × D × H 37 × 180 × 112 mm / 1.5 × 7.1 × 4.4 in Module weight 0.37 kg /0.8 lbs Power consumption 3.3 W
1.2 Technical specifications
1.2.1 NMT
Stimulation modes Train of four (TOF)
Measurement intervals for TOF and DBS
Measurement intervals for ST Manual; 1 s, 10 s, 20 s
Double burst (3.3) (DBS) Single twitch (ST) 50 Hz tetanus + post tetanic count (PTC)
Manual; 10 s, 12 s, 15 s, 20 s, 1 min, 5 min, 15 min
1.2.2 Stimulator
Stimulus pulse Square wave, constant current Pulse width 100, 200 or 300 µs Stimulus current range 10...70 mA with 5 mA steps Stimulus current accuracy 10 % or ±3 mA whichever is the greater Max. load 3 k Max. voltage 300 V
1.2.3 Regional block mode
Stimulation modes Single twitch Intervals 1 s, 2 s, 3 s Stimulus pulse Square wave, constant current Pulse width 40 µs Stimulus current range 0...5.0 mA with 0.1 mA steps Stimulus current accuracy 20 % or 0.3 mA whichever is the greater
2
Document No. 800 1018-1
Page 7

2 FUNCTIONAL DESCRIPTION

2.1 Measurement principle
2.1.1 Nerve stimulation
There are three stimulus modes in the NeuroMuscular Transmission Module: Train of Four (TOF), Double Burst 3,3 (DBS) and Single Twitch (ST).
In the Train of Four stimulus mode, four stimulation pulses are generated at 0.5 second intervals. The response is measured after each stimulus and the ratio of the fourth and first response of the TOF sequence is calculated (TOF%).
NOTE: If the first response does not exceed a certain signal level, TOF% is not calculated due to poor accuracy.
Double burst (3,3) stimulation includes two bursts with a 750 ms interval. Both bursts consist of three pulses separated by 20 ms intervals. The responses of both bursts are measured, and the ratio of the second and first response is calculated (DBS%). EMG responses are measured immediately after the first stimulus pulse of both bursts.
NeuroMuscular Transmission Module, M-NMT
Tetanic/PTC
In Single Twitch stimulation, one stimulation pulse is generated. The response is measured after the stimulus. In order to prevent decurarization of the stimulated area, the measurement is automatically stopped after 5 minutes stimulation in 1 sec cycle time.
Tetanic/PTC (Post Tetanic Count) can measure deeper relaxation than TOF. The tetanic stimulation is produced when Start is chosen under Tetanic/PTC. The length of stimulation is 5 seconds. The stimulation generates pulses with a frequency of 50 Hz and with a selected pulse width and current. After tetanic stimulation and a three second delay, Single Twitch stimulation is produced to detect the post tetanic count (PTC). PTC describes the number of responses detected after tetanic stimulation. If there is no response, the measurement will be stopped. If responses will not fade away, a maximum of 20 responses will be calculated. If more can be detected, the PTC value is displayed only as '> 20' and measurement will be stopped. If the TOF, DBS or ST measurement cycle was on when tetanic stimulation started, the cycle will continue after the PTC. After completing the PTC measurement during 1 minute TOF, DBS or another PTC measurement is not possible. This is to avoid erroneous readings due to post tetanic potentiation.
Document No. 800 1018-1
3
Page 8
Datex-Ohmeda S/5 monitors
A
µ
2.1.2 Response
Before each stimulation, the sequence offset, noise and threshold for the response detection is measured. Offset is a baseline of the noise measurement. Noise is calculated by the same algorithm as the response signal itself. The response detection threshold is calculated based on the noise, and if the response is not greater than the threshold then it is interpreted as no response.
EMG
Mechano
Sensor
Measurement
/D & P
Timing control
EMG leads off
Figure 2 Principle of response measurement
EMG measurement
The EMG response is measured as integrated muscle activity. The EMG measurement starts 3 ms after the stimulation and lasts 15 ms. The 3 ms delay helps to prevent the effect of stimulation artifact.
Mechano sensor measurement
Response is measured as movement of the thumb, which is the area of positive signal.
2.1.3 Regional block
A regional block cable can be used as a nerve locator in local anesthesia. A maximum current of
5.0 mA is given every, every other or every 3rd second. The response measurement is ocular.
2.2 Main components
2.2.1 NMT board
The NMT Circuit Board consists of the following functional sections:
constant current stimulator
measuring electronics for the EMG signals
4
Document No. 800 1018-1
Page 9
microprocessor for the stimulation and measuring control, and for counting the measuring
serial communication
The serial bus speed is 500 kbps and the bus itself is half duplex, i.e. data can be transferred in both directions but only one way at a time.
Serial communication
NeuroMuscular Transmission Module, M-NMT
results
Stimulator
Figure 3 Serial communication and opto isolation
The constant current stimulator generates pulses whose amplitude is independent of the load. The main components of the stimulator are a transformer, a capacitor and a transistor. The transformer produces a high voltage which charges the capacitor and the transistor adjusts the pulse width and amplitude of the current.
Power supply
Current
80C196KD
PWM
HSO
Patient
Constant current generator
adjustment
Current control
Continuous stimulation congestion
Figure 4 Stimulator block diagram
5
Document No. 800 1018-1
Page 10
Datex-Ohmeda S/5 monitors
2.3 Connectors and signals
2.3.1 Module bus connector
13
25
1
14
Table 1 Module bus connector (X1)
Pin No. I/O Signal
1 I RESET_RS485
2 I -15 VDC
3 I +15 VDIRTY
4 I +15 VDC
5 I/O -DATA_RS485
6 I/O DATA_RS485
7 - Ground & Shield
8 I -RESET_RS485
9 I CTSB
10 O RTSB
11 I RXDB
12 O TXDB
13 - Ground & Shield
14 I +32 VDIRTY
15 I GroundDIRTY
16 I CTSC
17 O RTSC
18 I RXDC
19 O TXDC
20 - ON/STANDBY
21 - PWM_ECG
22 - RXDD_RS232
23 - TXDD_RS232
24 I +5 VDC
25 I +5 VDC
6
Document No. 800 1018-1
Page 11
2.3.2 Front panel connector
NMT connector (NMT)
6
3
9
1
4
7
2
0
5
8
Pin No. Signal
1
2
3
4
5
6
7
8
9
10
EMG Signal +
EMG Signal -
Not Used
Stimulus +
Stimulus -
Ground
Not Connected
Sensor Identification
+5 V
Mechanical Signal
NeuroMuscular Transmission Module, M-NMT
Document No. 800 1018-1
7
Page 12
Datex-Ohmeda S/5 monitors

