Samsung HS40 Service manual

Page 1
HS40
Service Manual
Page 2
Version 1.00
HS40
Service Manual
SM-HS40-ENG-01
Page 3
Page 4
(Empty page)
Page 5
Safety Requirements

Classification

– Type of protection against electric shocks: Class I
– Degree of protection against electric shocks (when the patient is in physical contact):
Type BF or type CF mounting
– Degree of protection against the ingress of harmful liquids: General equipment
– Degree of safety of use in the presence of flammable anesthetic agent mixed with air,
oxygen, or nitrous ox ide: Not suitable for use near flammable anesthetic agent mixed with air, ox ygen, or nitrous oxide
– Mode of operation: Continuous operation

Safety standards the device conforms to

– Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and
Essential Performance [IEC 60601-1:2005/A1:2012]
– Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and
Essential Performance - Collateral Standard: Electromagnetic Compatibility – Requirements and Tests [IEC 60601-1-2:2007]
– Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and
Essential Performance - Collateral Standard: Usability [IEC 60601-1-6:2010]
– Medical Electrical Equipment - Part 2-37: Particular Requirements for the Basic Safety
and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment [IEC 60601-2-37:200 7]
– Medical Electrical Equipment - Part 1: General Requirements for Safety [IEC 60601-
1:1988, A1:1991, A2:1995]
– Medical Electrical Equipment - Part 1-1: General Requirements for Safety – Collateral
Standards: General Requirements for Medical Electrical Systems [IEC 60601-1-1:2000]
– Medical Electrical Equipment - Part 1-2: General Requirements for Safety – Collateral
Standards: Electromagnetic Compatibility – Requirements and Tests [IEC 60601-1- 2:2001, A1:2004]
– Medical Electrical Equipment - Part 1-4: General Requirements for Safety – Collateral
Standards: Programmable Electrical Medical Systems [IEC 60601-1-4:1996, A1:1999]
– Medical Electrical Equipment - Part 2-37: Particular Requirements for the Basic Safety
and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment
[IEC 60601-2-37:200 1, A1:2004, A2:2005]
– Medical Devices – Application of Risk Management to Medical Devices [ISO 14971:2007]
– Medical Electrical Equipment - Part 1: General Requirements for Safety [UL 60601-
1:2003]
– Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and
Essential Performance[CAN/CSA C22.2 No. 60601-1:14]
Page 6
– Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and
Essential Performance [ANSI /AAMI ES60601-1:2005/(R)2012, AND C1:2009 AND A2:2010(R)2012]
– Biological Evaluation of Medical Devices – Part 1: evaluation and testing [ISO 10993-1:
2009]
– Standard Means for Reporting the Acous tic Output of Medical Diagnostic Ultrasonic
Equipment [IEC 61157:2007]

Sta te me nts

Nemk o-CCL mark with the indicators “C” and “US” means that the product is certified for both the U.S. and Canadian markets, to the applicable U.S. and Canadian standards.
This mark certifies that the product conforms to applicable EEC standards and has been certified by the European certification agency.
This mark certifies that the product conforms to applicable EEC standards.
The GMP symbol indicates the Good Manufacturing Practice in accordance with the Korean quality management regulations .
Page 7
Precautions for Use
Be sure to read this Service Manual thoroughly to familiarize yourself with the operation of the product and the relevant safety information before attempting to use the product.
Keep this service manual near the product and refer to it when using the product.
Please familiarize yourself with the safety precautions in “Chapter 2. Safety ” and “Chapter
10. Maintenance” in particular.
This service manual does not inc lude diagnosis results or opinions. Also, check the reference information for the measured area of the body before using the application’s measurement results in any diagnosis.
This product is an ultrasound diagnosis device and cannot be used from the user’s PC. The manufacturer is not responsible for any problems that may be caused by s uch attempts.
This product must only be used by persons who have sufficient knowledge of and/or a qualification in clinical pathology. Unqualified persons are prohibited from us ing the product.
The manufacturer is not responsible for any damage to this product caus ed by user careless ness and/or neglect.
Product orders are based on individually agreed specifications and may not include all functions specified in this service manual.
Some functions or options, probes, and the like may not be used in certain countries.
All reference material on standards, regulations, and related revisions are valid at the time of
the public ation of this service manual.
Screen images in this service manual are examples and may differ from the actual screen or system.
The content of this service manual is subject to change without prior notice.
Products that are not manufactured by Samsung Medison are indicated with the trademarks
of their respective owners.
The following terms are used to highlight precautions that the user must pay particular attention to:
Page 8
DANGER
Disregarding this ins truction may result in death, serious injury, or other dangerous situations.
WARNING
Follow these ins tructions to prevent a serious accident or damage to property.
CAUTION
Follow these ins tructions to prevent
NOTE
The accompanying information covers installation, operation, and maintenance procedures that require careful attention from the user, but have little chance of leading
directly to a dangerous situation.
minor accidents and damage to property.
Page 9
Revision History
The revis ion history of this servic e manual is as follows:
VERSION DAT E REASON FOR CHANGE
1.00.00 2017.04.01 Initial Release
If You Need Assistance
If you need a service manual or any assistanc e with the product, pleas e c ontact the Samsung Medison customer service department or your local vendor.
Page 10
Patient Privacy Policy

Treatment of Patients’ Personal Information

One of Samsung Medison’s s ocial responsibilities is to recognize the importance of patient privacy and to handle and protect the personal information provided by our patients in an appropriate manner.
Patient information stored in repaired and demo products should be handled as follows:
Patient Information: Refers to information, such as IDs, names, DOBs, and images, which can be us ed to identify specific patients.
Patient Information Management: Personal information provided by patients mus t be managed carefully and solely for the purposes of repairs and demonstrations of Samsung Medison products, and they must not be used for unauthorized purposes, be lost, be revised, or be shared with a third party.

Management of Patient Information on Demo and Trade Products

Patient information stored on a product during the course of its use must be handled as the patient des ires. (delete or backup)
When completing a demonstration and/or collecting a trade product, patient information must be deleted if so requested by the client, to prevent its unintended distribution to outs ide parties .

Management of Patient Information on Repaired Hospital Products

Before removing the product from hospital premises for repairs, all patient information should be back ed up to a storage device such as external HDD or memory, and then handed over to the hospital.
Before removing the product from hospital premises, each patient should sign the patient privacy agreement.
Page 11
Request for Repairs and Patient Privacy Agreement
Date:
YY/MM/DD Name:
(Signature)
Customer Information
Hospital
Handler
Product Information
Product name
Software version
S/N
Symptom
Date: YYYY/MM/DD
Stored item
Outsourced product
Adminis tration
Handling date YY MM DD
Customer service representative
Please sign ins ide the box below if you agree to the following:
Data saved on memory devic es of products being repaired (e.g., hard disks) may become
lost during tests and repairs.
□ Data saved on a product being repaired must be back ed up by the customer.
Samsung will not be held responsible for any loss of data not backed up.
Samsung Medison does not use patient information for unauthorized purposes, modify
such information, or provide it to third parties.
Customer Signature
None Main unit HDD □ DVD/CD
USB Photo □ Other ( )
None
Provide details if applicable ( )
I entrust my product to ( ) for repair and agree to the policies above.
Page 12
Table of Contents 1
Table of Contents
Chapter 1 Introduction 1
1.1. Product Specification s............................................................................................. 2
1.2. Product Configuration ............................................................................................. 5
1.2.1. Monitor ........................................................................................................ 7
1.2.2. Control Panel ............................................................................................... 9
1.2.3. Cons ole ..................................................................................................... 16
1.2.4. Perip heral Devices ..................................................................................... 18
1.2.5. Probes....................................................................................................... 20
1.2.6. Accessories ............................................................................................... 22
1.2.7. Optional Acces sories .................................................................................. 23
1.2.8. Optional Functio ns ..................................................................................... 24
Chapter 2 Safety 1
2.1. Purpose of Use ........................................................................................................ 2
2.1.1. Contrain dications ......................................................................................... 2
2.2. Sa fe ty Information.................................................................................................... 3
2.2.1. Safety Symbols ............................................................................................ 3
2.2.2. Symbols ...................................................................................................... 7
2.2.3. Labels ......................................................................................................... 7
2.3. Electrical Safety ....................................................................................................... 8
2.3.1. Prevention of Electric Shocks ........................................................................ 8
2.3.2. ECG-Rel ated Informatio n ............................................................................ 10
2.3.3. ESD .......................................................................................................... 10
2.3.4. EMI ............................................................................................................11
2.3.5. EMC...........................................................................................................11
2.4. Mechanical Safety ...................................................................................................18
2.4.1. Moving the Eq uipm ent ................................................................................ 18
2.4.2. Precautions for Use .................................................................................... 19
2.5. Biological Safety .....................................................................................................22
2.5.1. ALARA Principl e......................................................................................... 22
2.6. Protecting the Environme nt ....................................................................................35
2.6.1. EU WEEE Directi ve .................................................................................... 35
2.6.2. EU REACH Regulation ............................................................................... 35
2.6.3. State of California Proposition 65 Warnin g ( US Only ) .................................... 35
Page 13
2 HS40 Service Manual
Chapter 3. Installing Product 1
3.1. Transporting .............................................................................................................3
3.1.1. Precautions During Trans po rt ....................................................................... 3
3.1.2. Brakes ........................................................................................................ 3
3.1.3. Precautions on Ramps ................................................................................. 3
3.1.4. Humi dity and Temperatu re ............................................................................ 4
3.2. Unpacking the Product .............................................................................................5
3.2.1. Dismantlin g the Product's Box ...................................................................... 5
3.2.2. Accessories ................................................................................................. 6
3.2.3. Release of the Locking Mecha nism ............................................................... 6
3.3. Installation Environment...........................................................................................7
3.3.1. Cauti on ....................................................................................................... 7
3.4. Installing the Product................................................................................................8
3.4.1. Installatio n Sa fety......................................................................................... 8
3.4.2. Connec ting P erip he rals ...............................................................................10
3.5. System Power ......................................................................................................... 13
3.5.1. Turnin g the Power On .................................................................................13
3.5.2. Shutting Down the System...........................................................................13
3.6. Sy stem Se tti ng s...................................................................................................... 14
3.6.1. General Sys tem Settings .............................................................................15
3.6.2. General ......................................................................................................15
3.6.3. Display .......................................................................................................19
3.6.4. Patient .......................................................................................................20
3.6.5. Quick Preset ...............................................................................................23
3.6.6. EzExam+....................................................................................................24
3.6.7. Imaging ......................................................................................................27
3.6.8. Application/Preset .......................................................................................30
3.6.9. Application..................................................................................................32
3.6.10. Measurement Settings.................................................................................33
3.6.11. Report ........................................................................................................49
3.6.12. Annotatio n ..................................................................................................52
3.6.13. Body Marker ...............................................................................................57
3.6.14. Customize ..................................................................................................60
3.6.15. Peripherals .................................................................................................64
3.6.16. Connecti vity ................................................................................................67
3.6.17. S e r vic e .......................................................................................................86
3.6.18. He
lp ...........................................................................................................86
Page 14
Table of Contents 3
Chapter 4. Product Inspection 1
4.1. Inspecting Function s ............................................................................................... 2
4.1.1. Basic Inspections ......................................................................................... 2
4.1.2. Detail ed Inspec tions ..................................................................................... 3
Chapter 5. Product Structure 1
5.1. Overview .................................................................................................................. 3
5.2. System Block Diagram............................................................................................. 5
5.2.1. System Block Diagram ................................................................................. 5
5.2.2. System Rack Design .................................................................................... 6
5.3. Basi c Structure of the Product................................................................................. 7
5.3.1. Electronic Structure ...................................................................................... 7
5.3.2. Ultraso un d System Part ................................................................................ 7
5.3.3. PC Part ....................................................................................................... 8
5.3.4. User Inte rface Part ....................................................................................... 8
5.3.5. Powe r Part ................................................................................................... 8
5.4. Ultrasound System Part ........................................................................................... 9
5.4.1. Main Functions of PSA (Pr ob e Select Assembly) ............................................ 9
5.4.2. Beam Forme r Board ....................................................................................11
5.4.3. Back End Board ......................................................................................... 15
5.5. PC Part ....................................................................................................................20
5.5.1. PC Mod ule................................................................................................. 20
5.5.2. Rear IO Board............................................................................................ 21
5.5.3. Softw ar e DS C ............................................................................................ 23
5.6. User Interface Part ..................................................................................................24
5.6.1. Control Panel ............................................................................................. 24
5.6.2. Main Monitor .............................................................................................. 26
5.6.3. To u c h-Sc re en............................................................................................. 27
5.6.4. Display Layout ........................................................................................... 28
5.7. Power Part ..............................................................................................................29
5.7.1. ADM .......................................................................................................... 29
5.7.2. Power Block Diagram ................................................................................. 30
5.7.3. Specificatio n .............................................................................................. 30
5.8. Interconnect Diagram..............................................................................................31
5.8.1. HS40 Interconnect Diagram ........................................................................ 31
5.8.2. Cable Description ....................................................................................... 32
Page 15
4 HS40 Service Manual
Chapter 6. Service Mode 1
6.1. Se rvice Mode ............................................................................................................3
6.1.1. How to Enter the Servic e Mode..................................................................... 3
6.2. Service Information ..................................................................................................4
6.3. Configuration ............................................................................................................5
6.3.1. Cloc k ........................................................................................................... 5
6.3.2. Keyboard/Region ......................................................................................... 5
6.3.3. TCP /IP ........................................................................................................ 6
6.3.4. Option ......................................................................................................... 7
6.3.5. User Acc ou nt and User Account Mana gem ent ............................................... 8
6.3.6. Printer ......................................................................................................... 9
6.4. System .................................................................................................................... 10
6.4.1. Upgra de Softw ar e .......................................................................................10
6.4.2. System Serial Num ber................................................................................. 11
6.4.3. Window Ex pl orer.........................................................................................12
6.4.4. Install R eco very System ..............................................................................13
6.5. Diagno stics............................................................................................................. 14
6.5.1. Control Panel Test .......................................................................................14
6.5.2. Keyboard Test.............................................................................................14
6.5.3. Power On Sel f Test .....................................................................................14
6.5.4. Built in Self Test ..........................................................................................14
6.5.5. Monitor Test ................................................................................................14
6.6. Backup and Restore ............................................................................................... 15
6.6.1. Backup .......................................................................................................15
6.6.2. Restore ......................................................................................................15
6.6.3. Log Backup ................................................................................................15
6.7. Report ..................................................................................................................... 16
6.7.1. Report ........................................................................................................16
6.8. Demo Play............................................................................................................... 17
6.8.1. Demo Play..................................................................................................17
6.9. Enter Installation Key.............................................................................................. 18
Page 16
Table of Contents 5
Chapter 7. Troubleshooting 1
7.1. Power Issue s............................................................................................................ 2
7.1.1. Power Does Not Turn On .............................................................................. 2
7.1.2. Power Does Not Turn Off .............................................................................. 2
7.1.3. Power Turns Off by Its el f............................................................................... 2
7.2. Monitor..................................................................................................................... 3
7.2.1. Nothi ng Is Displ ay ed o n the Scre en ............................................................... 3
7.2.2. Screen is Discolor ed ..................................................................................... 3
7.3. Error Message s ........................................................................................................ 4
7.3.1. Erro r Occurs Durin g B oot ing ......................................................................... 4
7.4. Image ....................................................................................................................... 4
7.4.1. 2D Mode: There is No IMAGE ECHO or IMA GE FORMAT .............................. 4
7.4.2. Lines (Noise) Appea r in 2D Mode Ima ge ........................................................ 4
7.4.3. M, C, PW, CW Mode Trouble ........................................................................ 4
7.5. USB Drive Unre cognized.......................................................................................... 5
7.5.1. Failed to r ecog nize the USB ( Rem ovable device) device ................................ 5
7.6. Trouble Shooting Tree ............................................................................................. 6
Chapter 8. Disassembly and Assembly 1
8.1. Preparation .............................................................................................................. 2
8.1.1. Preparati on .................................................................................................. 2
8.2. Product Di sassembly ............................................................................................... 3
8.2.1. Front Sid e Disassembly ................................................................................ 3
8.2.2. Rear Side Disassembly ................................................................................ 6
8.2.3. Control Panel Disass em bly ........................................................................... 7
8.2.4. Touch Panel Disassembly ............................................................................. 8
8.2.5. Track Ball .................................................................................................... 9
8.2.6. Speaker Disassembly ................................................................................. 10
8.2.7. Monitor Disassembly ...................................................................................11
8.2.8. Monitor A RM Disassem bly .......................................................................... 12
8.3. Installation of Optional Parts ..................................................................................13
8.3.1. Overvie w of Optional Parts ......................................................................... 13
8.3.2. Option Pa rts............................................................................................... 14
8.3.3. Installatio n of External Physical Keyboard .................................................... 16
8.3.4. Installatio n of Gel Warme r ........................................................................... 16
8.3.5. Installatio n of Endo Cavity Holder ................................................................ 17
8.3.6. Installatio n of Holder Chart .......................................................................... 18
8.3.7. I
nstallation of Rear Tray .............................................................................. 18
Page 17
6 HS40 Service Manual
8.3.8. Installatio n of Printer Tray (1/2) ....................................................................19
8.3.9. Installatio n o f Printer Tray (2/2) ....................................................................19
Chapter 9 Probes 1
9.1. Probes ......................................................................................................................2
9.1.1. Probe List .................................................................................................... 2
9.1.2. Ultraso un d Transmission Gel ........................................................................ 9
9.1.3. Sheaths ......................................................................................................10
9.1.4. Probe Safety Precautio ns ............................................................................ 11
9.1.5. Cleanin g an d Disinfecting th e Probe.............................................................13
9.2. Bi o p sy ..................................................................................................................... 22
9.2.1. Biopsy Kit Compon ents ...............................................................................22
9.2.2. Using the Bio psy Kit ....................................................................................23
9.2.3. Assembling the Biopsy Kit ...........................................................................26
9.2.4. Cleanin g an d Disinfecting th e Biopsy Kit.......................................................30
10. Maintenance 1
10.1. Operational Environment ..........................................................................................2
10.2. Product Maintena nce ................................................................................................3
10.2.1. Clea ning and Disinfecting ............................................................................. 3
10.2.2. Clea ning...................................................................................................... 3
10.2.3. Cleaning Air Filters ....................................................................................... 5
10.2.4. Accuracy Checks ......................................................................................... 6
10.3. Data Maintenance .....................................................................................................7
10.3.1. User Settings Backup ................................................................................... 7
10.3.2. Backing Up Patie nt Information ..................................................................... 7
10.3.3. Software...................................................................................................... 7
Chapter 11. Service Part List 1
11.1. Body Cover Parts......................................................................................................2
11.2. Control Panel Pa rts...................................................................................................5
Page 18
Chapter 1
1. Chapte r 1 I ntroduc tion

