The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result
in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept
any legal claims resulting from the data, figures or descriptions.
Firmware
Method data
Analysis specifications
Operating manual
Part of the process of consequently improving our products is the
continuous further development of the range of photometric tests
supplied and the meter firmware. All current data for the pHotoFlex /
pHotoFlex Turb can be found on the Internet under http://
www.WTW.com.
You can easily transfer new firmware to your instrument with the aid of
the AK 540/B cable and a PC. More detailed information can be found
in the appendix of the detailed operating manual on the CD-ROM
provided.
The compact pHotoFlex (Turb) handheld precision photometer enables you to carry out the following measurements quickly and reliably:
! Photometric measurements
– Concentration measurements
– Absorbance measurements
– Transmission measurements
! pH measurements
! Turbidity measurements (pHotoFlex Turb).
The pHotoFlex (Turb) handheld meter provides the maximum degree
of operating comfort, reliability and measuring certainty for all applications.
The proven MultiCal
®
calibration procedure supports you when calibrating for pH measurements and the AutoRead function enables precise pH measurements.
4
3
2
1
ba75491d0407/2006
1Keypad
2Display
3Cell shaft
(folded out for a 16 mm cell to be inserted)
4Socket field
7
OverviewpHotoFlex / pHotoFlex Turb
Note
If you need further information or application notes, you can obtain the
following material from WTW:
! Application reports
! Primers
! Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
1.2Keypad
Key functions
.
Select the measuring mode
<M> (long keystroke):
– Photometry
– Turbidity
– pH & ORP
Select the measured parameter within a measuring mode
<M> (short keystroke):
– pH & ORP: pH, ORP
– Photometry:
Concentration, Absorbance, % Transmission
– Turbidity: no measured parameters selectable
Start calibration (measuring modes, pH & ORP,
Turbidity)
Start zero adjustment or blank value measurement using the Photometry \ Adjustment menu
(measuring mode, Photometry)
<CAL/ZERO>
8
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOverview
In the Photometry measuring mode: Select a pro-
gram for concentration measurement
<PROG>
Open menus / confirm entries /
start measurement
.
<START/ENTER>
Call up the Configuration menu
(all settings are made here)
<MENU>
In the Photometry measuring mode, measured
parameter, Concentration
:
switch over between available citation forms
<FORM>
In the Photometry measuring mode, measured
parameter, Concentration
:
Switch over between available units <UNIT>
Switch the measuring instrument on/off
<ON/OFF>
PRT
Output display contents to RS232 interface (e.g.
print)
<PRT>
Open the Store menu <STO>,
Quick storing <STO> <STO>
Highlight menu items or selection
Set values
<▲>, <▼>
Switch to the next higher menu level /
cancel input
<ESC>
Note
Keys with an additional number printed on are assigned doubly.
This enables to directly enter numbers in special menus. Thus, you
can, for example, conveniently enter the date and time via the number
keys.
ba75491d0407/2006
9
OverviewpHotoFlex / pHotoFlex Turb
1.3Display
The graphic display shows all information of the current measurement
in the measured value display. The illumination enables to read the display even in the darkness.
Example
Photometry \ Concentration
1.29
2: A5/25 MCNH4
16 mm 0.20 - 8.00 mg/l
01.02.05 15:12
mg/l
1.4Socket field
-N
Measuring mode \ measured
parameter
Measured value (with unit)
Program and citation form
Diameter of the cell and
measuring range
Status line with date and time
10
Identifying the
connectors
12345
1pH electrode
2pH temperature sensor
3Power pack
4Contacts for operation on the LabStation
5RS232 serial interface
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOverview
1.5LabStation (optional)
With the LabStation, which is available as an accessory, you can conveniently use the pHotoFlex (Turb) in the laboratory.
Laboratory operation with the LabStation enables the following additional functions:
! With photometric measurements, the zero measurement is retained
even after switching the pHotoFlex (Turb) off and on again
! You can connect a barcode reader for the simplified calling up of
programs
! The LSdata software serves to easily enter user-defined programs
! Line power operation is possible to save the batteries or accumula-
tor pack
! The accumulator pack in the pHotoFlex (Turb) is automatically
charged as soon as the meter is placed in the LabStation.
ba75491d0407/2006
LabStation
Fig. 1-1LabStation
11
OverviewpHotoFlex / pHotoFlex Turb
12
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbSafety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating manual carefully before working with the meter. The operating manual must
always be available within the vicinity of the meter.
Target groupThe meter was developed for work in the field and in the laboratory.
We assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to take
when handling the chemicals of photometric test sets.
The personnel responsible for the commissioning, operation and maintenance must have the necessary qualifications for this work. If the personnel do not have the required skills they have to be instructed.
Furthermore, it must be ensured that the personnel read and completely understand the present operating manual.
Safety instructionsThe individual chapters of this operating manual use the following safe-
ty instructions to indicate various types of danger:
Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the instrument
or the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
ba75491d0407/2006
13
SafetypHotoFlex / pHotoFlex Turb
2.1Authorized use
This meter is authorized exclusively for carrying out the following measurements in the field and laboratory:
! Analysis of substances in water and aqueous solutions
using round cells
! Concentration measurement
! Absorbance and transmission measurement
! Measurement of pH value and ORP
! Turbidity measurement (pHotoFlex Turb only).
The technical specifications as given in chapter 7 T
ECHNICALDATA must
be observed. Only the operation and running of the meter according to
the instructions given in this operating manual is authorized. Any other
use is considered to be unauthorized.
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see chapter
7T
ECHNICALDATA).
It left the factory in a safe and secure technical condition.
Opening the photometer or adjustment, maintenance and repair work
must only be performed by specialist personnel authorized by the manufacturer.
The only exceptions to this are the activities described in chapter
5M
AINTENANCE, CLEANING, DISPOSAL. Non-compliance results in the
loss of warranty claims.
Follow the points listed below when operating the photometer:
! Follow the local safety and accident prevention regulations
Function and operating
safety
14
! Observe the enclosed instructions of reagents and accessories
! Observe the regulations when dealing with dangerous substances
! Follow the operating instructions at the workplace
! Use only original spare parts.
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during operation.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbSafety
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICALDATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
Safe operationIt is the responsibility of the operator to continuously observe the over-
all technical condition (externally recognizable deficits and damage as
well as alterations to the operational behavior) of the meter.
If safe operation is no longer possible, the instrument must be taken out
of service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
! has been damaged in transport
! has been stored under adverse conditions for a lengthy period of
time
! is visibly damaged
Obligations of the
operator
! no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
Caution
Danger of eye damage by visible and invisible LED radiation. In
the cell shaft there are light emitting diodes (LED) of the 1M class.
Do not look at the radiation using optical instruments.
With normal, authorized use there is no hazard.
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
! EEC directives for protective labor legislation
! National protective labor legislation
! Safety regulations
! Safety datasheets of the chemical manufacturers.
ba75491d0407/2006
15
SafetypHotoFlex / pHotoFlex Turb
16
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbCommissioning
3Commissioning
3.1Scope of delivery
! Handheld photometer, pHotoFlex Turb or pHotoFlex
! 4 batteries, 1.5 V type AA (in the battery compartment)
! Optional: Accumulator pack and power pack with Euro plug
and exchange plugs for USA, UK, and Australia
! Optional: LabStation
! 1 empty cell 16 mm
! 2 empty cells 28 mm
with label to mark the cell for turbidity measurements
! AMCO
! Microfiber cloth to clean the meter
! Compact operating manual and short operating manual
! CD-ROM with
®
-Clear turbidity standard (pHotoFlex Turb only)
Charging time of the
accumulator pack
– detailed operating manual
– photometry analysis manual with analysis specifications
– software to program user-defined methods
Note
The optional parts of the scope of delivery are available as accessories
(see section 8.1).
3.2Power supply
You can operate the meter either with batteries, accumulator pack or a
power pack. The power pack supplies the meter with low voltage
(9 V DC). At the same time, the accumulator pack is charged. The accumulator pack is charged even while the meter is switched off.
The LoBat display indicator appears when the batteries or accumulator
pack is nearly discharged.
approx. 36 hours.
Caution
The line voltage at the operating site must lie within the input voltage range of the original power pack (see chapter 7 T
TA).
ECHNICALDA-
ba75491d0407/2006
Caution
17
CommissioningpHotoFlex / pHotoFlex Turb
Use original power packs only (see chapter 7 TECHNICALDATA).
Note
The accumulator pack should not be completely discharged. If you do
not operate the instrument for a longer period of time you should charge
the accumulator pack every six months.
Automatic switchoffThe meter has an automatic switch-off function in order to save the bat-
teries or accumulator pack (see section 4.4).
Display illuminationDuring operation with batteries or accumulator pack the meter automat-
ically switches off the display illumination if no key is pressed for
30 seconds. The illumination is switched on with the next keystroke
again. The display illumination can also be switched off completely (see
section 4.4.2).
Note
Power pack and accumulator pack are available as an accessory (see
section 8.1).
