The material contained herein consists of information that is the property of LumaSense
Technologies and intended solely for use by the purchaser of the equipment described in this
manual. All specifications are subject to change without notice. Changes are made periodically
to the information in this publication, and these changes will be incorporated in new editions.
LumaSense Technologies prohibits the duplication of any portion of this manual or the use
thereof for any purpose other than the operation or maintenance of the equipment described
in this manual, without the express written permission of LumaSense Technologies.
Index .................................................................................................................................. 45
IGA 6/23 Advanced ManualContents · iv
1 General Information
1.1Information about the user manual
Congratulations on choosing the high quality and highly efficient IMPAC IGA 6/23 Advanced
pyrometer.
This manual provides important information about the instrument and can be used as a work of
reference for installing, operating, and maintaining your IGA 6/23 Advanced pyrometer. It is
important that you carefully read the information contained in this manual and follow all safety
procedures before you install or operate the instrument.
To avoid handling errors, keep this manual in a location where it will be readily accessible.
1.1.1Legend
Note: The note symbol indicates tips and useful information in this manual. All notes
should be read to effectively operate the instrument.
Attention: This sign indicates special information which is necessary for a correct
temperature measurement.
Warnings and Cautions: The general warnings and cautions symbol signifies the
potential for bodily harm or damage to equipment.
MBShortcut for Temperature range (in German: Messbereich)
1.1.2Terminology
The terminology used in this manual corresponds to the VDI- / VDE-directives 3511, Part 4.
1.2Safety
This manual provides important information on safely installing and operating the
IGA 6/23 Advanced pyrometer. Several sections of this manual provide safety warnings to avert
danger. These safety warnings are specified with a warning symbol. You must read and
understand the contents of this manual before operating the instrument even if you have used
similar instruments or have already been trained by the manufacturer.
It is also important to continually pay attention to all labels and markings on the instrument and
to keep the labels and markings in a permanent readable condition.
Warning: The pyrometer is only to be used as described in this manual. It is
recommended that you only use accessories provided by the manufacturer.
In addition, signs and markings on the device is to be observed and maintained in legible
condition.
1.3Limit of liability and warranty
All general information and notes for handling, maintenance, and cleaning of this instrument
are offered according to the best of our knowledge and experience.
LumaSense Technologies is not liable for any damages that arise from the use of any examples
or processes mentioned in this manual or in case the content of this document should be
IGA 6/23 Advanced ManualContents · 5
incomplete or incorrect. LumaSense Technologies reserves the right to revise this document and
to make changes from time to time in the content hereof without obligation to notify any
person or persons of such revisions or changes.
All instruments from LumaSense Technologies have a regionally effective warranty period.
Please check our website at http://info.lumasenseinc.com/warranty for up-to-date warranty
information. This warranty covers manufacturing defects and faults which arise during
operation, only if they are the result of defects caused by LumaSense Technologies.
The Windows compatible software was thoroughly tested on a wide range of Windows
operating systems and in several world languages. Nevertheless, there is always a possibility that
a Windows or PC configuration or some other unforeseen condition exists that would cause the
software not to run smoothly. The manufacturer assumes no responsibility or liability and will
not guarantee the performance of the software. Liability regarding any direct or indirect
damage caused by this software is excluded.
The warranty is VOID if the instrument is disassembled, tampered with, altered, or otherwise
damaged without prior written consent from LumaSense Technologies; or if considered by
LumaSense Technologies to be abused or used in abnormal conditions.
There are no user-serviceable components in the instrument:
·No adjustments may be made to the targeting laser. It is fixed at the factory.
·No adjustments may be made to the targeting laser's power level.
1.4Unpacking the Instrument
Before shipment, each instrument is assembled, calibrated, and tested at the LumaSense Factory.
When unpacking and inspecting your system components, you need to do the following:
1. Check all materials in the container against the enclosed packing list.
LumaSense Technologies cannot be responsible for shortages against the packing list
unless a claim is immediately filed with the carrier. Final claim and negotiations with the
carrier must be completed by the customer.
2. Carefully unpack and inspect all components for visible damage. If you note any damage
or suspect damage, immediately contact the carrier and LumaSense Technologies, Inc.
3. Save all packing materials, including the carrier’s identification codes, until you have
inspected all components and find that there is no obvious or hidden damage.
Note: LumaSense encourages you to register your product with us to receive updates,
product information, and special service offers:
http://info.lumasenseinc.com/registration.
.
1.5Service Request, Repair, or Support
Contact LumaSense Technologies Technical Support in case of a malfunction or service request.
Provide clearly stated details of the problem as well as the instrument model number and serial
number. Upon receipt of this information, Technical Support will attempt to locate the fault
and, if possible, solve the problem over the telephone.
If Technical Support concludes that the instrument must be returned to LumaSense Technologies
for repair, they will issue a Return Material Authorization (RMA) number.
Return the instrument upon receipt of the RMA number, transportation prepaid. Clearly
indicate the assigned RMA number on the shipping package exterior. Refer to Section 1.6,
Shipments to LumaSense for Repair, for shipping instructions.
IGA 6/23 Advanced ManualContents · 6
Technical Support can be contacted by telephone or email:
Santa Clara, California
·Telephone: +1 408 727 1600 or +1 800 631 0176
·Email: support@lumasenseinc.com
Frankfurt, Germany
·Telephone: +49 (0) 69 97373 0
·Email: support@lumasenseinc.com
Erstein, France
·Telephone +33 (0)3 88 98 98 01
·Email support@lumasenseinc.com
1.6Shipments to LumaSense for Repair
All RMA shipments of LumaSense Technologies instruments are to be prepaid and insured by
way of United Parcel Service (UPS) or preferred choice. For overseas customers, ship units airfreight, priority one.
The instrument must be shipped in the original packing container or its equivalent. LumaSense
Technologies is not responsible for freight damage to instruments that are improperly packed.
Contact us to obtain an RMA number (if one has not already been assigned by Technical
Support). Clearly indicate the assigned RMA number on the shipping package exterior.
Send RMA Shipments to your nearest technical service center:
Santa Clara, CaliforniaFrankfurt, Germany
LumaSense Technologies, Inc.
3301 Leonard Court
Santa Clara, CA 95054 USA
Telephone: +1 408 727 1600
+1 800 631 0176
Inoperable IMPAC pyrometers must be disposed of in compliance with local regulations for
electro or electronic material.
IGA 6/23 Advanced ManualContents · 7
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced ManualContents · 8
2 Introduction
50to1000°C (MB
10)
Sub Range:
Any range adjustable within the tem
perature range, minimum span
Spectral Ranges:
2 to 2.6
µm(main wavelength 2.3
µm)Resolution:
0.1 °C or 0.2 °F at interface;
< 0.0015%
of adjusted temperature range
Emiss
ivitye:
0.
050to1.000insteps of 1/1000
Transmittance
t
:
0.050 to
1.000
in steps of 1/1000
Response Time
t90:
0.5 ms (with dynamic adaptation at low signal levels)
Measurement Uncertainty:
< 1500 °C: 0.3% of reading in °C + 2 °C
Repeatability:
Umg
.
