The technical documentation supplied is also part of the
product. You should always keep this documentation
within reach.
These Operating Instructions describe your treatment
center with maximum possible equipment.
For daily care of your treatment center, please observe
the instructions provided in “Care and Maintenance by the Practice Team”.
To safeguard your warranty claims, please complete the
attached “Installation Report / Warranty Passport”
together with the service engineer immediately after the
installation of your treatment center.
Separate Operating Instructions with corresponding
instructions for care are enclosed with the dental
instruments, SIROLUX FANTASTIC, monitor, Cerec
Chairline, SIROCAM 3, tray and X-ray image viewer.
Prior to start-up, you must read all Operating Instructions
to familiarize yourself with the treatment center.
To prevent any personal injury or material damage,
pay special attention to any notes printed in bold type or
marked in one of the ways indicated below:
+
dental treatment
i
NOTE
CAUTION
WARNING
Additional information, explanation or supplement
The information provided under this keyword directly
concerns the functioning of the product and/or warns
against possible operating errors.
This warning symbol warns against possible hazards for
the product or the user.
In case you get stuck despite having thoroughly studied
the Operating Instructions, please contact your dental
depot.
Your C4+ Team
2D 3382.201.01.13.02
59 06 313 D 3382
Page 3
Sirona Dental Systems GmbH Table of Contents
bеЦдблЬ
Table of Contents
1Warning and safety information ................................................................................ 7
16.3Seat cushion C ................................................................................................................................. 101
Sirona Dental Systems GmbH1 Warning and safety information
Operating Instructions C4
+
1 Warning and safety information
Intended useThis dental treatment center is intended for diagnosis,
therapy and dental treatment of humans by properly
trained personnel.
This unit is not intended for operation in areas subject
to explosion hazards.
On-site installationThe on-site installation must have been performed
according to our requirements. The details are described
in the document “Installation Requirements”.
Maintenance and repairAs manufacturers of dental medical equipment and in
the interest of the operational safety of your system, we
stress the importance of having maintenance and repair of your treatment center performed only by ourselves or by agencies expressly authorized by us. Furthermore components influencing the safety of the unit
should always be replaced with original spare parts
upon failure.
When having such work done, we suggest that you
request a certificate stating the type and extent of work
performed, including statements concerning any modifications of the rated parameters or of the operating
ranges (if applicable), as well as the date, name of organization and signature.
Modifications of the systemModifications of this system which could impair the
safety of the system owner, patients or other persons
are prohibited by law!
For reasons of product safety, only original Sirona
accessories approved for this product, or accessories
from third parties approved by Sirona, may be used. The
user is responsible for dangers resulting from the use of
non-approved accessories.
If any equipment not approved by Sirona is connected, it
must comply with the applicable standards:
IEC 60950 for information technology equipment (e.g.
PCs), and IEC 60601-1 for medical electrical equipment.
Combination with other equipmentAny person who assembles or modifies a medical elec-
trical system complying with the standard IEC 60601-1-1
(safety requirements for medical electrical equipment)
by combining it with other equipment (e.g., by connecting it with a PC) is responsible for ensuring that the
requirements of this regulation are met to their full extent
for the safety of the patients, operators and environment.
In case of doubt, contact the manufacturer of the system
components.
59 06 313 D 3382
D 3382.201.01.13.02
7
Page 8
1 Warning and safety information Sirona Dental Systems GmbH
Operating Instructions C4
+
Electromagnetic compatibility (EMC)The C4
IEC 60601-1-2:2001.
Medical electrical equipment is subject to special precautionary measures with regard to EMC. It must be
installed and operated in compliance with the specifications provided in the document “Installation Requirements”.
Portable and mobile RF communication systems may
interfere with medical electrical equipment. For this reason, the use of mobile radio telephones in practice or
hospital environments must be prohibited.
With video systems:
Electromagnetic disturbances in the environment of the
+
C4
may result in reduced image quality. In such cases,
it is advisable to repeat the exposure before saving the
image.
In case of line voltage drops, image data stored in the
video camera may be lost. Place down the camera and
repeat the exposure.
Image artifacts may possibly occur as a result of
radio-frequency interference caused by the transmitters
of e.g. radio services or radio amateurs.
+
complies with the requirements of
Influence on cardiac pacemakersThe workstation includes strong magnets which can
affect cardiac pacemakers of patients, users and technical personnel.
Therefore, pease pay attention to the safety information
on pages 7 and 57.
Treatment of highly immunosuppressed patientsHighly immunosuppressed patients should not come in
contact with water from the treatment center. The use of
sterile solutions is recommended.
Quality of water / air supplyAir and water supply must meet the requirements speci-
fied in the installation instructions. Use only clean water.
Water quality complianceTo ensure compliance with the medical and national
legal requirements for water from treatment centers,
Sirona recommends equipping the treatment center with
a disinfection system. As owner of the treatment center,
you are responsible for the water quality and may have
to take alternative measures to ensure its compliance if
you operate the treatment center without a disinfection
system.
Please contact your specialized dealer or your relevant
dental association for the respective national requirements and measures.
Highlighting of warning and safety informationTo avoid personal injury and material damage, you must
also observe the warning and safety information provided in the present operating instructions. They are
highlighted by the captionNOTICE, CAUTION or
WARNING.
i
59 06 313 D 3382
8D 3382.201.01.13.02
Page 9
Sirona Dental Systems GmbH1 Warning and safety information
Operating Instructions C4
+
Symbols usedObserve accompanying documents (on name
plate of chair)
Ventilation slotsUnder no circumstances may the ventilation slots on the
unit be covered, since otherwise the air circulation will be
obstructed.
Do not spray with disinfectants or the like into ventilation
slots.
Vacuum systemThe suction removal of aluminum and other metal oxides
from blasting devices via the automatic separator and
the amalgam separator integrated in the treatment center is prohibited!
This would cause extreme wear and clogging of the suction and water paths.
A separate vacuum system must be used in connection
with metal oxide blasting devices.
Treatment centers equipped with a central wet suction
system are generally suitable for suction removal of the
above material. However, make sure to observe the
instructions provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in
connection with Sirona treatment centers. However, in
such cases, make sure that the system is subsequently
flushed with an adequate amount of water.
Patient chairPlease observe the maximum load capacity of the chair
of 135kg according to EN ISO 6875 (tested with a
four-fold safety factor according to IEC 60601-1)
Electric micromotors in dentist elementIntermittent operation: 1 min. ON – 7 min. OFF with a
motor current of 1.5A.
Drive motors for chair and backrest6% duty time, cycle time 250s.
Maintenance of the treatment centerDespite the outstanding quality of your treatment center
and regular care by the practice team, it is required in the
interest of operational safety that preventive maintenance be performed at predetermined intervals.
In order to guarantee the operational safety and reliability of your treatment center and to avoid damage
due to natural wear, you as the system owner must
have your system checked through regularly by an
authorized service engineer from your dental depot.
Furthermore, safety checks must be performed.
Please contact your dental depot to obtain a maintenance offer.
For details please refer to chapter 17 “Maintenance”
on page 106 .
Dismantling and reassemblyWhen dismantling and reassembling the system, pro-
ceed according to the installation instructions for new
installation in order to guarantee its functioning and stability.
59 06 313 D 3382
D 3382.201.01.13.02
9
Page 10
1 Warning and safety information Sirona Dental Systems GmbH
Operating Instructions C4
DisposalPlease observe the disposal regulations applicable in
your country.
Within the European Economic Area, this product is subject to Directive 2002/96/EC as well as the corresponding national laws. This directive requires environmentally
sound recycling/disposal of the product.
The product must not be disposed of as domestic refuse!
Please contact your dealer if final disposal of your product is required.
+
10D 3382.201.01.13.02
59 06 313 D 3382
Page 11
Sirona Dental Systems GmbH2 Tec hnical description
Operating Instructions C4
+
2 Technical description
Model designationC4
+
, C4+ Cart and C4+ Cabinet
Power supply230VAC, 50 Hz,
115VAC, 50/60Hz
100VAC, 50/60Hz
Nominal current4.5A at 230V
9.5A at 115V
11.5A at 100 V
Main unit fusefor 230VAC: T 6.3 A H , 250 VAC, REF 10 77 452
for 100/115 VAC: T 10 A H , 250 VAC, REF 10 77 460
DVGW: This unit complies with the technical rules and
safety and hygiene requirements for connection to the
drinking water supply, provided that a disinfection system is integrated.
This product bears the CE marking in accordance with
the provisions of the Council Directive 93/42/EEC of
June 14, 1993 concerning medical devices.
0123
59 06 313 D 3382
D 3382.201.01.13.02
11
Page 12
3 System versions Sirona Dental Systems GmbH
S
0
3.1 System overview C4
+
Operating Instructions C4
3 System versions
+
3.1System overview C4
1
20
21
22
2
4
3
5
6
18
+
87910
11 12
23
13 14
19
16
17
15
1
Assistant element
2
Holder 1: Mini L.E.D. polylight or 3rd suction hose or
SIROCAM 3 / SIROCAM C
3
Holder 2: SPRAYVIT in the assistant element or
SIROCAM 3 / SIROCAM C
4
Holder 3: Suction handpiece
5
Holder 4: Saliva ejector
6
Control panel on the assistant element
7
Rigid or swivelable cuspidor
8
Tu m bl e r f i ll er
9
Water unit with amalgam rotor, disinfection system,
automatic separator, wet suction
10
Patient chair, option: folding armrest
11
Holder 1: SPRAYVIT in the dentist element
12
Holder 2 and 3: Electric motor / highspeed handpiece burr drives
23
13
Holder 4: 3rd electric motor / highspeed handpiece
burr drive or SIROSONIC L scaler or Mini L.E.D.
polylight
14
Holder 5: 4th burr drive (only highspeed handpiece)
or SIROSONIC L or Mini L.E.D. polylight or
SIROCAM 3 / SIROCAM C
15
Additional instrument holder (only for SIROCAM 3 /
SIROCAM C)
16
Control panel on the dentist element
17
C4+ dentist element
18
Locking brake on the assistant element
19
Unit main switch
20
Flat headrest or manually adjustable headrest or
motorized headrest
21
4-way foot control of chair
22
Unit foot switch C+ or C
23
Foot button for SIROCAM 3 / SIROCAM Con the
assistant element
12D 3382.201.01.13.02
59 06 313 D 3382
Page 13
Sirona Dental Systems GmbH3 System versions
S
0
Operating Instructions C4
+
3.2 System overview C4+ Cart
3.2System overview C4
1234568791011 12 13 14
18
20
21
22
+
Cart
16 17
15
19
1
Assistant element
2
Holder 1: Mini L.E.D. polylight or 3rd suction hose or
SIROCAM 3 / SIROCAM C
3
Holder 2: SPRAYVIT in the assistant element or
SIROCAM 3 / SIROCAM C
4
Holder 3: Suction handpiece
5
Holder 4: Saliva ejector
6
Control panel on the assistant element
7
Rigid or swivelable cuspidor
8
Tu m bl e r f i ll er
9
Water unit with amalgam rotor, disinfection system,
automatic separator, wet suction
10
Patient chair, option: folding armrest
11
Holder 1: SPRAYVIT in the dentist element
12
Holder 2 and 3: Electric motor / highspeed handpiece burr drives
23
13
Holder 4: 3rd electric motor / highspeed handpiece
burr drive or SIROSONIC L scaler or Mini L.E.D.
polylight
14
Holder 5: 4th burr drive (only highspeed handpiece)
or SIROSONIC L or Mini L.E.D. polylight or
SIROCAM 3 / SIROCAM C
15
Additional instrument holder (only for SIROCAM 3 /
SIROCAM C)
16
Control panel on the dentist element
17
C4+ Cart – dentist element
18
Locking brake on the assistant element
19
Unit main switch
20
Flat headrest or manually adjustable headrest or
motorized headrest
21
4-way foot control of chair
22
Unit foot switch C+ or C
23
Foot button for SIROCAM 3 / SIROCAM Con the
assistant element
59 06 313 D 3382
D 3382.201.01.13.02
13
Page 14
3 System versions Sirona Dental Systems GmbH
S
0
1
3.3 System overview C4
+
CabinetOperating Instructions C4
+
3.3System overview C4
1234 56879
20
17
11
16
12
13
14
15
22
18
+
Cabinet
10
23
19
21
1
Assistant element
2
Holder 1: Mini L.E.D. polylight or 3rd suction hose or
SIROCAM 3 / SIROCAM C
3
Holder 2: SPRAYVIT in the assistant element or
SIROCAM 3 / SIROCAM C
4
Holder 3: Suction handpiece
5
Holder 4: Saliva ejector
6
Control panel on the assistant element
7
Rigid or swivelable cuspidor
8
Tu m bl e r f i ll er
9
Water unit with amalgam rotor, disinfection system,
automatic separator, wet suction
10
Patient chair, option: folding armrest
11
Holder 1: SPRAYVIT in the dentist element
12
Holder 2 and 3: Electric motor / highspeed handpiece burr drives
13
Holder 4: 3rd electric motor / highspeed handpiece
burr drive or SIROSONIC L scaler or Mini L.E.D.
polylight
14
Holder 5: 4th burr drive (only highspeed handpiece)
or SIROSONIC L or Mini L.E.D. polylight or
SIROCAM 3 / SIROCAM C
15
Additional instrument holder (only for SIROCAM 3 /
SIROCAM C)
16
Control panel on the dentist element
17
C4+ Cabinet – dentist element
18
Locking brake on the assistant element
19
Unit main switch
20
Flat headrest or manually adjustable headrest or
motorized headrest
21
4-way foot control of chair
22
Unit foot switch C+ or C
23
Foot button for SIROCAM 3 / SIROCAM Con the
assistant element
14D 3382.201.01.13.02
59 06 313 D 3382
Page 15
Sirona Dental Systems GmbH4 Controls and functional elements
AMALG
DESINF
Operating Instructions C4
+
4.1 Control panel on the dentist element
4 Controls and functional elements
4.1Control panel on the dentist element
Control panel
+
C4
40
100
SS
00
A
–
Mode
RPM
0.2
rpm x 1000
–
+
20
50
11
1
A/B
System status indication display and Mode
key
AMALG appears when the amalgam rotor needs to be
replaced.
