ERBA ELite 3 User Manual

Erba Lachema s.r.o., Karásek 2219/1d, 621 00 Brno, CZ
ELite 3
Hematology Analyzer
USER MANUAL
INS00066
Content
Page 2/72
Table of Contents
1. INTRODUCTION .................................................................................................. 5
1.1. Intended Use ................................................................................................. 5
1.2. The Instrument .............................................................................................. 5
1.2.1. Patient Testing ....................................................................................... 6
1.2.2. Reagents ............................................................................................... 7
1.2.3. Technical Operation ............................................................................... 7
1.2.4. Calibration .............................................................................................. 7
1.3. Instrument features ....................................................................................... 8
1.4. Parts of the Analyzer ................................................................................... 11
1.5. The measurement process ......................................................................... 11
1.5.1. Control Panels ..................................................................................... 12
1.5.2. Display ................................................................................................. 12
1.5.3. Touch screen ....................................................................................... 12
1.6. Control Material ........................................................................................... 13
1.7. Accessories ................................................................................................. 13
1.8. Specifications .............................................................................................. 14
2. INSTALLATION ................................................................................................. 15
2.1. General information ..................................................................................... 15
2.2. Environmental factors ................................................................................. 15
2.2.1. Electrical requirements ........................................................................ 15
2.2.2. Space requirements ............................................................................. 16
2.2.3. Peripherals ........................................................................................... 16
2.2.4. Reagents and waste handling .............................................................. 17
2.2.5. Maintenance ........................................................................................ 17
2.2.6. Cleaning ............................................................................................... 17
2.2.7. General points ...................................................................................... 17
2.3. Unpacking and installation .......................................................................... 18
2.3.1. Turning the Instrument ON, MAIN Menu .............................................. 20
2.3.2. Turning the Instrument OFF ................................................................. 21
2.3.3. Preparing for shipment ......................................................................... 21
2.3.4. Handling in Emergency ........................................................................ 22
2.3.5. Warning labels on the analyzer ............................................................ 23
3. MENU SYSTEM ................................................................................................. 24
3.1. General Information .................................................................................... 24
3.1.1. Navigating in the Menu System ........................................................... 24
Content
Page 3/72
3.1.2. Touch screen calibration ...................................................................... 24
3.1.3. Menu structure ..................................................................................... 25
4. OPERATING PRINCIPLES ............................................................................... 27
4.1. Impedance Method ..................................................................................... 27
4.2. Principle of HGB Measurement ................................................................... 27
4.3. Parameters ................................................................................................. 28
4.4. Absolute and Linearity Ranges of Parameters ............................................ 29
5. ROUTINE UTILIZATION and MEASUREMENT ............................................... 30
5.1. Sample handling ......................................................................................... 30
5.2. Sample analysis .......................................................................................... 33
5.2.1. Sample preparation .............................................................................. 33
5.2.2. Running a (new) sample ...................................................................... 33
5.2.3. Results ................................................................................................. 36
5.2.4. Warning flags ....................................................................................... 37
5.2.5. Parameter Limits (Normal ranges) ....................................................... 39
5.2.6. Blank Measurement ............................................................................. 39
5.2.7. Using Pre-diluted Mode ....................................................................... 40
6. DATABASE ....................................................................................................... 41
6.1. Database services....................................................................................... 42
6.2. The Filter / Select function .......................................................................... 43
6.3. Printing records ........................................................................................... 44
6.4. Manage records .......................................................................................... 45
7. MAINTENANCE ................................................................................................. 46
7.1. Cleaning ...................................................................................................... 46
7.2. Calibration ................................................................................................... 47
7.2.1. Calibration with factors ......................................................................... 48
7.2.2. Calibration by measurement ................................................................ 48
