This document contains proprietary information protected by copyright. All rights are reserved. No part of this
document may be photocopied, reproduced, or translated into another language without the prior consent of Agilent
Technologies. The information contained in this document is subject to change without notice.
Copyright 1995 -1998, 2000 Agilent Technologies, Inc.
Windows is a registered trademark of Microsoft Corporation.
Warranty
This Agilent Technologies software product is warranted against defects in materials and workmanship for a period
of 90 days from date of delivery. During the warranty period, Agilent Technologies will, at its option either repair or
replace parts which prove to be defective.
Agilent Technologies makes no express or implied warranty of any kind, including, but not limited to the implied
warranties of merchantability or fitness for a particular purpose, with regard to the program material contained
herein. Agilent Technologies shall not be liable for incidental or co nsequential damages in connection with or arising
out of the furnishing, performance, or use of this software.
Use of the supplied manual and software is restricted to Agilent ac source products only. The software is copyrighted
and may not be copied except for archival purposes, to replace a defective copy, or for program error verification.
Agilent Technologies warrants that this software designed for use with a personal computer, will execute its
programming instructions when properly installed on that personal computer. Agilent Technologies does not warrant
that the operation of the personal computer , software, or ac source will be uninterrupted or error free.
Limitation of Warranty
The foregoing warra nty shall not apply to defects resulting from: misuse, unauthorized modifica tion, operation or
storage outside the environmental specifications for the product, in-transit damage, improper maintenance, or defects
resulting from use of non-Agilent software, accessories, media, or such items not designed for use with the product.
Revisions
Revision codes on the Agilent 14761A HFTS software indicate the current revision. Minor changes to the software
such as bug fixes usually do not require a change to the manual. Therefore, the revision number of the software may
be higher than the software revision number shown below, yet the information in the manual still applies.
Software changes that require a change to the manual will be accompanied either by a new edition of the manual or
an update packet documenting the changes. In that case, the software revision number shown below will be updated
to agree with the revision number of the software.
This manual applies to software revision... A.05.03
Printing History
The manual printing date indicates the current edition. The printing date changes with each new edition or update.
Update packets or change sheets may be issued between editions to correct or add information. Minor corrections
incorporated at reprint do not cause a new edition.
August, 1995.........First Edition
October, 1995........Update 1
February, 1997.......Second Edition
June, 1997..............Third Edition
December, 1998......Fourth Edition
May, 2000...............Fifth Edition
2
Table of Contents
Notice2
Warranty2
Limitation of Warranty2
Revisions2
Printing History2
Table of Contents3
1 INTRODUCTION7
How to Use this Manual7
The Agilent 6800-Series AC Power Source/Analyzer8
The Agilent 14761A Harmonic/Flicker Test System Software8
EN 61000-3-2 and EN 60555 Part 2 Regulations9
Compliance Testing Implementation9
EN 61000-3-3 Regulation10
Compliance Testing Implementation10
2 INSTALLATION11
Install the Software11
Connect the Equipment11
Installing Interface Card Drivers for Windows 3.1 or 3.11 Systems12
Installing Interface Card Drivers for Windows 95 Systems12
Installing Interface Card Drivers for Windows NT 4.0 Systems12
Verify the Configuration12
Default Settings14
Getting Around in the Application14
Drop-down Menu14
Menu Tabs16
3 SETTING UP THE TEST17
Select a Test Type17
Create a Template19
Select the Test Setup Options for Quasi-stationary or Fluctuating Harmonics20
Standard Test Options20
Advanced Test Options21
Test Termination Options23
Select the Test Options for Voltage Fluctuations24
Standard Test Options24
Advanced Test Options25
Test Termination Options26
4 RUNNING THE TEST27
To Run a Pre-test for Quasi-stationary or Fluctuating Harmonics27
Pre-Te st Summary27
Pre-Test Power Statistics29
Pre-Test Source Harmonics30
To Run a Pre-test for Voltage Fluctuations31
Pre-Te st Summary31
Pre-test Measurements32
To Run a Test33
Test Status Information35
Validation Mode36
3
5 VIEWING TEST DATA37
Navigator Toolbar37
Viewing Quasi-stationary and Fluctuating Harmonics38
Graph Display38
Table Display41
Time-Series Display42
Statistics Display45
Probability Display (for Quasi-stationary or Fluctuating harmonics)46
Viewing 2.5 Minute Window Failures47
Viewing Voltage Fluctuations49
Pst Display49
Probability Display (for Voltage Fluctuations and Flicker)50
Distribution Display52
RMS Display53
Flicker Display54
Editing the Graph Attributes56
Copying Graphs and Tables to the Clipboard57
Using the Print Screen keyboard key57
Using the Copy Commands57
Viewing Reports58
Causes for Non-compliant EN 61000-3-2 or EN 60555-2 Tests58
Causes for Non-compliant EN 61000-3-3 Tests58
Short Form Report59
Long Form Report60
Remarks Report63
6 SEARCHING FOR SPECIFIC TEST DATA65
Searching for Data While the Test is Running65
Searching for Data After the Test has Completed65
Using the Span Control66
Using the Zoom Control67
Obtaining Detailed Failure and Error Information68
7 PRINTING69
Printing Graphs and Tables69
Using the Print Pre-test command69
Using the Print Graph/Table command69
Printing Reports69
From the File menu69
From the Report window70
Printing Reports to a File70
A SPECIFICATIONS73
Supported or Referenced EN 61000-3-2 and EN 60555 Part 2 Standards73
Supported or Referenced EN 61000-3-3 Standards73
PC Requirements73
Supported GPIB Interfaces73
Supported Equipment73
Equipment Specifications (IEC Mode)74
B GLOSSARY77
4
C IEC MODE COMMAND SUMMARY81
Introduction81
Using the SENSe:CURRent:ACDC:RANGe command81
CALCulate:INTegral:TIME83
CALCulate:SMOothing83
CALCulate:LIMit:UPPer84
FORMat85
FORMat:BORDer86
MEASure:ARRay:CURRent:HARMonic?87
MEASure:ARRay:VOLTage:FLUCtuations:ALL?88
MEASure:ARRay:VOLTage:FLUCtuations:FLICker?90
MEASure:ARRay:VOLTage:FLUCtuations:PST?91
SENSe:CURRent:PREFerence92
SENSe:WINDow92
SYSTem:CONFigure93
D CLASS DETERMINATION95
Class A Device Selected96
Class B Device Selected97
Class C Device Selected98
Class D Device Selected99
EN 60555 Part 2 Regulation Selected100
INDEX101
5
1
Introduction
How to Use this Manual
This manual describes the operation of the Agilent 14761A Harmonic/Flicker Test System (HFTS)
software when used in conjunction with the Agilent 6800-Series AC Power Source/Analyzers. Its
primary function is as a reference manual. If you have a question about a specific screen or how to
perform a certain task, simply turn to the appropriate section of the manual. The manual is organized
according to the procedure that you would follow if you were to run a compliance test. This manual
assumes that you are familiar with the EN 61000-3-2 and EN 61000-3-3 regulations and their
requirements. It also assumes that you are familiar with operating a personal computer in a Microsoft
Windows environment.
