Xylem WL16 User Manual

Global
Water
800-876-1172 • globalw.com
Covers WL16S and WL16U Data Loggers
11/30/09 01-404 Publication Number 38300112
WL16 Data Logger
Global
Instrumentation, Inc.
151 Graham Road
P.O. Box 9010
College Station, TX 78742-9010
T: 800-876-1172
Int’l: (979) 690-5560, F: (979) 690-0440
globalw@globalw.com
Water
Global
Water
800-876-1172 • globalw.com
I.
Checklist
3
II.
Inspection
3
III.
Warranty
3
IV.
WL16 Specifications
4
V.
System Requirements
6
VI.
Software Installation
6
VII.
Direct Connection
9
VIII.
Modem Connection
10
IX.
Main Screen
13
X.
On-Line Help Files
17
XI.
Get History
18
XII.
Setup Menus
19
XIII.
General Setup
20
XIV.
Analog Setup
23
XV.
Programming
27
XVI.
Modem Cable Diagram
28
XVII.
Adjusting the Zero Reference Level
29
XVIII. Maintenance 32
XIX.
Troubleshooting
32
XX.
Customer Support
33
Congratulations on your purchase of a Global Water WL16 Data Logger. This instrument has been quality t ested and approved for providing accurate and reliable measurements. We are confident that you will find the logger to be a valuable asset to your applications. Should you require assistance, our technical st aff will be happy to help.
Table of Contents
Copyright © Global Water Instrumentation, Inc. 2008
Global
Water
800-876-1172 • globalw.com
I. Checklist:
a. WL16S (Serial) or WL16U (USB) Data Logger b. WL16 Data Logger Manual c. WL16S comes with a RS-232 Serial Cable d. WL16U comes with a USB Cable, Type A to B e. Global Logger Interface Software on CDROM
II. Inspection:
a. Your Data Logger was carefully inspected and certified by our Quality
III. Warranty:
a. Global Water Instrumentation, Inc. warrants that its products are
b. The warranty begins on the date of your invoice.
Assurance Team before shipping. If any damage has occurred during shipping; please notify Global Water Instrumentation, Inc. and file a claim with the carrier involved.
free from defects in material and workmanship under normal use and service for a period of one year from date of shipment from factory. Global Water’s obligations under this warranty
are limited to, at Global Water’s option: (I) replacing or (II) repairing; any products determined to be defective. In no case shall Global Water’s liability exceed the products original purchase price. This warranty does not apply to any equipment that has been repaired or altered, except by Global Water Instrumentation, Inc., or which has been subject to misuse, negligence or accident. It is expressly agreed that this warranty will be in lieu of all warranties of fitness and in lieu of the warranty of merchantability.
Global
Water
800-876-1172 • globalw.com
IV. WL16 Specifications:
Logger:
Power Supply: Two 9VDC Alkaline batteries standard
8 VDC Min. to 24.0 VDC Absolute Maximum Standby Current: 65uA Typical
Logging Current: 5mA Typical + sensor current Operating Temperature Industrial, -40ºC to +85ºC (Battery may not apply) Analog Sensor Inputs:
4-20mA
Resolution: 12-Bit, 4096 Steps
1 channel + battery voltage monitor,
nd
2
channel reserved for temperature option
Sensor Warm-up Time: Programmable, 0-60 Seconds Sample Modes: Fixed Interval: Programmable from 1 Sec. to >1 Year
High Speed: 10 Samples per second
Logarithmic Sample Rate (Approximation)
Exception (Log only on deviation from previous reading) Storage Capacity: 81,759 Recordings for two analog inputs, battery voltage
monitor and date/time stamp Communication Ports: WL16S: RS-232 4-pin circular connector
WL16U: USB Type B
Selectable Baud Rates:
Auto Detect or fixed, rates at 9600, 19200, 28800,
38400, 57600,
115200
Housing: 1 7/8 diameter x 11 1/2 long, Stainless Steel and UV
protected
PVC
Weight: 450g (16oz), Logger only with batteries
Global
Water
800-876-1172 • globalw.com
Sensor:
Sensor Element: Silicone Diaphragm, Wet/Wet Transducer Range: Level: 0-3', 0-15', 0-30', 0-60', 0-120', 0-250'
Optional Temperature: 32°-122°F (0°-50°C) Linearity and Hysteresis: ±0.1% FS
Accuracy: Level: ±0.1% FS at constant temperature
±0.2% over 32° to 70°F range
Optional Temp: Smaller of 0.5°F or ±1% of reading Overpressure: 2 x full scale range
Resolution: Infinite (Analog) Outputs: 4-20mA ±1mA at full scale
Optional Temperature: 0-10mA ±1mA FS Supply Voltage: 10-36VDC Current Draw: Sum of sensor outputs. Warm Up Time: 10mS Min, 3 sec. recommended Operating Temperature: 0° (Not Frozen) to +185°F Compensated Range: 32° to 70°F submerged, automatic barometric
compensation
Housing:
Material: WL400: 304L Stainless Steel
WL400-S: 316
Size: WL400: 7.5” long x 0.82" diameter
SS
WL400-S: 9” long x 1.0” diameter Weight: WL400: 110g (4 oz)
WL400-S: 250g (9oz)
Cable:
Conductors: 4 each 22 AWG Jacket Material: 87A shore hardness Polyurethane Optional jacket: Fluorinated Ethylene Propylene (FEP) Teflon Cable O.D.:
7.8mm (0.307”) Vent tube: HD Polyethylene Shield: Aluminum Mylar Temperature range:
-30 to 85°C (-22 to 185°F) Weight: ~65g/m (0.7 oz/ft)
Global
Water
800-876-1172 • globalw.com
V. System Requirements:
Desktop or Laptop computer with:
1) Windows 98, ME, 2000, XP Vista or Windows 7 Operating Systems
2) CDROM Drive
3) RS-232 COM Port (for WL16S) or USB Port (for WL16U)
Note that the software provided with the WL16 data loggers is for running on desktop and laptop computers only. Software for many popular
Handheld PDA
devices is available from Global Water. Special cables may be required.
