Sirona C1+ Service manual

2I[EWSJ
'
7IVZMGI1ERYEP

)RKPMWL
IMPORTANT:
• It is essential that you take this Service Manual with you for every visit to a customer.
Furthermore, you must always have the spare parts list and wiring diagrams with you as well.
You can order additional copies of this Service Manual under
• Order Number 59 06 610 from our department DZA TU in Bensheim.
See reverse side of manual for address.
D 3386.076.01.01.02 11.2001
2I[EWSJ
'
7IVZMGI1ERYEP

)RKPMWL
Contents
1 Overview of modules and PCBs ................................................................................ 1 – 1
2 Important notes ........................................................................................................... 2 1
2.1 Technical data...................................................................................................... 2 2
2.2 Warnings and safety notes................................................................................... 2 – 2
2.3 Minimum configuration ......................................................................................... 2 – 2
2.4 Packing position for repackaging ......................................................................... 2 – 3
3 Wiring diagrams .......................................................................................................... 3 – 1
3.1 Wiring diagram of CAN bus.................................................................................. 3 – 2
3.2 EMERGENCY STOP wiring diagram ................................................................... 3 – 3
4 User-defined settings .................................................................................................. 4 – 1
5 Error catalog ................................................................................................................ 5 – 1
6 Reading service codes................................................................................................ 6 – 1
6.1 Reading the service codes on the dentist panel .................................................. 6 – 2
6.2 Reading service codes with PC connection to C1
+
.............................................. 6 – 4
7 Service code listings ................................................................................................... 7 – 1
7.1 Abbreviations / service code listings .................................................................... 7 – 2
7.2 Service code of PCB (SA) in connection box AK ................................................. 7 – 3
7.2.1SW version: 2.0 - 3.0
7.2.2SW version: 1.1- 1.3
7.3 Service messages of PCB (AJ) in dentist element AE......................................... 7 – 13
7.3.1SW version: 1.4 - 2.5
7.3.2SW version: 1.0 - 1.3
7.4 Service messages of PCB (PS) in dentist panel AP ............................................ 7 – 28
7.4.1SW version: 1.0 - 2.4
7.5 Service codes of PCB (TS) support arm in dentist element................................. 7 – 37
7.5.1SW version: 1.4 - 1.5
7.5.2SW version: 1.0, 1.2 - 1.3
7.6 Service messages of PCB (AJ) on the assistant panel........................................ 7 – 43
7.6.1SW version: 1.4 - 2.5
7.6.2SW version: 1.0 - 1.3
7.7 Service codes of PCB (PS) on the assistant panel .............................................. 7 – 55
7.7.1SW version: 1.0 - 2.4
7.8 Service codes of PCB (SS) chair control ............................................................. 7 – 64
7.8.1SW version: 1.1 - 2.5
7.9 Service codes of PCB (WS) in the water unit....................................................... 7 – 76
7.9.1SW version: 1.0, 1.1 - 1.9, 2.0 - 2.4, 3.0
59 06 628 D 3386 D 3386.076.01.01.02 11.2001
C1+

1 Overview of modules and PCBs

1 Overview of modules and PCBs
1 - 2 D 3386.077.01.01.02 11.2001
59 06 610 D 3386
Component PCB / Module
Dentist element (AE) PS = Control panel
AG = Base plate
IR = Flexible PCB
AVL = Supply for lamp C1
AD2 = Camera module
ACI = HF adapter, from C1
HF+ = HF module, from C1
AJ = Control module
AS = SPRAYVIT module
AL = Siromot module
AV = Distributor light module
AH = Solenoid valve and heater module
AU = SIROSON module
AO = HF modulation module, model 96
AC = Output stage for SIROTOM
Support arm (TA) TS = Support arm control
Assistant element (HE) PS = Control panel
AJ = Control module
AS = SPRAYVIT module
IR = Flexible PCB
HG = Base plate
Water unit (WE) WS = Control PCB
Patient chair (ST) SE = Chair output stage
SS = Control ST
NS = Power supply PCB
SA = Connection box control
NM = Monitor power supply
SVC = Video PCB
SL = Wiring PCB
KS = 4-way foot switch
Foot switch (FS) AF = Foot switch plate
1 Overview of modules and PCBs
)RKPMWL
1
STAEHEFS WETA
+
+
+
59 06 610 D 3386 D 3386.077.01.01.02 11.2001
1 - 3
1 Overview of modules and PCBs
1 - 4 D 3386.077.01.01.02 11.2001
59 06 610 D 3386
C1+

