Getinge Arjohuntleigh ALPHA RESPONSE Instructions For Use Manual

ALPHA RESPONSE
INSTRUCTIONS FOR USE
0086
...with people in mind
General Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About Alpha Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Mattress: Overlay and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Seat Cushion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Clinical Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Care of the patient when sitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Preparing the Systems for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installing the Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installing the Seat Cushion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installing the Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Controls, Alarms and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Mattress - Pump Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Installing the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Initial Inflation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mattress Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Stopping Therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
To Deflate the Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
To Store the Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Transport Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
CPR Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Seat Cushion - Pump Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Seat Cushion Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
To Deflate the Seat Cushion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Alpha Response System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Mattress Replacement, Overlay and Seat Cushion . . . . . . . . . . . . . . . . . . . . . . . . 25
Serial Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
(i)
Technical Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
(ii)

GENERAL SAFETY

Before you connect the system pump to a mains socket, read carefully all the installation instructions contained within this manual.
The system has been designed to comply with regulatory safety standards including:
EN60601-1:1990/A13:1996 and IEC 60601-1:1988/A2:1995.
UL60601-1, UL2601-1 and CAN/CSA C22.2 No. 601.1-M90.
EN60601-1:2006, EN60601-1-11:2010 and IEC 60601-1:2005.
AAMI/ANSI ES60601-1:2006 and CAN/CSA C22.2 No.60601.1(2008).
Safety Warnings
It is the responsibility of the care giver to ensure that the user can use this product safely.
Whilst the patient is unattended, the decision to use safety sides should be based on clinical assessment and in line with local policy.
Alignment of the bed frame, safety sides and the mattress should leave no gap wide enough to entrap a patient's head or body, or to allow egress to occur in a hazardous manner where entanglement with the mains power cable and tubeset or air hoses may result. Care should be exercised to prevent occurrence of gaps by compression or movement of the mattress. Death or serious injury may occur.
Make sure that the mains power cable and tubeset or air hoses are positioned to avoid causing a trip or other hazard, and are clear of moving bed mechanisms or other possible entrapment areas. The mains power cable of this pump is designed to allow movement of the bed, and should be fitted into the cable management flaps along the sides of the mattress, as described in this manual.
When using a seat cushion, the tubeset may be a trip hazard; position the tubeset so that it does not create a trip hazard, and always supervise babies and small children when the product is in use.
Electrical equipment may be hazardous if misused. There are no user-serviceable parts inside the pump. The pump's case must only be removed by authorised technical personnel. No modification of this equipment is allowed.
The mains power socket/plug must be accessible at all times. To disconnect the pump completely from the electricity supply, remove the plug from the mains power socket.
The CPR control and/or the CPR indicator tag must be visible and accessible at all times.
Disconnect the pump from the mains power socket before cleaning and inspecting.
Keep the pump away from sources of liquids and do not immerse in water.
Do not use the pump in the presence of uncontained flammable liquids or gasses.
The cover of this product is vapour permeable but not air permeable and may present a suffocation risk.
Do not use the mattress without a cover, it provides a protective barrier.
Bags supplied with this equipment may present a suffocation risk; to avoid the risk of suffocation keep the bags away from babies and small children.
There is no transport mode on the seat cushion.
Only the pump and mattress or seat combination as indicated by ArjoHuntleigh should be used. The correct function of the product cannot be guaranteed if incorrect pump and mattress or seat combinations are used.
(iii)
Precautions
For your own safety and the safety of the equipment, always take the following precautions:
Placing extra layers between the patient and the mattress potentially reduces the benefits provided by the mattress and should be avoided or kept to a minimum. As part of sensible pressure area care, it is advisable to avoid wearing clothing which may cause areas of localised high pressure due to creases, seams, etc. Placing objects in pockets should be avoided for the same reason.
Do not expose the system, especially the mattress, to naked flames, such as cigarettes, etc.
In the event of a fire, a leak in the seat or mattress could propagate the fire.
Do not use or store the system in direct sunlight.
Do not use phenol-based solutions to clean the system.
Make sure the system is clean and dry prior to use or storage.
Never use sharp objects or electrically heated under blankets on or under the system.
Store the pump and mattress in the protective bags supplied.
Pets and children must be supervised in the vicinity of the system.
Electromagnetic Compatibility (EMC)
This product complies with the requirements of applicable EMC Standards. Medical electrical equipment needs special precautions regarding EMC and needs to be installed in accordance with the following instructions:
The use of accessories not specified by the manufacturer may result in increased emissions by, or decreased immunity of, the equipment, affecting its performance.
Portable and mobile radio frequency (RF) communications equipment (e.g. mobile/cell phones) can affect medical electrical equipment.
If this equipment needs to be used adjacent to other electrical equipment, normal operation must be checked before use.
For detailed EMC information contact ArjoHuntleigh service personnel.
Environmental Protection
Incorrect disposal of this equipment and its component parts, particularly batteries or other electrical components, may produce substances that are hazardous to the environment. To minimise these hazards, contact ArjoHuntleigh for information on correct disposal.
Expected Service Life
The Alpha Response™ pump has an expected service life of seven years. To maintain the condition of the pump, have the pump serviced regularly according to the schedule recommended by ArjoHuntleigh.
Do NOT use unapproved accessories or attempt to modify, disassemble or otherwise misuse the Alpha extreme cases, death.
Response system. Failure to observe this caution could result in injury, or in
Design Policy and Copyright
® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. The content of this publication may not be copied either whole or in part without the consent of ArjoHuntleigh.
© ArjoHuntleigh 2013
(iv)

1. Introduction

About this Manual This manual is your introduction to the Alpha

Response™ system. Use it to initially set up the system and keep it as a reference for day-to-day routines and as a guide to maintenance. You must read and fully understand this manual before using the system.
If you have any difficulties in setting-up or using the Alpha Response system, contact your ArjoHuntleigh sales office, listed at the end of this manual.

