WarrantyWe guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result
in differences between this operating manual and your instrument.
Also, we cannot guarantee that there are absolutely no errors in this
manual. Therefore, we are sure you will understand that we cannot
accept any legal claims resulting from the data, figures or descriptions.
purchase.
The instrument warranty covers manufacturing faults that are
discovered within the warranty period.
The warranty does not cover components that are replaced during
maintenance work, e. g. batteries.
The warranty claim extends to restoring the instrument to readiness for
use but not, however, to any further claim for damages. Improper
handling or unauthorized opening of the instrument invalidates any
warranty claim.
To ascertain the warranty liability, return the instrument and proof of
purchase together with the date of purchase freight paid or prepaid.
The compact Lab 860 precision pH meter enables you to perform pH
measurements rapidly and reliably. The Lab 860 provides the
maximum degree of operating comfort, reliability and measuring
certainty for all applications.
The proven calibration procedures and special stability control function
(SC) support your work with the pH meter.
1Keypad
2Display
3Socket field
Note
The meter is also available as part of individual Sets of equipment.
You will find information on this and other accessories in the SI
Analytics GmbH laboratory catalog or via the Internet.
Note
If you need further information or application notes, you can obtain the
following material from SI Analytics GmbH:
z Application reports
z pH primers
z Safety datasheets.
ba75554e0501/2010
7
OverviewLab 860
STO
RCL
MODESC
CAL
MODE
CAL
SC
STO
STO
RCL
1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) indicates a keystroke in this operating
manual.
<On/Off>:
<On/Off
<MODE>:
<MODE
<CAL>:
<CAL
<SC>:
<SC
<STO>:
<STO
<RCL>:
<RCL
<▲>: Increment values, scroll
_>:
_>:
_>:
_>:
_>:
_>:
Switch meter on / off
Reset calibration values
Select measured parameter
Open setting menu for calibration and
measurement
Call up calibration procedure
Display calibration data
Activate / deactivate stability control
Set interval for data transmission
Store measured value
Set interval for automatic storage in
memory
Display / transmit measured values and
calibration records
Erase stored measured values
8
<▼>: Decrement values, scroll
<OK>:
_>:
<OK
Confirm entries
Open setting menu for system settings
ba75554e0501/2010
Lab 860Overview
8
88
8
8
1
8
8
8
°
C
°F
U pH
%
LoBat
Auto
Store
SC
TP
CalError
Time
Day.Month
Year
Baud
No.
Ident
mV/pH
RCL
SET
BUFFER
AutoCal
Status display
indicator
CalClock (calibration
interval and sensor
evaluation)
Measured value
display
Temperature display
and further function
indicators
1.2Display
Function display
indicators
[SET]During calibration: number of the buffer set
[BUFFER]During calibration: number of the buffer
[AutoCal]Calibration with automatic buffer recognition
[Auto][Store]Automatic storing is switched on
[CalError]An error occurred during calibration
[LoBat]With battery operation: batteries almost empty
[SC]Stability control is active
[RCL]Memory display / memory download
[TP]Temperature measurement active
ba75554e0501/2010
9
OverviewLab 860
1
2
3
4
5
6
1.3Socket field
Connectors:
1pH electrode
2Reference electrode
3Temperature sensor
4USB interface
5Power pack
6RS 232 interface
CAUTION
Only connect sensors to the meter that cannot return any voltages
or currents that are not allowed (> SELV and > current circuit with
current limiting).
Almost all sensors - in particular SI Analytics GmbH sensors fulfill these conditions.
10
ba75554e0501/2010
Lab 860Safety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating
manual before working with the meter.
The operating manual must always be available within the vicinity of the
instrument.
Target groupThe meter was developed for work in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.
Safety instructionsSafety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the level of danger:
Further notes
WARNING
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.
CAUTION
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the instrument
or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.
ba75554e0501/2010
11
SafetyLab 860
2.1Authorized use
This meter is authorized exclusively for pH and ORP measurements in
the laboratory.
The technical specifications as given in chapter 7 T
ECHNICALDATA
(page 59) must be observed. Only the operation and running of the
meter according to the instructions given in this operating manual is
authorized.
