Datex-Ohmeda S-5 User manual

Page 1
Datex-Ohmeda
Datex-Ohmeda S/5 Central, ViewStation and Network
Technical Reference Manual
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Datex-Ohmeda Inc. 3 Highwood Drive Tewksbury, MA 01876
Tel. (978) 640-0460 Fax (978) 640-0469
www.us.datex-ohmeda.com
0537
Conformity according to the Council Directive 93/42/EEC concerning Medical Devices.
All specifications are subject to change without notice.
Outside the USA, check local laws for any restriction that may apply.
September, 2000
Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 9 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.
Page 2
Datex-Ohmeda S/5 Central, ViewStation and Network Intended purpose
The Datex-Ohmeda Network transfers information between networked Datex-Ohmeda devices in the Datex­Ohmeda monitor network. It also allows information transfer between several Centrals. Within one Datex­Ohmeda monitor network it allows a networked device to display, store, print and otherwise process information received from other networked devices.
The Datex-Ohmeda S/5 Central maintains the network connections between the Datex-Ohmeda bedside monitors and other networked devices in Datex-Ohmeda monitor network. Furthermore, it coordinates the transfer of information between devices in the Datex-Ohmeda Network as well as between the Datex­Ohmeda Network and Hospital Information Systems (HIS).
The Datex-Ohmeda S/5 Central can be used for storing, printing, viewing or otherwise processing of information from several bedside monitors or other networked devices.
The Datex-Ohmeda S/5 ViewStation can be used for printing, viewing or otherwise processing of information from several bedside monitors or other networked devices.
The Datex-Ohmeda Network will be used for patients in the hospital and it is meant for continuous use.
The device is for use by qualified medical personnel only.
CE marking
The Datex-Ohmeda S/5 Central software, S-CNET99, and S/5 ViewStation software, S-VNET99, are
CE marked in accordance with the EU Medical Device Directive. The Datex-Ohmeda S/5 Central Patient Screen and Status Screen display units, computer and
keyboard are CE marked in accordance with the EU Directive for Electromagnetic Compatibility of Low Voltage equipment.
Responsibility of the manufacturer
Datex-Ohmeda Division, Instrumentarium Corp. is responsible for the safety, reliability and performance of the Datex-Ohmeda manufactured products only if:
assembly, operations, extensions, readjustments, modifications, service and repairs are carried out
by personnel authorized by Datex-Ohmeda. electrical installation complies with appropriate requirements.
equipment is used in accordance with the User’s Guide and serviced and maintained in accordance
with the Technical Reference Manual.
Storage and Transport
For allowed storage and transport conditions refer to the documentation delivered with the monitor.
Trademarks
Datex®,Ohmeda®, and other trademarks S/5, AS/3, CS/3, D-lite, Pedi-lite, D-fend, D-fend+, MemCard, ComWheel, EarSat, FingerSat, FlexSat, PatientO
Instrumentarium Corp. or its subsidiaries. All other product and company names are property of their respective owners.
All specifications subject to change without notice.
, Patient Spirometry and Tonometrics are property of
2
Copyright
Instrumentarium Corp., 1994-2000. All rights reserved.
This manual is protected by copyright. No part of this manual may be electronically copied, photocopied, reproduced, or translated to another language without the prior written consent of Datex-Ohmeda.
Page 3
Datex-Ohmeda S/5 Central, ViewStation and Network

TABLE OF CONTENTS

Datex-Ohmeda S/5 Central, ViewStation and Network
1 Introduction 1
1.1 Symbols ...............................................................................................................................................1
1.2 Abbreviations........................................................................................................................................1
1.2.1 Conventions used in this manual.................................................................................................2
1.3 Safety precautions................................................................................................................................2
2 System description 3
2.1 Datex-Ohmeda Monitor Network............................................................................................................3
2.1.1 Datex-Ohmeda S/5 Central ........................................................................................................4
2.1.2 Monitors ....................................................................................................................................4
2.1.3 Datex-Ohmeda S/5 Arrhythmia Workstation................................................................................5
2.2 Datex-Ohmeda Network (TCP/IP)...........................................................................................................5
2.2.1 ViewStation................................................................................................................................6
2.3 Network services...................................................................................................................................7
2.3.1 Central monitoring and remote viewing of patient sites.................................................................7
2.3.2 Monitor-to-monitor communication.............................................................................................7
2.3.3 Storage for the patient data files and preoperative data files ........................................................8
2.3.4 Storage for the record keeping menu files....................................................................................8
2.3.5 Storage for the user modes.........................................................................................................8
2.3.6 Printing and recording trends and snapshots ...............................................................................8
2.3.7 Access to other hospital data networks........................................................................................9
2.4 Time services ........................................................................................................................................9
2.5 Communication in the network ..............................................................................................................9
2.5.1 Data integrity .............................................................................................................................9
3 Cabling 11
3.1 General ..............................................................................................................................................11
3.1.1 Cabling standards ....................................................................................................................11
3.1.2 General restrictions ..................................................................................................................11
3.1.3 Recommended installation tools...............................................................................................11
3.2 Pre-installation testing ........................................................................................................................11
3.2.1 Site survey ...............................................................................................................................11
3.2.2 Site plan ..................................................................................................................................13
3.2.3 Site planning sheet...................................................................................................................14
3.2.4 Pre-installation test binder........................................................................................................14
3.3 Horizontal cabling ...............................................................................................................................14
3.3.1 Mon/IC-Net cables...................................................................................................................15
3.3.2 Mon-Net cables........................................................................................................................15
3.3.3 Wallboxes ................................................................................................................................16
3.3.4 Horizontal cables......................................................................................................................17
3.3.5 Horizontal patch panels............................................................................................................23
3.3.6 Patch cables ............................................................................................................................24
3.3.7 Hubs........................................................................................................................................25
3.4 Vertical cabling ...................................................................................................................................27
3.5 Cabling examples................................................................................................................................29
3.6 Post-installation testing.......................................................................................................................32
Document No. 8000633-1
i
Page 4
Technical Reference Manual
4 Installing Centrals and ViewStations 33
4.1 Components.......................................................................................................................................33
4.2 Unpacking..........................................................................................................................................33
4.3 Placing...............................................................................................................................................34
4.4 Connections .......................................................................................................................................34
5 Configuring 39
5.1 Setup files and tools ...........................................................................................................................39
5.1.1 When to use which setup software? ..........................................................................................39
5.2 Setup Procedure................................................................................................................................. 40
5.2.1 Network setup..........................................................................................................................41
5.2.2 View setup...............................................................................................................................44
5.2.3 System setup...........................................................................................................................46
6 Connecting networks 57 7 Installing a recorder or printer 59
7.1 Installing a recorder ............................................................................................................................59
7.2 Installing a printer...............................................................................................................................59
8 Software Installation 63
8.1 Operating system................................................................................................................................63
8.1.2 Installing the operating system .................................................................................................64
8.1.3 Installing NT Service Pack.........................................................................................................65
8.1.4 Installing driver for the network board........................................................................................65
8.1.5 Installing driver for display controller board ...............................................................................66
8.1.6 Configuring users .....................................................................................................................66
8.2 Installing the network software ............................................................................................................66
9 Functional Check 67
9.1 Checking the equipment and connections............................................................................................67
9.1.1 Checking the Patient Screen.....................................................................................................67
9.1.2 Checking the Status Screen......................................................................................................67
9.1.3 Checking the monitor-to-monitor communication......................................................................67
9.1.4 Checking the communication between the monitor networks .....................................................67
9.1.5 Checking the printer .................................................................................................................67
9.1.6 Checking the recorder ..............................................................................................................68
9.2 Checking the configuration..................................................................................................................68
10 Troubleshooting 81
10.1 Compatibility limitations .....................................................................................................................81
10.2 Network troubleshooting flowchart....................................................................................................... 82
10.3 Central/ViewStation troubleshooting chart...........................................................................................82
10.4 Error messages...................................................................................................................................84
10.5 Technical assistance...........................................................................................................................85
11 Planned maintenance 87 12 Datex-Ohmeda Products 89
12.1 Network Software, S-CNET99 and ViewStation Software S-VNET99.......................................................89
12.1.1Software architecture ...............................................................................................................89
12.1.2Directory Structures..................................................................................................................90
12.2 Start-up Procedure .............................................................................................................................93
ii
Document No. 8000633-1
Page 5
Datex-Ohmeda S/5 Central, ViewStation and Network
12.2.1After having exited the program.................................................................................................93
12.2.2After having switched off the Network Computer.........................................................................93
12.3 System/Installation menu...................................................................................................................93
12.3.1Alarms setup............................................................................................................................94
12.3.2Recorder setup.........................................................................................................................94
12.4 Service Pages .....................................................................................................................................95
12.4.1Viewing Service Pages ..............................................................................................................95
12.4.2Components page ....................................................................................................................95
12.5 Central Keyboard, K-CENTRALB ........................................................................................................ 103
12.5.1Technical specifications......................................................................................................... 103
12.5.2Functional description ........................................................................................................... 103
12.5.3Keyboard connector .............................................................................................................. 105
12.5.4Service and repair procedures................................................................................................ 105
12.5.5Disassembly and reassembly................................................................................................. 105
12.5.6Functional checkout .............................................................................................................. 110
12.5.7Service and repair procedures................................................................................................ 111
12.5.8Spare parts ........................................................................................................................... 111
12.5.9Exploded view of the keyboard ............................................................................................... 112
12.6 Datex-Ohmeda Display Controller Board, B-CDISP............................................................................. 113
12.6.1Technical Specifications ........................................................................................................ 113
12.6.2Functional description ........................................................................................................... 113
12.6.3Display connector ................................................................................................................. 115
12.6.4Jumper settings..................................................................................................................... 115
12.6.5Replacing the Display Controller Board, B-CDISP.................................................................... 116
12.6.6Service and repair procedures................................................................................................ 117
12.6.7Functional checkout .............................................................................................................. 117
12.7 Spare Parts...................................................................................................................................... 117
13 Other products 119
13.1 Network Computer, C-NTNET............................................................................................................. 119
13.2 Video displays, D-VIC17 and D-VIC15............................................................................................... 119
13.3 Uninterruptible power supply, UPS-500............................................................................................ 119
13.4 Central printer, P-CENTR................................................................................................................... 119
13.5 Strip-chart recorder, N-REC4 ............................................................................................................ 119
13.6 Hubs ............................................................................................................................................... 119
Document No. 8000633-1
iii
Page 6
Technical Reference Manual

TABLE OF FIGURES

Figure 2-1 Datex-Ohmeda Monitor Network.................................................................................................3
Figure 2-2 Datex-Ohmeda S/5 Central........................................................................................................4
Figure 2-3 Datex-Ohmeda Network (TCP/IP)................................................................................................5
Figure 3-1 Horizontal cabling ....................................................................................................................14
Figure 3-2 Mon/IC-Net Cable....................................................................................................................15
Figure 3-3 Mon-Net cable.........................................................................................................................15
Figure 3-4 Wallbox ...................................................................................................................................16
Figure 3-5 Horizontal cables .....................................................................................................................17
Figure 3-6 Horizontal patch panel .............................................................................................................23
Figure 3-7 Patch cable .............................................................................................................................24
Figure 3-8 Hub.........................................................................................................................................25
Figure 3-9 Vertical cabling ........................................................................................................................27
Figure 3-10 Vertical cables .........................................................................................................................28
Figure 3-11 Cabling example: A simple Datex-Ohmeda Monitor Network.......................................................29
Figure 3-12 Cabling example: A Simple Datex-Ohmeda Network (TCP/IP).....................................................30
Figure 3-13 Cabling example: A more complex Datex-Ohmeda Network (TCP/IP)..........................................31
Figure 3-14 PentaScanner and Super Injector..............................................................................................32
Figure 4-1 Main components of the S/5 Central and ViewStation...............................................................33
Figure 4-2 Typical Connectors on the Network Computer Rear Panel ...........................................................34
Figure 12-1 General architecture of S-CNETxx and S-VNETxx software ...........................................................89
Figure 12-2 Central Keyboard, K-CENTRALB, block diagram...................................................................... 103
Figure 12-3 Keyboard, K-CENTRALB, exploded view.................................................................................. 112
Figure 12-4 Display Controller Board, B-CDISP, general block diagram...................................................... 114
Figure 12-5 Display Controller, B-CDISP, jumper settings.......................................................................... 115

TABLE OF TABLES

Table 3-1 Separations for horizontal cables..............................................................................................18
Table 3-2 Typical fasteners and spacing intervals .....................................................................................19
Table 10-1 General troubleshooting...........................................................................................................82
Table 10-2 Central/ViewStation Troubleshooting........................................................................................82
Table 10-3 Screen messages.....................................................................................................................84
Table 12-1 Keyboard, K-CENTRALB, spare parts ...................................................................................... 111
iv
Document No. 8000633-1
Page 7

Part I

1 INTRODUCTION

1.1 Symbols

Attention. Consult accompanying documents. In this device the symbol means that the device must be supplied from an appropriate separating transformer when placed in the patient environment.
Fuse. Replace only with a fuse with same type and ratings. Equipotential terminal.
A LED marked with this symbol indicates recorder error.
Datex-Ohmeda S/5 Central, ViewStation and Network

1.2 Abbreviations

AM Datex-Ohmeda S/5 Anesthesia Monitor ARK Datex-Ohmeda S/5 Anesthesia Record Keeper ArrWS Datex-Ohmeda S/5 Arrhythmia Workstation CCM Datex-Ohmeda S/5 Critical Care Monitor CM Datex-Ohmeda S/5 Compact Anesthesia Monitor CMC Datex-Ohmeda S/5 Compact Critical Care Monitor CRC Cyclic redundancy check EMI Electromagnetic interference FTP Foiled twisted pair HIS Hospital information system ID Identification IM Information management RF Radio frequency TP Twisted pair UTP Unshielded twisted pair
Interface device connector.
Page scrolling symbol. The Central and ViewStations are automatically displaying the next page at configurable intervals.
Alarms silenced from the bedside monitor.
Submenu: there is another menu available under this selection
Document No. 8000633-1
1
Page 8
Technical Reference Manual
1.2.1 Conventions used in this manual
To help you find and interpret information easily, the manual uses consistent text formats for certain text types:
Command buttons are written in the following way:
Cancel.
Menu items are written in bold italic typeface, for example Screen 1 Setup. Menu access is described from top to bottom. For example, the selection of the Locations
item from the Screen 1 Setup submenu of the Setup menu, would be shown as Setup - Screen 1 setup - Locations.
File names, file paths and commands are written in Courier New typeface, for example
comm.exe.
Messages (alarm messages, informative messages) displayed on the screen are written
inside single quotes, for example ‘Printing...’
When referring to different sections in this manual, section names are written in italic
typeface and enclosed in double quotes, for example “Cleaning and Care.”

1.3 Safety precautions

WARNING A WARNING indicates that there is a possibility of injury to yourself or others.
Do not locate the S/5 Central, ViewStation, hubs or printers in a medically used room, they
are non-medical electrical equipment. If locating this type of equipment in this type of room is necessary, use an appropriate separating transformer.
Do not connect any other external equipment to the network than those specified by the
manufacturer.
Switch the Network Computer off before connecting or disconnecting any cables on the rear
panel.
Check the position of the voltage selector switch on the rear of the Network Computer before
connecting the power cable to the wall socket.
Connect the devices to a three-wire, grounded, hospital grade receptacle.
Replace the power cable if it is cracked, broken or otherwise damaged. The cable and the
plug must always be intact and undamaged.
Do not apply tension to the power cable.
Do not remove the grounding prong from the power plug.
Do not use extension cables or adapters of any type.
Only qualified personnel (distributor, cabling company or technical hospital staff) may carry
out the installation and service.
CAUTION A CAUTION indicates a condition that may lead to equipment damage or
malfunction.
Do not obstruct the ventilation holes of the Network Computer, displays, nor the hub.
Prevent accidental disconnection of all cables connected to the S/5 Central.
Do not immerse any liquid or allow liquid to enter the Network Computer, the device interior.
Wear a static control wrist strap when handling circuit boards and always hold the boards by
their non-conductive edges.
2
Document No. 8000633-1
Page 9

2 SYSTEM DESCRIPTION

Datex-Ohmeda Monitor Network is an Ethernet based clinical network which is controlled by the S/5 Central. The S/5 Central provides
central monitoring and remote viewing of patient sites.
monitor-to-monitor communication.
storage for patient data files and preoperative data files, record keeping menu files, and
user modes for the monitors.
printing of patient data
recording of waveform snapshots.
services for sharing printers.
communication with up to three other Centrals within the Datex-Ohmeda Network (TCP/IP)
access to other hospital data networks.
Datex-Ohmeda S/5 Central, ViewStation and Network
time services.