Intended Use The intended use of this product is to prevent and/or

manage pressure ulcers for patients up to 160 kg (352 lb).
About Alpha
Response
The Alpha Response
system should be used as part of a
prescribed plan of care (refer to “Indications” on page 4).
The Alpha Response system is a pressure redistributing mattress replacement, mattress overlay or seat cushion system designed to complement pressure ulcer treatment and prevention protocols. The product offers two therapeutic modes:
• Active
1
(Alternating) mode which periodically redistributes pressure away from vulnerable areas by inflating and deflating the cells beneath the body every 10 minutes.
• Reactive
1
(Constant Lower Pressure or CLP) mode where the cell pressure is reduced and held constant across the surface in order to lower the pressure exerted on the body.
The product also offers an additional option, Transport Mode, where therapy is interrupted and the mattress cells become static in order to assist with patient transport.
The mattresses can be used on standard hospital and normal domestic beds. Beds with divided sections for independent elevation of a patient's head and/or knees can be adjusted with these mattresses in position. The seat cushion can be used on standard hospital and normal domestic chairs.
1. International Pressure Ulcer Prevention and Treatment Guideline (2009). www.epuap.org
1
The pump has three settings for the patient weight range:
• Light: 40-70 kg
• Normal: 70-120 kg
• Heavy: 120-160 kg The pump will automatically detect whether a mattress
or seat cushion is connected.
If the backrest on the bed is raised (the patient is in a semi-recumbent position), the system detects the new position and automatically increases the pressure in the mattress cells to provide optimal pressure redistribution to the patient.

Pump

LCD Screen, Controls
and LED Indicator
A full technical description of the Alpha Response system can be found in the Service Manual, part number SER0021, available from your ArjoHuntleigh sales office.
Carry
Handle
Tubeset
Connector Bed-Hooks
Swing-out
2

Mattress: Overlay and Replacement

Detachable Cover3 Static Head Cells
14 Standard Cells
FOOT END
Tubeset
HEAD END
CPR
Control
Securing Straps
Cable Management Flap
Transportation Loop

Seat Cushion

Active (Alternating) Cells
Fixing Straps
Base Cover
Detachable Cover
Deflation Valves
(on underside for
packing and storage)
Tubeset
3

2. Clinical Applications

Indications The Alpha Response systems are indicated for the

prevention and/or management of all categories
1
of pressure ulcer, when combined with an individualised, comprehensive pressure ulcer protocol: for example, repositioning, nutritional support, skin care. Selection should be based upon a holistic assessment of the patient’s individual care needs.
These systems represent one aspect of a pressure ulcer management protocol; all other aspects of care should be considered by the prescribing clinician.
If existing wounds do not improve or the patient's condition changes the overall therapy regimen should be reviewed by the prescribing clinician.
The above are guidelines only and should not replace clinical judgement.
The Alpha Response mattress and cushion are designed for patients weighing up to

Contraindications Do not use Alpha Response systems for patients with

unstable spinal fractures.
160 kg (352 lb).

Cautions If patients have other unstable fractures, or conditions

which may be complicated by a soft or moving surface, advice should be sought from an appropriate clinician before use.
While the Alpha Response systems have been designed to manage patients up to the weight limits indicated above, those approaching this upper limit are likely to have additional care and mobility needs and may be better suited to a specialist bariatric system.
Active therapy (alternating) cushions may be unsuitable for patients with poor sitting posture or pelvic deformity; advice from a seating specialist should be sought.
Care of the patient
when sitting
Seated patients are at increased risk of pressure ulcers particularly if they are immobile or have wounds over the seating area. For optimal outcome, provide a pressure redistributing seat cushion in a chair which promotes a good sitting posture and has a level base seat to support the cushion, in addition to an individualised repositioning programme.
Mattress and cushion combinations may have different upper weight limits. Cushions should be used in combination with pressure-redistributing
mattresses to provide 24-hour therapy.
1. NPUAP/EPUAP International Pressure Ulcer Guideline, 2009.
4

3. Installation

WARNING
Bags supplied with this equipment may present a suffocation risk; to avoid the risk of suffocation keep the bags away from babies and small children.
Preparing the
Systems for Use
Remove the system from the packaging. You should have the following items:
Alpha Response pump including mains power cable.
Alpha Response mattress overlay, Alpha Response mattress replacement or Alpha Response seat cushion, which all have integral tubesets and top covers.
Do not use the mattress or seat cushion without a
cover.

Installing the Mattress

Caution
Do not use the mattress overlay directly on the bed frame.
Mattress Overlay 1. Place the overlay on top of the base mattress, with
the tubeset located near the foot end of the bed and the CPR at the head end. The mattress cover must be uppermost.
2. Attach the mattress to the bed frame using the four
strap assemblies on the underside of the mattress replacement:
• The strap assemblies are attached to four of the five pairs of anchor points, their position depending on the type of bed frame. One strap assembly must be at the head end of the mattress and a second must be at the foot end. The remaining two can be attached to any of the three pairs of anchor points in the middle of the mattress.
• Pass each half of the strap assembly under the base mattress, connect them together and pull the strap assembly tight.
5
Top View View on Underside of Mattress
Head
End
Base Mattress
Mattress Replacement 1. Remove the existing mattress from the bed frame
Strap Strap AssembliesTypical Strap Anchor Points
Foot
End
and check that there are no protruding bed springs or sharp objects on the bed frame surface.
2. Unroll the mattress onto the bed frame and ensure that the tubeset is located at the foot end of the bed and the CPR at the head end.
3. Attach the mattress to the bed frame using the eight fastener straps. The eight fastener straps can be moved to any of the ten anchor points on the bottom of the mattress replacement. This allows for attaching the mattress to different types of bed frame.
If the bed has divided sections for independent
elevation of a patient's head and/or knees, attach the mattress to the movable parts of the bed frame only.
One strap assembly must be at the head end of the
mattress and a second must be at the foot end.
WARNING
The CPR control and CPR indicator tag must be visible and accessible at all times.
Closing the CPR
Control
On the mattress overlay or mattress replacement, make sure the CPR control is closed:
• The CPR control consists of a connector in the side
of the mattress with a removable plug fitted.
• Make sure the CPR plug is fully pushed into the
connector on the mattress.
6
12

