The technical documentation supplied is also part of the
product. You should always keep this documentation
within reach.
These Operating Instructions describe your treatment
center with maximum possible equipment.
For daily care of your treatment center, please observe
the instructions provided in “Care and Maintenance by the Practice Team”.
To safeguard your warranty claims, please complete the
attached “Installation Report / Warranty Passport”
together with the service engineer immediately after the
installation of your treatment center.
Separate Operating Instructions with corresponding
instructions for care are enclosed with the dental
instruments, SIROLUX FANTASTIC, monitor, Cerec
Chairline, SIROCAM 3, tray and X-ray image viewer.
Prior to start-up, you must read all Operating Instructions
to familiarize yourself with the treatment center.
To prevent any personal injury or material damage,
pay special attention to any notes printed in bold type or
marked in one of the ways indicated below:
+
dental treatment
i
NOTE
CAUTION
WARNING
Additional information, explanation or supplement
The information provided under this keyword directly
concerns the functioning of the product and/or warns
against possible operating errors.
This warning symbol warns against possible hazards for
the product or the user.
In case you get stuck despite having thoroughly studied
the Operating Instructions, please contact your dental
depot.
Your C4+ Team
2D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH Table of Contents
bеЦдблЬ
Table of Contents
1Warning and safety information ................................................................................ 7
16.3Seat cushion C ................................................................................................................................. 101
Sirona Dental Systems GmbH1 Warning and safety information
Operating Instructions C4
+
1 Warning and safety information
Intended useThis dental treatment center is intended for diagnosis,
therapy and dental treatment of humans by properly
trained personnel.
This unit is not intended for operation in areas subject
to explosion hazards.
On-site installationThe on-site installation must have been performed
according to our requirements. The details are described
in the document “Installation Requirements”.
Maintenance and repairAs manufacturers of dental medical equipment and in
the interest of the operational safety of your system, we
stress the importance of having maintenance and repair of your treatment center performed only by ourselves or by agencies expressly authorized by us. Furthermore components influencing the safety of the unit
should always be replaced with original spare parts
upon failure.
When having such work done, we suggest that you
request a certificate stating the type and extent of work
performed, including statements concerning any modifications of the rated parameters or of the operating
ranges (if applicable), as well as the date, name of organization and signature.
Modifications of the systemModifications of this system which could impair the
safety of the system owner, patients or other persons
are prohibited by law!
For reasons of product safety, only original Sirona
accessories approved for this product, or accessories
from third parties approved by Sirona, may be used. The
user is responsible for dangers resulting from the use of
non-approved accessories.
If any equipment not approved by Sirona is connected, it
must comply with the applicable standards:
IEC 60950 for information technology equipment (e.g.
PCs), and IEC 60601-1 for medical electrical equipment.
Combination with other equipmentAny person who assembles or modifies a medical elec-
trical system complying with the standard IEC 60601-1-1
(safety requirements for medical electrical equipment)
by combining it with other equipment (e.g., by connecting it with a PC) is responsible for ensuring that the
requirements of this regulation are met to their full extent
for the safety of the patients, operators and environment.
In case of doubt, contact the manufacturer of the system
components.
59 06 313 D 3382
D 3382.201.01.13.02
7
1 Warning and safety information Sirona Dental Systems GmbH
Operating Instructions C4
+
Electromagnetic compatibility (EMC)The C4
IEC 60601-1-2:2001.
Medical electrical equipment is subject to special precautionary measures with regard to EMC. It must be
installed and operated in compliance with the specifications provided in the document “Installation Requirements”.
Portable and mobile RF communication systems may
interfere with medical electrical equipment. For this reason, the use of mobile radio telephones in practice or
hospital environments must be prohibited.
With video systems:
Electromagnetic disturbances in the environment of the
+
C4
may result in reduced image quality. In such cases,
it is advisable to repeat the exposure before saving the
image.
In case of line voltage drops, image data stored in the
video camera may be lost. Place down the camera and
repeat the exposure.
Image artifacts may possibly occur as a result of
radio-frequency interference caused by the transmitters
of e.g. radio services or radio amateurs.
+
complies with the requirements of
Influence on cardiac pacemakersThe workstation includes strong magnets which can
affect cardiac pacemakers of patients, users and technical personnel.
