The technical documentation supplied is also part of the
product. You should always keep this documentation
within reach.
These Operating Instructions describe your treatment
center with maximum possible equipment. They are valid
for both the C5
For care of your treatment center, please observe the
instructions "Care and Maintenance by the Practice Team".
To safeguard your warranty claims, please complete the
attached “Installation Report / Warranty Passport”
together with the service engineer immediately after the
installation of your treatment center.
Separate Operating Instructions and the corresponding
instructions for care have been attached to all dental
instruments, the SIROLUX FANTASTIC, the
SIROCAM 3, the tray and the X-ray image viewer.
Prior to start-up, you must read all Operating Instructions
to familiarize yourself with the treatment center.
To prevent any personal injury or material damage,
pay special attention to any notes printed in bold type or
marked in one of the ways indicated below:
+
and the C5+ Turn system versions.
+
, C5+ Turn dental
i
NOTE
ATTENTION
WARNING
Additional information, explanation or supplement
The information provided under this keyword directly
concerns the functioning of the product and/or warns
against possible operating errors.
This warning symbol warns against possible hazards for
the product or the user.
In case you get stuck despite having thoroughly studied
the Operating Instructions, please contact your dental
depot.
+
Your C5
Team
59 57 373 D 3415
2D 3415.201.01.09.02
Sirona Dental Systems GmbH Table of Contents
+
, C5+ Tur n
C5
Table of Contents
1Warning and safety information .............................................................................. 7
14.3Seat cushion C ................................................................................................................................. 99
Sirona Dental Systems GmbH1 Warning and safety information
Operating Instructions C5
+
, C5+ Tur n
1 Warning and safety information
Intended useThis dental treatment center is intended for diagnosis,
therapy and dental treatment of humans by properly
trained personnel.
This system is not intended for operation in areas sub-
ject to explosion hazards.
On-site installationThe on-site installation must have been performed
according to our requirements. The details are described
in the document “Installation Requirements”.
Maintenance and repairAs manufacturers of dental medical equipment and in
the interest of the operational safety of your system, we
stress the importance of having maintenance and repair of your treatment center performed only by ourselves or by agencies expressly authorized by us. Furthermore components influencing the safety of the system should always be replaced with original spare parts upon failure.
When having such work done, we suggest that you
request a certificate stating the type and extent of work
performed, including information about any modifications of the rated parameters or of the operating ranges
(if applicable), as well as the date, name of organization
and signature.
Modifications of the systemModifications of this system which could impair the
safety of the system owner, patients or other persons
are prohibited by law!
For reasons of product safety, only original Sirona
accessories approved for this product, or accessories
from third parties approved by Sirona, may be used. The
user is responsible for dangers resulting from the use of
non-approved accessories.
If any equipment not approved by Sirona is connected, it
must comply with the applicable standards:
IEC 60950 for information technology equipment (e.g.
PCs), and IEC 60601-1 for medical electrical equipment.
Combination with other equipmentAny person who assembles or modifies a medical elec-
trical system complying with the standard
IEC 60601-1-1 (safety requirements for medical electrical equipment) by combining it with other equipment
(e.g., by connecting it with a PC) is responsible for
ensuring that the requirements of this regulation are met
to their full extent for the safety of the patients, operators
and environment.
In case of doubt, contact the manufacturer of the system
components.
59 57 373 D 3415
D 3415.201.01.09.02
7
1 Warning and safety information Sirona Dental Systems GmbH
Operating Instructions C5
+
, C5+ Tur n
Electromagnetic compatibility (EMC)The C5
IEC 60601-1-2:2001.
Medical electrical equipment is subject to special precautionary measures with regard to EMC. It must be
installed and operated in compliance with the specifications provided in the document “Installation Requirements”.
Portable and mobile RF communication systems may
interfere with medical electrical equipment. For this reason, the use of mobile radio telephones in practice or
hospital environments must be prohibited.
With video systems:
Electromagnetic disturbances in the environment of the
+
C5
, C5+ Turn may result in reduced image quality. In
such cases, it is advisable to repeat the exposure before
saving the image.
In case of line voltage drops, image data stored in the
video camera may be lost. Place down the camera and
repeat the exposure.
Image artifacts may possibly occur as a result of radiofrequency interference caused by the transmitters of e.g.
radio services or radio amateurs.
+
, C5+ Turn complies with the requirements of
Influence on cardiac pacemakersThe workstation includes strong magnets which can
affect cardiac pacemakers of patients, users and technical personnel.