Installing the Seat Cushion

Cautions
• Do not use the seat cushion without a foam cushion beneath it.
• Always use the seat cushion with the protective top cover.
• Always use the seat cushion in the correct orientation.
• Avoid trailing cables - make sure that cables and tubes are positioned beneath the chair to avoid causing a hazard.
• Position the tubeset so it does not create a trip hazard and always supervise babies and small children when the product is in use.
1. Check that there are no sharp objects on the chair which may puncture the cushion.
2. Place the cushion on top of the chair surface. Stand in front of the chair and look towards it. Make sure that:
• The cells are uppermost.
• The tubeset appears from the front right corner of
the cushion.
• The cells in the seat cushion are in a horizontal
position across the chair, with the “V” shape pointing towards the front.
7
3. Secure the seat cushion to the chair by using the fixing straps as shown in the following illustrations.
4. If the chair is of the open sided construction, then fix the cushion as shown below:
5. If the chair is of the closed side type with a removable seat cushion, fix the seat cushion as shown below:
6. If the chair is of the closed side type with a non­removable seat cushion, then security will rely on the anti-slip base material of the seat cushion.
7. Place the protective cover over the seat cushion and ensure that the logo and the orientation icon, printed on the cover, are uppermost and at the front of the seat.
8. Zip the cover onto the seat cushion, taking care not to trap any material in the zip.
8

Installing the Pump 1. Position the pump, feet down, on any convenient

horizontal surface or alternatively suspend from the bed foot-rail by means of the swing-out hooks.
2. Insert the mains power plug into a suitable mains power socket.
Bed Foot-Rail
Mains Power Plug
Cable Management
in Mattresses
Swing-Out Hook
The mains power cable should be put through one of the cable management flaps which are on each side of the mattress, as follows:
1. Locate one of the cable management flaps.
2. If necessary, open the press studs along the flap.
3. Run the mains power cable along the side of the mattress securing the flap round the cable using the press studs.
Connecting the
Tubeset
Make sure the mattress/seat tubeset is not “kinked” or twisted, and push the tubeset connector firmly onto the pump until it clicks into place.
Make sure that the tubeset is securely connected to
the pump.
9
“Click”
12
Disconnecting the
Tubeset
Release
Button
To disconnect the tubeset at any time, push down the release button on the top of the tubeset connector and pull the tubeset connector away from the pump.
This will put the mattress into Transport Mode and will not deflate the mattress. To deflate the mattress, refer to “To Deflate the Mattress” on page 18.
There is no transport mode on the Alpha Response
seat cushion.
12
System Operation The system is now ready for use. Refer to “Controls,
Alarms and Indicators” on page 11 and “Mattress ­Pump Operation” on page 16 for day-to-day operating instructions.
10

