6.2Installation of the indicating/adjustment module .
6.3Adjustment system .....................
6.4Setup procedure .......................
6.5Menu schematic .......................
7 Maintenance and fault rectification
7.1Maintenance ..........................
7.2Fault rectification .......................
7.3Exchanging the electronics................
7.4Instrument repair .......................
2OPTISOUND 3030 C - 4 … 20 mA/HART
18
19
20
22
23
25
25
27
28
32
35
35
36
37
30511-EN-050914
Page 3
8 Dismounting
8.1Dismounting procedure ..................
8.2Disposal .............................
9 Supplement
9.1Technical data.........................
9.2Dimensions ...........................
9.3Certificates ...........................
Contents
38
38
39
43
45
Supplementary operating instructions manuals
Information:
OPTISOUND 3030 C is available in different versions and is
supplied specifically acc. to the order. Depending on the
selected version, supplementary operating instructions manuals come with the shipment. The supplementary operating
instructions are stated in paragraph "3. Product description".
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART3
Page 4
About this document
1 About this document
1.1 Function
This operating instructions manual has all the information you
need for quick setup and safe operation of OPTISOUND 3030C. Please read this manual before you start setup.
1.2 Target group
This operating instructions manual is directed to trained
personnel. The contents of this manual should be made
available to these personnel and put into practice by them.
1.3 Symbolism used
Information, tip, note
This symbol indicates helpful additional information.
Caution, warning, danger
This symbol informs you of a dangerous situation that could
occur. Ignoring this cautionary note can impair the person and/
or the instrument.
Ex applications
This symbol indicates special instructions for Ex applications.
lList
The dot set in front indicates a list with no implied sequence.
àAction
This arrow indicates a single action.
1Sequence
Numbers set in front indicate successive steps in a procedure.
4OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 5
For your safety
2 For your safety
2.1 Authorised personnel
All operations described in this operating instructions manual
must be carried out only by trained, specialised personnel
authorised by the operator. For safety and warranty reasons,
any internal work on the instruments must be carried out only
by personnel authorised by the manufacturer.
2.2 Appropriate use
OPTISOUND 3030 C is a sensor for continuous level
measurement.
2.3 Warning about misuse
Inappropriate or incorrect use of the instrument can give rise to
application-specific hazards, e.g. vessel overfill or damage to
system components through incorrect mounting or adjustment.
2.4 General safety instructions
OPTISOUND 3030 C is a high-tech instrument requiring the
strict observance of standard regulations and guidelines. The
user must take note of the safety instructions in this operating
instructions manual, the country-specific installation standards
(e.g. the VDE regulations in Germany) as well as all prevailing
safety regulations and accident prevention rules.
2.5 CE conformity
OPTISOUND 3030 C is in CE conformity with EMC (89/336/
EWG), fulfils the NAMUR recommendation NE 21 and is in CE
conformity with LVD (73/23/EWG).
Conformity has been judged acc. to the following standards:
l EMC:
- Emission EN 61326: 1997 (class A)
- Susceptibility EN 61326: 1997/A1: 1998
l LVD: EN 61010-1: 2001
2.6 Compatibility acc. to NAMUR NE 53
OPTISOUND 3030 C meets NAMUR recommendation NE 53.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART5
Page 6
For your safety
The parameter adjustment of the basic sensor functions is
independent of the software version. The range of available
functions depends on the respective software version of the
individual components.
The software version of OPTISOUND 3030 C can be
determined as follows:
l on the type label of the electronics
l via the indicating/adjustment module.
You can find all software histories on our website www.krohne.
com.
2.7 Safety information for Ex areas
Please note the Ex-specific safety information for installation
and operation in Ex areas. These safety instructions are part of
the operating instructions manual and come with the Exapproved instruments.
6OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 7
3 Product description
3.1 Configuration
Product description
Scope of delivery
Components
The scope of delivery encompasses:
l OPTISOUND 3030 C ultrasonic sensor
l optionally available with mounting strap or compression
flange
l Documentation
- this operating instructions manual
- Operating instructions manual "Indicating and adjust-
ment module" - optional
OPTISOUND 3030 C consists of the following components:
l Process fitting with transducer
l Housing with electronics
l Housing cover, optionally available with indicating/adjust-
ment module
The components are available in different versions.