7.3. Quality control ............................................................................................. 50
7.3.1. References........................................................................................... 51
7.3.2. Measure ............................................................................................... 51
7.3.3. Diagram ............................................................................................... 51
7.3.4. Database ............................................................................................. 52
8. Diagnostics ....................................................................................................... 53
8.1.1. Device Information ............................................................................... 53
8.1.2. Self Test ............................................................................................... 53
8.2. Reagent status ............................................................................................ 54
8.2.1. How to empty waste container ............................................................. 55
8.2.2. Neutralization of Waste ........................................................................ 55
Content
Page 4/72
9. Settings ............................................................................................................. 56
9.1. Printer settings ............................................................................................ 56
9.2. General settings .......................................................................................... 58
9.3. Measurement .............................................................................................. 58
9.3.1. Unit settings ......................................................................................... 59
9.3.2. Normal ranges ..................................................................................... 59
9.3.3. Profile ................................................................................................... 60
9.3.4. Settings ................................................................................................ 60
9.4. Date and Time ............................................................................................ 61
9.5. Multi user mode........................................................................................... 62
10. PRINTING .......................................................................................................... 65
10.1. Printouts ...................................................................................................... 65
11. Troubleshooting ............................................................................................... 67
11.1. Regular Troubleshooting Procedures .......................................................... 67
11.2. Weekly User Maintenance .......................................................................... 67
11.2.1. Cleaning needle washing head ............................................................ 67
12. Fluidic Schematics ........................................................................................... 68
13. Reagent consumption ..................................................................................... 69
14. Appendix A ....................................................................................................... 70
15. Contact .............................................................................................................. 71
16. Appendix- package inserts for reagents and cleaning solutions ................ 72
ELite 3, User Manual
Page 5/72
1. INTRODUCTION
1.1. Intended Use
This hematology analyzer is a fully automated cell counter designed for in vitro diagnostic use, developed for small clinics and point-of-care lab offices.
1.2. The Instrument
ELite 3 is a fully automated, bench-top hematology cell counter. It implements the so-called Coulter-method for counting cells passing through a small
aperture, and measures the hemoglobin content of red blood cells. The analyzer features a color graphical LCD display module with touch screen, and
has a separate START button. The software allows sending results to an external printer (via USB port), or can have
an optional, 58 mm built-in thermal printer module.
Its internal memory is capable of storing 1000 records with full histograms, and individual patient data. QC measurements are also stored in separate database. The software operating the instrument is easy to upgrade using a USB pen-drive. The instrument allows connecting to a host computer for uploading records stored in the memory through a USB B (slave) port. Archiving and restoring of records to and from USB pen-drive is also possible.
NOTE
:
If the equipment is used in a manner different from which the manufacturer specified, the protection provided by the equipment may be impaired. Misuse of equipment or use other than its intended purpose will invalidate conditions of warranty. The accuracy and precision may also be impaired.
ELite 3, User Manual
Page 6/72
1.2.1. Patient Testing
The analyzer can process 80 samples per hour in 3-part WBC differential mode. Samples can have individual sample data, and additional parameters.
You can print results to an external or to the optional built-in printer. The user can customize the report format.
The analyzer determines the following 20 hematology parameters, including 3-part WBC differential, from a 25 µl whole blood sample:
WBC total white blood cell count LYM lymphocytes count MON monocytes count GRA granulocytes count
LYM% lymphocytes percentage MON% monocyte percentage GRA% granulocytes percentage
HGB hemoglobin RBC red blood cell count HCT hematocrit MCV mean corpuscular volume MCH mean corpuscular hemoglobin MCHC mean corpuscular hemoglobin concentration RDWc v red cell distribution width* RDWsd red cell distribution width*
PLT platelet count PCT plateletcrit (trombocrit) MPV mean platelet volume PDWcv platelet distribution width* PDWsd red cell distribution width*
* RDW and PDW parameters have two forms of representation: CV and SD. Both parameters describe the distribution width, but from different aspects. User can select the units to use for displaying RDW and PDW parameters.
ELite 3, User Manual
Page 7/72
1.2.2. Reagents
Use only reagents supplied by the manufacturer with the analyzer, otherwise accuracy cannot be guaranteed. All reagents are stabilized and micro-filtered. Reagents are environmental friendly, azide-free and do not contain harmful ingredients. Please refer to the MSDS associated with reagent. Store reagents between +15 and +30°C. Do not use reagent beyond the expiration date printed on the container label. Discard opened container after 120 days. Do not use reagent once frozen. These reagents are ready for use and can be applied straight from the container; no special reagent preparation is necessary. Leave the reagent at room temperature for at least 12 hours before using.