2 Important notes

2 Important notes
2.1 Technical data
Model designation C1
Power supply 230V 50Hz,
Nominal current 4.5A at 230V,
+
115V 50/60Hz
9.5A at 115V
2.2 Warnings and safety notes
Caution! Prior to opening the unit, connecting a measuring instrument or replacing parts: Switch the unit OFF.
Operating safety: To guarantee the operating safety, the use of mobile radio telephones in the practice or hospital
area must be prohibited.
Troubleshooting: If you encounter difficulties search in the error catalog first and follow the procedure described
in it.
2.3 Minimum configuration
C1 serial no. < 3000
C1 component PCB Software version
Dentist panel / Assistant panel PS / Control panel >= 1.6
Dentist element / Assistant element AJ / Dentist element control >= 1.9
Water unit WS / Water unit control >= 1.9
C1 serial no. > 3000
C1 component PCB Software version
Water unit WS >= 2.1
+
serial no. > 10 000
C1
C1+ component PCB Software version
Dentist panel / Assistant panel PS / Control panel >= 2.3
Dentist element / Assistant element AJ / Dentist element control >= 2.5
Connection box SA / Connection box control >= 3.0
Patient chair SS / Chair control >= 2.5
Water unit WS / Water unit control >= 3.0
2 - 2 D 3386.076.01.01.02 11.2001
59 06 628 D 3386
2 Important notes
2.4 Packing position for repackaging
To repack the chair (e.g., for transport to a trade fair), move it to the position shown.
)RKPMWL
Cur
sor
Key
30
TONE
Auto
Ret
ROM
1.
2.
Invisible key
Packing key
880
2
Vid.
Mod
SAV
59 06 628 D 3386 D 3386.076.01.01.02 11.2001
To do this, press the following keys:
1. Invisible key in basic settings dialog 2
2. Packing position key
The chair now moves to the packing position automatically.
2 - 3
C1+

3 Wiring diagrams

3 Wiring diagrams
3.1 Wiring diagram of CAN bus
HF+
AP
Connection area Dentist element
AT
AE
Connection area
GA1 GA2
X1 X2
X1 X1X2X1
1 1A1
1 1
CAN H
CANHHEIN
AK
2 2B1
2 2
Track
X10
12X312
CAN L CAN L HE IN
X4 X3
B1 B1A1 A1A3 A3B3 B3
CAN L ST IN
CANHSTIN
CAN H ST OUT
CAN L ST OUT
25B11AA2A64B
X1 X5
215 6
CAN L ST IN
CANHSTIN
CAN H ST OUT
GA3
CAN L ST OUT
ST WE
A1
B1
Connection area Track
Connection area Assistant element
HE
HP
3 – 2 D 3386.076.01.01.02 11.2001
59 06 628 D 3386
3 Wiring diagrams
3.2 EMERGENCY STOP wiring diagram
)RKPMWL
3
PS
PCB
TS
PCB
X1.2
S5
X3.4B (C1/A)
X3.2
AJ
PCB
HG
PCB
GA2
GA1
AG
PCB
X1.5B
X4.5B
X1.3
X2.3
X1.3
X1.3
X4.5B X3.2X2.5 X1.5B
S3
The EMERGENCY STOP can be
triggered both by the safety switch
and the software.
X3.3A
X4.B
X4.A
AJ
PCB
X3.3B
X6.2A
X6.1B
SL
PCB
X1.2X2.5A (C1/A)
PS
PCB
X3.B X3.A
SS
PCB
X3.3
X3.2
X3.4
X3.1
WS
PCB
GA3
X5.3A
X3.3
X3.2A
X2.2A
S2
S2
SA
X11.2
X11.3
PCB
X4.2A
X1.2A
A
A
C
E
K
C
V1
E
K
59 06 628 D 3386 D 3386.076.01.01.02 11.2001
3 – 3
C1+

