Drager Babytherm 8000 Service Manual

Babytherm 8000 Technical Service Bulletin # 11
Re: Mattress heating, replacing Controller PCB, new
Software Version 1.04
Date: April 30, 2001
Reference Doc: Service Bulletin from Dräger Medizintechnik GmbH:
6132.200 Babytherm 8000 # 11, February 13, 2001
software version 1.04. If a new Controller PCB is used with the previous software version error code 15 will be generated.
Solution: When replacing the PCB Controller P/N 8290668, software version 1.04
must be installed.
Devices affected: All units Babytherm 8000 with installed software version 1.02.
When: With next replacement of Controller PCB P/N 8290668.
Reporting: N/A
Additional Info: N/A
Cost: Covered by Customer, post warranty.
Ordering Info: EPROM SW1.04 Babytherm 8000 P/N 2M 22 467
Distribution: Dräger Service Personnel and Authorized Service Organizations.
If you have any questions, please contact Technical Support by phone at 1-800-543-5047 or by fax at 1-215-721-5789.
Dräger Medical, Inc. Technical Product Manager
6132.200 Babytherm 8000 TSB No. 11 April 2001 1/1