The manual is organized as follows:
Chapter 1 provides an overview of the regulations and how compliance testing to these regulations is
implemented with the Agilent 14761A HFTS software.
Chapter 2 describes how to install the software and get it running. It describes some pitfalls to avoid so
that operation of the software will be glitchless and error-free.
Chapter 3 describes what you need to do to before you can run a test. There is certain information that
you need to provide about the Device Under Test as well as setting up the Agilent 6800-Series AC Power
Source/Analyzer.
Chapter 4 explains what happens when the pre-test and the compliance test is run.
Chapter 5 describes how to view the test data both while the test is running and after the test completes.
Chapter 6 describes the software tools available to find and display specific test data.
Chapter 7 explains how to print graphs and test reports.
Appendix A lists the product specifications
Appendix B is a glossary of terms related to the EN 61000-3-2 and EN 61000-3-3 Regulations.
Appendix C is for programmers. It describes the SCPI programming commands that implement the
harmonic/flicker tests.
Appendix D explains the class determination logic of the Agilent 14761A HFTS software, and how this
logic is used to set test limits.
7
1 - Introduction
The Agilent 6800-Series AC Power Source/Analyzers
The Agilent 6800-Series AC Power Source/Analyzers are specifically designed for testing products
compliant to the IEC low-frequency emissions regulations for quasi-stationary current harmonics,
fluctuating current harmonics, and voltage fluctuations and flicker. The following models provide up to
full power coverage of the single-phase regulatory requirements:
Modelrms Voltagerms Currentpeak CurrentVA
Agilent 6812B
Agilent 6841A
Agilent 6813B
Agilent 6842A
Agilent 6843A
230 Vrms (compliance)
300 Vrms (maximum)
230 Vrms (compliance)
300 Vrms (maximum)
230 Vrms (compliance)
300 Vrms (maximum)
3.3 Arms (compliance)
6.5 Arms (maximum)
7.6 Arms (compliance)
13 Arms (maximum)
16 Arms (compliance)
32 Arms (maximum)
40 A750 VA
80 A1750 VA
96 A (low range)
48 A (high range)
4800 VA
Each one-box test system contains the capabilities of a stand-alone ac source, power analyzer, flicker
meter, and line impedance network. Unlike multiple-box ac source and measurement configurations, the
power generation and measurement of the Agilent 6800-Series are controlled by a common internal
timebase, and are truly synchronized. This allows precise measurement of harmonics.
In addition to compliance testing, you can also use the Agilent 6800-Series AC Power Source/Analyzers
as standard ac sources. The units have a SYSTem:CONFigure command that that lets you switch
between IEC mode and Normal mode (the default), where the units behave as standard ac sources. When
you run the Agilent 14761A HFTS software, the units automatically switch from Normal to IEC mode.
The Agilent 14761A Harmonic/Flicker Test System Software
The Agilent 14761A Harmonic/Flicker Test System (HFTS) software application supports EN 61000-32,
EN 61000-3-3, and EN 60555 Part 2 compliance testing requirements. The Agilent 14761A HFTS
software provides an intuitive graphical user-interface from which you can:
ñSet up and run compliance-level tests. The setting up of many IEC details is facilitated through
the use of embedded standards expertise.
ñCollect real-time test data from the Agilent 6800-Series AC Power Source/Analyzer.
ñDisplay and monitor ongoing test results.
ñSave test results.
ñTerminate tests based on user-defined criteria.
ñAnalyze failures or marginal results using advanced test, display, and search options.
ñEvaluate the long-term test results compared to pass/fail criteria.
ñPrint reports and graphs.‘
See Appendix A for a complete list of supported standards and regulations.
NoteA selection in the Options/Default menu lets you compliance test to the older
EN 60555 Part 2 (IEC 555-2) regulation. The EN 60555-3 (IEC 555-3) regulation is no
longer a requirement and has officially been replaced by EN 61000-3-3.
8
Introduction - 1
EN 61000-3-2 and EN 60555 Part 2 Regulations
EN 61000-3-2 and EN 60555 Part 2 regulate the magnitude of harmonic currents drawn by products
from the ac line. For example, harmonic currents can occur as a result of high peak currents drawn by
switch-mode power supplies. Power companies, particularly in Europe, have led a regulatory initiative to
limit harmonic current generation at the product level because of a variety of undesirable effects on the
mains environment such as: interference with other equipment, overheating of conductors and power
factor correction networks, and power transmission losses. Personal computers, peripherals, and variable
speed motor drives are examples of the types of products addressed by the regulations.