4 User-defined settings

4 User-defined settings
The following user-defined settings must be made after replacing PCBs or software on the C1.
Component (PCB / software) User-defined settings
Dentist element PCB AJ
Also in case of software upgrade from version < 1.6 to version >= 1.6 and from version < 2.0 to version >= 2.0
Dentist panel AP Panel-specific settings:
Assistant element PCB AJ Also in case of software upgrade from version < 1.6 to version >= 1.6 and from version < 2.0 to version >= 2.0
Instrument settings for all function levels SPRAYVIT L: Instrument light, water and air temperature SIRONA motor: Speed, instrument light, direction of rotation, cooling media,
direct starter, foot switch control
Turb ine: Instrument light, cooling media SIROSON L: Intensity, instrument light SIROTOM: Cutting current, coagulation current POLYLUX: Exposure time, soft start (ON / OFF) Holder assignment for saline solution
(in case of missing Sprayvit light and water heater generate 20 pump strokes)
In case of camera integration: video mode
Brightness, contrast, keyboard sensitivity, audio signal volume, key clicking tone (ON / OFF), timer settings, simplified programs (ON/OFF), cursor (ON/OFF) SIROTOM operating and alarm tone, preselection of key position for direction of rotation/cooling medium/chip blower in instrument program. Sivision / bell switch-over version 1.7 or higher
In case of secondary monitor function, PC dialog (ON/OFF).
Different tones for each row of keys (ON/OFF).
SPRAYVIT L: Instrument light, water and air temperature POLYLUX: Exposure time
Functionality of the 4-way foot switch for Polylux and suction function version
2.0 or higher
Assistant panel HP Panel-specific settings:
Water unit PCB WS Also in case of software upgrade
Patient chair PCB SS
Connection box PCB SA Also in case of software upgrade
Brightness, contrast, keyboard sensitivity, audio signal volume, key clicking tone (ON / OFF), timer settings, simplified programs (ON/OFF) Sivision / bell switch-over version 1.7 or higher
SIROLUX: Brightness, reflector stage, SIROLUX switching condition in relation to chair position, tumbler filling quantity, tumbler filling (ON/OFF) when approaching S position, suction intensity, possibly time and date
Chair positions:
User, user size, treatment position (sitting/standing)
Track positions of dentist element Cuspidor flushing time, cuspidor flushing (ON/OFF) when approaching S position, 4-way foot switch functionality (switch-over headrest/backrest), Type of key for bell or free key (key/switch)
59 06 628 D 3386
4 – 2 D 3386.076.01.01.02 11.2001
Component (PCB / software) User-defined settings
4 User-defined settings
Support arm control PCB TS Also in case of software upgrade
Support arm position excluding track Intensity of saline solution pump
)RKPMWL
4
59 06 628 D 3386 D 3386.076.01.01.02 11.2001
4 – 3
C1+