Introduction

1.1. Product Specifications .............................................................................................. 2
1.2. Product Configuration ............................................................................................... 5
1.2.1. Monitor........................................................................................................... 7
1.2.2. Cont rol Pan el ................................................................................................. 9
1.2.3. Cons ole ....................................................................................................... 16
1.2.4. Peripheral D e vices ........................................................................................ 18
1.2.5. Prob es ......................................................................................................... 20
1.2.6. Accessories.................................................................................................. 22
1.2.7. Optional Access ories .................................................................................... 23
1.2.8. Optional Functions ........................................................................................ 24
Page 19
1 – 2 HS40 Service Manual

1.1. Product Specifications

Height: 1,620 mm (with Monitor)
Physical
Dimensions
Imaging Modes
Width: 520 mm
Depth: 730 mm (with Keyboard)
Weight: 54 kg (without acces sories)
Weight: Approx. 57 kg (with Safe Working Load)
2D Mode
M Mode
Color M Mode
Anatomical Mode
Color Doppler Mode
Pulsed W ave (PW) Spectral Doppler Mode
Continuous W ave (CW) Doppler Mode
Tissue Doppler Imaging (TDI) Mode
Tissue Doppler Wave (TDW) Mode
Power Doppler (PD) Mode
ElastoScan Mode
3D/4D/XI STIC imaging Mode
Dual Mode
Quad Mode
Combined Mode
Simultaneous Mode
Zoom Mode
Gray Scale 256 (8 bits )
Transmit focusing, maximum of eight points (four points simultaneously
Focusing
Probes
(Ty pe BF/IPX7)
selectable)
Digital dynamic receive focusing (continuous)
Linear Array
LA3-16AD
Curved Array
CA2-8AD, CF4-9
Phased Array
PN2-4
Endo Cavity
EVN4-9
3D
VN4-8, V5-9
Page 20
Pe ncil
DP2B
Chapter 1. Introduction 1 – 3
Probe
Connections
Monitor
ECG USB Type (Type CF)
Rear Panel
Input/Output
Connections
3 Probe Connectors, 4 Probe Connectors for option
CW Probe Connector
Main Monitor
Number of Pix el: 1920 x 1080
21.5 inch LCD Monitor (LED Backlight unit, hereafter referred to as “LCD monitor”)
Touch Screen Monitor
Number of Pix el: 1280 x 800
10.1 inch LCD Monitor (LED Backlight unit, hereafter referred to as “LCD monitor”)
Audio Output Port (Right/Left)
VGA monitor
LAN
USB Port
CW
HDMI output
HDMI Input (HDMI Input is currently not s upported.)
Image Storage
Application
Electrical
Parameters
Measurement
Packages
Signal
Processing
(Pre-process ing)
Maximum 45,000 frames for Cine memory
Maximum 14,000 Lines for Loop memory
Image filing s ystem
Obstetrics, Gynecology, Urology, Abdomen, Cardiac, Vascular, Small Parts, TCD, MSK, Pediatric
100 - 240VAC, 620VA, 50/60Hz
Abdomen, Obs tetrics, Gynecology, MSK, Pediatric, Small Parts, Urology, Vascular, Cardiac
* Refer to the ‘Chapter 8. Measurements and Calculations’ for additional information.
Acoustic Power Control
Analog TGC Control
Dynamic Aperture Control
Dynamic Apodization Control
Page 21
1 – 4 HS40 Service Manual
Dynamic LPF Control
Digital TGC Control
Slider TGC Control
Mode-Independent Gain Control
Black Hole/Noise Spike Filtering
Signal
Processing
(Pos t-
processing)
Measurement
1D Lateral/Axial Filtering
2D Edge/Blurring Filtering
Frame average
M/D Mode Sweep Speed Control
Zoom
Image View Area Control
Image Orientation (left/right and up/down)
Trackball operation of multiple cursors
• 2D mode: Linear measurements and area measurements using elliptical approximation or trace
• M mode: Continuous readout of distance, time, and slope rate
• Doppler mode: Velocity and trace
DVD Multi-Dr i ve
Digital B/W Video Printer
Digital Color Video Printer
USB Printer
Auxiliary
User Interface English, German, French, Spanish, Italian, Russ ian, Chinese, Portuguese
Pressure Limits
Humidity Limits
Temperature
Limits
DVD Recorder
Foot switch (IPX8)
USB Flash Memory Media
USB HDD
USB ECG
Monitor
Operating: 700 – 1060hPa
Storage: 700 – 1060hPa
Operating: 30 – 75%
Storage & Shipping: 20 – 90%
Operating: 10 – 35°C
Storage & Shipping: -25 – 60°C
Page 22
Chapter 1. Introduction 1 – 5

