Integra Camino User manual

0 (0)
Integra Camino User manual

Integra®

USER’S MANUAL

Camino® ICP Monitor

Intracranial Pressure and Temperature Monitor

Caution

Federal (USA) law restricts this device to sale by or on the order of a physician.

 

Publication Name

 

Integra Camino ICP Monitor User’s Manual

 

Integra Part Number 60903733 Rev. B

 

Effective Date September 2012

 

Trademark Acknowledgements

 

Integra, the Integra logo, and Camino are registered trademarks of Integra

 

LifeSciences Corporation or its subsidiaries in the United States and/or

 

other countries. Delta-Cal is a trademark of Utah Medical Products, Inc.

 

Sani-Cloth is a registered trademark of Professional Disposables

 

International, Inc. Aptimax, Sealsure, and Sterrad are registered trademarks

 

of Johnson and Johnson. Tyvec is a registered trademark of E.I. duPont.

 

Kimguard and One-Step are registered trademarks of Kimberly-Clark

 

Worldwide, Inc.

0086

Copyright Information

©2012 Integra LifeSciences Corporation. All rights reserved.

No part of this document may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means—electronic, mechanical, photocopying, recording, or otherwise—without the expressed, written consent of Integra LifeSciences Corporation. Additional copies of this document can be ordered from Integra LifeSciences Corporation.

Manufacturer:

Integra LifeSciences(Ireland) Limited

IDA Business and Technology Park

Sragh, Tullamore, County Offaly, Ireland

Distributed by:

Integra LifeSciences Corporation 311 Enterprise Drive

Plainsboro, NJ 08536 USA

Tel: 1-800-654-2873 (USA only) 1-(609) 275-0500

Fax: 1-609-275-5363

Internet Address: http://www.integralife.com

Made in Ireland. Printed in Ireland.

This page is intentionally left blank.

TABLE OF CONTENTS

 

List of Symbols and Abbreviations

 

Packaging and Label Symbols................................................................................................................

v

Software Symbols ..................................................................................................................................

vi

List of Abbreviations..............................................................................................................................

vii

Chapter 1: System Overview

 

Indications for Use / Intended Use ..........................................................................................................

1

Contraindications ....................................................................................................................................

1

Intended User..........................................................................................................................................

1

Intended Patient Population ....................................................................................................................

1

Description of the Integra Camino ICP Monitor.......................................................................................

2

Key Functions of Monitor................................................................................................................

2

Reviewing the User’s Manual..................................................................................................................

2

List of Warnings for Using the Monitor ....................................................................................................

3

Parts of the Monitor.................................................................................................................................

6

About the Front Panel.....................................................................................................................

6

About the Rear Panel .....................................................................................................................

7

About the Right Panel.....................................................................................................................

8

About the Bottom Panel..................................................................................................................

9

About the Left Panel .......................................................................................................................

9

Chapter 2: Setting Up System for the First Time

 

Procedures for Initial Setup...................................................................................................................

11

Chapter 3: Setting Up System for Clinical Use

 

Setting Up System for Clinical Use .......................................................................................................

19

Positioning the Monitor .................................................................................................................

19

Attaching to Equipment Pole (if applicable)..................................................................................

19

Powering the System On and Off .................................................................................................

20

Using the Battery for Power..........................................................................................................

21

Storing the Battery........................................................................................................................

22

About the Integra Catheters ..................................................................................................................

23

Connecting the Integra® Camino® Fiber Optic Catheters (110-4 Series) ..............................................

24

Connecting the Integra® Camino® Flex Catheters .................................................................................

27

Connecting to a Patient Bedside Monitor (if applicable) ......................................................................

30

i

Procedures for Synchronizing the Two Monitors .........................................................................

30

About Pressure and Temperature Measurements on the Patient Bedside Monitor.....................

33

Storing the System ...............................................................................................................................

33

Chapter 4: Monitoring the Patient’s ICP and Temperature

 

About the Touch Screen.......................................................................................................................

35

About the Synchronize to Monitor Button ....................................................................................

36

Reviewing the Status Bar.............................................................................................................

36

Verifying Status of Battery and AC Power ..................................................................................

36

Verifying Amount of Battery Charge Available.............................................................................

37

About the Alarms ..................................................................................................................................