Therefore, please pay attention to the safety information
on pages 29 and 52.
Treatment of highly immunosuppressed patientsHighly immunosuppressed patients should not come in
contact with water from the treatment center. The use of
sterile solutions is recommended.
Quality of water / air supplyAir and water supply must meet the requirements spec-
ified in the installation instructions. Use only clean
water.
Water quality complianceTo ensure compliance with the medical and national
legal requirements for water from treatment centers,
Sirona recommends equipping the treatment center with
a disinfection system. As owner of the treatment center,
you are responsible for the water quality and may have
to take alternative measures to ensure its compliance if
you operate the treatment center without a disinfection
system.
Please contact your specialized dealer or your relevant
dental association for the respective national requirements and measures.
Highlighting of warning and safety informationTo avoid personal injury and material damage, you must
also observe the warning and safety information provided in the present operating instructions. It is highlighted by the captionsNOTICE, CAUTION or
WARNING.
i
59 57 373 D 3415
8D 3415.201.01.09.02
Sirona Dental Systems GmbH1 Warning and safety information
Operating Instructions C5
+
, C5+ Tur n
Symbols usedObserve accompanying documents (on name
plate of chair)
Ventilation slotsUnder no circumstances may the ventilation slots on the
unit be covered, since otherwise the air circulation will be
obstructed.
Do not spray with disinfectants or the like into ventilation
slots.
Vacuum systemThe suction removal of aluminum and other metal oxides
from blasting devices via the automatic separator and
the amalgam separator integrated in the treatment center is prohibited!
This would cause extreme wear and clogging of the suction and water paths.
A separate vacuum system must be used in connection
with metal oxide blasting devices.
Treatment centers equipped with a central wet suction
system are generally suitable for suction removal of the
above material. However, make sure to observe the
instructions provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in
connection with Sirona treatment centers. However, in
such cases, make sure that the system is subsequently
flushed with an adequate amount of water.
Patient chairPlease observe the maximum load capacity of the chair
of 135kg according to EN ISO 6875 (tested with a
four-fold safety factor according to IEC 60601-1)
Electric micromotors in dentist elementIntermittent operation: 1 min. ON – 7 min. OFF with a
motor current of 1.5A.
Maintenance of the treatment centerDespite the outstanding quality of your treatment center
and regular care by the practice team, it is required in the
interest of operational safety that preventive maintenance be performed at predetermined intervals.
In order to guarantee the operational safety and reliability of your treatment center and to avoid damage
due to natural wear, you as the system owner must
have your system checked through regularly by an
authorized service engineer from your dental depot.
Furthermore, safety checks must be performed.
Please contact your dental depot to obtain a maintenance offer.
For details please refer to chapter 15 “Maintenance”
on page 105.
Dismantling and reassemblyWhen dismantling and reassembling the system, pro-
ceed according to the installation instructions for new
installation in order to guarantee its functioning and stability.
59 57 373 D 3415
D 3415.201.01.09.02
9
1 Warning and safety information Sirona Dental Systems GmbH
Operating Instructions C5
+
, C5+ Tur n
DisposalPlease observe the disposal regulations applicable in
your country.
Within the European Economic Area, this product is subject to Directive 2002/96/EC as well as the corresponding national laws. This directive requires environmentally
sound recycling/disposal of the product.
The product must not be disposed of as domestic refuse!
Please contact your dealer if final disposal of your product is required.
10D 3415.201.01.09.02
59 57 373 D 3415
Sirona Dental Systems GmbH2 Technical desc ription
Operating Instructions C5
+
, C5+ Tur n
2 Technical description
Model designationC5
+
, C5+ Turn
Power supply230VAC 50 Hz,
115VAC 50/60 Hz
100VAC 50/60 Hz
Nominal current4.5A at 230V
9.5A at 115V
11.5A at 100 V
Main unit fusefor 230VAC: T 6.3 A H , 250VAC, REF 10 77 452
Additional instrument holder (only for SIROCAM 3 /
SIROCAM C)
23
Foot pedal for swiveling the patient chair
13
3 Controls and functional elements Sirona Dental Systems GmbH
3.3 Control panel on the dentist elementOperating Instructions C5
+
, C5+ Tur n
3.3Control panel on the dentist element
Control panel
AMALG
DESINFAB
RPM
System status indication display and Mode
key
AMALG appears when the amalgam rotor needs to be
replaced.