1
2
3
Fig. 1: OPTISOUND 3030 C with plastic housing
1Housing cover, optionally available with indicating/adjustment module
2Housing with electronics
3Process fitting with transducer
3.2 Principle of operation
Area of application
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART7
OPTISOUND 3030 C is an ultrasonic sensor for continuous
level measurement. It is suitable for liquids and solids in
practically all industries.
Page 8
Product description
Physical principle
Power supply
Packaging
The transducer of the ultrasonic sensor emits short ultrasonic
pulses to the measured product. These pulses are reflected by
the product surface and received by the transducer as echoes.
The running time of the ultrasonic pulses from emission to
reception is proportional to the distance and hence to the level.
The determined level is converted into an appropriate output
signal and outputted as measured value.
Two-wire electronics 4 … 20 mA/HART for power supply and
measured value transmission on the same cable.
The power supply range can differ depending on the instrument version. The exact range is stated in the "Technical data"
in the "Supplement".
3.3 Adjustment
OPTISOUND 3030 C can be adjusted with two different
adjustment media:
l the indicating/adjustment module
l a HART handheld
The entered parameters are generally saved in OPTISOUND
3030 C, optionally also in the indicating/adjustment module.
3.4 Storage and transport
Your instrument was protected by packaging during transport.
Its capacity to handle normal loads during transport is assuredby a test acc. to DIN EN 24180.
The packaging of standard instruments consists of environ-
ment-friendly, recyclable cardboard. For special versions, PE
foam or PE foil is also used. Dispose of the packaging material
via specialised recycling companies.
Storage and transport temperature
8OPTISOUND 3030 C - 4 … 20 mA/HART
l Storage and transport temperature see "Supplement –
Technical data – Ambient conditions"
l Relative humidity 20 … 85 %
30511-EN-050914
Page 9
4 Mounting
4.1 General instructions
Mounting
Installation position
Moisture
Select an installation position you can easily reach for
mounting and connecting as well as later retrofitting of an
indicating and adjustment module. The housing can be rotated
by 330° without the use of any tools. You can also install the
indicating and adjustment module in four different positions
(each displaced by 90°).
Use the recommended cable (see chapter "Connecting to
power supply") and thighten the cable gland.
You can give your OPTISOUND 3030 C additional protection
against moisture penetration by leading the connection cable
downward in front of the cable entry. Rain and condensation
water can thus drain off. This applies mainly to mounting
outdoors, in areas where moisture is expected (e.g. by
cleaning processes) or on cooled or heated vessels.
Fig. 2: Measures against moisture penetration
Measuring range
The reference plane for the measuring range is the lower edge
of the transducer.
Make sure that a min. distance - the so called dead zone below the reference plane is maintained in which a measurement is not possible. The exact value of the dead zone is
stated in the Technical data in the Supplement.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART9
Page 10
Mounting
132
100%
0%
Fig. 3: Min. distance to the max. level
1Dead zone
2Reference plane
Information:
If the product reaches the transducer, buildup can form on it
over a period of time and later cause measurement errors.
132
100%
0%
Fig. 4: Measuring range (operating range) and max. measuring distance
1full
2empty (max. measuring distance)
3Measuring range
Pressure/Vacuum
Gauge pressure in the vessel does not influence OPTISOUND
3030 C. Low pressure or vacuum, however, damp the
ultrasonic pulses. This influences the measuring result,
10OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 11
Mounting
particularly if the level is very low. With pressures under -
0.2 bar (-20 kPa) you should use a different measuring
principle, e.g. radar or guided radar (TDR).
4.2 Mounting information
Installation position
When mounting OPTISOUND 3030 C, keep a distance of at
least 200 mm (7.9 in) to the vessel wall. If the sensor is
installed in the center of dished or rounded vessel tops,
multiple echoes can arise. These can, however, be faded out
by an appropriate adjustment (see chapter "Setup").
If this distance cannot be maintained, a false echo storage
should be carried out during setup. This applies particularly if
buildup on the vessel wall is expected. In this case, we
recommend repeating the false echo storage later on with
existing buildup.
1
2
> 200 mm
Fig. 5: Mounting on dished vessel tops
1Reference plane
2Vessel center or symmetry axis
In vessels with conical bottom it can be advantageous to
mount the sensor in the center of the vessel, as measurement
is then possible down to the lowest point of the vessel bottom.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART11
Page 12
Mounting
Fig. 6: Vessel with conical bottom
Socket
Preferably the transducer should be mounted flush to the
vessel top.