37

Monitoring the Patient’s ICP and Temperature ....................................................................................

37

Monitoring Trend Data..........................................................................................................................

39

Conditions that Reset Trend Data................................................................................................

40

Setting the High ICP Alarm Limit ..........................................................................................................

41

Specifying the High ICP Alarm Limit ............................................................................................

42

Disabling the High ICP Alarm ......................................................................................................

43

Silencing the High ICP Alarm Temporarily...................................................................................

43

Restoring Default High ICP Alarm Limit Values...........................................................................

43

Customizing the User Settings .............................................................................................................

44

Chapter 5: Responding to Physiological and Technical Alarms

 

About the Two Alarm Types .................................................................................................................

47

About the Technical Messages ............................................................................................................

47

Understanding the Alarm Symbols .......................................................................................................

48

How the Monitor Prioritizes the Alarms ...............................................................................................

48

Audio and Visual Indicators for Medium and Low Priority Alarms ...............................................

49

Priorities of Physiological and Technical Alarms .........................................................................

49

List of Priorities for Each Alarm....................................................................................................

50

Responding to the Physiological Alarm (ICP above alarm limit) ..........................................................

51

Responding to Technical Alarms..........................................................................................................

51

Responding to Irreversible System Failure Alarms......................................................................

51

Responding to ICP Catheter Failure Alarm..................................................................................

52

Responding to Temperature Sensor Failure Alarm .....................................................................

52

Responding to Low Battery Alarm ...............................................................................................

52

Responding to Monitor Overheating Alarm..................................................................................

53

Responding to Cooling Fan Failure Alarm ...................................................................................

53

Responding to Accuracy Range Alarm ........................................................................................

54

Responding to Battery Failure Alarm ...........................................................................................

55

ii

Chapter 6: Extracting Trend Data for Remote Evaluation

 

About Data Extraction ...........................................................................................................................

57

Extracting Data to USB Drive ...............................................................................................................

57

How the Monitor Stores Trend Data for Up to 5 Days ..................................................................

58

Extract Data via Digital Streaming ........................................................................................................

59

Conditions That Reset Trend Data During Recording...........................................................................

59

Chapter 7: Cleaning and Sterilizing the System

 

Cleaning the System and Components.................................................................................................

61

Cleaning Guidelines .....................................................................................................................

61

Sterilizing the Integra® Camino® Fiber Optic Catheter Cable ................................................................

63

Sterilizing the Integra® Camino® Flex Extension Cable .........................................................................

65

Sterilization Parameters ...............................................................................................................

65

About Single-Use Only Catheters .........................................................................................................

66

Disposal of the Monitor System and Components ................................................................................

66

Chapter 8: Troubleshooting the System

 

About the Troubleshooting Process ......................................................................................................

67

Responding to System Status Messages .............................................................................................

67

Responding to Problems During Use....................................................................................................

69

Responding to System Failure Messages.............................................................................................

73

Chapter 9: Testing and Preventive Maintenance

 

About These Procedures ......................................................................................................................

75

Use Fiber Optic Catheters for Each Test......................................................................................

75

Testing Pressure Input ..........................................................................................................................

76

Using a Graduated Drainage Bag ................................................................................................

76

Using a Pressure Simulator (preferred method)...........................................................................

78

Testing Pressure Output .......................................................................................................................

79

Testing Temperature Input....................................................................................................................

80

Testing Temperature Output .................................................................................................................

81

Testing High ICP Alarm Limit ................................................................................................................

84

Testing AC Power and Battery Charge .................................................................................................

85

AC Power .....................................................................................................................................

85

Low Battery Alarm ........................................................................................................................

85

Battery Charge .............................................................................................................................

86

Inserting A New Battery ........................................................................................................................

86

Determining Software Version ..............................................................................................................

87

iii

Testing Synchronizing to Patient Bedside Monitor ...............................................................................

87

Chapter 10: Contacting Integra for Technical Support and Annual Maintenance

 

About Technical Support ......................................................................................................................

89

About Annual Maintenance ..................................................................................................................

89

Appendix A: Technical Specifications

 

List of Technical Specifications ............................................................................................................

91

Classifications and Standards .............................................................................................................

93

Manufacturer’s Declaration Table.........................................................................................................

