Sanyo MDF-U537D, MDF-U537, MDF-U333 Service Manual

4 (2)

Service

 

Manual

Biomedical freezer

 

MDF-U333

 

MDF-U537

 

MDF-U537D

SANYO Electric Biomedical Co., Ltd.

MDF-U537

Effective models

This service manual is effective following models.

Model name

Model code

Voltage and Frequency

MDF-U537

823 168 51

115V

60Hz

 

823 168 52

220V

50Hz

 

823 168 53

220V

60Hz

 

823 168 54

230V

50Hz

 

823 168 55

240V

50Hz

 

823 168 56

115V

60Hz

 

823 168 57

110V

60Hz

 

824 168 58

230V

50Hz

 

824 168 59

220V

50Hz

 

824 168 60

240V

50Hz

 

825 168 61

 

 

 

825 168 62

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MDF-U537D

823 169 51

115V

60Hz

 

823 169 52

220V

50Hz

 

823 169 53

220V

60Hz

 

823 169 54

230V

50Hz

 

827 169 55

240V

50Hz

 

823 169 56

240V

50Hz

 

823 169 57

230V

50Hz

 

823 169 58

115V

60Hz

 

823 169 59

110V

60Hz

 

823 169 60

220V

50Hz

 

823 169 61

 

 

 

823 169 62

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MDF-U333

823 170 51

115V

60Hz

 

823 170 52

220V

50Hz

 

823 170 53

220V

60Hz

 

823 170 54

230V

50Hz

 

823 170 55

240V

50Hz

 

823 170 56

110V

60Hz

 

823 170 57

230V

50Hz

 

823 170 58

240V

50Hz

 

823 170 59

220V

50Hz

 

823 170 60

 

 

 

823 170 61

 

 

 

823 170 62

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sanyo MDF-U537D, MDF-U537, MDF-U333 Service Manual

Contents

Feature

 

 

1

Specif ication

MDF - 333

3

 

 

MDF - 537

5

 

 

MDF - 537D

7

Dim ensions

MDF - 333

9

 

 

MDF - 537

10

 

 

MDF - 537D

11

Ref rigeration

circuit

12

Ref rigeration circuit principle diagram MDF - 333

15

Ref rigeration circuit welding points MDF - 333

16

Ref rigeration circuit principle diagram MDF - 537

17

Ref rigeration circuit welding points MDF - 537

18

Ref rigeration circuit principle diagram MDF - 537D

19

Ref rigeration circuit welding points MDF - 537D

20

Electrical Parts

21

Specif ication

of sensor

24

Control

specif ication

25

W iring

diagram

34

Circuit

diagram

37

Components on P . C . B .

38

T est data

 

39

Parts loyout

 

49

User’smanual

 

50

Features

HFC refrigerant

Refrigerant circuit contains HFC refrigerant that effect to environment little.

Latest cooling system reduces negative factor to global environment. With ambient temperature at 35 , while temperature of inside chamber maintains -30 .

PCM structure

PCM structure reduces amount of organic solvent.

Temperature control function

Unified temperature control range for same categories realized to set -20 , which is suitable for reagent preservation.

Specification unified

Unified specification for same categories, which also realized to unify performance for high/low temperature alarm (SV is changeable in the range ±5 ±15 ), power failure alarm, remote alarm terminal, setting memorization by non-volatile memory.

Adopt new microprocessor system for display panel and keypad makes operation easily.

Self diagnosis function

Abnormal condition for temperature sensor is indicated by self-diagnosis system.

Validation

Control panel enables zero “0” adjust for validation.

New door latch mechanism

New door latch is adopted in order to close door surely for reagents. It is also available to use individual lock mechanism.

Insulation wall thickness

Unit has necessary insulation performance for each temperature range. Combination of PCM and frame form makes 610mm for frame size commonly.

Option

Internal temperature observation system is option. Unit has communication function corresponding to it.

1

Features

HFC refrigerant

Refrigerant circuit contains HFC refrigerant that effect to environment little.