2.1 Datex-Ohmeda Monitor Network

The Datex-Ohmeda monitor network may consist of up to 32 bedside monitors connected via a hub to the Central.
Figure 2-1 illustrates a typical configuration of the monitor network.
1
2
Figure 2-1 Datex-Ohmeda Monitor Network
(1) Bedside monitors connected to a hub (2) S/5 Central The hub (10Base T, 10 Mbps) is a multiport repeater and controls the information flow from all the
monitors to the Central.
Document No. 8000633-1
3
Page 10
Technical Reference Manual
2.1.1 Datex-Ohmeda S/5 Central
1
7
6
2
5
Figure 2-2 Datex-Ohmeda S/5 Central
(1) Patient Screen (LCD screen) (2) Network computer + two network boards (3) Patient screen (CRT screen) (4) Status Screen (5) Smart key (6) PC keyboard and mouse (7) S/5 Central keyboard
3
4
Patient Screen displays patient information such as waveforms, numeric fields, alarms, etc., and allows access to various system level operations. Patient Screen has built-in loudspeakers. Two Patient Screens can be connected to a S/5 Central.
Status screen displays system and network information, patient information, alarms and service pages in text. Status Screen is also used during configuration and when starting and exiting programs.
The system can be operated either by a PC keyboard and mouse or a specific S/5 Central keyboard, K-CENTRALB. Only one of the keyboards can be installed at a time. The PC keyboard is handy during installation and administrational tasks; the S/5 Central keyboard is meant for daily use of the S/5 Central in patient environment.
Additionally, there can be a strip-chart recorder, N-REC4, or a laser printer connected to the S/5 Central.
2.1.2 Monitors
Which Monitors?
The, monitors which can be connected to the Datex-Ohmeda Monitor Network are:
Datex-Ohmeda S/5 Anesthesia Monitor, S/5 Compact Anesthesia Monitor, S/5 Critical
Care Monitor and S/5 Compact Critical Care Monitor. These monitors require a network board, e.g. B-NET or B-UPINET to enable the connection to the network. NOTE: There are some limitations in data loading capabilities in monitors equipped with S-ANEXX software.
4
Document No. 8000633-1
Page 11
Datex-Ohmeda S/5 Central, ViewStation and Network
Datex-Ohmeda Cardiocap/5 with the options N-XNET and N-XDNET
Datex-Ohmeda S/5 Light Monitor (models F-LM1 and F-LMP1 with options N-LNET and L-
LDNET). NOTE: There are some limitations in the monitor-to-monitor communication and data loading capabilities in Light Monitor. Please see the Light Monitor’s manuals.
How many monitors?
The number of monitors, (4, 8, 12, …32) connected to the Datex-Ohmeda Monitor Network is set through the smart key. The smart key is a special plug inserted into the serial port of the S/5 Central. The maximum configuration is 32 monitors.
2.1.3 Datex-Ohmeda S/5 Arrhythmia Workstation
You can connect the Datex-Ohmeda S/5 Arrhythmia Workstation, to the Datex-Ohmeda Monitor Network. You can connect one monitor to one Arrhythmia Workstation only and 16 monitors to one workstation at the most. You can connect two Arrhythmia Workstations to a network with 32 monitors.
The Arrhythmia Workstation cannot use the same printer as Central; it requires its own.

2.2 Datex-Ohmeda Network (TCP/IP)

You can connect up to four Datex-Ohmeda Monitor Networks (4 x 32 monitors) together to form the Datex-Ohmeda Network (TCP/IP). The network uses the TCP/IP protocol, and the monitor networks are connected to it via a 100Mbps hub. It is possible to connect up to four S/5 ViewStations to the Datex-Ohmeda Network (TCP/IP).
1
2
3
Figure 2-3 Datex-Ohmeda Network (TCP/IP)
(1) Monitor network with S/5 Central (2) Datex-Ohmeda Monitor Network (TCP/IP) with S/5 ViewStation (3) Monitor network with S/5 Central and a recorder You can connect a laser printer to the S/5 Central or S/5 ViewStation or directly to the 100 Mbps
hub.
Document No. 8000633-1
5
Page 12
Technical Reference Manual
2.2.1 ViewStation
Datex-Ohmeda S/5 ViewStation consists of the same components as the Central, and functions similarly except:
It runs by the software S-VNETXX.
There are no monitors connected to it.
It cannot store patient data.
It has no smart key (N-ICXX).
It has only one network board (TCP/IP).
The ViewStation cannot work without a Central, but the Central works without a ViewStation.
The ViewStation is used for centralized monitoring and remote viewing of patient sites, for example in the office. A laser printer or recorder can be connected to ViewStation as well.
From a Central you can see only the monitors configured to it but from the ViewStation you can see any of the monitors connected to the Datex-Ohmeda Network (TCP/IP), up to four monitor networks. However, you can see a maximum of 32 monitors at a time.
ViewStations are connected to the 100 Mbps hub. NOTE: To enable connecting ViewStations to the system, the Central requires the software S-
CNET99 Ver 1.2, or newer.
6
Document No. 8000633-1
Page 13
Datex-Ohmeda S/5 Central, ViewStation and Network

2.3 Network services

The following services are provided by the S/5 Central.
2.3.1 Central monitoring and remote viewing of patient sites
Waveforms, numeric fields, alarms, and other patient information from the networked monitors are displayed at the Central on the Patient Screen. Real time waveform fields and numeric fields are transmitted to the Central after start up and the numeric fields are updated every 10 seconds. The Patient View is a copy of a single bedside monitor's display with a few exceptions, and the Multi View displays one or two vital waveforms from 8 or 16 beds simultaneously.
Exceptions in Patient View are: — Split screens are not displayed (minitrends or spirometry loops). — NMT or EEG data are not displayed. — In tonometry measurement, the PgCO — The spirometry loops are not displayed in the digit fields. — The text messages are not displayed — No other parameter messages than ‘Apnea’ and arrhythmia messages are displayed in the
corresponding waveform fields, and ‘Audio Off’ and ‘Alarms silenced’ alarm messages are displayed in the waveform fields.
Alarms are displayed both on the Patient Screen and on the Status Screen. Messages are transmitted to the Central as the start-up sequence is complete, and then the messages are transmitted whenever the alarm priority is altered. The priority change at the monitor is displayed at the Central within two seconds.
2.3.2 Monitor-to-monitor communication
Monitor-to-monitor communication is viewing other monitors at a monitor connected to the network using either view links or alarm links. Monitor-to-monitor communication is possible both within the Datex-Ohmeda Monitor Network and the Datex-Ohmeda Network (TCP/IP). In the TCP/IP network, the monitors must have software level S-XX99 or newer to communicate between monitors.
The number of simultaneous view links is limited to 16 simultaneously. Alarms, real time waveform fields, numeric fields and up to 60 minutes of trend data are transmitted. The numerical fields are updated every 10 seconds and the trend data once a minute.
2 digit field is displayed only.
Number of alarm links is not limited. Alarm messages are transmitted from one monitor to another via the Central as the alarm is initialized. Afterward the messages are transmitted whenever the alarm priority has changed.
NOTE: The Datex-Ohmeda S/5 Light Monitor can be connected to the network and viewed from other monitors, but other monitors cannot be viewed at the Light Monitor.
Document No. 8000633-1
7
Page 14
Technical Reference Manual
2.3.3 Storage for the patient data files and preoperative data files
You can collect patient data on the hard disk of the Central throughout the case, also during patient transport, or if the patient is moved to another monitor (data continuum). When the monitor is not connected to the network, you can save the data on the data card.
Patient data from the networked monitors is stored on the hard disk of the Central. The directory where the patient data is saved is generated when the patient is admitted or the case is started. The start time is stored in the status file, and the patient name or identification number is stored in the identification file. Up to 48 hours of trend data is stored in the trend files, and the record keeper events are stored in the event file. The end time of the case is added to the status file when the case is closed or the patient discharged. You can re-open the case within a configured time, max. 24 hours (if there is preoperative data only - no trend data - the time can be extended up to 96 hours).
NOTE: You can load data from the network to the monitors with the software S-ARKXX and S-ICUXX only and to Cardiocap/5 monitor, not to a monitors with the software S-ANEXX, or Light Monitor. Trend data from the Light Monitor is not collected to the network, but on a card only.
2.3.4 Storage for the record keeping menu files
The record keeping menu files are stored on the hard disk of the Central during network installation. You can load the menu pages to the monitors with the software S-ARKXX.
2.3.5 Storage for the user modes
You can use the predefined user modes with the monitors AM, CM, CCM and CMC that define the monitor setup. You can send the user modes from a monitor to the Central’s hard disk, and load them to other monitors.
By default, all the monitors with same software type and revision use same user modes. If part of the monitors needs specific user modes, make a group of the desired monitors, see section 1.
2.3.6 Printing and recording trends and snapshots
Either a laser printer or a strip-chart recorder, N-REC4 can be connected to the S/5 Central. On the laser printer the S/5 Central can print trends, waveform snapshots, anesthesia records and reports as defined in the networked monitors. For printer and recorder setup see section “Installing a recorder” andInstalling a printer.”
Also a network printer can be used, connected directly to the 100 Mbps hub. In that case the printer must be equipped with a network board (default in the printers sold by Datex-Ohmeda.)
Each monitor can have access to 9 printers at the maximum. You can start the central printing either from a monitor or the Central. Start the recorder manually
from the recorder or if it is connected to a Central it can be started automatically by an alarm. Automatic (alarm triggered) printing is not available at the ViewStation.
8
Document No. 8000633-1
Page 15
2.3.7 Access to other hospital data networks
You can convert information from other hospital networks, like admission data and laboratory results to record keeping events and send them to the monitors, and vice versa. However, you cannot connect the Datex-Ohmeda Monitor Network or Network (TCP/IP) to the hospital information system (HIS) directly, but via a Datex-Ohmeda IM server connected to the 100Mbps hub. For transferring information from the system to another, refer to the S/5 Anesthesia System Interface and Inlink manual, and to a guide for the specific datagate communication client.
For the manuals and tools required, contact your sales representative.

2.4 Time services

The Central provides time services for all connected monitors. Within the Datex-Ohmeda Network (TCP/IP) use preferably the primary Central or a Datex-Ohmeda IM Server as a time server for other Centrals. See 5.2.3 “System Setup” for details.

2.5 Communication in the network

The monitor registers to the monitor network by transmitting a registration message. The Central completes the registration sequence by transmitting a registration response.
Datex-Ohmeda S/5 Central, ViewStation and Network
The monitor transmits data packets to the Central with a maximum interval of 10 seconds. Similarly, the Central transmits data packets to the monitor with a maximum interval of 10 seconds. Even though there is no need to transmit data, dummy messages are transmitted to maintain the communication.
In case of an interruption in the communication, the monitor attempts to register to the Central every 10 seconds.
The data packet formula is that of standard Ethernet, but the contents are Datex-Ohmeda specific.

2.5.1 Data integrity

Data integrity means preventing or correcting possible communication errors. The error prevention approach in the network is cyclical redundancy check (CRC). In CRC, the transmitting device calculates a checksum for each data packet. The checksum is recalculated and compared at the receiving device, and if no errors are detected, an acknowledgment message can be sent back. If no acknowledgment message is received, the originating device retransmits the missing or corrupted packet.
Document No. 8000633-1
9
Page 16
Page 17

3 CABLING

3.1 General

3.1.1 Cabling standards
The international standards in network cabling are ISO/IEC 8802-3 (ANSI/IEEE 802.3), EIA/TIA­568 (Category 5) and EIA/TIA-TSB40. These standards define cable types, cable lengths, connector types, etc.
3.1.2 General restrictions
Configure the Network within the following general restrictions.
Up to 32 monitors can be connected to one monitor network.*
Only the devices specified by the manufacturer may be connected to the network.*
Arrhythmia Workstations can be connected to the network. A monitor can be connected to
one Arrhythmia Workstation only.
Up to four Centrals can be connected together through Datex-Ohmeda Network (TCP/IP)
with TCP/IP protocol.
Up to four ViewStations can be connected to the Datex-Ohmeda network.
Datex-Ohmeda S/5 Central, ViewStation and Network
* Refer to section System Description about permitted devices and their number.
3.1.3 Recommended installation tools
Ground wrist strap
Hand drill
Screwdrivers
525421-8 AMP stripper
558418-1 AMP connection tool/installation tool
HC-3013-A-T connection tool/installation tool
Cable scanner (reflectometer)
VX1301 stripper and the G30040 BNC pliers are needed if you will install coaxial cabling.

3.2 Pre-installation testing

The pre-installation testing includes a site survey and a site preparation.
3.2.1 Site survey
Before committing to a cabling system design, you must explore several important issues.
11
Document No. 8000633-1
Page 18
Technical Reference Manual
Architectural limitations
Wide distribution of equipment
High reliability access requirements (thick walls, etc.)
Inter building connections
New construction
Presents "ideal" design opportunities
Significantly lower labor costs for installation
Existing construction
Utility of current communications facilities
Flexibility and expandability of current facilities
Extent of alteration and additional requirements
Cost
Ensuring safety
Cost
Reliability
Expandability
Security
Observing building and electrical codes
Ensure availability and quality of electrical ground
Installed cost per meter (foot)
Installed cost per connection/node
Special considerations (problematic installations)
Armored cable for rodent protection
Hostile and friendly electrical environments
Physical cable protection
Cabling facilities are always under-designed and over-utilized.
Space in wiring closets and cable routs (c. 50 per cent)
Physical cable plant and facilities
Access to network
Protection of data
12
Document No. 8000633-1
Page 19
Site preparation
Consider the following for single connection changes as well as for large-scale initial installation:
Determining unusual installation problems
Determining the space in conduit, troughs or plenum
Evaluation of path alternatives
Specifying equipment locations
Understanding aesthetic requirements
General cabling standard compliance
Recommended participants in the site preparation:
Hospital system manager
Distributor’s site manager
Network planner or subcontractor
Construction or electrical foreman
Chief physician
Head nurse
The distributor’s site manager and the network planner or subcontractor are responsible for future and general cabling standard compliance, network constraints and specifications. The hospital system manager and the construction or electrical foreman are responsible for installation methods, structural concerns and general cabling standard compliance. The chief physician and the head nurse are responsible for the location of devices, access to information, aesthetics, flexibility (ability to accommodate moves and changes), and impact (disruption) during installation.
Datex-Ohmeda S/5 Central, ViewStation and Network
3.2.2 Site plan
Using a floor plan, carefully label where the Centrals, ViewStations, monitors and the hub will reside. Locate the Centrals and ViewStations in an office or central desk, the monitors in patient rooms, and the hub in a centrally located wiring closet or equipment room. Locate all devices within an appropriate distance of a power line source.
WARNING The Centrals, ViewStations, hubs, printers and recorders are non-medical
electrical equipment. Do not locate this type of equipment in a medically used room. If locating this type of equipment in this type of room is necessary, use an appropriate transformer.
Sketch out a map illustrating the horizontal cable from the Central and each monitor to the hub (star topology). If possible, use current communication facilities. Otherwise, use existing leads­through, risers and conduits.
Take into account electrical equipment that may generate high levels of EMI like motors and transformers. Also, avoid power wiring, water pipes, heating ducts, etc., that will affect the cabling system performance or the installation. Keep in mind that sensible placement of the cable allows future expansions and connectivity.
Make an approximation of each segment cable length. According to IEEE 802.3 specifications a twisted pair (TP) segment length may not exceed 100 m (328 ft.). The cable scanner, applied after the cable installation, determines the exact length of the segments. To facilitate your documentation, an example of a site plan is included in the appendices.
13
Document No. 8000633-1
Page 20
Technical Reference Manual
3.2.3 Site planning sheet
Using a site plan, transfer each segment cable length to the site planning sheet. Write down in which room you will locate the Central and the monitors, and to which hub port you will connect them. Further, add the type of the monitor and the location ID. Use a separate site planning sheet for each floor the network will extend to. To facilitate your documentation, an example of a site planning sheet is included in the appendices.
3.2.4 Pre-installation test binder
It is recommended to collect all documentation that you receive from the pre-installation testing in a pre-installation test binder. Include information about the hospital, distributor and subcontractor in the binder.