4. Controls, Alarms and Indicators

Control Panel

Weight Range Icons Activity Indicator
Weight/Mode Select Buttons
LCD Screen This displays the operating mode and status of the pump,
LCD Screen
Mattress/Seat Mode Icon
Select Button
as follows:
• Mattress/Seat Status.
• Weight Range (or Mattress Mode, if selected by the
Mode Selection).
LED Indicator
Run/Standby
Button
Mode
Selection
Icon
• Mode Selection.
• Activity Indicator.
Run/Standby Button Press the Run/Standby button to put the pump into the
Run mode; the LED indicator changes to green.
LED Indicator
To put the pump into Standby, press the Run/Standby button for approximately three seconds; this prevents accidental operation. The LCD screen goes blank and the LED indicator changes to amber.
After you put the pump in Standby, if you press the Run/ Standby button within approximately 15 seconds the
Run/Standby Button
pump goes straight to the Run mode and continues the previous therapy; if you wait more than 15 seconds the pump re-initialises and restarts the initial mattress/seat cushion inflation sequence.
If the mains power is disconnected from the pump while the pump is operating, the pump enters the Power Fail Alarm mode (refer to “Alarms” on page 15). Press and hold the Run/Standby button; the alarm stops and the pump switches off completely.
11
LED Indicator The multicolour LED adjacent to the Run/Standby
button indicates the status of the pump, as follows:
Amber (Constant) External power is applied to the
pump, but the pump is in Standby.
Green (Constant) The pump is in Run mode and
operating.
Red (Flashing) The pump has detected an alarm
condition.
Mattress/Seat Status There are five mattress/seat mode icons which can be
displayed, as follows:
Mattress Backrest Horizontal
Active (Alternating) Mode
Mattress Backrest Horizontal Reactive (CLP) Mode
Mattress Backrest Raised
Active (Alternating) Mode
Mattress Backrest Raised Reactive (CLP) Mode
Seat
Active (Alternating) Mode
Select Button The function of the Select button depends on the pump
control change being carried out and the icon that is displayed on the LCD screen directly above the button.
Mode Selection Icon This “double-arrow” icon indicates that the pump is in
normal therapy mode, and pressing the Select button below it will select either the Weight Range icons or the Mode icons.
When the Weight Range or Mode is being changed, the Mode Selection icon changes to a “tick” and flash. Pressing the Select button under the “tick” confirms the new selection.
If the Select button is not pressed for five seconds
when the flashing “tick” is displayed, the requested pump status change is ignored, the “tick” reverts back to the Mode Selection icon and the pump continues in its current state.
12
Weight Range There are three Weight Range icons displayed on the
LCD screen, the relative size of each “person” icon corresponding to the patient weight range. The selected weight range is indicated by the corresponding icon being solid and the other two icons as outlines.
Patient comfort and clinical judgement should be
used to select the correct weight range setting.
Light Weight Weight Range: 40-70 kg
Normal Weight Weight Range: 70-120 kg
Heavy Weight Weight Range: 120-160 kg
To change the Weight Range, do the following:
1. Press the Weight/Mode Select buttons to highlight the new Weight Range icon; the new icon is solid and flashing.
2. The Mode Selection icon changes:
3. Press the Select button to confirm the new Weight Range setting.
13
Mode If the Select button is pressed during therapy when the
Mode Selection icon is displayed, then the Weight Range icons are replaced by two Mode icons. The selected Mode is indicated by the corresponding icon being solid and surrounded by a square and the remaining icon as an outline:
Active (Alternating) Mode Backrest Horizontal
Reactive (CLP) Mode Backrest Horizontal
To change the Mode, do the following:
1. Make sure the Mode Selection icon is displayed.
2. Press the Select button, and the two Mode icons are then displayed.
3. Press the Weight/Mode Select buttons to highlight the new Mode icon. The new Mode icon is solid with a square border.
4. The new Mode icon and square flashes and the Mode Selection icon changes:
5. Press the Select button to confirm the new Mode.
6. Press the Select button again and the two Mode icons are replaced by the Weight Range icons.
Activity Indicator After the mattress has inflated and the system is in the
normal operating mode, an Activity Indicator icon is displayed in the top right corner of the LCD screen:
• The Activity Indicator rotates in a clockwise
direction to show that the pump is operating normally.
• The Activity Indicator stops rotating and starts
flashing if the pressure changes dramatically e.g. if
the patient moves heavily on the mattress or if the Weight Range is changed. Once the pump pressure has stabilised around its target pressure the Activity Indicator stops flashing and starts rotating again.
14

Alarms 1. When the pump detects an alarm condition:

• The corresponding visual alarm is displayed on the LCD screen, as detailed below.
• The LED indicator on the control panel flashes alternately red and green.
• An audible alarm is sounded, which increases in pitch if the alarm is ignored.
2. Press the Run/Standby button to stop the alarm.
3. Refer to Section 9, Page 26 “Troubleshooting” for the alarms, their possible causes and their remedies.
If the operation of the pump changes during use,
refer to Troubleshooting procedures on page 26 of this IFU before calling a service engineer or contacting your local ArjoHuntleigh sales office
.
15

5. Mattress - Pump Operation

These instructions cover the day-to-day operation of the system. Other operations, such as maintenance and repair, should only be carried out by suitably qualified personnel.
Refer to Section 4, Page 11 “Controls, Alarms and Indicators” for a
comprehensive description of the controls and indicators on the pump.
WARNING
Do not place the patient on the mattress until it is fully inflated

Installing the System Before using the Alpha Response system make sure the

system has been installed correctly in accordance with Section 3, Page 5 “Installation”.

Initial Inflation 1. Connect the pump to the mains power supply using

the supplied cable and switch on the pump.
2. When mains power is first connected to the pump:
• The LED indicator is illuminated amber to indicate that external power is applied to the pump but the pump is still in Standby.
• The LCD screen is blank.
3. Press the Run/Standby button to put the pump into the Run mode:
• The LED indicator changes to green.
• The LCD screen displays the ArjoHuntleigh logo
for 5 seconds, followed by the Alpha Response animated logo for five seconds.
• The pump carries out a self-test routine and
initialises itself.
4. At the end of this start-up sequence, the pump starts to inflate the mattress system.
If no tubeset is connected to the pump, or the tubeset
is not connected securely, the LCD screen displays a No Connection alarm, which alternates the No Mattress and No Seat screens, and an audible alarm is sounded. To clear the alarm, connect a mattress or seat tubeset to the pump.
16

Mattress Operation 1. Make sure the CPR control on the mattress is

closed.
2. At the start of the inflation sequence, the default mattress-inflation screen is displayed.
• The default patient weight range setting is
Normal, as indicated by the three Weight Range icons.
• The default cell inflation mode is Active
(Alternating) and the default setting for the bed frame is with the backrest horizontal, as indicated by the Mattress/Seat Mode icon.
• For the duration of the inflation sequence, the
patient body is replaced by an “egg timer” icon.
3. The pump may take up to 40 minutes to inflate a mattress to operating pressure.
Failure to fully inflate the mattress will result in a
low pressure alarm.
4. When the mattress inflation is complete, the default mattress-operating screen is displayed:
• The “egg timer” is replaced by the patient body
icon.
• The Activity Indicator is displayed and starts
rotating to indicate that the pump is in the Run mode.
5. The patient can now be put onto the mattress.
6. To change the Weight Range setting or put the mattress into Reactive (CLP) mode, refer to Section 4, Page 11 “Controls, Alarms and Indicators”.
These settings can also be changed while the
mattress is being inflated.
17
When changing between operating modes, patient’s
monitoring and repositioning program should be reviewed.
If the operation or performance of the pump
changes during use, stop using the system and refer to Section 9, Page 26 “Troubleshooting”.
Raising the Backrest
on the Bed Frame
1. If the backrest on the bed is raised (the patient is in a semi-recumbent position), the system detects the new position and automatically does the following:
• Increases the air pressure in the mattress cells to
provide optimal pressure redistribution to the patient.
• Changes the mattress icon from a “backrest
horizontal” icon to a “backrest raised” icon.
The screen below shows the mattress in Active
(Alternating) mode when the backrest is raised.
2. When the bed frame is put back to the “backrest horizontal” position, the system detects this and automatically does the following:
• Decreases the air pressure in the mattress cells to
the value before the backrest was raised.
• Changes the mattress icon back to the “backrest
horizontal” icon.