94

General Notes..............................................................................................................................

94

Appendix B: Integra Warranty

 

Warranty ..............................................................................................................................................

99

Index ..............................................................................................................................

103

iv

LIST OF SYMBOLS AND ABBREVIATIONS

Packaging and Label Symbols

Symbol

Definition

 

Symbol

Definition

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Follow instructions for use

 

 

 

 

 

 

 

 

 

 

 

 

Defibrillation-proof type CF applied part

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Catalogue number

 

 

 

 

 

 

 

 

 

 

 

 

Direct current

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Caution: Federal (USA) law restricts this

 

 

 

 

 

 

 

 

 

 

 

 

Class II equipment

 

 

 

device to sale by or on the order of a

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

physician.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Serial number

 

 

 

 

 

 

 

 

 

 

 

 

Waste Electronics and Electrical

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Equipment

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Batch code

 

 

 

 

 

 

 

 

 

 

 

 

USB connection

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Manufacturer

 

 

 

 

 

 

 

 

 

 

 

 

Safety information for transporting

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

lithium ion batteries.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Date of manufacture

 

 

 

CALIBRATION

Due date for annual maintenance

 

 

 

 

 

 

 

 

VERIFIED.

 

 

 

 

 

 

 

 

NEXT SCHEDULED

 

 

 

 

 

 

 

MAINTENANCE DUE

 

 

 

 

 

 

 

 

 

 

 

72903759 Rev 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+50°C

Temperature limitation

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-20°Cl

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

List of Symbols and Abbreviations • v

Software Symbols

Symbol

Description

 

Symbol

Definition

 

 

 

 

 

 

 

 

 

 

 

Active alarm

 

 

AC power not being used/available

 

 

 

 

 

 

Audio paused

 

 

Battery charge indicator

 

 

 

 

 

 

Inactive alarm

 

 

No battery connected or faulty battery

 

 

 

 

 

 

High ICP alarm limit

 

 

Battery being charged

 

 

 

 

 

 

AC power being used

 

 

System information panel

 

 

 

 

 

 

On/Off power

 

 

 

 

 

 

 

 

vi • List of Symbols and Abbreviations

List of Abbreviations

Abbreviation

Definition

 

 

 

 

AC

Alternating Current

 

 

ºC

Celsius

 

 

CAMCABL

Preamplification cable used with Integra fiber optic catheters

 

 

CSV

Comma-separated values

 

 

CT

Computer tomography

 

 

dB

Decibels

 

 

DC

Direct Current

 

 

DMM

Digital Multimeter

 

 

EtO

Ethylene oxide

 

 

ºF

Fahrenheit

FLEX

Flex Catheter

 

 

FLEXEXT

Flex Extension Cable

 

 

hPA

Hectopascal pressure unit

 

 

ICP

Intracranial pressure

 

 

ICT

Intracranial temperature

 

 

IPA

Isopropyl alcohol

 

 

LED

Light Emmitting Diode

 

 

mm

Millimeters

 

 

mmHg

Millimeters of mercury

 

 

MR

Magnetic resonance

 

 

OR

Operating Room

 

 

PMIO

Patient Monitor Input Output

 

 

TBI

Traumatic brain injury

 

 

USB

Universal Serial Bus

 

 

V

Volt

 

 

W

Watt

 

 

List of Symbols and Abbreviations • vii

This page is intentionally left blank.

viii • List of Symbols and Abbreviations

CHAPTER 1 SYSTEM OVERVIEW

Indications for Use / Intended Use.................................................................

1

Contraindications ............................................................................................

1

Intended User...................................................................................................

1

Intended Patient Population ...........................................................................

1

Description of the Integra Camino ICP Monitor............................................

2

Reviewing the User’s Manual .........................................................................

2

List of Warnings for Using the Monitor .........................................................

3

Parts of the Monitor.........................................................................................

6

Indications for Use / Intended Use

The Integra® Camino® ICP Monitor is indicated for use by qualified neurosurgeons or neurointensivists for measurement of intracranial pressure and temperature.

Contraindications

The Integra Camino ICP Monitor and its accessories are contraindicated for use in a Magnetic Resonance (MR) environment.

Intended User

The Integra Camino ICP Monitor is intended to be used by the following qualified medical and biomedical professionals:

A qualified neurosurgeon should perform the placement and handling of the catheters.

