Xylem WQL-pH User Manual

Operating manual
pH Logger WQL-pH
PC Program WQL-Log
ba75821de01 10/2009
DRAFT: Monday, 2. November 2009
pH logger and PC program
to display and evaluate
measurement data from the logger
The latest version of the present operating manual can be found on the Internet under www.WTW.com.
Copyright © Weilheim 2009, WTW GmbH
Reproduction in whole - or even in part - is prohibited without the express written permission of WTW GmbH, Weilheim. Printed in Germany.
WQL-pH Contents
Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
1.1 Logger WQL-pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
1.2 PC program WQL-Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
3.1 Preparing the logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
3.1.1 General information on the installation of the battery and electrode . . . . . . . . 85
3.1.2 Inserting the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
3.1.3 Installing the electrode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
3.1.4 Mounting suspension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
3.2 Installing the WQL-Log program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
3.2.1 PC requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
3.2.2 Installation routine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
3.3 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
3.4 Connection types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
3.5 Starting the WQL-Log program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
4 Setting the display (language etc.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
4.1 Setting the language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
4.2 Setting the temperature unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
5 Calibrating for pH measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
5.1 Buffer sets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
5.2 Calibration settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .107
5.3 Calibrating with AutoCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
5.4 Viewing the calibration records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
6 Measuring directly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
7 Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
7.1 Typical sequence of a logging job (checklist) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
7.2 Setting up a logging job - setting parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
7.3 Setting the parameters and starting the logging job . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
7.4 Signal LED to indicate the operating conditions of the logger . . . . . . . . . . . . . . . . . . . . 121
7.5 Reading in data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122
7.6 Overview of the data window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
7.7 Clearing the logger memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
8 Processing measurement data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
8.1 Querying data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
8.2 Exporting data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
ba75821e01 10/2009
77
Contents WQL-pH
9 Info menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133
9.1 Database information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
9.2 Device information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
9.3 Program info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
10 Resetting the logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
11 Maintenance, cleaning, storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
11.1 General maintenance instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
11.2 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
11.3 Replacing the electrode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
11.4 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
11.4.1 Battery service life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
11.4.2 Battery replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
11.5 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
12 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
12.1 Calibration and measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
12.2 Communication of the logger and PC program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
13 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
13.1 Measurement characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
13.2 Application characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
13.3 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
13.4 Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
78
ba75821e01 10/2009
WQL-pH Overview
111
2
3
4
7
5
6
89
10
Control panel

1 Overview

1.1 Logger WQL-pH

Structure of the
logger
1 Protective hood 2 Electrode 3 Coupling ring 4 Shaft 5 Sleeve 6 Cap 7 Battery compartment with battery 8 Key button 9 Signal LED 10 USB socket 11 Blind plug
ba75821e01 10/2009
79
Overview WQL-pH
Recommended
fields of
application
pH measurements in wells, bore holes, rivers, water bodies and other surface water, drinking water

1.2 PC program WQL-Log

The WQL-Log PC program serves to operate a connected WTW data logger of the WQL series type. The WQL-Log PC program automatically recognizes the logger when it is connected to the USB interface and controls the menus according to the type and measured parameter.
The PC program can carry out the following functions:
z Calibrate the electrode z Set up and start a logging job z Measure directly: The logger is connected to the PC via USB and the PC
program displays the current measured values.
z Read in measurement data of the logging job: Consistent data copying to
a PC-based database.
z Database-supported evaluation of measurement data, export functions
The current version of the WQL-Log PC program is available on the Internet on the site of the manufacturer of your device.
80
ba75821e01 10/2009
WQL-pH Overview
ba75821e01 10/2009
Fig. 1-1 Overview of the communication of the logger and PC program
81
Overview WQL-pH
82
ba75821e01 10/2009
WQL-pH Safety

2Safety

This component operating manual provides special instructions that must be followed during the operation of the logger and PC program.
Always keep this operating manual in the vicinity of the logger and PC program.
Symbols
used
indicates notes that draw your attention to special features.
indicates cross-references to other documents, e.g. application reports, operating manuals of electrodes, etc.
ba75821e01 10/2009
83
Safety WQL-pH
84
ba75821e01 10/2009
WQL-pH Commissioning

3 Commissioning

3.1 Preparing the logger

3.1.1 General information on the installation of the battery and electrode

z Normally, all screw joints of the logger housing can be opened and closed
by hand, without using any tools. If necessary, use a paper towel so you get a better grip on the parts.
z Install the battery and electrode in a clean and preferably dry environment.
Moisture in the screw joint can affect the functioning of the logger.
Please follow additionally the notes on commissioning in the operating manual of the electrode.
ba75821e01 10/2009
85
Commissioning WQL-pH
1
2
3

3.1.2 Inserting the battery

The logger is powered by a 3.6 V lithium battery, size AA, as provided in the scope of the delivery (see chapter 13 T
ECHNICAL DATAWQL-Log). Other
batteries of the same size such as alkaline manganese batteries do not provide the required operational voltage. Operating time, see section 11.4 B
ATTERY.
Fig. 3-1 Inserting the battery
1 Unscrew the cap (pos. 1 in Fig. 3-1).
2 Unscrew the sleeve (pos. 2).
3 Insert the battery When doing so, make sure that the battery is
correctly positioned (see label in the battery compartment).
4 Screw the sleeve on the shaft with the shorter
thread. At the rear end of the sleeve there is a thin O-ring inside (pos. 3). Make sure that this O-ring is evenly positioned in the groove and is not twisted.
86
ba75821e01 10/2009
WQL-pH Commissioning
5 Screw on the sleeve and cap again. In the screwed condition, no gap
may be visible at the joints.
After the battery has been inserted the signal LED flashes once per second. This means the logger has to be connected with the PC to set the date and time (see section 7.3 S
JOB and section 7.4 SIGNAL LED TO INDICATE THE OPERATING CONDITIONS OF THE LOGGER).
ETTING THE PARAMETERS AND STARTING THE LOGGING
ba75821e01 10/2009
87
Commissioning WQL-pH
1
1
3
2
4
5
6
2
7
8