EN 61000-3-2 and EN 60555 Part 2 actually cover two categories of harmonic currents: quasi-stationary
harmonics and fluctuating harmonics. Power supply manufacturers, for example, may be concerned
primarily with the first category while product manufacturers may be concerned with both categories.
Different measurement techniques are used for determining compliance for the two categories, with more
stringent testing requirements applied to testing fluctuating harmonics. In particular, compliance testing
for fluctuating harmonics requires non-stop harmonic analysis over extended periods of time. The EN
61000-3-2 regulation applies to class A, B, C, D, and motor-driven equipment.
The EN 60555 Part 2 regulation only applies to class A and class B equipment. This regulation is
selected in the Options/Defaults menu. See appendix D for information about class determination.
Compliance Testing Implementation
The Agilent 14761A HFTS software compliance tests to EN 61000-3-2 / EN 60555 Part 2 as follows:
ñIt initializes and programs the Agilent 6800-Series AC Power Source/Analyzer.
ñAfter the pre-test, the software displays one cycle of voltage and current, which represents the
average of all half-cycle measurements made during the pre-test, with Class D envelope
information superimposed over the waveshapes. Rms voltage, frequency, rms current, peak
current, real power, apparent power, power factor, voltage distortion, current distortion, and
percent in envelope are displayed as well. An additional display indicates if the harmonic-by-
harmonic voltage distortion is either “IN SPEC” or ”OUT OF SPEC” as per the regulations for
source distortion. This is based on the worst-case results obtained during the pre-test.
ñDuring the compliance test, the software produces a real-time bar graph that represents either the
absolute magnitudes of 40 harmonics, or the magnitudes of 40 harmonics expressed as a
percentage of the applicable limits. The graph display shows both the maximum measured value
and the value of the present data record. The value of the present data record is continually
updated while the test is running. A time-series graph is available to display test data for a
specific harmonic.
ñThe software also produces a table that displays either the absolute magnitudes of 40 harmonics,
or the magnitudes of 40 harmonics expressed as a percentage of the applicable limits in real-time.
The table shows both the peak data values and the present data values. Note that you can also
display the data of any individual harmonic versus time, either as an absolute magnitude or as a
percentage of the allowable limits.
ñWhen testing fluctuating harmonics, the software uses a 2.5 minute sliding window. Within this
window harmonic peaks of up to 150% of the steady-state harmonic limits are allowed, provided
that these samples do not total more than 10% of the total samples (or 15 seconds of time) within
any 2.5 minute observation period. Samples that contribute to a 2.5 minute window failure must
fall between 100% and up to 150% of the steady-state harmonic limits. Values greater than 150%
of the limits are counted as individual failures.
9
1 - Introduction
EN 61000-3-3 Regulation
EN 61000-3-3 regulates the magnitude, rate, and time-duration of voltage fluctuations and flicker
caused by products connected to the ac line. Voltage fluctuations are created by time-varying current
drains working against branch circuit impedance that exists in all power distribution networks. Flicker
occurs when an incandescent lamp changes in intensity due to the frequency and amplitude of voltage
fluctuations in the same branch circuit.
Because flicker is annoying, and for certain individuals presents a health hazard, the regulation seeks to
regulate flicker generation to an imperceptible level. A specialized instrument called a flickermeter,
which is built into the Agilent 6800-Series AC Power Source/Analyzer, is used to measure flicker in
terms of human perceptibility. A perceptibility level of 1 represents the threshold of perception for the
average individual.
Note that the EN 60555-3 (IEC 555-3) regulation is no longer a requirement and has officially been
replaced by EN 61000-3-3.
Compliance Testing Implementation
The Agilent 14761A HFTS software compliance tests to EN 61000-3-3 as follows:
ñIt initializes and programs the Agilent 6800-Series AC Power Source/Analyzer.
ñAfter the pre-test, the software shows one cycle of voltage and current, which represents the
average of all half-cycle measurements made during the pre-test, with Class D envelope
information superimposed over the waveshapes. Note that the envelope waveshape is provided
for information only, the regulation does not require Equipment Under Test (EUT)
classification. Therefore, no Percent in Envelope information or Volt THD IN SPEC/OUT of
SPEC information is provided. However, information about rms voltage, frequency, rms current,
peak current, real power, apparent power, power factor, voltage distortion, and current distortion
is provided.
ñWhen the compliance test is run, the software generates a bar graph which displays the short-
term flicker (Pst) compared to the predetermined test limits at the end of each integration period.
The graph shows the maximum measured value and the present value .
ñDuring the compliance test, the software produces a bar graph that displays the selected time
range of rms voltage values. This graph effectively shows the time-series of rms voltage
variations produced by the varying load currents flowing through the reference impedance.
ñProbability and distribution graphs reflect a statistical view of the raw data used to generate the
Pst graph at the end of each integration period. Note that long term flicker (Plt) data is generated
during post processing after the test has completed . This information is included in the long
form report.
ñSummary values for maximum rms voltage deviation (Dmax), steady-state voltage change (Dc),
and the time interval during which the voltage deviation exceeds the prescribed limit (Dt), are
displayed at the bottom of each display. These are the maximum values for each parameter
within the presently displayed integration period.
ñThe software also displays the time-series for instantaneous flicker, which can be useful for
diagnosing faults in the equipment under test.
10
Installation
Install the Software
NOTE:A README.TXT file is included on the installation disks. It contains product updates or
corrections that are not documented in this manual. Use any text editor to read this file.
The Agilent 14761A HFTS software (p/n 5063-2363) comes on two disks. When installed, it requires 5
Mbytes of hard disk space, 8 Mbytes of RAM, and 512 Kbytes of conventional memory. It also requires
the correct interface card drivers to be installed on your PC. See “Installing Interface Card Drivers”.