Designated qualified hospital staff (e.g. neurosurgeon, nurse, intensivist, trauma physician, or physician’s assistant) should perform the operation of the monitor.

Chapter 9 provides instructions for testing and maintaining the monitor. The procedures in this chapter are intended to be performed by the hospital’s biomedical engineering staff.

Intended Patient Population

Patients undergoing treatment with this monitor under the cranial applications are expected to have had a TBI, undergone a major neurosurgical procedure, or some other traumatic, ischemic or hemorrhagic incident requiring controlled monitoring of ICP and brain temperature.

Chapter 1 • System Overview • 1

Description of the Integra Camino ICP Monitor

The Integra Camino ICP Monitor is a compact, portable device that provides tools for continuously determining and monitoring intracranial pressure (ICP) and intracranial temperature (ICT) directly in the brain, depending on which catheters are connected to the system. This monitor supports the following catheters:

Series of Integra® Camino® Fiber Optic Catheters (110-4 series) for measuring both ICP and temperature.

Integra® Camino® Flex Catheter for measuring ICP values.

All Integra catheters measure their respective values at the tip of the catheter. This design eliminates the need for a fluid-filled system to communicate pressure (and carry pressure waves) to an external transducer.

Key Functions of Monitor

During clinical use, the Integra Camino ICP Monitor provides several key functions to facilitate the process for monitoring and analyzing patient data:

Touch screen interface for evaluating patient ICP/ICT data and setting patient parameters (see page 35)

Physiological alarm that activates if the patient’s Mean ICP value exceeds a user-specified limit for more than 5 seconds (see page 41)

Rechargeable lithium ion battery that supplies power to monitor during patient transport (see page 21)

Storage of patient’s ICP trend data for up to 5 days (see page 39)

Outputs for transferring patient data to a patient bedside monitor (see page 30)

Outputs for extracting patient data to remote media types via USB drive or digital streaming (see page 57)

For instructions on using the Integra catheters, see the directions for use supplied with each respective catheter.

Reviewing the User’s Manual

Integra recommends that all physicians, nurses, and technicians who will be using, operating, and maintaining the Integra Camino ICP Monitor review this user’s manual prior to using the system. If there are additional questions after reading this manual, contact Integra.

2 • Chapter 1 • System Overview

List of Warnings for Using the Monitor

Failure to observe one or more of the following warnings could compromise patient safety or result in measurement errors.

Warnings

Use of the Integra Camino ICP Monitor is restricted to one patient at a time.

The Integra Camino ICP Monitor and its accessories are contraindicated for use in a Magnetic Resonance (MR) environment.

Always verify the high ICP alarm limit is set appropriately for each patient prior to treatment.

Selecting the Alarm Off feature on the Alarm panel will disable the high ICP alarm

limit indefinitely. Use caution if this feature is selected. To re-enable this alarm, press the Alarm On and Accept buttons.

No modification of the Integra Camino ICP Monitor is allowed.

The Integra Camino ICP Monitor is a sensitive electronic device. When using the monitor, always handle with care. If damage is suspected, contact Integra.

Read the user’s manual from the patient bedside monitor’s manufacturer before connecting the Integra Camino ICP Monitor to a patient bedside monitor.

To prevent injury to the patient, user, or other persons, or damage to the monitor, always verify that the monitor is clamped securely to the equipment pole.

To reduce the risk of electric shock, do not disassemble the Integra Camino ICP Monitor. Refer all servicing to qualified service personnel at Integra.

To prevent electrical shock, only use the AC power adapter supplied by Integra (REF # MONPWR). Using a different AC power adapter may not provide protection against electric shock.

Danger - Possible explosion hazard if used in the presence of flammable anaesthetics.

Only use Integra supplied accessories on the Integra Camino ICP Monitor. This applies in particular to catheters, catheter cables, battery, AC power adapter, and USB-to-R232 adapter cable.

If the Integra Camino ICP Monitor loses power and shuts down while it is connected to a patient bedside monitor, do not use the ICP values on the patient bedside monitor for patient measurements; the ICP values on the patient bedside monitor will be invalid.

Connect the monitor to an AC power supply immediately if the low battery alarm is activated.