DESINF is flashing when DENTOSEPT P must be
refilled.
A appears when user A is selected.
B appears when user B is selected.
22
59 06 313 D 3382
D 3382.201.01.13.02
Mode key with + / – keys for programming basic set-
tings.
In addition to the settings in the Mode dialog, the following direct settings can be made with the
•Intensity of the instrument lights
•Intensity of the operation light
•Tumbler filling time
•Cuspidor flushing time
+ / – keys:
15
Page 16
4 Controls and functional elements Sirona Dental Systems GmbH
4.1 Control panel on the dentist elementOperating Instructions C4
Main functions – dark blue keys
Quick setting keys for instrument intensity and
40
100
speed
for electric motors and SIROSONIC L ultrasonic scaler
+
A
INSTR. LIGHT
20
0.2
rpm x 1000
%
50
1
Spray ON / OFF
on burr drives
Chip blower
on burr drives
Counterclockwise rotation
for electric motor
Instrument light ON / OFF
Pressing this key for some time displays the light intensity settings dialog.
Mouth rinsing position chair program.
SS
with last position memory function (freely programmable)
Chair program 0
Entry/exit position (freely programmable)
00
Chair programs 1 and 2
11
16D 3382.201.01.13.02
22
(freely programmable)
i
NOTICE
The four program keys are also used for reprogramming
the chair programs.
59 06 313 D 3382
Page 17
Sirona Dental Systems GmbH4 Controls and functional elements
Operating Instructions C4
+
4.1 Control panel on the dentist element
Additional functions – light blue keys
Preselection of user A or B
A/B
All instruments must be in place.
You cannot switch between the two users after picking
up an instrument
.
Saving the instrument settings
The instrument to be programmed must have been
picked up.
X-ray image viewer ON/OFF.
For SIVISION 3 also WHITE SCREEN activation
Composite key ON / OFF
for SIROLUX operation light, reduced light intensity
8,000 lux for composite fillings
<
A
OPERATING LIGHT
A
CUP FILL TIMER
A
BOWL FLUSH TIMER
SIROLUX operation light ON / OFF
%
for normal light intensity > 24,000 lux
Pressing this key for a while displays the light intensity
settings dialog:
OPERATING LIGHT
Tumbler filling
Pressing this key for a while displays the settings dialog:
CUP FILL TIMER
Cuspidor flushing
Pressing this key for a while displays the settings dialog:
BOWL FLUSH TIMER
Freely selectable function
e.g. call key, always functions as button
freely available relay 230 V, 6 A
(connected by the service engineer)
Freely selectable function
59 06 313 D 3382
D 3382.201.01.13.02
The function can be preset in the Mode dialog as button
or switch.
freely available relay 230V, 6 A
(connected by the service engineer).
17
Page 18
4 Controls and functional elements Sirona Dental Systems GmbH
4.1 Control panel on the dentist elementOperating Instructions C4
Chair functions – gray keys
for manual chair settings:
Tilt backrest backward or forward.
Height adjustment down / up
Additional functions of the control
panel video keys
i
NOTICE
Apart from the previously described functions, the keys
of the dentist element have additional functions in connection with the video application. These are described
in chapter 14 starting on page 71 .
+
18D 3382.201.01.13.02
59 06 313 D 3382
Page 19
Sirona Dental Systems GmbH4 Controls and functional elements
Operating Instructions C4
+
4.2 Control panel on the assistant element
4.2Control panel on the assistant element
Control panel
11
00
22
SS
Main functions – dark blue keys
Mouth rinsing position chair program.
with memory function (freely programmable)
Chair program 0
Entry/exit position (freely programmable)
Chair programs 1 and 2
(freely programmable)
i
NOTICE
The four program keys are also used for reprogramming
the chair programs.
59 06 313 D 3382
D 3382.201.01.13.02
19
Page 20
4 Controls and functional elements Sirona Dental Systems GmbH
4.2 Control panel on the assistant elementOperating Instructions C4
Additional functions – light blue keys
Tumbler filling
ON / OFF
Cuspidor flushing
ON / OFF
X-ray image viewer ON / OFF
for SIVISION 3 also WHITE SCREEN activation
Treatment center sanitation
+
SAN
L
This key can be used to start the treatment center sanitation program (see instructions “Care and Cleaning
by the Practice Team”).
Freely selectable function
The function can be preset in the Mode dialog as button
or switch.
Light ON/OFF
With SPRAYVIT picked up:
•Instrument light
With SPRAYVIT in holder:
•Operation light
20D 3382.201.01.13.02
59 06 313 D 3382
Page 21
Sirona Dental Systems GmbH5 Putting the system into operation
Operating Instructions C4
+
5 Putting the system into operation
Initial startup
The disinfection system adds a disinfectant to the water
(1:100) to prevent the formation of microorganisms in
the water system.
Prior to initial startup of your treatment center, sanita-tion must be performed.
If the service engineer has skipped the sanitation procedure after installation of your treatment center on the
basis of an agreement with you, please perform sanitation yourself as described in the separate instructions
“Care and Cleaning by the Practice Team”.
Sanitation takes approx. 24 hours.
Main switch
Switch ON the main switch ( l ).
A
READY
10:04:27
ON
OFF
During power-up, the treatment center automatically
performs a self-test.
CAUTION
For safety reasons, always switch the treatment center
OFF (O) after consulting hours. This cuts off the air and
water supply as well as the line voltage.
Ready for operation
After the end of the self-test, the message READY the
time and the last user selected (A or B) appear on the
display.
After the treatment center is switched on, the cursor is
between the chair program keys 1 and 2. If a program
LED is illuminated, then the chair was in a programmed
position before switching off. If no program LED is illuminated, then the chair was in a non-programmed position
before switching off.
The unit is now ready for operation.
59 06 313 D 3382
D 3382.201.01.13.02
21
Page 22
5 Putting the system into operation Sirona Dental Systems GmbH
A
MAINTENANCE
IN 30 DAYS
Operating Instructions C4
Display of next maintenance date
the next maintenance date is less than 30 days away,
this message appears each time the unit is switched on.
MAINTENANCE
IN 30 DAYS
(see MAINTENANCE Mode dialog).
After this date has expired, the following message is displayed:
MAINTENANCE
REQUIRED
The display message then disappears as soon as an
instrument is removed from its holder.
i
NOTICE
This message can be reset only by the service technician after completion of maintenance work.
+
right
A
B
READY
10:04:27
wrong
A/B
CAUTION
The dental chair has a maximum load capacity of 135kg
according to EN ISO 6875 (tested with a four-fold safety
factor according to IEC 60601-1).
The patient’s arms and legs must rest on the upholstery
of the chair!
Preselecting the user
The treatment center offers the possibility of managing
two different chair and instrument programs for two
users.
Preselect user A or B using the A/B key, with all instruments in place.
The preselected user is shown in the top left corner of
the display. All settings which have been stored for that
user are then activated.
22D 3382.201.01.13.02
59 06 313 D 3382
Page 23
Sirona Dental Systems GmbH6 Foot switch
Operating Instructions C4
+
6.1 C+ foot switch
6 Foot switch
6.1C
+
foot switch
1.
B
C
2.
A
D
This foot switch enables hands-free operation by means
of the cursor control, but also manual operation by
means of the keys.
The foot switch has a 4-way foot control plate for cursor
navigation, which works independently of the pedal.
Functions
Step on foot pedal – instrument picked up
Activation of the instrument.
If appropriate, intensity control relative to pedal
movement (if “speed controller foot switch” has been
preset under Mode).
1. Shift 4-way foot control plate
With cursor control enabled
(Mode dialog 11.8):
A to the front – cursor upward
B to the back – cursor downward
C to the left – cursor to the left
D to the right – cursor to the right
Cursor navigation along the marked path and release of the corresponding key function.
3.
4.
2. Actuate left button – all instruments in place
Programmed movement of the chair to the mouth
rinsing position S or the last treatment position (according to starting situation).
Actuate left button – instrument picked up
Toggling between spray ON and spray OFF, or
SIVISION function.
3. Actuate right button – all instruments in place
Programmed movement of the chair to the entry/exit
position 0.
Actuate right button – instrument picked up
Chip blower active for duration of actuation, or
SIVISION function.
59 06 313 D 3382
D 3382.201.01.13.02
23
Page 24
6 Foot switch Sirona Dental Systems GmbH
6.1 C+ foot switchOperating Instructions C4
Automatic cursor control with C+ foot switch
2.1. / 3.
A yellow diode on the dentist element indicates the position of the cursor. The cursor path is marked by lines.
1. After you switch on the system, or when the chair
has reached the entry/exit position 0, the cursor is
between the chair program keys 1 and 2.
2. When you pick up an instrument or activate any of
the instruments, the cursor jumps to the horizontal
navigation path (position between 1st and 2nd row of
keys) in the instrument block.
3. After you have placed the instrument in its holder,
the cursor jumps back to its position between the
chair program keys 1 and 2.
CEREC Chairline foot switch
This foot switch includes all of the functions of the
+
foot switch.
C
+
It also features an additional pedal for operation of the
CEREC Chairline.
For details please refer to the CEREC Chairline Operating Instructions, REF 60 46 028.
24D 3382.201.01.13.02
59 06 313 D 3382
Page 25
Sirona Dental Systems GmbH6 Foot switch
Operating Instructions C4
+
6.2 C foot switch
6.2C foot switch
This foot switch enables hands-free operation by con-
A*
C
trolling all functions with the entire pedal surface. However, it is also possible to control the unit manually with
the keys.
For cursor navigation, place your foot fully on the pedal
surface.
Functions (factory settings)
i
NOTICE
The foot switch functions A and B can be reversed by the
service engineer upon request.
B*
D
With cursor control enabled (Mode dialog 11.8)
Actuate foot switch in direction A – cursor upward.
Actuate foot switch in direction B – cursor downward.
Swing foot switch to the left C – cursor to the left.
Swing foot switch to the right D – cursor to the right.
Cursor navigation along the marked path and release of
the corresponding key function.
Instrument start
The cursor must be between the spray and the chip
blower key.
Actuate the foot switch in direction A after picking up the
instrument.
With cursor control disabled (Mode dialog
11.8)
Actuate foot switch in direction A:
With instrument picked up – Activation of the instru-
ment.
Actuate foot switch in direction B:
With instrument picked up – No function
59 06 313 D 3382
D 3382.201.01.13.02
Swing foot switch to the left C:
With instrument picked up – Spray ON/OFF for burr
instruments or SIVISION function.
With instruments in place – SIVISION function.
Swing foot switch to the right D:
With instrument picked up – Chip blower ON for burr
instruments or SIVISION function.
With instruments in place – SIVISION function.
25
Page 26
6 Foot switch Sirona Dental Systems GmbH
6.3 Selecting the chair programOperating Instructions C4
Automatic cursor control with C foot switch
+
3.
1.
2.
6.3Selecting the chair program
A yellow diode on the dentist element indicates the position of the cursor. The cursor path is marked by lines.
1. After you switch on the system, or when the chair
has reached the entry/exit position 0, the cursor is
between the chair program keys 1 and 2.
2. When you pick up an instrument or activate any of
the instruments, the cursor jumps to a position between the spray and chip blower keys in the first in-
strument block.
If the cursor is in the quick setting key column at the
very left, it also jumps to a position between the
spray and chip blower keys in the instrument block
when you swing the foot switch to the right (D).
3. After you have placed the instrument in its holder,
the cursor jumps back to its position between the
chair program keys S and 0.
Activate the desired chair program 1 or 2. The chair
moves to the programmed treatment position.
Move the dentist element by hand to the desired treat-
1
ment position. Use the handle (1) on the dentist element
to do this.
The dentist element can be fixed by the service engineer
at a height which is convenient for you.
26D 3382.201.01.13.02
59 06 313 D 3382
Page 27
Sirona Dental Systems GmbH7 Program selection
Operating Instructions C4
+
7.1 Safety
7 Program selection
7.1Safety
CAUTION
The patient’s arms and legs must be resting on the chair
upholstery during the program run!
CAUTION
Make sure that no obstacles (e.g. window wings, drawers, devices, …) extend into the movement range.