5 Error catalog

5 Error catalog
Compo­nent
Dentist element
Dentist element
Dentist element
Dentist element
Dentist element
Dentist element
Description of problem Loss of instrument
programs.
After removing the instruments, the factory settings are displayed for all levels and holders. Sprayvit light switched off.
Service code message: AE 7F
Instrument failure
Dentist panel closes the instrument dialog. Instrument functions Motor Scaler Turbine SIROTOM including media spray, water and instrument light are shut off, if active. After depositing and removing again, normal function.
Instrument failure
Instruments Motor Scaler SIROTOM do not function when activated. Media spray, water and instrument light are switched when activated. After depositing and removing again, same error. Can only be remedied by switching on/off.
Instrument failure SIRONA motor
The instrument Sirona motor interrupts its function or does not function when activated. Media spray, water and instrument light remain switched on or are switched when activated. After depositing and removing again, normal function.
Service code message: AE 00 6B
Sirona motor runs on for too long
Scaler switches off again after 1–2 seconds.
Scaler dialog is closed.
Cause Corrective action
Software error PCB AJ of version <2.0
Error while reading hose coding following, e.g., electromagnetic interference or contact problems on hose coupling.
Software version PCB AJ in dentist element <1.8.
Software error on PCB AJ in dentist element, in case of very short instrument removal, e.g., due to oscillating motion of the thin SIROTOM hose.
Software version PCB AJ in dentist element <1.8.
The software of PCB AL (Siromot module) detects open motor load circuits in very smooth motors.
Software version PCB AL in dentist element <3.6.
Software PCB AL in dentist element <3.7
Water on the flange due to leaks in hose or heater.
Residual moisture due to thermal disinfection
Replace AJ software. Part No.: 54 33 375
Replace AJ software. Part No.: 54 33 375 Check and correct (if necessary) the position of the SIROTOM flange (if available). Do not overlap or cross the cables.
Replace AJ software. Part No.: 54 33 375
Replace AL software. Part No.: 54 33 383
Replace AL software. Part No.: 54 33 383
Remedy leaks.
Remedy moisture.
5 – 2 D 3386.076.01.01.02 11.2001
59 06 628 D 3386
5 Error catalog
Compo­nent
Dentist element
Dentist element
Dentist element
Dentist element
Dentist element
Dentist element
Description of problem Performance of
scaler is reduced or fluctuating, or no performance.
High-frequency whistling.
Whistling tone from dentist element when activating a bur instrument
Motor/turbine, immediate cooling/ driving air
After removing the motor, turbine cooling/driving air is immediately on, or after releasing the foot switch cooling/
driving air and spray, if any, are not switched off.
After depositing the instrument everything is switched off.
Occurs on all motors/turbines,
possibly only from time to time
Motor/turbine, low air outlet.
After removing the motor, turbine or after releasing the foot switch, low air outlet. After depositing the instrument everything is switched off.
Does not occur on exactly one motor/ turbine.
No air and no water on one bur instrument.
Dripping instruments.
Cause Corrective action
Old sealing compound of piezo is brittle. Electrical contacts in handpiece are broken.
Pressure setting too low
MV11 Module
Solenoid valve MV11 is caught.
Defective membrane in the module of the instrument for which the error does not occur.
MV21-X on module defective (X – holder no.)
White break on “auto valve of module due to excessive screwing torque or spring force too low.
Replace handpiece
Replace membrane and check input pressure, pressure =
4.3 bar +/- 0.1 (in case Sprayvit flow is active)
Replace MV11. Replace module.
Replace solenoid valve MV11.
Replace membrane and check input pressure. Pressure = 4.3 bar +/- 0.1 (in case SPRAYVIT flow is active).
Replace MV21.X.
Auto valve has been reworked, replace auto valve, do not tighten too hard. Check filter in heater.
English
5
59 06 628 D 3386 D 3386.076.01.01.02 11.2001
Filter tissue of heater is dissolved.
Precipitation in the heater element.
Check filter in heater and replace it, if necessary, and rinse water pipes. (Defective filter and information to Bensheim)
Replace heater and rinse water pipes.
5 – 3
5 Error catalog
Compo­nent
Dentist element
Dentist element
