DIALOG
®
EVOLUTION
+
HEMODIALYSIS SYSTEM FEATURES WITH ADIMEA™ DOSE MONITOR DIALOG
®
EVOLUTION
+
HEMODIALYSIS SYSTEM FEATURES
Beginning of treatment Middle of treatment End of treatment
Adimea™
Real-time Dialysis Dose Monitoring System
Provides a continuous, real-time measurement of dialysis dose (Kt/V or URR). Ultraviolet light is used to measure the
change in the concentration of uremic substances in the spent dialysate. Dialog+ graphically displays the Adimea curve, and
the user can interpret if the target Kt/V will not be achieved, and, thus, interventive actions may be taken to improve the treatment
outcome within the prescribed dose.
Adimea can be used to measure the dialysis dose for every treatment. With this information, the user can take steps to ensure the
prescribed dose is achieved.
No additional disposable is required. Only the patient’s pre-dialysis weight and target Kt/V value need to be entered.
BBraunUSA.com/Adimea
Streamline® Airless System Set
The airless pressure-transmitting assembling is designed to eliminate
blood-air interface
Simplied tubing design
Locksite® needleless access
Designed for Comfort
Automatic Blood Pressure Measurement (ABPM)
The ABPM in the Dialog system can measure blood pressure during
ination which may add to a patient’s comfort during therapy.
Diacap® Ultra Filter
Produces highly pure dialysate to reduce the risk of contaminated dialysis uid and its
associated complications. Diacap® Ultra is the recommended dialysate lter for use on
the Dialog machine.
Bicarbonate Cartridge Holder
Allows the use of the Solcart B® bicarbonate cartridges for dispensing of bicarbonate.
Patient Therapy Card Reader
The therapy card stores the patient’s machine settings to allow for quick and
accurate setup. Adimea data is also stored on the Therapy Card so past Adimea
results can be viewed while the patient is being treated. Adimea Kt/V trends can be
reviewed so the user can see how the patient’s dialysis dose has changed over time.
Advanced Technical Support & Maintenance Mode (TSM)
Allows technician to test internal components
May reduce troubleshooting and preventive maintenance time
Provides Service History Information
Graphical display provides overview of sub-system components
Conguration screen allows technician to set default treatment
values and modes
Advancement in Technology
Convenient circuit board placement
Access to internal components
Dual control system designed to provide redundancy and security
Heat exchanger and heating system for improved power eciency
Disinfection Cycles
Chemical, Thermal, Decalcication, Citric/Thermal Cycles with
programmable dwell times and user-denable short programs.
Concentrate wands are loaded into machine and disinfected during
the selected cycle.
User Interface Characteristics
High Resolution (1024 x 768) Touch Screen combined with o
screen Hard Keys.
Icon Driven User interface
On-screen HELP functions
Automatic Dialyzer Rinsing Program
On-Line Trending Graphs
On-Line Technical Data
Dening Features
User-denable Proles for UF, Na, Bicarb, Dialysate Flow,
Temperature, and Heparin
Patient Therapy Card Reader for patient-specic parameters for use
in subsequent treatments
Auto On/O timer - machine will automatically turn o after
completion of a disinfection cycle and automatically turn on at the
programmed time
BIC Holder for Solcart B® cartridges
Quieter Machine
The Dialog+® Hemodialysis System is equipped in the following way to
help reduce sound emissions:
Pumps are mounted on rubber brackets.
Machine fan does not cycle during treatment; only during
disinfection cycles.
Reducing sound emission provides a quieter environment for both
the patient and the dialysis sta during treatment.
Patient care begins with Dialog+®
Icon-Based Machine Operation
Standby Mode
Stops dialysate ow at the end of preparation
to prevent waste of water and concentrates.
Particularly useful when a patient is not
immediately ready for treatment.
Technical Overview
An overview of machine’s sensors, measuring
devices, and valves is available in real time.
This allows the technician to troubleshoot
possible machine issues while the problem is
actually occurring. This screen is designed to
allow for eective phone support.
Auto On-and-O
Dialog+ can be programmed to automatically
turn on and start pre-programmed selections
such as the disinfection cycle mode and start
preparation mode for therapy.
Proling Capability
UF, Na, Bicarb, Dialysate Flow, Temperature,
and Heparin Parameters can all be proled.
Event Timer
An individualized event, such as a time to
give patient medication, can be programmed
to sound a reminder alarm during treatment.
Call History and Trend Data
Graphic representation of dierent
parameters of therapy course. Can hold
up to 20 previous treatments.
List of Stored Adimea™ Curves
Stores data for up to 12 therapies without
the patient card reader, and up to 50
therapies with the card.
For more information, contact a B. Braun Renal
Therapies Division representative at 1-800-848-2066.
BBraunUSA.com/Dialog

Technical Data
Physical size of machine:
Dialysate Temp. Range:
Proling:
Conductivity Range:
Proling:
Dialysate Flow Range:
Proling:
Ultraltration Range:
Proling:
Extracorporeal Circuit Blood
Flow Range:
Settings:
Heparin Flow Range:
Settings:
Proling:
66 in. H x 20 in. W x 25 in. D
33 - 40° C
User-denable, Linear, Exponential
12.5 - 16 mS/cm
Total Sodium and Bicarbonate
User-denable, Linear, Exponential
300 - 800 ml/min
User-denable, Linear, Exponential
0 - 4,000 ml/hr
User-denable, 20 Pre-set Proles
50 - 600 ml/min
10 ml/min increments
1 - 10 ml/hr
0.1 or 0.5 ml/hr increments
User-denable, Linear, Exponential
Venous Pressure Range:
20 - 390 mmHg.
Pre- or Post-pump monitoring
Arterial Pressure Range:
-400 - +400 mmHg.
Air Detection Red Sensor: Ultrasonic sensor on venous line initiates
an automatic heparin bolus at the
beginning of dialysis when blood is
detected in extracorporeal circuit. Stops
blood pump at the end of dialysis. Venous
line — at air detector