A
BACKREST
10:04:27
Safety stop
A built-in safety circuit stops the chair movement in the
following situations:
•The toeboard collides with an obstacle.
•The backrest collides with an obstacle.
•The motorized headrest (if available) collides with an
obstacle
•The swivelable cuspidor is swiveled in during chair
movement.
•The assistant element is pushed upward or downward by an obstacle.
At the same time, a double beep signal sounds.
The chair then moves automatically a little upward until
the obstacle is cleared (except for the swivelable cuspidor and assistant element).
With instruments in place, the last activated safety
switch is indicated in plain text on the display:
•TILTING PART for motorized headrest
•BACKREST
•TOEBOARD
•ASSISTANT ELEMENT (is also displayed when the
cuspidor safety switch is activated)
59 06 313 D 3382
D 3382.201.01.13.02
The message CUSPIDOR appears after 10 seconds if
the automatic return movement of the cuspidor to its
position is inhibited.
27
Page 28
7 Program selection Sirona Dental Systems GmbH
7.2 Program selectionOperating Instructions C4
7.2Program selection
Keys for selecting the programmed patient positions.
+
SS
11
00
22
00
SS
With swivelable cuspidor:
Before the chair moves to the selected program position,
the swiveled-in cuspidor automatically returns to its
starting position.
Factory settings:
The operation light switches OFF automatically before
the chair moves to the entry/exit position 0 or the mouth
rinsing position S.
The operation light switches ON automatically after the
chair has reached patient position 1 or 2.
Changing the factory settings:
When programming the patient positions you can preset
whether the operation light should switch OFF or ON in
the different chair programs 0, S, 1 or 2 (see page 36).
Program key 0
Intended for the entry/exit position.
Program key S with memory function.
In the S program, you can program the tumbler filling
and / or the cuspidor flushing functions (see Mode dia-
log 11.9 and 11.10).
When you press this key a second time, the treatment
center returns to its previous position (memory function).
Example: You have programmed this key for the mouth
rinsing position. When you press this key, the treatment
center moves to the programmed position.
Memory function: When you press this key a second
time, the treatment center returns to its previous position.
28D 3382.201.01.13.02
59 06 313 D 3382
Page 29
Sirona Dental Systems GmbH7 Program selection
Operating Instructions C4
+
7.3 MOVEMENT STOP!
7.3MOVEMENT STOP!
You can stop the movement of the chair to a programmed position as follows:
•By pressing one of the chair-related keys on the
dentist or assistant element.
•By actuating the 4-way foot control.
•By pushing the 4-way switches on a motorized
headrest in any direction.
All movements of the treatment center are stopped
immediately.
At the same time, a double beep signal sounds.
Please note that when you press the program key of
the program which is just running a second time, the
movement will NOT stop!
59 06 313 D 3382
D 3382.201.01.13.02
29
Page 30
8 Patient positions Sirona Dental Systems GmbH
8.1 Flat headrestOperating Instructions C4
8 Patient positions
8.1Flat headrest
The headrest must be adjusted manually.
+
2
1
8.2Articulated headrest
To do this, take the load off the backrest and pull out or
push in the headrest using the handle (1).
The head pad (2) is held magnetically.
DANGER
A strong magnet is integrated in the underside of the
head cushion (2) which could affect a nearby cardiac
pacemaker.
Therefore, please avoid bringing the magnet in close
proximity of a cardiac pacemaker of a patient, user or
technical personnel.
In addition, direct contact of the head cushion with
magnetic cards can lead to a data loss.
The headrest must be adjusted manually.
To do this, take the load off the backrest and pull out or
push in the headrest using the handle (1).
1
2
30D 3382.201.01.13.02
Tilt the headrest backward using the locking button (2).
DANGER
Tilt the headrest forward by simply pulling it out.
59 06 313 D 3382
Page 31
Sirona Dental Systems GmbH8 Patient positions
A
Operating Instructions C4
+
8.3 MultiMotion headrest
8.3MultiMotion headrest
Adjustment to patient's height
The headrest can be adjusted to the patient's height by
pulling out or pushing in the headrest extension. This
adjustment is usually required at the beginning of treatment!
i
NOTICE
Before positioning the patient on the headrest, make
sure that the head support has been optimally adjusted
to the patient's height. This simplifies all subsequent
work with the MultiMotion considerably, since readjustment to the patient's height can thus be omitted when
changing over from mandibular to maxillary treatment.
The MultiMotion enables you to adjust the patient's head
in a way that optimally supports viewing of areas of the
mouth which are difficult to access.
Adjustment of hyperextension
Mandibular position
The mandibular position can be set by pulling the operating handle (A). The anatomical movement of the
arched extension keeps the patient's head in the support.
Pull the headrest out of the guide by pulling
handle A.
i
NOTICE
You can reduce the adjustment noise by pressing
release A.
59 06 313 D 3382
D 3382.201.01.13.02
31
Page 32
8 Patient positions Sirona Dental Systems GmbH
A
B
B
8.3 MultiMotion headrestOperating Instructions C4
Maxillary position
Take the load off the headrest by lifting it slightly.
Press release A on the handle.
Let the headrest slide down slowly.
Let go of the release when the headrest reaches the
desired position.
+
Rotation and tilt
The MultiMotion enables rotation of the patient's head
about the longitudinal axis of his body as well as lateral
tilting of his head.
The MultiMotion is set to the desired position as follows:
1. Press and hold one or both of the two side control
elements B.
The rotation and tilt adjustment locks of the headrest
are now released.
2. Set the headrest to the desired position.
3. Let go of the control elements B.
The supporting surface is fixed in place.
Make sure that the headrest is indeed fixed in place
after you let go of the control elements!
32D 3382.201.01.13.02
59 06 313 D 3382
Page 33
Sirona Dental Systems GmbH8 Patient positions
C
E
D
F
Operating Instructions C4
+
8.3 MultiMotion headrest
Removing the MultiMotion
For certain treatments (e.g. of children) it may be expedient to remove the MultiMotion completely in order to
obtain better access to the patient. The patient's head
will then be supported on the contact surface of the
headrest extension F.
To remove the MultiMotion, proceed as follows:
1. Press release button C.
2. Pull the complete headrest out of the headrest ex-
tension.
3. Cover the opening of the headrest extension with
cover cap E.
4. If a children's head pad D is available, place it on
the headrest extension. The head pad is held magnetically.
i
NOTICE
The children's head pad (D) is not included in the scope
of supply, however, is available from your dental dealer.
DANGER
The children's head pad (D) contains a strong magnet on its bottom side. This magnet could affect any
cardiac pacemaker located nearby.
Therefore, do not allow patients, users and technical
personnel with a cardiac pacemaker to be located
near this magnet.
Furthermore, direct contact of the head pad with
magnetic cards can lead to data loss.
i
NOTICE
Deposit the removed MultiMotion in a safe place where it
cannot fall onto the floor.
Inserting the MultiMotion
To insert the MultiMotion, proceed as follows:
Check the guide of the removed headrest for con-
tamination. Remove any contamination.
Remove the cover cap E.
DANGER
Reinsert the MultiMotion in the guide from above until it
audibly locks in place.
Pull on the headrest again to make sure that it is
locked securely in place.
59 06 313 D 3382
D 3382.201.01.13.02
33
Page 34
8 Patient positions Sirona Dental Systems GmbH
8.4 Adjusting the patient positions manuallyOperating Instructions C4
8.4Adjusting the patient positions manually
You can start the motorized chair movements manually
with the keys in the right block on the dentist element
control panel.
Backrest tilt
You can tilt the backrest also with the 4-way foot control
on the chair base.
Lever to the left:
The backrest tilts backward
Lever to the right:
The backrest tilts forward
Height adjustment
You can adjust the height also with the 4-way foot control
on the chair base.
+
Lever up:
The chair moves up
Lever down:
The chair moves down
34D 3382.201.01.13.02
59 06 313 D 3382
Page 35
Sirona Dental Systems GmbH8 Patient positions
Operating Instructions C4
+
8.5 Motorized headrest
8.5Motorized headrest
Moving the headrest in/out
Move the headrest out/in by pushing the upper 4-way
switch at the back of the headrest up/down.
Tilting the headrest
Tilt the headrest backward/forward by pushing the upper
4-way switch at the back of the headrest to the left/right.
Backrest tilt
You can tilt the backrest with the lower 4-way switch at
the back of the headrest.
•Lever to the left:
The backrest tilts backward.
•Lever to the right:
The backrest tilts forward.
Height adjustment
You can adjust the height with the lower 4-way switch at
the back of the headrest.
The four factory-set programs can be reprogrammed individually by user A and user B.
SS
00
•Programs 1 and 2
•Entry/exit program 0
•Mouth rinsing program S
+
11
22
1. Move the chair to the desired treatment position
pressing the different setting keys (see page 30).
2. To make the SIROLUX FANTASTIC operation light
switch on or remain off when the chair reaches the
programmed treatment position, you must switch the
lamp ON or OFF now. This setting is then also programmed.
CAUTION
In order to avoid damage to the dentist and assistant elements and to the chair upholstery, make sure that these
elements do not protrude into the movement range of the
chair.
3. To save the program settings, press the desired program key approx. 3 seconds until an audible signal
sounds and the LED of the corresponding key lights
up.
Programming is completed now.
i
NOTICE
Programming is not possible by activating the program
keys with the foot switch (cursor control). Thus programming errors are avoided.
by
36D 3382.201.01.13.02
59 06 313 D 3382
Page 37
Sirona Dental Systems GmbH9 Dentist element and instrument functions
2
k
g
m
a
x
.
2
k
g
m
a
x
.
2
k
g
m
a
x
.
2
k
g
m
a
x
.
Operating Instructions C4
+
9.1 Dentist element
9 Dentist element and instrument
functions
9.1Dentist element
Maximum load capacity
The maximum load of the dentist element is 2kg
(4.4 lbs).
Dentist element C4+ Cart
2
2
m
m
a
a
k
k
.
.
x
x
g
g
4.4 lbs
4.4 lbs
4.4 lbs
m
.
a
x
2
g
k
4.4 lbs
m
.
a
x
2
g
k
The maximum load of the dentist element C4+Cart is 2kg
(4.4 lbs).
CAUTION
Risk of toppling over! Move cart slowly and only
on even floors without any obstacles!
CAUTION
Take care to distribute any additional loads evenly
on the dentist element!
Dentist element C4+ Cabinet
The maximum load of the dentist element C4+ Cabinet is
2kg (4.4 lbs).
59 06 313 D 3382
D 3382.201.01.13.02
37
Page 38
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.2 General instrument functionsOperating Instructions C4
The dentist element C4+ Cabinet lets you save on space
by swiveling it under the work surface after use.
CAUTION
When stowing the cabinet away, please take care that
nothing is crushed and that no objects lying on the dentist element are knocked off.
CAUTION
When moving the chair (via program or manually), make
sure that the patient or the patient chair does not under
any circumstances come into contact with the support
arm or the dentist element.
+
9.2General instrument functions
3
C
Spray
You can switch the spray cooling ON/OFF by pressing
the spray key on the dentist element control panel. If
spray is preselected, the green LED in the key is illuminated.
+
With the C
you can switch the spray for the picked-up instrument
ON/OFF by stepping on the left foot switch button (3).
With the C foot switch
the cursor jumps directly between the spray and chip
blower keys when you pick up an instrument. You can
switch the spray cooling ON/OFF by swinging the foot
switch to the left (C).
foot switch
38D 3382.201.01.13.02
59 06 313 D 3382
Page 39
Sirona Dental Systems GmbH9 Dentist element and instrument functions
A
%
Operating Instructions C4
+
9.2 General instrument functions
Spray amount
The spray amount is preset at the factory. However, it
can be adjusted by means of the control valve at the bottom front part of the dentist element.
To make the adjustment loosen the ring (5) counterclockwise, adjust the spray by turning the screw (6) and
5
screw the ring (5) tight again.
This setting is then valid for all burr drives.
6
Chip blower
C+ foot switch
4
With a C
picked-up instrument by stepping on the right foot switch
button (4). As long as you keep the button pressed, an
air jet escapes from the nozzle of the burr instrument.
+
foot switch, you activate the chip blower for a
INSTR. LIGHT
C foot switch
With a C foot switch, the cursor jumps directly between
D
the spray and chip blower keys when you pick up an
instrument. When you swing the foot switch to the right
(D), an air jet escapes from the nozzle of the burr instrument for the duration of actuation. When the chip blower
is active, the green LED in the key is illuminated.
Instrument light ON / OFF
With this key on the dentist element control panel, you
preset the instrument light for the picked-up instrument.
If light ON is preset, the green LED in the key is illuminated.
When you press the key for more than 3 seconds, the
INSTR. LIGHT settings dialog is displayed. Here you
can set a light intensity between 60% and 100% with the
– / + keys.
i
NOTICE
If more than one instrument is picked up, only the instrument which was picked up first is operative.
59 06 313 D 3382
D 3382.201.01.13.02
39
Page 40
9 Dentist element and instrument functions Sirona Dental Systems GmbH
A
RPM
9.3 Electric motorOperating Instructions C4
9.3Electric motor
Speed setting
When you pick up an electric motor, the programmed
speed of the motor in RPM (revolutions per minute)
appears on the display.
