ZOLL AED Plus Trainer User Manual

For further information about the functioning of the clinical AED Plus, see the AED Plus Operator’s Guide and Administrator’s Guide.
The Remote Controller lights indicate the current training settings while the Trainer2 simulates the actual AED Plus behavior.
The Remote Controller will turn itself off after 10 seconds. The Remote Controller will remember the last scenario setting which reappears the next time the View/Enter Button is pressed.
NOTE
SETUP and TEST
1
SELECT SCENARIO SETTINGS
2
Scenario 3: SHOCK - NO SHOCK
Scenario 2: NO SHOCK - SHOCK
Scenario 1: SHOCK - SHOCK
Manual Scenario Mode
CONTROLS / INDICATORS
Handheld Remote Controller:
PUSH HARDER prompt
(only when CPR Metronome is beeping).
NO SHOCK advised on next heart analysis
(manual mode only).
Pads attached/not attached to victim’s bare chest
(toggle-switch).
View/Enter scenario
Show current scenario selection (press once) Start scenario (press again)
Move Up
to next scenario.
Scenario 4: NO SHOCK - NO SHOCK
Set-up Controls:
Training Controls:
GOOD COMPRESSIONS prompt
(only when CPR Metronome is beeping).
Pause (resume) current scenario.
End current scenario.
NOTE
Install 4 “D” alkaline batteries in the AED Plus Trainer2.
Trainer2 unit: Action:
Press On/Off Button to turn Trainer2 on. Green light turns on, indicating Trainer2 is
ready for use.
Remote Controller:
Install 2 “AA” batteries in handheld remote. Turn the Remote Controller on by pushing the View/Enter Button.
Point Remote Controller at unit and press View/Enter Button again. Green light on Remote Controller will show Manual Scenario Mode is selected.
Manual Scenario Mode View/Enter Button
Trainer2 unit: Action:
Turn Trainer2 unit off (hold down On/Off Button). Setup test is now complete. Trainer2 unit and
Remote Controller are ready for use.
Trainer2 unit:
Action:
Press the On/Off Button to turn on the Trainer2.
Action:
Remote Controller: Action:
Use Move Up Button to select desired scenario.
Use CPR Metronome On/Off Button to turn Real CPR Help
®
(beeping) on or off for this
scenario.
Press the View/Enter Button once to transmit the selected Training Scenario to the Trainer2 unit.
CPR PROMPTS
Remote Controller: Action:
The Instructor uses the Remote Controller to control prompts during a Training Scenario’s CPR period. The instructor can use the following buttons to simulate the use of an AED Plus defibrillator with Real CPR Help.
Remote Controller: Action:
In all scenarios (manual and automatic), the rescue can be paused by pressing the Pause/Resume Button once. Press it a second time to resume the scenario.
Trainer2 unit: Action:
The Push Harder and Good Compressions Buttons are only valid during the CPR intervall. The red light on the Trainer2’s status window emits one long flash if the prompt is currently valid.
If the Push Harder or Good Compressions Buttons are pushed when the Trainer2 is not beeping, the unit’s red light emits three quick
flashes, indicating that the prompt is currently
invalid.
In all scenarios (manual and automatic), the rescue can be ended by pressing the End Button once.
NOTE
On Off
On Off
On Off
The buttons above the line on the handheld Remote Controller are Set-up Controls, which the instructor uses to select a training scenario (an LED lights next to the selected scenario). The buttons below the line are Training Controls, which the instructor uses to introduce events into the training scenario.
CPR Metronome On/Off Button
Enables the CPR metronome to beep for the selected Training Scenario. The LED above the button lights to indicate that the CPR metronome is enabled. You should enable the CPR metronome whenever CPR-D padz are connected to the Trainer2.
When a CPR period occurs in the Training Scenario, the CPR metronome begins to beep at 60 compressions per minute (CPM), then automatically accelerates to 100 CPM after a few seconds. Metronome beeps are independent of the student's actual compression rate and continue until either the CPR interval ends or the the instructor ends the scenario.
ZOLL AED Plus
On Off
®
®
Trainer2:
Press to turn on Trainer2. Press again to start current scenario (as selected using Remote Controller). Press and hold down to turn off Trainer2.
Green light remains on = Trainer2 is on and ready for use. Green light off = Trainer2 is turned off or has no battery power. Red Light steady = Last command from Remote Controller accepted. Red Light 3 flashes = Last command from Remote Controller is invalid.
3
SCENARIO CONTROLS
Remote Controller: Action:
Analysis - VF
3a
VF (Ventricular Fibrillation-shock advised) is the default simulated heart rhythm in Manual Scenario.
For the next analysis to result in NO SHOCK ADVISED, push the No Shock button.
Once the No Shock button has been pushed, the next
NOTE
simulated heart analysis will advise NO Shock. This setting cannot be reversed for the next analysis.
In all scenarios the Pads Button must be pressed once to proceed with analysis. If pressed again, it will begin to prompt for attaching pads.
CPR Prompts
Push Harder Button Pressing this button
the Trainer2 to issue the PUSH HARDER voice prompt. Instructors should press this button when a student's chest compressions are too shallow.
Good Compressions Button Pressing this button during the CPR period causes
the Trainer2 to issue the GOOD COMPRESSIONS voice prompt. Intructors should press this button to provide students with positive feedback when they respond to the PUSH HARDER prompt by improving the depth of chest compressions
Accurately imitating the student’s AED Plus unit requires
NOTE
instructors be familiar with the protocol currently followed by the AED Plus for PUSH HARDER and GOOD COMPRESSIONS
during the CPR period causes
prompts.