0.15% of reading in °C + 1 °C
Noise Equivalent
Temperature /
NETD
2.1Appropriate use
The IMPAC IGA 6/23 is a short wave infrared temperature measuring device with digital signal
processing. It is used for non-contact temperature measurements on metals, ceramics, graphite,
etc. with a temperature range between 50 and 1800 °C.
2.2Scope of delivery
Pyrometer, PC adjustment, and evaluation software InfraWin, works certificate, and operating
instructions.
Note: A connection cable is not included with the instrument and has to be ordered
separately (see Chapter 8, Reference numbers).
Note: During operation the
instruments will warm up and might
reach an intrinsic temperature of up
to 58 °C
0 to 70 °C at housing
Storage Temperature:-20 to 80 °C
Weight:0.6 kg
Housing:Stainless steel
CE-Label:According to EU directives about electromagnetic immunity
Interface
Connection:12-pin connector
distance
emissivity, sub range, ambient temperature compensation, wait time tw,
settings for maximum value storage, address, baud rate, transmittance,
response time t90, 0 to 20 mA or 4 to 20 mA analog output range, °C/°F
Readable via interface: measured value, internal temperature of the
unit, measuring distance
Communication
Baud rate: 1200 Bd to 115.2 kBd (on request RS232, not addressable)
10 ms; 50 ms; 250 ms; 1 s; 5 s; 25 s), via interface, automatically with the
next measuring object, hold-function
Electrical
clear
isolated from each other
IGA 6/23 Advanced ManualIntroduction · 10
Note: The calibration / adjustment of the instruments was carried out in accordance
with VDI/VDE directive “Temperature measurement in industry, Radiation
thermometry, Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4.
For additional details on this directive, see http://info.lumasenseinc.com/calibration or
order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany.
2.4Dimensions
IGA 6/23 Advanced with the view
finder
IGA 6/23 Advanced with laser
sighting
2.5Physical User Interface
1 12-pin connector
2 Digital display
3 Optional sighting
4 Screw to adjust the
focus
5 LED distance display
6 LED indicator light
7 Adjusting for the view
finder
IGA 6/23 Advanced
with laser sighting
IGA 6/23 Advanced ManualIntroduction · 11
IGA 6/23 Advanced
with the view finder
Warning
The stainless steel housing with integrated water cooling air purge
The vacuum pickup
KF 16
with window allows the user to easily mount
Tube support
: If selected, the IGA 6/23 Advanced laser light
option is a Class 2 and emits laser light.
To minimize the risk of eye injury, do not look directly into
the beam, and do not point the laser beam of the pilot light
into the eyes of another person.
2.6Accessories (optional)
Numerous accessories guarantee easy installation of the pyrometer. The following overview
shows a selection of suitable accessories. You can find the entire accessory program with all
reference numbers in Chapter 8, Reference numbers.
2.6.1Mounting
Mounting brackets are available for safe mounting and alignment of
the pyrometer to the measuring object.
Mounting angle
2.6.2Cooling Jacket
allows use of the pyrometer above the maximum permissible ambient
temperature. The pyrometer can be operated in ambient temperatures
up to 180 °C.
Water cooling jacket with
integrated air purge
2.6.3Air Purge
The air purge protects the lens from contamination of dust and
moisture. It has to be supplied with dry and oil-free pressurized air and
generates an air stream shaped like a cone.
2.6.4Vacuum Pickup
the pyrometer on vacuum systems.
2.6.5Scanning Mirror
The scanning mirror unit SCA 5 allows the measured object to be
scanned over a certain range. The measuring beam of the pyrometer
moves straight in one line across the object and collects temperature
data of this line. This is useful when used in combination with the
maximum value storage (peak picker) to measure objects which move
out of the target area. The scanning angle of the mirror is 0 to 12° and
the scanning frequency 0 to 5 Hz. Both values are easily adjustable at
the scanner.
Air purge
Vacuum Pickup
Scanning Mirror
2.6.6Flange System
IGA 6/23 Advanced ManualIntroduction· 12
with air purge
nozzle and
flange
The flange system is a modular mounting system to fix the pyrometer on furnaces, vacuum
Schematic drawing of the flange system
Pyrometer
Ceremaic
Instrument’s
chambers, etc.
600 x 24, closed
sighting tube
It can consist of e.g. mounting support, tube support with air purge and flange and an open or
closed ceramic sighting tube. The mounting support can be equippedwith a quartz window for
vacuum applicationsIt may consist of an equipment rack, flange, and an open or closed ceramic
IGA 6/23
Advanced
support
tube. The equipment rack can be equipped for vacuum applications with a fused silica.
2.7Transport, Packing, Storage
With faulty shipping, the instrument can be damaged or destroyed. To transport or store the
instrument, please use the original box or a box padded with sufficient shock-absorbing
material. For storage in humid areas or shipment overseas, the device should be placed in
welded foil (ideally along with silica gel) to protect it from humidity.
The pyrometer is designed for a storage temperature of -20 to 80 °C with non-condensing
conditions. Storing the instrument out of these conditions can cause damage or result in
malfunction of the pyrometer.
IGA 6/23 Advanced ManualIntroduction · 13
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced ManualIntroduction · 14
3 Controls and Installation
12-pin connector
Pin assignment
3.1Electrical Installation
The IGA 6/23 Advanced is powered by a voltage of 24 V DC ± 25% (ripple < 50 mV). It is
important to ensure correct polarity when connecting the device to the power supply.
To meet the electromagnetic requirements (EMV), a shielded connecting cable must be used.
LumaSense offers connecting cables, which are not part of the standard scope of delivery. The
shield of the connecting cable has to be connected only on the pyrometer’s side. If the
connecting cable is extended, the shield of the extension also needs to be extended. The shield
must be open on the power supply side (switch board), to avoid ground loops.
The connecting cable has wires for the power supply, interface, analog output, switch contact,
and external clearing of the maximum value storage via contact and 12-pin connector (see
Chapter 8, Reference numbers).
Attention:
Once the instrument has been connected to the power supply, it is immediately ready for use.
Although it does not need to be warmed up, it does need to run for approximately 15 to 30
minutes before achieving full accuracy. The instrument can be switched off by interrupting the
power supply or unplugging the electrical connector.
When connecting the power supply, ensure the polarity is correct.
3.1.1Pin assignment of the connector
(view of the connector pins)
IGA 6/23 Advanced ManualControls and Installation · 15
Pin Color Function
K
white
+24 V DC
power supply
A
brown
0 V DC
power supply
L
green
+ I
Ausg.
analog output
B
yellow
–
I
Ausg.
analog
output
Targeting light activate / deactivate via
External clearing of max. value storage
G
red
DGND
(GND
for interface
)
F
black
B1 (RS485) or
RxD (RS232)
C
violet
A1 (RS48
5) or
TxD (RS232)
D
gray-
pink
B2 (RS485) (
bridged with
F)
E
red-blue
A2 (RS485) (
bridged with
C)
Screen only for cable extension,
H gray
J pink
M orange
external switch (bridged with K)
(bridge to K) or hold function*
don’t connect at the switchboard
*The connector pin J can be used for two different functions:
1. External clearing of the maximum value storage: When the pyrometer is in
operating mode, pin J can be used for external clearing of maximum value storage.