DESINF starts flashing when DENTOSEPT P must be
refilled.
A appears when user A is selected.
B appears when user B is selected.
Mode key with + / – keys for programming basic set-
tings.
In addition to the settings in the Mode dialog, the following direct settings can be made with the
•Intensity of the instrument lights
•Intensity of the operation light
•Tumbler filling time
•Cuspidor flushing time
•Chair conversion position (C5
+ / – keys:
+
Turn only)
14D 3415.201.01.09.02
59 57 373 D 3415
Sirona Dental Systems GmbH3 Controls and functional elements
Operating Instructions C5
+
, C5+ Turn3.3 Control panel on the dentist element
Main functions – dark blue keys
Quick setting keys for instrument intensity
40
100
and speed
for electric motors and SIROSONIC L ultrasonic scaler
A
INSTR. LIGHT
20
0.2
rpm x 1000
%
50
1
Spray ON / OFF
on burr drives
Chip blower
on burr drives
Counterclockwise rotation
for electric motor
Instrument light ON / OFF
Pressing this key for a while displays the light intensity
settings dialog.
59 57 373 D 3415
D 3415.201.01.09.02
11
00
22
SS
Mouth rinsing position chair program.
with last position memory function (freely programmable)
Chair program 0
Entry/exit position (freely programmable)
Chair programs 1 and 2
(freely programmable)
i
NOTICE
The four program keys are also used for reprogramming
the chair programs.
15
3 Controls and functional elements Sirona Dental Systems GmbH
3.3 Control panel on the dentist elementOperating Instructions C5
+
, C5+ Tur n
Additional functions – light blue keys
Preselection of user A or B
A/B
All instruments must be in place.
You cannot switch between the two users after picking
up an instrument
.
Saving the instrument settings
The instrument to be programmed must have been
picked up.
X-ray image viewer ON / OFF.
Composite key ON / OFF
for SIROLUX operation light, reduced light intensity
8,000 lux for composite fillings
<
A
OPERATING LIGHT
A
CUP FILL TIMER
A
BOWL FLUSH TIMER
SIROLUX operation light ON / OFF
%
for normal light intensity > 24,000 lux
Pressing this key for a while displays the light intensity
settings dialog:
OPERATING LIGHT
Tumbler filling
Pressing this key for a while displays the settings dialog:
CUP FILL TIMER
Cuspidor flushing
Pressing this key for a while displays the settings dialog:
BOWL FLUSH TIMER
Freely selectable function
e.g. call key, always functions as button
freely available relay 230V, 6 A
(connected by the service engineer).
Freely selectable function
The function can be preset in the Mode dialog as button
or switch.
freely available relay 230V, 6 A
(connected by the service engineer).
Support arm brake
Unlocks the pneumatic support arm brake of the dentist
element while being pressed.
59 57 373 D 3415
16D 3415.201.01.09.02
Sirona Dental Systems GmbH3 Controls and functional elements
Operating Instructions C5
+
, C5+ Turn3.3 Control panel on the dentist element
Chair functions – gray keys
for manual chair settings:
Tilt backrest backward or forward.
Height adjustment down / up
Additional functions of the control
panel video keys
i
NOTICE
Apart from the previously described functions, the keys
of the dentist element have additional functions in connection with the video application. These are described
in chapter 18 starting on page 53.
59 57 373 D 3415
D 3415.201.01.09.02
17
3 Controls and functional elements Sirona Dental Systems GmbH
3.4 Control panel on the assistant elementOperating Instructions C5
+
, C5+ Tur n
3.4Control panel on the assistant element
Control panel
11
00
22
SS
Main functions – dark blue keys
Mouth rinsing position chair program.
with memory function (freely programmable)
Chair program 0
Entry/exit position (freely programmable)
Chair programs 1 and 2
(freely programmable)
i
NOTICE
The four program keys are also used for reprogramming
the chair programs.
18D 3415.201.01.09.02
59 57 373 D 3415
Sirona Dental Systems GmbH3 Controls and functional elements
Operating Instructions C5
+
, C5+ Turn3.4 Control panel on the assistant element
Additional functions – light blue keys
Tumbler filling
ON / OFF
Cuspidor flushing
ON / OFF
X-ray image viewer ON / OFF
for SIVISION 3 also WHITE SCREEN activation
Treatment center sanitation
SAN
L
This key can be used to start the treatment center sanitation program (see instructions “Care and Cleaning
by the Practice Team”).