1.2. Product Configuration

This product consists of monitor, control panel, console, peripheral devices and probes.
Monitor
Monitor arm
Touch Panel
Speaker
Control Panel & Handle
USB port
Probe holder
Lift & Lift Lever
ECG port
Outlet for internal peripheral
devices
[Figure 1.1 Front of the Product]
Probe port
CW Port
Wheel
Brake
Filter
Page 23
1 – 6 HS40 Service Manual
Storage compartments
Ventilation
Cable holder
Rear panel
ID label
Power connection
[Figure 1.2 Back of the Product]
Page 24
Chapter 1. Introduction 1 – 7
1.2.1. Monitor
Ultrasound images and other information are displayed on the color LCD monitor.
1.2.1.1. Screen Layout
The monitor displays ultrasound images, operation menus and a variety of other information.
[Figure 1.3 Monitor Display]
Title Area
Displays patient information, hospital n ame, a pplicatio n, frame rate, dept h, probe i nform ation, acoustic output information, and the current date and time.
Image Area
Displays ultrasound images. Image information, annotation, and measurement information are also displayed.
Thumbnail Area
Images s aved by pressing the preset Store button are shown in the thumbnails. When you save Single sc reens, up to 4 images are shown in a list; for Quad screens, up to 12 images are shown. Click it with the pointer to enlarge the preview image.
User Information and Status Information Area
Information that is useful to the user, such as current system status , image information, selectable items, etc., is displayed.
User Key (User Defined Key) Area
Settings for User Defined Key s, including the positions of Set and Ex it buttons, are displayed. You can change the setting of each button in Setup > Customize > Buttons.
Page 25
1 – 8 HS40 Service Manual
For information on User Key Setup, please refer to ‘Chapter 3. Utilities’.
NOTE
TIP
Principles of Operation of the Diagnostic Ultrasound System
Medical ultrasound images are created by digital memory and computer when they convert the high-frequency wave signals that are transmitted and received by the probe.
As ultrasound waves propagate through the human body, they generate reflected signals whenever they encounter a change in density. For example, reflected signals are generated when signals pass from fatty tissues to muscle tissues. Reflected signals return to the probe, where they are converted into electronic signals. The reflected signals are amplified and processed by analog and digital circuits that have filters for various frequencies and response time options. Then they are again converted into high-frequency elec tronic signals, and saved as a series of digital image signals. The monitor displays the image signals stored on the storage device in real time. The entire process of transmitting, receiving, and processing signals is controlled by the computer.
Page 26
1.2.2. Control Panel
The system can be controlled by using the control panel.
Chapter 1. Introduction 1 – 9
[Figure 1.4 Control Panel]
The control panel consists of a soft menus, buttons, dials, dial-buttons, a slider, and a trackball.
The dial-button can be used both as a dial and a button.
Page 27
1 – 10 HS40 Service Manual
1.2.2.1. Functions of the Control Panel
The following are the descriptions and instructions for the controls on the control panel. For more information on controls with multiple functions, s ee ‘Chapter 3. Utilities’ and later in this
manual.
Button Turns the system on/off.
On/Off
End Exam
SonoView
Report
EzExam
Button
Finishes the exam of the currently selec ted patient and resets the related data.
Button Displays the screen for viewing and managing stored images.
Button
Displays the Report screen that shows the measurement results of the current application and other information.
When you press the dial-button, the menu items that are available in the current scan mode are shown on screen.
Dial-button
Rotating the EzEx am dial-button to the right s elects the menu one row above the current menu selected, and rotating it to the left selec ts the menu one row below.
Button Allows the user to enter a Body Marker over an image.
Adjusts the angle of the sample volume in Spectral Doppler mode. It is also used to adjust the Arrow’s angle or the probe
Dial-button
angle for a Body Marker.
In 3D View, turning this dial-button will rotate the image around the x-axis.
Dial-button
Dial-button
Enter or exit PW Spectral Doppler Mode. Rotate this dial­button to adjust the Gain value.
In 3D View, turning this dial-button will rotate the image around the y-axis.
Enter or exit Color Doppler Mode. Rotate this dial-button to adjust the Gain value.
In 3D View, turning this dial-button will rotate the image around the z-axis.
Page 28
Chapter 1. Introduction 1 – 11
Dial-button
Button
Button
Button Enter or exit Power Doppler Mode.
Button Use this button to enter or exit 3D or 4D Mode.
Button
Press this button to enter 2D Mode. Rotate this dial-button to adjust the Gain value.
Enter or exit M Mode. Rotate this dial-button to adjust the Gain value.
Enter or exit CW Spectral Doppler Mode. Available only with the phased array probe.
Press this button to turn the Quick Sc an function on. The ‘Q scan’ mark will appear at the top of the image. This can be used in applications for all probes.
Button
Button
Switch Adjusts the scanning depth of the image.
Switch Moves the focus to the target area for observation.
Dial-button
Compares two independent images. The active image mode is shown on the left.
Compares two independent images. The activated image mode is shown on the right.
Makes the Zoom Box appear. In order to exit Zoom Mode, press the Ex i t button.
Page 29
1 – 12 HS40 Service Manual
Set/Exit
Clear
Pointer
Change
Button
Button
Button
Button
Button
Button
Pauses an image being scanned or reactivates a paused image.
Stands for User Key ; the user c an assign functions they need to each button. The function for each button can be ass igned in Setup > Customize > Buttons.
The user can select, set up, and use Set/Set, Se t/Ex it, or Exit/Set. The function for each button can be assigned in Setup > Customize > Buttons.
– Set: Selects an item or value using the trackball.
Alternatively, this can be us ed to change the function of the trackball.
– Exit: Exits the function currently being used and returns to
the previous screen.
Deletes t ext, Arrows, Body Markers, measurement results, etc., display ed on an image.
In Scan Mode, an arrow-shaped pointer appears on the screen.
Changes to a different function that can be s upported by the current trackball setting.
Calculator
Trackball
Button Allows the user to place text over an image.
Button Starts measurements by application.
Trackball
Moves the cursor on the sc reen. Also, scrolls through Cine images.
Page 30
Chapter 1. Introduction 1 – 13
For further details about setting up Quick Preset, please refer to ‘Setup > System > Quick Preset’ in ‘Chapter 3. Utilities’.
1.2.2.2. Touch Screen
The touch s creen is an operating tool that can be touched by the user to input data. The functio ns that are available in the current mode are shown in the form of buttons or a dial-button.
Touch Screen Layout
[Figure 1.5 Touch Screen Display]
① These buttons are always display ed on the touch screen. Buttons that are in use are
shown in blue and buttons that cannot be used are deactivated.
Patient
Probe
Quick Preset
1–4
Setup
NOTE
The Patient Information screen will appear, where y ou can select a Patient ID in the list or enter new patient information.
It displays the Probe Selection screen where you can select and modify the probe and application.
It shows up to 4 Preset buttons c onfigured at Setup.
The general s ystem settings that do not have a direct bearing on imaging are explained below.
Page 31
1 – 14 HS40 Service Manual
CAUTION
Too great a difference in the gain value settings of adjacent TGC sliders may cause stripes in an image.
② Displays the mode that is c urrently used.
③ This is the Keyboard button. Clicking this button will activate the On-screen Keyboard.
④ This is the TGC area. Adjust TGC Slide when TGC Curve is activated on the touch
screen. Move the slider to the right to increase Gain and brighten the image. TGC stands for Time Gain Compensation.
⑤ Menu Area: The menu items that are available in the current input mode are shown in the
form of buttons. The user can access the desired menu item by pressing the corresponding button. The menu currently in use is shown in blue. To change the values
in the menu, use the and buttons.
⑥ Soft Menu Area: The Soft Menu items that are available in the current input mode are
shown. Press or rotate the dial-buttons right below each menu.
TIP
When there are Two Soft Menus
When there are two menus available - upper and lower, both menus can be adjusted with the corresponding dial-button. Or tap the button for the menu you want to use on the touch screen and then us e the dial-button.
Page 32
1.2.2.3. Adjusting the Control Panel
CAUTION
You can move the control panel up and down only while pulling the lever at the side of the control panel.
Do not apply excessive force to the control panel.
Use the handles on the console when moving the product.
Adjusting the Height
Chapter 1. Introduction 1 – 15
[Figure 1.6 Control Panel]
NOTE
Page 33
1 – 16 HS40 Service Manual
1.2.3. Console
The console consists of two parts - the inner and outer units. The interior of the console mainly contains devices that produce ultrasound images. On the exterior of the console are various
connectors, probe holders, storage compartments , handles, and wheels, etc .
1.2.3.1. Rear Panel
A monitor and other peripheral devices, like a printer, are connected via the rear panel at the ba ck of the system.
① Audio port (Output): Used for audio signal output.
(Input: Not supported.)
② S-VHS Port (Output): Provides an S-VHS
connection for a VCR.
③ RGB Port (Output): Provides analog signal output
to be displayed on the monitor.
④ USB port: Used to connect to peripheral USB
devices.
⑤ Network port: Connect to a network. You can
transfer patient information to another server via DICOM network.
[Figure 1.7 Rear Panel]
⑥ HDMI port (Output): Outputs digital signals to the
monitor. (Input: Not supported.)
⑦ Trig port (In/Output): Not used.
Page 34
Chapter 1. Introduction 1 – 17
1.2.3.2. Power Connection Part
The power connection part is loc ated at the bottom on the rear panel.
[Figure 1.8 Power Connection Part]
① Power switch: Turns on the power of the product, and cuts off power if overcurrent or
overvoltage occurs.
② Power inlet: Accepts the power cord, which connects to an external power supply.
③ Power fuse: Cuts off power if overcurrent or overvolta ge occ urs.
④ Equipotential Terminal: This should be connected to the equipotential connection part in the
exam room.
1.2.3.3. Probe Holder
Probe holders are mounted at the left and right-hand sides of the control panel.
The holders can be removed for easy cleaning.
TIP
Hockey Stick Probe Holde r
Insert the Hockey stick probe holder into the probe holder to mount. It is provided as an option.
Page 35
1 – 18 HS40 Service Manual
1.2.4. Peripheral Devices
Peripheral devices can be connected to their corresponding ports on the left/right or rear sides of the console, as needed.
Do not install peripheral devic es that are not listed in this user manual in the
patient environment. If y ou install an unlisted device in the patient environment, it may cause an electrical hazard.
CAUTION
Do not c onnect additional external peripheral devices to the auxiliary socket
outlet. Doing s o may decrease safety level.
[Figure 1.9 Patient Environment]
Refer to the user manual of the peripheral device for its operating information.
NOTE
1.2.4.1. Internal Peripheral Devices
These are peripheral devices mounted in the sys tem.
DV D-Multi
DVD±R, DVD±RW, DVD-RAM, CD-R, CD-RW
Soli d Sta te Drive
SSD 512G
Page 36
Chapter 1. Introduction 1 – 19
CAUTION
When using a peripheral device via a USB port, always turn the power off before connecting/disconnecting the device. Connection/disconnection of USB devices while the power is on may lead to malfunction of the s ystem and the USB devices.
• Please check the port that the printer uses before connec ting. Printers should be connected to the printer port while the USB printer should be c onnected to
1.2.4.2. External Peripheral Devices
These are peripheral devices that can be connec ted for use when needed and are connected via the USB port located at the rear panel.
TIP
The following products are recommended:
Digital Video Printer
Black and White: Sony UP-D897 (UP-897MD, UP-X898MD, UP-D898MD), Mitsubishi
Color: Sony UP-D25MD, Mitsubis hi CP30DW
US B Pri nter
Samsung CLP-620NDK, Samsung ML-2950
DV D Re corde r
Sony DVO-1000MD
CAUTION
• The console’s USB ports are located on the rear panels of both the control panel and the console.
• W e recommend that you connect USB storage devices (flash memory media, etc.) to the ports on the c ontrol panel, and other USB peripheral devices to the rear panel for convenience.
P95DE (P95DW, P95D, P95DE)
You must install a printer and drivers that are compatible with Microsoft
Windows 7 (Englis h version). Contact Samsung Medison customer support division for inquiries about printer driver installation.
When connecting the printer, ensure that the printer is configured under
Microsoft W indows or system setup and has been chosen as the default printer.
the USB port.
Foot Switch
3 Pedals HID Type
To configure the foot s witch function, go to Setup > Customize > Buttons > Foot Switc h.
You can individually set the Left, Middle, and Right foot s witches to one of the following: Exi t, Freeze, U1, U2, U3, U4, Update, Single, or Quad.
Page 37
1 – 20 HS40 Service Manual
CAUTION
Connecting or removing the Foot Switch while the product is turned on may cause the sys tem to malfunction.
For information on probes, refe Manual.
If the probe foreign objects.
M i sc.
Flash Memory Media
The system cannot recognize USB 1.1 flash memory. Remove the flash
memory from the console and equip again with an appropriate device.
Regarding file formats that are not ordinarily saved: Please chec k first to see
NOTE
if it is possible to save the file format on a desktop PC before trying to save the file on flash memory.
Do not use flas h memory media which contain anti-virus programs or are
defective. Otherwis e, the product may fail to work properly.
1.2.5. Probes
Probes are devices that generate ultrasound waves and process reflected wave data for the purpose of image formation.
r to ‘Chapter 9. Probes’ and the Reference
NOTE
1.2.5.1. Connecting Probes
Be sure to connec t or disconnect probes when the power is off to ensure the safety of the syste m and the probes.
1. Connect probes to the probe connectors on the front panel of the system. Up to four (options inclusive) (five including CW) probes may be connected.
2. Lift up the probe’s lockdown switch and disconnect the probe.
3. Connect the probe to the probe port.
4. Push down the probe’s lock down switch to lock it in place.
does not connect properly, try reconnecting it after removing any
NOTE
Page 38
[Figure 1.10 Probe Connector]
Chapter 1. Introduction 1 – 21
Page 39
1 – 22 HS40 Service Manual
Accessories c an be
1.2.6. Accessories
An accessory box containing the items below is supplied with the product.
different according to the country.
NOTE
[Figure 1.11 Accessories]
Page 40
Chapter 1. Introduction 1 – 23
1.2.7. Optional Accessories
The following optional accessories are provided with the product.
Page 41
1 – 24 HS40 Service Manual
1.2.8. Optional Functions
HS40 products have the following options based on the version:
• 4D
• 3D XI™
• DICOM
• EzExam+
• MultiVision
• 5D NT
• CW Function
• Cardiac Measurement
• Strain+
For further information about optional functions, please refer to the relevant chapters in this manual.
XI STIC
SEE Stream (RU region only)
Needle Mate
Realistic Vue
Mobile Export
5D Follicle
Auto IMT+
ElastoScan
Panoramic
Page 42
Chapter 2
2. Sa fety Chapte r 2 S afety

Safety

2.1. Purpose of Use ...........................................................................................................2
2.1.1. Cont rain dications ............................................................................................2
2.2. Safe ty Informa tion ......................................................................................................3
2.2.1. Safety Symbols ...............................................................................................3
2.2.2. Symbols .........................................................................................................7
2.2.3. Lab els ............................................................................................................7
2.3. Electrical Safety..........................................................................................................8
2.3.1. Prevention of Electric Shocks ...........................................................................8
2.3.2. ECG-Rel ated Information ............................................................................... 10
2.3.3. ESD ............................................................................................................. 10
2.3.4. EMI .............................................................................................................. 11
2.3.5. EMC............................................................................................................. 11
2.4. Mechanical Safety .................................................................................................... 18
2.4.1. Movi ng the Equipm ent ................................................................................... 18
2.4.2. Precautions for Use ....................................................................................... 19
2.5. Biological Safety....................................................................................................... 22
2.5.1. ALARA Principl e ............................................................................................ 22
2.6. Protecting the Environment...................................................................................... 35
2.6.1. EU WEEE Directive ....................................................................................... 35
2.6.2. EU REACH Reg ulation .................................................................................. 35
2.6.3. State of Cali fornia Prop osition 65 Warning (US Only) ....................................... 35
Page 43
2 – 2 HS40 Service Manual
For detailed information on applications and presets, please refer to ‘Chapter Introduction’ and ‘Chapter
CAUTION

2.1. Purpose of Use

Ultrasound diagnostic system and probes were designed for obtaining ultrasound images and analyzing body fluid.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal
(Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
9. Probes’ in this user manual.
NOTE
2.1.1. Contraindications
This product must not be used for ophthalmological applications, or any other use that involves the ultrasound beam passing through the eye.
Federal law restric ts this devic e to sale by or on the order of a physician.
The method of application or use of the device is described in the manual
‘Chapter 6. Starting Diagnos is’ and ‘Chapter 7. Diagnosis Modes’.
1.
Page 44
Chapter 2. Safety 2 – 3

