6.2Installation of the indicating/adjustment module .
6.3Adjustment system .....................
6.4Setup procedure .......................
6.5Menu schematic .......................
7 Maintenance and fault rectification
7.1Maintenance ..........................
7.2Fault rectification .......................
7.3Exchanging the electronics................
7.4Instrument repair .......................
2OPTISOUND 3030 C - 4 … 20 mA/HART
18
19
20
22
23
25
25
27
28
32
35
35
36
37
30511-EN-050914
8 Dismounting
8.1Dismounting procedure ..................
8.2Disposal .............................
9 Supplement
9.1Technical data.........................
9.2Dimensions ...........................
9.3Certificates ...........................
Contents
38
38
39
43
45
Supplementary operating instructions manuals
Information:
OPTISOUND 3030 C is available in different versions and is
supplied specifically acc. to the order. Depending on the
selected version, supplementary operating instructions manuals come with the shipment. The supplementary operating
instructions are stated in paragraph "3. Product description".
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART3
About this document
1 About this document
1.1 Function
This operating instructions manual has all the information you
need for quick setup and safe operation of OPTISOUND 3030C. Please read this manual before you start setup.
1.2 Target group
This operating instructions manual is directed to trained
personnel. The contents of this manual should be made
available to these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution, warning, danger
This symbol informs you of a dangerous situation that could
occur. Ignoring this cautionary note can impair the person and/
or the instrument.
Ex applications
This symbol indicates special instructions for Ex applications.
lList
The dot set in front indicates a list with no implied sequence.
àAction
This arrow indicates a single action.
1Sequence
Numbers set in front indicate successive steps in a procedure.
4OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual
must be carried out only by trained, specialised personnel
authorised by the operator. For safety and warranty reasons,
any internal work on the instruments must be carried out only
by personnel authorised by the manufacturer.
2.2 Appropriate use
OPTISOUND 3030 C is a sensor for continuous level
measurement.
2.3 Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards, e.g. vessel overfill or damage to
system components through incorrect mounting or adjustment.
2.4 General safety instructions
OPTISOUND 3030 C is a high-tech instrument requiring the
strict observance of standard regulations and guidelines. The
user must take note of the safety instructions in this operating
instructions manual, the country-specific installation standards
(e.g. the VDE regulations in Germany) as well as all prevailing
safety regulations and accident prevention rules.
2.5 CE conformity
OPTISOUND 3030 C is in CE conformity with EMC (89/336/
EWG), fulfils the NAMUR recommendation NE 21 and is in CE
conformity with LVD (73/23/EWG).
Conformity has been judged acc. to the following standards:
l EMC:
- Emission EN 61326: 1997 (class A)
- Susceptibility EN 61326: 1997/A1: 1998
l LVD: EN 61010-1: 2001
2.6 Compatibility acc. to NAMUR NE 53
OPTISOUND 3030 C meets NAMUR recommendation NE 53.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART5
For your safety
The parameter adjustment of the basic sensor functions is
independent of the software version. The range of available
functions depends on the respective software version of the
individual components.
The software version of OPTISOUND 3030 C can be
determined as follows:
l on the type label of the electronics
l via the indicating/adjustment module.
You can find all software histories on our website www.krohne.
com.
2.7 Safety information for Ex areas
Please note the Ex-specific safety information for installation
and operation in Ex areas. These safety instructions are part of
the operating instructions manual and come with the Exapproved instruments.
6OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
3 Product description
3.1 Configuration
Product description
Scope of delivery
Components
The scope of delivery encompasses:
l OPTISOUND 3030 C ultrasonic sensor
l optionally available with mounting strap or compression
flange
l Documentation
- this operating instructions manual
- Operating instructions manual "Indicating and adjust-
ment module" - optional
OPTISOUND 3030 C consists of the following components:
l Process fitting with transducer
l Housing with electronics
l Housing cover, optionally available with indicating/adjust-
ment module
The components are available in different versions.
1
2
3
Fig. 1: OPTISOUND 3030 C with plastic housing
1Housing cover, optionally available with indicating/adjustment module
2Housing with electronics
3Process fitting with transducer
3.2 Principle of operation
Area of application
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART7
OPTISOUND 3030 C is an ultrasonic sensor for continuous
level measurement. It is suitable for liquids and solids in
practically all industries.
Product description
Physical principle
Power supply
Packaging
The transducer of the ultrasonic sensor emits short ultrasonic
pulses to the measured product. These pulses are reflected by
the product surface and received by the transducer as echoes.
The running time of the ultrasonic pulses from emission to
reception is proportional to the distance and hence to the level.
The determined level is converted into an appropriate output
signal and outputted as measured value.
Two-wire electronics 4 … 20 mA/HART for power supply and
measured value transmission on the same cable.
The power supply range can differ depending on the instrument version. The exact range is stated in the "Technical data"
in the "Supplement".
3.3 Adjustment
OPTISOUND 3030 C can be adjusted with two different
adjustment media:
l the indicating/adjustment module
l a HART handheld
The entered parameters are generally saved in OPTISOUND
3030 C, optionally also in the indicating/adjustment module.
3.4 Storage and transport
Your instrument was protected by packaging during transport.
Its capacity to handle normal loads during transport is assuredby a test acc. to DIN EN 24180.
The packaging of standard instruments consists of environ-
ment-friendly, recyclable cardboard. For special versions, PE
foam or PE foil is also used. Dispose of the packaging material
via specialised recycling companies.
