Xylem ISE Electrodes User Manual

Instruction manual for the
y
r
SCHOTT
®
Instruments electrodes for measuring the Na
+
concentration
®
SI Analytics GmbH Hattenbergstrasse 10 55122 Mainz Germany Telefon +49 (0) 6131/66 5119 Telefax +49 (0) 6131/66 5001 e-mail: support@si-analytics.com
www.si-anal
electrode plug
refilling hole
diaphragm senso
tics.com
Preparation and general
If there is an irrigation cap over the sensor and diaphragm, this should be removed. It contains irrigation fluid (sodium chlorid solution 0.1 mol/l). The electrode is ready to perfor m a measurement. Electrodes stored dry must be irrigated for 24 hours in irrigation fluid. Lacking electrolyt solution 3 M potassium chlorid solution (L 3008) should be refilled. The filling level of the electrode solution should always be at least 5 cm above the level of the measurement solution.
Attention!
Electrolyte solution and irrigation fluid are different. The closure of the refilling hole must be opened during
measurement. The membrane and diaphragm must be immersed in the measurement solution. Please also note the instruction manual of the measuring device.
For calibration following standard solutions are needed:
0.1 M NaCl
Dissolve 5.85 g sodium chloride in approximately 800 ml distilled water and fill it up to 1 l.
0.001 M NaCl
Dilute 10 ml of the 0.1 M NaCl solution up to 1 l with distilled water.
Calibration
The pNa electrode should be connected to the pH-meter lik e a pH electrode. Switch the pH-meter in the mV range. For calibration begin with the lowest concentration and without rinsing the next solution is measured. The resulting mV value is noted and a calibration curve printed. The pNa electrode follows the Nernst equation. The temperature of the solution must be taken into account.
Example calibration curve
mV
mV1
mV2
Measurement
The measured solution should be slightly alcalic, b ut at least have a pH of 6. Is this not the case, the pH can be adjusted with sodium free bases like ammonia water or triethanolamine.
Attention!
Do not take sodium hydroxide solution for adjusting the pH. The response of the pNa electrode is slow. For a stable
signal it takes from 10-40 min. A shorter waiting time can be reached when the electrode is rinsed before measuring with the new solution or an alcalic buffer solution is added.
Cleaning
Dirt and contamination on the sensor and diaphragm le ad to measurement inaccuracies. They can be removed with the following procedure:
Coatings and deposits can be removed with diluted mineral acids (e.g. diss. hydrochloric acid)
Organic contaminations with the aid of suitable solvents
Grease with tensid solution or alcohol
Proteins with hydrochloric pepsine solution (cleaning
solution L 510).
During maintenance there is to pay attention to:
After cleaning, rinse off the electrodes with distilled water, do not rub dry.
Platinum diaphragms must not be treated by any mechanical treatment. Rinsing (e.g. vacuuming) may be followed by chemical cleaning (e.g. with diluted hydrochloric acid).
The glass membrane can be clea ned by r ubbing off with a fluff-free cloth soaked in ethanol.
log a
P 1
P 2
1
log a2
log a
Storage and maintenance
Electrodes should be stored at a temperature between 0°C and 40°C. Depending on the storage conditions (temperature and air humidity) the irrigation fluid in the cap may dry out prematurely. pNa electrodes stored dr y must be irrigated for minimum 24 hour in sodium chlorid so lution
0.1 mol/l before measurement. Crystals in the electrolyte area can be dissolved by warming the electrode in a water bath. The electrolyte solution should then be replaced.
The correct potassium chlorid concentration must be observed.
Quality
Every electrode must fulfill the strict quality requireme nts of final testing. The working life depends mainly on the us age conditions. Extreme measuring conditions as for example high or frequently fluctuating temperatures, strong acids and caustic solutions, protein and heavily contaminated solutions, sulfides, bromides and iodides shorten the electrode’s working life. Hydrofluoric acid, sodium hydroxide and hot phosphoric acid attack glass.
Technical details Na 61
pNa range: 0 ... 6 Temperature range: -10 ... +80°C Length: 170 mm Sensor: Na sensitive glass Diaphragm: platinum Reference systeme: Silamid® Material: glass, d = 12 mm Zero point: pNa = 2.0 Electrolyt: 3 M KCl Irrigation fluid: 0.1 M NaCl Connection cable: e.g. L 1 A Temperature sensor: no
Order number
Combination electrode Na 61 285100026 Connection cable L1A 285122456 Electrolyte 3M KCl L 3008 285138505 Cleaning solution L 510 285138538
Further information
Further information can be found in the labratory catalo gue of SI Analytics GmbH.
Subject to technical amendment.
GA_pNa_us_T814047_090603
Loading...