1. Place Disk #1 in the A drive of your computer.
2. In the Windows Program Manager, run A:\SETUP
3. Follow the directions on your screen. During the installation procedure, you will be asked to
specify an installation directory on your hard disk (the default is C:\HFTS5).
4. You will also be asked to select default voltage and frequency settings for the equipment that you
will be testing (choose either 120 Vac/ 60 Hz or 230 Vac/ 50 Hz). This option may help in avoiding
the accidental applications of damaging voltages to the equipment under test. This option may
changed at any time in the Standard Test Options window.
2
NOTE:Before you run the Agilent 14761A HFTS software, make sure that:
ñNo LAN software is running
ñNo screen savers are running. Press CtrlEsc to check the Windows Task List. Also
check the Control Panel Desktop application.
ñIf you have Power Management Software running on your PC, make sure that any power
management modes such as Sleep or Standby are turned off while the tests are running.
Connect the Equipment
To use the Agilent 14761A HFTS software, install or connect the following equipment to your computer.
1.Make sure that an Agilent GPIB or a National GP-IB interface card with the appropriate drivers has
been installed in your computer. See “Installing Interface Card Drivers”.
2 Install the Agilent 6800-Series AC Power Source/Analyzer and connect it to the computer.
3.Set up the AC Power Source/Analyzer for remote sensing at the input terminals of the equipment
under test. Failure to do this will result in inaccurate output voltage programming.
4.Turn on your computer and Agilent 6800-Series AC Power Source/Analyzer. Setup is complete.
However, in order for this software to operate, you must install a driver for your GPIB card.
11
2 - Installation
Installing Interface Card Drivers for Windows 3.1 or 3.11 Systems
ñIf you are using an Agilent 82335 GPIB card, install the driver software from the I/O Libraries
CD-ROM (E2094) that came with the card. Open the \win31\disk1 directory on the CD-ROM
and run setup.exe. If you do not have this CD-ROM, you may purchase it through a local Agilent
Sales and Support office.
ñIf you are using an Agilent 82340, 82341, or 82350 GPIB card, the CD-ROM that came with the
card (Agilent E2094F) may contain driver software only for Windows 95 and Windows NT. In
that case you should obtain the latest version of this CD-ROM (Rev F.01.02 or later), which also
contains drivers for Windows 3.1. Open the \win31\disk1 directory on the CD-ROM and run
setup.exe.
ñIf you are using a National Instruments GPIB card, install the associated Windows 3.x driver.
You may have received a disk containing this driver when you purchased the card. Also, at the
time this guide was written, the newest drivers were available from www.ni.com. Install the
latest version of NI-488.2.
Installing Interface Card Drivers for Windows 95 Systems
ñIf you are using an Agilent 82335, 82340, 82341, or 82350 GPIB card, install the driver software
from the I/O Libraries CD-ROM that came with the card. If this CD-ROM is an E2094F or later,
run setup.exe in the CD-ROM’s root directory. If the CD-ROM is an E2094E, open the
\win95nt\disk1 directory on the CD-ROM and run setup.exe. If you do not have this CD-ROM,
you may purchase it through a local Agilent Sales and Support office.
ñIf you are using a National Instruments GPIB card, install the associated Windows 95/98 driver.
You may have received a disk containing this driver when you purchased the card. Also, at the
time this guide was written, the newest driver was available from www.ni.com. Install the latest
version of NI-VISA or NI-488.2.
Installing Interface Card Drivers for Windows NT 4.0 Systems
ñIf you have an Agilent 82335 GPIB card, this card will NOT operate under Windows NT. You
should purchase an Agilent 82340, 82341, or 82350 GPIB card through a local Agilent Sales and
Support office.
ñIf you are using an Agilent 82340, 82341, or 82350 GPIB card, install the software from the I/O
Libraries CD-ROM that came with the card. If this CD-ROM is an Agilent E2094F or later, run
setup.exe in the CD-ROM’s root directory. If this CD-ROM is an E2094E, open the
\win95nt\disk1 directory on the CD-ROM and run setup.exe. If you do not have this CD-ROM,
you may purchase it through a local Agilent Sales and Support office.
ñIf you are using a National Instruments GPIB card, install the associated Windows NT driver.
You may have received a disk containing this driver when you purchased the card. Also, at the
time this guide was written, the newest driver was available from www.ni.com. Install the latest
version of NI-VISA or NI-488.2.
Verify the Configuration
Once your equipment and software are installed, click on the Start button and select Programs |
Regulatory Test Solution | Harmonic and Flicker Emissions. (For Windows 3.1, click on the HFTS
12
Installation - 2
icon to run the software.) When the software is run, it automatically searches for an interface card and
queries the Agilent 6800-Series AC Power Source/Analyzer for its model number. Verify that your
configuration is correct as follows:
1. Select the New button in the Welcome Window
2. In the Options menu, select the Configure... command. This displays a Configuration dialog box
on the screen. The Configuration Dialog box contains the following fields:
ñThe Interface Name identifies the interface session when using an SICL interface. The
application determines the default interface name by checking which interface driver is installed
in your system. If “HP” is the interface type, the Interface Name list box displays all of the SICL
names presently configured in the Windows 95 system. (SICL interface names are assigned to
the ac source in the interface card’s configuration utility.) For Windows NT systems, the list box
displays hpib7 as the default interface name. If “National” is the interface type, the Interface
Name box displays the following pre-defined GPIB names: GPIB0, GPIB1, GPIB2,or GPIB3.
ñThe I/O Address displays the address of the Agilent 6800-Series AC Power Source/Analyzer.
The instrument address can be read or set using the front panel Address key.
ñThe Interface Type is either HP, National, or File only.
ñThe HFTS Model # is the instrument model that you are controlling.