2.2. Safety Information

Please read the following safety information before using this product. It is relevant to the ultrasound s ystem, the probes, the recording devices, and any of the optional equipment.
The product is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of the medical device.
2.2.1. Safety Symbols
The International Electro Technical Commiss ion (IEC) has established a set of symbols for medical electronic equipment, which c lass ify a connection or warn of potential hazards. The classifications and symbols are shown below.
Symbols Description
WARNING: The acc ompanying information must be followed to prevent serious accidents and/or damage to property.
CAUTION: The accompanying information helps to prevent minor accidents and/or damage to property.
Refer to the user manual.
Follow the user manual.
CAUTION: Risk of electric shock
Type BF applied part (Classification based on degree of protection against electric haz ard)
Defibrillation-proof type CF applied part (Classification bas ed on degree of protection against electric hazard)
Power On/Off
Page 45
2 – 4 HS40 Service Manual
Symbols Description
Power On
Power Off
V~
Power On for part of the product
Power Off for part of the product
Alternating current voltage source
Direct current voltage source
Dangerous voltage (Indicates dangerous voltages over 1000V AC or 1500V DC)
Protective earth (ground)
Equipotentiality
Data output port
Data input port
Data Input/Output port
Page 46
Symbols Description
IPX 1
IPX 7
IPX 8
Chapter 2. Safety 2 – 5
Output port
Input port
Print remote output
Foot Switch Port
ECG port
USB port
Network port
Microphone Port
Probe port
Protected against vertically falling water drops
Protected against the effects of temporary immersion in water
Protected against the effects of continuous immersion in water
CAUTION: Elec trostatic sensitive devices (ESD)
Do not sit on the product.
Page 47
2 – 6 HS40 Service Manual
Symbols Description
Do not push the product.
Do not lean against the product.
DANGER: Do not place your fingers, hands , or any other part of your body in this spac e.
Do not place objects on the monitor that are heavier than the weight specified by the symbol.
Do not place heavy objec ts on the monitor while the monitor mount is folded.
Do not push down on the monitor while the monitor mount is folded.
Page 48
Chapter 2. Safety 2 – 7
2.2.2. Symbols
Symbols Description
Authorised Representative In The European Community
Manufacturer
2.2.3. Labels
Phrases containing the words ‘warning’ and/or ‘caution’ are display ed on the product’s surface in order to protect it.
Page 49
2 – 8 HS40 Service Manual

2.3. Electrical Safety

This equi pment is categorized as a Cl ass I device with Typ e BF or Ty pe CF (ECG) appli ed parts .
As for US requirement, the LEAKAGE CURRENT might be measured from a
center-tapped circuit when the equipment connects in the United States to 240V supply system.
CAUTION
2.3.1. Prevention of Electric Shocks
In a hospital environment, hazardous current c an form due to potential differences between exposed conductive parts and connected devices. The solution to the problem is consistent equipotential bonding. Medical equipment is connected with connecting leads made up of
sockets which are angled to the equipotential bonding network in medical rooms.
To help assure grounding reliability, connect to a “hospital grade” or “hospital
only” grounded power outlet.
[Figure 2.1 Equipotential Bonding]
Additional equipment c onnected to medical electrical equipment must comply with the respective IEC standards (e.g. IEC 60950/EN 60950 for data processing equipment, IEC 60601­1/EN 60601-1 for medical devices). Furthermore, all configurations shall comply with the requirements for medical electrical systems (see IEC 60601-1-1/EN 60601-1-1). Any body connecting additional equipment to signal input and output ports of medical electrical equipment should make sure that the equipment complies with IEC 60601-1-1/EN 60601-1-1.
Page 50
Chapter 2. Safety 2 – 9
cluding all of its externally mounted
• Never open the cover of the product. Hazardous voltages are present inside.
uipment. HF surgical equipment may
Electric shock may result if this system, in
recording and monitoring devices, is not properly grounded.
WARNING
CAUTION
All internal adjustments and replacements must be made by qualified Samsung Medison Customer Support Department personnel.
Always check the product’s housing, cables, cords, and plugs before using
the product. Disc onnect the power source and do not use the equipment if the housing is damaged such as cracked, and chipped, or if the cable is worn.
Always disc onnect the system from the wall outlet prior to cleaning the
system.
All patient contac t devices, s uch as probes and ECG leads, must be
removed from the patient prior to application of a high voltage defibrillation pulse.
The use of flammable anesthetic gas or oxidizing gases (N2O) should be
avoided. There is a risk of explosion.
Avoid installing the s ystem in such a way that it is difficult for the operator to
disconnect it from the power source.
Do not us e together with HF surgical eq
be damaged, which may result in fire.
The system must only be c onnected to a supply mains with protective earth
to avoid risk of elec tric shock.
The system has been designed for 100-240VAC; you should selec t the input
voltage of any connected printer and VCR. Prior to connecting a peripheral power c ord, verify that the voltage indicated on the power cord matches the voltage rating of the peripheral device.
An isolation transformer protects the system from power surges. The
isolation transformer continues to operate when the sy stem is in standby.
Do not immerse the cable in liquids. Cables are not waterproof.
Make sure that the inside of the system is not exposed to or flooded with
liquids. In such cases, fire, electric shock, injury, or damage to the product may occur.
The auxiliary socket outlets installed on this sys tem are rated 100-240VAC
with maximum total load of 150VA. Use these outlets only for supplying power to equipment that is intended to be part of the ultrasound sys tem. Do not connect additional multiple-socket outlets or extension cords to the system.
Do not c onnect any peripheral devices that are not listed in this manual to
the auxiliary socket outlet of the system. It may cause an electrical hazard.
Do not touch SIP/SOP and the patient simultaneously. There is a risk of
electric shock from leakage current.
Page 51
2 – 10 HS40 Service Manual
• This device is not intended to provide a primary ECG monitoring function, and
2.3.2. ECG-Related Information
therefore does not have means of indicating an inoperative electrocardiograph.
WARNING
Do not use ECG electrodes with HF surgical equipment. HF surgical
equipment may be damaged, which may result in fire.
Do not use ECG electrodes during c ardiac pacemaker procedures or any
procedures that involve other ty pes of electrical s timulators.
Do not us e ECG leads and electrodes in an operating room.
2.3.3. ESD
Electrostatic disc harge (ESD), commonly referred to as a static shock, is a naturally oc curring phenomenon. ESD is most prevalent during conditions of low humidity, which can be caused by heating or air conditioning. The s tatic shock or ESD is a discharge of the electrical energy build­up from a charged individual to a less or non-charged individual or object. An ESD occurs when an individual with an electrical energy build-up comes into c ontact with conductive objects suc h as metal doorknobs, file cabinets, computer equipment, and even other individuals.
The level of electrical energy discharged from a system user or patient to an
ultrasound system can be significant enough to cause damage to the system or probes.
CAUTION
Always perform the pre-ESD preventive procedures before using connectors
marked with the ESD warning label.
– Apply anti-static spray to carpets or linoleum.
– Use anti-static mats.
– Ground the product to the patient table or bed.
It is highly recommended that the user be given training on ESD-related
warning symbols and preventive procedures.
Page 52
Chapter 2. Safety 2 – 11
CAUTION
In cases where EMI is causing disturbances, it may be necessary to relocate this system.
2.3.4. EMI
Although this system has been manufactured in c omplianc e with existing EMI (ElectroMagnetic Interface) requirements, use of this system in the presence of an electromagnetic field can cause degradation of the ultrasound image or product damage.
If this occ urs often, Samsung Medison suggests a review of the environment in which the system is being used, to identify possible sources of radiated emissions. These emissions could be from other elec trical devices used within the same room or an adjacent room. Communication devices such as cellular phones and pagers can cause these emissions. The existence of radios, TVs, or microwave transmission equipment nearby can also c ause
interference.
2.3.5. EMC
The testing for EMC (Electromagnetic Compatibility) of this system has been performed according to the international standard for EMC with medic al devic es (IEC 60601-1-2). This IEC standard was adopted in Europe as the European norm (EN 60601-1-2).
2.3.5.1. Guidance and Manufacturer ’s Declaration – Electromagnetic Emission
This product is intended for use in the electromagnetic environment spec ified below. The customer or the user of this product should ensure that it is used in such an environment.
Em i ssi on Te st Compliance Electromagnetic Environment – Guidance
RF Emission
CISPR 11
RF Emission
CISPR 11
Harmonic Emission
IEC 61000-3-2
Flicker Em ission
IEC 61000-3-3
Group 1
Class A
Class A
Complies
The Ultrasound System uses RF energy only for its internal functions . Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The Ultrasound System is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Page 53
2 – 12 HS40 Service Manual
CAUTION
When connecting other customer user’s responsibility to ensure the electromagnetic compatibility of the system.
WARNING
The use of cables, probes, and peripherals other than those spec ified may result in increased emission or decreased Immunity of the Ultra
2.3.5.2. Approved Cables, Probes and Peripherals for EMC
Cables
Cables connec ted to this product may affect its emissions; us e only the cable types and lengths listed in the table below.
Cable Typ e Length
VGA Shielded Normal
USB Shielded Normal
LAN(RJ45) Twisted pair Any
S-Video Shielded Normal
Foot Switch Shielded 2.99m
Audio R.L Shielded Normal
Parallel Shielded Normal
HDMI Shielded Normal
ECG 3-lead Shielded Normal
Probes
The image probe used with this product may affect its emission. The probe listed in ‘Chapter
9. Probes’ when used with this product, have been tested to comply with the group1 Clas s A emission as required by International Standard CISPR 11.
Peripherals
Peripherals used with this product may affect its emiss ions.
-supplied accessories to the system, it is the
sound System.
Page 54
Chapter 2. Safety 2 – 13
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601 Test
Level
±6KV Contact
±8KV Air
±2KV
for power supply lines
±1KV
for input/output lines
±1KV differential mode
±2KV common mode
Compliance Level
±6KV Contact
±8KV Air
±2KV
for power supply lines
±1KV
for input/output lines
±1KV differential mode
±2KV common mode
Electromagnetic
Environment –
Guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
<5% Uт for 0.5 cycle
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4­11
Power frequency (50/60Hz) magnetic field
IEC 61000-4-8
NOT E: Uт is the AC mains voltage prior to applic ation of the test level.
(>95% dip in Uт)
40% Uт for 5 cycle
(60% dip in Uт)
70% Uт for 25 cycle
(30% dip in Uт)
<5% Uт for 5 s
(<95% dip in Uт)
3 A/m 3 A/m
<5% Uт for 0.5 cycle
(>95% dip in Uт)
40% Uт for 5 cycle
(60% dip in Uт)
70% Uт for 25 cycle
(30% dip in Uт)
<5% Uт for 5 s
(<95% dip in Uт)
Mains power quality should be that of a typical commercial or hospital environment. If the user of this product requires continued operation during mains power interruptions , it is recommended that this product be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typic al location in a typic al commercial or hospital environment.
Page 55
2 – 14 HS40 Service Manual
a
Immunity
Te st
Conducted RF
IEC 61000-4­6
Radiated RF
IEC 61000-4­3
IEC 60601 Test
Level
3 Vrms
150 kHz to 80MHz
3 V/m
80 MHz to
2.5GHz
Compliance
Level
Electromagnetic Environment –
Guidance
3V Portable and mobile RF c ommunications
equipment should not be used closer to any part of the Ultrasound System, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended Separation Distanc e
80MHz to 800MHz
800MHz to 2.5GHz
3V/m Where P is the transmitter’s maximum
output power rating in watts (W ) according to the transmitter’s manufacturer, and d is the recommended separation distanc e in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site surve y , compliance level in each frequency range.
b
a
should be les s than the
Interference may occur in the vic inity of equipment marked with the following symbol:
NOT E 1) At 80MHz and 800MHz, the higher frequency range applies.
NOT E 2) These guidelines may not apply in all s ituations . Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To ass ess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in whic h the Ultrasound System is used exceeds the applicable RF compliance level above, the Ultrasound System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Ultrasound System, or using a shielded location with a higher RF shielding effectiveness and filter attenuation.
b
Over the frequency range 150kHz to 80MHz, field s trengths s hould be less than 3V/m.
Page 56
Chapter 2. Safety 2 – 15
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73 1 1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
CAUTION
If the s ystem is connected to other customer l equipment will work correctly in the presence of electromagnetic emiss ion phenomena.
2.3.5.3. Recommended Distance Between Wireless Communication Device and this Product
This product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this product can help to prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and this product as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum
Separation Distance According to Frequency of Transmitte r (m)
Output Pow er of
Transmitte r
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2.5GHz
(W)
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated by us ing the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOT E 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOT E 2) These guidelines may not apply in all s ituations . Electromagnetic propagation is affec ted by absorption and reflection from structures, objects and people.
2.3.5.4. Electromagnetic Environment – Guidance
It is recommended to use ultrasound sys tems in shielded locations offering RF shielding effectiveness, with shielded cables. Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than 3V/m.
It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to ensure that they meet the minimum specification.
ocal area network (LAN), Samsung Medison cannot guarantee that the remote
-supplied equipment, such as a
Page 57
2 – 16 HS40 Service Manual
2.3.5.5. Avoiding Electromagnetic Interference
Typic al interference on Ultrasound Imaging Systems varies depending on Electromagnetic phenomena. Please refer to the following table:
Imaging Mode
2D
M
Color
ESD1 RF2 Power Line3
Change of operating mode, sy stem settings, or system reset.
Brief flashes in the displayed or recorded image.
For sector imaging probes, white radial bands or flashes in the centerlines of the image.
For linear imaging probes, white vertical bands, sometimes more pronounced on the sides of the image.
Increase in the image background noise or white M mode lines.
Color flashes , radial or vertical bands, increase in background noise, or changes in color image.
White dots , dashes, diagonal lines, or diagonal lines near the center of the image.
White dots , dashes, diagonal lines, or increase in image background nois e
Color flashes, dots, dashes, or changes in the color noise level.
Horizontal lines in the
Doppler
spectral display or tones, abnormal noise in the audio, or both.
1. ESD caused by discharging of electric charge build-up on insulated surfaces or pers ons.
2. RF energy from RF transm itting equipment s uch as portable phones, hand-held radios, wireless devices, commercial radio and TV, and so on.
3. Conducted interference on powerlines or connected cables caused by other equipm ent, such as switching power supplies, electrical controls, and natural phenomena such as lightning.
Vertical lines in the spectral display, popping type nois e in the audio, or both.
Page 58
Chapter 2. Safety 2 – 17
A medical device can either generate or receive electromagnetic interference. The EMC standards describe te sts for both emitted and received interference.
Samsung Medison’s ultrasound products do not generate electromagnetic interference in excess of standard levels es tablished for such devices.
An Ultrasound System is des igned to receive signals at radio frequency and is therefore susceptible to interference generated by RF energy sources. Examples of other sources of interference are medical devices, information technology products, and radio and television transmission towers. Tracing the source of radiated interference can be a difficult task . Customers should c onsider the following in an attempt to locate the source:
• Is the interference intermittent or constant?
• Does the interference show up only with one transducer operating at the same frequency or with several transducers?
• Do two different transducers operating at the same frequency have the same problem?
• Is the interference present if the system is moved to a different location in the facility?
The answers to these questions will help determine if the problem resides with the s ystem or the scanning environment. After you answer the questions, contact the Samsung Medison Customer Support Department.
Page 59
2 – 18 HS40 Service Manual
WARNING
Special caution should be taken during movement. the product may result in product damage or personal injury.
WARNING
Be aware of the c astors, especially when moving the system. Samsung Medison recommends that you exercise caution when moving the product up or down ramps.

