For further information about the functioning of
the clinical AED Plus, see the AED Plus Operator’s Guide
and Administrator’s Guide.
The Remote Controller lights indicate the current training
settings while the Trainer2 simulates the actual AED Plus behavior.
The Remote Controller will turn itself off after 10 seconds.
The Remote Controller will remember the last scenario setting which
reappears the next time the View/Enter Button is pressed.
NOTE
SETUP and TEST
1
SELECT SCENARIO SETTINGS
2
Scenario 3: SHOCK - NO SHOCK
Scenario 2: NO SHOCK - SHOCK
Scenario 1: SHOCK - SHOCK
Manual Scenario Mode
CONTROLS / INDICATORS
Handheld Remote Controller:
PUSH HARDER prompt
(only when CPR Metronome is beeping).
NO SHOCK advised on next heart analysis
(manual mode only).
Pads attached/not attached to victim’s bare chest
(toggle-switch).
View/Enter scenario
Show current scenario selection (press once)
Start scenario (press again)
Move Up
to next scenario.
Scenario 4: NO SHOCK - NO SHOCK
Set-up Controls:
Training Controls:
GOOD COMPRESSIONS prompt
(only when CPR Metronome is beeping).
Pause (resume) current scenario.
End current scenario.
NOTE
Install 4 “D” alkaline batteries in the AED Plus Trainer2.
Trainer2 unit: Action:
Press On/Off Button to turn Trainer2 on.
Green light turns on, indicating Trainer2 is
ready for use.
Remote Controller:
Install 2 “AA” batteries in handheld remote.
Turn the Remote Controller on by pushing the
View/Enter Button.
Point Remote Controller at unit and press
View/Enter Button again. Green light on Remote
Controller will show Manual Scenario Mode is
selected.
Manual Scenario Mode
View/Enter Button
Trainer2 unit: Action:
Turn Trainer2 unit off (hold down On/Off Button).
Setup test is now complete. Trainer2 unit and
Remote Controller are ready for use.
Trainer2 unit:
Action:
Press the On/Off Button to turn on the Trainer2.
Action:
Remote Controller: Action:
Use Move Up Button to select desired scenario.
Use CPR Metronome On/Off Button to turn
Real CPR Help
®
(beeping) on or off for this
scenario.
Press the View/Enter Button once to transmit the
selected Training Scenario to the Trainer2 unit.
CPR PROMPTS
Remote Controller: Action:
The Instructor uses the Remote Controller to control prompts during a
Training Scenario’s CPR period. The instructor can use the following buttons
to simulate the use of an AED Plus defibrillator with Real CPR Help.
Remote Controller: Action:
In all scenarios (manual and automatic), the rescue
can be paused by pressing the Pause/Resume
Button once. Press it a second time to resume the
scenario.
Trainer2 unit: Action:
The Push Harder and Good Compressions
Buttons are only valid during the CPR intervall.
The red light on the Trainer2’s status window emits
one long flash if the prompt is currently valid.
If the Push Harder or Good Compressions
Buttons are pushed when the Trainer2 is not
beeping, the unit’s red light emits three quick
flashes, indicating that the prompt is currently
invalid.
In all scenarios (manual and automatic), the rescue
can be ended by pressing the End Button once.
NOTE
On
Off
On
Off
On
Off
The buttons above the line on the handheld Remote Controller are Set-up
Controls, which the instructor uses to select a training scenario (an LED lights
next to the selected scenario). The buttons below the line are Training Controls,
which the instructor uses to introduce events into the training scenario.
•
CPR Metronome On/Off Button
Enables the CPR metronome to beep for the selected
Training Scenario. The LED above the button lights to
indicate that the CPR metronome is enabled. You should
enable the CPR metronome whenever CPR-D padz
are connected to the Trainer2.
When a CPR period occurs in the Training Scenario, the
CPR metronome begins to beep at 60 compressions
per minute (CPM), then automatically accelerates to
100 CPM after a few seconds. Metronome beeps are
independent of the student's actual compression rate
and continue until either the CPR interval ends or the
the instructor ends the scenario.
•
•
•
•
•
ZOLL AED Plus
On
Off
®
®
Trainer2:
Press to turn on Trainer2.
Press again to start current scenario
(as selected using Remote Controller).
Press and hold down to turn off Trainer2.
Green light remains on = Trainer2 is on and ready for use.
Green light off = Trainer2 is turned off or has no battery power.
Red Light steady = Last command from Remote Controller accepted.
Red Light 3 flashes = Last command from Remote Controller is invalid.
3
SCENARIO CONTROLS
Remote Controller: Action:
Analysis - VF
3a
VF (Ventricular Fibrillation-shock advised) is the
default simulated heart rhythm in Manual Scenario.
For the next analysis to result in NO SHOCK ADVISED,
push the No Shock button.
Once the No Shock button has been pushed, the next
NOTE
simulated heart analysis will advise NO Shock. This setting cannot be
reversed for the next analysis.
In all scenarios the Pads Button must be pressed
once to proceed with analysis. If pressed again, it
will begin to prompt for attaching pads.
CPR Prompts
Push Harder Button
Pressing this button
the Trainer2 to issue the PUSH HARDER voice prompt.
Instructors should press this button when a student's
chest compressions are too shallow.
Good Compressions Button
Pressing this button during the CPR period causes
the Trainer2 to issue the GOOD COMPRESSIONS voice
prompt. Intructors should press this button to provide
students with positive feedback when they respond
to the PUSH HARDER prompt by improving the depth
of chest compressions
Accurately imitating the student’s AED Plus unit requires
NOTE
instructors be familiar with the protocol currently followed by the
AED Plus for PUSH HARDER and GOOD COMPRESSIONS
during the CPR period causes
prompts.