ñThe HFTS Serial # is where you enter the serial number of instrument that you are
controlling.(this is optional)
ñEnter the date the instrument was last calibrated (this is optional)
ñEnter the calibration approvel name (this is optional)
Pressing the Discover button causes the software to search for an interface card and query the Agilent
6800-Series AC Power Source/Analyzer for its model number. Any equipment that is located is then
displayed in the appropriate fields in the configuration dialog box.
Pressing Write as Defaults updates the HFTS.INI file with the information that has been entered in this
dialog box either by the user or by the Discover feature. These default values are used during subsequent
start-ups of the Agilent 14761A HFTS software to confirm system configuration.
Press OK to accept the changes, or Cancel to cancel the changes and exit the dialog box.
13
2 - Installation
Default Settings
In the Options menu, select Defaults to view the default settings. The following settings are configurable:
ñThe line voltage and frequency that will be applied to the equipment under test.
ñThe quasi-stationary/fluctuating harmonics regulations version that you will be testing to.
Select either EN 61000-3-2 or the older EN 60555 Part 2 regulation.
Press OK to accept the defaults, or Cancel to cancel the changes and exit the dialog box.
Getting Around in the Application
There are two ways to access the various functions of this application:
1. From the drop-down menus at the top of the screen, or
2. From the menu tabs at the bottom of the screen.
The menu tabs at the bottom of the screen are the same as the View menu commands.
Drop-down Menu
The following commands are located in the drop-down menus at the top of the screen.
File
New...
New From Template...
Open...
Save
Save As...
Lock Test File
Edit Template..
Save As Template...
Print Pre-test
Print Report...
Print Graph/Table...
Printer Setup...
Exit
Lets you select a new test
Lets you select a new test but applies an existing set-up template
Opens an existing test
Saves the presently opened test
Saves the present test under a new name
Sets the attribute of a completed test file to read only
Lets you edit a template
Saves only the template information of the presently opened test
Lets you print the currently displayed pre-test screen
Lets you print the currently displayed report
Lets you print the currently displayed graph or table
Lets you change the printer setup
Exits the application
14
Edit
Copy Table
Copy Graph
View
Main
Test Setup >
Pre-test
Test
Display >
Report >
Options
Local Lockout
Validation Mode
Configure...
Defaults...
Installation - 2
Copies highlighted information from the table to the Clipboard
Copies the presently displayed graph to the Clipboard
Lets you select the type of test to run
Lets you select the test set-up parameters
Lets you run the pre-test
Lets you run the test
Lets you view specific test results
Lets you view the test report
Lets you disable the front panel keys of the Agilent 6800-Series AC Power
Source/Analyzer.
Lets you run a special mode used to validate current harmonics testing
methodology.
Lets you specify the I/O slot address of the interface, the instrument address,
the interface type, and the Agilent 6800-Series AC Power Source/Analyzer
model number.
Lets you view and change the default values for line voltage, frequency, and
test standards.
Mains
Auto On/Auto Off
Auto On/Manual Off
Manual On/Auto Off
Manual On/Manua l O ff
Help
Contents
Using Help
About
Mains ON/ Mains Off
Automatically controls the output of the Agilent 6800-Series AC Power
Source/Analyzer while the test is running.
Lets you manually turn off the output of the Agilent 6800-Series AC Power
Source/Analyzer from the Test window. Provides automatic turn-on at the
beginning of pretests and tests.
Lets you manually turn on the output of the Agilent 6800-Series AC Power
Source/Analyzer from the Test window. Provides automatic turn-off at the
completion of pretests and tests.
Lets you manually control the output of the Agilent 6800-Series AC Power
Source/Analyzer from the Test window.
Accesses the Help contents.
Explains how to use the on-line help.
Displays the Agilent 14761A HFTS software revision.
Indicates the present state of the ac source output.
15
2 - Installation
Menu Tabs
The menu tabs that are located at the bottom of the screen let you easily access the primary functions of
the Agilent 14761A HFTS software. These functions are also available from the file menus. Altogether,
there are five groups of menu tabs:
The Main level group accesses the following functions:
The > symbol in the tab label indicates a lower-level of tabs, which access a number of additional
screens. Once at the lower level, the /\ tab returns you to the Main Level. When active, the [last tab] in
the series jumps to the last lower-level tabs that you had previously accessed.
The following tabs appear in the Test Setup> group. Each tab accesses its respective function.
The following tabs appear in the Pre-test> group. Each tab accesses its respective function.
The following tabs appear in the Display> group for Quasi-stationary or Fluctuating Harmonic testing.
Each tab accesses its respective function.
The following tabs appear in the Display> group for Voltage Fluctuation testing. Each tab accesses its
respective function.
The following tabs appear in the Report> group. Each tab accesses its respective function.
16
3
Setting Up the Test
If you have not already done so, click on the Agilent HFTS icon to run the software. Now you can set up your
test. If the Agilent HFTS icon does not appear on your computer, go back to chapter 2 and install the Agilent
14761A HFTS software.
Select a Test Type
When you first run the Agilent 14761A HFTS software, the Welcome window appears.
Select a test type from the Welcome Window.
1.Click on either Quasi-stationary Harmonics, Fluctuating Harmonics, or Voltage Fluctuations
NoteTo select a different test after you get past the Welcome window, use the drop-down File
menu commands
2.Choose to configure a new test, run an existing test, or configure a new test based on an existing
template. Note that these selections are also available in the File menu.
ñSelecting New opens the Main window, in which you start entering test setup information about
the test that you will be running. If you want to run other tests of the same type using the same
setup information, consider saving this information to a template file once the setup has been
established.