Dentist element
Dentist element
Dentist element
Dentist element
Support arm control
Support arm control
Description of problem No instrument dialog
when removing instruments.
SPRAYVIT 4000 occasionally without water.
(Air is still available)
Moist air on SPRAYVIT.
Some water is ejected when the air key is pressed.
SPRAYVIT L handpiece becomes hot.
Both heater cartridges heat up when only one key is pressed. Air heater and SPRAYVIT L heat up immediately after removing the instrument. Valve body is burnt.
Assistant element Polylight cannot be selected using the dentist panel.
Software PCB AJ assistant element version 2.0
Chair and AE travel to position “0” when an instrument is removed.
Cursor is switched off (M1 mode), panel in chair program, “FS down activated
Dentist element does not reach programmed position.
Program key must be pressed several times.
Service code message in AT/HP: 40 04
(RESET AT) – frequent in C1/96).
Support arm does not swivel in joint 1, Motor 1 is running
(Serial no.: <3000)
Cause Corrective action
Instrument in holder 5 or 6 has not been deposited properly. Instrument tray bent (holder is not aligned with light barrier)
Transparency of ZEG hose
Contact problem or humidity on base of air heater in valve body. Consequence: SPRAYVIT is not detected by dentist element.
Damaged o-rings on Sprayvit nozzle, due to sterilization and lack of maintenance.
Water in SPRAYVIT hose.
Software error in PCB AJ version 2.0
Software error dentist panel / assistant panel
PCB TS reset because watchdog tolerance is too high. PCB TS:
Serial no. >3000:
Hardware <2.1, Software <1.5
Serial no. <3000:
Hardware <1.3
Pertinax pinion of joint 1 defective
Slipping clutch of joint 1 skids
Information for the customer.
Install a new instrument tray
Hose change in progress
Temporary measure: Clean air heater (remove also solder residue) and tighten contact terminals.
Replace o-rings on Sprayvit nozzle.
Replace hose and replace valve body, if necessary.
Replace AJ software. Part No.: 54 33 375
No solution as of yet. Serial no. <3000: Software version on PCB PS >= 1.8 Serial no. >3000: Software version on PCB PS >= 1.7
Replacement of PCB TS.
Part No.: 46 96 931 - Serial no. >3000
Part No.: 46 85 231 - Serial no. <3000
Install new motor 1 with metal pinion
Replace slipping clutch of joint 1
5 – 4 D 3386.076.01.01.02 11.2001
59 06 628 D 3386
5 Error catalog
Compo­nent
Support arm control
Support arm control
Support arm control
Support arm control
Support arm control
Description of problem Support arm does
not swivel in joint 1, Motor 1 is running
Furthermore, a knocking noise can be heard (serial no.: >3000).
Support arm motors run on.
Support arm motors keep running for a few seconds after reaching the programmed position.
Support arm motor 1 does not reach programmed position.
Program key must be pressed twice
Unit head drops when an additional load is applied.
The unit head keeps falling back to the lowest position.
Cause Corrective action
Gear Z70 is damaged, because motor 1 has not been inserted deep enough or grub screw is missing in lower support arm bearing.
Pot. target value is outside the movement range
Slipping clutch of joint 1 has not been tightened enough at factory Only serial no.: >3000
Slipping clutch of joint 1 too weak only serial no.: <3000
Spring weight compensation not properly set.
Slipping clutch torn off.
Replace lower support arm bearing entirely. Rep. no. 5433623
Check the fastening and fitting of the potentiometers, then try to travel to the corresponding program positions again and press the save key.
Tighten slipping clutch with a screwdriver, as necessary. Factory setting: 20 +- 5N
Replace slipping clutch of joint 1. Installation instructions currently being drafted
Tighten the spring using allen screw on spring arm. Remarks: The slipping clutch of the height adjustment is exclusively required to dampen the spring force. It is preset at the factory and can be adjusted slightly in exceptional cases. Generally, the following applies: Always tighten the spring for weight - force adjustment.
Dismount the slipping clutch. When the driving pin is not hammered in deep enough, drive it in and then mount back the slipping clutch with an undamaged groove.
English
5
Support arm control
Assistant element
Assistant element
Unit head does not swivel, motor is running.
Polylight cannot be started.