2.4. Mechanical Safety

2.4.1. Moving the Equipment
Careless transportation of
• Before transporting the product, check that the brakes on the wheels are unlock ed. Also, make sure to retract the monitor arm completely so that it is secured in a stationary position.
• Always use the handles on the console and move the product slowly.
This product is designed to resist shocks. However, exces sive shock, for example, if the product falls over, may cause serious damage.
If the system operates abnormally after repositioning, please c ontact the Samsung Medison Customer Support Department.
2.4.1.1. Foot Lock
Brakes are mounted to the wheels of the console. To lock the brakes, press the bottom part of the brake with your foot. To unlock them, press the part labeled OFF at the top of the brake with your foot.
You can use the brakes to control the movement of the product. We recommend that you lock the brakes when using the product.
2.4.1.2. Precautions on Ramps
Always mak e sure that the control panel is facing the direction of movement.
Leaving the CART unattended on an inclined surface may cause the CART to topple, even if you engage the foot lock . Do not rest the product on ramps .
Page 60
2.4.2. Precautions for Use
the
Do not press the control panel excess ively.
Never attempt to modify the product in any way.
Check the operational safety when using the product after a prolonged break
CAUTION
in service.
Make sure that other objects, such as metal pieces , do not enter the system.
Do not block the ventilation slots.
Do not pull on the power cord to unplug the product. Doing so might damage
the cord and cause the product to short-circuit, or the cord itself to break. Unplug the cord by pulling on the plug itself.
Excessive bending or twisting of cables on patient-applied parts may caus e
failure or intermittent operation of the sys tem.
Improper cleaning or sterilization of a patient-applied part may cause
permanent damage.
Servicing the product, including repairs and replacement of parts, must be
done by qualified Samsung Medison service personnel. Assuming that the product is used in ac cordance with the guidelines contained in this manual and maintained by qualified service personnel, the expec ted service life of product is approximately 7 years.
Chapter 2. Safety 2 – 19
Please refer to ‘Chapter 10. Maintenance’ for detailed information on protection, cleaning, and disinfecting the equipment.
Page 61
2 – 20 HS40 Service Manual
CAUTION
2.4.2.1. Caution for Using Monitor
When adjusting the height or position of the monitor, be careful of the space in the middle of the monitor arm. Catching your fingers or other body parts in it may result in injury.
[Figure 2.2 Safety Note for Monitor]
When moving the product, be sure to keep the monitor arm locked.
When rotating the monitor, be sure to releas e the lock first. Rotating the
monitor while it is locked may cause damage to the internal cables.
Page 62
Chapter 2. Safety 2 – 21
CAUTION
When you adjust the heigh track mark) and oil film may be left on the lift. They are the result of the lift's movements, and not an indication the product is defective.
2.4.2.2. Caution for Using Control Panel
Do not push the control panel with excessive force or lean on it
Do not sit on the control panel or exert ex cessive force on it.
Do not lift the product using the handles on the control panel.
When adjusting the height or position of the control panel, pay attention to the space between the bottom of the control panel and the lever. Catching your fingers or other body parts in it may result in injury.
[Figure 2.3 Safety Note for Control Panel]
Pull the height adjustment lever to adjust the height of the control panel.
t of the control panel of this product, a rail mark (or
NOTE
Page 63
2 – 22 HS40 Service Manual
• be harmful to the patient. If there is no medical benefit, minimize the exposure