VI. Software Installation:
Global Logger Installation: This software is for use on Windows based desktop and laptop computers only. Insert the software installation CD and run the Setup
program. This will install the Global Logger application in the Program Files folder and automatically put the Global Logger icon on the desktop.
Serial communication (WL16S Only
): If you are using the RS-232 serial port to
access the data logger, no further action is needed. You can start using the logger by opening the Global Logger application program. Skip the next section.
USB Driver Installation (WL16U Only): With the installation CD in the drive
and DC power applied to the logger, plug the USB cable into the data logger to
CDROM
open the ADD New Hardware Wizard. The menus that follow will look different for
different operation systems but the procedure is the same. Do not allow the Hardware Wizard to automatically search for the driver. Instead, specify the location (CDROM). If a generic driver gets, or has been previously installed, it must
be uninstalled before proceeding by using the Add or Remove Programs section of the Control Panel. Instead, click the Search External Media box or Specify a
Location box and install the driver from the CDROM. Running the Setup program that installs the Global Logger software also installs a folder with the driver in it. This location can also be specified if the CDROM is not available; although, it is recommended that the driver be installed at the same time the Global Logger is installed.
Global
Water
800-876-1172 • globalw.com
WL16U Only The following menus apply to the Windows XP operating system.
1) Do not let Windows search for software. Click “No, not at this time” when asked if Windows Update can search for software.
Click Next.
2) Do not let Windows automatically install the software. Click “Install from a specific location”.
Click Next.
3) Specify the location of the driver. Make sure the Global Logger CDROM is in the CD drive and click “Search removable media” only.
Click Next.
Global
Water
800-876-1172 • globalw.com
WL16U Only
4) The Hardware Wizard searches for the driver.
If Windows advises that the software is not certified, click:
“Continue Anyway”.
This only means that Microsoft has not licensed the software. Second party software is allowed and no harm will come to your computer.
5) The first half of the installation is complete. The Hardware Wizard must make two passes.
Click Finish.
6) Two drivers might be required so the Hardware Wizard may repeat the exact same process again. As before; a: do not let the Windows Update do a search, b: tell Windows to search from a specific location instead of automatically installing software, c: tell Windows to search from removable media (CDROM).
7) The installation is complete. Click Finish.
Global
Water
800-876-1172 • globalw.com
8) To confirm the installation, look in the Device Manager and find the USB Serial Port. The COM setting following that will show the virtual COM port that has been automatically assigned to the driver. As you will see in the following sections, this COM port will be automatically detected by the Global Logger software so it is not necessary to note which port is being used by the driver.
VII. Direct Connection:
Important: This data logger supports an Auto Baud Rate Mode which automatically detects the communication speed. However; this communication software also supports loggers which may be programmed in a Fixed Baud Rate Mode. This is useful in telemetry applications, particularly when using Global Water data loggers with our Global Access radio modem software. The upcoming discussion assumes that the logger is programmed in auto detect mode. If you cannot connect to the logger and suspect that the logger may be using a fixed rate mode; try each connection speed separately, especially 38400 which is required when using the Global Access software.
There are 3 ways to connect to the WL16 data Loggers, direct connection through the RS-232 serial port (WL16S only) or USB port (WL16U only), as well as dial out through an installed modem (requires WL16S only with external serial modem).
Global
Water
800-876-1172 • globalw.com
- 10 -
To make a direct connection to the logger, click the “Direct” option and from the “Port” menu, select the port you wish to use.
For USB access (WL16U) select the USB serial port at the COM port the driver was assigned, which is automatically shown in the list as “Global Logger USB Device”.
Serial connection (WL16S) is generally done using the serial cable provided with the logger through the 9-pin COM1 connector at the back of the computer although COM2 can also be used.
Next, select the desired baud rate from the “Baud Rate” menu. Generally, the fastest rate of 115,200 is used although; long cables or other factors may require a slower setting. Click “Connect” to make the connection. If a connection is not made, an error
message will appear. Check the battery, communication cable,
installation
VIII. Modem Connection (WL16S Only):
and COM port settings.
driver
Access to the WL16S can be made through dial-out to a remote modem by clicking the “Modem” option. The connection from modem to logger requires an RS-232 serial modem, USB modems are not supported as an interface to the logger; therefore, the WL16U logger cannot be accessed by modem. Also, some switch settings must be made to the remote modem and an initialization file may be required to properly configure the host modem.