3.3 Horizontal cabling

Use horizontal cabling in cabling bounding to one floor as illustrated in Figure 3-1.
7
8
6
5
4 3 2
1
2
11
9
8
10
Figure 3-1 Horizontal cabling
(1) ViewStation (2) Mon/IC-Net Cable (3) Wallbox (4) Horizontal cable (5) Horizontal Patch Panel (6) Patch Cable (7) 100 Mbps hub (8) Shelf (9) 10 Mbps hub (10) Mon-Net Cable (11) S/5 Central
14
Document No. 8000633-1
Page 21
3.3.1 Mon/IC-Net cables
Mon/IC-Net cables are work area cables and carry signals between Centrals or ViewStations and wallboxes both in the monitor network and in the TCP/IP network. The cable type is category 5 unshielded twisted pair, designed to be thin and flexible. The cable is terminated with RJ-45 connectors at both ends, and the coupling method is straight through. Figure 3-2 illustrates a Mon/IC-Net Cable.
Figure 3-2 Mon/IC-Net Cable
Datex-Ohmeda S/5 Central, ViewStation and Network
Connecting the Mon/IC-Net cable to Central/ViewStation
1. For the monitor network: connect one cable end to the network connector marked MON on
the rear panel of the Central. For the Datex-Ohmeda Network: connect one cable end to the network connector marked TCP/IP on the rear panel of the Central/ViewStation.
2. Connect the other end to the connector on the wallbox.
3.3.2 Mon-Net cables
Mon-Net cables are work area cables and carry signals between monitors and wallboxes. The cable type is category 5 unshielded twisted pair, designed to be thin and flexible. The cable is terminated with RJ-45 connectors at both ends, and the coupling method is straight through.
The Mon-Net cable includes an identification plug. The plug contains an EEPROM chip with a unique value to identify the location of the monitor. This location ID is necessary for the network setup. Figure 3-3 illustrates a Mon-Net Cable.
Figure 3-3 Mon-Net cable
15
Document No. 8000633-1
Page 22
Technical Reference Manual
Connecting the Mon-Net cable to monitors
1. Connect the connector with the identification plug, to the connector on the rear panel of the
monitor.
2. Connect the identification plug to the connector on the rear panel of the monitor (in the
Network Board, B-NET).
3. Connect the other end to the connector on the wallbox.
3.3.3 Wallboxes
Wallboxes transfer signals from Mon-Net cables and Mon/IC-Net cables to horizontal cables. The wallbox consists of one slot suitable for an RJ-45 connector. Figure 3-4 illustrates a wallbox.
Figure 3-4 Wallbox
Labeling the wallbox
Clearly label all wallboxes on both the attachment plate and the center plate. The labeling method must be flexible to adapt to changes in the wiring. If a labeling method already exists within the hospital, use it.
Installing the wallbox
1. Drill two holes in the wall at a height of 1.2-1.3 m (3.9-4.3 ft.) for the surface mount box
N
E
T
fastening screws. Pull the horizontal cable through the box. Fasten the box to the wall with two fastening screws. NOTE: If you use a wall device outlet box, you do not need the surface mount box.
16
Document No. 8000633-1
Page 23
Datex-Ohmeda S/5 Central, ViewStation and Network
2. Finish the horizontal cable and attach the RJ-45 connector to the attachment plate. Fasten
the plate to the surface mount box with two fastening screws.
3. Place the cover ring and the adaptation ring upon the attachment plate. Pierce the center
plate and fasten the plate to the attachment plate with two fastening screws.
3.3.4 Horizontal cables
Horizontal cables carry signals between wallboxes and horizontal patch panels. The term "horizontal" is used because the cable runs horizontally along the floor(s) or ceiling(s).
The cable type is category 5 twisted pair, either unshielded or foiled. The unshielded twisted pair (UTP) is thin and flexible, while the foiled twisted pair (FTP) cable is thick and fixed. The cable jacket can optionally be standard PVC or plenum. Standard PVC cable is flame resistant to 75 °C while plenum cable is fire retardant up to temperatures of 200 °C. Use plenum cable instead of standard PVC in areas where a high degree of fire resistance is required such as heating, ventilation and air conditioning (HVAC) ducts. Technical data sheets for the horizontal cables are included in the appendices. Figure 3-5 illustrates horizontal cables.
Figure 3-5 Horizontal cables
FTP
UTP
17
Document No. 8000633-1
Page 24
Technical Reference Manual
Labeling the horizontal cables
Clearly label all horizontal cables on both the local and remote end. The labeling method must be flexible to adapt to changes in the wiring. If a labeling method already exists within the hospital, use it.
Routing the horizontal cables
Route the cables where they will be concealed and protected from possible damage.
Do not route the cables in a common cabling enclosure. However, you may ground them
together in a common cabling enclosure.
Route the cables at an adequate distance from AC power lines, air conditioning systems
and diathermy units, etc., to minimize RF and AC interference. Never route long runs of cables parallel to AC power lines. Table 3-1 illustrates overall separations for horizontal cables.
Avoid routing the cables outside the building especially when prewiring. If you wish to route
cables in this fashion, slope holes bored through exterior walls upward from the outside. Keep the size of the holes as small as possible and seal the exterior surface after you have installed the cables.
Route the cables without any sharp angles. The bend radius of a UTP type cable should not
be less than 8.3 cm (3.3 in.) while the bend radius for a FTP type cable should not be less than 9.3 cm (3.7 in.).
Route the cables and install the mounting hardware (roughing-in) after installation of the
electrical power system, but before thermal insulation and wallboards are installed when performing prewiring.
Table 3-1 Separations for horizontal cables
Voltage source Type of wire Separation Alternative
Electric Supply Bare wire (any voltage), or open wire
over 300 V Electric Supply Open wire not over 300 V 5 cm (2 in) (*) Electric Supply Wire in separate conduit, in armored or
non-metallic shield or ground wire Radio and television Antenna lead-in or ground wire 10 cm (4 in) (*) Signal or control All types None N/A CATV Coaxial cables with grounded shielding None N/A Telecommunication Drop wire, all types 5 cm (2 in) (*) Neon signs Wire from transformer 15 cm (6 in) None Lighting systems Lighting rods and wires 1.8 m (6 ft) (**)
1.5 m (5 ft) None
None N/A
(*) When you cannot obtain minimum separations, provide additional protection by means of plastic tube,
wire guard, or two layers of vinyl tape extending 5 cm (2 in.) beyond each side of the wiring being crossed.
(**) If you cannot obtain separations of 1.8 m (6 ft.), separations of least 10 cm (4 in.), are permissible if:
you make all telecommunication, power, and lighting rod ground connections to a common
you use separately driven ground rods for telecommunications, power, and lighting and
18
Document No. 8000633-1
metallic cold water pipe that is properly grounded, or
bond the ground rods together.
Page 25
Pulling the horizontal cables
Pull the cables using 2.5 m (8.2 ft.) long pipes and pull several cables simultaneously.
Avoid damaging the cables when pulling them. Pulling tension on a TP cable may not
exceed 11.3 kg (25 lb.) to prevent stretching the conductors.
Be particularly careful when pulling the cables through joists and studs (especially metal
studs) to prevent damage to the cable shield.
Install a "pull-string" with the cable runs (e.g., inside wall cavities between floors) to
facilitate pulling of future cable runs.
After the cable is pulled, cut off 1 m (3.3 in.) of the cable end to eliminate possible damage.
Fastening the horizontal cables
Fasten the cable securely without adversely affecting the insulation.
Use wire clamps, wire staples, bridle rings and drive rings as fasteners when fastening and
also when changing direction of the cable to hold the cable without any sharp angles. Place fasteners according to spacing intervals in Table 3-2.
Table 3-2 Typical fasteners and spacing intervals
Datex-Ohmeda S/5 Central, ViewStation and Network
Fastener Horizontal spacing Vertical spacing Spacing from corner
Wire clamp 40 cm (16 in) 40 cm (16 in) 5 cm (2 in) Wire staples 18 cm (7 in) 18 cm (7 in) 5 cm (2 in) Bridle rings 1.2 m (4 ft) 2.5 m (8 ft) 5 to 22 cm
Drive rings 1.2 m (4 ft) 2.5 m (8 ft)
Finishing UTP cables
Finishing UTP cables for wallboxes and patch panels.
1. Strip off approximately 5.0 cm (2 in.) of insulation from the cable end.
(2 to 8.5 in) 5 to 22 cm
(2 to 8.5 in)
19
Document No. 8000633-1
Page 26
Technical Reference Manual
2. Fold the stripped portion upwards and place the cable between the connector blocks.
3. Separate the data wires. Slide orange pair and brown pair, then blue pair and green pair
through opposite openings above the connector blocks. Leave the pairs twisted. The maximal untwisting must be less than 13 mm (0.5 in.).
4. Pass the wire pairs along the outer surface of the connector blocks. Lay the wires over the
connector pins according to the color code and press the wires into respective pins. Cut off excess wire and remove the fragments.
5 6 7 8
1
2
3 4
6
2345
Block
1 2 3 4 5 6 7 8
WHT/BLU
BLU
WHT/ORG
ORG
WHT/GRN
GRN
WHT/BRN
BRN
1
Jack
Terminal
5 4 3 6 1 2 7 8
8
7
5. Push back the excess length of cable into the wall cavity.
20
Document No. 8000633-1
Page 27
Finishing the FTP cables
Finishing FTP cables for wallboxes:
1. Strip off approximately 8 cm (3.1 in.) of insulation and cable shield from the cable end and
cut off the drain wire.
2. Finish the FTP Cables for Wallboxes as described above about UTP cables starting from the
step 2.
Finishing the FTP cable for Patch Panels
1. Strip off approximately 8.0 cm (3.1 in.) of insulation and shield from the cable end.
Separate the drain wire and isolate the wire with tape.
Datex-Ohmeda S/5 Central, ViewStation and Network
2. Fold the stripped portion upwards and place the cable between the connector blocks.
3. Separate the data wires. Slide orange pair and brown pair, then blue pair and green pair
through opposite openings above the connector blocks. Leave the pairs twisted. The maximal untwisting must be less than 13 mm (0.5 in.).
21
Document No. 8000633-1
Page 28
Technical Reference Manual
4. Pass the data wires along the outer surface of the connector blocks. Lay the wires over the
connector pins according to the color code and press the wires into respective pins. Cut off excess wire and remove the fragments.
5
6
7 8
1 2 3
4
1
2
4567
3
8
Block
1 2 3 4 5 6 7 8
WHT/BLU
BLU
WHT/ORG
ORG
WHT/GRN
GRN
WHT/BRN
BRN
Jack
Terminal
5 4 3 6 1 2 7 8
5. Unscrew the GND connector screw on the patch panel. Wrap the drain wire(s) around the
screw in the direction the screw tightens and tighten the screws. The cable shield must make electrical contact with an earth reference at one point only.
6. Push back the excess length of cable into the wall cavity.
22
Document No. 8000633-1
Page 29
3.3.5 Horizontal patch panels
Horizontal patch panels are used with hubs to transfer signals from horizontal cables to patch cables. The panel constitutes of 16 slots suitable for RJ-45 connectors. Figure 3-6 illustrates a horizontal patch panel.
Figure 3-6 Horizontal patch panel
Datex-Ohmeda S/5 Central, ViewStation and Network
Labeling the patch panel
Clearly label all horizontal patch panel slots. The labeling method must be flexible to adapt to changes in the wiring. If a labeling method already exists within the hospital, use it.
Installing the patch panel
The horizontal patch panel is equipped for direct mounting to 19" rack mounting hardware. Fasten the panel to the rack according to the following procedures.
1. Line up the holes on the edges of the panel with the holes on the rack. Attach the panel to
the rack with four fastening screws. Always locate the panel beneath the hub to leave hub signal lights visible.
2. Finish the horizontal cables and complete the panel by firmly pressing the RJ-45 connectors
into reserved slots of the panel.
23
Document No. 8000633-1
Page 30
Technical Reference Manual
3.3.6 Patch cables
Patch cables are equipment cables that carry the signals between horizontal patch panels and hubs. The cable type is category 5 unshielded twisted pair, designed to be thin and flexible. The cable is terminated with RJ-45 connectors at both ends, and the coupling method is straight through. Figure 3-7 illustrates a patch cable.
Figure 3-7 Patch cable
Connecting the patch cable
1. Connect one cable end to the connector on the patch panel.
2. Connect the other end to the connector on the front panel of the hub.
24
Document No. 8000633-1
Page 31
3.3.7 Hubs
Datex-Ohmeda S/5 Central, ViewStation and Network
All patch cables terminate in a hub that is functioning as a multiport repeater. Four different hubs are recommended, and they can also be purchased locally: — HUB-8 and HUB-16 are 10 Mbps hubs for the monitor network — HUB-8FAST and HUB-16FAST are 100Mbps hubs for Datex-Ohmeda Network. Both hub types can be connected with category 5 cabling. NOTE: if you purchase locally a hub with selectable speed 10/100 Mbps, make sure you set the
speed to 10 Mbps when installed in the monitor network and to 100 Mbps in the TCP/IP network. NOTE: The Datex-Ohmeda monitors use 10 Mbps network boards, and 10 Mbps hubs should be
used in the monitor networks. If a 10/100 Mbps hub with automatic speed detection is used instead, the S/5 Central Monitor Network Board had to be configured to use 10 Mbps exclusively. Use the 3COM NIC Diagnostic program for making the configuration. The program is available at the 3COM EtherCD CD-ROM that has been supplied to you with the Central computer.
Figure 3-8 Hub
Connecting Hubs together
To enlarge the network, connect hubs together with a cross over cable through the RJ-45 ports or with a terminated ThinNet coaxial cable through the BNC ports.
NOTE: The total number of the monitors connected to one monitor network (one Central) must not exceed 32.
25
Document No. 8000633-1
Page 32
Technical Reference Manual
Installing the hub and the self
To install the hub to a 19" rack, a shelf is required.
1. Line up the holes on the edges of the shelf with the holes on the rack. Attach the shelf to the
rack with the fastening screws, included in the accessory kit.
2. Place the hub on the shelf.
3. Connect the power cable to the power socket on the rear panel of the hub and to the wall
socket.
26
Document No. 8000633-1
Page 33

3.4 Vertical cabling

Use vertical cabling to connect different floors to each other as illustrated in Figure 3-9.
Datex-Ohmeda S/5 Central, ViewStation and Network
1
3
4
5 6 7
8
2
9
1
Figure 3-9 Vertical cabling
(1) 10 Mbps hub (2) Patch cable (3) Shelf (4) Horizontal patch panel (5) Horizontal cable (6) Wallbox (7) Mon/IC-Net cable (8) Mon-Net cable (9) Vertical cable
27
Document No. 8000633-1
Page 34
Technical Reference Manual
Vertical cables
Vertical cables carry signals between patch panels. The term "vertical" is used because the cable runs vertically along the ceiling(s) of a building. The cable type is category 5 foiled twisted pair (FTP) or Belden 9907 coaxial. Figure 3-10 illustrates vertical cables.
Figure 3-10 Vertical cables
NOTE: If the environment is electrically noisy, you can use optical fiber as a vertical cable. For further information, consult your local distributor.
FTP
Coaxial
Finishing coaxial cables
1. Strip off approximately 1.5 cm (0.6 in.) from the cable end and slide the sleeve of the BNC
connector over the cable and past the point of stripping.
2. Peel back the braided portion so that none of the strands are near the center conductor.
Slide the gold pin over the center conductor. Make sure that all the wires of the center conductor are seated inside the gold pin. Crimp the gold pin.
28
Document No. 8000633-1
Page 35
Datex-Ohmeda S/5 Central, ViewStation and Network
3. Slide the BNC ferrule over the gold pin till it rests firmly against the rolled back braided
portion. No part of the braided shield must pass inside the ferrule.
4. Slide the sleeve up over the ferrule of the BNC connector. All of the braided shield should be
in between the ferrule and the sleeve. Crimp down the sleeve.

