ZOLL, E Series, and RescueNet are registered trademarks, and SurePower Charger Station is a trademark of ZOLL Medical Corporation. All other
trademarks and registered trademarks are property of their respective owners.
Preface.............................................................................................................................................................. v
Additional Reference Material ........................................ .... ... ... ................................................................................. vi
Conventions................................................................................................................................................................ vi
Service Policy Warranty............................................................................................................................................ vii
Technical Service for International Customers ..................................................................................................... vi ii
Ch a p t e r 1 Maintenance Tests ........................................................................................................................ 1-1
Before You Begin the Maintenance Tests............................................................................................................... 1-2
Equipment You Need to Perform the Maintenance Tests...................................................................................... 1-2
Equipment You Need for the E Series Options Maintenance Tests ..................................................................... 1-3
Physical Inspection of the Unit............................................................................................................................. 1-4
Front Panel Button Test....................................................................................................................................... 1-6
3, 5, and 12 Leads Test....................................................................................................................................... 1-9
Power Supply Test (Optional)............................................................................................................................ 1-10
Leakage Current Test........................................................................................................................................ 1-14
Paddles Test (If applicable)............................................................................................................................... 1-15
Calibrating Pulses on Strip Chart Test............................................................................................................... 1-17
Notch Filter Test ................................................................................................................................................ 1-18
Defibrillator Self Test ......................................................................................................................................... 1-21
Advisory Message Test ..................................................................................................................................... 1-27
Pacer Test ......................................................................................................................................................... 1-28
SpO2 Monitor Test for SpO2 Option.................................................................................................................. 1-31
EtCO2 Monitor Test (for EtCO2 Option)............................................................................................................ 1-33
NIBP Monitor Test .............................................................................................................................................. 1-37
NIBP Volume Leak Test with Bio-Tek NIBP Analyzer........................................................................................ 1-39
NIBP Volume Leak Test with Fluke Biomedical Cufflink NIBP Analyzer....................................................... ... .. 1-41
Ch a p t e r 2 Troubleshooting ............................................................................................................................... 2-1
ZOLL E Series Error Messages ............................................................................................................................... 2-5
Ch a p t e r 3 Replacement Parts........................................................................................................................... 3-1
Field Replacement Parts.......................................................................................................................................... 3-5
Ch a p t e r 4 Functional Description.................................................................................................................... 4-1
Main System Board............................. .... ................................................ ... ... .... ... ... ... .... ... ........................................ 4-2
Main System Board Functions ............................................................................................................................... 4-4
Power Supply ......................................................................................................................................................... 4-5
ECG Front End ...................................................................................................................................................... 4-6
Multifunction Electrode (MFE)/PADS (System Board and High Voltage Module) ................................................. 4-6
CPU and EPU ........................................................................................................................................................ 4-6
High Voltage Module................................................................................................................................................. 4-7
Defibrillator Charging and Discharging ......... .... ..................................................................................................... 4-7
High Voltage Capacitor Monitor ............................................................................................................................. 4-8
Pacer/Defibrillator Control Signals ......................................................................................................................... 4-9
System Interconnect Module..................................................................................................................................4-11
Front Panel and Controls PWBA ......................................................................................................................... 4-11
E Series Options..................................................................................................................................................... 4-12
12 Lead Option .................................................................................................................................................... 4-12
Ap p e n d i x A Overview ..........................................................................................................................................A-1
9650-1450-01 Rev. Aiii
E Series Service Manual
Preface
ZOLL® Medical Corporation’s E Series® Service Manual is intended for the service technician whose responsibility is to identify malfunctions and/or
make repairs at the subassembly level. The ZOLL E Series Service Manual has five main sections and one appendix.
Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to review this section thoroughly before attempting to use or service
the E Series unit.
Chapter 1—Maintenance Tests explains how to check the defibrillator’s performance using a series of recommended checkout procedures to be
conducted every six months.
Chapter 2—Troubleshooting provides a listing of the procedures and error messages to help the service technician detect faults and repair them.