If the reflective properties of the medium are good, you can
mount OPTISOUND 3030 C on a socket piece. You will find
recommended values of the socket heights in the following
illustration. The socket end should be smooth and burr-free, if
possible also rounded. Carry out a false echo storage.
max.
h
45°
d
Fig. 7: Deviating socket dimensions
d
100 mm/4"
150 mm/6"
200 mm/8"
250 mm/10"
h
max.
300 mm
400 mm
500 mm
600 mm
30511-EN-050914
12OPTISOUND 3030 C - 4 … 20 mA/HART
Page 13
Mounting
Sensor orientation
In liquids, direct the sensor as perpendicular as possible to the
product surface to achieve optimum measurement results.
Fig. 8: Orientation in liquids
To optimally align the sensor to solid products, a swivelling
holder (mounting strap) can be implemented.
Fig. 9: Orientation in solids
To reduce the min. distance to the medium, you can also
mount OPTISOUND 3030 C with a beam deflector. By doing
this, it is possible to fill the vessel nearly to maximum. This
arrangement is particularly suitable for open vessels such as e.
g. overflow basins.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART13
Page 14
Mounting
Fig. 10: Beam deflector
~200
~400x400
°
5
4
Vessel installations
The ultrasonic sensor should be installed at a location where
no installations cross the ultrasonic beam.
Vessel installations such as for example, ladders, limit
switches, heating spirals, struts etc. can cause false echoes
superimposed on the useful echo. Make sure when planning
your measuring site that the ultrasonic signals have "free
access" to the measured product.
If there are existing vessel installations, a false echo storage
should be carried out during setup.
If large vessel installations such as struts or supports cause
false echoes, these can be attenuated through supplementary
measures. Small, inclined sheet metal or plastic baffles above
the installations scatter the ultrasonic signals and avoid direct
false echoes.
Fig. 11: Cover smooth profiles with deflectors
Agitators
If there are agitators in the vessel, a false echo storage should
be carried out with the agitators in motion. This ensures that
the interfering reflections from the agitators are saved with the
blades in different positions.
14OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 15
Fig. 12: Agitators
Mounting
Inflowing material
Foam
Do not mount the instruments in or above the filling stream.
Make sure that you detect the product surface and not the
inflowing product.
Fig. 13: Inflowing liquid
Through the action of filling, stirring and other processes in the
vessel, dense foams which considerably damp the emitted
signals may form on the product surface.
If foams are causing measurement errors, the sensor should
be used in a standpipe or, alternatively, the more suitable
guided radar sensors (TDR) should be used.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART15
Page 16
Mounting
Guided radar is not influenced by foam generation and is
particularly suitable for such applications.
Air flow
Standpipe measurement
If there are strong air currents in the vessel, e.g. due to strong
winds in outdoor installations, or because of air turbulence, e.
g. through cyclone exhausting, you should mount OPTI-SOUND 3030 C in a standpipe or use a different measuring
principle, e.g. radar or guided radar (TDR).
By using a standpipe (surge or bypass tube), the influence of
vessel installations, foam generation and turbulence is
excluded.
Standpipes must extend all the way down to the requested
min. level, as measurement is only possible within the tube.
max.
1
min.
Fig. 14: Standpipe in a tank
1Vent hole ø 5 … 10 mm
OPTISOUND 3030 C can be used from tube diameters of
100 mm.
Avoid large gaps and thick welding joints when connecting the
tubes. Generally carry out a false echo storage.
Measurement in a standpipe is not recommended for very
adhesive products.
16OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 17
Connecting to power supply
5 Connecting to power supply
5.1 Preparing the connection
Note safety instructions
Take note of safety
instructions for Ex
applications
Select power supply
Selecting the connection ca-
ble
Always observe the following safety instructions:
l Connect only in the complete absence of line voltage
l If overvoltages are expected, overvoltage arresters should
be installed.
In hazardous areas you should take note of the appropriate
regulations, conformity and type approval certificates of the
sensors and power supply units.
Power supply and current signal are transmitted via the same
two-wire connection cable. The power supply range can differ
depending on the instrument version. The exact range is
stated in the "Technical data" in the Supplement.
Provide a reliable separation between the supply circuit and
the mains circuits acc. to DIN VDE 0106 part 101.