When external clearing is selected from the tclear dropdown menu, pin J is connected
for a short time to pin K (power supply voltage) to clear the stored maximum value. The
function “external clearing” is triggered when the clear time is set to “extern”.
2. Hold function: When the hold function mode is activated, the current temperature
reading is frozen as long as pin J and pin K are connected. (See section 4.6 for clear time
for the maximum value storage).
3.1.2Connecting the pyrometer to a PC
The pyrometer is equipped with an RS485 serial interface. With the RS485, long transmission
distances can be realized and the transmission is, to a large extent, free of problems. The RS485
also allows several pyrometers to be connected in a bus system.
If an RS485 connection is not available at the PC, it can be accomplished using the RS485 to USB
connector. When using a RS485 to USB adapter, make sure that the adapter is fast enough to
receive the pyrometer’s answer to an instruction of the master. Most of the commonly used
adapters are too slow for fast measuring equipment, so it is recommended to use the
RS485 ó USB converter “USB nano“ (ref. no. 3 852 600).
3.1.3Connection to RS485
Half-duplex mode: A1 and A2 as well as B1 and B2 are bridged in the 12-pin round connector of
the connecting cable, to prevent reflections due to long stubs. It also safeguards against the
interruption of the RS485 bus system, should a connecting plug be pulled out. The master labels
mark the connections on the RS485 converter.
The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the
cable. Adjusted values are from 1200 to 115200 Bd.
IGA 6/23 Advanced ManualControls and Installation · 16
Terminator
Pyrometer 1
e.g. Address 00
Pyrometer 2
e.g. Address 01
Pyrometer 32
e.g. Address 31
RS485 Bus System
3.1.4Connection of Additional Units
For temperature indication of the pyrometer, LumaSense offers pure indicators (series DA 4000).
LumaSense also offers indicators with features to change pyrometer parameters (DA 6000 and
DA 6000-N) as well as a fast digital PID controller PI 6000.
Digital DisplayParameterizing IndicatorDigital Controller
Additional analyzing instruments, including LED digital displays only need to be connected to a
power supply and the analog outputs of the pyrometer (exception: the digital display DA 6000
can also be connected with its serial interface, whereas the digital display DA 6000-N has to be
connected with its serial interface).
Connection of additional units
Other instruments like an analog controller or printer can be connected to the display in a series
as shown above (total load of resistance max. 500 Ohm).
IGA 6/23 Advanced ManualControls and Installation · 17
3.2Sighting
Viewfinder Adjustment Ring
The IGA 6/23 Advanced can be purchased with Through-Lens Sighting (viewfinder) or with a
Laser Targeting Light. These sighting options allow you to easily align the pyrometer to the
measuring object.
3.2.1Viewfinder
The IGA 6/23 Advanced can be equipped with a viewfinder which offers through-lens sighting.
The viewfinder is true-sided and parallax-free. A circle marks the position of the measuring spot,
but not the exact spot size.
The viewfinder is equipped with an adjustable eye-protection filter, which allows you to view
hot objects without exposing your eye to high intensity light. When you are viewing hot objects,
turn the adjustment ring on the viewfinder so the filter will let in less light.
Warning: To ensure eye protection, make sure you use the shaded filter when
viewing objects with temperatures of 1500 °C or higher.
Note: You can turn the adjustment ring on the viewfinder in both a clockwise and
counterclockwise direction to change the filter from light to dark.
When you are viewing low temperature objects, turn the viewfinder adjustment ring so the
filter will let in more light.
3.2.2Laser Targeting Light
The IGA 6/23 Advanced can be equipped with a laser targeting light to assist with aligning the
pyrometer to the measuring object. The laser targeting light is a visible red light with a
wavelength between 630 and 680 nm and a maximum power of 1 mW. The laser is classified as
product of laser class II.
The laser spot marks the center of the measuring spot on the target. The diameter does not
correspond to the spot size. The smallest diameter of the laser spot approximately indicates the
measuring distance.
Warning:
that emits radiation.
To reduce the risk of injury to the eyes, do not look directly
into the targeting laser and do not point the targeting laser
into anyone's eyes.
The instrument is equipped with a class II laser
IGA 6/23 Advanced ManualControls and Installation · 18
Never look directly into the laser beam. The beam and spot can be watched safely from side.
Push button for
Also, make sure that the beam will not be reflected into eyes of people by mirrors or shiny
surfaces.
The laser targeting light can be switched on and off by pressing the button of the rear cover of
the housing.
Laser Targeting Light
Note: The laser warning signs on the pyrometer should be easily viewable at all times,
even after it has been installed.
The laser targeting light can also be switched on and off by using an external contact (see
section 3.1.1 Pin Assignment) or through the InfraWin software. If it is not switched off by one
of the above-mentioned methods, it will be switched off automatically after approximately
2 minutes.
To prevent damage to the laser, the targeting light is also switched off automatically if the
internal temperature of the pyrometer exceeds 61 °C. It can only be used again once the
temperature falls below 61 °C.
3.3 Optics
3.3.1Spot Sizes
The IGA 6/23 Advanced has a Vario optics, which can be manually adjusted at all distances
between 210 mm and 5000 mm.
The table of Spot Sizes in Relation to Measuring Distance shows examples of the pyrometer’s
spot size M [mm] in relation to the measuring distance a [mm] (min. 90% of the radiation
intensity). Increasing or decreasing the measuring distance will change the spot size.
Note: Since these are 1-color (mono) pyrometers, they can measure objects at any
distance (whether focused or non-focused). However, the object has to be bigger than,
or at least as big as, the spot size of the pyrometer in the measuring distance.
IGA 6/23 Advanced ManualControls and Installation · 19
Aperture D for all temperature ranges is 13 to 15 mm with the aperture being the effective
Measuring
Spot Diame
ter
Spot Diameter
Spot Diameter
210
4.2
2.1
0.6
30063
0.9
500105
1.5
800168
2.3
1300
26133.7
2000
40205.8
5000
1005015
diameter of the lens. This is dependent on the optical setting. The largest value applies to a very
small measuring distance, while the minimum value applies to the largest measuring distance.
Distance
a [mm]
Examples of Spot Sizes in Relation to Measuring Distance
Note: Effective aperture D for all temperature ranges is 13 mm (focused to longest
distance) to 15 mm (focused to shortest distance).
Focused spot sizes between the listed distances can be found by linear interpolation between
the listed values.
M [mm]
MB 10
M [mm]
MB 13
M [mm]
MB 18
3.3.2Deviation from the focused measuring distance
Spot sizes for non-focused distances (shorter or longer than the focused distance) may be
calculated by using the formula below.