3 SERVICE PROCEDURES

3.1 General service information

Field service of the NeuroMuscular Transmission Module, M-NMT, is limited to replacing faulty circuit boards or mechanical parts. Faulty circuit boards should be returned to Datex-Ohmeda for repair.
Datex-Ohmeda is always available for service advice. Please provide the unit serial number, full type designation, and a detailed description of the fault.
The Datex-Ohmeda NMT Stimulator (order code 871251) is recommended for functional checks.
CAUTION Only trained personnel with appropriate equipment should perform the tests and repairs outlined in
this section. Unauthorized service may void warranty of the unit.
8
Document No. 800 1018-1
Page 13

3.2 Service check

These instructions include complete procedures for a service check. The service check is recommended to be performed after any service repair. However, the service check procedures can also be used for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (Appendix A) which should be filled in when performing the procedures.
NeuroMuscular Transmission Module, M-NMT
The mark
the procedure.
The procedures are designed for monitors with S/5 monitor software of revision 01. However, most of the procedures also apply to monitors, which contain some other monitor software type/revision.
? in the instructions means that the check form should be signed after performing
3.2.1 Recommended tools
Tool Order No. Notes
NMT simulator 871251
M-NMT ElectroSensor
M-NMT MechanoSensor
M-NMT Sensor Cable
3 k resistor
Screwdriver
Detach the module box by removing the two screws from the back of the module. Be careful
with the loose latch and spring locking pin.
1. Check internal parts:
screws are tightened properly
cables are connected properly
the EMC cover is attached properly
there are no loose objects inside the module
?
2. Check external parts:
the front cover and the front panel sticker are intact
connectors are intact and are attached properly
the module box, latch and spring locking pin are intact
?
Refit the module box and check that the latch moves properly.
9
Document No. 800 1018-1
Page 14
Datex-Ohmeda S/5 monitors
Turn the monitor on and wait until the normal monitoring screen appears.
Configure the monitor screen so that information regarding the NMT measurement is shown,
for example:
Monitor Setup - Screen 1 Setup - Digit Fields - Field 4 - Nmt
Preset the NMT measurement settings:
Others - NMT - Stimulus Mode - TOF
3. Plug in the module. Check that it goes in smoothly and locks up properly
Set Cycle Time - 10 sec NMT Setup - Current - S(70 mA)
Pulse Width - 200 Stim. Beep Volume - 2
µµµµ
S
?
4. Check that the module is recognized, i.e. the NMT header with related information appear in
the chosen digit field.
?
5. Enter the service menu:
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
Take down the information regarding NMT software by selecting SCROLL VERS and turning the ComWheel.
?
6. Enter the NMT module service menu:
Parameters - NMT
Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not increasing faster than by 50 per second. Check that the module’s memories have passed the internal memory test, i.e. “RAM”, “ROM” and “EEPROM” all state OK.
10
Document No. 800 1018-1
?
7. Check the front panel START-UP and STOP/CONTINUE membrane keys.
Press each key for at least one second and check that the key being pressed is identified, i.e. the color under the corresponding text on the menu turns blue.
?
Page 15
NeuroMuscular Transmission Module, M-NMT
8. Check that the message “Cable off” is shown in the digit field and that “Cable” in the service
menu states OFF.
Plug the M-NMT Sensor Cable with the M-NMT ElectroSensor into front panel connector NMT. Check that the message in the digit field changes to “Measurement OFF” and “Cable” on the service menu states EMG and ELECTR. OFF.
?
9. Perform the stimulus current test.
Connect a 3 k resistor between the ElectroSensor’s stimulus electrode leads (brown and white).