After removing the instrument the timer in the assistant panel does not switch over to the Polylight values and the Polylight cannot be started.
Polylight light power is not in green range
when measured with Translux Test.
Slipping clutch has burst due to excessive tightening
The grub screw used to attach the pinion to the back-geared motor 3 is loose.
Software error PCB AJ software, assistant element. Polylight detection in case of an old Polylight (without electronics) using fan current defective. Software version on PCB AJ, Assistant element <1.9.
Fiber optic is not properly inserted or locked in place in the module.
Lamp has air leaks so that the quartz tube has become opaque.
Lamp has not been inserted properly.
Dirt in filter of lamp module, at light entrance or exit of fiber optic and of handpiece.
Replace slipping clutch. Rep. no. 3319501
Tighten the grub screw and secure with Loctite. (The back-geared motor must be dismounted to do this)
Replace AJ software (basic retrofit kit).
Pull out the fiber optic and insert it again, until it catches.
Replace the halogen lamp. Only use selected Osram Xenophot HLX bulbs.
Lamp must lie flat against the heat sink, place the lug on the lamp in the notch intended for this purpose.
Clean all interfaces with ethanol
59 06 628 D 3386 D 3386.076.01.01.02 11.2001
5 – 5
5 Error catalog
Compo­nent
Assistant element
Water unit
Water unit
Water unit
Water unit
Water unit
Water unit
Description of problem Short total failure.
SIROLUX (unit model), film viewers, if any, are shut off and must be switched on again. The suction and separator function are interrupted for 2–4 seconds and start up by themselves again.
Service code message in AP/HP: 40 08
Amalgam separator full message too early.
Panel display amalgam and continuous tone, rotor not filled yet (total weight of rotor <500 g)
Amalgam separator goes over to fault mode, loud noises and vibration from amalgam separator.
Amalgam is flashing and intermittent beeping tone.
Service code message: WE 06
Amalgam separator goes over to fault mode.
Amalgam is flashing and intermittent beeping tone. (loud drive noise, rustling or standstill)
Service code WE: 5C
Drive below target speed
There is barely any water arriving to the tumbler filling.
Tumbler filling does not work from time to time.
Vibrations of the amalgam separator
Cause Corrective action
Software error on PCB WS, which can lead to a RESET of the module when filling the water tank and the disinfectant injection.
Serial no. <3000:
Software version on PCB WS <1.9.
Serial no. >3000:
Software version on PCB WS <2.1.
Software version on PCB WS <2.0.
Driver bolt defective.
Bearing damage on motor (audible) due to leak in pump
Motor is blocking
Bearing in lower part is defective
Rotor axle is defective
Drain is blocked up
Suction from lower part is blocked up
Media pipe is clogged due to precipitation in the heater element.
Software version on PCB WS <2.4.
Float switch in mixing tank is caught.
Amalgam separator has been mounted stressed
Replace AJ software. Part No.: 54 33 391
Replace WS software. Part No.: 54 33 391
Replace amalgam drive.
Replace amalgam drive.
Replace the amalgam drive.
Check the rotors operation (must run smoothly) replace lower part, if necessary.
Replace rotor
Check siphon and drain, clean them if necessary. Check the hose path
Clean suction channel in lower part of amalgam separator.
Replace heater, clean media pipes, if necessary.
A solution is being worked out.
Replace float switch.
Position all connections to amalgam separator so as to leave 1mm clearance in all directions from amalgam separator,
Defective rotor or flyweight
Suction from lower part is blocked up
Transport safety device has not been removed.
Replace rotor
Clean suction channel
Remove the transport safety device
59 06 628 D 3386
5 – 6 D 3386.076.01.01.02 11.2001
5 Error catalog
Compo­nent
Water unit
Water unit
Water unit
Water unit
Water unit
Water unit
Water unit
Description of problem Tumbler falls into
cuspidor
Cuspidor does not drain off
intermittent beeping tone
Service code WE: 07
Water escapes from water unit
Suction does not switch off or is delayed
Water unit is deformed in left­handed version
Disinfectant consumption too high
Flushing water quantity:
too low
too high
Cause Corrective action
Slope of cuspidor too steep
Vibrations of the amalgam separator
Clogged drain, def. water detector, air pressure too low
Leaks on tumbler plug adapter
Water escapes from mixing tank overflow, Clogged or badly positioned drain