Formula for Calculating Spot Sizes
The InfraWin software also includes a Spot Size Calculator that calculates the data for the
non-focused regions, if you enter the values of aperture D, focused measurement distance a,
and focused measuring field diameter M as found in the above table (see section 3.3.1).
3.3.3Finding the right measuring distance
The measuring distance can be set using the Focus Adjustment Screw on the back of the device.
The focused distance value can be adjusted at all distances between 210 mm and 5000 mm. To
focus, turn the focus adjustment set screw to make the target image appear sharp and clear.
Note: The optics are manually focusable with a measuring distance of a = 210 to
5000 mm.
IGA 6/23 Advanced ManualControls and Installation · 20
The LED Distance Indicator Light (labeled mm) will turn red and the approximate focused
measuring distance in mm will automatically be shown on the Digital Display for a few seconds
after making an adjustment using the Focus Adjustment Set Screw.
Note: Turning the focus adjustment screw counterclockwise will shorten the
measuring distance.
Turning the focus adjustment screw clockwise will lengthen the measuring distance.
IGA 6/23 Advanced ManualControls and Installation · 21
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced ManualControls and Installation · 22
4 Settings / parameter descriptions
The pyrometer is equipped with a wide range of settings for optimal adaptation to the required
measuring condition and to measure the temperature correctly.
All settings can be read and set only in the pyrometer parameters window of the
software InfraWin. Adjusting the settings at the instrument is not possible (detailed
description of the software see Chapter 5, Software InfraWin).
Selecting the pyrometer parameters window shows the current settings of the
pyrometer. Changing a value can either be done by typing a value in an input box or by
selecting a preset value from the list field.
4.1 Factory Settings
Temperature display (°C / °F) = °C
Emissivity (ε) = 100%
Transmittance (W) = 100%
Exposure time (t90) = min
Dynamical adaptation dm = 4
Clear time (t
max / min = maximum value storage preset
Analog output (out) = 0 ... 20 mA
Sub range same as basic temperature range
Ambient temperature compensation = auto
Address = 00
Baud rate = 19200 Bd
Clear
) = off
4.2 Temperature Display
The Digital Display on the back of the pyrometer will show the temperature measurement in
either °C or °F. You can select which temperature scale you wish to use through the InfraWin
software or by using the UPP Data Format commands.
The Measuring Value can also be viewed at any time through the InfraWin Software or by using
the UPP Data Format commands.
4.3Emissivity ε
Different materials have different emissivities ranging between 0% and 100%. The emissivity is
also dependent upon on the surface condition of the material, the spectral range of the
pyrometer, and the measuring temperature. The emissivity setting of the pyrometer has to be
adjusted accordingly.
Emissivity settings between 5% and 100% can be established through the InfraWin software or
by using the UPP Data Format commands.
IGA 6/23 Advanced ManualSettings / parameter descriptions · 23
Measuring object
Emissivity [%]
Emissivity [%]
“Black body furnace“
100
Steel, molten
20 to 30
Extruded Aluminum
13
Nickel
15 to 20
Brass
18
Titanium, non
-
oxidized
50
Brass oxidized (tarnished)
65 to 70
Titanium, oxidized
75 to 80
Copper
5
Molybdenum
10
Cop
per, oxidized
70 to 80
Molybdenum, oxidized
75 to 80
Inconel
30
Black Carbon
95
Inconel, oxidized
85
Graphite
80 to 90
Oxidized Iron
85 to 90
Stoneware, glazed
60
Steel rolling scale
80 to 88
Porcelain rough
80 to 90
Dependency of the response time on th
e temperature value
MB10Temperature value
60°C120°C150
°C
t90min
2 ms
1 ms
0.5 ms
MB13Temperature value
100°C150°C200
t90min
2 ms
1 ms
0.5 ms
MB18Temperature value
200°C250°C300
t90min
4 ms
1 ms
0.5 ms
Note: Emissivity e can be set from 5% to 100% in steps of 0.1%
(at 2.3 µm)
Measuring object
(at 2.3 µm)
4.4Transmittance t
Transmittance is a parameter that can compensate for signal loss due to external windows etc.
For example, if the emissivity of the material is 0.6 and the transmittance of an additional
window is 0.9, then the product would be 0.54 which is well inside the allowed range.
The product of transmittance and emissivity (t x e) must not be less than 5%.
Note: Transmittance t can be set from 5% to 100% in steps of 0.1%
4.5Response Time (t90)
The response time t90 is the time interval for the analog output of the pyrometer to go from a
low temperature value up to 90% of the temperature step to a high value when measuring an
abrupt increase from said low to said high temperature.
The dynamic exposure time adjustment prolongs the exposure time at the lower range limit.
With the InfraWin software and the UPP Data Format commands it is possible to configure the
dynamic exposure time.
°C
°C
Independently of this, the pyrometer performs a measurement every 25 µs and updates the
analog output. Slower response times can be used to achieve a constant temperature reading
for measuring objects that have rapidly fluctuating temperatures.
The response time is set using the InfraWin software or by using the UPP Data Format commands.
When the setting is set to min., the IGA 6/23 Advanced operates using a time constant of
IGA 6/23 Advanced ManualSettings / parameter descriptions · 24
approximately 0.5 ms. The response time can be extended to 1 ms; 3 ms; 5 ms; 10 ms; 50 ms;
The integrated maximum value storage is activated when the parameter t
is set to something other than “OFF” or “HOLD”.
If the maximum value storage is switched on, the highest last temperature
value will always be displayed and stored. As such, it may be beneficial to
periodically clear and reset the stored maximum values in order to obtain
new temperature readings.
CLEAR
)
clear
off
10 ms
50 ms
250ms
1 s
5 s
This storage also has to be cleared at regular intervals when fluctuating
object temperatures cause the display or the analog outputs to change too
rapidly or when the pyrometer is not constantly viewing an object to be
25 s
extern
auto
measured.
4.6.1Single and Double Storage Modes
Depending upon the selected settings, the maximum value storage will either work in single
storage mode or in double storage mode.
Single Storage
Mode:
Single storage mode is used when you want to reset the stored value using
an external impulse via one contact closure from an external relay (such as
between two measured objects). The relay contact is connected directly to
the pyrometer between pins J and K. This mode allows a new value to be
established after each impulse from the reset signal. Single storage mode
also comes into effect when the Clear Peak Memory t
is set to AUTO.
clear
Double Storage
Mode:
Note: The maximum value storage setting coincides with adjustments made to the
response time.
The response time setting (working like a low-pass filter) is applied first. After that,
the maximum storage is processed. So when using both, the maximum storage takes
the peak of the signal that was previously smoothed by the response time filter.
Double storage mode comes into effect when selecting one of the reset
intervals. This mode utilizes two memories. With the first memory, the
highest measured value is held and is deleted alternately in the time
interval set (clear time). The other memory retains the maximum value
throughout the next time interval. The disadvantages of fluctuations in the
display with the clock frequency are thereby eliminated.
IGA 6/23 Advanced ManualSettings / parameter descriptions · 25
4.6.2Clear Time Settings
Einstellu
n
gen:
The following settings are available through the InfraWin software or by using the UPP data
format commands.