Connecting the power
pack (optional)
4
2
3
1
1If necessary, replace the Euro plug (1) on the power pack (2)
by the country-specific plug suitable for your country.
2Connect the plug (3) to the socket (4) of the photometer.
3Connect the power pack to an easily accessible power socket.
18
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbCommissioning
3.3Connecting the LabStation
Note
The LabStation is available as an accessory (see section 8.1).
In order to use the functions of the LabStation for operation in the laboratory, connect the LabStation and place the pHotoFlex (Turb) in the
LabStation.
Connecting the LabSta-
tion (optional)
6
5
4
2
3
1
1If necessary, replace the Euro plug (1) on the power pack (2)
by the country-specific plug suitable for your country.
2Connect the plug (3) to the socket (4) of the LabStation.
3Connect a PC, printer or barcode reader to the socket (5) of the
LabStation as necessary.
ba75491d0407/2006
Note
When the pHotoFlex (Turb) is operated in the LabStation, a barcode
reader can be connected. A barcode reader enables to select program
numbers in a simplified manner.
Recommended barcode readers can be found in section 8.2.
For the interface of your meter, set the same baud rate that is set for
the barcode reader (see section 4.4.3).
The baud rate of the barcode reader is given in the operating manual
of your barcode reader.
19
CommissioningpHotoFlex / pHotoFlex Turb
4Connect the power pack to an easily accessible power socket.
5Place the pHotoFlex (Turb) in the LabStation.
LabStation
6Connect a pH electrode to the socket on the pHotoFlex (Turb)
through the cut-out (6) as necessary.
20
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbCommissioning
3.4Initial commissioning
Perform the following activities:
! For
– accumulator operation: insert the accumulator pack (see
section 5.1.2)
– line power operation and charging the accumulator pack: connect
the power pack (see section 3.2)
– operation with LabStation: connect the LabStation and place the
pHotoFlex (Turb) in the LabStation (see section 3.3)
! Switch on the meter (see section 4.1)
! Set the language as necessary (see section 4.3.3)
! Set the date and time as necessary (see section 4.3.4)
Note
When you set the language, date and time according to the mentioned
sections of this operating manual you will quickly become familiar with
the simple operation of the pHotoFlex (Turb).
ba75491d0407/2006
21
CommissioningpHotoFlex / pHotoFlex Turb
22
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4Operation
4.1Switching on the meter
Switching onPress the <ON/OFF> key.
For 30 seconds, Start menu appears with a selection of the measuring
modes. The measuring mode last selected is highlighted.
The status line indicates the meter designation and the version number
of the software.
Start
Photometry
Turbidity
pH & ORP
i pHotoFlex V 0.24
After a few seconds, the meter automatically switches to the measuring
mode and measured parameter used last.
The measured value display appears (here, e.g. measuring mode Pho-
tometry).
Photometry \ Concentration
i Select program
with <PROG>
01.02.05 15:12
With <M> (long pressure) change the measuring mode.
With <M> (short pressure) toggle between the different measured parameters in the selected measuring mode.
Switching offPress the <ON/OFF> key.
Automatic switchoffThe meter has an automatic switchoff function in order to save the bat-
teries or accumulator pack (see section 4.4). The automatic switchoff
switches the meter off if no key is pressed for an adjustable period.
ba75491d0407/2006
The automatic switchoff is not active
! if the power is supplied by the power pack (optional),
! if the power is supplied by the LabStation (optional),
! if the Timer or Analysis timer function is on.
23
OperationpHotoFlex / pHotoFlex Turb
Display illumination
with battery and accu-
mulator pack operation
During operation with batteries or accumulator pack the meter automatically switches off the display illumination if no key is pressed for
30 seconds. The illumination is switched on again with the next keystroke.
4.2Inserting a cell
To be able to insert cells in the pHotoFlex (Turb), the cell shaft has to
be prepared to take in a cell.
1Push the dust cover (1) upward.
The cell shaft for 28 mm cells is open.
! Insert a 28 mm cell (see below)
! Insert a 16 mm cell (see page 25)
1
Inserting a 28 mm cell
2Insert the cell so that it is positioned on the bottom of the cell
shaft.
The cell is ready to be measured.
3For turbidity measurement:
Align the cell (see section 4.7.2).
24
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Inserting a 16 mm cell
1Put the fold-out cell shaft (2) in an upright position until it locks
into place.
2
2Pull up the height adapter (3).
The cell shaft is extended.
3
3Open the external light cover (4) of the cell shaft.
4
ba75491d0407/2006
25
OperationpHotoFlex / pHotoFlex Turb
4Insert the 16 mm cell (marking points forward) so that it is po-
sitioned on the bottom of the cell shaft.
5Close the external light cover (4).
The cell is ready to be measured.
Note
For optimum measurement results, the cell must always be covered by
the external light cover. Otherwise, external light can falsify the measurement result.
26
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.3General operating principles
This section contains basic information on the operation of the
pHotoFlex (Turb).
Operating elements,
display
Operating modes,
navigation
An overview of the operating elements and the display is given in
section 1.2 and section 1.3.
An overview of the operating modes of the pHotoFlex (Turb) and the
navigation through menus and functions can be found in section 4.3.1
and section 4.3.2.
4.3.1Operating modes
The instrument has the following operating modes:
! Measurement
The display indicates measurement data in the measured value display
! Calibration
The display indicates a calibration process with calibration information,
or a process to carry out a zero adjustment
! Data transmission
The meter transmits measuring datasets or calibration records to the
serial interface
! Configuration
The display indicates a menu with further menus, settings and functions
4.3.2Navigation
Measured value displayIn the measured value display, you can
! select a measuring mode with <M> (long pressure)
! select a measured parameter in the active measuring mode (e. g.
pH <−> mV) with <M> (short pressure)
! open the menu with <MENU>
! switch to the superordinate Start menu with <ESC>.
ba75491d0407/2006
27
OperationpHotoFlex / pHotoFlex Turb
Menus and dialogsThe menus for settings and dialogs in courses contain further sub-
menus. The selection is made with the <▲><▼> keys.
The current selection is displayed in reverse video.
! Menus
The name of the menu is displayed at the upper edge of the frame.
Menus are opened by confirming with <START/ENTER>. Example:
Configuration
Photometry
Turbidity
pH & ORP
System
Info
! Settings
Settings are indicated by a colon. The current setting is displayed on
the right-hand side. With <START/ENTER>, the selection of the
possible settings is opened. Subsequently, the setting can be
changed with <▲><▼> and <START/ENTER>.
Example:
System
Language:Deutsch
Beep:
Illumination
Contrast
Temperature unit
Switchoff time
:On
:
:
:
Off
48 %
°C
30 min
! Functions
Functions are designated by the name of the function. They are immediately carried out by confirming with <START/ENTER>.
Example: display the Calibration record function
(in the pH & ORP / Calibration menu).
pH & ORP
Calibration record
Cal. type: AutoCal
TEC
Calibration interval
Unit for slope
:i 2.00 4.01 7.00 10.01
:
:
007 d
mV/pH
28
! Messages
Information or instructions are marked by the i symbol. They can-
not be selected.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Example:
pH & ORP
i Buffer recognition TEC
i Immerse sensor in buffer 1
The i symbol indicates
info texts, e.g.
messages, notes or instructions
Set temperature:25 °C
Continue
Note
The principles of navigation are explained in the two following sections
by reference of examples:
! Setting the language (section 4.3.3)
! Setting the date and time (section 4.3.4).
ba75491d0407/2006
29
OperationpHotoFlex / pHotoFlex Turb
4.3.3Navigation example 1: Setting the language
Note
The following example describes in the language of the country how to
set the language. On delivery, English is set as the language in the
pHotoFlex (Turb). During initial commissioning, the language is set in
the menu, Configuration / System / Language.
1In the measured value display:
Open the Configuration menu with <MENU>.
The instrument is in the configuration mode.
2Select the System menu with <▲><▼>.
The current selection is displayed in reverse video.
3Open the System menu with <START/ENTER>.
4Select the Language menu with <▲><▼>.
The current selection is displayed in reverse video.
System
Language:Deutsch
Store
Display
Reset
Interface
Continue ...
5Open the setting of theLanguagewith <START/ENTER>.
System
Language:
Store
Display
Reset
Interface
Continue ...
Deutsch
6Select the required language with <▲><▼>.
30
7Confirm the setting with <START/ENTER>.
The setting is active. The menu is displayed in the selected language.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
8To make further settings, switch to the next higher menu level
with <ESC>.
or
Switch to the measured value display with <M> (short pres-
sure).
The instrument is in the measurement mode.
4.3.4Navigation example 2: Setting the date and time
The meter has a clock with a date function. The date and time are indicated in the status line of the measured value display. When storing
measured values and calibrating, the current date and time are automatically stored as well.
Numerals are generally entered via the number keys.
Setting the date, time
and date format
The correct setting of the date and time and date format is important for
the following functions and displays:
! Current date and time
! Calibration date
! Identification of stored measured values.
Therefore, check the time at regular intervals.