MV 33 is caught Replace MV 33
Lack of stiffness in left­handed version
Metering valve is leaking Replace the metering valve
Control nozzle not properly adjusted
C
B
1.
2.
Reduce slope of cuspidor, if possible
Order silicone mat free of charge Part No.: 54 46.443
Use of tumblers with hollow bottom
See: Vibrations of the amalgam separator Check drain and air pressure;
Clean water detector or replace, if necessary;
If everything is in order, reduce the flow of the cuspidor flushing
Replace sealing ring in plug adapter Part No.: 70 36 189 (red o-rings)
Check slope of drain, Clean the siphon
Retrofit reinforcing plate Part No.: 47 08 272
Part No.: 14 35.580
1. Adjusting cuspidor flushing:
For right-handed version – Remove side panel A of water unit and adjust control nozzle B until the bottom of the cuspidor is flushed evenly. Remount back side panel A of water unit.
For left-handed version – Remove cover C. Lift rubber cover and adjust control nozzle B using a coin until the bottom of the cuspidor is flushed evenly. Refit the cover.
2. Check the swivel-back motion of the cuspidor:
English
5
Patient chair
Patient chair
59 06 628 D 3386 D 3386.076.01.01.02 11.2001
Leaks in connection box
No chair motion after switching on.
Serial no.: <3000 Chair output stage version: Hardware PCB SE 1.0
Service code message: ST 03 37 (also 01 01)
A
B
1.
Counter nut on ceramic valve is loose
Output stage relay is caught.
– Swivel the cuspidor into the movement range of the
chair
– Press the manual chair button and start the chair
program consecutively
The cuspidor must always swivel back into the starting position automatically
Tighten the nut, fix by a lock nut and secure with locking varnish
Replace PCB SE - Version 1.1.
5 – 7
5 Error catalog
Compo­nent
Patient chair
Patient chair
Patient chair
Patient chair
Description of problem No chair motion after
switching on
only in cooled treatment center
Service code message: ST 02 00 to 02 05
Chair does not reach programmed position.
Program key must be pressed several times, particularly when travelling to S or 0 position. No service code message in ST. Software PCB SS <1.8.
Chair does not reach programmed position.
Program key must be pressed several times, or chair does not leave position. EMERGENCY STOP relay is switching.
Service code message in AP/HP: 40 04
(RESET AT)
No service code message in ST.
(more frequent for serial no.: >3000)
Chair does not reach programmed position. Service code message: 08 0X; X = 0...2
Cause Corrective action
Temperature sensibility of PCB SS, Software version on PCB SS <2.0
EMERGENCY STOP activated by EMERGENCY STOP sensory mechanism.
Defective or badly adjusted safety switch for foot part, or defective cable. Service code ST 09 11, EMERGENCY STOP activated by safety switch for foot part.
Defective safety switch for backrest, or defective cable or backrest mechanics jammed. Service code ST 09 13, EMERGENCY STOP activated by free lift switch of backrest.
PCB TS is RESET due to excessively high watchdog tolerance.
PCB TS: C1/96, Serial no. >3000:
Hardware: <2.1, software: <1.5
C1/A: Serial no. <3000:
Hardware: <1.3
Broken wire in the motor sensor circuit (L16) Serial no. <3000.
Contacts pushed back on connector X15 on motor circuit (L16).
Replace the software on PCB SS
Replace software for diagnostics PCB SS, version >= 2.0 to recognize source of EMERGENCY STOP. Serial no.: <3000, Part No.: 54 33 680 Serial no.: >3000, Part No.: 54 33 409
Adjust or replace the switch or cable. (Switch must only switch shortly before the upper end stop and must not be charged in the lower end stop).
Replace the switch or cable. Eliminate the stiffness of the backrest.
Replace PCB TS.
Serial no.: >3000, Part No.: 46 96 931
Serial no.: <3000, Part No.: 46 85 231
Replace motor sensor circuit (L16).
Check contacts.
Patient chair
Patient chair
Chair does not reach programmed position. Service code message: 08 03
Chair does not travel to programmed position.
Potentiometer 0–2 defective.
Potentiometer 3 no longer moves into gear due to worn sword rollers
Unintentional call of the factory programs, mostly with foot switch. (Display with icon in odontogram)
Switch the potentiometers between each other and check if error is reproduced, replace the potentiometer, if necessary
Install roller kit Part No.: 54 33 870
Information to customers.
59 06 628 D 3386
5 – 8 D 3386.076.01.01.02 11.2001
Loading...
+ 88 hidden pages