Bear in mind the following factors regarding supply voltage:
l the reduction of the output voltage of the power supply unit
under nominal load (with a sensor current of 20.5 mA or
22 mA in case of fault signal)
l the influence of additional instruments in the circuit (see
load values in Technical data).
OPTISOUND 3030 C is connected with standard two-wire
cable without screen. An outer cable diameter of 5 ... 9 mm
ensures the seal effect of the cable entry. If electromagnetic
interference is expected, we recommend the use of screened
cable.
Cable screening and grounding
Connect the cable screen on both ends to ground potential. In
the sensor, the screen must be connected directly to the
internal ground terminal. The ground terminal outside on the
housing must be connected to the potential equalisation (low
impedance).
If potential equalisation currents are expected, the connection
on the evaluation side must be made via a ceramic capacitor
(e.g. 1 nF, 1500 V). The low frequency potential equalisation
currents are thus suppressed, but the protective effect against
high frequency interference signals remains.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART17
Page 18
Connecting to power supply
Select connection
cable for Ex applica-
tions
Take note of the corresponding installation regulations for Ex
applications. In particular, make sure that no potential equalisation currents flow over the cable screen. In case of
grounding on both sides this can be achieved by the use of a
capacitor or a separate potential equalisation.
5.2 Connection procedure
Proceed as follows:
1 Unscrew the housing cover
2 If an indicating and adjustment module is installed, remove
it by turning it slightly to the left.
3 Loosen compression nut of the cable entry
4 Remove approx. 10 cm (4 in) of the cable mantle, strip
approx. 1 cm (0.4 in) insulation from the ends of the
individual wires
5 Insert the cable into the sensor through the cable entry
6 Lift the opening levers of the terminals with a screwdriver
(see following illustration)
7 Insert the wire ends into the open terminals according to
the wiring plan
8 Press down the opening levers of the terminals, you will
hear the terminal spring closing
9 Check the hold of the wires in the terminals by lightly
pulling on them
10 Connect the screen to the internal ground terminal and the
external ground terminal to potential equalisation
11 Tighten the compression nut of the cable entry, the seal
ring must completely encircle the cable
12 Screw the housing cover back on
The electrical connection is finished.
18OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 19
Housing overview
Connecting to power supply
Fig. 15: Connection steps 6 and 7
5.3 Wiring plans, single chamber housing
44
1
Fig. 16: Material versions, single chamber housing
1Plastic
2Aluminium
3Stainless steel
4Filter element for pressure compensation
23
4
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART19
Page 20
Connecting to power supply
Electronics and connection
compartment
Display
I2C
Wiring plan
1
125678
2
Fig. 17: Electronics and connection compartment, single chamber housing
1Spring-loaded terminals for power supply
2Ground terminal for connection of the cable screen
12 5
Display
678
I2C
1
Fig. 18: Wiring plan, single chamber housing
1Power supply/Signal output
After OPTISOUND 3030 C is connected to power supply, the
instrument carries out a self-test for approx. 30 seconds. The
following steps are carried out:
l Internal check of the electronics
l Indication of the instrument type, the firmware version as
well as the sensor TAGs (sensor name)
l Output signal jumps briefly (approx. 10 seconds) to the set
fault current
22OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 23
Connecting to power supply
Then the actual measured value is displayed and the
corresponding current is transmitted to the cable.
1)
1)
The values correspond to the actual level as well as to the settings already
carried out, e.g. default setting.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART23
Page 24
Setup with the indicating/adjustment module
6 Setup with the indicating/adjustment
module
6.1 Short description
The indicating/adjustment module is used for measured value
display, adjustment and diagnosis. It can be mounted in all
sensors of the OPTISOUND instrument family, in the single as
well as double chamber housing (optionally in the electronics
or connection compartment).
6.2 Installation of the indicating/adjustment
module
The indicating/adjustment module can be inserted or removed
at any time. It is not necessary to interrupt the power supply.
To mount, proceed as follows:
1 Unscrew the housing cover
2 Place the indicating/adjustment module in the desired
position on the electronics (you can choose any one of four
different positions - each displaced by 90°)
3 Press the indicating and adjustment module onto the
electronics and turn it to the right until it snaps in.
4 Screw housing cover with inspection window tightly back
on
Removal is carried out in reverse order.
The indicating/adjustment module is powered by the sensor,
an additional connection is not necessary.
24OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 25
Setup with the indicating/adjustment module
Fig. 23: Installation of the indicating/adjustment module
Note:
If you intend to retrofit OPTISOUND 3030 C with an indicating/
adjustment module for continuous measured value indication,
a higher cover with an inspection glass is required.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART25
Page 26
Setup with the indicating/adjustment module
6.3 Adjustment system
Key functions
Adjustment system
1
1.1
2
3
Fig. 24: Indicating and adjustment elements
1LC display
2Indication of the menu item number
3Adjustment keys
l [OK] key:
- move to the menu overview
- confirm selected menu
- edit parameter
- save value
l [–>] key to select:
- menu change
- list entry
- editing position
l [+] key:
- modify value of a parameter
l [ESC] key:
- interrupt input
- jump to the next higher menu
The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual
menu items. The functions of the individual keys are shown in
the above illustration. Approx. 10 minutes after the last
pressing of a key, an automatic reset to measured value
indication is triggered. Any values not confirmed with [OK] will
not be saved.
30511-EN-050914
26OPTISOUND 3030 C - 4 … 20 mA/HART
Page 27
Setup with the indicating/adjustment module
6.4 Setup procedure
Address setting HART-Multi-
drop
Parameter adjustment
In HART-Multidrop mode (several sensors on one input) the
address must be set before continuing with the parameter
adjustment. You will find a detailed description in the operating
instructions manual of the indicating/adjustment module.
HART mode
Standard
Address 0
Because OPTISOUND 3030 C is a distance measuring
instrument, its primary task is to measure the distance from the
sensor to the product surface. In order to indicate the actual
filling level, the measured distance must be allocated to a
specific height percentage. To make this adjustment, the full
and empty distances in the vessel are entered. If these values
are not known, it is also possible to carry out the adjustment
with other distances, e.g. 10 % and 90 %. The origin of these
distance values is always the lower edge of the flange (with
flange versions) or the lower edge of the transducer (all other
versions).
The actual level is then calculated on the basis of these
entered values. At the same time, the operating range of the
sensor is limited from maximum range to the requested range.
The real product level during this adjustment is not important,
because the min./max. adjustment is always carried out
without changing the product level. These settings can be
made ahead of time without the instrument having to be
installed.
In the main menu item "Basic adjustment", the individual
submenu items should be selected one after the other and
provided with the correct parameter values.
Start your parameter adjustment with the following menu items
of the basic adjustment:
Carrying out min. adjustment
Proceed as follows:
1
Move from the measured value display to the main menu
by pushing [OK].
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART27
Page 28
Setup with the indicating/adjustment module
▶
Basic adjustment
Display
Diagnostics
Service
Info
2 Select the menu item "Basic adjustment" with [–>] and
confirm with [OK]. Now the menu item "Min. adjustment" is
displayed.
Min. adjustment
0.00 %
=
5.000 m(d)
3 Prepare the % value for editing with [OK] and set the
cursor to the requested position with [–>]. Set the
requested percentage value with [+] and save with [OK].The cursor jumps now to the distance value.
4 Enter the appropriate distance value in m (corresponding
to the percentage value) for the empty vessel (e.g. distance
from the sensor to the vessel bottom).
5 Save the settings with [OK] and move to "Max. adjustment"
with [–>].
4.000 m(d)
Carrying out max. adjustment
Proceed as follows:
Max. adjustment
100.00 %
=
1.000 m(d)
2.000 m(d)
1 Prepare the % value for editing with [OK] and set the
cursor to the requested position with [–>]. Set the
requested percentage value with [+] and save with [OK].The cursor jumps now to the distance value.
2 Enter the appropriate distance value in m (corresponding
to the percentage value) for the full vessel. Keep in mind
that the max. level must lie below the dead band.
3 Save the settings with [OK] and move to "Medium
selection" with [–>].
Medium selection
Each product has different reflective properties. In addition,
there are various interfering factors which have to be taken into
account: agitated product surfaces and foam generation (with
liquids); dust generation, material cones and echoes from the
28OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 29
Setup with the indicating/adjustment module
vessel wall (with solids). To adapt the sensor to these different
conditions, you should first select in this menu item "Liquid" or
"Solid".
Medium
Liquid
With solids, you can also choose between "Powder/Dust",
"Granular/Pellets" or "Ballast/Pebbels".
Through this additional selection, the sensor is adapted
perfectly to the product and measurement reliability, particularly in products with bad reflective properties, is considerably
increased.