Note
After a fall of the supply voltage (empty batteries or accumulator pack),
the date and time are reset to 01.01.2003, 00:00 hours.
The data format can be switched from the display of day, month, year
(dd.mm.yy) to the display of month, day, year (mm/dd/yy or mm.dd.yy).
1In the measured value display:
Open the Configuration menu with <MENU>.
The instrument is in the configuration mode.
2Select and confirm the System / Continue ... / Date/time menu
with <▲><▼> and <START/ENTER>.
ba75491d0407/2006
Date/time
Time:
:
Date
Date format
14:53:40
30.10.03
:dd.mm.yy
31
OperationpHotoFlex / pHotoFlex Turb
3Select and confirm the Time menu with <▲><▼> and
<START/ENTER>.
A display for the entry of numerals with the number keys opens
up.
Time
14:53:40
4Enter the time using the number keys.
The digit to be changed is displayed underlined.
Note
In the case of wrong entries, you can cancel the procedure with <ESC>.
After canceling with <ESC>, it is possible to enter all digits once again.
The new digits are only taken over by confirming with <START/EN-TER>.
5Confirm the setting with <START/ENTER>.
The time is set.
6Set the current Date as necessary. The setting is made similar-
ly to that of the time.
7Change the date format as necessary.
8To make further settings, switch to the next higher menu level
with <ESC>.
or
Switch to the measured value display with <M> (short pres-
The automatic switchoff
switches the meter off if
no entry is made for a
specified period of time
(Switchoff time). This
saves the batteries or accumulator pack.
Continue ... /
Beep
On
Off
Switch on/off the beep on
keystroke
4.4.1Measured value memory
In the Measured value memory menu, you find functions to display and
edit the stored measurement datasets:
! Display the measurement datsets on the screen (Display)
! Download the measurement datsets to the RS232 interface (RS232
download)
! Set up filter rules for the stored measurement datsets (Data filter)
! Erase all stored measurement datsets (Delete)
Settings/functions
! Information on the number of occupied memory locations
The settings can be found in the Configuration / System / Measured val-ue memory menu.
To switch to the Configuration menu, press the <MENU> key.
Menu itemSetting/func-
Description
tion
Display-Displays in pages all mea-
surement datasets that correspond to the filter settings.
Further options:
! Scroll through the
datasets with <▲><▼>.
! Output the displayed
dataset to the interface
with <PRT>.
! Quit the display with
<ESC>.
36
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Menu itemSetting/func-
Description
tion
RS232 download-Downloads to the interface
all measurement datasets
that correspond to the filter
settings. The download is
ordered according to the
date and time.
The process can take several minutes. To terminate the
process prematurely, press
<ESC>.
Data filtersee
section 4.8.2
Allows to set filter criteria in
order to display and download datasets to the interface.
Delete-Erases the entire contents
of the measuring data memory, independent of the filter
settings.
Note:
All calibration data remains
stored when performing this
action.
All details on the subjects of memory and stored data is found in
section 4.8.2.
ba75491d0407/2006
37
OperationpHotoFlex / pHotoFlex Turb
4.4.2Display
In the Configuration / System / Display menu, you set the display fea-
tures:
! Switching on/off the display illumination (Illumination)
! Display contrast (Contrast)
The settings can be found in the Configuration / System / Display menu.
To switch to the Configuration menu, press the <MENU> key.
Settings
Menu itemSettingDescription
IlluminationAuto offThe display illumination is
automatically switched off if
no key has been pressed
for 30 seconds.
On
Off
Switches the display illumination on or off permanently (see section 4.5.8)
Contrast0 ... 100 %Changes the display con-
trast
Brightness0 ... 100 %Changes the display bright-
ness
4.4.3Interface
In the Interface menu, you set the features of the interface:
! Transmission speed (Baud rate)
! Output format (Output format)
38
Settings
The settings can be found in the Configuration / System / Interface
menu.
To switch to the Configuration menu, press the <MENU> key.
Menu itemSettingDescription
Baud rate1200, 2400,
4800, 9600,
Baud rate of the data
interface
19200
Output formatASCII
CSV
Output format for data
transmission
For details, see section 4.9
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.4.4Date/time
In the Configuration / System / Continue ... / Date/time menu, you set
the system clock:
! Current time (Time)
! Current date (Date)
! Format of the date display (Date format)
The settings can be found in the Configuration / System / Continue ...
Date/time menu.
To switch to the Configuration menu, press the <MENU> key.
Settings
Menu itemSettingDescription
Timehh:mm:ssEnter the time with the
number keys
DateEnter the date with the
number keys
Date formatdd.mm.yy
Settings of time and date.
mm.dd.yy
mm/dd/yy
ba75491d0407/2006
39
OperationpHotoFlex / pHotoFlex Turb
4.5Photometry
4.5.1General information
Photometric measurements serve to determine chemical substances in
liquid samples. For this determination, the substance to be determined
has to be present in a form that is suitable for photometric measurement. At the same time, possible disturbing factors have to be excluded.
Before measurement, the sample has to be pretreated in order to bring
the substance to be determined into the form that is suitable for measurement and at the same time exclude disturbing factors.
Pretreatment of the sample is described in the analysis specification.
In a simple case, pretreatment can be to dissolve a solid substance in
water; it can, however, also include chemical conversions, e. g. a digestion.
The chemicals required in the analysis specification are available as
test sets.
Note
Suitable analysis specifications for test sets can be found in the photometry analysis manual (on CD-ROM).
There you will also find further instructions on handling chemicals and
on how to proceed when applying the analysis specifications.
Methods and the corresponding method data for many test sets are
stored as programs
in the pHotoFlex (Turb). A program number is as-
signed to each program.
By entering the program number or by using a barcode reader the
stored method data is loaded.
You can look up an overview of the available methods in the photometry analysis manual and display it on the screen of the pHotoFlex (Turb)
(see section 4.5.7).
You can measure the following parameters with the pHotoFlex (Turb):
! Concentration [mg/l]
! % Transmission []
40
! Absorbance []
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Clean the cells before filling them with sample and also before
measuring as necessary (see section 5.2.2).
The cells must be absolutely clean and free of scratches.
2For measurement, place the pHotoFlex (Turb) on a horizontal
surface.
4.5.2Settings for photometric measurements
For photometric measurements, the following settings are available in
the Configuration / Photometry menu:
! Setting the measured parameter
! Displaying a list of all programs
! Setting the dilution factor
Settings
! Switching on or off the analysis timer
! Resetting the settings for photometric measurements
The settings can be found in the Configuration / Photometry menu.
To switch to the Configuration menu, press the <MENU> key.
Menu itemSettingDescription
Measured parameter
Concentration
% Transmission
Absorbance
Measured parameters in
the Photometry measuring
mode
ProgramsDisplay all programs with
the corresponding program data (see
section 4.5.7).
DilutionSet the dilution factor (see
section 4.5.10)
Analysis timerOn
Off
Switch on/off the analysis
timer (see section 4.5.8)
ResetReset all settings for the
Photometry measuring
mode (see section 4.10.3)
ba75491d0407/2006
41
OperationpHotoFlex / pHotoFlex Turb
4.5.3Measuring the concentration
1Press the <M> key (long pressure) repeatedly until the Pho-
tometry measuring mode is selected.
2Press the <M> key (short pressure) repeatedly until the mea-
sured variable, Konzentration is selected.
First concentration measurement with the pHotoFlex (Turb)
Photometry \ Concentration
i Select program
with <PROG>
01.02.05 15:12
Second and all further
concentration measurements
Photometry \ Concentration
i Select program
with <PROG>
or with ▲
1: A5/25 MCNH4-N
16 mm0.20 - 6.51 mg/l
▼
01.02.05 15:12
Note
From the second concentration measurement, the data of the program
last used is automatically displayed here.
With <▲><▼> you can quickly switch between the ten programs last
used.
To select a program, you can also read in the program number of an
analysis specification with a barcode reader (see section 8.2). The following step 3 is then skipped. You can directly start measurement.
The program number of the test is given in the analysis specification,
on the list of available programs and on the packing of some tests (under the barcode).
42
3Open the Program number display with <PROG>,
enter the required program number with the number keys and
confirm with <START/ENTER>.
or (from the second concentration measurement):
Select a program out of the last ten programs with <▲><▼>.
The program data is displayed.
Note
If a program number is selected that requires a measured blank value,
the menu automatically guides to the blank value measurement.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Photometry \ Concentration
i Insert sample
i Start measurement
with <START>
1: A5/25 MCNH4-N
16 mm0.20 - 6.51 mg/l
01.02.05 15:12
4Insert the cell (see section 4.2).
5Start the measurement with <START/ENTER>.
Measurement is started. The result is displayed.
Photometry \ Concentration
[BV]
A blank value measured by
the user is used
0.74mg/l
1: A5/25 MCNH4-N
16 mm0.20 - 6.51 mg/l
01.02.05 15:12
4.5.4Blank value (reagent blank value)
A blank value is required for every concentration measurement.