Therefore, pease pay attention to the safety information
on pages 7 and 57.
Treatment of highly immunosuppressed patientsHighly immunosuppressed patients should not come in
contact with water from the treatment center. The use of
sterile solutions is recommended.
Quality of water / air supplyAir and water supply must meet the requirements speci-
fied in the installation instructions. Use only clean water.
Water quality complianceTo ensure compliance with the medical and national
legal requirements for water from treatment centers,
Sirona recommends equipping the treatment center with
a disinfection system. As owner of the treatment center,
you are responsible for the water quality and may have
to take alternative measures to ensure its compliance if
you operate the treatment center without a disinfection
system.
Please contact your specialized dealer or your relevant
dental association for the respective national requirements and measures.
Highlighting of warning and safety informationTo avoid personal injury and material damage, you must
also observe the warning and safety information provided in the present operating instructions. They are
highlighted by the captionNOTICE, CAUTION or
WARNING.
i
59 06 313 D 3382
8D 3382.201.01.13.02
Sirona Dental Systems GmbH1 Warning and safety information
Operating Instructions C4
+
Symbols usedObserve accompanying documents (on name
plate of chair)
Ventilation slotsUnder no circumstances may the ventilation slots on the
unit be covered, since otherwise the air circulation will be
obstructed.
Do not spray with disinfectants or the like into ventilation
slots.
Vacuum systemThe suction removal of aluminum and other metal oxides
from blasting devices via the automatic separator and
the amalgam separator integrated in the treatment center is prohibited!
This would cause extreme wear and clogging of the suction and water paths.
A separate vacuum system must be used in connection
with metal oxide blasting devices.
Treatment centers equipped with a central wet suction
system are generally suitable for suction removal of the
above material. However, make sure to observe the
instructions provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in
connection with Sirona treatment centers. However, in
such cases, make sure that the system is subsequently
flushed with an adequate amount of water.
Patient chairPlease observe the maximum load capacity of the chair
of 135kg according to EN ISO 6875 (tested with a
four-fold safety factor according to IEC 60601-1)
Electric micromotors in dentist elementIntermittent operation: 1 min. ON – 7 min. OFF with a
motor current of 1.5A.
Drive motors for chair and backrest6% duty time, cycle time 250s.
Maintenance of the treatment centerDespite the outstanding quality of your treatment center
and regular care by the practice team, it is required in the
interest of operational safety that preventive maintenance be performed at predetermined intervals.
In order to guarantee the operational safety and reliability of your treatment center and to avoid damage
due to natural wear, you as the system owner must
have your system checked through regularly by an
authorized service engineer from your dental depot.
Furthermore, safety checks must be performed.
Please contact your dental depot to obtain a maintenance offer.
For details please refer to chapter 17 “Maintenance”
on page 106 .
Dismantling and reassemblyWhen dismantling and reassembling the system, pro-
ceed according to the installation instructions for new
installation in order to guarantee its functioning and stability.
59 06 313 D 3382
D 3382.201.01.13.02
9
1 Warning and safety information Sirona Dental Systems GmbH
Operating Instructions C4
DisposalPlease observe the disposal regulations applicable in
your country.
Within the European Economic Area, this product is subject to Directive 2002/96/EC as well as the corresponding national laws. This directive requires environmentally
sound recycling/disposal of the product.
The product must not be disposed of as domestic refuse!
Please contact your dealer if final disposal of your product is required.
+
10D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH2 Tec hnical description
Operating Instructions C4
+
2 Technical description
Model designationC4
+
, C4+ Cart and C4+ Cabinet
Power supply230VAC, 50 Hz,
115VAC, 50/60Hz
100VAC, 50/60Hz
Nominal current4.5A at 230V
9.5A at 115V
11.5A at 100 V
Main unit fusefor 230VAC: T 6.3 A H , 250 VAC, REF 10 77 452
for 100/115 VAC: T 10 A H , 250 VAC, REF 10 77 460
DVGW: This unit complies with the technical rules and
safety and hygiene requirements for connection to the
drinking water supply, provided that a disinfection system is integrated.
This product bears the CE marking in accordance with
the provisions of the Council Directive 93/42/EEC of
June 14, 1993 concerning medical devices.