Cat. No. Product name Volume
HEM00004 Erba Diluent- Diff 20 l HEM00008 Erba Lyse- Diff 1 l HEM00001 Erba Cleaner 1 l
Additional cleaning solutions:
HEM00017 Erba Clenz 1 l HEM00016 Erba Hypoclean 1 l
For package insert for reagents and cleaning solutions please see Appendix
1.2.3. Technical Operation
As the cell counter is a fully automated instrument, operating it requires minimal training or technical support. Operator interaction is reduced to the following:
Perform a Blank Measurement in case the instrument is not used for a specific time
Enter sample and/or patient data
Insert the sample to be analyzed into the sample holder
Print results either one-by-one, or in groups by selecting records from the
database
Perform simple weekly maintenance, as described later in this description (7.1.2.).
1.2.4. Calibration
EELLiittee 33
arrives to your laboratory factory-calibrated and ready to use. However, calibration needs updating whenever you find that the results have slightly changed, or a different or new control material is used. With each control material you receive for the instrument, you will find a control sheet listing the parameters the instrument should match. Perform these calibrations as explained in a later chapter (7.2).
ELite 3, User Manual
Page 8/72
1.3. Instrument features
Figures 1 and 2 show front and rear view of the analyzer, with controls and
connectors.
Figure 1. Front view
Color LCD
touchscreen
START button
Built-in
thermal
printer
Sample holder
with
interchangeable
adapters
Front panel USB
socket
Status indicator
ELite 3, User Manual
Page 9/72
Figure 2. Rear view
Instrument label
(S/N, manuf.data)
Reagent
connectors
Power switch
USB A
connectors
External grounding connector
USB B
connector
Built-in
thermal
printer
Reagent Lock
connector
Power source
connector
ELite 3, User Manual
Page 10/72
To replace paper in the printer:
- open the paper lid (pull the lid upwards by the handle)
- remove central plastic roller of old paper roll
- unwind new paper roll, so that the “starting edge”
is coming from down under towards you
- gently drop the new roll into the holder of the printer, and hold the “starting edge” with your hand, and make sure it comes out on the front of the printer
- close the lid, making sure that the paper is captured between the lid and the front of the printer
The analyzer works with an external power supply. The power supply module has a so-called auto range input, allowing operation on 230V or 115V power system. The power supply unit complies CE and UL safety certifications.
The input socket is a standard power cable connection and the output is a DC jack.
CAUTION! Use only the provided power supply with the instrument:
“GlobeTek Electronics Corp.” Model ID: GT-81081-6012-T3
ELite 3, User Manual
Page 11/72
1.4. Parts of the Analyzer
The hematology analyzer is composed of three main units:
Fluidic System: Performs sampling, diluting, mixing, and lysing
functions. Generates the regulated vacuum used for moving cells through the aperture during the counting process.
Data Processing System: Counts, measures and calculates blood cell
parameters, generates and stores numerical results and histograms.
Control Panel: Features an LCD display, touch screen, START
button, status LED, and USB port interfaces.
1.5. The measurement process
For the Schematics of the fluidics system, see Section 13.
Sample aspiration and dilution:
Stages of the blood testing process
a.
25 µl of anti-coagulated (K3-EDTA) whole blood sample is aspirated into the
sampling needle, and mixed with 4 ml of diluent and stored in the chamber
(MIX dilution).
b.
25 µl of the MIX dilution is aspirated into and stored in the needle during WBC
measurement and hemoglobin analysis
c.
Lysing reagent is added to the mix dilution held in the WBC chamber for
WBC differential analysis. This amount of lysing reagent is patient type
dependent and the operator can change it.
d.
After WBC counting, HGB reading and washing process, 4 ml of diluent is
added to the second dilution (using the 25 μl of mix dilution stored in the
needle).
e.
This portion is analyzed for RBC count, PLT count and their parameters.
f.
Another washing process prepares the unit for the next analysis.
Table 1.