Freely selectable function
The function can be preset in the Mode dialog as button
or switch.
Light ON/OFF
With SPRAYVIT picked up:
•Instrument light
With SPRAYVIT in holder:
•Operation light
59 57 373 D 3415
D 3415.201.01.09.02
19
4 Putting the system into operation Sirona Dental Systems GmbH
Operating Instructions C5
+
, C5+ Tur n
4 Putting the system into operation
Initial startup
The disinfection system adds a disinfectant to the water
(1:100) to prevent the formation of microorganisms in
the water system.
Prior to initial startup of your treatment center, sanita-tion must be performed.
If the service engineer has skipped the sanitation procedure after installation of your treatment center on the
basis of an agreement with you, please perform sanitation yourself as described in the separate instructions
“Care and Cleaning by the Practice Team”.
Sanitation takes approx. 24 hours.
Main switch
Switch ON the main switch ( l ).
A
READY
10:04:27
ON
OFF
During power-up, the treatment center automatically
performs a self-test.
CAUTION
For safety reasons, always switch the treatment center
OFF (O) after consulting hours. This cuts off the air and
water supply as well as the line voltage.
Ready for operation
After the end of the self-test, the message READY the
time and the last user selected (A or B) appear on the
display.
After the treatment center is switched on, the cursor is
between the chair program keys 1 and 2. If a program
LED is illuminated, then the chair was in a programmed
position before switching off. If no program LED is illuminated, then the chair was in a non-programmed position
before switching off.
The unit is now ready for operation.
20D 3415.201.01.09.02
59 57 373 D 3415
Sirona Dental Systems GmbH4 Putting the system into operation
A
MAINTENANCE
IN 30 DAYS
Operating Instructions C5
+
, C5+ Tur n
Display of next maintenance date
the next maintenance date is less than 30 days away,
this message appears each time the unit is switched on.
MAINTENANCE
IN 30 DAYS
(see MAINTENANCE Mode dialog).
After this date has expired, the following message is displayed:
MAINTENANCE
REQUIRED
The display message then disappears as soon as an
instrument is removed from its holder.
i
NOTICE
This message can be reset only by the service technician
after completion of maintenance work.
CAUTION
right
wrong
The dental chair has a maximum load capacity of 135kg
according to EN ISO 6875 (tested with a four-fold safety
factor according to IEC 60601-1).
The patient’s arms and legs must rest on the upholstery
of the chair!
Height adjustment of the dentist element
In order to adjust the height of the dentist element, you
1
1
must press one of the two keys (1) to release the brake.
1
1
Preselecting the user
A/B
The treatment center offers the possibility of managing
two different chair and instrument programs for two
users.
59 57 373 D 3415
D 3415.201.01.09.02
21
4 Putting the system into operation Sirona Dental Systems GmbH
Operating Instructions C5
+
, C5+ Tur n
Preselect user A or B using the A/B key, with all instruments in place.
A
B
The preselected user is shown in the top left corner of
the display. All settings which have been stored for that
user are then activated.
READY
10:04:27
22D 3415.201.01.09.02
59 57 373 D 3415
Sirona Dental Systems GmbH5 Foot switch
Operating Instructions C5
+
, C5+ Tur n5.1 C+ foot switch
5 Foot switch
5.1C
+
foot switch
1.
B
C
2.
A
D
This foot switch enables hands-free operation by means
of the cursor control, but also manual operation by
means of the keys.
The foot switch has a 4-way foot control plate for cursor
navigation, which works independently of the pedal.
Functions
Step on foot pedal – instrument picked up
Activation of the instrument.
If appropriate, intensity control relative to pedal
movement (if “speed controller foot switch” has been
preset under Mode).
1. Shift 4-way foot control plate
With cursor control enabled
(Mode dialog 10.8):
A to the front – cursor upward
B to the back – cursor downward
C to the left – cursor to the left
D to the right – cursor to the right
Cursor navigation along the marked path and release of the corresponding key function.
3.
4.
2. Actuate left button – all instruments in place
Programmed movement of the chair to the mouth
rinsing position S or the last treatment position (according to starting situation).
Actuate left button – instrument picked up
Toggling between spray ON and spray OFF, or
SIVISION function.
3. Actuate right button – all instruments in place
Programmed movement of the chair to the entry/exit
position 0.