You can set the speed with the quick setting keys on the
left side of the dentist element control panel.
If the cursor is in the intensity setting block, you can set
the speed also with the foot switch using the forward/
40
20
100
50
backward cursor.
When you press the corresponding key briefly (< 0.5s)
then the value in RPM indicated next to the relevant key
appears on the display:
Key1–200 rpm
Key50–20000 rpm
Key100–40000 rpm
+
0.2
rpm x 1000
1
When you set the speed with the cursor, the displayed
value is incremented/decremented in steps of 10000
when you actuate the foot switch briefly (< 1s) forward or
backward.
At the values of 200, 20000 and 40000, the green LED
in the corresponding key lights up.
When you press and hold down a key for > 0.5s, you
adjust the speed in predefined increments:
Increments of200:from200– 2000 rpm
Increments of400:from2000– 10000 rpm
Increments of1000: from10000 – 40000 rpm
When you press a key whose value is >
on the display for > 0.5s, the speed is increased.
When you press a key whose value is less than the value
shown on the display for > 0.5s, the speed is reduced.
When you actuate the foot switch forward or backward
for > 0.5s, the displayed speed value is increased or
reduced in predefined increments.
When you set intermediate values, the green key LED
does not light up.
the value shown
Counterclockwise rotation key
Counterclockwise rotation key, only for electric
motor.
When this key is activated the green LED in the key is
iluminated.
40D 3382.201.01.13.02
59 06 313 D 3382
Page 41
Sirona Dental Systems GmbH9 Dentist element and instrument functions
A
Operating Instructions C4
+
After you start the electric motor with the foot switch, an
audible warning signal sounds 6 times.
You can activate counterclockwise rotation of the electric motor also with the foot switch:
•4-way foot control plate to the left – counterclockwise
rotation
•4-way foot control plate to the right – clockwise rotation
The corresponding motor must have been picked up
before.
9.4 Highspeed handpiece
9.4Highspeed handpiece
A
9.5SIROSONIC L scaler
bar
PSI
Starting the highspeed handpiece
When you pick up the highspeed handpiece, 0 bar or PSI
(according to the presetting in Mode dialog 10.13)
appears on the display. When you start or operate the
highspeed handpiece, the current air pressure in bar or
PSI always appears on the display.
Tip protector
The tip protector is used as a tool for screwing in instrument tips.
CAUTION
While the handpiece is in its holder, the tip protector must
remain fitted to prevent injuries.
Intensity setting
When you pick up the SIROSONIC L handpiece, the
programmed intensity value appears on the display.
59 06 313 D 3382
D 3382.201.01.13.02
41
Page 42
9 Dentist element and instrument functions Sirona Dental Systems GmbH
A
9.5 SIROSONIC L scalerOperating Instructions C4
You set the intensity in increments of 1 / 50 / 100 by
pressing the quick setting keys briefly (< 0.5s).
40
100
If the cursor is in the intensity setting block, you can set
the intensity also with the foot switch using the forward/
backward cursor.
+
20
0.2
rpm x 1000
50
1
When you set the intensity with the cursor, the displayed
intensity value is incremented/decremented in steps of
25 when you actuate the foot switch briefly (< 1s) forward or backward.
At the values of 1, 50 and 100, the green LED in the corresponding key lights up.
When you press and hold down a key for > 0.5s, you
adjust the intensity in predefined increments:
Increments of 1 from 1 – 4
Increments of 5 from 10 – 100
When you press a key whose value is >
on the display for > 0.5s, the intensity is increased.
When you press a key whose value is less than the value
shown on the display for > 0.5s, the intensity is reduced.
When you actuate the foot switch forward or backward
for > 0.5s, the displayed intensity value is increased or
reduced in predefined increments.
When you set intermediate values, the green key LED
does not light up.
The cooling water flow is always switched on automatically.
the value shown
Endodontics setting
An intensity range of 1 – 4 is provided for endodontics
mode. It can be adjusted in increments of 1.
ENDO
The message ENDO appears on the display.
For safety reasons, we recommend that you program
the set value for endodontics. Otherwise, the previously
programmed value becomes active again when the
instrument is placed back in its holder in the meantime.
CAUTION
When you exceed the value of 4 in endodontic therapy,
temperature damage and needle breakages can occur.
42D 3382.201.01.13.02
59 06 313 D 3382
Page 43
Sirona Dental Systems GmbH9 Dentist element and instrument functions
A
A
RPM
Operating Instructions C4
+
9.6 SPRAYVIT
9.6SPRAYVIT
If the SPRAYVIT is the only instrument that has been
picked up, you may switch the instrument light on or off.
You can also set the brightness.
The instrument light switches on when you activate the
SPRAYVIT (if preset).
The instrument light switches off after a time lag of 10s
when the SPRAYVIT is no longer activated.
When you place the SPRAYVIT back in its holder, the
instrument light switches off immediately.
If the SPRAYVIT is activated together with another
instrument, the SPRAYVIT light is not switched on.
DANGER
After changing hoses, press the water button of the
SPRAYVIT repeatedly until water comes out! Only
then can you begin treatment.
9.7Saving the instrument programs
40
100
A/B
The factory-set instrument programs can be changed
individually by user A and user B.
Select the corresponding user, i.e. A or B, pick up an
instrument and set it according to your wishes.
Example for electric motor:
•40,000 rpm
•CW rotation
•with cooling spray
•with instrument light
To save the settings, press key A/B for some time
(approx. 3 s). An audible signal sounds. These settings
are then activated whenever the instrument is picked up.
You cannot switch between the two users after picking up an instrument.
i
NOTICE
If counterclockwise rotation was programmed, the motor
is reset to clockwise rotation after you switch the treatment center off and back on.
59 06 313 D 3382
D 3382.201.01.13.02
43
Page 44
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.8 Placing treatment instruments in their holdersOperating Instructions C4
9.8Placing treatment instruments in their holders
A ball stopper for an unoccupied instrument holder
comes with the dentist element.
Plug the ball stopper (1) into the unoccupied instrument
holder to prevent a treatment instrument from being
placed inadvertently in this holder.
To block further unoccupied instrument holders, you can
order more ball stoppers (REF 58 99 575).
1
+
44D 3382.201.01.13.02
59 06 313 D 3382
Page 45
Sirona Dental Systems GmbH9 Dentist element and instrument functions
Operating Instructions C4
+
9.9 Surgery with saline solution (NaCl)
9.9Surgery with saline solution (NaCl)
Preparation of the unit
•Hook the NaCl bottle (1) in place.
•Attach the peristaltic pump hose set (2).
9
1
8
7
2
6
5
4
3
•Push the short end of the hose (3) with the cannula
through the stopper and into the NaCl bottle as illustrated.
•The regulator in the hose clip (4) must be in the topmost position (completely open).
•Run the long hose (5) along the corresponding motor hose up to the contra-angle handpiece and fasten it with the clips (6).
•Fit the coupling (7) onto the hose and connect it with
the thin silicone hose (8).
•Connect the spray clip (9) with the thin hose and attach it to the contra-angle handpiece.
i
NOTICE
The one-way NaCl pump must be secured with a cap after it is attached to the drive (10). Only then is proper operation of the pump guaranteed. The safety cap is supplied with every one-way pump.
10
Pump hose set
The peristaltic pump hose set (2) is a disposable arti-
cle and can be purchased in packs of 10 pcs. as consumable material under article number F 58707 directly
from the manufacturer.
2
A
RPM
NACL
Ordering address:Satelec
Industriestr. 9
D-40822 Mettmann, Germany
i
NOTICE
How to preset a certain instrument holder for saline solution is described under “Basic treatment center settings
in the Mode dialog” 11.2 on page 57.
CAUTION
The pump flow rate must be at least 70ml/min for reasons of safe cooling and to prevent a pressure rise in the
hose (risk of bursting). This is not always given when using third-party ultrasonic tips.
Switching the NaCl pump on / off
After picking up the respective handpiece, you can
switch the NaCl pump ON/OFF by pressing the spray
key (or by stepping on the left button on the C+ foot
switch or by swinging the pedal plate to the left on the C
foot switch).
59 06 313 D 3382
D 3382.201.01.13.02
45
Page 46
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.10 Satelec Mini L.E.D. curing lightOperating Instructions C4
When the green LED of the key lights up, the NaCl pump
is ON.
NACL appears on the display in the instrument dialog of
the instrument to which the NaCl pump has been
assigned.
9.10Satelec Mini L.E.D. curing light
Safety information
CAUTION
Always use the glare shield!
The light beams emitted by this instrument may be hazardous and may never be directed to the eyes of any person, even if this person wears goggles with protective
lenses. The light may be directed exclusively to the treatment area in the oral cavity.
+
right
DO NOT STARE INTO BEAM OR VIEW
L.E.D. RADIATION
DIRECTLY WITH
OPTICAL INSTRUMENTS
CLASS 2M LASER PRODUCT
P < 1 mW, λ = 420nm - 480nm
IEC 60825-1 Ed. 1.2
!
wrong
!
L.E.D. beam
exit aperture
CAUTION
Do not stare into the beam path with the glass rod removed (Class 2M laser product).
Viewing the L.E.D. beam exit aperture with certain optical instruments (e.g. a magnifying lens) within a distance
of 100mm may constitute an eye hazard.
46D 3382.201.01.13.02
59 06 313 D 3382
Page 47
Sirona Dental Systems GmbH9 Dentist element and instrument functions
Operating Instructions C4
+
9.10 Satelec Mini L.E.D. curing light
Additional safety information
Any condensation occurring in the handpiece of the Mini
L.E.D. may cause impairments (e.g. fogging of the
L.E.D.). If the handpiece is moved from a cool environment to a warm room, always wait for it to reach room
temperature before putting it into operation.
Curing lights must not be used on persons who are suffering from or have in the past been afflicted by photobiological reactions (including solar urticaria and erythropoetic porphyria). Nor should they be used on persons
currently being treated with any medicine which
increases one’s sensitivity to light (including methoxsalene and chlorotetracycline).
Any persons who have in the past suffered from retinal
or eye lens disorders or have undergone eye surgery,
especially for gray cataracts, must consult their medical
eye specialist before using the light or undergoing treatment with the Mini L.E.D. Caution is advisable even if the
patient gives his or her consent, since the light intensity
can cause accidents. It is especially advisable to always
wear appropriate protective goggles (UV filter).
Installation
1. Connect the supply cable to the treatment center.
4
2
3
1
2. Screw the handpiece on.
3. Attach the sterilized light guide to the handpiece.
Make sure to insert the light guide correctly.
The light guide engages with a “click”.
4. Slip the glare shield onto the light guide. It protects
your eyes against reflected light.
59 06 313 D 3382
D 3382.201.01.13.02
47
Page 48
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.10 Satelec Mini L.E.D. curing lightOperating Instructions C4
Installation options
•in the dentist element, holder 4 or 5
•in the assistant element, holder 1
+
1
7
4
6
5
Handpiece and accessories
1Handpiece
2Light guide
3Supply cable
3
4Status LEDs
5Glare shield
6ON/OFF switch
7Function button
2
Operating modes
The Mini L.E.D. has three different modes. To select the
desired menu, press the function button.
4.1 Status LED
The status LED signals the following states:
4.1
4.2
4.3
4.4
ColorMode
greenNormal operation
red, flashingOverheat protection
4.2 Quick curing mode
In this mode the Mini L.E.D. operates at full power for 10
seconds.
LED 4.2 shows your selection.
0510
Sec.
In this mode, the light intensity output is approximately
as follows:
59 06 313 D 3382
48D 3382.201.01.13.02
Page 49
Sirona Dental Systems GmbH9 Dentist element and instrument functions
Operating Instructions C4
+
9.10 Satelec Mini L.E.D. curing light
•1100mW/cm² (± 10%) with the standard light guide,
dia. 7.5 mm.
•2000mW/cm² (± 10%) with the optional booster light
guide, dia. 5.5mm.
4.3 Pulsed curing mode
This mode is selected by pressing the function button.
The light operates at full power in “pulsed” mode; it emits
012.5
Sec.
radiation in 10 consecutive light pulses of 1s each.
Between the individual pulses there is a 250 ms break.
LED 4.3 shows your selection.
4.4 “Soft start” mode
This mode is selected by pressing the function button.
The “soft start” mode provides:
05101520
Sec.
•A 10-second “soft start” from 0 to 1100 mW/cm², or
from 0 to 2000 mW/cm² with the (optional) “booster
light guide”, dia. 5.5 mm.
•Full power during 10 seconds.
LED 4.4 shows your selection.
Handling
After you switch on the treatment center, the Mini L.E.D.
is by default in the quick curing mode (4.2).
If you change the operating mode of the Mini L.E.D.
while the treatment center is in operation, the new mode
remains valid even after placing the Mini L.E.D. in its
holder.
After you have selected a mode, the Mini L.E.D. is ready
for operation.
Position the light guide as close as possible to the composite material surface you want to photopolymerize.
Make sure that the light guide never touches the material
to be cured, as this may cause damage to the light guide
and reduce its effectiveness.
•Never work without the glare shield (5).
•Never look into the light which is reflected by the
tooth surface.