Stopping Therapy To stop the therapy, press and hold the Run/Standby

button for three seconds to put the pump into Standby.
• The LED indicator changes to amber.
• the LCD screen goes blank.
• If the pump is to be completely isolated from the
mains, remove the plug from the mains power socket.
To Deflate the
Mattress
To deflate the mattress:
1. Stop the therapy and put the pump into Standby.
2. Disconnect the tubeset from the pump (refer to “Disconnecting the Tubeset” on page 10).
3. Activate the CPR control at the head end of the mattress to deflate it (refer to “To Activate the CPR Control” on page 20)
18

To Store the Mattress Following deflation:

1. Start rolling the mattress from the foot end towards the head end; stop after ONE turn of the mattress.
2. Bring the tubeset connector over the mattress and secure.
3. Continue to roll the mattress from the foot end towards the head end.
Make sure the mattress is dry before rolling it up.
3
1
2

Transport Mode To transport a patient who is lying on the Alpha

Response mattress:
1. Stop the therapy and put the pump into Standby.
2. Disconnect the tubeset from the pump (refer to “Disconnecting the Tubeset” on page 10). This will automatically put the mattress into transport mode.
3. Put the tubeset connector through the transportation loop to prevent any damage to the tubeset while the bed is being moved.
The patient will remain supported for up to eight
hours on the mattress.
4. To resume normal operation, reconnect the tubeset to the pump and restart the therapy.
Tubeset Connector
Transportation Loop
19
IMPORTANT
IN THE EVENT OF CARDIAC ARREST
CPR Control In the event of a patient suffering cardiac arrest and CPR
(Cardio-Pulmonary Resuscitation) needing to be administered, activate the CPR control to rapidly deflate the mattress:
To Activate the

CPR Control

Use a quick, firm pull on the CPR tag to remove the CPR plug. The air is rapidly evacuated from the mattress.
The plug is fastened to the mattress by a strap. If the pump is operating when the CPR is activated,
the Low Pressure alarm may be activated.
12
WARNING
The CPR control and CPR indicator tag must be visible and accessible at all times.
To Close the CPR Control
Push the CPR plug into the connector on the mattress.
Make sure the plug is securely fitted.
12
20

6. Seat Cushion - Pump Operation

General Refer to the "Initial Inflation" section in

Section 5, Page 16 “Mattress - Pump Operation”.
WARNING
Do not place the patient on the seat cushion until it is fully inflated
Seat Cushion
Operation
1. At the start of the inflation sequence, the default seat-inflation screen is displayed.
• The default patient weight range setting is
Normal, as indicated by the three Weight Range icons.
• The only cell inflation mode is Active
(Alternating), as indicated by the Mattress/Seat Mode icon.
• For the duration of the inflation sequence, the
patient body is replaced by an “egg timer” icon.
2. The pump may take up to 5 minutes to inflate a seat cushion to operating pressure.
Failure to fully inflate the seat cushion will result in
a low pressure alarm.
3. When the seat cushion inflation is complete, the default seat-operating screen is displayed; the “egg timer” is replaced by the patient body icon.
4. The patient can now sit on the seat cushion.
5. To change the Weight Range setting, refer to Section 4, Page 11 “Controls, Alarms and Indicators”.
This setting can also be changed while the seat
cushion is being inflated.
21
To Deflate the Seat
Cushion
To deflate the seat cushion, do the following:
1. Press and hold the Run/Standby button for three seconds to put the pump into Standby.
• The LED indicator changes to red.
• The LCD screen goes blank.
• If the pump is to be completely isolated from the
mains, remove the plug from the mains power socket.
2. Disconnect the tubeset from the pump (refer to “Disconnecting the Tubeset” on page 10).
3. Depress the two valves on the underside of the seat cushion to release the remaining air from it.
There is no transport mode on the Alpha Response
seat cushion.
22

7. Decontamination

The following processes are recommended, but should be adapted to comply with the local or national guidelines (Decontamination of Medical Devices) which may apply within the Healthcare Facility or the country of use. If you are uncertain, you should seek advice from your local Infection Control Specialist.
The Alpha Response system should be routinely decontaminated between patients and at regular intervals while in use; as is good practice for all reusable medical devices.
WARNING
Remove the electrical supply to the pump by disconnecting the mains power cord from the mains power supply before cleaning. Protective clothing should always be worn when carrying out decontamination procedures.
Caution
Do not use Phenol-based solutions or abrasive compounds or pads during the decontamination process as these will damage the surface coating. Do not boil or autoclave the cover.
Avoid immersing electrical parts in water during the cleaning
process. Do not spray cleaning solutions directly onto the pump.
To Clean Clean all exposed surfaces and remove any organic
debris by wiping with a cloth moistened with a simple (neutral) detergent and water. Dry thoroughly.
Do not allow water or cleaning solutions to collect on the surface of the pump.
Chemical Disinfection To protect the integrity of the cover we recommend a
chlorine-releasing agent, such as sodium hypochlorite, at a strength of 1,000ppm available chlorine (this may vary from 250ppm to 10,000ppm depending on local policy and contamination status).
Wipe all cleaned surfaces with the solution, then wipe with a cloth moistened in water and dry thoroughly.
Alcohol based disinfectants (strength 70%) may be used as an alternative.
Ensure the product is dry before storage.
If an alternative disinfectant is selected from the wide variety available we recommend that suitability for use is confirmed with the chemical supplier prior to use.
DO NOT WRING/MANGLE, AUTOCLAVE OR USE PHENOLIC BASED SOLUTIONS TO LAUNDER THE TOP COVER.
23
For information on the mattress top cover, including laundering guidelines, refer to “COVER SPECIFICATION” on page 29.
Re-use with multiple
patients
Professional hygiene maintenance is required before re-use of the system with a different patient.
24