0123
59 06 313 D 3382
D 3382.201.01.13.02
11
3 System versions Sirona Dental Systems GmbH
S
0
3.1 System overview C4
+
Operating Instructions C4
3 System versions
+
3.1System overview C4
1
20
21
22
2
4
3
5
6
18
+
87910
11 12
23
13 14
19
16
17
15
1
Assistant element
2
Holder 1: Mini L.E.D. polylight or 3rd suction hose or
SIROCAM 3 / SIROCAM C
3
Holder 2: SPRAYVIT in the assistant element or
SIROCAM 3 / SIROCAM C
4
Holder 3: Suction handpiece
5
Holder 4: Saliva ejector
6
Control panel on the assistant element
7
Rigid or swivelable cuspidor
8
Tu m bl e r f i ll er
9
Water unit with amalgam rotor, disinfection system,
automatic separator, wet suction
10
Patient chair, option: folding armrest
11
Holder 1: SPRAYVIT in the dentist element
12
Holder 2 and 3: Electric motor / highspeed handpiece burr drives
23
13
Holder 4: 3rd electric motor / highspeed handpiece
burr drive or SIROSONIC L scaler or Mini L.E.D.
polylight
14
Holder 5: 4th burr drive (only highspeed handpiece)
or SIROSONIC L or Mini L.E.D. polylight or
SIROCAM 3 / SIROCAM C
15
Additional instrument holder (only for SIROCAM 3 /
SIROCAM C)
16
Control panel on the dentist element
17
C4+ dentist element
18
Locking brake on the assistant element
19
Unit main switch
20
Flat headrest or manually adjustable headrest or
motorized headrest
21
4-way foot control of chair
22
Unit foot switch C+ or C
23
Foot button for SIROCAM 3 / SIROCAM Con the
assistant element
12D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH3 System versions
S
0
Operating Instructions C4
+
3.2 System overview C4+ Cart
3.2System overview C4
1234568791011 12 13 14
18
20
21
22
+
Cart
16 17
15
19
1
Assistant element
2
Holder 1: Mini L.E.D. polylight or 3rd suction hose or
SIROCAM 3 / SIROCAM C
3
Holder 2: SPRAYVIT in the assistant element or
SIROCAM 3 / SIROCAM C
4
Holder 3: Suction handpiece
5
Holder 4: Saliva ejector
6
Control panel on the assistant element
7
Rigid or swivelable cuspidor
8
Tu m bl e r f i ll er
9
Water unit with amalgam rotor, disinfection system,
automatic separator, wet suction
10
Patient chair, option: folding armrest
11
Holder 1: SPRAYVIT in the dentist element
12
Holder 2 and 3: Electric motor / highspeed handpiece burr drives
23
13
Holder 4: 3rd electric motor / highspeed handpiece
burr drive or SIROSONIC L scaler or Mini L.E.D.
polylight
14
Holder 5: 4th burr drive (only highspeed handpiece)
or SIROSONIC L or Mini L.E.D. polylight or
SIROCAM 3 / SIROCAM C
15
Additional instrument holder (only for SIROCAM 3 /
SIROCAM C)
16
Control panel on the dentist element
17
C4+ Cart – dentist element
18
Locking brake on the assistant element
19
Unit main switch
20
Flat headrest or manually adjustable headrest or
motorized headrest
21
4-way foot control of chair
22
Unit foot switch C+ or C
23
Foot button for SIROCAM 3 / SIROCAM Con the
assistant element
59 06 313 D 3382
D 3382.201.01.13.02
13
3 System versions Sirona Dental Systems GmbH
S
0
1
3.3 System overview C4
+
CabinetOperating Instructions C4
+
3.3System overview C4
1234 56879
20
17
11
16
12
13
14
15
22
18
+
Cabinet
10
23
19
21
1
Assistant element
2
Holder 1: Mini L.E.D. polylight or 3rd suction hose or
SIROCAM 3 / SIROCAM C
3
Holder 2: SPRAYVIT in the assistant element or
SIROCAM 3 / SIROCAM C
4
Holder 3: Suction handpiece
5
Holder 4: Saliva ejector
6
Control panel on the assistant element
7
Rigid or swivelable cuspidor
8
Tu m bl e r f i ll er
9
Water unit with amalgam rotor, disinfection system,
automatic separator, wet suction
10
Patient chair, option: folding armrest
11
Holder 1: SPRAYVIT in the dentist element
12
Holder 2 and 3: Electric motor / highspeed handpiece burr drives
13
Holder 4: 3rd electric motor / highspeed handpiece
burr drive or SIROSONIC L scaler or Mini L.E.D.