59 06 313 D 3382
D 3382.201.01.13.02
49
Page 50
9 Dentist element and instrument functions Sirona Dental Systems GmbH
9.10 Satelec Mini L.E.D. curing lightOperating Instructions C4
To start the polymerization cycle, briefly press the ON/
OFF switch (6). This is confirmed by an audible signal.
+
6
In each mode, an audible signal sounds every 5 seconds.
The end of the cycle is also indicated by an audible signal. You may interrupt the cycle at any time though by
lightly pressing the ON/OFF switch.
• The tumbler must always be placed in the center of the
area provided for it.
• If an empty tumbler is standing below the tumbler filler
when the treatment center is switched on with the main
switch, the tumbler will not be filled automatically. To activate automatic tumbler filling, remove the tumbler briefly and then put it back again.
• The Mode dialog “CUP FILL ON S YES – NO” must be
set in such a way that “NO” flashes.
• Hydrocolloid connection is not possible if tumbler filling
with automatic sensor control is installed.
When you press this key briefly, the flushing function is
switched ON and the cuspidor is flushed for a preset
period of time.
When you press the key again while the cuspidor is
being flushed, flushing stops.
Only for cuspidor flushing key of the dentist element:
Pressing this key for > 3 seconds displays the settings
dialog
BOWL FLUSH TIMER
You can set a flushing time between 20 and 420 seconds (in steps) with the –
10.5 Purge / Auto Purge function (purging the water paths)
Auto Purge (APURGE) function
The Auto Purge function offers possibility of automated
purging of all water carrying instruments in the dentist
element, of the assistant Sprayvit and of the tumbler filling unit.
All of the instruments inserted in the sanitation bowl will
be purged when APURGE is activated. If the instruments remain in the sanitation bowl after the treatment
center is switched off, the purging process will automatically be started again the next time the treatment center
is switched on.
We recommend the APURGE function in the following
cases:
•APURGE function before beginning and at the end
of the work day:
A complete purge cycle is performed following activation. The instruments remain in the sanitation
bowl and the treatment center can be switched off.
On the next day, the APURGE function is automatically performed again on all of the instruments remaining in the sanitation bowl as well as on the tumbler filling unit immediately after the treatment center
is switched on.
Then you can prepare the treatment center for daily
practice operation.
SELECT MODE
PURGE / APURGE / SAN
1
•APURGE function after longer periods of disuse:
A complete purge cycle is performed following activation. The instruments remain in the sanitation
bowl and the treatment center can be switched off.
Now you can switch on the treatment center briefly
every day and then switch it off again after the purging process.
Setting the APURGE function
Activate the selection dialog box by pressing quick setting key 1 (> 3 s) when all of the instruments are deposited in their holders.
You will be prompted to choose between PURGE/APURGE and SANITATION1.
Select APURGE with the
flash.
– or + key. APURGE must
59 06 313 D 3382
D 3382.201.01.13.02
1. The SANITATION function is described in the instructions
10.5 Purge / Auto Purge function (purging the water paths)Operating Instructions C4
If you press quick setting key 1 briefly (< 3 s), the following prompt will appear if all instruments are in their holders:
+
SELECT ITEMS
FOR PURGING
1
wrong
SELECT ITEMS FOR PURGING
Remove the instruments to be purged from the dentist
and assistant elements and insert them in the adapters
of the sanitation bowl as far as they will go.
CAUTION
Be careful not to snap off the instrument hoses when inserting the instruments.
The maximum water flow must be set on the instruments.
Press quick setting key 1 again (< 3 s), the instruments
are then purged one after the other.
The purging time in seconds set previously in the Mode
dialog (11.14) runs down for each instrument individually.
An audible signal sounds after the last instrument has
been flushed.
You now can continue in either of the following ways:
•
Leave the instruments in the sanitation bowl:
right
The treatment center can then be switched off.
The instruments will automatically be purged as
soon as the treatment center is switched on
again.
The APURGE function remains activated.
•
The instruments are returned to their holders:
The treatment center can be prepared for daily
practice operation.
The APURGE function is completed.
i
NOTICE
When removing the sanitation bowl, please notice that it
is filled with water.
56D 3382.201.01.13.02
59 06 313 D 3382
Page 57
Sirona Dental Systems GmbH11 Basic treatment center settings in the Mode dialog
A
NACL – PUMP
OFF
A
NACL – PUMP
ON FOR HANGER 2
Operating Instructions C4
+
11.1 Mode key
11 Basic treatment center settings in the
Mode dialog
11.1Mode key
When you press the Mode key, you can access and
change various basic treatment center settings in a certain order.
They usually apply both to user A and B.
The corresponding display disappears automatically
after approx. 10 seconds. To exit the Mode dialog more
quickly, press the Mode key for > 3 seconds.
The flashing text is selected and saved when you exit the
Mode dialog.
You can select the desired setting by pressing the
+ key.
Key lock
The key lock can be activated and deactivated again by
pressing the - and + keys simultaneously.
The key lock affects the entire control panel of the dentist
element
i
NOTICE
The key lock can be deactivated only as described above. Switching the treatment center off and on again has
no effect on this state.
11.2Mode: NACL PUMP OFF or ON for instrument holder ..
Saline solution for the desired burr instrument
or the SIROSONIC L scaler (instrument holder
2 − 5)
NACL – PUMP
OFF
Here you can assign the NaCl pump to a certain instrument using the
– / + keys.
– or
NACL – PUMP
ON FOR HANGER 2
59 06 313 D 3382
D 3382.201.01.13.02
For preparation of the NaCl pump, see page 45 .
i
NOTICE
NACL appears on the display in the instrument dialog of
the instrument to which the NaCl pump has been assigned.
i
NOTICE
If you want to assign the NaCl pump to the SIROSONIC L
scaler, always select holder 5, regardless of the scaler’s
position in the instrument rack.
57
Page 58
11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
A
MAINTENANCE
360 8
A
SERVICE ONLY
START – STOP
A
FC CONTROL MODE
ON/OFF – VARIABLE
A
TIME – HOURS : MIN
11 : 21
11.3 Mode: MAINTENANCEOperating Instructions C4
11.3Mode: MAINTENANCE
Display of next maintenance date
This dialog displays the number of days until the next
maintenance date. You should have maintenance perfor-
MAINTENANCE
360 8
med by your service engineer at least once annually.
The instrument running time is indicated in the lower
right-hand corner of the display (for service engineers
only).
As soon as the time remaining until the next maintenance date is less than 30 days, the message is automatically displayed each time the unit is switched on.
The display message then disappears as soon as an
instrument is removed from its holder.
11.4Mode: SERVICE ONLY START − STOP
For service engineers only!
+
SERVICE ONLY
START – STOP
11.5Mode: FC CONTROL MODE ON/OFF − VARIABLE
Direct starter or speed controller foot switch
Presetting of the foot switch control mode for the instrument start.
FC CONTROL MODE
ON/OFF – VARIABLE
With direct starter (ON/OFF) presetting, the bar display always jumps immediately to the maximum value
when you step on the foot switch. You are working with
the displayed maximum value.
With speed controller foot switch (VARIABLE) pre-setting, the bar display depends on the pedal position.
You are working in a range between 0 and the displayed
maximum value.
This setting can be chosen individually for user A or
B!
11.6Mode: TIME − HOURS : MIN
Setting the time
TIME – HOURS : MIN
11 : 21
58D 3382.201.01.13.02
Here you can set the time of day in hours and minutes.
The hours display flashes first and can be adjusted with
the
– / + keys.
59 06 313 D 3382
Page 59
Sirona Dental Systems GmbH11 Basic treatment center settings in the Mode dialog
A
DAY : MON : YEAR
21 : JUN : 00
A
FOOT – CONTROL
NORMAL – EXTENDED
A
BOWL FLUSH ON S
YES – NO
A
CUP FILL ON S
YES – NO
Operating Instructions C4
+
When you press the Mode key a second time, the
minute display starts flashing and can be adjusted in the
same way.
11.7 Mode: DAY : MON : YEAR
11.7Mode: DAY : MON : YEAR
Setting the date
Here you can set the year, month and day.
DAY : MON : YEAR
21 : JUN : 00
The setting is made in the same way as for Mode
TIME − HOURS : MIN in the order of year, month, day.
The date is not displayed during operation of the unit.
It is required only for internal treatment center functions,
e.g. for service information.
11.8Mode: FOOT – CONTROL NORMAL − EXTENDED
Switching the cursor on/off
If you want to control the treatment center only manually,
you can switch the cursor control off.
FOOT – CONTROL
NORMAL – EXTENDED
The then differing functions of the relevant foot switch
are described in the “Foot switch” section.
This setting can be chosen individually for user A or
B!
11.9Mode: BOWL FLUSH ON S YES − NO
Cuspidor flushing with chair program S
If you want the cuspidor flushing to switch on in program
S of the treatment chair.
BOWL FLUSH ON S
YES – NO
11.10Mode: CUP FILL ON S YES − NO
Tumbler filling function with chair program S
If you want the tumbler to be filled in program S of the
treatment chair.
CUP FILL ON S
YES – NO
59 06 313 D 3382
D 3382.201.01.13.02
59
Page 60
11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
11.11 Mode: WHITE BALANCE (only with an integrated SIROCAM 3)Operating Instructions C4
11.11Mode: WHITE BALANCE (only with an integrated SIROCAM 3)
White balance for the SIROCAM 3
A
TAKE VCAM
This Mode dialog is only displayed in connection with an
integrated SIROCAM 3.
The camera handpiece contains data making it possible
to restore the factory-set color definitions regardless of
the dental treatment center where the camera is used.
Thus the camera can be used at different dental treatment centers.
+
A
WHITE BALANCE
HOME – USER – EXIT
For adjustments, e.g. if you wish to set individual colors,
you must perform a while balance.
Select the “TAKE VCAM” display in the Mode dialog.
When you pick up the camera, the dialog for performing
the white balance is displayed.
There are 3 options available:
USER - individual white balance
HOME - factory settings (can be restored at any
time)
EXIT- to exit the WHITE BALANCE dialog.
The display shows the current camera status, i.e., if an
individual white balance was performed the last time,
USER is displayed in bold type.
If the factory settings are valid, HOME is displayed in
bold type.
The selected item is flashing.
Select USER with the
Set the slide switch (2) of the camera to its middle position (overview exposure).
– / + keys. USER starts flashing.
2
A
Then hold the camera so that its lens is located 2–5 cm
above a color pattern sheet of your own choice (A) in the
live mode.
The screen must be completely filled by the focused
color area.
•Press the Mode key to start the white balance (takes
several seconds).
If the white balance has been completed successfully,
USER is displayed in bold type and an audible confirmation signal sounds.
If the white balance has not been successful, USER will
not be displayed in bold type. You may repeat the white
balance as often as desired.
Different color patterns will produce different results with
respect to the color settings. For example, a color area
with a high amount of blue will reduce the amount of blue
color in the video image.
59 06 313 D 3382
60D 3382.201.01.13.02
Page 61
Sirona Dental Systems GmbH11 Basic treatment center settings in the Mode dialog
A
TAKE INSTRUMENT
A
MAX LIGHT VOLT
3.9 V WARNING
Operating Instructions C4
+
11.12 Instrument settings via Mode dialog: REMOVE INSTRUMENT.
If you want to reset the camera to its factory settings,
select HOME with the
pressing the Mode key, HOME is displayed in bold type.
To exit this Mode dialog, select EXIT and press the
Mode key, or put the camera back in place. Automatic
exit from the dialog is disabled as soon as USER or
HOME is selected. This allows you to repeat the white
balance as often as necessary until you obtain the
desired result.
– / + keys. When you confirm by
11.12Instrument settings via Mode dialog: REMOVE INSTRUMENT.
This Mode dialog enables the following settings
• Instrument light voltage
and
• Maximum torque of SL motors
Setting the instrument light voltage
You can change the voltage setting for instruments in
this Mode dialog.
First a message prompting removal of an instrument
appears.
Then the current voltage setting of the (removed) instrument is displayed.
MAX LIGHT VOLT
3.9 V WARNING
Using the -/+ keys, you can adjust the instrument light
voltage from 3.0 to 5.6V in steps of 0.1V.
Starting at 3.9V, WARNING also appears.
After depositing the instrument, please remove and set
the next instrument.
Observe the voltage specified by the instrument
manufacturer!
i
NOTICE
The voltage (3.6V) is already preselected for Sirona instruments.
59 06 313 D 3382
D 3382.201.01.13.02
61
Page 62
11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
A
TAKE INSTRUMENT
A
MAX. TORQUE
YES – NO
A
# FUNCTION
TOGGLE – MOMENTARY
A
PURGE TIME
120 SEC
11.13 Mode: # FUNCTION TOGGLE - MOMENTARYOperating Instructions C4
Setting the torque of the electric motors
This Mode dialog is used to set the automatic controller
action of the SL or SL ISO electric motors. Individual settings are possible for each motor holder.
First a message prompting removal of an instrument
appears.
After removing the instrument, select the setting dialog
by pressing the Mode key again:
MAX TORQUE
MAX. TORQUE
YES – NO
If Yes is flashing:
The motor will attain a higher torque at speeds > 15,000
rpm.
If No is flashing:
The running smoothness of the motor will be higher at
speeds > 15,000 rpm.
i
NOTICE
The new setting becomes effective only after the corresponding motor is removed again.