Latest cooling system reduces negative factor to global environment. With ambient temperature at 35 , while temperature of inside chamber maintains -30 .

PCM structure

PCM structure reduces amount of organic solvent.

Temperature control function

New cooling system realizes to set -20 , which is suitable for reagent preservation.

Specification unified

Unified specification for same categories, which also realized to unify performance for high/low temperature alarm (SV is changeable in the range ±5 ±15 ), power failure alarm, remote alarm terminal, setting memorization by non-volatile memory.

Adopt new microprocessor system for display panel and keypad makes operation easily.

Self diagnosis function

Abnormal condition for temperature sensor is indicated by self-diagnosis system.

Validation

Control panel enables zero “0” adjust for validation.

New door latch mechanism

New door latch is adopted in order to close door surely for reagents. It is also available to use individual lock mechanism.

Draw box for reagent preservation

100mm for depth of box makes things easily to arrange and take in & out. (Adopted in bottom chamber of MDF-U537/U537D)

Compressor protection function

Sensor detects compressor’s warming up to control unit running.

2

Specification

MDF-U333

Items

Specification

Model description

Biomedical freezer

Model Number

 

MDF-U333

Outer dimension (Max)

W 614 × D 733 × H 1620 mm

Inner dimension

W 490 × D 485 × H 1290 mm

Effective chamber capacity

274 lit.

Exterior

 

Zinc galvanized steel w/powder polyester coating (PCM)

Interior

 

Polystyrene resin

Insulation

 

Rigid polyurethane foam

Outer door

 

1 pc, Zinc galvanized steel w/powder polyester coating (PCM)

Latch

 

1 pc

Door lock

 

1 pc

 

 

Basket

(L) 4 pcs Inner dimension: W446×D369×H220 mm

Accessories

 

(S) 1 pc (for bottom column) Inner dimension: W487×D221×H155mm

 

 

 

 

Others

Defrosting spatula 1 pc, key 1 set, 4 card holders

Caster

 

4 pcs

Leveling legs

 

2 pcs (with caster)

Access hole

 

φ 30mm. 1 place in left side.

Weights

 

81 Kg

Evaporator

 

Tube on sheet type (function both evaporator and shelf)

Condenser

 

Wire and tube type

Compressor

 

Hermetic type 225W

Refrigerant

 

HFC refrigerant

 

R-134a

 

 

 

Cooling performance

-30 (internal 1/2h air temperature, AT 35 , no load)

Internal temperature control

-20 ~-30

range

 

(AT 5 ~35 no load)

Power supply

 

Local voltage

3

MDF-U333

 

Items

 

Specification

Temperature controller

Microprocessor controlled system with non-volatilized memory.

Temp. set range : -18 -40

 

 

Temperature sensor

Thermistor

 

Temperature display

LED digital display (green colored) by 1 .

Range : -50 50

 

 

 

High/Low

Changeable SV±5 15 . (Initial SV±10 )

 

ALARM lamp brinks and buzzer sounds intermittently

 

Temperature

 

(about 15 min. delay)

 

 

Alarm

 

Remote alarm contact: turns ON during power failure alarm.

Power failure

ALARM lamp

brinks, buzzer sounds intermittently, remote alarm

 

 

contact turns ON during power failure alarm.

 

 

 

Remote alarm

Remote alarm terminal 3P: contact capacity DC30v, Max2A, N.O.

 

contact

–COM. (During power failure, temperature alarm: ON)

 

 

ALARM lamp

 

 

 

BUZZER: Buzzer key

 

 

ALARM TEST: Alarm test key

Operation

Control panel

SET: To switch the SV and PV

 

: To shift the set digit

panel

 

: To change the SV

 

 

DEF: Defrost key

 

Rear

Power switch (Breaker switch)

 

Battery switch: Power failure alarm switch

 

 

 

 

 

Key lock function

Press key for five seconds

L0: Unlock

L1: Lock

 

 

Self diagnosis function

When each sensor is in fault,

Error code and internal temperature are displayed alternately.

 

 

Remote alarm contact turns ON with buzzer sounds.