polylight
14
Holder 5: 4th burr drive (only highspeed handpiece)
or SIROSONIC L or Mini L.E.D. polylight or
SIROCAM 3 / SIROCAM C
15
Additional instrument holder (only for SIROCAM 3 /
SIROCAM C)
16
Control panel on the dentist element
17
C4+ Cabinet – dentist element
18
Locking brake on the assistant element
19
Unit main switch
20
Flat headrest or manually adjustable headrest or
motorized headrest
21
4-way foot control of chair
22
Unit foot switch C+ or C
23
Foot button for SIROCAM 3 / SIROCAM Con the
assistant element
14D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH4 Controls and functional elements
AMALG
DESINF
Operating Instructions C4
+
4.1 Control panel on the dentist element
4 Controls and functional elements
4.1Control panel on the dentist element
Control panel
+
C4
40
100
SS
00
A
–
Mode
RPM
0.2
rpm x 1000
–
+
20
50
11
1
A/B
System status indication display and Mode
key
AMALG appears when the amalgam rotor needs to be
replaced.
DESINF is flashing when DENTOSEPT P must be
refilled.
A appears when user A is selected.
B appears when user B is selected.
22
59 06 313 D 3382
D 3382.201.01.13.02
Mode key with + / – keys for programming basic set-
tings.
In addition to the settings in the Mode dialog, the following direct settings can be made with the
•Intensity of the instrument lights
•Intensity of the operation light
•Tumbler filling time
•Cuspidor flushing time
+ / – keys:
15
4 Controls and functional elements Sirona Dental Systems GmbH
4.1 Control panel on the dentist elementOperating Instructions C4
Main functions – dark blue keys
Quick setting keys for instrument intensity and
40
100
speed
for electric motors and SIROSONIC L ultrasonic scaler
+
A
INSTR. LIGHT
20
0.2
rpm x 1000
%
50
1
Spray ON / OFF
on burr drives
Chip blower
on burr drives
Counterclockwise rotation
for electric motor
Instrument light ON / OFF
Pressing this key for some time displays the light intensity settings dialog.
Mouth rinsing position chair program.
SS
with last position memory function (freely programmable)
Chair program 0
Entry/exit position (freely programmable)
00
Chair programs 1 and 2
11
16D 3382.201.01.13.02
22
(freely programmable)
i
NOTICE
The four program keys are also used for reprogramming
the chair programs.
59 06 313 D 3382
Sirona Dental Systems GmbH4 Controls and functional elements
Operating Instructions C4
+
4.1 Control panel on the dentist element
Additional functions – light blue keys
Preselection of user A or B
A/B
All instruments must be in place.
You cannot switch between the two users after picking
up an instrument
.
Saving the instrument settings
The instrument to be programmed must have been
picked up.
X-ray image viewer ON/OFF.
For SIVISION 3 also WHITE SCREEN activation
Composite key ON / OFF
for SIROLUX operation light, reduced light intensity
8,000 lux for composite fillings
<
A
OPERATING LIGHT
A
CUP FILL TIMER
A
BOWL FLUSH TIMER
SIROLUX operation light ON / OFF
%
for normal light intensity > 24,000 lux
Pressing this key for a while displays the light intensity
settings dialog:
OPERATING LIGHT
Tumbler filling
Pressing this key for a while displays the settings dialog:
CUP FILL TIMER
Cuspidor flushing
Pressing this key for a while displays the settings dialog:
BOWL FLUSH TIMER
Freely selectable function
e.g. call key, always functions as button
freely available relay 230 V, 6 A
(connected by the service engineer)
Freely selectable function
59 06 313 D 3382
D 3382.201.01.13.02
The function can be preset in the Mode dialog as button
or switch.
freely available relay 230V, 6 A
(connected by the service engineer).
17
4 Controls and functional elements Sirona Dental Systems GmbH
4.1 Control panel on the dentist elementOperating Instructions C4
Chair functions – gray keys
for manual chair settings:
Tilt backrest backward or forward.