Chapter 3—Replacement Parts List displays a complete list of ZOLL part numbers for field replaceable parts available for the E Series unit, allowing
the service person to identify and order replacement parts from ZOLL.
Chapter 4—Functional Description provides technical descriptions for the E Series major subassembly modules.
Appendix A—E Series interconnect diagrams and maintenance checklists.
Safety Considerations
The following section describes general warnings and safety considerations for operators and patients. Service technicians should review the safety
considerations prior to servicing any equipment and read the manual carefully before attempting to disassemble the unit. Only qualified personnel should
service the E Series unit.
Federal (U.S.A.) law restricts this unit for use by or on the order of a physician.
Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug Administration (FDA) under section 510(K) of the Medical
Device Act to obtain approval to market is based upon the use of ZOLL accessories such as disposable electrodes, patient cables and batteries. The use of
external pacing/defibrillation electrodes and adapter units from sources other than ZOLL is not recommended. ZOLL makes no representations or
warranties regarding the performa nce or ef fectivenes s of its product s when used in conjunction with pacing/d efibrillation elect rodes and adapter units fro m
other sources. If unit failure is attributable to pacing/defibrillation electrodes or adapter units not manufactured by ZOLL, this may void ZOLL's warranty.
Only qualified personnel should disassemble the E Serie s unit.
WARNING! This unit can generate up to 2250 volts with sufficient current to cause lethal shocks.
9650-1450-01 Rev. Av
E Series Service Manual
All persons near the equipment must be warned to STAND CLEAR prior to discharging the defibrillator.
Do not discharge the unit’s internal energy more than three times in one minute or damage to the unit may result.
Do not discharge a battery pack except in a ZOLL SurePower™
Charger Station or compatible ZOLL Battery Charging/Testing unit.
Do not use the E Series in the presence of flammable agents (such as gasoline), oxygen-rich atmospheres, or flammable anesthetic s. Using the un it near the
site of a gasoline spill may cause an explosion.
Do not use the unit near or within puddles of water.
Note:The E Series is protected against interference from radio frequency emissions typical of two-way radios and cellular phones (digital and analog)
used in emergency service/public safety activities. Users of the E Series should assess the unit’s performance in their typical environment of use
for the possibility of radio frequency interference from high-power sources. Radio Frequency Interference (RFI) may be observed as shifts in
monitor baseline, trace compression, or transient spikes on the display.
Additional Reference Material
In addition to this guide, there are several other components to the ZOLL E Series documentation. They include:
• E Series Operator’s Guide - A comprehensive reference work that describes all the user tasks needed to operate the E Series.
• E Series Configuration Guide - Describes the E Series features and functions whose operation can be customized by authorized users.
• E Series Operator’s Guide Option Insert: 12-Lead ECG Monitoring- Describes using the 12-lead ECG monitoring option with the E Series un it.
• E Series Operator’s Guide Option Insert: End Tidal Carbon Dioxide (EtCO
• E Series Operator’s Guide Option Insert: Non-Invasive Blood Pressure (NIBP)- Describes using the NIBP option with the E Series unit.
• E Series Operator’s Guide Option Insert: Pulse Oximetry (SpO
• E Series Operator’s Guide Option Insert: Non-Interpretive 12-Lead ECG Monitoring- Describes using the non-interpretive 12-lead ECG monitoring
option with the E Series unit.
• E Series Operator’s Guide Option Insert:12-Lead Reperfusion Therapy Algorithm - Describes the reperfusion therapy algorithm option for the E Series
)- Describes using the SpO2 option with the E Series unit.
2
unit.
)- Describes using the EtCO2 option with the E Series unit.
2
Conventions
WARNING!Warning statements describe conditions or actions that can result in personal injury or death.
vi9650-1450-01 Rev. A
E Series Service Manual
CautionCaution statements describe conditions or actions that can result in damage to the unit.
Note:Notes contain additional information on using the defibrillator.