Power switch

Breaker switch

Compressor protection

Overload relay

Option

Chart recorder: MTR-G85

Mounting kit: MPR-S7(for bottom mounting)

 

 

: MDF-S740T(for top mounting)

4

Specification

MDF-U537

Items

Specification

Model description

Biomedical freezer

Model Number

 

MDF-U537

Outer dimension (Max)

W 804 × D 772 × H 1802 mm

Inner dimension

W 658 × D 607 × H 1272 mm

Effective chamber capacity

482 lit.

Exterior

 

Zinc galvanized steel w/powder polyester coating (PCM)

Interior

 

Polystyrene resin

Insulation

 

Rigid polyurethane foam

Outer door

 

Zinc galvanized steel w/powder polyester coating (PCM)

 

2 pcs (upper and lower compartment)

 

 

 

Latch

 

2 pcs

Door lock

 

1 pc

 

 

 

(M) 6 pcs (upper compartment)

 

 

Basket

Inner dimension: W290×D536×H136 mm

Accessories

 

(S) 8 pcs (lower compartment)

 

 

 

 

 

Inner dimension: W290×D536×H100mm

 

 

Others

Defrosting spatula 1 pc, key 1 set, 14 card holders

Caster

 

4 pcs

Leveling legs

 

2 pcs (with caster)

Access hole

 

φ 30mm. 1 place on rear.

Weights

 

131 Kg

Evaporator

 

Tube on sheet type (function both evaporator and shelf)

Condenser

 

Wire and tube type

Compressor

 

Hermetic type 350W

Refrigerant

 

HFC refrigerant

 

R-407D

 

 

 

Cooling performance

-30 (internal 1/2h air temperature, AT 35 , no load)

Internal temperature control

-20 ~-30

range

 

(AT 5 ~35 no load)

Power supply

 

Local voltage

5

MDF-U537

 

Items

 

Specification

Temperature controller

Microprocessor controlled system with non-volatilized memory.

Temp. set range : -18 -40

 

 

Temperature sensor

Thermistor 2P

Temperature display

LED digital display (green colored) by 1 .

Range : -50 50

 

 

 

High/Low

Changeable SV±5 15 . (Initial SV±10 )

 

ALARM lamp brinks and buzzer sounds intermittently

 

Temperature

 

(about 15 min. delay)

 

 

Alarm

 

Remote alarm contact: turns ON during power failure alarm.

Power failure

ALARM lamp

brinks, buzzer sounds intermittently, remote alarm

 

 

contact turns ON during power failure alarm.

 

 

 

Remote alarm

Remote alarm terminal 3P: contact capacity DC30v, 2A, N.O. –COM.

 

contact

(During power failure, temperature alarm: ON)

 

 

ALARM lamp

 

 

 

BUZZER: Buzzer key

 

 

ALARM TEST: Alarm test key

Operation

Control panel

SET: To switch the SV and PV

 

: To shift the set digit

panel

 

: To change the SV

 

 

DEF: Defrost key

 

Rear

Power switch (Breaker switch)

 

Battery switch: Power failure alarm switch

 

 

 

 

 

Key lock function

Press key for five seconds

L0: Unlock

L1: Lock

 

 

Self diagnosis function

When each sensor is in fault,

Error code and internal temperature are displayed alternately.

 

 

Remote alarm contact turns ON with buzzer sounds.

Power switch

Breaker switch

 

 

Overload relay

 

 

Detect temperature near compressor by thermistor sensor to control

Compressor protection

compressor during overload.

 

 

Alarm: Error code and internal temperature are displayed alternately.

 

 

Remote alarm contact turns on with buzzer sounds.

Option

Chart recorder: MTR-G85

Mounting kit: MPR-S7(for bottom mounting)

 

 

: MDF-S740T(for top mounting)

6

Specification

MDF-U537D

Items

 

Specification

Model description

Biomedical freezer

Model Number

 

MDF-U537D

 

Outer dimension (Max)

W 804 × D 772 × H 1802 mm

Inner dimension

Upper chamber

W 658 × D 607 × H 589 mm

Lower chamber

W 658 × D 607 × H 603 mm

 

 

 

Effective chamber capacity

452 lit.