Chapter 1 • System Overview • 3

Warnings

When using the battery:

-Do not heat above 80 °C.

-Do not open battery.

-Do not dispose of in fire.

-Do not short circuit as battery may ignite, explode, leak, or get hot causing personal injury.

-Replace battery with same part number only (REF # BAT1001).

-Use of another battery may present a risk of fire or explosion.

To prevent injury to the patient, user, or other persons, make sure that the battery cover is closed securely during monitor use.

Do not connect an Integra Camino Fiber Optic Catheter and an Integra Camino Flex Catheter simultaneously to the monitor. The Integra Camino ICP Monitor is designed to report ICP measurements with either the Flex Catheter or Fiber Optic Catheter connected to the monitor, but not both together.

To prevent possible patient injury resulting from incorrect ICP measurements, always perform the steps listed on page 23 before implanting a new Integra Camino Fiber Optic Catheter into the patient. In particular, always make sure to adjust the new Fiber Optic Catheter’s ICP value to zero before implantation while the catheter is in the air. Never attempt to re-zero a catheter while the catheter is inside the patient.

Once the Integra Camino Fiber Optic Catheter has been zeroed to the Integra Camino ICP Monitor, do not replace the Fiber Optic Catheter Cable being used for patient measurement. Replacing the Fiber Optic Catheter Cable with another cable after the catheter has already been zeroed may result in inaccurate patient measurements.

Once the Integra Camino Fiber Optic Catheter has been zeroed to the Integra Camino ICP Monitor, do not transfer this zeroed catheter to any other monitor. Transferring a zeroed catheter to a different monitor may result in inaccurate ICP measurements.

To prevent possible patient injury resulting from incorrect ICP measurements, always perform each of the steps listed on page 26 before implanting a new Integra Camino Flex Catheter into the patient. In particular, always leave the Flex Catheter in the air until the monitor completes the autozero (i.e. initialization) process successfully.

Once the Integra Camino Flex Catheter has been initialized (autozeroed) by the Integra Camino ICP Monitor, do not replace the Integra Camino Flex Extension Cable being used for patient measurement. Replacing the Flex Extension Cable with another cable after the Flex Catheter has already been initialized may result in inaccurate ICP measurements.

Once the Integra Camino Flex Catheter has been initialized (i.e. autozeroed) by the Integra Camino ICP Monitor, do not transfer this initialized catheter to any other monitor. Transferring an initialized catheter to a different monitor may result in inaccurate ICP measurements.

4 • Chapter 1 • System Overview

Warnings

The Integra Camino ICP Monitor will only store the Mean ICP data from the most recent 5 days. All stored trend data older than 5 days will be lost. If monitoring is continued for more than 5 days, placement of a new catheter under sterile conditions is recommended. Note that replacing a catheter will reset the trend data. Please extract any data that you wish to retain prior to replacing the catheter.

Do not autoclave or immerse the Integra Camino ICP Monitor in liquid as damage may occur. If the monitor is exposed to liquids, turn off the unit, remove the AC power adapter, dry the unit thoroughly, and send to biomed staff for evaluation before reapplying power.

Only use the cleaning agents listed in Chapter 7 for cleaning and disinfecting the Integra Camino ICP Monitor system. Using solvents or cleaning agents not listed in Chapter 7 may damage the plastic exterior of the Integra Camino ICP Monitor.

Chapter 1 • System Overview • 5

Parts of the Monitor

The Integra Camino ICP Monitor contains hardware, software, and electrical components that support specific Integra catheters for monitoring the patient’s ICP and temperature. The following section provides information on the different parts of the monitor.

About the Front Panel

1

2

3

4

 

The front panel contains:

Number

Item

Description

 

 

 

 

 

 

1

Handle

Handle used for carrying the monitor.

2

Power Status

Green LED button that indicates the monitor is

 

 

being powered by the AC power adapter. Note

 

 

that this button does not illuminate if the monitor

 

 

is being powered by the battery.

3

Power Button

Turns the monitor on and off. This button is

 

 

illuminated when the power is on.

4

Touch Screen

Provides software tools for viewing data and

 

 

controlling parameters for monitoring the

 

 

patient’s ICP and ICT levels.