+
i
NOTICE
11.13Mode: # FUNCTION TOGGLE − MOMENTARY
Free function key # as button or switch
In this mode you can preset whether you want to use the
# FUNCTION
TOGGLE – MOMENTARY
free function key
17).
# as a button or as a switch (see page
11.14Mode: PURGE TIME... SEC
Setting the purge time for flushing the water
paths
(Only for dentist element)
PURGE TIME
120 SEC
You can preset a purging time between 60 and 120 seconds for each instrument with the – /+ keys for the
purge function (see page 54).
62D 3382.201.01.13.02
59 06 313 D 3382
Page 63
Sirona Dental Systems GmbH11 Basic treatment center settings in the Mode dialog
A
PRESSURE
PSI/BAR
A
SPRACHE/LANGUAGE
ENG/D/I/F/E
A
POLYLUX WITH CFS
YES – NO
A
SUCTION WITH CFS
YES – NO
Operating Instructions C4
+
11.15 Mode: PRESSURE PSI / BAR
11.15Mode: PRESSURE PSI / BAR
Indication of pressure in PSI or bar
In this mode, you can select whether you want the driving air pressure of the highspeed handpiece to be dis-
PRESSURE
PSI/BAR
played in PSI or bar.
11.16Mode: SPRACHE / LANGUAGE ENG / D / I / F / E
Language setting
In this mode, you can select the desired display language: ENG=English, D=German, I=Italian, F=French
SPRACHE/LANGUAGE
ENG/D/I/F/E
and E=Spanish.
11.17Mode: POLYLUX WITH CFS YES – NO
This dialog has no function in connection with the Mini
L.E.D. curing light.
Set NO.
POLYLUX WITH CFS
YES – NO
11.18Mode: SUCTION WITH CFS YES – NO
Interrupting or restarting the suction flow with
the 4-way foot control
If YES is flashing, you can interrupt or restart the suc-
SUCTION WITH CFS
YES – NO
tion flow of the large suction hose (not at the saliva ejector hose) by pushing the 4-way foot control on the chair
in any direction. When you put the large suction hose in
its holder while the suction flow is interrupted, the suction flow is automatically restarted when you pick up the
hose again.
59 06 313 D 3382
D 3382.201.01.13.02
63
Page 64
11 Basic treatment center settings in the Mode dialog Sirona Dental Systems GmbH
Presetting a white screen for the SIVISION
monitor
If YES flashes –
WHITESCREEN
YES – NO
The “X-ray image viewer” key switches ON/OFF both
the X-ray image viewer and a WHITE SCREEN on the
SIVISION 3 monitor.
If NO flashes –
The “X-ray image viewer” key only switches the X-ray
image viewer ON and OFF.
If an X-ray image viewer and SIVISION 3 are installed,
always select “NO”.
11.20Mode: A/B VIDEO MODE ON – OFF
Locking the video mode switchover
This mode is only displayed in connection with an integrated SIVISION 3.
A / B VIDEO MODE
ON – OFF
A/B
If ON is flashing –
You can use the “A/B” key to switch between VIDEO
mode and PC mode after picking up the SIVISION 3
camera.
+
11.21Mode: BLOW OUT SPRAY
BLOW OUT SPRAY
YES – NO
If OFF is flashing –
You cannot use the “A/B” key to switch between VIDEO
mode and PC mode for SIVISION 3.
This prevents the selected mode from being changed
inadvertently.
Prevents drop formation
You can activate the instrument blow-out function in this
setting option. This removes any residual water (one to
two drops) by blowing air through the instrument heads.
Just press the
OFF.
The selected item flashes.
i
NOTICE
This function is switched OFF in the factory setting.
i
NOTICE
Sirona recommends activating this function especially
for the C3+ treatment center.
– or + key to switch this function ON or
64D 3382.201.01.13.02
59 06 313 D 3382
Page 65
Sirona Dental Systems GmbH12 Assistant element
Operating Instructions C4
+
12.1 Assistant element
12 Assistant element
12.1Assistant element
The assistant element can be swiveled to the side and to
the front.
To do this grasp the assistant element laterally at the
handpiece rack (1).
The ease of the swivel movement towards the front can
be set with the knob (2).
CAUTION
Make sure that the backrest does not collide with the
assistant element if you have positioned the latter in the
1
movement range of the backrest.
Also make sure that the assistant element is not posi-
2
tioned above the patient when the chair moves upward
(on contact with the patient, a safety switch stops the
movement).
Removing a handpiece rack with closed suction handpiece holders is described in the document “Care and Cleaning by the Practice Team”.
DANGER
The suction hose handpieces contain magnets. In
the case of body contact of the suction hose handpiece in the vicinity of a cardiac pacemaker, this can
have a disturbing effect on the functioning of the cardiac pacemaker.
Therefore avoid bringing the suction hose hand-
1
piece into the vicinity of a cardiac pacemaker.
2
Furthermore, direct contact of the suction hose
handpiece with magnetic cards can lead to data loss.
Suction hose handpieces with integrated magnets
are marked with the following symbol:
59 06 313 D 3382
D 3382.201.01.13.02
65
Page 66
12 Assistant element Sirona Dental Systems GmbH
A
SUCTION WITH CFS
YES – NO
12.2 Suction handpieces on the assistant elementOperating Instructions C4
12.2Suction handpieces on the assistant element
The vacuum system switches on when you pick up a
suction hose from the holder.
CAUTION
Observe the warning and safety information on page 9
(vacuum system).
You can angle the suction handpiece by turning it.
CAUTION
With a wet suction system, never pour water into the cuspidor while the unit is switched off.
CAUTION
If the vacuum system switches off by accident, immediately remove the suction cannula from the patient’s
mouth.
If the unit has no second saliva ejector hose for the surgical cannula, the thick suction hose must be used for
surgical aspiration.
+
SUCTION WITH CFS
YES – NO
To insert a surgical cannula, please attach the adapter
supplied.
If the unit is equipped with a second saliva ejector hose,
the surgical cannula can be inserted directly into the
saliva ejector handpiece.
You can interrupt or restart the suction flow of the large
suction hose also by pushing the 4-way foot control on
the chair in any direction.
You can preset this under Mode 11.18 . When you put
the large suction hose in its holder while the suction flow
is interrupted, the suction flow is automatically restarted
when you pick up the hose again.
66D 3382.201.01.13.02
59 06 313 D 3382
Page 67
Sirona Dental Systems GmbH12 Assistant element
Operating Instructions C4
+
CAUTION
Make sure to hold the suction cannula in such a way that
the cannula opening cannot attach itself to the oral mucous membranes by accident.
i
NOTICE
The factory-set suction power can be adjusted in the water unit by a service engineer.
12.3 SPRAYVIT on the assistant element
12.3SPRAYVIT on the assistant element
If the SPRAYVIT is the only instrument that has been
picked up, you may switch the instrument light on or off.
L
Its brightness is set to an optimum value and cannot be
changed.
The instrument light switches on when you activate the
SPRAYVIT (if preset).
The instrument light switches off after a time lag of 10s
when the SPRAYVIT is no longer activated.
When you place the SPRAYVIT back on the instrument
rack, the instrument light switches off immediately.
DANGER
After changing hoses, press the water button of the
SPRAYVIT repeatedly until water comes out! Only
then can you begin treatment.
12.4Satelec Mini L.E.D. curing light on the assistant element
For safety instructions and operation, see page 46.
The cuspidor on the water unit can be swiveled manually
by approx. 110mm towards the patient chair.
Collision protection
Certain chair movements are locked as long as the cuspidor is swiveled in. When you start a programmed
movement S, 0, 1 or 2, the cuspidor first moves back to
its starting position before the chair moves to the programmed position.
When you press a key for the individual adjustment of
the chair, i.e. height adjustment up/down or backrest tilt backward/forward, the cuspidor also swivels back
first, and if the corresponding key is still pressed, the
chair moves.
You can make individual adjustments of the motorized
headrest even when the cuspidor is swiveled in.
+
If the cuspidor is swiveled in while the chair is moving,
the program is aborted. This is signaled by a double
beep.
13.2Adjusting the water amount for the cuspidor
•Open the flap (1) on the base of the water unit.
•Adjust the amount of water with the valve (2).
2
1
1
68D 3382.201.01.13.02
59 06 313 D 3382
Page 69
Sirona Dental Systems GmbH13 Water unit
Operating Instructions C4
+
13.3 Amalgam rotor
13.3Amalgam rotor
The message AMALG appears on the display when the
amalgam rotor needs replacement, i.e., when having
reached a filling level of 95% (see instructions “Care and Cleaning by the Practice Team”).
A
AMALG
RPM
A continuous sound indicates that the rotor is full and
no longer functioning properly.
An intermittent sound indicates a system error. The
water supply is interrupted. Please contact your service
engineer.
i
NOTICE
The rotor inside the amalgam separator must be replaced at least once a year, even if the AMALG message
does not light up. Otherwise, unpleasant odors can develop.
59 06 313 D 3382
D 3382.201.01.13.02
69
Page 70
13 Water unit Sirona Dental Systems GmbH
13.4 Disinfection system (optional)Operating Instructions C4
13.4Disinfection system (optional)
When the message DESINF flashes on the dentist ele-
ment display during treatment, you should refill 1 liter of
DENTOSEPT P at the latest after the end of the treatment day.
A
DESINF
RPM
When the message flashes, it is not possible to perform
sanitation.
Remove the cover behind the cuspidor and fill in 1 liter
of DENTOSEPT P.
The filler trough should always be kept clean. Always
make sure that no tap water enters the filler trough!
For Canada and Japan only:
1 carton = 6 x 1 liter, REF 59 69 535
Microbiological water test (after an operating pause of
more than 1 week)
Sanitation of the unit
Perform sanitation every 4 weeks. This is the only way
to effectively combat the formation of biofilm in the water
paths (see instructions “Care and Cleaning by the
Practice Team”).
70D 3382.201.01.13.02
59 06 313 D 3382
Page 71
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.1 System overview of SIVISION 3
14 SIVISION 3
14.1System overview of SIVISION 3
SIVISION on the lamp support tube
• Rotatable
•swivelable
Monitor on tray arm
• Rotatable
•swivelable
59 06 313 D 3382
D 3382.201.01.13.02
Monitor on swivel arm
• Rotatable
•swivelable
• Height adjustable
71
Page 72
14 SIVISION 3 Sirona Dental Systems GmbH
14.1 System overview of SIVISION 3Operating Instructions C4
CAUTION
Make sure that the patient does not collide with the monitor during movements of the dental treatment chair
and/or operation of the cuspidor. Swivel the monitor out
of the collision area if necessary.
SIVISION 3 monitor versions
• 15“ Monitor
• with glass pane
+
• 19“ Monitor
• with glass pane
• with loudspeaker
CAUTION
The loudspeaker socket of the monitor may be connected only to a device which complies with IEC 60950 (e.g.
PC) or IEC 60601-1, and under no circumstances e.g. to
a stereo system etc.
72D 3382.201.01.13.02
59 06 313 D 3382
Page 73
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.2 Camera versions
14.2Camera versions
SIROCAM 3
• Integrated 4-view image memory
• 3-fold adjustable focusing range
• Titanium case
SIROCAM C
• Integrated 4-view image memory
• 2-fold adjustable focusing range
• Plastic case
59 06 313 D 3382
D 3382.201.01.13.02
73
Page 74
14 SIVISION 3 Sirona Dental Systems GmbH
14.3 SIROCAM 3 intraoral camera (extra equipment)Operating Instructions C4
14.3SIROCAM 3 intraoral camera (extra equipment)
The SIROCAM 3 intraoral camera combined with SIVISION 3 is a perfect system for effective communication
with the patient.
+
4
3
*
t 1 / t 2
2
1
The camera images cannot be used for diagnosis!
The SIROCAM 3 intraoral camera is a sensitive optical
instrument and must therefore always be handled with
care.
The camera illuminates the tooth areas to be imaged
with 7 white-light LEDs (1).
These are arranged around the entrance prism (2).
The LEDs and the lens are protected by a scratch-proof
antiglare glass cover.
CAUTION
* = Intermittent camera operation:
t1= 1 min on, t2 = 3 min off.
The camera heats up during operation.
To prevent overheating of the camera, the camera must
be put down for a time interval t2 = 3 min after an operating time of t1 = 1 min.
The camera will heat up even when it is placed in its
holder, as one of the states “black screen”, “white
screen” or “quad image” is always active.
The slide switch (3) is used for presetting the depth of
focus.
The camera button (4) is intended for controlling different SIVISION 3 functions.
i
NOTICE
For camera operation and further information, please refer to the operating instructions of the SIROCAM 3.
i
NOTICE
The camera handpiece contains data making it possible
to restore the factory-set color definitions regardless of
the dental treatment center where the camera is used.
Thus the camera can be used at different dental treatment centers. To customize the color settings, you can
perform a white balance as long as the camera is active,
(see Mode dialog, section 11.11).
74D 3382.201.01.13.02
59 06 313 D 3382
Page 75
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
1
+
Hygienic protective covers for SIROCAM 3
Hygienic protective covers are available for the camera.