2.5. Biological Safety

For safety instructions concerning probes and biopsies, refer to ‘Chapter 9. Probes’.
Ultrasound waves may have damaging effects on cells and, therefore, may
time and maintain the ultrasound wave output level at low. Please refer to the
WARNING
2.5.1. ALARA Principle
Guidance for the use of diagnostic ultrasound is defined by the “As Low As Reasonably Achievable” (ALARA) principle. The decision as to what is reasonable has been left to the
judgment and insight of qualified personnel. No set of rules can be formulated that would be sufficiently complete to dictate the correct response to every circumstance. By keeping ultrasound exposure as low as possible, while obtaining diagnostic images, users can minimize ultrasonic bioeffects.
ALARA principle.
Do not us e the system if an error mes sage appears on the video display
indicating that a hazardous condition exists. Note the error c ode, turn off the power to the system, and call Samsung Medison customer service department.
Do not us e a system that exhibits erratic or inconsistent functioning.
Discontinuities in the scanning sequenc e are indicative of a hardware failure that should be corrected before use.
The system limits the maximum contact temperature to 43 degrees Celsius,
and the ultrasonic wave output observes American FDA regulations.
Since the threshold for diagnostic ultrasound bioeffects is undetermined, it is the sonographer’s responsibility to control the total energy transmitted into the patient. The sonographer must reconcile exposure time with diagnos tic image quality. To ensure diagnostic image quality and limit exposure time, the Ultrasound System provides controls that can be manipulated during the exam to optimize the results.
The ability of the user to abide by the ALARA principle is important. Advances in diagnostic ultrasound, not only in the technology but also in its applications, have resulted in the need for increased and improved information to guide the user. The output indices are designed to provide that important information.
There are a number of variables, which affect the way in which the output display indices can be used to implement the ALARA principle. These variables include mass , body size, location of the bone relative to the focal point, attenuation in the body, and ultrasound exposure time. Exposure time is an especially useful variable, because the user controls it. The ability to limit the index values over time s upports the ALARA principle.
2.5.1.1. Applying ALARA
The sys tem imaging mode us ed depends upon the information needed. 2D mode and M mode imaging provide anatomical information, while Doppler, Power, and Color imaging provide information about blood flow. Scanned modes, like 2D mode, Power, or Color, disperse or scatter the ultrasonic energy over an area, while an unscanned mode, like M mode or Doppler, concentrate the ultrasonic energy. Understanding the nature of the imaging mode being used allows the sonographer to apply the ALARA principle with informed judgment. The probe frequency, system set-up values, scanning techniques , and operator experience aid the
Page 64
Chapter 2. Safety 2 – 23
sonographer in meeting the ALARA principle. The decision as to the amount of acoustic output is, in the final analysis, up to the system operator. This decision must be based on the following factors: type of patient, type of exam, patient history, ease or difficulty of obtaining diagnostically useful information, and the potential localized heating of the patient due to probe surface temperatures. Prudent use of the system occurs when patient exposure is limited to the lowest index reading for the shortest amount of time necessary to achieve acceptable diagnostic results.
Although a high index reading does not mean that a biological effect is actually oc curring, it should be tak en seriously. Every effort should be made to reduce the pos sible effects of a high index reading. Limiting exposure time is an effective way to accomplish this goal.
There are several system controls that the operator can use to adjust the image quality and limit the acoustic intensity. These controls are related to the techniques that an operator might use to implement ALARA and can be divided into three categories: direct, indirect, and receiver control.
2.5.1.2. Direct Controls
Application selection and the output intens ity control directly affect acoustic intensity. There are different ranges of allowable intensity or output depending on your selection. Selecting the correct range of ac oustic intensity for the applic ation is one of the first things required during any exam. For example, peripheral vascular intensity levels are not recommended for fetal exams . Some systems automatically select the proper range for a particular procedure, while others require manual selection. Ultimately, the user bears the responsibility for proper clinical us e. Samsung Medison systems provide both automatic and user-definable settings.
Output has direct impact on acoustic intensity. Once the applic ation has been es tablished, the output control can be used to increase or decrease the output intensity. The output c ontrol allows you to select intensity levels less than the defined maximum. Prudent use dic tates that you select the lowest output intensity consistent with good image quality.
2.5.1.3. Indirect Controls
The indirect controls are those that have an indirect effect on the acoustic intens ity. These controls affect the imaging mode, pulse repetition frequency, focus depth, puls e length, and probe selection.
The choice of imaging mode determines the nature of the ultrasound beam. 2D mode is a scanning mode, Doppler is a s tationary or unscanned mode. A stationary ultrasound beam concentrates energy on a single loc ation. A moving or scanned ultrasound beam dis perses the energy over a wide area and the beam is only c oncentrated on a given area for a fraction of the time necessary in unscanned mode.
The pulse repetition frequency or rate refers to the number of ultrasound bursts of energy over a specific period of time. The higher the pulse repetition frequency, the more pulses of energy in a given period of time. Several controls affect pulse repetition frequency: focal depth, display depth, sample volume depth, color sensitivity, number of focal zones, and sector width controls.
The focus of the ultrasound beam affects the image resolution. To maintain or increase resolution at a different focus requires a variation of output over the focal zone. This variation of output is a function of s ystem optimization. Different exams require different focal depths. Setting the focus to the proper depth improves the resolution of the structure of interest.
Pulse length is the time during which the ultrasonic burst is turned on. The longer the pulse, the greater the time-average intensity value. The greater the time-average intensity, the greater the likelihood of temperature increase and cavitations. Pulse length, burst length or puls e duration is
Page 65
2 – 24 HS40 Service Manual
the output pulse duration in Pulsed Doppler. Increasing the Doppler sample volume, increases the pulse length.
Probe selection affects intensity indirectly. Tissue attenuation changes with frequency. The higher the probe operating frequency, the greater the attenuation of the ultrasonic energy. Higher probe operating frequencies require higher output intensity to scan at an increased depth. To scan deeper at the same output intensity, a lower probe frequency is required. Using more gain and output beyond a point, without corresponding increases in image quality, can mean that a lower frequency probe is needed.
Receiver Controls
Receiver controls are used by the operator to improve image quality. These c ontrols have no effect on output. Receiver controls only affect how the ultrasound echo is received. These controls include gain, TGC, dynamic range, and image processing. The important thing to remember, relative to output, is that receiver controls should be optimized before increasing output. For example; before increasing output, optimize gain to improve image quality.
2.5.1.4. Additional Considerations
Ensure that scanning time is kept to a minimum, and ensure that only medically required scanning is performed. Never compromise quality by rushing through an ex am. A poor exam will require a follow-up, which ultimately increases the scanning time. Diagnostic ultrasound is an important tool in medicine, and, like any tool, should be used efficiently and effectively.
2.5.1.5. Output Display Features
The sys tem output display com prises two basic indices: a mecha nical index and a th erm al index. The thermal index consists of the following indices: soft tissue (TIs), bone (TIb) and cranial bone (TIc). One of these three thermal indices will be displayed at all times. The thermal index to be displayed is determined by the system preset or user choice, depending upon the application.
The mechanic al index is continuously displayed over the range of 0.0 to 1.9, in increments of
0.1. The thermal index consists of the three indices, and only one of these is displayed all the time. Each diagnostic probe application has a default selection. The TIb or TIs is continuously displayed over its range of 0.0 to maximum output, based on the probe and application, in increments of 0.1.
The default s etting of the application-spec ific nature is also an important factor of index selection. A default setting is a system control state which is preset by the manufacturer or the operator. The sys tem has default index settings for the probe applic ation. The default s ettings are applied automatically by the ultrasound system when the power is turned on, new patient data is entered into the system database, or a c hange of application takes place.
The decision as to which of the three thermal indices to display should be based on the following criteria:
Appropriate index for the application: TIs is used for imaging s oft tissue, and TIb for a focus at or near a bone. Some factors might create artificially high or low thermal index readings (e.g. presence of fluid or bone, or the flow of blood). A highly attenuating tissue path, for example, may cause the potential for local zone heating to be lower than the thermal index indicates.
The selection of scanne d modes or unscanned mod es o f operation als o affect the thermal index. For scanned modes, heating tends to be near the surface; for unscanned modes, the potential for heating tends to be deeper in the focal zone.
Page 66
Chapter 2. Safety 2 – 25
Always limit ultrasound exposure time. Do not rush the scanning. Ensure that the indices are kept to a minimu m, and th at expos ure time is limited without compromising dia gnostic sensitivity.
Mechanical Index (MI) Display
Mechanical bioeffects are threshold phenomena that occur when a certain level of output is exceeded. The threshold level varies, however, with the type of tissue. The potential for mechanical biologic al effects varies with peak pressure and ultrasound frequency. The MI accounts for these two factors. The higher the MI value, the greater the likelihood of mechanical bioeffects occurring. However, there is no specific MI value that means that a mechanical bioeffect will actually occur. The MI should be used as a guide for implementing the ALARA principle.
The rmal Index (TI) Display
The TI informs the user of potential temperature increase occurring on the body surface, within body tissue, or at the point of focus of the ultrasound beam on bone. The TI is an estimate of the temperature inc rease in specific body tissues . The actual amount of any temperature rise is influenced by factors such as tissue type, vascularity, and mode of operation. The TI should be used as a guide for implementing the ALARA principle.
The bone thermal index (TIb) informs the user about potential heating at or near the focus after the ultrasound beam has passed through soft tissue or fluid, such as the skeletal structure of a 2-3 month old fetus. The cranial bone thermal index (TIc) informs the user about the potential heating of bone at or near the surface, for example, the cranial bone. The soft tissue thermal index (TIs) informs the user about the potential for heating within soft homogeneous tissue. TIc is displayed when you selec t a trans-cranial application.
You can select the TI to display at Setup > Imaging > Display.
Mechanical and Thermal Indices Display Precision and Accuracy
The Mechanic al and Thermal Indices on the system are precise to 0.1 units.
The MI and TI display accuracy estimates for the sys tem are given in the manual's Acoustic Output Tables. These accuracy estimates are based on the variability ranges of probes and systems, inherent acoustic output modeling errors, and the measurement variability, as described below.
The display ed values should be interpreted as relative information to help the system operator achieve the ALARA principle through prudent use of the s ystem. The values should not be interpreted as actual physical values of investigated tiss ue or organs. The initial data that is used to support the output display is derived from laboratory measurements bas ed on the AIUM measurement standard. The measurements are then put into algorithms to calculate the displayed output values.
Many of the assumptions used in the process of measurement and calculation are conservative in nature. Over-estimation of actual in situ exposure, for the vast majority of tiss ue paths, is built into the measurement and calculation process. For example, the acoustic output values measured underwater are derated us ing a conservative, industry standard, attenuation coefficient of 0.3dB/cm-MHz.
Conservative values for tissue characteristics were selected for use in the TI models. Conservative values for tissue or bone absorption rates , blood perfusion rates, blood heat capacity, and tis sue thermal conductivity were s elected.
Page 67
2 – 26 HS40 Service Manual
A steady state temperature rise is assumed in the industry standard TI models, and the assumption is made that the ultrasound probe is held steady in one position long enough for a steady state to be reached.
A number of factors are considered when estimating the ac curacy of display values: Hardware deviation, algorithm accuracy, and measurement deviation. Deviation among probes and systems in particular is an important factor. Probe deviation results from piezoelectric crystal efficiencies, process-related impedance differences, and sensitive lens focusing parameter variations. Differences in the system pulse voltage control and efficiencies are also a contributor to variability. There are inherent unc ertainties in the algorithms used for estimating acoustic output values over the range of poss ible system operating conditions and pulse voltages. Inaccuracies in laboratory meas urements are related to differences in hydrophone calibration and performance, positioning, alignment and digitization tolerances, and variability among test operators.
The conservative assumptions of the output es timation algorithms of linear propagation, at all depths, through a 0.3dB/cm-MHz attenuated medium are not taken into account in the calculation of the accuracy estimate displayed. Neither linear propagation nor uniform attenuation at the 0.3dB/cm-MHz rate occurs in underwater measurements, or in most tissue paths in the body. In the body, different tissues and organs have dis similar attenuation characteristics. In water, there is almos t no attenuation. In the body, and particularly in underwater measurements, non-linear propagation and saturation losses oc cur as pulse voltages increase.
The display accuracy es timates take into account the variability ranges of probes and systems, inherent acoustic output modeling errors, and the measurement variability. Display accuracy estimates are measured according to AIUM measurement standards but not based on errors caused during the measurement or inherent errors. They are also independent of the effects of non-linear loss on the measured values.
2.5.1.6. Control Effects – Controls Affecting the Indices
As various system controls are adjusted, the TI and MI values may change. This will be most apparent as the Power control is adjusted; however, other system controls will also affect the on-screen output values.
Power
Power controls the system acoustic output. Two real-time output values are on the screen: a TI and a MI. They change as the system responds to Power adjustments.
In combined modes, such as simultaneous Color, 2D mode, and Pulsed Doppler, the total TI is the sum of TIs of the individual modes. Each mode is a vital c ontributor to this total; the displayed MI will be from the mode with the largest peak pressure.
2.5.1.7. 2D Mode Controls
2D Mode Size
Narrowing the sector angle may increase the frame rate. This will increase the TI. Pulse voltage may be automatically adjusted down with software controls to k eep the TI below the system maximum. A decrease in pulse voltage will decrease MI.
Page 68
Chapter 2. Safety 2 – 27
Zoom
Increasing the zoom magnification may increase frame rate. This will increase the TI. The number of focal zones may also increase automatically to improve the resolution. This ac tion may change the MI, sinc e the peak intensity can occur at a different depth.
Number of Focal Zones
Increasing the number of focal zones may change both the TI and MI by automatically changing the frame rate or focal depth. Lower frame rates decrease the TI. The MI displayed will correspond to the focal zone with the largest peak intens ity.
Focus
In general, higher MI values will occur when the focal depth is near the natural focus of the probe (transducer).
2.5.1.8. Color and Power Controls
Color Sensitivity
Increasing the color sensitivity may increase the TI. More time is required for the scanning of color images. Color pulses are the dominant pulse type in this mode.
Color Sector Width
Narrower color sector width will increase the color frame rate, and so the TI will increase. The system may automatically decrease the pulse voltage to stay below the system maximum. A decrease in pulse voltage will decrease the MI. If Pulsed Doppler is also enabled, th en Pulse d Do ppl er will remai n the d omin ant mode and the TI cha ng e will be small.
Color Sector Depth
Deeper color sector depth may automatically decrease color frame rate, or select a new color focal zone or color pulse length. The TI will change due to the combination of these effec ts. Generally, the TI will decrease with increased color sector depth. MI will correspond to the peak intensity of the dominant pulse type, which is a color pulse. However, if Pulsed Doppler is als o enabled, then Pulsed Doppler will remain the dominant mode and the TI change will be small.
Sca le
Using the Scale control to increase the color veloc ity range may increase the TI. The system will automatically adjust the pulse voltage to stay below the system maximum. A decrease in pulse voltage will also decrease MI.
2D Mode Size
A narrower 2D mode sector width in Color imaging will increase color frame rate. The TI will increase. MI will not change. If Pulsed Doppler is also enabled, then Pulsed Doppler will remain the dominant mode and the TI change will be small.
Page 69
2 – 28 HS40 Service Manual
2.5.1.9. M Mode and Doppler Controls
Simultaneous and Update Methods
Use o f combi nation mod es affects both th e TI an d MI th ro ugh t he combi nation of pulse types. During Simultaneous mode, the TI is an ancillary element. During auto-update and duplex, the TI will display the dominant puls e type. The display ed MI will be from the mode with the largest peak pressure.
Sample Volume Depth
When Doppler sample volume depth is inc reased, the Doppler PRF may automatically decrease. A decrease in PRF will decrease the TI. The sys tem may also decrease the pulse voltage to remain below the system maximum. A decrease in pulse voltage will decrease MI.
2.5.1.10. Other
2D, Color, M Mode, PW, and CW Modes
When a new imaging mode is selected, both the TI and the MI will change to their default settings. Each mode has a corresponding pulse repetition frequency and maximum intensity point. In combined or simultaneous modes, the TI is the sum of the contribution from the modes enabled, and the MI is the value for the focal zone of the mode with the largest derated intensity. If a mode is turned off and then reselected, the system will return to the previously selected settings.
Probe
Each probe model available has unique s pecific ations for the contact area, beam shape, and center frequency. Defaults are initialized when you select a probe. Samsung Medison’s factory defaults vary according to the probe, applic ation, and mode. Defaults have been chosen below the FDA limits for intended use.
De pth
An increase in the 2D mode depth will automatically decrease the 2D mode frame rate. This would decrease the TI. The system may also automatically choose a deeper 2D mode focal depth. A change of focal depth may change the MI. The MI display ed is that of the zone with the largest peak intensity.
Application
Acoustic output defaults are set when you select an application. Factory defaults vary with probe, application, and mode. Defaults have been chosen below the FDA limits for intended use.
Page 70
Chapter 2. Safety 2 – 29
2.5.1.11. Related Guidance Documents
For more information about ultrasonic bioeffects and related topics, refer to the following;
• Medical Ultrasound Safety (AIUM, 2014).
• AIUM Consensus Report on Potential Bioeffects of Diagnostic Ultrasound: Executive Summary, J. Ultrasound in Medicine, 2008, Vol. 27, Num. 4.
• W FUMB. Symposium on Safety of Ultrasound in Medicine: Conclusions and Recommendations on Thermal and Non-thermal Mechanisms for Biological Effects. Ultrasound in Med. & Biol; 1998, 24: Supplement 1.
• Bioeffects and Safety of Diagnostic Ultrasound (AIUM, 1993)
• Guidelines for the safe use of diagnostic ultrasound equipment. (BMUS, 2009)
• Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (U.S. FDA - 2008)
• Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment. (IEC, 2007)
• Acoustic Output Labeling Standard for Diagnostic Ultrasound Equipment (AIUM, 2008)
• Standard Means for the Reporting of the Acoustic Output of Medical Diagnostic Ultrasonic Equipment. (IEC, 2007)
• Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices On Diagnostic Ultrasound Equipment (AIUM/NEMA, 2004)
• Ultrasonics - Field characterization -Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields (IEC, 2005)
• Measurement and Characterization of Medical Ultrasonic Fields up to 40 MHz. (IEC, 2007)
• Ultrasonics-Power Measurements- Radiation Force Balances and Performance Requirements. (IEC, 2006)
• Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment. (AIUM/NEMA,
2004).
Page 71
2 – 30 HS40 Service Manual
)23.0( alf
e
)069.0( lf
e
2.5.1.12. Acoustic Output and Measurement
Since the first usage of diagnostic ultrasound, the poss ible human biological effects (bioeffects) of ultrasound exposure have been studied by various scientific and medical institutions. In October 1987, the American Institute of Ultrasound in Medicine (AIUM) ratified a report prepared by its Bioeffects Committee (Bioeffect Considerations for the Safety of Diagnostic Ultrasound, J Ultrasound Med., Sept. 1988: Vol.7, No.9 Supplement), sometimes referred to as the Stowe Report, which reviewed available data on possible effects of ultrasound exposure. Another report, “Bioeffects and Safety of Diagnostic Ultrasound”, dated January 28, 1993, provides more up to date information. In addition, periodically updated reports on biological effects , results, and guidelines on safe usage have been published by groups s uch as the WFUMB (World Federation of Ultrasound in Medicine and Biology), AIUM, and BMUS.
The Acoustic output for this system has been measured and calc ulated in accordance with the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices On Diagnostic Ultrasound Equipment (AIUM/NEMA, 2004) and Acoustic Output Meas urement Standard for Diagnostic Ultrasound Equipment (AIUM/NEMA, 2004).
2.5.1.13. In Situ, Derated, and Water Value Intensities
All intensity parameters are measured in water. Since water does not absorb acoustic energy, these water measurements represent the largest possible value. Biological tissue absorbs acoustic energy. The true value of the intensity at any point depends on t he amount and type of tiss ue, and the frequency of the ultrasound that passes through the tissue. The intensity value in the tissue, In Situ, has been estimated using the following formula:
In Situ = Water
Where: In Situ = In Situ Intens ity Value
Water = Water Value Intensity
e = 2.7183
a = Attenuation Factor
Tissue a (dB/cm-MHz)
Brain .53
Heart .66
Kidney .79
Liver .43
Muscle .55
l = Skin line to measurement depth (cm)
f = Center frequency of the transducer/system/mode combination (MHz)
Since the ultrasonic path during an examination is likely to pass through varying lengths and types of tis sue, it is diffic ult to estimate the true In Situ intensity. An attenuation factor of 0.3 is used for general reporting purposes; therefore, the In Situ value is commonly reported using this formula:
In Situ (derated) = W ater
Since this value is not the true in situ intensity, the term “derated” is used.
Page 72
Chapter 2. Safety 2 – 31
peak rarefactional acoustic pressure calculated at a specific distance
The maximum derated and max imum water values do not always occur under the same operating c onditions. Therefore, the reported max imum water and derated values may not be related to the In Situ (derated) formula. For example, a multi-zone array transducer has the greatest water value intensities at its deepest zone. The same transducer may have its largest derated intensity at one of its shallowest focal zones.
2.5.1.14. Terms and Symbols Related to Acoustic Output and Measurement
The terms and symbols used in the acoustic output tables are defined in the following paragraphs.
-12dB OUTPUT BEAM AREA, ultrasonic beam area induced by -12dB
A
aprt
output beam size (unit: c m
at max. Ipi EQUIVALENT BEAM DIAMETER, the acoustic beam's diameter at the
d
eq
location where the PULSE-INTENSITY INTEGRAL is maximal, expressed as an equivalent beam area (unit: cm)
EQUIVALENT BEAM DIAMETER, the acoustic beam's diameter at Zb
d
eq(zb)
location, expressed as an equivalent beam area (unit: cm
2
)
2
)
Dim of A
-12dB OUTPUT BEAM DIMENSIONS, the dimensions of an ultrasound
aprt
beam (whose pulse beam width is -12dB) from a specific direction that is perpendicular to the transducer output plane and the beam alignment axis (unit: cm)
ACOUSTIC WORKING FREQUENCY, the arithmetic average of f1 and
f
awf
f2 that are farthest from each other among the frequencies of the pressure spectrum of the acoustic signal whose amplitudes are lower than the peak amplitude, i.e. maximum amplitude, by 3dB (unit: MHz)
Focal Length The focal length in a direction parallel to the beam alignment axis in the
defined operational state of the ultrasound s ystem (unit: cm)
at max. MI
I
pa,α
(z)
I
ta, α
The average attenuated pulse s trength at the location where Mechanical Index (MI) is maximal (unit: W/cm
Attenuated average strength over time at a specific focal length (z) (unit: mW/cm
2
)
2
)
MI MECHANICAL INDEX, a variable representing potential cavitations within
the human body (unit: N/A)
P OUTPUT POWER, the average power over time of an ultrasonic
transducer's emissions into a free field through a specified medium suc h as water (unit: mW)
(z) ATTENUATED OUTPUT POWER, the power of ultrasonic output
P
α
calculated at a s pecific distance from the transducer after attenuation occurs (unit: mW)
ATTENUATED PEAK-RAREFACTIONAL ACOUSTIC PRESSURE, the
P
r,α
Page 73
2 – 32 HS40 Service Manual
after attenuation occurs (unit: MPa)
Pr at max. Ipi The peak rarefactional acoustic press ure at the location where PULSE-
INTENSITY INTEGRAL is maximal (unit: MPa)
prr PULSE REPETITION RATE, the inverse number of the time interval
between two contiguous acoustic pulses (unit: Hz)
TIB BONE THERMAL INDEX, a thermal index for a focal zone formed near a
bone after the ultrasound beam passes through soft tissue, e.g. applied to a fetus (in the 2nd or 3rd trimester) or to the head of a neonate (through the fontanel) (unit: N/A)
TIC CRANIAL BONE THERMAL INDEX, a thermal index for an ultrasound
beam entering the body and passing through a bone, e.g. the skull of children or adults (unit: N/A)
Soft tissue thermal index in scanning mode (unit: N/A)
TIS
scan
TIS
PULSE DURA TION (unit: us)
t
d
z_at_max _Ipi,
DEPTH FOR BONE THERMAL INDEX (unit: cm)
z
b
BREAK-POINT DEPTH, which is the EQUIVALENT APERTURE
z
bp
Soft tis sue thermal index in non-scanning mode (unit: N/A)
non-scan
Location where PULSE-INTENSITY INTEGRAL is maximal (unit: cm)
α
DIAMETER multiplied by 1.5 (unit: cm)
DEPTH FOR SOFT-TISSUE THERMAL INDEX, the distanc e from a
z
s
plane where the product of minimum attenuated output power, ATTENUATED SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY, and 1 cm
2
is maximal at the distance range that is equal to, or greater than, the equivalent aperture diameter multiplied by 1.5, when the beam dimension of -12dB output is defined along the beam alignment axis (unit: cm)
Page 74
Chapter 2. Safety 2 – 33
2.5.1.15. Acoustic Measurement Precision and Uncertainty
The Acoustic Measurement Precision and Acoustic Measurement Uncertainty are described below.
Qua ntity Precision Total Uncerta i nty
Ipi,α (attenuated pulse intensity integral)
P (acoustic power) 6.2% ±19%
P
(attenuated rarefaction pressure) 5.4% ±15%
r,α
f
(acoustic working frequency) < 1% ±4.5%
awf
3.2% +21% to -24%
Systematic Uncertainties
For the pulse intensity integral, derated rarefaction pressure Pr.3, center frequency, and pulse duration, the analysis includes considerations of the effects on accuracy of:
Hydrophone calibration drift or errors.
Hydrophone/Amp frequency response.
Spatial averaging.
Alignment errors.
Voltage measurement accuracy, including.
– Osc illos cope vertical accuracy.
– Osc illos cope offset accuracy.
– Oscilloscope clock accuracy.
– Osc illos cope Digitization rates.
– Noise.
The acoustic power is measured using a Radiation Force for systematic uncertainties through the use of calibrated NIST acoustic power sources.
We also refer to a September 1993 analysis conducted by the working group of the IEC technical committee 87 and prepared by K. Beissner, as a first supplement to IEC publication
1161.
The document includes analys is and discussion of the sources of error/measurement effects due to:
– Balance system calibration.
– Absorbing (or reflec ting) target suspension mechanisms.
– Linearity of the balance system.
– Extrapolation to the moment of switching the ultrasonic transducer (compensation for
ringing and thermal drift).
– Target imperfections.
Page 75
2 – 34 HS40 Service Manual
– Absorbing (reflecting) target geometry and finite target size.
– Target misalignment.
– Ultrasonic transducer mis alignment.
– Water temperature.
– Ultrasonic attenuation and acoustic streaming.
– Coupling or shielding foil properties.
– Plane-wave assumption.
– Environmental influences.
– Excitation voltage measurement.
– Ultrasonic transducer temperature.
– Effects due to nonlinear propagation and saturation loss.
2.5.1.16. Training
The users of this ultrasound system must familiarize themselves with the ultrasound system to optimiz e the performance of the device and to detect possible malfunctions. It is recommended that all users receive proper training before us ing the device. You can receive training on the use of the product from the Samsung Medison service department, or any of the customer support centers worldwide.
Page 76
Chapter 2. Safety 2 – 35
WARNING