Dilution rates used: Measurement times:
MIX dilution
1:160 WBC count 12 seconds
RBC dilution
1:32 000 HGB measurement 3 seconds
WBC dilution 1:196 (depends on lyse amount)
RBC/PLT count 12 seconds
ELite 3, User Manual
Page 12/72
1.5.1. Control Panels
START button
Pressing and releasing the START button triggers an analysis cycle.
Status indicator
A two-color (red/green) LED (light emitting diode) is located above START button. Its actual color indicates the status of the analyzer.
LED color Analyzer status
Green
The analyzer is ready to measure sample. Analysis can be initiated by pressing START button.
Red blinking
Blood sample can be removed when the LED blinks red 3 times and the instrument beeps 3 times.
Red
The analyzer is currently performing an analysis. No new measurement can be started.
Yellow
The analyzer is performing a maintenance process.
Yellow blinking
The instrument is in stand-by and display light is off. Hit the screen to have SW wake up from stand-by.
1.5.2. Display
The display is 320 x 240 dots, high contrast backlit high-color graphic LCD module, with integrated touch screen.
1.5.3. Touch screen
The LCD screen has a touch-sensitive foil on the front surface. If the operator touches the LCD active area gently, the analyzer can recognize it and identify the position where the screen was pressed. By touching (slightly pressing) one small spot on the touch screen, the SW will activate the function/menu/key that the corresponding area represents.
ELite 3, User Manual
Page 13/72
1.6. Control Material
The analyzer allows continuous monitoring of measurement performance with Erba Hematology Control (Low, Normal, High) - control blood. This must match the types
of samples usually run on the instrument. Specification for this material (assay values and allowed tolerances along with expiry date) is always packed with the approved control material.
1.7. Accessories
Below is a list of accessories shipped with your analyzer. This list can also be referred as the EELLiittee 33 - pack”
EELLiittee 33
Hematology Analyzer User's Manual (this booklet) Reagent Tubing kit (with colored tubes)
Diluent tube (green) Lyse tube (yellow) Cleaner tube (blue)
Waste tube (red) Cleaning Tube Kit. Caps for reagent containers (matching connector colors). Waste Container (20 L). External power supply and power cable. Sample tube adapters. Optional: spare thermal roll-paper.
Reagent Tubing Kit: Cleaning Tube Kit:
ELite 3, User Manual
Page 14/72
1.8. Specifications
Sample volume
25 μl of whole blood in normal 3-part or non-diff mode 50 μl of whole blood in pre-diluted mode
Chambers
2 counting chambers for diluting whole blood and counting
Reagent system
Isotonic Diluent, Lyse, Cleaner
Aperture diameter
70μm (RBC/PLT), 100μm (WBC)
Throughput
80 tests/hour
Characteristics Parameter:
Accuracy
Reproducibility
(CV)
Carry-over
sample to
sample
Test range Unit
WBC
3% 3% <1% 4.0-20.0 103/μl
RBC
3%
2%
<1% 4.0-15.0 106/μl
HCT
3%
3%
<1% 25.0-50.0 %
MCV
2%
1%
N/A 60-100 fl
HGB
2%
2%
<1% 9-16 g/dl
PLT
5%
5%
<3% or <20 200-900 103/μl
Sampling method
Open tube system with automatic sample rotor.
Sample types
Human (general), male, female, baby, toddler and child.
Clog prevention
High-voltage pulse on aperture in each analysis cycle, chemical cleaning and high pressure back-flush of the aperture using Cleaner reagent.
Cleaning procedure
High-voltage burst of the aperture, high-pressure back-flush, chemical cleaning of the aperture using Cleaner reagent.
Calibration
1- or more-measurement automatic and manual (factors) calibration of WBC, HGB, RBC, PLT, MCV (or HCT), RDW, and MPV.