Start the test by highlighting START CURR. TEST on the service menu and pressing the ComWheel. Check that the test was successful with all three test currents, i.e. the “Current test (mA):” on the menu states 30 OK, 50 OK and 70 OK.
?
Connect the M-NMT ElectroSensor leads to the NMT simulator. Set the switch on the
simulator to “Fade off” and turn the knob to “max”. Check “Cable” on the service menu now states only EMG.
10. Start NMT measurement (TOF) by pressing the START-UP key on the module.
When the message “Supramax search” changes to “Setting reference” in the digit field, check that the supramaximal current detected is less than 70 mA, i.e. the “Current set” value on the service menu is less than 700.
?
11. Check that the module gives four successive stimulus pulses with 10 second intervals. A
small asterisk () should be shown in the digit field during each of the stimulus pulses and simultaneous sound signals should be heard from the loudspeaker.
Check that on the service menu the values for “T1%”, “T2%”, “T3%”, “T4%” and “Ratio%” are all within 950-1059.
Check also that in the digit field the “TOF%” value is within 95-105, “Count” is 4 and “T1%” is within 95-105.
?
12. Check that the “Noise” value on the service menu stays under 50.
?
13. Change the stimulus pulse width to 100 µs through the NMT service menu:
NMT Setup - Pulse Width - 100
µµµµ
S
Document No. 800 1018-1
11
Page 16
Datex-Ohmeda S/5 monitors
Check that the “TOF%” value is still within 95-105, “Count” is 4 and “T1%” is within 95-105 in the digit field.
Check the same parameters with a stimulus pulse width of 300 µs.
?
14. Turn the knob on the NMT simulator to “0”.
Check that on the service menu the values for “T1%”, “T2%”, “T3%”, “T4%” turn to 0 and the “Ratio%” states - - -. In the digit field “TOF%” should also state - - - , and “Count” and “T1%” should show 0.
Turn the NMT simulator knob back to “max”.
?
15. Change the stimulus mode to Double Burst Stimulation (DBS) through the service menu:
NMT Setup - Stimulus Mode - DBS
Check that the module now gives only two stimulus pulses with a 10 seconds interval.
Check that on the service menu the values for “T1%”, “T2%”, and “Ratio%” are still within 950-1059. In the digit field the “DBS%” value should be within 95-105, “Count” is 2 and “T1%” is within 95-105.
?
16. Change the stimulus mode to Single Twitch Stimulation (ST):
NMT Setup - Stimulus Mode - ST
Check that the module starts to give only one stimulus pulse with a 1 second interval. Note the time when the ST stimulation started.
Check that on the service menu the value for “T1%” is within 950-1059. In the digit field the “Count” value should be 1 and “T1%” within 95-105. Let the monitor continue to give single twitch stimulation.
?
17. Check that the NMT measurement stops and the message “Measurement OFF” appears in
the digit field for NMT five minutes after the start of the ST stimulation.
12
Document No. 800 1018-1
?
18. Replace the M-NMT ElectroSensor with the M-NMT MechanoSensor and check that “Cable”
on the service menu states PIEZO.
?
Page 17
19. Perform an electrical safety check and a leakage current test.
?
20. Check that the module functions normally after performing the electrical safety check.
?
21. Clean the module with suitable detergent.
?
Fill in all necessary documents.
3.3 Disassembly and reassembly
Disassemble the NeuroMuscular Transmission Module, M-NMT, in the following way. See the exploded view of the module.
1. Remove the two screws from the back of the module.
NeuroMuscular Transmission Module, M-NMT
2. Pull the module box slowly rearward and detach it from main body. Be careful with the loose
latch and spring locking pin.
3. Detach the NMT board by removing the two screws located near the front panel frame,
disconnect the cables and pull out the front panel frame.
To reassemble the module, reverse the order of the disassembly steps.
CAUTION When reassembling the module, make sure that the cables are reconnected
properly.
13
Document No. 800 1018-1
Page 18
Datex-Ohmeda S/5 monitors

