Ideal 61-361, TEST PRO 360 Manual

5 (3)

#61-361

IDEAL Test Pro®

360 Series True RMS

Multimeter

WARNING!

1.DO NOT UNDER ANY CIRCUMSTANCES EXCEED THESE RATINGS:

-Voltage is not to exceed 600V AC or DC.

-Resistance, Capacitance, Logic, and Continuity functions are not to be performed on circuits capable of delivering greater than 500V AC or DC.

-Current measurements are not to be performed on circuits capable of delivering greater than 600V AC on insulated conductors, 250V AC on uninsulated conductors.

2.To avoid electrical shock hazards and/or damage to the meter:

-Do not exceed the voltage ratings for the meter. Use caution when measuring voltage.

-Do not use during electrical storms. AC power sources with inductive loads or electrical storms may result in high voltage. High energy transients can damage meter and present a dangerous shock hazard.

-Turn off the power to the circuit or device being measured before taking resistance and capacitance measurements. Fully discharge all capacitors before measuring.

3.Ensure meter is in proper working order before using. Visually inspect meter for damage. Performing a continuity check can verify proper operation. If the meter reading goes from overload to zero, this typically means the meter is in proper working order.

4.Visually inspect leads for damage before using. Replace if insulation is damaged or leads appear suspect.

5.Never ground yourself when taking electrical measurements. Do not touch exposed metal pipes, outlets, fixtures, etc. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any other approved insulating material. Keep your fingers behind the finger guards on the probes. Work with others.

6.Before beginning all unknown measurements, set meter to the highest range possible.

Page 2

WARNING! (cont.)

7.Before breaking a circuit for testing, turn off the power to the circuit. When disconnecting from a circuit, disconnect the hot lead first, then the common lead.

8.Disconnect the meter from the circuit before turning off any indicator, including motors, transformers,and solenoids.

Overload Protection

VAC + VDC

200m Vrange

500VDC/350VAC

 

 

for 15 sec

 

>200m Vrange

600V AC/DC

AAC + ADC

mA input

0.5A/250V

 

10A input

10A/600V

Ohms ()

 

500V AC/DC

Diode

 

500V AC/DC

Continuity

 

500V AC/DC

Capacitance (MFD)

 

500V AC/DC

Frequency (Hz)

 

500V AC/DC

Phase Rotation (RST)

 

500V AC/DC

Unit of Measure Multipliers

For your reference, the following symbols are often used to

make measurement easier:

 

Symbol

Verbal

Multiplier

M

mega

x1,000,000

k

kilo

x1,000

m

milli

÷1,000

µ

micro

÷1,000,000

AC Voltage (VAC)

PHASE ROTATI

HZ-MFD-

VAC 200

PHASE ROTATION-TRUE RMS

600

 

HZ-MFD-AUTO OFF

 

2000m

VAC

600

600

VDC

200

200

200m

2000m

 

20

200m

 

 

 

 

 

2000m

 

 

 

 

RST

RST

 

 

200m

 

Hz

 

 

ADC

 

 

 

 

 

200M

 

 

AAC

 

 

 

 

 

20M

 

 

200

 

 

 

 

 

200k

 

 

2k

 

 

20k

20k

MFD

 

 

2k 200

 

 

2A

 

 

 

 

MFD

 

CAT. LLL 600V

 

 

 

COM

 

V

 

 

 

!

 

 

R/L1

 

600V

 

 

 

 

MAX

 

 

 

 

350V/15

 

 

 

 

SEC FOR

 

 

 

 

200mV

 

 

 

 

RANGE

 

 

Black

 

Red

 

 

 

 

 

 

 

Function

Range

Resolution

Accuracy

AC Voltage

200.0 mV

.1 mV

±(2.0% + 4)

 

2000 mV

1 mV

 

50 to 500Hz

 

200.0 V

.1V

 

all ranges

 

600 V

1 V

 

 

To Measure AC Voltage:

1.Plug the test leads into the meter inputs as indicated in the above diagram.

2.Select the proper range to be used within the VAC Function.

3.Connect the meter in parallel with the load or circuit.

4.Measure AC voltage.

Page 3

Page 4

Ideal 61-361, TEST PRO 360 Manual

DC Voltage (VDC)

TION-TRUE RMS

-AUTO OFF

 

 

 

600

VDC

PHASE ROTATION-TRUE RMS

 

 

HZ-MFD-AUTO OFF

 

 

VAC

600

600

VDC

2000m

200

200

2000m

 

20

 

200m

 

 

 

 

 

2000m

 

 

 

 

 

RST

 

 

 

ADC

Hz

 

 

ADC

 

 

 

 

 

 

 

200M

 

 

 

 

20M

 

 

200

 

 

 

 

 

200k

 

 

2k

 

 

20k

20k

MFD

 

 

2k 200

 

 

2A

 

 

 

 

MFD

 

CAT. LLL 600V

 

 

 

 

 

+ -

 

COM

 

V

R/L1

 

!

 

 

 

600V

 

 

 

 

MAX

 

 

 

 

350V/15

 

 

 

 

SEC FOR

 

 

 

 

200mV

 

 

 

 

RANGE

 

 

Red

Black

Function

Range

Resolution

Accuracy

DC Voltage

200.0 mV

.1 mV

 

 

2000 mV

1 mV

± (1.2% +1)

 

20.00 V

.01 V

all ranges

 

200.0 V

.1 V

 

 

600 V

1 V

 

To Measure DC Voltage:

Do not attempt to measure voltage in circuits capable of delivering more than 600V.

1.Plug the leads into meter inputs as indicated in the diagram above.

2.Select the proper range to be used within the VDC Function.

3.Connect the meter in parallel with the load or circuit.

4.Measure DC voltage.

Resistance (Ohms)

PHASE ROTATION-TRUE RMS

HZ-MFD-AUTO OFF

 

 

VAC 200

200 VDC

 

 

600

600

 

 

2000m

20

 

 

200m

 

 

 

2000m

 

 

RST

200m

 

 

Hz

ADC

 

 

 

Hz

 

200M

AAC

 

 

20M

200

200M

 

 

 

200k

2k

 

 

20M

 

20k

20k

 

2k 200

MFD

 

 

2A

 

 

 

MFD

CAT. LLL 600V

20k

 

 

 

COM

V

2k

200

 

!

 

 

 

 

 

R/L1

600V

 

 

 

MAX

 

 

 

350V/15

 

 

 

SEC FOR

 

 

 

200mV

 

 

 

RANGE

 

Black

 

 

 

Red

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Function

Range

Resolution

 

Accuracy

Resistance

200.0

 

 

.1

 

± (1.5% + 4)

 

2.000k

 

.001k

 

± (1.5% + 4)

 

20.00k

 

.01k

 

± (1.5% + 4)

 

200.0k

 

 

.1k

 

± (1.5% + 4)

 

20.00M

 

.01M

 

± (2.5% + 4)

 

200.0M

 

.1 M

 

± (5.0% +20)

To Measure Resistance:

Resistance is measured in Ohms.

1.Turn the power off to the circuit or device that is to be measured and discharge all capacitors before attempting a measurement.

2.Plug the red test lead into the “V-” meter input.

3.Plug the black test lead into the “Com” meter input.