B.Braun Infusomat FM Error code

İNFUSOMAT FMS ERROR CODE LİST
100 defective RAM memory 101 UMOT cannot be switched on 102 UMOT still switched on despite overvoltage 103 UMOT still switched on despite MOTEIN=0 104 UMOT still switched on despite undervoltage 105 ON/OFF key pressed longer than 14 sec 106 defective air sensor (calibration value?) 107 defective program memory 108 defective program flow 109 different number of pump head cycles 110 different keyboard gaps between 80c838 and 68hc11 111 different program versions between 80c535 and 68hc11 112 defective program flow 113 testbit!=0 out of switch-on test 116 defective program memory - text 117 defective program memory - text does not match with program 118 reset during active operation 119 defective ROM
FF01 dummy for test FF02 battery not present / missing battery current FF03 defective RAM memory FF04 defective program memory FF05 defective program memory FF06 calibration data error from EEPROM FF07 pump head cycle not plausible FF08 failure / inaccuracy of system clock FF09 failure 100msec system clock FF10 reset during active operation FF12 no dynamic pressure sensor signal (EDB) FF16 defective membrane keyboard (from IFMC02003 on)
İNFUSOMAT FM ERROR CODE LİST
100 defective RAM memory U13 101 UMOT cannot be switched on 102 UMOT still switched on despite overvoltage 103 UMOT still switched on despite MOTEIN=0 104 UMOT still switched on despite undervoltage 105 ON/OFF key pressed longer than 14 sec 106 defective air sensor (calibration value ? ) 107 defective LED display 108 defective LED display 109 defective LED display 110 defective LED display - RDTS 111 defective LED display 112 defective LED display 113 defective LED display 114 defective program memory U10, U11, U15 115 defective program flow 116 MS has frequency without drive 117 different number of pump head cycles 118 different keypad gaps between 80c535 and 68hc11 119 mode different in 68hc11 120 changed operating mode 121 defective tube adapter 122 different program versions between 68hc11 and 80c535 123 defective local program flow 124 testbit=0 out of switch-on test 125 PR is not active though 80c535 active PR and 68hc11 active PR 126 PR is active though 80c535 not active PR and 68hc11 active PR 127 text not loaded / defective program memory U10, U11, U15 128 text does not match with program / defective program memory
U10, U11, U15
F01 dummy for test F02 battery not present / missing battery current F03 defective RAM memory U17 F04 defective program memory U21, U22, U31 F05 defective program memory U21, U22, U31
Loading...
+ 1 hidden pages