3.5 Cabling examples

Central
Recorder
Office
Figure 3-11 Cabling example: A simple Datex-Ohmeda Monitor Network.
All monitors and the Central are connected to the same 10 Mbps hub.
Central is situated in the office and equipped with a recorder.
Two monitors are in the operating room.
Two monitors are in the recovery room.
Monitor
OR
Hub (10 Mbps) Patch panel
Monitor
Monitor
Recovery
Wallbox
Monitor
29
Document No. 8000633-1
Page 36
Technical Reference Manual
Hub (100 Mbps) Patch panel
Hub (10 Mbps)
Wallbox
Central
Printer
Office
Monitor
ICU
Monitor
OR
Monitor
Monitor
Hub (10 Mbps)
Monitor Central
Monitor Monitor
Recovery
ArrWs
Recorder
Office
Printer
Figure 3-12 Cabling example: A Simple Datex-Ohmeda Network (TCP/IP)
Two Centrals, one in each department's office and one arrhythmia workstation
Two monitor networks: 1
All monitors and ArrWS are connected to 10 Mbps hub.
Centrals are connected both to the10 Mbps and 100 Mbps hub.
A laser printer is connected directly to the 100 Mbps hub.
30
Document No. 8000633-1
st
one with two monitors in operating room and two in recovery
room, and a printer connected to the Central.
nd
one with three monitors in ICU, and arrhythmia workstation and the central with a
The 2 strip-chart recorder in the office.
Page 37
Datex-Ohmeda S/5 Central, ViewStation and Network
Printer
View
Station
Central
Recorder
Office
Monitor Monitor
OR
Hub (10 Mbps)
Monitor Monitor
Recovery
D-O
IM server
HIS
Hub (100 Mbps) Patch panel
Wallbox
Hub (10 Mbps)
Monitor
ICU
Monitor
Monitor
ArrWs
Central
Recorder
Office
Printer
View
Station
Figure 3-13 Cabling example: A more complex Datex-Ohmeda Network (TCP/IP)
Two monitor networks, both Centrals have their own recorders. All the monitors are
connected to the 10 Mbps hubs and Centrals are connected both to the 10 Mbps and 100 Mbps hubs
Two ViewStations connected to a 100 Mbps hub.
Two laser printers connected to 100 Mbps hub
The Datex-Ohmeda Network (TCP/IP) can be connected to the hospital information system
(HIS) via a Datex-Ohmeda Information management server via the 100 Mbps hub
31
Document No. 8000633-1
Page 38
Technical Reference Manual

3.6 Post-installation testing

The post-installation testing includes a physical cabling control and a functional cabling control.
Physical cabling control
The physical cabling control is necessary to discover any physical faults within the cabling. Perform the control by strictly comparing the original plan (documentation) to the actual cabling.
Functional cabling control
The functional cabling control is necessary to discover any functional faults within the cabling. Perform the control using a cable scanner suitable for testing Category 5 cable systems. The tester should give a comprehensive diagnostic view of the network by measuring the following set of features.
Near- end crosstalk
Attenuation
Length
Attenuation to crosstalk ratio
Impedance
Loop resistance
Capacitance
It should also be possible to save and print the measurement results.
TM
E.g. PentaScanner
and Super InjectorTM by Microtest, see Figure 3-14, fulfill these needs.
Figure 3-14 PentaScanner and Super Injector
For performing the tests, finding faults and their solutions, refer to the scanner’s documentation.
Certification report
Make sure you have a complete record of the cabling tests and the certification that cablings function properly.
Attach a copy of the test report to this folder or to a separate installation folder.
32
Document No. 8000633-1
Page 39
Datex-Ohmeda S/5 Central, ViewStation and Network

4 INSTALLING CENTRALS AND VIEWSTATIONS

4.1 Components

The main components of the Central and ViewStation are illustrated in the following figure:
1
7
Figure 4-1 Main components of the S/5 Central and ViewStation
(1) Patient Screen, D-CFLT15 (2) Network Computer, C-NTNET (3) Patient Screen, D-VIC17 (4) Status Screen, D-VIC15 (5) PC keyboard and a mouse (6) S/5 Central Keyboard, K-CENTRALB Network computer is equipped with two commercial network boards and a Datex-Ohmeda specific
display controller board, B-CDISP. Also the operating system, Windows NT, and the network software (S-CNET99 for Centrals, S-VNET99 for ViewStations) are pre-installed. The Central is equipped with the smart key, N-ICXX, where XX defines the number of the beds which can be connected to the Central.
6
2
5
3
4
Additionally the package should include the User’s Reference Manual, power cable, connection cables, and back-up disks of the operating system and the software
NOTE: Old display board, B-CDHI, cannot be used with the S/5 Central. NOTE: Network Manager configuration does not include Patient Screen, S/5 Central Keyboard nor
B-CDISP board.

4.2 Unpacking

1. Confirm that the packing boxes are undamaged. If any of the boxes is damaged, contact the
2. Open the top of the boxes and carefully unpack all components.
shipper.
33
Document No. 8000633-1
Page 40
Technical Reference Manual
3. Confirm that all components are undamaged. If any of the components is damaged,
contact the shipper.
4. Confirm that all components you have ordered are included. If any of the components is
excluded, contact your local distributor.

4.3 Placing

When placing the Central or ViewStation, consider the following aspects.
Light: bright lights may disturb observing the information displayed.
Space: leave space for ventilation to prevent the device from heating.
Power requirements: use extra separating transformer in medical areas.
Electromagnetic and radio frequency interference.
Variations in atmospheric pressure.
Refer also to the technical specifications delivered with the computer and displays .
WARNING The Central and the hub are non-medical electrical equipment. Do not locate
this type of equipment in a medically used room. If locating this type of equipment in a medically used room is necessary, it shall be supplied from an additional transformer providing at least basic isolation (isolating or separating transformer).

4.4 Connections

NOTE: The rear panel connectors may be slightly different in different PC models.
1
Figure 4-2 Typical Connectors on the Network Computer Rear Panel
2
3
4
5
6
10
9
8
7
(1) Power cord plug (2) Display data cable connector for Status Screen
34
Document No. 8000633-1
Page 41
(3) Datex-Ohmeda Network (TCP/IP) cable connector (4) Monitor Network (Mon) cable connector (not in the ViewStations) (5) Audio cable connector (6) Display data cable connector for Patient Screen(s) (7) Serial port for the Smart Key (not in the ViewStations) (8) Parallel port for printer or recorder cable connector (9) Mouse cable connectors (10) Keyboard cable connector
NOTE: If you have two independent Patient Screens, you need two D-CDISP boards.
Connecting the Central to the line power source
Preferable use an uninterruptible power supply (UPS):
1. Connect the UPS input power cable to the UPS power inlet and to the power outlet.
2. Connect the UPS output power cable to the power inlet on the rear panel of the Network
Computer and to the UPS power outlet.
Datex-Ohmeda S/5 Central, ViewStation and Network
If the devices are placed in the medical room, connect them using an appropriate transformer.
Connecting the keyboards
Connect the keyboard cable to the keyboard connector on the rear panel of the Network
Computer, see Figure 4-2.
It is possible to connect only one keyboard at a time. The PC keyboard is used during installation, and to operate Network Manager, and the S/5 Central keyboard, K-CENTRALB, is used to operate S/5 Central in daily use.
The keyboard is selected in the System Setup. If the keyboard is changed remember change the setting, too.
Connecting the mouse
Connect the mouse cable to the mouse connector on the rear panel of the Network
Computer.
Connecting the loudspeakers
Connect the loudspeakers (part of the Patient Screen) to the audio connector on the rear
panel of the Network Computer.
NOTE: The audio knobs at the lower part of display’s front panel are covered to prevent the accidental silencing of the alarm sounds. Make sure you do not turn the sound off when making other display adjustments.
WARNING Always make sure that audio cables are properly connected. WARNING Never turn the volume off from the alarm loudspeakers.
Document No. 8000633-1
35
Page 42
Technical Reference Manual
Connecting the Patient Screen
1. Connect the display data cable to any display connector on the Display Controller Board, B-
CDISP, on the rear panel of the Network Computer.
2. Connect the display power cable to the wall socket. In medical area use an appropriate
separating transformer.
The B-CDISP board has two connectors. You can connect another Patient Screen to the same board as a slave screen (the same information is displayed on both of the screens). To get two independent Patient Screens you need two B-CDISP boards, and one of them must be configured as primary and the other as secondary display board. See Part II for details of the product.
In the System Setup, you must select whether you use the CRT or LCD display.
Connecting the Status Screen
1. Connect the display data cable to the display connector on the rear panel of the Network
Computer.
2. Connect the display power cable to the display power outlet on the rear panel of the
Network Computer. Status Screen pages and columns are configured in the System Setup. NOTE: Status Screen should be configured for 1024 x 768 pixel resolution with large fonts. Use
the NT’s Display Setup to configure the Status Screen.
Connecting the Smart Key
The Smart Key is pre-installed.
Connect the Smart Key to the serial port on the rear panel of the Network Computer.
Connecting the Central to the Datex-Ohmeda Monitor Network
Connect the Mon/IC-Net Cable to the network connector (indicated “Mon.”) on the rear
panel of the Network Computer, and to 10 Mbps hub. Configure the monitor network with the program. Configure the monitor views using the Datex-Ohmeda View Setup program.
Connecting the Central/ViewStation to the Datex-Ohmeda Network (TCP/IP)
Connect the Central to the Datex-Ohmeda network (TCP/IP) using the network board
indicated “TCP/IP”.
Connect the ViewStations to the 100Mbps hub.
Configure the Datex-Ohmeda Monitor Network (TCP/IP) with Network Setup and views seen from each ViewStation with the View Setup.
Connecting a recorder
1. Connect the recorder cable to the parallel connector on the rear panel of Network
Computer, see Figure 4-2, and to the connector on the recorder’s rear panel.
2. Connect the recorder’s power cable to the power inlet and to the wall socket.
36
Document No. 8000633-1
Page 43
Connecting a printer
Local printer
1. Connect the printer cable to the connector on the rear panel of Network Computer, and to
2. Connect the printer’s power cable to the power inlet and to the wall socket.
Network printer
If the printer is going to be shared by several Centrals and/or ViewStations, it is recommended to connect the printer to Datex-Ohmeda network (TCP/IP). This installation requires that the printer is equipped with a network board. The printers purchased from Datex-Ohmeda have the network board installed. If you use a locally purchased printer, install the network board (HP JetDirect) according to the instructions provided by the printer manufacturer.
Datex-Ohmeda S/5 Central, ViewStation and Network
the connector on the printer’s rear panel. Secure the connections with finger screws.
37
Document No. 8000633-1
Page 44
Page 45

5 CONFIGURING

Before configuring the Datex-Ohmeda S/5 Centrals and ViewStations all cablings must be done and tested, and all the equipment must be installed, see chapters 0 and 0. If any changes in the operation system configuration are required, see chapter 7.

5.1 Setup files and tools

Several configuration files define the configuration of the monitor network: —
NETWORK.AIC defines the network configuration of the system. This file has to be the
same for each Central and ViewStation that are connected to the same Datex-Ohmeda Network (TCP/IP).
MONVIEWS.AIC defines which monitors are shown on the Patient and Status Screens on
each Central and ViewStation. Each Central and ViewStation uses its own
MONVIEWS.AIC file. MONVIEWS.AIC file is edited with the View Setup software.
SETUP.AIC defines the general configuration of each Central and ViewStation. Each
Central and ViewStation uses its own System Setup software.
Datex-Ohmeda S/5 Central, ViewStation and Network
NETWORK.AIC file is edited with the Network Setup software.
SETUP.AIC file. SETUP.AIC file is edited with the
These files are located in the will be copied to
D:\CENTRAL\PROGRAM directory during the system startup.
D:\CENTRAL\SETUP folder of the Central or ViewStation. The files
After each part of the configuration is complete, print a hard copy of the setup by clicking
File and Print.
Save the hard copy in the installation folder.
After the entire configuration is complete, restart the Central/ViewStation, and run the
DOCCheckup software (see page 68).
5.1.1 When to use which setup software?
Network Setup software has to be used every time when:
a new network is installed.
a new Central or ViewStation is added in the Datex-Ohmeda Network (TCP/IP).
a monitor is added in the Datex-Ohmeda Monitor Network.
a monitor is removed from the Datex-Ohmeda Monitor Network.
networks have to be connected to/removed from each other.
For instructions, see chapter 5.2.1. View Setup software has to be used every time when:
a new network is installed, configuring the view setup for the first time
a monitor is added to a view or page seen at the Central or ViewStation.
a monitor is removed from a view or page seen at the Central or ViewStation.
For instructions, see chapter 5.2.2. System Setup software has to be used every time when:
a new network is installed.
39
Document No. 8000633-1
Page 46
Technical Reference Manual
a laser printer or recorder is installed.
alarm settings are changed.
display settings are changed.
password is changed.
monitor-to-monitor communication is enabled/disabled.
patient data path is changed.
menu paths for record keeping functions are changed.
For general instructions, see chapter 5.2.3 and for recorder installation, see chapter 1 and for printer installation, see chapter 7.2.
NOTE: Settings are checked during the system startup only, and become active only after the Central or ViewStation software has been restarted.