User interface
Easy-to-use, menu driven user interface with touch-screen and separate START button, status LED
Languages available
English, Spanish, Portuguese, French, Russian, Indonesian, German, Italian, Czech
Data capacity
1000 results, with RBC, PLT, and WBC 3-part histogram
Host computer interface
USB B port
Data back-up method
USB mass storage device
Software upgrade method
via USB A port using USB mass storage device
Printer interface
USB with support for HP printers (DeskJet, LaserJet, PCL3, PS, LIDIL)
Built-in printer
Axiohm thermal printer module, 58 mm wide roll paper, full report with histograms
Display
320x240 -dots, high-contrast, backlit, color graphics LCD (liquid crystal display)
User interface
Full-LCD Touch-screen + separate START button, red/green state LED
External keyboard
USB keyboard via USB A port
Power requirement
12VDC, 5A, 60W max. operating power
Power supply unit
External, auto-ranging power unit for 100-120 or 200-240 VAC, 50–60Hz
Operating temperature
59–86 °F (15–30 °C). Optimal temperature is 77 °F (25 °C)
Dimensions (W x D x H)
12.6 x 10.2 x 14.4 in (320 x 260 x 365 mm)
Net weight
12 kg
Storage tempera
ture
3 – 35°C
Transport temperature
5 – 40°C
ELite 3, User Manual
Page 15/72
2. INSTALLATION
2.1. General information
This chapter provides instructions for the installation of EELLiittee 33 hematology analyzer. The procedures described below must be followed correctly to ensure proper operation and service. Please carefully read and follow all instructions in this User’s Manual before operating the analyzer.
This hematology analyzer is a precision instrument: handle with care. Dropping or other improper handling of the instrument will disturb calibrated mechanic and electronic components and/or cause damage.
CAUTION: Always handle the instrument with care.
2.2. Environmental factors
Operate EELLiittee 33 within the ambient temperature range of 15 - 30°C and relative humidity of 45% - 85%. The optimum operating temperature is 25°C.
Avoid using the instrument in areas of extreme high or low temperatures or where it is exposed to direct sunlight. If kept at a temperature less than 10°C, the instrument should be allowed to sit for an hour at the correct room temperature before use.
Reagents should be stored at a temperature range of 18 - 30°C. Place the instrument in a well-ventilated location. Do not place it near potentially
interfering devices capable of emitting radio frequencies (e.g. radio or television receiver, radars, centrifuge, X-ray devices, fans, etc.).
Operation at an altitude over 3000 meters (9000 ft) is not recommended, because the throughput will be degraded.
Instrument is safe for transient voltages to INSTALLATION CATEGORY II and POLLUTION DEGREE 2.
Environmental and electrical characteristics provide accuracy and precision of the instrument and maintain a high level of operational safety for lab personnel.
2.2.1. Electrical requirements
EELLiittee 33
comes with an approved power cord, appropriate for your power system.
Proper use of the appropriate power cord assures adequate grounding of the system.
WARNING: Improper grounding of the analyzer bypasses important safety
features and may result in electrical hazard.
ELite 3, User Manual
Page 16/72
2.2.2. Space requirements
It is important to install the instrument in a suitable location. A poor location can adversely affect its performance. Consider the following space requirements:
- Select a location near a power source and close to a suitable drain.
- Place the unit on a clean and level surface.
- Leave at least 0.5m (20in) space on both sides and above the instrument to
access pneumatics and (optional) built in printer. Provide a minimum of 0.2m (8in) between the rear panel and the wall to allow for heat dissipation and tube clearance.
- Install the reagents in a suitable place that will make your work easy. The best place is on the ground, below the supporting desk of the instrument. The pneumatic system is capable of aspirating reagents from containers being 1m (3ft) below the reagent inputs. Make sure the reagent tubes are not bent, broken, twisted or blocked in between the desk the instrument is on and the wall behind. Such circumstances can result in instrument operation failure.
- DO NOT PLACE the reagents above the instrument, as there can be a risk of falling and spilling.
WARNING: Install the unit on a table or workbench. If the unit was installed
without a supporting desktop under the unit, there is a possibility that the analyzer could accidentally fall.