8. Routine Maintenance

Alpha Response System

Maintenance The equipment has been designed to be virtually
maintenance-free between service periods.
Servicing ArjoHuntleigh will make available on request service
manuals, component parts lists and other information necessary for ArjoHuntleigh trained personnel to repair the system.
Service Period ArjoHuntleigh recommend that the Alpha Response
system should be serviced by an ArjoHuntleigh authorised service agent, after 12 months running time has elapsed.
The Service symbol illuminates to indicate that the pump is ready for a service (refer to “Alarms” on page 15).

Pump

General Care,
Maintenance and
Inspection
Check all electrical connections and power cable for signs of excessive wear or damage.
If the pump has been subjected to abnormal treatment, e.g. immersed in water or dropped, the unit must be returned to an authorised service centre.
Biofilter The internal biofilter can be run continuously for 12
months before it requires autoclaving or replacement. The biofilter can only be replaced by a service engineer.

Mattress Replacement, Overlay and Seat Cushion

General Care Remove the top cover and inspect for signs of wear or
any tears.
Check all zips are secure.
Check integrity of all connectors, including cell to manifold connections.
Ensure all cell fasteners are correctly connected to the mattress base sheet and are not loose or damaged.

Serial Labels

Pump The serial number for the pump is on the label on the
back of the pump case.
Mattress The mattress serial label can be found just inside the
base cover above the tubeset.
Seat Cushion The seat cushion serial label can be found just inside the
front of the base cover.
Quote these serial numbers when requesting service.
25

9. Troubleshooting

The following table provides a troubleshooting guide for the Alpha Response systems in the event of malfunction.
LCD Screen Possible Cause Remedy
Pump is inflating the mattress replacement or mattress overlay.
Mattress Inflating
Pump is inflating the seat cushion. The “egg timer” is replaced by the patient
Seat Inflating
1. No tubeset connected to pump.
2. Tubeset fitted but not connected securely.
The “egg timer” is replaced by the patient body icon when mattress inflation is complete.
body icon when seat cushion inflation is complete.
1. Connect a mattress or seat tubeset to the pump.
2. Remove and reconnect tubeset, making sure it is securely pushed onto the pump until a “click” is heard.
No Mattress/Seat
(graphics alternate)
Low Pressure
High Pressure
Power Fail
Hardware Fail
Service Indicator (top right of LCD
screen)
1. The tubeset is not connected properly.
2. CPR control not fully closed
3. There is a leak in the system.
1. Tubeset is “kinked” or blocked.
2. Pump has detected an internal fault.
External mains power supply has been removed while the pump is operating.
Pump has detected an internal fault.
Pump needs a service: After 12 months run time, the
spanner icon is illuminated. After a further 3 months run time,
the spanner icon starts flashing.
1. Remove and reconnect tubeset, making sure it is securely pushed onto the pump until a “click” is heard.
2. Make sure the CPR plug is fully pushed into the CPR grommet.
3. Call service engineer.
1. Check and remove any “kinks” or blockages in the tubeset.
2. Disconnect the pump from the electrical supply and call service engineer.
Reconnect the mains power supply to the pump.
Disconnect the pump from the electrical supply and call service engineer.
Call the service engineer.
If correct operation or performance of the pump is not restored by troubleshooting
procedures, stop using the system immediately and call the service engineer.
26