5.2 Setup Procedure

Since the Centrals and ViewStations rely on the network setup file to get information on the network configuration of the Datex-Ohmeda Network, it is very important that all Centrals and ViewStations always have access to identical when configuring the network configuration:
NETWORK.AIC files. The following procedure is recommended
1. Define the first Central you install as a Primary Central of the system.
2. Use the Network Setup always on the Primary Central only, and then copy the updated
NETWORK.AIC files in the CENTRAL\SETUP folders of the other Centrals and
ViewStations in the system.
3. After the network setup file has been copied to other Centrals and ViewStations, make
ViewSetup and SystemSetup adjustments for each Central and ViewStation.
4. Run the DOCCheckUp software on each Central and ViewStation (see chapter 9.2).
40
Document No. 8000633-1
Page 47
5.2.1 Network setup
The Network Setup allows you to
define network properties, i.e. the network name and type, and PC name or IP address
add and remove monitors to/from a network
make groups to enable use of different user modes in monitors e.g. in operation room and
recovery
connect monitor networks together
NOTE: Use the Network Setup always on the Primary Central only. Remember to copy the updated
NETWORK.AIC file in other Centrals and ViewStations after the editing.
To start the setup program:
Click Central Setup icon and select Network Setup.
Datex-Ohmeda S/5 Central, ViewStation and Network
Selecting the Primary Central
By default the CENTRAL_1 is the primary Central. To change it:
1. Click the pull down menu beside the window Primary Central.
2. Select the primary Central from the list.
41
Document No. 8000633-1
Page 48
Technical Reference Manual
Adding Centrals or ViewStations to the network
You can add up to four Centrals and ViewStations to the network:
1. Click
2. Give the descriptive name for the network, e.g. the
3. Select the network type, either Central or ViewStation,
4. Give the PC name (CENTRAL_1, etc.), or the IP
The PC name and IP address are defined during Windows NT installation. Make sure you use the same PC names or IP addresses. Preferably use the PC name instead of fixed IP address!
IP port is fixed (5554). NOTE: One of the Centrals in the network should be
defined as Central_1. All Centrals are defined to be Central_1 when they are sent from the factory. The PC names of the Centrals and ViewStations can be changed in Windows NT by clicking Control Panel - Network - Identification -Change.
If you want to edit these pieces of information, double click the desired item on the list.
Add Network.
department name.
by clicking the pull down menu.
address.
To remove a Central or ViewStation:
Click Remove Network.
Adding, removing and editing monitor groups
Grouping the monitors enables you to store and load specific user modes for a group of monitors from the network. By default, all the monitors with the same software (e.g. S-ARK97) use the same user modes. If the monitors with the same software are desired to have different user modes, different groups must be defined, and the user modes defined during the monitor configuration must be saved in to the network (refer to the monitor’s manuals).
Monitors with different software versions, e.g. S-ARK95, S-ARK97 and S-ICU97 cannot use the same user modes.
To add a group:
1. Click Add Groups.
2. Give the group number and name. You can edit the group name and number by double clicking them. To remove a group:
Select Remove Group.
42
Document No. 8000633-1
Page 49
Adding, removing, and editing monitor locations
All the monitors in the network must be identified in the Network Setup to enable networking. Add all the monitors to all Centrals, when configuring the primary Central:
1. Highlight the Central of those added
in the 'Networks' field.
Datex-Ohmeda S/5 Central, ViewStation and Network
2. Click
3. Give the monitor ID number. This
4. Name the monitor.
5. Select the group which the monitor
If you want to edit these pieces of information, double click the desired item on the list. To remove a monitor:
Click Remove Monitor.
Add Monitor in the field of
‘Monitors’.
number is defined by the location ID plug of the Mon-Net cable, which must be connected to the ID connector on the monitor’s Network Board, B-NET.
belongs to.
Connecting networks together
Connecting the networks together enables the monitor-to-monitor communication between monitors residing in different monitor networks, and defines which Centrals can be viewed from a ViewStation.
1. Highlight the Central/ViewStation in
the 'Networks' field.
2. Click Select Connections.
3. Select which of the Centrals are
connected to the Central/ViewStation in question.
43
Document No. 8000633-1
Page 50
Technical Reference Manual
Saving the network setup
1. Select File-Save to save the configuration into the
NETWORK.AIC file.
2. Preview the setup by selecting the Print PreView.
3. Print the document of the setup and file it for the later
troubleshooting etc. purpose.
4. Exit the Network Setup by selecting Exit.
5.2.2 View setup
From a Central you can view only the monitors connected to the same 10 Mbps hub. From a ViewStation you can view any of the monitors in those networks connected together with it, see above. To enable viewing, configure the the monitor views with the ViewSetup software.
The ViewStation can have up to 20 user specific configurations, e.g. for different departments or patient groups. The maximum number of monitors in each configuration is 32, that are shown as up to 10 Multi View pages on Patient Screen. On Status Screen the pages are seen as a row of tabs. On one page you can only select monitors from the same monitor network. A Central can only have one configuration.
Click Central Setup icon and select ViewSetup.
44
Document No. 8000633-1
Page 51
Datex-Ohmeda S/5 Central, ViewStation and Network
The system suggests the name ‘Default’ for the first configuration. You can edit the name and other items by double clicking the item.
1. Click Add Config to make a new configuration and name it.
2. Highlight the desired Central. The list of monitors connected to it appears.
3. Click Add Page and name it.
4. Highlight the page and click Add Monitor to add the desired monitor to the page. The
monitor appears on the rightmost column, and its color in the middle column is changed to red. All monitors on a page must belong to the same monitor network.
To remove a monitor from the page:
1. Highlight the monitor in the rightmost column.
2. Click Remove monitor. To remove a configuration or page:
1. Highlight the configuration/page to be removed
2. Click the Remove Config./Page.
Saving the View Setup
1. After configuring, click File - Save. Also print the setup by clicking Print to make a hard copy
of the configuration for later troubleshooting etc. purpose.
2. Exit the View Setup program.
45
Document No. 8000633-1
Page 52
Technical Reference Manual
5.2.3 System setup
Passwords and user names
1. Click Central Setup icon and select SystemSetup.
2. Give the user name and password.
If you are installing the system, use ‘Install’ password with the default password ‘CCCCC’.
This password entitles to edit all the setup options. For the most frequently changed items,
like Patient Screen or printer settings, use ‘Setup’ password, default is ‘BBBBB’. It entitles
you to enter any other option group except "System Setup". To change the password:
Click the Change Password button.
NOTE: In the following lists of configurable options, all items are available for a Central setup. For the Network Manager configuration, only the items marked with NM are available, and for ViewStation, only the items marked with VS.
If the Network Manager configuration is used. It must be set on when the network software is installed, see page 66.
46
Document No. 8000633-1
Page 53
System Setup
Datex-Ohmeda S/5 Central, ViewStation and Network
System Setup sholud be used only when installing the system and requires the Install password.
General
NM
NM, VS
NM, VS
NM
NM, VS
NM, VS
Hospital name Keyboard
Central language
Record keeper language
Time format Use time server
Give the hospital name. The name appears in the printed anesthesia records.
Select the keyboard type and layout. The default value is ‘K-CENTRALB,
International’.
Select the system language. The default value is ‘English’.
For managing Record Keeper language, see ARK Configuration Software
documents. The default value is ‘1’.
Select the time format. The default value is 24 hours.
Select if the time server is used. The default value is ‘NO’.
Using the time server sets the same time for all the Centrals in Datex-Ohmeda Network (TCP/IP), and thus, for all the monitors in the entire network. If the network is connected to a Datex-Ohmeda IM server, use it as a time server. If no other IM server is available, use the primary Central as a time server.
47
Document No. 8000633-1
Page 54
Technical Reference Manual
NM, VS
NM
Time server address
Mon. network board name
Name the server which works as a time server. This selection can be changed
through ‘Components’ service page on the run.
If the network board is changed e.g. because of the failure, run the DOCCheckUp to
find out the network board name, and select the name for the board no 1 (monitor network). Default value is \Device\El90x2. For the use of DOCCheckUp, see chapter 9.2
NM
Cluster number
Give identification number (1-8) for each Central for patient data addressing. This
address is required when patient trend data is passed between the Centrals. See 'Menu server - Other options - External Central' settings later in this chapter, see page 55. Default value is 1.
Mode server No options available.
Monitor-to-monitor
Enable or disable monitor-to-monitor communication. Default value is ‘Yes’.
communication
Passwords
NM, VS
Exit password
Give the password required to close down the network. The default password is
‘AAAAA’.
NM, VS
Setup password
Give the password required to enter the System Setup Installation menu on
Central's /ViewStation (Patient Screen). The default password is ‘BBBBB’.
NM, VS
Service password
Give the password required to enter the service pages (Status Screen) and System
Setup Service menu (Patient Screen). The default password is ‘CCCCC’.
Log Files Files are discussed in Part II, in the Software introduction.
48
Document No. 8000633-1
Page 55
Patient Screens
Datex-Ohmeda S/5 Central, ViewStation and Network
General
VS
VS
VS
VS
VS
VS
Patient screen 1, 2 type
Page scroll
Select the Patient Screen type, 'CRT' for the display D-VIC17, or 'LCD' for D-
CFLT15. The default value is 'CRT'. Requires the Install password.
Select the interval for the automatic page scroll. The default value is 10 s.
interval Stop rotation Select ‘Yes’ (default) to stop the automatic page scroll, when the screen is selected. Screen mode
auto change
When there are two Patient Screens in the system: Select whether the inactive (non-
focused) Patient Screen will display the maximum number of monitors (Multi View) when the Patient View is selected on the active (focused) screen. The default value is ‘Yes’.
Multi View page end action
Restore cursor location
Select the cursor action when the last patient of a Multi View page is reached by
turning the ComWheel (Stop/Change Page/Change Screen/Go Around) The default value is ‘Change page’.
Select ‘Yes’ (default) to return the cursor to the same place when the screen is
selected again. Requires the Install password.
49
Document No. 8000633-1
Page 56
Technical Reference Manual
VS
NM
Waveforms
VS
VS
VS
VS
Hide patient’s id (screen 1,
Select whether patient name and ID are hidden on the Patient Screen or not. The
default value is ‘No’.
screen 2) Hide patient
names on
Select whether patient name and ID are hidden on the Status Screen or not. The
default value is ‘No’.
Status Screen
ECG grid Select whether a grid is used for the ECG curve in the Multi View. The distance
between grid points corresponds to approximately 400 ms for ECG Speed selection ‘Low’, 200 ms for ECG Speed selection ‘Medium’ and 100 ms for ECG Speed selection ‘High’. The default value is ‘No’.
ECG speed, Patient view
Set the ECG sweep speed for Patient View, 8 and 16 patient modes. The default
value is ‘Medium’. /8 patient mode / 16 patient mode
Expiratory flow
Select ‘Yes’ (default) to display the expiratory flow positive in the spirometry.
displayed positive
Overlapping pressures
Select if the pressure waveforms are combined, i.e., displayed overlapping each
other, or not. The default value is ‘No’.
Alarms
NM, VS
VS
VS
VS
Displayed alarms
Alarm sounds enabled
ISO alarm sounds
Audible alarms always on
Select which type of alarms are displayed on the Multi View and the Status Screen.
The default value is ‘Yellow and red alarms’.
Select if the alarm sound is enabled. The default value is ‘YES’. Requires the Install
password.
Select if the alarm sounds are according to ISO standard (Yes) or Datex-Ohmeda
specific (No). The default value is ‘YES’. Requires the Install password.
Default value ‘YES’ means that the alarms are enabled in Central even when the
audio alarms have been turned off in the monitor. Exception: Silencing/suspending
alarms for 2 minutes at the monitor, silences also the alarms at the Central.
‘NO’ means that Central’s audible alarms follow the settings made at the monitor.
Requires the Install password.
VS
Silencing individual monitors
Select ‘YES’ to enable silencing desired monitor locations’ alarms permanently on
the Central/ViewStation. ‘NO’ disables this selection. NOTE: Affects only Central's
alarms, not the monitor's. Requires the Install password.
Fixed View No options available
Other options No options available
50
Document No. 8000633-1
Page 57
Printers
Datex-Ohmeda S/5 Central, ViewStation and Network
Read also chapter 1.
General
VS
VS
Default printer Select the default printer, the default value is 'Printer 1'. Anesthesia
record copies
VS
Snapshot in printer
Laser printers
VS
Printer 1 (2 etc) name
VS
Printer 1 (2 etc) port
Select how many copies of anesthesia record are printed. Default value is 1.
Print snapshot in printer if the recorder is not responding. Default value is 'No'.
Requires the Install password.
Give a descriptive name for the printer you are using. The name appears on the
monitor when printing.
Select the printer the operating system is using.
51
Document No. 8000633-1
Page 58
Technical Reference Manual
Recorder
VS, NM
VS, NM
VS, NM
VS, NM
VS, NM
NM
NM
Recorder type Select 4 inch recorder. (The 2-inch alternative is for later use.)
NOTE: Select ‘No recorder’ when laser printer is used. 2-inch
Select the recorder port for a 2-inch recorder. Requires the Install password.
recorder port
Recorder
Select the speed of 1, 6.25, 12.5, 25 and 50 mm/sec, the default is ‘Auto scale’.
speed* Record time
before manual
Manual record length*
Recorder
Define how long a time before a key press is printed when the Recorder
Start/Stop button is pressed.
Set the total length of the manually started strip chart. The default value is 30
seconds.
Define the waveform scales. The default value is ‘Auto scaling’
ECG/BP/Pleth scale
Recorder trace 1-4
Define the waveforms recorded.
Trace 1: the default value is ECG + heart rate Trace 2: the default value is Invasive blood pressure 1 Trace 3: the default value is Invasive blood pressure 2 Trace 4: the default value is Plethysmograph
Alarm triggered printing
NM
NM
NM
Items marked with an asterisk * can also be adjusted in the System - Installation - Recorder setup menu of the Central’s Patient Screen. The settings made through menu are not permanent but valid only until the system is restarted. See Part II for further information.
Autotrigger on*
Record time before alarm
Alarm triggered record length*
Select ‘Yes’ to start a printer or recorder automatically when a serious alarm occurs.
Define how long a time before the alarm will be included in printing.
Set the total length of the alarm triggered strip chart. The default value is 20 seconds.
52
Document No. 8000633-1
Page 59
Menu server
Datex-Ohmeda S/5 Central, ViewStation and Network
Menu server is used with Centrals and Network Managers only, not ViewStations.
File paths
NM
NM
NM
Patient data file system active
Path 1 to menus Path 2 to menus
Patient data path
Select if the patient data is stored or not. The default value is ‘Yes’.
Name the folder which contains the record keeping menu files.
The directory indicated by ‘Path 1 to menus’ contains the original record keeping menu files created by the ARK Configuration software. These files are automatically copied to the directory indicated by ‘Path 2 to Menus’ when the network software is started. ‘Path 1 to menus’ and ‘Path 2 to menus’ should be the same, the default being
D:\CENTRAL\MENUS, unless the Central is are connected to (TCP/IP) network.
Name the directory which contains the patient data and preoperative data files. The
default value is
D:\CENTRAL\PATDATA.
NOTE: You can select from the existing directories only, or use the default path. The default path is created automatically when the network software is started for the first time after setup.
53
Document No. 8000633-1
Page 60
Technical Reference Manual
NM
Archive patient data path
NM
Disk cleaner enabled
Time settings
NM
Preop data storage time (days)
NM
Postop data storage time (days)
NM
Patient data storage time (days)
Give the directory where the patient data is archived. The default is
D:\CENTRAL\PATDATA\ARC
Select if the automatic disk cleaner is enabled according to the ‘Patient data
storage time’ and ‘Preop data storage time’. The default value is ‘Yes’
Disable the disk cleaner only if the Database Outlink software is run regularly to
archive the patient data files. Requires the Install password.
Give the storage time for the preoperative data files (1-8 days). Default value is five
days. After this time the files are erased by the built-in disk cleaner.
NOTE: The preoperative cases do not have any measured parameters or timed events. The preoperative case can be created by entering only untimed events through Record Keeper Keyboard or by feeding in a preop case from the hospital information system. Preoperative cases can be loaded to the monitor from the 'Preop Data / net' -menu created using the ARK Configuration Software. The menu number is 165.
Give the storage time for the postoperative data files (1-8 days). Default value is five
days. After this time the files are erased by the built-in disk cleaner.
Give the storage time for the patient data files (2-90 days). As the time expires the
files will be deleted automatically. The default value is 7 days, and it is recommended
not to be more than 14 days. After this time the files are erased by the disk cleaner.
NM
Lab data storage time (days)
NM
Preop list time range (hours)
NM
Patient list time range (hours)
NM
Lab list time range (hours)
NM
User ID timespan
Inlink options
NM
Inlink incoming message path
NM
Path to case list
Give the storage time for the laboratory result files (1-8 days). Default value is five
days. After this time the files are erased by the built-in disk cleaner.
Give the access time (1-96 hours) for the preoperative data files (containing patient
name, id, demographics, etc.). The default value is 48 hours.
Give the access time for the patient data files (1-24 hours). The default value is 24
hours.
Give the access time (1-96 hours) for the laboratory result files. The default value is
12 hours.
Set the length of time ( 1-30000 seconds) when the ARK user id is valid after last
documentation. Default is 300s. Requires the Install password.
These options are for Datex-Ohmeda Inlink users’, refer also to documentation delivered with it. Requires the Install password.
For Inlink users only, refer to its documentation. Requires the Install password.
For Inlink users only, refer to its documentation. Requires the Install password.
NM
Outmsg path (1-9) for exported data
54
Document No. 8000633-1
For Inlink users only, refer to its documentation. Requires the Install password.
Page 61
Datex-Ohmeda S/5 Central, ViewStation and Network
NM
Event codes
For Inlink users only, refer to its documentation. Requires the Install password. (1-9) for exported data
Search options Anesthesia Record Keeping related search options, refer to the record keeping
documentation and Datex-Ohmeda S/5 Anesthesia Pathway Builder User's
Reference Manual. Requires the Install password.
Other options
NM
NM
Preop data sort criteria
Patient data sort criteria
Select the order (time or alphabetical) of the menu lines in the ‘Preop Data / net’
menu (165). The default value is ‘Alphabetical order’.
Select the order (time or alphabetical) of the menu lines in the ‘Patient from Net’
menu. The default value is ‘Time order’.
NOTE: This selection is available only when S-ARKXX or S-ICUXX softwares are used.
NM
NM
Lab data sort criteria
Check patient name before load
Select the order (menu or alphabetical) of lab data displayed. The default value is
‘Alphabetical order’.
Select whether patient data identification information (patient name and ID) is
checked before loading patient data from Central to monitor. Default value is ‘Yes’.
If the value of this option is set to ‘Yes’, and the identification of loaded patient data
does not match to the identification defined at the monitor, a message ‘Loading
Failed. Wrong patient.’ is displayed in the message field of the monitor.
NM
NM
NM
External Central 1, 2, 3
External load action
Enhanced
Give the cluster number and path to the patient data files to enable patient data
transfer between Centrals. Requires the Install password.
Select whether to make a back-up or not when patient data files are loaded from
another monitor network. Default is ‘Make back-up’.
For Inlink users only, refer to its documentation. Requires the Install password. patient lists on
NM
Data origin
For Inlink users only, refer to its documentation. Requires the Install password. filter
NM
NM
Data type filter For Inlink users only, refer to its documentation. Requires the Install password. Data time
For Inlink users only, refer to its documentation. Requires the Install password. range filter
Configuring the Status Screen
The Status Screen uses the same configurations and pages as Patient Screen, which are defined in ViewSetup. In addition, there is also a page called ‘All’, where all the monitor locations are displayed independently from the pages they are configured on.
Additionally, you can define pages for information that is displayed on the Status Screen only, like event data from the record keeping (e.g. personnel, medication etc.). There can be up to 10 different pages, each containing up to three columns, see below.
Pages under Service tab cannot be configured. The service pages are introduced in the Part II of this manual.
Document No. 8000633-1
55
Page 62
Technical Reference Manual
Pages
Define the name and up to three columns to be shown on the page:
1. Select Status Screen - Pages and click Add page.
2. Select the columns to be displayed on the page by double clicking the desired column in
To remove a column from a page:
Double click the column in the ‘Current Columns’ box
Columns
If there is not a suitable column in the list of available ones, you can make one of your own:
Select Status Screen - Columns and click Add column.
Name the column according to your purpose.
Give class information, or upper and lower limit of the code range according to the codes of
Tick ‘Show Time Stamp’ if you want to display the time of the event.
56
Document No. 8000633-1
the ‘Available Columns’ box.
the Record Keeper, so that information find its way to the desired column. For further information on the code ranges, refer to the Datex-Ohmeda Record Keeping documentation or technical support.
Page 63

6 CONNECTING NETWORKS

1. All Centrals and ViewStations to be connected must have a different PC name.
All Centrals are shipped as Central_1, and one Central, with the status of primary Central, should use that name. Other Centrals can be renamed under Windows NT Control Panel ­Network - Identification: click Change button. See page 64.
2. Enable time services, and use a Datex-Ohmeda IM server, or the primary Central, as a time
server, see page 47.
Datex-Ohmeda S/5 Central, ViewStation and Network
3. Make sure you have copied a similar
made appropriate configurations for each of them. Especially make sure you have defined which Centrals and ViewStations are connected together, see Select connections, page 43.
4. Make sure you have defined individual cluster numbers for each Central in System setup -
General, see page 48, and given these cluster numbers and patient data addresses in the in System setup - Menu server - Other options, see page 55.
5. Make sure that you have enabled monitor-to-monitor communication, in System Setup -
General, see page 48.
NETWORK.AIC to all Centrals and ViewStations, and
57
Document No. 8000633-1
Page 64
Page 65
Datex-Ohmeda S/5 Central, ViewStation and Network

7 INSTALLING A RECORDER OR PRINTER

You can connect either a laser printer or a strip-chart recorder to a Central or ViewStation. Printers and recorders are configured as part of the system setup, as shown on the page 51.