17
3 - Setting Up the Test
ñSelecting New From Template opens a dialog box that lets you select a template file which
contains previously-configured test setup information. This template information will be copied
into your new data file. The template itself is not changed unless you make subsequent changes
to your setup and save them into the template using “Save as Template.”. When you run the test,
the test data will then be added to the information obtained from the template. You can modify
the template defaults and set up data as required for the new test or use the values supplied by the
template. The following file extensions are used to identify template files:
.stt template files for quasi-stationary harmonic tests
.flt template files for fluctuating harmonic tests
.fkt template files for voltage fluctuation tests
The default directory for template files is C:\hfts\template
ñSelecting Open Existing File opens a dialog box that lets you select a data file which contains
setup information and may or may not contain pre-test as well as test data. This lets you examine
previously run test data or replace old test data with data obtained from a new test run. The
following file extensions are used to identify data files:
.sta data files for quasi-stationary harmonic tests
.flu data files for fluctuating harmonic tests
.flk data files for voltage fluctuation tests
The default directory for data files is C:\hfts
NoteRunning a pre-test or a test after opening an existing file will cause all existing data to be
overwritten with new data. You can prevent this from occurring by checking the Make
test data read-only option in the Test window before running a test, or by using the
Lock Test File command in the File menu.
18
Setting Up the Test - 3
Create a Template
Template files contain setup information for the test that you will be running. Using templates can save
valuable time by eliminating the need to enter repetitive data each time you set up a test.
Create a template by entering information into the Main and the Test Setup windows. Use the Tab key
to move among the different fields.
1. Enter the following information into the text boxes of the Main window:
your name
the company name
a brief description of the test procedure that you will be running
a device ID such as a model number
a test ID such as a serial number or test number that uniquely identifies the test
2. Use the menu tabs on the bottom of the screen to access the Test Setup windows. Note that you
can also access the Test Setup windows from the View Menu.
3. Starting with the Standard Test Setup window, select and configure each of the following Test
Setup windows as needed: Standard, Advanced, and Termination
4. Select Save as Template in the File menu to create a template file. Once created, the template
file may be used to quickly create new tests by adding or editing only the information that must
be changed for each test. This is useful if you are conducting a series of tests or testing similar
equipment that is subject to a standardized test.
19
3 - Setting Up the Test
Select the Test Setup Options for Quasi-stationary or
Fluctuating Harmonics
Standard Test Options
Standard options specify the line voltage, line frequency, test time, and device class for compliance tests.
They are selected in the Standard window.
ñSelect the line voltage that will be applied to the equipment under test.
Selecting Variable lets you enter a value other than 120 Vac or 230 Vac.
ñSelect the line frequency (50 or 60 Hz) that will be applied to the equipment under test.
ñSelect the test duration in seconds, minutes, hours, or days.
The # of Records field displays the number of data records that will be taken during the test time
that you have specified. Note that the relationship between time and the number of records per
second is a function of line frequency and measurement window type. The measurement window
type is specified in the Advanced Test Setup window, and is set to a default of Rectangular.
ñSelect from one of the following device classes for your equipment under test to which the EN
61000-3-2 regulation applies:
Class A: All equipment except that stated in one of the remaining three classes.
Class B: Portable electrical tools, which are hand held during normal operation and used for a
short time (a few minutes) only.
Class C: Lighting equipment, including dimming devices.
20
Setting Up the Test - 3
Class D: Equipment having an input current with a "special wave shape" (e.g. equipment with
off-line capacitor-rectifier ac input circuitry and switch-mode power supplies) and an active input
power ≤ 600 W. The active power is defined as Watts. For the Class D mA/W limits to apply, the
active power must also be greater than 75 W.
Motor Driven Device: Check this box if the equipment you are testing is a motor-driven device.
This will cause class A limits to be used in accordance with EN 61000-3-2 regardless of the input
current waveshape.
NoteThe EN 60555 Part 2 regulation only applies to Class A and Class B equipment. The
Agilent 14761A HFTS software will reflect this based on the regulations selection made
in the Options/Defaults menu. See appendix D for information about class
determination.
Advanced Test Options
The advanced test options may be used to modify test conditions outside the range permitted for
compliance tests. They are selected in the Advanced window. Generally, you would not change the
factory default settings if you are doing compliance testing because the changes may invalidate the test.
As an exception to the above statement, the Current Measurement Range setting and the Current Limit
settings may be changed without invalidating the compliance test as long as the changed settings do not
cause current limiting to occur during the test. Current limiting may occur during the time period when
power is initially applied but before the test is run. This time delay may be specified in the Advanced
Setup window for tests, and in the Pre-test window for pre-tests. Current limit errors are detected and
displayed if they occur during the test. Note that you can also set test limit overrides < 100% without
automatically flagging the test as non-compliant.
21
3 - Setting Up the Test
ñSelect the rms and peak current limit of the Agilent 6800-Series AC Power Source/Analyzer that
you will be applying to the equipment under test. (The Agilent Model 6843A only has rms
current limit control, therefore, the peak current limit control does not appear.) Generally, it is
best to leave these parameters set to their maximum default values. Although the function of the
current limit is to protect the equipment under test, if the ac source goes into current limit during
a normal test the test will be invalid. Default maximum current limit values are automatically
established based on the model number that you entered into the Main window.
NoteThe Agilent 6800-Series AC Power Source/Analyzer can supply brief peak currents that
exceed its current capability. If this occurs for an extended time with Agilent
6812B/6841A and 6813B/6842A models, the units may activate an internal protection
circuit (the SOA limit) to turn the output off. If the equipment that you are testing causes
the Agilent 6800-Series AC Power Source/Analyzer to turn its output off, you may need
to lower the peak current limit setting. This will clip the output current of the Agilent
6800-Series AC Power Source/Analyzer during inrush but if properly set, will not
interfere with the normal running of the pre-test and tests.
ñSelect the measurement window type (Rectangular or Hanning)
ñSelect the peak current measurement range of the Agilent 6800-Series AC Power
Source/Analyzer
Low Range
High Range
6812B
6841A
8 A peak8 A peak9.6 A peak
80 A peak80 A peak96 A peak
6813B
6842A
6843A
Because of its better measurement resolution, use the low range when testing low power
equipment. However, be careful that the equipment under test never exceeds the peak currents
indicated in the above chart or the test results will be invalid.