 

Exterior

 

Zinc galvanized steel w/powder polyester coating (PCM)

Interior

 

Polystyrene resin

 

Insulation

 

Rigid polyurethane foam

Outer door

 

Zinc galvanized steel w/powder polyester coating (PCM)

 

2 pcs (upper and lower compartment)

 

 

 

Latch

 

2 pcs

 

Door lock

 

1 pc

 

 

 

 

(M) 6 pcs (upper compartment)

 

 

Basket

Inner dimension: W290×D536×H136 mm

Accessories

 

(S) 8 pcs (lower compartment)

 

 

 

 

 

Inner dimension: W290×D536×H100mm

 

 

Others

Defrosting spatula 1 pc, key 1 set, 14 card holders

Caster

 

4 pcs

 

Leveling legs

 

2 pcs (with caster)

Access hole

 

φ 30mm. 2 places on rear.

Weights

 

136 Kg

 

Evaporator

 

Tube on sheet type (function both evaporator and shelf)

Condenser

 

Wire and tube type

Compressor

 

Hermetic type 350W

Refrigerant

 

HFC refrigerant

 

 

R-407D

 

 

 

 

 

Cooling performance

-30 (internal 1/2h air temperature, AT 35 , no load)

Internal temperature control

-20 ~-30

 

range

 

(AT 5 ~35 no load)

Power supply

 

Local voltage

 

7

MDF-U537D

 

Items

 

Specification

Temperature controller

Microprocessor controlled system with non-volatilized memory.

Temp. set range : -18 -40

 

 

Temperature sensor

Thermistor 2P (upper and lower compartment)

 

 

LED digital display (green colored) by 1 .

Temperature display

Range : -50 50

 

 

Selectable to display upper or lower compartment by switch.

 

High/Low

Changeable SV±5 15 . (Initial SV±10 )

 

ALARM lamp brinks and buzzer sounds intermittently

 

Temperature

 

(about 15 min. delay)

 

 

Alarm

 

Remote alarm contact: turns ON during power failure alarm.

Power failure

ALARM lamp

brinks, buzzer sounds intermittently, remote alarm

 

 

contact turns ON during power failure alarm.

 

 

 

Remote alarm

Remote alarm terminal 3P: contact capacity DC30v, 2A, N.O. –COM.

 

contact

(During power failure, temperature alarm: ON)

 

 

ALARM lamp

 

 

 

BUZZER: Buzzer key

 

 

ALARM TEST: Alarm test key

 

Control panel

SET: To switch the SV and PV

Operation

: To shift the set digit

 

 

: To change the SV

panel

 

 

DEF: Defrost key

 

 

 

 

TOP/BOTTOM: Switch key to display upper or lower compartment.

 

Rear

Power switch (Breaker switch)

 

Battery switch: Power failure alarm switch

 

 

 

 

 

Key lock function

Press key for five seconds

L0: Unlock

L1: Lock

 

 

Self diagnosis function

When each sensor is in fault,

Error code and internal temperature are displayed alternately.

 

 

Remote alarm contact turns ON with buzzer sounds.

Power switch

Breaker switch

 

 

Overload relay

 

 

Detect temperature near compressor by thermistor sensor to control

Compressor protection

compressor during overload.

 

 

Alarm: Error code and internal temperature are displayed alternately.

 

 

Remote alarm contact turns on with buzzer sounds.