OFFWhen set to OFF, the maximum value storage is switched off and all new
temperature values are measured but not stored.
10 ms…25,0 sIf the clear time is set between 10 ms and 25.0 s, the maximum value is
held in double storage mode. After the entered time, the value will be
cleared alternately from one of the storages, while the value of the other
storage is shown.
EXTERN
AUTOThe AUTO mode is used for discontinuous measuring tasks. For example,
HOLDThe HOLD function allows you to freeze the current temperature reading at
With the external clearing function, the storage operates in single storage
mode. The values are immediately cleared from the storage by contacting
the wires connected to pins J and K, if the EXTERN mode was selected.
when objects are being transported on a conveyer belt and pass the
measuring beam of the pyrometer only for a few seconds. In this case, the
maximum value for each object has to be obtained.
With the AUTO mode, the maximum value is stored until a new hot object
appears in the measuring beam. The temperature, which has to be
recognized as “hot” is defined by the low limit of the adjusted sub range.
The stored maximum value will be deleted once the temperature of the new
hot object exceeds the low limit of the sub range by 1% (transition in
positive direction) or by at least 2 °C. This is also valid if the sub range equals
the basic range.
any moment. This feature is activated using an external switch that has been
connected between connector pins J and K.
The temperature reading will remain frozen as long as the contact remains
closed.
4.7Analog Output
The analog output has to be selected according to the signal input of the
connected instrument (controller, PLC, etc.). If 4 to 20 mA is set, the analog
output gives 3.9 mA for temperatures below lower range limit.
0 … 20 mA
4 … 20 mA
4.8Sub Range
You have the opportunity to choose a sub range (minimum span 50 °C) within the basic
measuring range of the pyrometer. This sub range corresponds to the analog output.
Example: Range 150…1800 °C, Sub Range 925…975 °C.
The sub range setting also affects the maximum value storage when the Clear Peak Memory
t
is set to AUTO. For more information on the t
clear
Note: Settings for Sub Range - any range adjustable within the temperature range
with a minimum span of 50 °C.
IGA 6/23 Advanced ManualSettings / parameter descriptions · 26
AUTO setting, refer to section 4.6.
clear
4.9Device Address
Settings:
When connecting several pyrometers to one serial interface with RS485, it is
necessary for each instrument to have its own device address for
communication purposes. First, it is necessary to connect each instrument
separately to give it an address. After that, all instruments can be connected
and addressed individually.
Note: Only via own communication program with interface command (not
possible with InfraWin, because InfraWin automatically detects a connected
pyrometer): If parameters should be changed simultaneously on all pyrometers, the
global Address 98 can be used. This allows you to program all pyrometers at the
same time, regardless of the addresses that have already been assigned. If the address
of a pyrometer is unknown, it is possible to communicate with it using the global
Address 99 (connect only one pyrometer).
4.10 Focused Distance
The focused distance value can be
adjusted at all distances between 210 mm
and 5000 mm using the Focus Adjustment
Screw on the back of the device.
The LED Distance Indicator Light will turn
red and the focused measuring distance in
mm will automatically be shown on the
Digital Display within a few seconds of
making an adjustment using the Focus
Adjustment Set Screw.
00
97
.
.
.
The focused distance can be viewed at
any time through the InfraWin software or by using the UPP Data Format commands.
4.11 Baud Rate
The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the
cable. A maximum cable length for 19200 Bd with RS485 is 2 km. The baud rate is reduced by
50% if the transmission distance is doubled.
Baud Rates can be set at 1200, 2400, 4800, 9600, 19200, 38400, 57600, or 115200.
4.12 Pyrometer Internal Temperature
The internal temperature of the pyrometer can be read through the PC interface using the
InfraWin software or by using the UPP Data format commands. It is a few degrees higher than
the ambient temperature due to the heat generated by the electronics.
When using the Laser Targeting Light, the targeting light is switched
off automatically if the internal temperature of the pyrometer exceeds
61 °C. This safety feature is used to prevent damage to the laser. It can
only be used again once the temperature falls below 61 °C.
IGA 6/23 Advanced ManualSettings / parameter descriptions · 27
4.13 Ambient Temperature Compensation
Settings:
The compensation of the ambient temperature can be set for temperatures
within the basic measuring range. This compensation is only used for very
few special applications. The standard setting of this parameter is “auto",
because the temperature of the air around the pyrometer is normally the
ambient temperature of the measured object. Should the measured object be
placed in an area with a higher wall temperature (e.g. inside a furnace), the measurement might
be falsified (probably too high temperature indication). This influence can be compensated by
presetting the ambient temperature of the object (presetting within the measuring range of the
instrument). After switching over to “man” the corresponding data field T
is activated so that
amb
the ambient temperature value can be entered.
Attention: It has to be considered that this method only improves the results if the
ambient temperature at the place of the measured object is always constant and the
emissivity is well known.
auto
00 °C (32 °F)
.
.
.
70 °C (158 °F)
IGA 6/23 Advanced ManualSettings / parameter descriptions · 28
5 Software InfraWin
The operating and analyzing InfraWin software is included with delivery of the pyrometer. In
addition to allowing you to make parameter adjustments, the InfraWin software also provides
temperature indication, data logging, and measurement analysis features.
This section gives an overview about the functions of the software. It also provides a description
of the individual icons found in the program's help menu. Click on the F1 button after loading
InfraWin or click on the ? in the menu bar.
The following descriptions refer to software InfraWin version 5.0. The latest version is available
for free as download from the homepage www.lumasenseinc.com.
5.1 Connecting the pyrometer to a PC
The program InfraWin can operate up to two devices. Two devices using RS485 may be operated
simultaneously by the same interface, if two different addresses have been properly entered
(see section 4.8 Device Address for more information).
5.2 Installation
To install the InfraWin software, select setup.exe from the InfraWin-CD or from the downloaded
and unpacked zip file from the internet and then follow the installation instructions.
5.3 Program start
The first time you load InfraWin 5, you will be prompted to select a default language. The
InfraWin software is available in German, English, Spanish, French, Portuguese, and Chinese.
Once installed, click Language/Languages if you would like to select another language.
On the start page, the screen shows the following start menu icons.
Opens a saved file
Storage of measured values for further processing
Starts measurement for the selected device
Setting of the parameters of the instruments
Time interval between two measurements
Listing of measured or stored values in tabular form
Processing of measured (stored) readings in graph form
Processing of measured (stored) readings in a text file
Calculation of spot sizes in various measuring distances
Only if available: controls the programmable controller PI 6000
Search I/O module
IGA 6/23 Advanced ManualSoftware InfraWin · 29
5.4Basic settings
All preset values for the device
can be displayed and
modified, if necessary under the
Devices/Parameters window.
Changing an existing pyrometer
setting can be accomplished by typing
a value in an input box or by selecting
a preset value from the list field.
Choose the correct settings for your
application from the displayed
options.
This window contains the parameter
settings described in Chapter 4,
Parameters.