6 • Chapter 1 • System Overview

About the Rear Panel

1

5

2

6

3

4

The rear panel contains:

Number

Item

Description

 

 

 

 

 

 

1

USB Port

Connection port for extracting trend data via

 

 

USB transfer or digital streaming.

2

AC Power Adapter Port

Connection port for the AC power cord.

3

PMIO Port

Connection port for PMIO cable. This cable

 

 

is used to connect the Integra Camino ICP

 

 

Monitor to a patient bedside monitor.

4

Pole Clamp

Clamping system for securing monitor to an

 

 

equipment pole.

5

Air Vent

Grated opening that allows air being

 

 

circulated by the internal cooling fan to leave

 

 

the monitor.

6

Cable Strap

Rubber strap used to secure AC power

 

 

adapter during transport.

Chapter 1 • System Overview • 7

About the Right Panel

1

2

3

The right panel contains:

Number

Item

Description

 

 

 

 

 

 

1

Temperature Port

Connection port for the temperature connector on

 

 

the Fiber Optic Catheter Cable.

2

Pressure Port

Connection port for the ICP connector on the

 

 

Integra Camino Flex Extension Cable (pressure

 

 

only).

3

Pressure Port

Connection port for the ICP connector on the

 

 

Fiber Optic Catheter Cable.

8 • Chapter 1 • System Overview

About the Bottom Panel

1

2

The bottom panel contains:

Number

Item

Description

 

 

 

 

 

 

1

Battery Door Cover

Removable cover for accessing/replacing the

 

 

14.4V lithium ion battery.

2

Air Vent

Grated opening that allows air being circulated

 

 

by the internal cooling fan to leave the monitor.

About the Left Panel

The left panel does not contain any usable connector ports or buttons.

Chapter 1 • System Overview • 9

This page is intentionally left blank.

10 • Chapter 1 • System Overview

CHAPTER 2 SETTING UP SYSTEM FOR THE FIRST TIME

Procedures for Initial Setup

Step 1: Unpack the System (REF # CAM02)

Remove the contents from the Integra® Camino® ICP Monitor shipping box and verify the following items are included.

Warning

Only use Integra supplied accessories on the Integra Camino ICP

Monitor. This applies in particular to catheters, catheter cables, battery,

AC power adapter, and USB-to-R232 adapter cable.

Content Description

Integra Camino ICP Monitor

Quantity = 1

AC power adapter (18 V DC , 1.67 A, 30 W)

REF # MONPWR

Quantity = 1

14.4 V lithium ion battery

REF # BAT1001

Quantity = 1

Chapter 2 • Setting Up System for the First Time • 11

Content Description

Integra® PMIO patient bedside monitor cable

REF # PMIOMPM

Quantity = 1

Integra® Camino® Fiber Optic Catheter Cable

REF # CAMCABL

Quantity = 1

Integra® Camino® Flex Extension Cable

REF # FLEXEXT

Quantity = 1

USB-to-RS232 adapter cable

REF # EXPORTCAB

Quantity = 1

User manual

REF # UM-CAM02XX

(where XX refers to the country’s language code)

Quantity = 1

After unpacking the contents, inspect the shipment for any signs of damage or loss. If any damages are discovered, notify the carrier, the supplier, and retain all shipping cartons for examination.

12 • Chapter 2 • Setting Up System for the First Time

Step 2: Install the Battery

Perform the following steps with a Phillips screwdriver and the Integra-supplied 14.4 V lithium ion battery.

1.Make sure the monitor is unplugged and turned off.

2.Turn the monitor upside down so the handle is facing downward.

3.Remove the 2 screws to take off the battery cover.

4.Verify the Integra logo on the battery is facing up and insert the battery:

A. Align the battery’s connectors on the front of the battery to the connector slots on the monitor.

B. Slide the battery’s connector into the monitor’s connector slots until the battery is fully inserted into place.

5. Secure the battery cover to the monitor by re-inserting the two small screws.

Chapter 2 • Setting Up System for the First Time • 13

Step 3: Prepare the AC Power Adapter with Region-Specific Plug

1.Remove the AC power adapter from the package and attach the region-specific adapter plug to the backside of the AC power adapter.

 

 

 

 

 

 

 

US

UK

Europe

Australia

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Region-Specific Adapter Plugs

Backside of AC power adapter

A.Align the two slots on the adapter plug over the two prongs on the AC power adapter.