For some programs (methods) for concentration measurement, the
blank values are already stored in the photometer. They are used automatically. For all other programs, the blank value has to be determined separately before the first measurement.
Each stored reagent blank value can be replaced by a blank value determined by the user.
Note
You will find more information on blank values in the photometry analysis manual.
A blank value is always stored for the program that has just been called
up. It remains stored until it is erased (menu item, Delete blank value)
or overwritten.
The Reset function erases all blank values measured by the user and
restores the blank values stored in the factory.
ba75491d0407/2006
If a blank value measured by the user is stored for a program, this blank
value is used for measurement. The usage of the blank value measured by the user is documented and also indicated in the measured
value display.
43
OperationpHotoFlex / pHotoFlex Turb
Measuring the blank
value
1Press the <M> key (long pressure) repeatedly until the Pho-
tometry measuring mode is selected.
2Press the <M> key (short pressure) repeatedly until the mea-
sured variable, Concentration is selected.
3Select a program with <PROG> as necessary.
Note
The following measurement of the blank value applies only to the selected program. If no program is selected, the message i No program selected. appears on the display.
4Open the adjustment menu with <CAL/ZERO>.
Photometry \ Adjustment
Zero adjust.
Measure blank value
Delete blank value
5Using <▲><▼> and <START/ENTER>, select and start the
Measure blank value function.
The menu-guided blank value measurement starts.
Follow the instructions on the display.
Measure blank value
i 2: A5/25 MC
Cell = 16 mm
i
i Insert blank sample
Start measurement
6Insert a cell with blank sample (see section 4.2).
7Start the measurement of the blank value with <START/EN-
TER>.
After measuring, the result of the blank value measurement is
displayed and stored.
The result is displayed as i Blank value measurement suc-cessful! or i Blank value measurement erroneous!
44
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Measure blank value
i 2: A5/25 MC
i Cell = 16 mm
i Absorbance = 0.301
i Blank value mea-
surement
successful!
8Confirm the result with <START/ENTER>.
The blank value measurement is completed.
The meter is ready to measure.
or:
Discard the result with <ESC>.
Subsequently, carry out a new blank value measurement.
4.5.5Measuring the absorbance/transmission
Note
The transmission measurement is not described separately in the following example as it operates in exactly the same way as the absorbance measurement. The result of the measurement is displayed in % Transmission.
1Press the <M> key (long pressure) repeatedly until the Pho-
tometry measuring mode is selected.
2Press the <M> key (short pressure) repeatedly until the mea-
sured variable, Absorbance or % Transmission is selected.
Photometry \ Absorbance
i Select cell
with
16 mm610 nm
▲ ▼
01.02.05 15:12
ba75491d0407/2006
3Select the cell diameter with <▲><▼> and confirm with
<START/ENTER>.
4Select the wavelength with <▲><▼> and confirm with
<START/ENTER>.
45
OperationpHotoFlex / pHotoFlex Turb
Photometry \ Absorbance
i Insert sample
i Start measurement
with <START>
16 mm610 nm
01.02.05 15:12
5Clean the cell (see section 5.2.2).
6Insert the cell (see section 4.2).
7Start the measurement with <START/ENTER>.
The measurement result is displayed when the measurement
is finished.
Photometry \ Absorbance
0.532
16 mm610 nm
01.02.05 15:12
4.5.6Zero adjustment
The zero adjustment, i. e. measuring and storing the absorbance of a
cell filled with water, is necessary after the meter is switched on.
Additionally, we recommend to carry out a zero adjustment if the ambient temperature has changed.
Only perform the zero adjustment against distilled water in an optically
perfect cell. The zero adjustment must be performed separately for
each cell type.
1Press the <M> key (long pressure) repeatedly until the Pho-
tometry measuring mode is selected.
46
2Press the <M> key (short pressure) repeatedly until the mea-
sured variable, Concentration is selected.
3Press the <CAL/ZERO> key.
The menu for adjustment measurements opens up.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Photometry \ Adjustment
Zero adjust.
Measure blank value
Delete blank value
4Using <▲><▼> and <START/ENTER>, select and start the
Zero adjust. function.
The menu-guided zero adjustment starts.
Follow the instructions on the display.
Zero adjust.
i Insert zero
cell (dist. water)
Cell16 mm
Start measurement
5Insert the cell (see section 4.2)
6Set another cell with <▲><▼> and <START/ENTER> as nec-
essary.
7Start the measurement of the zero adjustment with <START/
ENTER>.
After measuring, the result of the zero adjustment is displayed
and stored.
i Zero adjust. successful! (successful zero adjustment) or
i Calibration error! (zero adjustment not successful)
is displayed as the result.
The zero adjustment is completed.
Note
If i Calibration error! was displayed as the calibration result, a note au-
tomatically reminds you of another zero adjustment before the next
measurement.
Measuring is not possible without a valid zero adjustment.
ba75491d0407/2006
8Confirm the result with <START/ENTER>.
The zero adjustment is completed.
The meter is ready to measure.
47
OperationpHotoFlex / pHotoFlex Turb
4.5.7Programs
Displaying program
data
You can view the most important data of all methods.
The method data is ordered according to the program number.
1Open the Configuration / Photometry / Programs menu.
The display shows the most important data of the selected program.
Programs
1: A5/25 MC
0.20 - 8.00 mg/l
NH4-N
16 mm
i Scroll with
▲ ▼
Program number, test, code
Measuring range with unit
Citation form
Diameter of the cell
Note
This data is also to be found in the photometry analysis manual in the
overview of the test sets and in the individual analysis specifications for
the test sets.
Updating programsUnder http://www.WTW.com on the Internet, you can always find the
latest software version with the newest programs and method data for
your pHotoFlex (Turb) (see A
PPENDIX: FIRMWAREUPDATE).
User-defined programsUser-defined programs (methods) can be stored under program num-
bers between 900 and 999. You can store up to 100 user-defined programs (see section 4.13).
48
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.5.8Analysis timer
Measuring according to analysis specifications often means there are
waiting periods between the individual steps.
These waiting periods (time intervals) are stored in the instrument with
the program data for each program. The active Analysis timer function
automatically reminds you to observe these time intervals by means of
the menu guidance.
If you want to manually enter time intervals, use the Timer function (see
section 4.5.9).
The Analysis timer with the required time interval is automatically displayed at the due point.
Start the Analysis timer with the <START/ENTER> key.
It is not possible to shorten the time intervall.
An acoustic signal sounds when the adjusted time interval has expired.
The Analysis timer function is switched on or off in the Configuration /
Photometry/Analysis timer menu.
This setting generally applies to all measurements with methods according to analysis specification.
4.5.9Timer
When measuring according to analysis specifications, waiting periods
often have to be kept between individual steps of the method.
With the Timer function you manually set a time interval.
If you want to be automatically reminded of the given time interval, use
the Analysis timer function (see section 4.5.8).
The timer is displayed in the measured value display. It always displays
the remaining time of the adjusted time interval.
When the adjusted time interval has expired, the timer indicates
00:00:00 and an acoustic signal sounds.
The Timer function is started in the Configuration / Timer menu by en-
tering a time interval.
ba75491d0407/2006
49
OperationpHotoFlex / pHotoFlex Turb
4.5.10 Measuring diluted samples
If the concentration of a test sample exceeds the measuring range of a
method, you can dilute the sample by a factor 1 ... 99 so that the concentration of the diluted test sample is within the measuring range of
the method (see photometry analysis manual). Thus a valid measurement is possible.
After entering the factor for the dilution the meter converts the concentration to that of the undiluted sample.
The display then indicates the measured value of the undiluted sample.
Entering the factor of
the dilution
1Select the program for which a dilution factor is to be entered.
2Open the Configuration / Photometry / Dilution menu.
The current factor of the dilution is displayed.
Dilution
Water + sample0 + 1
i PROG 1
3Open the display for the entry of numerals with <START/EN-
TER>.
4Enter the factor of the dilution with the number keys.
The factor has to be a whole number between 0 ... 99.
5Confirm the factor with <START/ENTER>.
6Exit the Dilution menu with <ESC>.
For the following measurements with the selected program, the
concentration of the undiluted sampled is displayed as the
measurement result.
50
The entered dilution factor is only valid for the selected program. The
dilution factor is erased if:
! the photometer is switched off
! a different program number is selected
! the factor 0 is entered in the Dilution menu.
If a dilution factor is active, it is indicated on the display during measurement in the form [x + 1].
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.6pH value / ORP voltage
4.6.1General information
You can measure the following variables:
! pH value [ ]
! ORP [mV]
Caution
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result. The
RS232 interface is not galvanically isolated.
Temperature
measurement
For reproducible pH measurements, it is essential to measure the temperature of the test sample.
You have the following possibilities for measuring the temperature:
! Automatic measurement of the temperature by a temperature sen-
sor (NTC30 or Pt 1000) integrated in electrode.
! Manual determination and input of the temperature.