Actuate right button – instrument picked up
Chip blower active for duration of actuation, or SIVISION function.
59 57 373 D 3415
D 3415.201.01.09.02
23
5 Foot switch Sirona Dental Systems GmbH
5.2 C foot switchOperating Instructions C5
Automatic cursor control with C+ foot switch
2.1. / 3.
A yellow diode on the dentist element indicates the position of the cursor. The cursor path is marked by lines.
1. After you switch on the system, or when the chair
has reached the entry/exit position 0, the cursor is
between the chair program keys 1 and 2
2. When you pick up an instrument or activate any of
the instruments, the cursor jumps to the horizontal
navigation path (position between 1st and 2nd row
of keys) in the instrument block.
3. After you have placed the instrument in its holder,
the cursor jumps back to its position between the
chair program keys 1 and 2.
+
, C5+ Tur n
5.2C foot switch
B*
C
A*
D
This foot switch enables hands-free operation by controlling all functions with the entire pedal surface. However, it is also possible to control the unit manually with
the keys.
For cursor navigation, place your foot fully on the pedal
surface.
Functions (factory settings)
i
NOTICE
The foot switch functions A and B can be reversed by the
service engineer upon request.
With cursor control enabled (Mode dialog 10.8)
Actuate foot switch in direction A – cursor upward.
Actuate foot switch in direction B – cursor downward.
Swing foot switch to the left C – cursor to the left.
Swing foot switch to the right D – cursor to the right.
Cursor navigation along the marked path and release of
the corresponding key function.
24D 3415.201.01.09.02
59 57 373 D 3415
Sirona Dental Systems GmbH5 Foot switch
Operating Instructions C5
+
, C5+ Tur n5.2 C foot switch
Instrument start
The cursor must be between the spray and the chip
blower key.
Actuate the foot switch in direction A after picking up the
instrument.
With cursor control disabled (Mode dialog
10.8)
Actuate foot switch in direction A:
With instrument picked up – Activation of the instru-
ment.
Actuate foot switch in direction B:
With instrument picked up – No function
Swing foot switch to the left C:
With instrument picked up – Spray ON/OFF for burr
instruments or SIVISION function.
With instruments in place – SIVISION function.
Swing foot switch to the right D:
2.
1.
3.
With instrument picked up – Chip blower ON for burr
instruments or SIVISION function.
With instruments in place – SIVISION function.
Automatic cursor control with C foot switch
A yellow diode on the dentist element indicates the position of the cursor. The cursor path is marked by lines.
1. After you switch on the system, or when the chair
has reached the entry/exit position 0, the cursor is
between the chair program keys 1 and 2.
2. When you pick up an instrument or activate any of
the instruments, the cursor jumps to a position between the spray and chip blower keys in the first instrument block.
If the cursor is in the quick setting key column at the
very left, it also jumps to a position between the
spray and chip blower keys in the instrument block
when you swing the foot switch to the right (D).
3. After you have placed the instrument in its holder,
the cursor jumps back to its position between the
chair program keys S and 0.
59 57 373 D 3415
D 3415.201.01.09.02
25
6 Program selection Sirona Dental Systems GmbH
6.1 SafetyOperating Instructions C5
+
, C5+ Tur n
6 Program selection
6.1Safety
CAUTION
The patient’s arms and legs must be resting on the chair
upholstery during the program run as well as during manual chair movement!
CAUTION
Make sure that no obstacles (e.g. window wings, drawers, devices, …) extend into the movement range.
Safety stop
A built-in safety circuit stops the chair movement in the
following situations:
•The toeboard collides with an obstacle.
•The backrest collides with an obstacle.
•The motorized headrest (if installed) collides with an
obstacle
•The swivelable cuspidor is swiveled in during chair
movement (C5
•The assistant element is pushed upward or downward by an obstacle.
•The cuspidor is pushed upward or downward by an
obstacle (C5+ only).
At the same time, a double beep signal sounds.
The chair then moves automatically a little upward until
the obstacle is cleared (except for the swivelable cuspidor, assistant element, cuspidor).
+
only).
CAUTION
Safety switches on the C5+
Risk of collision!
When moving the chair, make sure that the patient or
armrest does not come into contact with the water unit or
the assistant element.
