Druck DPS5000 Specifications

Page 1
DPS5000
Digital Pressure Sensing Platform
The new DPS5000 from GE, part of the UNIK5000
family, oers integrated digital electronics to enhance the performance level of the UNIK 5000 Pressure Sensing Platform to levels unmatched by traditional analogue sensors. It features an I2C digital interface, over which fully compensated readings of Pressure and Temperature are sent, as well as control of many functions of the device.
With 40 years of pressure measurement experience, our eld-proven Druck technology is at the heart of the new platform, resulting in a range of high quality, high stability pressure sensors.
Bespoke as Standard
Custom-built from standard components, manufacturing sensors to your requirements is fast and simple; each UNIK 5000 is a “bespoke” pressure sensing solution, but with the short lead times and competitive pricing you would expect from standard products.
Expertise
We have the people and the knowledge to support your needs for accurate and reliable product performance; our team of experts can help you make the right sensor selection, guiding you and providing the help and tools you need. It is important that you ensure that the sensor materials and performance selected are suitable for your application.
Features
Ranges from 70 mbar to 100 bar
Total accuracy to ±0.1 % FS
Stainless steel construction
3 V supply voltage
Low power
I2C digital output
Sleep mode
Hazardous area certications
Excellent long-term stability
bhge.com
Page 2
DPS5000 Specications
Operating Pressure Ranges
Gauge Ranges
Zero-based ranges:
70 mbar
200 mbar
350 mbar
700 mbar
1 bar
2 bar
3.5 bar
7 bar
10 bar
20 bar
28 bar
35 bar
70 bar
100 bar
Absolute Ranges
0.8 to 3 bar
1 to 7 bar
2 to 12 bar
3 to 28 bar
5 to 40 bar
10 to 70 bar
18 to 90 bar
15 to 100 bar
Over Pressure
2 × pressure range for absolute ranges 4 × pressure range for gauge ranges
Containment Pressure
6 × pressure range for gauge ranges (200 bar maximum)
200 bar for absolute ranges
Supply
Supply voltage
2.7 to 3.6 Vdc Current consumption <50 µA Standby <2 mA average during acquisition
Output/Communications
I2C Slave Device, 100 kHz maximum data rate
Power-on Time
30 ms to acquisition from standby
Insulation Resistance
500 Vdc ≥ 100 MΩ
Performance
Pressure Performance
Calibrated accuracy over the calibrated temperature range including zero and span setting and the eects of non-linearity, hysteresis and repeatability
Gauge Reference
±0.1 % FS Increases pro-rata for pressure ranges below
700 mbar
Absolute Reference
±0.2 % of reading (RDG)
Temperature Performance
Accuracy over the calibrated temperature range ±3°C
Long Term Stability
±0.05 % FS /year typical ±0.1 % FS maximum at reference conditions Increases pro-rata for pressure ranges below
700 mbar
Physical Specications
Environmental Protection
Internal version – Not applicable External version – IP68 to 700 mH2O
Operating Temperature Range
-40°C to +85°C (-40°F to +185°F)
Pressure Media
Fluids compatible with stainless steel 316L and Hastelloy C276
Statement in accordance with the European Pressure Equipment Directive
Refer to document K0581 for product classication and regulatory information
Enclosure Materials
Stainless steel (body)
Pressure Connectors
PA G1/4 female PB G1/4 male at PE 1/4 NPT female PF 1/4 NPT male PG 1/8 NPT male Other connectors may be available. Contact BHGE to
discuss your requirement.
Page 3
Electrical Connector
Option Code Description IP rating
0 Internal None (ying leads) -
3 External Polyurethane cable IP68 to 700 mH2O
Wiring Details
Option 0 Internal Option 3
External
Pin Colour
1 Red Red Supply +ve
2 Yellow Orange I2C SDA
3 Green Black I2C SCL
4 Blue White 0 V
5 Orange Yellow Do not
6 Black Blue Do not
- - Screen Case
Function
Connect
Connect
General Certications
RoHS 2011/65/EU CRN Certied OF 13650.513467890YTN for
pressure ranges up to and including 100 bar
CE Conformity
ATEX 2014/34/EU (Optional)
EMC Directive 2014/30/EU (External version only)
BS EN 61326-1:2013 BS EN 61000-6-1: 2007 BS EN 61000-6-3: 2007 + A1:2011
Hazardous Area Approvals
External Version (Apparatus)
ATEX II 1G IECEx Ex ia IIC T4 Ga (-40°C ≤ Ta ≤ 80°C)
Internal Version (Component)
ATEX II 1G IECEx Ex ia IIC Ga
Ordering Information
1) Select part number
Main Product Family DPS Product Diameter and Material
50 25mm Stainless Steel
Electrical Connection
0 None (ying leads) 3 Polyurethane cable Note 1
Electronics Option
D Digital I2C bus
Compensated Temperature Range
TC -40° to +85°C (-40° to +185°F)
Accuracy
A3 Premium
Calibration
CC Full Thermal
Hazardous Area Certication H0 None
H1 IECEx/ATEX Intrinsically Safe ‘ia” Group IIC
Pressure Connector
PA G1/4 Female PB G1/4 Male Flat PE 1/4 NPT Female PF 1/4 NPT Male PG 1/8 NPT Male
DPS 50 D - TC - A3 - CC - -
Ordering Notes
Note 1: Cable length 3 m
Typical Order Example:
DPS503D-TC-A3-CC-H1-PF 20 bar gauge
Page 4
Mechanical Drawings
Internal Version
External Version
Notes:
(1) All dimensions in millimetres.
bhge.com
© 2015 Baker Hughes, a GE company – All rights reserved.
Baker Hughes reserves the right to make changes in specications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your BHGE representative for the most current information. The Baker Hughes logo is a trade mark of Baker Hughes, a GE company. The GE Monogram is a trademark of the General Electric Company.
920-660B