The meter recognizes whether a suitable electrode is connected and
automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring mode:
Temperature
sensor
Resolution of the
temp. display
Temperature of the test
sample
yes0.1 °Cautomatic measurement
-1°Cmanual measurement
and entry
Preparatory activitiesPerform the following preparatory activities when you want to measure:
ba75491d0407/2006
1Connect a pH or ORP electrode to the meter.
2Press the <M> key (long pressure) repeatedly until the pH &
ORP measuring mode is selected.
3Press the <M> key (short pressure) repeatedly until the mea-
sured parameter, pH or ORP is selected.
4Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without a temperature sensor.
5Calibrate or check the meter with the electrode.
51
OperationpHotoFlex / pHotoFlex Turb
4.6.2Measuring the pH value
1Perform the preparatory activities according to section 4.6.1.
2Immerse the pH electrode in the test sample.
AutoRead
(Drift control)
CriteriaWith identical measurement conditions, the following applies:
pH & ORP
\
pH
6.94
25.0
[AR]
3Press the <M> key (short pressure) repeatedly until the mea-
sured variable, pH is selected.
The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured parameter flashes until a stable measured value is available.
Measured
parameter
pH valueBetter than 0.01> 30 seconds
°C
01.02.05 15:12
ReproducibilityResponse time
52
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.6.3Measuring the ORP voltage
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution.
1Perform the preparatory activities according to section 4.6.1.
2Submerse the ORP electrode in the sample.
pH & ORP
[AR]
AutoRead
(drift control)
CriteriaWith identical measurement conditions, the following applies:
The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured parameter flashes until a stable measured value is available.
Measured
parameter
ORP voltagebetter than 1 mV> 30 seconds
\
ORP
157
25.0
3Press the <M> key (short pressure) repeatedly until the mea-
sured parameter, ORP is selected.
mV
°C
01.02.05 15:12
ReproducibilityResponse time
OverviewFor pH and ORP measurements, the following settings are available in
ba75491d0407/2006
4.6.4Settings for pH and ORP measurements
the Configuration / pH & ORP menu:
! Measured parameter
! Calibration record (display, print)
! Selecting the calibration type
! Entering the Calibration interval
! Selecting the Unit for slope
! Selecting the Temperature unit
! Reset
53
OperationpHotoFlex / pHotoFlex Turb
Settings/functionsThe settings can be found in the Configuration / pH & ORP menu.
To switch to the Configuration menu, press the <MENU> key.
Menu itemPossible
setting
Measured parameterpH & ORP
mV
Calibration /
-Displays the calibration
Calibration record
Calibration /
Cal. type
Calibration /
TEC
NIST/DIN
1 ... 999 dCalibration interval for the
Calibration interval
Calibration /
Unit for slope
mV/pH
%
Description
record of the last calibration.
Buffer sets to be used for
pH calibration.
For details, see
section 4.6.5.
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measured value display.
Unit of the slope.
The % display refers to the
Nernst slope of
-59.16 mV/pH (100 x determined slope/Nernst
slope).
54
Man. temperature-20 ... +130 °CEntry of the manually de-
termined temperature. For
measurements without
temperature sensor only.
Temperature unit°C, °FDegrees Celsius
Degrees Fahrenheit
ResetReset all settings for the
pH & ORP measuring
mode (see section 4.10.3)
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.6.5Calibration
Why calibrate?pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is displayed. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the meter. Thus, you
should calibrate at regular intervals.
When to calibrate?! After connecting another electrode
! When the sensor symbol flashes:
– after the calibration interval has expired
– after voltage interruption (e.g. empty batteries, empty
accumulator pack)
Buffer sets for
calibration
You can use the buffer sets quoted in the table for an automatic calibration. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into account during
calibration.
Buffer setName on the
display
pH values at
25 °C
Technical buffer solutionsTEC2.00
4.01
7.00
10.01
NIST/DIN buffer solutionsNIST/DIN1.679
4.006
6.865
9.180
12.454
Note
The buffers are selected in the Configuration / pH & ORP / Cal. type
menu, see section 4.6.4).
ba75491d0407/2006
55
OperationpHotoFlex / pHotoFlex Turb
Calibration pointsCalibration can be performed using one, two or three buffer solutions in
any order (single-point-, two-point or three-point calibration). The meter
determines the following values and calculates the calibration line as
follows:
Determined val-
Displayed calibration data
ues
1-pointAsy! Asymmetry = Asy
! Slope = Nernst slope
(-59.16 mV/pH at 25 °C)
2-pointAsy
Slp.
3-pointAsy
Slp.
! Asymmetry = Asy
! Slope = Slp.
! Asymmetry = Asy
! Slope = Slp.
The calibration line is calculated
by linear regression.
Note
You can display the slope in the unit, mV/pH or % (see section 4.6.4).
AutoReadThe calibration procedure automatically activates the AutoRead func-
tion.
The current AutoRead measurement can be terminated at any time
(accepting the current value).
Calibration recordWhen finishing a calibration, the new calibration values are displayed
Displaying and down-
loading calibration data
to interface
56
as an informative message (
i symbol) first. Then you can decide
whether you want to take over these values of the new calibration or
whether you want to continue measuring with the old calibration data.
After accepting the new calibration values the calibration record is displayed.
You can view the data of the last calibration on the display. Subsequently, you can download the displayed calibration data to the interface, e. g. to a printer or PC, with the <PRT> key.
The calibration record of the last calibration can be found under the
Configuration / pH & ORP / Calibration / Calibration record menu item.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Sample printout of a
record
31.10.03 16:13
pHotoFlex Ser. no. 12345678
Calibration pH & ORP
Calibration date 31.10.03 16:13:33
Calibration interval 7 d
AutoCal TEC
Buffer 1 4.01
Buffer 2 7.00
Buffer 3 10.01
Voltage 1 184.0 mV 24.0 °C
Voltage 2 3.0 mV 24.0 °C
Voltage 3 -177.0 mV 24.0 °C
Slope -60.2 mV/pH
Asymmetry 4.0 mV
Sensor +++
Calibration evaluationAfter calibrating, the meter automatically evaluates the calibration. The
asymmetry and slope are evaluated separately. The worse evaluation
of both is taken into account. The evaluation appears on the display
and in the calibration record.
Sample display
Calibration
record
Asymmetry
[mV]
Slope
[mV/pH]
+++-15 ... +15-60.5 ... -58
++- 20 ... +20-58 ... -57
+-25 ... + 25-61 ... -60.5
--30 ... +30-62 ... -61
Clean the electrode according
to the electrode operating
manual
--------
Eliminate the error according to
chapter 6 W
HATTODOIF...
< -30 or
> 30
or
-57 ... -56
or
-56 ... -50
... -62 or
... -50
ba75491d0407/2006
57
OperationpHotoFlex / pHotoFlex Turb
Preparatory activitiesPerform the following preparatory activities when you want to calibrate:
1Connect the pH electrode to the meter.
The pH measured value display is displayed on the screen.
2Keep the buffer solutions ready. Adjust the temperature of the
buffer solutions, or measure the current temperature if you
measure without a temperature sensor.
4.6.6Carrying out the TEC and NIST/DIN calibration procedures
The two calibration procedures only differ in the usage of different buffer sets (see section 4.6.5). Make sure the Cal. type is correctly set in
the pH & ORP / Calibration menu (see section 4.6.4).
For this procedure, use any one, two or three WTW technical buffer solutions in ascending or descending order.
The TEC calibration is described below. With the NIST/DIN calibration,
the NIST/DIN buffer recognition and different nominal buffer values are
displayed. Apart from that, the procedure is identical.
Note
The TEC calibration for pH 10.01 is optimized for the WTW technical
buffer solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions
can lead to an erroneous calibration. The correct buffer solutions are
given in the WTW catalog or on the Internet.
1Press the <M> key (short pressure) repeatedly until the mea-
sured parameter, pH or ORP is selected.
2Start the calibration with <CAL/ZERO>.
The calibration display appears.
pH & ORP \ Calibration
i Buffer recognition TEC
i Immerse sensor in buffer 1
Continue
58
3Immerse the electrode in buffer solution 1.
4If the Set temperature menu item appears, measure and enter
the temperature of the buffer manually (measurement without
temperature sensor).
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
5Using <▲><▼>, select Continue and press <START/EN-
TER>. The buffer is measured.
The measured value is checked for stability (AutoRead).
pH & ORP \ Calibration
i Buffer value = 7.00
i U = 3 mV
i Temperature =
Terminate AutoRead
24.8 °C
6Wait for the end of the AutoRead measurement or accept the
calibration value with <START/ENTER>.
The calibration display for the next buffer appears.
pH & ORP \ Calibration
i Buffer recognition TEC
i Immerse sensor in buffer 2
Exit with one point calibration
Continue
7For single-point calibration, select Exit with one point calibra-
tion with <▲><▼> and confirm with <START/ENTER>.
The calibration is completed as a single-point calibration.
The new calibration values are displayed as an informative
message (
i).
You have the following options:
! Accept the new calibration values with <START/ENTER>.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
! To switch to the measured value display without
accepting
the new calibration values, press <M> (short pressure) or
<ESC>.