10. Technical Specification

PUMP
Model: Alpha Response
Part Numbers: 464001 - UK
464009AU - AUS
Supply Voltage: 100 - 230 V
Supply Frequency: 50 - 60 Hz
Power Input: 24 - 36 VA
Size: 240 (L) x 210 (H) x 135 mm (D) [9.49 x 8.27 x 5.3”]
Weight: 3.2 kg (7.05 lb)
Case Material: ABS Plastic
Plug Fuse Rating: 5A to BS1362 (UK ONLY)
Degree of protection against electric shock:
Degree of protection against liquid ingress:
Mode of operation: Continuous
Mains Connected: Class II, Double Insulated with Functional Earth Type BF
IP21 - Protection against ingress of solid objects more than 12.5mm diameter and water droplets falling vertically.
PUMP SYMBOLS
The operator must read this document (Instructions for Use) before use. Note: This symbol is blue on the product label.
With respect to electric shock, fire and mechanical hazards only in accordance with CAN/ CSA-C22.2 No. 60601.1 (2008). MEDICAL EQUIPMENT
Refer to this document (Instructions for Use) for
i
a description of the product classification (3rd Edition).
SN:
Run/Standby
Button
Double Insulated Type BF
Serial Number
Ref:
Do not dispose of in domestic refuse
Model number
Refer to this document (Instructions for Use) for a description of the product classification (2nd Edition).
Manufacturer: This symbol is accompanied by the name and the address of the manufacturer.
27
PUMP ENVIRONMENTAL INFORMATION
Condition Temperature Range Relative Humidity Atmospheric Pressure
Operating +5°C to +40°C
(+41°F to +104°F)
Storage (Long Term) +5°C to +40°C
(+41°F to +104°F)
Storage (Short Term) -20°C to +65°C
(-4°F to +149°F)
30% to 75%
(non-condensing)
20% to 95%
(non-condensing)
20% to 95%
(non-condensing)
700hPa to 1060 hPa
700 hPa to 1060 hPa
500 hPa to 1060 hPa
If the pump is stored in conditions outside the “Operating” ranges, allow time for its temperature
to stabilise to normal before use.
One of the effects of prolonged exposure to high temperatures is to increase the self-discharge of
the internal battery; this will reduce the duration of power fail alarms. The pump will fully charge the battery over a 24-hour period when the pump is connected to a mains power supply.
MATTRESS
Description Size (Height x Length) Cell Material
Mattress Replacement (MR) 205mm (H) x 2090mm (L) Polyurethane PU Laminate
Mattress Overlay (OL) 115mm (H) x 2090mm (L) Polyurethane PU Laminate
Base Cover
Material
MATTRESS SIZE INFORMATION
Part No. Description Width (mm)
465001ADV Alpha Response Mattress Replacement (Advantex) 886
465001DAR Alpha Response Mattress Replacement (Dartex) 886
465002ADV Alpha Response Mattress Replacement, Narrow (Advantex) 806
465002DAR Alpha Response Mattress Replacement, Narrow (Dartex) 806
465003ADV Alpha Response Mattress Overlay (Advantex) 886
465003DAR Alpha Response Mattress Overlay (Dartex) 886
465004ADV Alpha Response Mattress Overlay, Narrow (Advantex) 806
465004DAR Alpha Response Mattress Overlay, Narrow (Dartex) 806
SEAT CUSHION
Part Number: 465005DAR
Length: 470mm
Width: 455mm
Height: 50mm
Cell Material: Polyurethane
28
CLEANING SYMBOLS
95
Do Not Use Phenol-based cleaning Solutions
Do not iron Tumble dry at 80-85°C
Use solution diluted to 1000 ppm of Available Chlorine
Wash at 95°C (203°F) MAX
130
Tumble dry at 130°C (266°F)
Wipe surface with damp cloth
COVER SPECIFICATION
Feature
Removable Cover Ye s Yes
Moisture Vapour Permeable Yes Yes
Standard Cover (Dartex)
®
Advantex
®
Air Permeable No No
Low Friction Yes 18% lower
Water Resistant / Repellent Ye s Yes
Infection Control Material coating is
Antimicrobial
Fire Retardant BS 7175: 0,1 & 5 BS 7175: 0,1 & 5
2-Way Stretch Yes Some
Washing Conditions
Drying Conditions Tumble Dry up to 130°C (266°F) or
Life Span 50 Wash Cycles (minimum) 50 Wash Cycles (minimum)
Application Area Acute and Homecare Acute and Homecare
(a)
MAX 95°C (203°F) MAX 95°C (203°F)
(b)
Air Dry
Material coating is
Antimicrobial
Tumble Dry ONLY
at 80-85°C (176°F-185°F)
a. The top cover may be washed. The temperature in the washing cycle may be up to 95°C (203°F);
however it is recommended that you check your local policy to determine the time/ temperature ratio required to achieve thermal disinfection.
b. The top cover may be tumble dried or air dried. The temperature in the drying cycle may be up
to 130°C (266°F); however it is recommended that you check your local policy to determine the time/temperature ratio required.
c. The top cover should be tumble dried only at a temperature of 80-85°C (176-185°F); however it
is recommended that you check your local policy to determine the time/temperature ratio required.
(c)
29
Guidance and manufacturer’s declaration - electromagnetic emissions
The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment - guidance
RF emissions
CISPR - 11
RF emissions
CISPR - 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Group 1 The pump uses RF energy only for its internal
function. therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B The pump is suitable for use in all
establishments including domestic establishments and those directly connected to
Class A
Complies
the public low-voltage power supply network that supplies buildings used for domestic purposes.
Recommended separation distances between portable and mobile RF communications equipment and the pump
The pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the pump as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum
output power of
transmitter