3.1.3 Installing the electrode

88
Fig. 3-2 Installing the electrode
1 Unscrew the protective hood (pos. 1 in Fig. 3-2).
2 Unscrew the coupling ring (pos. 2).
ba75821e01 10/2009
WQL-pH Commissioning
3 Pull out the blind plug (pos. 3).
4 Prepare the electrode (pos. 4):
– Make sure that both O-rings (pos. 5 and 6 ) are mounted, clean
and dry.
– Remove the watering cap (pos. 7) and dab dry the shaft of the
electrode with a clean paper towel. – Position the coupling ring (pos. 2) on the shaft of the electrode. – Plug on the watering cap (pos. 7) to protect the electrode for the
time of the installation.
5 Insert the thus prepared electrode (pos. 8) in the electrode
receptacle: – Position the electrode at the opening and, against the pressure of
the air cushion inside, slowly push it in approx. 2 cm, until the air
escapes and the counter pressure is noticeably reduced. – Then push in the electrode as far as it will go and turn it until the
electrical plug connection locks in place using light pressure. The
electrode is completely inserted when it cannot be turned any
more.
6 Screw on the coupling ring (pos. 2) by hand to the limit.
7 For calibrating and measuring remove the watering cap.
8 Screw on the protective hood (pos. 1 in Fig. 3-2).
The protection hood can be screwed on for transport or storage even if the watering cap is plugged on. Please follow the notes on the storage of the electrode in the operating manual of the electrode.
ba75821e01 10/2009
89
Commissioning WQL-pH

3.1.4 Mounting suspension

To attach the logger to a rope or chain, a suitable shackle is provided whose bolt fits through the cross hole in the cap. As an alternative, you can screw some different sling gear into the M6 threaded hole at the cap end (e.g. M6 eye bolt).
Corrosion
protection
In water, metal (e.g. zinc-plated) parts of the mounting suspension such as thimbles, shackles or swivel connectors create voltages (chemically caused according to the electrochemical series), which can cause corrosion of the stainless steel of the logger or the above mentioned suspension parts.
Lightning
protection
Electrically conductive ropes or chains harbor the danger of the logger being damaged by lightning.
Recommendation For reasons of corrosion and lightning protection, we recommend to use a
nonconductive suspension, e.g. one with low-wear, resistant polypropylene ropes.
90
ba75821e01 10/2009
WQL-pH Commissioning

3.2 Installing the WQL-Log program

3.2.1 PC requirements

The WQL-Log PC program requires the following system components:
Hardware
requirements
Software
requirements
z Computer with Intel or Pentium III 500 MHz processor or higher (1 GHz or
quicker recommended.)
z At least 192 MB RAM (512 MB recommended) z Hard disk with at least 600 MB available memory z CD-ROM drive z One free USB interface for each measuring instrument to be connected.
z 32-bit
-operating system Windows 2000 Service Pack 4, Windows 7,
Windows Server 2003 and higher, Windows Vista or Windows XP

3.2.2 Installation routine

The software is subject to continuous development. The current version of the WQL-Log program is available for download on the Internet under http:/ /www.WTW.com.
The following files and/or directories are on the supplied CD-Rom:
Installing the
driver
z The installation file for the PC program and the database server z A directory with the installation program for the driver of the USB
interface
z A directory with the operating manual for the pH logger and the PC
program.
1 Insert the enclosed CD-Rom in the corresponding disk drive of the
PC.
2 Open the directory, USB_VCP_driver.
3 Execute the file, CP210xVCPInstaller with a double click.
ba75821e01 10/2009
91
Commissioning WQL-pH
4 Follow the user guidance (click Install).
5 Confirm the installation with OK. A virtual COM Port has been created
with the installation.
Installing the PC
program and
database server
6 Execute the WQLLog_Vxxxx.exe file with a double click (xxxx =
current version number). The installation wizard appears.
92
7 Follow the user guidance (click OK).
ba75821e01 10/2009
WQL-pH Commissioning
8 Check off the checkbox, I accept the licensing terms and conditions.
9Click Next. An information window appears with the components
required for the installation.
10 Click Install. The installation is carried out. Follow the user guidance
(click Next) until the Authentication Mode window appears.
ba75821e01 10/2009
93
Commissioning WQL-pH
11 Select the Mixed Mode option.
12 In the Enter password field, enter the password (WTW!2009).
13 Repeat the password entry (WTW!2009) in the Confirm password
field.
The password WTW!2009 must be entered twice. If a different password is entered, the PC program is not granted access to the database.
14 Click Next. Then follow the user guidance (the default settings should
not be changed) until the following display appears:
94
ba75821e01 10/2009
WQL-pH Commissioning
15 Here you can change the location where the PC program should be
stored (click Change... ). When the required location is displayed, click Next.
ba75821e01 10/2009
16 Here you find information on the installation. Click Install.
95
Commissioning WQL-pH
17 Click Finish to complete the installation.
96
ba75821e01 10/2009
WQL-pH Commissioning

3.3 Initial commissioning

1 Start the WQL-Log program (the relevant icon is on the desktop).
The Login window appears.
2 If the WQL-pH logger is connected to the PC via a USB cable and
should communicate with the WQL-Log PC program: Confirm with OK. The WQL-Log program searches for the connection with the logger.
3 Confirm with OK.
During the initial commissioning, the Select database window appears.
ba75821e01 10/2009
97
Commissioning WQL-pH
4 Confirm the option, Neue Datenbank anlegen (Create new database)
with OK. The window to save the newly created database appears.
98
5 Change the location and name of the database as necessary.
Click Speichern (Save). The database is stored and the message, Datenbank erfolgreich angelegt (Database was successfully created) appears.
6 Confirm with OK. A restart message appears:
ba75821e01 10/2009
WQL-pH Commissioning
7 Confirm with OK. The program is terminated.
If the database cannot be created in the preselected directory (error message, "Access denied"), select the following directory as the location:
(Windows program directory)\Microsoft SQL Server\ MSSQL.1\MSSQL\Data\
Example:
C:\Programs\Microsoft SQL Server\MSSQL.1\MSSQL\Data\
If you are in any doubt contact your system administrator.
ba75821e01 10/2009
99
Commissioning WQL-pH