ñSelect a test limit override for the harmonic failure threshold. This is specified as a percent of
the limit as defined by the applicable regulation. For Class C, test limits are also a function of the
measured (or user-specified) power factor and the line current at the fundamental frequency. If
necessary, you can override the power factor measured in the pre-test. For Class D, test limits are
also a function of real power (or watts). If necessary you can also override the watts measured in
the pre-test. Overrides are useful to prevent minor changes in measured values from changing the
test limits each time the test is run.
ñSelect a measurement delay from the time power is applied to the equipment under test until the
time that the Agilent 6800-Series AC Power Source/Analyzer starts measuring data. According
to EN 61000-3-2, this delay must be 10 seconds or less for compliance level fluctuating harmonic
tests. Note that the measurement delay has no consequential impact on test results. This is
because during the "delay to start of test" portion of both the pre-test and main test, test results
are not recorded and current limiting events are ignored. However, once the actual acquisition of
measurement data begins, current limit events are again treated as errors that invalidate the test.
Since high inrush currents usually occur only during the first few mains cycles following
application of mains voltage, a properly set current limit threshold will not be crossed once the
"delay to start of test" period has ended.
22
Setting Up the Test - 3
Test Termination Options
Termination options let you select the conditions that will terminate the test. They are selected in the
Termination window.
NoteThe following conditions terminate any test automatically if the Agilent 6800-Series AC
Power Source/Analyzer operating limits are exceeded: Rail fault condition,
Overtemperature, Overvoltage, and SOA. Refer to the applicable ac source User’s
Guide for an explanation of these conditions.
ñSelect termination on errors. One error will terminate the test when this is selected:
UNR - when the output becomes unregulated.
ñSelect termination when the current limit is exceeded. The two current limit conditions that will
terminate the test when this occurs are: CL - when the output goes into current limit mode, and
CP - when the output current limit protection has tripped.
ñSelect termination upon exceeding a specified number of failures. When testing fluctuating
harmonics and this box is checked, you can specify failures:
of any type,
with a 2.5 minute window
without a 2.5 minute window.
ñSelect termination on exceeding a specified number of watch events. Each event that satisfies the
selected criteria is counted as one event occurrence.
23
3 - Setting Up the Test
If termination On Watch Events is enabled, you can select the following Watch Items criteria:
ñAll harmonics greater than the specified percent of threshold
ñOdd harmonics greater than the specified percent of threshold
ñOdd≤ 19 harmonics greater than the specified percent of threshold
ñEven harmonics greater than the specified percent of threshold
ñNth, a specific Harmonic Number that is greater than the specified percent of threshold
Select the Test Options for Voltage Fluctuations
Standard Test Options
Standard options specify the line voltage, line frequency, and test time for compliance tests. They are
selected in the Standard window.
ñSelect the line voltage that you will be applying to the equipment under test.
Selecting Variable lets you enter a value other than 120 Vac or 230 Vac.
ñSelect the line frequency (50 or 60 Hz) that will be applied to the equipment under test.
24
Setting Up the Test - 3
ñSelect the number of integration periods to be acquired during the test. The Resulting Time
field displays the test duration based on the number of integration periods as well as the Pst
integration time that you have specified in the Advanced window. The Pst integration time is set
to a default of 10 minutes with the Advanced options disabled.
Advanced Test Options
ñSelect the rms and peak current limit of the Agilent 6800-Series AC Power Source/Analyzer that
you will be applying to the equipment under test. Generally, it is best to leave these parameters
set to their maximum default values. Although the function of the current limit is to protect the
equipment under test, if the unit does go into current limit during a normal test, the test will be
invalid.
ñSelect the returned data type (All, Pst/Rms Summary, or Rms/Flicker Time Series)
ñSelect the Pst (or short-term) flicker integration time.
ñSelect a test limit override for the Pst/Plt Thresholds (short-term/long-term flicker) and Rms
Thresholds. This is specified as a percent of the specification in the applicable regulation.
ñSelect a measurement delay from the time power is applied to the equipment under test until the
time that the Agilent 6800-Series AC Power Source/Analyzer starts measuring data.
25
3 - Setting Up the Test
Test Termination Options
NoteThe following conditions terminate any test automatically if the Agilent 6800-Series AC
Power Source/Analyzer operating limits are exceeded: Rail fault condition, Overtemperature,
Overvoltage, and SOA. Refer to the ac source User’s Guide for more information.
ñSelect termination on errors. One error will terminate the test when this is selected:
UNR - when the output becomes unregulated.
ñSelect termination when the current limit is exceeded. Two current limit conditions will
terminate the test when this occurs: CL - when the output goes into current limit mode; CP when the output current limit protection has tripped.
ñSelect termination on exceeding a specified number of Pst failures
ñSelect termination on exceeding a specified number of Rms voltage failures. You can specify
Rms voltage failures: of any type
Dmax failures
Dc failures
Dt failures.
ñSelect termination on exceeding a specified number of watch events. Each event that satisfies the
selected criteria is counted as one event occurrence.
If termination On Watch Events is enabled, you can select the following Watch Items criteria:
ñPst values greater than the specified perceptibility units
ñInstantaneous Flicker values greater than the specified perceptibility units
26
4
Running the Test
Before actual test data can be acquired, the pre-test must be run. This is accomplished in the Pre-Test
window. As its name implies, the Pre-Test window centralizes functions that, following test setup, need
to be performed prior to running an actual test. This includes running preliminary tests to measure
quantities that are subsequently used to set test limits for Class C and Class D devices. Once the pre-test
data is available, the main test can be started. This is accomplished in the Test window. Test status
information is displayed in the Status bar on the bottom of the screen. Note that you must always have
pre-test results present before running the main test.