W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 2.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
150 kHz to 80 MHz
d = 1.2P
80 MHz to 800 MHz
m
d = 1.2P
800 MHz to 2.5 GHz
d = 2.3P
30
Guidance and manufacturer’s declaration - electromagnetic immunity
The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should assure that it is used in such an environment.
Immunity
Test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
IEC 60601 Test
Level
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2.5GHz
Compliance
Level
3 V
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the pump, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2P
d = 1.2P 80 MHz to 800 MHz
d = 2.3P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a, should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the pump is used exceeds the applicable RF compliance level above, the pump should be observed to verify normal operation. If abnormal operation is observed, additional measures may be necessary, such as reorientating or relocating the pump.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
31
32
AUSTRALIA ArjoHuntleigh Pty Ltd 78, Forsyth Street O’Connor AU-6163 Western Australia Tel: +61 89337 4111 Free: +1 800 072 040 Fax: + 61 89337 9077
BELGIQUE / BELGIË ArjoHuntleigh NV/SA Evenbroekveld 16 BE-9420 ERPE-MERE Tél/Tel: +32 (0) 53 60 73 80 Fax: +32 (0) 53 60 73 81 E-mail: info@arjohuntleigh.be
BRASIL Maquet do Brasil Equipamentos Médicos Ltda Rua Tenente Alberto Spicciati, 200 Barra Funda, 01140-130 SÃO PAULO, SP - BRASIL Fone: +55 (11) 2608-7400 Fax: +55 (11) 2608-7410
CANADA ArjoHuntleigh 90 Matheson Boulevard West Suite 300 CA-MISSISSAUGA, ON, L5R 3R3 Tel/Tél: +1 905 238 7880 Free: +1 800 665 4831 Institutional Free: +1 800 868 0441 Home Care Fax: +1 905 238 7881 E-mail: info.canada@arjohuntleigh.com
ČESKÁ REPUBLIKA ArjoHuntleigh s.r.o. Hlinky 118 CZ-603 00 BRNO Tel: +420 549 254 252 Fax: +420 541 213 550
DANMARK ArjoHuntleigh A/S Vassingerødvej 52 DK-3540 LYNGE Tel: +45 49 13 84 86 Fax: +45 49 13 84 87 E-mail: dk_kundeservice@arjohuntleigh.com
DEUTSCHLAND ArjoHuntleigh GmbH Peter-Sander-Strasse 10 DE-55252 MAINZ-KASTEL Tel: +49 (0) 6134 186 0 Fax: +49 (0) 6134 186 160 E-mail: info-de@arjohuntleigh.com
ESPAÑA ArjoHuntleigh Ibérica S.L. Ctra. de Rubí, 88 1ª planta - A1 08173 Sant Cugat del Vallés ES- BARCELONA 08173 Tel: +34 93 583 11 20 Fax: +34 93 583 11 22 E-mail: info.es@arjohuntleigh.com
FRANCE ArjoHuntleigh SAS 2 Avenue Alcide de Gasperi CS 70133 FR-59436 RONCQ CEDEX Tél: +33 (0) 3 20 28 13 13 Fax: +33 (0) 3 20 28 13 14 E-mail: info.france@arjohuntleigh.com
HONG KONG ArjoHuntleigh (Hong Kong) Ltd 1510-17, 15/F, Tower 2 Kowloon Commerce Centre 51 Kwai Cheong Road Kwai Chung HONG KONG Tel: +852 2207 6363 Fax: +852 2207 6368
INTERNATIONAL ArjoHuntleigh International Ltd ArjoHuntleigh House Houghton Hall Park Houghton Regis UK-DUNSTABLE LU5 5XF Tel: +44 (0) 1582 745 800 Fax: +44 (0) 1582 745 866 E-mail: international@arjohuntleigh.com
ITALIA ArjoHuntleigh S.p.A. Via di Tor Vergata 432 IT-00133 ROMA Tel: +39 (0) 6 87426211 Fax: +39 (0) 6 87426222 E-mail: Italy.promo@arjohuntleigh.com
NEDERLAND ArjoHuntleigh Nederland BV Biezenwei 21 4004 MB TIEL Postbus 6116 4000 HC TIEL Tel: +31 (0) 344 64 08 00 Fax: +31 (0) 344 64 08 85 E-mail: info.nl@arjohuntleigh.com
NEW ZEALAND ArjoHuntleigh Ltd 41 Vestey Drive Mount Wellington NZ-AUCKLAND 1060 Tel: +64 (0) 9 573 5344 Free Call: 0800 000 151 Fax: +64 (0) 9 573 5384 E-mail: nz.info@ArjoHuntleigh.com
NORGE ArjoHuntleigh Norway AS Olaf Helsets vei 5 N-0694 OSLO Tel: +47 22 08 00 50 Faks: +47 22 08 00 51 E-mail: no.kundeservice@arjohuntleigh.com
ÖSTERREICH ArjoHuntleigh GmbH Dörrstrasse 85 AT-6020 INNSBRUCK Tel: +43 (0) 512 204 160 0 Fax: +43 (0) 512 204 160 75
POLSKA ArjoHuntleigh Polska Sp. z o.o. ul. Ks Piotra Wawrzyniaka 2 PL-62-052 KOMORNIKI (Poznan) Tel: +48 61 662 15 50 Fax: +48 61 662 15 90 E-mail: arjo@arjohuntleigh.com
PORTUGAL ArjoHuntleigh em Portugal MAQUET Portugal, Lda. (Distribudor Exclusivo) Rua Poeta Bocage n.º 2 - 2G PT-1600-233 Lisboa Tel: +351 214 189 815 Fax: +351 214 177 413 E-mail: Portugal@arjohuntleigh.com
SUISSE / SCHWEIZ ArjoHuntleigh AG Fabrikstrasse 8 Postfach CH-4614 HÄGENDORF Tél/Tel: +41 (0) 61 337 97 77 Fax: +41 (0) 61 311 97 42
SUOMI Oy Vestek AB Martinkuja 4 FI-02270 ESPOO Puh: +358 9 8870 120 E-mail: info@vestek.fi
SVERIGE ARJO Scandinavia AB Hans Michelsensgatan 10 SE-211 20 MALMÖ Tel: +46 (0) 10 494 7760 Fax: +46 (0) 10 494 7761 E-mail: kundservice@arjohuntleigh.com
UNITED KINGDOM ArjoHuntleigh UK ArjoHuntleigh House Houghton Hall Park Houghton Regis UK-DUNSTABLE LU5 5XF Tel: +44 (0) 1582 745 700 Fax: +44 (0) 1582 745 745 E-mail: sales.admin@arjohuntleigh.com
USA ArjoHuntleigh Inc. 2349 W Lake Street Suite 250 US-Addison, IL 60101 Tel: +1 630 307 2756 Free: +1 800 323 1245 Institutional Free: +1 800 868 0441 Home Care Fax: +1 630 307 6195 E-mail: us.info@arjohuntleigh.com
REV 14: 07/2014
www.arjohuntleigh.com
www.arjohuntleigh.com
ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö, Sweden
GETINGE GROUP is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh focuses on patient mobility and wound management solutions. GETINGE provides solutions for infection control within healthcare and contamination prevention within life sciences. MAQUET specializes in solutions, therapies and products for surgical interventions and intensive care.
464933EN_02: 07/2014
Loading...