3.4 Connection types

The PC program WQL-Log works with two types of connection: z Work with database
(the logger and PC program do not communicate)
z Connected to logger
(the logger and PC program communicate)
Working with the
database
Connected to the
logger
In this type of connection it is only possible to process the measurement data stored in the database and to change the system settings.
When the logger is logged on and connected to the WQL-Log PC program, the program and the logger communicate with each other. The measurement datasets present in the logger are read in to the database and are thus available for processing.
You can parameterize the logger according to your requirements and set up a logging job. When the logging job is started, the connection between the logger and the WQL-Log PC program is cut. The cable connection between the logger and PC does not have to be disconnected for this.
100
ba75821e01 10/2009
WQL-pH Commissioning

3.5 Starting the WQL-Log program

1 Start the WQL-Log program.
The Login window appears.
Connection to
logger
2 If the WQL-pH logger is connected to the PC via a USB cable and
should communicate with the WQL-Log PC program: Confirm with OK. The WQL-Log program searches for the connection with the logger.
3 Confirm with OK. The data window appears. During the initial
commissioning it does not yet contain data to be graphically displayed.
ba75821e01 10/2009
101
Commissioning WQL-pH
Working with the
database
4 If you want to work with the database:
Select Work with database and confirm with OK. The data window appears with the logger data that was last read in (during the initial commissioning it does not yet contain any data).
102
ba75821e01 10/2009
WQL-pH Setting the display (language etc.)

4 Setting the display (language etc.)

With the Display menu can set the language and temperature unit for the WQL-Log PC program and the WQL-pH logger.

4.1 Setting the language

1 In the Display menu, select the menu item, Language.
2 The options Deutsch (German) and Englisch (English) appear.
3 Select the required language with a mouse click.
The setting is immediately active.

4.2 Setting the temperature unit

1 In the Display menu, select the menu item, Temperature unit.
2 The options, ° C (degrees Celsius) and ° F (degrees Fahrenheit)
appear.
3 Select the required temperature unit with a mouse click.
The setting is immediately active.
ba75821e01 10/2009
103
Setting the display (language etc.) WQL-pH
104
ba75821e01 10/2009
WQL-pH Calibrating for pH measurements

5 Calibrating for pH measurements

Why calibrate? During the operation of a pH electrode, the slope and asymmetry of the
electrode change with time. Calibrating determines the current slope and asymmetry (zero point) of the electrode and stores them in the logger.
When to calibrate? Always calibrate:
z During the initial commissioning z Before starting a logging job z After installing another electrode
Calibration procedure, AutoCal :
AutoCal is a fully automatic single-point or two-point calibration using the datasets for
buffer solutions selected in the Buffer sets menu. The buffer solutions are automatically recognized. Depending on the setting (see page 116), the PC program displays the relevant nominal buffer value or the current electrode voltage in mV. The calibration can be terminated after the first buffer solution. This corresponds to a single-point calibration. For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
AutoRead When calibrating with AutoCal, the AutoRead function is automatically
activated. The AutoRead function checks the stability of the measured pH and temperature signal. The stability has a considerable effect on the reproducibility of the measured value.
The current AutoRead measurement can be terminated at any time (accepting the current value).
In the delivery condition, the WQL-pH logger is equipped with default calibration values (see chapter 10 R
ESETTING THE LOGGER). To achieve
accurate measured values it is essential to calibrate with the electrode before logging.
ba75821e01 10/2009
105
Calibrating for pH measurements WQL-pH

5.1 Buffer sets

You can use the buffer sets quoted in the table for an automatic calibration. The pH values are valid for the specified temperature values. The temperature dependence of the pH values is taken into account during the calibration.
Buffer set Menu
designation
WTW technical buffer
TEC 2.00
solutions
pH values
4.01
at temp.
25 °C
7.00
10.01
*
Standard buffer solutions according to DIN 19266
NIST/DIN 1.679
4.006
25 °C
6.865
9.180
12.454
The TEC calibration for pH 10.01 is optimized for the WTW technical buffer solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions can lead to an erroneous calibration. The correct buffer solutions are given in the WTW catalog or on the Internet.
106
ba75821e01 10/2009
WQL-pH Calibrating for pH measurements

5.2 Calibration settings

Buffer sets The following buffer sets can be selected:
z NIST/DIN z TEC
1 In the Device menu, select the menu item, Calibrate.
2 The calibration window appears.
Selecting the
buffer dataset
You can select the buffer dataset NIST/DIN or TEC.
3 In the Buffer set menu, select and confirm the required buffer
set with a mouse click. The selected setting is immediately active.
ba75821e01 10/2009
During calibration and direct measurement make sure that neither the calibration or measurement solution nor the logger shaft are electrically grounded (e.g. with a metal stand). Otherwise, this could result in erroneous values.
107
Calibrating for pH measurements WQL-pH