Any other use is considered unauthorized.
2.2General safety instructions
This instrument is constructed and tested in compliance with the IEC
1010 safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during
operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICALDATA (page 59).
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty
functioning of the instrument. In this event, wait until the temperature of
the instrument reaches room temperature before putting the instrument
back into operation.
CAUTION
The meter is only allowed to be opened by personnel authorized
by SI Analytics GmbH.
12
ba75554e0501/2010
Lab 860Safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
of service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
ba75554e0501/2010
13
SafetyLab 860
14
ba75554e0501/2010
Lab 860Commissioning
3Commissioning
3.1Scope of delivery
z Lab 860 laboratory meter
z Power pack
z 4 batteries 1.5 V Mignon type AA
z USB cable with A plug and B plug
z Transparent cover
z Operating manual
z CD-ROM with USB driver
3.2Initial commissioning
Perform the following activities:
Insert batteries
z Insert batteries
z Switch on the meter
z Set the date and time
z Connect the power pack (for line power operation only).
1Open the battery compartment (1) on the underside of the
meter.
2Place four batteries (type Mignon AA) in the battery
compartment.
3Close the battery compartment (1).
The date (day) flashes in the display.
4Set the date and time according to page 44.
ba75554e0501/2010
15
CommissioningLab 860
1
CAUTION
Make sure that the poles of the batteries are the right way round.
The
±signs on the batteries must correspond to the ± signs in the
battery compartment.
Only use leakproof alkaline manganese batteries.
Switching on the meter
Setting the date and
time
Connecting the power
pack
1Switch on the meter with <On/Off>.
A display test is briefly displayed.
2See page 44.
You can either operate the measuring instrument with batteries or with
the plug-in power supply. The plug-in power supply supplies the
measuring instrument with low voltage (12 VDC). This saves the
batteries.
CAUTION
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see page 59).
CAUTION
Use original power packs only (see page 59).
3Insert the plug into the socket of the pH meter.
4Connect the original power pack to an easily accessible power
outlet.
Note
You can carry out measurements without the power pack.
16
ba75554e0501/2010
Lab 860Operation
4Operation
4.1Switching on the meter
1Place the meter on a flat surface and protect it from intense
light and heat.
2Press the <On/Off>
A display test is briefly displayed.
Subsequently, the meter switches to the measuring mode
(measured value display).
Note
The meter has an energy saving feature to avoid unnecessary battery
depletion during battery operation.
The energy saving feature switches off the meter if no key was pressed
during the specified interval (setting the switch-off interval, see
page 44).
The energy saving feature is not active:
z if the meter is supplied via the power pack or the USB interface
z if a PC is connected (with communication cable to RS232 interface)
z if the printer cable is connected (for external printers).
key.
ba75554e0501/2010
17
OperationLab 860
4.2General operating principles
This section contains basic information on the operation of the Lab 860.
4.2.1Operating modes
The instrument has the following operating modes:
z Measurement
The display indicates the measurement data in the measured value
display
z Calibration
The display guides you thru a calibration procedure with calibration
information
z Storage
The meter stores measuring data automatically or manually
z Transmitting data
The meter transmits measuring data and calibration records to a
serial interface automatically or manually.
z Configuration
The system menu or a sensor menu with submenus, settings and
functions is displayed
4.2.2Operation
KeysThe meter is operated via keys. The keys can have different functions
with long or short keystrokes.
FunctionsGenerally, with a short keystroke a function is carried out. A long
keystroke opens a setting menu.
In a setting menu, settings are selected with the <▲><▼> keys.
A setting is confirmed with <OK>. With confirming, the setting is
finished and the next setting is displayed.
RepresentationIn this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<OK
_>).
18
ba75554e0501/2010
Lab 860Operation
4.3Measuring
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect an electrode to the meter.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3Calibrate or check the meter with the electrode.
4Select the measured parameter with <MODE>.
Note
Incorrect calibration of pH electrodes leads to incorrect measured
values. Calibrate regularly before measuring.
CAUTION
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result.
The RS232 and USB interfaces are not galvanically isolated.
Temperature sensorYou can measure with or without a temperature sensor. If a
temperature sensor is connected, it is indicated on the display by TP.
