GE Infrastructure
Sensing
Features P P RTD °F (°C) TC mA/V Hz
Indicator (measure pressure) !!
Calibrator (measure or source) !! ! ! ! ! !
Thermometer (dual input T1, T2, T1 - T2) !
mA measure with 24 V loop power !! !! !
Switch test !! !! !
HART resistor !! !! !
Programmable step and ramp output !! ! ! ! !
Hold, scaling, max/min/avg, filter, scaling, !!!! ! ! ! ! !!
tare
25 pressure units, flow scaling, leak test !!!! ! ! ! ! !!
1000 points data memory, RS232 !
Applications
Measurement and monitoring !!!! ! !! ! !!
Indicator, controller and recorder testing !!!! !! ! ! !
Transmitter maintenance and calibration !! !! !
Process loop set-up and maintenance !! !! !
Switch, trip and safety system testing !! !! !
!With IDOS UPM connected
DPI 800
DPI 832
DPI 841
DPI 842
DPI 802
DPI 811
DPI 812
DPI 820
DPI 821
DPI 822
IDOS
Specifications
DPI 150 Pressure Indicator
IDOS Flexibility
Intelligent Digital Output Sensor (IDOS)
Universal Pressure Modules are available from
10 in H2O to 10,000 psi (25 mbar to 700 bar).
Total Flexibility
IDOS modules can be used with any compatible instrument; for example, a DPI 812 RTD loop calibrator can
become a fully featured pressure calibrator.
Instrument Range Expansion
Simply achieved by adding modules
DPI 800 Series IDOS UPM DPI 150
Compatible Products
The DPI 800 Series of robust , highly rel iable and simple
to use hand-held indicators and calibrators address a
broad range of applications with the following features:
Features Applications
• Barometer* • Sensor calibration
• IEEE-488* and RS232 • Instrument calibration
• Analog output, V and mA* • Test/measurement
• Airspeed indication • System monit oring
• Hold, maximum/minimum and average • Autmated production test
• Filter, scaling and tare and calibration
• 25 pressure units plus 5 user • Pressure data logging
• Leak test • Leak testing
*Optional feature
Pressure Range G/D G A Media *Accuracy %FS
+
_
SP
±10 in H2O (25 mbar) ! "#0.1 0.03
.±1, 3, 5, or 10 psi ! "#0.075 0.03
(70, 200, 350,
or 700 mbar)
5 psi (350 mbar) ! " 0.1 N/A
-15 to 15 or 30 psi ! "#0.05 0.01
(-1 to 1 or 2 bar)
30 psi (2 bar) ! " 0.075 N/A
-15 to 50, 100, 150 ! ! 0.05 0.01
or 300 psi (-1 to 3.5,
7, 10 or 20 bar)
100, 300 psi ! ! 0.075 N/A
(7, 20 bar)
500, 1000, 1500, ! ! 0.05 0.01
2000 or 3000 psi
(35, 70, 100, 135,
200 bar)
5000 or 10,000 psi ! ! 0.05 N/A
(350 or 700 bar)
Sealed gauge
G = gauge, A = absolute, G/D = gauge/differential; calibrated referenced to atmosphere maximum line pressure 30 psi (2 bar). !Stainless steel compatibility "Non-cor-
rosive gas/fluid #Non-corrosive gas. (N/A = not available). Accuracy assumes regular
zero correction.
A high precision, simple to use bench or panel mounted
indicator suitable as a secondary standard