Datex-Ohmeda Inc.
3030 Ohmeda Drive
53707-7550 MADISON, WIS
USA
Tel. +1-608-221 1551, Fax +1-608-222 9147
www.us.datex-ohmeda.com
All specifications are subject to change without notice.
Document No. 800 1018-1
June 2001
Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.
Table of contents
TABLE OF CONTENTS
NeuroMuscular Transmission Module, M-NMT
TABLE OF CONTENTSi
TABLE OF FIGURESii
Introduction1
1Specifications2
1.1 General specifications ..............................................................................................................................2
2.2 Main components.....................................................................................................................................4
2.3 Connectors and signals.............................................................................................................................6
2.3.1 Module bus connector......................................................................................................................6
2.3.2 Front panel connector.......................................................................................................................7
3Service procedures8
3.1 General service information.......................................................................................................................8
3.2 Service check ...........................................................................................................................................9
6.1 Spare part list.........................................................................................................................................19
This section provides information for the maintenance and service of the NeuroMuscular
Transmission Module, M-NMT. The M-NMT module is a single width plug-in module designed for
use with the S/5 Anesthesia and Compact Anesthesia Monitors. The module contains peripheral
nerve stimulation and response measurement which supports electromyography EMG. The module
can also be used as a nerve locator for regional nerve blocking with a regional block cable.
However, in this case there is no response measurement.
NeuroMuscular Transmission Module, M-NMT
Figure 1NeuroMuscular Transmission Module, M-NMT
1
Document No. 800 1018-1
Datex-Ohmeda S/5 monitors
1 SPECIFICATIONS
1.1 General specifications
Module size, W × D × H37 × 180 × 112 mm / 1.5 × 7.1 × 4.4 in
Module weight0.37 kg /0.8 lbs
Power consumption3.3 W
1.2 Technical specifications
1.2.1 NMT
Stimulation modesTrain of four (TOF)
Measurement intervals for TOF and DBS
Measurement intervals for ST Manual; 1 s, 10 s, 20 s
Double burst (3.3) (DBS)
Single twitch (ST)
50 Hz tetanus + post tetanic count (PTC)
Stimulus pulseSquare wave, constant current
Pulse width100, 200 or 300 µs
Stimulus current range10...70 mA with 5 mA steps
Stimulus current accuracy10 % or ±3 mA whichever is the greater
Max. load3 kΩ
Max. voltage300 V
1.2.3 Regional block mode
Stimulation modesSingle twitch
Intervals1 s, 2 s, 3 s
Stimulus pulseSquare wave, constant current
Pulse width40 µs
Stimulus current range0...5.0 mA with 0.1 mA steps
Stimulus current accuracy20 % or 0.3 mA whichever is the greater
2
Document No. 800 1018-1
2 FUNCTIONAL DESCRIPTION
2.1 Measurement principle
2.1.1 Nerve stimulation
There are three stimulus modes in the NeuroMuscular Transmission Module: Train of Four (TOF),
Double Burst 3,3 (DBS) and Single Twitch (ST).
In the Train of Four stimulus mode, four stimulation pulses are generated at 0.5 second intervals.
The response is measured after each stimulus and the ratio of the fourth and first response of the
TOF sequence is calculated (TOF%).
NOTE: If the first response does not exceed a certain signal level, TOF% is not calculated due to
poor accuracy.
Double burst (3,3) stimulation includes two bursts with a 750 ms interval. Both bursts consist of
three pulses separated by 20 ms intervals. The responses of both bursts are measured, and the
ratio of the second and first response is calculated (DBS%). EMG responses are measured
immediately after the first stimulus pulse of both bursts.
NeuroMuscular Transmission Module, M-NMT
Tetanic/PTC
In Single Twitch stimulation, one stimulation pulse is generated. The response is measured after
the stimulus. In order to prevent decurarization of the stimulated area, the measurement is
automatically stopped after 5 minutes stimulation in 1 sec cycle time.
Tetanic/PTC (Post Tetanic Count) can measure deeper relaxation than TOF. The tetanic stimulation
is produced when Start is chosen under Tetanic/PTC. The length of stimulation is 5 seconds. The
stimulation generates pulses with a frequency of 50 Hz and with a selected pulse width and
current. After tetanic stimulation and a three second delay, Single Twitch stimulation is produced
to detect the post tetanic count (PTC). PTC describes the number of responses detected after
tetanic stimulation. If there is no response, the measurement will be stopped. If responses will not
fade away, a maximum of 20 responses will be calculated. If more can be detected, the PTC value
is displayed only as '> 20' and measurement will be stopped. If the TOF, DBS or ST measurement
cycle was on when tetanic stimulation started, the cycle will continue after the PTC. After
completing the PTC measurement during 1 minute TOF, DBS or another PTC measurement is not
possible. This is to avoid erroneous readings due to post tetanic potentiation.
Document No. 800 1018-1
3
Datex-Ohmeda S/5 monitors
A
µ
2.1.2 Response
Before each stimulation, the sequence offset, noise and threshold for the response detection is
measured. Offset is a baseline of the noise measurement. Noise is calculated by the same
algorithm as the response signal itself. The response detection threshold is calculated based on
the noise, and if the response is not greater than the threshold then it is interpreted as no response.
EMG
Mechano
Sensor
Measurement
/D & P
Timing control
EMG leads off
Figure 2Principle of response measurement
EMG measurement
The EMG response is measured as integrated muscle activity. The EMG measurement starts 3 ms
after the stimulation and lasts 15 ms. The 3 ms delay helps to prevent the effect of stimulation
artifact.
Mechano sensor measurement
Response is measured as movement of the thumb, which is the area of positive signal.
2.1.3 Regional block
A regional block cable can be used as a nerve locator in local anesthesia. A maximum current of
5.0 mA is given every, every other or every 3rd second. The response measurement is ocular.
2.2 Main components
2.2.1 NMT board
The NMT Circuit Board consists of the following functional sections:
• constant current stimulator
• measuring electronics for the EMG signals
4
Document No. 800 1018-1
• microprocessor for the stimulation and measuring control, and for counting the measuring
• serial communication
The serial bus speed is 500 kbps and the bus itself is half duplex, i.e. data can be transferred in
both directions but only one way at a time.
Serial communication
NeuroMuscular Transmission Module, M-NMT
results
Stimulator
Figure 3Serial communication and opto isolation
The constant current stimulator generates pulses whose amplitude is independent of the load. The
main components of the stimulator are a transformer, a capacitor and a transistor. The transformer
produces a high voltage which charges the capacitor and the transistor adjusts the pulse width and
amplitude of the current.
Power supply
Current
80C196KD
PWM
HSO
Patient
Constant current
generator
adjustment
Current
control
Continuous
stimulation
congestion
Figure 4Stimulator block diagram
5
Document No. 800 1018-1
Loading...
+ 21 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.