Datex-Ohmeda S-5 Service Manual M-NMT

Datex-Ohmeda

S/5TM NeuroMuscular Transmission Module, M-NMT (Rev. 02)

Technical Reference Manual Slot

USA Tel. +1-608-221 1551, Fax +1-608-222 9147
www.us.datex-ohmeda.com
All specifications are subject to change without notice.

Document No. 800 1018-1

June 2001

Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.

Table of contents

TABLE OF CONTENTS
NeuroMuscular Transmission Module, M-NMT
TABLE OF CONTENTS i
TABLE OF FIGURES ii
Introduction 1
1 Specifications 2
1.1 General specifications ..............................................................................................................................2
1.2 Technical specifications............................................................................................................................2
1.2.1 NMT.................................................................................................................................................2
1.2.2 Stimulator........................................................................................................................................2
1.2.3 Regional block mode........................................................................................................................2
2 Functional description 3
2.1 Measurement principle .............................................................................................................................3
2.1.1 Nerve stimulation.............................................................................................................................3
2.1.2 Response.........................................................................................................................................4
2.1.3 Regional block .................................................................................................................................4
2.2 Main components.....................................................................................................................................4
2.2.1 NMT board.......................................................................................................................................4
2.3 Connectors and signals.............................................................................................................................6
2.3.1 Module bus connector......................................................................................................................6
2.3.2 Front panel connector.......................................................................................................................7
3 Service procedures 8
3.1 General service information.......................................................................................................................8
3.2 Service check ...........................................................................................................................................9
3.2.1 Recommended tools ........................................................................................................................9
3.3 Disassembly and reassembly..................................................................................................................13
4 Troubleshooting 14
4.1 Troubleshooting chart .............................................................................................................................14
4.2 Troubleshooting flowchart .......................................................................................................................15
5Service Menu 16
5.1 NMT menu..............................................................................................................................................17
6 Spare parts 19
6.1 Spare part list.........................................................................................................................................19
6.1.1 NMT Module, M-NMT rev. 00...........................................................................................................19
6.1.2 NMT Module, M-NMT rev. 01...........................................................................................................19
6.1.3 NMT Module, M-NMT rev. 02...........................................................................................................19
6.1.4 Front panel stickers for AS/3 modules (square buttons) ...................................................................20
6.1.5 Front panel stickers for S/5 modules (round buttons).......................................................................20
7 Earlier revisions 21
APPENDIX A 23
Document No. 800 1018-1
i
Datex-Ohmeda S/5 monitors
Service check form A-1

TABLE OF FIGURES

Figure 1 NeuroMuscular Transmission Module, M-NMT ....................................................................................... 1
Figure 2 Principle of response measurement ......................................................................................................4
Figure 3 Serial communication and opto isolation...............................................................................................5
Figure 4 Stimulator block diagram......................................................................................................................5
Figure 5 Module troubleshooting flowchart .......................................................................................................15
ii
Document No. 800 1018-1

INTRODUCTION

This section provides information for the maintenance and service of the NeuroMuscular Transmission Module, M-NMT. The M-NMT module is a single width plug-in module designed for use with the S/5 Anesthesia and Compact Anesthesia Monitors. The module contains peripheral nerve stimulation and response measurement which supports electromyography EMG. The module can also be used as a nerve locator for regional nerve blocking with a regional block cable. However, in this case there is no response measurement.
NeuroMuscular Transmission Module, M-NMT
Figure 1 NeuroMuscular Transmission Module, M-NMT
1
Document No. 800 1018-1
Datex-Ohmeda S/5 monitors

1 SPECIFICATIONS

1.1 General specifications
Module size, W × D × H 37 × 180 × 112 mm / 1.5 × 7.1 × 4.4 in Module weight 0.37 kg /0.8 lbs Power consumption 3.3 W
1.2 Technical specifications
1.2.1 NMT
Stimulation modes Train of four (TOF)
Measurement intervals for TOF and DBS
Measurement intervals for ST Manual; 1 s, 10 s, 20 s
Double burst (3.3) (DBS) Single twitch (ST) 50 Hz tetanus + post tetanic count (PTC)
Manual; 10 s, 12 s, 15 s, 20 s, 1 min, 5 min, 15 min
1.2.2 Stimulator
Stimulus pulse Square wave, constant current Pulse width 100, 200 or 300 µs Stimulus current range 10...70 mA with 5 mA steps Stimulus current accuracy 10 % or ±3 mA whichever is the greater Max. load 3 k Max. voltage 300 V
1.2.3 Regional block mode
Stimulation modes Single twitch Intervals 1 s, 2 s, 3 s Stimulus pulse Square wave, constant current Pulse width 40 µs Stimulus current range 0...5.0 mA with 0.1 mA steps Stimulus current accuracy 20 % or 0.3 mA whichever is the greater
2
Document No. 800 1018-1