Service Policy Warranty
In North America: Consult your purchasing agreement for terms and conditions associated with your warranty. Outside of North America, consult ZOLL
authorized representative.
In order to maintain this warranty, the instructions and procedures contained in this manual must be strictly followed. For additional information, please
call the ZOLL Technical Service Department 1-800-348-9011 in North America.
Technical Service
If the ZOLL E Series unit requires service, contact the ZOLL Technical Service Department:
Telephone: 1-978-421-9655; 1-800-348-9011 (US only)
Fax 1-978-421-0010
Have the following information available for the Technical Service representative:
• Unit serial number.
• Description of the problem.
• Department where equipment is used.
• Sample chart recorder strips documenting the problem, if applicable.
• Purchase Order to allow tracking of loan equipment.
• Purchase Order for a unit with an expired warranty.
If the unit needs to be sent to ZOLL Medical Corporation, obtain a Service Request number from the Technical Service representative. Return the unit in
its original container to:
ZOLL Medical Corporation
269 Mill Road
Chelmsford, Massachusetts 01824-4105
Attn: Technical Service Department, SR #
9650-1450-01 Rev. Avii
E Series Service Manual
Technical Service for International Customers
International customers should return the unit in its original co ntainer to the nearest authorized ZOLL Medical Corporation Service Center. To locate an
authorized service center, contact the International Sales Department at ZOLL Medical. See back cover of this manual.
viii9 650-1450-01 Rev. A
E Series Service Manual
Chapter 1
Maintenance Tests
Overview
The E Series has two checkout procedures: the operator’s shift checklist and the extensive six-month maintenance tests checkout procedures.
Because the E Series units must be maintained ready for immediate use, it is important for users to conduct the Operator’s Shift Checklist procedure at the
beginning of every shift. This proce dure c an be c omplet ed in a few mi nutes an d re quires no additional test equipment. (See the ZOLL E
Guide for the Operator’s Shift Checklist.)
A qualified biomedical technician must perform a more thorough maintenance test checkout every six months to ensure that the functions of the E Series
unit work properly. This chapter describes the step by step procedures for performing the six
back of this document (ZOLL E
Series Maintenance Tests Checklist) to record your results of the maintenance tests.
month maintenance test checkout. Use the checklist at the
Series Operator’s
This chapter describes the following maintenance tests:
• 1. Physical Inspection of the Unit
• 2. Front Panel Button Test
• 3. 3, 5, and 12 Leads Test
• 4. Power Supply Test
• 5. Leakage Current Test
• 6. Paddles Test
• 7. Heart Rate Display Test
• 8. Calibrating Pulses on Strip Chart Test
• 9. Notch Filter Test
• 10. Heart Rate Alarm Test
• 11. Defibrillator Self Test
• 12. Synchronized Cardioversion Test
• 13. Shock Test
• 14. Summary Report Test
9650-1450-01 Rev. A1-1
E Series Service Manual
• 15. Advisory Message Test
• 16. Pacer Test
• 17. SpO
• 18. EtCO
• 19. Barometric Pressure Calibration Chec k
• 20. CO2 Accuracy Check
• 21. NIBP Monitor Test
• 22. NIBP Volume Leak Test with Fluke Biomedi cal NIBP Analyzer
• 23. NIBP Volume Leak Test with Fluke Biomedical Cufflink NIBP Analyzer
• 24. NIBP Transducer Calibration Test
Monitor Test
2
Monitor Test
2
Before You Begin the Maintenance Tests
• Assemble the tools or specialized testing equipment listed in the “Equipment You Need to Perform the Maintenance Tests” section shown below.
• Keep an extra fully charged ZOLL E Series compatible battery available.
• Schedule an hour to conduct the entire maintenance test.
• Photocopy the checklist at the back of this document and use the copy to record your results. As you conduct each step of a procedure, mark the
Pass/Fail/NA check boxes on your checklist and then save it for your maintenance file.