Enter the requested parameter via the appropriate keys, save
your settings and jump to the next menu item with the [–>] key.
Vessel shape
Damping
Apart from the medium also the vessel form can influence the
measurement. To adapt the sensor to these conditions, this
menu item offers (depending on either liquid or solid is
detected) different options. For Liquid these are Storage tank,
Stilling tube, Open vessel or Stirred vessel, for Solid Silo or
Bunker.
Vessel shape
Storage tank
Enter the requested parameter via the appropriate keys, save
your settings and jump to the next menu item with the [–>] key.
To suppress fluctuation in the measured value display, e.g.
caused by a turbulent product surface, an integration time can
be set. This time can be between 0 and 999 seconds. Please
note that the reaction time of the entire measurement will be
longer and the sensor will react to quick changes of the
measured value with a corresponding delay. In general, a time
of a few seconds is sufficient to smooth the measured value
display.
Damping
0 s
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART29
Page 30
Setup with the indicating/adjustment module
Enter the requested parameter via the appropriate keys, save
your settings and jump to the next menu item with the [–>] key.
Linearization curve
Sensor-TAG
A linearization is necessary for all vessels in which the vessel
volume does not increase linearly with the level – e.g. in a
cylindrical or spherical tank -–and the indication or output of
the volume is requested. Corresponding linearization curves
are preprogrammed for these vessels. They represent the
correlation between the level percentage and vessel volume.
By activating the appropriate curve, the volume percentage of
the vessel is displayed correctly. If the volume should not be
displayed in percent but e.g. in l or kg, a scaling can be also
set in the menu item "Display".
Linearization curve
Linear
Enter the requested parameter via the appropriate keys, save
your settings and jump to the next menu item with the [–>] key.
In this menu item you can enter an unambiguous designation
for the sensor, e.g. the measurement loop name or the tank or
product designation. In digital systems and in the documentation of larger plants, a singular designation should be entered
for exact identification of individual measuring sites.
Sensor-TAG
Sensor
With this menu item, the Basic adjustment is finished and you
can now jump to the main menu with the [ESC] key.
False signal suppression
High sockets or vessel installations, such as e.g. struts or
agitators as well as buildup and weld joints on the vessel walls
cause interfering reflections which can impair the measurement. A false signal suppression detects, marks and saves
these false signals, so that they are no longer taken into
account for the level measurement. This false signal suppression should be created with empty vessel so that all
potential interfering reflections will be detected.
30OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 31
Setup with the indicating/adjustment module
False signal suppression
Change now?
Proceed as follows:
1 Move from the measured value display to the main menu
by pushing [OK].
2 Select the menu item "Service" with [–>] and confirm with
[OK]. Now the menu item "False signal suppression" is
displayed.
3 Confirm "False signal suppression - Change now" with
[OK] and select in the below menu "Create new". Enter the
actual distance from the sensor to the product surface. All
false signals in this area are detected by the sensor and
saved after confirming with [OK].
Note:
Check the distance to the product surface, because if an
incorrect (too large) value is entered, the existing level will be
saved as false signal. The filling level would then no longer be
detectable in this area.
Optional settings
Additional adjustment and diagnosis options such as e.g.
scaling, simulation or trend curve presentation are shown in
the following menu schematic. You will find a detailed
description of these menu items in the operating instructions
manual of the indicating and adjustment module.
Output mode 4-20 mA
Failure mode: <3.6mA
Min-current 3.8mA
Language
Deutsch
Date of last change using PC
26. July 2005
4.3
4.7
5.3
Simulation
Start simulation?
HART mode
Standard
Address 0
Sensor details
Display now?
4.4
4.8
5.4
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART33
Page 34
Maintenance and fault rectification
7 Maintenance and fault rectification
7.1 Maintenance
When used as directed in normal operation, OPTISOUND
3030 C is completely maintenance-free.
7.2 Fault rectification
Causes of malfunction
Fault rectification
Checking the 4 … 20 mA sig-
nal
OPTISOUND 3030 C offers maximum reliability. Nevertheless
faults can occur during operation. These may be caused by the
following, e.g.:
l Sensor
l Process
l Power supply
l Signal processing.
The first measures to be taken are to check the output signal
and evaluate fault messages via the indicating/adjustment
module. The procedure is described below.
Connect a hand-held multimeter with a suitable measuring
range acc. to the wiring plan.