7.1 Installing a recorder

1. Connect the recorder cable to the parallel connector on the rear panel of Network
Computer, see Figure 4-2, and to the connector on the recorder’s rear panel.
2. Connect the recorder’s power cable to the power inlet and to the wall socket.

7.2 Installing a printer

You can install a printer:
locally to a Central/ViewStation/Network Manager
locally to a Central/ViewStation/Network Manager but shared for other
Centrals/ViewStations/Network Manager as well
directly to the network hub (100 Mbps), thus shared for Centrals, Network Managers and
ViewStations in the network
NOTE: If the Central or ViewStation, which the printer is connected to, is turned off or goes down, the printer is not available for other Centrals/ViewStations.
Following installation instructions are introductory only. Follow the instructions provided by the manufacturer of the operating system, printer and network board.
After installation, configure the printer to work with the Datex-Ohmeda Central or Network with the Central’s.
Installing a local printer
1. Connect the printer cable to the connector on the rear panel of Network Computer, and to
the connector on the printer’s rear panel. Secure the connections with finger screws.
2. Connect the printer’s power cable to the power inlet and to the wall socket.
Adding a local printer to the operating system
1. Turn on the printer and computer, start the Windows NT and log on as a central
administrator.
2. Click Start in the lower left corner of the screen and select Settings - Printers - Add Printer.
The operating system starts a program to configure the printer connection.
3. Select My computer. After each step move forward by clicking Next.
4. Select the printer port where you attached the printer cable (typically LPT1).
5. Select the manufacturer and model from the list. If your printer is not on the list, click Have
disk button and insert the installation disk which came along with the printer.
6. Select the driver with PCL language.
Document No. 8000633-1
59
Page 66
Technical Reference Manual
7. Select whether the printer will be local or shared with other Centrals in the network. If it will
be shared, give it a share name, e.g. CENTRAL_1 Printer. Use the same name throughout the configuration.
8. Print a test page. After successful installation the printer icon with the printer name appears into Printers folder.
Installing a network printer
If the printer is going to be shared by several Centrals and/or ViewStations, it is recommended to connect the printer to the hub. This installation requires that the printer is equipped with a network board. The printers purchased from Datex-Ohmeda have the network board installed. If you use a locally purchased printer, install the network board (HP JetDirect) according to the instructions provided by the printer manufacturer.
Preparations
1. Connect the printer cable to the 100 Mbps TCP/IP network hub, and to the connector on the
printer’s rear panel.
2. Connect the printer’s power cable to the power inlet and to the wall socket.
3. Print the JetDirect configuration page from the printer’s menu, and notice the printer’s
hardware address, you’ll need it later on.
4. Use preferably default Datex-Ohmeda IP addresses. IP addresses can be configured
manually, refer to printer’s manual.
The Datex-Ohmeda default IP addresses are:
PRINTER_1 192.168.1.50 PRINTER_2 192.168.1.51 PRINTER_2 192.168.1.52
Adding a network printer to the operating system
1. Insert the HP JetDirect CD ROM to the CD ROM drive.
2. Click the ‘Install Network Printer’ icon. The installation program starts.
3. Select ‘Specify a printer by address’.
4. Enter the hardware address you got during preparations.
5. Set mode to TCP/IP.
6. To select the driver click ‘Have Disk’ button and insert the disk provided by the printer
manufacturer.
7. Give the printer name.
8. Select that the printer is used shared and give the share name (by default the same as
9. Print a test page. After the successful installation the printer icon appears to the Printers folder.
60
Document No. 8000633-1
above).
Page 67
Configuring the printer
1. Click Central Setup icon and select SystemSetup.
2. (If you are configuring a locally connected printer and you have had a recorder installed
earlier to this Central, go to System setup and Printers- Recorder - Recorder type and select No recorder, page 48)
3. Select Printers - Laser Printers - Printer 1, 2, etc. port.
4. Click the little square beside the Current Value box and select the printer.
5. Select the size and orientation of the paper, and the paper source, and save with ‘OK’.
6. Go to Printers - Laser Printers - Printer 1, 2,... name, and give a descriptive name. This
name is seen on the bedside monitors.
7. Click File and select Save, and exit the Setup program.
Datex-Ohmeda S/5 Central, ViewStation and Network
61
Document No. 8000633-1
Page 68
Page 69

8 SOFTWARE INSTALLATION

The operating system, the network software, and the setup software are installed by the manufacturer. Re-installation may be required, e.g. after changing the hard disk of the Network Computer.
Before installation, make sure you have all the necessary drivers available:
network board driver
audio driver
display controller board driver
printer driver, if a printer is installed

8.1 Operating system

All the Network Computers shipped from the manufacturer are equipped with Windows NT operating system with Service pack 6, or newer. If the hard disk of the computer is changed the operating system has to be reinstalled.
Datex-Ohmeda S/5 Central, ViewStation and Network
User names and passwords
To enter the operating system requires a password which depends on what kind of rights each user needs to have. The default passwords are
User User id Password
administrator, e.g. IT department of the hospital
central administrator, person(s) responsible for Central configuration and installation, e.g. IT person of a department
user, e.g. doctors and nurses using the Central in daily routines
Make sure that the administrator and central administrator passwords are changed frequently to prevent misuse. Do not change the central user password, because this password is required for starting the Central software.
You must log on as an administrator or central administrator to change settings under operating system. Remember to change the user again to central user after the configuration is done.
To log on as a different user press and hold the shift key while the operating system starts up.
administrator admin
central admin centraladmin
central user centraluser
63
Document No. 8000633-1
Page 70
Technical Reference Manual
8.1.2 Installing the operating system
Following installation instructions are introductory only. Follow the instructions provided by the operating system manufacturer.
1. Place the Windows NT 4.0 workstation CD-ROM into the CD-ROM drive and turn the
computer on.
2. If you have some version of NT already installed, choose to install a fresh copy.
3. Check that setup program finds the right hardware and software components, e.g.:
Computer: Standard PC
Display: Auto Detect
Keyboard: XT, AT, or Enhanced Keyboard (83-104)
Keyboard Layout: US
Device: Logitech Mouse Port Mouse
No Changes: The above list matches my computer.
3. Make sure that your disk will be divided into three parts C:, D: and E: in the following way:
C: 500Mb FAT D: 2Gb NTFS E: rest NTFS
4. Install the Windows NT on the E:\WINNT
5. Remove any floppy or CD disks from the drives and restart the computer.
Windows NT setup
1. Insert the Windows NT Workstation CD-ROM into CD-ROM drive when prompted, and
2. In ‘Setup Options’ select Custom Setup.
3. Fill in the prompted information: name, organization and product ID-number from the
4. Name the computer, use one of the following:
5. Change the central administrator password, max 14 characters. The factory installed
perform the rest of the setup.
‘Certificate of the Authenticity’.
Central_1 Central_2 Central_3 Central_4 ViewStation_1 ViewStation_2 ViewStation_3 ViewStation_4
password is “centraladmin”. NOTE: The administrator account allows maximum access to your computer’s recourses. Therefore the administrator password is an important piece of information and should be stored in a safe place.
6. Under ‘Select Components’, select the Notepad under ‘Accessories’ deselect all the other
7. During Networking installation, select the options ‘This computer will participate on a
64
Document No. 8000633-1
components to prevent misuse.
network’ and ‘Wired to the network’.
Page 71
Datex-Ohmeda S/5 Central, ViewStation and Network
8. Select the network adapter, and install the corresponding driver from the disk.
9. Select ‘TCP/IP protocol’ for the network TCP/IP protocols.
10. Select default services in the ‘Network Services’.
11. Give the prompted IP addresses, the default ones are:
Central_1 192.168.1.30 Central_2 192.168.1.31 Central_3 192.168.1.32 Central_4 192.168.1.33 ViewStation_1 192.168.1.40 ViewStation_2 192.168.1.41 ViewStation_3 192.168.1.42 ViewStation_4 192.168.1.43
12. Define the workgroup for the workstation as ‘DONET’.
13. Finish setup. Now the Windows NT is installed and configured. Check the final tunings:
1. Set Date/Time Properties according to your location and check that time and date are right.
2. Change in ‘Display Properties’ the desktop area to 1024 by 768 pixels; large fonts.
3. Prepare the Emergency repair disk.
4. Remove all disks from the drives and restart the computer with the administrator password
which you just gave.
8.1.3 Installing NT Service Pack
1. Insert the Microsoft Windows NT Service Pack 6 CD ROM into the CD ROM drive of the
Network Computer.
2. Accept the License Agreement, but do not backup files to uninstall new Service Pack.
3. Click Install.
4. Restart the computer.
8.1.4 Installing driver for the network board
Installing the above mentioned programs install also the network board for Datex-Ohmeda Network. To install the network board for monitor network do the following.
1. In the operating system select Start - Settings - Control Panel - Network - Adapters.
2. Click Add and Have Disk.
3. Insert the driver disk which
4. Select Intel(R) PRO Adapter and click OK.
5. In the Network folder select Bindings.
6. Select Show bindings for all adapters.
7. Click Disable and Close.
8. Restart the computer.
65
Document No. 8000633-1
Page 72
Technical Reference Manual
8.1.5 Installing driver for display controller board
Run SETUP.EXE file to install the driver for the display controller board.
8.1.6 Configuring users
1. In the operating system select Start - Programs - Administrative Tools -User Manager.
2. Define the following group memberships
User Account: Group Membership Default Password
Central Admin Administrators centraladmin Central User Users centraluser

8.2 Installing the network software

The network software, S-CNETXX and S-VNETXX is pre-installed by the manufacturer. To reinstall:
1. Close all other applications.
2. Log in as a central administrator.
3. Make backup copies of the files networc.aic, monviews.aic and setup.aic to a diskette.
4. Insert the Network Setup disk in the disk drive.
5. Use NT Explorer, or select Start and Run, and start SETUP.EXE. The 'Welcome!'
display should appear.
6. Follow instructions displayed. The button 'Next' means approval, 'Back' takes you to the
former view, and 'Cancel' interrupts the installation.
If the network software is already installed
If a former version of the software is found, select 'Next' to reinstall the software, or cancel
the installation. Follow the instructions displayed. If the software is installed in the computer the former version exists, the old files are moved in a subfolder 'Old Central Setup'.
Central or Network Manger?
Select according to your system between a complete Central with Patient Screen(s) and
Network Manager configuration.
Final check
1. Click the box that enables reading the Readme file and read the latest release notes.
2. Restart the Central and log in as a central user.
66
Document No. 8000633-1
Page 73
Datex-Ohmeda S/5 Central, ViewStation and Network

9 FUNCTIONAL CHECK

9.1 Checking the equipment and connections

Turn on the power to the computer. Confirm that all networked monitors register to the network.
9.1.1 Checking the Patient Screen
1. Confirm that the Multi View is activated.
2. Press the Setup key. Confirm that the setup menu appears.
3. Rotate the ComWheel. Confirm that the highlight moves within the menu.
4. Press the Normal Screen key. Confirm that the menu disappears.
5. Rotate the ComWheel. Confirm that the cursor moves respectively on the display.
6. Push the ComWheel. Confirm that the Patient View is activated.
7. Press the Multi View key. Confirm that the Multi View is activated.
9.1.2 Checking the Status Screen
1. Confirm that the user pages are activated.
2. Click the tabs with the mouse. Confirm that the pages are changed.
3. Click the system menu in the upper left corner of the screen. Confirm that the menu
appears.
9.1.3 Checking the monitor-to-monitor communication
1. Make sure the monitors are connected to the wall boxes and turned on.
2. Press
3. Confirm that you receive information from the other monitor you chose, both the Patient
Patient Data on a monitor’s command board, and select Other Patients.
view and alarms.
9.1.4 Checking the communication between the monitor networks
Repeat the procedure above but make sure that the monitors are in different monitor
networks.
9.1.5 Checking the printer
1. Press Patient Data.
2. Select the information you want to print and press ComWheel.
3. Make sure that the desired printing appears to the desired printer.
4. Make sure that Headings of the printings are right.
67
Document No. 8000633-1
Page 74
Technical Reference Manual
9.1.6 Checking the recorder
1. Press Recorder Start/Stop to start a manual strip.
2. Confirm that the recorder prints the desired waveforms and headings.