• Perform the tests in the order presented.
• Perform all the steps of each test procedure.
• Complete all the steps of the procedure before evaluating the test results.
Equipment You Need to Perform the Maintenance Tests
For testing purposes, you can substitute an equivalent device.
• ZOLL Medical Electrode Adapter from Fluke Biomedical (part number 3010-0378).
• Fluke Biomedical Impulse 4000 Defibrillator Analyzer with 1.06 software or higher.
• Fluke Biomedical
• ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201).
• Stop watch.
1-29650-1450-01 Rev. A
601 Pro Series International Safety Analyzer.
• Standard series II PC flash memory cards.
• PCMCIA card reader and PC.
• RescueNet
• Phillips #1 screwdriver.
• Phillips #2 screwdriver.
• Flatblade screwdriver.
• Needle nose pliers without teeth.
• Wooden cuticle sticks with bev eled edges (or similar non-conducting implements).
®
Code Review V3.31 or higher from ZOLL Data Systems.
Equipment You Need for the E Series Options Maintenance Tests
E Series Service Manual
• Fluke Biomedical
• SpO
• EtCO
• Paddles (if used).
• Printer Paper.
• Battery.
• AC line cord.
• 3 lead, 5 lead and 12 lead ECG cables. (12 lead cable needed if 12 lead option is installed.)
• Fluke Biomedical BP Pump NIBP Monitor Analyzer (For NIBP units only) with NIBP cable and cuff (if NIBP option is installed), or
• Fluke Biomedical Cufflink Analyzer (if NIBP option is installed)
cable and sensor (if option is installed).
2
cable, and CAPNOSTAT 5 Mainstream cable with airway adapter, or CAPNOSTAT 5 Sidestream cable with cannula (if option is installed).
2
Index 2PFE SpO2 Simulator or equivalent. (For SpO2 units only.)
Note:The Fluke Biomedical BP Pump NIBP Monitor Analyzer and the Fluke Biomedical BP Pump NIBP Analyzer use different technologies for
testing NIBP monitors and therefore, the manual provides two di fferent procedures for performing the NIBP Volume Leak test with each of these
types of test equipment.
9650-1450-01 Rev. A1-3
E Series Service Manual
1.0 Physical Inspection of the Unit
Tools Needed: None.
Test Setup: None.
Observe this...Pass/Fail/NA
1.1Housing
Is the unit clean and undamaged?
1.2Does the unit show signs of excessive wear?
1.3Does the handle work properly?
1.4Does the recorder door open and close properly?
1.5Are input connectors clean and undamaged?
1.6Are there any cracks in the housing?
1.7Do the front panel or selector switches have any damage or cracks?
1.8Are there any loose housing parts?
1.9Paddles
Do the adult and pedi plates have major scratches or show signs of damage?
1.10Do the adult shoes slide on and off easily to expose the covered pedi plates?
1.1 1Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
1.12Cables
Are all cables free of cracks, cuts, exposed or broken wires?
o o
o o
o o
o o
o o
o o
o o
o o
o o o
o o o
o o o
o o
1.13Are all bend/strain reliefs undamaged and free of excessive cable wear?
1.14Battery
Is the ZOLL battery fully charged?
1-49650-1450-01 Rev. A
o o
o o
Observe this...Pass/Fail/NA
E Series Service Manual
1.15Is the battery seated in the battery well correctly?
Record your results on the Maintenance Test Checklist.
o o
9650-1450-01 Rev. A1-5
E Series Service Manual
2.0 Front Panel Button Test
Tools Needed: None.
Test Setup:
1. Install strip chart paper into the recorder tray.
2. Install the battery in the unit or connect the A/C power cord to the unit and then plug the cord into an electrical outlet.
3. Connect the universal cable and ECG cable (3 lead, 5 lead, or 12 lead) to the ZOLL simulator, or Fluke Biomedical Impulse 4000 Analyzer (or
equivalent).