2.2.3. Peripherals
Connect external peripherals only when both the instrument and the peripheral device are off. Possible peripherals are:
- external printer
- the printer must be recommended by authorized technician
- the printer must be approved and listed
- the printer must have a CE mark
- external keyboard
- the external keyboard must be approved
- the external keyboard must have a USB port or suitable adapter
- link to host computer via USB port
- serial link cable must be approved by technician
- USB B port (linking to host computer) requires a USB A-B cable and
USB driver software (contact service for availability)
ELite 3, User Manual
Page 17/72
2.2.4. Reagents and waste handling
Handle reagents according to national or international regulations.
WARNING! Reagents may cause corrosion and skin irritation. If any of the
liquids leaked onto the cover of analyzer or the furniture, wipe it off immediately. In case of skin contact, rinse the liquid off with plenty of water.
Waste generated by the unit is biohazard material. Handling and disposal must happen according to regulations regarding reagent systems. See Section 7.3.2.
WARNING! Waste contains poisonous substances (because of chemical
content) and human origin substances meaning biohazard. These substances are representing potential danger to environment. For this reason, safe handling of the waste liquid is very important.
2.2.5. Maintenance
The user should check the following components weekly:
- bottom of washing head for salt build up – should be wiped off with a damp cloth or wiper
- tubing system – by opening the side door and look for any liquid leakage. If you experience leakage, contact authorized technician.
WARNING! The power supply unit and internal electronic boards must NOT be
opened or serviced by the user!
2.2.6. Cleaning
Clean the instrument and its power supply – in off state – on the outside only, using a damp cloth with a soft detergent. DO NOT let liquids get inside these units.
2.2.7. General points
The manufacturer guarantees work safety reliability and general characteristics under the following conditions only:
- services and repairs are performed by an authorized technician
- the electrical system of the laboratory follows national and/or international
regulations
- the system is operated according to instructions contained herein
ELite 3, User Manual
Page 18/72
2.3. Unpacking and installation
1. Carefully remove the analyzer from the shipping carton. Inspect the
instrument for any visible signs of damage incurred during shipping. Would you find any damage, file a claim with the carrier or your distributor immediately. Check the accessories received against the packing list. Contact Service if anything is missing.
CAUTION! Prior to initial operation, allow the instrument to reach room
temperature (approx. 2 hours). Rapid temperature changes in an operating unit can lead to water condensation, which may damage electronic parts, and cause malfunction.
2. Place the instrument on a firm work surface in the designated work area,
near an appropriate AC electrical outlet. The power outlet connection MUST be grounded.
NOTE Before making connections: Make sure that all power is in “OFF”
state before connections (printer, external keyboard) are made. Carefully read all literature accompanying the instrument and its accessories. Pay particular attention to the operating procedures for the external printer.
3. Keyboard and external printer
Attach the keyboard cable to one of the USB A ports on the back of the instrument. Attach both ends of the printer cable to the appropriate ports on the printer and EELLiittee 33. Attach the AC adapter to the printer (if required) and plug it into an AC outlet.
4. Host Computer
The instrument has a built-in USB B port that allows connection to a host computer. You can export results, including histograms. USB B I/O settings are located in Settings menu.
For installation instructions for communication, please, contact Service.
5. Power supply
Connect the power supply to the instrument. Attach power cord outlet to the external power supply of EELLiittee 33 and plug the other end into a properly grounded AC outlet.
WARNING: Do not switch on the analyzer before connecting external power
supply to it and to the AC outlet, as well as before connecting an external printer or a keyboard to the analyzer.
ELite 3, User Manual
Page 19/72
6. Reagent Containers
Place the reagent containers near the instrument, to an accessible location. Do not place the containers to a higher position than that of
EELLiittee 3
3, because would a tube come off its connector, the fluids spill out. Use the supplied connecting tubes and special bottle caps. Be sure that the color on each tube and cap match. You can, for example, place the reagent containers below the desk the analyzer is installed on, as the instrument has sufficient power to draw the liquids from a lower location.
All containers should be left open (do not block the small air vent hole on
the special container caps) in order to provide free airflow.
Figure 5. Reagent connections
WARNING! Reagents may cause corrosion and skin irritation. If any of liquids
leaked to cover of analyzer or the furniture, wipe it off immediately. In case of skin contact, rinse the liquid with plenty of water.