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the asymmetry of the electrode.
ba75491d0407/2006
59
OperationpHotoFlex / pHotoFlex Turb
Continuing for two-point
calibration
(Cal. typeTEC)
8Thoroughly rinse the electrode with distilled water.
9Immerse the electrode in buffer solution 2.
10If the Set temperature menu item appears, measure and enter
the temperature of the buffer manually (measurement without
temperature sensor).
11Using <▲><▼ >, select Continue and press <START/EN-
TER>. The buffer is measured.
The measured value is checked for stability (AutoRead).
pH & ORP \ Calibration
i Buffer value = 10.01
i U = -177 mV
i Temperature =24.8 °C
Terminate AutoRead
12Wait for the end of the AutoRead measurement or Terminate
AutoRead with <START/ENTER> and take over the calibration
value.
The calibration display for the next buffer appears.
pH & ORP \ Calibration
i Buffer recognition TEC
i Immerse sensor in buffer 3
Exit with 2 point calibration
Continue
13For two-point calibration, select Exit with 2 point calibration with
<▲><▼> and confirm with <START/ENTER>.
The calibration is completed as a two-point calibration.
The new calibration values are displayed as an informative
message (
i).
You have the following options:
! Accept the new calibration values with <START/ENTER>.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
! To switch to the measured value display without
accepting
the new calibration values, press <M> (short pressure) or
<ESC>.
60
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Continuing for three-
point calibration
(Cal. typeTEC)
14Thoroughly rinse the electrode with distilled water.
15Immerse the electrode in buffer solution 3.
16If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with <▲><▼> and <START/ENTER>
in the Set temperature setting.
17Using <▲><▼>, select Continue and press <START/EN-
TER>. The buffer is measured.
The measured value is checked for stability (AutoRead).
pH & ORP \ Calibration
i Buffer value = 4.01
i U = 184 mV
i Temperature =24.8 °C
Terminate AutoRead
18Wait for the end of the AutoRead measurement or Terminate
AutoRead with <START/ENTER> and take over the calibration
value.
The new calibration values are displayed as an informative
message (
i).
You have the following options:
! Accept the new calibration values with <START/ENTER>.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
ba75491d0407/2006
! To switch to the measured value display without
accepting
the new calibration values, press <M> (short pressure) or
<ESC>.
61
OperationpHotoFlex / pHotoFlex Turb
4.7Turbidity
4.7.1General information
Venting the sampleAir bubbles in the sample affect the measuring result to a massive ex-
tent because they have a large scattering effect on the incident light.
Larger air bubbles cause sudden changes in the measured values
whereas smaller air bubbles are recorded by the instrument as turbidity. Therefore, avoid or remove air bubbles:
Avoiding or removing
air bubbles
! During sampling, ensure all movement is kept to a minimum
! If necessary, vent the sample (ultrasonic baths, heating or adding a
surface-active substance to reduce the surface tension)
4.7.2Aligning and marking a cell
Even completely clean quality cells exhibit tiny directional differences
in their light transmittance. Therefore, if you want to achieve accurate
and reproducible measurement results, it is necessary to always align
the sample cells and cells for calibration standards in the same way
(see section 2130 of the "Standard Methods for the Examination of Water and Wastewater", 19th edition).
To do so, the optimum alignment of the cell is determined.
Aligning the cell
1Press the <M> key (long pressure) repeatedly until the Turbid-
ity measuring mode is selected.
2Clean the cell (see section 5.2.2).
3Insert the cell (see section 4.2).
4Align the cell:
! Press and hold the <START/ENTER> key.
! Slowly and in small steps turn the cell by one complete ro-
tation (by 360 °).
After each step wait for a short time until the displayed measured value is stable.
! Turn the cell back to the position with the lowest measured
value.
Note
To keep the drift as low as possible, the time for aligning the cell while
pressing and holding the <START/ENTER> key is limited to 30 sec-
onds. After this time, the meter starts measuring automatically.
5Release the <START/ENTER> key.
Measurement starts. The measured value is displayed.
62
Marking a cellTo be able to quickly bring a cell into the optimum position, it is helpful
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
to mark the optimum position of the cell once it is determined. This
shortens each measurement or calibration procedure with this cell considerably.
The marking can, e. g., be done on a label on the cap of the cell.
6Mark the optimum position of the cell.
The cell is prepared for the shortened measuring and calibration procedures.
4.7.3Measuring turbidity
Caution
Never pour any liquids directly into the cell shaft. Always use a
cell for measurement. The meter only measures precisely if the
cell is closed with the black light protection cap (WTW cells).
Measuring
Note
The outside of the cell always has to be clean, dry, and free of fingerprints and scratches. Clean the cell before starting to measure (see
section 5.2.2). Only hold the cells by the top or by the black light protection cap.
1Press the <M> key (long pressure) repeatedly until the Turbid-
ity measuring mode is selected.
2Rinse out a clean cell with the sample to be measured:
Pour approximately 10 ml sample into the cell. Close the cell
and rotate it several times before throwing the sample away.
3Repeat the rinsing procedure twice more.
4Fill the cell with the sample to be measured (approx. 15 ml).
Close the cell with the black light protection cap.
5Clean the cell (see section 5.2.2).
6Insert the cell (see section 4.2).
ba75491d0407/2006
63
OperationpHotoFlex / pHotoFlex Turb
7Align the cell:
! Marked cell
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and for a short time hold the <START/ENTER>
key until the measured value is displayed.
! Unmarked cell (see page 62)
– Press and hold the <START/ENTER> key.
– Slowly and in small steps turn the cell by one complete
rotation (by 360 °). After each step wait for a short time
until the displayed measured value is stable.
– Turn the cell back to the position with the lowest mea-
sured value.
Note
To keep the drift as low as possible, the time for aligning the cell while
pressing and holding the <START/ENTER> key is limited to 30 sec-
onds. After this time, the meter automatically starts measuring or calibrating.
Display if the
measuring range is
exceeded
8Release the <START/ENTER> key.
Measurement starts. The measured value is displayed.
Turbidity
157.0
9Repeat the steps 2 to 8 for further samples.
If the measured value is outside the measuring range of the
pHotoFlex Turb, it is indicated on the display:
Turbidity
> 1100
FNU
NTU
01.02.05 15:12
FNU
NTU
64
01.02.05 15:12
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.7.4Calibration
When to calibrate?! Regularly every 90 days
! With a temperature change
Calibration procedures
and calibration
For the menu-guided three-point calibration you need the following
three calibration standards in the mentioned order:
standards
Standard no.NTU/FNU
11000
210,0
30,02
Preparing calibrationPerform the following preparatory activities when you want to calibrate:
1Keep the cells with the required calibration standards ready
and mark them as necessary (see page 62).
2Clean the cell (see section 5.2.2).
3Insert the cell (see section 4.2).
Caution
Never open the cells with the calibration standards.
Carrying out calibration
ba75491d0407/2006
1Press the <M> key (long pressure) repeatedly until the Turbid-
ity measuring mode is selected.
2Press the <CAL/ZERO> key.
The menu-guided calibration begins.
Follow the instructions on the display.
Turbidity \ Calibration
i Insert standard
1000 FNU/NTU
i Press and hold <START>
i Align sample
3Insert the cell with the displayed calibration standard (here e.g.
1000 NTU/FNU) in the cell shaft (see section 4.2).
65
OperationpHotoFlex / pHotoFlex Turb
4Align the cell:
! Marked cell:
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and time hold the <START/ENTER> key until the
measured value is displayed.
! Unmarked cell (see page 62)
– Press and hold the <START/ENTER> key.
– Slowly and in small steps turn the cell by one complete
rotation (by 360 °).
– After each step wait for a short time until the displayed
measured value is stable.
– Turn the cell back to the position with the lowest mea-
sured value.
Turbidity \ Calibration
i Turb.= 1000 FNU/NTU
i Start calibration
by releasing <START>
5Release the <START/ENTER> key.
Measurement of the calibration standard begins.
Note
Before measuring the third calibration standard of 0.02 NTU/FNU you
can exit the calibration with <ESC> at any time.
The new calibration data is discarded. The old calibration data is further
used.
6Repeat the steps 4 - 6 with the calibration standards, 10.0 NTU/
FNU and 0.02 NTU/FNU.
After measuring the 0.02 NTU/FNU calibration standard, the
calibration result (i Calibration successful! or i Calibration er-ror!) is displayed and stored.
Calibration is completed.
66
7Confirm the calibration result with <START/ENTER>.
The display shows instructions for the first measurement.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Turbidity
i Insert sample
i Press and hold <START>
i Align sample
Note
If i Calibration error! was displayed as the calibration result, a note ap-
pears on the display to recalibrate before measuring.
Should a valid calibration not be possible the meter also offers to continue measuring with the last valid calibration data.
ba75491d0407/2006
67
OperationpHotoFlex / pHotoFlex Turb
4.8Storing
The meter has 1000 storage locations for measurement datsets.