2.6. Protecting the Environment

For dis posal of the system or accessories that have come to the end of their
service lives, contact the vendor or follow appropriate disposal procedures.
You are respons ible for complying with the relevant regulations for disposing
CAUTION
2.6.1. EU WEEE Directive
Thi s marking on the product, acces sories or literature indicates that the product and its electronic accessories (e.g. charger, headset, USB cable) should not be disposed of with other household waste at the end of their working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate these items from other types of waste and recycle them responsibly to promote the sustainable reuse of material resources.
of wastes .
The lithium ion battery used in the product must be replaced by a Samsung
Medison service engineer or an authorized dealer.

Correct Disposal of This Product (Waste Electrical & Electronic Equipment)

Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can take these items for environmentally safe recycling.
Business users s hould contact their supplier and check the terms and conditions of the purchase contract. This product and its electronic acc essories should not be mixed with other commercial wastes for disposal.
2.6.2. EU REACH Regulation
For information on Samsung’s environmental commitments and product specific regulatory obligations e.g. REACH visit:
Samsung.com/uk/aboutsamsung/samsungelec tronics/corp or atecitiz enship/d ata_corner.html
2.6.3. State of California Proposition 65 Warning (US Only)
This product contains chemicals known to the State of California to cause cancer, birth defects or other reproductive harm.
Page 77
Chapter 3
3. Insta lli ng Pr oduct Chapte r 3 . Insta lling Pr oduct

Installing Product

3.1. Transporting ......................................................................................................... 3
3.1.1. Precautions During Trans po rt ..........................................................................3
3.1.2. Brakes ...........................................................................................................3
3.1.3. Precautions on Ramps ...................................................................................3
3.1.4. Humi dity and Temperatu re ..............................................................................4
3.2. Unpacking the Product...................................................................................... 5
3.2.1. Dismantlin g the Product's Box .........................................................................5
3.2.2. Accessories ...................................................................................................6
3.2.3. Release of the Locking Mecha nism .................................................................6
3.3. Installation Environment................................................................................... 7
3.3.1. Cauti on..........................................................................................................7
3.4. Installing the Product ........................................................................................ 8
3.4.1. Installatio n Sa fety ...........................................................................................8
3.4.2. Connec ting P erip he rals ................................................................................ 10
3.5. System Power .................................................................................................... 13
3.5.1. Tu rnin g the Power On................................................................................... 13
3.5.2. Shutting Down the System ............................................................................ 13
3.6. System Settings ................................................................................................ 14
3.6.1. General Sy stem Settings .............................................................................. 15
3.6.2. General ....................................................................................................... 15
3.6.3. Display ........................................................................................................ 19
3.6.4. Patient ......................................................................................................... 20
3.6.5. Quick Preset ................................................................................................ 23
Page 78
3 – 2 HS40 Service Manual
3.6.6. EzExam+..................................................................................................... 24
3.6.7. Imaging ....................................................................................................... 27
3.6.8. Application/Preset ........................................................................................ 30
3.6.9. Application................................................................................................... 32
3.6.10. Measurement Settings.................................................................................. 33
3.6.11. Report ......................................................................................................... 49
3.6.12. Annotatio n ................................................................................................... 52
3.6.13. Body Marker ................................................................................................ 57
3.6.14. Customize ................................................................................................... 60
3.6.15. Peripherals .................................................................................................. 64
3.6.16. Connecti vity ................................................................................................. 67
3.6.17. S e r vic e ........................................................................................................ 86
3.6.18. Help ............................................................................................................ 86
Page 79
Chapter 3. Installing Product 3 – 3
WARNING
The product weighs more than 100 kg. Be extra careful when transporting it. Careless transportation of the product may result in product damage or personal injury.
WARNING
Be especially aware of the castors when moving the system. Samsung Medison rec down ramps in advance of any significant reloc ation.

3.1. Transporting

This product is a finely tuned piec e of medical electronic equipment; careful attention is required when transporting it.
3.1.1. Precautions During Transport
This product and its packaging are designed to protect the product from physical impacts. However, drops and ex ternal impacts may c ause serious damage to the product.
Before transporting the product, check that the wheel brakes are unlocked. Also, be sure to retract the monitor arm completely s o that it is secured in a stationary position.
Always use the handles at the back of the console and move the product slowly.
3.1.2. Brakes
You can use the brakes to control the movement of the product. The brake pedal is located at the center of the console pedal and when pressed brakes all the wheels simultaneously. To lock the brakes, press the front part of the brake pedal with your foot. To unloc k the brakes, press the back of the pedal.
We recommend that you lock the brakes when using the product.
3.1.3. Precautions on Ramps
Always mak e sure that the control panel is facing the direction of movement. If you leave the product on an incline, the product may fall over even if the brakes are engaged. Do not rest the product on a ramp.
ommends that you practic e transporting the product up or
Page 80
3 – 4 HS40 Service Manual
3.1.4. Humidity and Temperature
Table 3-1. The Product's Humidity and Temperature Tolerance shown below illustrates the
appropriate temperature and humidity ranges for transporting, s toring, and operating the product.
Category Temperature [°C] Humi dity [%]
Transporting -25 – 60 20 – 90
Storage -10 – 50 20 – 90
Operating 10 – 35 30 – 75
[Table 3-1. The Product’s Humidity and Temperature Tolerance]
Page 81
Chapter 3. Installing Product 3 – 5

3.2. Unpacking the Product

3.2.1. Dismantling the Product's Box
1. Open the box.
2. Remove the protective cover.
3. Take out the probe box and accessory box and store them in a safe place.
4. Unlock the brakes on the wheels.
5. Grab the rear handle on the product and move it to the place where you wish to install it.
[Figure 3.1 Dismantling the Product's Box]
Page 82
3 – 6 HS40 Service Manual
3.2.2. Accessories
An accessory box containing the items below is supplied with the product. If it does not match your order, please contact the Samsung Medison cus tomer servic e department or your local vendor.
3.2.3. Release of the Locking Mechanism
When this product is being transported, the monitor arm must be secured in place by the locking mechanism to ensure safety. Release the locking mechanis m by following this procedure before using the product:
1. Pull down the loc king mechanism on the center of the monitor arm to release the lock.
[Figure 3.2 Release of the Locking Mec hanism]
Page 83
Chapter 3. Installing Product 3 – 7
CAUTION
Placing the system near generators, X result in screen noise and abnormal visual images.
Sharing the power source with other electrical devices may also produce noise.

3.3. Installation Environment

3.3.1. Caution
When installing the product, please pay attention to the following: For more information on use and setup, please refer to the accompanying manual for this product.
-ray machines, or broadcast cables may
Optimal conditions for the sys tem are a temperature of 10-35° and a humidity of 30-75%.
Avoid ex cess humidity.
Avoid direct sunlight.
Avoid excessive fluctuations in temperature.
Avoid installing the product near a heating appliance.
Avoid dus ty and/or poorly ventilated locations.
Avoid locations that are subject to vibration.
Avoid locations where chemical substances or harmful gases are present.
Page 84
3 – 8 HS40 Service Manual
NOTE
The product may be shipped with the power cable already connected to the console.
CAUTION