Do this...Observe this...Pass/Fail/NA
2.1Turn the selector switch to MONITOR.
(For AED units, turn the selector switch to ON
and select Manual mode.)
2.2Press the LEAD button; three times for the 3
lead cable and seven times for the 5 lead cable.
2.3Connect the 12 lead cable to unit and simulator.
Press the LEAD button and select the lead for
each of the 12 lead settings.
2.4Set the simulator to NSR of 120 BPM. To check
the size of the ECG waveform, press the SIZE
button.
2.5Press the ALARM SUSPEND button.Alarm symbol changes from disabled to enabled. If the alarm sounds, press
2.6Press the RECORDER button.The strip chart paper moves out of the unit from the paper tray. Check that the
2.7Open the paper compartment door.
Press RECORDER button.
Listen for 4 beep tones. PADS and MONITOR display on the monitor.
NOTE: PADS is a factory default setting.
Each time you press the LEAD button, a different lead number appears under
the LEAD heading on the display.
PADS, I, II, III will display a 3 lead ECG cable if connected or no ECG cable is
connected.
PADS, I, II, III, AVR, AVL, AVF, V1 will display a 5 lead ECG cable.
A 12 Lead cable will display PADS, I, II, III , AVR, A VL, AVF, VI, V2, V3, V4, V4,
V5, V6.
As you press the SIZE button five times (0.5, 1.0, 1.5, 2.0, 3.0), note that the
size of the ECG waveform appropriately changes on the display.
the ALARM SUSPEND button to turn it off. The alarm will only be suspended
for 90 seconds at this point. Press and hold the ALARM SUSPEND button for
3 seconds to disable alarms.
correct time, date, ECG lead annotation and waveform are recorded on the
paper. (Set Time and Date, if necessary.)
CHECK RECORDER message appears on the monitor.
o o
o o
o o o
o o o
o o o
o o o
o o o
1-69650-1450-01 Rev. A
E Series Service Manual
Do this...Observe this...Pass/Fail/NA
2.8Close the paper compartment door.
Press RECORDER button.
2.9Press RECORDER button.Strip chart paper stops flowing out of paper tray.
2.10Press the VOLUME softkey.
To increase the volume of the beep, press the
Inc. softkey.
2.11To decrease the volume of the beep, press the
Dec. softkey.
2.12Press the CONTRAST button. Contrast menu displays.
2.13Press the CONTRAST button.
To increase the contrast of the display, press
the Inc. softkey.
2.14To decrease the contrast of the display, press
the Dec. softkey.
Strip chart paper flows out of paper tray. Verify that the CHECK RECORDER
message no longer displays.
The volume bar graph displays.
Audible beep when the QRS wave displays. The bar graph increases on the
display indicating an increase in volume. This action does not increase the
volume which is normal.
Note: The QRS tone is on or off. There is no gradual change in volume. If
equipped, voice prompts are gradual. Note: The voice volume has 5 settings.
Setting 3 is in the mid-range.
The bar graph decreases on the display indicating a decrease in volume. The
volume shuts off at the last bar; otherwise, the volume is the same as originally
set.
Background light and characters display.
The contrast increases on the monitor display.
Thebar graph increases on the display indicating an increase in contrast.
The bar graph decreases on the display indicating a decrease in contrast. The
display contrast changes.
o o o
o o o
o o
o o
o o
o o o
o o o
o o o
o o
2.15Press the SUMMARY button.Summary menu displays on the monitor showing the summary report options.
2.16Press the CODEMARKER button.Code marker menu displays.
2.17Connect a/c current and install the battery. Turn
the unit off.
9650-1450-01 Rev. A1-7
The CHARGER ON indicator lights.
The amber or green lights illuminate.
Note: If both lights flash ON/OFF, the unit is defective or no battery is installed.
o o o
o o o
o o o
E Series Service Manual
Do this...Observe this...Pass/Fail/NA
2.18If applicable, connect d/c current and install the
battery. Turn the unit off.