5.3 Calibrating with AutoCal

Always check the selected buffer set before calibrating. From the respective buffer set you can use any 2 buffers in any order.
We recommend to always calibrate with the protective hood screwed on so the electrode is protected. Use a stand as necessary.
Preparatory
activities
1 Connect the WQL-pH logger (with installed electrode) to a USB
interface of your PC via the USB cable.
2 Keep the buffer solutions ready.
3 In the Device menu, select the menu item, Calibrate.
The calibration window appears.
Calibration
The following example covers the AutoCal calibration using WTW technical buffer solutions.
4 Immerse the logger (with the electrode installed) in the first
buffer solution.
108
5 Use the Continue button to start the measurement.
ba75821e01 10/2009
WQL-pH Calibrating for pH measurements
6 The AutoRead running... note appears and the electrode voltage U
(mV) and temperature are displayed. The prompt for buffer 2 appears as soon as a stable value is recognized.
You can prematurely terminate the AutoRead function manually with Accept at any time. If the AutoRead function is prematurely terminated, the current measurement data are accepted immediately.
ba75821e01 10/2009
109
Calibrating for pH measurements WQL-pH
7 If you want to carry out a single-point calibration: Press Complete.
This completes the single-point calibration; the calibration record is displayed (with the Nernst slope of -59.2 mV/pH at 25 °C).
Or continue with the two-point calibration:
8 Thoroughly rinse the logger and electrode with deionized water.
9 Immerse the logger in the second buffer solution.
10 Start the measurement with Continue.
The AutoRead running... note appears and the electrode voltage U (mV) and temperature are displayed. The calibration record with the value of the slope (mV/pH) and the value of the asymmetry (mV) is displayed as soon as a stable value is recognized.
Calibration record
110
11 Accept the calibration with Accept. The message Calibration
successful appears and the new calibration will now be used for
measurement. If you press Cancel, the logger discards the new calibration and will continue to use the previous calibration values.
The calibration line is determined by linear regression.
You can view the calibration records at any time (see section 5.4 V
THE CALIBRATION RECORDS).
ba75821e01 10/2009
IEWING
WQL-pH Calibrating for pH measurements
Calibration
evaluation
After calibrating, the current system condition is evaluated. The asymmetry and slope are evaluated separately. The probe symbol shows the worse case evaluation of both.
.
Display Asymmetry [mV] Slope
[mV/pH]
-15 ... +15 -60.5 ... -58
-20 ... +20 -58 ... -57
-25 ... +25 -61 ... -60.5 or
-57 ... -56
Clean the electrode according to the electrode operating manual
-30 ... +30 -62 ... -61 or
-56 ... -50
ba75821e01 10/2009
111
Calibrating for pH measurements WQL-pH

5.4 Viewing the calibration records

The last 10 calibration records are stored in the PC program and can be recalled. Proceed as follows:
1 In the Device menu, select the menu item, Calibrate.
2 The calibration window appears.
3 Click the menu item, Calibration record. A list with the existing
calibration records appears.
4 Click the required record. The record is displayed.
5 Return to the calibration window with Back.
112
ba75821e01 10/2009
WQL-pH Measuring directly

6 Measuring directly

In conjunction with the pH logger, the WQL-Log PC program can directly measure and display the pH value or corresponding electrode voltage of a solution. The directly measured values are not stored in the logger.
During calibration and direct measurement make sure that neither the calibration or measurement solution nor the logger shaft are electrically grounded (e.g. with a metal stand). Otherwise, this could result in erroneous values.
Proceed as follows when you want to measure directly:
1 Connect the WQL-pH logger to the PC with the USB cable.
2 In the Device/Login menu, connect the logger to the WQL-Log PC
program.
3 If there are any logged data, import them with the query (direct
measurement is not possible without importing the logged data ).
4 In the Device menu, click the Measure menu item.
5 Immerse the WQL-pH logger (with installed electrode) in the test
sample.
6 The measured value window pops up with the pH or mV value
(depending on the setting) and the temperature value of the test sample.
ba75821e01 10/2009
7 Terminate the direct measurement with Quit.
113
Logging WQL-pH

7 Logging

Listed below are the typical operating steps with which to set up and carry out a logging job. The chapters where the operating steps are described in detail are also mentioned.

7.1 Typical sequence of a logging job (checklist)

1 Connect the logger to the PC with the USB cable.
2 Start the WQL-Log program (see section 3.5 S
PROGRAM).
TARTING THE WQL-LOG
3 Register the logger (see section CONNECTION TO LOGGER ).
4 If there are measured values in the logger, read them in (see
chapter 7.5 R
EADING IN DATA).
5 Clear the logger memory (see section 7.6 OVERVIEW OF THE DATA
WINDOW).
6 Calibrate the logger with the electrode (see chapter 5 C
FOR PH MEASUREMENTS).
ALIBRATING
7 Set up the logging job and send it (see section 7.3 SETTING THE
PARAMETERS AND STARTING THE LOGGING JOB).
8 Disconnect the USB cable from the logger.
9 If necessary, start the logger with the key button.
10 Check the operation of the logger based on the behavior of the signal
LED (see section 7.4 S
CONDITIONS OF THE LOGGER).
IGNAL LED TO INDICATE THE OPERATING
11 Tightly close the logger with the cap.
114
12 Install the logger at the measuring location.
13 After completion of the logging job: Connect the logger to the PC with
the USB cable (see step 1 etc.).
ba75821e01 10/2009
WQL-pH Logging

7.2 Setting up a logging job - setting parameters

You can set the following parameters for a logging job (i.e. the job to determine and record measured values according to your requirements over a certain period of time):
Tab Setting Setting options
(default setting in bold)
Recording Interval z 1 second, 5 seconds,
10 seconds, 30 seconds
z 1 minute, 5 minutes, 10 minutes,
15 minutes, 30 minutes
z 1 hour, 2 hours, 3 hours,
6 hours, 9 hours, 12 hours, 24 hours
Start z Immediately
(when the setting mode is quit and the logger is separated from the PC program)
Stop z Time period
Measured value ID
pH/mV Measuring
mode
Time Automatic
time adjustment
z Logger key button
(when the key button at the control panel of the logger is pressed)
(1 hour ... 365 days)
z Memory full
(the loggers stops logging only when the memory is full, i.e. when 600,000 datasets have been stored)
You can assign a measured value ID (consisting of up to 20 alphanumerical characters, no special characters) e.g. the name of the measuring site). Default setting: default
z pH z mV
z Yes (date/time for the logger is
automatically synchronized with that of the PC)
ba75821e01 10/2009
z No
(date/time for the logger is set manually)
115
Logging WQL-pH