9.2 Checking the configuration

1. Start the DOCCheckUp program by clicking it in the D:\CENTRAL\PROGRAMS folder.
2. See the
See Appendix for an example of a log printout.
DOCCheckUp Introduction
Datex-Ohmeda Central CheckUp Utility (DOCCheckUp) is a utility that helps system administrators to find and fix errors during DOC system installation and configuration changes.
The DOCCheckUp could be used every time before system startup or only after system installation and configuration changes.
Tasks performed by DOCCheckUp:
1. Check platform
2. Check computer name
3. Check device drivers
4. Check current account
5. Check network boards
6. Check status of TCP/IP and Windows Socket (WinSock) 2.0
7. Check directory structures
8. Check central system configuration (
CENTRAL.LOG file created by it, and check that there are no fatal error messages.
CENTRAL.AIC)
9. Check startup configuration (
10. Check Datex-Ohmeda networks configuration (
11. Check view configurations (
12. Check general configurations (
13. Check system licenses In the Appendix E, there is an example of the Central.log created by the DOCCheckup.
Task Descriptions
Check Platform:
Requirements:
Windows NT WorkStation/Server 4.0
Service Pack 6 or higher
68
Document No. 8000633-1
STARTUP.AIC). Sets ARP table entries static.
NETWORK.AIC)
MONVIEWS.AIC)
SETUP.AIC)
Page 75
Datex-Ohmeda S/5 Central, ViewStation and Network
Error message Reason
“GetVersionEx(): … “ probably earlier version of Windows NT (3.5) “Unsupported: Microsoft Windows 3.1” Windows 3.1 is not supported “Unsupported: Microsoft Windows 95” Windows 95 is not supported “Unsupported: Microsoft Windows 98” Windows 98 is not supported “Unknown version of Microsoft Windows” Can not find Windows version “Unsupported: Microsoft Windows NT … ” Probably not proper Service Pack.
Success Meaning
“Microsoft Windows NT Service Pack 6 ” Windows NT 4.0 with Service Pack 6
Check computer name:
Computer name defined in Control Panel -> Network -> Identification -> Computer Name
Error message Reason
“GetComputerName(): … “ probably earlier version of Windows NT (3.5)
Success Meaning
“Computer Name: … ”
Check current account
The current account that is used for start the DOC system has to have administrator
privileges
Computer name is defined.
This value will be used as a local Datex-Ohmeda network
name.
A later check of Control Panel -> Network -> Protocols ->
TCP/IP Protocol -> Properties -> DNS -> HostName should give the same value.
Error message Reason
“GetUserName(): … “ probably earlier version of Windows NT (3.1)
“Account … does not have sufficient privileges.”
Current account does not have administrator privileges, that means that later on the system startup will failed.
Success Meaning
List of required privileges:
- All the listed privileges are needed.
- “Bypass traverse checking”
- “Shut down the system”
- “Act as a part of the opeating system”
- “Increase quotas”
- “Replace a process levell token”
- “Change the system time“
- “Load and unload device drivers“
69
Document No. 8000633-1
Page 76
Technical Reference Manual
Check network boards
Requirements:
Two (2) network adapters have to be installed in Control Panel -> Network -> Adapters
They should be 3COM EtherLink XL network adapters. The one connected to Monitor
Network (MON adapter) should not be bound to TCP/IP or WINS client protocols.
TCP/IP protocol has to be installed.
If DOC is placed on standalone computer, then Control Panel -> Network -> Protocols ->
TCP/IP Protocol -> Properties -> DNS -> Domain is empty, Default Gateway, Primary WINS Server contains the local IP address, DNS for Windows Resolution is enabled and NetBios Interface Service should be installed.
Error message Reason
“No network adapter present“
No network adapter has been installed, or the installed network adapters do not
function.
“RegOpenKeyEx():“ No value found in
Software\Microsoft\Windows NT\CurrentVersion\NetworkCards or
System\CurrentControlSet\Services\Parameters\Tcpip registry keys.
Success Meaning
Adapter …: Number of the adapter
Service Name: … Service name of the adapter
Symbolic Link: Symbolic link of the adapter. If this adapter is used for AS/3 network, the same
value should be defined in MONGATE_BOARD (SETUP.AIC).
Description: Description of the adapter
Manufacturer: Manufacturer of the adapter
Using for TCP/IP Networking This adapter probably is used for TCP/IP networking
Can be used for monitor networking
This adapter probably can be used for AS/3 networking
70
Document No. 8000633-1
Page 77
Check status of TCP/IP and Windows Socket (WinSock) 2.0
Requirements:
TCP/IP protocol has to be installed.
Error message Typical reason
Datex-Ohmeda S/5 Central, ViewStation and Network
“WSAStartup(): …“ “gethostname(): …” “gethostbyname(): …
Too many open files. Too many open sockets: Each implementation may have a maximum number of
socket handles available, either globally, per process or per thread. Too many processes:
A Windows Sockets implementation may have a limit on the number of applications that may use it simultaneously. WSAStartUp may fail with this error if the limit has been reached.
Host not found: No such host is known. The name is not an official hostname or alias, or it cannot be found in the database(s) being queried. This error may also be returned for protocol and service queries, and means the specified name could not be found in the relevant database.
Invalid service provider version number. A service provider returned a version number other than 2.0.
Valid name, no data record of requested type The requested name is valid and was found in the database, but it does not have the correct associated data being resolved for. The usual example for this is a hostname -> address translation attempt (using gethostbyname which uses the DNS (Domain Name Server), and an MX record is returned but no A record ­indicating the host itself exists, but is not directly reachable.
This is a non-recoverable error This indicates some sort of non-recoverable error occurred during a database lookup. This may be because the database files (e.g. BSD-compatible HOSTS, SERVICES or PROTOCOLS files) could not be found, or a DNS request was returned by the server with a severe error.
Unable to initialize a service provider. Either a service provider's DLL could not be loaded or the provider's function failed.
Network subsystem is unavailable The Windows Sockets implementation cannot function at this time because the underlying system it uses to provide network services is currently unavailable. Check:
that the appropriate Windows Sockets DLL file is in the current path,
that they are not trying to use more than one Windows Sockets
implementation simultaneously. If there is more than one WINSOCK DLL on your system, be sure the first one in the path is appropriate for the network subsystem currently loaded.
the Windows Sockets implementation documentation to be sure all
necessary components are currently installed and configured correctly.
Document No. 8000633-1
71
Page 78
Technical Reference Manual
Error message Typical reason
“WSAStartup(): …“ “gethostname(): …” “gethostbyname(): …
Non-authoritative host not found
This is usually a temporary error during hostname resolution and means that the
local server did not receive a response from an authoritative server. A retry at
some time later may be successful.
cont.
Success Meaning
“Initializing Winsock 2.0 … -> OK”
“Hostname … -> OK” - Found HostName in Control Panel -> Network -> Protocols -> TCP/IP
“IP address … -> OK” - Successful retrieve IP address from HostName “Stop WinSock 2.0 … -> OK” - Successful shutdown Windows Socket 2.0
Check directory structures:
The following directories should exist:
D:\
D:\CENTRAL\GSPUI
D:\CENTRAL\GSPUI\GSP
D:\CENTRAL\GSPUI\TEXTS
D:\CENTRAL\MONITORGATE
D:\CENTRAL\PRINTERSERVICES
D:\CENTRAL\PROGRAM
D:\CENTRAL\SETUP
D:\CENTRAL\STATUSSCREEN
D:\CENTRAL\TOOLS
- Winsock 2.0 initialization successful
Protocol -> Properties -> DNS -> HostName and it is the same like Computer Name
Warning if the following directories do not exist
D:\CENTRAL\B2BMANAGER
D:\CENTRAL\MENUSERVER
D:\CENTRAL\MODESERVER
Error message Reason
“Warning … not exist“ - The directory does not exist, but it is not critical error
“Error … not exist“ - The directory does not exist and it is a critical error.
Success Meaning
“… -> OK” - Directory exists
72
Document No. 8000633-1
Page 79
Check Central configuration (CENTRAL.AIC)
Datex-Ohmeda S/5 Central, ViewStation and Network
Requirements:
CENTRAL.AIC and CENTRAL.OPT must exist and be well configured.
Error message Reason
“Loading \CENTRAL\PROGRAM\CENTR AL.OPT failed - line=…: …. “
“Loading \CENTRAL\PROGRAM\CENTR AL.AIC failed - line=…: …. “
1. File is used by other process
Probably the file is open and locked by other application
2. File pointer was moved
The file was changed by other application during loading
3. Confuse read and write functions
File content is changed by other application
4. I/O error: Error occurs with disk-,file access/read/write
5. Missing the closing string mark: The option value is defined between “ ”
6. The comment character ([]) is missing
Characters [ and ] is used for comment, but must be used in pair.
7. The read text is longer then 32 or 255
The text defined without specified length is longer than 32 or 255 (depends on option)
8. Read an invalid char: Invalid character found in the file.
9. Missing keyword or value string: Keyword should be defined, but not
found in the file, or the value string is not defined at all.
10. Missing '=' character: Character = is used for assignment, but not found
after the option name.
11. Missing ';' sign: Option definition should end with character ;, but not
found
12. Missing '{' or '}: Character { or } is found without pair
13. Incorrect keyword: Check OPT file.
14. Incorrect type of the keyword: Check OPT file.
15. Incorrect value: The value probably is defined in incorrect type or
enumeration
16. Required keyword is missing
17. Keyword should be found in AIC
18. Required keyword repeated: Keyword should be repeated in AIC
“… not accessible, log option -
- Logfile is not accessible, the log option might be turned off earlier.
> off”
“Startup file … not accessible” - Central startup file defined in a STARTUP_EXE option can not be found “Working directory … not
accessible”
- The WORKING_DIRECTORY of STARTUP_EXE, COMPONENT_INIT or
COMPONENT_AVAIL option is not accessible, probably does not exist.
“File to be read … not
accessible”
- Checking of existence of file defined in FILE_READ failed, probably does
not exist or wrong path
73
Document No. 8000633-1
Page 80
Technical Reference Manual
Error message Reason
“Component … not accessible” - Checking of existence of NAME defined in COMPONENT_INIT or
COMPONENT_AVAIL failed, probably does not exist or wrong path
Success Meaning
“Loading Central Setup ... ->
- Successful load CENTRAL.OPT and CENTRAL.AIC
Loaded”
“Checking Central Setup ... ->
- No error found
OK”
Check startup configuration (STARTUP.AIC)
Requirements:
STARTUP.AIC and STARTUP.OPT must exist and well-defined.
DOCSTARTUP.EXE must exist in D:\CENTRAL\PROGRAM
Error message Reason
“CreateProcess() …“ - DOCSTARTUP.EXE not found or wrong directory “-> Error: FILE_COPY(…,…) …“ - File copy operation failed “-> Warning: FILE_COPY(…,…) …“ - File copy operation failed, but not critical error “-> Error: PATH_CHK(…) …” - Path does not exist “-> Warning: PATH_CHK(…) …” - Path does not exist, but not critical error “CreateProcess(D:\CENTRAL\TOOLS\ACFO..” - ACFO utility not found
Success Meaning
“Checking StartUp Setup … -> OK” - Startup configuration correct
Check Windows NT Address Resultion Protocol (ARP) tables
Error message Reason
“Warning: ARP table – entries could not be made static.”
— Address resolution protocol tables could not be made.
Success Meaning
“OK: ARP table entries made static.” — Address resolution protocol table entries successfully
made.
74
Document No. 8000633-1
Page 81
Datex-Ohmeda S/5 Central, ViewStation and Network
Check Datex-Ohmeda network configuration (NETWORK.AIC)
Requirements:
NETWORK.AIC
and NETWORK.OPT must exist in D:\CENTRAL\SETUP and well-
defined.
Error message Reason
“Loading
See Check central supervising configuration \CENTRAL\PROGRAM\NETWORK.OPT failed ­line=…: …. “
“Loading \CENTRAL\PROGRAM\NETWORK.AIC failed - line=…: …. “
“…No network defined” No network defined in the NETWORK.AIC
“Network …'s object is empty” NETWORK definition found, but empty
“Network …'s IPAddress not found” NETWORK definition found, but MonGate option not
“Network …'s name not found” NETWORK definition found, but Name option not
“Network …'s type not found” Type option of NETWORK definition not found
“No monitor belong to this Central” Central type DOC should have monitors
“There are monitors which belong to this
ViewStation type DOC should not have any monitor ViewStation”
“Network …'s port not found” MonGate_Port option of NETWORK definition not found
“Network …'s ID not found” ID option of NETWORK definition not found
“Interconnecting Network …'s IP not found” Network to be connected to the local Central has no IP
address
“Interconnecting Network …'s Port Number not found”
“No connection backward” or “Connection from … to … is one-way”
“Local Central Network not found, probably invalid IP address or wrong interconnection”
Network to be connected to the local Central has no port
number
Connection between networks is only one-way. It should be
two-way
Invalid IP address, incorrect interconnection or IP address can
not be resolved (no name server or NetBios interface is not
installed)
“Interconnecting from … to Network ID … not found”
The defined network ID can not be found in the NETWORK.AIC
file, that network could be removed before.
“Validating IP: … “ TCP/IP or WinSock error. See Check status of TCP/IP and
Windows Socket (WinSock) 2.0
Success Meaning
“Loading Central Network Setup … -> Loaded” - File loading successful “Initializing Winsock 2.0 ... -> OK” - TCP/IP and WinSock 2.0 present
Document No. 8000633-1
75
Page 82
Technical Reference Manual
Success Meaning
“Local central network name: …” - Name of the local central “Local central network type: … “ and
- Type and successful validation of central type
“Validating network type ... -> OK”
“Local monitors ... -> OK” - Number of local monitors “Local central network IP address: …” - IP address of the local central “Local central network port number: …” - Port number of the local central “Local central network ID: … ” - Network ID of the local central “Interconnecting Network …'s IP: …” - The IP address of the central interconnect to the local one “Interconnecting Network …'s Port Number: …” - The port number of the central interconnect to the local
one
“Validating IP: … -> OK” - Successful validation of the IP address “Validating Interconnection with Network … ->
- Successful validation of interconnection with other DOC
OK“
“Validating Interconnection of network … ->
- Interconnection of that DOC is correct
OK”
Check view configurations (MONVIEWS.AIC)
Requirements:
network.aic and network.opt must exist in D:\central\setup and well-defined.
monviews.aic and monviews.opt must exist in D:\central\setup and well-defined.
Error message Reason
“Loading \CENTRAL\SETUP\NETWORK.OPT
failed - line=…: …. “
“Loading \CENTRAL\SETUP\NETWORK.AIC
failed - line=…: …. “
“Loading \CENTRAL\SETUP\MONVIEWS.OPT
failed - line=…: …. “
“Loading \CENTRAL\ SETUP\MONVIEWS.AIC
failed - line=…: …. “
“No network defined” - No NETWORK defined in NETWORK.AIC “No view configuration defined” - No view configuration (VIEW_CONFIG) defined in
“View configuration not found” - VIEW_CONFIG defined, but empty “Invalid page object in view configuration” - PAGE is defined, but probably empty “Invalid monitor object” - MONITOR is defined, but empty “Central station should show local monitors
in Default view configuration”
“Monitors in the same page should belong to
the same network”
See Check central supervising configuration
See Check central supervising configuration
MONVIEWS
- Central type DOC should not show remote monitors in the
Default view configuration
- The monitors that belong to the same page should belong
to the same DOC network
76
Document No. 8000633-1
Page 83
Error message Reason
Datex-Ohmeda S/5 Central, ViewStation and Network
“Network ID … not found”
- Monitor defined to belong to a DOC, but the network not
found in NETWORK.AIC by ID, probably removed or the monitor is moved to other network.
“Monitor … not found” - The monitor defined in a view configuration, but do not
belong to the network with the ID specified in NETWORK.AIC, probably wrong NETWORK ID or the monitor has been removed from the NETWORK.AIC
“Page … contains no monitor” - Page, if defined, should contains at least one monitor “No page defined in view configuration”
- View configuration, if defined, should contains at least one
page
“Default view configuration not found” - There is no Default configuration in MONVIEWS.AIC
Success Meaning
“Loading Central Network Setup … ->
Loaded”
“Loading Central View Setup … -> Loaded” - File loading successful “Processing view configuration … ” - A view configuration is under processing “Processing page … “ - A page is under processing “Monitor … ID …, Group ID …, Network ID … -
> OK.”
- File loading successful
- Successful validation of a monitor and its name, plug id,
group id and network id.
Check general configuration ( SETUP.AIC)
Requirements:
SETUP.AIC
and SETUP.OPT must exist in D:\CENTRAL\SETUP and well defined.
Error message Reason
“Loading \CENTRAL\PROGRAM\SETUP.OPT
failed - line=…: …. “
“Loading \CENTRAL\PROGRAM\SETUP.AIC
failed - line=…: …. “
“No setup defined” - SETUP.AIC probably empty “AS/3 network board is not specified in general
setup”
“Lookup AS/3 Network Adapter … in the Registry
-> Not found”
“The NDISHOOK driver isn't in the path” - NDISHOOK.SYS not found in D:\CENTRAL\PROGRAM “User probably doesn't have sufficient access
rights to load drivers.”
“The NDISHOOK.SYS driver is 'partially'
unloaded”
- See Check central supervising configuration
- MONGATE_BOARD option not found
- The adapter is not defined, not installed, is used for
TCP/IP or not SMC8432 or not 3COM EtherLink XL
- Confused error, probably Check current account at
“Load and unload device drivers“ failed.
- Run W32NNTSC.EXE (D:\CENTRAL\MONITORGATE) to
remove the driver, also check that W32NNTSC.EXE should be run as a STARTUP_EXE in CENTRAL.AIC
77
Document No. 8000633-1
Page 84
Technical Reference Manual
Error message Reason
“Testing AS/3 Network Adapter & Cable ->
- Adapter installation might failed
failed”
“Network cable probably not connected” - No traffic on the AS/3 network “Time synchronization option is not defined” - TIME_ADJUSTMENT_ENABLED option is not defined “Time server IP is not defined” - TIME_ADJUSTMENT_ENABLED is TRUE, but
TIME_SERVER_ADDRESS is not defined
“Validating IP: …” - Validating Time Server IP failed. See Check status of
TCP/IP and Windows Socket (WinSock) 2.0
“Server does not support Time Protocol” - Time Server does not support Time Protocol (TCP/IP,
port 13)
“Server refuses Time Protocol Request” - Time Server refuses Time Protocol Request (TCP/IP,
port 13, connection establishing request)
“Time Protocol Request failed” - Connection to Time Server success, but no answer “DEFAULT_PRINTER is empty” - DOC system should have a default printer “DEFAULT_RECORDER value must be
RECORDER”
“Default printer connected to no port” - Default Printer needs to connected to a valid port
- DEFAULT_RECORDER option should have the value
RECORDER.
“No printer installed” - No printer installed on the machine “Default printer … not found” - Described printer not installed or incorrect printer port “… has no name” or “… has no port” - Valid printer should have name and port “… and … option has the same value” - SINGLE_VIEW_WAVE and NUMBERFIELD options
should have different values or OFF
“… user not defined” - Service user and R&D user should be defined “… no password to ...” and “invalid … password” - Password missing or invalid
Success Meaning
“Lookup AS/3 Network Adapter \Device\El90x2
in the Registry ... -> Found.”
“Open AS/3 Network Adapter \Device\El90x2 ...
-> OK.”
“AS/3 Network Adapter Address
00.50.04.79.38.6F -> OK”
- Monitor network adapter device name found in registry
- Monitor network adapter opened successfully
- Found valid monitor network adapter Ethernet address
“Testing AS/3 Network Adapter & Cable ... ->
- Network adapter and cable checking successful
OK.”
“Close AS/3 Network Adapter \Device\El90x2 ... - Network adapter closed
78
Document No. 8000633-1
Page 85
Datex-Ohmeda S/5 Central, ViewStation and Network
Success Meaning
-> OK.”
“Time synchronization OFF” - Time synchronization with a time server disabled “Check Path <path name> = <path> -> OK.” - Setup path checked and exists “-> Warning: <path name> = <path> not exist.” - Setup path checked but not found; probably not
needed since only warning given
“Default recorder <recorder name> ... -> OK.” - Recorder definition found “Check Path REC_SPOOL_DIR=<recorder data
- Recorder data spool directory found
spool directory>... -> OK.”
“Default printer <printer name>... -> OK.” - Default printer definition found “Printer PRNN_NAME attached to port <printer
port name> ... -> OK.”
“Check if Recorder type is <valid printer type>... -
- Printer N connection to port < printer port name > has
been defined
- Recorder definition is valid
> OK.”
“Check if Recorder port is LPT1 ... -> OK.” - Recorder port definition is valid “Check Service User Account ... -> OK.” - Service user account definition is valid “Check Service Password ... -> OK.” - Service user password definition is valid “Check Exit Password ... -> OK.” - Exit password definition is valid “Check Exit Password ... -> OK.” - Exit password definition is valid
Check system licenses
Requirements:
SMART-KEY or licensed password.
Error message Reason
“The system has no license” No Smart-Key installed
Success Meaning
“Number of licensed monitors is N -> OK.” - Found Smart Key “Check number of licensed monitors ... -> OK.” - Number of licenses is valid
79
Document No. 8000633-1
Page 86
Page 87