5.4.1Open/Save
The open and save buttons enable you to save customized configurations, which can
then be opened and reused at any time.
5.4.21 measur…
The 1 measur… function shows the current measuring temperature in the
pyrometer parameters window for approximately 1 second.
5.4.3Print
The print icon will generate a screenshot of the parameter form and allow you to
send it a printer.
5.4.4Close
The close icon will close the pyrometer parameters window.
5.4.5Test
Clicking on the Test
icon will open a
window that allows you to directly
communicate with the pyrometer
using the UPP (Universal
Pyrometer Protocol) Data Format
commands.
After entering an interface
command (00 is the adjusted
address ex works, ms is the
command reading temperaturevalue) and clicking on Send, the
window shown to the right will be
opened.
This window already shows the
answer of the pyrometer in 1/10°. The actual temperature reading is 49.0 °C.
Len indicates the length of the answered data string, incl. Carriage Return (Chr(13)).
IGA 6/23 Advanced ManualSoftware InfraWin · 30
In the lower part of the window, the connection with the preset baud rate can be checked. Here
the command was sent 100 times with 19200 baud. It has taken 1.100 seconds without
transmission errors.
Emi: AutoFind: If the true temperature of the measured object is known, you can calculate the
emissivity of the measured object using the Emi: AutoFind function:
A measured temperature is displayed with
the current set emissivity (in this example
100%) (here: 116 °C).
If you press Emi: Autofind a window will
open which allows you to enter the "true"
temperature (here 130 °C).
Once the temperature entry has been entered
and confirmed with Calculate, InfraWin will
then calculate the emissivity which occurs
with the new temperature (here 79.0%). This
is displayed immediately and can be used for
further temperature measurement.
5.5 Measurement online trend
The measurement function
allows you to access a number
of input tabs located on the left side of the
screen. The main or home tab is the Output
Screen. You can toggle the input tabs on and
off by clicking them. This window displays:
·temperature as graphical diagram
·internal temperature of the instrument
·current temperature
·quantity of the measured values and
file size of the current measurement
The example shows a sample reading over the period of approximately 15 seconds with a
temperature range between 400 and 800 °C. The final temperature (at the end of the reading) is
431 °C.
The Mark Zone button allows you to color mark a temperature range for easier recognition.
The Threshold button allows you to set a temperature value as a baseline to prevent recording
values above or below the baseline temperature. This allows you to keep the output file size
small.
IGA 6/23 Advanced ManualSoftware InfraWin · 31
The Scaling Trend button allows you to scale temperature trend view.
Note: The measuring values of “measurement online trend” are automatically saved
as "standard.i12". Should you need to edit the data later, you need to save the file as
another .i12-file because old values are over-written when a new measurement is
taken.
Files from older program versions (.i10-files) can be opened and saved as .i12.
5.6 Listing (analyzing)
For analyzing
the measured
values in this field, all measured
data appears in a numeric list.
The column between time and
temperature provides a time
resolution of milliseconds. The
value specifies the time in seconds
after midnight (0:00 h). The
amount of data depends on the
frequency that readings were taken
(settings at 10.9 PC sampling rates).
As the amount of data increases, so does the amount of storage space required to save it. In
order to save room, all .i12 data files are stored by a binary code.
5.7 Output .TXT file (analyzing)
The same file as under Output listing may be converted into a text file and
can be easily opened, for example with Microsoft Excel. With the standard
import settings, Excel automatically formats the columns accordingly (with tabs as separators).
5.8 Trend output (analyzing)
The graph’s curve depicts the
temperature change over time
within the specified temperature
range.
Additionally, other information
appears in this window; such as
recorded time (x-axis) and
temperature in degrees (y-axis) as
well as the time and temperature
at the vertical cursor line which can
be dragged with the mouse.
Selecting the Trend output initially
causes all the saved data to be
displayed.
If the data exceeds an amount that can be reasonably represented, you may “Zoom“ in on a
partial segment using the mouse (such as the segment represented in the example). Under
“Total” you can return to the representation of the entire curve.
IGA 6/23 Advanced ManualSoftware InfraWin · 32
Note: The last reading is saved in the standard.i12 file and automatically appears in
this form upon opening Listing or Trend output.
Selecting file open with another file, the previous file will be overwritten and replaced by the standard.i12 file.
5.9 PC sampling rate (time interval between two
measurements)
This function sets a time interval. After
each interval, one measured value is
stored on the PC. Longer time intervals will result in
creating smaller stored file sizes. This function is mainly
used for long term measurements.
5.10 Spot size calculator
The InfraWin Spot Size Calculator
calculates the data for the nonfocused regions. To calculate data, enter the values
of aperture D, focused measurement distance a,
and focused measuring field diameter M as found
in the documentation relating to the optics
supplied with your specific pyrometer.
After entering the aperture and the main spot size,
the input of interim values calculates spot sizes in
different measuring distances of the fixed optics.
5.11 Search I/O Module
The I/O Module allows
accessories to connect to
the software and is used to trigger
measurement externally or to send a signal
(like a relay) under certain conditions.
IGA 6/23 Advanced ManualSoftware InfraWin · 33
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced ManualSoftware InfraWin · 34
6 Maintenance
6.1Cleaning the front window
Since the device does not contain parts that require regular maintenance, the only regular
maintenance required is periodic inspection of the front window for build-up of foreign
particiles. If allowed to build up, the particles can influence the energy received by the
instrument.
The IGA 6/23 Advanced window is not water soluable and can be cleaned with standard lens
tissue dampened with a commercially available glasses or camera lens cleaning solution. Use a
soft blower/brush (available in camera stores) to remove any grit on the window before you rub
the lens with lens tissue and solution.
Attention: NEVER CLEAN THE IGA 6/23 Advanced WINDOW WITH A DRY TISSUE
OF ANY KIND! The only time dry lens tissue may be used is to dry a window which has
already been cleaned with wet lens tissue.
6.2Calibration
LumaSense calibrated your IGA 6/23 Advanced pyrometer at the factory and delivered your
instrument with a Works Certificate. Normally we advise against changing the factory set
calibration. If you believe that the calibration may have changed, perhaps because your
operating environment is severe, an approximate field (on-site) calibration is possible. You have
the choice of an on-site calibration or arranging a more precise calibration at the LumaSense
factory.
6.2.1Laboratory Calibration
Contact LumaSense for information about calibration at the LumaSense factory. For most of our
customers who do not have large numbers of infrared thermometers in service, we recommend
that our laboratory be chosen to do calibration. When you have many infrared thermometers,
you may find it most convenient and economical to have your own calibration laboratory.
LumaSense has a variety of blackbody calibration sources including very economical ones.
Among these you may find the source most suitable for your own laboratory.
6.2.2On-Site Calibration
See Chapter 9 Troubleshooting before attempting to perform your own calibration. The
instrument was calibrated at the factory to its original accuracy as stated.
Nearly all erroneous temperature readings are caused by application problems such as:
·Reflections: Unexpectedly high readings may be caused by the IGA 6/23 Advanced
“seeing” a reflection of another hotter source, especially if the target emissivity is less
than 0.8.