Option

Chart recorder: MTR-G85

Mounting kit: MPR-S7(for bottom mounting)

 

 

: MDF-S740T(for top mounting)

8

Dimensions

MDF-U333

9

Dimensions

MDF-U537

10

Dimensions

MDF-U537D

11

Refrigeration circuit

MDF-U333

 

 

 

Unit

Parts description

 

Specification

 

 

 

 

 

 

Evaporator

Type

 

Tube on sheet

 

 

Material

 

Aluminum (φ 8.0)

 

Dehydrator

Type of molecular sieves

 

A-XH9 134 stamped

 

Charged q’ty

 

8±1g

 

Capillary tube

Resistance

 

0.686MPaG 7Kg G

 

Dimension

 

φ 1.8×(φ 0.65)×2000mm

 

Resistance indicator color

 

White

 

Condenser assembly A

Type

 

Wire and tube

 

 

Dimension

 

w480×P40×30 columns

Condenser assembly B

Type

 

Wire and tube

 

 

Dimension

 

w480×P40×30 columns

 

 

 

Note: Lower 2 columns are used

 

 

 

in desuper circuit.

 

 

 

 

 

Frame pipe

Material

 

C1220T-1/2H

 

 

Dimension

 

φ 4.0× 0.5

 

Evacuation and vacuum

Evacuation time: over than

20min. by 2 ports (high pressure and low

pressure

pressure side). (N2 exchange)

 

 

When charged, vacuum pressure is lower than 1.0torr

 

pump capacity 300L/min

 

Refrigerant and oil

Refrigerant : R-134a

Charged q’ty : 226±5g

 

Oil : CF 32

 

Charged q’ty :

230±10cc

 

Oil additive: n-pentane

 

Charged q’ty :

14g

Refrigeration circuit spec.

Refer to following pages

 

 

 

and welding points

 

 

 

 

12

Refrigeration circuit

MDF-U537

 

 

 

 

 

 

Unit

Parts description

 

Specification

 

 

 

 

 

 

 

 

 

Evaporator

 

Type

9lines×1column×5pcs

Upper (1)

 

 

 

 

 

 

 

 

Lower (1)

 

 

 

 

 

 

 

 

Material

Aluminum (φ 8.0)

Bottom (3)

 

 

 

 

 

 

 

 

Dry core

 

Type

D-S032T

 

 

 

 

Manufacturer

SANYO

 

 

 

 

Solid core

50cm3

 

 

 

Capillary tube

 

Resistance

0.725MPaG

7.4Kg G

(Total number: 2)

 

 

Dimension

φ 1.8×L4500

 

 

Resistance indicator color

Blue

 

 

 

Condenser

assembly

Type

Wire and tube

front (Total number: 4)

Dimension

w480×P40×6 columns

Frame pipe

 

Material

C1220T

 

 

 

 

Dimension

φ

4.0

×

0.5

×L

 

 

 

 

 

7470

Evacuation and

vacuum

Evacuation time: over than

20min. by 2 ports (high pressure and low

pressure

 

pressure side). (N2 exchange)

 

 

 

 

 

 

When charged, vacuum pressure is lower than 1.0torr

 

 

pump capacity 300L/min

 

 

 

 

Refrigerant and oil

Refrigerant : R-407D

 

Charged q’ty : 301±5g

 

 

Oil : CF 32

 

Charged q’ty : 280±10cc

 

 

Oil additive: n-pentane

 

Charged q’ty: 19g

Refrigeration circuit spec.

Refer to following pages

 

 

 

 

 

and welding points

 

 

 

 

 

 

13

Refrigeration circuit

MDF-U537D

 

 

 

Unit

Parts description

 

Specification

 

 

 

 

 

Evaporator assembly

Type

9lines×1column×5pcs

Upper (1)

 

 

 

 

 

Lower (1)

 

 

 

 

 

 

 

 

Bottom (2)

 

Material

Aluminum (φ 8.0)

 

Lower entrance (1)

 

 

 

Dry core

 

Type

D-S032T

 

 

 

Manufacturer

SANYO

 

 

 

Solid core

50cm3

 

Capillary tube

 

 

Upper compartmnt

Lower compartmnt

 

 

Resistance

10.3Kg/ G

9.9Kg/ G

 

 

Dimension

φ 1.8×L4000

φ 1.8×L3000

 

 

Resistance indicator color

Yellow

Red

Condenser

assembly

Type

Wire and tube

 

front (Total number: 4)

Dimension

w480×P40×6 columns

Frame pipe

 

Material

C1220T

 

 

 

Dimension

φ 4.0× 0.5×L7470

Evacuation and

vacuum

Evacuation time: over than

20min. by 2 ports (high pressure and low

pressure

 

pressure side). (N2 exchange)

 

 

 

When charged, vacuum pressure is lower than 1.0torr

 

 

pump capacity 300L/min

 

Refrigerant and oil

Refrigerant : R-407D

Charged q’ty : 357±5g

 

 

Oil : CF 32

Charged q’ty :

280±10cc

 

 

Oil additive: n-pentane

Charged q’ty :

23g

Refrigeration circuit spec.