Height adjustment down / up
Additional functions of the control
panel video keys
i
NOTICE
Apart from the previously described functions, the keys
of the dentist element have additional functions in connection with the video application. These are described
in chapter 14 starting on page 71 .
+
18D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH4 Controls and functional elements
Operating Instructions C4
+
4.2 Control panel on the assistant element
4.2Control panel on the assistant element
Control panel
11
00
22
SS
Main functions – dark blue keys
Mouth rinsing position chair program.
with memory function (freely programmable)
Chair program 0
Entry/exit position (freely programmable)
Chair programs 1 and 2
(freely programmable)
i
NOTICE
The four program keys are also used for reprogramming
the chair programs.
59 06 313 D 3382
D 3382.201.01.13.02
19
4 Controls and functional elements Sirona Dental Systems GmbH
4.2 Control panel on the assistant elementOperating Instructions C4
Additional functions – light blue keys
Tumbler filling
ON / OFF
Cuspidor flushing
ON / OFF
X-ray image viewer ON / OFF
for SIVISION 3 also WHITE SCREEN activation
Treatment center sanitation
+
SAN
L
This key can be used to start the treatment center sanitation program (see instructions “Care and Cleaning
by the Practice Team”).
Freely selectable function
The function can be preset in the Mode dialog as button
or switch.
Light ON/OFF
With SPRAYVIT picked up:
•Instrument light
With SPRAYVIT in holder:
•Operation light
20D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH5 Putting the system into operation
Operating Instructions C4
+
5 Putting the system into operation
Initial startup
The disinfection system adds a disinfectant to the water
(1:100) to prevent the formation of microorganisms in
the water system.
Prior to initial startup of your treatment center, sanita-tion must be performed.
If the service engineer has skipped the sanitation procedure after installation of your treatment center on the
basis of an agreement with you, please perform sanitation yourself as described in the separate instructions
“Care and Cleaning by the Practice Team”.
Sanitation takes approx. 24 hours.
Main switch
Switch ON the main switch ( l ).
A
READY
10:04:27
ON
OFF
During power-up, the treatment center automatically
performs a self-test.
CAUTION
For safety reasons, always switch the treatment center
OFF (O) after consulting hours. This cuts off the air and
water supply as well as the line voltage.
Ready for operation
After the end of the self-test, the message READY the
time and the last user selected (A or B) appear on the
display.
After the treatment center is switched on, the cursor is
between the chair program keys 1 and 2. If a program
LED is illuminated, then the chair was in a programmed
position before switching off. If no program LED is illuminated, then the chair was in a non-programmed position
before switching off.
The unit is now ready for operation.
59 06 313 D 3382
D 3382.201.01.13.02
21
5 Putting the system into operation Sirona Dental Systems GmbH
A
MAINTENANCE
IN 30 DAYS
Operating Instructions C4
Display of next maintenance date
the next maintenance date is less than 30 days away,
this message appears each time the unit is switched on.
MAINTENANCE
IN 30 DAYS
(see MAINTENANCE Mode dialog).
After this date has expired, the following message is displayed:
MAINTENANCE
REQUIRED
The display message then disappears as soon as an
instrument is removed from its holder.
i
NOTICE
This message can be reset only by the service technician after completion of maintenance work.
+
right
A
B
READY
10:04:27
wrong
A/B
CAUTION
The dental chair has a maximum load capacity of 135kg
according to EN ISO 6875 (tested with a four-fold safety
factor according to IEC 60601-1).
The patient’s arms and legs must rest on the upholstery
of the chair!
Preselecting the user
The treatment center offers the possibility of managing
two different chair and instrument programs for two
users.
Preselect user A or B using the A/B key, with all instruments in place.
The preselected user is shown in the top left corner of
the display. All settings which have been stored for that
user are then activated.
22D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH6 Foot switch
Operating Instructions C4
+
6.1 C+ foot switch
6 Foot switch
6.1C
+
foot switch
1.
B
C
2.
A
D
This foot switch enables hands-free operation by means
of the cursor control, but also manual operation by
means of the keys.
The foot switch has a 4-way foot control plate for cursor
navigation, which works independently of the pedal.