7.3 Setting the parameters and starting the logging job

To change one or several of the above listed settings proceed as follows:
1 Connect the WQL-pH logger to the PC with the USB cable.
2 Start the WQL-Log program.
The Login window appears.
3 Confirm with OK. The WQL-Log program searches for the connection
with the logger.
4 Confirm with OK.
5 If the logger contains new data:
The prompt to read in the data appears.
116
ba75821e01 10/2009
WQL-pH Logging
6 Confirm the reading in of the data with OK.
The reading process can take some time (depending on the number of datasets to be read in).
7 When the reading of the measured values is completed, a message
appears (confirm with OK) and the prompt to erase the measured value memory.
8 Press OK or Cancel.
The data window appears with a graphic of the read-in data, no matter whether or not the data memory was erased.
ba75821e01 10/2009
117
Logging WQL-pH
9 In the Device menu, select the menu item, Set up logging job. A note
on calibration appears.
10 Confirm with OK. The Logging tab appears, the first tab of the
window, Set up logging job.
118
11 Interval: To set the logging interval, click the arrow on the right side
of the setting field. A drop-down menu appears with the possible intervals. Select the required interval with a mouse click.
12 Start: To set the start of the logging, click the arrow on the right side
of the setting field. A drop-down menu appears with the starting times. Select the required starting time with a mouse click.
13 Stop: To set the end of the logging, click the arrow on the right side
of the first setting field. A drop-down menu appears with the possible settings. Select the required starting time with a mouse click. If Time period was selected:
14 To set the time period of the logging, click the arrow on the right side
of the second setting field. A drop-down menu appears with the possible time periods. Select the required time period with a mouse click.
ba75821e01 10/2009
WQL-pH Logging
15 Enter the measured value ID (up to 20 alphanumerical characters, no
special characters).
16 Click the pH/mV tab.
17 Select the required mode for the logging, pH or mV.
18 Click the Time tab.
19 Select Yes or No for the automatic date/time adjustment. If No was
selected, click the arrows on the right side of the selection fields and set the date and time for the logger.
In the Logging tab, the memory capacity required with the selected settings is displayed at the bottom (number x of 600,000 possible datasets).
Starting the
logging job
20 To start the logging job, click the button, Send logging job. The
message, The settings have been successfully transferred appears.
21 Confirm with OK. A window pops up with an overview of the logging
job just sent.
ba75821e01 10/2009
22 Confirm with OK. A note appears with the message that the logger
will now be disconnected from the PC program and, depending on the setting, the information that the logging is started immediately or that the logging is started when the logger key button is pressed.
119
Logging WQL-pH
The setting routine for the logging job can be canceled at any time with the Cancel button. In this case the logger does not start the logging job; instead, the data window appears.
120
ba75821e01 10/2009
WQL-pH Logging

7.4 Signal LED to indicate the operating conditions of the logger

Next to the key button on the control panel of the logger there is a signal LED. When the logger is not connected to the WQL-Log PC program, this red LED indicates the different operating conditions of the logger as follows:
Mode of flashing Operating condition
One short flashing per second z The logging job is completed.
z The power supply of the logger
was interrupted. The logger has to be connected to the PC program in order to set the date and time.
Two short flashings, then a pause of 3 seconds
The logger is ready to log and can be started with the logger key button.
A short flashing every 15 seconds The logging job is running (the logger
is logging).
Notes
z The flashing stops after 2 minutes to save energy. Pressing the logger
key button starts the flashing for 2 minutes again.
z Important: The logger can be started with the logger key button only while
the signal LED is flashing (if starting the logger with the key button was set during the setting up of the logging job).
ba75821e01 10/2009
121
Logging WQL-pH

7.5 Reading in data

Depending on the logging job, the WQL-pH logger measures for a certain period of time and at certain intervals the pH value or mV value and the temperature of a solution. The data is stored in the logger. The WQL-Log PC program has to read in and store in a database the data before the data can be displayed and processed.
If the logger contains new data, the prompt to read in the data appears automatically when the logger is connected to the PC program. If data stored in the logger are not read in, the logger cannot be connected to the PC program (data protection!).
To read in the data of the logging job, proceed as follows:
1 Connect the WQL-pH logger to the PC with the USB cable.
2 Start the WQL-Log program.
The Login window appears.
3 Confirm with OK. The WQL-Log program searches for the connection
with the logger.
122
ba75821e01 10/2009
WQL-pH Logging
4 Confirm with OK.
5 If the logger contains new data:
The prompt to read in the data appears.
6 Confirm the reading in of the data with OK.
The reading process can take some time (depending on the number of datasets to be read in).
7 When the reading of the measured values is completed, a message
appears (confirm with OK) and the prompt to erase the measured value memory.
ba75821e01 10/2009
8 Press OK or Cancel.
The data window appears with a graphic of the read-in data, no matter whether or not the data memory was erased.
123
Logging WQL-pH
When the logger is connected to the WQL-Log PC program, any existing new measured values must be read in. If data stored in the logger are not read in, the logger cannot be connected to the PC program (data protection!).
124
ba75821e01 10/2009
WQL-pH Logging
Number of selected datasets
Name and series number of the connected logger
Menu bar
Measured value ID
Name and series number of the logger that provided the data

7.6 Overview of the data window

The measurement datasets read in from the logger are stored in a database and graphically displayed in a data window.
The data window is structured as follows:
ba75821e01 10/2009
125
Logging WQL-pH

7.7 Clearing the logger memory

The query to clear the logger memory always appears after the data from the logger have been read in to the database. If the logger memory is not cleared after the data has been read in, the query to clear the data reappears during the setting up of a new logging job. At this point of time the logger memory has to be cleared so the logging job can be set up.
You can also clear the logger memory manually.
There is no danger of clearing data that have not been read in as the data stored in the logger have to be read in (compulsory guide) when the logger is connected. If data stored in the logger are not read in, the logger cannot be connected to the PC program.
Proceed as follows to clear the logger memory:
1 In the Device menu, select the menu item, Memory/Clear memory.
2 A security prompt appears. Confirm the security prompt with OK.
3 The data is cleared. A message appears after the clearing that
informs you that the memory was cleared. Confirm the message with OK.
126
ba75821e01 10/2009
WQL-pH Processing measurement data