4 TROUBLESHOOTING

4.1 Troubleshooting chart
Trouble Cause Treatment
Check the stimulus electrodes.
EMG electrode off.
Supramax. not found. Loose electrodes or loose stimulus
Response too weak. Loose stimulus electrodes.
Ref. not stable. Patient is relaxated.
Loose electrodes or loose stimulus clip.
clip.
Stimulus electrodes attached to wrong place.
Patient is relaxated.
Measuring electrodes attached to wrong place.
Patient is relaxated.
Movement artifact.
Change or attach the electrodes or clip.
Change or attach the electrodes or clip.
Change the place of the stimulus electrode.
Change or attach the electrodes.
Change the place of the meas. electrodes.
14
Document No. 800 1018-1
Page 19
4.2 Troubleshooting flowchart
Possible fault in NMT module
Open NMT Service Menu
Check Timeouts, Bad checksums,
Bad c-s by mod
no
OK?
yes
Check RAM,ROM,EEPROM
OK?
yes
Test stimulus current
OK?
yes
Check that sensor is identified
OK?
yes
Connect ElectroSensor to NMT
stimulator and start measurement
no
no
Does another module
work in the same slot?
Fault not in NMT module
Replace NMT board
Does the test pass with
another sensor?
yes
Replace the original sensor Connect the connector
Does the test pass with
another sensor?
no
Does the test pass with
another NMT sensor
no
Does the test pass with
another NMT input
no
Replace NMT board
no
cable?
board?
no
Does the test pass with
another NMT sensor
yes
Replace the original
NMT sensor cable
cable?
NeuroMuscular Transmission Module, M-NMT
yes
no yesno
Are the NMT input board
wires connected?
no yes
yes
yes
yes
Replace NMT board
Does the test pass
with another NMT
input board?
Replace the original
NMT input board
Replace the original sensor
Replace the original NMT sensor
cable
Replace the original NMT input
board
no
Replace NMT board
Responses OK?
yes
Check front panel key functions
OK?
no
no
Does the test pass with
another sensor?
no
Does the test pass with
another NMT sensor
cable
no
Does the test pass with
another NMT input
board?
no
Replace NMT board
Is the keypad
connected?
yes
Does the test pass with
another keypad?
no
Replace NMT board
Figure 5 Module troubleshooting flowchart
yes
yes
yes
no
yes
Replace the original sensor
Replace the original NMT sensor
cable
Replace the original NMT input
board
Connect the keypad
Replace the original keypad
15
Document No. 800 1018-1
Page 20
Datex-Ohmeda S/5 monitors