You can transmit measured values (datasets) to the data storage with
the <STO> key.
Each storage process transmits the current dataset to the interface at
the same time.
The number of storage locations that are still free is displayed in the
Store menu. The number of storage locations that are occupied is displayed in the System \ Measured value memory menu.
! and the following measurement data depending on the selected
measuring mode
Measuring
Measurement data
mode
Photometry:! Program number
! Measured value
! Citation form
! Use of a blank value (BV)
! Dilution (x +1)
Turbidity:! Measured value
pH & ORP:! Measured value (pH/mV)
! Measured temperature value (°C/°F)
! AutoRead info (AR)
AR appears with the measured value if the AutoRead criterion was met while storing (stable measured value). Otherwise, the AR display is missing.
68
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.8.1Storing measurement datsets
Proceed as follows to transmit to the data storage and simultaneously
output to the interface a measurement dataset:
1Press the <STO> key.
The Store display appears.
Store (996 free)s
i 02.02.2005 11:24:16
0.00 mg/l PO4-P BV
PROG 1
ID:
Store (ID: 1:
1
2Using <▲><▼>, <START/ENTER> and the number keys,
change and confirm the ID number (ID) as necessary (0 ...
999).
3Using <START/ENTER> or <STO>, confirm Store.
The dataset is stored. The instrument switches to the measured value display.
Note
A measurement dataset is stored quickly by twice pressing <STO>. It
is stored with the ID last set.
If the storage is fullYou can erase the entire storage (see section 4.8.5), or overwrite the
oldest dataset with the next storing procedure.
A security prompt appears before a dataset is overwritten.
ba75491d0407/2006
69
OperationpHotoFlex / pHotoFlex Turb
4.8.2Filtering measurement datsets
The functions to display and download stored measurement datsets
(see section 4.4.1) refer to all stored measurement datsets that correspond to the adjusted filter criteria.
The settings can be found in the Configuration / System / Measured val-ue memory /Data filter menu.
To switch to the Configuration menu, press the <MENU> key.
Data filter
Menu itemSetting/functionDescription
FilterFilter criteria:
No filterData filter switched off
IDSelection according to ID
number
PROGSelection according to pro-
gram
DateSelection according to pe-
riod
ID + PROGSelection according to pro-
gram and ID number
ID + DateSelection according to pe-
riod and ID number
PROG + DateSelection according to pro-
gram and date
ID + PROG + DateSelection according to ID,
program and date
70
IDEntry of filter criteria
PROG
These menu items are
made visible by selecting
Date
the filter criteria in the Filter
menu.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.8.3Displaying measurement datsets
You can read out stored datasets on the display. Only those datasets
are displayed that correspond to the selected filter criteria (see
section 4.8.2).
Start reading out the data on the display in the menu, Configuration / System / Measured value memory / Display.
Display of a dataset
02.02.2005 11:24:16
ID: 1
7.000 (pH) 24.8 °CAR
i Scroll with
▲ ▼
Further datasets that correspond to the filter criteria are displayed with
the <▲><▼> keys.
Quitting the displayTo quit the display of stored measurement datasets, you have the fol-
lowing options:
! Switch directly to the measured value display with <M> (short pres-
sure).
! Leave the display and switch to the higher menu with <ESC> or
<START/ENTER>.
4.8.4Downloading the measurement datsets to the RS232
interface
ba75491d0407/2006
You can download stored datasets to the RS232 interface. Only those
datasets are downloaded that correspond to the selected filter criteria
(see section 4.8.2).
The datasets are downloaded in the adjusted output format (see
section 4.9.3).
The data download to the interface is started in the menu, Configura-tion / System / Measured value memory / RS232 download.
71
OperationpHotoFlex / pHotoFlex Turb
4.8.5Erasing stored measurement datasets
You can erase the stored measurement datsets altogether if you do no
longer need them.
Erasing all measurement datsets is done in the menu, Configuration / System / Measured value memory / Delete.
Note
Erasing individual datasets is not possible. If all storage locations are
occupied, however, it is possible to overwrite the oldest dataset at a
time. A security prompt appears before a dataset is overwritten.
4.9Transmitting data (RS 232 interface)
Via the RS 232 interface, you can transmit data to a PC or an external
printer.
4.9.1Connecting a PC/external printer
Connect the interface to the devices via the AK540/B (PC) or AK540/S
(external printer) cable.
Caution
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
Baud ratecan be selected from:
1200, 2400, 4800, 9600, 19200
The baud rate must agree with the baud rate set on
the PC/printer.
HandshakeRTS/CTS
PC only:
Paritynone
Data bits8
72
Stop bits1s
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
Socket assignment
3
4
6
1
5
1 RTS
2
2 RxD
3 TxD
4 DTR
5 SG
RS 232
REC
6 CTS
4.9.2Configuring the RS232 interface
For an error-free data transmission, the RS232 interface should be set
to the same transmission speed (Baud rate) on the pHotoFlex (Turb)
and PC/printer.
You can set the following values for the baud rate on the
pHotoFlex (Turb) : 1200, 2400, 4800, 9600, 19200.
The baud rate is selected in the menu, Configuration / System / Inter-face / Baud rate.
4.9.3Selecting the output format of datasets
Output format, ASCII
For downloading data to the interface you can select the output format.
It is selected in the menu, Configuration / System / Interface / Output format.
The ASCII output format delivers formatted datasets.
The CSV output format delivers datasets separated by ";".
pHotoFlex Ser. no. 12345678
31.10.04 09:56:20
ID: 1
10.01 (pH) AR
25 °C
________________________________________
The following table shows which data are transmitted to the interface in
which way:
DataOperation / description
Current
measured value
! Press <PRT>.
! Simultaneously with every manual storage
process.
Stored measured
values
! Display stored dataset and press <PRT>.
! All datasets according to the filter criteria via
the RS232 download function
(see section 4.8.2).
Calibration record
(pH electrode)
! Calibration record of a pH electrode with
<PRT> (after calling up from the storage or at
the end of a calibration).
! Calibration record of a pH electrode at the
end of a calibration procedure is printed automatically.
Note
With the <PRT> key you output the data that is being shown on the display to the interface (displayed measured values, stored measurement
datsets, calibration record).
ba75491d0407/2006
75
OperationpHotoFlex / pHotoFlex Turb
4.10Reset
You can reset (initialize) all system and measurement settings.
Note
For turbidity measurement, there are no resettable settings.
4.10.1 Resetting the system settings
With the System / Reset function, all resettable settings are reset.
! Settings for pH & ORP (see section 4.10.3)
! Settings for Photometry (see section 4.10.2)
! System settings
System settingDefault settings
Baud rate4800 Baud
Output formatASCII
IlluminationAuto off
Contrast50 %
Brightness50 %
Switchoff time30 min
BeepOn
Measuring modePhotometry
76
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.10.2 Resetting the photometer settings
With the Photometry / Reset function, all photometer settings are reset.
SettingDefault settings
PROG0
Cell size16 mm
Measured parameterKonzentration
Wavelength436 nm
Blank valuesall erased
4.10.3 Resetting the pH settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
The following settings for pH measurements are reset to the default
settings with the Reset function:
SettingDefault settings
Cal. type TEC
Calibration interval7 d
Measured parameterpH & ORP
Asymmetry (Asy)0 mV
Slope (Slp.)-59.16 mV/pH
Temperature, manual25 °C
Temperature°C
ba75491d0407/2006
77
OperationpHotoFlex / pHotoFlex Turb
4.11Meter information
The following meter information is listed in the Configuration / Info
menu:
! Model designation
! Software version
! Version number(s) of the stored program data
! Series number of the meter
Info
i Model: pHotoFlex
i Software: V 0.35
i Prog.
i Ser. no.: 04280003
: V 0.07
4.12Software update
With a software update you obtain the current software with all new programs and method data (see appendix).
A software update comprises
! new instrument software
! new programs (methods)
! revisions of existing methods
The current software version can be found on the Internet under
www.WTW.com.
The proceeding for updating the software can be found in the appendix
(see A
PPENDIX: FIRMWAREUPDATE).
78
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
4.13Administrating user-defined methods
User-defined programs can be
! entered
! read out
! erased.
To store user-defined programs (methods) in the meter, determine a
calibration line for your method yourself and transmit the data of this
calibration line to the pHotoFlex (Turb).
Administration of the method data can be carried out in the LSdata software (see operating manual LSdata) or with the aid of a terminal program (section 4.13.1).
4.13.1 Administrating user-defined programs with a terminal
program
Generally, a terminal program serves to establish a connection to a device on a data interface and to communicate with the device via a console on the display.
Thus, it is also possible to send command lines.
Terminal programs are available for different operating systems by different manufacturers. Windows (version 95 to XP) contains the "HyperTerminal" terminal program. It is in the program menu under
accessories.
For more detailed information please refer to the user information of the
terminal program.
Note
The CD-ROM contains a configuration file for the HyperTerminal program. By double-clicking the configuration file (*.ht), the HyperTerminal
with the required pre-settings for data exchange with the meter is
opened.