Refer to following pages

 

 

and welding points

 

 

 

14

Refrigeration circuit principle diagram

MDF-U333

15

Refrigeration circuit welding points

MDF-U333

16

Refrigeration circuit principle diagram

MDF-U537

17

Refrigeration circuit welding points

MDF-U537

18

Refrigeration circuit principle diagram

MDF-U537D

19

Refrigeration circuit welding points

MDF-U537D

20

Electrical Parts

MDF-U333

 

AC110V,60Hz

AC115V,60Hz

AC220-240V , 50/60Hz

Compressor

Type

C-2SN220L0W

C-2SN220L2W

C-2SN220L5W

 

Code

80771120

80771122

80771125

 

Rated voltage (50/60Hz)

100 / 100 , 110

-- / 115

220,240 / 220

 

Winding resistance C-S(Aux)

5.8Ω

7.2Ω

30.6Ω

 

C-R(Main)

1.6Ω

2.0Ω

7.8Ω

Starting relay

Type

AMVL-180A

AMVL-180A

AMVL-300A

 

Pick up voltage(50Hz)

AC91V 111V

AC91V 111V

AC185V 217V

 

(60Hz)

AC108 128V

AC108 128V

AC215 232V

 

Drop out voltage(50Hz)

AC30 65V

AC30 65V

AC60 120V

 

(60Hz)

AC30 75V

AC30 75V

AC69 132V

Overload relay

Type

P24KU

P19KU

P10MU

 

Action to the temp. (No current)

 

 

 

 

OFF

150±10

150±10

130±8

 

ON

85±10

85±10

69±10

 

Action to the current (AT25 )

24.0A

19.5A

10A

 

Operation time

6 15 seconds

6 15 seconds

6 15 seconds

 

Non-Action to the temp. (80

 

 

 

 

 

6.1A more than

5.4A more than

2.5A more than

 

Non-Action

30 minutes

30 minutes

30 minutes

 

Action

8.0A within 15 minutes

7.4A within 15 minutes

3.4A within 15 minutes

Starting capacitor

Rating

100μF,160VAC

100μF,160VAC

30μF,300VAC

Running capacitor

Rating

25μF,220VAC

20μF,220VAC

5μF,400VAC

Temp. control relay

Type

G4F-11123T

G4F-11123T

G4F-11123T

 

Contact capacity

AC220V,20A

AC220V,20A

AC220V,20A

 

Coil

DC12V

DC12V

DC12V

Switching power supply

Type

ZWS10-12/J

ZWS10-12/J

ZWS10-12/J

 

Input

AC100-240V

AC100-240V

AC100-240V

 

50/60Hz,0.3A

50/60Hz,0.3A

50/60Hz,0.3A

 

 

 

Rated output

DC12V,0.85A

DC12V,0.85A

DC12V,0.85A

Temperature sensor

Type

502AT-1

502AT-1

502AT-1

 

Rating

5kΩ,25

5kΩ,25

5kΩ,25

P.C.B.