ELite 3, User Manual
Page 20/72
2.3.1. Turning the Instrument ON, MAIN Menu
a. In case you use an external printer (for information, read manual
shipped with the printer) connect it and turn it on.
b. Turn the analyzer on using the power switch on the rear panel. The
‘ON’ position is marked by the ‘I’ symbol.
After turning on power, there will not be LCD activity for a few seconds, but the status LED goes on. During start-up, the following screen is displayed.
When SW is loaded, Main menu is displayed. Tap a touch-screen item to go activate the
menu element.
CAUTION! Wait 5 minutes before initiating any measuring process to allow
the instrument to reach the optimal working temperature.
In some cases, a priming cycle is necessary prior to sample introduction. The instrument will perform priming cycle automatically if additional liquid in the tubing system is required.
Run a priming cycle in case of:
installation
extended time out of use
replacement of any component related to the Fluidic System
replacement of reagents with instrument turned on
ELite 3, User Manual
Page 21/72
2.3.2. Turning the Instrument OFF
DO NEVER turn off the analyzer by simply flipping the power switch on the rear panel. Doing so may result in erroneous operation during later use. It can be so, because the instrument uses diluent. This liquid is an isotonic saline solution containing salt. Would it not be washed out of special units of the instrument or would chambers not be filled with this solution may lead to dust condensation or salt build up. Therefore always follow the instructions below when switching the instrument off.
In Main menu, select SHUTDOWN. The following screen appears.
EXIT SHUT DOWN Select Shutdown.
The analyzer will perform the necessary steps to prevent failure to the pneumatic system, and then gives a tone indicating that it is safe to shut it off.
Turn off power using the rear power switch.
Turn off the instrument using the power switch on the rear panel. The ‘OFF’ position is marked by ‘O’ symbol.
2.3.3. Preparing for shipment
Use the second item in the Shutdown menu when the instrument is to be shipped or left unused for a longer time (more than 1 week). The instrument will ask you to use the cleaning tube kit and 100ml of distilled water.
Follow the instructions appearing on the display.
SHUTDOWN Preparing for shipment (2)
Remove tubing connectors, so the system can drain itself.
Leave the waste connector attached.
ELite 3, User Manual
Page 22/72
Next, you should connect the cleaning tube kit to the reagent inputs, submerging the free end in a bottle containing at least 100 ml of distilled water.
Cleaning tube kit connected The analyzer will flush any remaining reagents from the system into the waste container.
As a next step, the analyzer asks you to
remove the cleaning tube kit. Leave Waste connected.
When finished, the analyzer prompts you to power off the system. Remove the
waste connector after shutting down.
2.3.4. Handling in Emergency
In case of emergency situation - like instrument catching on fire (short-circuit, etc.) - cut off power immediately by disconnecting mains power or DC input line, and use a fire-extinguisher if necessary.
ELite 3, User Manual
Page 23/72
2.3.5. Warning labels on the analyzer
Label Meaning Explanation
Biohazard
Sample and waste are
potentially infectious
material
Corrosive
Reagents may cause
corrosion or skin irritation.
Warning General warning of injury.
Sharp
needle
warning
Sampling needle may
cause injury.
ELite 3, User Manual
Page 24/72
3. MENU SYSTEM
3.1. General Information
This chapter contains information about the structure and usage of the software implemented menu structure.
This integrated software controls instrument operations including calculation and evaluation of measured data, displaying results and information screens, storage and recalling of data.
3.1.1. Navigating in the Menu System
The instrument uses a menu system to initiate actions and to access settings. Navigate in the menu system by simply touching the LCD at the item you want to
open/activate. From any submenu, the Home button will go back to Main menu, while Back moves one step back in the menu tree.
3.1.2. Touch screen calibration
In case you experience uncertainties during tapping the touch screen (you tap a specific location, still the required function key is not being activated) you will have to calibrate the touch panel.
Tap and gently hold any location on the touch screen. (Make sure not to press it hard, otherwise you can damage the screen.) After cca 10 seconds, a calibrating screen appears. Tap the reference points one after another. If you made an error, you will hear an error beep, and the process restarts.
If the calibration was successful, you will return to the original screen.
Loading...
+ 53 hidden pages