Functions
Step on foot pedal – instrument picked up
Activation of the instrument.
If appropriate, intensity control relative to pedal
movement (if “speed controller foot switch” has been
preset under Mode).
1. Shift 4-way foot control plate
With cursor control enabled
(Mode dialog 11.8):
A to the front – cursor upward
B to the back – cursor downward
C to the left – cursor to the left
D to the right – cursor to the right
Cursor navigation along the marked path and release of the corresponding key function.
3.
4.
2. Actuate left button – all instruments in place
Programmed movement of the chair to the mouth
rinsing position S or the last treatment position (according to starting situation).
Actuate left button – instrument picked up
Toggling between spray ON and spray OFF, or
SIVISION function.
3. Actuate right button – all instruments in place
Programmed movement of the chair to the entry/exit
position 0.
Actuate right button – instrument picked up
Chip blower active for duration of actuation, or
SIVISION function.
59 06 313 D 3382
D 3382.201.01.13.02
23
6 Foot switch Sirona Dental Systems GmbH
6.1 C+ foot switchOperating Instructions C4
Automatic cursor control with C+ foot switch
2.1. / 3.
A yellow diode on the dentist element indicates the position of the cursor. The cursor path is marked by lines.
1. After you switch on the system, or when the chair
has reached the entry/exit position 0, the cursor is
between the chair program keys 1 and 2.
2. When you pick up an instrument or activate any of
the instruments, the cursor jumps to the horizontal
navigation path (position between 1st and 2nd row of
keys) in the instrument block.
3. After you have placed the instrument in its holder,
the cursor jumps back to its position between the
chair program keys 1 and 2.
CEREC Chairline foot switch
This foot switch includes all of the functions of the
+
foot switch.
C
+
It also features an additional pedal for operation of the
CEREC Chairline.
For details please refer to the CEREC Chairline Operating Instructions, REF 60 46 028.
24D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH6 Foot switch
Operating Instructions C4
+
6.2 C foot switch
6.2C foot switch
This foot switch enables hands-free operation by con-
A*
C
trolling all functions with the entire pedal surface. However, it is also possible to control the unit manually with
the keys.
For cursor navigation, place your foot fully on the pedal
surface.
Functions (factory settings)
i
NOTICE
The foot switch functions A and B can be reversed by the
service engineer upon request.
B*
D
With cursor control enabled (Mode dialog 11.8)
Actuate foot switch in direction A – cursor upward.
Actuate foot switch in direction B – cursor downward.
Swing foot switch to the left C – cursor to the left.
Swing foot switch to the right D – cursor to the right.
Cursor navigation along the marked path and release of
the corresponding key function.
Instrument start
The cursor must be between the spray and the chip
blower key.
Actuate the foot switch in direction A after picking up the
instrument.
With cursor control disabled (Mode dialog
11.8)
Actuate foot switch in direction A:
With instrument picked up – Activation of the instru-
ment.
Actuate foot switch in direction B:
With instrument picked up – No function
59 06 313 D 3382
D 3382.201.01.13.02
Swing foot switch to the left C:
With instrument picked up – Spray ON/OFF for burr
instruments or SIVISION function.
With instruments in place – SIVISION function.
Swing foot switch to the right D:
With instrument picked up – Chip blower ON for burr
instruments or SIVISION function.
With instruments in place – SIVISION function.
25
6 Foot switch Sirona Dental Systems GmbH
6.3 Selecting the chair programOperating Instructions C4
Automatic cursor control with C foot switch
+
3.
1.
2.
6.3Selecting the chair program
A yellow diode on the dentist element indicates the position of the cursor. The cursor path is marked by lines.
1. After you switch on the system, or when the chair
has reached the entry/exit position 0, the cursor is
between the chair program keys 1 and 2.
2. When you pick up an instrument or activate any of
the instruments, the cursor jumps to a position between the spray and chip blower keys in the first in-
strument block.
If the cursor is in the quick setting key column at the
very left, it also jumps to a position between the
spray and chip blower keys in the instrument block
when you swing the foot switch to the right (D).
3. After you have placed the instrument in its holder,
the cursor jumps back to its position between the
chair program keys S and 0.
Activate the desired chair program 1 or 2. The chair
moves to the programmed treatment position.