Storage and transport temperature
8OPTISOUND 3030 C - 4 … 20 mA/HART
l Storage and transport temperature see "Supplement –
Technical data – Ambient conditions"
l Relative humidity 20 … 85 %
30511-EN-050914
4 Mounting
4.1 General instructions
Mounting
Installation position
Moisture
Select an installation position you can easily reach for
mounting and connecting as well as later retrofitting of an
indicating and adjustment module. The housing can be rotated
by 330° without the use of any tools. You can also install the
indicating and adjustment module in four different positions
(each displaced by 90°).
Use the recommended cable (see chapter "Connecting to
power supply") and thighten the cable gland.
You can give your OPTISOUND 3030 C additional protection
against moisture penetration by leading the connection cable
downward in front of the cable entry. Rain and condensation
water can thus drain off. This applies mainly to mounting
outdoors, in areas where moisture is expected (e.g. by
cleaning processes) or on cooled or heated vessels.
Fig. 2: Measures against moisture penetration
Measuring range
The reference plane for the measuring range is the lower edge
of the transducer.
Make sure that a min. distance - the so called dead zone below the reference plane is maintained in which a measurement is not possible. The exact value of the dead zone is
stated in the Technical data in the Supplement.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART9
Mounting
132
100%
0%
Fig. 3: Min. distance to the max. level
1Dead zone
2Reference plane
Information:
If the product reaches the transducer, buildup can form on it
over a period of time and later cause measurement errors.
132
100%
0%
Fig. 4: Measuring range (operating range) and max. measuring distance
1full
2empty (max. measuring distance)
3Measuring range
Pressure/Vacuum
Gauge pressure in the vessel does not influence OPTISOUND
3030 C. Low pressure or vacuum, however, damp the
ultrasonic pulses. This influences the measuring result,
10OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Mounting
particularly if the level is very low. With pressures under -
0.2 bar (-20 kPa) you should use a different measuring
principle, e.g. radar or guided radar (TDR).
4.2 Mounting information
Installation position
When mounting OPTISOUND 3030 C, keep a distance of at
least 200 mm (7.9 in) to the vessel wall. If the sensor is
installed in the center of dished or rounded vessel tops,
multiple echoes can arise. These can, however, be faded out
by an appropriate adjustment (see chapter "Setup").
If this distance cannot be maintained, a false echo storage
should be carried out during setup. This applies particularly if
buildup on the vessel wall is expected. In this case, we
recommend repeating the false echo storage later on with
existing buildup.
1
2
> 200 mm
Fig. 5: Mounting on dished vessel tops
1Reference plane
2Vessel center or symmetry axis
In vessels with conical bottom it can be advantageous to
mount the sensor in the center of the vessel, as measurement
is then possible down to the lowest point of the vessel bottom.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART11
Mounting
Fig. 6: Vessel with conical bottom
Socket
Preferably the transducer should be mounted flush to the
vessel top.
If the reflective properties of the medium are good, you can
mount OPTISOUND 3030 C on a socket piece. You will find
recommended values of the socket heights in the following
illustration. The socket end should be smooth and burr-free, if
possible also rounded. Carry out a false echo storage.
max.
h
45°
d
Fig. 7: Deviating socket dimensions
d
100 mm/4"
150 mm/6"
200 mm/8"
250 mm/10"
h
max.
300 mm
400 mm
500 mm
600 mm
30511-EN-050914
12OPTISOUND 3030 C - 4 … 20 mA/HART
Mounting
Sensor orientation
In liquids, direct the sensor as perpendicular as possible to the
product surface to achieve optimum measurement results.
Fig. 8: Orientation in liquids
To optimally align the sensor to solid products, a swivelling
holder (mounting strap) can be implemented.
Fig. 9: Orientation in solids
To reduce the min. distance to the medium, you can also
mount OPTISOUND 3030 C with a beam deflector. By doing
this, it is possible to fill the vessel nearly to maximum. This
arrangement is particularly suitable for open vessels such as e.
g. overflow basins.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART13
Mounting
Fig. 10: Beam deflector
~200
~400x400
°
5
4
Vessel installations
The ultrasonic sensor should be installed at a location where
no installations cross the ultrasonic beam.
Vessel installations such as for example, ladders, limit
switches, heating spirals, struts etc. can cause false echoes
superimposed on the useful echo. Make sure when planning
your measuring site that the ultrasonic signals have "free
access" to the measured product.
If there are existing vessel installations, a false echo storage
should be carried out during setup.
If large vessel installations such as struts or supports cause
false echoes, these can be attenuated through supplementary
measures. Small, inclined sheet metal or plastic baffles above
the installations scatter the ultrasonic signals and avoid direct
false echoes.
Fig. 11: Cover smooth profiles with deflectors
Agitators
If there are agitators in the vessel, a false echo storage should
be carried out with the agitators in motion. This ensures that
the interfering reflections from the agitators are saved with the
blades in different positions.
14OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Fig. 12: Agitators
Mounting
Inflowing material
Foam
Do not mount the instruments in or above the filling stream.
Make sure that you detect the product surface and not the
inflowing product.
Fig. 13: Inflowing liquid
Through the action of filling, stirring and other processes in the
vessel, dense foams which considerably damp the emitted
signals may form on the product surface.
If foams are causing measurement errors, the sensor should
be used in a standpipe or, alternatively, the more suitable
guided radar sensors (TDR) should be used.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART15
Loading...
+ 33 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.