3.4. Installing the Product

3.4.1. Installation Safety
The power soc ket and equipotential terminal are located at the back of the product.
[Figure 3.3 Power Socket and Equipotential Terminal]
① Power switch: Supplies power to the entire system or cuts off power Turn off the power.
② Power inlet: Accepts the power cord, which connects to an external power supply.
③ Power fuse: Cuts off the power s upply if overcurrent or overvoltage occurs.
④ Equipotential potential: This should be connected to the equipotential connector in the exam
room.
3.4.1.1. Power Cord Connection
Prior to connecting a power cord, verify that the voltage indicated on the power cord matches the voltage rating of the installation point.
If the product needs to be transported or stored for an extended period, the
temperature and humidity of the environment must be c hecked.
A sudden change in temperature may cause condensation and lead to
product failure.
Page 85
Chapter 3. Installing Product 3 – 9
Refer to “Table 3-2. Operational Temperature of Product” before turning the product on.
Temperature [°C] -20 -15 -10 -5 0 5 10 – 35 45 50 55 60
Waiting time
[Hour]
16 10 8 6 4 2
[Table 3.2 Operational Temperature of the Product]
Use
immediately
2 4 6 10
3.4.1.2. Connecting the Equipotential Terminal
In a hospital environment, dangerous electric currents may occur as a result of the potential difference between a contactable conductive part and c onnected equipment in treatment rooms. The solution to the problem is cons istent equipotential bonding. The equipotential terminal of an item of medical equipment must be connected to the equipotential bonding network in its treatment room, as shown in the figure below.
3.4.1.3. Probe Connection
Be sure to turn off the power before connecting or disconnecting a probe to ensure the safe use of the system and the probes.
1. Connect the probes to the probe connectors on the front panel of the sys tem. Up to four (five
including CW ) probes may be connected. The CW probe should only be connected to its own dedicated connector.
2. To install it, turn the connec tor turning handle clockwise.
[Figure 3.4 Probe Connector]
Page 86
3 – 10 HS40 Servic e Manual
NOTE
For more information on the recommended peripheral devices,
3. Installing the Product".
3.4.2. Connecting Peripherals
Do not install peripheral devices within the patient environment that are not
listed in this Service Manual. If you install an unofficial device in the patient environment, it may cause an electrical hazard.
CAUTION
Do not connect additional external peripheral devic es to the auxiliary socket
outlet. Doing so may increase safety risks.
[Figure 3.5 Patient Environment]
refer to "Chapter
Page 87
Chapter 3. Installing Product 3 – 11
CAUTION
When using a peripheral device via a before connecting/disconnecting the device. Connecting or disconnecting a USB device while the power is turned on may cause the system and/or the USB device to malfunction.
3.4.2.1. External Peripheral Devices
These are peripheral devices that can be connec ted for use when needed and are connected via the USB port located on the rear panel.
USB port, always turn the power off
To remove the removable disk, go to Utility > Storage manager.
USB ports are located both on the console and its rear panel.
We recommend that you connect USB storage devices (flash memory
NOTE
USB Vide o Printer
The following products are recommended:
media, etc.) to the ports on the console, and other USB peripheral devices to the rear panel, for convenience.
BW: Mits ubishi P-95DE, Sony UP-D897, Sony UP-X898MD, Sony UP-D898MD
[Figure 3.6 B/W Printer Installation Screen]
Color: Mitsubishi CP-30DW, Sony UP-D25MD
Page 88
3 – 12 HS40 Servic e Manual
[Figure 3.7 Color Printer Installation Screen]
You must install a printer and driver compatible with the English version of
Microsoft W indows 7. Contact the Samsung Medison cus tomer service department if you have questions about printer driver installation.
CAUTION
W hen connecting the printer ensure that it is configured under Microsoft
Windows or system setup and has been chosen as the default printer.
Please check which port the printer uses before connecting. Printers should
be connected to the printer port while USB printers should be connected to the USB port.
Other
Flash Memory Media
NOTE
The system cannot recognize USB 1.1 flash memory. Remove the flash
memory from the console and equip again with an appropriate device.
W hen using a flash memory device which supports functions other than
saving files, please check first to see if it is possible to save the file on a desktop PC.
Do not use flash memory media which contains anti-virus programs or is
defective. Otherwis e, the product may fail to work properly.
Page 89
Chapter 3. Installing Product 3 – 13
CAUTION
Make sure to connect the probes and peripheral devices that will be used before powering on the sy stem. If you attempt to connect them during system use, it may lead to patient injury or irreparable damage to the console.
CAUTION
Before starting the diagnos is, you must register the pati
• in Intel PC Main Board of the

3.5. System Power

Boot up the sy stem for use.
3.5.1. Turning the Power On
Press the On/Off button when the power is off. Booting begins, and the product logo appears on the screen. When booting is completed, the 2D mode screen will appear in End Exam status.
ent information.
If the power switc h near the power connection port on the rear panel of the
product has been switc hed off, wait for 10 seconds before turning on the product.
NOTE
Do not press keyboard keys or buttons while booting is in progress. Doing so
may cause the system to malfunction.
If y ou turn on the power after turning it off suddenly, the sy stem may turn on
and off momentarily. This is a feature of the built­product, and does not indicate a system error.
3.5.2. Shutting Down the System
Press the On/Off button while us ing the s ystem to initiate shutdown. Click Shut Down on the screen to shut down the system, or click Cancel to cancel.
If the system does not turn off, keep the On/Off button depressed for five
seconds or longer to force a shutdown.
If the ultrasound sys tem is unplugged while working or turned off by forcibly
CAUTION
shutting down, the files may get corrupted and it may result in sys tem failure or data loss .
To ens ure that the product is safely cut off from electric al power, set the
power switch at the rear of the product to the Off position after using the product.
Press the On/Off button twice to turn off the s ystem. The Select Next
Operation window appears when you press the On/Off button once. The system will turn off if you press the button again.
Page 90
3 – 14 HS40 Servic e Manual

3.6. System Settings

General system settings that do not have direct bearing on imaging are explained. The setup can be modified depending on your specific needs or preferences.
1. On the touch screen, tap the button, enter the Utility Menu, and tap the Setup button.
2. The Se tup button will appear on the touch s creen. Select a tab that contains items you want to specify.
TIP
3. Specify the settings for each item.
4. Save settings and exit. Click Ex it on the monitor or tap the Ex i t button on the touch screen to switch to Scan Mode.
Selecting a Tab
You can select the tab you want in either one of two ways. Select the method that suits you.
• Use the trackball and the Set button to select a tab.
• Tap the corresponding button on the touch s creen.
[Figure 3.8 Utility, Setup – Touch Screen]
Page 91
Chapter 3. Installing Product 3 – 15
3.6.1. General System Settings
In the Setup screen, select S y st e m tab. Or tap S yste m on the touch screen.
From this tab, y ou can configure the general system settings .
3.6.2. General
In the Setup screen, select General tab in the Syste m category.
[Figure 3.9 Setup – System – General]
3.6.2.1. Location
You can specify the information that is display ed in the title area on the s creen.
Institute
Enter the name of the hospital/institution where the product is ins talled. Special characters can be entered.
Language
Select the system language. (Supported languages: English, German, French, Spanish, Italian, Russ ian, Simplified Chinese, Portugues e (Brazilian)
TIP
• For Russian, the Russia Restricted Mode option must be disabled.
• The option must be disabled by qualified Samsung Medison s ervice personnel.
Page 92
3 – 16 HS40 Servic e Manual
• installed. If it is not set, the date and time can be automatic ally synchronized
3.6.2.2. Date and Time
Set the date and time for use in the s ystem.
Before starting diagnosis, please confirm the time, date and time zone and
modify the settings if necessary.
Make s ure to set the date and time to the time z one in which the sy stem is
CAUTION
to the time z one of the country of the manufacturer.
You cannot change the date and time when a patient ID has been
registered. To change the date and time, you should finish the current diagnosis by pressing the End Ex a m button on the control panel.
NOTE
W hen the system time and date are changed, the change is not reflected in
the time and date indicated in previous diagnosis results.
Date Format
Specify the date format. Select a date format using the combo button. The date format that you specify will be applied to various date fields in Patient Information.
Time Format
Specify the time format. Press the combo button to select the preferred display format (12 Hour or 24 Hour).
Date and Time
Select settings for date and time. When you press this button, the Date and Time screen appears.
1. Set the date and time by using the trackball and the Set button on the control panel.
2. Press Save to apply the settings . Tap Cancel or press Exit button on the control panel to cancel.
Time Zone
Specify the local time zone where the system is used. W hen y ou press this button, the Time Zone Settings screen appears.
1. Press Time Zone.
2. Specify the time z one by using the trackball and the Set button for the combo button.
3. Press OK to apply the settings . Tap Cancel to cancel.
Page 93
Chapter 3. Installing Product 3 – 17
3.6.2.3. Trackball Speed
Scan Mode
Specify the trackball speed in Scanning Mode as Slow, Normal, or Fast.
Measurement
Used to specify the track ball’s speed during measurement. Select Slow, Normal, or Fas t.
Slower speeds allow for more precise measurements.
3.6.2.4. Video
Format
Choose between NTSC or PAL.
3.6.2.5. Beep Sound
Touch Screen
Set the sound level when using the touch screen. Turn it on or off by using the trackball.
Control Panel
Set the sound level of sound when using buttons and dial-buttons in the control panel. You can turn it on or off by using the trackball.
In particular, for Set/Exit, Freeze, Save, and Othe r buttons and dial-buttons, you can turn them on or off.
3.6.2.6. Account
Register a user ID and password.
User Login
Set the user account (login) function. If the user log-in is set to on, it can be used for the following areas:
– Screen Saver
– Booting up the Sy stem
User Account Manage r
This is the exclusive administrator function for approval and management of accounts. The Account List window will be enabled.
– Create: Fill out the User ID, Password, and Name fields. Then click the OK button to
create a new ID.
– Change Password: Change password.
– Delete: Delete the selected ID.
Page 94
3 – 18 HS40 Servic e Manual
– Close: Close Settings.
[Figure 3.10 Setup – User Account Manager]
Login
You can set a User Ac count ID after logging in to the Admin account. Please contact the service engineer to learn how to reset user acc ount information including the Admin account.
The Admin account c an be configured when you run ‘User Account
Manager’ at Setup for the first time.
The Admin account c annot be deleted.
NOTE
Logout
Once the user account function is activated, you cannot load exams without
logging in.
The pas sword must be 8 to 16 characters and composed of at least three of
the following:
– English alphabet upper case
– English alphabet lower cas e
– Numbers
– Special characters
Close the Admin account.
Page 95
Chapter 3. Installing Product 3 – 19
3.6.3. Display
In the Setup screen, select Display tab in the S yste m category.
Configure the settings for displaying images.
[Figure 3.11 Setup – System – Display]
3.6.3.1. Display
System Logo
Choose one of SAMSUNG or HS40.
Direction Marker
Set the Direction Marker. Among SAMSUNG HS40, HS40, and S Marker, select a desirable shape.
Screen Saver
Select the wait time (in minutes) for the screen saver. Use the track ball to set it off or 5, 10, 30 or 60 minutes .
System Pow er Button
– Ask me what to do: Set a pop-up to appear requesting confirmation before shutting down
the system.
– Shut Down: The power will turn off.
Page 96
3 – 20 HS40 Servic e Manual
Option
– Auto Freeze: Automatic ally switch to freeze state. Set the freeze time between 0 and 60
minutes.
– Prompt for Save on Exit: Use the checkbox to s elect whether to save or not.
– Check On: Set a mess age to recheck whether to save the changes when c losing the
setup.
– Check Off: Set it to automatic ally save changes when closing the setup.
– Boot up Caps Lock On: W hen the checkbox is selected, Caps Lock is turned on after
completing the system booting.
Touch Screen Brightness
– By using the trackball, adjust brightness of the touch screen between 0 and 100 in the
units of 10.
– Adjust it by pressing – and + buttons.
3.6.4. Patient
In the Setup screen, select the Patient tab in the S y st e m category.
[Figure 3.12 Setup – System – Patient]
Page 97
Chapter 3. Installing Product 3 – 21
3.6.4.1. Patient Data
Name Display
– Last, First Middle: Dis play the name of the patient in the order last name, first name, then
middle name.
– First Last Middle: Display the name of the patient in the order first name, last name, then
middle name.
Title Display
Show or hide patient information in the Title area of the screen.
– None: Patient information is not displayed.
– Date of Birth: Displays the patient’s date of birth.
– Age: Displays the patient’s age.
– Gender: Selec t this checkbox to display the patient’s gender. Note that the patient's
gender can only be displayed if their date of birth or age is also displayed.
Title LMP/GA/EDD Display
– LMP/GA: The last menstruation date and estimated gestation weeks are shown on the
screen.
– EDD/GA: The expected delivery date and estimated gestation weeks are shown on the
screen.
Middle Name Input Box
When On is checked, a text input field will be display ed enabling you to enter a middle name when you create a Patient record.
Save Patient Page as First Image
If On is checked, Patient information is saved as the first image.
Other ID
When a Patient record is created, a GUI for text input is displayed for you to enter the Other ID.
– Other ID Format: Any, Numbers, Letters and Numbers, and NHS Numbers are available.
Except Any, inputs in each format type can only be entered as text.
Page 98
3 – 22 HS40 Servic e Manual
3.6.4.2. User Defined List
You may enter information related to studies beforehand, so that you can assign the information easily when entering patient information. Use the Change, Delete All, Delete, Up, and Dow n buttons to add, edit, or delete information; y ou may create a list of up to 20 for each item.
Operator: You can save the name of the operator who scanned the patient.
Diag. Physician: The name of the phys ician who diagnosed the patient c an be saved.
Ref. Physician: You can save the name of the referring physician.
Indication: You can save information about the patient's medical history.
Description: You can save up to 20 diagnostic memos per application.
Page 99
Chapter 3. Installing Product 3 – 23
3.6.5. Quick Preset
In the Setup screen, select Quick Pre se t tab in the S yste m category.
[Figure 3.13 Setup – System – Quick Preset]
3.6.5.1. Quick Pre set
After s electing the Probe, Application, and Preset connected to the port, save them to Quick Preset. Up to 4 sets can be saved. By using Preview, you can preview the saved preset.
In the diagnosis mode, the saved Probe and Preset are shown as a button on the touch screen.
Page 100
3 – 24 HS40 Servic e Manual
3.6.6. EzExam+
In the Setup screen, select the Ez Ex a m+ tab in the S yste m category.
EzExam+ is a macro function that saves your frequently used exams as a protocol and runs them together. This feature streamlines the diagnosis process.
[Figure 3.14 Setup – System – EzExam+]
3.6.6.1. Protocol List
Protocol List: Provides a protocol list for the application. EzExam+ in the Default List cannot be deleted or edited and the addition of new EzExam+ is not allowed.
① Category: A list of available applications is shown. If previously configured EzExams are
available, they will be s hown in the Preset Lists when an application is selected. Selection of ABD, CARD, GYN, MSK, OB, PED, SMP, TCD, URO, or VAS is available.
② New Protocol: W hen New Protocol window appears, enter a name and press OK. Tap
Cancel to cancel.
③ Edit: You can select any protoc ol in the protoc ol list.
④ Rename: You can edit the name of any protocol in the protocol list.
⑤ Delete: Delete a protocol selected from the protocol list.
⑥ Import: Import the protocol data.
⑦ Export: Exports the protocol selected from the protocol list.
⑧ Move Up Button: Moves the protocol selected from the protocol list up one row.
⑨ Move Down Button: Move the protocol selected from the protocol list down one row.
Loading...