10 TROUBLESHOOTING

If the monitor is connected to the monitor network, the symbol is displayed
on the monitor’s screen. NOTE: In Compact Monitor, the battery operation symbol overrides the network symbol.
If none of the monitors can be connected to the network, the fault is probably related to the
cabling between the Central's network board and hub.
If only one monitor can be logged into the network, the fault is probably related to the smart
key: the smart key is probably faulty, or not inserted.
If all the monitors, except one, can be connected to the network, the fault is probably
related to the cabling between the monitor and hub, or in the monitor, network board or hub.
Use also DOCCheckup utility to find if the configuration is properly done, see page 68.
The steps for the network troubleshooting are outlined in a flowchart in Table 10-1.
Datex-Ohmeda S/5 Central, ViewStation and Network

10.1 Compatibility limitations

You can only connect the monitors with the software S-XXX95, or newer to the network with the S/5 Central. If there are monitors using the record keeping software, they must be of the same revision.
You can only admit a patient from the Central (give the patient ID) with monitors equipped with a S­XXX97 software, or newer.
If the monitors are equipped with L-ARK99(A) or S/5 ARK software license, they require the S­CNET99, ver.1.1, or newer, network software.
For monitor-to-monitor communication over different networks, the monitor software level S-XX99 is required.
To enable connecting ViewStations to the system, the Central requires the software S-CNET99 Ver
1.2, or newer. NOTE: Japanese monitors (softwares S-ANE97-31 or S-ICU97-31) must not be connected to the
same network with other monitors of any other language or revision, because their incompatibility may cause errors in the network communication.
81
Document No. 8000633-1
Page 88
Technical Reference Manual

10.2 Network troubleshooting flowchart

Table 10-1 General troubleshooting
Network communication problem?
A) All monitors of a network affected?
B) Some monitors of the network affected?
Examine the network architecture and topology using the original plan documentation as a reference. Confirm that the network configuration is appropriate (use DOCCheckup document)
Check the Mon/IC -Net cable and its connecetions. Replace if necessary
Check the Central's/ViewStations patch cable and its connections. Replace if necessary.
Test the horizontal cabling using a cable scanner.
Check the Mon-Net cable and its connections. Replace if necessary.
Check that the monitor's patch cable and its connections. Replace if necessary.
Check that the smart key is installed and of right kind (defines the number of the monitors, max. 32). Also confirm that the com port of the computer is functioning properly.
All monitors of a network affected, proceed A)
Only one or some monitors affected, proceed B).
Proceed to Central's/ViewStation's troubleshooting, see the table below
Proceed to monitor troubleshooting, see the corresponding Technical Reference Manual.
Test the horizontal cabling using the cable scanner.

10.3 Central/ViewStation troubleshooting chart

For the Central or ViewStation troubleshooting refer to the table above.
Table 10-2 Central/ViewStation Troubleshooting
What if… Maybe Try
Central/ViewStation
Nothing functions Central is not connected to wall
socket.
Keyboard keys have no effect on the Patient Screen.
The keyboard cable may be loose or defective.
The controller board may be defective.
Unplug and replug the power cord and check that it is intact.
Confirm that the Keyboard-IC Cable is firmly connected. If necessary, replace the cable.
If necessary, replace the board.
82
Document No. 8000633-1
Page 89
Datex-Ohmeda S/5 Central, ViewStation and Network
What if… Maybe Try
Function keys have no effect on the screen.
Alphanumeric keys have no effect on the screen.
ComWheel has no effect on the screen.
The cable to the membrane keypad may be loose.
The membrane keypad may be defective.
The controller board may be defective.
The cable to the controller board may be loose.
The alphanumeric keyboard may be defective.
The controller board may be defective.
The cable to the controller board may be loose.
The rotary wheel may be defective. If necessary, replace the wheel.
The controller board may be defective.
Confirm that the cable is firmly connected.
If necessary, replace the keypad.
If necessary, replace the board.
Confirm that the cable is firmly connected.
If necessary, replace the keyboard.
If necessary, replace the board.
Confirm that the cable is firmly connected.
If necessary, replace the board.
The system will not start, or halts immediately after startup.
Patient Screen doesn’t function. The display power cable or the
Patient Screen includes distortion
The configuration settings may be defective.
The network board may be loose or defective.
display data cable may be loose or defective.
The display controller board(s), B­CDISP, may be loose or defective.
During software installation the Network Manager configuration is selected.
Patient Screen may be defective. If necessary, replace the display.
The display controller board may be defective.
Check CENTRAL.LOG file for startup report. Revise the network setup. If necessary, restore the default setup.
Confirm that the network board is firmly connected. If necessary, replace the board.
Confirm that the display power cable and the display data cable are firmly connected. If necessary, replace the cables.
Confirm that the board is firmly connected. If necessary, replace the board.
Reinstall the Network software (S­CNETXX or S-VNETXX).
If necessary, replace the board.
83
Document No. 8000633-1
Page 90
Technical Reference Manual
What if… Maybe Try
Patient Screen includes regular stripes.
Status Screen doesn’t function. The display power cable, or the
Record Keeping
The record keeping menus cannot be found, or they are in the wrong language, or in two languages.
For the monitor troubleshooting, refer to its documentation.
The Display Controller, B-CDISP may be defective.
display data cable may be loose or defective.
The display power cable, or the display data cable may be loose or defective.
Menus are not in the specified directory.
If necessary, replace the board.
Confirm that the display power cable and the display data cable are firmly connected. If necessary, replace the cables.
Confirm that the display power cable and the display data cable are firmly connected. If necessary, replace the cables.
Start the Network Setup software and open the General Setup window. Ensure that the ‘Path 1 to Menus’ indicates a valid directory.
NOTE: ‘Path 1 to Menus’ and ‘Path 2 to Menus’ should be similar unless ‘Path 1 to Menus’ indicates a directory in the office network.

10.4 Error messages

Following messages may be displayed on the Patient or Status Screens.
Table 10-3 Screen messages
Message Cause Solution
Disk is almost full The patient data disk is
almost full. The warning will be given approximately one week in advance.
Printer error (with an explanatory note)
The printer/recorder is out of paper.
Printer is off-line.
The printer is not ready or cannot be accessed.
Printer/recorder is not connected properly.
Recorder’s writing head is cold or hot
The value of the patient data storage time has to be reduced in the setup. Contact the network administrator.
Add paper.
Ensure that the printer power cord is properly connected and the power is ON, and on-line.
Check that the cables are connected properly and that the power is ON.
Check the connecting cable
Wait for the recorder warm up or cool down.
84
Document No. 8000633-1
Page 91

10.5 Technical assistance

For technical assistance, do not hesitate to consult your authorized Datex-Ohmeda distributor. Be prepared to describe the problem in detail and to provide the distributor with the following information.
Type and serial number of the Central or ViewStation. The serial number can be found on a
sticker on the small lever on the front panel.
Serial number of the Display Controller Board(s), B-CDISP. The serial number can be found
on a sticker on the edge of the board.
Printouts of NETWORK.AIC, SETUP.AIC, MONVIEWS.AIC,
STATSCRN.AIC, STARTUP.AIC, CENTRAL.AIC and MONITORS.CFG.
The files can be printed through service pages of the Status Screen.
Printouts of version data from networked monitors through service pages of the Status
Screen.
Error logs, especially CENTRAL.LOG, through service pages of the Status Screen.
Datex-Ohmeda S/5 Central, ViewStation and Network
85
Document No. 8000633-1
Page 92
Page 93

11 PLANNED MAINTENANCE

Perform the following steps at least once a year and always after service. You need an allen wrench and documents for the disk scanning program,
Step Procedure
Check that all the components of the Central/ViewStation and displays are clean and undamaged. Make sure that the cooling holes are not obstructed.
Check that all Mon/IC-net and Mon-net cable connections are secure, and that no cables are pulled taut. Make sure that the wallboxes are clean and undamaged.
Check that the system starts up normally, and that all active network devices register to the network.
Check that the display contrast and brightness, as well as the vertical and horizontal position, are adequately adjusted. For readjustments, please refer to the display’s documentation.
Datex-Ohmeda S/5 Central, ViewStation and Network
Check that the Central Keyboard keys function properly by pressing the keys one by one.
Check that the ComWheel functions properly by turning the ComWheel clockwise and counterclockwise.
Check that the system clock shows the correct time. For readjustments, please refer to the section Installation of this manual.
Print the service pages and check that the contents of the pages is OK.
Remove all temporary files that might be stored on the hard disk.
Back up
STARTUP.AIC, CENTRAL.AIC MENUBLD.AIC file if it is stored on the hard disk.
Run a disk scanning program. For further instructions, please refer to documentation delivered with the computer.
Shut the system down, and disconnect it from the power outlet. Open the system unit cover and remove dust accumulated around the power supply unit.
For planned maintenance of printers, power supplies, and hubs, please refer to the manuals delivered with the device.
SETUP.AIC, STATSCRN.AIC, MONVIEWS.AIC,
and
NETWORK.AIC
files. Also back up
87
Document No. 8000633-1
Page 94
Page 95
Datex-Ohmeda S/5 Central, ViewStation and Network

PART II Product Service guide

12 DATEX-OHMEDA PRODUCTS

12.1 Network Software, S-CNET99 and ViewStation Software S-VNET99

The Network Software, S-CNETxx, and the ViewStation Software, S-VNETxx, consist of several functional components.
12.1.1 Software architecture
Not available in S-VNETXX
k
r
r
o
t
o
Services
i
w
n
t
o
e
M
n
Menu
Files
Menu
Patient Screen
Status Screen
Monitor
Gate
TCP/IP Network
Server
&
Patient
Logistics
Patient
Data
Figure 12-1 General architecture of S-CNETxx and S-VNETxx software
The main components are:
Monitor Gate component manages communication between the software components and
monitors, as well as between components themselves.
Patient Screen component provides user interface for patient data and alarms.
Status Screen component provides character-based user interface for system status.
Menu Server and Patient Logistics is responsible for storing patient data in disk drive,
retrieving stored data to monitors, and providing menu descriptions for anesthesia record keepers when needed. Not in S-VNETxx.
Services component:
Bed-to-bed service provides monitor-to-monitor communication service for monitors. Not in S-VNETxx.
Printer services provide monitors access to global printing resources Mode services are used for storing and retrieving monitor settings. Not in S-VNETxx.
Document No. 8000633-1
89
Page 96
Technical Reference Manual
12.1.2 Directory Structures
When the Central/ViewStation is installed at the factory, the software is located in the D:\Central directory, and more specific directory structure can be found in the readme.txt file.
NOTE: Be sure to edit only the files that are meant to be edited.
Configuration and other general folders
D:\ CENTRAL\
B2BMANAGER\
GSPUI\
MENUS
MENUSERVER\
MODESERVER\
MONITORGATE
PRINTERSERVICES\
PROGRAM\
SETUP\
STATUSSCREEN\
TOOLS\
The folder for the monitor-to-monitor communication component files.
The folder for the Patient Screen files.
See more specific explanation below.
The folder for the record keeping menu server
The folder for the user mode files.
The folder for communication between the components
The folder for the printer service files.
The folder for the network software program files.
The folder for files created/needed during system setup.
The folder for the Status Screen files.
The folder for tool program files.
90
Document No. 8000633-1
Page 97
Patient data folders
The following files and directories are created when the network software is started for the first time after setup. If these default directories are not used, the new directories must be created before the network setup.
If the network will communicate with the Datex-Ohmeda IM network, the following files and directories must be copied to the IM server.
Datex-Ohmeda S/5 Central, ViewStation and Network
D:\ CENTRAL
\
PATDATA\
111\ 112\ 113\
111\52302T01.LCK
111\52302T01\
PATDATA\ARC\
CASEBCFG\
CASEBNDX.DAT
Default directory for patient data files and pre-operatory data files.
Examples of monitor location level directories. There is one directory for each monitor location. The name of the directory is the monitor ID plug number.
Lock file for case 52302T01. Indicates that the case directory is reserved for the monitor with the ID number 111, which is saving data to the corresponding case directory. The lock file is removed when the case is closed. Case directory. The data of one case/monitoring period is saved to a single directory. The case directory can, for example, contain the following files: 52302T01.EVT Contains record keeping events. 52302T01.PAT Contains patient identification information. 52302T01.PHY Contains physiological data. 52302T01.PH1 Contains physiological data. 52302T01.PH2 Contains physiological data. 52302T01.PH3 Contains physiological data. 52302T01.STA Contains case status information (start and end times of the case). ACT_CASE.LOG Contains information on all subcases in this case directory.
The structure of this directory corresponds to the patient data directory.
The cases older than setup value ‘Patient list time range’ and preoperative cases (cases containing no trend data) older than ‘Preop list time’ are moved to this directory.
Directory for archived record keeper configuration data needed during post-processing.
Contains information on record keeper configuration data in CASEBCFG\ -directory.
91
Document No. 8000633-1
Page 98
Technical Reference Manual
Menu folder
D:\ CENTRAL\
MENUS\
ENG\ENGUFRM.0A1 ENG\ENGED1.TXT GER\GERUFRM.0A1 GER\GERED1.TXT FRE\FREUFRM.0A1 FRE\FREED1.TXT DUT\DUTUFRM.0A1 DUT\DUTED1.TXT ITA\ITAUFRM.0A1 ITA\ITAED1.TXT SPA\SPAUFRM.0A1 SPA\SPAED1.TXT DAN\DANUFRM.0A1 DAN\DANED1.TXT
Default menu file set for a configuration without S/5 Record Keeper.*
Directory for the S/5 Record Keeper menus
* NOTE: If user-configured menus (also called casebase) are used, all these files should be removed.
Mode folder
D:\ CENTRAL\
FIN\FINUFRM.0A1 FIN\FINED1.TXT SWE\SWEUFRM.0A1 SWE\SWEED1.TXT NOR\NORUFRM.0A1 NOR\NORED1.TXT Z_UNITS.TXT D_UNITS.TXT
MODESERVER\ MODES\
\AS3ARK\v800\G1 , G2, G3...
\CS3\v700\...
A folder where different user modes are saved
92
Document No. 8000633-1
Page 99
Datex-Ohmeda S/5 Central, ViewStation and Network

12.2 Start-up Procedure

12.2.1 After having exited the program
When the computer is on and the operating system is running:
Double click the Datex-Ohmeda Central icon on the screen.
12.2.2 After having switched off the Network Computer
When the computer has been turned off after exiting program, to restart:
1. Switch on the Network Computer by pressing the
2. Switch on the display units (Patient Screens, Status Screen) and the printer.
3. In the lower left corner of the screen select Start - Programs - D-O Network - Central. During the start-up, following steps are taken:
1. Checking the system configuration
are listed to the page, see appendices.
2. Essential parts of the record keeping configuration are saved to the
for postprocessing applications, and if the setup option ‘Path 1 to menus’ differs from the ‘Path 2 to menus’, the configuration files are copied from the first path to the second one.
CENTRAL.LOG file, which can be viewed and printed in the File View

12.3 System/Installation menu

Most typical installation settings are collected to the installation menu of the Central’s software, so that they can be changed after initial installation on the run, that is, without switching off the network software. These changes are valid only as long as the software is running.
1. Press
2. Give the password (default is BBBBB, this may have been changed during installation). In the menu, you can save the desired 'Normal screen' to which the display returns after the
Normal Screen key is pressed, select the printer to which the Central prints of those configured
into the system and access Alarms Setup and Recorder Setup submenus.
Setup key and select System and Installation.
Power ON/OFF switch.
. If a critical error occurs, the system will not start. Errors
CASEBFG -directory
System
Installation
Save Normal Screen
Select Printer Net1
Alarms Setup
Recorder Setup
Previous menu
93
Document No. 8000633-1
Page 100
Technical Reference Manual
12.3.1 Alarms setup
1. Select Alarms Setup.
2. Set audible alarms on or off, select which priority alarms are shown in the Patient View, and
adjust the alarm volume.
3. Confirm the changes by selecting Previous Menu or by pressing Normal Screen.
12.3.2 Recorder setup
1. Select Recorder Setup.
2. Set autoalarm recording on/off, select paper speed, select the strip length (20 or 30
seconds) for alarm recording, and the length of the manual recording (30 s or continuous).
94
Document No. 8000633-1
Loading...