2.Declaration of Conformity ........................................................................................................................ 5
3.Introduction and Specifications ............................................................................................................... 6
4.Communications and Getting started ..................................................................................................... 7
4.1. Serial Interface ......................................................................................................................................... 7
4.2. Command Language ................................................................................................................................ 7
4.4. Getting Started ......................................................................................................................................... 8
5.Hardware and Wiring ................................................................................................................................. 9
8.1. System Diagnosis Commands – ID, IV, IS, SR, RS .............................................................................. 13
8.1.1.ID Get Device Identity ................................................................................................................................... 13
8.1.2.IV Get Firmware Version ............................................................................................................................... 13
8.1.3.IS Get Device Status ..................................................................................................................................... 13
8.1.5.RS Read Serial Number ................................................................................................................................ 13
8.2.1.CE Read TAC* Counter / Open Calibration Sequence ................................................................................. 14
8.2.2.CM Set Maximum Output Value ..................................................................................................................... 14
8.2.3.DS Set Display Step Size .............................................................................................................................. 14
8.2.4.DP Set Decimal Point Position ...................................................................................................................... 14
8.2.5.CZ Set Calibration Zero Point ....................................................................................................................... 14
8.2.6.CG Set Calibration Gain (Span) ..................................................................................................................... 15
8.2.7.ZT Enable / Disable Zero Tracking ................................................................................................................ 15
8.2.8.ZI Initial Zero Range ...................................................................................................................................... 15
8.2.9.FD Reset to Factory Default Settings ............................................................................................................ 15
8.2.10.CS Save the Calibration Data ....................................................................................................................... 15
8.3. Motion Detection Commands – NR, NT ................................................................................................. 16
8.3.1.NR Set No-motion Range ............................................................................................................................... 16
8.3.2.NT Set No-motion Time Period ..................................................................................................................... 16
8.4. Filter Setting Commands – FL, FF, UR .................................................................................................. 17
8.4.2.FF Moving Average ....................................................................................................................................... 17
8.5.1.SZ Set Zero ................................................................................................................................................... 18
8.5.2.RZ Reset Zero Point...................................................................................................................................... 18
8.5.3.ST Set Tare ................................................................................................................................................... 18
8.6.1.GG Get Gross Value ...................................................................................................................................... 19
8.6.2.GN Get Net Value .......................................................................................................................................... 19
8.6.3.GT Get Tare Value ........................................................................................................................................ 19
8.6.4.GS Get ADC Sample Value ........................................................................................................................... 19
8.6.5.GF Get Filtered Value ................................................................................................................................... 19
8.6.6.GW Get Data String “Net, Gross and Status“ ................................................................................................. 19
8.6.7.GA Get Triggered Average Value .................................................................................................................. 19
LDU 68.1 & 68.2 Technical Manual, Rev. 12 January 2010
8.7.1.SG Send Gross Value continuously ............................................................................................................... 20
8.7.2.SN Send Net Value continuously .................................................................................................................. 20
8.7.3.SF Send Filtered Value continuously (Display Mode only) ............................................................................ 20
8.7.4.SW Send Data String “Net, Gross and Status“ continuously .......................................................................... 20
8.7.5.SA Send Triggered Average Value automatically ......................................................................................... 20
8.8. Commands for External I/O Control – IN, IO, IM ................................................................................... 21
8.8.1.IN Read the Status of the Input Channels ..................................................................................................... 21
8.8.2.IO Read / Set the Status of the Output Channels .......................................................................................... 21
8.8.3.IM Control of the logic outputs by the host application .................................................................................. 21
8.9. Setpoint Output Commands – Sn, Hn, An ............................................................................................. 22
8.9.1.S0 / S1 Setpoint Value ................................................................................................................................... 22
8.9.2.H0 / H1 Setpoint Hysteresis and Switching Action ......................................................................................... 22
8.9.3.A0 / A1 Allocation of Gros or Net Value ......................................................................................................... 22
8.10.3.CL Close Device Address n .......................................................................................................................... 23
8.10.4.DX Half-duplex or Full-duplex ....................................................................................................................... 23
8.10.6.OP Open Device ............................................................................................................................................ 23
8.11.Save Calibration and Setup Data Commands – CS, WP, SS, GI, PI ................................................ 24
8.11.1.WP Save the Setup Parameters .................................................................................................................... 24
8.12.2.MT Measuring Time ....................................................................................................................................... 25
8.12.3.GA Get Triggered Average Value .................................................................................................................. 25
The model LDU 68.x is a precise digital amplifier for weighing and force measurements with strain gauge (SG)
sensors. The LDU 68.2 can be used for industrial applications,the LDU 68.1 can be used in legal for trade as
well as for industrial applications. The device features full multi-drop communications capability and can be
programmed via a straightforward ASCII command set.
The LDU XX.X series and the amplifier DAS 72.1 with on-board digital display, use the same command set.
You can connect up to 32 SG amplifiers of either the LDU XX.X series or DAS 72.1 type onto a single RS 485
bus. The LDU 68.x with its accurate A to D converter and an internal sample rate of up to 90 measurement
values per second, is particularly suitable for static or dynamic measurements and control purposes.
Specifications LDU 68.1 LDU 68.2
Accuracy class III Test certificate according OIML R76 10 000 Teile -
Linearity< 0.002 % FS
Excitation
Analogue input range
Minimum input sensitivity0.1 μV / d
Certified accuracy according OIML R76 1.0 μV / e Resolution > ± 130 000 counts (input); ± 100 000 counts (output)
Conversion rate90 values/second internal, up to 90 values/second external
Digital filterIIR filter 0.2 to 5 Hz; Bessel; in 8 steps adjustable
Calibration By software via ASCII commands, sehr einfach durchführbar
Communication interface
Standard weighing functionsGross, tare, net, zero, etc.