Press the View/Enter Button once to turn on the Remote Controller.
3b
PAUSE/END SCENARIO
Additional Remote Controllers are available. Contact ZOLL Medical Corporation or your distributor.
1. Keep Trainer2 and Remote Controller clean, undamaged, and free of excessive wear.
2. Verify that there are no cracks or loose parts in Trainer2 and Remote Controller.
3. Replace Training electrodes as necessary.
MAINTENANCE
Trainer2 does not respond (green light in status window is not lit) when On/Off Button is pushed
TROUBLESHOOTING
Clean the Trainer2 unit and remote controller with soft, damp cloth using soap and water, then wipe dry immediately.
Do not immerse any parts in water. Do not clean with ketones (e.g., MEK, acetone). Avoid using abrasives on any surface.
Do not sterilize the Trainer2 or the Remote Controller.
CLEANING
TRAINING USE ONLY - NOT FOR CLINICAL USE
REF 9650-0090-01 Rev. C
Product intended for indoor use only.
Do not attempt to plug AED Plus or AED Pro simulator into connector of Trainer2.
This device is for training only. It will not deliver defibrillation energy.
Replace all four D batteries in Trainer2 unit.
Remote Controller does not respond (scenario
selection is not lit) when View/Enter Button is pushed.
Replace both AA batteries in Remote Controller.
With new batteries in both, Trainer2 fails to respond to Remote Controller.
AED Plus Trainer2 problem/ display:
Action:
Contact ZOLL Technical Service.
Do not attempt to take Trainer2 or Remote Controller apart.
Trainer2 does not beep or react to Remote Controller during CPR interval.
Re-set scenario and make certain that CPR metronome light is on.
Trainer2 announces breathing prompts, OPEN AIRWAY and CHECK BREATHING.
Unit is configured for ERC. Change to AHA configuration. (See Switching AHA/ERC Configuration section of this Operator’s Guide).
SWITCHING AHA/ERC CONFIGURATION
American Heart Association (AHA) configuration: No prompts for breathing.
European Resuscitation Council (ERC) configuration: Contains breathing prompts OPEN AIRWAY and CHECK BREATHING.
To switch between either AHA or ERC configuration, do the following:
With the Trainer2 unit off, hold down the Shock Button. Press On/Off Button once. Green light will illuminate at status window to indicate AHA cofiguration (prompt sequence will contain no breathing prompts).
Press Shock Button again. Green light will turn off and red light will illuminate at status window to indicate ERC configuration (prompt sequence will contain breathing ).
Press Shock Button again to revert to AHA configuration (green light on and red light off).
When configured as desired (green = AHA, red = ERC), hold down On/Off Button to turn unit off to store configuration.
On Off
SCENARIOS
®
Simulated heart rhythm: Student Response:
1
2
3
4
VF 1 VF 2 VF 3 VF 4 Done
Simulated heart rhythm: Student Response:
ASYSTOLE 1 ASYSTOLE 2 VF 1 VF 2 Done
Simulated heart rhythm: Student Response:
VF 1 VF 2 NSR 1 NSR 2 Done
Simulated heart rhythm: Student Response:
ASYSTOLE 1 ASYSTOLE 2 ASYSTOLE 3 ASYSTOLE 4 Done
Shock CPR Shock CPR
No Shock CPR Shock CPR
Shock CPR No Shock CPR
No Shock CPR No Shock CPR
Automated External Defibrillator Trainer2
AED Plus
Trainer2
Operator’s Guide
AHA/ERC 2010
ACCESSORIES
You can order the following Trainer2 accessories from the ZOLL Customer Service Department or your distributor:
Item: REF
1. CPR-D padz® Training Electrodes (1 Set) 8900-0804-01
2. CPR-D padz Training Gel (5 Pair) 8900-0803-01
3. stat-padz
4. pedi-padz
5. Trainer2 Remote Controller 8008-0007
6. Trainer Carrying Case 8000-0375-01
7. Trainer Demo Manikin 8000-0835-01
® Training Electrodes (1 Set) 8900-0805-01
® II Training Electrodes (case of 6) 8900-000861-01
WARNINGS
Remove battery compartment doors only to replace batteries.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in accompanying documents.
Portable and Mobile RF Communications equipment can affect Medical Electrical Equipment.
NOTE
SYMBOLS
The following symbols appear on the ZOLL AED Trainer2:
Authorized representative in the European Community.
Return to a collection site intended for waste electrical and electronic equipment (WEEE). Do not dispose of in unsorted trash.
Manufacturer.
Temperature limitation.
ZOLL, AED Plus, Real CPR Help, CPR-D padz, stat-padz and pedi-padz are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries.
Copyright 2012 by ZOLL Medical Corporation
Conformité Européene - Complies with Low Voltage Directive 73/23/EEC.
Caution.
REF Catalogue Number.
. All rights reserved.
For technical support or repair:
ZOLL Medical Corporation
269 Mill Road Chelmsford, MA USA 01824-4105
Attn:
Technical Service Department
Tel: 978-421-9655 800-348-9011 Fax: 978-421-0001
ZOLL Medical Europe
International Service: Contact your local distributor.
© 2012 ZOLL Medical Corporation
Newtonweg 18 6662 PV ELST The Netherlands
Tel: +31 (0) 481 366410 Fax: +31 (0) 481 366411
Loading...