5 SERVICE MENU

1. Press the Monitor Setup key.
2. Select Install/Service (password 16-4-34).
3. Select Service (password 26-23-8).
4. Select Parameters.
5. Select NMT.
16
Document No. 800 1018-1
Page 21
5.1 NMT menu
NeuroMuscular Transmission Module, M-NMT
Start Curr. test Start Curr. test is a test where the module itself checks the difference between a given current and
the measured current. A 3 k before starting the test. All the currents checked will be displayed on the service data screen. After the current value, the status of the test is shown. If the test is not passed, send the module back to the factory for calibration.
T1%/ref T1% / ref gives a selection for changing settings related to the M-NMT MechanoSensor. A setting
of 3 should always be used for normal monitoring. The settings 1 and 2 are for research purposes only.
Record Data Record Data prints out the displayed service data and circuit board information (id., serial number,
and software id.) onto the Recorder Module, M-REC.
resistance should be connected between the stimulus electrodes
Service Data Cable shows the type of cable being used.
Cable check value shows the bit amount. From the following chart you can check the value for
each cable:
EMG -100...+100
Cable OFF >1950
Piezo 900...1100
Regional block 600....800
Module Keys checks the function of the module keys. A blue background appears at the back of the text when a key is pressed for more than one second.
Stimulus mode shows the selected stimulus. Stimulus mode can be changed using the NMT Setup menu.
Measurement indicates ON/OFF.
T1%, T2% shows the measured response. A value of 1000 corresponds to 100%.
17
Document No. 800 1018-1
Page 22
Datex-Ohmeda S/5 monitors
Absolute T1, T2 shows the voltage measured from the A/D converter.
Noise indicates the interference just before the measurement. A typical value is <10.
Offset is an average of the noise measurement. A typical value is 510.
Curr set is the selected current, a value of 700 corresponds to 70 mA.
Curr meas is the measured current, a value of 700 corresponds to 70 mA.
Pulses indicates pulses the module has produced.
Piezo probe T1% and ref. search shows information related to the MechanoSensor settings.
Timeouts is a cumulative number that indicates how many times the module has not responded to
the monitor's inquiry.
Bad checksums is a cumulative number that indicates how many times communication from the module to the monitor has failed. Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. The monitor starts counting these items at power up and resets to zero at power off. The non-zero values do not indicate a failure, but the continuous counting (more than 50 per second) indicates either serial communication failure, or module not in place. Also, other modules can cause communication errors that cause these numbers to rise.
RAM indicates the state of the RAM memory. ROM indicates whether the checksum in the EPROM is in accordance with the software calculated
value. EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error).
18
Document No. 800 1018-1
Page 23