Digital inputs
Digital outputs2x open collector outputs, < 30 V DC, max. 200 mA
Temperature effect on zero
Temperature effect on span
Operation temperature range–15 °C to +55 °C–15 °C to +50 °C
Storage temperature range –30 °C to +70 °C
Enclosure
Dimensions and weight
Power supply12 ... 24 V DC ±10 %, < 60 mA, not galvanically isolated
Available accesoires
EMCOIML R-76:2006 and DIN EN 45 501:1992/AC1993
5 V DC, load cells 350 – 1 150 Ohm (87,5 – 1 150 Ohm at supply
voltage up to max. 12 V DC); 6 wire technique
±11 mV (bipolar; for weighing applications, force and torque
measurements)
RS485 oder RS422; Voll-Duplex oder Halb-Duplex;
9600 ... 115200 Baud; busfähig bis zu 32 Einheiten
2x opto-isolated inputs, 10 ... 30 V DC max. 3 mA,
Status via software
Tinned steel enclosure, IP40 protection,
special IP65 housing on request
82 x 31 x 6 mm, weighs approx. 30 g;
with adaptor board 99 x 41 x 12 mm, approx. 50 g
LDU 68.1 & 68.2 Technical Manual, Rev. 12 January 2010
Page 6 of 28
4. Communications and Getting started
4.1. Serial Interface
Communicating with the LDU 68.1 digitizer is carried out via the RS422/RS485 port.
The data format is the familiar 8/N/1 structure (8 data bits, no parity, 1 stop bit).
Available baud rates via the RS422/RS485 port are as follows: 9 600, 19 200, 38 400, 57 600 or 115 200 baud.
RS422:
Connection using a 4 wire technique
Point-to-Point connection, i.e. no bus communication possible
Half duplex setup (DX=0)
RS485:
Connection using 2- or 4-wire techniques
Multi-drop connection possible, up to 32 LDU XX.X
Half or Full Duplex (DX=0 or DX=1) possible
(RS232):
The optional adapter board UA 77.1 is available which has an RS485/RS232 converter built in.
4.2. Command Language
The command set of LDU XX.X series is based on a simple ASCII format (2 capital letters). This enables the
user to setup the device, get results or check parameters.
Example: LDU XX.X with the address or channel number 1 is connected via the RS 485 port to a bus system.
You want to get the net weight.
In this manual means: Space “_“ and Enter (CR/LF) ““
Master (PC / SPS) sends Slave (LDU XX.X) responds Meaning
OP 1
OK
GN
N+123.45
Net weight with algebraic sign / floating point
The command OP_2 opens the communication channel to LDU XX.X device #2. Now device #2 acknowledges
that it is active and responds to any commands on the bus. Communication with device #2 will be closed by
another OP command (for another device on the bus e.g. OP_5) or by the command CL_2.
Each OP_X command implies a CL command to all other devices on the bus except device #X. This makes the
address structures easier and the system performance improves.
Open device #1
Device # 1 ready
Get net weight
LDU 68.1 & 68.2 Technical Manual, Rev. 12 January 2010
Page 7 of 28
4.3. Baud Rate / Device Address
Baud Rate:
For baud rate setup use command BR, see chapter 8.10
Factory default: 9 600 baud
Device Address:
For address setup use command AD, see chapter 8.10
Factory default: Address 0
Setting the device address to 0 will set the continuously active mode, where the device becomes permanently
active, and will listen and respond to any command on the bus, without the need for an OP xxx command.
Note: The LDU XX.X series has sets of special solder pads on the under side of the PCB (see chapter 5.1). If
the solder pads are bridged then the LDU will be in a special configuration mode to set up the baud rate and
device address. This function should be used only when baud rate or address is unknown.
After power up, the LDU XX.X will enter a special baud rate search mode – waiting for a space character (0x20)
to be received. The time duration of this character is measured by the LDU and its baud rate timing will be set
accordingly – i.e. the baud rate of the terminal used will be the baud rate used by the LDU subsequently. The
device address will also be set to ”0“
4.4. Getting Started
You will require:
PC or PLC with either a RS422 or RS485 communication port
If using a PC / PLC with RS232 port, a RS422/485 to RS232 converter will be required (option UA 77.1)
Interconnecting cabling - confirm that all relevant pins are used – see the wiring diagram at the end of
this section
A load cell / scale with test weights or a load cell simulator
A 12-24 VDC power supply capable of delivering approximately 100mA for each LDU and load cell
One or more LDU 68.1
A suitable ASCII communication software *
Refer to the following wiring diagram in chapter 5.
* You can easily communicate between a PC and an LDU using programs such as Procomm, Telemate, Kermit
or HyperTerminal (included in Windows).
Also the DOP software with graphical user interface and oscilloscope function, running under Windows
2000/XP is available as freeware. This software is included on the Flintec product CD-ROM (2008 or later).
LDU 68.1 & 68.2 Technical Manual, Rev. 12 January 2010
Page 8 of 28
.
5.
H
W
W
W
h
r
W
r
r
n
1
2
2
2
e
f
ardwa
e and
iring
5.1.
5.2.
iring
ith Unit
Adapte
UA 77.
(RS23
)
5.3.
LDU 68
ith Unit
.1 & 68.2 Tec
Adapte
nical Manual
UA 73.
Rev. 12 Ja
(RS42
uary 2010
/ RS485 Full-/Half-Dupl
x)
Page 9 o
28
Loading...
+ 19 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.