To Run a Pre-test for Quasi-stationary or Fluctuating
Harmonics
Pre-Test Summary
1. Select the Pre-test menu tab on the bottom of the screen to access the Pre-test Summary
window. You can also access the Pre-test window from the View/Pre-test menu.
2. If desired, change the time delay from when power is applied until the pre-test is run. The default
pre-test time is 10 seconds.
27
4 - Running the Test
3. It is also possible to increase the duration of the pre-test. One purpose of running a longer pre-
test is to obtain more data over which the test results are averaged. Additional data more
accurately characterizes the worst-case operation of the equipment under test (see “Pre-test
Measurements” for more information).
4. Press the Run Pre-test button to run the pre-test.
NoteRunning a pre-test from a file that already contains data will not only overwrite the pre-
test data, but will also delete the existing test data.
5. Use the Print Screen command located in the File menu to print the Pre-test window.
For Quasi-stationary and Fluctuating Harmonic tests, the pre-test checks if you have correctly classified
your equipment as Class D versus any other class in the Standard Test Setup window. It displays the
input-current waveshape of the equipment under test, which lets you examine the waveshape for “fit”
into the Class D special waveshape window.
ñThe Pre-test graph displays one cycle of voltage and current, superimposed on the Class D
waveshape window. The current cycle that is superimposed in the Class D waveshape window is
the half-cycle having the highest peak value within a single 16-cycle snapshot taken at the end of
the pre-test interval (see “one-shot measurements” later in this chapter for more information).
Press the Power Statistics tab if you need to configure the active power requirements for
determining Class D limits, or if you need to display additional power measurement information.
ñThe Class Indicator displays a green background if the data collected from the equipment under
test matches the Class selection. If the data collected from the equipment under test does not
match the Class selection, the Class indicator displays a red background. In this case you should
return to the Standard Test Setup window, select the correct class, and rerun the pre-test. Refer to
appendix D for information about the class determination logic of the Agilent 14761A HFTS
software.
NOTE:Class D equipment must have input power > 75 W and ≤ 600 W, and 95% of its input-
current waveshape must fall within the Class D waveshape window. For EN 61000-3-2
testing, devices generating Class D waveforms but NOT meeting all Class D input power
criteria will be tested using Class A limit values for odd harmonics. If testing against
limits for even harmonics is also desired, select Class A. Testing a device that meets the
Class D "special waveshape" with Class A selected, results in compliant test reports if
the "Motor Driven Device" checkbox, available for EN 61000-3-2 testing, is also
selected.
ñThe Percent in Envelope indicator displays the percentage of waveform data within the Class D
waveshape standard if Class A or Class D has been selected.
ñThe Voltage THD “IN SPEC” or “OUT OF SPEC” indicator summarizes the results of testing to
determine if the output voltage of the ac source is within the harmonic voltage limits as specified
in EN 61000-3-2/EN 60555-2. The voltage THD results are based on measurements of
individual voltage harmonics up to the 40th harmonic, and a comparison of these measurements
to the limits specified in the regulations. Select the Source Harmonics tab to display the
individual worst-case voltage harmonics that occurred during the pre-test.
ñThe Measured Values area on the right side of the display summarizes nine of the most common
mains input characteristics (see “Pre-test Measurements” for more information).
28
Pre-Test Power Statistics
Running the Test - 4
This screen provides additional information about the active input power of the device under test. It
displays the following power measurements, which are calculated at the end of the pre-test:
- Maximum Power in watts (default selection for setting Class D limits)
- Mean Power in watts
- Standard Deviation in watts
- specified Percentile in watts
The Percentiles area displays the results of a statistical evaluation of the active input power during the
pre-test. Percentile values are displayed according to the selected scale increment. Scale increments of
10, 5, or 1 may be specified. A scale increment of 1 displays all percentile values.
If you are testing Class D equipment, the Use to set limits area is active, letting you specify what type of
Watts measurement will be used by the Agilent 14761A HFTS software as the basis for calculating Class
D test limits. Select from: Maximum Power (the default), Mean Power, or specified Percentile. This
provides you with greater flexibility for specifying Class D test limits.
Note:Based on your selection, you can see what the Class D test limits will be by acccessing
the Display Table window. The Class D limits are shown in the Limits column.
You can also override the pre-test watts measurement later in the Advanced Test Setup
window in the Test Limit Overrides section.
29
4 - Running the Test
Pre-Test Source Harmonics
This screen displays the pre-test voltage data for harmonics 1-40 on a worst-case basis. The Limit(%)
column indicates the permissible limits for each harmonic expressed as percentages of the fundamental
voltage. For 230 V/50 Hz power distribution systems, these are:
- No limits for the 1st harmonic (the fundamental)
- 0.9% of V
- 0.4% of V
- 0.3% of V
- 0.2% of V
- 0.1% of V
- 0.1% of V
fundamental
fundamental
fundamental
fundamental
fundamental
fundamental
for 3rd harmonic
for 5th harmonic
for 7th harmonic
for 9th harmonic
for even harmonics from 2 to 10
for all harmonics from 11 to 40
The Limit(Abs) column indicates the permissible limits for each harmonic expressed as absolutes (in
volts). These values are obtained by multiplying the %limits divided by 100 times the measured
minimum value for the fundamental component. There are no limits for the fundamental.
The Max(%) column indicates the maximum pre-test values for harmonics 2 through 40 expressed as
percentages. These values are obtained by dividing the maximum harmonic values by the minimum
fundamental value and multiplying by 100. Failures are indicated in red.
The Max(Abs) column indicates the maximum pre-test values for harmonics 2 through 40 expressed as
absolutes (in volts). These values are simply the maximum values of the voltage harmonic data returned
during the pre-test. Failures are indicated in red.
30
Loading...
+ 76 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.