Note
The pH meter automatically recognizes the type of the temperature
sensor used. Therefore, you can connect electrodes with an NTC30 or
Pt1000.
The temperature measurement is absolutely essential for a
reproducible pH measurement. If the measurement is made without a
temperature sensor, proceed as follows:
1Measure the current temperature using a thermometer.
2Set the temperature value with <▲><▼>.
Note
When calibrating without temperature sensor, also set the current
temperature of each buffer solution manually (see <▲><▼>).
ba75554e0501/2010
19
OperationLab 860
8
4
2
2
9
9
6
pH
°
C
TP
4.3.1Measuring the pH value
1Perform the preparatory activities according to page 19.
2Immerse the pH electrode in the test sample.
3Using <MODE>, scroll as necessary until the measured
parameter pH is displayed.
Stability control SC
(drift control)
The stability control function (drift control) checks the stability of the
measurement signal. The stability has a considerable effect on the
reproducibility of the measured value.
For identical measurement conditions, the following criteria apply:
Drift within 15 sec < 0.02 pH units.
1If necessary, call up the measured variable pH with <MODE>.
2With <SC>, activate the stability control function.
The SC function display indicator appears.
The current measured value is frozen (hold function).
3Start measurement with stability control with <OK>.
SC flashes until a stable measured value is reached.
This measured value is downloaded to the interface.
4If necessary, start the next measurement with stability control
with <OK>.
5To terminate the stability control function: Press the <MODE>
or <SC>
key.
20
Note
The current measurement with stability control can be terminated at
any time (accepting the current value) by pressing <OK>.
ba75554e0501/2010
Lab 860Operation
3
6
1
8
2
4
U
°
C
TP
mV
4.3.2Measuring the ORP
The meter can, in conjunction with an ORP electrode, measure the
ORP (mV) of a solution.
1Perform the preparatory activities according to page 19.
2Immerse the ORP electrode in the test sample.
3If necessary, call up the measured parameter U with <MODE>.
4Wait for a stable measured value.
Note
ORP electrodes are not calibrated. However, you can check ORP
electrodes using a test solution.
ba75554e0501/2010
21
OperationLab 860
4.4Calibration
Why calibrate?pH electrodes age. This changes the zero point (asymmetry) and slope
of the pH electrode. As a result, an inexact measured value is
displayed. Calibration determines the current values of the zero point
and slope of the electrode and stores them. Thus, you should calibrate
at regular intervals.
The calibration data is stored in the meter.
When do you have to
calibrate?
Buffer sets for
calibration
z After connecting another electrode
z If the CalClock has expired and flashes
You can use the buffer sets quoted in the table for an automatic
calibration. The pH values are valid for the specified temperature
values. The temperature dependence of the pH values is taken into
account during calibration.
No.Buffer set*pH valuesat
1SI Analytics GmbH DIN buffers
according to DIN 19266/NBS
1.679
4.006
25 °C
6.865
9.180
12.454
2SI Analytics GmbH Technical
buffers according to DIN 19267
2.000
4.010
25 °C
7.000
10.011
3Merck1*4.000
20°C
7.000
9.000
22
4Merck2 *1.000
6.000
8.000
13.000
5Merck3 *4.660
6.880
9.220
6DIN 19267 *1.090
4.650
6.790
9.230
7Mettler Toledo USA *1.679
4.003
7.002
10.013
ba75554e0501/2010
20°C
20°C
25 °C
25 °C
Lab 860Operation
No.Buffer set*pH valuesat
8Mettler Toledo TEC *1.995
4.005
7.002
9.208
9Fisher *2.007
4.002
7.004
10.002
10Fluka BS *4.006
6.984
8.957
11Radiometer *1.678
4.005
7.000
9.180
12Baker *4.006
6.991
10.008
13Metrohm *3.996
7.003
8.999
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
14Beckman *4.005
25 °C
7.005
10.013
15Hamilton Duracal *4.005
25 °C
7.002
10.013
16Precisa *3.996
25 °C
7.003
8.999
*Brand names or trade names are trademarks of their respective
owners protected by law (see page 67).
ba75554e0501/2010
23
Loading...
+ 51 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.