2.19Remove the battery.Note that both charge lights (green and amber) flash alternately.
2.20Replace the battery and the turn unit on.Note that the yellow charge light illuminates.
2.21Press the ANALYZE button.The SELECT DEFIB MODE message appears on the monitor. (For manual
2.22Move the selector switch to DEFIB. Select 2J.
Press the CHARGE button.
2.23Press and hold the ENERGY SELECT down
arrow.
2.24Press and release the ENERGY SELECT up
arrow 19 times.
2.25Press the CHARGE button.Note the display shows the unit charged up to 200J and the SHOCK button
2.26Press the SHOCK button.The unit discharges and the SHOCK button is no longer lit. A 15 second strip
The CHARGER ON indicator lights.
The amber or green lights illumi nate.
The yellow light indicates the battery is being charged. The green light
indicates the battery is fully charged to present capacity.
If both lights flash ON/OFF, the unit is defective or no battery is installed.
devices.)
The display shows that the unit is charging. The SHOCK button lights when the
unit is charged. Ready tone for DEFIB sounds.
Unit discharges internally and selected energy decrements to 1J.
chart automatically prints, displaying the number of joules delivered (if
equipped with recorder and configured to print event).
o o o
o o
o o
o o o
o o o
o o o
o o o
o o o
o o o
Record your results on the Maintenance Test Checklist.
1-89650-1450-01 Rev. A
3.0 3, 5, and 12 Leads Test
Tools Needed: 3 lead, 5 lead, and 12 lead cables. Test each cable separately.
Test Setup:
1. The E Series unit must be config ured to displ ay ECG LEAD OFF message.
2. Connect the lead wires appropriate to each test to the Fluke Biomedical Impulse 4000 or equival en t (Symbio CS1201).
Do this...Observe this...Pass/Fail/NA
E Series Service Manual
3.1Turn the selector switch to MONITOR. Select
leads.
3.2Disconnect one lead from the simulator. The ECG LEAD OFF message displays within 3 seconds (if configured).
3.3Reconnect the lead. Repeat step 3.2 with the
remaining leads.
3.4Repeat 3.2 and 3.3 for 5 lead and 12 lead
cables.
Record your results on the Maintenance Tests Checklist.
NO ECG LEADS OFF message displayed.
Wait for ECG LEAD OFF message to clear from the display (if configured).
NOTE: If heart rate alarm sounds, press and hold the ALARM SUSPEND
button for 4 seconds to disable the alarms.
NOTE: When testing the 12 lead cable, the ECG LEAD OFF message displays
when you pull off a limb lead. When you pull off a V lead, the ECG VX LEAD OFF message displays where “X” is the number between 1 and 6.
o o o
o o o
o o o
o o o
9650-1450-01 Rev. A1-9
E Series Service Manual
4.0 Power Supply Test (Optional)
Tools Needed:
• 2 red miniature alligator to miniature alligator leads.
• 1 black miniature alligator to miniature alligator test lead.
• DC power supply (15 Amp minimum).
• 0.1Ω 1%resistor (¼W or greater).
• 1000Ω 1% ¼W resistor.
• Fluke 75 multimeter or equivalent.
Test Setup:
1. Make sure the unit an d power supply are turned off.
2. Connect one end of the black lead to the “-” terminal in the battery well.
3. Connect the other end of the black lead to the “-” terminal of the power supply.
4. Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it. Connect the other end of the red lead
to the “+” terminal of the power supply.
5. Set the power supply voltage to 7V.
CautionBe sure to connect the power supply properly to the E Series battery well terminals or damage to the unit may result. Do NOT raise the
power supply voltage above 12V.
1-109650-1450-01 Rev. A
5HG
E Series Service Manual
?
%ODFN
Do this...Observe this...Pass/Fail
4.1Turn the selector switch to MONITOR.
(For AED units, turn the selector switch to ON
and select Manual mode.)
4.2Turn the unit off.