8 Processing measurement data

8.1 Querying data

The data recorded by the logger and stored in the database can be displayed as a graphic via the Query menu item. Proceed as follows:
Querying current
data
1 In the Query menu, click the menu item, New.
The New query window appears.
2The Standard tab appears when a logger is connected.
ba75821e01 10/2009
3 To graphically display the data last imported from the connected
logger, click Apply. The data appear as a graphic, broken according to pH value or voltage value and temperature.
127
Processing measurement data WQL-pH
The series number of the connected logger is displayed in the bottom left corner of the data window (see section 7.6 O
VERVIEW OF THE DATA WINDOW).
128
ba75821e01 10/2009
WQL-pH Processing measurement data
Querying user-
defined data
1 Click the User-defined tab in the New query window. The User-
defined tab appears immediately in the Query menu if no logger is
connected.
2 If you only want to query datasets with a certain measured value ID,
check off the box for Measured value ID and then click the arrow next to the selection field for the measured value ID. Select the required measured value ID.
3 In the Sensors field, select one or two measured values to be
queried.
4 If you only want to query datasets with a certain date, check off the
box for Time period and then click the arrow next to the selection fields for the time (from and to). Select the required date in both fields. Next to the date fields are the time fields where you enter the time (hours, minutes, seconds) either with the arrow keys or by entering the number with the keyboard.
5Click Apply to display the selected datasets.
ba75821e01 10/2009
129
Processing measurement data WQL-pH
6 The queried data appear as a graphic.
Always make the settings for the user-defined query from the top down , as the data are filtered in this sequence.
130
ba75821e01 10/2009
WQL-pH Processing measurement data

8.2 Exporting data

The data displayed as a graphic with the aid of the Query menu can be exported to a *.csv file. In this form the data can be read and processed with Microsoft Excel and many other spreadsheets.
If the number of datasets to be exported is more than 60,000 and the data should be processed with Microsoft Excel 2007 or older, export the data with the function, "Export as *.csv file (splitted)", as Excel versions 2007 and older can store only 60,000 datasets in one worksheet. The number of datasets is given in the header of the graphic, next to the series number of the logger that provided the data.
Proceed as follows to export the datasets displayed as a graphic:
1 In the Export menu, click the menu item, Export as *.csv file or Export
as *.csv file (splitted). The Save as window appears.
ba75821e01 10/2009
2 Save the data to be exported in the required directory and with the
required file name. The WQL-Log PC program suggests a file name.
131
Processing measurement data WQL-pH
If the number of datasets to be exported is more than 60,000 and the data is exported with the function, Export as *.csv file (splitted), the WQL-Log program automatically divides the data into files with max. 60,000 datasets and adds the number 1, 2 etc. to the suggested file names.
132
ba75821e01 10/2009
WQL-pH Info menus

9 Info menus

9.1 Database information

The data read in from the logger are saved in a database. You can query the name, location, type and utilization (in %) of this database.
To do so, proceed as follows:
Database
information
1 In the File menu, click the menu item, Database info.
The Database info window appears.
2 Confirm with OK. The Database info window disappears.
ba75821e01 10/2009
The size of a database file can be 4 GB max. When this amount of data is stored, a new database file must be created. The new database file is created in the Device/Create new database menu.
100,000 datasets need approx. 6.5 MB. Thus the database can store approx. 15 million datasets.
133
Info menus WQL-pH

9.2 Device information

If a logger is connected to the PC, you can query the device name, series number and firmware version of the connected logger via the Device/Info menu item.
To do so, proceed as follows:
Device information
1 In the Device menu, click the menu item, Info.
The Device info window appears.
2 Confirm with OK. The Device info window disappears.
134
ba75821e01 10/2009
WQL-pH Info menus

9.3 Program info

In the Help/About WQL-Log menu you can look up which version of the WQL- Log PC program was installed.
To do so, proceed as follows:
1 In the Help menu, click the About WQL-Log menu item.
The About WQL-Log
Program info
window appears.
2 Confirm with OK. The About WQL-Log window disappears.
ba75821e01 10/2009
135
Resetting the logger WQL-pH

10 Resetting the logger

You can reset to the default condition the logger and the PC program with its settings.
In the default condition, the settings are as follows:
Configuration parameters Setting
Logging interval 1 second
Start of the logging Logger key button
End of the logging Time period
1 hour
Measured value ID None
Measuring mode pH
Date, time 01.01.2008, 00:00
Resetting to
default condition
Calibration Default values
(Nernst slope = -59,2 mV/pH at 25 °C, asymmetry = 0 mV)
To reset the logger and PC program to the default condition, proceed as follows:
1 In the Device menu, click the menu item, Reset. A security prompt
appears.
2 Confirm with OK. The settings are reset to the default condition.
136
ba75821e01 10/2009
WQL-pH Maintenance, cleaning, storage

11 Maintenance, cleaning, storage

11.1 General maintenance instructions

General
information
z Normally, all screw joints of the logger housing can be opened and closed
by hand, without using any tools. If necessary, use a paper towel so you get a better grip on the parts. Should the coupling ring for the electrode be stuck, for example after a long-term logging job, you can use a wrench (wrench size 16 mm) on the hexagon.
z In the screwed condition, no gaps may be visible at the joints. z Change the electrode in a clean and preferably dry environment.
Thoroughly dry all parts. Moisture in the screw joint can affect the functioning of the logger.
z Prior to unscrewing any screw joints, clean the exterior of the logger (see
section 11.2 E
XTERIOR CLEANING) and dry it thoroughly.
z Prior to reassembly, clean all internal and external threads thoroughly.
Contaminated threads can get stuck by-and-by. Normally, the threads can be screwed together without using any effort and without noticeable grinding (sand particles).
z Clean all O-rings prior to reassembly. Dirt, e.g. fibers, on the O-rings can
affect the tightness.