They can be ordered under REF 59 38 654. Package
size: 500 pcs.
How to use the hygienic protective covers:
1. Hold the hygienic protective cover with its paper side
2. Pull off the upper protective foil with the blue stripe.
14.3 SIROCAM 3 intraoral camera (extra equipment)
face-down and insert the camera with the optical
system downward into the cover.
2
3
4
3. Then pull off the backing paper.
4. Finally, stretch the hygienic protective cover over the
camera from front to rear.
CAUTION
For optimum image quality, there must not be any folds in
the transparent area of the cover above the lens window.
CAUTION
Use only the hygienic protective covers offered by Sirona.
Prior to each application on a new patient, the camera
must be covered with a new hygienic protective cover.
59 06 313 D 3382
D 3382.201.01.13.02
75
Page 76
14 SIVISION 3 Sirona Dental Systems GmbH
14.4 SIROCAM C intraoral camera (extra equipment)Operating Instructions C4
14.4SIROCAM C intraoral camera (extra equipment)
The SIROCAM C intraoral camera combined with SIVISION 3 is a perfect system for effective communication
with the patient.
The camera images cannot be used for diagnosis!
3
*
t 1 / t 2
2
1
The SIROCAM C intraoral camera is a sensitive optical
instrument and must therefore always be handled with
care.
The camera illuminates the tooth areas to be imaged
with 6 white-light LEDs (1).
These are arranged around the single-lens optical system (2).
The LEDs and the lens are protected by a scratch-proof
antiglare glass cover.
CAUTION
* = Intermittent camera operation:
t1= 1 min on, t2 = 3 min off.
The camera heats up during operation.
To prevent overheating of the camera, the camera must
be put down for a time interval t2 = 3 min after an operating time of t1 = 1 min.
+
When the camera is in LIVE IMAGE mode, it is possible
to toggle between close range and far range by pressing
button (3) on the camera.
Button released – close range 5 – 15mm
Button pressed – far range from 15mm to smile line.
76D 3382.201.01.13.02
59 06 313 D 3382
Page 77
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.5 Operating modes of SIVISION 3
Hygienic protective covers for SIROCAM C
Hygienic protective covers are available for the camera.
They can be ordered under REF 59 14 705. Package
size: 500 pcs.
How to use the hygienic protective covers:
1. Hold the hygienic protective cover with its paper side
face-down and insert the camera with the optical
system downward into the cover.
2. Pull off the upper protective foil with the blue stripe.
3. Then pull off the backing paper.
14.5Operating modes of SIVISION 3
4. Finally, stretch the hygienic protective cover over the
camera from front to rear.
CAUTION
For optimum image quality, there must not be any folds in
the transparent area of the cover above the lens window.
CAUTION
Use only the hygienic protective covers offered by Sirona.
Prior to each application on a new patient, the camera
must be covered with a new hygienic protective cover.
SIVISION 3 with SIROCAM 3 or SIROCAM C can be
used basically in two different operating modes:
59 06 313 D 3382
D 3382.201.01.13.02
•VIDEO mode
The camera image is directly displayed on the
SIVISION monitor.
The quad image memory of the camera is activated.
•PC mode
The camera image is first processed on the PC and
then the image generated on the PC is displayed on
the SIVISION monitor.
77
Page 78
14 SIVISION 3 Sirona Dental Systems GmbH
14.5 Operating modes of SIVISION 3Operating Instructions C4
SWITCHING TO PC MODE
+
A
VIDEOCAM
PC MODE
A
VIDEOCAM
VIDEO MODE
A/B key on the control panel
Only with:
•camera picked up and activated,
•manually or by cursor (using the foot switch),
“VIDEOCAM” and the operating mode “PC
MODE” now appear on the display.
The image from the PC monitor, e.g. SIDEXIS/VI-
DEXIS, appears on the SIVISION monitor. The camera switches to LIVE IMAGE.
This mode remains active even after you switch OFF
and back ON the treatment center.
SWITCHING TO VIDEO MODE
A/B key on the control panel
Only with:
•camera picked up and activated,
•manually or by cursor (using the foot switch),
A/B
A
A/B VIDEO MODE
ON – OFF
“VIDEOCAM” and the operating mode “VIDEO
MODE” now appear on the display.
The SIVISION monitor displays the camera image.
This mode remains active even after you switch OFF
and back ON the treatment center.
This toggling function of the “mode switchover” key can
be disabled in Mode dialog 11.20 Mode: A/B VIDEO
MODE ON – OFF to prevent inadvertent switching.
78D 3382.201.01.13.02
59 06 313 D 3382
Page 79
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)
14.6SIROCAM 3 / SIROCAM C integrated in the dentist element
(video)
Operation in video mode
In this version the SIROCAM 3 / SIROCAM C intraoral
camera is installed in holder 5 of the dentist element or
in an additional holder on the dentist element.
The camera can be controlled with the unit foot switch if
no other instrument is activated (picked up and controllable with foot switch) when you pick up the camera.
If the camera is the active instrument “VIDEOCAM”
and the operating mode “VIDEO MODE” appear on
the display.
A
VIDEOCAM
VIDEO MODE
59 06 313 D 3382
D 3382.201.01.13.02
79
Page 80
14 SIVISION 3 Sirona Dental Systems GmbH
14.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)Operating Instructions C4
SAVING AN IMAGE
In VIDEO mode, the camera can save 4 still images consecutively (indexed with I, II, III and IIII).
+
A 5th view shows a QUAD IMAGE in which these 4 still
images are tiled on the screen.
Picking up the camera
If a STILL IMAGE or QUAD IMAGE was displayed while
the camera was in place, this view is retained.
If a BLACK SCREEN or a WHITE SCREEN was displayed while the camera was in place, the view preceding the BLACK or WHITE SCREEN, i.e. a LIVE, STILL
or QUAD IMAGE, reappears.
80D 3382.201.01.13.02
59 06 313 D 3382
Page 81
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
C
+
+
14.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)
TOGGLING BETWEEN STILL IMAGE AND LIVE
IMAGE
Foot switch C+ or C
With camera picked up and cursor being in the instruments key block.
With the C+ foot switch:
Step on the foot pedal.
With the C foot switch:
C
Actuate the foot switch in direction A – cursor upward.
Toggling between a STILL IMAGE and a LIVE IMAGE.
C
+
C
SWITCHING TO THE NEXT IMAGE
Spray ON / OFF key on the control panel
Manually or by cursor (using the foot switch) with instruments in place or camera picked up.
With C+ foot switch also left foot switch button (only with
camera picked up).
i
NOTICE
If the camera is in the LIVE IMAGE mode, the LIVE IMAGE is stored in the current memory location when you
switch to the next image.
With the camera in place, a BLACK SCREEN is stored in
the current memory location when you switch to the next
image with empty memory.
In SINGLE IMAGE mode, switching to the next memory
location.
•If a STILL IMAGE is stored there, it will be displayed.
•If no STILL IMAGE is stored there, a LIVE IMAGE
will appear with the camera picked up.
In QUAD IMAGE mode, switching to memory location I, II, III or IIII for image selection.
59 06 313 D 3382
D 3382.201.01.13.02
The number of the currently selected image lights up
in the upper right corner of the image.
81
Page 82
14 SIVISION 3 Sirona Dental Systems GmbH
14.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)Operating Instructions C4
TOGGLING BETWEEN QUAD IMAGE AND SINGLE IMAGE
Chip blower key on the control panel
Manually or by cursor (using the foot switch) with instruments in place or camera picked up.
With C+ foot switch also right foot switch button (only
with camera picked up).
+
C
+
C
Toggling between a QUAD IMAGE and a SINGLE
IMAGE.
When you switch to a SINGLE IMAGE, the image previously selected in the QUAD IMAGE is displayed.
CLEARING THE MEMORY
Electric motor CCW rotation key on the control
panel
Only with the camera picked up and activated, manually
or by cursor (using the foot switch).
Clears all image memory locations of the camera.
The LIVE IMAGE mode for image I appears.
59 06 313 D 3382
82D 3382.201.01.13.02
Page 83
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)
Placing back the camera
•If a LIVE IMAGE was displayed, a BLACK SCREEN
appears.
C
+
C
•A STILL IMAGE or QUAD IMAGE, BLACK
SCREEN or WHITE SCREEN remain displayed.
SWITCHING TO THE NEXT IMAGE WITH THE CURSOR
Special function only with cursor:
The cursor is in the chair programs key block, level 1.
With the C
4-way foot control plate forward.
With the C foot switch:
Actuate the foot switch in direction A – cursor upward.
•Switching through the images:
+
foot switch:
STILL IMAGE I, STILL IMAGE II, STILL IMAGE III,
STILL IMAGE IIII, QUAD IMAGE, STILL IMAGE I,
STILL IMAGE II ...
59 06 313 D 3382
D 3382.201.01.13.02
83
Page 84
14 SIVISION 3 Sirona Dental Systems GmbH
14.6 SIROCAM 3 / SIROCAM C integrated in the dentist element (video)Operating Instructions C4
WHITE SCREEN
X-ray image viewer ON / OFF key on the control
panel
•Switch X-ray image viewer or WHITE SCREEN on
the SIVISION monitor ON/OFF.
Presetting possible in the Mode menu (11.19):
•Only X-ray image viewer ON/OFF if NO flashes
or
•X-ray image viewer and WHITE SCREEN on the
SIVISION monitor ON/OFF if YES flashes.
i
NOTICE
WHITESCREEN - NO should be selected if an X-ray image viewer is installed. If no X-ray image viewer is installed, select WHITESCREEN - YES.
A
+
WHITESCREEN
YES – NO
BLACK SCREEN
“100” quick setting key on the control panel
Manually or by cursor (using the foot switch) with instruments in place or camera picked up. Cursor forward if
the cursor LED next to the quick setting keys lights up.
•Toggling between any view and a BLACK SCREEN.
•The previous view reappears when you press this
key again.
84D 3382.201.01.13.02
59 06 313 D 3382
Page 85
Sirona Dental Systems GmbH14 SIVISION 3
SIVISION
SIVISION
SIVISION
Operating Instructions C4
+
14.7 SIROCAM 3 / SIROCAM C integrated in the dentist element (PC)
14.7SIROCAM 3 / SIROCAM C integrated in the dentist element (PC)
Operation in PC mode using
a PC connection and SIDEXIS/VIDEXIS
The monitor is controlled by the PC in this case.
The functional scope of the PC can be used in this mode.
If the camera is the active instrument, “VIDEOCAM”
and the operating mode “PC MODE” appear on the display.
A
VIDEOCAM
PC MODE
Picking up the camera
•The LIVE IMAGE window opens.
C
+
C
TOGGLING BETWEEN STILL IMAGE AND LIVE
IMAGE
Foot switch C
With camera picked up and cursor being in the instruments key block.
With the C + foot switch:
Step on the foot pedal.
With the C foot switch:
Actuate the foot switch in direction A – cursor upward.
Toggling between a STILL IMAGE and a LIVE IMAGE.
+
or C
59 06 313 D 3382
D 3382.201.01.13.02
85
Page 86
14 SIVISION 3 Sirona Dental Systems GmbH
14.7 SIROCAM 3 / SIROCAM C integrated in the dentist element (PC)Operating Instructions C4
USING THE FUNCTION KEYS
Other functions, e.g. the SIDEXIS/VIDEXIS functions
Tile, Next Image, Full Frame, Zoom In, Zoom Out,
Rotate Image as well as other PC controls can be
assigned as desired to the following function keys on the
dentist element control panel:
•Spray
•Chip blower
•Electric motor CCW rotation
•Instrument light
•Quick setting keys 50 and 1
•Cursor forward, if the cursor is between the spray and chip blower keys (with C foot switch:
only if the camera is in its holder)
•Cursor forward, if the cursor is between the S
and 0 chair program keys
+
•(With C
foot switch – cursor forward. With C foot
switch – Actuate foot switch in direction A – cursor
upward)
•with instruments in place or camera picked up.
•Call key
•and # key (if not already used for relay functions)
The factory settings as well as the configuration procedure are described in the “Installation manual for PC software“,
REF 59 11 719 Operator’s Manual.
+
86D 3382.201.01.13.02
59 06 313 D 3382
Page 87
Sirona Dental Systems GmbH14 SIVISION 3
SIVISION
SIVISION
Operating Instructions C4
+
14.7 SIROCAM 3 / SIROCAM C integrated in the dentist element (PC)
SAVING AN IMAGE
Key example: Spray ON / OFF key on the control
panel
For this purpose, a patient must be registered in
SIDEXIS/VIDEXIS.
Manually or by cursor (using the foot switch) with instruments in place or camera picked up.
+
With C
camera picked up).
foot switch also left foot switch button (only with
C
+
C
•Save the STILL IMAGE.
With the camera picked up, a LIVE IMAGE appears.
Placing back the camera
59 06 313 D 3382
D 3382.201.01.13.02
•The LIVE IMAGE window closes.
The displayed STILL IMAGE or any saved images
remain open.
87
Page 88
14 SIVISION 3 Sirona Dental Systems GmbH
14.7 SIROCAM 3 / SIROCAM C integrated in the dentist element (PC)Operating Instructions C4
WHITE SCREEN
X-ray image viewer ON / OFF key on the control
panel
•Switch X-ray image viewer or WHITE SCREEN on
the SIVISION monitor ON/OFF.