·Spot Sizes: Ensure the IGA 6/23 Advanced is using a proper focused distance and
takes into account the spot size in relation to measuring distance. The pyrometer can
measure objects at any distance. However, the object has to be bigger than or at least as
big as the spot size of the pyrometer in the measuring distance.
·Obscured window: Clean the window. Refer to the cleaning window section 6.1.
IGA 6/23 Advanced ManualMaintenance · 35
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced ManualMaintenance · 36
7 Data format UPP (Universal Pyrometer
Description
Command
Parameters
0 = 1200 Baud 1 = 2400 Baud 2 = 4800 Baud
lowed 8 = 115200 Baud
Protocol)
Software commands can be exchanged directly with the pyrometer through an interface using
suitable communication software or by using the “Test” function located in the “Pyrometer
Parameters” window of the InfraWin software package.
The data exchange occurs in ASCII format with the following transmission parameters:
•The data format is: 8 data bits, 1 stop bit, even parity (8,1,e) no handshake;
•The device responds to the entry of a command with output (such as the measuring
value) + CR (Carriage Return, ASCII 13), to pure entry commands with “ok” + CR, or “no”
+ CR.
•Every command starts with the 2-digit device address AA followed by two lower case
command letters and finished with CR.
Example Read Command: Entry: “00em“ + <CR>
The emissivity setting (e) of the device with the address 00 is returned.
Answer: “0970“ + <CR> means Emissivity = 0.97 or 97.0%
•The ASCII parameter “X” indicates a change to be made in a parameter. When used, the
command contains the new value.
Example Write Command: Entry: “00emXXXX“ + <CR>
The parameter used for the emissivity setting (e) with the address 00 is changed.
Answer: “00em0853“ + <CR> changes the Emissivity to 0.853 or 85.3%
·A “?” after the lower case command letters answers with the limits of the respective
settings (only at setting commands, not at query commands).
Example Read Limits Command: Entry: “00em?“ + ˜CR!
Answer: Could be 00501000 + <CR>, which means that E can vary between 0.050 and
1.000 (or 5% and 100%)
Analog OutputAAasXX=0 (0 to 20 mA) X=1 (4 to 20 mA)
Reference Number AAbnOutput: XXXXXX (hex 6-digit)
Baud rate (set)AAbrXX = 0 to 6 or 8 (dec.)
3 = 9600 Baud 4 = 19200 Baud 5 = 38400 Baud
6 = 57600 Baud 7 = is not al
Emissivity e
Transmittance t of
window
AAemXXXX
AAem
AAetXXXX
AAet
XXXX = 0050 to 1000e=0.050 to 1.000
Answer: DDDD 4 decimial digits 0050 to 1000
XXXX = 0050 to 1000e=0.050 to 1.000
Answer: DDDD 4 decimial digits 0050 to 1000
IGA 6/23 Advanced ManualData format UPPÒ· 37
Description
Command
Parameters
Response time t90 AAezXX=0 to 9
Sub range (set)
0=min 1=1 ms 2=3 ms 3=5 ms
4=10 ms 5=50 ms 6=250 ms
7=1 s 8=3 s 9=10 s
Dynamic exposure
time
AAdmXX=0 to 9
0 = off, 1 = min, 9 = max
Temp display °C or °F AAfhXOutput: X=0 Display in °C
X=1 Display in °F
Device addressAAgaXXXX=(00 to 97)
00 to 97=regular device addresses
99=global address with response
Internal temperature
(read)
Ambient temperature
compensation
AAgt
Answer: DDD 3 decimal digits (000 to 098 °C or
032 to 210 °F)
AAtm
gt=current temp. tm=maximum temp. (memory)
AAutXXXX XXXX = value of ambient temperature, 4-digit, hex
XXXX e.g. FFEC corresponds to -20 degrees
-99dez = FF9Dhex means: automatic, no manual
compensation
LaserAAlaX
AAla
Wait time
AAtwXX
X=0 off X=1 on
Answer: 1-digit: “0” or “1”
XX = 00 … 99
(decimal, in bit time of the current baud rate)
Software simulation
AAlx
Clears maximum storage
of external clearance
Clear peak memory
t
clear
AAlzXX=0 to 9
0=OFF 1=10 ms 2=50 ms 3=250 ms
4=1 s 5=5 s 6=25 s
7=EXTERN 8=AUTO 9=HOLD
Basic range (read) AAmbAnswer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
Sub range (read) AAmeAnswer: XXXXYYYY
2x4 hex-digit for lower and upper range limit (°C or °F)
1. AAm1XXXXYYYY
AAm2AAm2 confirms the change (auto reset)
2.
XXXXYYYY=2x4 hex-digit for lower and upper sub range
limit (°C or °F)
Measuring value AAmsAnswer: QQQQQ (88880=Overflow)
5 decimal digit (in °C or °F, last digit is 1/10 °C or °F)
Device typeAAnaOutput: “IGA 6/23 Advanced”
(16 ASCII-characters)
IGA 6/23 Advanced ManualData format UPPÒ· 38
DescriptionCommand Parameters
Read Parameters AapaAnswer: 15 decimal digits
DD.............: Emissivity (see em)
..D............: Response time (see ez)
...D...........: Clear peak memory (see lz)
....D..........: Analog output (see as)
.....DD........: Internal temperature (see gt)
.......DD......: Device address (see ga)
.........D.....: Baud rate (see br )
..........01000
: Fixed value
Serial numberAAsnOutput: XXXXX (hex 5-digit)
Device type /
Software version
Communication
Module/ software
version in detail
Software version in
detail
AAveAnswer: VVMMJJ VV=27
MM=Month JJ = Year of software version
AAvctt.mm.jj XX.YY
tt = day; mm = month; jj = year;
XX.YY = Software version
AAvstt.mm.jj XX.YY
tt = day; mm = month; jj = year;
XX.YY = Software version
Note: the lett er “l” means the lower case letter of “L”.
Additional instruction for the RS485 interface:
Requirements to the master system during half-duplex operation:
1. After an inquiry, the bus should be switched into a transmission time of 3 bits (some
older interfaces are not fast enough for this).
2. The pyrometer's response will follow after 5 ms, at the latest.
3. If there is no response, there is a parity or syntax error and the inquiry has to be
repeated.
4. After receiving the response, the master has to wait at least 1.5 ms before a new
command can be entered.
IGA 6/23 Advanced ManualData format UPPÒ· 39
To ensure consistent document formatting, this page was intentionally left blank.
IGA 6/23 Advanced ManualData format UPPÒ· 40
8 Reference Numbers
8.1Reference numbers instrument
Temperature Range With Laser Targeting With Viewfinder
50 to 1000 °C (MB 10)3 914 2103 914 220
75 to 1300 °C (MB 13)3 914 2503 914 260
150 to 1800 °C (MB 18)3 914 2903 914 300
Ordering note:
A connection cable is not included in scope of delivery and has to be ordered separately.