Slots on

 

adapter plug

Prongs on AC power adapter

 

B. Push the adapter plug down over the two prongs until the entire base of the plug sits flush against the AC power adapter.

14 • Chapter 2 • Setting Up System for the First Time

C. Gently slide the adapter plug forward until it snaps into place.

2.To remove the adapter plug, press the release button on the AC power adapter and pull the plug away from the adapter.

Press the release button to remove adapter plug

Step 4: Plug the Monitor into AC Power

Perform the following steps with the Integra-supplied AC power adapter:

1.Place the monitor on a flat surface.

2.Connect the AC power adapter to the Integra Camino ICP Monitor:

A.On the back of the monitor, attach the connector end of the AC power adapter into the port labeled Input: 18 V .

Connector end of

AC power adapter

B. Insert the plug end of the AC power adapter into a grounded AC wall outlet.

3.Turn on the monitor. On the front of the monitor, press the power button; the Integra Camino ICP Monitor will display the Integra logo before initiating the monitor setup process.

Chapter 2 • Setting Up System for the First Time • 15

4.After the initial setup process completes, the Integra Camino ICP Monitor sounds a one-second startup tone and displays the Main panel on the touch screen.

Caution

The purpose of the startup tone verifies that the audio alarms are functioning correctly. If this tone does not sound during the startup process, contact Integra for service.

Step 5: Set the Time and Date

1. On the touch screen, press the Settings tab and select Set Time and Date.

2.On the displayed panel, press the desired field (hour, minutes, date, month, or year) and use the arrows to specify the appropriate setting. Note that you may adjust each of these settings prior to accepting them in the following step.

3.Press Accept; the Integra Camino ICP Monitor will display the selected time/date on the touch screen.

Step 6: Specify the Language

1. On the touch screen, press the Settings tab (fourth tab from the left)

Settings tab

2. Press Set Language.

Set Language button

16 • Chapter 2 • Setting Up System for the First Time

3.In the displayed Language: menu, use the arrows to select the desired language.

4.Press Accept; the Integra Camino ICP Monitor will display all of the text that appears on the touch screen in the selected language.

Step 7: Use AC Power to Charge the Battery to Full Capacity

1.Turn off the monitor. On the front of the monitor, press the power button.

2.Keep the Integra Camino ICP Monitor on AC power with the monitor turned off for 5 hours. This will re-charge the battery to full capacity.

3.After 5 hours, turn on the monitor by pressing the power button.

4.On the touch screen, view the battery power symbol on the status bar to verify the symbol displays four green bars; this indicates the battery has full charge.

Chapter 2 • Setting Up System for the First Time • 17

This page is intentionally left blank.

18 • Chapter 2 • Setting Up System for the First Time

CHAPTER 3 SETTING UP SYSTEM FOR CLINICAL USE

Setting Up System for Clinical Use..............................................................

19

About the Integra Catheters .........................................................................

23

Connecting the Integra® Camino® Fiber Optic Catheters (110-4 Series) 24

Connecting the Integra® Camino® Flex Catheters ....................................

27

Connecting to a Patient Bedside Monitor (if applicable) ...........................

30

Storing the System ........................................................................................

33

Setting Up System for Clinical Use

The following section contains instructions for positioning the monitor, powering the monitor, and connecting catheters to the monitor prior to clinical use.

Positioning the Monitor

The Integra® Camino® ICP Monitor is intended to be positioned on a hard flat surface or securely clamped to an equipment pole or bed support next to the patient. The distance between the patient and the monitor is restricted by the length of the Integra catheter cable. It is not permitted to extend the Integra catheter cable with non-Integra extension cables.

Caution

To prevent the monitor from overheating:

Do not block the air vents on the rear and bottom of the unit

Do not place the monitor on a soft or uneven surface, which may result in blockage of the monitor’s air vents. For example, do not place the monitor on the bed during patient transport.

Attaching to Equipment Pole (if applicable)

The Integra Camino ICP Monitor includes a clamp for attaching the monitor to an equipment pole. To attach:

1. On the rear of the monitor, fit the pole clamp around the equipment pole.

Chapter 3 • Setting Up System for Clinical Use • 19