Presetting possible in the Mode menu (11.19):
– only X-ray image viewer ON/OFF if NO flashes
or
+
A
WHITESCREEN
YES – NO
– X-ray image viewer and WHITE SCREEN on the SIVISION monitor ON/OFF if YES flashes.
i
NOTICE
WHITESCREEN - NO should be selected if an X-ray image viewer is installed. If no X-ray image viewer is installed, select WHITESCREEN - YES.
BLACK SCREEN
“100” quick setting key on the control panel
Manually or by cursor (using the foot switch) with instruments in place or camera picked up. Cursor forward if
the cursor LED next to the quick setting keys lights up.
•Toggling between any view and a BLACK SCREEN.
•The previous view reappears when you press this
key again.
i
NOTICE
WHITE SCREEN/BLACK SCREEN is only possible
when “AUTO” is set on the monitor (see SIVISION 3 “Operating Instructions for the Flat-Screen Monitor”)
88D 3382.201.01.13.02
59 06 313 D 3382
Page 89
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.8 SIROCAM 3 / SIROCAM C integrated in the assistant element (video)
14.8SIROCAM 3 / SIROCAM C integrated in the assistant element (video)
Operation in video mode
If a PC is connected, PC mode must be switched OFF!
See section 14.10 „Switching between VIDEO mode and PC mode“
In this version the SIROCAM 3 / SIROCAM C intraoral
camera is installed in holder 1 or 2 of the assistant element.
It is operated exclusively with a separate foot button in
SIROCAM 3
A
C
n
O
o
R
r
I
I
S
s
the chair base and the camera button.
Picking up the camera
SIROCAM C
•If a STILL IMAGE or QUAD IMAGE was displayed
while the camera was in place, this view is retained.
•If a BLACK SCREEN was displayed while the camera was in place, a LIVE IMAGE appears.
•If a WHITE SCREEN was displayed while the camera was in place, the view preceding the WHITE
SCREEN, i.e. a LIVE, STILL or QUAD IMAGE, reappears.
59 06 313 D 3382
D 3382.201.01.13.02
89
Page 90
14 SIVISION 3 Sirona Dental Systems GmbH
14.8 SIROCAM 3 / SIROCAM C integrated in the assistant element (video)Operating Instructions C4
GENERATING STILL IMAGES
•Briefly touch the foot button when the camera has
been picked up. The image number I appears in the
upper right corner of the image.
•When you touch the foot button again briefly, a LIVE IMAGE reappears.
•To freeze the next STILL IMAGE briefly touch the
foot button again. Image number II appears in the
upper right corner of the image.
•Use the same method to generate STILL IMAGEIII
and IIII.
•When you briefly touch the foot button after STILL IMAGE IIII has appeared, the QUAD IMAGE (the 4
single images in a tiled view) is displayed on the
screen.
+
90D 3382.201.01.13.02
59 06 313 D 3382
Page 91
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.8 SIROCAM 3 / SIROCAM C integrated in the assistant element (video)
SELECTING / OVERWRITING IMAGES
•When a QUAD IMAGE is displayed with the camera
picked up, you can select a specific image from the
QUAD IMAGE by briefly touching the foot button.
The image number of the selected image appears in
SIROCAM C
the upper right corner of the image.
SIROCAM 3
•When you briefly press the camera button, the selected image is displayed on the screen as a single
STILL IMAGE. When you press the button again,
the view changes to the QUAD IMAGE, etc.
59 06 313 D 3382
D 3382.201.01.13.02
•When you briefly touch the foot button while the selected single STILL IMAGE is displayed on the
screen, a LIVE IMAGE appears. It can be frozen by
touching the foot button again and overwrites the selected single STILL IMAGE at this memory location.
91
Page 92
14 SIVISION 3 Sirona Dental Systems GmbH
14.8 SIROCAM 3 / SIROCAM C integrated in the assistant element (video)Operating Instructions C4
DELETING ALL IMAGES (from the camera memory)
1.
With a QUAD IMAGE displayed and the camera picked
up.
+
SIROCAM 3
2.
•Press and hold down the foot button and then
press the camera button.
2.
SIROCAM C
•A LIVE IMAGE appears.
92D 3382.201.01.13.02
59 06 313 D 3382
Page 93
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.8 SIROCAM 3 / SIROCAM C integrated in the assistant element (video)
Placing back the camera
•If a LIVE IMAGE was displayed, a BLACK SCREEN
appears.
•A STILL IMAGE or a QUAD IMAGE remains dis-
played.
WHITE SCREEN
To activate a WHITE SCREEN, the camera must be
placed in its holder.
When you briefly press the camera button now, a
WHITE SCREEN is displayed on the monitor. When you
press the button again, the view changes to a QUAD IMAGE, STILL IMAGE or BLACK SCREEN, etc.
59 06 313 D 3382
D 3382.201.01.13.02
93
Page 94
14 SIVISION 3 Sirona Dental Systems GmbH
14.9 SIROCAM 3 / SIROCAM C integrated in the assistant element (PC)Operating Instructions C4
14.9SIROCAM 3 / SIROCAM C integrated in the assistant element
(PC)
Operation in PC mode using a PC connection
and SIDEXIS/VIDEXIS
To use the PC mode, SIDEXIS/VIDEXIS must be open.
The PC mode must be activated on the PC! See section
14.10 “Switching between VIDEO mode and PC
mode”.
SIROCAM 3
A
C
n
O
o
R
r
I
I
S
s
SIROCAM C
Picking up the camera
•A LIVE IMAGE opens.
+
TOGGLING BETWEEN LIVE IMAGE AND STILL
IMAGE
•Briefly touch the foot button when the camera has
been picked up.
•When you briefly touch the foot button again (< 1s),
a LIVE IMAGE reappears.
94D 3382.201.01.13.02
59 06 313 D 3382
Page 95
Sirona Dental Systems GmbH14 SIVISION 3
SIVISION
Operating Instructions C4
+
14.9 SIROCAM 3 / SIROCAM C integrated in the assistant element (PC)
SAVING A STILL IMAGE IN SIDEXIS/VIDEXIS
•Save the displayed STILL IMAGE by pressing the
foot button for some time (> 1s) when the camera
has been picked up.
•A LIVE IMAGE opens again automatically.
Placing back the camera
•The LIVE IMAGE is closed.
•A STILL IMAGE remains displayed.
59 06 313 D 3382
D 3382.201.01.13.02
95
Page 96
14 SIVISION 3 Sirona Dental Systems GmbH
14.10 Switching between VIDEO mode and PC modeOperating Instructions C4
USING THE CAMERA BUTTON AND THE FOOT BUTTON AS FUNCTION KEYS
The camera must be placed in its holder.
Further functions, such as WHITE SCREEN, BLACK
SCREEN or the SIDEXIS/VIDEXIS functions
Tile, Next Image, Full Frame, Zoom In, Zoom Out,
Rotate Image as well as other PC controls can be
assigned as desired to the camera button and the foot
button.
The factory settings as well as the configuration procedure are described in the “Installation manual for PC software“, REF 59 11 719 Operator’s Manual.
+
14.10Switching between VIDEO mode and PC mode
Only with SIROCAM 3 / SIROCAM C in the assistant
element and a connected PC
Video mode
PC mode
Systray
The system can be used in two operating modes:
•Video mode
The video image from the camera is directly displayed on the SIVISION monitor.
•PC mode
The video image is processed on the PC. The SIVISION monitor shows the PC user interface.
To toggle between the two modes, the popup menu of
the Siucom program on the PC is used.
The SIROCAM 3 / SIROCAM C camera is always in
video mode after the treatment center is switched on or
the camera is connected to the assistant element.
When the Siucom program is running (executes simultaneously with SIDEXIS/VIDEXIS) and the entry
“PC_MODE_ON” has been activated (factory setting),
the SIROCAM 3 / SIROCAM C camera automatically
switches to the PC mode.
The SIROCAM 3 / SIROCAM C camera remains in PC mode until one of the following occurs: video mode is
activated on the PC, the camera is disconnected from
the treatment center or the treatment center is switched
off.
96D 3382.201.01.13.02
59 06 313 D 3382
Page 97
Sirona Dental Systems GmbH14 SIVISION 3
Operating Instructions C4
+
14.10 Switching between VIDEO mode and PC mode
Switching to video mode
Right-click the Siucom icon in the Systray (usually in the
lower right corner of the screen).
In the Siucom popup menu which then opens, click
“PC_MODE_OFF”. Confirm by clicking OK in the
„SIROCAM being deactivate“ message box.
These settings will remain saved even after you switch
off the PC.
The SIVISION monitor must also be set to video mode
(S Video Port; see monitor operating instructions).
Switching to PC mode
Right-click the Siucom icon in the Systray (usually in the
lower right corner of the screen).
In the Siucom popup menu which then opens, click
“PC_MODE_ON”. Confirm by clicking OK in the „SIRO-
CAM being activated“ message box.
These settings will remain saved even after you switch
off the PC.
The SIVISION monitor must also be set to PC mode
(Analog Port; see monitor operating instructions).
After switching the treatment center or the PC off and on
again, it may be necessary to reset the monitor to PC mode.
59 06 313 D 3382
D 3382.201.01.13.02
97
Page 98
14 SIVISION 3 Sirona Dental Systems GmbH
14.11 SIVISION3 – second monitor function without SIROCAM3/ SIROCAM COperating Instructions C4
14.11SIVISION3 – second monitor function without SIROCAM3 /
SIROCAM C
The monitor integrated in the treatment center serves as
primary or secondary monitor for the externally connected PC.
The following function keys of the control panels as well
as foot switch functions are available for controlling PC
functions:
•Spray
•Chip blower
•Electric motor CCW rotation
•Instrument light
•Call key (if not already used for relay functions)
•# key (if not already used for relay functions)
•All quick setting keys: 100, 50 and 1
•Cursor forward, if the cursor is between the
spray and chip blower keys
•Cursor forward, if the cursor is between the S
and 0 chair program keys
+
(With C
switch – Actuate foot switch in direction A – cursor
upward.)
All instruments must be in place when you do so!
foot switch – cursor forward. With C foot
+
as well as the function keys:
•X-ray image viewer
They can be used to control e.g. SIDEXIS/VIDEXIS
functions such as Tile, Next Image, Full Frame, Zoom In, Zoom Out, Rotate Image, but also other PC applications.
The factory settings as well as the configuration procedure are described in the “Installation manual for PC
software“, REF 59 11 719 Operator’s Manual.
98D 3382.201.01.13.02
59 06 313 D 3382
Page 99
Sirona Dental Systems GmbH15 CEREC Chairline (Option)
Operating Instructions C4
+
15.1 CEREC Chairline
15 CEREC Chairline (Option)
15.1CEREC Chairline
CEREC Chairline is used for 3-dimensional acquisition
of the preparation, the occlusion and the antagonist.
For details on its function and operation, please refer to
the CEREC Chairline Operating Instructions
(REF 60 46 028) and Operator Manual (REF 59 56 458).
CEREC Chairline foot switch
A Bluetooth radio module for transmitting the CEREC
foot switch signal to the CEREC PC is integrated in the
CEREC Chairline foot switch.
For details refer to Chapter 5.
Model designation of radio moduleBlue RS+I (Stollmann Co.)
Frequency2.4GHz - 2.4835GHz (ISM band)
Bluetooth class2
Rangeapprox. 10m
ApprovalsCE conformity according to R&TTE Directive (1999/5/EC)
Part 15 of the FCC Rules,
RSS-210 of Industry Canada
59 06 313 D 3382
D 3382.201.01.13.02
99
Page 100
16 Extra equipment and accessories Sirona Dental Systems GmbH
16.1 Media block on the patient chairOperating Instructions C4
16 Extra equipment and accessories
16.1Media block on the patient chair
Extra equipment
Additional equipment which is approved for dental use
(e.g. by EMS) may be connected to a media block on the
base of the chair (observe manufacturers’ specifications
for the additional equipment).
Fuse
Line voltage
230V
Water
Air
The media block is equipped with quick couplings for air
and water and an outlet with a separate fuse (6.3A) for
non-heating equipment.
For treatment centers with a disinfection system, the
DVGW approval (EN1717/ DIN1988) remains valid also
with an integrated media block. In this case, the additional equipment connected to the treatment center also
complies with the requirements of EN1717 / DIN1988.
+
Technical data and requirements•Connected equipment is treated with disinfected wa-
ter if a disinfection system is installed. During regular
operation, this is 0.1‰ to 0.2‰ hydrogen peroxide,
during sanitation 1.4%.
CAUTION
Before connecting any additional equipment, check to
make sure that it is suitable for hydrogen peroxide.
If the connected equipment is not suitable for the higher
sanitation concentration, Sirona recommends to disconnect it from the treatment center during sanitation and to
perform a separate sanitation in accordance with the
manufacturer’s specifications.
Sirona does not assume any liability for possible damage to the connected equipment!
Specification of pressure and flow conditions
PressureFlow
Water2.2 ± 0.2 barmax. 300 ml/min
Air4.4 ± 0.5 barmax. 70 Nl /min
100D 3382.201.01.13.02