4.3Adjust the power supply voltage to 10.3V and
turn the selector switch to MONITOR (for AED
units, turn the selector switch to ON).
4.4Low Battery Test
Set voltage to 9.9V.
%DWWHU\:HOO
The unit should not turn on.
The unit should turn on.
No LOW BATTERY message displays.
$PS
6XSSO\
?
o o
o o
o o
4.5Set voltage to 9.4V.LOW BATTERY message displays within 30 seconds.
4.6Shut Down Voltage Test
Set voltage to 8.5V.
Record your results on the Maintenance Tests Checklist.
9650-1450-01 Rev. A1-11
Unit should shut off within 30 seconds.
o o
o o
E Series Service Manual
Test Setup:
1. Remove red lead from power supply and connect to 0.1Ω resistor.
2. Connect other end of resistor to “+” terminal of power supply usi ng a second red lead.
3. Connect multimeter across the resistor.
4. Set voltage scale (if DVM is not autoranging) to 220 mV
5HG
%DWWHU\:HOO
?
'00
?
%ODFN
Do this...Observe this...Pass/Fail/NA
4.7System Current Test
Set power supply to 10.3V.
4.8Turn the selector switch to MONITOR.
(For AED units, turn the selector switch to ON
and select Manual mode.)
4.9Turn unit off.
Record your results on the Maintenance Tests Checklist.
$PS
6XSSO\
?
Voltage across resistor should be 116 mV or less (< 1.16 A of ON current).
o o o
1-129650-1450-01 Rev. A
Test Setup for Off Current Test:
1. Remove 0.1Ω resistor and replace with 1K Ω.
2. Connect DMM across resistor.
3. Set voltage scale to DCV.
4. Measure voltage across resistor.
Do this...Observe this...Pass/Fail
E Series Service Manual
4.10Off Current Test
Measure across resistor with unit turned off.
Record your results on the Maintenance Tests Checklist.
Voltage should be less than 450 mV (<450 μA of current).
o o
9650-1450-01 Rev. A1-13
E Series Service Manual
5.0 Leakage Current Test
Tools Needed: See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Test Setup: See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with accessories: MFC,
external paddles, and anterior/posterior paddles.
Maximum Leakage Acceptance Limits
Normal ConditionSingle Fault Condition*
ECG10μΑ50μΑ
MFC100μΑ100μΑ
Earth500μΑ1000μΑ
*Single fault considered AC mains on applied part.
1-149650-1450-01 Rev. A
6.0 Paddles Test (If applicable)
Tools Needed: None.
Test Setup: If applicable, connect the universal cable to the paddles and place the paddles in paddle wells.
Do this...Observe this...Pass/Fail/NA
E Series Service Manual
6.1Turn the selector switch to DEFIB. Press and
hold the ENERGY DOWN button on the
sternum paddle.
6.2Press and release the ENERGY UP button on
the sternum paddle for each setting.
6.3Press and release the RECORDER button on
the sternum paddle.
6.4Select 30J using the paddle ENERGY button.
Press the CHARGE button on the Apex paddle.
6.5Press and release the APEX SHOCK button.No discharge.
6.6Press and release the STERNUMSHOCK
button.
6.7Press and hold both paddles SHOCK buttons.The unit discharges. The TEST OK message displays and the red LED turns
Record your results on the Maintenance Tests Checklist.
The recorder turns on. Press and release again to turn off.
The unit charges to 30J, then the red LED charge indicator illuminates and the
charge tone sounds. (Note that the front panel shock button does not
illuminate).
No discharge.
off. The recorder runs.
o o
o o
o o o
o o
o o
o o
o o
9650-1450-01 Rev. A1-15
E Series Service Manual
7.0 Heart Rate Display Test
Tools Needed:
• Calibrated ECG simulator with 60Hz sine wave output capability.
• Mini-phone plug for measuring output signal from 1 Volt ECG OUT jack (optional).
• ECG Cable (3 or 5 leads).