Press the View/Enter Button once to turn on the
Remote Controller.
3b
PAUSE/END SCENARIO
Additional Remote Controllers are available.
Contact ZOLL Medical Corporation or your distributor.
1. Keep Trainer2 and Remote Controller clean, undamaged,
and free of excessive wear.
2. Verify that there are no cracks or loose parts in Trainer2 and
Remote Controller.
3. Replace Training electrodes as necessary.
MAINTENANCE
Trainer2 does not respond
(green light in status window
is not lit) when On/Off
Button is pushed
TROUBLESHOOTING
Clean the Trainer2 unit and remote controller with soft, damp
cloth using soap and water, then wipe dry immediately.
Do not immerse any parts in water.
Do not clean with ketones (e.g., MEK, acetone).
Avoid using abrasives on any surface.
Do not sterilize the Trainer2 or the Remote Controller.
CLEANING
TRAINING USE ONLY - NOT FOR CLINICAL USE
REF 9650-0090-01 Rev. C
Product intended for indoor use only.
Do not attempt to plug AED Plus or AED Pro
simulator into connector of Trainer2.
This device is for training only. It will not deliver defibrillation energy.
Replace all four D batteries
in Trainer2 unit.
Remote Controller does not
respond (scenario
selection is not lit)
when View/Enter
Button is pushed.
Replace both AA batteries in
Remote Controller.
With new batteries in both,
Trainer2 fails to respond to
Remote Controller.
AED Plus Trainer2 problem/
display:
Action:
Contact ZOLL
Technical Service.
Do not attempt to take Trainer2 or Remote Controller apart.
Trainer2 does not beep or
react to Remote Controller
during CPR interval.
Re-set scenario and make
certain that CPR metronome
light is on.
Trainer2 announces breathing
prompts, OPEN AIRWAY and
CHECK BREATHING.
Unit is configured for ERC.
Change to AHA configuration.
(See Switching AHA/ERC
Configuration section of this
Operator’s Guide).
SWITCHING AHA/ERC CONFIGURATION
American Heart Association (AHA) configuration: No prompts for
breathing.
European Resuscitation Council (ERC) configuration: Contains breathing
prompts OPEN AIRWAY and CHECK BREATHING.
To switch between either AHA or ERC configuration, do the following:
With the Trainer2 unit off, hold down the Shock Button. Press On/Off Button
once. Green light will illuminate at status window to indicate AHA
cofiguration (prompt sequence will contain no breathing prompts).
Press Shock Button again. Green light will turn off and red light will illuminate
at status window to indicate ERC configuration (prompt sequence will
contain breathing ).
Press Shock Button again to revert to AHA configuration (green light on
and red light off).
When configured as desired (green = AHA, red = ERC), hold down On/Off
Button to turn unit off to store configuration.
On
Off
SCENARIOS
®
Simulated heart rhythm: Student Response:
1
2
3
4
VF 1
VF 2
VF 3
VF 4
Done
Simulated heart rhythm: Student Response:
ASYSTOLE 1
ASYSTOLE 2
VF 1
VF 2
Done
Simulated heart rhythm: Student Response:
VF 1
VF 2
NSR 1
NSR 2
Done
Simulated heart rhythm: Student Response:
ASYSTOLE 1
ASYSTOLE 2
ASYSTOLE 3
ASYSTOLE 4
Done
Shock
CPR
Shock
CPR
No Shock
CPR
Shock
CPR
Shock
CPR
No Shock
CPR
No Shock
CPR
No Shock
CPR
Automated External Defibrillator Trainer2
AED Plus
Trainer2
Operator’s Guide
AHA/ERC 2010
ACCESSORIES
You can order the following Trainer2 accessories from the ZOLL
Customer Service Department or your distributor:
Item: REF
1. CPR-D padz® Training Electrodes (1 Set) 8900-0804-01
2. CPR-D padz Training Gel (5 Pair) 8900-0803-01
3. stat-padz
4. pedi-padz
5. Trainer2 Remote Controller 8008-0007
6. Trainer Carrying Case 8000-0375-01
7. Trainer Demo Manikin 8000-0835-01
® Training Electrodes (1 Set) 8900-0805-01
® II Training Electrodes (case of 6) 8900-000861-01
WARNINGS
Remove battery compartment doors only to replace batteries.
Medical Electrical Equipment needs special precautions regarding
EMC and needs to be installed and put into service according to the
EMC information provided in accompanying documents.
Portable and Mobile RF Communications equipment can affect
Medical Electrical Equipment.
NOTE
SYMBOLS
The following symbols appear on the ZOLL AED Trainer2:
Authorized representative in the European Community.
Return to a collection site intended for waste electrical and electronic
equipment (WEEE). Do not dispose of in unsorted trash.
Manufacturer.
Temperature limitation.
ZOLL, AED Plus, Real CPR Help, CPR-D padz, stat-padz and pedi-padz are trademarks or
registered trademarks of ZOLL Medical Corporation in the United States and/or other countries.
Copyright 2012 by ZOLL Medical Corporation
Conformité Européene - Complies with Low Voltage Directive 73/23/EEC.
Caution.
REF Catalogue Number.
. All rights reserved.
For technical support or repair:
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA USA 01824-4105
Attn:
Technical Service Department
Tel: 978-421-9655 800-348-9011
Fax: 978-421-0001
ZOLL Medical Europe
International Service: Contact your
local distributor.
© 2012 ZOLL Medical Corporation
Newtonweg 18
6662 PV ELST
The Netherlands
Tel: +31 (0) 481 366410
Fax: +31 (0) 481 366411