6 SPARE PARTS

6.1 Spare part list
NOTE: Accessories are listed in the Patient Monitor Supplies and Accessories.
6.1.1 NMT Module, M-NMT rev. 00
NeuroMuscular Transmission Module, M-NMT
6
4
1
3
2
5
7
8
10
9
11
Item Description Order No. Item Description Order No.
1 Module box (single width) 886167 8 NMT board, M-NMT (Rev. 00-01) *887487
2 Latch 879181 9 EMC cover, M-NMT 886320
3 Spring pin 879182 10 Cross cylinder-head screw M3x12 628700
4 Cross recess screw M3x8 black 616215 11 NMT input board 887184
5 Front panel unit, M-NMT 887186 12 Membrane keypad 880101
6 Cross cylinder-head screw M3x6 61721 13 Front panel sticker, see 6.1.3 -
7 Metal frame 879184 - - -
12
13
The front panel unit includes the connector and input board. * this part is recommended for stock
6.1.2 NMT Module, M-NMT rev. 01
No new spare parts.
6.1.3 NMT Module, M-NMT rev. 02
No new spare parts.
19
Document No. 800 1018-1
Page 24
Datex-Ohmeda S/5 monitors
6.1.4 Front panel stickers for AS/3 modules (square buttons)
Front panel stickers that are related to the Compact Module type and adaptation:
Adaptation codes: DA=Danish, DE=German, EN=English, ES=Spanish, FI=Finnish, FR=French, IT=Italian, JA=Japanese, NL=Dutch, NO=Norwegian, PT=Portuguese, SV=Swedish
Item 13.
M-NMT (rev. 00-01)
Adaptation
DA 892217
DE 886003
EN 886002
ES 887498
FI 888873
FR 886004
IT 887542
JA 888318
NL 887370
NO 893572
PT 895240
SV 887369
Order No.
6.1.5 Front panel stickers for S/5 modules (round buttons)
Front panel stickers that are related to the Compact Module type and adaptation:
Adaptation codes: DA=Danish, DE=German, EN=English, ES=Spanish, FI=Finnish, FR=French, IT=Italian, JA=Japanese, NL=Dutch, NO=Norwegian, PT=Portuguese, SV=Swedish
M-NMT (rev. 02)
Adaptation
DA 898846
DE 898837
EN 898836
ES 898840
FI 898843
FR 898838
IT 898841
JA 8000383
NL 898839
NO 898845
PT 898842
SV 898844
Order No.
20
Document No. 800 1018-1
Page 25

7 EARLIER REVISIONS

This manual also supports earlier module revision 00.
NeuroMuscular Transmission Module, M-NMT
21
Document No. 800 1018-1
Page 26
Datex-Ohmeda S/5 monitors
22
Document No. 800 1018-1
Page 27
APPENDIX A, Service Check Form, M-NMT

APPENDIX A

23
Document No. 800 1018-1
Page 28
Datex-Ohmeda S/5 monitors
24
Document No. 800 1018-1
Page 29
APPENDIX A, Service Check Form, M-NMT

SERVICE CHECK FORM

NeuroMuscular Transmission Module, M-NMT
Customer
Service
Service engineer Date
OK = Test OK N.A. = Test not applicable Fail = Test Failed
OK N.A. Fail OK N.A. Fail
1. Internal parts 2. External parts
3. Installation 4. Recognition
5. Module software
6. Communication and
memories
8. ElectroSensor
recognition
10. Supramaximal current < 70 mA
NMT
Module type S/N
7. Membrane keys
9. Stimulus current test
11. TOF measurement with NMT simulator
T1%
T2% 950-1059
T3% 950-1059
T4% 950-1059
Ratio% 950-1059
TOF% 95-105
Count 4
T1% 95-105
12. Noise < 50
950-1059
Document No. 800 1018-1
A-1(2)
Page 30
Datex-Ohmeda S/5 monitors
13. Stimulus pulse width
100 µs 300 µs Allowed range
TOF% 95-105
Count 4
T1% 95-105
OK N.A. Fail
14. No response
15. DBS measurement with NMT simulator
T1%
T2% 950-1059
Ratio% 950-1059
DBS% 95-105
Count 2
T1% 95-105
16. ST measurement with NMT simulator
T1%
Count 1
T1% 95-105
17. Automatic measurement off 18. MechanoSensor
recognition
19. Electrical safety check 20. Functioning after
electrical safety check
21. Final cleaning
950-1059
950-1059
Notes
Used Spare Parts
Signature
A-2(2)
Document No. 800 1018-1
Loading...