Type

DF-333

DF-333

DF-333

Battery

Type

5HR-AAC

5HR-AAC

5HR-AAC

 

Rating

6V 1100MAH

6V 1100MAH

6V 1100MAH

Battery switch

Type

SLE6A2-5

SLE6A2-5

SLE6A2-5

 

Rating

AC250V 4A

AC250V 4A

AC250V 4A

Breaker switch

Type

BAM215131

BAM215131

BAM215131

 

Rating

AC250V 15A

AC250V 15A

AC250V 15A

21

Electrical Parts

MDF-U537

 

AC110-115V,60Hz

AC220V,50/60Hz

AC230V-240V,50Hz

Compressor

Type

C-2SN350L0R

C-2SN350L5S

C-2SN350L5S

 

Code

807764201

807763251

807763251

 

Rated voltage (50/60Hz)

 

 

 

 

Winding resistance C-S(Aux)

5.8Ω

30.6Ω

30.6Ω

 

C-R(Main)

1.6Ω

7.8Ω

7.8Ω

Starting relay

Type

AMVL-180A

AMVL-300A

AMVL-300A

 

Pick up voltage(50Hz)

AC91V 111V

AC185V 217V

AC185V 217V

 

(60Hz)

AC108 128V

AC215 232V

AC215 232V

 

Drop out voltage(50Hz)

AC30 65V

AC60 120V

AC60 120V

 

(60Hz)

AC30 75V

AC69 132V

AC69 132V

Overload relay

Type

P18AU

P10MU

P10MU

 

Action to the temp. (No current)

 

 

 

 

OFF

120±8

130±8

130±8

 

ON

69±10

69±10

69±10

 

Action to the current (AT25 )

18.5A

10A

10A

 

Operation time

6 15 seconds

6 15 seconds

6 15 seconds

 

Non-Action to the temp. (80

 

 

 

 

 

4.0A more than

2.5A more than

2.5A more than

 

Non-Action

30 minutes

30 minutes

30 minutes

 

Action

5.5A within 15 minutes

3.4A within 15 minutes

3.4A within 15 minutes

Starting capacitor

Rating

100μF,160VAC

30μF,300VAC

30μF,300VAC

Running capacitor

Rating

25μF,220VAC

5μF,400VAC

5μF,400VAC

Temp. control relay

Type

G4F-11123T

G4F-11123T

G4F-11123T

 

Contact capacity

AC220V,20A

AC220V,20A

AC220V,20A

 

Coil

DC12V

DC12V

DC12V

Solenoid valve

Type

G2R-1A-T

G2R-1A-T

G2R-1A-T

control relay

Contact capacity

AC250V,10A

AC250V,10A

AC250V,10A

 

Coil

DC12V

DC12V

DC12V

Switching power supply

Type

ZWS10-12/J

ZWS10-12/J

ZWS10-12/J

 

Input

AC100-240V

AC100-240V

AC100-240V

 

50/60Hz,0.3A

50/60Hz,0.3A

50/60Hz,0.3A

 

 

 

Rated output

DC12V,0.85A

DC12V,0.85A

DC12V,0.85A

Temperature sensor

Type

502AT-1

502AT-1

502AT-1

 

Rating

5kΩ,25

5kΩ,25

5kΩ,25

Solenoid valve control

Type

502AT-1

502AT-1

502AT-1

sensor

Rating

5kΩ,25

5kΩ,25

5kΩ,25

Compressor protect sensor

Type

502AT-1

502AT-1

502AT-1

 

Rating

5kΩ,25

5kΩ,25

5kΩ,25

P.C.B.

Type

DF-537

DF-537

DF-537

Battery

Type

5HR-AAC

5HR-AAC

5HR-AAC

 

Rating

6V 1100MAH

6V 1100MAH

6V 1100MAH

Battery switch

Type

SLE6A2-5

SLE6A2-5

SLE6A2-5

 

Rating

AC250V 4A

AC250V 4A

AC250V 4A

Breaker switch

Type

BAM215131

BAM215131

BAM215131

 

Rating

AC250V 15A

AC250V 15A

AC250V 15A

Solenoid valve coil

Type

NEVAC120V

NEVAC220V

NEVAC240V

 

Rating

120V 50/60Hz

220V 50/60Hz

240V 50/60Hz

Condensing fan motor

Type

FU2-A051B1MP

FU2-A051B5MP

FU2-A051B5MP

 

Rating

AC110-115V 50/60Hz

220-240V 50/60Hz

220-240V 50/60Hz

 

Thermal fuse

130

130

130

22

Electrical Parts

MDF-U537D

 