8.2Reference numbers accessories
3 820 320 Special connection cable with angled connector and additional targeting light push
3 826 510 PI 6000: PID programmable controller, very fast, for digital IMPAC pyrometers
3 826 720 USB to RS485 adapter cable, 1.8 m long
3 834 210Mounting support (adjustable)
3 835 160Air purge unit, aluminum
3 835 590 90° mirror for Series 5 and Series 6, quartz glass window
3 837 230Water cooling jacket (heavy duty) with integrated air purge
3 843 250ROT5 scanning mirror attachment up to 70°
Connection cable, 5 m, with straight connector
Connection cable, 10 m, with straight connector
Connection cable, 15 m, with straight connector
Connection cable, 20 m, with straight connector
Connection cable, 25 m, with straight connector
Connection cable, 30 m, with straight connector
Connection cable, 5 m, with right angle connector
Connection cable, 10 m, with right angle connector
Connection cable, 15 m, with right angle connector
Connection cable, 20 m, with right angle connector
Connection cable, 25 m, with right angle connector
Connection cable, 30 m, with right angle connector
(All connection cables
include a short
adapter cable with a
9-pin D-SUB
connector. This
connector may be
used in combination
with the RS485 to USB
adapter.)
3 843 430Scanning mirror attachment up to 12°
3 846 260Instrument's support (Series 5 & 6)
3 846 290Instrument's support (Series' 5 & 6) with fused silica window
3 846 590 Vacuum support KF16 with quartz glass window
3 852 290
IGA 6/23 Advanced ManualReference Numbers · 41
Power supply NG DC for DIN rail mounting; 100 to 240 V AC Þ 24 V DC, 1 A
3 852 550
Power supply NG 2D for DIN rail mounting; 85 to 265 V AC Þ 24 V DC, 600 mA with
two settable limit switches
3 890 640 DA 4000-N: LED-digital display to be built into the switchboard, 230 V AC
3 891 210 DA 4000-N: LED-digital display to be built into the switchboard, 115 V AC
3 890 650 DA 4000: like DA 4000-N, but additionally with 2 limit switches, 230 V AC
3 891 220 DA 4000: like DA 4000-N, but additionally with 2 limit switches, 115 V AC
3 890 570 DA 6000-N digital display to allow adjustment of the Pyrometer through the RS485
interface
3 890 530 DA 6000: like DA 6000-N with analog input and 2 limit switches for the RS485
interface
3 890 630 LD24-UTP; large digital indicator, 57 mm height of digits
3 834 210 Adjustable mounting support (Series 5 & 6)
IGA 6/23 Advanced ManualReference Numbers · 42
9 Troubleshooting
No analog output
Reversed leads to
IGA 6/23 Advanced
The
IGA 6/23 Advanced
is
Reversed leads to other instruments
Open circuit in lines connecting all
Total burden higher than 500
Ohms
Insufficient supply voltage (See wiring
Temperature reading
Pyrometer output setting differs from
Check the 0 to 20 mA or
Temperature readings
Target size too small
3.2
Incorrect alignment of the pyrometer
Realign pyrometer to achieve
Dirty window or lens
6.1
Target temperature too low.
Incorrect slope adjustments
Faulty
IGA 6/23 Advanced
Contact LumaSense. See
Temperature readings
Targets hotter than expected. (This
Verify target temperature
The m
easurement is influenced by
Use mechanical accessories to
If target emissivity is less than 1,
Block reflection
paths to see if
Incorrect spectral response for
See Technicial Specifications
SymptomProbable CauseComments
even if the display
shows a temperature
above lower range
limit
differs between builtin display and external
instrumentation
are too low
in the current loop
instruments in current loop
diagram)
external instrument’s input setting
Target is not at correct focus distance
to the object
diode protected if power leads
are reversed.
See section 3.1
4 to 20 mA on both sides
Check the sub range setting
on pyrometer against the
range setting on the external
instrument
See section
With small target/close focus
models, aiming may require XY positioning table for fine
adjustment during aiming
the maximum temperature
signal
See Section
section 1.5
are too high
IGA 6/23 Advanced ManualTroubleshooting · 43
often happens with first time infrared
Pyrometer users)
reflections of hot machine parts
reflections from nearby hot objects
will influence readings
application. IGA 6/23 Advanced may
be “seeing” through target
with contact type method.
(Thermocouples may not
accurately determine surface
temperature due to “heat
sink” effect)
avoid the influence of the
interfering radiation
readings drop to expected
temperature
Consult LumaSense for
analysis.
SymptomProbable CauseComments
Temperat
ure readings
Faulty
IGA 6/23 Advanced
Contact LumaSense. See
Electrical noise in lead wires caused
3.1
Lens or window clouding up
Increase or decrease air purge
Improper grounding of cable shield
IGA6/23
Advanced
mounting not
Noi
sy readings (fast
Target moves more than 95% out of
Application may require
Target may be changing temperature
Spe
ed of response may be
Target may be reflecting nearby
Loose connection in wiring
Electrical noise lead w
ires caused by
Cable run into in same conduit as AC
Flame or reflections of flame may be
Extremely dense steam, smoke, thick
Non-uniform emissivity of large
Slower response speed is
Faulty
IGA 6/23 Advanced
Contact LumaSense. See
Laser Targeting Light
The instrument‘s internal
Use cooling jacket
are too high
fluctuations)
by intense magnetic fields and/or
improper selection of interconnecting
cable and/or cable run in same
conduit as AC power lines
and/or IGA 6/23 Advanced housing (or
cooling jacket)
secure
field of view. Conveyor type
applications will also cause
fluctuations as targets move by the
field of view
quickly. (Thermocouples do not
respond as fast as the
IGA 6/23 Advanced)
section 1.5
See section
rate and/or investigate other
cause.
Maximum to be stored. Apply
t
settings. See section 4.6
clear
slowed down
Apply t90 settings. See section
4.5
fails
objects that have fluctuating
temperatures
intense magnetic fields, improper
selection of inter-connecting cable
power lines
entering the field of view
dust, personnel or moving machinery
intermittently blocking 95% of
optical path of IGA 6/23 Advanced
moving targets
required
section 1.5
temperature is higher than 61 °C
Note: The wavelength band of the IGA 6/23 reacts at low measuring temperatures
(below 130 °C) to incandescent lamps or very bright daylight (not valid for fluorescent
tube). For a correct measurement strong external light to the measured object should
be avoided.
IGA 6/23 Advanced ManualTroubleshooting · 44
Index
1
1 measure… 30
A
Accesories 12
Air Purge 12
Analog Output 26
Aperture 20
Appropriate use 9
B
Basic settings 30
Baud Rate 27
C
Calibration 35
Laboratory 35
On-Site 35
Cleaning the front window 35
Clear Peak Memory 25
Clear Time Settings 26
Close 30
Connecting the pyrometer to a PC 16, 29
Connection cable 9
Connection of Additional Units 17
Connection to RS485 16
Cooling Jacket 12
D
Data format UPP 37
Device Address 27
Device Settings