59 57 373 D 3415
26D 3415.201.01.09.02
Sirona Dental Systems GmbH6 Program selection
Operating Instructions C5
+
, C5+ Tur n6.2 Program selection
With instruments in place, the last activated safety
A
BACKREST
10:04:27
switch is indicated in plain text on the display:
•TILTING PART for motorized headrest
•BACKREST
•TOEBOARD
•ASSISTANT ELEMENT (is also displayed when the
cuspidor safety switch is activated)
The message CUSPIDOR appears after 10 seconds if
the automatic return movement of the cuspidor to its
position is inhibited.
Position of the dentist element
CAUTION
When moving the chair (via program or manually), make
sure that the patient does not under any circumstances
come into contact with the support arm or the dentist element.
6.2Program selection
SS
11
59 57 373 D 3415
D 3415.201.01.09.02
00
22
Keys for selecting the programmed patient positions.
•Activate the desired chair program 1 or 2. The chair
moves to the programmed treatment position.
With swivelable cuspidor: Before the chair moves to
the selected program position, the swiveled-in cuspidor
automatically returns to its starting position.
Factory settings: The operation light switches OFF
automatically before the chair moves to the entry/exit
position 0 or the mouth rinsing position S.
The operation light switches ON automatically after the
chair has reached patient position 1 or 2.
Changing the factory settings: When programming
the patient positions you can preset whether the operation light should switch OFF or ON in the different chair
programs 0, S, 1 or 2 (see page 35).
27
6 Program selection Sirona Dental Systems GmbH
6.3 MOVEMENT STOP!Operating Instructions C5
Program key 0
Intended for the entry/exit position.
+
, C5+ Tur n
00
Program key S with memory function.
SS
In the S program, you can program the tumbler filling
and / or the cuspidor flushing functions (see Mode dia-
log 10.9 and 10.10).
When you press this key a second time, the treatment
center returns to its previous position (memory function).
Example: You have programmed this key for the mouth
rinsing position. When you press this key, the treatment
center moves to the programmed position.
Memory function: When you press this key a second
time, the treatment center returns to its previous position.
6.3MOVEMENT STOP!
You can stop the movement of the chair to a programmed position as follows:
•By pressing one of the chair-related keys.
•By actuating the 4-way foot control.
•By pushing the 4-way switches on a motorized
headrest
All movements of the treatment center are stopped
immediately.
Please note that when you press the program key of
the program which is just running a second time, the
movement will NOT stop!
28D 3415.201.01.09.02
59 57 373 D 3415
Sirona Dental Systems GmbH7 Patient positions
Operating Instructions C5
+
, C5+ Tur n7.1 Flat headrest
7 Patient positions
7.1Flat headrest
The headrest must be adjusted manually.
2
1
7.2Articulated headrest
1
2
To do this, take the load off the backrest and pull out or
push in the headrest using the handle (1).
The head pad (2) is held magnetically.
The headrest must be adjusted manually.
To do this, take the load off the backrest and pull out or
push in the headrest using the handle (1).
Tilt the headrest backward using the locking button (2).
59 57 373 D 3415
D 3415.201.01.09.02
Tilt the headrest forward by simply pulling it out.
WARNING
A strong magnet is integrated in the underside of the
head cushion (2) which could affect a nearby cardiac
pacemaker.
Therefore, please avoid bringing the magnet in close
proximity of a cardiac pacemaker of a patient, user
or technical personnel.
In addition, direct contact of the head cushion with
magnetic cards can lead to a data loss.
29
7 Patient positions Sirona Dental Systems GmbH
A
7.3 MultiMotion headrestOperating Instructions C5
+
, C5+ Tur n
7.3MultiMotion headrest
Adjustment to patient's height
The headrest can be adjusted to the patient's height by
pulling out or pushing in the headrest extension. This
adjustment is usually required at the beginning of treatment!
i
NOTICE
Before positioning the patient on the headrest, make
sure that the head support has been optimally adjusted
to the patient's height. This simplifies all subsequent
work with the MultiMotion considerably, since readjustment to the patient's height can thus be omitted when
changing over from mandibular to maxillary treatment.
The MultiMotion enables you to adjust the patient's head
in a way that optimally supports viewing of areas of the
mouth which are difficult to access.
Adjustment of hyperextension
Mandibular position
The mandibular position can be set by pulling the operating handle (A). The anatomical movement of the
arched extension keeps the patient's head in the support.
Pull the headrest out of the guide by pulling
handle A.
i
NOTICE
You can reduce the adjustment noise by pressing
release A.
30D 3415.201.01.09.02
59 57 373 D 3415
Loading...
+ 78 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.