AC110-115V,60Hz

AC220V,50/60Hz

AC230V-240V,50Hz

Compressor

Type

C-2SN350L0R

C-2SN350L5S

C-2SN350L5S

 

Code

807764201

807763251

807763251

 

Rated voltage (50/60Hz)

 

 

 

 

Winding resistance C-S(Aux)

5.8Ω

30.6Ω

30.6Ω

 

C-R(Main)

1.6Ω

7.8Ω

7.8Ω

Starting relay

Type

AMVL-180A

AMVL-300A

AMVL-300A

 

Pick up voltage(50Hz)

AC91V 111V

AC185V 217V

AC185V 217V

 

(60Hz)

AC108 128V

AC215 232V

AC215 232V

 

Drop out voltage(50Hz)

AC30 65V

AC60 120V

AC60 120V

 

(60Hz)

AC30 75V

AC69 132V

AC69 132V

Overload relay

Type

P18AU

P10MU

P10MU

 

Action to the temp. (No current)

 

 

 

 

OFF

120±8

130±8

130±8

 

ON

69±10

69±10

69±10

 

Action to the current (AT25 )

18.5A

10A

10A

 

Operation time

6 15 seconds

6 15 seconds

6 15 seconds

 

Non-Action to the temp. (80

 

 

 

 

 

4.0A more than

2.5A more than

2.5A more than

 

Non-Action

30 minutes

30 minutes

30 minutes

 

Action

5.5A within 15 minutes

3.4A within 15 minutes

3.4A within 15 minutes

Starting capacitor

Rating

100μF,160VAC

30μF,300VAC

30μF,300VAC

Running capacitor

Rating

25μF,220VAC

5μF,400VAC

5μF,400VAC

Compressor relay

Type

G4F-11123T

G4F-11123T

G4F-11123T

 

Contact capacity

AC220V,20A

AC220V,20A

AC220V,20A

 

Coil

DC12V

DC12V

DC12V

Temp. control relay

Type

G2R-1A-T

G2R-1A-T

G2R-1A-T

 

Contact capacity

AC250V,10A

AC250V,10A

AC250V,10A

 

Coil

DC12V

DC12V

DC12V

Switching power supply

Type

ZWS10-12/J

ZWS10-12/J

ZWS10-12/J

 

Input

AC100-240V

AC100-240V

AC100-240V

 

50/60Hz,0.3A

50/60Hz,0.3A

50/60Hz,0.3A

 

 

 

Rated output

DC12V,0.85A

DC12V,0.85A

DC12V,0.85A

Temperature sensor

Type

502AT-1

502AT-1

502AT-1

 

Rating

5kΩ,25

5kΩ,25

5kΩ,25

Solenoid valve control

Type

502AT-1

502AT-1

502AT-1

sensor

Rating

5kΩ,25

5kΩ,25

5kΩ,25

Compressor protect sensor

Type

502AT-1

502AT-1

502AT-1

 

Rating

5kΩ,25

5kΩ,25

5kΩ,25

P.C.B.

Type

DF-537D

DF-537D

DF-537D

Battery

Type

5HR-AAC

5HR-AAC

5HR-AAC

 

Rating

6V 1100MAH

6V 1100MAH

6V 1100MAH

Battery switch

Type

SLE6A2-5

SLE6A2-5

SLE6A2-5

 

Rating

AC250V 4A

AC250V 4A

AC250V 4A

Breaker switch

Type

BAM215131

BAM215131

BAM215131

 

Rating

AC250V 15A

AC250V 15A

AC250V 15A

Solenoid valve coil

Type

NEVAC120V

NEVAC220V

NEVAC240V

(2 pcs.)

Rating

120V 50/60Hz

220V 50/60Hz

240V 50/60Hz

Condensing fan motor

Type

FU2-A051B1MP

FU2-A051B5MP

FU2-A051B5MP

 

Rating

AC110-115V 50/60Hz

220-240V 50/60Hz

220-240V 50/60Hz

 

Thermal fuse

130

130

130

23