Move the dentist element by hand to the desired treat-
1
ment position. Use the handle (1) on the dentist element
to do this.
The dentist element can be fixed by the service engineer
at a height which is convenient for you.
26D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH7 Program selection
Operating Instructions C4
+
7.1 Safety
7 Program selection
7.1Safety
CAUTION
The patient’s arms and legs must be resting on the chair
upholstery during the program run!
CAUTION
Make sure that no obstacles (e.g. window wings, drawers, devices, …) extend into the movement range.
A
BACKREST
10:04:27
Safety stop
A built-in safety circuit stops the chair movement in the
following situations:
•The toeboard collides with an obstacle.
•The backrest collides with an obstacle.
•The motorized headrest (if available) collides with an
obstacle
•The swivelable cuspidor is swiveled in during chair
movement.
•The assistant element is pushed upward or downward by an obstacle.
At the same time, a double beep signal sounds.
The chair then moves automatically a little upward until
the obstacle is cleared (except for the swivelable cuspidor and assistant element).
With instruments in place, the last activated safety
switch is indicated in plain text on the display:
•TILTING PART for motorized headrest
•BACKREST
•TOEBOARD
•ASSISTANT ELEMENT (is also displayed when the
cuspidor safety switch is activated)
59 06 313 D 3382
D 3382.201.01.13.02
The message CUSPIDOR appears after 10 seconds if
the automatic return movement of the cuspidor to its
position is inhibited.
27
7 Program selection Sirona Dental Systems GmbH
7.2 Program selectionOperating Instructions C4
7.2Program selection
Keys for selecting the programmed patient positions.
+
SS
11
00
22
00
SS
With swivelable cuspidor:
Before the chair moves to the selected program position,
the swiveled-in cuspidor automatically returns to its
starting position.
Factory settings:
The operation light switches OFF automatically before
the chair moves to the entry/exit position 0 or the mouth
rinsing position S.
The operation light switches ON automatically after the
chair has reached patient position 1 or 2.
Changing the factory settings:
When programming the patient positions you can preset
whether the operation light should switch OFF or ON in
the different chair programs 0, S, 1 or 2 (see page 36).
Program key 0
Intended for the entry/exit position.
Program key S with memory function.
In the S program, you can program the tumbler filling
and / or the cuspidor flushing functions (see Mode dia-
log 11.9 and 11.10).
When you press this key a second time, the treatment
center returns to its previous position (memory function).
Example: You have programmed this key for the mouth
rinsing position. When you press this key, the treatment
center moves to the programmed position.
Memory function: When you press this key a second
time, the treatment center returns to its previous position.
28D 3382.201.01.13.02
59 06 313 D 3382
Sirona Dental Systems GmbH7 Program selection
Operating Instructions C4
+
7.3 MOVEMENT STOP!
7.3MOVEMENT STOP!
You can stop the movement of the chair to a programmed position as follows:
•By pressing one of the chair-related keys on the
dentist or assistant element.
•By actuating the 4-way foot control.
•By pushing the 4-way switches on a motorized
headrest in any direction.
All movements of the treatment center are stopped
immediately.
At the same time, a double beep signal sounds.
Please note that when you press the program key of
the program which is just running a second time, the
movement will NOT stop!
59 06 313 D 3382
D 3382.201.01.13.02
29
8 Patient positions Sirona Dental Systems GmbH
8.1 Flat headrestOperating Instructions C4
8 Patient positions
8.1Flat headrest
The headrest must be adjusted manually.
+
2
1
8.2Articulated headrest
To do this, take the load off the backrest and pull out or
push in the headrest using the handle (1).
The head pad (2) is held magnetically.
DANGER
A strong magnet is integrated in the underside of the
head cushion (2) which could affect a nearby cardiac
pacemaker.
Therefore, please avoid bringing the magnet in close
proximity of a cardiac pacemaker of a patient, user or
technical personnel.
In addition, direct contact of the head cushion with
magnetic cards can lead to a data loss.
The headrest must be adjusted manually.
To do this, take the load off the backrest and pull out or
push in the headrest using the handle (1).
1
2
30D 3382.201.01.13.02
Tilt the headrest backward using the locking button (2).
DANGER
Tilt the headrest forward by simply pulling it out.