Test Setup:
1. Turn the selector switch to MONITOR. Press LEAD button until “I” displays.
2. Connect the ECG leads to the Fluke Biomedical Impulse 4000 or equivalent.
3. Connect the ECG cable to the unit.
Do this...Observe this...Pass/Fail/NA
7.1Set the ECG Simulator to 120BPM.The Heart Rate displays as 120 +/- 2 bpm
Record your results on the Maintenance Tests Checklist.
o o o
1-169650-1450-01 Rev. A
8.0 Calibrating Pulses on Strip Chart Test
Tools Needed: None
Test Setup: None.
Do this...Observe this...Pass/Fail/NA
8.1Press the RECORDER button.
E Series Service Manual
8.2Press and hold SIZE button to activate the
calibration signal.
Record your results on the Maintenance Tests Checklist.
The strip chart displays a signal of 300 ppm with an amplitude of 10 mm +/- 1
mm. The signal also appears on the video display.
o o o
9650-1450-01 Rev. A1-17
E Series Service Manual
9.0 Notch Filter Test
Tools Needed: Fluke Biomedical Impulse 4000 (or equivalent).
Test Setup:
1. Connect the ECG cable to the Fluke Biomedical Impulse 4000 or equivalent.
2. Connect the ECG cable to the unit.
Do this...Observe this...Pass/Fail/NA
9.1Turn the selector switch to MONITOR mode.
(For AED units, turn the selector switch to ON
and select Manual mode.)
9.2Select lead I, size 3x.
Select 60Hz (or 50 Hz for a 50Hz unit) on the
Fluke Biomedical Impulse 4000.
9.3Press RECORDER button.Verify that the waveform amplitude on the strip chart is less than 1.5 mm.
9.4Turn the ECG simulator off.
Record your results on the Maintenance Tests Checklist.
1-189650-1450-01 Rev. A
o o o
10.0 Heart Rate Alarm Test
Tools Needed: Fluke Biomedical Impulse 4000
Do this...Observe this...Pass Fail/NA
E Series Service Manual
10.1Turn the selector switch to MONITOR mode.
(For AED units, turn the selector switch to ON
and select Manual mode.)
Connect the ECG leads to the Fluke Biomedical
Impulse 4000. Set the simulator to 120 BPM
and the defibrillator to lead II.
10.2Press ALARMS.The alarm menu displays.
10.3Press SELECT PARAM softkey until ECG HR
displays.
10.4Press INC> for state.Cursor scrolls through ENABLE, AUTO and DISABLE.
10.5Press DEC>for state.Cursor scrolls through ENABLE, DISABLE, AND AUTO.
10.6Press INC> until ENABLE displays.ENABLE displays.
10.7Set LOW limit to 30, HIGH limit to 150 then,
press the RETURN softkey.
10.8Press ALARM SUSPEND button.No alarm sounds.
10.9Remove a lead wire from the Fluke Biomedical
Impulse 4000.
10.10Reattach ECG Lead wire to Fluke Biomedical
Impulse 4000 and hold the ALARM SUSPEND
button on unit for 4 seconds.
10.11Press the ALARM SUSPEND button.Alarm is enabled. Alarm symbol (without “X”) displays.
Lead II message displays.
NSR ECG at 120 BPM +/- 2 displayed.
Cursor scrolls through parameters.
MONITOR displays.
The alarm symbol flashes and the heart symbol stops flashing. The ECG LEAD
OFF alarm tone sounds. Recorder prints a stripchart showing a low heart rate,
if enabled.
The alarm symbol has an “X through it.
The heart symbol flashes with each QRS wave.
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
o o o
10.12Set simulator to 160 BPM or higher. Heart Rate Value is highlighted, alarm tone sounds, the alarm and the heart
symbol both flash.
10.13Press the ALARM SUSPEND button in the unit.Alarm is suspended for 90 seconds. The alarm symbol has an “X” through it.
The heart symbol flashes with each QRS wave.
9650-1450-01 Rev. A1-19
o o o
o o o