? 4 … 20 mA signal not stable
l level fluctuations
à set integration time via the indicating/adjustment
module
? 4 … 20 mA signal missing
l incorrect connection to power supply
à Check connection acc. to chapter "Connection proce-
dure" and correct, if necessary, acc. to chapter "Wiring
plans"
l no power supply
à check cables for line break, repair, if necessary
l supply voltage too low or load resistance too high
à check and adapt, if necessary
34OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 35
Maintenance and fault rectification
? Current signal greater than 22 mA or less than 3.6 mA
l Electronics module defective
à Exchange instrument or return it for repair
In Ex applications, the regulations for the wiring of intrinsically
safe circuits must be observed.
Fault messages via the indicating/adjustment module
? E013
l no measured value available
à sensor in boot phase
à sensor does not find an echo, e.g. because of faulty
installation or incorrect parameter adjustment
? E017
l adjustment span too small
à Carry out a fresh adjustment and increase the distance
between min. and max. adjustment
? E036
l no operable sensor software
à carry out a software update or return instrument for
repair
? E041
l hardware error, electronics defective
à Exchange instrument or return it for repair
7.3 Exchanging the electronics
If the electronics module is defective, it can be replaced by the
user.
In Ex applications, only instruments and electronics modules
with appropriate Ex approval may be used.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART35
Page 36
Maintenance and fault rectification
If there is no electronics module available on site, it can beordered from the responsible Krohne agency.
7.4 Instrument repair
If a repair is necessary, please proceed as follows:
You can download a return form from our Internet homepagehttp://www.krohne-mar.com/fileadmin/media-lounge/PDF-
Download/Specimen_e.pdf.
By doing this you help us carry out the repair quickly and
without having to call back for needed information.
l Print and fill out one form per instrument
l Clean the instrument and pack it damage-proof
l Attach the completed form and possibly also a safety data
sheet to the instrument.
36OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 37
Dismounting
8 Dismounting
8.1 Dismounting procedure
Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures,
corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power
supply" and carry out the listed steps in reverse order.
8.2 Disposal
OPTISOUND 3030 C consists of materials which can be
recycled by specialised recycling companies. We have
purposely designed the electronic modules to be easily
separable. Mark the instrument as scrap and dispose of it
according to national government regulations (e.g. in G ermany
acc. to electronic scrap ordinance).
Materials: see "Technical data"
If you cannot dispose of the instrument properly, please
Fig. 26: Voltage diagram
1HART load
2meaningless
3Voltage limit
4Supply voltage
Electrical protective measures
Protection
- Plastic housingIP 66/IP 67
- Alu and stainless steel housingIP 66/IP 68 (0.2 bar)
5)
Overvoltage categoryIII
Protection classII
5)
Requirement to maintain the protection is the suitable cable.
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART41
Page 42
Supplement
9.2 Dimensions
Housing
M20x1,5/
½ NPT
~ 116mm
9
(4
/16")
ø 84mm
(3
5
/16")
M20x1,5
")
64
/
31
114mm
(4
~ 69mm
23
(2
M20x1,5/
½ NPT
~ 69mm
23
(2
/32")
/32")
ø 77mm
1
(3
/32")
")
32
/
13
112mm
(4
M20x1,5/
½ NPT
ø 77mm
1
(3
/32")
1234
")
64
/
39
117mm
(4
~ 87mm (3 27/64")
M20x1,5/
½ NPT
ø 84mm
5
(3
/16")
")
32
/
23
120mm (4
Fig. 27: Housing versions in protection IP 66/IP 67 and IP 66/IP 68, 0.2 bar (with integrated indicating and adjustment
module the housing height or width increases by 9 mm/0.35 in)
Fig. 28: OPTISOUND 3030 C
1Mounting strap
2Compression flange
3Dead zone: 0.6 m (2 ft)
4Meas. range: in liquids up to 15 m (49.2 ft), in solids up to 7 m (23 ft)
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART43
Page 44
Supplement
9.3 Certificates
CE declaration of conformity
Fig. 29: CE declaration of conformity
44OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 45
Manufacturer declaration
Supplement
Fig. 30: Manufacturer declaration
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART45
Page 46
Supplement
46OPTISOUND 3030 C - 4 … 20 mA/HART
30511-EN-050914
Page 47
Supplement
30511-EN-050914
OPTISOUND 3030 C - 4 … 20 mA/HART47
Page 48
Subject to change without notice
30511-EN-050914
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.