11.2 Exterior cleaning

Clean the logger with tapwater and a soft sponge or brush. Remove the protective hood. The electrode should be cleaned with a soft toothbrush or paintbrush under running tap water.
For the cleaning of the electrode please also read the operating manual of the electrode.
ba75821e01 10/2009
137
Maintenance, cleaning, storage WQL-pH

11.3 Replacing the electrode

When replacing the electrode the general maintenance instructions in section 11.1 must be followed.
The electrode is replaced in the same way as it was installed during the commissioning (see section 3.1.3 I
NSTALLING THE ELECTRODE).

11.4 Battery

11.4.1 Battery service life

The logger has an energy saving feature. The service life of the battery depends very much on the usage of the logger, especially on the measuring interval. The service life to be expected of a full battery (2600 mAh) can be estimated by means of the following table:
Measuring interval Service time to be expected
1 sec 3 months 5 sec 5 months 10 sec 7 months 30 sec 11 months 1 min 13 months 5 min 15 months 10 min and longer 16 months
When inserting the battery, note down the installation time on the battery to be able to estimate the remaining operational lifetime. When a battery is inserted, the logger consumes a small amount of energy even when it is not working. Therefore, we recommend to remove the battery during longer measuring breaks.

11.4.2 Battery replacement

138
When replacing the battery the general maintenance instructions in section 11.1 must be followed.
The battery is replaced in the same way as it is inserted during the commissioning (see section 3.1.2 I
NSERTING THE BATTERY).
ba75821e01 10/2009
WQL-pH Maintenance, cleaning, storage
Data preservation After the battery was removed all stored measured values are retained. Only
the date and time are reset to the default condition (01.01.2008, 00:00) when the power supply is interrupted.
We recommend to connect the logger to the PC for the time of the battery replacing process so the logger is supplied with power via the USB connection while there is no battery inserted. Thus the date and time will be retained. Otherwise, the date and time will be set the next time the logger is connected to the PC with the WQL-Log PC program. Depending on the setting in the WQL-Log program, this is either done automatically when the connection is established (synchronization with the system time of the PC), or manually.
State of the logger after the power supply was interrupted (battery empty or removed, see also section 7.4 S
CONDITIONS OF THE LOGGER):
IGNAL LED TO INDICATE THE OPERATING
Case 1: Logging
job active at the
time of
interruption
Case 2: No logging
job at the time of
interruption
Case 3: Battery
removed for
storage
z The logging job is terminated. z The time is reset to the default condition. z All data logged up to this point of time are retained and are automatically
saved on the PC the next time the logger is connected to the PC.
z The terminated logging job can be continued with a corresponding new
logging job. To do so, the time must be set anew if the automatic time synchronization is switched off.
z The time is reset to the default condition. z Prior to setting up the next logging job, you have to set the time if the
automatic time synchronization is switched off.
z The time is reset to the default condition. z Prior to setting up the next logging job, you have to set the time if the
automatic time synchronization is switched off.
ba75821e01 10/2009
139
Maintenance, cleaning, storage WQL-pH
12

11.5 Storage

For longer storage periods we recommend to leave the electrode installed and to remove the battery and store it separately. Plug the watering cap filled with the suitable reference electrolyte on the clean electrode (see electrode operating manual).
If the logger is stored without the electrode, close the electrode receptacle with the enclosed blind plug to protect the plug connector from dirt and moisture. When doing so make sure that both O-rings of the blind plug are in place (pos. 1 and 2 in Fig. 11-1).
Fig. 11-1 Blind plug
Store the logger in a dry place while observing the storing conditions according to chapter 11.5 S
TORAGE. Low temperatures usually delay the
aging of the electrode.
140
ba75821e01 10/2009
WQL-pH What to do if...

12 What to do if...

12.1 Calibration and measuring

Lengthy stability
check with
AutoRead
Implausible
measured values
Logger does not
react
Cause Remedy
No stable measured value Provide stable measuring conditions
(e.g. temperature)
Cause Remedy
– Electrode not installed or
defective
– No calibration performed – Calibrate
– Watering cap still on the
electrode
– Electrode contaminated – Clean electrode
– Liquid penetrated into the plug
connector
Cause Remedy
– Battery depleted – Insert new battery (see
– Check electrode and electrode
connection
– Pull off watering cap and calibrate
– Rinse the plug connector with
deionized water, clean and dry it
section 11.4)
System cannot be
calibrated
ba75821e01 10/2009
Cause Remedy
– Slope of the electrode not within
tolerance (see section 5.3)
– Slope of the electrode too low – Replace electrode
– Asymmetry of the electrode too
high
– Condition the electrode – If the slope is still outside the
tolerance, replace the electrode
– Check the buffer solutions
– Replace electrode
141
What to do if... WQL-pH

12.2 Communication of the logger and PC program

Reading in of
measured values
is aborted
Cause Remedy
– Communication problem – Repeat the read in process
Data logging of the
logger aborted
Cause Remedy
– Power supply interrupted – Connect the logger to the PC
program and set up logging job.
– Make sure that the logger is
supplied with power (battery)
142
ba75821e01 10/2009
WQL-pH Technical data