The pHotoFlex (Turb) can administrate the database of user-defined
programs via command lines.
Storing user-defined
programs
1With the aid of the AK 540/B interface cable, connect the
pHotoFlex (Turb) 540 to the serial interface (COM port) of the
PC (see section 8.1.1).
2Make sure the pHotoFlex (Turb) is switched on.
3Start the terminal program on the PC.
4Configure the connection settings for the COM interface as
necessary.
Enter a command line according to the following scheme in the terminal
program:
U.500#11,Program No,Model,Cuvette [mm],Wave length [nm],MRB,
MRE,E0,Slope,Resolution,Formula,Unit
The individual data sections of the command line are separated by
commas. The dot "." has to be used as a decimal separator within a
data section.
80
5Enter the command line.
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbOperation
6Finish the command line with Enter.
The data is transmitted to the pHotoFlex (Turb).
Note
If a program is already stored under the selected number, the programming procedure is canceled. To store the program under the selected
number, first erase the program stored under the number.
After successful transmission, the terminal program writes "!>".
If the transmission failed, the terminal program writes "!>".
Erasing user-defined
programs
To erase user-defined programs, enter a command line according to
the following scheme in the terminal program:
Erase all user-defined
programs
Erase one userdefined program
Command lineU.520U.521#1,Program No
ExampleU.520U.521#1,900
7Enter the command line.
8Finish the command line with Enter.
The data is transmitted to the pHotoFlex (Turb).
The requested data is displayed on the terminal as the result.
After successful transmission, the requested data is displayed on the
terminal.
If the transmission failed, the terminal program writes "!>".
Reading out user-
defined programs
ba75491d0407/2006
To read out user-defined programs, enter a command line according to
the following scheme in the terminal program:
Read out all userdefined programs
Read out one userdefined program
Command lineU.510U.511#1,Program No
ExampleU.510U.511#1,900
9Enter the command line.
81
OperationpHotoFlex / pHotoFlex Turb
10Finish the command line with Enter.
The data is transmitted to the pHotoFlex (Turb).
The requested data is displayed on the terminal as the result.
After successful transmission, the terminal program writes "!>".
If the transmission failed, the terminal program writes "!>".
The meter is almost maintenance-free.
The only maintenance task is replacing the batteries or accumulator
pack.
5.1.1Inserting/exchanging the batteries
Caution
Make sure that the poles of the batteries are the right way round.
The ± signs on the batteries must correspond to the ± signs in the
battery compartment.
1
2
1Open the battery compartment:
– Unscrew the two screws (1) on the underside of the meter,
– Remove the lid of the battery compartment (2).
ba75491d0407/2006
2If necessary, take four old batteries out of the battery compart-
ment.
3Insert four batteries (3) in the battery compartment.
4Close the battery compartment and fix it with the screws.
Occasionally wipe the outside of the meter with a damp, lint-free cloth.
Disinfect the housing with isopropanol as required.
Caution
The housing components are made out of synthetic materials
(polyurethane, ABS and PMMA). Thus, avoid contact with acetone
and similar detergents that contain solvents. Remove any splashes immediately.
5.2.1Cleaning the cell shaft
If liquid is in the cell shaft (e.g. due to a spilled cell), clean the cell shaft
as follows:
Caution
Cells can contain poisonous or corrosive substances. If the content is released follow the danger warnings on the cell. If necessary, take corresponding protective measures (protective
goggles, protective gloves etc.).
1Switch the pHotoFlex (Turb) off and pull out the power plug.
2Rinse the cell shaft with distilled water.
5.2.2Cleaning the cells
Cells have to be clean, dry, and free of fingerprints and scratches.
Therefore, clean them regularly:
1Clean the cells inside and out with hydrochloric acid or labora-
tory soap.
2Rinse out several times with distilled water.
3Let them dry in the air.
4Only hold the cells by the top or by the light protection cap so
that the optical path is not impaired.
5Before measuring, clean the cell with the enclosed cleaning
cloth.
Note
Scratches in the glass change the optical characteristics of the cell and
falsify the measured value. For this reason, never use scratched cells!
PackingThis meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the meter from transport damages.
Batteries/accumulator
pack
NiMH
MeterDispose of the meter without the batteries and accumulator pack as
Remove the batteries or accumulator pack from the meter (see
section 5.1).
Dispose of the batteries or accumulator pack at a suitable facility according to local legal requirements. It is illegal to dispose of the accumulator pack with household refuse.
electronic waste at an appropriate collection point.
86
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbWhat to do if...
6What to do if...
6.1General errors
Display, LoBat
Instrument does not
react to keystroke
RS232 interface
does not react
CauseRemedy
– The batteries or accumulator
pack are largely depleted
CauseRemedy
– Software error
– Operating condition undefined
or EMC load unallowed
CauseRemedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Insert new batteries
– Charge the accumulator pack
(see section 3.2)
– Processor reset:
Press the <START/ENTER>
and <PRT> key
simultaneously.
– Processor reset:
Press the <START/ENTER>
and <PRT> key
simultaneously.
Error message,
0, 8, 16, 16384
ba75491d0407/2006
CauseRemedy
Error
– Instrument error– Repeat measurement
– Meter defective,
send meter to WTW for repair
and quote the error number
87
What to do if...pHotoFlex / pHotoFlex Turb
6.2Photometry
Measuring range
undercut or exceeded
Obviously incorrect
measured values
CauseRemedy
– Program not suitable– Select program with suitable
measuring range
– Dilute the sample
CauseRemedy
– Measurement disturbed by
– Close the external light cover.
external light
– Cell not correctly inserted– Insert the cell so that it is
positioned on the bottom of
the cell shaft.
– Cell contaminated– Clean the cell
– Cell shaft contaminated– Clean the cell shaft
– Dilution set incorrectly– Set the dilution
– Selected program unsuitable– Select other program
– Zero measurement incorrect– Perform zero measurement
Measuring range
undercut or exceeded
– Blank value incorrect– Remeasure the blank value
– pH value not stable– Measure with air excluded if
necessary
– Temperature not stable– Temper if necessary
Electrode + test sample:
– Conductivity too low
– Use suitable electrode
(e.g. in ultrapure water)
– Temperature too high– Use suitable electrode
ba75491d0407/2006
– Organic liquids– Use suitable electrode
89
What to do if...pHotoFlex / pHotoFlex Turb
Obviously incorrect
measured values
Sensor symbol flashes
CauseRemedy
pH electrode:
– Not connected– Connect electrode
– Cable broken– Replace cable or electrode
– pH electrode unsuitable– Use suitable electrode
– Temperature difference
between buffer and test
– Adjust temperature of buffer
or sample solutions
sample too high
– Measurement procedure not
– Follow special procedure
suitable
CauseRemedy
– Calibration interval expired– Recalibrate the measuring
system
Error message
Measured values
obviously incorrect
Measured value display
< 0.01 FNU
6.4Turbidity
CauseRemedy
– Cell not correctly inserted– Lock cell into place
– Cell contaminated– Clean the cell
– Calibration too old– Carry out calibration
CauseRemedy
– Measured value outside the
measuring range
– not possible
90
ba75491d0407/2006
pHotoFlex / pHotoFlex TurbTechnical data
7Technical data
7.1General data
7.1.1pHotoFlex (Turb)
Dimensionsapprox. 236 x 86 x 117 mm
Weightapprox. 0.6 kg (without batteries)
Mechanical structureType of protectionIP 67
Electrical safetyProtective classIII
Test certificatescETLus, CE, FCC
Storage- 25 °C ... + 65 °C
Ambient conditions
Operation0 °C ... + 50 °C
Climatic class2
Allowable relative
humidity
Power
supply
Yearly mean:
30 days /year:
other days:
Batteries4 x 1.5 V, type AA
Operating time with
battery operation
Accumulator pack
(optional)
Power pack
Charging device
(optional)
75 %
95 %
85%
approx. 5000 measurements
5 x 1.2 V nickel metal hydride (NiMH),
type AA
FRIWO FW7555M/09, 15.1432.500-00
Friwo Part. No. 1883259
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1,5 A
Connection max. overvoltage category II
Primary plugs contained in the scope of delivery: Euro, US, UK and Australian.
ba75491d0407/2006
91
Technical datapHotoFlex / pHotoFlex Turb
Serial
interface
Guidelines
and norms used
Connection of the cable AK 540/B or AK 540/S
Baud rateadjustable:
1200, 2400, 4800, 9600, 19200 Baud
TypeRS232
Data bits8
Stop bits2
ParityNone
HandshakeRTS/CTS
Cable length max. 15 m
EMCEC guideline 89/336/EEC
EN 61326-1/A3:2003
FCC Class A
Instrument safetyEEC guideline 73/23/EEC
EN 61010-1 :2001
UL3101-1
CAN/CSA-C22.2 No. 1010.1-92
Climatic classVDI/VDE 3540
IP protectionEN 60529:1991
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is
operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to
correct the interference at his own expense.