2 FUNCTIONAL DESCRIPTION

2.1 Measurement principle
2.1.1 Nerve stimulation
There are three stimulus modes in the NeuroMuscular Transmission Module: Train of Four (TOF), Double Burst 3,3 (DBS) and Single Twitch (ST).
In the Train of Four stimulus mode, four stimulation pulses are generated at 0.5 second intervals. The response is measured after each stimulus and the ratio of the fourth and first response of the TOF sequence is calculated (TOF%).
NOTE: If the first response does not exceed a certain signal level, TOF% is not calculated due to poor accuracy.
Double burst (3,3) stimulation includes two bursts with a 750 ms interval. Both bursts consist of three pulses separated by 20 ms intervals. The responses of both bursts are measured, and the ratio of the second and first response is calculated (DBS%). EMG responses are measured immediately after the first stimulus pulse of both bursts.
NeuroMuscular Transmission Module, M-NMT
Tetanic/PTC
In Single Twitch stimulation, one stimulation pulse is generated. The response is measured after the stimulus. In order to prevent decurarization of the stimulated area, the measurement is automatically stopped after 5 minutes stimulation in 1 sec cycle time.
Tetanic/PTC (Post Tetanic Count) can measure deeper relaxation than TOF. The tetanic stimulation is produced when Start is chosen under Tetanic/PTC. The length of stimulation is 5 seconds. The stimulation generates pulses with a frequency of 50 Hz and with a selected pulse width and current. After tetanic stimulation and a three second delay, Single Twitch stimulation is produced to detect the post tetanic count (PTC). PTC describes the number of responses detected after tetanic stimulation. If there is no response, the measurement will be stopped. If responses will not fade away, a maximum of 20 responses will be calculated. If more can be detected, the PTC value is displayed only as '> 20' and measurement will be stopped. If the TOF, DBS or ST measurement cycle was on when tetanic stimulation started, the cycle will continue after the PTC. After completing the PTC measurement during 1 minute TOF, DBS or another PTC measurement is not possible. This is to avoid erroneous readings due to post tetanic potentiation.
Document No. 800 1018-1
3
Datex-Ohmeda S/5 monitors
A
µ
2.1.2 Response
Before each stimulation, the sequence offset, noise and threshold for the response detection is measured. Offset is a baseline of the noise measurement. Noise is calculated by the same algorithm as the response signal itself. The response detection threshold is calculated based on the noise, and if the response is not greater than the threshold then it is interpreted as no response.
EMG
Mechano
Sensor
Measurement
/D & P
Timing control
EMG leads off
Figure 2 Principle of response measurement
EMG measurement
The EMG response is measured as integrated muscle activity. The EMG measurement starts 3 ms after the stimulation and lasts 15 ms. The 3 ms delay helps to prevent the effect of stimulation artifact.
Mechano sensor measurement
Response is measured as movement of the thumb, which is the area of positive signal.
2.1.3 Regional block
A regional block cable can be used as a nerve locator in local anesthesia. A maximum current of
5.0 mA is given every, every other or every 3rd second. The response measurement is ocular.
2.2 Main components
2.2.1 NMT board
The NMT Circuit Board consists of the following functional sections:
constant current stimulator
measuring electronics for the EMG signals
4
Document No. 800 1018-1
microprocessor for the stimulation and measuring control, and for counting the measuring
serial communication
The serial bus speed is 500 kbps and the bus itself is half duplex, i.e. data can be transferred in both directions but only one way at a time.
Serial communication
NeuroMuscular Transmission Module, M-NMT
results
Stimulator
Figure 3 Serial communication and opto isolation
The constant current stimulator generates pulses whose amplitude is independent of the load. The main components of the stimulator are a transformer, a capacitor and a transistor. The transformer produces a high voltage which charges the capacitor and the transistor adjusts the pulse width and amplitude of the current.
Power supply
Current
80C196KD
PWM
HSO
Patient
Constant current generator
adjustment
Current control
Continuous stimulation congestion
Figure 4 Stimulator block diagram
5
Document No. 800 1018-1
Loading...
+ 21 hidden pages