13 Technical data

13.1 Measurement characteristics

Measuring
Potentiometric measurement by means of combination electrode
principle
pH measurement Measuring range ** 0.000 ... +20.000
Resolution 0.001 Accuracy **
0.005 ± 1 digit
mV measurement Measuring range ** -1000.0 ... +1000.0 mV
Resolution 0,1 mV
Temperature
measurement
Accuracy **
Temperature sensor Automatic recognition of the temperature
0.2 mV ± 1 digit
sensor of the electrode:
NTC 30 (30 k:at 25 °C / 77 °F) or Pt1000 Measuring range ** -5.0 ... 105 °C (+23 ... 221 °F) Resolution 0.1 K Accuracy **
0.1 K ± 1 digit
** Note: The measuring ranges and accuracy exclusively refer to the measuring electronics of the logger without the electrode. For the measuring ranges and accuracy of the operable logger, the specifications of the electrode and buffer solutions have also to be taken into account.
Logger function Measuring interval adjustable:
1 / 5 / 10 / 30 s
1 / 5 / 10 / 15 / 30 min
1 / 2 / 3 / 6 / 9 / 12 / 24 h Capacity of the data
600.000 measurement datasets
memory
ba75821e01 10/2009
143
Technical data WQL-pH
Thread M6

13.2 Application characteristics

Measuring
Allowed temperature range 0 ... 60 °C (32 ... 140 °F)
medium
Allowed pH range 2 ... 12
Pressure
IP 68 (1 x 10
resistance
The WQL-pH logger meets all requirements according to article 3(3) of 97/ 23/EC ("pressure equipment directive").
Storage
Recommended storing type Electrode installed with the watering cap put
conditions
Storage temperature -25 ... 65 °C (-13 ... 149 °F)

13.3 General data

Test certificates CE
6
Pa or 10 bar)
on, or electrode receptacle closed with blind plug, battery removed
(electrode removed)
Dimensions
(in mm)
Weight 860 g (with SensoLyt
Electrode
SMEK socket
connection
Material Metal parts coming into
contact with the sample: – Shaft
–Sleeve – Coupling ring – Protective hood –Cap
Battery compartment Brass, gold-plated Screws Stainless steel V2A
®
WQL electrode)
Stainless steel 1.4571*
144
Plug connector for electrode PEEK, contacts gold-plated
ba75821e01 10/2009
WQL-pH Technical data
Material
(continued)
Guidelines
and norms used
Housing of the control panel PVC-U Key button Silicone Seals FPM (Viton) Blind plug POM Shackle Stainless steel V4A *
* Stainless steel can be susceptible to corrosion at chloride concentrations of
500 mg/l and more.
EMC EC directive 2004/108/EC
EN 61326
FCC Class A Meter safety EC directive 2006/95/EC
EN 61010-1 Climatic class VDI/VDE 3540 Pressure resistance EC directive 97/23/EC IP protection class EN 60529

13.4 Electrical data

Power
supply
Electrical safety Protective class III
USB interface Type USB 1.1
Battery Lithium thionyl chloride battery 3.6 V,
size AA, 2600 mAh Operational life At least 3 months, depending on measuring
interval
Baud rate 38400 Socket type Mini USB Cable length max. 3 m
ba75821e01 10/2009
145
Technical data WQL-pH
146
ba75821e01 10/2009
WQL-pH Lists
Index
A
AutoCal . . . . . . . . . . . . . . . . . . . . . . . . .105
AutoRead . . . . . . . . . . . . . . . . . . . . . . .105
B
Buffer solutions
according to DIN 19266
Buffer 10,01 Trace . . . . . . . . . . . . . .106
WTW Technical buffers . . . . . . . . . . .106
. . . . . . . . . . .106
C
Calibration . . . . . . . . . . . . . . . . . . . . . . .105
AutoCal . . . . . . . . . . . . . . . . . . . . . .108
Calibration evaluation . . . . . . . . . . . .111
Calibration record . . . . . . . . . . . . . . .110
Clearing the memory . . . . . . . . . . . . . . .126
Connected to logger (type of connection) .100
D
Data filter . . . . . . . . . . . . . . . . . . . . . . .129
Data preservation . . . . . . . . . . . . . . . . . .139
Data window . . . . . . . . . . . . . . . . . . . . .125
Default settings . . . . . . . . . . . . . . . . . . .136
E
Exporting data . . . . . . . . . . . . . . . . . . . .131
Measuring mode . . . . . . . . . . . . . 115
Start . . . . . . . . . . . . . . . . . . . . . 115
Stop . . . . . . . . . . . . . . . . . . . . . 115
Start . . . . . . . . . . . . . . . . . . . . . . . . 116
Logging job setting routine . . . . . . . . . . .118
M
Measuring directly . . . . . . . . . . . . . . . . . 113
O
Operating conditions of the logger (LED) . 121
Q
Query . . . . . . . . . . . . . . . . . . . . . . . . . . 127
R
Reading in of measured values . . . . . . . . 122
Recording . . . . . . . . . . . . . . . . . . . . . . . 115
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . 136
S
Sensor symbol . . . . . . . . . . . . . . . . . . . 111
Single-point-calibration . . . . . . . . . . . . . 110
Size of the database . . . . . . . . . . . . . . . 133
Starting the logging job . . . . . . . . . . . . . 116
System messages . . . . . . . . . . . . . . . . . 141
F
Fields of application . . . . . . . . . . . . . . . . .80
Filtering of data . . . . . . . . . . . . . . . . . . .129
Firmware version . . . . . . . . . . . . . . . . . .134
Flashing of the logger . . . . . . . . . . . . . . .121
G
Grounded test sample . . . . . . . . . . . . . .113
L
Logging job
Checklist
Overview of the settings . . . . . . . . . .119
Setting parameters . . . . . . . . . . . . . .115
Automatic time adjustment . . . . . .115
Interval . . . . . . . . . . . . . . . . . . . .115
Measured value ID . . . . . . . . . . .115
ba75821e01 10/2009
. . . . . . . . . . . . . . . . . . . . .114
T
Time adjustment . . . . . . . . . . . . . . . . . . 115
Trouble shooting . . . . . . . . . . . . . . . . . . 141
Two-point calibration . . . . . . . . . . . . . . . 110
U
Utilization of the database . . . . . . . . . . . 133
W
What to do if... . . . . . . . . . . . . . . . . . . . . 141
Working with the database (type of connection)
100
147
Lists WQL-pH
148
ba75821e01 10/2009
Loading...