Sanyo MLS-2781L, MLS-3751L Manual

0 (0)
Sanyo MLS-2781L, MLS-3751L Manual

In-LabTop Loading Autoclaves

n o i t a z i l i r e t s

MLS-3751L

MLS-3781L

MLS-3751L

50 liters (1.8 cu. ft.)

MLS-3781L

75 liters (2.6 cu. ft.)

38.0 inches

inches

 

29.4

18.8

inches

18.8

inches

 

 

 

Sterilization Temperatures

+115°C to +135°C

In-lab autoclaves (personal-use mobile units)

Space efficient, convenience minded, portable autoclaves offer sterilization on-demand.

Researchers will no longer waste valuable time and energy when limited to using a centralized building autoclave. Installation and maintenance of central autoclaves are not only costly but time consuming. Sanyo offers two convenient solutions. These portable autoclaves are designed for individual lab use and can conveniently be moved from one lab to another. Model MLS-3751L has a low enough profile to be stored under a lab bench when not in use.

innovation performance reliability support

innovation performance reliability support

s t e r i l i

MOBILE AUTOCLAVE UNITS

Personal Series In-Lab Autoclaves

SANYO MLS-Series top-loading autoclaves are a popular method of sterilization for today’s research laboratories. Self-contained and easy to use, these reliable units are ideal for a wide range of applications, including liquid culture media preparation, labware and waste sterilization. Designed to meet good laboratory practice criteria in biotechnology and research laboratories, SANYO MLSSeries portable autoclaves deliver high pressure steam with speed, efficiency and reliability.

Small Footprint with Swing-Up Door

The door opens upwards, so no side opening space is required. This makes the 14.56 inch (37 centimeters) wide chamber easy to access. The MLS-3781L offers 75liter capacity with the same footprint as the MLS-3751L (50 liters) and is only 18.8 inches wide.

Double Interlocking Structure

To ensure the safe operation of this high temperature, high-pressure autoclave, the chamber and the open/close hand-pull lever are controlled by a double interlock system dependent on temperature and pressure sensors. This way, it is impossible to open the autoclave until temperature and pressure levels are safe.

z a t i o n

DESIGN

 

RELIABILITY & CONTROL

Human-oriented design

SANYO has included a host of features to make this autoclave user friendly. There is a heat-insulating resin housing and the corners are gently rounded. The control panel is mounted on top so it is easier to see and use. There is a clip-on drainage hose and an exhaust tank that’s easy to fill and empty, eliminating the need for fixed plumbing.

Microprocessor Temperature Control

Sterilizing temperature is controlled by the microprocessor to within +2°C of the

set temperature in the range of 115°C to 135°C.

Process Voice Notification

A new feature the MLS series includes is voice notification of the system process. Each step of the process is notified via a pre-recorded voice message, allowing the end user to hear the process as it is happening from across the room.

Integrated Cool Down Fan

Included as a standard feature, the cooldown fan decreases cool down process times significantly. This enables access to sterilized contents much faster.

Air Vent Control

The exhaust valve release temperature can be set to allow automatic release after the sterilizing cycle has been completed. The setting range is from +0°C to 45°C above boiling point to suit the items being sterilized (patent pending).

Optional Process Printer (Model MLS-37PR)

Process documentation is available via an optional built-in type thermal printer. The printer prints out the process run results of temperature, pressure, vs. time. Each printout is date and time stamped for future reference.

Optional Object Temperature Sensor (Model MLS-37SB)

Measures temperature of items in the autoclave chamber and sets the sterilization timer accordingly. Suitable for materials that heat up more slowly than the chamber, such as resins. After installation, the autoclave control system can be switched between this sensor and the main sensors.